Safety and Preparation for Use1-3
Specifications1-5
Abridged Command List1-7
GETTING STARTED
Your First Measurements2-1
The Basic Lock-in2-3
Displays and Traces2-7
Outputs, Offsets and Expands2-13
Scans and Sweeps2-17
Using the Disk Drive2-23
Aux Outputs and Inputs2-31
Trace Math2-35
SR850 BASICS
What is a Lock-in Amplifier?3-1
What Does a Lock-in Measure?3-3
The SR850 Functional Diagram3-5
Reference Channel3-7
Phase Sensitive Detectors3-9
Time Constants and DC Gain3-11
DC Outputs and Scaling3-13
Dynamic Reserve3-17
Signal Input Amplifier and Filters3-19
Input Connections3-21
Intrinsic (Random) Noise Sources3-23
External Noise Sources3-25
Noise Measurements3-27
OPERATION
FRONT PANEL 4-1
Power On/Off and Power On Tests4-1
Video Display4-1
Soft Keys4-2
Keypad4-2
Spin Knob4-2
Disk Drive4-2
Front Panel BNC Connectors4-2
SCREEN DISPLAY4-5
Default Display4-5
Data Traces4-6
Single/Dual Trace Displays4-7
Bar Graphs4-9
Polar Graphs4-10
Strip Charts4-11
Trace Scans, Sweeps and Aliasing4-13
Settings and Input/Output Monitor4-15
Menu Display4-15
Status Indicators4-16
KEYPAD4-19
Normal and Alternate Keys4-19
Menu Keys4-19
Additional Menus4-20
Entry Keys4-20
START/CONT and PAUSE/RESET4-20
CURSOR4-21
ACTIVE DISPLAY4-21
MARK4-21
CURSOR MAX/MIN4-21
AUTO RESERVE4-22
AUTO GAIN4-22
AUTO PHASE4-22
AUTO SETUP4-22
AUTOSCALE4-22
PRINT to a PRINTER4-23
PRINT to a FILE4-23
HELP4-23
LOCAL4-23
REAR PANEL 4-25
Power Entry Module4-25
IEEE-488 Connector4-25
RS232 Connector4-25
Parallel Printer Connector4-25
PC Keyboard Connector4-25
Rear Panel BNC Connectors4-26
Aux Inputs (A/D Inputs)4-26
Aux Outputs (D/A Outputs)4-26
X and Y Outputs4-26
Signal Monitor Output4-26
Trigger Input4-27
TTL Sync Output4-27
Preamp Connector4-27
USING SRS PREAMPS4-27
MENUS
Menu Guide5-1
Default Settings5-2
Reference and Phase Menu 5-3
Input and Filters Menu5-7
Gain and Time Constant Menu5-9
Output and Offset Menu5-15
Trace and Scan Menu5-17
Display and Scale Menu5-21
Aux Outputs Menu5-25
Cursor Setup Menu5-29
Edit Mark Menu5-31
Math Menu5-33
Disk Menu5-41
System Setup Menu5-49
Table of Contents
PROGRAMMING
GPIB Communications
6-1
RS232 Communications
6-1
Status Indicators and Queues
6-1
Command Syntax
6-1
Interface Ready and Status
6-2
GET (Group Execute Trigger)
6-2
DETAILED COMMAND LIST
6-3
Reference and Phase
6-4
Input and Filter
6-6
Gain and Time Constant
6-7
Output and Offset
6-9
Trace and Scan
6-10
Display and Scale
6-11
Cursor
6-13
Mark
6-14
Aux Input and Output
6-15
Math
6-16
Store and Recall
6-18
Setup
6-19
Print and Plot
6-21
Front Panel and Auto Functions
6-22
Data Transfer
6-23
Interface
6-28
Status Reporting
6-29
STATUS BYTE DEFINITIONS
6-30
Serial Poll Status Byte
6-30
Service Requests
6-31
Standard Event Status Byte
6-29
LIA Status Byte
6-32
Error Status Byte
6-32
PROGRAM EXAMPLES
Microsoft C, Nationall Instr GPIB
6-33
QUICKBASIC, Nationall Instr GPIB
6-39
TESTING
Introduction
7-1
Preset
7-1
Serial Number
7-1
Firmware Revision
7-1
General Installation
7-2
Necessary Equipment
7-3
If A Test Fails
7-3
PERFORMANCE TESTS
Self Tests
7-5
DC Offset
7-7
Common Mode Rejection
7-9
Amplitude Accuracy and Flatness
7-11
Amplitude Linearity
7-13
Frequency Accuracy
7-15
Phase Accuracy
7-17
Sine Output Amplitude
7-19
DC Outputs and Inputs
7-21
Input Noise
7-23
PERFORMANCE TEST RECORD
7-25
CIRCUITRY
Circuit Boards
8-1
Video Driver and CRT
8-1
CPU Board
8-3
Power Supply Board
8-5
DSP Logic Board
8-7
Analog Input Board
8-9
PARTS LISTS
Power Supply Board
8-11
Analog Input Board
8-13
DSP Logic Board
8-19
CPU Board
8-25
Chassis Assembly
8-29
Miscellaneous Parts
8-32
SCHEMATIC DIAGRAMS
CPU Board
Power Supply Board
DSP Logic Board
Analog Input Board
SAFETY AND PREPARATION FOR USE
WARNING
Dangerous voltages, capable of causing injury or death, are present in
this instrument. Use extreme caution whenever the instrument covers
are removed. Do not remove the covers while the unit is plugged into a
live outlet.
CAUTION
This instrument may be damaged if operated
with the LINE VOLTAGE SELECTOR set for the
wrong AC line voltage or if the wrong fuse is
installed.
LINE VOLTAGE SELECTION
The SR850 operates from a 100V, 120V, 220V, or
240V nominal AC power source having a line frequency of 50 or 60 Hz. Before connecting the power cord to a power source, verify that the LINE
VOLTAGE SELECTOR card, located in the rear
panel fuse holder, is set so that the correct AC input voltage value is visible.
Conversion to other AC input voltages requires a
change in the fuse holder voltage card position
and fuse value. Disconnect the power cord, open
the fuse holder cover door and rotate the fuse-pull
lever to remove the fuse. Remove the small printed circuit board and select the operating voltage
by orienting the printed circuit board so that the
desired voltage is visible when pushed firmly into
its slot. Rotate the fuse-pull lever back into its normal position and insert the correct fuse into the
fuse holder.
LINE FUSE
Verify that the correct line fuse is installed before
connecting the line cord. For 100V/120V, use a 1
Amp fuse and for 220V/240V, use a 1/2 Amp fuse.
LINE CORD
The SR850 has a detachable, three-wire power
cord for connection to the power source and to a
protective ground. The exposed metal parts of the
instrument are connected to the outlet ground to
protect against electrical shock. Always use an
outlet which has a properly connected protective
ground.
SERVICE
Do not attempt to service or adjust this instrument
unless another person, capable of providing first
aid or resuscitation, is present.
Do not install substitute parts or perform any unauthorized modifications to this instrument. Contact
the factory for instructions on how to return the instrument for authorized service and adjustment.
1-3
1-4
1-5
SPECIFICATIONS
SIGNAL CHANNEL
Voltage Inputs
Single-ended (A) or differential (A-B).
Current Input
106 or 108 Volts/Amp.
Full Scale Sensitivity
2 nV to 1 V in a 1-2-5-10 sequence (expand off).
Input Impedance
Voltage:
10 MΩ+25 pF, AC or DC coupled.
Current:
1 kΩ to virtual ground.
Gain Accuracy
±1% from 20°C to 30°C (notch filters off).
Input Noise
6 nV/√Hz at 1 kHz (typical).
Signal Filters
60 (50) Hz and 120(100) Hz notch filters (Q=4).
CMRR
90 dB at 100 Hz (DC Coupled).
Dynamic Reserve
Greater than 100 dB (with no signal filters).
Harmonic Distortion
<-90 dB to 10 kHz, <-80 dB to 100 kHz.
REFERENCE CHANNEL
Frequency Range
1 mHz to 102 kHz
Reference Input
TTL (rising or falling edge) or Sine.
Sine input is1 MΩ, AC coupled (>1 Hz). 400 mV pk-pk minimum signal.
Digital displays have no zero drift on all dynamic reserves.
Analog outputs: <5 ppm/°C for all dynamic reserves.
Time Constants
10 µs to 30 s (reference > 200 Hz). 6, 12, 18, 24 dB/oct rolloff.
up to 30000 s (reference < 200 Hz). 6, 12, 18, 24 dB/oct rolloff.
Synchronous filtering available below 200 Hz.
4 mVrms to 5 Vrms (into a high impedance load) with 2 mV resolution.
(2 mVrms to 2.5 Vrms into 50Ω load).
Amplitude Accuracy
1%
Amplitude Stability
50 ppm/°C
Outputs
Sine output on front panel. TTL sync output on rear panel.
When using an external reference, both outputs are phase locked to the
external reference.
SR850 DSP LOCK-IN AMPLIFIER
SR850 DSP Lock-In Amplifier
1-6
INPUTS AND OUTPUTS
Channel 1 Output
X, R, θ, or Trace 1-4. Traces are defined as A•B/C or A•B/C2 where A, B,
and C are selected from the quantities Unity, X, Y, R, θ, Xnoise, Ynoise,
Rnoise, Aux Inputs 1 through 4, or Frequency.
Output Voltage: ±10 V full scale. 10 mA max output current.
Channel 2 Output
Y, R, θ, or Trace 1-4. Traces are defined as A•B/C or A•B/C2 where A, B,
and C are selected from the quantities Unity, X, Y, R, θ, Xnoise, Ynoise,
Rnoise, Aux Inputs 1 through 4, or Frequency.
Output Voltage: ±10 V full scale. 10 mA max output current.
X and Y Outputs
Rear panel outputs of cosine (X) and sine (Y) components.
Output Voltage: ±10 V. 10 mA max output current.
Aux. Outputs
4 BNC Digital to Analog outputs.
±10 V full scale, 1 mV resolution. May be set to a fixed voltage or swept in
amplitude (linear or log). 10 mA max output current.
Aux. Inputs
4 BNC Analog to Digital inputs.
Differential inputs with1 MΩ input impedance on both shield and center
conductor. ±10 V full scale, 1 mV resolution.
Trigger Input
TTL trigger input triggers each data sample and/or start of scan.
Monitor Output
Analog output of signal amplifiers (before the demodulator).
DISPLAYS
Screen Format
Single or dual display.
Displayed Quantities
Each display may show one of the traces.
Traces are defined as A•B/C or A•B/C2 where A, B and C are selected from
the quantities Unity, X, Y, R, θ, Xnoise, Ynoise, Rnoise, Aux Inputs 1
through 4, or Frequency.
Display Types
Large numeric readout with bar graph, polar graph, and strip chart.
Chart Data Buffer
64k data points may be stored and displayed on strip charts. The buffer can
be configured as a single trace with 64k points, 2 traces with 32k points
each, or 4 traces with16k points each. The internal data sample rate ranges
from 512 Hz down to 1 point every 16 seconds. Samples can also be
triggered.
ANALYSIS FUNCTIONS
Smoothing
5 - 25 point Savitsky-Golay smoothing of trace regions.
Curve Fits
Line, Exponential, and Gaussian fits of trace regions.
Calculator
Arithmetic, trigonometric, and logarithmic calculations on trace regions.
Statistics
Mean and standard deviation of trace regions.
GENERAL
Monitor
Monochrome CRT. 640H by 480V resolution.
Adjustable brightness and screen position.
Interfaces
IEEE-488, RS232 and Printer interfaces standard.
All instrument functions can be controlled through the IEEE-488 and RS232
interfaces. A PC keyboard input is provided for additional flexibility.
Preamp Power
Power connector for SR550 and SR552 preamplifiers.
Hardcopy
Screen dumps to dot matrix and HP LaserJet compatible printers. Data
plots to HP-GL compatible plotters (via RS232 or IEEE-488). Screens can
also be saved to disk as PCX image files.
Disk
3.5 inch DOS compatible format, 720 kbyte capacity.
Storage of data and setups.
Power
60 Watts, 100/120/220/240 VAC, 50/60 Hz.
Dimensions
17"W x 6.25"H x 19.5"D
Weight
40 lbs.
Warranty
One year parts and labor on materials and workmanship.
1-7
COMMAND LIST
VARIABLES
i,j,k,l,m
Integers
f
Frequency (real)
x,y,z
Real Numbers
s
String
REFERENCE and PHASE
page
description
PHAS (?) {x}
6-4
Set (Query) the Phase Shift to x degrees.
FMOD (?) {i}
6-4
Set (Query) the Reference Source to Internal (0), Sweep (1) , or External (2).
FREQ (?) {f}
6-4
Set (Query) the Reference Frequency to f Hz.Set only in Internal reference mode.
SWPT (?) {i}
6-4
Set (Query) the Internal Sweep Type to Linear (0) or logarithmic (1).
SLLM (?) {f}
6-4
Set (Query) the Start Frequency to f Hz.Set only in Internal Sweep mode.
SULM (?) {f}
6-4
Set (Query) the Stop Frequency to f Hz.Set only in Internal Sweep mode.
RSLP (?) {i}
6-4
Set (Query) the External Reference Slope to Sine(0), TTL Rising (1), or TTL Falling (2).
HARM (?) {i}
6-5
Set (Query) the Detection Harmonic to 1 ≤ i ≤ 32767 and i•f ≤ 102 kHz.
SLVL (?) {x}
6-5
Set (Query) the Sine Output Amplitude to x Vrms. 0.004 ≤ x ≤5.000.
INPUT and FILTER
page
description
ISRC (?) {i}
6-6
Set (Query) the Input Configuration to A (0), A-B (1) , or I (2).
IGAN (?) {i}
6-6
Set (Query) the Current Conversion Gain to 1 MΩ (0) or 100 MΩ (1).
IGND (?) {i}
6-6
Set (Query) the Input Shield Groungind to Float (0) or Ground (1).
ICPL (?) {i}
6-6
Set (Query) the Input Coupling to AC (0) or DC (1).
ILIN (?) {i}
6-6
Set (Query) the Line Notch Filters to Out (0), Line In (1) , 2xLine In (2), or Both In (3).
GAIN and TIME CONSTANT
page
description
SENS (?) {i}
6-7
Set (Query) the Sensitivity to 2 nV (0) through 1 V (26) rms full scale.
RMOD (?) {i}
6-7
Set (Query) the Dynamic Reserve Mode to Max (0), Manual (1), or Min (2).
RSRV (?) {i}
6-7
Set (Query) the Dynamic Reserve to i
th
reserve. Set will switch to Manual Reserve Mode.
OFLT (?) {i}
6-7
Set (Query) the Time Constant to 10 µs (0) through 30 ks (19).
OFSL (?) {i}
6-8
Set (Query) the Low Pass Filter Slope to 6 (0), 12 (1), 18 (2) or 24 (3) dB/oct.
SYNC (?) {i}
6-8
Set (Query) the Synchronous Filter to Off (0) or On below 200 Hz (1).
OUTPUT and OFFSET
page
description
FOUT (?) i {, j}
6-9
Set (Query) the CH1 (i=1) or CH2 (2) Output Source to XY,R,
θ
,Trace 1, 2, 3, 4 (j=0...6).
OEXP (?) i {, x, j}
6-9
Set (Query) the X, Y, R (i=1,2,3) Offset to x percent and Expand to j. -105.00 ≤ x ≤ 105.00 and
1 ≤ j ≤ 256.
AOFF i
6-9
Auto Offset X, Y, R (i=1,2,3).
TRACE and SCAN
page
description
TRCD (?) i {, j, k, l, m}
6-10
Set (Query) the Definition of Trace i (1-4) to j•k/l and Store (m=1) or Not Store (0). j, k, l select
1, X, Y, R, q, Xn, Yn, Rn, Aux 1, Aux 2, Aux 3, Aux4, or F (j,k,l = 0...12). l=13-24 selects X
2
through F2.
SRAT (?) {i}
6-10
Set (Query) the Sample Rate to 62.5 mHz (0) through 512 Hz (13) or Trigger (14).
SLEN (?) {x}
6-10
Set (Query) the Scan Length to x seconds.
SEND (?) {i}
6-10
Set (Query) the Scan Mode to 1 Shot (0) or Loop (1).
TRIG
6-10
Software trigger command. Same as trigger input.
DISPLAY and SCALE
page
description
ASCL
6-11
Auto Scale the active display.
ADSP (?) {i}
6-11
Set (Query) the active display to Full (0), Top (1) or Bottom (2). Full screen is always active.
SMOD (?) {i}
6-11
Set (Query) the Screen Format to Single (0) or Up/Down dual (1) display mode.
MNTR (?) {i}
6-11
Set (Query) the Monitor Display to settings (0) or Input/Output (1).
DTYP (?) i {, j}
6-11
Set (Query) theFull (i=0), Top (i=1) or Bottom (i=2) Display Type to Polar (j=0), Blank (j=1),
Bar (j=2) or Chart (j=3).
SR850 DSP Lock-In Amplifier
SR850 DSP Lock-In Amplifier
1-8
DTRC (?) i {, j}
6-11
Set (Query) theFull (i=0), Top (i=1) or Bottom (i=2) Display Trace to trace j (1,2,3,4).
DSCL (?) {x}
6-11
Set (Query) theFull (i=0), Top (i=1) or Bottom (i=2) Display Range to x.
DOFF (?) {x}
6-11
Set (Query) theFull (i=0), Top (i=1) or Bottom (i=2) Display Center value to x.
DHZS (?) {i}
6-12
Set (Query) theFull (i=0), Top (i=1) or Bottom (i=2) Display Horizontal Scale to 2 ms (0) through
200 ks (32) per div.
RBIN? i
6-12
Query the bin number at the right edge of the Full (i=0), Top (i=1) or Bottom (i=2) chart display.
CURSOR
page
description
CSEK (?) {i}
6-13
Set (Query) the active display Cursor Seek mode to Max (0), Min (1) or Mean (2).
CWID (?) {i}
6-13
Set (Query) the active display Cursor Width to Off (0), Narrow (1), Wide (2) or Spot (3).
CDIV (?) {i}
6-13
Set (Query) the active display Chart Divisions to 8 (0), 10 (1) or None (2).
CLNK (?) {i}
6-13
Set (Query) the Cursor Control Mode to Linked (0) or Separate (1).
CDSP (?) {i}
6-13
Set (Query) the active display Cursor Readout to Delay (0), Bin (1), Fsweep (2) or Time (3).
CMAX
6-13
Move active chart cursor to max or min. Same as pressing [CURSOR MAX/MIN] key.
CURS? i
6-13
Query the cursor horz,vert position of Full (0), Top (1) or Bottom (2) chart display.
CBIN (?) {i}
6-13
Set (Query) the center of the cursor region in the active chart display. i is the bin number.
MARK
page
description
MARK
6-14
Places a mark in the data buffer. Same as pressing [MARK] key.
MDEL
6-14
Delete the nearest mark to the left of the cursor. Same as pressing <Marker Delete> softkey.
CNXT
6-14
Move active chart cursor to next mark to the right.
CPRV
6-14
Move active chart cursor to next mark to the left.
MACT?
6-14
Query the number of active marks. Also returns the active mark numbers.
MBIN? i
6-14
Query the bin number of mark #i.
MTXT (?) i {,s}
6-14
Set (Query) the label text for mark #i.
AUX INPUT/OUTPUT
page
description
OAUX ? i
6-15
Query the value of Aux Input i (1,2,3,4).
AUXM(?) i{, j}
6-15
Set (Query) the Output Mode of Aux Output i (1,2,3,4). j selects Fixed (0), Log (1) or Linear (2).
AUXV (?) i {, x}
6-15
Set (Query) voltage of Aux Output i (1,2,3,4) to x Volts. -10.500 ≤ x ≤ 10.500. Fixed Output
Mode only.
SAUX (?) i {, x, y, z}
6-15
Set (Query) the Aux Output i (1,2,3,4) Sweep Limits to Start (x), Stop (y) and Offset (z) voltages. 0.001 ≤ x,y ≤ 21.000 and
-10.500
≤ z ≤ 10.500.
TSTR (?) {i}
6-15
Set (Query) the Trigger Starts Scan? mode to No (0) or Yes (1).
MATH
page
description
SMTH i
6-16
Smooth the data within the active chart using 5 (0), 11 (1), 17 (2), 21 (3), 25 (4) point width.
COPR (?) {i}
6-16
Set (Query) the Calculator Operation to +, -, x, /, sin, cos, tan, √x, x
2
, log, 10x (i=0...10).
CALC
6-17
Do the Calculation selected by COPR with the argument set by CTRC or CARG.
CAGT (?) {i}
6-17
Set (Query) the Calculation Argument Type to Trace (0) or Constant (1).
CTRC (?) {i}
6-17
Set (Query) the Trace Argument to Trace i (1,2,3,4).
CARG (?) {x}
6-17
Set (Query) the Constant Argument value to x.
FTYP (?) {i}
6-17
Set (Query) the Fit Type to Linear (0), Exponential (1) or Gaussian (2).
FITT i, j
6-17
Fit the data within the chart region between i% and j% from the left edge. 0 ≤ i,j ≤ 100.
PARS ? i
6-17
Query the fit parameters a (0), b (1), c (2) or t0 (3).
STAT i, j
6-17
Statistically analyze the data within the chart region between i% and j% from the left edge.
0 ≤ i,j ≤ 100.
SPAR ? i
6-17
Query the Statistical results mean (0), standard dev (1), total (2) or delta time (3).
STORE AND RECALL FILE
page
description
FNAM (?) {s}
6-18
Set (Query) the current File Name to string s.
SDAT
6-18
Save the Active Display's Trace Data to the file specified by FNAM.
SASC
6-18
Save the Active Display's Trace Data in ASCII format to the file specified by FNAM.
SR850 DSP Lock-In Amplifier
1-9
SSET
6-18
Save the Settings to the file specified by FNAM.
RDAT
6-18
Recall the Trace Data from the file specified by FNAM to the active display's trace buffer.
RSET
6-18
Recall the Settings from the file specified by FNAM.
SETUP
page
description
OUTX (?) {i}
6-19
Set (Query) the Output Interface to RS232 (0) or GPIB (1).
OVRM (?) {i}
6-19
Set (Query) the GPIB Overide Remote state to Off (0) or On (1).
KCLK (?) {i}
6-19
Set (Query) the Key Click to Off (0) or On (1).
ALRM (?) {i}
6-19
Set (Query) the Alarms to Off (0) or On (1).
THRS (?) {i}
6-19
Set (Query) the Hours to 0≤ i ≤ 23.
TMIN (?) {i}
6-19
Set (Query) the Minutes to 0 ≤ i ≤ 59.
TSEC (?) {i}
6-19
Set (Query) the Seconds to 0 ≤ i ≤ 59.
DMTH (?) {i}
6-19
Set (Query) the Month to 1 ≤ 1 ≤ 12.
DDAY (?) {i}
6-19
Set (Query) the Day to 1 ≤ 1 ≤ 31.
DYRS (?) {i}
6-19
Set (Query) the Year to 0 ≤ 1 ≤ 99.
PLTM (?) {i}
6-19
Set (Query) the Plotter Mode to RS232 (0) or GPIB (1).
PLTB (?) {i}
6-19
Set (Query) the Plotter Baud Rate to 300 (0), 1200 (1), 2400 (2), 4800 (3), 9600 (4).
PLTA (?) {i}
6-19
Set (Query) the Plotter GPIB Address to 0 ≤ i ≤ 30.
PLTS (?) {i}
6-19
Set (Query) the Plot Speed to Fast (0) or Slow (1).
PNTR (?) {i}
6-19
Set (Query) the Trace Pen Number to 1 ≤ i ≤ 6.
PNGD (?) {i}
6-20
Set (Query) the Grid Pen Number to 1 ≤ i ≤ 6.
PNAL (?) {i}
6-20
Set (Query) the Alphanumeric Pen Number to 1 ≤ i ≤ 6.
PNCR (?) {i}
6-20
Set (Query) the Cursor Pen Number to 1 ≤ i ≤ 6.
PRNT (?) {i}
6-20
Set (Query) the Printer Type to Epson (0), HP (1) or File (2).
PRINT AND PLOT
page
description
PRSC
6-21
Print the screen. Same as the [PRINT] key.
PALL
6-21
Plot the display(s).
PTRC
6-21
Plot the trace(s) only.
PCUR
6-21
Plot the cursor(s) only.
FRONT PANEL CONTROLS
AUTO FUNCTIONS
page
description
STRT
6-22
Start or continue a scan. Same as [START/CONT] key.
PAUS
6-22
Pause a scan. Does not reset a paused or done scan.
REST
6-22
Reset the scan. All stored data is lost.
ASCL
6-11
Auto Scale the active display.
ATRC (?) {i}
6-22
Set (Query) the active display to Top (0) or Bottom (1). Full screen is always active.
AGAN
6-22
Auto Gain function. Same as pressing the [AUTO GAIN] key.
ARSV
6-22
Auto Reserve function. Same as pressing the [AUTO RESERVE] key.
APHS
6-22
Auto Phase function. Same as pressing the [AUTO PHASE] key.
AOFF i
6-22
Auto Offset X,Y or R (i=1,2,3).
CMAX
6-22
Move Cursor to Max or Min. Same as pressing the [CURSOR MAX/MIN] key.
DATA TRANSFER
page
description
OUTP? i
6-23
Query the value of X (1), Y (2), R (3) or
θ
(4). Returns ASCII floating point value.
OUTR? i
6-23
Query the value of Trace i (1,2,3,4). Returns ASCII floating point value.
SNAP?i,j{k,
l,m,n}
6-23
Records the values of 2,3,4,5, or 6 parameters at a single instant in time.
OAUX? i
6-23
Query the value of Aux Input i (1,2,3,4). Returns ASCII floating point value.
SPTS? i
6-23
Query the number of points stored in Trace i (1,2,3,4).
TRCA? i,j,k
6-23
Read k≥1 points starting at bin j≥0 from trace i (1,2,3,4) in ASCII floating point.
TRCB? i,j,k
6-23
Read k≥1 points starting at bin j≥0 from trace i (1,2,3,4) in IEEE binary floating point.
TRCL? i,j,k
6-24
Read k≥1 points starting at bin j≥0 from trace i (1,2,3,4) in non-normalized binary floating point.
FAST (?) {i}
6-25
Set (Query) Fast Data Transfer Mode On (2) for Windows programs, On (1) for DOS programs,
or Off (0). On will transfer binary X and Y every sample during a scan over the GPIB interface.
SR850 DSP Lock-In Amplifier
1-10
SERIAL POLL STATUS BYTE
(6-28)
bit
name
usage
0
SCN
No data is being acquired
1
IFC
No command execution in progress
2
ERR
Unmasked bit in error status byte set
3
LIA
Unmasked bit in LIA status byte set
4
MAV
The interface output buffer is non-empty
5
ESB
Unmasked bit in standard status byte set
6
SRQ
SRQ (service request) has occurred
7
Unused
STANDARD EVENT STATUS BYTE
(6-29)
bit
name
usage
0
INP
Set on input queue overflow
1
Unused
2
QRY
Set on output queue overflow
3
Unused
4
EXE
Set when command execution error occurs
5
CMD
Set when an illegal command is received
6
URQ
Set by any key press or knob rotation
7
PON
Set by power-on
LIA STATUS BYTE
(6-29)
bit
name
usage
0
RESRV
Set when a RESRV overload is detected
1
FILTR
Set when a FILTR overload is detected
2
OUTPT
Set when a OUTPT overload is detected
3
UNLK
Set when reference unlock is detected
4
RANGE
Set when detection freq crosses 200 Hz
5
TC
Set when time constant is changed
6
TRIG
Set when unit is triggered
7
PLOT
Set when a plot is completed
ERROR STATUS BYTE
(6-30)
bit
name
usage
0
Prn/Plt Err
Set when an printing or plotting error occurs
1
Backup Error
Set when battery backup fails
2
RAM Error
Set when RAM Memory test finds an error
3
Disk Error
Set when a disk error occurs
4
ROM Error
Set when ROM Memory test finds an error
5
GPIB Error
Set when GPIB binary data transfer aborts
6
DSP Error
Set when DSP test finds an error
7
Math Error
Set when an internal math error occurs
STATUS BYTE DEFINITIONS
STRD
6-25
Start a scan after 0.5sec delay. Use with Fast Data Transfer Mode.
INTERFACE
page
description
*RST
6-26
Reset the unit to its default configurations.
*IDN?
6-26
Read the SR850 device identification string.
LOCL(?) {i}
6-26
Set (Query) the Local/Remote state to LOCAL (0), REMOTE (1), or LOCAL LOCKOUT (2).
OVRM (?) {i}
6-26
Set (Query) the GPIB Overide Remote state to Off (0) or On (1).
TRIG
6-26
Software trigger command. Same as trigger input.
STATUS
page
description
*CLS
6-27
Clear all status bytes.
*ESE (?) {i} {,j}
6-27
Set (Query) the Standard Event Status Byte Enable Register to the decimal value i (0-255).
*ESE i,j sets bit i (0-7) to j (0 or 1). *ESE? queries the byte. *ESE?i queries only bit i.
*ESR? {i}
6-27
Query the Standard Event Status Byte. If i is included, only bit i is queried.
*SRE (?) {i} {,j}
6-27
Set (Query) the Serial Poll Enable Register to the decimal value i (0-255). *SRE i,j sets bit i (0-
7) to j (0 or 1). *SRE? queries the byte, *SRE?i queries only bit i.
*STB? {i}
6-27
Query the Serial Poll Status Byte. If i is included, only bit i is queried.
*PSC (?) {i}
6-27
Set (Query) the Power On Status Clear bit to Set (1) or Clear (0).
ERRE (?) {i} {,j}
6-27
Set (Query) the Error Status Enable Register to the decimal value i (0-255). ERRE i,j sets bit i
(0-7) to j (0 or 1). ERRE? queries the byte, ERRE?i queries only bit i.
ERRS? {i}
6-27
Query the Error Status Byte. If i is included, only bit i is queried.
LIAE (?) {i} {,j}
6-27
Set (Query) the LIA Status Enable Register to the decimal value i (0-255). LIAE i,j sets
bit i (0-7) to j (0 or 1). LIAE? queries the byte, LIAE?i queries only bit i.
LIAS? {i}
6-27
Query the LIA Status Byte. If i is included, only bit i is queried.
GETTING STARTED
The sample measurements described in this section are designed to acquaint the first time user with the
SR850 DSP Lock-In Amplifier. Do not be concerned that your measurements do not exactly agree with these
exercises. The focus of these measurement exercises is to learn how to use the instrument.
It is highly recommended that the first time user step through some or all of these exercises before attempting
to perform an actual experiment.
The experimental procedures are detailed in two columns. The left column lists the actual steps in the experiment. The right column is an explanation of each step.
Key Types
There are two types of front panel keys which will be referred to in this
manual. Hardkeys are those keys with labels printed on them. Their function is determined by the label and does not change. Hardkeys are referenced by brackets like this - [HARDKEY]. The softkeys are the six gray
keys along the right edge of the screen. Their function is labelled by a
menu box displayed on the screen next to the key. Softkey functions
change depending upon the situation. Softkeys will be referred to as the
<Soft Key> or simply the Soft Key.
Hardkeys
The keypad consists of five groups of hardkeys. The ENTRY keys are
used to enter numeric parameters which have been highlighted by a softkey. The MENU keys select a menu of softkeys. Pressing a menu key
will change the menu boxes which are displayed next to the softkeys.
Each menu groups together similar parameters and functions. The
CONTROL keys start and stop actual data acquisition, select the cursor
and toggle the active display. These keys are not in a menu since they
are used frequently and while displaying any menu. The SYSTEM keys
output the screen to a printer and display help messages. These keys
can also be accessed from any menu. The AUTO keys perform auto
functions such as Auto Gain and Auto Phase.
Softkeys
The SR850 has a menu driven user interface. The 6 softkeys to the right
of the video display have different functions depending upon the information displayed in the menu boxes at the right of the video display. In general, the softkeys have two uses. The first is to toggle a feature on and
off or to choose between settings. The second is to highlight a parameter
which is then changed using the knob or numeric keypad. In both cases,
the softkey selects the parameter which is displayed adjacent to it.
Knob
The knob is used to adjust parameters which have been highlighted by a
softkey. Many numeric entry fields may be adjusted with the knob. In
addition, many parameters are adjusted only with the knob. These are
typically parameters with a limited set of values, such as sensitivity or
time constant. In these cases, the parameter is selected by a softkey.
The [CURSOR] key will set the knob function to scrolling the cursor
within the active chart display.
2-1
YOUR FIRST MEASUREMENTS
Getting Started
2-2
The input impedance of the lock-in is 10 MΩ. The
Sine Out has an output impedance of 50Ω. Since
The Basic Lock-in
THE BASIC LOCK-IN
This measurement is designed to use the internal oscillator to explore some of the basic lock-in functions.
You will need BNC cables.
Specifically, you will measure the amplitude of the Sine Out at various frequencies, sensitivities, time constants and phase shifts. The "normal" lock-in display will be used throughout this exercise.
1. Disconnect all cables from the lock-in. Turn
the power on while holding down the [
(backspace) key. Wait until the power-on tests
are completed.
2. Connect the Sine Out on the front panel to the
A input using a BNC cable.
←]
When the power is turned on with the backspace
key depressed, the lock-in returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The display is the "normal" lock-in display. The
lock-in setup is displayed across the top of the
screen. The sensitivity, reserve, time constant,
prefilters and input configuration are all easily visible. Watch how these indicators change as you
change parameters. The upper numeric readout
and bar graph shows the value of X (Rcosθ) and
the lower graph shows the the value of Y (Rsinθ).
the Sine Output amplitude is specified into a high
impedance load, the output impedance does not
affect the amplitude.
The lock-in defaults to the internal oscillator reference set at 1.000 kHz. The reference mode (Intrnl)
and frequency are displayed at the bottom of the
screen. In this mode, the lock-in generates a synchronous sine output at the internal reference
frequency.
3. Press [AUTO PHASE]
4. Press [REF/PHASE]
The sine amplitude is 1.000 Vrms and the
sensitivity is 1 V(rms). Since the phase shift of the
sine output is very close to zero, the upper display
(X) should read close to 1.000 V and the lower display (Y) should read close to 0.000 V.
Automatically adjust the reference phase shift to
eliminate any residual phase error. This should set
the value of Y to zero.
Display the Reference and Phase menu. The
phase shift (displayed in the top menu box) should
be close to zero.
2-3
5. Press the <Rotate 90 deg> softkey.
50 Ω load).
The Basic Lock-in
This adds 90° to the reference phase shift. The
value of X drops to zero and Y becomes minus the
magnitude (-1.000 V).
Press the <deg.> softkey.
Use the knob to adjust the phase shift until Y
Highlight the phase shift.
The knob can be used to adjust parameters which
are continuous, such as phase, amplitude and frequency. The final phase value should be close to
zero again.
Phase shifts can also be entered numerically.
The knob is useful for small adjustments while
numeric entry is easier when changing to a precise value or by a large amount.
Use the Auto Phase function to return Y to zero
and X to the amplitude.
Highlight the internal reference frequency menu.
Enter 12.345 kHz in exponential form. The meas-
ured signal amplitude should stay within 1% of 1 V
and the phase shift should stay close to zero (the
value of Y should stay close to zero).
Parameters can be entered in real or integer form
as well. In this case, the frequency is changed to
1.000 kHz.
7. Press <Sine Output>
Use the knob to adjust the amplitude.
Press [.] [0] [1] [ENTER]
8. Press [GAIN/TC]
Press [AUTO GAIN]
The internal oscillator is crystal synthesized with
25 ppm of frequency error. The frequency can be
set with 5 digit or 0.1 mHz resolution, whichever is
greater.
Highlight the sine output amplitude.
As the amplitude is changed, the measured value
of X should equal the sine output amplitude.
The amplitude may be entered numerically also.
The sine amplitude can be set from 4 mV to 5 V
rms into high impedance (half the amplitude into a
Display the Gain and Time Constant menu.
The Auto Gain function will adjust the sensitivity so
2-4
9. Press <Sensitivity>
Use the knob to change the sensitivity to
50 mV.
Change the sensitivity back to 20 mV.
10. Press <Time Constant>
Use the knob to change the time constant to
300 µs.
Change the time constant to 3 ms.
The Basic Lock-in
that the measured magnitude (R) is a sizable percentage of full scale.
Highlight the full scale sensitivity.
Parameters which are discrete values, such as
sensitivity and time constant, can only be changed
with the knob. Numeric entry is not allowed for
these parameters.
Highlight the time constant.
The values of X and Y become noisy. This is
because the 2f component of the output (at 2 kHz)
is no longer attenuated completely by the low pass
filters.
Let's leave the time constant short and change the
filter slope.
11. Press the <Filter db/oct.> softkey until 6 dB/oct
is selected.
Press <Filter db/oct.> again to select
12 dB/oct.
Press <Filter db/oct.> twice to select
24 db/oct.
Press <Filter db/oct> again to select 6 db/oct.
12. Press [REF/PHASE]
Press <Ref. Frequency>
Press [5] [0] [ENTER]
Parameters which have their available options displayed within the menu box are selected by pressing the corresponding softkey until the desired
option is chosen.
The X and Y outputs are somewhat noisy at this
short time constant and only 1 pole of low pass
filtering.
The outputs are less noisy with 2 poles of filtering.
With 4 poles of low pass filtering, even this short
time constant attenuates the 2f component reasonably well and provides steady readings.
Let's leave the filtering short and the outputs noisy
for now.
Display the Reference and Phase menu.
Highlight the internal reference frequency.
Enter 50 Hz for the reference frequency. With a
3 ms time constant and only 6 db/oct of filtering,
the output is totally dominated by the 2f component at 50 Hz.
2-5
The Basic Lock-in
13. Press [GAIN/TC]
Press <Synchronous> to select <200 Hz.
Display the Gain and Time Constant menu again.
This turns on synchronous filtering whenever the
detection frequency is below 200 Hz.
Synchronous filtering effectively removes output
components at multiples of the detection frequency. At low frequencies, this filter is a very effective
way to remove 2f without requiring extremely long
time constants.
The outputs are now very quiet and steady, even
though the time constant is very short. The
response time of the synchronous filter is equal to
the period of the detection frequency (20 ms in this
case).
This concludes this measurement example. You
should have a feeling for the basic operation of the
menus, knob and numeric entry. Basic lock-in
parameters have been introduced and you should
be able to perform simple measurements.
2-6
The input impedance of the lock-in is 10 MΩ. The
tors have either a 50Ω or 600Ω output impedance.
Displays and Traces
DISPLAYS and TRACES
This measurement is designed to use the internal oscillator and an external signal source to explore some of
the display types. You will need a synthesized function generator capable of providing a 100 mVrms sine
wave at 1.000 kHz (the DS345 from SRS will suffice), BNC cables and a terminator appropriate for the generator function output.
Specifically, you will display the lock-in outputs when measuring a signal close to, but not equal to, the internal reference frequency. This setup ensures changing outputs which are more illustrative than steady outputs.
The displays will be configured to show X, Y, R and θ in bar graph and polar formats. The example Scans and
Sweeps demonstrates the use of the chart graph.
1. Disconnect all cables from the lock-in. Turn
the power on while holding down the [←]
(backspace) key. Wait until the power-on tests
are completed.
2. Turn on the function generator, set the fre-
quency to 1.0000 kHz (exactly) and the amplitude to 500 mVrms.
Connect the function output (sine wave) from
the synthesized function generator to the A
input using a BNC cable and appropriate
terminator.
When the power is turned on with the backspace
key depressed, the lock-in returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The display is the "normal" lock-in display. The
lock-in setup is displayed across the top of the
screen. The sensitivity, reserve, time constant,
prefilters and input configuration are all easily visible. Watch how these indicators change as you
change parameters. The upper numeric readout
and bar graph shows the value of X (Rcosθ) and
the lower graph shows the the value of Y (Rsinθ).
generator may require a terminator. Many generaUse the appropriate feedthrough or T termination if
necessary. In general, not using a terminator
means that the function output amplitude will not
agree with the generator setting.
The lock-in defaults to the internal oscillator reference set at 1.000 kHz. The reference mode (Intrnl)
and frequency are displayed at the bottom of the
screen. In this mode, the internal oscillator sets
the detection frequency.
3. Press [REF/PHASE]
The internal oscillator is crystal synthesized so
that the actual reference frequency should be very
close to the actual generator frequency. The X and
Y displays should read values which change very
slowly. The lock-in and the generator are not
phase locked but they are at the same frequency
with some slowly changing phase.
Display the Reference and Phase menu.
2-7
Displays and Traces
Highlight the internal oscillator frequency.
By setting the lock-in reference 0.2 Hz away from
the signal frequency, the X and Y outputs are
0.2 Hz sine waves (difference between reference
and signal frequency). The X and Y output displays should now oscillate at about 0.2 Hz (the
accuracy is determined by the crystals of the generator and the lock-in).
Display the Display and Scale menu.
The SR850 collects data in the form of traces.
There are 4 definable traces and only these trace
quantities may be displayed. The default definition
of these traces is X, Y, R and θ for traces 1, 2, 3
and 4.
The Display Scale menu box shows the display
parameters for the Full (screen), Top or Bottom
(split screen) displays. In this case, the Top display parameters are shown.
Each display shows one of the data traces. The
Top display defaults to showing trace 1 which has
a default definition of X. Thus the top bar graph
shows the X output.
Trace 3 has a default definition of R so showing
trace 3 on the top graph will display the quantity R.
R is phase independent so it shows a steady value
(close to 0.500 V).
[AUTO SCALE] automatically scales the active
display. The top display is the active display (as
indicated by the inverse trace identifier at the
upper left of the display).
To modify the bottom graph, you must display the
bottom graph's parameters in the Display Scale
menu box. This also makes the bottom display the
active display (for autoscaling). The trace indicator
(at the upper left of each display) is highlighted on
the active display.
The bottom display defaults to trace 2 (Y).
Trace 4 is θ. The phase between the reference
and the signal changes by 360° every 5 sec
(0.2 Hz difference frequency).
Press <Ref. Frequency>
Use the knob to change the frequency to
999.80 Hz.
4.
Press [DISPLAY/SCALE]
Press <Type/Trace> twice to highlight the displayed trace number.
Use the knob to change the trace number to 3.
Press [AUTO SCALE]
5.
Press <Full, Top or Bottom> to select Bottom.
Press <Type/Trace> twice to highlight the displayed trace number.
Use the knob to change the trace number to 4.
2-8
Displays and Traces
[AUTO SCALE] automatically scales the active
display. In this case, the trace data is moving and
autoscaling may not do a very satisfactory job.
To manually set the graph scale, you set the range
(±) and center value (@). The graph displays a
scale equal to the center value plus and minus the
range.
In this case, set the bar graph to ±180°. The bar
graph should be a linear phase ramp at 0.2 Hz.
The monitor display at the top of the screen monitors either the lock-in settings (sensitivity, time
constant, etc.) or the measured lock-in inputs and
outputs (X, Y, R, θ and Aux In 1-4).
The Input/Output monitor allows you to see all of
the measured quantities, even if they are not
shown on the larger displays.
The screen is now setup for a single display. The
default display type for the full screen display is a
polar graph.
The polar graph plots the quantities X and Y on an
X-Y axis. The resulting vector has a length equal
to the magnitude R and has a phase angle relative
to the positive X axis equal to θ. In this case, since
the phase is rotating at the difference frequency,
the vector rotates at 0.2 Hz.
Display the Reference and Phase menu.
Highlight the internal oscillator frequency.
As the internal reference frequency gets closer to
the signal frequency, the rotation gets slower and
slower. If the frequencies are EXACTLY equal,
then the phase is constant.
By using the signal source as the external reference, the lock-in will phase lock its internal oscillator to the signal frequency and the phase will be a
constant.
Highlight the reference source.
Select external reference mode. The lock-in will
phase lock to the signal at the Reference Input.
Press [AUTO SCALE]
Press the <± Range> softkey. This is the fifth
softkey from the top.
Press [1] [8] [0] [ENTER]
6.
Press <Monitor> to select Input/Output.
7.
Press <Format> to select Single.
8.
Press [REF/PHASE]
Press <Ref. Frequency>
Use the knob to adjust the frequency slowly to
try to stop the rotation of the signal vector.
9.
Use a BNC cable to connect the TTL SYNC
output from the generator to the Reference
Input of the lock-in.
Press <Ref. Source>
Use the knob to select External.
2-9
With a TTL reference signal, the slope needs to be
set to either rising or falling edge.
The signal vector on the polar graph will not rotate
since the phase is a constant. The actual phase
depends upon the phase difference between the
function output and the sync output from the
generator.
The external reference frequency (as measured by
the lock-in) is displayed at the bottom of the
screen. The reference mode is shown as Ext+ for
external TTL rising edge. The LOCK indicator
should be on (successfully locked to the external
reference).
Display the Display and Scale menu.
Change the screen to dual display mode again.
Display the Trace and Scan menu.
Trace 3 is defined as R and is displayed on the top
graph. Let's change the definition of trace 3 to
something else.
Traces are defined as A•B/C. The quantities A, B,
and C are selected from the various quantities
measured by the lock-in.
Trace 3 has now been redefined to be X. The top
graph now displays X. The trace definition is
shown at the upper left of each graph.
Traces may be defined to be ratios and products
of 2 or 3 quantities.
Trace 3 is now defined as X/R and is equal to
cosθ, independent of the signal amplitude. The
traces can be defined to display the most useful
quantities for a given experiment. Trace data may
be stored in the data buffers. Scans and chart
graphs will be discussed in a later example.
Change the reference phase shift to check that
trace 3 displays cosθ.
Automatically adjust the measured phase shift to
zero. The top display should show cos0° or 1.
Highlight the phase shift.
Press <Ref. Slope> to select Rising Edge.
10.
Press [DISPLAY/SCALE]
Press <Format> to select Up/Down.
Press the second softkey, next to the trace
definition, to highlight the R.
Use the knob to change the A parameter from
R to X.
Press the second softkey to highlight the
denominator (C) of the trace definition.
Use the knob to change the C parameter from
1 to R.
12.
Press [REF/PHASE]
Press [AUTO PHASE]
Press <deg.>
Displays and Traces
2-10
Displays and Traces
Using the keypad, enter a phase shift which is
45° greater than the displayed phase shift.
At a measured phase shift of 45°, trace 3 should
equal cos45° or 0.707.
This concludes this measurement example. You
should have a feeling for the basic operation of the
display types and trace definitions.
2-11
Displays and Traces
2-12
The input impedance of the lock-in is 10 MΩ. The
Sine Out has an output impedance of 50Ω. Since
Outputs, Offsets and Expands
OUTPUTS, OFFSETS and EXPANDS
This measurement is designed to use the internal oscillator to explore some of the basic lock-in outputs. You
will need BNC cables and a digital voltmeter (DVM).
Specifically, you will measure the amplitude of the Sine Out and provide analog outputs proportional to the
measurement. The effect of offsets and expands on the displayed values and the analog outputs will be
explored.
1. Disconnect all cables from the lock-in. Turn
the power on while holding down the [←]
(backspace) key. Wait until the power-on tests
are completed.
2. Connect the Sine Out on the front panel to the
A input using a BNC cable.
When the power is turned on with the backspace
key depressed, the lock-in returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The display is the "normal" lock-in display. The
lock-in setup is displayed across the top of the
screen. The sensitivity, reserve, time constant,
prefilters and input configuration are all easily visible. Watch how these indicators change as you
change parameters. The upper numeric readout
and bar graph shows the value of X (Rcosθ) and
the lower graph shows the the value of Y (Rsinθ).
the Sine Output amplitude is specified into a high
impedance load, the output impedance does not
affect the amplitude.
The lock-in defaults to the internal oscillator reference set at 1.000 kHz. The reference mode (Intrnl)
and frequency are displayed at the bottom of the
screen. In this mode, the lock-in generates a synchronous sine output at the internal reference
frequency.
3. Connect the CH1 output on the front panel to
the DVM. Set the DVM to read DC Volts.
The sine amplitude is 1.000 Vrms and the
sensitivity is 1 V(rms). Since the phase shift of the
sine output is very close to zero, the upper display
(X) should read close to 1.000 V and the lower display (Y) should read close to 0.000 V.
The CH1 output defaults to X. The output voltage
is simply (X/Sensitivity - Offset)xExpandx10V. In
this case, X = 1.000 V, the sensitivity = 1 V, the
offset is zero percent and the expand is 1. The
output should thus be 10 V or 100% of full scale.
2-13
4. Press [REF/PHASE]
Outputs, Offsets and Expands
Display the Reference and Phase menu.
Press <Sine Output>
Press [.] [5] [ENTER]
5. Press [OUTPUT/OFFSET]
Press <Auto Offset>
Highlight the sine output amplitude.
Enter an amplitude of 0.5 V. The top display
should show X=0.5 V and the CH1 output should
be 5 V on the DVM.
Display the Output and Offset menu. This menu
chooses which measured parameters or traces
are output on CH1 and CH2. In addition, the X, Y
and R offsets and expands are programmed in this
menu.
The Offset & Expand menu box displays the offset
and expand of either X, Y or R. In this case, the X
offset and expand is displayed. The <X,Y,R> soft-
key selects the which offset and expand is
displayed.
Auto Offset automatically adjusts the Xoffset (or Y
or R) such that X (or Y or R) becomes zero. The
offset should be about 50% in this case. Offsets
are useful for making relative measurements. In
analog lock-ins, offsets were generally used to
remove DC output errors from the lock-in itself.
The SR850 has no DC output errors and the offset
is not required for most measurements.
Press <Offset>
Press [4] [0] [ENTER]
Press <Expand>
Set the expand to 10 using the knob or the
numeric entry keys.
The offset affects both the displayed value of X
and any analog output proportional to X. The CH1
output should be zero in this case.
The highlighted OFFST indicator turns on at the
bottom left of the top display to indicate that the
displayed trace is affected by an offset.
Highlight the X offset.
Enter an offset of 40% of full scale. The output off-
sets are a percentage of full scale. The percent-
age does not change with the sensitivity. The dis-
played value of X should be 0.100 V (0.5 V - 40%
of full scale). The CH1 output voltage is
(X/Sensitivity - Offset)xExpandx10V.
CH1 Out = (0.5/1.0 - 0.4)x1x10V = 1 V
Highlight the X expand.
With an expand of 10, the display has one more
digit of resolution (100.XX mV).
2-14
Outputs, Offsets and Expands
The highlighted EXPD indicator turns on at the
bottom left of the top display to indicate that the
displayed trace is affected by an expand.
The CH1 output is
(X/Sensitivity - Offset)xExpandx10V. In this case,
the output voltage is
CH1 Out = (0.5/1.0 - 0.4)x10x10V = 10V
The expand allows the output gain to be increased
by up to 256. The output voltage is limited to
10.9 V and any output which tries to be greater will
turn on the OUTPT overload indicator at the
bottom left of the screen.
With offset and expand, the output voltage gain
and offset can be programmed to provide control
of feedback signals with the proper bias and gain
for a variety of situations.
Offsets do add and subtract from the displayed
values while expand increases the resolution of
the display.
6. Connect the DVM to the X output on the rear
panel.
7. Connect the DVM to the CH1 output on the
front panel again.
Press <Expand>
Press [1] [ENTER]
The X and Y outputs on the rear panel always provide voltages proportional to X and Y (with offset
and expand). The X output voltage should be
10 V, just like the CH1 output.
The front panel outputs can be configured to
output different traces quantities while the rear
panel outputs always output X and Y.
Outputs proportional to X and Y (rear panel, CH1
or CH2) have 100 kHz of bandwidth. The CH1 and
CH2 outputs, when configured to be proportional
to R, θ, or a trace (even a trace defined as X or Y)
are updated at 512 Hz and have a 200 Hz bandwidth. It is important to keep this in mind if you use
very short time constants.
Let's change CH1 to a trace.
First, set the X expand back to 1.
Press <Offset>
Press [0] [ENTER]
Set the X offset back to 0.0%.
X Should be 0.500 V again and the CH1 output
should be 5.0 V.
2-15
Outputs, Offsets and Expands
Press <Source>
Use the knob to select Trace1.
8. Press [TRACE/SCAN]
Press the second softkey, next to the trace
definition, to highlight the X.
Use the knob to change the numerator from X
to 1.
Press the second softkey twice to highlight the
denominator (C) of the trace definition.
Use the knob to change the denominator from
1 to X.
Highlight the CH1 source. The CH1 output is pro-
portional to this source.
CH1 can be proportional to X, R, θ, or Trace 1-4.
Choose Trace 1. Trace 1 has a default definition of
X so the CH1 output should remain 5.0 V (but its
bandwidth is only 200 Hz instead of 100 kHz).
Display the Trace and Scan menu.
Traces are defined as A•B/C. The quantities A, B,
and C are selected from the various quantities
measured by the lock-in.
Trace 1 is defined as X by default. Let's change it
to 1/X.
Trace 1 is now 1•1/1 and the top display shows
1.000
Change the denominator.
Trace 1 is now defined as 1/X. The top display
shows Trace 1. The trace definition is shown at
the upper left of the top display. The trace units
are shown at the bottom center of the top display
(1/V).
Remember, X was 0.5V. Thus, 1/X is 1/0.5 = 2.0
(1/V). The display should show 2.0 (or very close).
Displays use the actual measured quantities to
calculate the value of a trace. If X was 5 mV, the
value of Trace 1 would be 1/5 mV or 200 (1/V).
Traces are calculated using Volts, degrees, and
Hz for the units of A, B and C.
The CH1 output voltage is 0.2V. This is because
trace output voltages are calculated using the
output voltages of the A, B and C quantities rather
than their displayed values. In this case, X=0.5V.
As an analog output voltage, this would be 5.0 V
(1/2 scale of 1V full scale sensitivity). The 1/X
output voltage is 1/5.0V or 0.2 V.
See the DC Outputs and Scaling discussion in the
Lock-In Basics section for more detailed
information.
2-16
The input impedance of the lock-in is 10 MΩ. The
Sine Out has an output impedance of 50Ω. Since
Scans and Sweeps
SCANS and SWEEPS
This measurement is designed to use the internal oscillator to explore some of the basic lock-in functions.
You will need BNC cables.
Specifically, you will measure the response of the line notch filters by sweeping the internal reference frequency and measuring the sine output. Traces and strip chart displays will be used to record X, Y, R and θ as the
signal is swept through the input notch filters.
1. Disconnect all cables from the lock-in. Turn
the power on while holding down the [←]
(backspace) key. Wait until the power-on tests
are completed.
2. Connect the Sine Out on the front panel to the
A input using a BNC cable.
When the power is turned on with the backspace
key depressed, the lock-in returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The display is the "normal" lock-in display. The
lock-in setup is displayed across the top of the
screen. The sensitivity, reserve, time constant,
prefilters and input configuration are all easily visible. Watch how these indicators change as you
change parameters. The upper numeric readout
and bar graph shows the value of X (Rcosθ) and
the lower graph shows the the value of Y (Rsinθ).
the Sine Output amplitude is specified into a high
impedance load, the output impedance does not
affect the amplitude.
The lock-in defaults to the internal oscillator reference set at 1.000 kHz. The reference mode (Intrnl)
and frequency are displayed at the bottom of the
screen. In this mode, the lock-in generates a synchronous sine output at the internal reference
frequency.
3. Press [INPUT/FILTERS]
Press the <Line Notches> softkey until Both filters are selected.
The sine amplitude is 1.000 Vrms and the
sensitivity is 1 V(rms). Since the phase shift of the
sine output is very close to zero, the upper display
(X) should read close to 1.000 V and the lower display (Y) should read close to 0.000 V.
Display the Input and Filters menu. This menu
allows the input configuration to be changed.
With the line notch filters engaged, signal inputs at
60 (50) Hz and 120 (100) Hz are removed. Note
that the line filter indicators at the top of the screen
are both on.
2-17
4. Press [DISPLAY/SCALE]
Scans and Sweeps
Display the Display and Scale menu. This menu
configures and scales the different screen displays
and graphs.
Press the <Type/Trace> softkey to select the
Trace number.
Use the knob to select Trace 3.
5. Press [REF/PHASE]
Press <Ref. Source>
Use the knob to select Internal Sweep.
6. Press <Sweep Menu>
Press <Start>
Press [4] [0] [ENTER].
Highlight the trace number for the Top Bar graph.
The SR850 acquires and displays data in the form
of traces. The default definitions for the 4 traces
are X, Y, R and θ. These definitions may be
changed.
Display the magnitude R (Trace 3) on the top
graph.
Display the Reference and Phase menu.
Choose internal reference frequency sweep. In
this mode, the SR850 will sweep the internal oscillator from a start to a stop frequency.
Set the sweep start and stop frequencies in this
submenu. Sweeps may be linear or logarithmic. In
this case, let's use a linear sweep.
Highlight the start frequency.
Set the start frequency to 40 Hz. The reference
changes to 40.000 Hz as shown in the frequency
readout at the bottom center of the screen.
Press <Stop>
Press [1] [6] [0] [ENTER].
7. Press [GAIN/TC]
Press <Time Constant> and use the knob to
select 10 ms.
Press <Synchronous> to select <200 Hz.
8. Press [TRACE/SCAN]
Highlight the stop frequency.
Set the stop frequency to 160 Hz.
Display the Gain and Time Constant menu.
Choose a short time constant so that the frequen-
cy can be swept in a reasonably short time.
Since the sweep frequencies are below 200 Hz,
you can take advantage of the synchronous filter
to remove the 2f component of the output without
using a long time constant.
Display the Trace and Scan menu. This menu
allows the trace definitions to be changed and the
scan (sweep) to be configured. Trace data is sampled and stored in the buffers at the sample rate.
Swept parameters (reference frequency in this
case) are also updated at the sample rate (imme-
2-18
Scans and Sweeps
diately after the data is sampled). The scan time
sets the amount of time the buffer will store and
the length of any sweep.
In this measurement, let's leave the trace definitions equal to the defaults and just change the
sample rate and scan time.
Press <Sample Rate>
Use the knob to select 32 Hz.
Press <Scan Length>
Press [1] [0] [0] [ENTER]
Press <1 Shot/Loop> to select 1 Shot.
Highlight the sample rate.
The trace data will be sampled at 32 Hz and
stored in the buffer. After each data point is
recorded, the reference frequency will be updated.
This determines the resolution of our data along
the time axis.
Highlight the scan length.
Set the scan length to 100 seconds (1:40). This
configures the data buffer to hold 3200 data points
(32 Hz sample rate x 100 second scan length).
The scan length is also the sweep time for the
internal frequency sweep. Sweeps are always
coordinated with the data acquisition. In this case,
the sweep range is 120 Hz (40 Hz to 160 Hz) and
will take 100 seconds and be updated 3200 times.
Each stored data point will represent a frequency
increment of 0.0375 Hz (120/3200).
Scans can repeat (Loop) or stop when finished (1
Shot). Let's take a single scan of data.
Now you're finally ready to start the scan.
9. Press [START/CONT]
The [START/CONT] key starts the scan and
sweep. The scan indicator at the bottom left corner
of the screen switches from STOP to Run 1 (single
scan in progress). The reference frequency read-
out at the bottom center displays the frequency
while sweeping.
As the frequency passes through 60 (50) Hz and
120 (100) Hz, the value of R should drop close to
zero as the signal sweeps through each input
notch filter.
When the scan is complete, the scan indicator
switches to DONE and the frequency should be
160 Hz.
Now let's try displaying the data in a more mean-
ingful way.
2-19
Loading...
+ 249 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.