• REVIEW AND UNDERSTAND ALL SAFETY PROCEDURES IN THE “HANDLE WITH CARE”
BOOKLET FORM # TW-10 P/N 7950-8052 BEFORE ATTEMPTING TO INSTALL, OPERATE
OR PERFORM MAINTENANCE ON THIS LN
• DO NOT ATTEMPT TO USE OR MAINTAIN ANY LIQUID NITROGEN FREEZER UNTIL YOU
READ AND UNDERSTAND THESE INSTRUCTIONS.
• DO NOT PERMIT UNTRAINED PERSONS TO USE OR MAINTAIN THIS UNIT.
• IF YOU DO NOT FULLY UNDERSTAND THESE INSTRUCTIONS, CONTACT YOUR SUPPLIER
FOR FURTHER INFORMATION.
• BEFORE ATTEMPTING TO OPERATE THIS CONTROLLER WITH ANY CRYOSCIENCE BY
TAYLOR-WHARTON™ LABS & K SERIES UNITS, YOU MUST READ THE SEPARATE
OPERATING AND SAFETY MANUAL PROVIDED WITH THAT TAYLOR-WHARTON UNIT.
*CS SERIES Control Systems are used on all Taylor-Wharton LABS units.
The unit featured is only one of the cryostorage systems available in the CryoScience by Taylor-Wharton™ lines.
Nitrogen is an inert, colorless, odorless, and tasteless gas making up four-fifths of
the air you breathe – and can be very dangerous. Air is roughly one-fifth oxygen.
Liquid nitrogen is at a temperature of -196°C (-320°F) under normal atmospheric
pressure. Cryogenic freezers are used in LN2 service only.
Extreme Cold - Cautionary Statement
Accidental contact of liquid nitrogen or cold issuing gas with the skin or eyes may
cause a freezing injury similar to frostbite. Handle the liquid so it won’t splash or
spill. Protect your eyes and cover the skin where the possibility of contact with the
liquid, cold pipes and equipment, or cold gas exists. Safety goggles or a face shield
should be worn when operating this equipment. Insulated gloves that can be easily
removed and long sleeves are recommended for arm protection. Trousers without
cuffs should be worn outside boots or over the shoes to shed spilled liquid.
Keep Equipment Area Well Ventilated
Although nitrogen is non-toxic and non-flammable, it can cause asphyxiation in
a confined area without adequate ventilation. Any atmosphere not containing
enough oxygen for breathing can cause dizziness, unconsciousness, or even death.
Nitrogen, a colorless, odorless, and tasteless gas that cannot be detected by the
human senses, will be inhaled normally as if it were air. One (1) liter of liquid
nitrogen is equivalent to 696 liters of nitrogen gas. Without adequate ventilation, the
expanding nitrogen will displace the normal air resulting in death.
WARNING:
The following safety
precautions are for
your protection.
Before installing,
operating, or
maintaining this unit
read and follow all
safety precautions
in this section
and in reference
publications. Failure
to observe all safety
precautions can
result in property
damage, personal
injury, or possibly
death.
WARNING:
Maintain adequate
ventilation to
prevent asphyxiation
hazard (see Safety
Precautions).
Liquid Nitrogen System
The liquid nitrogen supply pressure at the inlet to the refrigerator should be in
the range of 10 psig (0.7 bar/69 kPa) to 20 psig (1.4 bar/138 kPa) for optimum
performance. Higher operating pressures will increase transfer losses and create
excessive turbulence of the liquid in the refrigerator, which can generate false signals
to the liquid level controller causing the refrigerator to under-fill. In “liquid phase”
storage applications, excessive turbulence can cause splashing which could result in
personal injury and/or damage to the refrigerator. When installing piping or fill hose
assemblies, make certain a suitable safety relief valve is installed in each section
of plumbing between any two isolation points. Trapped liquefied gas will expand
greatly as it warms and may burst hoses or piping causing damage or personal
injury. A relief valve is installed in the refrigerator plumbing to protect the line
between the customer supplied shut-off valve and the refrigerator solenoid valve.
Relief valves can be piped to the outside of the building.
Electrical
• This product is not intended for a life support function.
• This product is intended to be used in hospitals and clinics.
• This product has no Radio Transmitter (Intentional Radiator) functions.
• This product is not intended for electromagnetic shielded rooms only.
• This product does not intentionally apply RF energy for its function.
• This product does not intentionally receive RF energy for its function.
Caution:
When installing field
fabricated piping,
make certain a
suitable safety relief
valve is installed
in each section of
piping between any
two isolation points.
WARNING:
Inlet pressure should
not exceed 22 psig
(1.5 bar/152 kPa).
Higher pressures
could result
in damage to
equipment.
WARNING:
Electrical shock can
kill. Do not attempt
any service on these
units without first
disconnecting the
electrical power cord.
The liquid level controllers used with these refrigerators operate from 12 VDC.
Disconnect the electrical power cord from the outlet before attempting any service.
3
CS SERIES
For more detailed information concerning safety precautions and safe practices to
be observed when handling cryogenic liquids consult CGA publication P-12 “Safe
Handling of Cryogenic Liquids” available from the Compressed Gas Association:
CGA website: www.cganet.com; CGA customer service +1.703.788.2700; or email
customerservice@cganet.com.
GENERAL INFORMATION
The CS SERIES Control System can monitor and control both the liquid nitrogen
level and the vapor temperature range in the cryostorage unit you have selected. CS
SERIES Control Systems are designed to work with Taylor-Wharton LABS Cryostorage
Systems. The features are designed to provide a safe environment for samples while
at the same time tracking all relevant information associated with the freezer. This
control provides a complete historical record of the environment in your unit and
therefore, the environment in which your samples have been stored in this system.
This controller features a vacuum fluorescent display. The addition of a liquid
nitrogen supply and inventory control racks for systematic retrieval of stored product
completes the total Cryostorage System.
Taylor-Wharton LABS Cryostorage Systems are designed for applications where
extremely low temperature storage of biological products is required. They are also
appropriate for industrial or other applications where liquid nitrogen temperatures
and high capacity are needed.
Before beginning installation or operation of this CS SERIES Control System, make
sure that you read and understand this manual as well as the operating and safety
instructions for the cryostorage unit you will be using with this controller.
4
CS SERIES Control System Specifications
Control Type: LN2 Level Control & Temperature Control
Level Measurement Sensor Type: 8-Thermistor Fixed
4-Thermistor Adjustable
Range: 8 inch range (8-Thermistor)
Low, Normal, High (4-Thermistor)
Redundancy Multiple discrete points
CS SERIES
Temperature Measurement: Sensor Type:
Accuracy of 1.5% with Resolution of 0.1°c
Number of Channels: 2
Temperature Display Units: °C, F, K, R
Electrical: Input Voltage: 100-240 VAC
Input Current (max: 1.75 A
Input Current (continuous): 0.5 A
Power Consumption (max): 21 W
Power Consumption (continuous): 6 W Input
Frequency: 50/60 Hz
Output: 12 VDC
Control Input Voltage: 12 VDC
Power Cord: Available for all countries
Battery: Rating: 12 Volt, 18Ah
Type: Absorbent Glass Mat (AGM) Sealed Non-
Spillable
Short Protection: Installed PCB with thermal fuse
Battery Cover: Vinyl
Solenoid Valve Input Voltage: 12 VDC
Input Current 0.96 amps
Number of Communication Ports: 3
User Interface: Display Type Vacuum Fluorescent
Display: (VFD)
Buttons:11
Level, Temperature and Alarm
Information
“At a Glance” Status LED status wheel
Filling 1 LED
Menu Access 1 LED
Power LED
Control Tests: Power Up Self Test Control System Check
Thermistor Status
Battery voltage
Control voltages
Temperature Circuit
Alarms: Low Level Alarm: Always enabled
High Level Alarm: Always enabled
Sensor Error Alarm: Always enabled
High Temperature Alarm: (T/C #1, T/C #2)
Programmable
5
CS SERIES
Alarms (continued...): Low Temperature Alarm: (T/C #1, T/C#2)
Programmable
Thermocouple Calibration: Alarm Always enabled
Thermocouple Open Alarm: Always enabled
Power Failure (Remote only): Always enabled
Low LN2 Supply Alarm: Programmable
Battery Mode Warning: Always enabled
Lid Open Too Long Alarm: Programmable
Valve Stuck Open Alarm: Programmable
Unauthorized Access Warning: Programmable
LN2 Usage Warning: Programmable
Low Battery Voltage: Always enabled
Temperature Alarm Delay: Programmable
Audible Alarm: Always enabled
Audible Alarm Re-trigger: Programmable
Visual Alarm Indicator: Always enabled
Remote Alarm Delay: Programmable
Buttons: Power: Turns power on/off
Fill/Defog: Open Solenoid Valve
Stop: Close Solenoid Valve
Menu: Access Menu
Mute: Silence audible
Enter: Save a setting or select a menu choice
Back/Exit: Leave a setting unchanged or back out
of menu
Up Arrow: Scroll the menu system or increase a
value
Down Arrow: Scroll the menu system or decrease
a value
Left Arrow: Scroll horizontal menu
Right Arrow: Scroll horizontal menu
Data Collection: Temperature
Level
Alarms
Memory: 4 Mb
Dimensions: Display Width: 9.5 in. (241 mm)
Display Height: 2.0 in. (50.8 mm)
Display Depth: 1.31 in. (33.3 mm)
Display Weight: 0.625 lbs (.28 kg)
Main Control Width: 8.875 in. (225.4 mm)
Main Control Height: 6.688 in. (169.9 mm)
Main Control Depth: 1 in. (25.4 mm)
Main Control Weight: 1.0 lbs (.45 kg)
Battery Width: 7.25 in. (184.2 mm)
Battery Height: 6.375 in. (161.9 mm)
Battery Depth: 3.25 in. (82.6 mm)
Battery Weight: 12.4 lbs (5.6 kg)
6
DELIVERY AND RETURNS
Unpacking and Inspection
Inspect shipping containers for external damage.
All claims for damage (apparent or concealed)
or partial loss of shipment must be made in
writing within five (5) days from receipt of goods.
If damage or loss is apparent, please notify the
appropriate parties as indicated below:
Domestic LTL Shipments – The customer
should notify and file the appropriate damage
claims with the carrier. All products are shipped
FOB Origin.
CS SERIES
Figure 1.0 Crated LABS 38K
delivery, and Taylor-Wharton must also be notified. Confirm with Taylor-Wharton
Customer Service the filing procedures for any UPS damage claims.
International Shipments – Any damage and/or claims are to be filed with the
carrier. Insurance agent(s) and customs’ brokers should also be notified.
In all cases, Taylor-Wharton should be notified so we can assist if needed
in filing damage claims.
Open the shipping containers; a packing list is included with the system to simplify
checking that all components, cables, accessories, and manuals were received.
Please use the packing list to check off each item as the system is unpacked. Inspect
for damage. Be sure to inventory all components supplied before discarding any
shipping materials. If there is damage to the system during transit, be sure to file
proper claims promptly. Please advise Taylor-Wharton of such filings. In case of parts
or accessory shortages, advise Taylor-Wharton immediately. Taylor-Wharton cannot
be responsible for any missing parts unless notified within 10 days of shipment.
Domestic UPS Shipments – Any damage
should be noted and reported to shipper upon
Freight Damage Procedures
Any freight damage claims are your responsibility. Cryostorage Systems
are delivered to your carrier from Taylor-Wharton’s dock in new
condition; when you receive our product you may expect it to be in that
same condition. For your own protection, take time to visually inspect each
shipment in the presence of the carrier’s agent before you accept delivery. If any
damage is observed, make an appropriate notation on the freight bill. Then, ask the
driver to sign the notation before you receive the equipment. You should decline to
accept containers that show damage which might affect serviceability.
Repackaging for Shipment
If it is necessary to return any part of the system for repair or replacement, a
Material Return Authorization (MRA) number must be obtained from an authorized
factory representative before returning the equipment to our service department.
Contact your distributor for return authorization. When returning equipment for
service, the following information must be provided before obtaining an MRA:
A. System model and serial number, and controller model and unit, if available.
B. User’s name, company, address, and phone number
C. Malfunction symptoms
7
CS SERIES
CAUTION:
If using a power
source other than
North American
110/220 VAC,
contact TaylorWharton Customer
Service to make
sure you are using
the appropriate
converter. Use of an
unapproved power
source converter may
cause permanent
damage to the unit.
If possible, the original packing material should be retained for reshipment. If not
available, consult Taylor-Wharton for shipping and packing instructions. It is the
responsibility of the customer to assure that the goods are adequately packaged for
return to the factory. All refrigerators returned to Taylor-Wharton must be clean and
sterile before return. Refer to the operating manual of the freezer you are using for
cleaning instructions.
INSTALLATION
Getting Unit into Service
Your Cryostorage System comes with complete instructions for how you should
remove the unit from the crate and put it into service. Read both this manual and
your Cryostorage System’s manual before beginning any installation. Make sure to
follow any required procedures and safety guidelines when you are connecting your
Liquid Nitrogen source.
The CS SERIES Control System is designed to be operated at normal room
temperatures 15° C to 27° C (60° F to 80° F) at a relative humidity level below 50%.
The humidity level should be maintained such that the electronics are not exposed to
condensation.
The Taylor-Wharton Cryostorage freezer should be positioned such that the all sides
of the unit are easily accessible and the user can easily connect/disconnect the
power cord from the wall socket.
Proper ventilation MUST BE adequate to sustain life for those working with or
maintaining this equipment.
Electrical
WARNING:
Electrical shock can
kill. Do not attempt
any service on these
units without first
disconnecting the
electrical power cord.
WARNING:
Maintain adequate
ventilation to
prevent asphyxiation
hazard (see Safety
Precautions).
WARNING:
If the fill fails to
stop for any reason,
quickly close the
liquid supply valve
to prevent overfilling
until the cause of
the problem can be
determined.
The liquid level controllers used with these refrigerators operate at 12 VDC. The
external transformer has a 100/240 VAC primary. Disconnect the electrical power
cord from the outlet before attempting any service.
Electromagnetic Compatibility (EMC)
Although this equipment conforms to the intent of the 2004/108/EC EMC Directive,
all medical equipment may produce electromagnetic interference or be susceptible
to electromagnetic interference. The following are guidance and manufacturer’s
declarations regarding EMC for the CS SERIES Control System.
The CS SERIES Control System needs special precautions regarding EMC and needs
to be installed and put into service according to the EMC information provided in
the following pages.
Portable and Mobile RF communications equipment can affect the performance
of the CS SERIES Control System. Please use the guidelines and recommendations
specified in the EN Compliance tables found on pages 43-46.
Other Medical Equipment or Systems can produce electromagnetic emissions and
therefore can interfere with the functionality of the CS SERIES Control System. Care
should be used when operating the CS SERIES Control System adjacent to or stacked
with other equipment. If adjacent or stacked use is necessary, the CS SERIES Control
System should initially be observed to verify normal operation in the configuration in
which it will be used.
The electrical cables, external power supplies and accessories listed or referenced
in this manual have been shown to comply with the test requirements listed in the
8
CS SERIES
EN Compliance tables found on pages 43-46. Care should be taken to use only
manufacturer-recommended cables, power supplies and electrical accessories
with the CS SERIES Control System. If a third-party supplier offers cables, external
power supplies and electrical accessories for use with the CS SERIES Control System
and they are not listed or referenced in this manual, it is the responsibility of that
third-party supplier to determine compliance with the standards and tests in the EN
Compliance tables found on pages 43-46.
The use of electrical cables and accessories other than those specified in this manual
or referenced documents may result in increased electromagnetic emissions from
the CS SERIES Control System or decreased electromagnetic immunity of the CS
SERIES Control System.
Power Supply Connection
Connect the power supply to your LABS Cryostorage System and then plug the
power supply into a surge-protected 110/220 VAC outlet.
Validation
Some organizations require that equipment be validated periodically. If information
is needed on the proper techniques to validate this equipment,please contact your
supplier.
WARNING:
This equipment
is intended for
use by healthcare
professionals. As
with all electrical
medical equipment,
this equipment
may cause radio
interference or
may disrupt the
operation of nearby
equipment. It may
be necessary to take
mitigation measures
such as re-orienting
or relocating the
CS SERIES Control
System unit or
shielding the
location.
OPERATION
These instructions are for operators experienced with cryogenic equipment. Before
operating the system, become familiar with the safety precautions in this manual
and in reference publications. Make certain all applicable provisions set forth in the
Installation Section have been followed before placing a system in operation. Study
this manual thoroughly. Know the location and function of all system components.
Initial Fill
The Cryostorage System, using the CS SERIES Controller, comes preset from the
factory. The liquid nitrogen supply pressure at the inlet to the refrigerator should be
in the range of 10 psig (0.7 bar/69 kPa) to 20 psig (1.4 bar/138 kPa) for optimum
performance. Higher operating pressures will increase transfer losses and create
excessive turbulence of the liquid in the refrigerator which can generate false signals
to the liquid level controller causing the refrigerator to under fill. In “liquid phase”
storage applications, excessive turbulence can cause splashing which could result in
personal injury.
For complete instructions for the initial fill refer the Quick Start Guide that comes
with your cryogenic freezer model.
9
CS SERIES
Control Components
CS SERIES Control System Components
The CS SERIES Control System for the Taylor-Wharton Cryostorage freezer consists of
the following components:
• Main Control Module
• VFD (Vacuum Florescent Display) Module
• Shielded Display Cable
• Wiring Harness Assembly
• Power Supply
• Thermocouple Assembly (Optional Second Thermocouple)
• Sensor Assembly
• Cryogenic Solenoid Valve
• Remote Alarm Plug
• Lid Switch Assembly
• 12 Volt, 18aH battery
• Solenoid Valve Assembly with Freezeguard
• Strainer
The CS SERIES Control System is assembled onto the freezer at the factory and
completely tested. Refer to theQuick Start Guidefor freezer set-up. Start operation
of the control system by plugging the power cord into the wall outlet. Press and hold
the POWER button for 2 seconds. The CS SERIES Control System will go through
a short startup routine and then start operation. Refer to Figure 2.0 below CS200
Controller Display for a description of the controller’s navigation buttons.
CS200 Controller Display Guide
Status WheelVFD Graphics Display
Shows information
about the Freezer
Fill Button
Used to Open
the Valve
Stop Button used
to Close the Valve
Valve Indicator lights
if the Valve is Open
Figure 2.0 CS200 Controller Display
Enter button “selects”
a Menu choice or setting
Mutes the
Audible Alarm
Directional Buttons
Increment/decrement values or allow
movement through the menu system
Power Indicator
Power On/Off
Back Button moves
out of Menu System
or leaves current
Setting unchanged
Enter/Exit
Menu System
Menu Indicator
10
Operational Theory
The CS SERIES Control System automatically maintains the Liquid Nitrogen (LN2)
level and monitors temperature in the Cryogenic freezer. Operational conditions
are monitored and any alarm is triggered if necessary. Operations data is stored in
memory on the control board.
#8
#7
High Level Alarm#6
CS SERIES
#5Stop Fill
#4
#3Start Fill
#2Low Level Alarm
#1
Figure 3.0 Thermister LN2 Level
The CS CONTROL SYSTEM uses thermistors to measure the LN2 level within
the vessel. A thermistor is a thermal resistor and its resistance changes as the
temperature changes. When a thermistor is submerged in LN2, its resistance will be
significantly greater than its resistance at room temperature. The control can detect
this resistance change and determine the level of the LN2 within the freezer. The CS
CONTROL SYSTEM is designed to work with an 8 thermistor assembly. There are
four thermistors that can be selected that will maintain the LN2 level. These selected
thermistors correspond to Low Alarm, Start Fill, Stop Fill and High Alarm. When
the LN2 level drops below the Start Fill thermistor, the control opens a solenoid
valve allowing LN2 to enter the vessel. This continues until the Stop Fill thermistor is
submerged in LN2 at which point the solenoid valve is closed preventing the flow of
additional LN2 into the vessel. The Low Alarm, Start Fill, Stop Fill and High Alarm
settings can all be changed by the user through the menu system.
11
CS SERIES
The LN2 level is indicated on the display and is determined by the number of
thermistors submerged in LN2 according to the table below:
The CS SERIES Control System monitors a number of conditions and provides an
alarm if a problem is detected. The alarms are listed below:
Note: Please see
description of the offset.
Level displayed may vary
depending on the offset
setting.
Table 2.0 Alarm Conditions
ALARMPROBLEM DETECTED
Low Level Alarm
High Level AlarmLN2 level is too high. Thermistor #6 on the sensor assembly
Sensor Error AlarmA problem exists with the level sensor. The control detects an
High Temperature AlarmThe temperature detected is warmer than the high
Low Temperature AlarmThe temperature detected is colder than the low temperature
Thermocouple Calibration
Alarm
Thermocouple Open AlarmA problem exists with the temperature sensor
Power FailureNo Power.
Low LN2 Supply AlarmA problem may exist with the LN2 supply connected to the
LN2 Use WarningThe consumption of LN2 has increased and should be
LN2 level is too low. Thermistor #2 on the sensor assembly is
not submerged in LN2.
is submerged in LN2.
open sensor circuit meaning that the sensor is unplugged or
the sensor assembly has been damaged. A faulty sensor is
denoted as “O” for open.
temperature alarm setting.
alarm setting.
The calibration data is incorrect.
(thermocouple). The control detects an open circuit meaning
that the sensor is unplugged or the sensor assembly has
been damaged.
freezer. This alarm occurs if the freezer does not fill within
the designated amount of time determined by the setting
on the control. Possible reasons include an empty supply
cylinder, low head pressure in the supply cylinder or a closed
shut off valve.
checked.
12
Table 2.0 Alarm Conditions
ALARMPROBLEM DETECTED
Operating in Battery Mode
Alarm Warning
Low Battery VoltageThe voltage on the battery is low.
Lid Open Too Long AlarmThe lid has been opened for a period which is longer than
Valve Stuck Open AlarmThe valve is stuck open.
Unauthorized Access
Warning
Temp Alarm DelayThis is the amount of time after a warm temperature is
Audible Alarm RetriggerThe audible alarm is retriggered if the error condition that
Remote Alarm RelayThe control provides a relay to provide an external signal
The power from the power supply has been disrupted and
the control system is operating on battery power.
the designated alarm setting.
The lid has been opened and an incorrect or no
identification has been entered.
detected before the control goes into alarm.
caused it is not corrected. The retrigger time can be adjusted
by the user.
that an alarm condition has occurred. The user can set the
remote alarm timer that determines the amount of time an
error must be active before the relay is triggered.
CS SERIES
All alarms include the following:
• The flashing status wheel flashes to signal an error condition
• An audible tone sounds
• The error detected is displayed and scrolled on the screen
• The remote alarm relay changes state to provide a dry contact output signal
Lid Switch
The LABS Lid Switch (Figure 4.0) is attached to the hinge and determines whether or
not the lid is open on the freezer. This also allows the control to determine whether
to activate the Auto Defog, Quick Chill or Lid Alarm features.
Lid Switch
Figure 4.0 LABS Lid Switch
Solenoid Valve
These units are designed to work with 12 VDC solenoid valve (see Figure 17.0 LABS
Plumbing Assembly on page 37).
Thermocouples
Type T thermocouples monitor the temperature in the freezer. The user may choose
to use NONE, 1 or 2 thermocouples with this control at any time. (The unit comes
complete with one Thermocouple)
13
CS SERIES
Remote Alarm Connector
Power Supply
A 12 VDC power supply is supplied for the CS SERIES Control System. The system is
supplied with a transformer compatible with common household (North American)
100/240 VAC. (For other power outlets contact Taylor-Wharton Customer Service.)
UL approval for the system as a whole is not required since the control operates on
low voltage. If your power source differs, or is subject to disruption or line surges
due to other equipment on line, consult your Taylor-Wharton representative.
Remote Alarm
If an error condition occurs after a user defined period of time, a remote alarm
can be initiated. This is accomplished by connecting a remote device to the remote
alarm jack on the rear electrical panel. The 3-pin jack on the back of the unit
provides continuity between pin #2 (common) and pin #3 in the normal condition.
Continuity between pin #1 and pin #2 is provided in an error condition.
Figure 5.0 Remote Alarm Plug Connection
Operating Parameters
When materials are immersed in liquid nitrogen, they will assume the temperature
of the liquid (-196°C/-320°F). When material is stored in the vapor phase over the
liquid, the liquid nitrogen is still a very cold refrigerant, but the refrigerator’s interior
temperature increases as product is stored higher over the liquid. This temperature
differential is not significant for many biological storage applications, and is affected
by the amount of product stored in the refrigerator, the type size and material of
inventory control system, and the liquid level in the unit.
The liquid level in the refrigerator is determined by the position of the of the
Thermistor Assembly in the sensor tube. These sensors are set at installation to
maintain a specific liquid level. A filling cycle is initiated when the level falls below
the Start Fill sensor and is completed when the Stop Fill sensor is reached. This
filling cycle repeats when the level fall below the Start Fill sensor. Sensor Probe
assignments may be changed on the controller keypad to define new start and stop
levels, and these levels may be set independently to vary the liquid level deference
between fills. Prior to the initial fill of the refrigerator, a determination should be
made whether vapor phase or liquid phase storage will be utilized.
All units are supplied with an eight thermistor assembly and a freeze-guard sensor
unless otherwise specified. The LABS factory setting positions will maintain liquid
level within a distance of 2 in. (50.8 mm) from the bottom of the operating tray on
the LABS 20K, 38K, 40K, 80K units, and 3 in. (76.2 mm) from the bottom of the
operating tray on the LABS 94K.
14
Temperature Monitoring
The CS SERIES Control System uses a type T thermocouple to monitor the
temperature in the vessel. The thermocouple is factory-installed near the top of
the Taylor-Wharton Cryostorage freezer vapor chamber and the temperatures are
measured at that point.
The CS SERIES Control System accommodates a second user-installed thermocouple
for temperature measurement at a secondary location. This is also a Type T
thermocouple.
The control provides a High Temperature Alarm for each thermocouple which can
be selected by the user. If the temperature exceeds the temperature alarm set point,
the status wheel flashes and an audible alarm is triggered.
Liquid Phase Storage
Liquid phase storage is normally utilized when -196 is required to maintain stored
product viability and the storage medium is adequate for storage in liquid nitrogen.
In a typical liquid phase storage system, the liquid level sensors are positioned to
maintain the liquid level at or below the top level of the inventory control system.
During operation, the upper levels of the inventory control system will at times
become exposed as the liquid level fluctuates.
CS SERIES
Care must be taken to ensure that the liquid level remains below the bottom of the
refrigerator lid. Exposure to liquid nitrogen may result in physical damage to the lid.
Additionally, operating the refrigerator with high liquid levels characteristic of liquid
phase storage may result in turbulence during fill cycles. Caution must be exercised
if the refrigerator lid is opened during a fill, and appropriate safety equipment
should always be worn.
The Taylor-Wharton Cryostorage Systems are set at factory for vapor phase storage.
Maintenance
To insure proper operation and maintain excellent performance of the TaylorWharton Cryostorage freezer, an annual maintenance schedule should be followed
for the CS SERIES Control System. This would include the following:
Table 3.0 Annual Maintenance Schedule
CS SERIES Control
System
Harness Assembly
Battery
Solenoid Valve
Lid Switch
Level Sensor Assembly
Thermocouple Assembly
Power Supply
Examine for exposure to moisture, wear and tear,
connector problems, and cracks or other damages
to the faceplate or buttons. In addition, periodic
firmware updates may be important.
Examine for damage to the cable and damage to
connectors.
Examine connection cable and connector for
damage.Examine vinyl cover for damage.
Replace every 2 years.
Examine wires and connector for damage.
Replace every 3 years.
Examine lid switch pickup and wires for damage.
Replace if necessary.
Examine for damage to wires and connector.
Examine for damage to wires and connector.
Examine for damage to power supply and power
cords.
15
CS SERIES
In addition, inspection and preventive maintenance should also be performed on
the freezer and its mechanical parts. Refer to owner’s manual for details.
If any intermittent alarm or defect with the Cryo-Storage System is observed or
suspected, it should be investigated and remedied immediately even if this falls
outside of the normal maintenance schedule.
CONTROLLER OPERATION
This section of the operating manual is for Taylor-Wharton approved equipment that
uses the CS SERIES Control System.
Introduction
The CS200 Control System, temperature and LN2 level controller is designed for
easy operation and uninterrupted reliable service. This controller will maintain the
selected liquid level range of LN2 in your refrigerator as well as provide audible
and visual alarms for any non conforming conditions that may occur. An Alarm
is any condition outside the activated preset limits on the control, such as an
open sensor circuit or temperature alarm. “System Events” are lid openings and
closings, solenoid valve openings and closing, and operation of the controller’s
relay for remote alarm indication. System Events, Alarms and Temperature “Data”
can be downloaded.
The System should require no additional attention to maintain liquid level if an
adequate supply source of liquid nitrogen is maintained. If your protocol calls for
you to “top-off” the Cryostorage System at the end of a workday or workweek,
press the FILL button. The unit will fill to the upper allowable liquid level and stop
automatically. You may choose to manually stop the fill by pressing the STOP
button at anytime during the fill.
Operation Data
The CS200 CONTROL SYSTEM stores data related to the operation of the TaylorWharton Cryo-Storage freezer. This data includes date, time, LN2 level, temperature,
system events and error conditions. This can be useful for audit purposes, operation
analysis and preventive maintenance.
Communications
The CONTROL SYSTEM has been designed with advanced communications
capabilities. This allows for the transfer of data out of the control where the data
can more easily be used. Please check with your supplier for available protocols and
compatible products.
Normal Fill Cycle
When the refrigerator is filled and the controller is operating, the START FILL and
LOW ALARM sensors are immersed in liquid nitrogen. Their resistance values
are interpreted by the controller as “in liquid”. At the same time, the STOP FILL
and the HIGH ALARM sensors are above the liquid pool, informing the control
that these sensors are in vapor. As liquid nitrogen evaporates, the liquid level
in the refrigerator drops slowly until the START FILL sensor is above the liquid
and sends a different signal to the controller. After a delay sufficient to ensure
the signal, the controller interprets this condition as low liquid and opens the fill
solenoid, admitting more liquid nitrogen from the supply source.
16
The refrigerator will fill slowly. The fill continues until the STOP FILL sensor sends
the controller a signal that it is now in liquid. The controller will close the liquid
supply valve to stop the fill. As liquid evaporates, the display will indicate the
liquid is at a normal level as the cycle begins again.
Control Setting Adjustments
Level
The standard sensor assembly that is installed on a freezer consists of a circuit
board with thermistors installed. The assembly has a maximum range of 6 inches
(150mm). Thermistor assignments can be changed through the menu system and
the operating range can be changed by either raising or lowering the circuit board
(offset) within the freezer.
If a wider range is needed between the start fill and stop fill thermistor, please
contact your supplier for an alternate sensor assembly.
Features & Settings
Some control settings can be changed through the menu system.
Enter and Exit the menu system by pressing:
Move down through the menu system by pressing:
CS SERIES
Move up through the menu system by pressing:
Select a menu choice or lock in a setting by pressing:
Back out of the menu system by pressing:
When changing settings, single button presses will increment/decrement a value
one step at a time. Pressing and holding a button will allow for rapid change of a
setting.
The menu system incorporates icons to show the user settings which can be changed
and which are locked.
Locked menu choice:
Accessible menu choice:
Also, the menu system includes a scroll bar on the right side of the display. The
scroll bar will indicate the current location in the menu system and will indicate if
additional menu choices are available with the display of up and down arrows on
the scroll bar.
The control will continue to monitor all sensors and conditions while the user
accesses the menu system. If no activity is detected for 30 seconds, the control will
automatically return to the main operational screen.
17
CS SERIES
View Only
The View Only menu selection allows the user to view settings but will not allow
changes to any settings. This restriction is designated with a lock symbol. Refer to
Figure 8.1 Menu System - View Only on page 22 to navigate the “View Only” menu.
Change Settings
The Change Settings menu selection allows the user to change operational settings
for the control system. Refer to Figure 8.2 Menu System - Change Settings on page
23 to navigate through any changes you need to make.
Below are the Menu Settings that are available in both the “View
Only” and “Change Settings” modes.
•TEMPERATURE
– Thermocouple Select: Allows the user to turn on/off thermocouples for
temperature measurement.
– Calibrate
•Calibrate Thermocouple 1: Allows the user to calibrate the control for use
with thermocouple #1.
•Calibrate Thermocouple 2: Allows the user to calibrate the control for use
with thermocouple #2.
•Restore Calibration: Resets the calibration to the data stored during factory
calibration.
•Set Ambient: Ambient temperature is set at the factory and should not be
changed in the field.
– Test Temperature System
•Check Thermocouples: Gives status of temperature sensors.
•Test Temp Alarms: Allows the user to manually test temperature alarms for
the thermocouples.
– Control by Temperature
•Temp Control On/Off: Enables/Disables temperature control.
•Temp Control Range: Allows the user to set the range for temperature control.
– Temperature Alarms
•High Temperature Alarm: Allows the user to set the high temperature
alarm for both thermocouples. Settings [00° C to -1900° C]
•Low Temperature Alarm: Allows the user to set the low temperature alarm
for both thermocouples. Settings [00° C to -1900° C]
•Time with no Temp alarm: Displays the time since the last high or low
temperature alarm.
18
– Temperature Units: Allows the user to choose the units of measure for
temperature.
•LEVEL
– Thermistor Status: Gives the user the status for the 8 thermistors on the level
sensor assembly and the freezeguard sensor installed at the valve.
– Sensor Positions: Allows the user to assign the thermistors for low level alarm,
start fill, stop fill, high level alarm and the sensor offset.
– Valve Open Duration: Allows the user to set the fixed times for valve open
and valve closed durations.
– Sensor Type: Allows the user to set the sensor type as a 4 thermistor sensor or
an 8 thermistor sensor. Also allows the user to disable the freezeguard sensor.
– Level Sensor Alarm
•Sensor Error Mute: Allows the user to mute the audible alarm for a sensor
error.
•Freezeguard Error Mute: Allows the user to mute the audible alarm for a
freezeguard sensor error.
•SYSTEM
– Inch/Metric: Allow the user to switch between English and Metric level
measurement systems.
– Date/Time:
•Set Date/Time: Allows the user to set the date and the time. Use the up/
down arrows to change settings. Use the left/right arrows to move between the
date and time fields.
CS SERIES
Figure 6.0 Set Date/Time Menu
•Date Format: Allows the user to change between U.S. and International format.
– Lid Functions
•Lid Switch: Permits the user to enable/disable the lid switch feature.
•Quickchill Timer: Enables the user to set the amount of time that the valve
will stay open after the lid is closed.
•Manual Defog Timer: Allows the user to set the amount of time that the
valve stays open when the Fill/Defog button is pressed and the LN2 level is
within normal range.
•Auto Defog Timer: Allows the user to set the amount of time that the valve
stays open. Triggered by the lid switch when the lid is opened and the LN2 level
is within normal range.
•User Access On/Off: Allows the user to enable/disable this feature.
– User ID’s: Allows the user to set user identification for freezer access.
•Lid Open Valve Off: Allows the user to enable this feature which insures that
solenoid is closed when the lid is opened.
– System Alarms
•Test Alarm: Allows the user to manually check the audible, visual and remote alarm.
•LN2 Supply Alarm Delay: Allows the user to set the time to wait before
an error condition occurs if the freezer does not fill in a timely manner. If this
alarm occurs the supply should be checked to insure that the supply valve on
turned on, a sufficient supply of LN2 (at least 10 psi) is available in the supply
cylinder. Settings [30, 45, 60, 75, 90 min] or [2, 3, 4 hours]
19
CS SERIES
•Remote Alarm Delay: Allows the user to set the time before the remote
alarm relay is triggered after an error condition occurs. Settings [Immediate,
30, 60, 90 min] or [2,3, 4, 6 hours]
•Lid Open Alarm Delay: Allow the user to set the time before an alarm
condition occurs if the lid is opened too long.
•Valve Stuck Alarm: Allow the user to enable/disable this feature.
•Audible Re-trigger: Allows the user to set the time before the audible alarm
is re-triggered after an alarm has been acknowledged. Settings [15, 30, 60,
120, 240, 480 min] or [1 day]
– Display
•Display Brightness: Allows the user to adjust brightness for the display.
Settings [1 to 15]
•Screen Saver Time: Allows the user to adjust the time before the screen
saver starts working. Settings [0 to 60 min]
– Freezer ID: Allows the user the capability to change the freezer identification.
•SCHEDULED EVENTS
– Scheduled Fills
•Set Time: Allows the user to set the day and time for the next filling operation.
•Next Fill: Displays the next scheduled fill.
– LN2 Supply Reminder
•Set Reminder: Allows the user to set the day which the control will display a
reminder to check the supply of LN2.
•Clear Reminder: Allows the user to acknowledge that the LN2 supply has
been checked.
•Next Reminder: Displays the next scheduled reminder.
Below are the Menu Settings that are available only in the “Change
Settings” mode.
•SECURITY
– Settings Password: Allows the user to set a password to allow setting changes.
– Power Password: Allows the user to set the password which secures power
button operation.
•CONTACT SUPPORT
– Distributor: Contact information for the distributor.
– Taylor-Wharton: Contact information for Taylor-Wharton.
•SYSTEM INFORMATION
– Displays the control serial number, the firmware version and the freezer
identification.
20
Temperature
The temperature in the Cryostorage freezer is measured at the location of the
thermocouple. The installation location is different depending on the model of
the freezer but usually the thermocouple is positioned level with the top storage
box. This may mean that temperatures displayed may be slightly warmer than the
temperature experienced by the samples or product stored in the freezer.
Temperature Control
The CS SERIES Control System has the capability to control the vapor temperature
at the top of the freezer. The temperature is controlled at the location of
thermocouple 1.
To maintain temperature, the CS200 SERIES Control System bubbles N2 gas
through the pool of LN2 in the bottom of the freezer. The nitrogen gas evaporates
some of the LN2 causing a cooling effect reducing the temperature in the freezer.
Since a pool of LN2 is important in the process of temperature control, an adequate
level is always maintained by the control systems and the level control always takes
precedence over temperature control. Temperature control will not work if the LN2
level is above High Level Alarm.
Enabling the temperature control feature will always increase the use of LN2.
The increased consumption will depend on a number of factors including the
temperature to be maintained, the range of the controlled temperature, the ambient
temperature and the length of supply hose connected to the freezer and the
frequency which users open the lid or otherwise introduce heat.
Battery Operation
The CS200 SERIES Control System is designed to operate with or without an optional
battery backup system if required. Simply plug the battery into the appropriate
connector. The control will recognize that the battery has been connected and will
activate the battery backup mode.
CS SERIES
LED Status Wheel
The CS SERIES Control System offers an innovative concept called a status wheel. In
normal operation, the LED’s on the status wheel light to show a slow rotation of the
wheel. If an error condition occurs, the control will alert the user with an additional
visual alarm from the status wheel. The rotation can be accelerated, flashed or even
illuminated in the opposite direction.
ConditionLED Flash Property
NormalClockwise pattern, 1 LED per second
General ErrorClockwise pattern, rapid rotation, flash
Figure 7.0 LED Status Wheel Flash Patterns
Lid Switch
The lid switch consists of a magnet and a pickup installed on the lid. The control can
determine when the lid is opened because the magnet moves out of range of the
pickup and the circuit becomes open. A number of features are associated with the
lid switch such as Auto Defog, Quickchill and the Lid Open Too Long Alarm. If the
lid switch is deactivated, these features are also disabled.
21
CS SERIES
MENU
VIEW ONLY
TEMPERATURE
LEVEL
SYSTEM
SCHEDULED EVENTS
THERMOCOUPLE SELECT
TEST TEMPERATURE SYSTEM
CONTROL BY TEMPERATURE
TEMPERATURE ALARMS
TEMPERATURE UNITS
THERMISTOR STATUS
SENSOR POSITIONS
VALVE OPEN DURATION
SENSOR TYPE
LEVEL SENSOR ALARM
INCH/METRIC
DATE/TIME
LID FUNCTIONS
SYSTEM ALARMS
DISPLAY
FREEZER ID
SCHEDULED FILLS
LN2 SUPPLY REMINDER
T/C 1 ON/OFF T/C ON/OFF
CHECK THERMOCOUPLES
VALIDATE CALIBRATION
TEST TEMP ALARMS
TEMP CONTROL ON/OFF
TEMP CONTROL RANGE
HIGH TEMPERATURE ALARM
LOW TEMPERATURE ALARM
TIME WITH NO TEMP ALARM
CELCIUS FAHRENHEIT
KELVIN RANKINE
123456789 LID SWITCH
123456789 0
02 03 04 05
SENSOR OFFSET 0 INCH
OPEN 0 CLOSE 0
NUMBER OF THERMISTORS
FREEZEGUARD
SENSOR ERROR MUTE
FREEZEGUARD ERRROR MUTE
INCH METRIC
SET DATE/TIME
DATE FORMAT
LID SWITCH
QUICKCHILL TIMER
MANUAL DEFOG TIMER
AUTO DEFOG TIMER
USER ACCESS
LID OPEN VALVE OFF
TEST ALARMS
LN2 SUPPLY ALARM DELAY
REMOTE ALARM DELAY
LID OPEN ALARM DELAY
VALVE STUCK ALARM
AUDIBLE RE-TRIGGER
DISPLAY BRIGHTNESS
SCREEN SAVER TIME
FREEZER 1
SET TIME
NEXT FILL
SET REMINDER
CLEAR REMINDER
NEXT REMINDER
T/C 1 OK T/C
2 OFF
OFF
OFF ON
-150C -165C
THERMOCOUPLE 1
THERMOCOUPLE 2
THERMOCOUPLE 1
THERMOCOUPLE 2
TEMP 1: 0
4 8
OFF ON
ON OFF
ON OFF
03 -03 -12 15 06
U.S. INTERNATIONAL US
MM/DD/YY INT: DD/MM/YY
OFF ON
20
20
20
USER ACCESS ON/OFF
USER IDs
USER ID ON/OFF
OFF ON
AUDIBLE
VISUAL
REMOTE
30 45 60 75 90 MIN 2 3
4 HOURS
IMMEDIATE 30 60 90 MIN 2
3 4 5 HOURS
10 MIN
OFF ON
15 30 60 120 240 480 MIN 1
DAY
0
20
SET DAY SET TIME
DAILY 17:00
03-12-12 14:37
OFF SUN MON TUES WED
THRS FRI SAT
YES
3/3/2012
-100
-100
-180
-180
OFF ON
USER 1
USER 2
USER 3
OFF ON
OFF ON
OFF ON
OFF ON
22
Figure 8.1 Menu System - View Only
CS SERIES
CHANGE SETTINGS
CONTACT SUPPORT
SYSTEM INFORMATION
TEMPERATURE
LEVEL
SYSTEM
SCHEDULED EVENTS
SECURITY
DISTRIBUTOR
PAGER DIGITAL
TAYLOR-WHARTON
SERIAL # 12345
VERSION 1.00
FREEZER ID FRZ01
THERMOCOUPLE SELECT
CALIBRATE
TEST TEMPERATURE SYSTEM
CONTROL BY TEMPERATURE
TEMPERATURE ALARMS
TEMPERATURE UNITS
THERMISTOR STATUS
SENSOR POSITIONS
VALVE OPEN DURATION
SENSOR TYPE
LEVEL SENSOR ALARM
INCH/METRIC
DATE/TIME
LID FUNCTIONS
SYSTEM ALARMS
DISPLAY
FREEZER ID
SCHEDULED FILLS
LN2 SUPPLY REMINDER
SETTINGS PASSWORD
POWER PASSWORD
CONTACT INFO
CONTACT INFO
CONTACT INFO
T/C 1 ON/OFF T/C ON/OFF
CALIBRATE THERMOCOUPLE 1
CALIBRATE THERMOCOUPLE 2
RESTORE CALIBRATION
SET AMOUNT
CHECK THERMOCOUPLES
VALIDATE CALIBRATION
TEST TEMP ALARMS
TEMP CONTROL ON/OFF
TEMP CONTROL RANGE
HIGH TEMPERATURE ALARM
LOW TEMPERATURE ALARM
TIME WITH NO TEMP ALARM
CELCIUS FAHRENHEIT
KELVIN RANKINE
123456789 LID SWITCH
123456789 0
02 03 04 05
SENSOR OFFSET 0 INCH
OPEN 0 CLOSE 0
NUMBER OF THERMISTORS
FREEZEGUARD
SENSOR ERROR MUTE
FREEZEGUARD ERROR MUTE
INCH METRIC
SET DATE/TIME
DATE FORMAT
LID SWITCH
QUICKCHILL TIMER
MANUAL DEFOG TIMER
AUTO DEFOG TIMER
USER ACCESS
LID OPEN VALVUE OFF
TEST ALARMS
LN2 SUPPLY ALARM DELAY
REMOTE ALARM DELAY
LID OPEN ALARM DELAY
VALVE STUCK ALARM
AUDIBLE RE-TRIGGER
DISPLAY BRIGHTNESS
SCREEN SAVER TIME
FREEZER ID #1
SET TIME
NEXT FILL
SET REMINDER
CLEAR REMINDER
NEXT REMINDER
0 0 0 0
0 0 0 0
PUT T/C 1 IN ICE WATER (0C)
PUT T/C 1 IN LN2 (-196C)
DONE
PUT T/C 2 IN ICE WATER (0C)
PUT T/C 2 IN LN2 (-196C)
DONE
ABORT LOAD
35.3
T/C 1 OK T/C 2
OFF
OFF
OFF ON
-150C -155C
THERMOCOUPLE 1
THERMOCOUPLE 2
THERMOCOUPLE 1
THERMOCOUPLE 2
TEMP1: 0
4 0
OFF ON
ON OFF
ON OFF
00 -00 -12 15:00
US INTERNATIONAL
US: MM/DD/YY INTL: DD/MM/YY
OFF ON
30
30
33
USER ACCESS ON/OFF
USER IDs
USER ID ON/OFF
OFF ON
AUDIBLE
VISUAL
REMOTE
30 45 60 75 90 MIN 2
3 4 HOURS
IMMEDIATE 30 60 90 MIN 2
3 4 5 HOURS
10 MIN
OFF ON
15 30 60 120 240 480 MIN 1
DAY
0
30
SET DAY SET TIME
DAILY 17:00
03-13-12 14:37
OFF SUN MON TUES WED
THURS FRI SA
YES
10-13-12T
-100
-100
-130
-130
OFF ON
USER 1
USER 2
USER 3
OFF ON
OFF ON
OFF ON
OFF ON
Figure 8.2 Menu System - Change Settings
23
CS SERIES
Factory Defaults
Thermocouple #1 On
Thermocouple #2 Off
Control by Temperature Off
High Temperature Alarm #1 -100C
High Temperature Alarm #2 -100C
Low Temperature Alarm #1 Off
Low Temperature Alarm #2 Off
Temperature Units Celsius
Sensor Position Freezer dependant
Valve Open Duration Off
Sensor Type 8 thermistor sensor
Freezeguard feature On
Sensor Error Mute Off
Freezeguard Error Mute Off
Level Units of Measure inch
Date/Time Factory Set for Eastern Standard Time
Lid Switch On
Quickchill Timer 30 seconds
Manual Defog Timer 30 seconds
Auto Defog Timer 30 seconds
User Access Off
Lid Open-Valve Off Off
LN2 Supply Alarm Delay 30 minutes
Remote Alarm Delay 30 minutes
Lid Open Alarm Delay 10 minutes
Valve Stuck Alarm On
Audible Retrigger 30 minutes
Display Brightness 10
Screen Saver Time 30 minutes
Freezer ID Not set
Scheduled Fills Off
LN2 Supply Reminder Off
Settings Password Off
Power Password Off
24
Troubleshooting
If the Taylor-Wharton Cryostorage freezer with the CS200 SERIES Control System
installed experiences problems or appears that it is not operating at optimum
efficiency, please contact your distributor for assistance. The CS SERIES Control
System has incorporated state of the art diagnostic tools to assist in the identification
and correction of any issues that may arise.
Interconnection Block Diagram
CS SERIES
Figure 9.0 Interconnection Block Diagram
25
CS SERIES
4 wiresposition 1P1P1
4 wiresposition 2P1P2
10 wiresposition 1P2P1
10 wiresposition 2P2P2
BlueP1P1Valve
RedP1P1
Yellow + White/ YellowP2P1Remote A larm
Orange + Whit e/OrangeP2P1Remote Alarm
Brown + White/B rownP2P 1Remote Alarm
Yellow/GreenP1P1Power
WhiteP1P1Power
Harness Control End
White/Blac kP2P1
Red + BlackP2P1RS-232
White/RedP2P1
BlackP 2P2
BrownP2P2
RedP2P2Sens or
OrangeP2P2
YellowP2P2
P2P2
whitePower Connector
Yellow/Green
Valve Connector
Red
Blue
White/Bla
ckP2P1DB-9 Connector
Connector Box EndWhite/RedP2P1
Red + BlackP2P1
White/Yellow + YellowP2P1
White/Orange + OrangeP2P1Remote Alarm Connector
White/Brown + B rownP2P 1
BlackP2P2
BrownP 2P2
RedP2P2Sensor Assembly
OrangeP2P2
YellowP2P2
Pod 1
Pod 2
2
1
5
4
3
2
1
3
2
1
6
5
4
3
2
1
P1P1
P1P1
P1P1
P1P1
2
3
5
1
2
3
1
2
3
4
5
Wiring Diagram
Figure 10.0 CS Series Control System Harness Wiring Diagram
26
Installation & Setup
Main Harness Connector
Connection to Main Control
The CS SERIES Control System consists of the following components.
• Main Control
• VFD Display Module
• Wiring Harness Assembly
• Power Supply
• Thermocouple Assembly
• Lid Switch Assembly
• Sensor Assembly
• Cryogenic Solenoid Valve
• Remote Alarm Plug
CS SERIES
• LN
2
• Plumbing
Connect the wiring harness assembly to the main control. The connector is keyed
and can only be plugged in one way.
Display ConnectorT/C #1T/C #2
Figure 11.0 Main Control Back Panel
Connect the thermocouple plug of the harness assembly into the control at the
thermocouple plug labeled T/C #1. One of the blades on the plug is slightly wider
insuring that it is plugged in correctly. The copper blade should plug into copper
colored plug. Finally, connect the display cable into the main control and the display.
Figure 12.0 Connection to Main Control
27
CS SERIES
Connector to Main ControlConnector for Level Sensor
Valve and Freeze Guard
Connect the level sensor assembly into the pigtail with the round connector
terminating at the control end of the harness assembly. The connector is a round
locking connector that is keyed so it can only be plugged in one way.
Connect the solenoid valve and the lid switch into the appropriate connectors on
the connector box. These are located on the circuit board inside the box and are
labeled. The solenoid connector is a 6 pin locking connector and the lid switch is a
2 pin locking connector. Each is keyed so they can only be plugged in one way.
Cryowire PortPower
Battery
Lid Switch Connector
28
Cryowire Port
Harness Connector
to Main Control
Remote Alarm
Figure 14.0 Harness Assembly & Back Panel
Sensor Connector
Connect the Remote Alarm plug into the panel at the end of the wiring harness
Remote Alarm ConnectorCryowire Dual Communication Ports
Power ConnectorBattery ConnectorCryowire Communication Port
assembly. This is keyed so that it can only be plugged in one way.
Figure 15.0
Finally, connect the barrel plug of the power supply and the battery into the
appropriate receptacles in the connector box.
CS SERIES
Figure 16.0
Diagnostic Menu
Need diagnostic menu
External Connector Ratings
Designated UseMax rated voltage/current ratingsConnector type
General ErrorClockwise pattern, rapid rotation, flash
ConditionLED Flash Pattery
29
CS SERIES
Temperature Thermocouple Select
The chamber temperature is monitored with 1 or 2 Type T thermocouples. The
thermocouple is placed in the chamber to monitor temperature level at the top
of the inventory system. Factory installation includes one thermocouple inside of
the LABS thermocouple tube or inside of the thermocouple tube at an elevation to
match the height of standard racks. A second Type T thermocouple may be added to
monitor another location inside the chamber. Both thermocouples can be activated/
deactivated through the menu system.
Temperature Calibration
The CS SERIES Control System uses a type T thermocouple to measure temperature
within the CryoStorage freezer. The temperature curve for a thermocouple is
nonlinear so it is important that the CS SERIES Control System have a good
calibration to provide accurate temperature readings. The temperature is traceable
to the National Institute of Standards and Technology (NIST) ITS-90 Thermocouple
Database.
There are three important reference points needed for calibration:
Ambient Temperature:
Ice Water: 0°C
Liquid Nitrogen (LN2): -196°C
The ambient temperature is the temperature measured inside the control box and
is used to provide temperature compensation adjustment. This is calibrated at the
factory and should not be adjusted in the field.
Ice water and LN2 provide the reference points on the temperature curve. If these
two points are calibrated correctly then all other points (temperatures) on the curve
are correct.
The accuracy of a thermocouple is +/- 1°C or +/- 1.5% of the reading, whichever is
greater.
To check a calibration, submerge the thermocouple in ice water slurry and then LN2.
If readings are within accuracy specifications noted above, then the temperature
circuit on the control is properly calibrated. If not, follow the steps below to calibrate
the temperature.
30
Table 4.0 Temperature Calibration Instruction
Access the menu by pressing
Highlight CHANGE SETTINGS and press
Highlight TEMPERATURE and press
Highlight CALIBRATE and press
Highlight CALIBRATE THERMOCOUPLE 1. The display will read Dip
the thermocouple into ice water.
Submerge the thermocouple into an ice water bath. The bar graph
will fluctuate. When a good stable reading in obtained, the bar
graph will decrease to 1 or 2 bars. At this point, press to lock in the
value
The display will read Dip the thermocouple into LN2.
Submerge the thermocouple into LN2. The bar graph will fluctuate.
When a good stable reading in obtained, the bar graph will
decrease to 1 or 2 bars. At this point, press to lock in the value.
If the calibration values fall within the expected range, DONE! Will
be displayed on the screen.
CS SERIES
Test Temperature System
The temperature circuitry can be checked at any time through the menu system.
This check will tell if the thermocouples are working or if they are “open” (broken
or unplugged). If a thermocouple is not connected to the control it will check as
“open.” If a thermocouple is “Disabled” through the menu system, it will not show
up on the check.
Test Level Sensors
The sensor assembly can be tested through the menu system. The sensor diagnostics
indicates the sensor number and the status (whether in liquid or gas). If the control
is set for an eight-thermistor it will indicate 9 sensors in the diagnostics. If it is set
for a four-sensor assembly, it will indicate 4 sensors. The status is indicated with
either an “O” for open, a “G” for gas or an “L” for liquid. This is an easy means to
tell if sensors are in or out of liquid or if a new sensor assembly is needed (open or
defective sensors).
The sensor type can be set through the menu system. The sensor type selection
should match the sensor type that is being used in the system. This is a 4-sensor, a
8-sensor (Freeze-Guard) or an 8-sensor array. If the sensor assembly is unplugged
and the main control is still on, the display will indicate that a sensor error has
occurred. In addition the level indicated would be 8” on an 8-sensor assembly,
or “High Alarm” on a 4-sensor assembly. This occurs because the control cannot
differentiate between a very high resistance (when a thermistor is in LN2) and an
infinite resistance (when an open circuit appears in the level sensing circuitry).
31
CS SERIES
Alarms and Error Conditions
The CS SERIES Control System control tracks many different conditions in the freezer
and therefore, has a full complement of alarms associated with these different
conditions. As alarms occur, they cause an audible beep as well as a flashing red
light on the same wheel. A remote alarm relay is also triggered following a userdesignated period of time, after the error condition occurs, if it is not corrected. In
addition, the error condition is displayed on the top line until the error condition is
corrected. When an error does occur, the audible alarm may be muted by pressing
the designated button. The audible alarm will then be silent until activated by a new
error condition. A red light will continue to flash until all errors are corrected. The
remote alarm will be activated if the power is interrupted.
The High Temperature Alarm for Thermocouple #1 can be set through the menu
system. This alarm is activated if the temperature rises above the designated
temperature. The alarm temperature can range from 0ºC to -190ºC. It can also be
disabled.
High Temperature Alarm for Thermocouple #2 can be set through the menu system.
This alarm is activated if the temperature rises above the designated temperature.
The alarm temperature can range from 0ºC to -190ºC and it can also be disabled.
System Alarms
A Low LN2 Supply Alarm can be set through the menu system. This alarm is activated
if the solenoid valve is not closed within a designated time period after a fill starts.
The solenoid valve can be closed either automatically (the LN2 level reaches the
STOP FILL sensor) or manually (the stop fill button is pressed). To stop the timer
which activates this alarm. The possible choices for this alarm are None, 15, 30, 45,
60 minutes, 2 or 3 hours. This alarm does not correct itself until the fill is stopped
(the solenoid closes).
The Sensor Error Alarm can be set through the menu system. This alarm is activated
if the control detects a sensor error such as an open sensor. The possible choices
are ENABLE or DISABLE. An open sensor can be confirmed through the TEST LEVEL
SENSORS option in the menu system.
A Remote Alarm Timer can be set through the menu system. This is the amount of
time allowed to pass before the remote alarm relay is triggered if an error condition
is not corrected. The possible choices are None, Immediate, 30 minutes, 60 minutes
or 2 hours.
Lid Open Too Long Alarm can be set through the menu system. This is the amount of
time the lid can be open before it triggers an alarm condition. The possible choices
are None, 1, 2, 5 or 10 minutes.
32
Thermocouple Alarm can be set through the menu system. This alarm is activated
if either thermocouple experiences an open circuit. Your choices are ENABLE or
DISABLE.
Test Alarms
Audible, visual and remote alarms can be tested at any time by the user through the
menu system. Follow the instructions on the display to hear the audible “chirping”
indicator of an alarm or to see the red status wheel flash or to trigger an immediate
relay closure of the remote alarm.
Logging
On board memory logging function is one of the most powerful and useful features
of the CS SERIES Control System control. It provides a historical record for not only
your freezer but also a complete record of the environment in which specimens were
stored. Four separate logs are kept in the control:
1. System log – System logs are events that occur in the system including lid
opening/closing, LN2 filling, Quick-Chill, Defog.
2. Error log – Error logs are outside the activated preset limits detected by the
system.
3. Temperature #1 log and Temperature #2 log (for use with optional thermocouple
#2) – The two temperature logs are simply records of the temperatures recorded by
the two thermocouples in the system.
The system and the error log each hold 4096 events while the combined
temperature logs hold an additional 32,768 temperature events. All the logs are
kept in non-volatile memory, meaning that the information is saved regardless of
whether the control has power.
When an event (system, error or temperature) occurs, the control does two things
with the data:
CS SERIES
1. It enters the beginning or the conclusion of an event in the internal memory of
the controller.
2. It sends the event data out the serial port of the freezer.
The control consolidates the events in the internal log (combining “start event” /
”conclude event” information to provide one event with duration), however, when
the data is sent out the serial port no consolidation of data occurs.
As an example, a fill would provide one log entry in the internal log of the control,
indicated as follows: Fill occurred on 01/07/2012 @ 8:07 for 24 minutes. The
same data coming out of the serial port would cause two entries in a computer or
printer and would be indicated as follows:
Fill Started on 01/07/12 @ 8:07
… (elapsed time)
Fill occurred on 01/07/12 @ 8:07 for 24 minutes
Menu Access causes the control to make some decisions on logging an event and
they are handled in the following manner: Temperature is immediately logged (if it
is enabled) and then a fresh time period is started when the Menu system is exited.
System and error logs are placed in a suspended state until the control exits the
menu system and timing is started again. Duration of system and error logs then are
total time of the event less any time that the user was in the menu system.
When logs are dumped to the serial port, the oldest events are sent first. The control
operates on the FIFO (First In First Out) method. If the maximum number of entries
is reached, the oldest are lost to make way for the newest entries.
Dump Logs
DUMP SYSTEM LOGS is accessible through the menu system of the control. This
option sends data from the system logs out the serial port of the freezer. When
this option is chosen, the display reports how many system logs are stored. While
the data is being sent to the serial port, it can be paused or completely cancelled
through the menu system.
33
CS SERIES
DUMP ERROR LOGS is accessible through the menu system of the control. This
option sends data from the error logs to the serial port of the freezer. When this
option is chosen, the display reports how many error logs are in the system. While
the data is being sent out the serial port, it can be paused or completely cancelled
through the menu system.
DUMP TEMP LOG #1 and TEMP LOG #2 are accessible through the menu system
of the control. This option sends data from the temperature logs to the serial port of
the freezer. When this option is chosen, the display reports how many temperature
logs are in the system. While the data is being sent out the serial port, it can be
paused or completely cancelled through the menu system.
Error Logs
SENS. ERR. LOGGING is accessible through the menu system of the control. This
menu choice turns on/off the logging of all sensor errors. The choices are ENABLE
or DISABLE. Records data in the error log.
LOW LN2 SUPPLY LOGGING is accessible through the menu system of the control.
This menu choice turns on/off the logging of the low LN2 supply error. The choices
are ENABLE or DISABLE. Records data in the error log.
REMOTE ALARM LOGGING is accessible through the menu system of the control.
The menu choice turns on/off the logging of the remote alarm activation. The
choices are ENABLE OR DISABLE. Records data in the error log.
OPEN THERMOCOUPLE LOGGING is accessible through the menu system of the
control. This menu choice turns on/off the logging of the thermocouple open alarm.
The choices are ENABLE and DISABLE. Records data in the error log.
HIGH TEMP #1 LOG is accessible through the menu system of the control.
This menu choice turns on/off the logging of the high temperature alarm for
Thermocouple #1. The choices are ENABLE and DISABLE. Records data in the error
log.
HIGH TEMP #2 LOG is accessible through the menu system of the control. This menu
choice turns on/off the logging of the high temperature alarm for Thermocouple
#2. The choices are ENABLE and DISABLE. Records data in the error log.
System Logs
FILL LOGGING is accessible through the menu system of the control. This menu
choice turns on/off the logging of tank filling operations. The choices are ENABLE or
DISABLE. Records data in the system log.
LID ACTION LOGGING is accessible through the menu system of the control. This
menu choice turns on/off the logging of lid openings and closings. The choices are
ENABLE or DISABLE. Records data in the system log.
USER ACCESS LOGGING is accessible through the menu system of the control. This
menu choice turns on/off the logging of user access codes, which are requested
when the lid is opened. Records data in the system log.
34
Temperature Logs
Temperature Logging Rates for thermocouple #1 (T/C #1 LOGGING) and
thermocouple #2 (T/C #2 LOGGING) are accessible through the menu system of
the control. This menu choice adjusts the rate at which temperatures are logged for
the two thermocouples. The possible choices are Disabled, 15, 30 minutes, 1, 2, 4,
6, 12 or 24 hours. Records data in the temperature logs.
Erase Logs
ERASE LOGS is accessible through the menu system of the control. This menu choice
erases any of the four logs found in the control. Please note that once a log has
been erased, it is gone forever.
ERASE SYSTEM LOGS).
ERASE ERROR LOGS.
ERASE TEMPERATURE LOG #1.
ERASE TEMPERATURE LOG #2.
Display Brightness
DISPLAY BRIGHTNESS can be set through the menu system of the control. This
option changes the intensity of the display. The possible choices are 25%, 50%, 75%
and 100%.
Making Adjustments to the CS SERIES Control System Sensor
Assembly
Table 5.0: CS SERIES Control System Sensor Assembly Factory Settings
CS SERIES
CRYOSTORAGE
SYSTEM
LABS
20K, 38K, 40K, 80K
LABS 94K5 in. (127mm)6 in. (152mm)9 in. (228mm)
LOW LEVEL
ALARM
2 in. (50mm)3 in. (76mm)5 in. (127mm)
START
FILL
STOP
FILL
(at carousel)
(at carousel)
HIGH LEVEL
ALARM
6 in. (152mm)
10 in. (254mm)
The sensor assembly is preset at the factory for vapor phase storage. If adjustments
need to be made, the following procedure will simplify the process.
The CS SERIES Control System control installed on the Taylor-Wharton Cryostorage
units operates with specially designed software to match the design characteristics of
your refrigerator. Refer to the chart below to see the versions and their difference.
To make adjustments to a sensor assembly in a refrigerator filled with LN2, the
following procedure can be used:
• Measure the LN2 liquid level in the refrigerator.
• Take this measured level and subtract the offset to determine how many sensors
should be in liquid.
a. Remove sensors.
b. Fill to correct level and turn off Liquid supply.
c. Eg. Level is to be 15 in. (381 mm) and sensor number 6 is currently and will
remain the Stop Fill. 15 in. (381 mm) – 6 in. (152.4 mm) = 9 in. (238.6
mm) offset. Set Offset to 9 in (228mm).
• Go to “Test Level Sensor” through the CS SERIES Control System menu. “L”
means a sensor is in liquid while “G” means a sensor is in gas.
• Move the sensor up or down so that the appropriate numbers of sensors are in
liquid (read “L”).
• Return to the CS SERIES Control System main screen and the level indicated
should match the physically measured reading.
The Sensor Offset, the START FILL and the STOP FILL can all be set through the CS
SERIES Control System menu system.
35
CS SERIES
Removing/Installing the Solenoid Valve
The LN2 and power must be turned off before beginning work on the solenoid valve.
For LABS Units
Disconnect only the solenoid valve lead connection from the back of controller
board.
To remove the solenoid valve, loosen the compression fitting that connects the
plumbing tubing to fill tube. Unscrew the two (2) mounting screws that hold the
solenoid valve to the solenoid bracket. Then remove the solenoid valve and its
associated plumbing. Disconnect the plumbing from the inlet and outlet side of the
solenoid valve.
To install a new solenoid valve, attach the connecting plumbing to the inlet and
outlet connections of the valve using Teflon tape on tapered threads (3 wraps).
Attach the compression fitting to the fill tube first and then connect the compression
fitting to the elbow that is connected to the outlet side of the solenoid valve. Position
the solenoid valve onto the solenoid valve bracket and tighten the two (2) mounting
screws. Attach the solenoid valve lead connection to the controller board.
Controller Electrical Tests
If a controller is removed from the refrigerator for service, the liquid refrigerant
level must be maintained manually to protect stored product. The fill solenoid valve
will be inoperative with the controller removed. A flexible fill line terminated with
a phase separator may be used periodically through the open refrigerator lid to
conduct manual fill operations until automatic operation is restored.
Sensor Probes. The sensor probes used to detect liquid level by Taylor-Wharton
controllers are temperature-sensitive resistors called “thermistors.” Their resistance
to electrical current flow changes greatly with their temperature.
• At Room Temperature ....2 ohms to 10 ohms
• In Cold Nitrogen Gas ....8 K ohms to 18 K ohms
• In Liquid Nitrogen .........18 K ohms to 35 K ohms
Controller Logic. The liquid level controllers read the values of the sensors as
indications of liquid level. The function of a sensor, and its value, are interpreted by
the controller to set normal operating, fill, fill termination, and alarm conditions. In
addition, the refrigerator temperature is monitored and an alarm is triggered if the
temperature raises above a pre-determined point.
• From cold gas to liquid ..Resistance greater than 22K ohms.
• From liquid to cold gas ..Resistance less than 16K ohms.
36
• Defective Sensor ............ Resistance greater than 50K ohms, or less that 5 ohms.
Plumbing Assembly
Freeze-Guard
Sensor
CS SERIES
Solenoid Valve
Male Branch
Tee
90 deg Elbow
90 deg Elbow
Male Branch
Tee
Reducer
Strainer
Female Connector
Relief Valve
Figure 17.0 LABS Plumbing Assembly
Relief Valve Tube
TROUBLESHOOTING
Symptoms
The key to troubleshooting your CS SERIES Control System and your cryostorage
system is to determine which component in the system is the source of the problem.
Determine if the problem is occurring in any of the following subsystems: Supply
Vessel, Transfer Line, Power Source, Temperature, Level Sensing, Security, Lid Switch,
Solenoid Valve, Control Display, Alarm System, Communications. After determining
which sub-system is having the problem, isolate the problem further by performing
sub-system tests. Once the problem is isolated and defined, it will be easier to solve.
Controller Will Not Turn ON
1. Press POWER button. If display is blank and dark go to next step.
2. Check all connections. Start with the connections to the controller. Pay special
attention to jacks labeled “POWER”. Power cord must be plugged into an outlet
providing AC voltage between 100 and 130 to deliver AC voltage 21 to 30 to
the back of the freezer. Or the 100-240 50/60Hz European standard.
37
CS SERIES
Indicates High Liquid Level
1. Determine actual liquid level using a dipstick. Select MENU, LEVEL SENSING,
TEST LEVEL SENSORS. An “L” or “G” will indicate individual thermistor status. “L”
meaning that the thermistor is submerged in liquid and “G” indicates that the
thermistor is in cold. A fill solenoid valve that freezes (or sticks) open will typically
empty the supply cylinder. Replace solenoid valve if it has failed even once.
2. Liquid level is sensed by thermistors located in a sensor tube. If the sensor tube
is blocked or iced at the top, the liquid level in the sensor tube may not rise and
fall at the same rate as the liquid level in the freezing chamber. Make sure the
sensor tube is not obstructed.
3. The pool of LN2 can become turbulent during a fill. Bubbling and splashing can
be amplified by the rack arrangement. The deeper the pool, the more turbulent
the surface of the pool will be during a fill. The turbulence of the pool surface
can splash on the X-High thermistor and cause a false high alarm. Rearrange
the racks to reduce the turbulence and splashing.
4. Confirm that sensor assembly is responding to changing liquid level with a diptest.
• Close liquid supply valve at source.
• Mark the sensor assembly at top of tube to assure back to the original
position.
• Remove sensor assembly from sensor tube. DO NOT FORCE. Fill solenoid
valve should open and LOW LEVEL alarm should be activated.
• Select MENU, LEVEL SENSING, TEST LEVEL SENSORS. An “L” or “G” will
indicate individual thermistor status (L = Liquid and G = Gas).
• Dip each thermistor in succession into LN2 Observe the controller display
noting that each thermistor changes from “G” to “L” as each is submerged.
Response time may vary.
• Return to the main menu and submerge the START FILL THERMISTOR IN
LIQUID. Note that the low level alarm ceases, fill solenoid valve is still open.
Control is flashing FILLING.
• Manually press STOP button and note that the fill solenoid valve closes.
• Press FILL button to re-open fill solenoid valve.
• Continue to lower the sensor until the STOP FILL thermistor is immersed in
the LN2 . The fill should stop after a confirming (CHECK) SPLASH GUARD
period.
38
• Simulate an over fill by lowering the HIGH LEVEL Alarm thermistor into the
LN2. HIGH LEVEL alarm should sound with 10 seconds.
• Re-install sensor assembly and thermocouple into sensor tube as before.
Open supply valve on LN2 supply.
Indicates Low LN2 Supply
When the START FILL thermistor is uncovered, the controller calls for the fill solenoid
valve to open. If the STOP FILL thermistor is not covered with LN2 in the predetermined amount of time, the controller is programmed to interpret this as a LN2
supply shortage.
1. Check contents gauge and pressure gauge of supply cylinder. Both liquid
contents and pressure (15 to 22 psi - 0.7bar/69kPa to 1.4 bar/138 kPa) are
required to complete a fill.
2. Confirm that no other transfer hose or apparatus is attached to the supply
cylinder. Either could compromise adequate tank pressure required to complete
a standard fill in 30 minutes.
• Check the distance that the LN2 must travel to reach the freezer. Observe
the time it takes for Liquid LN2 to reach the Freezer through the usual piping
conditions (Pipe-Temperature at start fill). Liquid should be entering the
chamber within 4 minutes under normal (usual) line temperature conditions.
• Distances over 6 feet (1.83 m) without a gas by-pass are discouraged due to
excessive boil off.
• A “Keep-Cold” or a “Keep-Full” device are almost always needed on an LN2
pipeline, whether the pipe is Vacuum Jacketed or Foam Insulated.
CS SERIES
3. Confirm that the solenoid valve is open when a fill is called for.
• If no flow is detected, the solenoid valve is not getting the signal to open, or
it is opening and there is a blockage in the line. Check the connections on
the leads near the solenoid itself, as well as the connection at the controller.
• Confirm that the wires have not been pinched, creating a short circuit.
4. If a longer fill time is desired, change the setting by selecting: MENU, SYSTEM
ALARMS, LN2 SUPPLY ALARM.
Indicates Open Sensor
1. Normally, this message is associated with a loose plug or connector. Check all
connections.
2. To determine this select: MENU, LEVEL SENSING, TEST LEVEL SENSORS. An “L”,
“G”, or “O” (open) will indicate individual thermistor status.
3. Thermocouple is not repairable. Replace from stock.
Temperature Reading 10 to 20 Degrees Warm
• Prepare an ice water slurry with crushed ice and tap water. Dip or pour LN2
into a styrofoam cup to prepare an LN2 bath. Calibrate the controller. Select
MENU, TEMPERATURE, CALIBRATE TEMPERATURE. Follow the on-screen
instructions. Hold the thermocouple in each bath until the control completes its
self-calibration.
• Make sure the thermocouple is clean and dry before and after each bath.
• Reposition the thermocouple at the elevation the customer wants to monitor or
control.
39
CS SERIES
Fill Solenoid Cycles On and Off
CS SERIES Control System features a timer function whereby the valve open duration
is limited, allowed to shut for a short warm-up and then opened again. To confirm
or change the valve open duration:
• Select MENU, USER OPTIONS, CONTROL OPTIONS, FREEZE-GUARD
OPTIONS, VALVE OPEN DURATION.
If the controller Freeze-Guard function detects that the valve is stuck in the open
position (i.e., continues to fill after the valve is de-energized), it will attempt to DeIce the valve by causing a rapid cycling of the valve in an attempt to dislodge the
blockage.
• To activate or de-activate this feature: Select MENU, USER OPTIONS,
CONTROL OPTIONS, FREEZE-GUARD OPTIONS, VALVE DE-ICING.
Repeated cycling of the Solenoid Valve, caused by excessive turbulence, power/low
battery, bad connection, or circuit board corrosion, could be attributable to moisture
invasion on the controller circuit board.
To troubleshoot proper operation of the fill valve:
1. Check gasket seals.
2. Access control board and allow time to fully dry prior to reinstallation.
3. Confirm the pressure of the supply vessel is less than 22 psi (1.4 bar/138kPa).
4. Watch the interface panel on the controller to verify if the Start and Stop Fill
cycles are repeating. You should see and hear each cycle.
Fill Solenoid Makes Excessive Humming Noise
During normal operation, the fill solenoid valve will make a soft humming noise. If
the noise is excessively loud, turn the fill valve on and off using the control panel. If
a soft humming noise is heard, the valve should continue to operate reliably. If the
humming noise is excessive, replace solenoid valve. You should change the solenoid
valve every 3 years.
Display = “Check”
The word CHECK will flash during periods of Splash-Guard protection.
Lid Open Alarm
A lid switch is located near the front hinge. The proximity switch should actuate when
the hinge is opened and again when it is shut.
40
• Confirm that there is a ½ in. (12 mm) gap between the sensors.”
QCF (Quick Chill Feature) Will Not Operate
See LID OPEN ALARM
To activate or adjust times: Select MENU, USER OPTIONS, CONTROL OPTIONS,
LID/DEFOG SETTINGS, QUICK CHILL.
Defog Feature Will Not Operate
See LID OPEN ALARM.
To activate or adjust times: Select MENU, USER OPTIONS, CONTROL OPTIONS,
LID/DEFOG SETTINGS, AUTO DE-FOG.
Push Buttons Will Not Respond
Moisture may have migrated into the sealed faceplate. Call Taylor-Wharton for a
Return Authorization. Refer to the Service and Maintenance section of this manual to
get the contact information you will need to return any equipment.
Liquid Level Readout is Incorrect
Liquid level is determined by a 8-thermistor assembly inside a protective sensor
tube. The user may adjust the desired START FILL elevations and STOP FILL
elevations for the pool of nitrogen with the confines of the 5 inches (127mm). START
FILL cannot be assigned to thermistor number 1. STOP FILL cannot be assigned to
thermistor number 8. The factory installs this eight thermistor assembly all the way to
the floor of the sensor tube, which positions thermistor number 1 at 1 in. (25.4 mm)
above the floor (i.e.; the first thermistor is offset from the floor zero inches).
CS SERIES
If you would like to stop the fill at a depth greater than seven inches from the
floor, the sensor assembly must physically be raised inside the sensor tube. The
distance that it has been raised becomes the new OFFSET and the controller must
be notified of this offset from the floor. The factory settings are START FILL at 3
inches and STOP FILL at 5 inches, with an OFFSET of zero.
Power Failure Alarm
If power is interrupted, a note of the time and duration will be recorded in the log.
If the unit is not connected to a battery back-up or an external alarm, then no local
alarm will be sounded unless a high temperature is detected after power resumes. A
Remote Alarm will be triggered anytime the power is interrupted.
REPLACEMENT PARTS
A complete list of Replacement Parts and Accessories for the LABS Cryostorage
Systems is available from the following Taylor Wharton Customer
Care Centers:
Europe Americas Asia/Australia
+421 (0)55-727-7157 +251-443-8680 +61 (0)2 6040-2533
In U.S. and Canada:
800-898-2657
41
CS SERIES
SERVICE AND MAINTENANCE HISTORY LOG
Fill in top section at installation. Copy this form each time service is required.
Fill bottom section with service notes to keep a complete log of each freezer service and maintenance history.
End User Company Name _________________________________ LOG NO ______________________________
Service Contract/NameCompany ___________________________________________________________________
Service Contract Phone Number/Fax ________________________________________________________________
Cryostorage System Model/Serial Number ___________________________________________________________
Controller Serial Number__________________________________________________________________________
Control Version Number __________________________________________________________________________
In-service Date ___________________________________________________________________________________
Describe Conditions – Actual Describe Conditions – Controller Reading
Liquid Level – via Dipstick ______________________ Liquid Level – Per Controller _______________________
Level Sensor Type _____FG _____ 8T _____4T _____ Liquid Level Setting _____ HIGH _____LOW
Lid _____ Open _____ Closed Lid _____Open _____ Closed
Filling _____ Yes _____ No Filling ______Yes______No
Temperature _________________________________ Temperature _____________________________________
LN2 Supply ______________ ltr______________ psi LN2 Supply Alarm ___ On ______ Off
Note: Ice Build-up _____ a little _____a lot Remote Alarm ______ On ______ Off
Note: Gasket condition _____ Seals _____ Leaks Audible Alarm ______ On ______ Off
Display Lights _____ On _____ Off
Taylor-Wharton Technical Services - Phone: In USA & Canada 800-898-2657
NOTE: All Taylor-Wharton Cryostorage Systems must be cleaned and sterilized prior to return to Taylor-Wharton for repair or maintenanceand
must be accompanied by a written statement to this effect. Any Cryostorage Systems or CS SERIES Control System received without this
statement will be returned to the sender, freight collect. Contact Customer Service by telephone: 800-898-2657 In USA & Canada or
+421 (0)55-727-7124 Kosice or +49 (0)4841-985139 Husum. Or email: salesandservice@taylorwharton.com.
42
Appedix
Certifications & Listings
This product complies with the following standards and directives:
• CB Scheme Report and Certificate
• IEC 61010-1
• UL 61010-1
• CAN/CSAC22.2#61010-1
• 93/42/EEC Medical Device Directive for the European Union
Guidance and Manufacturer’s Declaration – Electromagnetic Emissions
The CS CONTROL SYSTEM is intended for use in the electromagnetic environment
specified below. The customer or the end user of the CE1000 should assure that it is used
in such an environment.
only for its internal function. Therefore, its RF
emissions are very low and are not likely to cause
any interference in nearby electronic equipment.
Class BThe CS CONTROL SYSTEM is suitable for
use in all establishments, including domestic
Class A
Complies
establishments and those directly connected to
the public low-voltage power supply network that
supplies buildings used for domestic purposes.
43
CS SERIES
Table 7.0
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The CS CONTROL SYSTEM is intended for use in the electromagnetic environment
specified below. The customer or the end user of the CS CONTROL SYSTEM unit should
assure it is used only in such an environment.
Immunity TestEN/IEC 60601
Test Level
Electromagnetic
± 6kV contact
Discharge (ESD)
± 8kV air
EN/IEC 61000-4-2
Electrical fast
transient/burst
± 2kV for
power
supply lines
EN/IEC 61000-4-4
± 1kV for
input/output
lines
Surge
± 1kV
differential
EN/IEC 61000-4-5
mode (lineline)
Compliance
Level
± 6kV contact
± 8kV air
± 2kV for
power
supply lines
± 1kV for
input/output
lines
± 1kV
differential
mode (lineline)
Intneded Electromagnetic
Environment
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines
EN/IEC 61000-4-11
± 2kV
common mode
(line-earth)
<5% UT
(>95% dip in
UT) for 0.5
cycle
40% UT (60%
dip in UT) for
5 cycles
± 2kV
common mode
(line-earth)
<5% UT
(>95% dip in
UT) for 0.5
cycle
40% UT (60%
dip in UT) for 5
cycles
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the CS CONTROL SYSTEM
requires continued operation
during power mains interruptions,
it is recommended that the CS
CONTROL SYSTEM be powered
from an uninterruptible power
Power frequency
(50/60Hz) magnetic
field
70% UT (30%
dip in UT) for
25 cycles
<5% UT
(>95% dip
in UT) for 5
seconds
3A/m3A/mPower frequency magnetic fields
70% UT (30%
dip in UT) for
25 cycles
<5% UT
(>95% dip
in UT) for 5
seconds
supply or a battery.
should be at levels characteristic
of a typical location in a
typical commercial or hospital
EN/IEC 61000-4-8
environment.
Note UT is the a.c. mains voltage prior to application of the test level.
44
Table 8.0
Guidance and Manufacturer’s Declaration – Electromagnetic Immunity
The CS CONTROL SYSTEM is intended for use in the electromagnetic environment
specified below. The customer or the end user of the CS CONTROL SYSTEM should assure
it is used in such an environment.
Immunity TestEN/IEC 60601
Test Level
3Vrms
Conducted RF
EN/IEC 61000-4-6
Radiated RF
EN/IEC 61000-4-3
150kHz to
80MHz
3V/m
80MHz to
2.5GHz
Compliance
Level
3Vrms
150kHz to
80MHz
3V/m
80MHz to
2.5GHz
Intended Electromagnetic
Environment
Portable and mobile RF
communications equipment
should be used no closer to any
part of the CS CONTROL SYSTEM,
including cables, than the
recommended separation distance
calculated from the equation
applicable to the frequency of the
transmitter.
Recommended separation
distance
d = 1.2√P
d = 1.2√P 80MHz to 800 MHz
d = 2.3√P 800MHz to 2.5GHz
where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d
is the recommended minimum
separation distance in meters (m).
Field strengths from fixed RF
transmitters, as determined
by an electromagnetic site
surveya, should be less than
the compliance level in each
frequency range.
b
Interference may occur in the
vicinity of equipment marked with
the following symbol:
CS SERIES
Graphic needed.
NOTE 1: At 80MHz and 800MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from objects, structures and people.
a
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast and TV
broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic
environment due to fixed RF transmitters, an electromagnetic site survey should be
considered. If the measured field strength in the location in which the CS CONTROL
SYSTEM is used exceeds the applicable RF compliance level above, the CS CONTROL
SYSTEM should be observed to verify normal operation. If abnormal performance is
observed, additional measures may be necessary, such as reorienting or relocating the
CS CONTROL SYSTEM.
b
Over the frequency range 150kHz to 80MHz, field strengths should be less than 3V/m.
45
CS SERIES
Table 9.0
Recommended separation distances between portable and mobile RF communications
equipment and the CS CONTROL SYSTEM.
The CS CONTROL SYSTEM is intended for use in an electromagnetic environment in which
radiated RF disturbances are controlled. The customer or the user of the CS CONTROL
SYSTEM can help prevent electromagnetic interference by maintaining a minimum distance
between the portable and mobile RF communications equipment (transmitters) and the CS
CONTROL SYSTEM as recommended below, according to the maximum output power of
the communications equipment.
Rated maximum output
rding to frequency
of transmitter in
meters (m) power of
transmitter in watts (W)
0.01.12.12.23
0.1.38.38.73
1.01.21.22.3
103.83.87.3
100121223
For transmitters rated at a maximum output power not listed above, the recommended
separation distance d in meters (m) can be estimated using the equation applicable to the
frequency of the transmitter, where P is the maximum output power rating of the transmitter
in watts (W) according to the transmitter manufacturer.
NOTE 1: At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Separation distance acco Rated maximum output rding to
frequency of transmitter in meters (m)
150kHz to 80MHz
d = 1.2√P
80MHz to 800MHz
d = 1.2√P
800MHz to 2.5GHz
d = 2.3√P
46
Declaration of Conformity
EC DECLARATION OF CONFORMITY
The undersigned representing the manufacturer
Pacer Digital Systems, Inc.
Attn: Kevin Oeff
8658 Castle Park Drive
Suite 103
Indianapolis, IN 46256
USA
Herewith declared that the Product: LN2 Level Control for Cryostorage System
Model/Type ref.: CS CONTROL SYSTEM
is in conformity with the Essential requirements of the following EC Directives when
subject to correct installation, maintenance and use conforming to its(their)intended
purpose, to the applicable regulations and standards, to our operation and
maintenance manual.
93/42/EEC Medical Device Directive
2004/108/EC EMC Directive
2006/95/EC Low Voltage Directive
CS SERIES
and that the Standards and/or technical specifications referenced below have been
applied:
• EN 60601-1:2006 Edition 3 Medical Electrical Equipment – General
Requirements for basic safety and essential performance.
• IEC 60601-1-2: 2007 Edition 3 Medical Electrical Equipment – General
Requirements for basic safety and essential performance – Collateral standard :
Electromagnetic Compatibility
Signature: Kevin Oeff Digitally signed by Kevin Oeff
DN: cn=Kevin Oeff, o=Pacer Digital Systems, Inc., ou,
email=kevinoeff@pacerdigital.com, c=US
Date: 2012.05.24 13:48:04 -04’00’
Position: President
Date: 24 May 2012
Place: Indianapolis, IN USA
Warranty
Taylor-Wharton warrants that each of its electronic control products will be free
from defects in material and workmanship, in the normal service for which the
product was manufactured, for a period of two years from the date of purchase.
Taylor-Wharton at its option will either repair or replace any item covered under this
warranty.
This warranty is void if the product is used for any other purpose than that for which
it was designed, including but not limited to connection with third party systems.
47
CS SERIES
This warranty is also void if the product is in any way altered or repaired by others.
Taylor-Wharton shall not be liable under this warranty, or otherwise, for defects
caused by negligence, abuse or misuse of this product, corrosion, fire or the effects
of normal wear.
The remedies set forth herein are exclusive. Taylor-Wharton shall not be liable for
any indirect or consequential damages including, without limitation, damages
relating to lost profits or loss of products, resulting from the delivery, use or
failure of the product or for any other cause. By accepting delivery of the product,
the purchaser acknowledges that this limitation of remedies is reasonable and
enforceable. In no case shall Taylor-Wharton’s liability exceed the purchase price for
the product.
Labels
Figure 18.1 Main Rating Label
Figure 18.2 Control & Display Label
48
CS SERIES
Cryowire
Network
Power
12 VDC, 3A
Battery
12V, 2.5A
4075 Hamilton Blvd.
Theodore, AL 36582
1-800-898-2657
www.taylorwharton.com
Figure 18.3 Connector Box Label
Cryowire
Network
Cryowire
Local
Remote
Alarm
NO
C
NC
Figure 18.4 Battery Label - (Optional)
49
CS SERIES
50
CS SERIES
51
The Americas
Phone: + 251-443-8680
Phone: + 800-898-2657 in USA & Canada