TDS1000- and TDS2000-Series
Digital Storage Oscilloscopes
071-1076-01
This document supports firmware version 1.00
and above.
Warning
The servicing instructions are for use by
qualified personnel only. To avoid personal
injury, do not perform any servicing unless you
are qualified to do so. Refer to all safety
summaries prior to performing service.
Tektronix products are covered by U.S. and foreign patents, issued and pending. Information in this publication supercedes
that in all previously published material. Specifications and price change privileges reserved.
Tektronix, Inc., P.O. Box 500, Beaverton, OR 97077
TEKTRONIX and TEK are registered trademarks of Tektronix, Inc.
WARRANTY
Tektronix warrants that the products that it manufactures and sells will be free from defects in materials and
workmanship for a period of three (3) years from the date of purchase from an authorized Tektronix distributor. If
any such product proves defective during this warranty period, Tektronix, at its option, either will repair the
defective product without charge for parts and labor, or will provide a replacement in exchange for the defective
product. Batteries are excluded from this warranty.
In order to obtain service under this warranty, Customer must notify Tektronix of the defect before the expiration
of the warranty period and make suitable arrangements for the performance of service. Customer shall be
responsible for packaging and shipping the defective product to the service center designated by Tektronix,
shipping charges prepaid, and with a copy of customer proof of purchase. Tektronix shall pay for the return of the
product to Customer if the shipment is to a location within the country in which the Tektronix service center is
located. Customer shall be responsible for paying all shipping charges, duties, taxes, and any other charges for
products returned to any other locations.
This warranty shall not apply to any defect, failure or damage caused by improper use or improper or inadequate
maintenance and care. Tektronix shall not be obligated to furnish service under this warranty a) to repair damage
resulting from attempts by personnel other than Tektronix representatives to install, repair or service the product;
b) to repair damage resulting from improper use or connection to incompatible equipment; c) to repair any
damage or malfunction caused by the use of non-Tektronix supplies; or d) to service a product that has been
modified or integrated with other products when the effect of such modification or integration increases the time
or difficulty of servicing the product.
THIS WARRANTY IS GIVEN BY TEKTRONIX WITH RESPECT TO THE LISTED PRODUCTS IN
LIEU OF ANY OTHER WARRANTIES, EXPRESS OR IMPLIED. TEKTRONIX AND ITS VENDORS
DISCLAIM ANY IMPLIED WARRANTIES OF MERCHANTABILITY OR FITNESS FOR A
PAR TICULAR PURPOSE. TEKTRONIX’ RESPONSIBILITY TO REP AIR OR REPLACE DEFECTIVE
PRODUCTS IS THE SOLE AND EXCLUSIVE REMEDY PROVIDED TO THE CUSTOMER FOR
BREACH OF THIS W ARRANTY. TEKTRONIX AND ITS VENDORS WILL NOT BE LIABLE FOR ANY
INDIRECT, SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES IRRESPECTIVE OF
WHETHER TEKTRONIX OR THE VENDOR HAS ADVANCE NOTICE OF THE POSSIBILITY OF
SUCH DAMAGES.
Table of Contents
Specifications
Operating Information
Theory of Operation
General Safety Summaryvii...................................
Service Safety Summaryxi....................................
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
v
Table of Contents
vi
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
General Safety Summary
Review the following safety precautions to avoid injury and prevent damage to
this product or any products connected to it.
Only qualified personnel should perform service procedures.
Injury Precautions
Use Proper Power Cord. To avoid fire hazard, use only the power cord specified
for this product.
Avoid Electric Overload. To avoid electric shock or fire hazard, do not apply a
voltage to a terminal that is outside the range specified for that terminal.
Avoid Overvoltage. To avoid electric shock or fire hazard, do not apply potential
to any terminal, including the common terminal, that varies from ground by more
than the maximum rating for that terminal.
Avoid Electric Shock. To avoid injury or loss of life, do not connect or disconnect
probes or test leads while they are connected to a voltage source.
Ground the Product. This product is grounded through the grounding conductor
of the power cord. To avoid electric shock, the grounding conductor must be
connected to earth ground. Before making connections to the input or output
terminals of the product, ensure that the product is properly grounded.
Connect the Probe Properly. The probe ground lead is at ground potential. Do not
connect the ground lead to an elevated voltage.
Do Not Operate Without Covers. To avoid electric shock or fire hazard, do not
operate this product with covers or panels removed.
Use Proper Fuse. To avoid fire hazard, use only the fuse type and rating specified
for this product.
Do Not Operate in Wet/Damp Conditions. To avoid electric shock, do not operate
this product in wet or damp conditions.
Do Not Operate in an Explosive Atmosphere. To avoid injury or fire hazard, do not
operate this product in an explosive atmosphere.
Product Damage
Precautions
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Use Proper Power Source. Do not operate this product from a power source that
applies more than the voltage specified.
Provide Proper Ventilation. To prevent product overheating, provide proper
ventilation.
Do Not Operate With Suspected Failures. If you suspect there is damage to this
product, have it inspected by qualified service personnel.
vii
General Safety Summary
Symbols and Terms
Terms in this Manual. These terms may appear in this manual:
WARNING. Warning statements identify conditions or practices that could result
in injury or loss of life.
CAUTION. Caution statements identify conditions or practices that could result in
damage to this product or other property.
Terms on the Product. These terms may appear on the product:
DANGER indicates an injury hazard immediately accessible as you read the
marking.
WARNING indicates an injury hazard not immediately accessible as you read the
marking.
CAUTION indicates a hazard to property including the product.
Symbols on the Product. The following symbols may appear on the product:
DANGER
High Voltage
Protective Ground
(Earth) Terminal
ATTENTION
Refer to Manual
Double
Insulated
viii
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
General Safety Summary
Probe Safety
A guard around the probe body provides a finger barrier for protection from
electric shock.
Finger guard
WARNING. To avoid electric shock when using the probe, keep fingers behind the
guard on the probe body.
To avoid electric shock while using the probe, do not touch metallic portions of
the probe head while it is connected to a voltage source.
Connect the probe to the oscilloscope and connect the ground terminal to ground
before you take any measurements.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
ix
General Safety Summary
x
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Service Safety Summary
Only qualified personnel should perform service procedures. Read this Service
Safety Summary and the General Safety Summary before performing any service
procedures.
Do Not Service Alone. Do not perform internal service or adjustments of this
product unless another person capable of rendering first aid and resuscitation is
present.
Disconnect Power. To avoid electric shock, disconnect the main power by means
of the power cord or, if provided, the power switch.
Use Care When Servicing With Power On. Dangerous voltages or currents may
exist in this product. Disconnect power, remove battery (if applicable), and
disconnect test leads before removing protective panels, soldering, or replacing
components.
To avoid electric shock, do not touch exposed connections.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
xi
Service Safety Summary
xii
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Preface
Related Manuals
The service manual for the TDS1000- and TDS2000-Series Digital Storage
Oscilloscopes provides instructions to verify the performance of, calibrate,
troubleshoot, and repair the oscilloscopes to the module level.
Unless noted otherwise, the term “oscilloscope” refers to all of the models in the
TDS1000 and TDS2000 series.
These manuals contain additional documentation for the oscilloscopes:
Extension module
User manual
Language
English071-1064-XX071-0409-XX071-1075-XX
French071-1065-XX*071-0483-XX
German071-1067-XX*071-0485-XX
Italian071-1066-XX*071-0484-XX
Spanish071-1068-XX*071-0482-XX
Portuguese071-1070-XX*071-0486-XX
Japanese071-1069-XX*071-0488-XX
Korean071-1073-XX*071-0491-XX
Simplified Chinese071-1071-XX*071-0489-XX
Traditional Chinese071-1072-XX*071-0490-XX
Russian071-1074-XX*071-0487-XX
*These manuals contain a language overlay for the front-panel controls.
part number
instructions part
number
Programmer manual
part number
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
xiii
Preface
Contacting Tektronix
Phone1-800-833-9200*
AddressTektronix, Inc.
Department or name (if known)
14200 SW Karl Braun Drive
P.O. Box 500
Beaverton, OR 97077
USA
Web sitewww.tektronix.com
Sales support1-800-833-9200, select option 1*
Service support1-800-833-9200, select option 2*
Technical supportEmail: techsupport@tektronix.com
*This phone number is toll free in North America. After office hours, please leave a
voice mail message.
Outside North America, contact a Tektronix sales office or distributor; see the
Tektronix web site for a list of offices.
Product End-of-Life Handling
Components that Contain Mercury. The cold cathode fluorescent tube located in
the liquid crystal display backlight contains trace amounts of mercury. When you
are ready to reclaim the oscilloscope, you must properly transfer it according to
local regulations concerning mercury-containing equipment, or ship the
oscilloscope to the Tektronix Recycling Operations (RAMS). You can contact
Tektronix for the RAMS shipping address and instructions.
1-800-833-9200, select option 3*
1-503-627-2400
6:00 a.m. -- 5:00 p.m. Pacific time
xiv
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Specifications
g
q
These specifications apply to all TDS1000- and TDS2000-series oscilloscopes.
To verify that an oscilloscope meets specifications, it must first meet the
following conditions:
HThe oscilloscope must have been operating continuously for ten minutes
within the specified operating temperature.
HYou must perform the Do Self Cal operation, accessible through the Utility
menu, if the operating temperature changes by more than 5 °C.
HThe oscilloscope must be within the factory calibration interval of one year.
Specifications begin in Table 1--1. All specifications are guaranteed unless noted
“typical.” Specifications that are marked with the n symbol are checked in the
chapter Performance Verification.
Table 1- 1: Oscilloscope specifications
Acquisition
Acquisition ModesSample, Peak Detect, and Average
Acquisition Rate, typicalUp to 180 waveforms per second, per channel (Sample acquisition mode, no measurements)
Single SequenceAcquisition ModeAcquisition Stops After
Sample, Peak DetectSingle acquisition, all channels simultaneously
AverageN acquisitions, all channels simultaneously, N is selectable from
4, 16, 64, and 128
Inputs
Input CouplingDC, AC, or GND
Input Impedance, DC
Coupled
P2200 Probe
Attenuation
Supported Probe
Attenuation Factors
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maximum VoltageOvervoltage Category*Maximum Voltage
Between Signal and Common
npu
Channel-to-ChannelTDS1002 and TDS2002TDS1012, TDS2012, TDS2014, TDS2022, and TDS2024
Common Mode Rejection,
typical
CAT I and CAT II300 V
CAT III150 V
RMS
RMS
Derate at 20 dB/decade above 100 kHz to 13 V peak AC at 3 MHz and above.
For non-sinusoidal waveforms, peak value must be less than 450 V. Excursion above 300 V should be
less than 100 ms duration.
RMS signal level including any DC component removed through AC coupling must be limited to 300 V.
ese
uesare e
100:1 at 60 Hz
20:1at30MHz
ge to theos
100:1 at 60 Hz
20:1at50MHz
ope m
result.
Measured on MATH Ch1 -- Ch2 waveform, with test signal applied between signal and common of both
channels, and with the same VOLTS/DIV and coupling settings on each channel
Measured on MATH Ch3 -- Ch4 waveform for 4-channel models
Channel-to-Channel Crosstalk TDS1002 and TDS2002TDS1012, TDS2012, and
TDS2014
≥ 100:1 at 30 MHz≥ 100:1 at 50 MHz≥ 100:1 at 100 MHz
Measured on one channel, with test signal applied between signal and common of the other channel,
and with the same VOLTS/DIV and coupling settings on each channel
--
.
TDS2022 and TDS2024
1- 2
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Specifications
SampleandAveragemodes
g
5s/div**),t
ypica
l
y
p
n
DCGainAccurac
y
n
DCMeasurement
Table 1- 1: Oscilloscope specifications (Cont.)
Vertical
Digitizers8-bit resolution (except when set to 2 mV/div), each channel sampled simultaneously
VOLTS/DIV Range2 mV/div to 5 V/div at input BNC
Position Range2 mV/div to 200 mV/div, ±2V
> 200 mV/div to 5 V/div, ±50 V
nAnalog Bandwidth in
Sample and Average modes
at BNC or with P2200 probe,
DC Coupled
Analog Bandwidth in PeakTDS1002 and TDS2002TDS1012, TDS2012, TDS2014, TDS2022, and TDS2024
Detect mode (50 s/div to
**
Selectable Analog Bandwidth
Limit, typical
Lower Frequency
Limit, AC Coupled
Rise Time at BNC,
typical
Peak Detect
Response**
nDC Gain Accuracy
nDC Measurement
Accuracy, Average Acquisition
Mode
TDS1002 and TDS2002TDS1012, TDS2012, and
TDS2014
60 MHz{*
20 MHz* (when vertical scale is set to < 5 mV)
50 MHz{*75 MHz{*
20 MHz* (when vertical scale is set to < 5 mV)
20 MHz*
≤ 10 Hz at BNC
≤ 1 Hz when using a 10X passive probe
TDS1002 and TDS2002TDS1012, TDS2012, and
<5.8ns<3.5ns<2.1ns
Captures 50% or greater amplitude of pulses ≥12 ns wide typical (50 s/div to 5 s/div) in the center 8
vertical divisions
±3% for Sample or Average acquisition mode, 5 V/div to 10 mV/div
±4% for Sample or Average acquisition mode, 5 mV/div and 2 mV/div
Measurement TypeAccuracy
Average of ≥ 16 waveforms with
vertical position at zero
Average of ≥ 16 waveforms with
vertical position not at zero
100 MHz{*200 MHz{*
TDS2014
±(3% × reading + 0.1 div + 1 mV) when 10 mV/div or greater is
selected
±[3% × (reading + vertical position) + 1% of vertical position +
Add 2 mV for settings from 2 mV/div to 200 mV/div
Add 50 mV for settings from > 200 mV/div to 5 V/div
Volts Measurement
Repeatability,
Average Acquisition Mode
{ When vertical scale is set to ≥ 5mV.
*Bandwidth reduced to 6 MHz with a 1X probe.
** The oscilloscope reverts to Sample mode when the SEC/DIV (horizontal scale) is set from 2.5 s/div to 5 ns/div on 1 GS/s models,
or from 2.5 s/div to 2.5 ns/div on 2 GS/s models. The Sample mode can still capture 10 ns glitches.
Delta volts between any two averages of ≥ 16 waveforms acquired
under same setup and ambient
conditions
±(3% × reading + 0.05 div)
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
1- 3
Specifications
p
g
/
g
g
n
TriggerSensitivity,
Table 1- 1: Oscilloscope specifications (Cont.)
Horizontal
Sample Rate RangeTDS2022 and TDS2024TDS1002, TDS1012, TDS2002, TDS2012,and TDS2014
5S/sto2GS/s5S/sto1GS/s
Waveform Interpolation(sin x)/x
Record Length2500 samples for each channel
SEC/DIV RangeTDS2022 and TDS2024TDS1002, TDS1012, TDS2002, TDS2012, and TDS2014
2.5ns/divto50s/div,ina1,2.5,5
sequence
nSample Rate and Delay
Time Accuracy
Delta Time Measurement
Accuracy (Full Bandwidth)
Position RangeTDS1002, TDS1012, TDS2002, TDS2012, TDS2014, TDS2022, and TDS2024
Not Equal Guardbandt ≤ 330 ns: guardband = ±16.5 ns
Trigger Frequency Counter
Readout Resolution6 digits
Accuracy (typical)±51 ppm including all frequency reference errors and ±1 count errors
Frequency RangeAC coupled, 10 Hz minimum to rated bandwidth
Signal SourcePulse Width or Edge Trigger modes: all available trigger sources
Trigger when < (Less than), > (Greater than), = (Equal), or ≠ (Not Equal);
Positive pulse or Negative pulse
Not Equal: If the pulse is narrower than the specified width, the trigger point is the trailing edge.
Otherwise, the oscilloscope triggers when a pulse continues longer than the time specified as the Pulse
Width.
Less than: The trigger point is the trailing edge.
Greater than (also called time-out trigger): The oscilloscope triggers when a pulse continues longer than
the time specified as the Pulse Width.
t ≤ 330 ns: guardband = ±16.5 ns
165 ns < t ≤ 330 ns: guardband = --16.5 ns/+33 ns
t ≤ 165 ns: guardband = ±16.5 ns
The Frequency Counter measures trigger source at all times, including when the oscilloscope acquisition
is halted due to changes in the run status, or acquisition of a single shot event has completed.
Pulse Width Trigger mode: The oscilloscope counts pulses of signifi cant magnitude inside the 250 ms
measurement window that qualify as triggerable events, such as narrow pulses in a PWM pulse train if
set to < mode and the width is set to a relatively small time.
Edge Trigger mode: The oscilloscope counts all edges of sufficient magnitude and correct polarity.
Video Trigger mode: The Frequency Counter does not operate.
Measurements
CursorsVoltage difference between cursors (∆V)
Time difference between cursors (∆T)
Reciprocal of ∆TinHertz(1/∆T)
Automatic MeasurementsFrequency, Period, Mean, Pk-Pk, Cycle RMS, Min, Max, Rise Time, Fall Time, Pos Width, Neg Width
1- 6
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Specifications
Table 1- 2:
Oscilloscope general specifications
Display
Display Type145 mm (5.7 in) diagonal liquid crystal
Display Resolution320 horizontal by 240 vertical pixels
Display ContrastAdjustable, temperature compensated
Backlight Intensity, typical65 cd/m
2
Probe Compensator Output
Output Voltage, typical5Vinto≥ 1MΩ load
Frequency, typical1 kHz
Power Source
Source Voltage
100 -- 120 VAC
120 -- 240 VAC
( 10%) from 45 Hz through 440 Hz, CAT II
RMS
( 10%) from 45 Hz through 66 Hz, CAT II
RMS
Power ConsumptionLess than 30 W
Fuse1 A, T rating, 250 V
Environmental
TemperatureOperating0 °Cto+50°C
(32 °F to 122 °F)
Nonoperating-- 4 0 °Cto+71°C
(--40 °F to 159.8 °F)
Cooling MethodConvection
Humidity+40 °C or below
≤ 90% relative humidity
(+104 °F or below)
+41° Cto+50° C
≤ 60% relative humidity
(106 °F to 122 °F)
AltitudeOperating and Nonoperating3,000 m (10,000 ft)
Random VibrationOperating0.31 g
from 5 Hz to 500 Hz, 10 minutes on
RMS
each axis
Nonoperating2.46 g
from 5 Hz to 500 Hz, 10 minutes on
RMS
each axis
Mechanical ShockOperating50 g, 11 ms, half sine
Mechanical
SizeHeight151.4 mm (5.96 in)
Width323.8 mm (12.75 in)
Depth124.5 mm (4.90 in)
Weight (approximate)When packaged for domestic shipment3.6kg(8.0lbs)
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
1- 7
Specifications
Table 1- 3: Oscilloscope EMC certifications and compliances
European UnionMeets the intent of Directive 89/336/EEC for Electromagnetic Compatibility. Compliance was
demonstrated to the following specifications as listed in the Official Journal of the European
Communities:
EN 61326, EMC requirements for Class A electrical equipment for measurement, control, and laboratory
1,2
use
IEC 61000--4--2, Electrostatic discharge immunity (Performance criterion B)
IEC 61000--4--3, RF electromagnetic field immunity (Performance criterion A)
IEC 61000--4--4, Electrical fast transient/burst immunity (Performance criterion B)
IEC 61000--4--5, Power line surge immunity (Performance criterion B)
IEC 61000--4--6, Conducted RF immunity (Performance criterion A)
4
IEC 61000--4--11, Voltage dips and interruptions immunity (Performance criterion B)
EN 61000--3--2, AC power line harmonic emissions
Australia/New ZealandMeets the i ntent of Australian EMC Framework as demonstrated to the following specification:
AS/NZS 2064.1/2
U.S.A.Emissions comply with FCC Code of Federal Regulations 47,
Part 15, Subpart B, Class A Limits
1
Emissions that exceed the levels required by this standard may occur when this equipment is connected to a test object.
2
To ensure compliance to the standards listed above, attach only high quality shielded cables to this oscilloscope. High quality
shielded cables typically are braid and foil types that have low impedance connections to shielded connectors at both ends.
3
The increase in trace noise while subjected to a test field (3 V/m over the frequency range 80 MHz to 1 GHz, with 80% amplitude
modulation at 1 kHz) is not to exceed 2 major divisions peak--to--peak. Ambient conducted fields may induce triggering when the
trigger threshold is offset less than 1 major division from ground reference.
4
The increase in trace noise while subjected to a test field (3 V/m over the frequency range 150 kHz to 80 MHz, with 80%
amplitude modulation at 1 kHz) is not to exceed 1 major division peak--to--peak. Ambient conducted fields may induce triggering
when the trigger threshold is offset less than 0.5 major divisions from ground reference.
3
1- 8
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Specifications
Table 1- 4: Oscilloscope safety certifications and compliances
CertificationsCAN/CSA C22.2 No. 1010.1-92
UL3111-1, First Edition
CSA Certified Power CordsCSA Certification includes the products and power cords appropriate for use in the North America power
network. All other power cords supplied are approved for the country of use.
Pollution Degree 2Do not operate in environments where conductive pollutants may be present .
CAT ISignal levels in special equipment or parts of equipment, telecommunications, electronics
Table 1- 5: Oscilloscope general certifications and compliances
Russian FederationThis product was certified by the GOST ministry of Russia to be in compliance with all appli cable EMC
regulations.
Peoples Republic of ChinaThis product has received the Chinese Metrology Certification (CMC).
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
1- 9
Specifications
1- 10
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Operating Information
TDS1000- and TDS2000-series oscilloscopes are two- or four-channel instruments in small, lightweight, benchtop chassis that you can use to take
ground-referenced measurements.
In addition to a list of general features, this chapter covers:
HHow to install your product
HHow to add extended functions
HHow to perform a brief functional check
HHow to use the self-calibration routine
HHow to restore factory default settings
For more detailed information about oscilloscope operation, refer to your user
manual.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
2--1
Operating Information
General Features
ModelChannelsBandwidthSample rateDisplay
TDS1002260 MHz1.0 GS/sMonochrome
TDS10122100 MHz1.0 GS/sMonochrome
TDS2002260 MHz1.0 GS/sColor
TDS20122100 MHz1.0 GS/sColor
TDS20144100 MHz1.0 GS/sColor
TDS20222200 MHz2.0 GS/sColor
TDS20244200 MHz2.0 GS/sColor
HContext-sensitive Help system
HColor or monochrome LCD display
HSelectable 20 MHz bandwidth limit
H2500-point record length for each channel
HAutoset Menu
HProbe Check Wizard
HCursors with readouts
HTrigger frequency readout
HEleven automatic measurements
HWaveform averaging and peak detection
HDual time base
HMath Fast Fourier Transform (FFT)
HPulse Width trigger capability
HVideo trigger capability with line-selectable triggering
HExternal trigger
HSetup and waveform storage
HVariable persistence display
2--2
HRS-232, GPIB, and Centronics ports with the optional TDS2CMA Commu-
nications Extension Module
HUser interface in ten user-selectable languages
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Installation
Operating Information
Power Cord
Use only power cords designed for your oscilloscope. Use a power source that
delivers 90 to 264 VAC
must deliver 90 to 132 VAC
, 45 to 66 Hz. If you have a 400 Hz power source, it
RMS
, 360 to 440 Hz.
RMS
Refer to page 10--7 in the Replaceable Parts chapter for a list of power cords
available from Tektronix. All power cords must be rated for > 3 Amps, 3 wire,
and have an 18 AWG (1.0 mm) Safety Ground lead, or larger.
Securing cable
Power cable
Figure 2--1: Routing the power cord and security cable
Security Loop
Use the built-in cable channels to secure both your oscilloscope and extension
module to your location.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
2--3
Operating Information
Extension Modules
You can increase the feature set of your oscilloscope by inserting an extension
module. Refer to page 10--7 for more information. Figure 2--2 shows the proper
way to install and remove a module without bending the connecting pins.
For more information on extension modules, see your oscilloscope user manual.
CAUTION. Electrostatic discharge (ESD) can damage components in the
extension module and the oscilloscope. Do not operate your instrument with the
extension module connector exposed.
Module release tab
Remove
extension
module
Install
extension
module
Figure 2--2: Installing and removing an extension module
2--4
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Functional Check
Operating Information
Perform this quick functional check to verify that your oscilloscope is operating
correctly.
ON/OFF
button
PASSED
CH1
PROBE COMP
1. Turn on the oscilloscope.
Wait until the display shows that all power-on tests passed. Push the DEFAULT
SETUP button. The default Probe menu attenuation setting is 10X.
2. Set the switch to 10X on the P2200 probe and connect the oscilloscope probe to
channel 1. To do this, align the slot in the probe connector with the key on the
CH1 BNC, push to connect, and twist to the right to lock the probe in place.
Attach the probe tip and reference lead to the PROBE COMP connectors.
3. Push the AUTOSET button. Within a few seconds, you should see a square
wave in the display (approximately 5 V at 1 kHz peak-to-peak).
Move the probe to the CH2 BNC. Repeat step 3. For four-channel models,
repeat step 3 for CH3 and CH4. A similar test, described below, can be
performed on the EXT TRIG connector, although the amplitude scale factor is
undefined.
Connect the probe to the EXT TRIG BNC. Attach the probe tip and reference
lead to the PROBE COMP connectors. Push TRIG MENU and push the top
option button to select Edge triggering. Then, push the second option button to
select Ext as the source. Finally, push and hold the front-panel TRIG VIEW
button. Once again, you should see a square wave in the display.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
2--5
Operating Information
SEC/DIV50
0
Self Calibration
Default Setup
The self-calibration routine lets you quickly optimize the oscilloscope signal path
for maximum measurement accuracy. You can run the routine at any time but you
should always run the routine if the ambient temperature changes by 5 _Cor
more.
To compensate the signal path, disconnect any probes or cables from the channel
input connectors. Then, press the UTILITY button and select the Do Self Cal
option. Follow the on-screen instructions to proceed.
The table below lists the state of the oscilloscope after you press DEFAULT
SETUP.
NOTE. When you push the DEFAULT SETUP button, the oscilloscope displays
the CH1 waveform and removes all other waveforms.
Table 2--1: Default settings
Menu or systemOption, button or knobDefault setting
ACQUIRESample, Peak Detect, and
Average
Averages: 4, 8, 16, 64, and
128
RUN/STOPRUN
CURSORTypeOff
SourceCH1
Horizontal (voltage)+/-- 3.2 divs
Vertical (time)+/-- 4 divs
DISPLAYTypeVectors
PersistOff
FormatYT
HORIZONTALWindowMain
Trig KnobLevel
POSITION0.00 s
Sample
16
s
Window Zone
2--6
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
50 s
Operating Information
Sourc
e
CH1
Windo
w
Hanning
Table 2--1: Default settings (Cont.)
Menu or systemDefault settingOption, button or knob
MATHOperationCH1 -- CH2
FFT operation:
Window
FFT Zoom
MEASURESourceCH1
TypeNone
TRIGGER ModeTypeEdge
Edge TRIGGERTypeEdge
SourceCH1
SlopeRising
ModeAuto
CouplingDC
LEVEL0.00 V
Video TRIGGERTypeVideo
SourceCH1
PolarityNormal
SyncAll Lines
StandardNTSC
Pulse Width TRIGGERTypePulse
Hanning
X1
SourceCH1
When=
Set Pulse Width1.00 ms
PolarityPositive
ModeAuto
CouplingDC
Vertical system,
all channels
CouplingDC
BW Limit
Volts/DivCoarse
Probe10X
InvertOff
POSITION0.00 divs (0.00 V)
VOLTS/DIV1.00 V
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Off
2--7
Operating Information
The DEFAULT SETUP button does not reset the following:
HLanguage option
HSaved setup files
HSaved reference waveform files
HDisplay contrast
HCalibration data
HPrinter setup
HRS-232 setup
HGPIB setup
2--8
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Theory of Operation
This chapter describes the electrical operation of the TDS1000- and
TDS2000-series oscilloscopes to the module level. It describes the basic
operation of each functional circuit block shown in Figures 3--1 and 3--2. Of
necessity, the descriptions for the two and four channel units, and the color and
monochrome units, are slightly different.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
3--3
Theory of Operation
Main Board
The main board of a four-channel oscilloscope is essentially 2, two-channel
oscilloscopes tied together through a common microprocessor, and some special
interconnects to support combining the display and trigger systems. For this
reason, the focus of the main board discussion will be the two-channel system,
with differences noted as necessary.
At a minimum, the main board contains attenuators, an amplifier ASIC, a
digitizer/trigger system ASIC, a signal processing/display/system services ASIC,
RAM, flash PROM, a microprocessor, and special power supplies. For a
four-channel oscilloscope, the attenuators and ASICs are duplicated. Most of the
other aspects of the circuitry remain unchanged.
Acquisition System
Signals from the CH 1, CH 2 and other input connectors pass through attenuators
and an AC-coupling switch to the amplifier ASIC. The EXT TRIG input has an
abbreviated version of this path, lacking some of the attenuator settings and the
AC coupling switch.
The amplifier ASIC contains buffers and variable gain amplifiers, as well as
filters that provide 20 MHz bandwidth limiting. The task of the amplifier ASIC is
to convert from a 1 MΩ single-ended environment in the front end to a much
lower impedance differential (and thus less noise-sensitive) environment for the
acquisition process. The amplifier ASIC assures that the input signal is amplified
to approximately the correct level to allow the fullest possible use of the digitizer.
The acquisition ASIC contains samplers and peak detectors for each input
channel, a common amplifier, an A/D converter, and the trigger logic. The
digitized waveform samples are transferred to the processing and display ASIC.
In four-channel systems, the two acquisition ASICs are interconnected so that a
trigger on one ASIC can cause a trigger on the other.
The processor system adds the microprocessor and flash PROM to the processing
and display system. The processor system interprets the front-panel control
changes detected by the display ASIC, provides control parameters based upon
user setting requests, computes waveform measurements, and manages the
extension module interface. Saved setups, waveforms, and calibration constants
are stored in nonvolatile memory (NVRAM). The processor system shares
DRAM with the display system.
3--4
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Theory of Operation
Processing and Display
System
Input Signal Interface
The processing and display system consists of the display ASIC, DRAM, and
system oscillator. Digitized acquisition samples are received by the display ASIC
and stored in DRAM. Once data are received by the display ASIC, various
corrections are applied, display rasterization is performed, and the waveform is
placed into a display buffer. At the same time, the waveform is being read from
the display buffers and written to the LCD. Additional circuitry in the display
ASIC supports scanning the front panel, handling DRAM refresh, providing the
processor clock, and performing various memory mapping tasks required by all
elementary microprocessor based systems. In a four-channel system, the two
display ASICs are interconnected so that one ASIC may provide display
information for the second.
The processing and display system handles some of the computational tasks.
Other tasks are performed by the processor system. Since all array processing is
performed in the processing and display system, no computations can be
performed that involve data from two different channel sets. Thus, subtracting
channel 3 data from channel 2 is prohibited. Channel 1 and Channel 2 data may
be combined in all of the supported ways.
BNC connectors are mounted on the main board for all signal inputs. The signal
inputs are compatible with the supplied P2200 probes.
Probe Compensation
External Trigger
Main Board Power
The PROBE COMP and ground terminals are provided for probe adjustment.
The EXT TRIG channel is processed on the chain containing the highest
numbered normal input channel.
To support various functions on the main board, a number of secondary power
supplies are generated. For the amplifier and acquisitions ASICs, the main board
creates a +2.5 V and --2.5 V supply. The +2.5 V supply is derived from the
+3.3 V logic supply. The --2.5 V supply is derived from the --4 V supply. A three
terminal regulator provides +5 V for option modules and the LCD display.
An additional power supply provides the LCD bias voltage which ranges from
+19 V to +28 V, depending on contrast setting and display type. This +28 V
supply has a temperature sensor on the front-panel board that varies the output
voltage of the supply to maintain contrast over a wide temperature range.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
3--5
Theory of Operation
Power Supply
Display Module
The main power supply module for the TDS1000- and TDS2000-series
oscilloscopes is a wide input range universal supply. It is capable of providing
about 17 W of power for the oscilloscope while allowing the input to run from
about +90 V to +264 V. Input frequency ranges from 47 Hz to 440 Hz, which
allows operation in virtually all countries in the world and in a number of
off-grid environments such as military power on aircraft.
The secondary supplies from the power supply and the approximate current
draws are listed in the table on page 6--35 with the associated connector pins on
J101.
For information on voltages used in the oscilloscope, refer to the Troubleshooting
section of this manual.
Front Panel
Two-Channel
Oscilloscopes
The display module is a standard passive liquid crystal display (LCD). The
monochrome unit is patterned with 320 columns by 240 rows, and has the
associated drivers and backlight. The backlight is based on a 5 mA side light
fluorescent tube.
The color unit is patterned with 320 columns x 3 sub-columns by 240 rows and
has associated drivers and backlight. The backlight is based upon a 5 mA top
light fluorescent tube which runs a higher voltage and is longer to provide
sufficient extra light to overcome the higher loss associated with the color filters.
You can access all of the switches, position encoders, and LEDs on the front-panel board of two-channel oscilloscopes. Additionally, an IC on the front-panel
board provides buffering and multiplexing of switch signals to the main board.
Two signals and a sense line are provided by the main board to support the front
panel. One of these lines resets the scan; a second clocks the scan to the next
position; and the sense line receives the current state of the selected switch or
encoder position.
For the encoders, some amount of debouncing occurs inside the front-panel IC.
All key debouncing is handled in the display ASIC on the main board.
3--6
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Theory of Operation
Four-Channel
Oscilloscopes
LEDs
Extension Modules
The front-panel board of the four channel units is effectively two panels in
parallel. The left side of the board is largely handled by the display ASIC for
channels 1 and 2. The right side of the board is handled by the channel 3 and 4
ASIC. Separate front-panel ICs support these data paths.
The LEDs are lighted to indicate an alternative use for the associated front-panel
knob. For more information, see the User Manual for this oscilloscope.
The LEDs are controlled by latching the value of the Channel 1--2 front panel
scan counter when the appropriate scan value is set.
Optional extension modules add the capability for the microprocessor to interface
with IEEE-488 (GPIB) controllers and RS-232 controllers, and provide a PC-like
parallel printer interface. These modules contain the necessary chips to provide
standard RS-232 and IEEE-488 states. The parallel port interface is controlled by
software without significant hardware assistance.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
3--7
Theory of Operation
3--8
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Performance Verification
c
OscilloscopeCalibration
This chapter contains performance verification procedures for the specifications
marked with the n symbol. The following equipment, or a suitable equivalent,
is required to complete these procedures.
Equipment Required
Description
DC Voltage Source17.5mVto7V,±0.5%
Leveled Sine Wave Generator 50 kHz and 200 MHz, ±3%
Time Mark Generator10 ms period, ±10 ppm
Minimum
requirements
accuracy
amplitude accuracy
accuracy
Examples
Wavetek 9100 Universal
Calibration System with
Os
illoscopeCalibration
Module (Option 250)
Fluke 5500A Multi-product
Calibrator with Oscilloscope
Calibration Option (Option
5500A-SC)
50 Ω BNC CableBNCmaletoBNCmale,
≈ 1 m (36 in) long
50 Ω BNC CableBNCmaletoBNCmale,
≈ 25 cm (10 in) long
50 Ω Feedthrough
Termination
Dual Banana to BNC AdapterBanana plugs to BNC femaleTektronix part number
BNC T AdapterBNC male to dual BNC
BNC male and female
connectors
female connectors
Tektronix part number
012-0482-XX
Tektronix part number
012-0208-XX
Tektronix part number
011-0049-XX
103-0090-00
Tektronix part number
103-0030-00
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
4- 1
Performance Verification
curacy
curacy
curacy
curacy
Test Record
Serial
number
TestPassedFailed
Self Test
Oscilloscope testsLow limitTest resultHigh limit
Channel 1 DC Gain Ac-5mV/div33.6 mV36.4 mV
Procedure performed byDate
200 mV/div1.358 V1.442 V
2V/div13.58 V14.42 V
Channel 2 DC Gain Ac-5mV/div33.6 mV36.4 mV
200 mV/div1.358 V1.442 V
2V/div13.58 V14.42 V
Channel 3 DC Gain Ac-
1
5mV/div33.6 mV36.4 mV
200 mV/div1.358 V1.442 V
2V/div13.58 V14.42 V
Channel 4 DC Gain Ac-
1
5mV/div33.6 mV36.4 mV
200 mV/div1.358 V1.442 V
2V/div13.58 V14.42 V
Channel 1 Bandwidth2.12 V—
Channel 2 Bandwidth2.12 V—
Channel 3 Bandwidth
Channel 4 Bandwidth
1
1
2.12 V—
2.12 V—
Sample Rate and Delay Time Accuracy-- 2 d i v s+2 divs
Channel 1 Edge Trigger SensitivityStable trigger—
Channel 2 Edge Trigger SensitivityStable trigger—
Channel 3 Edge Trigger Sensitivity
Channel 4 Edge Trigger Sensitivity
1
1
Stable trigger—
Stable trigger—
External Edge Trigger SensitivityStable trigger—
1
Channels 3 and 4 are only on four channel oscilloscopes.
4- 2
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Performance Verification Procedures
Before beginning these procedures, two conditions must first be met:
HThe oscilloscope must have been operating continuously for ten minutes
within the operating temperature range specified.
HYou must perform the Self Calibration operation described below. If the
ambient temperature changes by more than 5 °C, you must perform the Self
Calibration operation again.
The time required to complete the entire procedure is approximately one hour.
WARNING. Some procedures use hazardous voltages. To prevent electrical shock,
always set voltage source outputs to 0 V before making or changing any
interconnections.
Performance Verification
Self Test
Self Calibration
Check DC Gain Accuracy
This internal procedure is automatically performed every time the oscilloscope is
powered on. No test equipment or hookups are required. Verify that no error
messages are displayed before continuing with this procedure.
The self calibration routine lets you quickly optimize the oscilloscope signal path
for maximum measurement accuracy. You can run the routine at any time, but
you should always run the routine if the ambient temperature changes by 5 _Cor
more.
1. Disconnect any probes or cables from the CH1 and CH2 input connectors.
2. Press the UTILITY button and select the Do Self Cal option to start the
routine. The routine takes approximately one minute to complete.
3. Verify that self calibration passed.
This test checks the DC gain accuracy of all input channels.
1. Set the DC voltage source output level to 0V.
2. Set up the oscilloscope using the following steps:
Press menu buttonSelect menu optionSelect setting
1. DEFAULT SETUP——
2. CH 1Probe1X
3. ACQUIREAverage16
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
4- 3
Performance Verification
Press menu buttonSelect settingSelect menu option
4. MEASURE
SourceAn unchecked channel
TypeMean
3. As shown below, connect the oscilloscope channel selected in the table to the
DC voltage source.
DC voltage
source
+--
Dual banana
to BNC
adapter
BNC cable
Digitizing oscilloscope
4. For each VOLTS/DIV setting listed below, perform the following steps:
a. Set the DC voltage source output level to the positive voltage listed and
then record the mean measurement as V
pos
.
b. Reverse the polarity of the DC voltage source and then record the mean
measurement as V
c. Calculate V
diff=Vpos
neg
.
-- V
and then compare V
neg
to the accuracy
diff
limits in the table.
VOLTS/DIV settingDC voltage source output levelsAccuracy limits for V
5mV/div+17.5 mV, --17.5 mV33.6 mV to 36.4 mV
diff
4- 4
200 mV/div+700 mV, --700 mV1.358 V to 1.442 V
2V/div+7.00 V, -- 7.00 V13.58 V to 14.42 V
5. Set DC voltage source output level to 0V.
6. Disconnect the test setup.
7. Repeat steps 1 through 6 until all input channels have been checked.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Performance Verification
Check Bandwidth
This test checks the bandwidth of all input channels.
1. Set up the oscilloscope using the following steps:
Press menu buttonSelect menu optionSelect setting
DEFAULT SETUP——
CH 1Probe1X
ACQUIREAverage16
TRIGGERCouplingNoise
MEASURE
SourceAn unchecked channel
TypePk-Pk
2. As shown below, connect the oscilloscope channel selected in the table to the
leveled sine wave generator.
Leveled
sine wave
generator
Digitizing oscilloscope
Output
50 Ω feedthrough
BNC cable
terminator
3. Set the oscilloscope VOLTS/DIV to 500 mV/div.
4. Set the oscilloscope SEC/DIV to 10 s/div.
5. Set the leveled sine wave generator frequency to 50 kHz.
6. Set the leveled sine wave generator output level so the peak-to-peak
measurement is between 2.98 V and 3.02 V.
7. Set the leveled sine wave generator frequency to:
H60 MHz if you are checking a TDS1002/2002
H100 MHz if you are checking a TDS1012/2012/2014
H200 MHz if you are checking a TDS2022/2024.
8. Set the oscilloscope SEC/DIV to 10 ns/div.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
4- 5
Performance Verification
9. Check that the peak-to-peak measurement is ≥2.12 V.
10. Disconnect the test setup.
11. Repeat steps 1 through 10 until all input channels have been checked.
Check Sample Rate and
Delay Time Accuracy
This test checks the time base accuracy.
1. Set up the oscilloscope using the following steps:
Press menu buttonSelect menu optionSelect setting
DEFAULT SETUP——
CH 1Probe1X
2. Connect the oscilloscope to the time mark generator as shown below.
Time mark
generator
Output
BNC cable
Digitizing oscilloscope
50 Ω feedthrough
terminator
4- 6
3. Set the time mark generator period to 10 ms.
4. Set the oscilloscope VOLTS/DIV to 500 mV/div.
5. Set the oscilloscope Main SEC/DIV to 1ms/div.
6. Press SET LEVEL TO 50%.
7. Use the vertical POSITION control to center the test signal on screen.
8. Use the horizontal POSITION control to set the position to 10.00 ms.
9. Set the oscilloscope SEC/DIV to 250 ns/div.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Performance Verification
10. Check that the rising edge of the marker crosses the center horizontal
graticule line within ±2 divisions of center graticule.
NOTE. One division of displacement from graticule center corresponds to
a 25 ppm time base error.
11. Disconnect the test setup.
Check Edge Trigger
Sensitivity
This test checks the edge trigger sensitivity for all input channels.
1. Set up the oscilloscope using the following steps:
Press menu buttonSelect menu optionSelect setting
DEFAULT SETUP——
CH 1Probe1X
TRIGGERModeNormal
ACQUIRESample—
MEASURE
SourceAn unchecked channel
TypePk-Pk
2. As shown below, connect the oscilloscope channel selected in the table to the
leveled sine wave generator.
Leveled
sine wave
generator
Digitizing oscilloscope
Output
BNC cable
3. Set the oscilloscope VOLTS/DIV to 500 mV/div.
4. Set the oscilloscope SEC/DIV to 25 ns/div.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
50 Ω feedthrough
terminator
4- 7
Performance Verification
5. Set the leveled sine wave generator frequency to 10 MHz.
6. Set the leveled sine wave generator output level to approximately 500 mV
p-p
so that the measured amplitude is approximately 500 mV. (The measured
amplitude can fluctuate around 500 mV.)
7. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
8. Set the leveled sine wave generator frequency to
H60 MHz if you are checking a TDS1002/2002
H100 MHz if you are checking a TDS1012/2012/2014
H200 MHz if you are checking TDS2022/2024.
9. Set the oscilloscope SEC/DIV to 10 ns/div.
10. Set the leveled sine wave generator output level to approximately 750 mV
p-p
so that the measured amplitude is approximately 750 mV. (The measured
amplitude can fluctuate around 750 mV.)
11. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
12. Change the oscilloscope setup using the following step:
Press menu buttonSelect menu optionSelect setting
TRIGGERSlopeFalling
13. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
14. Change the oscilloscope setup using the following step:
Press menu buttonSelect menu optionSelect setting
TRIGGERSlopeRising
15. Disconnect the test setup.
16. Repeat steps 1 through 15 until all input channels have been checked.
4- 8
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Performance Verification
Check External Edge
Trigger Sensitivity
This test checks the edge trigger sensitivity for the external trigger.
1. Set up the oscilloscope using the following steps:
Press menu buttonSelect menu optionSelect setting
DEFAULT SETUP——
CH 1Probe1X
TRIGGER
ACQUIRESample—
MEASURE
SourceExt
ModeNormal
SourceCH1
TypePk-Pk
2. Connect the oscilloscope to the leveled sine wave generator as shown below.
Leveled
sine wave
generator
Digitizing oscilloscope
50 Ω
Output
BNC cable
BNC
cable
BNC T connector
feedthrough
terminator
3. Set the leveled sine wave generator frequency to
H60 MHz if you are checking a TDS1002/2002
H100 MHz if you are checking a TDS1012/2012/2014
H200 MHz if you are checking TDS2022/2024.
4. Set the oscilloscope VOLTS/DIV to 100 mV/div.
5. Set the oscilloscope SEC/DIV to 10 ns/div.
6. Set the leveled sine wave generator output level to approximately 200 mV
so that the measured amplitude is approximately 200 mV. (The measured
amplitude can fluctuate around 200 mV.)
p-p
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
4- 9
Performance Verification
7. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
8. For TDS2022 and TDS2024 models, set the sine wave generator frequency
to 200 MHz.
9. For TDS2022 and TDS2024 models, set the sine wave generator output level
to approximately 350 mV
p-p
.
10. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
11. Change the oscilloscope setup using the following step:
Press menu buttonSelect menu optionSelect setting
TRIGGERSlopeFalling
12. Press SET LEVEL TO 50%. Adjust TRIGGER LEVEL as necessary and
then check that triggering is stable.
13. Change the oscilloscope setup using the following step:
Press menu buttonSelect menu optionSelect setting
TRIGGERSlopeRising
14. Disconnect the test setup.
4- 10
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Adjustment Procedures
This chapter contains adjustment procedures for the TDS1000- and
TDS2000-series oscilloscopes.
Only qualified personnel should perform service procedures. Read the ServiceSafety Summary and the General Safety Summary at the beginning of this manual
before performing any service procedures. Also refer to the chapter OperatingInformation for information about using the TDS1000 and TDS2000 oscilloscopes.
NOTE. The voltage references inside the TDS1000 and TDS2000 oscilloscopes
are very stable over time and should not require routine updates. Before
performing any procedure in this chapter, first verify that the oscilloscope does
not meet specifications. Refer to the chapter Performance Verification to verify
the specifications.
Required Equipment
The equipment described in Table 5--1, or a suitable equivalent, is required to
complete these procedures.
Table 5--1: Required equipment
DescriptionMinimum requirementsExamples
DC Voltage Source-- 2 0 V t o 2 0 V , ±0.1% accuracy Wavetek 9100 Universal Calibration
System with Oscilloscope
Calibration Module (Option 250)
Leveled Sinewave
Generator
50 Ω BNC Cable
50 Ω BNC Cable (two)
50 Ω Feedthrough
Termination
60 kHz and 200 MHz, ±0.6%
amplitude accuracy
BNCmaletoBNCmale,
≈ 1 m (36 in) long
BNCmaletoBNCmale,
≈ 25 cm (10 in) long
BNC male and female
connectors
Fluke 5500A Multi-product
Calibrator with Oscilloscope
Calibration Option
(Option 5500A-SC)
Tektronix part number 012-0482-00
Tektronix part number 012-0208-00
Tektronix part number 011-0049-01
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
5--1
Adjustment Procedures
Table 5--1: Required equipment (Cont.)
DescriptionExamplesMinimum requirements
Dual Banana to BNC
Adapter
Banana plugs to BNC femaleTektronix part number 103-0090-00
BNC TOne male and two female
BNC connectors
Tektronix part number 103-0030-00
5--2
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Adjustment Procedure
Adjustment Procedures
The adjustment procedure consists of 70 steps for the two-channel oscilloscope,
and 106 steps for the four-channel model. Each step requires an external voltage
source for new calibration constants.
Before performing the adjustment procedure, you must warm up the oscilloscope
for at least ten minutes in an ambient temperature between 20 °C and 30 °C.
Adjustments performed prior to warm-up or outside this temperature range may
result in poor performance.
If all steps in the procedure are completed successfully, a “Pass” message is
displayed and the new calibration constants take affect. If any step fails, the
procedure is aborted and the current calibration is not affected. You can choose to
abort the procedure at any step by selecting FCAL ABORT without affecting the
current calibration.
NOTE. While performing the adjustment procedure, keep in mind that some of the
steps take a significant amount of time (up to 5 minutes) to complete.
Enable the Service Menu
The equipment setups, shown in Figure 5--1, are required to complete the
adjustment procedure. Table 5--2 on page 5--6 lists the steps in the procedure
and the signal requirement for each step.
You must enable the Service menu to perform the adjustment procedure. To do
this:
1. Power on the oscilloscope.
2. Push the front-panel MEASURE button to access the MEASURE menu.
3. Push the top option button to access the Measure 1 menu.
4. Push and hold the front-panel SINGLE SEQ button.
5. Push and hold the front-panel AUTOSET button.
6. Wait at least two seconds.
7. Release the SINGLE SEQ button.
8. Release the AUTOSET button. A message appears in the lower left corner of
the screen stating “Service mode ON.”
9. Push the front-panel UTILITY button. The last item in the Utility menu is
now Service.
After the adjustment is complete, disable the Service menu by pushing the
UTILITY front-panel button, and the Service and Service Mode Off option
buttons.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
5--3
Adjustment Procedures
DC voltage setup 1
DC voltage source
+--
Dual banana to
BNC adapter
BNC cable
BNC cable
Two Channel
Digitizing oscilloscope
10-inch
BNC cables
BNC ‘T’ connector
Four Channel
Digitizing oscilloscope
10-inch
BNC cables
10-inch
BNC cable
BNC ‘T’
connector
10-inch BNC cable
10-inch
BNC cables
BNC ‘T’
connectors
DC voltage setup 2
Sinewave generator setup
Leveled sine
wave generator
Output
50 Ω feedthrough terminator
(Table 5--2 and 5--3 list
DC voltage source
+--
Two Channel
Digitizing oscilloscope
proper channel)
Dual banana to
BNC adapter
Two Channel
Digitizing oscilloscope
BNC cable
Edge setup
Square wave
generator
Four Channel
Digitizing oscilloscope
Output
50 Ω feedthrough terminator
(Table 5--2 and 5--3 list
proper channel)
BNC cable
Figure 5--1: Adjustment setups
5--4
BNC cable
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Adjustment Procedures
Adjustment Procedure
Use this procedure to load new calibration constants.
1. Enable the Service menu as described on page 5--3.
2. Push the Service option button.
3. Push the Factory Cal option button.
4. Push the Warmup Timer option button.
5. Push the Start Timer option button to ensure a 10-minute warm up. After 10
minutes passes, the oscilloscope displays a large black or green cross.
6. Press the front-panel UTILITY button and push the Service option button.
7. Push the Factory Cal option button.
8. Push the FCAL INIT option button to initiate the factory calibration routine.
You are now in step 1 of the procedure.
9. As indicated by Step 1 in Table 5--2, connect a 0.0 V
source to the channel
DC
inputs.
10. Press FCAL STEP to load the calibration constant for step 1.
11. Follow the screen prompts along with Table 5--2 for the remainder of the
routine, applying the proper signal for each step.
NOTE. During some steps, the oscilloscope may appear to be idle for several
minutes while it is processing information internally. You must wait for the screen
prompt to continue.
If any step in the procedure fails, the procedure terminates. Failure will result
if an incorrect signal source is connected. If this happens, you must start the
procedure over from the beginning.
If the procedure completes successfully, a “Pass” message is displayed and
the new adjustment takes effect.
12. Compensate the signal path by performing the self-calibration routine.
Disconnect any probes or cables from the channel input connectors. Then,
press the UTILITY button and select Do Self Cal to confirm that you are
ready to proceed.
13. Disable the Service menu option by pushing the UTILITY front-panel
button, and the Service and Service Mode Off option buttons.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
5--5
Adjustment Procedures
g
Table 5--2: Adjustment steps
Step
Two-channel
models
Refer to the DC voltage setup 1 diagram
11DC Voltage 10.0 V
22DC Voltage 1--20.0 V
33DC Voltage 1-- 2 . 0 V
44DC Voltage 1-- 1 . 6 V
55DC Voltage 1-- 1 . 2 V
66DC Voltage 1-- 0 . 8 V
77DC Voltage 1-- 0 . 4 V
88DC Voltage 1--0.32 V
99DC Voltage 1-- 0 . 2 V
1010DC Voltage 1--0.16 V
1111DC Voltage 1--0.08 V
1212DC Voltage 1--0.04 V
1313DC Voltage 1--0.03 V
1414DC Voltage 1--0.02 V
1515DC Voltage 1--0.015 V
1616DC Voltage 10.015 V
1717DC Voltage 10.02 V
1818DC Voltage 10.03 V
1919DC Voltage 10.04 V
2020DC Voltage 10.08 V
2121DC Voltage 10.16 V
2222DC Voltage 10.2 V
2323DC Voltage 10.32 V
2424DC Voltage 10.4 V
2525DC Voltage 10.8 V
2626DC Voltage 11.2 V
2727DC Voltage 11.6 V
2828DC Voltage 12.0 V
2929DC Voltage 120.0 V
3030DC Voltage 10.0 V
Step
Four-channel
models
SignalSignal source settingInput
CH1 through
CHx
5--6
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Table 5--2: Adjustment steps (Cont.)
S
i
O
hmBWL
V
5050SinewaveGenerator,50OhmBW
L,2.5
V
p-p
5151SinewaveGenerator,50OhmBW
L,5
V
p-p
Adjustment Procedures
Step
Two-channel
models
Step
Four-channel
models
Refer to the DC voltage setup 2 diagram
3131DC Voltage 2-- 5 . 0 V
3232DC Voltage 2-- 1 . 0 V
3333DC Voltage 21.0 V
3434DC Voltage 25.0 V
Refer to the Sinewave generator setup diagram. BWL refers the maximum bandwidth of the oscilloscope.
3535Sinewave Generator, 50 Ohm5 kHz, 1 V
3636Sinewave Generator, 50 Ohm1 MHz, 100 mV
3737Sinewave Generator, 50 Ohm1MHz,1V
3838Sinewave Generator, 50 Ohm50 kHz, 1 V
3939Sinewave Generator, 50 Ohm20 MHz, 1 V
4040
newaveGenerator, 50
4141Sinewave Generator, 50 OhmBWL, 50 mV
4242Sinewave Generator, 50 OhmBWL, 100 mV
4343Sinewave Generator, 50 OhmBWL, 250 mV
4444Sinewave Generator, 50 OhmBWL, 500 mV
4540Sinewave Generator, 50 OhmBWL, 1 V
4646Sinewave Generator, 50 Ohm5 kHz, 2.5 V
4747Sinewave Generator, 50 Ohm1MHz,2.5V
4848Sinewave Generator, 50 Ohm50 kHz, 2.5 V
4949Sinewave Generator, 50 Ohm20 MHz, 2.5 V
,25m
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
p-p
InputSignal source settingSignal
EXT TRIG
CH1
5252Sinewave Generator, 50 OhmBWL, 1.6 V
5353Edge, 50 Ohm. Refer to Edge setup
1 kHz, 0 to --800 mV
p-p
diagram.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
5--7
Adjustment Procedures
S
i
O
hmBWL
V
6969SinewaveGenerator,50OhmBW
L,2.5
V
p-p
7070SinewaveGenerator,50OhmBW
L,5
V
p-p
Table 5--2: Adjustment steps (Cont.)
Step
Two-channel
models
5454Sinewave Generator, 50 Ohm5 kHz, 1 V
Step
Four-channel
models
p-p
5555Sinewave Generator, 50 Ohm1 MHz, 100 mV
5656Sinewave Generator, 50 Ohm1MHz,1V
5757Sinewave Generator, 50 Ohm50 kHz, 1 V
5858Sinewave Generator, 50 Ohm20 MHz, 1 V
5959
newaveGenerator, 50
6060Sinewave Generator, 50 OhmBWL, 50 mV
,25m
p-p
p-p
p-p
p-p
p-p
6161Sinewave Generator, 50 OhmBWL, 100 mV
6262Sinewave Generator, 50 OhmBWL, 250 mV
6363Sinewave Generator, 50 OhmBWL, 500 mV
6464Sinewave Generator, 50 OhmBWL, 1 V
6565Sinewave Generator, 50 Ohm5 kHz, 2.5 V
6666Sinewave Generator, 50 Ohm1MHz,2.5V
p-p
p-p
p-p
6767Sinewave Generator, 50 Ohm50 kHz, 2.5 V
6868Sinewave Generator, 50 Ohm20 MHz, 2.5 V
InputSignal source settingSignal
CH2
p-p
p-p
p-p
p-p
p-p
p-p
5--8
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Table 5--2: Adjustment steps (Cont.)
S
i
O
hmBWL
V
86SinewaveGenerator,50OhmBW
L,2.5
V
p-p
87SinewaveGenerator,50OhmBW
L,5
V
p-p
Adjustment Procedures
Step
Two-channel
models
NA
Step
Four-channel
models
71Sinewave Generator, 50 Ohm5 kHz, 1 V
p-p
72Sinewave Generator, 50 Ohm1 MHz, 100 mV
73Sinewave Generator, 50 Ohm1MHz,1V
74Sinewave Generator, 50 Ohm50 kHz, 1 V
75Sinewave Generator, 50 Ohm20 MHz, 1 V
76
newaveGenerator, 50
77Sinewave Generator, 50 OhmBWL, 50 mV
,25m
p-p
p-p
p-p
p-p
p-p
78Sinewave Generator, 50 OhmBWL, 100 mV
79Sinewave Generator, 50 OhmBWL, 250 mV
80Sinewave Generator, 50 OhmBWL, 500 mV
81Sinewave Generator, 50 OhmBWL, 1 V
82Sinewave Generator, 50 Ohm5 kHz, 2.5 V
83Sinewave Generator, 50 Ohm1MHz,2.5V
p-p
p-p
p-p
84Sinewave Generator, 50 Ohm50 kHz, 2.5 V
85Sinewave Generator, 50 Ohm20 MHz, 2.5 V
InputSignal source settingSignal
CH3
p-p
p-p
p-p
p-p
p-p
p-p
88Sinewave Generator, 50 OhmBWL, 1.6 V
89Edge, 50 Ohm. Refer to Edge setup
1 kHz, 0 to --800 mV
diagram.
p-p
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
5--9
Adjustment Procedures
S
i
O
hmBWL
V
W
L
p-p
105SinewaveGenerato
r,50OhmBW
L,2.5
V
p-p
W
L,5V
106SinewaveGenerato
r,50OhmBW
L,5
V
p-p
Table 5--2: Adjustment steps (Cont.)
Step
Two-channel
models
NA
Step
Four-channel
models
90Sinewave Generator, 50 Ohm5 kHz, 1 V
p-p
91Sinewave Generator, 50 Ohm1 MHz, 100 mV
92Sinewave Generator, 50 Ohm1MHz,1V
93Sinewave Generator, 50 Ohm50 kHz, 1 V
94Sinewave Generator, 50 Ohm20 MHz, 1 V
95
newaveGenerator, 50
96Sinewave Generator, 50 OhmBWL, 50 mV
,25m
p-p
p-p
p-p
p-p
p-p
97Sinewave Generator, 50 OhmBWL, 100 mV
98Sinewave Generator, 50 OhmBWL, 250 mV
99Sinewave Generator, 50 OhmBWL, 500 mV
100Sinewave Generator, 50 OhmBWL, 1 V
101Sinewave Generator, 50 Ohm5 kHz, 2.5 V
102Sinewave Generator, 50 Ohm1MHz,2.5V
p-p
p-p
p-p
103Sinewave Generator, 50 Ohm50 kHz, 2.5 V
InputSignal source settingSignal
CH4
p-p
p-p
p-p
p-p
p-p
104Sinewave Generator, 50 Ohm20 MHz, 2.5 V
105Sinewave Generator, 50 OhmB
106Sinewave Generator, 50 OhmB
,2.5V
-
p-p
5--10
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Preparation
This chapter contains the information needed to do periodic and corrective
maintenance on the oscilloscope. The following sections are included:
HPreparation tells you how to get ready to do oscilloscope maintenance.
HPreventing ESD provides general information on preventing damage to
internal modules when doing maintenance.
HInspection and Cleaning provides procedures for inspecting the oscilloscope
and cleaning its external and internal modules.
HRemoval and Installation Procedures provides procedures for the removal
and installation of modules.
HTroubleshooting provides information for isolating failed modules.
HRepackaging Instructions provides packaging information for shipment or
storage.
Preventing ESD
Only qualified personnel should perform service procedures. Read the Service
Safety Summary and the General Safety Summary sections of this manual beforeperforming any service procedures. Also refer to the chapter Operating Information and your oscilloscope user manual for information about using the oscillo-
scope.
When performing any service which requires internal access to the oscilloscope,
adhere to the following precautions. These precautions will help you avoid
damaging internal modules and their components due to electrostatic discharge
(ESD).
CAUTION. Static discharge can damage any semiconductor component in this
oscilloscope.
HMinimize handling of static-sensitive modules.
HTransport and store static-sensitive modules in their static protected
containers or on a metal rail. Label any package that contains static-sensitive
modules.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--1
Maintenance
Inspection and Cleaning
HDischarge the static voltage from your body by wearing a grounded antistatic
wrist strap while handling these modules.
HService static-sensitive modules only at static-free workstations. Do not
allow anything capable of generating or holding a static charge on the
workstation surface.
HHandle circuit boards by the edges when possible.
HDo not slide the modules over any surface.
HAvoid handling modules in areas that have a floor or work-surface covering
capable of generating a static charge.
This section describes how to inspect for dirt and damage, and how to clean the
oscilloscope. Inspection and cleaning are done as preventive maintenance which,
when done regularly, may prevent malfunctions and enhance reliability.
General Care
Inspection and Cleaning
Procedures
Preventive maintenance consists of visually inspecting and cleaning the
oscilloscope, and using general care when operating it. Perform preventive
maintenance more often if you operate the oscilloscope in a severe environment.
A good time to perform preventive maintenance is just before adjustment of the
oscilloscope.
For optimum performance, follow these recommendations:
HProtect the oscilloscope from adverse weather conditions. The oscilloscope is
not waterproof.
HDo not store or leave the oscilloscope where the liquid crystal display (LCD)
will be exposed to direct sunlight or high humidity for long periods of time.
HThe front and rear cases help keep dust out of the oscilloscope and must be in
place during normal operation.
HTo avoid damage to the oscilloscope or probes, do not expose them to any
sprays, liquids, or solvents.
The collection of dirt on internal components can cause them to overheat and
fail. Dirt also provides an electrical conduction path that could cause an oscilloscope failure, especially under high-humidity conditions. Inspect and clean the
oscilloscope and probes as often as operating conditions require.
6--2
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Inspection, Exterior. Inspect the outside of the oscilloscope for damage, wear, and
missing parts. Oscilloscopes that appear to have been dropped or otherwise
abused should be checked thoroughly to verify correct operation and performance. Immediately repair defects that could cause personal injury or lead to
further damage to the oscilloscope.
Cleaning Procedure, Exterior. To clean the oscilloscope exterior and probes:
1. Remove loose dust on the outside of the oscilloscope and probes with a
lint-free cloth. Use care to avoid scratching the clear glass display shield.
2. Use a soft cloth dampened with water to clean the oscilloscope, or use a 75%
isopropyl alcohol solution for more efficient cleaning.
CAUTION. To prevent getting moisture inside the oscilloscope during external
cleaning, use only enough liquid to dampen the cloth or applicator.
To avoid damage to the surface of the oscilloscope or probes, do not use any
abrasive or chemical cleaning agents.
Inspection, Interior. Inspect and clean the interior if it appears that the oscilloscope may have been damaged or exposed to an unusually dirty environment.
To inspect and clean the interior, disassemble the oscilloscope by completing
each of the disassembly procedures in the order given starting on page 6--5.
Inspect the internal portions of the oscilloscope for damage and wear using
Table 6--1 as a guide. Repair defects immediately.
Table 6--1: Internal inspection check list
ItemInspect forRepair action
Front and
Rear Case
Circuit boardsLoose, broken, or corroded solder connec-
Solder
connections
Wiring and
cables
Cracks or deformations. Scratched lettering
or display filter. Loose connectors or labels.
tions. Burned circuit boards. Burned,
broken, or cracked circuit-run plating.
Cold solder or rosin joints.Resolder joint and clean with
Loose plugs or connectors. Burned, broken,
or frayed wiring.
Repair or replace defective
hardware.
Remove failed module and
replace with a new module.
isopropyl alcohol.
Firmly seat connectors. Repair or
replace modules with defective
wires or cables.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--3
Maintenance
Cleaning Procedure, Interior. To clean the oscilloscope interior, perform the
following steps:
1. Blow off dust with dry, low-pressure, deionized air (approximately 9 psi).
2. Remove any remaining dust with a lint-free cloth dampened in isopropyl
alcohol (75% solution) and then wipe with a lint-free cloth dampened with
warm deionized water. A cotton-tipped applicator is useful for cleaning in
narrow spaces and on circuit boards.
STOP. If, after performing steps 1 and 2, a module is clean upon inspection, skip
steps 3 and 4. Proceed with step 5.
If steps 1 and 2 do not remove all the dust or dirt, all modules in the oscilloscope
except the Display Module and Front Panel Module may be spray washed using
a solution of 75% isopropyl alcohol.
Extensive cleaning of the Display Module is difficult because water must not be
trapped between the display shield and the display. The polarizers integrated into
the display degrade with long-term exposure to water. The display shield and
display must be replaced as an assembly if significant moisture or debris become
trapped inside. Cotton tipped applicators and lint-free rags moistened in
isopropyl alcohol and/or deionized water offer the best cleaning with the least
opportunity for damage to the assembly.
The Front Panel module can be damaged by deposition of film or debris inside
the knob encoders. To avoid this, care must be taken to not drip liquid or blow
debris into the encoders.
Proceed with step 3.
3. Spray wash dirty parts w ith the 75% isopropyl alcohol solution and wait 60
seconds for the majority of the alcohol to evaporate.
4. Use hot (49 _Cto60_C / 120 _F to 140 _F) deionized water to thoroughly
rinse the parts.
5. Dry all parts with low-pressure, deionized air.
6. Dry all components and assemblies in an oven or drying compartment using
low-temperature (52 _Cto66_C / 125 _F to 150 _F) circulating air.
Lubrication. Periodic lubrication is not required for the oscilloscope.
6--4
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Removal and Installation Procedures
This section describes how to remove and install the major mechanical and
electrical modules. It provides the following information:
HTools Required on page 6--6 describes the tools needed to perform the
procedures.
HList of Modules on page 6--5 describes where to locate a list of replaceable
modules.
HSummary of Procedures on page 6--6 lists the procedures for removal and
installation of modules.
HRear Feet, on page 6--7, begins a series of procedures that describe the
removal and installation of modules.
Maintenance
Preparation
Read this section to prepare for module removal and installation.
WARNING. Only qualified personnel should perform service procedures. Before
doing this or any other procedure in this manual, read the General SafetySummary and Service Safety Summary found at the beginning of this manual.
Always disconnect the power cord before disassembling any portion of the
oscilloscope.
To prevent damage to electrical components, read Preventing ESD on page 6--1.
HSee Figure 10--1 on page 10--6 for an overview of how the entire oscilloscope
is assembled.
HRead the Summary of Procedures on page 6--6 to understand which
procedure to use to access a given module. Then read Tools Required on
page 6--6 for a list of tools needed to remove and install modules in the
oscilloscope.
HIf you are disassembling the oscilloscope for cleaning, refer to Inspection
and Cleaning on page 6--2 for cleaning instructions.
List of Modules
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
The Mechanical Parts List chapter provides a list of all replaceable modules.
Any electrical or mechanical module, assembly, or part listed in the parts list is
referred to as a module.
6--5
Maintenance
Summary of Procedures
Complete disassembly is best achieved by following the procedures in the order
given in Table 6--2 .
Table 6--2: List of procedures
ProcedureBegins on page
Rear Feet6--7
Flip Feet6--9
Front-Panel Knobs6--10
Power Button6--10
Rear Case6--10
Front Feet6--14
Power Supply Module6--15
Internal Assembly6--17
Display Cable6--19
Front-Panel Cable6--21
Main Board Module6--22
Display Module6--24
Front-Panel Module6--26
Tools Required
Keypad6--28
Front Case6--30
Use the following tools to remove and replace all modules in the oscilloscope:
HTorque-limiting screwdriver, long shank, 12 in-lb (1.3 N
⋅m) range with
Torx T-15 tip
H9/16 inch deep socket with wide center hole (S
⋅K Hand Tool Corp., S-K
#40818, or equivalent)
HTorque wrench to 40 in-lb (4.5 N
⋅m)
HScrewdriver with 1/4 inch flat blade
HPliers (all-purpose and needle-nose)
HSharp knife (X-ACTO
knife or equivalent)
6--6
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Rear Feet
You will need a flat-blade screwdriver to remove the rear feet.
Removal. To remove the rear feet, refer to Figure 6--1 and follow these steps:
1. Place the oscilloscope face down on a soft surface such as an anti-static mat.
2. Use a flat-blade screwdriver to pry out the foot, or twist the foot as you are
removing it so that it disengages from the guides. Use care not to mar the
surface of the rear case.
Figure 6--1: Removing the rear feet
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--7
Maintenance
Installation. To install the rear feet, refer to Figure 6--2 and follow these steps:
1. Place the oscilloscope face down on a soft surface such as an anti-static mat.
2. Align the slots of the rear foot with the guides in the rear case and press the
foot in.
6--8
Figure 6--2: Installing the rear feet
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Flip Feet
You will need a pair of needle-nose pliers to remove the flip feet.
Removal. To remove the flip feet, refer to Figure 6--3 and follow these steps:
1. Place the oscilloscope face down on a soft surface such as an anti-static mat.
2. Grip the flip foot with the pliers as close to the notches as possible and
squeeze gently on the foot.
3. Twist slightly to swing one hinge free of the slot.
Hinge
To remove foot, squeeze at notches
and swing one hinge free of slot
Figure 6--3: Removing and installing the flip feet
To install foot, place hinge of
foot into one slot, then
swing other hinge into place
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--9
Maintenance
Installation. To install the flip feet, refer to Figure 6--3 and follow these steps:
1. Place the oscilloscope face down on a soft surface such as an anti-static mat.
2. Hold the flip foot with the smooth side facing down.
3. Gently squeeze the foot so that the hinge on one side fits into the slot, and
twist slightly to snap into the slot on the other side.
Front-Panel Knobs
Power Button
Rear Case
Remove each front-panel knob by firmly grasping the knob (with pliers if
necessary) and pulling it away from the front panel.
NOTE. Wrap the plier jaws with tape (such as electrical) to avoid marring the
knob surface.
Install front-panel knobs by aligning the keyed knob with the shaft and pressing
the knob onto the shaft.
Remove the power button by firmly grasping the button (with pliers if necessary)
and pull it straight out of the cabinet.
NOTE. Wrap the plier jaws with tape (such as electrical) to avoid marring the
button surface.
Install the button by pressing it onto the power button shaft until it snaps into
place.
You will need a torque-limiting Torx T-15 screwdriver and pliers for this
procedure.
6--10
Removal. To remove the rear case, refer to Figure 6--4 and follow these steps:
1. Remove the power button using the procedure on page 6--10.
2. If installed, remove the extension module from the rear of the oscilloscope.
For instructions, refer to the TDS1000- and TDS2000-Series User Manual.
CAUTION. To avoid damage, do not remove the rear case with an extension
module installed.
3. Place the oscilloscope face down on a soft surface such as an anti-static mat.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
4. Remove the four screws located on the rear case. Tilt the handle to gain
better access to the top two screws.
5. Lift the rear case off of the oscilloscope.
Remove rear case
screws (4)
ESD Grounding ears
Figure 6--4: Removing and installing the rear case
Rear case
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--11
Maintenance
Installation. Use this procedure to install the rear case.
1. Place the oscilloscope face down on a soft surface (such as an anti-static
mat).
2. Place the rear case onto the assembled oscilloscope and ensure that:
HAll of the edges align with the front case.
HThe ESD grounding ears align with the rear case slots. If not, they must
be bent into the correct position. Figure 6--5 shows proper alignment.
HThe notches on the edge of the main board module align with ridges on
the rear case near the extension module connector slot. Figure 6--5 shows
the location of the notches.
CAUTION. Failure to properly align the ESD grounding ears may result in ESD
damage to the extension module and/or the oscilloscope. Refer to Figure 6--5.
Be careful not to pinch any wires or other parts between the front and rear cases.
3. Install the four screws in the rear case. Tilt the handle to better access the top
two holes.
CAUTION. Avoid cross threading the screws that secure the rear case to the front
case. An easy way to avoid this is to slip the screw into the hole and, using slight
pressure, slowly rotate the screw counterclockwise until you feel the screw slip
into the screw threads. Then tighten the screw clockwise. A properly started
screw will have very little resistance going in.
4. Install the power button as described on page 6--10.
6--12
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Main board notches
Figure 6--5: Aligning the oscilloscope rear case
Rear case
slots (4)
ESD grounding ears (4)
ESD grounding ears are properly
aligned with the rear case slots when
they are visible in approximately one
half the width of the slots.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--13
Maintenance
Front Feet
You will need a torque-limiting Torx T-15 screwdriver and pliers for this
procedure.
Removal. First, remove the power button and rear case using the procedures on
page 6--10. Then, slide the front feet out of the front case as shown in Figure 6--6.
Installation. Slide the feet into the slots provided in the front case as shown in
Figure 6--6. Use the installation procedures for each module removed to
reassemble the oscilloscope.
Front foot
6--14
Figure 6--6: Removing and installing the front feet
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Power Supply Module
Use a torque-limiting Torx T-15 screwdriver and pliers for this procedure.
Removal. To remove the module, refer to Figure 6--7 and follow these steps:
1. Remove the power button and rear case using the procedures on page 6--10.
2. Disconnect the following wires:
HThe ground wire on the power supply module from the chassis ground
lug.
HThe two-conductor backlight cable on the power supply module.
HThe power supply ribbon cable on the main board at J101. Press down on
the latch while pulling up on the cable.
3. Remove the four screws connecting the power supply module to the chassis
and lift the module from the internal assembly.
Remove screws (4)
Power supply
ribbon cable
Ground wire
Backlight connector
(monochrome LCD)
Backlight cable
(color LCD)
Figure 6--7: Removing the power supply module
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--15
Maintenance
Installation. To install the power supply module, refer to Figure 6--8 and follow
these steps:
1. Align and place the power supply module into the internal assembly.
2. Install the four screws. Do not put a screw into the empty hole in Figure 6--8.
3. Reconnect the following wires:
HThe ground wire on the power supply module to the chassis ground lug.
You may need to use pliers to secure the wire in place.
HThe two-conductor backlight cable on the power supply module.
Connect to the black socket for a monochrome LCD, and to the white
socket for a color LCD.
HThe power supply ribbon cable to the main board at J101. Twist the cable
four revolutions to keep it from contact with the boards and chassis.
4. Use the installation procedure for each module removed to reassemble the
oscilloscope.
Power supply module
Power supply ground wire
Chassis ground lug
Install screws (4)
Empty hole
NO SCREW
Backlight connector
(monochrome LCD)
Power supply
ribbon cable
Backlight cable
(color LCD)
6--16
Figure 6--8: Installing the power supply module
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Internal Assembly
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. To remove the assembly, refer to Figure 6--9 and follow these steps:
1. Remove the front-panel knobs, power button, and rear case using the
procedures on page 6--10.
2. Remove the four screws attaching the internal assembly to the front case.
3. Lift the entire internal assembly (consisting of the chassis, all of the boards,
and the display screen) out of the front case.
NOTE. The switch keypad will most likely remain inside the front case. It does not
need to be removed with the internal assembly.
Remove screws (4)
Figure 6--9: Removing and installing the internal assembly
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Internal
assembly
Front case
6--17
Maintenance
Installation. To install the internal assembly, refer to Figure 6--9 and follow these
steps:
1. Use the installation procedure for each module removed to reassemble the
oscilloscope.
2. Place the internal assembly into the front case.
3. Install the four screws to attach the internal assembly to the front case.
6--18
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Display Cable
You will need a torque-limiting Torx T-15 screwdriver and pliers for this
procedure.
Removal. To remove the display cable, refer to Figure 6--10 and follow these
steps:
1. Remove the power button and rear case using the procedures on page 6--10.
2. Untwist the cable tie to release the display cable.
3. Disconnect the display cable at J201 on the main board by pulling straight up
from the connector.
4. Disconnect the display cable on the display module by pulling it through the
horizontal slot on the chassis.
Main board J201
Figure 6--10: Removing the display cable
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Display cable
Display cable slot on
the chassis
6--19
Maintenance
Installation. To install the display cable, refer to Figure 6--11 and follow these
steps:
1. Remove the display module using the procedure on page 6--24.
2. Attach the display cable to the appropriate connector on the display module
at the location shown in Figure 6--11: 12-pin for the monochrome display,
and 15-pin for the color display.
3. With the display module in place but not connected, thread the display cable
through the appropriate slot on the chassis.
4. Twist the cable two revolutions to keep it from contact with the boards and
chassis, and secure using the left cable tie.
5. Connect the display cable at J201 on the main board by pushing the cable
straight down into the connector.
6. Install the four screws to attach the display module to the internal assembly.
Display
module
Display cable
Figure 6--11: Installing the display cable
Internal
assembly
Main
board
6--20
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Front-Panel Cable
You will need a torque-limiting Torx T-15 screwdriver and pliers for this
procedure.
Removal. To remove the front-panel cable, refer to Figure 6--12 and follow these
steps:
1. Remove the power button and rear case using the procedures on page 6--10.
2. Untwist the cable tie to release the front-panel cable.
3. Disconnect the front-panel cable at J202 on the main board by pulling
straight up from the connector.
4. Disconnect the front-panel cable at J1 on the front-panel module.
Front-panel cable
Main board
J202
Front-panel
module J1
Figure 6--12: Removing and installing the front-panel cable
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--21
Maintenance
Installation. To install the front-panel cable, refer to Figure 6--12 and follow these
steps:
1. Connect the front-panel cable to the front-panel module at J1 through the
appropriate slot on the chassis.
2. Twist the cable two revolutions to keep it from contact with the boards and
chassis, and secure using the right cable tie.
3. Connect the front-panel cable at J202 on the main board by pushing the cable
straight down into the connector.
Main Board Module
You will need a torque-limiting Torx T-15 long shank screwdriver, a torque
wrench, and a 9/16 inch deep socket for this procedure.
CAUTION. The main board module is particularly sensitive to ESD. Read the
Preventing ESD section on page 6--1 before removing or installing the main
board.
Removal. To remove the main board module, refer to Figure 6--13 and follow
these steps:
1. Remove the power button and rear case using the procedures on page 6--10.
2. Remove the internal assembly from the front case using the procedure on
page 6--17.
3. Disconnect the following wires from the main board by pulling straight up
from the connector:
HThe power supply ribbon cable at J101. Press down on the latch while
pulling up on the cable.
HThe display cable at J201.
6--22
HThe front-panel cable at J202.
4. Use the screwdriver to remove the two screws attaching the main board
module to the internal assembly.
5. Use the socket to remove the nuts and washers from the BNC connectors.
6. Pull the board out from the front slots of the internal assembly.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Front-panel connector
Display connector
Power supply connector
Internal assembly
Figure 6--13: Main board removal
Main board module
Remove Main board
screws (2)
Remove hex nuts and lock washers
Installation. To install the main board module, refer to Figure 6--13 and follow
these steps:
1. Working from the bottom of the internal assembly, place the front of the
main board module into the slots on the assembly (near the the BNC
connectors).
2. Use the deep socket to attach the washers and nuts to each BNC connector so
that the main board is securely attached to the chassis. Torque the nuts to 40
in-lbs using the torque wrench.
3. Install the two screws attaching the main board module to the internal
assembly.
4. Reconnect the following wires on the main board:
HThe power supply ribbon cable at J101.
HThe display cable at J201.
HThe front-panel cable at J202.
5. Use the installation procedure for each module removed to reassemble the
oscilloscope.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--23
Maintenance
Display Module
You will need a torque-limiting Torx T-15 screwdriver and pliers for this
procedure.
Removal. To remove the display module, refer to Figure 6--14 and follow these
steps:
1. Remove the power button and rear case using the procedures on page 6--10.
2. Remove the internal assembly from the front case using the procedure on
page 6--17.
3. Disconnect the display cable using the procedure on page 6--19.
4. Disconnect the backlight cable from the power supply module.
5. Remove the four screws attaching the display module to the internal
assembly. The location of the screws as shown in Figure 6--14 will vary
between the color and monochrome displays.
CAUTION. The display shield is attached to the display module. Removing the
shield will destroy the attaching gasket.
Power supply
module
Backlight
cable
Display
module
Display
ribbon cable
Figure 6--14: Removing the display module
Internal
assembly
Main
board
6--24
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Installation. To install the display module, refer to Figure 6--15 and follow these
steps:
1. Reconnect the display cable at J201 on the main board by pushing the cable
straight down into the connector.
2. Route the display ribbon cable through the opening in the chassis. For more
information, see the display cable procedure on page 6--19.
3. Route the backlight cable through the opening in the internal assembly.
4. Install the four screws to attach the display module to the internal assembly.
The location of the screws will vary between color and monochrome
displays.
5. Reconnect the backlight cable to the power supply module. Connect to the
black socket for a monochrome LCD, and to the white socket for a color
LCD.
6. Use the installation procedure for each module removed to reassemble the
oscilloscope.
Power supply
module
Backlight
cable
Display
module
Display
ribbon cable
Figure 6--15: Installing the display module
Internal
assembly
Main
board
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--25
Maintenance
Front-Panel Module
You will need a torque-limiting Torx T-15 screwdriver and a flat-blade screwdriver for this procedure.
Removal. To remove the front-panel module, refer to Figure 6--16 and follow
these steps:
1. Remove the power button and rear case using the procedures on pages 6--10.
2. Remove the entire internal assembly from the front case using the procedure
on page 6--17.
3. Disconnect the front-panel ribbon cable at J202 on the main board by pulling
straight up from the connector.
4. Remove the four screws attaching the front-panel module to the internal
assembly.
5. Pull the module away from the internal assembly.
Internal
assembly
Front-panel
ribbon cable
Front-panel module
Figure 6--16: Removing the front-panel module
Main
board
6--26
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Installation. To install the front-panel module, refer to Figure 6--17 and follow
these steps:
1. Reconnect the front-panel cable at J202 on the main board by pushing the
cable straight down into the connector.
2. Route the front-panel cable through the opening in the internal assembly. For
more information, see the procedure on page 6--21.
3. Install the four screws to attach the front-panel module to the internal
assembly.
4. Use the installation procedure for each module removed to reassemble the
oscilloscope.
Internal
assembly
Front-panel
ribbon cable
Front-panel module
Figure 6--17: Installing the front-panel module
Main
board
module
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--27
Maintenance
Keypad
You will need a torque-limiting Torx T-15 screwdriver for this procedure.
Removal. To remove the switch keypad, refer to Figure 6--18 and follow these
steps:
1. Remove the power button and rear case using the procedures on page 6--10.
2. Remove the entire internal assembly from the front case using the procedure
on page 6--17.
3. Lift the keypad out of the front case.
CAUTION. Keep the electrical contacts on the back of the keypad clean. Skin oils
and dust will inhibit good electrical contact.
Use care to avoid touching
the electrical contacts
Figure 6--18: Removing and installing the keypad
Keypad
Front case
6--28
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
Installation. To install the switch keypad, refer to Figure 6--18 and follow these
steps:
1. Place the front case face down on a soft surface.
2. Align the switch keypad into the front case using the guide posts and the
holes in the keypad.
CAUTION. Keep the electrical contacts on the back of the switch pad clean. Skin
oils and dust will inhibit good electrical contact.
3. Check that all buttons are properly seated into place before proceeding.
4. Place the entire internal assembly into the front case.
5. Use the installation procedure for each module removed to reassemble the
oscilloscope.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--29
Maintenance
Front Case
You will need a torque-limiting Torx T-15 screwdriver for this procedure.
Removal. To remove the front case, follow these steps:
1. Remove the power button and rear case using the procedures on page 6--10.
2. Remove the entire internal assembly from the front case using the procedure
on page 6--17.
3. Remove the switch keypad using the procedure on page 6--28.
Installation. To install the front case, refer to Figure 6--9 on page 6--17 and follow
these steps:
1. Place the switch keypad into the front case.
2. Place the entire internal assembly into the front case.
3. Use the installation procedure for each module removed to reassemble the
oscilloscope.
6--30
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Troubleshooting
Maintenance
This section contains information and procedures to help you isolate a defective
module in the oscilloscope.
WARNING. Before doing this or any other procedure in this manual, read the
General Safety Summary and Service Safety Summary found at the beginning of
this manual. Also, to prevent possible injury or damage to electrical components,
read Preventing ESD on page 6--1.
Adjustment After Repair
Required Tools and
Equipment
Troubleshooting Tree
If you replace the main board or power supply, you must adjust the oscilloscope
after repair. Refer to the chapter Adjustment Procedures.
You may need the the following tools and equipment to troubleshoot the
oscilloscope:
Tools and equipmentExample
Oscilloscope with 1X/10X and 100X probesTektronix TDS1000- or TDS2000-series
oscilloscope with standard accessory 1X/10X
probe (P2200) and optional 100X probe
(P5100 or P5200)
DMM3.5 or greater digit DMM
To isolate a problem to a specific troubleshooting procedure, use the troubleshooting tree in Figures 6--19, 6--20, and 6--21. Detailed procedures follow the
tree.
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--31
Maintenance
Put scope or
frequency counter
on probe comp
1 kHz?
Yes
Verify LCD
connector and
cable values
No
Replace main
board
No
Run probe check
on each channel
Turn on power
Does screen
lighten?
Yes
See
boot up
screen?
Yes
Errors?
No
Front panel
problems?
No
No
Yes
Yes
From Figure 6-21
A
To Figure 6-21
B
C
To Figure 6-20
D
To Figure 6-20
No
Yes
Replace display
or display cable
Probe
check ok?
Connect
external trigger
to probe comp
Set trigger
source to EXT
Press set 50%
Good
voltages?
Replace main
board
Figure 6--19: Oscilloscope troubleshooting tree (1 of 3)
6--32
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Yes
No
Replace main
board
Trigger
indicator ON?
Yes
Good scope
No
Replace main
board
Maintenance
C
From Figure
6-19
D
From Figure
6-19
Able to run
front panel
tests?
Yes
Run front panel
tests in service
menu
<5key
problem
Yes
No
No
Check cabling to
main board.
Replace cable;
front panel or main
board
Check cabling to
main board;
Replace cable or
main board
Module
errors?
No
Verify power
supply voltages
Power
supplies OK?
Yes
Replace main
board
Replace power
supply
Yes
No
supply Voltages
No
Reinstall module
Disconnect
power supply
Verify power
Power
supplies OK?
Replace front
panel board
Figure 6--20: Oscilloscope troubleshooting tree (2 of 3)
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Yes
Replace main
board
6--33
Maintenance
To Figure 6-19
From Figure 6-19
A
B
Power cord
attached?
Open case
Fuse OK?
Yes
Backlight cable
attached?
Using a high
voltage 100X
probe test voltages
on backlight
connectors
No
Yes
No
No
Yes
Attach power
cord
Reinstall case
Replace power
supply
Attach backlight
cable
~50 kHz
>700 V
Detach display
~50 kHz
>700 V
Replace power
supply
pk--pk
No
pk--pk
No
?
?
Figure 6--21: Oscilloscope troubleshooting tree (3 of 3)
Yes
Yes
Replace display
6--34
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Maintenance
PROBE COMP Output
Troubleshooting the
Power Supply
When the oscilloscope is on and operating properly, the PROBE COMP output
should generate a square wave, approximately 5 V in amplitude, at a 1 kHz
frequency. Use the oscilloscope and set the Attenuation switch to 10X on the
P2100 probe to probe this output.
As shown in the table below, there are two cases of defects. Either the 1 kHz
signal is on and the power supply and main board are both active and functionally capable of doing acquisition, or one of those two modules is not performing
correctly. While a large number of different possible frequencies exist, all
involve detected failures on the main board due to main board failure or power
supply failure.
Signal at PROBE COMPPossible problem
1 kHz signal, no displayRefer to Troubleshooting the Display (page 6--36)
Non-1 kHz signalRefer to Troubleshooting the Main Board (page 6--43)
No signalRefer to Troubleshooting the Power Supply (page 6--35)
To troubleshoot the power supply, follow these steps:
1. Remove the rear case using the procedure Rear Case on page 6--10.
2. Use the test oscilloscope to measure the voltages from the power supply
module at J101 on the main board module. The table below lists the voltages
you should expect to see.
Maximum
Current
SupplyVoltage range*
35 V30 VDCto 40 VDCfloating.
Pin 2 connected to +3.3 V
on the main board.
+6 V5.5 V to 6.5 V0.7 A43, 5, 8,10+5 V
+3.3 V3.0 V to 3.6 V. Requires
minimum load to maintain
regulation.
-- 4 V-- 5 .0 V t o -- 3 . 5 V0.8 A93, 5, 8,10-- 2 . 5 V
Line trigger--2 V to 6 V open circuit.
±1 diode drop when attached to the main board.
Draw
15 mA12+28 V LCD
1.5 A6, 73, 5, 8,103.3 V, +2.5 V
1mA113, 5, 8,10Line trigger
J101
power
pins
J101
return
pins
Derived
supplies
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
6--35
Maintenance
Maximum
Current
Supply
Backlight
(monochrome or
color)
*
With 3 W minimum load; still functions with no load.
Voltage range*
>500 V
, ∼50 kHzNANANA
pk-pk
Draw
J101
power
pins
J101
return
pins
Derived
supplies
3. If all of the voltages are present, the main board is probably defective.
Replace it.
4. If all or some of the voltages are missing, turn off the oscilloscope and
disconnect the cable at connector J101 on the main board module.
5. Turn on the oscilloscope and check the voltages at the loose end of the cable
disconnected from J101, checking for the same voltages as in step 2.
6. If all of the voltages are missing, check the line fuse on the power supply
module. If the fuse is defective, replace the power supply module.
7. If the fuse is okay, and all or some of the voltages are missing, the power
supply module is probably defective. Replace it.
Troubleshooting the
Display
To troubleshoot a nonfunctional display, follow these steps:
1. Remove the rear case using the procedure Rear Case on page 6--10.
2. Turn on the oscilloscope and ensure that the display lights up. If it does not,
either the display or power supply is defective. If this procedure is not
conclusive, continue with Troubleshooting the Power Supply on page 6--35
NOTE. Without a functional display module, the simplest way to verify that the
oscilloscope is on and functioning is to check for a signal at the PROBE COMP
output. The PROBE COMP output generates approximately a 5 V, 1 kHz square
wave.
3. Use the test oscilloscope and set the Attenuation switch to 10X on the P2200
probe to probe the display cable connector at J201 on the main board. The
following table describes the signals you should expect to see. See the
6--36
TDS1000- and TDS2000-Series Digital Storage Oscilloscopes Service Manual
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.