Temperature Limitation, Operating: 10°C to 40°C, Storage: -15°C to 50°C
Humidity Limitation, 10% - 90%
Refer to instruction manual/ booklet
Disposal: Contact local distributor who will take the necessary steps according to your national
market.
Do Not Iron
Damp Wipe Only
Chlorinated Bleach: concentration less than or equal to 1000 ppm chlorinate or 70% alcohol
Do Not Tumble Dry
Do Not Dry Clean
Allow to Completely Air Dry
Manufacturer
Protected against solid foreign objects of 12,5 mm and greater; Protection against vertically falling
water drops
Authorized representative in the European community
Catalogue Number (model)
Serial Number
Minimum inflation level of the mattress
Maximum inflation level of the mattress
CPR
Do Not Open with Cutter
Return To Table of Contents
4476004- 5210 V1.5www.stryker.com
Introduction
WARNING/CAUTION/NOTE DEFINITION
The words WARNING, CAUTION and NOTE carry special meanings and should be carefully reviewed.
WARNING
Alerts the reader about a situation which, if not avoided, could result in death or serious injury. It may also describe
potential serious adverse reactions and safety hazards.
CAUTION
Alerts the reader of a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the
user or patient or damage to the equipment or other property. This includes special care necessary for the safe and
effective use of the device and the care necessary to avoid damage to a device that may occur as a result of use or
misuse.
NOTE
Provides special information to make maintenance easier or important instructions clearer.
www.stryker.com476004-5210 V1.55
Return To Table of Contents
Technical Specification
ItemSpecification
Power SupplyAC230V 50Hz, 0.05A ( for 230V system )
Fuse RatingT1AL, 250V
Dimension (L x W x H)25 x 12.5 x 8.5 cm / 9.8” x 4.9” x 3.3”
Weight1.4 kg / 3.1 lbs
Cycle Time8 min/60 Hz, 9.6 min/50 Hz
Atmospheric Pressure 700 hPa to 1013.25 hPa
• Operation: 10°C to 40°C (50°F to 104°F)
Temperature
Environment
Humidity
Classification
• Storage: -15°C to 50°C (5°F to 122°F)
• Shipping: -15°C to 70°C (5°F to 158°F)
• Operation: 10% to 90% non-condensing
• Storage: 10% to 90% non-condensing
• Shipping: 10% to 90% non-condensing
• Class II, Type BF, IP21
• Applied Part: Air Mattress
• Not suitable for use in the presence of a flammable anesthetic mixture (No AP or APG protection)
Air MattressSpecification
ModelP100
Model Number2880
Flame Retardant StandardsEN 597-1 and EN 597-2
Safe Working Load135 kg / 297 lbs
Dimension (L x W x H)193 x 89 x 18 cm / 76”x 35” x 7.1”
Weight5.75 kg / 12.7 lbs
Return To Table of Contents
6476004- 5210 V1.5www.stryker.com
Introduction
This manual is designed to assist with the operation and maintenance of the P100 Powered Support Surface. Carefully
read this manual thoroughly before using or beginning maintenance on the support surface. To ensure safe operation
of this equipment, it is recommended that methods and procedures are established for educating and training staff on
the safe operation of the support surface.
CONTRAINDICATIONS
Air support therapy is not recommended when spinal stability is a concern. This support surface is not intended to
support a patient in the prone position.
INTENDED USE OF PRODUCT
P100 is a constant low pressure powered support surface intended to provide pressure redistribution to aid in the
prevention and treatment of pressure ulcers. The system consists of a control unit combined with an alternating air
cell mattress. The air cells redistribute the patient’s weight over the surface and aid in the reduction of tissue interface
pressure. It is recommended that the product be operated by personnel who are qualified to perform general nursing
procedures and have received adequate training in the prevention and treatment of pressure ulcers.
This support surface is intended to be used with human patients in a general hospital, nursing home or homecare
environment and for patients at risk of developing pressure ulcers, as well as those who require therapy for pre- existing
pressure ulcers. The safe working load for P100 is 135 kg /297 lbs; the patient must not exceed safe working load
specified by the support surface, frame, and accessories. Patients shall meet the minimum age requirement of 2 years
old.
P100 shall be used with a mattress cover at all times.
The support surface is not intended to be a sterile product nor is it intended to include a measuring function.
EXPECTED SERVICE LIFE
The products are intended to offer safe and reliable operation when in use or installed according to the instructions
provided by Stryker Medical. Stryker Medical recommends that the system be inspected and serviced by authorized
technicians if there are any signs of wear or concerns with device function and indication on products. Otherwise,
service and inspection of the devices generally should not be required. P100 has an expected service life of 2 years.
www.stryker.com476004-5210 V1.57
Return To Table of Contents
Introduction
CONTACT INFORMATION
Contact Stryker Customer Service or Technical Support at: (800) 327-0770 or (269) 324-6500.
Stryker Medical
3800 E. Centre Avenue
Portage, MI 49002
USA
Please have the serial number (A) of your Stryker product available when calling Stryker Customer Service or Technical
Support. Include the serial number in all written communication.
PRODUCT SERIAL NUMBER LOCATION/IDENTIFICATION
The serial number (A) is located at the mattress cover near
foot right corner of the mattress as shown in Figure 1. Also
on the foam crib and air cell with stamp on it, to reference the
serial number, unzip the cover about one foot to access the
foam crib and air cell. The serial number is also located at the
bottom casing of control unit.
▼
Figure 1
Format:
2880
MYYM M-SSSSS
Model Number Legend (X)
2880 P100
P100 Mattress
Foam and Air Cell
• M = Mattress
• YY = Year
• MM = Month
• SSSSS = Sequence (Numeric)
Month Legend (MM)
January01
February02
March03
April04
May05
June06
July07
August08
September09
October10
A
Year Legend (YY)
201212
201313
201414
201515
201616
Return To Table of Contents
8476004- 5210 V1.5www.stryker.com
November11
December12
Summary of Safety Precautions
WARNING
• Check patient’s skin regularly. Consult physician if any redness or skin break occurs. Serious injury could result if
the patient’s skin condition is left untreated.
• Do not place the control unit in the patient’s bed, in contact with the patient, or under sheets or other coverings.
• Doing so could cause serious injury or could affect control unit performance.
• Do not use in the presence of a flammable anesthetic mixture or with oxygen (O2) or nitrous oxide (N2O).
• Verify bed side rails are compatible with bed frame and existing mattress. A risk assessment must be performed
by a suitably qualified person, especially when side rails are prescribed, to ensure that the bed meets the IEC
60601-2-52 bed standard.
• Use with appropriate top sheet and minimize layers of bedding between patient and mattress.
• Assess patient’s risk of entrapment according to protocols and monitor accordingly.
• Close supervision is necessary when this product is used on or near children. Electrical burns or choking may
result from a child swallowing a small part detached from the device.
• Use this product only for its intended use as described in this manual.
• Do not operate product if the power cord or plug has been damaged.
• Keep the cord away from heated surfaces.
• Never block any air openings of this product or place it on soft surfaces, such as a bed or couch, where openings
may be blocked. Keep the air opening free of lint, hair, and other similar particles.
• Never drop or insert any object into any opening or hose.
• Do not modify this equipment without the authorization of the manufacturer.
• Mattress covers have passed skin sensitization and skin irritation tests. However, if you suspect that you may have
had or are having an allergic reaction, please consult a physician immediately.
• The power cord to the Control Unit should be positioned to avoid a strangulation hazard and/or damage to the cord.
Careful consideration is required when routing the power cable. It is recommended that placing the cord under the
bed frame and attaching it to an electrical outlet at the head of bed.
• Serious injury or death can result from the use (potential entrapment) or non-use (potential patient falls) of siderails
or other restraints. The safe use of the support surface is maximized when used in conjunction with siderails;
there may be an increased risk of falls when siderails are not present. Local policies regarding the use of siderails
should be taken into account. Whether and how to use siderails is a decision that should be based on each patient’s individual needs and should be made by the physician, operators, and responsible parties.
• The risk of entrapment can develop when the support surface is placed on bed frames that leave gaps of even a
few inches between the support surface and the headboard, footboard, and siderails. The support surface is NOT
to be used when such gaps are present.
• When cleaning the support surface, ensure that no liquid is allowed to seep into the zipper area and watershed
cover barrier (underside); fluids allowed to come in contact with the zipper may leak into the support surface.
• Do not expose the mattress to excessive moisture. Personal injury or equipment damage could occur.
• The use of quaternaries containing glycol ethers and/or accelerated hydrogen peroxides may compromise the
cover integrity and legibility.
• Be aware of devices or equipment placed on the top of the support surface. Damage to the surface may occur due
to the weight of the equipment, heat generated by the equipment, or sharp edges on the equipment.
• Do not put overlays or accessories inside the cover. Doing so may reduce pressure redistribution performance.
• It is the responsibility of the caregiver team to evaluate the appropriate CPR protocol to be used with the surface.
• If there is a possibility of electro-magnetic interference with mobile phones, please increase the distance (3.3m)
between devices or turn off the mobile phone.
NOTE
The P100 support surface must be used with a mattress cover at all times. The support surface cover may interact
with all external skin.
www.stryker.com476004-5210 V1.59
Return To Table of Contents
CONTROL UNIT FRONT
Figure 2
◄
1. Power Switch On/Off
2. Front Panel
1
2
CONTROL UNIT REAR
Figure 3
◄
3
4
5
3. Hanger
4. Air Hose Port
5. Power Cord
P100
Product Description
Figure 4
◄
6. P100 Mattress
7. Foam and Air Cell
8. Air Hose
CONTROL PANEL
9
10
6
Figure 5
◄
9. Pressure Adjust Knob
Pressure adjust knob controls the air pressure output. When turning
clockwise, the output pressure will increase. Vice versa for decreasing air
pressure. Please consult your care giver for a suitable setting.
10. Power Switch
To turn on/off the control unit:
a. Power ON/OFF Switch on side of unit.
b. COMFORT CONTROL DIAL adjusts for patient comfort.
Soft(
Firm (
7
)—Minimum inflation level of the mattress
)—Maximum inflation level of the mattress.
8
Return To Table of Contents
10476004-5210 V1.5www.stryker.com
12
Instructions
1. Place control unit on flat surface or suspend control unit on end of bed using attached hooks. See Figure 2 and
Figure 3. Remove the plug to disconnect the device.
2. Position the mattress on bed frame.
3. Connect the hose assembly between the mattress air cell and the control unit. Unscrew the cap from the air
valve of the mattress and screw the adaptor from control unit onto the air valve tightly.
4. Plug the power cord and adjust the pressure control knob to highest setting for quick inflation and turn control
unit on using green on/off switch. The unit will take approximately 40 minutes to inflate the mattress.
5. After installation, make sure the flap is not folding upwards to avoid fluid seeping through mattress cover.
NOTE
Make sure the control unit is suitable for the local power voltage and frequency.
6. Position patient on the mattress and adjust the pressure control knob for patient comfort.
7. A Hand Check must be performed every 8 hours to verify proper operation of the device. See Figure 6.
8. To perform the Hand Check:
With the patient on his or her back, slide hand, flat and palm up, between the overlay and the mattress. Hand
should be directly under the air cell that is under the patient’s buttocks (or other bony area). See Figure 6.
Figure 6
▼
HAND CHECK
Properly inflatedUnder-inflated
Wait for full inflation of the air cell directly above hand. If the patient’s body is not in direct contact with hand, the system
is operating correctly. If, during full inflation of air cell, the patient’s body is in direct contact with flat hand, the system
is not operating properly. Adjust the pressure control to a higher setting. Wait 10 minutes and repeat the Hand Check.
If the Hand Check fails, check that the hoses are not kinked or pinched. If repeated Hand Check fails and hoses are
not kinked, contact Stryker for further instruction.
WARNING
Deflate before CPR or CPR could be ineffective.
To deflate mattress for CPR:
Disconnect the hoses from the control unit. See Figure 7. The air cell will deflate in approximately 20 seconds.
Proceed with CPR procedures.
Figure 7
▼
www.stryker.com476004-5210 V1.511
Return To Table of Contents
Cleaning and Disinfection
The control unit housing, tubing, and mattress should be cleaned between patients.
• To clean, use water and a clean cloth to wipe down the Control Unit, power cord, hoses, mattress top cover,
middle layer and bottom cover. Do not clean the foam. Do not use abrasive cleaners on the mattress. Note:
Blood and other body fluids must be thoroughly cleaned from all surfaces before applying disinfectants.
• Apply disinfectants to the external surfaces of the control unit, hoses and mattress top cover, middle layer and
bottom cover by wiping. Stryker recommends a chlorine-based solution with a concentration less than or equal to
1000 ppm or 70% alcohol twice a week.
• It is not recommended to disinfect the internal parts of the mattress on a regular basis, but only as needed for
particular instance, the air cell and the middle layer of the cover could be wiped with a cloth and disinfectants as
recommended above.
• Wipe down the mattress with a clean, dry cloth to remove any excess of disinfectant.
• If other detergent or other cleaning agent is used, choose one that will not have adverse chemical effects on the
surface of the plastic case of the control unit, mattress cover and any other component of the device.
• When cleaning the support surface, ensure that no liquid is allowed to seep into the zipper area and watershed
cover barrier (underside); fluids allowed to come in contact with the zipper may leak into the support surface.
• Avoid dust and proximity to dusty areas.
• All components should be air dried thoroughly before use.
WARNING
• Do not use phenolic based products for cleaning.
• Do no dry the mattress in direct sunlight.
Return To Table of Contents
12476004-5210 V1.5www.stryker.com
Troubleshooting
ProblemSolution
Loss of powerCheck if the plug is connected to mains.
Patient is bottoming out
Air cells fail to inflate
No air produced from some air
outlets of the air tube connector
Pressure setting might be inadequate for the patient. Adjust comfort range 1 to 2
levels higher and wait for a few minutes for best comfort.
Make sure the air hose is not kinked, cracked, or split. Verify that the power
switch is illuminated, signifying the control unit has power. Verify that the air
hoses are fully inserted with a positive connection.
This is normal since there is alternating mode. Air outlets take turns to produce
air during their cycle time.
www.stryker.com476004-5210 V1.513
Return To Table of Contents
Service Information
COVER REPLACEMENT
Tools Required: None
Procedure:
1. Disconnect the hose assembly between the mattress air cell and control unit.
2. Unzip the top cover.
3. Remove the air cell.
4. Unzip the middle layer at the patient’s right side and then remove the foam from the bottom part.
5. Discard the old cover.
6. Place the new cover, unzipped and open the top cover and middle layer.
7. Carefully slide the foam to the bottom part and zip the middle layer to close.
8. Carefully place the air cell on the top part and zip the cover to close.
9. Verify proper operation of the unit before returning it to service.
AIR CELL REPLACEMENT
Tools Required: None
Procedure:
1. Disconnect the hose assembly from the air valve of mattress.
2. Unzip the top cover.
3. Remove and discard the old air cell.
4. Place the new air cell and zip the cover to close
CONTROL UNIT REPLACEMENT
Tools Required: None
Procedure:
1. Disconnect the plug from mains power and hose.
2. Discard the old control unit.
3. Place the new control unit and connect the plug to mains power and hose.
HOSE REPLACEMENT
Tools Required: None
Procedure:
1. Disconnect the hose from control unit and mattress.
2. Discard the old hose.
3. Connect the new hose to control unit and mattress.
Return To Table of Contents
14476004-5210 V1.5www.stryker.com
Preventive Maintenance
Preventative maintenance should be performed annually, at a minimum. A preventative maintenance program should
be established for all Stryker Medical equipment. Preventative maintenance may need to be performed more frequently
based on the usage level of the product.
CHECKLIST
_______Cover zipper opens and closes properly and has no visible damage.
_______No tears, rips, holes, cracks, or other openings in the mattress cover.
_______Check labels for legibility, proper adherence, and integrity.
_______Support surface cover straps and snaps are intact and are not damaged.
_______Straps properly secure the support surface assembly to the crib.
_______Foam and other components have not degraded or come apart.
_______Check main power cord and do not plug if there is an abrasion or excessive wear.
_______Check airflow from the air hose.
_______Check the air hose if there is kink or breaks.
_______Verify proper operation of the unit before returning it to service.
The parts and accessories listed on this page are currently available for purchase. Some of the parts identified on the
assembly drawing parts in this manual may not be individually available for purchase. Please call Stryker Customer
service USA at 1-800-327-0770 for availability and pricing.
Part NamePart Number
P100 mattress cover assembly2880-030-100
Air cell assembly2880-030-400
P100 control unit2880-030-500
Air hose, PVC, P1002880-030-520
Return To Table of Contents
16476004-5210 V1.5www.stryker.com
Appendix A: EMC Information
GUIDANCE AND MANUFACTURER’S DECLARATION- ELECTROMAGNETIC EMISSIONS:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make
sure it is used in such an environment.
Voltage fluctuations / Flicker
emissions
IEC61000-3-3
WARNING
1. The device should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is
necessary, the device should be observed to verify normal operation in the configuration in which it will be used.
2.Use of accessories, transducers and cables other than those specified or provided by the manufacturer of this
equipment could result in increased electromagnetic emissions or decreased electromagnetic immunity of this
equipment and result in improper operation.
3. Portable RF communications equipment (including peripherals such as antenna cables and external antennas)
should be used no closer than 30 cm (12 inches) to any part of the Pump, including cables specified by the
manufacturer. Otherwise, degradation of the performance of this equipment could result.
Group1
Class B
Class A
Complies
The device uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not likely
to cause any interference in nearby electronic equipment
The device is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network
www.stryker.com476004-5210 V1.517
Return To Table of Contents
Appendix A: EMC Information
GUIDANCE AND MANUFACTURER’S DECLARATION- ELECTROMAGNETIC IMMUNITY:
This device is intended for use in the electromagnetic environment specified below. The user of this device should make
sure it is used in such an environment.
Basic EMC
standard
Electrostatic
Discharge (ESD)
IEC61000-4-2
Electrical fast
transient/ burst
IEC61000-4-4
Immunity Test LevelsCompliance
HOME HEALTHCARE
Levels
ENVIRONMENT
±8kV contact
±15kV air
±2kV for power supply line
±1kV for input/output line
±8kV contact
±15kV air
±2kV for power
supply line
±1kV for input/
output line
Electromagnetic
Environment-Guidance
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30 %.
Mains power quality should be that
of a typical commercial or hospital
environment
Surge
IEC61000-4-5
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
Power frequency
(50/60Hz) magnetic
field
IEC61000-4-8
Conducted RF
IEC 61000-4-6
± 1 kV line(s) to
line(s)
± 2 kV line(s) to earth
Voltage Dips:
i) 100% reduction for 0.5
period,
ii) 100% reduction for 1 period,
iii) 30% reduction for 25/30
period,
Voltage Interruptions:
100% reduction for 250/300
period
± 1 kV line(s) to
line(s)
(1)
230V (U
)
T
Voltage Dips:
i) 100% reduction
for 0.5 period,
ii) 100% reduction
for 1 period,
iii) 30% reduction
for 25/30 period,
Voltage
Mains power quality should be that
of a typical commercial or hospital
environment.
Mains power quality should be that
of a typical commercial or hospital
environment. If the user of this
device requires continued operation
during power mains interruptions, it
is recommended that the device be
powered from an uninterruptible power
supply or a battery.
Interruptions:
100% reduction
for 250/300
period
30 A/m30 A/mPower frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
3 Vrms
0,15 MHz – 80 MHz
6 Vrms in ISM and amateur
radio bands between
0,15 MHz and 80 MHz
80 % AM at 1 kHz
(4)
6VrmsPortable and mobile RF
communications equipment should
be used no closer to any part of this
device, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Return To Table of Contents
18476004-5210 V1.5www.stryker.com
Appendix A: EMC Information
Radiated RF EM
Fields
IEC61000-4-3
10 V/m 80 MHz to 2,7 GHz
80 % AM at 1 kHz
385-6000 MHz, 9-28V/m, 80%
AM(1kHz) pulse mode and
other modulation
10V/mRecommended separation distance
d=√P 150kHz to 80MHz
d=0.6√P 80MHz to 800MHz
d=1.2√P 800 MHz to 2.7G Hz
Where P is the maximum output
power rating of the transmitter
in watts (W) according to the
transmitter manufacturer and d is the
recommended separation distance in
meters (m).
b
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey ,a should be
less than the compliance level in each
frequency ranged.
Interference may occur in the vicinity
of equipment marked with the following
symbol:
NOTE 1: UT is the a.c. mains voltage prior to the application of the test level
NOTE 2: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 3: These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and
reflection from structures, objects and people
NOTE 4: The ISM (industrial, scientific and medical) bands between 0.15 MHz and 80 MHz are 6.765 MHz to 6.795
MHz; 13.553 MHz to 13.567 MHz; 26.957 MHz to 27.283 MHz; and 40.66 MHz to 40.70 MHz. The amateur radio
bandsbetween 0.15 MHz and 80 MHz are 1.8 MHz to 2.0 MHz, 3.5 MHz to 4.0 MHz, 5.3 MHz to 5.4 MHz, 7 MHz to
7.3 MHz,10.1 MHz to 10.15 MHz, 14 MHz to 14.2 MHz, 18.07 MHz to 18.17 MHz, 21.0 MHz to 21.4 MHz, 24.89 MHz to
24.99MHz, 28.0 MHz to 29.7 MHz and 50.0 MHz to 54.0 MHz.
a)Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land
mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically
with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site
survey should be considered. If the measured field strength in the location in which the device is used exceeds
the applicable RF compliance level above, the device should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as reorienting or relocating the device.
b)Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 10 V/m.
www.stryker.com476004-5210 V1.519
Return To Table of Contents
Warranty
LIMITED WARRANTY
Stryker Medical Division, a division of Stryker Corporation, warrants to the original purchaser P100 Powered Support
Surface to be free from defects in material and workmanship for a period of one (1) years for the support surface
assembly and the control unit after date of delivery under normal use*. Stryker’s obligation under this warranty is
expressly limited to supplying replacement parts and labor for, or replacing, at this option, any product which is, in the
sole discretion of Stryker, found to be effective. If requested by Stryker, products or parts for which a warranty claim
is made shall be returned repaid to the factory. Any improper use or any alteration or repair by others in such manner
as in Stryker’s judgment affects the product materially and adversely shall void this warranty. Any repair of Stryker
products using parts not provided or authorized by Stryker shall void this warranty. No employee or representative of
Stryker is authorized to change this warranty in any way.
CONDITIONS AND LIMITATIONS
Stryker Medical’s P100 Powered Support Surface is designed for an expected service life as listed below under normal
use conditions, and with appropriate periodic maintenance as described in the operations/maintenance manual for
each device.
This statement constitutes Stryker’s entire warranty with respect to the aforesaid equipment. Stryker makes no other
warranty or representation, either expressed or implied, except as set forth herein. There is no warranty of
merchantability and there are no warranties of fitness for any particular purpose. In no event shall Stryker be
liable here under for incidental or consequential damages arising from or in any manner related to sales or use of
any such equipment. This warranty does not extend to, nor cover:
• Normal wear and tear; or
• Damage or product failure due to causes beyond Stryker’s control such as, but not limited to abuse, theft, fire,
flood, wind , lightning, freezing, clogging of mattress pores due to tobacco smoke, unusual atmosphere conditions, material degradation due to exposure to moisture; or
• Damage to support surface or support surface handles through the use of the support surface for patient transfer or transport.
* Normal use is defined as normal hospital or facility usage. Damages arising from abnormal use such as those caused
by needle punctures, burns, chemicals, negligent use or improper care or improper cleaning or staining resulting from
it are exempt from warranty coverage.
TO OBTAIN PARTS AND SERVICE
Stryker products are supported by a nationwide network of dedicated Stryker Field Service Representatives. These
representatives are factory trained, available locally, and carry a substantial spare parts inventory to minimize repair
time. Simply call your local representative or call Stryker Customer Service USA at 1-800-327-0770.
RETURN AUTHORIZATION
Merchandise cannot be returned without approval from the Stryker Customer Service Department. An authorization
number will be provided which must be printed on the returned merchandise. Stryker reserves the right to charge
shipping and restocking fees on return merchandise. Special, modified, or discontinued items not subject to return.
DAMAGED MERCHANDISE
ICC Regulations require that claims for damaged merchandise must be made with the carrier within fifth (15) days of
receipt of merchandise. Do not accept damaged shipments unless such damage is noted on the delivery receipt at the time of receipt. Upon prompt notification, Stryker will file a freight claim with the appropriate carrier for damages
incurred. Claim will be limited in amount to the actual replacement cost. In the event that this information is not received
by Stryker within the fifteen (15) day period following the delivery of the merchandise, or the damage was not noted on
the delivery receipt at the time of receipt, the customer will be responsible for payment of the original invoice in full.
Claims for any short shipment must be made within thirty (30) days of invoice.
INTERNATIONAL WARRANTY CLAUSE
This warranty reflects U.S. domestic policy. Warranty outside the U.S. may vary by country. Please contact your local
Stryker Medical representative for extra information.
Return To Table of Contents
20476004-5210 V1.5www.stryker.com
www.stryker.com476004-5210 V1.521
Return To Table of Contents
Stryker Medical
3800 E. Centre Avenue
Portage, Michigan 49002
USA
Stryker European Operations B.V.
Herikerbergweg 110
Amsterdam
1101 CM
Netherlands
www.stryker.com
Support motorisé P100
2880
Manuel d’utilisation et d’entretien
2017/12476004 - 5210 V1.5www.stryker.com
Table des matières
Symboles et définitions .................................................................................4
Élimination : contactez votre distributeur local qui prendra les dispositions nécessaires conformément
aux règles de votre marché national.
Ne pas repasser
Nettoyer uniquement avec un chiffon humide
Blanchiment au chlore : concentration en chlore inférieure ou égale à 1000 ppm ou 70 % en alcool
Ne pas sécher au sèche-linge
Ne pas nettoyer à sec
Laisser complètement sécher à l’air libre
Fabricant
Protégé contre les corps étrangers solides de 12,5 mm et plus ; protection contre les chutes de
gouttes d’eau verticales
Représentant agréé dans la Communauté européenne
Numéro de référence (modèle)
Numéro de série
Niveau de gonflage minimum du matelas
Niveau de gonflage maximum du matelas
Réanimation cardio-respiratoire (RCR)
Ne pas ouvrir avec un couteau
Retour à la Table des matières
4476004- 5210 V1.5www.stryker.com
Introduction
DÉFINITION DES TERMES AVERTISSEMENT/ATTENTION/REMARQUE
Les termes AVERTISSEMENT, ATTENTION et REMARQUE ont une signification particulière et doivent faire l’objet
d’une attention particulière..
AVERTISSEMENT
Avertit le lecteur d’une situation qui, si elle n’est pas évitée, pourrait avoir des conséquences mortelles ou entraîner des
blessures graves. Il peut également décrire des effets indésirables potentiellement graves et des risques pour la sécurité.
ATTENTION
Avertit le lecteur d’une situation potentiellement dangereuse qui, si elle n’est pas évitée, peut entraîner des blessures
légères ou modérées chez l’utilisateur ou le patient ou endommager l’équipement ou d’autres biens. Cela comprend le
fait de veiller à utiliser l’appareil de manière efficace et en toute sécurité et de veiller à éviter les dommages qui peuvent
se produire sur un appareil en raison de l’utilisation ou de la mauvaise utilisation qui en est faite.
REMARQUE
Fournit des informations spécifiques pour faciliter l’entretien ou clarifier des instructions importantes.
www.stryker.com476004-5210 V1.55
Retour à la Table des matières
Spécification technique
Dispositif de commandeSpécification
Alimentation électriqueCA 230 V 50 Hz, 0,05 A (pour une installation en 230 V)
Calibre du fusibleT1AL, 250V
Dimensions (L x l x h)25 x 12.5 x 8.5 cm
Poids1,4 kg
Temps de cycle8 min/60 Hz, 9.6 min/50 Hz
Environnementt
Classification
Pression
atmosphérique
Température
Humidité
de 700 hPa à 1013,25 hPa
• Fonctionnement : de 10°C à 40°C
• Stockage de -15°C à 50°C
• Transport : de -15°C à 70°C
• Fonctionnement : de 10 % à 90 % sans condensation
• Stockage de 10 % à 90 % sans condensation
• Transport : de 10 % à 90 % sans condensation
• Classe II, Type BF, IP21
• Partie appliquée : Matelas à air
• Non approprié à une utilisation en présence d’un mélange anesthésique inflammable (pas de protection de type AP ou APG)
Matelas à airSpécification
ModèleP100
Numéro du modèle2880
Normes ignifugesEN 597-1 et EN 597-2
Charge maximale d'utilisation135 kg
Dimensions (L x l x h)193 x 89 x 18 cm
Poids5,75 kg
Retour à la Table des matières
6476004- 5210 V1.5www.stryker.com
Introduction
Ce manuel a été conçu pour aider à l’utilisation et à l’entretien du support motorisé P100. Veuillez lire attentivement ce
manuel avant d’utiliser ou de commencer l’entretien du support. Pour assurer un fonctionnement en toute sécurité de
cet équipement, il est recommandé d’établir des méthodes et procédures pour l’instruction et la formation du personnel
concernant une utilisation en toute sécurité du support.
CONTRE-INDICATIONS
La thérapie sur support à air n’est pas recommandée en cas de problèmes de stabilité rachidienne. Ce support n’est
pas prévu pour soutenir un patient couché sur le ventre.
UTILISATION PRÉVUE DU PRODUIT
P100 est un support motorisé à faible pression constante, conçu pour permettre une répartition de la pression afin
d’aider à la prévention et au traitement des escarres de décubitus. Le système comprend un dispositif de commande
associé à un matelas à cellules d’air à air alterné. Les cellules d’air répartissent le poids du patient sur toute la surface
et aident à réduire la pression d’interface au niveau des tissus. Il est recommandé que le produit soit utilisé par du
personnel qualifié pour effectuer des procédures générales de soins infirmiers et correctement formé en matière de
prévention et de traitement des escarres de décubitus.
Ce support est destiné à être utilisé pour des patients humains dans un hôpital général, une maison de santé ou un
environnement de soins à domicile, et pour des patients présentant des risques de développement d’escarres de
décubitus, ainsi que pour ceux nécessitant une thérapie pour des escarres de décubitus préexistants. La charge
maximale d’utilisation pour le support P100 est de 135 kg ; le patient ne doit pas dépasser la charge maximale
d’utilisation spécifiée pour le support, le cadre et les accessoires. Les patients doivent être âgés au minimum de 2 ans.
P100 doit être à tout moment utilisé avec une housse de matelas.
Le support n’est pas conçu comme un produit stérile et n’est pas non plus destiné à inclure une fonction de mesure.
DURÉE DE VIE PRÉVUE
Les produits sont conçus pour offrir un fonctionnement fiable et en toute sécurité lorsqu’ils sont utilisés ou installés
conformément aux instructions fournies par Stryker Medical. Stryker Medical recommande de confier l’inspection et
l’entretien du système à des techniciens agréés en cas d’apparition de signes d’usures ou de problèmes concernant une
fonction de l’appareil et une indication sur les produits. À part dans ces cas, les appareils ne devraient généralement
pas nécessiter d’entretien ou d’inspection. La durée de vie prévue du support P100 est de 2 ansservice and inspection
of the devices generally should not be required. P100 has an expected service life of 2 years.
www.stryker.com476004-5210 V1.57
Retour à la Table des matières
Introduction
NOUS CONTACTER
Contactez le service clientèle ou l’assistance technique de Stryker aux numéros suivants : +1 (800) 327-0770 or (269)
324-6500.
Stryker Medical
3800 E. Centre Avenue
Portage, MI 49002
USA
Veuillez vous munir du numéro de série (A) de votre produit Stryker lorsque vous appelez le service clientèle ou
l’assistance technique de Stryker. Indiquez le numéro de série dans toute communication écrite..
LOCALISATION/IDENTIFICATION DU NUMÉRO DE SÉRIE DU PRODUIT
Le numéro de série (A) est situé au niveau de la housse du
matelas près du coin droit au pied du matelas, comme indiqué
sur la Figure 1. Le numéro de série est également apposé sur
la base en mousse et cellule d’air ; pour le consulter, ouvrez la
fermeture éclair de la housse d’environ 30 cm pour accéder à
la base en mousse et à la cellule d’air. Le numéro de série est
également situé sur la partie inférieure du boîtier du dispositif
de commande.
Matelas P100
Mousse et cellule d’air
▼
Figure 1
A
Format:
2880
M AA M M-SSSSS
Légende du numéro du modèle (X)
2880 P100
• M = Matelas
• AA = Année
• MM = Mois
• SSSSS = Séquence (Numérique)
Légende du mois
(MM)
Janvier01
Février02
Mars03
Avril04
Mai05
Juin06
Juillet07
Août08
Septembre09
Octobre10
Novembre11
Légende de
l'année (AA)
201212
201313
201414
201515
201616
Retour à la Table des matières
8476004- 5210 V1.5www.stryker.com
Décembre12
Loading...
+ 234 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.