Please check for change information
at the rear of this manual.
Fifth Edition: January 1995
Copyright T ektronix, Inc. 1994, 1995. All rights reserved.
T ektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
Printed in the U.S.A.
T ektronix, Inc., P.O. Box 1000, Wilsonville, OR 97070–1000
TEKTRONIX and TEK are registered trademarks of T ektronix, Inc.
WARRANTY
T ektronix warrants that this product will be free from defects in materials and workmanship for a period of three (3) years
from the date of shipment. If any such product proves defective during this warranty period, T ektronix, at its option, either
will repair the defective product without charge for parts and labor, or will provide a replacement in exchange for the
defective product.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration of the
warranty period and make suitable arrangements for the performance of service. Customer shall be responsible for
packaging and shipping the defective product to the service center designated by T ektronix, with shipping charges prepaid.
T ektronix shall pay for the return of the product to Customer if the shipment is to a location within the country in which the
T ektronix service center is located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any
other charges for products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. T ektronix shall not be obligated to furnish service under this warranty a) to repair damage resulting
from attempts by personnel other than T ektronix representatives to install, repair or service the product; b) to repair
damage resulting from improper use or connection to incompatible equipment; or c) to service a product that has been
modified or integrated with other products when the effect of such modification or integration increases the time or
difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THIS PRODUCT IN LIEU OF ANY
OTHER WARRANTIES, EXPRESSED OR IMPLIED. TEKTRONIX AND ITS VENDORS DISCLAIM ANY
IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
TEKTRONIX’ RESPONSIBILITY TO REPAIR OR REPLACE DEFECTIVE PRODUCTS IS THE SOLE AND
EXCLUSIVE REMEDY PROVIDED TO THE CUST OMER FOR BREACH OF THIS WARRANTY. TEKTRONIX
AND ITS VENDORS WILL NOT BE LIABLE FOR ANY INDIRECT , SPECIAL, INCIDENTAL, OR
CONSEQUENTIAL DAMAGES IRRESPECTIVE OF WHETHER TEKTRONIX OR THE VENDOR HAS
ADVANCE NOTICE OF THE POSSIBILITY OF SUCH DAMAGES.
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it.
Only qualified personnel should perform service procedures.
While using this product, you may need to access other parts of the system. Read
the General Safety Summary in other system manuals for warnings and cautions
related to operating the system.
Injury Precautions
Use Proper Power Cord
Avoid Electric Overload
Ground the Product
Do Not Operate Without
Covers
Use Proper Fuse
Do Not Operate in
Wet/Damp Conditions
Do Not Operate in
Explosive Atmosphere
To avoid fire hazard, use only the power cord specified for this product.
To avoid electric shock or fire hazard, do not apply a voltage to a terminal that is
outside the range specified for that terminal.
This product is grounded through the grounding conductor of the power cord. To
avoid electric shock, the grounding conductor must be connected to earth
ground. Before making connections to the input or output terminals of the
product, ensure that the product is properly grounded.
To avoid electric shock or fire hazard, do not operate this product with covers or
panels removed.
To avoid fire hazard, use only the fuse type and rating specified for this product.
To avoid electric shock, do not operate this product in wet or damp conditions.
To avoid injury or fire hazard, do not operate this product in an explosive
atmosphere.
CSA Certification includes the products and power cords appropriate for use in
the North America power network. All other power cords supplied are approved
for the country of use.
This is the Performance Verification and Specifications for the TDS 684A and
7XXA Oscilloscopes. It contains procedures suitable for determining if each
instrument functions, was adjusted properly, and meets the performance
characteristics as warranted.
Also contained in this document are technical specifications for these oscilloscopes.
The following documents are related to the use or service of the digitizing
oscilloscope:
HThe TDS 684A, TDS 744A, & TDS 784A User Manual (Tektronix part
number 070-8991-XX).
HThe TDS Family Programmer Manual (Tektronix part number
070-8709-XX) describes using a computer to control the digitizing oscilloscope through the GPIB interface.
HThe TDS 684A, TDS 744A, & TDS 784A Reference (Tektronix part number
070-8999-XX) gives you a quick overview of how to operate your digitizing
oscilloscope.
HThe TDS 684A, TDS 744A, & TDS 784A Service Manual (Tektronix part
number 070-8992-XX) provides information for maintaining and servicing
the digitizing oscilloscope to the module level.
Two types of Performance Verification procedures can be performed on this
product; Brief Procedures and Performance Tests. You may not need to perform
all of these procedures, depending on what you want to accomplish.
HTo rapidly confirm that the oscilloscope functions and was adjusted properly,
just do the brief procedures under Self Tests, which begin on page 1–5.Advantages: These procedures are quick to do, require no external
equipment or signal sources, and perform extensive functional and accuracy
testing to provide high confidence that the oscilloscope will perform
properly. They can be used as a quick check before making a series of
important measurements.
HTo further check functionality, first do the Self Tests just mentioned; then do
the brief procedures under Functional Tests that begin on page 1–7.Advantages: These procedures require minimal additional time to perform,
require no additional equipment other than a standard-accessory probe, and
more completely test the internal hardware of the oscilloscope. They can be
used to quickly determine if the oscilloscope is suitable for putting into
service, such as when it is first received.
HIf more extensive confirmation of performance is desired, do the Perform-
ance Tests, beginning on page 1–15, after doing the Functional and Self Tests
just referenced.
Advantages: These procedures add direct checking of warranted specifica-
tions. They require more time to perform and suitable test equipment is
required. (See Equipment Required beginning on page 1–16.)
If you are not familiar with operating this oscilloscope, read the TDS 684A, TDS
744A, & TDS 784A Reference (070-8999-XX) or the TDS 684A, TDS 744A, &
TDS 784A User Manual (070-8991-XX). These contain instructions that will
acquaint you with the use of the front-panel controls and the menu system.
Throughout these procedures the following conventions apply:
HEach test procedure uses the following general format:
HEach procedure consists of as many steps, substeps, and subparts as required
Title of Test
Equipment Required
Prerequisites
Procedure
to do the test. Steps, substeps, and subparts are sequenced as follows:
1. First Step
a. First Substep
HFirst Subpart
HSecond Subpart
b. Second Substep
2. Second Step
HIn steps and substeps, the lead-in statement in italics instructs you what to
do, while the instructions that follow tell you how to do it, as in the example
step below, “Initialize the oscilloscope” by doing “Press save/recall SETUP.
Now, press the main-menu button...”.
Initialize the oscilloscope: Press save/recall SETUP. Now, press the
main-menu button Recall Factory Setup; then the side-menu button OKConfirm Factory Init.
HWhere instructed to use a front-panel button or knob, or select from a main
or side menu, or verify a readout or status message, the name of the button or
knob appears in boldface type: “press SHIFT; then UTILITY, press the
main-menu button System until Cal is highlighted in the pop-up menu.
Verify that the status message is Pass in the main menu under the VoltageReference label.”
STOP. The symbol at the left is accompanied by information you must read to do
the procedure properly.
1–2
HRefer to Figure 1–1: “Main menu” refers to the menu that labels the seven
menu buttons under the display; “side menu” refers to the menu that labels
The Self Tests use internal routines to confirm basic functionality and proper
adjustment. No test equipment is required to do these test procedures.
The Functional Tests utilize the probe-compensation output at the front panel as
a test-signal source for further verifying that the oscilloscope functions properly.
A probe, such as the P6245, is required to do these test procedures.
This procedure uses internal routines to verify that the oscilloscope functions and
was adjusted properly. No test equipment or hookups are required.
Verify Internal Adjustment,
Self Compensation, and
Diagnostics
Equipment
Required
PrerequisitesPower on the Digitizing Oscilloscope and allow a 20 minute warm-up
1. Verify that internal diagnostics pass: Do the following substeps to verify
passing of internal diagnostics.
a. Display the System diagnostics menu:
HPress SHIFT; then press UTILITY.
HRepeatedly press the main-menu button System until Diag/Err is
highlighted in the pop-up menu.
b. Run the System Diagnostics:
HFirst disconnect any input signals from all four channels.
HPress the main-menu button Execute; then press the side-menu
button OK Confirm Run Test.
c. Wait: The internal diagnostics do an exhaustive verification of proper
oscilloscope function. This verification will take up to three and a half
minutes on some models. At some time during the wait, a “clock” icon
(shown at left) will appear on-screen. When the verification is finished,
the resulting status will appear on the screen.
None
before doing this procedure.
d. Confirm no failures are found: Verify that no failures are found and
e. Confirm the three adjustment sections have passed status:
HPress SHIFT; then press UTILITY.
HHighlight Cal in the pop-up menu by repeatedly pressing the
main-menu button System. See Figure 1–2.
HVerify that the word Pass appears in the main menu under the
following menu labels: Voltage Reference, Frequency Response,
and Pulse Trigger. See Figure 1–2.
f.Run the signal-path compensation: Press the main-menu button Signal
Path; then press the side-menu button OK Compensate Signal Paths.
g. Wait: Signal-path compensation may take five minutes on the
TDS 7XXA or fifteen minutes on the 684A to run. While it progresses, a
“clock” icon (shown at left) is displayed on-screen. When compensation
completes, the status message will be updated to Pass or Fail in the
main menu. See step h.
h. Confirm signal-path compensation returns passed status: Verify that the
word Pass appears under Signal Path in the main menu. See Figure 1–2.
1–6
Highlight Cal
1
3
Verify Pass
2
Verify Pass
Figure 1–2: Verifying Adjustments and Signal-Path Compensation
2. Return to regular service: Press CLEAR MENU to exit the system menus.
The purpose of these procedures is to confirm that the oscilloscope functions
properly. The only equipment required is one of the standard-accessory probes
and, to check the file system, a 3.5 inch, 720 K or 1.44 Mbyte floppy disk.
CAUTION. The optional P6245 probes that can be used with this oscilloscope
provide an extremely low loading capacitance (<1 pF) to ensure the best
possible signal reproduction. These probes should not be used to measure
signals exceeding ±8 volts, or errors in signal measurement will be observed.
Above 40 volts, damage to the probe may result. To make measurements beyond
±10 volts, use either the P6139A probe (good to 500 volts), or refer to the
catalog for a recommended probe.
STOP. These procedures verify functions; that is, they verify that the oscilloscope
features operate. They do not verify that they operate within limits.
Therefore, when the instructions in the functional tests that follow call for you to
verify that a signal appears on-screen “that is about five divisions in amplitude”
or “has a period of about six horizontal divisions,” etc., do NOT interpret the
quantities given as limits. Operation within limits is checked in Performance
Tests, which begin on page 1–15.
STOP. DO NOT make changes to the front-panel settings that are not called out
in the procedures. Each verification procedure will require you to set the
oscilloscope to certain default settings before verifying functions. If you make
changes to these settings, other than those called out in the procedure, you may
obtain invalid results. In this case, just redo the procedure from step 1.
When you are instructed to press a menu button, the button may already be
selected (its label will be highlighted). If this is the case, it is not necessary to
press the button.
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground (typically black) to PROBE COMPENSA-TION GND. If using a P6245 probe, you may want to attach a Y-lead
connector and two SMD KlipChips as shown in Figure 1–3.
1–8
SignalGnd
Figure 1–3: Universal Test Hookup for Functional Tests – TDS 684A Shown
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
2. Verify that all channels operate: Do the following substeps — test CH 1
first, skipping substep a and b since CH 1 is already set up for verification
and as the trigger source from step 1.
a. Select an unverified channel:
HPress WAVEFORM OFF to remove the channel just verified from
display.
HPress the front-panel button that corresponds to the channel you are
to verify.
HMove the probe to the channel you selected.
b. Match the trigger source to the channel selected:
HPress TRIGGER MENU.
HPress the main-menu button Source.
HPress the side-menu button that corresponds to the channel selected
(Ch2, Ch3, or Ch4).
c. Set up the selected channel:
HSet the vertical SCALE to 200 mV.
HSet the horizontal SCALE to 200 s. Press CLEAR MENU to
remove any menu that may be on the screen.
d. Verify that the channel is operational: Confirm that the following
statements are true.
HThe vertical scale readout for the channel under test shows a setting
of 200 mV, and a square-wave probe-compensation signal about
2.5 divisions in amplitude is on-screen. See Figure 1–1 on page 1–3
to locate the readout.
HThe vertical POSITION knob moves the signal up and down the
screen when rotated.
HTurning the vertical SCALE knob counterclockwise decreases the
amplitude of the waveform on-screen, turning the knob clockwise
increases the amplitude, and returning the knob to 200 mV returns
the amplitude to about 2.5 divisions.
e. Verify that the channel acquires in all acquisition modes: Press SHIFT;
then press ACQUIRE MENU. Use the side menu to select, in turn, each
of the three hardware acquire modes and confirm that the following
statements are true. Refer to the icons at the left of each statement as you
confirm those statements.
a. One period of the square-wave probe-compensation signal is about five
horizontal divisions on-screen for the 200 s horizontal scale setting (set
in step 1c).
b. Rotating the horizontal SCALE knob clockwise expands the waveform
on-screen (more horizontal divisions per waveform period), counterclockwise rotation contracts it, and returning the horizontal scale to
200 s returns the period to about five divisions.
c. The horizontal POSITION knob positions the signal left and right
on-screen when rotated.
3. Remove the test hookup: Disconnect the probe from the channel input and
the probe-compensation terminals.
Verify the Main and
Delayed Trigger Systems
Equipment
Required
PrerequisitesNone
One probe such as the P6245 or P6139A
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground to PROBE COMPENSATION GND. See
Figure 1–3 on page 1–8.
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
c. Modify default settings:
HSet the vertical SCALE to 200 mV.
HSet the horizontal SCALE for the M (main) time base to 200 s.
HPress TRIGGER MENU.
HPress the main-menu button Mode & Holdoff.
HPress the side-menu button Normal.
HPress CLEAR MENU to remove the menus from the screen.
HVerify that the trigger READY indicator on the front panel flashes
about once every second as the waveform is updated on-screen.
4. Remove the test hookup: Disconnect the probe from the channel input and
the probe-compensation terminals.
Verify the File System
Equipment
Required
PrerequisitesNone
One probe such as the P6245 or P6139A
One 720 K or 1.44 Mbyte, 3.5 inch DOS-compatible disk.
You can use a disk of your own or you can use the Programming
Examples Software 3.5 inch disk (Tektronix part number 063-1 134-XX)
contained in the TDS Family Programmer Manual (Tektronix part
number 070-8709-XX).
1. Install the test hookup and preset the oscilloscope controls:
a. Hook up the signal source: Install the probe on CH 1. Connect the probe
tip to PROBE COMPENSATION SIGNAL on the front panel;
connect the probe ground to PROBE COMPENSATION GND. See
Figure 1–3 on page 1–8.
b. Insert the test disk: Insert the disk in the disk drive to the left of the
monitor.
HPosition the disk so the metal shutter faces the drive.
HPosition the disk so the stamped arrow is on the top right side. In
other words, place the angled corner in the front bottom location.
HPush the disk into the drive until it goes all the way in and clicks
into place.
c. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
d. Modify default settings:
HSet the vertical SCALE to 200 mV.
HSet the horizontal SCALE for the M (main) time base to 200 s.
Notice the waveform on the display now shows two cycles instead of
five.
HPress CLEAR MENU to remove the menus from the screen.
HPress SETUP.
HPress the main-menu button Save Current Setup; then press the
side-menu button To File.
HTurn the general purpose knob to select the file to save. Choose
(or ). With this choice, you will save a file
starting with , then containing 5-numbers, and a extension.
For example, the first time you run this on a blank, formatted disk or
on the Example Programs Disk, the oscilloscope will assign the
name to your file. If you ran the procedure again, the
oscilloscope would increment the name and call the file
.
HPress the side-menu button Save To Selected File.
2. Verify the file system works:
HPress the main-menu button Recall Factory Setup and the side-menu
button OK Confirm Factory Init to restore the 500 s time base and
the five cycle waveform.
HPress the main-menu button Recall Saved Setup; then press the
side-menu button From File.
HTurn the general purpose knob to select the file to recall. For example, if
you followed the instructions above and used a blank disk, you had the
oscilloscope assign the name TEK00000.SET to your file.
HPress the side-menu button Recall From Selected File.
HVerify that Digitizing Oscilloscope retrieved the saved setup from the
disk. Do this by noticing the horizontal SCALE for the M (main) time
base is again 200 s and the waveform shows only two cycles just as it
was when you saved the setup.
3. Remove the test hookup:
HDisconnect the probe from the channel input and the probe-compensa-
tion terminals.
HRemove the disk from the disk drive. Do this by pushing in the tab at the
This section contains a collection of procedures for checking that the TDS 684A
and 7XXA Digitizing Oscilloscopes perform as warranted.
The procedures are arranged in four logical groupings: Signal Acquisition System
Checks, Time Base System Checks, Triggering System Checks, and Output Ports
Checks. They check all the characteristics that are designated as checked in
Specifications. (The characteristics that are checked appear in boldface type
under Warranted Characteristics in Specifications.)
STOP. These procedures extend the confidence level provided by the basic
procedures described on page 1–5. The basic procedures should be done first,
then these procedures performed if desired.
Prerequisites
The tests in this section comprise an extensive, valid confirmation of performance and functionality when the following requirements are met:
HThe cabinet must be installed on the Digitizing Oscilloscope.
HYou must have performed and passed the procedures under Self Tests, found
on page 1–5, and those under Functional Tests, found on page 1–7.
HA signal-path compensation must have been done within the recommended
calibration interval and at a temperature within ±5_ C of the present
operating temperature. (If at the time you did the prerequisite Self Tests, the
temperature was within the limits just stated, consider this prerequisite met.)
HThe Digitizing Oscilloscope must have been last adjusted at an ambient
temperature between +20_ C and +30_ C, must have been operating for a
warm-up period of at least 20 minutes, and must be operating at an ambient
temperature between +4_ C and either +45_ C for the TDS 684A or +50_ C
for the TDS 7XXA. (The warm-up requirement is usually met in the course
of meeting the first prerequisite listed above.)
These procedures use external, traceable signal sources to directly check
warranted characteristics. The required equipment list follows this introduction.
16. Generator, Video SignalProvides PAL compatible outputsTektronix TSG 121Used to Test Video Option 05
17. Oscillator, Leveled Sine
wave Generator
18. Pulse GeneratorTektronix part number PG 502 Used to Test Video Option 05
19. Cable, Coaxial
(two required)
20. Terminator, 75
(two required)
1
Requires a TM 500 or TM 5000 Series Power Module Mainframe.
2
Warning: The optional P6245 probes that may be used with this oscilloscope provide an extremely low loading
capacitance (<1 pF) to ensure the best possible signal reproduction. These probes should not be used to measure
signals exceeding ±8 V , or errors in signal measurement will be observed. Above 40 V, damage to the probe may result.
To make measurements beyond ±8 V, use either the P6139A probe (good to 500 V), or refer to the catalog for a recommended probe.
60 Hz. Sine waveTektronix part number SG 502 Used to Test Video Option 05
75 , 36 in, male to male BNC
connectors
Impedance 75 ; connectors: female
BNC input, male BNC output
Programming Examples
Software Disk (Tektronix part
number 063-1134-XX) that
comes with the TDS Family
Programmer Manual (Tektronix part number
070-8709-XX)
Tektronix part number
012-1338-00
Tektronix part number
01 1-0102-01
Checking File System Basic
Functionality
Equipped Instruments Only
Equipped Instruments Only
Equipped Instruments Only
Used to Test Video Option 05
Equipped Instruments Only
Used to Test Video Option 05
Delay Between Channels
Delay Between ChannelsN/A____________________50 ps
Time Base System
Long Term Sample Rate/
Delay Time @ 100 ns/10.0 ms
Trigger System Accuracy
Pulse-Glitch or Pulse-Width,
Hor. scale ≤ 1 ms
Lower Limit
Upper Limit
Pulse-Glitch or Pulse-Width,
Hor. scale > 1 ms
Lower Limit
Upper Limit
Main Trigger, DC Coupled, Positive Slope9.9393 V____________________10.1147 V
Main Trigger, DC Coupled, Negative Slope9.9393 V____________________10.1147 V
Delayed Trigger, DC Coupled, Positive Slope9.9393 V____________________10.1 147 V
Delayed Trigger, DC Coupled, Negative Slope9.9393 V____________________10.1147 V
CH1 Sensitivity, 50 MHz, MainPass/Fail____________________Pass/Fail
CH1 Sensitivity , 50 MHz, DelayedPass/Fail____________________Pass/Fail
CH1 AUX Trigger InputPass/Fail____________________Pass/Fail
CH1 Sensitivity , full bandwidth, MainPass/Fail____________________Pass/Fail
CH1 Sensitivity , full bandwidth, DelayedPass/Fail____________________Pass/Fail
Output Signal Checks
MAIN TRIGGER OUTPUT, 1 MW
These procedures check those characteristics that relate to the signal-acquisition
system and are listed as checked under Warranted Characteristics in Specifica-
tions.
Performance Tests
Check Offset Accuracy
(Zero Setting)
Equipment
Required
PrerequisitesThe oscilloscope must meet the prerequisites listed on page 1–15
1. Preset the instrument controls:
a. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
HPress CLEAR MENU to remove the menus from the screen.
b. Modify the default settings:
HPress SHIFT; then press ACQUIRE MENU.
HOn the TDS 684A, press the main-menu button Mode; then press
the side-menu button Average 16.
HOn the TDS 7XXA, press the main-menu button Mode; then press
the side-menu button Hi Res.
None
HPress CURSOR.
HPress the main-menu button Function; then press the side-menu
button H Bars.
HPress CLEAR MENU.
HBe sure to disconnect any input signals from all four channels.
2. Confirm input channels are within limits for offset accuracy at zero offset:
Do the following substeps — test CH 1 first, skipping substep a since CH 1
is already set up to be checked from step 1.
a. Select an unchecked channel: Press WAVEFORM OFF to remove the
channel just confirmed from the display. Then, press the front-panel
button that corresponds to the channel you are to confirm.
Figure 1–4: Measurement of DC Offset Accuracy at Zero Setting
g. Test all channels: Repeat substeps a through f for all input channels.
3. Disconnect the hookup: No hookup was required.
WARNING. The generator is capable of outputting dangerous voltages. Be sure to
set the DC calibration generator to 0 volts before connecting, disconnecting,
and/or moving the test hookup during the performance of this procedure.
Equipment
Required
PrerequisitesThe oscilloscope must meet the prerequisites listed on page 1–15
Two dual-banana connectors (Item 6)
One BNC T connector (Item 7)
One DC calibration generator (Item 9)
Two precision coaxial cables (Item 5)
1. Install the test hookup and preset the instrument controls:
a. Hook up the test-signal source:
HSet the output of a DC calibration generator to 0 volts.
HConnect the output of a DC calibration generator through a
dual-banana connector followed by a 50 W precision coaxial cable to
one side of a BNC T connector. See Figure 1–5.
HConnect the Sense output of the generator through a second
dual-banana connector followed by a 50 W precision coaxial cable to
the other side of the BNC T connector. Now connect the BNC T
connector to CH 1. See Figure 1–5.
Output Sense
DC Calibrator
Dual Banana to
BNC Adapters
Figure 1–5: Initial T est Hookup
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
c. Modify the default settings:
50 Coaxial Cables
Digitizing Oscilloscope
BNC T
Connector
1–30
HPress SHIFT; then press ACQUIRE MENU.
HPress the main-menu button Mode; then press the side-menu button
Average 16.
2. Confirm input channels are within limits for DC accuracy at maximum offset
and position: Do the following substeps — test CH 1 first, skipping substep
2a since CH 1 is already selected from step 1.
HUse the keypad to set vertical offset to the positive-polarity setting
listed in the table for the current vertical scale setting. The baseline
level will remain off screen.
HSet the generator to the level and polarity indicated in the table for
the vertical scale, position, and offset settings you have made. The
DC test level should appear on screen. (If it doesn’t return, the DC
accuracy check is failed for the current vertical scale setting of the
current channel.)
e. Measure the test signal: Press CLEAR MENU. Read the measurement
results at the Mean measurement readout. See Figure 1–6.
Turn on the
measurement called
mean and read the
results here.
Figure 1–6: Measurement of DC Accuracy at Maximum Offset and Position
f.Check against limits:
HCHECK that the readout for the measurement Mean readout on
screen is within the limits listed for the current vertical scale and
position/offset/generator settings. Enter value on test record.
HRepeat substep d, reversing the polarity of the position, offset, and
generator settings as is listed in the table.
1–32
HCHECK that the Mean measurement readout on screen is within the
limits listed for the current vertical scale setting and position/offset/
generator settings. Enter value on test record.
HRepeat substeps c through f until all vertical scale settings, listed in
Table 1–3, are checked for the channel under test.
g. Test all channels: Repeat substeps a through f for all four channels.
3. Disconnect the hookup:
a. Set the generator output to 0 V.
b. Disconnect the cable from the generator output at the input connector of
the channel last tested.
Check Analog Bandwidth
Equipment
Required
PrerequisitesSee page 1–15
One high-frequency leveled sine wave generator and its leveling head
(Item 12)
Two10X attenuators (Item 1)
1. Install the test hookup and preset the instrument controls:
a. Initialize the oscilloscope:
HPress save/recall SETUP. Press the main-menu button Recall
Factory Setup; then press the side-menu button OK Confirm
Factory Init.
b. Modify the default settings:
HPress TRIGGER MENU. Press the main-menu button Coupling.
HPress the side-menu button Noise Rej.
HTurn the horizontal SCALE knob to 50 ns. Press SHIFT; then press
ACQUIRE MENU.
HPress the main-menu button Mode; then press the side-menu button
Average 16.
HPress MEASURE. Press the main-menu button High–Low Setup;
then press the side-menu button Min–Max.
c. Hook up the test-signal source: Connect, through its leveling head, the
sine wave output of a high-frequency leveled sine wave generator to
CH 1. Set the output of the generator to a reference frequency of 6 MHz.
See Figure 1–7.
e. Display the test signal: Do the following subparts to first display the
reference signal and then the test signal.
HPress MEASURE; then press the main-menu button Select
Measrmnt for CHx.
HPress the side-menu button more until the menu label Pk-Pk
appears in the side menu (its icon is shown at the left). Press the
side-menu button Pk-Pk.
HRepeatedly press the side-menu button –more– until Frequency
appears in the side menu (its icon is shown at the left). Press the
side-menu button Frequency.
HPress CLEAR MENU.
HSet the generator output so the CHx Pk-Pk readout equals the
reference amplitude in Table 1–4 that corresponds to the vertical
scale set in substep d.
HPress the front-panel button SET LEVEL TO 50% as necessary to
trigger a stable display. At full bandwidth, which for the TDS
684A/784A = 1 GHz and for the 744A = 500 MHz frequency, you
may also want to make small, manual adjustments to the trigger
level. You can use the TRIGGER LEVEL knob to do this.
STOP. Checking each channel’s bandwidth at all vertical scale settings is time
consuming and unnecessary. You may skip checking the remaining vertical scale
settings in Table 1–4 (that is, skip the following substep, h) if this digitizing
oscilloscope has performed as follows:
HPassed the 100 mV vertical scale setting just checked in this
procedure.
HPassed the Verify Internal Adjustment, Self Compensation, and
Diagnostics procedure found under Self Tests, on page 1–5.
NOTE. Passing the signal path compensation confirms the signal path for all
vertical scale settings for all channels. Passing the internal diagnostics ensures
that the factory-set adjustment constants that control the bandwidth for each
vertical scale setting have not changed.
h. Check remaining vertical scale settings against limits (optional):
Check Delay Between
Channels
HIf desired, finish checking the remaining vertical scale settings for
the channel under test by repeating substeps d through g for each of
the remaining scale settings listed in Table 1–4 for the channel under
test.
HWhen doing substep e, skip the subparts that turn on the CHx Pk-Pk
measurement until you check a new channel.
HInstall/remove 10X attenuators between the generator leveling head
and the channel input as needed to obtain the six division reference
signals listed in the table.
i.Test all channels: Repeat substeps a through g for all four channels.
3. Disconnect the hookup: Disconnect the test hook up from the input
connector of the channel last tested.
Equipment
Required
PrerequisitesSee page 1–15
One medium-frequency leveled sine wave generator (Item 1 1)
One precision coaxial cable (Item 5)
One 50 terminator (Item 4)
One dual-input coupler (Item 8)
2. Confirm all four channels (CH 1 through CH 4) are within limits for channel
delay:
a. Set up the generator: Set the generator frequency to 250 MHz and the
amplitude for about six divisions in CH 1.
Hint: As you are adjusting the generator amplitude, push SET LEVEL
TO 50% frequently to speed up the updating of the waveform amplitude
on screen.
b. The horizontal SCALE should already be set to 500 ps. On the TDS
784A, now set it to 200 ps. On the TDS 744A, push the front-panel
ZOOM button, press the side-menu On button, set the horizontal
SCALE to 250 ps, and be sure the vertical scale factor is kept at 1.0X
and the horizontal scale factor is 2.0X.
c. Save a CH 2 waveform: Press CH 2. Be sure the vertical scale factor is
kept at 1.0X. Then press save/recall WAVEFORM. Now, press the
main-menu button Save Wfm; then press the side-menu button
To Ref 2.
d. Save CH 3 waveform:
HMove the coupler from CH 2 to CH 3, so that CH 1 and CH 3 are
driven. Press WAVEFORM OFF. Press CH 3. Be sure the vertical
scale factor is kept at 1.0X. Then press the side-menu button
To Ref 3.
e. Display all test signals:
HPress WAVEFORM OFF to remove CH 3 from the display.
HDisplay the live waveform. Move the coupler from CH 3 to CH 4,
so that CH 1 and CH 4 are driven. Press CH 4 to display. Be sure
the vertical scale factor is kept at 1.0X. See Figure 1–10 on page
1–40.
HDisplay the reference waveforms. To do this, press the front-panel
button MORE. Press the main-menu buttons Ref 2 and Ref 3. You
may notice their overlapping ground reference indicators. See Figure
1–10 on page 1–40.
f.Measure the test signal:
HLocate the time reference points for these waveforms. Do this by
first identifying the point where the rising edge of the left-most
waveform crosses the center horizontal graticule line. Next, note the
corresponding time reference point for the right-most waveform. See
Figure 1–10 on page 1–40.
HPress the main-menu button Function; then press the side-menu
button V Bars.
HPress CLEAR MENU.
HAlign one V bar cursor to the time reference point of the left-most
waveform edge and the other cursor to the time reference point of the
right-most waveform edge by rotating the General Purpose knob.
(Press SELECT to switch between the two cursors.) See Figure
1–10 on page 1–40.
HRead the measurement results at the D: cursor readout, not the @:
readout on screen.
Read results.
Display the waveforms.
4
1
2
Locate the time reference
points for these waveforms.
Align each cursor to the time
3
reference points
Figure 1–10: Measurement of Channel Delay – TDS 684A Shown
1–40
g. Check against limits: CHECK that the cursor readout on screen is
≤100 ps for the TDS 684A or ≤50 ps for the TDS 744A/784A.
h. If the channel skew is within the limits, enter time on the test record and
proceed to step 3. Otherwise, proceed with steps i through p.
i.Use the cursors to measure the skew from CH1 to CH2, CH1 to CH3,
and CH1 to CH4. Write down these three numbers in the first measure-
ment column of Table 1–5. Note that these numbers may be either
positive or negative.
j.Repeat the procedure from step 1.c through 2.e., 4th bullet item.
k. Again use the cursors to measure the skew from CH1 to CH2, CH1 to
CH3, and CH1 to CH4. Write down these numbers in the second
measurement column of Table 1–5. Note that these numbers may be
either positive or negative.
l.Add the first CH1 to CH2 skew measurement to the second CH1 to CH2
skew measurement and divide the result by 2. Use Table 1–5.
m. Add the first CH1 to CH3 skew measurement to the second CH1 to CH3
skew measurement and divide the result by 2. Use Table 1–5.
n. Add the first CH1 to CH4 skew measurement to the second CH1 to CH4
skew measurement and divide the result by 2. Use Table 1–5.
o. Check against limits: CHECK that the largest of the three results from
steps l, m, and n is between –100 ps and + 100 ps for the TDS 684A or
between –50 ps and + 50 ps for the TDS 744A/784A.
p. Enter time on the test record.
T able 1–5: Delay Between Channels Worksheet
Add First and
Coupling
CH1 to CH2
skew
CH1 to CH3
skew
CH1 to CH4
skew
First
Measurement
Second
Measurement
Second
Measurements
Divide Sum
by 2
3. Disconnect the hookup: Disconnect the cable from the generator output at
These procedures check those characteristics that relate to the Main and Delayed
time base system and are listed as checked under Warranted Characteristics in
Specifications.
Check Accuracy for
Long-Term Sample Rate,
Delay Time, and Delta
Time Measurements
Equipment
Required
PrerequisitesSee page 1–15
Time Mark
Generator
One time-mark generator (Item 13)
One 50 , precision coaxial cable (Item 5)
Digitizing Oscilloscope
Output
50 Coaxial Cable
Figure 1–11: Initial Test Hookup
1. Install the test hookup and preset the instrument controls:
a. Hook up the test-signal source: Connect, through a 50 W precision
coaxial cable, the time-mark output of a time-mark generator to CH 1.
Set the output of the generator for 10 ms markers.
1–42
b. Initialize the oscilloscope:
HPress save/recall SETUP. Press the main-menu button Recall
Factory Setup. Press the side-menu button OK Confirm Factory
Init.
c. Modify the initialized front-panel control settings:
HSet the vertical SCALE to 500 mV.
HPress VERTICAL MENU; then press the main-menu button
Coupling. Press the side-menu button to toggle it to the 50
setting.
HPress SET LEVEL TO 50%.
HUse the vertical POSITION knob to center the test signal on screen.
1. Install the test hookup and preset the instrument controls:
a. Initialize the instrument:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
b. Modify the default setup:
HPress VERTICAL MENU.
HPress the main-menu button Coupling; then press the side-menu
button to select 50 coupling.
HSet the horizontal SCALE to 10 ns on the TDS 684A and 12.5 ns on
the TDS 7XXA.
c. Hook up the test-signal source: Connect the output of a
medium-frequency leveled sine wave generator (Item 11) to CH 1. Do
this through a 50 W precision coaxial cable, followed by a 10X
attenuator. See Figure 1–13.
Performance Tests
Medium-Frequency
Sine Wave Generator
Output
10X Attenuator
50 Coaxial Cable
Digitizing Oscilloscope
Figure 1–13: Initial T est Hookup
2. Confirm the trigger system is within time-accuracy limits for pulse-glitch or
pulse-width triggering (Horizontal Scale ≤1 s):
a. Display the test signal: Set the output of the sine wave generator for a
100 MHz, five-division sine wave on screen. Press SET LEVEL TO50%.
HPress HORIZONTAL MENU.
HPress the main-menu button Time Base.
HPress the side-menu buttons Delayed Only and Delayed Trigger-
able.
HSet D (delayed) horizontal SCALE to 500 s.
b. Select the Delayed trigger system:
HPress SHIFT; then press the front-panel DELAYED TRIG button.
HPress the main-menu button Level.
c. Measure the test signal: Press the side-menu button SET TO 50%.
Read the measurement results in the side (or main) menu below the label
Level.
d. Check against limits: Do the following subparts in the order listed.
HCHECK that the Level readout in the side menu is within 9.863 V to
10.137 V, inclusive, for the TDS 684A or is within 9.9393 V to
10.1147 V, inclusive, for the TDS 7XXA.
HEnter voltage on test record.
HPress the main-menu button Slope; then press the side-menu button
for negative slope. See icon at left. Press the main-menu button
Level. Repeat substep c.
HCHECK that the Level readout in the side menu is within 9.863 V to
10.137 V, inclusive, for the TDS 684A or is within 9.9393 V to
10.1147 V, inclusive, for the TDS 7XXA.
HEnter voltage on test record.
4. Disconnect the hookup:
a. First set the output of the DC calibration generator to 0 volts.
b. Disconnect the cable from the generator output at the input connector of
One medium-frequency leveled sine wave generator (Item 1 1)
One high-frequency leveled sine wave generator (Item 12)
Two precision 50 coaxial cables (Item 5)
One 10X attenuator (Item 1)
One BNC T connector (Item 7)
One 5X attenuator (Item 2)
1. Install the test hookup and preset the instrument controls:
a. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
b. Modify the initialized front-panel control settings:
HSet the horizontal SCALE for the M (main) time base to 20 ns on
the TDS 684A or 25 ns on the TDS 7XXA.
HPress HORIZONTAL MENU; then press the main-menu button
Time Base.
HPress the side-menu button Delayed Only; then press the side-menu
button Delayed Triggerable.
HSet the horizontal SCALE for the D (delayed) time base to 20 ns on
the TDS 684A or 25 ns on the TDS 7XXA; then press the side-menu
button Main Only.
HPress TRIGGER MENU; then press the main-menu button Mode
& Holdoff. Press the side-menu button Normal.
HPress VERTICAL MENU; then press the main-menu button
Coupling. Press the side-menu button to select the 50 W setting.
HPress SHIFT; then press ACQUIRE MENU. Press the main-menu
button Mode; then press the side-menu button Average 16.
c. Hook up the test-signal source:
HConnect the signal output of a medium-frequency sine wave
generator to a BNC T connector. Connect one output of the T
connector to CH 1 through a 50 W precision coaxial cable. Connect
triggered on the selected slope (positive or negative). This display
should not have its trigger point switching between opposite slopes,
nor should it roll across the screen. At horizontal scale settings of
2 ms/division and faster, TRIG’D will remain constantly lighted. It
will flash for slower settings.
HPress TRIGGER MENU; then press the main-menu button Slope.
HPress SET LEVEL TO 50%. Adjust the TRIGGER LEVEL knob
so that the TRIG’D light is on. Set the level to near the middle of
the range where the TRIG’D light is on. CHECK that the trigger is
stable for the test waveform on both the positive and negative slopes.
Use the side menu to switch between trigger slopes.
HEnter pass/fail result for main trigger on the test record.
HLeave the Main trigger system triggered on the positive slope of the
waveform before continuing to the next step.
1–54
Check if stable trigger.
Figure 1–18: Measurement of Trigger Sensitivity — 50 MHz Results Shown on a TDS
684A Screen
c. Check Delayed trigger system for stable triggering at limits: Do the
following subparts in the order listed.
HPress HORIZONTAL MENU; then press the main-menu button
Time Base. Press the side-menu button Delayed Only; then press
Delayed Triggerable in the same menu.
HPress SHIFT; then press DELAYED TRIG. Press the main-menu
button Level.
HPress the side-menu button SET TO 50%.
CHECK that a stable trigger is obtained for the test waveform for
both the positive and negative slopes of the waveform. Use the
TRIGGER LEVEL knob to stabilize the Main trigger. Use the
general purpose knob to stabilize the Delayed trigger. Press the
main-menu button Slope; then use the side menu to switch between
trigger slopes. See Figure 1–18.
HEnter pass/fail result for delayed trigger on the test record.
HLeave the Delayed trigger system triggered on the positive slope of
the waveform before continuing to the next step. Also, return to the
main time base: Press HORIZONTAL MENU; then press the
main-menu button Time Base. Press the side-menu button MainOnly.
HPress CLEAR MENU.
3. Confirm the AUX Trigger input:
a. Display the test signal:
HRemove the 10X attenuator and reconnect the cable to CH 1.
HSet the test signal amplitude for about 2.5 divisions on screen.
HNow fine adjust the generator output until the CH 1 Amplitude
readout indicates the amplitude is 250 mV. (Readout may fluctuate
around 250 mV.)
b. Check the AUX trigger source for stable triggering at limits: Do the
following in the order listed.
HUse the definition for stable trigger from step 2.
HPress TRIGGER MENU; then press the main-menu button Source.
HPress the side-menu button –more– until the side-menu label
HPress SET LEVEL TO 50%. CHECK that a stable trigger is
obtained for the test waveform on both the positive and negative
slopes. Press the main-menu button Slope; then use the side menu to
switch between trigger slopes. Use the TRIGGER LEVEL knob to
stabilize the trigger if required.
HEnter the pass/fail result on the test record.
HLeave the Main trigger system triggered on the positive slope of the
waveform before proceeding to the next check.
HPress the main-menu button Source; then press the side-menu button
–more– until CH 1 appears. Press CH 1.
4. Confirm that the Main and Delayed trigger systems are within sensitivity
limits (full bandwidth, for TDS 684A & TDS 784A = 1 GHz, for TDS 744A
= 500 MHz):
a. Hook up the test-signal source: Disconnect the hookup installed in
step 1. Connect, through its leveling head, the signal output of a
high-frequency leveled sine wave generator to CH 1.
b. Set the Main and Delayed Horizontal Scales:
HSet the horizontal SCALE to 500 ps for the M (Main) time base.
HPress HORIZONTAL MENU. Now press the main-menu button
Time base; then press the side-menu button Delayed Triggerable.
HPress the side-menu button Delayed Only.
HSet the horizontal SCALE to 500 ps for the D (Delayed) time base.
NOTE. You just checked the trigger sensitivity. If desired, you may repeat steps 1
through 4 for the other channels (CH2, CH3, and CH4).
5. Disconnect the hookup: Disconnect the cable from the channel last tested.
The procedure that follows checks those characteristics of the output signals that
are listed as checked under Warranted Characteristics in Specifications. The
oscilloscope outputs these signals at its front and rear panels.
Check Outputs — CH 3
Main and Delayed Trigger
Equipment
Required
PrerequisitesSee page 1–15. Also, the Digitizing Oscilloscope must have passed
Two 50 precision cables (Item 5)
One calibration generator (Item 10)
Check DC Voltage Measurement Accuracy on page 1–29.
1. Install the test hookup and preset the instrument controls:
Calibration
Generator
To MAIN
TRIGGER
OUT
Digitizing Oscilloscope
50 Coaxial Cables
Figure 1–19: Initial T est Hookup
a. Hook up test-signal source 1:
HConnect the standard amplitude output of a calibration generator
through a 50 W precision coaxial cable to CH 3.
HSet the output of the calibration generator to 0.500 V.
b. Hook up test-signal source 2: Connect the Main Trigger Out at the rear
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
d. Modify the initialized front-panel control settings:
HSet the horizontal SCALE to 200 ms.
HPress SHIFT; then press ACQUIRE MENU.
HPress the main-menu button Mode; then press the side-menu button
Average.
HSelect 64 averages.
2. Confirm Main and Delayed Trigger outputs are within limits for logic levels:
a. Display the test signal:
HPress WAVEFORM OFF to turn off CH 1.
HPress CH 2 to display that channel.
HSet the vertical SCALE to 1 V.
HUse the vertical POSITION knob to center the display on screen.
b. Measure logic levels:
HPress MEASURE; then press the main-menu button Select
Measurement for Ch2.
HSelect high and low measurements. To do this, repeatedly press the
side-menu button –more– until High and Low appear in the side
menu (their icons are shown at the left). Press both side-menu
buttons High and Low.
c. Check Main Trigger output against limits:
HCHECK that the Ch2 High readout is ≥2.5 volts and that the Ch2
Low readout is ≤0.7 volts. See Figure 1–20.
HEnter high and low voltages on test record.
HPress VERTICAL MENU; then press the main-menu button
Coupling. Now press the side-menu button to toggle it to the
50 W setting.
1–58
HCHECK that the Ch2 High readout is ≥1.0 volt and that the Ch2
Figure 1–20: Measurement of Main Trigger Out Limits
d. Check Delayed Trigger output against limits: See Figure 1–20.
HMove the precision 50 W cable from the rear-panel Main Trigger
Output BNC to the rear-panel Delayed Trigger Output BNC.
HCHECK that the Ch2 High readout is ≥1.0 volt and that the Ch2
Low readout ≤0.25 volts.
HEnter high and low voltages on test record.
HPress the side-menu button to select the 1 MW setting.
HPress CLEAR MENU.
HCHECK that the Ch2 High readout is ≥2.5 volts and that the Ch2
Low readout is ≤0.7 volts.
HEnter high and low voltages on test record.
3. Confirm CH 3 output is within limits for gain:
a. Measure gain:
HMove the precision 50 W cable from the rear-panel DELAYED
TRIGGER OUTPUT BNC to the rear-panel SIGNAL OUT BNC.
HPress the main-menu button Source.
HPress the side-menu button Ch3.
HSet vertical SCALE to 100 mV.
HPress SET LEVEL TO 50%.
HPress MEASURE; then press the main-menu button Select
Measrmnt for Ch2.
HRepeatedly press the side-menu button –more– until Pk-Pk appears
in the side menu (its icon is shown at the left). Press the side-menu
button Pk-Pk.
HPress CLEAR MENU.
b. Check against limits:
HCHECK that the readout Ch2 Pk-Pk is between 80 mV and 120 mV,
inclusive, for the TDS 684A or is between 88 mV and 132 mV,
inclusive, for the TDS 7XXA.
Check Probe
Compensator Output
HEnter voltage on test record.
HPress VERTICAL MENU; then press the side-menu button to
toggle to the 50 W setting.
HPress CLEAR MENU.
HCHECK that the readout Ch2 Pk-Pk is between 40 mV and 60 mV,
inclusive, for the TDS 684A or is between 44 mV and 66 mV,
inclusive, for the TDS 7XXA.
HEnter voltage on test record.
4. Disconnect the hookup: Disconnect the cables from the channel inputs and
the rear panel outputs.
Equipment
Required
PrerequisitesSee page 1–15. Also, the Digitizing Oscilloscope must have passed
One female BNC to clip adapter (Item 3)
Two dual-banana connectors (Item 6)
One BNC T connector (Item 7)
Two 50 precision cables (Item 5)
One DC calibration generator (Item 9)
Check Accuracy For Long-T erm Sample Rate, Delay Time, and Delta
Time Measurements on page 1–42.
1. Install the test hookup and preset the instrument controls:
a. Hook up test-signal:
HConnect one of the 50 W cables to CH 1. See Figure 1–21.
HConnect the other end of the cable just installed to the female
BNC-to-clips adapter. See Figure 1–21.
HConnect the red clip on the adapter just installed to the PROBE
COMPENSATION SIGNAL on the front panel; connect the blackclip to PROBE COMPENSATION GND. See Figure 1–21.
Digitizing Oscilloscope
Black Lead
to GND
Performance Tests
Female BNC to
Clip Adapter
50 Coaxial Cable
Figure 1–21: Initial T est Hookup
b. Initialize the oscilloscope:
HPress save/recall SETUP.
HPress the main-menu button Recall Factory Setup.
HPress the side-menu button OK Confirm Factory Init.
c. Modify the initialized front-panel control settings:
HSet the horizontal SCALE to 200 ms.
HPress SET LEVEL TO 50%.
HUse the vertical POSITION knob to center the display on screen.
HPress SHIFT; then press ACQUIRE MENU.
HPress the main-menu button Mode; then press the side-menu button
HPress SAVE/RECALL WAVEFORM; then press the main-menu
button Save Wfm Ch 1.
HPress the side-menu button To Ref 1 to save the probe compensation
signal in reference 1.
HDisconnect the cable from CH 1 and the clips from the probe
compensation terminals.
HPress MORE; then press the main-menu button Ref 1 to displayed
the stored signal.
HPress CH 1.
d. Hook up the DC standard source:
HSet the output of a DC calibration generator to 0 volts.
HConnect the output of a DC calibration generator through a
dual-banana connector followed by a 50 W precision coaxial cable to
one side of a BNC T connector. See Figure 1–23.
HConnect the Sense output of the generator through a second
dual-banana connector followed by a 50 W precision coaxial cable to
the other side of the BNC T connector. Now connect the BNC T
connector to CH 1. See Figure 1–23.
DC Calibrator
Dual Banana to
BNC Adapters
50 Coaxial Cables
Figure 1–23: Subsequent T est Hookup
e. Measure amplitude of the probe compensation signal:
HPress SHIFT; then press ACQUIRE MENU. Press the side-menu
button AVERAGE then enter 16 using the keypad or the general
purpose knob.
PrerequisitesSee page 1–15. These prerequisites include running the signal path
PAL signal source (Item 16)
60 Hz. sine wave generator (Item 17)
Pulse generator (Item 18)
Two 75 cables (Item 19)
Two 75 terminators (Item 20)
One BNC T connector (Item 7)
50 cable (Item 5)
50 terminator (Item 4)
compensation routine.
1. Set up digitizing oscilloscope to factory defaults by completing the following
steps:
a. Press save/recall SETUP.
b. Press the main-menu Recall Factory Setup.
c. Press the side-menu OK Confirm Factory Init.
d. Wait for the Clock Icon to leave the screen.
e. CONFIRM the digitizing oscilloscope is setup as shown below.
2. Set up digitizing oscilloscope for TV triggers by completing the following
steps:
a. Press TRIGGER MENU.
b. Press the main-menu Type pop-up until you select Video.
c. Press the main-menu Standard pop-up until you select 625/PAL.
d. Press the main-menu Line.
e. Use the keypad to set the line number to 7 (press 7, then ENTER).
f.Press VERTICAL MENU.
g. Press the main-menu Bandwidth.
h. Select 250 MHz from the side menu.
i.Press the main-menu Fine Scale.
j.Use the keypad to set the fine scale to 282mV (press 282, SHIFT, m,
then ENTER).
k. Press HORIZONTAL MENU.
l.Press the main-menu Horiz Scale.
m. Use the keypad to set the horizontal scale to 200 ns (press 200, SHIFT,
n, then ENTER).
3. Check Jitter vs. Signal Amplitude
a. Set up equipment for Jitter Test. See Figure 1–25.
HConnect one of the rear panel composite outputs marked COMPST
on the TSG121 through a 75 W cable and a 75 W terminator to the
CH1 input of the TDS.
HPress the 100% FIELD control (the fourth TSG121 front-panel
button from the left) of the PAL signal source.
PAL Signal
Source
COMPST
75 Cable
Digitizing Oscilloscope
75
Terminator
Figure 1–25: Jitter T est Hookup
b. CHECK that the oscilloscope lights up its front panel TRIG’D LED and
it displays the waveform on screen. See Figure 1–26.
Figure 1–26: Jitter T est Displayed Waveform – TDS 684A Shown
c. Press SHIFT; then press ACQUIRE MENU.
d. Press the main-menu Mode.
e. Select the side-menu Average. It should be already set to 16.
f.Press the main-menu Create Limit Test Template.
g. Press the side-menu V Limit.
h. Use the keypad to set V Limit to 180 mdiv (press 180, SHIFT, m, then
ENTER)
i.Press the side-menu OK Store Template.
j.Press MORE.
k. Press the main-menu Ref1.
l.Press CH1.
m. Press SHIFT; then press ACQUIRE MENU.
n. Press the main-menu Limit Test Setup.
o. Toggle the side-menu Limit Test to ON.
p. Toggle the side-menu Ring Bell if Condition Met to ON.
q. Press the main-menu Mode.
r. Press the side-menu Envelope.
s.Use the keypad to set envelope to use 100 acquisitions (press 100, then
ENTER).
t.Press the main-menu Stop After button.
u. Press the side-menu Single Acquisition Sequence.
v. Confirm that the oscilloscope successfully makes 100 acquisitions. If not
successful, the oscilloscope bell will ring. When the word Run in the top
left corner of the display changes to STOP, the test is complete. See
Figure 1–27.
Stop shows test complete
Figure 1–27: Jitter Test When Completed – TDS 684A Shown
w. Press the main-menu Limit Test Setup.
x. Toggle the side-menu Ring Bell if Condition Met to OFF.
y. Toggle the side-menu Limit Test to OFF.
1–68
4. Check Triggered Signal Range.
Set up oscilloscope for Triggered Signal Test.
a. Press MORE.
b. Press WAVEFORM OFF.
c. Press HORIZONTAL MENU.
d. Use the keypad to set horizontal scale (/div) to 50 s (press 50, SHIFT,
m, then ENTER).
e. Press SHIFT; then press ACQUIRE MENU.
f.Press the main-menu Stop After.
g. Press the side-menu RUN/STOP button only.
h. Press the main-menu Mode.
i.Press the side-menu Sample.
j.Press RUN/STOP.
k. Press VERTICAL MENU.
l.Use the keypad to set fine scale to 300 mV (press 300, SHIFT, m, then
ENTER).
m. CONFIRM that the TRIG’D LED stays lighted and that the waveform
on screen is stable. That is, it does not move horizontally or vertically.
Also, CONFIRM that the waveform on the screen has one positive pulse
n. Use the keypad to set the fine scale to 75 mV (press 75, SHIFT, m, then
ENTER).
o. CONFIRM that the TRIG’D LED stays lighted and that the waveform
on screen is stable. That is, it does not move horizontally or vertically.
Also, CONFIRM that the waveform on the screen has one positive pulse
and a number of negative pulses. See Figure 1–29.
p. Disconnect all test equipment (TSG121) from the digitizing oscillo-
scope.
5. Check 60 Hz Rejection.
a. Set up oscilloscope for 60 Hz Rejection Test.
HUse the keypad to set the Ch1 Fine Scale to 282 mV (press 282,
SHIFT m, then ENTER).
HPress WAVEFORM OFF.
HPress CH2.
HPress VERTICAL MENU.
HUse the keypad set the fine scale to 2 V (press 2, then ENTER).
HPress HORIZONTAL MENU.
HUse the keypad to set the horizontal scale (/div) to 5 ms (press 5,
SHIFT, m, then ENTER).
b. Set up 60 Hz signal generator (SG 502).
HConnect the output of the SG 502 to the CH2 input through a 50 W
COMPST on the TSG 121) to a 75 W cable and a 75 W terminator.
Connect both signals to the CH1 input through a BNC T. See
Figure 1–32.
HPress VERTICAL MENU.
HIf needed, press the main-menu Fine Scale.
HUse the keypad to set fine scale to 500 mV (press 500, SHIFT, m,
then ENTER).
HConnect another composite signal connector of the PAL signal
source (labeled COMPST on the TSG 121) through a 75 W cable
and a 75 W terminator to the CH2 input. See Figure 1–32.
PAL Signal
Source
Signal Generator
COMPST
75 Cable
Output
COMPST
50 Cable
75 Cable
Figure 1–32: Subsequent 60 Hz Rejection T est Hookup
HCONFIRM that the TRIG’D LED stays lighted and that the
waveform on screen is stable. In other words, be sure the waveform
does not move horizontally or vertically. Also, confirm that the
waveform on the screen has one positive pulse and a number of
negative pulses. See Figure 1–33.
Digitizing Oscilloscope
BNC T
Connector
75
Terminators
HDisconnect all test equipment from the digitizing oscilloscope.
Figure 1–33: 60 Hz Rejection Test Result – TDS 684A Shown
6. Check Line Count Accuracy.
a. Set up oscilloscope for Line Count Accuracy Test.
HPress WAVEFORM OFF.
HPress CH1.
HPress HORIZONTAL MENU.
HPress the main-menu Record Length.
HPress the side-menu –more– until you see the appropriate menu.
HPress the side-menu 5000 points in 100divs.
HPress the main-menu Horiz Scale (/div).
HUse the keypad to set the horizontal scale to 200 ns (press 200,
SHIFT, n, then ENTER).
b. Check Line Count Accuracy.
HConnect a composite output signal from the rear of the PAL signal
source (labeled COMPST on the TSG 121) to the CH1 input
through a 75 W cable and a 75 W terminator. See Figure 1–34.
HPress the side-menu Fields.
HUse the keypad to set the number of fields to 1 (press 1, then
ENTER).
HPress the side-menu Sync Width.
HUse the keypad to set the width to 400 ns (press 400, SHIFT, n,
then ENTER).
HPress the side-menu –more– 1 of 2. Then press V1 Start Time.
HUse the keypad to set V1 start time to 10.10 s (press 10.10, SHIFT,
m, then ENTER).
HPress the side-menu V1 Stop Time.
HUse the keypad to set V1 stop time to 10.50 s (press 10.50, SHIFT,
m, then ENTER).
HPress the main-menu Type pop-up to select Edge.
HPress HORIZONTAL MENU.
HPress the main-menu Record Length.
HSelect the side-menu 1000 points in 20div. If needed, first press the
side-menu –more– until you see the appropriate side-menu item.
HTurn the HORIZONTAL POSITION knob to position the trigger
‘T’ two divisions to the left of the center screen.
HPress MEASURE.
HIf needed, press the main-menu Select Measrmnt.
HPress the side-menu Negative Width.
HPress the side-menu Period.
b. Set up the pulse generator (PG502) for Sync Duty Cycle Test.
HSet PULSE DURATION to 50 ns.
HSet PERIOD to 10 s.
HSet OUTPUT (VOLTS) to –1 for LOW LEVEL and +1 for HIGH
LEVEL.
1–78
HDepress the COMPLEMENT button.
HBe sure BACK TERM is depressed (in).
HTurn the pulse generator PULSE DURATION variable control to
adjust the negative pulse so the oscilloscope’s CH1 – Width
measurement displays 400ns +/–10 ns.
HTurn the HORIZONTAL SCALE knob to set the oscilloscope time
base to 5ms/div.
HTurn the pulse generator PERIOD variable control to adjust the
period until the oscilloscope CH1 Period measurement reads
21.000ms –25/+50 ns. See Figure 1–39. Read note shown below.
NOTE. The pulse duration and period adjustments are critical in making this
measurement. If the pulse duration and/or the duty cycle are not stable, the
FLEXFMT function may not function. You must take care when making these
adjustments.
This section begins with a general description of the traits of the TDS 684A and
7XXA Digitizing Oscilloscopes. Three sections follow, one for each of three
classes of traits: nominal traits, warranted characteristics, and typical character-istics.
The TDS 684A and 7XXA Digitizing Oscilloscopes are portable, four-channel
instruments suitable for use in a variety of test and measurement applications and
systems. Table 2–1 lists key features.
T able 2–1: Key Features of the TDS 684A and 7XXA Oscilloscopes
FeatureTDS 684ATDS 7XXA
Digitizing rate, maximum5 GS/s on each channel
simultaneously
Analog bandwidth1 GHzTDS 744A: 500 MHz
TDS 744A: 2 GS/s
TDS 784A: 4 GS/s
TDS 784A: 1 GHz
ChannelsFour, each with 8-bit resolution
Record lengths, maximum15,000 samples50,000 samples
(500,000 with option 1M)
Acquisition modesSample, envelope, and
average
Trigger modesInclude: edge, logic, and pulse.
Video trigger , with option 05, modes include:
NTSC, SECAM, PAL, HDTV, and FlexFormat.
DisplayColor for distinguishing among waveforms, measurements, and
associated text
Storage1.44 Mbyte, 3.5 inch, DOS 3.3-or-later floppy disk.
NVRAM storage for saving waveforms, hardcopies, and setups
I/OFull GPIB programmability.
Hardcopy output using GPIB, RS-232, or Centronics ports
Use a combination of front-panel buttons, knobs, and on-screen menus to control
the many functions of the oscilloscope. The front-panel controls are grouped
according to function: vertical, horizontal, trigger, and special. Set a function you
adjust often, such as vertical positioning or the time base setting, directly by its
own front-panel knob. Set a function you change less often, such as vertical
coupling or horizontal mode, indirectly using a selected menu.
Menus
Indicators
General Purpose Knob
GUI
Pressing one (sometimes two) front-panel button(s), such as vertical menu,
displays a main menu of related functions, such as coupling and bandwidth, at
the bottom of the screen. Pressing a main-menu button, such as coupling,
displays a side menu of settings for that function, such as AC, DC, or GND
(ground) coupling, at the right side of the screen. Pressing a side-menu button
selects a setting such as DC.
On-screen readouts help you keep track of the settings for various functions,
such as vertical and horizontal scale and trigger level. Some readouts use the
cursors or the automatic parameter extraction feature (called measure) to display
the results of measurements made or the status of the instrument.
Assign the general purpose knob to adjust a selected parameter function. More
quickly change parameters by toggling the SHIFT button. Use the same method
as for selecting a function, except the final side-menu selection assigns the
general purpose knob to adjust some function, such as the position of measurement cursors on screen, or the setting for a channel fine gain.
The user interface also makes use of a GUI, or Graphical User Interface, to make
setting functions and interpreting the display more intuitive. Some menus and
status are displayed using iconic representations of function settings, such as
those shown here for full, 250 MHz and 20 MHz bandwidth. Such icons allow
you to more readily determine status or the available settings.
Signal Acquisition System
The signal acquisition system provides four, full-featured vertical channels with
calibrated vertical scale factors from 1 mV to 10 V per division. All channels can
be acquired simultaneously.
Each of the full-featured channels can be displayed, vertically positioned, and
offset, can have their bandwidth limited (250 MHz or 20 MHz) and their vertical
coupling specified. Fine gain can also be adjusted.