Safety and Preparation for Use1-3
Specifications1-5
Abridged Command List1-7
GETTING STARTED
Your First Measurements2-1
The Basic Lock-in2-3
X, Y, R and θ2-7
Outputs, Offsets and Expands2-9
Storing and Recalling Setups2-13
Aux Outputs and Inputs2-15
SR830 BASICS
What is a Lock-in Amplifier?3-1
What Does a Lock-in Measure?3-3
The SR830 Functional Diagram3-5
Reference Channel3-7
Phase Sensitive Detectors3-9
Time Constants and DC Gain3-11
DC Outputs and Scaling3-13
Dynamic Reserve3-15
Signal Input Amplifier and Filters3-17
Input Connections3-19
Intrinsic (Random) Noise Sources3-21
External Noise Sources3-23
Noise Measurements3-25
OPERATION
Power On/Off and Power On Tests4-1
Reset4-1
[Keys]4-1
Spin Knob4-1
Front Panel BNC Connectors4-2
Key Click On/Off4-2
Front Panel Display Test4-2
Display Off Operation4-2
Keypad Test4-3
Standard Settings4-4
FRONT PANEL
Signal Input and Filters4-5
Sensitivity, Reserve, Time Constants4-7
CH1 Display and Output4-12
CH2 Display and Output4-15
Reference4-18
Auto Functions4-21
Setup4-23
Interface4-24
Warning Messages4-26
REAR PANEL
Power Entry Module4-27
IEEE-488 Connector4-27
RS232 Connector4-27
Aux Inputs (A/D Inputs)4-27
Aux Outputs (D/A Outputs)4-27
X and Y Outputs4-27
Signal Monitor Output4-28
Trigger Input4-28
TTL Sync Output4-28
Preamp Connector4-28
Using SRS Preamps4-29
PROGRAMMING
GPIB Communications5-1
RS232 Communications5-1
Status Indicators and Queues5-1
Command Syntax5-1
Interface Ready and Status5-2
GET (Group Execute Trigger)5-2
DETAILED COMMAND LIST5-3
Reference and Phase5-4
Input and Filter5-5
Gain and Time Constant5-6
Display and Output5-8
Aux Input and Output5-9
Setup5-10
Auto Functions5-11
Data Storage5-12
Data Transfer5-15
Interface5-19
Status Reporting5-20
STATUS BYTE DEFINITIONS
Serial Poll Status Byte5-21
Service Requests5-22
Standard Event Status Byte5-22
LIA Status Byte5-23
Error Status Byte5-23
PROGRAM EXAMPLES
Microsoft C, Nationall Instr GPIB5-25
USING SR530 PROGRAMS5-31
Table of Contents
TESTING
Introduction6-1
Preset6-1
Serial Number6-1
Firmware Revision6-1
Test Record6-1
If A Test Fails6-1
Necessary Equipment6-1
Front Panel Display Test6-2
Keypad Test6-2
PERFORMANCE TESTS
Self Tests6-3
DC Offset6-5
Common Mode Rejection6-7
Amplitude Accuracy and Flatness6-9
Amplitude Linearity6-11
Frequency Accuracy6-13
Phase Accuracy6-15
Sine Output Amplitude6-17
DC Outputs and Inputs6-19
Input Noise6-21
Performance Test Record6-23
CIRCUITRY
Circuit Boards7-1
CPU and Power Supply Board7-3
DSP Logic Board7-5
Analog Input Board7-7
PARTS LISTS
DSP Logic Board7-9
Analog Input Board7-15
CPU and Power Supply Board7-21
Front Panel Display Boards7-24
Miscellaneous7-30
SCHEMATIC DIAGRAMS
CPU and Power Supply Board
Display Board
Keypad Board
DSP Logic Board
Analog Input Board
SAFETY AND PREPARATION FOR USE
WARNING
Dangerous voltages, capable of causing injury or death, are present in
this instrument. Use extreme caution whenever the instrument covers
are removed. Do not remove the covers while the unit is plugged into a
live outlet.
CAUTION
This instrument may be damaged if operated
with the LINE VOLTAGE SELECTOR set for the
wrong AC line voltage or if the wrong fuse is
installed.
LINE VOLTAGE SELECTION
The SR830 operates from a 100V, 120V, 220V, or
240V nominal AC power source having a line frequency of 50 or 60 Hz. Before connecting the power cord to a power source, verify that the LINE
VOLTAGE SELECTOR card, located in the rear
panel fuse holder, is set so that the correct AC input voltage value is visible.
Conversion to other AC input voltages requires a
change in the fuse holder voltage card position
and fuse value. Disconnect the power cord, open
the fuse holder cover door and rotate the fuse-pull
lever to remove the fuse. Remove the small printed circuit board and select the operating voltage
by orienting the printed circuit board so that the
desired voltage is visible when pushed firmly into
its slot. Rotate the fuse-pull lever back into its normal position and insert the correct fuse into the
fuse holder.
LINE FUSE
Verify that the correct line fuse is installed before
connecting the line cord. For 100V/120V, use a 1
Amp fuse and for 220V/240V, use a 1/2 Amp fuse.
LINE CORD
The SR830 has a detachable, three-wire power
cord for connection to the power source and to a
protective ground. The exposed metal parts of the
instrument are connected to the outlet ground to
protect against electrical shock. Always use an
outlet which has a properly connected protective
ground.
SERVICE
Do not attempt to service or adjust this instrument
unless another person, capable of providing first
aid or resuscitation, is present.
Do not install substitute parts or perform any unauthorized modifications to this instrument. Contact
the factory for instructions on how to return the instrument for authorized service and adjustment.
1-3
1-4
SR830 DSP LOCK-IN AMPLIFIER
SPECIFICATIONS
SIGNAL CHANNEL
Voltage InputsSingle-ended (A) or differential (A-B).
Current Input106 or 108 Volts/Amp.
Full Scale Sensitivity2 nV to 1 V in a 1-2-5-10 sequence (expand off).
Input ImpedanceVoltage: 10 MΩ+25 pF, AC or DC coupled.
Current: 1 kΩ to virtual ground.
Gain Accuracy±1% from 20°C to 30°C (notch filters off).
Input Noise6 nV/√Hz at 1 kHz (typical).
Signal Filters60 (50) Hz and 120(100) Hz notch filters (Q=4).
CMRR90 dB at 100 Hz (DC Coupled).
Dynamic ReserveGreater than 100 dB (with no signal filters).
Harmonic Distortion-80 dB.
REFERENCE CHANNEL
Frequency Range1 mHz to 102 kHz
Reference InputTTL (rising or falling edge) or Sine.
<0.1°/°C to 100 kHz
Harmonic DetectDetect at Nxf where N<19999 and Nxf<102 kHz.
Acquisition Time(2 cycles + 5 ms) or 40 ms, whichever is greater.
DEMODULATOR
Zero StabilityDigital displays have no zero drift on all dynamic reserves.
Analog outputs: <5 ppm/°C for all dynamic reserves.
Time Constants10 µs to 30 s (reference > 200 Hz). 6, 12, 18, 24 dB/oct rolloff.
up to 30000 s (reference < 200 Hz). 6, 12, 18, 24 dB/oct rolloff.
Synchronous filtering available below 200 Hz.
Harmonic Rejection-80 dB
INTERNAL OSCILLATOR
Frequency1 mHz to 102 kHz.
Frequency Accuracy25 ppm + 30 µHz
Frequency Resolution4 1/2 digits or 0.1 mHz, whichever is greater.
Distortionf<10 kHz, below -80 dBc. f>10 kHz, below -70 dBc.1 Vrms amplitude.
Output Impedance50 Ω
Amplitude 4 mVrms to 5 Vrms (into a high impedance load) with 2 mV resolution.
(2 mVrms to 2.5 Vrms into 50Ω load).
Amplitude Accuracy1%
Amplitude Stability50 ppm/°C
OutputsSine output on front panel. TTL sync output on rear panel.
When using an external reference, both outputs are phase locked to the
external reference.
1-5
SR830 DSP Lock-In Amplifier
DISPLAYS
Channel 14 1/2 digit LED display with 40 segment LED bar graph.
X, R, X Noise, Aux Input 1 or 2. The display can also be any of these
quantities divided by Aux Input 1 or 2.
Channel 2 4 1/2 digit LED display with 40 segment LED bar graph.
Y, θ, Y Noise, Aux Input 3 or 4. The display can also be any of these
quantities divided by Aux Input 3 or 4.
OffsetX, Y and R may be offset up to ±105% of full scale.
ExpandX, Y and R may be expanded by 10 or 100.
Reference4 1/2 digit LED display.
Display and modify reference frequency or phase, sine output amplitude,
harmonic detect, offset percentage (X, Y or R), or Aux Outputs 1-4.
Data Buffer16k points from both Channel 1 and Channel 2 display may be stored
internally. The internal data sample rate ranges from 512 Hz down to 1
point every 16 seconds. Samples can also be externally triggered. The data
buffer is accessible only over the computer interface.
INPUTS AND OUTPUTS
Channel 1 OutputOutput proportional to Channel 1 display, or X.
Output Voltage: ±10 V full scale. 10 mA max output current.
Channel 2 OutputOutput proportional to Channel 2 display, or Y.
Output Voltage: ±10 V full scale. 10 mA max output current.
X and Y OutputsRear panel outputs of cosine (X) and sine (Y) components.
Output Voltage: ±10 V full scale. 10 mA max output current.
Aux. Outputs4 BNC Digital to Analog outputs.
±10.5 V full scale, 1 mV resolution. 10 mA max output current.
Aux. Inputs4 BNC Analog to Digital inputs.
Differential inputs with1 MΩ input impedance on both shield and center
conductor. ±10.5 V full scale, 1 mV resolution.
Trigger InputTTL trigger input triggers stored data samples.
Monitor OutputAnalog output of signal amplifiers (before the demodulator).
GENERAL
InterfacesIEEE-488 and RS232 interfaces standard.
All instrument functions can be controlled through the IEEE-488 and RS232
interfaces.
Preamp PowerPower connector for SR550 and SR552 preamplifiers.
Power40 Watts, 100/120/220/240 VAC, 50/60 Hz.
Dimensions17"W x 5.25"H x 19.5"D
Weight30 lbs.
WarrantyOne year parts and labor on materials and workmanship.
1-6
SR830 DSP Lock-In Amplifier
COMMAND LIST
VARIABLESi,j,k,l,mIntegers
fFrequency (real)
x,y,zReal Numbers
sString
REFERENCE and PHASEpage description
PHAS (?) {x}5-4Set (Query) the Phase Shift to x degrees.
FMOD (?) {i}5-4Set (Query) the Reference Source to External (0) or Internal (1).
FREQ (?) {f} 5-4Set (Query) the Reference Frequency to f Hz.Set only in Internal reference mode.
RSLP (?) {i}5-4Set (Query) the External Reference Slope to Sine(0), TTL Rising (1), or TTL Falling (2).
HARM (?) {i} 5-4Set (Query) the Detection Harmonic to 1 ≤ i ≤ 19999 and i•f ≤ 102 kHz.
SLVL (?) {x}5-4Set (Query) the Sine Output Amplitude to x Vrms. 0.004 ≤ x ≤5.000.
INPUT and FILTERpage description
ISRC (?) {i}5-5Set (Query) the Input Configuration to A (0), A-B (1) , I (1 MΩ) (2) or I (100 MΩ) (3).
IGND (?) {i}5-5Set (Query) the Input Shield Grounding to Float (0) or Ground (1).
ICPL (?) {i}5-5Set (Query) the Input Coupling to AC (0) or DC (1).
ILIN (?) {i}5-5Set (Query) the Line Notch Filters to Out (0), Line In (1) , 2xLine In (2), or Both In (3).
GAIN and TIME CONSTANT page description
SENS (?) {i}5-6Set (Query) the Sensitivity to 2 nV (0) through 1 V (26) rms full scale.
RMOD (?) {i}5-6Set (Query) the Dynamic Reserve Mode to HighReserve (0), Normal (1), or Low Noise (2).
OFLT (?) {i}5-6Set (Query) the Time Constant to 10 µs (0) through 30 ks (19).
OFSL (?) {i}5-6Set (Query) the Low Pass Filter Slope to 6 (0), 12 (1), 18 (2) or 24 (3) dB/oct.
SYNC (?) {i}5-7Set (Query) the Synchronous Filter to Off (0) or On below 200 Hz (1).
DISPLAY and OUTPUTpage description
DDEF (?) i {, j, k}5-8Set (Query) the CH1 or CH2 (i=1,2) display to XY, Rθ, XnYn, Aux 1,3 or Aux 2,4 (j=0..4)
and ratio the display to None, Aux1,3 or Aux 2,4 (k=0,1,2).
FPOP (?) i {, j}5-8Set (Query) the CH1 (i=1) or CH2 (i=2) Output Source to X or Y (j=1) or Display (j=0).
OEXP (?) i {, x, j}5-8Set (Query) the X, Y, R (i=1,2,3) Offset to x percent ( -105.00 ≤ x ≤ 105.00)
and Expand to 1, 10 or 100 (j=0,1,2).
AOFF i5-8Auto Offset X, Y, R (i=1,2,3).
AUX INPUT/OUTPUTpage description
OAUX ? i5-9Query the value of Aux Input i (1,2,3,4).
AUXV (?) i {, x}5-9Set (Query) voltage of Aux Output i (1,2,3,4) to x Volts. -10.500 ≤ x ≤ 10.500.
SETUP page description
OUTX (?) {i}5-10Set (Query) the Output Interface to RS232 (0) or GPIB (1).
OVRM (?) {i}5-10Set (Query) the GPIB Overide Remote state to Off (0) or On (1).
KCLK (?) {i}5-10Set (Query) the Key Click to Off (0) or On (1).
ALRM (?) {i}5-10Set (Query) the Alarms to Off (0) or On (1).
SSET i5-10Save current setup to setting buffer i (1≤i≤9).
RSET i5-10Recall current setup from setting buffer i (1≤i≤9).
AUTO FUNCTIONS page description
AGAN5-11Auto Gain function. Same as pressing the [AUTO GAIN] key.
ARSV5-11Auto Reserve function. Same as pressing the [AUTO RESERVE] key.
APHS5-11Auto Phase function. Same as pressing the [AUTO PHASE] key.
AOFF i5-11Auto Offset X,Y or R (i=1,2,3).
1-7
SR830 DSP Lock-In Amplifier
DATA STORAGEpage description
SRAT (?) {i}5-13Set (Query) the DataSample Rate to 62.5 mHz (0) through 512 Hz (13) or Trigger (14).
SEND (?) {i}5-13Set (Query) the Data Scan Mode to 1 Shot (0) or Loop (1).
TRIG5-13Software trigger command. Same as trigger input.
TSTR (?) {i}5-13Set (Query) the Trigger Starts Scan modeto No (0) or Yes (1).
STRT5-13Start or continue a scan.
PAUS5-13Pause a scan. Does not reset a paused or done scan.
REST5-14Reset the scan. All stored data is lost.
DATA TRANSFER page description
OUTP? i5-15Query the value of X (1), Y (2), R (3) or θ (4). Returns ASCII floating point value.
OUTR? i5-15Query the value of Display i (1,2). Returns ASCII floating point value.
SNAP?i,j{,k,l,m,n}5-15Query the value of 2 thru 6 paramters at once.
OAUX? i5-16Query the value of Aux Input i (1,2,3,4). Returns ASCII floating point value.
SPTS? 5-16Query the number of points stored in Display buffer.
TRCA? i,j,k5-16 Read k≥1 points starting at bin j≥0 from Display i (1,2) buffer in ASCII floating point.
TRCB? i,j,k5-16Read k≥1 points starting at bin j≥0 from Display i (1,2) buffer in IEEE binary floating point.
TRCL? i,j,k5-17Read k≥1 points starting at bin j≥0 from Display i (1,2) buffer in non-normalized binary floating
point.
FAST (?) {i}5-17Set (Query) Fast Data Transfer Mode On (1) or Off (0).On will transfer binary X and Y every
sample during a scan over the GPIB interface.
STRD5-18Start a scan after 0.5sec delay. Use with Fast Data Transfer Mode.
INTERFACE page description
❋RST5-19Reset the unit to its default configurations.
❋IDN?5-19Read the SR830 device identification string.
LOCL(?) {i}5-19Set (Query) the Local/Remote state to LOCAL (0), REMOTE (1), or LOCAL LOCKOUT (2).
OVRM (?) {i}5-19Set (Query) the GPIB Overide Remote state to Off (0) or On (1).
TRIG5-19Software trigger command. Same as trigger input.
STATUSpage description
❋CLS5-20Clear all status bytes.
❋ESE (?) {i} {,j}5-20Set (Query) the Standard Event Status Byte Enable Register to the decimal value i (0-255).
❋ESE i,j sets bit i (0-7) to j (0 or 1). ❋ESE? queries the byte. ❋ESE?i queries only bit i.
❋ESR? {i}5-20Query the Standard Event Status Byte. If i is included, only bit i is queried.
❋SRE (?) {i} {,j}5-20Set (Query) the Serial Poll Enable Register to the decimal value i (0-255). ❋SRE i,j sets bit i (0-
7) to j (0 or 1). ❋SRE? queries the byte, ❋SRE?i queries only bit i.
❋STB? {i}5-20Query the Serial Poll Status Byte. If i is included, only bit i is queried.
❋PSC (?) {i}5-20Set (Query) the Power On Status Clear bit to Set (1) or Clear (0).
ERRE (?) {i} {,j}5-20Set (Query) the Error Status Enable Register to the decimal value i (0-255). ERRE i,j sets bit i
(0-7) to j (0 or 1). ERRE? queries the byte, ERRE?i queries only bit i.
ERRS? {i}5-20Query the Error Status Byte. If i is included, only bit i is queried.
LIAE (?) {i} {,j}5-20Set (Query) the LIA Status Enable Register to the decimal value i (0-255). LIAE i,j sets
bit i (0-7) to j (0 or 1). LIAE? queries the byte, LIAE?i queries only bit i.
LIAS? {i}5-20Query the LIA Status Byte. If i is included, only bit i is queried.
1-8
STATUS BYTE DEFINITIONS
SR830 DSP Lock-In Amplifier
SERIAL POLL STATUS BYTE (5-21)
bitnameusage
0SCNNo data is being acquired
1IFCNo command execution in progress
2ERRUnmasked bit in error status byte set
3LIAUnmasked bit in LIA status byte set
4MAVThe interface output buffer is non-empty
5ESBUnmasked bit in standard status byte set
6SRQSRQ (service request) has occurred
7Unused
STANDARD EVENT STATUS BYTE (5-22)
bitnameusage
0INPSet on input queue overflow
1Unused
2QRYSet on output queue overflow
3Unused
4EXESet when command execution error occurs
5CMDSet when an illegal command is received
6URQSet by any key press or knob rotation
7PONSet by power-on
LIA STATUS BYTE (5-23)
bitnameusage
0RSRV/INPT Set when on RESERVE or INPUT overload
1FILTR Set when on FILTR overload
2OUTPTSet when on OUTPT overload
3UNLKSet when on reference unlock
4RANGESet when detection freq crosses 200 Hz
5TCSet when time constant is changed
6TRIGSet when unit is triggered
7Unused
ERROR STATUS BYTE (5-23)
bitnameusage
0Unused
1Backup Error Set when battery backup fails
2RAM ErrorSet when RAM Memory test finds an error
3Unused
4ROM ErrorSet when ROM Memory test finds an error
5GPIB ErrorSet when GPIB binary data transfer aborts
6DSP ErrorSet when DSP test finds an error
7Math ErrorSet when an internal math error occurs
1-9
SR830 DSP Lock-In Amplifier
1-10
GETTING STARTED
YOUR FIRST MEASUREMENTS
The sample measurements described in this section are designed to acquaint the first time user with the
SR830 DSP Lock-In Amplifier. Do not be concerned that your measurements do not exactly agree with these
exercises. The focus of these measurement exercises is to learn how to use the instrument.
It is highly recommended that the first time user step through some or all of these exercises before attempting
to perform an actual experiment.
The experimental procedures are detailed in two columns. The left column lists the actual steps in the experiment. The right column is an explanation of each step.
[Keys]Front panel keys are referred to in brackets such as [Display] where
'Display' is the key label.
KnobThe knob is used to adjust parameters which are displayed in the
Reference display.
2-1
Getting Started
2-2
The Basic Lock-in
THE BASIC LOCK-IN
This measurement is designed to use the internal oscillator to explore some of the basic lock-in functions.
You will need BNC cables.
Specifically, you will measure the amplitude of the Sine Out at various frequencies, sensitivities, time constants and phase shifts.
1. Disconnect all cables from the lock-in. Turn
the power on while holding down the [Setup]
key. Wait until the power-on tests are
completed.
2. Connect the Sine Out on the front panel to the
A input using a BNC cable.
When the power is turned on with the [Setup] key
pressed, the lock-in returns to its standard default
settings. See the Standard Settings list in the
Operation section for a complete listing of the
settings.
The Channel 1 display shows X and Channel 2
shows Y.
The lock-in defaults to the internal oscillator reference set at 1.000 kHz. The reference mode is indicated by the INTERNAL led. In this mode, the
lock-in generates a synchronous sine output at the
internal reference frequency.
The input impedance of the lock-in is 10 MΩ. The
Sine Out has an output impedance of 50Ω. Since
the Sine Output amplitude is specified into a high
impedance load, the output impedance does not
affect the amplitude.
The sine amplitude is 1.000 Vrms and the
sensitivity is 1 V(rms). Since the phase shift of the
sine output is very close to zero, Channel 1 (X)
should read close to 1.000 V and Channel 2 (Y)
should read close to 0.000 V.
3. Press [Auto Phase]
4. Press [Phase]
5. Press the [+90°] key.
Automatically adjust the reference phase shift to
eliminate any residual phase error. This should set
the value of Y to zero.
Display the reference phase shift in the Reference
display. The phase shift should be close to zero.
This adds 90° to the reference phase shift. The
value of X drops to zero and Y becomes minus the
magnitude (-1.000 V).
2-3
The Basic Lock-in
Use the knob to adjust the phase shift until Y
is zero and X is equal to the positive
amplitude.
Press [Auto Phase]
6. Press [Freq]
Use the knob to adjust the frequency to
10 kHz.
Use the knob to adjust the frequency back to
1 kHz.
The knob is used to adjust parameters which are
shown in the Reference display, such as phase,
amplitude and frequency. The final phase value
should be close to zero again.
Use the Auto Phase function to return Y to zero
and X to the amplitude.
Show the internal oscillator frequency in the
Reference display.
The knob now adjusts the frequency. The measured signal amplitude should stay within 1% of 1 V
and the phase shift should stay close to zero (the
value of Y should stay close to zero).
The internal oscillator is crystal synthesized with
25 ppm of frequency error. The frequency can be
set with 4 1/2 digit or 0.1 mHz resolution, whichever is greater.
7. Press [Ampl]
Use the knob to adjust the amplitude to
0.01 V.
8. Press [Auto Gain]
9. Press [Sensitivity Up] to select 50 mV full
scale.
Change the sensitivity back to 20 mV.
10. Press [Time Constant Down] to change the
time constant to 300 µs.
Show the sine output amplitude in the Reference
display.
As the amplitude is changed, the measured value
of X should equal the sine output amplitude. The
sine amplitude can be set from 4 mV to 5 V rms
into high impedance (half the amplitude into a 50
Ω load).
The Auto Gain function will adjust the sensitivity so
that the measured magnitude (R) is a sizable percentage of full scale. Watch the sensitivity indicators change.
Parameters which have many options, such as
sensitivity and time constant, are changed with up
and down keys. The sensitivity and time constant
are indicated by leds.
The values of X and Y become noisy. This is
because the 2f component of the output (at 2 kHz)
is no longer attenuated completely by the low pass
filters.
Press [Time Constant Up] to change the time
constant to 3 ms.
Let's leave the time constant short and change the
filter slope.
2-4
11. Press the [Slope/Oct] key until 6 dB/oct is
selected.
The Basic Lock-in
Parameters which have only a few values, such as
filter slope, have only a single key which cycles
through all available options. Press the corresponding key until the desired option is indicated
by an led.
The X and Y outputs are somewhat noisy at this
short time constant and only 1 pole of low pass
filtering.
The outputs are less noisy with 2 poles of filtering.
Press [Slope/Oct] again to select 12 dB/oct.
Press [Slope/Oct] twice to select 24 db/oct.
Press [Slope/Oct] again to select 6 db/oct.
12. Press [Freq]
Use the knob to adjust the frequency to
55.0 Hz.
13. Press [Sync Filter]
With 4 poles of low pass filtering, even this short
time constant attenuates the 2f component rea-
sonably well and provides steady readings.
Let's leave the filtering short and the outputs noisy
for now.
Show the internal reference frequency on the
Reference display.
At a reference frequency of 55 Hz and a 6 db/oct,
3 ms time constant, the output is totally dominated
by the 2f component at 100 Hz.
This turns on synchronous filtering whenever the
detection frequency is below 200 Hz.
Synchronous filtering effectively removes output
components at multiples of the detection frequen-
cy. At low frequencies, this filter is a very effective
way to remove 2f without using extremely long
time constants.
The outputs are now very quiet and steady, even
though the time constant is very short. The
response time of the synchronous filter is equal to
the period of the detection frequency (18 ms in this
case).
This concludes this measurement example. You
should have a feeling for the basic operation of the
front panel. Basic lock-in parameters have been
introduced and you should be able to perform
simple measurements.
2-5
The Basic Lock-in
2-6
X, Y, R and θ
X, Y, R and θ
This measurement is designed to use the internal oscillator and an external signal source to explore some of
the display types. You will need a synthesized function generator capable of providing a 100 mVrms sine
wave at 1.000 kHz (the DS335 from SRS will suffice), BNC cables and a terminator appropriate for the generator function output.
Specifically, you will display the lock-in outputs when measuring a signal close to, but not equal to, the internal reference frequency. This setup ensures changing outputs which are more illustrative than steady outputs.
The displays will be configured to show X, Y, R and θ.
1. Disconnect all cables from the lock-in. Turn
the power on while holding down the [Setup]
key. Wait until the power-on tests are
completed.
2. Turn on the function generator, set the frequency to 1.0000 kHz (exactly) and the amplitude to 500 mVrms.
Connect the function output (sine wave) from
the synthesized function generator to the A
input using a BNC cable and appropriate
terminator.
When the power is turned on with the [Setup] key
pressed, the lock-in returns to its standard settings. See the Standard Settings list in the
Operation section for a complete listing of the
settings.
The Channel 1 display shows X and Channel 2
shows Y.
The input impedance of the lock-in is 10 MΩ. The
generator may require a terminator. Many generators have either a 50Ω or 600Ω output impedance.
Use the appropriate feedthrough or T termination if
necessary. In general, not using a terminator
means that the function output amplitude will not
agree with the generator setting.
The lock-in defaults to the internal oscillator reference set at 1.000 kHz. The reference mode is indicated by the INTERNAL led. In this mode, the
internal oscillator sets the detection frequency.
The internal oscillator is crystal synthesized so
that the actual reference frequency should be very
close to the actual generator frequency. The X and
Y displays should read values which change very
slowly. The lock-in and the generator are not
phase locked but they are at the same frequency
with some slowly changing phase.
3. Press [Freq]
Use the knob to change the frequency to
999.8 Hz.
Show the internal oscillator frequency on the
Reference display.
By setting the lock-in reference 0.2 Hz away from
the signal frequency, the X and Y outputs are
0.2 Hz sine waves (frequency difference between
reference and signal). The X and Y output displays
2-7
X, Y, R and θ
should now oscillate at about 0.2 Hz (the accuracy
is determined by the crystals of the generator and
the lock-in).
4. Press [Channel 1 Display] to select R.
5. Press [Channel 2 Display] to select θ.
6. Press [Freq]
Use the knob to adjust the frequency slowly to
try to stop the rotation of the phase.
7. Use a BNC cable to connect the TTL SYNC
output from the generator to the Reference
Input of the lock-in.
The default Channel 1 display is X. Change the
display to show R. R is phase independent so it
shows a steady value (close to 0.500 V).
The default Channel 2 display is Y. Change the
display to show θ. The phase between the reference and the signal changes by 360° approximately every 5 sec (0.2 Hz difference frequency).
The bar graph in this case is scaled to ±180°. The
bar graph should be a linear phase ramp at
0.2 Hz.
Show the internal oscillator frequency.
As the internal reference frequency gets closer to
the signal frequency, the phase rotation gets
slower and slower. If the frequencies are
EXACTLY equal, then the phase is constant.
By using the signal generator as the external reference, the lock-in will phase lock its internal oscillator to the signal frequency and the phase will be a
constant.
Press [Source] to turn the INTERNAL led off.
Press [Trig] to select POS EDGE.
Select external reference mode. The lock-in will
phase lock to the signal at the Reference Input.
With a TTL reference signal, the slope needs to be
set to either rising or falling edge.
The phase is now constant. The actual phase
depends upon the phase difference between the
function output and the sync output from the
generator.
The external reference frequency (as measured by
the lock-in) is displayed on the Reference display.
The UNLOCK indicator should be OFF (successfully locked to the external reference).
The displays may be stored in the internal data
buffers at a programmable sampling rate. This
allows storage of 16000 points of both displays.
2-8
Outputs, Offsets and Expands
OUTPUTS, OFFSETS and EXPANDS
This measurement is designed to use the internal oscillator to explore some of the basic lock-in outputs. You
will need BNC cables and a digital voltmeter (DVM).
Specifically, you will measure the amplitude of the Sine Out and provide analog outputs proportional to the
measurement. The effect of offsets and expands on the displayed values and the analog outputs will be
explored.
1. Disconnect all cables from the lock-in. Turn
the power on while holding down the [Setup]
key. Wait until the power-on tests are
completed.
2. Connect the Sine Out on the front panel to the
A input using a BNC cable.
When the power is turned on with the [Setup] key
pressed, the lock-in returns to its standard settings. See the Standard Settings list in the
Operation section for a complete listing of the
settings.
The Channel 1 display shows X and Channel 2
shows Y.
The lock-in defaults to the internal oscillator reference set at 1.000 kHz. The reference mode is indicated by the INTERNAL led. In this mode, the
lock-in generates a synchronous sine output at the
internal reference frequency.
The input impedance of the lock-in is 10 MΩ. The
Sine Out has an output impedance of 50Ω. Since
the Sine Output amplitude is specified into a high
impedance load, the output impedance does not
affect the amplitude.
The sine amplitude is 1.000 Vrms and the
sensitivity is 1 V(rms). Since the phase shift of the
sine output is very close to zero, Channel 1 (X)
should read close to 1.000 V and Channel 2 (Y)
should read close to 0.000 V.
3. Connect the CH1 OUTPUT on the front panel
to the DVM. Set the DVM to read DC Volts.
4. Press [Ampl]
Use the knob to adjust the sine amplitude to
0.5 V.
The CH1 output defaults to X. The output voltage
is simply (X/Sensitivity - Offset)xExpandx10V. In
this case, X = 1.000 V, the sensitivity = 1 V, the
offset is zero percent and the expand is 1. The
output should thus be 10 V or 100% of full scale.
Display the sine output amplitude.
Set the amplitude to 0.5 V. The Channel 1 display
should show X=0.5 V and the CH1 output voltage
should be 5 V on the DVM (1/2 of full scale).
2-9
Outputs, Offsets and Expands
5. Press [Channel 1 Auto Offset]
X, Y and R may all be offset and expanded separately. Since Channel 1 is displaying X, the
OFFSET and [Expand] keys below the Channel 1
display set the X offset and expand. The display
determines which quantity (X or R) is offset and
expanded.
Auto Offset automatically adjusts the X offset (or Y
or R) such that X (or Y or R) becomes zero. In this
case, X is offset to zero. The offset should be
about 50%. Offsets are useful for making relative
measurements. In analog lock-ins, offsets were
generally used to remove DC output errors from
the lock-in itself. The SR830 has no DC output
errors and the offset is not required for most
measurements.
The offset affects both the displayed value of X
and any analog output proportional to X. The CH1
output voltage should be zero in this case.
The Offset indicator turns on at the bottom of the
Channel 1 display to indicate that the displayed
quantity is affected by an offset.
Press [Channel 1 Offset Modify]
Use the knob to adjust the X offset to 40.0%
Press [Channel 1 Expand] to select x10.
Show the Channel 1 (X) offset in the Reference
display.
Change the offset to 40% of full scale. The output
offsets are a percentage of full scale. The percentage does not change with the sensitivity. The displayed value of X should be 0.100 V (0.5 V - 40%
of full scale). The CH1 output voltage is
(X/Sensitivity - Offset)xExpandx10V.
CH1 Out = (0.5/1.0 - 0.4)x1x10V = 1 V
With an expand of 10, the display has one more
digit of resolution (100.00 mV full scale).
The Expand indicator turns on at the bottom of the
Channel 1 display to indicate that the displayed
quantity is affected by a non-unity expand.
The CH1 output is
(X/Sensitivity - Offset)xExpandx10V. In this case,
the output voltage is
CH1 Out = (0.5/1.0 - 0.4)x10x10V = 10V
The expand allows the output gain to be increased
by up to 100. The output voltage is limited to
10.9 V and any output which tries to be greater will
2-10
Outputs, Offsets and Expands
turn on the OVLD indicator in the Channel 1
display.
With offset and expand, the output voltage gain
and offset can be programmed to provide control
of feedback signals with the proper bias and gain
for a variety of situations.
Offsets add and subtract from the displayed
values while expand increases the resolution of
the display.
6. Connect the DVM to the X output on the rear
panel.
7. Connect the DVM to the CH1 OUTPUT on the
front panel again.
Press [Channel 1 Output] to select Display.
Press [Channel 1 Display] to select R.
The X and Y outputs on the rear panel always provide voltages proportional to X and Y (with offset
and expand). The X output voltage should be
10 V, just like the CH1 output.
The front panel outputs can be configured to
output different quantities while the rear panel outputs always output X and Y.
NOTE:
Outputs proportional to X and Y (rear panel, CH1
or CH2) have 100 kHz of bandwidth. The CH1 and
CH2 outputs, when configured to be proportional
to the displays (even if the display is X or Y) are
updated at 512 Hz and have a 200 Hz bandwidth.
It is important to keep this in mind if you use very
short time constants.
CH1 OUTPUT can be proportional to X or the display. Choose Display. The display is X so the CH1
output should remain 10.0 V (but its bandwidth is
only 200 Hz instead of 100 kHz).
Let's change CH1 to output R.
The X and Y offset and expand functions are
output functions, they do NOT affect the calculation of R or θ. Thus, Channel 1 (R) should be 0.5V
and the CH1 output voltage should be 5V (1/2 of
full scale).
The Channel 1 offset and expand keys now set
the R offset and expand. The X offset and expand
are still set at 40% and x10 as reflected at the rear
panel X output.
See the DC Outputs and Scaling discussion in the
Lock-In Basics section for more detailed information on output scaling.
2-11
Outputs, Offsets and Expands
2-12
Storing and Recalling Setups
STORING and RECALLING SETUPS
The SR830 can store 9 complete instrument setups in non-volatile memory.
1. Turn the lock-in on while holding down the
[Setup] key. Wait until the power-on tests are
completed. Disconnect any cables from the
lock-in.
2. Press [Sensitivity Down] to select 100 mV.
Press [Time Constant Up] to select 1 S.
3. Press [Save]
Use the knob to select setup number 3.
Press [Save] again.
When the power is turned on with the [Setup] key
pressed, the lock-in returns to its standard settings. See the Standard Settings list in the
Operation section for a complete listing of the
settings.
Change the lock-in setup so that we have a nondefault setup to save.
Change the sensitivity to 100 mV.
Change the time constant to 1 second.
The Reference display shows the setup number
(1-9).
The knob selects the setup number.
Press [Save] again to complete the save opera-
tion. Any other key aborts the save.
The current setup is now saved as setup number
3.
4. Turn the lock-in off and on while holding down
the [Setup] key. Wait until the power-on tests
are complete.
5. Press [Recall]
Use the knob to select setup number 3.
Press [Recall] again.
Change the lock-in setup back to the default
setup. Now let's recall the lock-in setup that we
just saved.
Check that the sensitivity and time constant are 1V
and 100 ms (default values).
The Reference display shows the setup number.
The knob selects the setup number.
Press [Recall] again to complete the recall opera-
tion. Any other key aborts the recall.
The sensitivity and time constant should be the
same as those in effect when the setup was
saved.
2-13
Storing and Recalling Setups
2-14
Aux Outputs and Inputs
AUX OUTPUTS and INPUTS
This measurement is designed to illustrate the use of the Aux Outputs and Inputs on the rear panel. You will
need BNC cables and a digital voltmeter (DVM).
Specifically, you will set the Aux Output voltages and measure them with the DVM. These outputs will then be
connected to the Aux Inputs to simulate external DC voltages which the lock-in can measure.
1. Disconnect all cables from the lock-in. Turn
the power on while holding down the [Setup]
key. Wait until the power-on tests are
completed.
2. Connect Aux Out 1 on the rear panel to the
DVM. Set the DVM to read DC volts.
3. Press [Aux Out] until the Reference display
shows the level of Aux Out 1( as indicated by
the AxOut1 led below the display).
Use the knob to adjust the level to 10.00 V.
Use the knob to adjust the level to -5.00 V.
When the power is turned on with the [Setup] key
pressed, the lock-in returns to its standard settings. See the Standard Settings list in the
Operation section for a complete listing of the
settings.
The 4 Aux Outputs can provide programmable
voltages between -10.5 and +10.5 volts. The outputs can be set from the front panel or via the
computer interface.
Show the level of Aux Out 1 on the Reference
display.
Change the output to 10V. The DVM should display 10.0 V.
Change the output to -5V. The DVM should display -5.0 V.
The 4 outputs are useful for controlling other
parameters in an experiment, such as pressure,
temperature, wavelength, etc.
4. Press [Channel 1 Display] to select AUX IN 1.
5. Disconnect the DVM from Aux Out 1. Connect
AuxOut 1 to Aux In 1 on the rear panel.
Change the Channel 1 display to measure Aux
Input 1.
The Aux Inputs can read 4 analog voltages. These
inputs are useful for monitoring and measuring
other parameters in an experiment, such as pressure, temperature, position, etc.
We'll use Aux Out 1 to provide an analog voltage
to measure.
Channel 1 should now display -5 V (Aux In 1).
2-15
Aux Outputs and Inputs
6. Press [Channel 2 Display] to select AUX IN 3.
7. Connect Aux Out 1 to Aux In 3 on the rear
panel.
Change the Channel 2 display to measure Aux
Input 3.
Channel 2 should now display -5 V (Aux In 3).
The Channel 1 and 2 displays may be ratio'ed to
the Aux Input voltages. See the Basics section for
more about output scaling.
The displays may be stored in the internal data
buffers at a programmable sampling rate. This
allows storage of not only the lock-in outputs, X,Y,
R or θ, but also the values of the Aux Inputs. See
the Programming section for more details.
2-16
WHAT IS A LOCK-IN AMPLIFIER?
SR830 BASICS
Lock-in amplifiers are used to detect and measure
very small AC signals - all the way down to a few
nanovolts! Accurate measurements may be made
even when the small signal is obscured by noise
sources many thousands of times larger.
Lock-in amplifiers use a technique known as
phase-sensitive detection to single out the component of the signal at a specific reference frequency
AND phase. Noise signals at frequencies other
than the reference frequency are rejected and do
not affect the measurement.
Why use a lock-in?
Let's consider an example. Suppose the signal is a
10 nV sine wave at 10 kHz. Clearly some amplification is required. A good low noise amplifier will
have about 5 nV/√Hz of input noise. If the amplifier
bandwidth is 100 kHz and the gain is 1000, then
we can expect our output to be 10 µV of signal
(10 nV x 1000) and 1.6 mV of broadband noise
(5 nV/√Hz x √100 kHz x 1000). We won't have
much luck measuring the output signal unless we
single out the frequency of interest.
If we follow the amplifier with a band pass filter
with a Q=100 (a VERY good filter) centered at
10 kHz, any signal in a 100 Hz bandwidth will be
detected (10 kHz/Q). The noise in the filter pass
band will be 50 µV (5 nV/√Hz x √100 Hz x 1000)
and the signal will still be 10 µV. The output noise
is much greater than the signal and an accurate
measurement can not be made. Further gain will
not help the signal to noise problem.
Now try following the amplifier with a phasesensitive detector (PSD). The PSD can detect the
signal at 10 kHz with a bandwidth as narrow as
0.01 Hz! In this case, the noise in the detection
bandwidth will be only 0.5 µV (5 nV/√Hz x √.01 Hz
x 1000) while the signal is still 10 µV. The signal to
noise ratio is now 20 and an accurate measurement of the signal is possible.
What is phase-sensitive detection?
Lock-in measurements require a frequency reference. Typically an experiment is excited at a fixed
frequency (from an oscillator or function generator)
and the lock-in detects the response from the
experiment at the reference frequency. In the diagram below, the reference signal is a square wave
at frequency ωr. This might be the sync output
from a function generator. If the sine output from
the function generator is used to excite the experiment, the response might be the signal waveform
shown below. The signal is V
where V
The SR830 generates its own sine wave, shown
as the lock-in reference below. The lock-in reference is VLsin(ωLt + θ
The SR830 amplifies the signal and then multiplies
it by the lock-in reference using a phase-sensitive
detector or multiplier. The output of the PSD is
simply the product of two sine waves.
V
psd
The PSD output is two AC signals, one at the difference frequency (ωr - ωL) and the other at the
sum frequency (ωr + ωL).
If the PSD output is passed through a low pass
filter, the AC signals are removed. What will be
left? In the general case, nothing. However, if ω
equals ωL, the difference frequency component
will be a DC signal. In this case, the filtered PSD
output will be
V
psd
is the signal amplitude.
sig
Reference
θ
Signal
Lock-in Reference
= V
= 1/2 V
= 1/2 V
sigVL
1/2 V
sin(ωrt + θ
cos([ωr - ωL]t + θ
sigVL
cos([ωr + ωL]t + θ
sigVL
cos(θ
sigVL
ref
sig
θ
).
ref
sig
sig
)sin(ωLt + θ
- θ
ref
sig
)
sin(ωrt + θ
)
ref
- θ
+ θ
ref
) -
ref
sig
sig
sig
)
)
r
3-1
SR830 Basics
This is a very nice signal - it is a DC signal proportional to the signal amplitude.
Narrow band detection
Now suppose the input is made up of signal plus
noise. The PSD and low pass filter only detect signals whose frequencies are very close to the lockin reference frequency. Noise signals at frequencies far from the reference are attenuated at the
PSD output by the low pass filter (neither ω
ω
ref
nor ω
noise+ωref
are close to DC). Noise at fre-
noise
quencies very close to the reference frequency will
result in very low frequency AC outputs from the
PSD (|ω
noise-ωref
| is small). Their attenuation
depends upon the low pass filter bandwidth and
roll-off. A narrower bandwidth will remove noise
sources very close to the reference frequency, a
wider bandwidth allows these signals to pass. The
low pass filter bandwidth determines the bandwidth of detection. Only the signal at the reference
frequency will result in a true DC output and be
unaffected by the low pass filter. This is the signal
we want to measure.
Where does the
lock-in reference come from?
We need to make the lock-in reference the same
as the signal frequency, i.e. ωr = ωL. Not only do
the frequencies have to be the same, the phase
between the signals can not change with time, otherwise cos(θ
be a DC signal. In other words, the lock-in reference needs to be phase-locked to the signal
reference.
sig
- θ
) will change and V
ref
psd
will not
sync) which is always phase-locked to the reference oscillator.
Magnitude and phase
Remember that the PSD output is proportional
to V
cosθ where θ = (θ
sig
difference between the signal and the lock-in reference oscillator. By adjusting θ
equal to zero, in which case we can measure V
-
(cosθ=1). Conversely, if θ is 90°, there will be no
output at all. A lock-in with a single PSD is called a
single-phase lock-in and its output is V
This phase dependency can be eliminated by
adding a second PSD. If the second PSD multiplies the signal with the reference oscillator shifted
by 90°, i.e. VLsin(ωLt + θ
tered output will be
V
V
psd2
psd2
= 1/2 V
~ V
sig
sigVL
sinθ
Now we have two outputs, one proportional to
cosθ and the other proportional to sinθ. If we call
the first output X and the second Y,
X = V
cosθY = V
sig
these two quantities represent the signal as a
vector relative to the lock-in reference oscillator. X
is called the 'in-phase' component and Y the
'quadrature' component. This is because when
θ=0, X measures the signal while Y is zero.
- θ
sig
ref
sin(θ
sig
). θ is the phase
ref
we can make θ
ref
+ 90°), its low pass fil-
- θ
ref
)
sig
sinθ
sig
sig
cosθ.
Lock-in amplifiers use a phase-locked-loop (PLL)
to generate the reference signal. An external reference signal (in this case, the reference square
wave) is provided to the lock-in. The PLL in the
lock-in locks the internal reference oscillator to this
external reference, resulting in a reference sine
wave at ωr with a fixed phase shift of θ
. Since
ref
the PLL actively tracks the external reference,
changes in the external reference frequency do
not affect the measurement.
All lock-in measurements
require a reference signal.
In this case, the reference is provided by the excitation source (the function generator). This is
called an external reference source. In many situations, the SR830's internal oscillator may be used
instead. The internal oscillator is just like a function generator (with variable sine output and a TTL
By computing the magnitude (R) of the signal
vector, the phase dependency is removed.
R = (X2 + Y2)
1/2
= V
sig
R measures the signal amplitude and does not
depend upon the phase between the signal and
lock-in reference.
A dual-phase lock-in, such as the SR830, has two
PSD's, with reference oscillators 90° apart, and
can measure X, Y and R directly. In addition, the
phase θ between the signal and lock-in reference,
can be measured according to
θ = tan-1 (Y/X)
3-2
WHAT DOES A LOCK-IN MEASURE?
SR830 Basics
So what exactly does the SR830 measure?
Fourier's theorem basically states that any input
signal can be represented as the sum of many,
many sine waves of differing amplitudes, frequencies and phases. This is generally considered as
representing the signal in the "frequency domain".
Normal oscilloscopes display the signal in the
"time domain". Except in the case of clean sine
waves, the time domain representation does not
convey very much information about the various
frequencies which make up the signal.
What does the SR830 measure?
The SR830 multiplies the signal by a pure sine
wave at the reference frequency. All components
of the input signal are multiplied by the reference
simultaneously. Mathematically speaking, sine
waves of differing frequencies are orthogonal, i.e.
the average of the product of two sine waves is
zero unless the frequencies are EXACTLY the
same. In the SR830, the product of this multiplication yields a DC output signal proportional to the
component of the signal whose frequency is exactly locked to the reference frequency. The low pass
filter which follows the multiplier provides the averaging which removes the products of the reference
with components at all other frequencies.
The SR830, because it multiplies the signal with a
pure sine wave, measures the single Fourier (sine)
component of the signal at the reference frequency. Let's take a look at an example. Suppose the
input signal is a simple square wave at frequency
f. The square wave is actually composed of many
sine waves at multiples of f with carefully related
amplitudes and phases. A 2V pk-pk square wave
can be expressed as
frequencies is removed by the low pass filter following the multiplier. This "bandwidth narrowing" is
the primary advantage that a lock-in amplifier provides. Only inputs at frequencies at the reference
frequency result in an output.
RMS or Peak?
Lock-in amplifiers as a general rule display the
input signal in Volts RMS. When the SR830 displays a magnitude of 1V (rms), the component of
the input signal at the reference frequency is a
sine wave with an amplitude of 1 Vrms or
2.8 V pk-pk.
Thus, in the previous example with a 2 V pk-pk
square wave input, the SR830 would detect the
first sine component, 1.273sin(ωt). The measured
and displayed magnitude would be 0.90 V (rms)
(1/√2 x 1.273).
Degrees or Radians?
In this discussion, frequencies have been referred
to as f (Hz) and ω (2πf radians/sec). This is
because people measure frequencies in cycles
per second and math works best in radians. For
purposes of measurement, frequencies as measured in a lock-in amplifier are in Hz. The equations
used to explain the actual calculations are sometimes written using ω to simplify the expressions.
Phase is always reported in degrees. Once again,
this is more by custom than by choice. Equations
written as sin(ωt + θ) are written as if θ is in
radians mostly for simplicity. Lock-in amplifiers
always manipulate and measure phase in
degrees.
S(t) = 1.273sin(ωt) + 0.4244sin(3ωt) +
0.2546sin(5ωt) + ...
where ω = 2πf. The SR830, locked to f will single
out the first component. The measured signal will
be 1.273sin(ωt), not the 2V pk-pk that you'd measure on a scope.
In the general case, the input consists of signal
plus noise. Noise is represented as varying signals
at all frequencies. The ideal lock-in only responds
to noise at the reference frequency. Noise at other
3-3
Loading...
+ 144 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.