Safety and Preparation for Use iii
Specifications v
Abridged Command List1-9 viii
GETTING STARTED
Your First Measurement 1-1
Analyzing a Sine Wave 1-2
Second Measurement Example 1-6
Amplifier Noise Level 1-7
Using Triggers and the Time Record 1-10
Using the Disk Drive 1-14
Using Data Tables 1-20 Using Limit Tables 1-23
Using Trace Math 1-27
Using the Source 1-31
Sine 1-32
Two Tone 1-36
Noise 1-38
Chirp 1-41
Things to Watch Out For 1-46
ANALYZER BASICS
What is an FFT Spectrum Analyzer? 2-1
Frequency Spans 2-2
The Time Record 2-3
Measurement Basics 2-4
Display Type 2-5
Windowing 2-6
Averaging 2-7
Real Time Bandwidth and Overlap 2-8
Input Range 2-9
The Source 2-10
OPERATION
Front Panel3-1
Power On/Off 3-1
Reset 3-1
Video Display 3-1
Soft Keys 3-2
Keypad 3-2
Spin Knob 3-2
Disk Drive 3-2
BNC Connectors 3-2
Screen Display 3-3
Data Display 3-3
Single/Dual Trace Displays 3-3
Marker Display 3-5
Menu Display 3-5
Status Indicators 3-5
Keypad 3-7 Normal and Alternate Keys 3-7
Menu Keys 3-7
Entry Keys 3-8
START and PAUSE/CONT 3-8
MARKER 3-9
ACTIVE TRACE 3-9
AUTO RANGE 3-9
AUTOSCALE 3-9
SPAN UP/DOWN 3-9
MARKER ENTRY 3-9
MARKER MODE 3-9
MARKER REF 3-9
MARKER CENTER 3-9
MARKER MAX/MIN 3-10
PRINT 3-10
HELP 3-10
LOCAL 3-10
Rear Panel3-11
Power Entry Module 3-11
IEEE-488 Connector 3-11
RS232 Connector 3-11
Parallel Printer Connector 3-11
PC Keyboard Connector 3-11
MENUS
Frequency Menu 4-1
Measure Menu 4-3
Display Menu 4-15
Marker Mode Menu 4-17
Input Menu 4-19
Scale Menu 4-25
Analyze Menu 4-27
Average Menu 4-43
Source Menu 4-47
System Menu 4-55
Store/Recall Menu 4-71
Default Settings 4-79
PROGRAMMING
GPIB Communications 5-1
RS232 Communications 5-1
Status Indicators and Queues 5-1
Command Syntax 5-1
Interface Ready and Status 5-2
Detailed Command List 5-3
Frequency Commands 5-4
Measurement Commands 5-5
Display and Marker Commands 5-6
Scale Commands 5-8
Input Commands 5-9
Analysis Commands 5-10
Data Table Commands 5-11
Limit Table Commands 5-12
Averaging Commands 5-13
Source Commands 5-14
Print and Plot Commands 5-15
i
TABLE OF CONTENTS
System Setup Commands 5-16 Store and Recall Commands 5-18 Trace Math Commands 5-19
Front Panel Control Commands 5-20
Data Transfer Commands 5-21
Interface Commands 5-23
Status Reporting Commands 5-24
Status Byte Definitions 5-25
Serial Poll Status Byte 5-25
Serial Polls 5-25
Service Requests (SRQ) 5-26
Standard Event Status Byte 5-26 FFT Status Byte 5-27
Error Status Byte 5-27
Program Examples Microsoft C, Nat'l Instruments GPIB 5-28
BASIC, Nat'l Instruments GPIB 5-31
TESTING
Introduction 6-1
Preset 6-1
Serial Number 6-1
Firmware Revision 6-1
General Installation 6-1
Necessary Equipment 6-3
If A Test Fails 6-3
Performance Tests
Self Tests 6-4
DC Offset 6-5
Common Mode Rejection 6-7
Amplitude Accuracy and Flatness 6-8
Amplitude Linearity 6-11
Anti-Alias Filter Attenuation 6-13
Frequency Accuracy 6-14
Phase Accuracy 6-15
Harmonic Distortion 6-17
Noise and Spurious Signals 6-19
Sine Source 6-21
Analog Input Board
Overview 7-7
Input Amplifier 7-7
Gain Stages and Attenuators 7-7
Anti-Alias Filter 7-7
A/D Converter 7-8
I/O Interface 7-8
Power 7-8
Parts Lists
CPU Board 7-9
Power Supply Board 7-13
DSP Logic Board 7-16 Analog Input Board 7-20
Chassis Assembly 7-28
Miscellaneous 7-30
Schematic Diagrams
CPU Board
Power Supply Board
DSP Logic Board
Analog Input Board
ii
SR770 FFT SPECTRUM ANALYZER
SAFETY AND PREPARATION FOR USE
WARNING
Dangerous voltages, capable of causing injury or death, are present in this instrument. Use extreme
caution whenever the instrument covers are removed. Do not remove the covers while the unit is
plugged into a live outlet.
CAUTION
This instrument may be damaged if operated
with the LINE VOLTAGE SELECTOR set for the
wrong AC line voltage or if the wrong fuse is
installed.
LINE VOLTAGE SELECTION
The SR770 operates from a 100V, 120V, 220V, or
240V nominal AC power source having a line
frequency of 50 or 60 Hz. Before connecting the
power cord to a power source, verify that the LINE
VOLTAGE SELECTOR card, located in the rear
panel fuse holder, is set so that the correct AC
input voltage value is visible.
Conversion to other AC input voltages requires a
change in the fuse holder voltage card position
and fuse value. Disconnect the power cord, open
the fuse holder cover door and rotate the fuse-pull
lever to remove the fuse. Remove the small
printed circuit board and select the operating
voltage by orienting the printed circuit board so
that the desired voltage is visible when pushed
firmly into its slot. Rotate the fuse-pull lever back
into its normal position and insert the correct fuse
into the fuse holder.
LINE FUSE
Verify that the correct line fuse is installed before
connecting the line cord. For 100V/120V, use a 1
Amp fuse and for 220V/240V, use a 1/2 Amp fuse.
LINE CORD
The SR770 has a detachable, three-wire power
cord for connection to the power source and to a
protective ground. The exposed metal parts of the
instrument are connected to the outlet ground to
protect against electrical shock. Always use an
outlet which has a properly connected protective
ground.
SERVICE
Do not attempt to service or adjust this instrument
unless another person, capable of providing first
aid or resuscitation, is present.
Do not install substitute parts or perform any
unauthorized modifications to this instrument.
Contact the factory for instructions on how to
return the instrument for authorized service and
adjustment.
iii
SR770 FFT SPECTRUM ANALYZER
iv
SR770 FFT SPECTRUM ANALYZER
SPECIFICATIONS
FREQUENCY
Measurement Range 476 µHz to 100 kHz, baseband and zoomed. Spans 191 mHz to 100 kHz in a binary sequence.
Center Frequency Anywhere within the measurement range subject to span and range
limits.
Accuracy 25 ppm from 20°C to 40°C.
Resolution Span/400
Window Functions Blackman-Harris, Hanning, Flattop and Uniform. Real-time Bandwidth 100 kHz
SIGNAL INPUT
Number of Channels 1
Input Single-ended or true differential
Input Impedance 1 MΩ, 15 pf
Coupling AC or DC
CMRR 90 dB at 1 kHz (Input Range < -6 dBV)
80 dB at 1 kHz (Input Range <14 dBV)
50 dB at 1 kHz (Input Range ≥ 14 dBV)
Noise 5 nVrms/√Hz at 1 kHz typical, 10 nVrms/√Hz max.
(-166 dBVrms/√Hz typ., -160 dBVrms/√Hz max.)
AMPLITUDE
Full Scale Input Range -60 dBV (1.0 mVpk) to +34 dBV (50 Vpk) in 2 dB steps.
Dynamic Range 90 dB typical
Harmonic Distortion No greater than -80 dB from DC to 100 kHz. (Input Range ≤ 0 dBV)
Spurious Input range ≥ -50 dBV: No greater than -85 dB below full scale below 200 Hz.
No greater than -90 dB below full scale to 100 kHz.
Input Sampling 16 bit A/D at 256 kHz
Accuracy ± 0.3 dB ± 0.0
Averaging RMS, Vector and Peak Hold.
Linear and exponential averaging up to 64k scans.
TRIGGER INPUT
Modes Continuous, internal, external, or external TTL.
Internal Level: Adjustable to ±100% of input scale.
Positive or Negative slope.
Minimum Trigger Amplitude: 10% of input range. External Level: ±5V in 40 mV steps. Positive or Negative slope.
Impedance: 10 kΩ Minimum Trigger Amplitude: 100 mV.
External TTL Requires TTL level to trigger (low<.7V, high>2V).
Post-Trigger Measurement record is delayed by 1 to 65,000 samples (1/512 to 127
time records) after the trigger.
Delay resolution is 1 sample (1/512 of a record).
Pre-Trigger Measurement record starts up to 51.953 ms prior to the trigger.
Delay resolution is 3.9062 µs. Phase Indeterminacy <2°
2% of full scale (excluding windowing effects).
v
SR770 FFT SPECTRUM ANALYZER
DISPLAY FUNCTIONS
Display Real, imaginary, magnitude or phase spectrum.
Measurements Spectrum, power spectral density, time record and 1/3 octave.
Analysis Band, sideband, total harmonic distortion and trace math.
Trace Math Add, subtract, multiply, and divide with a constant, ω (2πf), or another
Graphic Expand Display expands up to 50x about any point in the display.
MARKER FUNCTIONS
Harmonic Marker Displays up to 400 harmonics of the fundamental.
Delta Marker Reads amplitude and frequency relative to defined reference.
Next Peak/Harmonic Locates nearest peak or harmonic to the left or right.
Data Tables Lists Y values of up to 200 user defined X points.
Limit Tables Automatically detects data exceeding up to 100 user defined upper and
lower limit trace segments.
SOURCE OUTPUT
Amplitude Range 0.1 mVpk to 1.000 Vpk
Amplitude Resolution 1 mVpk (Output>100 mVpk); 0.1 mVpk (Output ≤ 100.0 mVpk)
DC Offset <10.0 mV (typical)
Output Impedance < 5 Ω; ±50 mA peak output current.
SINE Amplitude Accuracy ±1% (0.09 dB) of setting, 0 Hz to 100 kHz, 0.1 Vpk to 1.0 Vpk,
high impedance load.
Frequency Resolution 15.26 mHz (1 kHz/65536)
Harmonics, Sub-Harmonics, 0.1 Vpk to 1 Vpk 0 to 10 kHz <-80 dBc
10 kHz to 100 kHz <-70 dBc
Spurious Signals <-100 dBV (typical, line frequency related)
TWO TONE Amplitude Accuracy ±1% (0.09 dB) of setting, 0 Hz to 100 kHz, 0.1 Vpk to 0.5 Vpk,
high impedance load.
Frequency Resolution 15.26 mHz (1 kHz/65536)
Harmonics, Sub-Harmonics 0.1 Vpk to 0.5 Vpk 0 to 10 kHz <-80 dB below larger tone
10 kHz to 100 kHz <-70 dB below larger tone
Spurious Signals <-100 dBV (typical, line frequency related)
WHITE NOISEOutput is 0 Hz to 100 kHz at all measurement spans.
Flatness <0.25 dB pk-pk (typical), <1.0 dB pk-pk (max)
(5000 rms averaged spectra, Source Cal on).
PINK NOISEOutput is 0 Hz to 100 kHz at all measurement spans.
Flatness <4.0 dB pk-pk, 20 Hz - 20 kHz
(measured using 1/3 octave analysis, Source Cal on).
CHIRP Output is equal amplitude sine waves at each frequency bin of the
measurement span.
Flatness Measured spectra (all spans, Source Cal on)
<0.05 dB pk-pk (typical), <0.2 dB pk-pk (max), Amplitude=1.0 Vpk.
Phase Auto Phase function calibrates to current phase spectrum.
GENERAL
Monitor Monochrome CRT. 640H by 480V resolution.
Adjustable brightness and screen position.
vi
SR770 FFT SPECTRUM ANALYZER
Interfaces IEEE-488, RS232 and Printer interfaces standard.
All instrument functions can be controlled through the IEEE-488 and
RS232 interfaces. A PC keyboard input is provided for additional
flexibility.
Hardcopy Screen dumps and table and setting listings to dot matrix and HP
LaserJet compatible printers. Data plots to HP-GL compatible plotters
(via RS232 or IEEE-488).
Disk 3.5 inch DOS compatible format, 720 kbyte capacity. Storage of data,
setups, data tables, and limit tables.
Power 60 Watts, 100/120/220/240 VAC, 50/60 Hz.
Dimensions 17"W x 6.25"H x 18.5"D
Weight 36 lbs.
Warranty One year parts and labor on materials and workmanship.
vii
SR770 FFT SPECTRUM ANALYZER
COMMAND LIST
VARIABLES g Trace0 (0), Trace1 (1), or Active Trace (-1)
i,j Integers
f Frequency (real)
x,y Real Numbers
s String
FREQUENCYpage
SPAN (?) {i} 5-4 Set (Query) the Frequency Span to 100 kHz (19) through 191 mHz (0).
STRF (?) {f} 5-4 Set (Query) the Start Frequency to f Hz.
CTRF (?) {f} 5-4 Set (Query) the Center Frequency to f Hz.
OTYP (?) {i} 5-4 Set (Query) the number of bands in Octave Analysis to 15 (0) or 30 (1).
OSTR (?) {i} 5-4 Set (Query) the Starting Band in Octave Analysis to -2 ≤ i ≤ 35.
WTNG (?) {i} 5-4 Set (Query) the Weighting in Octave Analysis to none (0) or A-weighting (1).
MEASUREMENTpage
MEAS (?) g {,i} 5-5 Set (Query) the Measurement Type to Spectrum (0), PSD (1), Time (2), or
DISP (?) g {,i} 5-5 Set (Query) the Display to LogMag (0),LinMag (1), Real (2), Imag (3), or Phase
UNIT (?) g {,i} 5-5 Set (Query) the Units to Vpk or deg (0),Vrms or rads (1), dBV (2), or dBVrms
VOEU (?) g {,i} 5-5 Set (Query) the Units to Volts (0), or EU (1).
EULB (?) g {,s} 5-5 Set (Query) the EU Label to string s.
EUVT (?) g {,x} 5-5 Set (Query) the EU Value to x EU/Volt.
WNDO (?) g {i} 5-5 Set (Query) the Window to Uniform (0), Flattop (1), Hanning (2), or BMH (3).
DISPLAY and MARKER page
ACTG (?) {i} 5-6 Set (Query) the Active Trace to trace0 (0) or trace1 (1).
FMTS (?) g {,i} 5-6 Set (Query) the Display Format to Single (0) or Dual (1) trace.
GRID (?) g {,i} 5-6 Set (Query) the Grid mode to Off (0), 8 (1), or 10 (2) divisions.
FILS (?) g {,i} 5-6 Set (Query) the Graph Style to Line (0) or Filled (1).
MRKR (?) g {,i} 5-6 Set (Query) the Marker to Off (0), On (1) or Track (2).
MRKW (?) g {,i} 5-6 Set (Query) the Marker Width to Norm (0), Wide (1), or Spot (2).
MRKM (?) g {,i} 5-6 Set (Query) the Marker Seeks mode to Max (0), Min (1), or Mean (2).
MRLK (?) {i} 5-6 Set (Query) the Linked Markers to Off (0) or On (1).
MBIN g,i 5-6 Move the marker region to bin i.
MRKX? 5-6 Query the Marker X position.
MRKY? 5-6 Query the Marker Y position.
MRPK 5-6 Move the Marker to the on screen max or min. Same as [MARKER MAX/MIN]
MRCN 5-6 Make the Marker X position the center of the span. Same as [MARKER
MRRF 5-6 Turns Marker Offset on and sets the offset equal to the marker position.
MROF (?) {i} 5-6 Set (Query) the Marker Offset to Off (0) or On (1).
MROX (?) {x} 5-7 Set (Query) the Marker Offset X value to x.
MROY (?) {x} 5-7 Set (Query) the Marker Offset Y value to x.
PKLF 5-7 Move the marker to the next peak to the left.
PKRT 5-7 Move the marker to the next peak to the right.
MSGS s 5-7 Display message s on the screen and sound an alarm.
SCALEpage
TREF (?) g {,x} 5-8 Set (Query) the Top Reference to x.
BREF (?) g {,x} 5-8 Set (Query) the Bottom Reference to x.
YDIV (?) g {,x} 5-8 Set (Query) the Vertical Scale (Y/Div) to x.
AUTS g 5-8 AutoScale graph g. Similar to the [AUTO SCALE] key.
EXPD (?) g {,i} 5-8 Set (Query) the Horizontal Expand to no expand (5), 128, 64, 30, 15, or 8 bins
ELFT (?) g {,i} 5-8 Set (Query) the Left Bin when expanded to bin i.
XAXS (?) g {,i} 5-8 Set (Query) the X Axis scaling to Linear (0) or Log (1).
description
description
Octave (3).
(4).
(3).
description
key.
CENTER] key.
description
(4-0).
viii
SR770 FFT SPECTRUM ANALYZER
INPUT page description
ISRC (?) {i} 5-9 Set (Query) the Input to A (0) or A-B (1).
IGND (?) {i} 5-9 Set (Query) the Input Grounding to Float (0) or Ground (1).
ICPL (?) {i} 5-9 Set (Query) the Input Coupling to AC (0) or DC (1).
IRNG (?) {i} 5-9 Set (Query) the Input Range to i dBV full scale. -60 ≤ i ≤ 34 and i is even.
ARNG (?) {i} 5-9 Set (Query) the Auto Range mode to Manual (0) or Auto (1).
AOFF 5-9 Perform Auto Offset calibration.
AOFM (?) {i} 5-9 Set (Query) the Auto Offset Mode to Off (0) or On (1).
TMOD (?) {i} 5-9 Set (Query) the Trigger Mode to Cont (0), Int (1), Ext (2), Ext TTL(3), or
source (4).
TRLV (?) {x} 5-9 Set (Query) the Trigger Level to x percent. -100.0 ≤ x ≤ 99.22.
TDLY (?) {i} 5-9 Set (Query) the Trigger Delay to i samples. -13300≤ i ≤ 65000.
ARMM (?) {i} 5-9 Set (Query) the Arming Mode to Auto (0) or Manual (1).
ARMS 5-9 Manually arm the trigger.
ANALYSISpage
ANAM (?) g {,i} 5-10 Set (Query) the real time Analysis to None (0), Harmonic (1), Sideband (2), or
CALC? g,i 5-10 Query result i (0 or 1) of the latest real time analysis.
FUND (?) g {,f} 5-10 Set (Query) the Harmonic Fundamental to frequency f Hz.
NHRM (?) g {,i} 5-10 Set (Query) the Number of Harmonics to 0 ≤ i ≤ 400.
NHLT 5-10 Move the Marker or Center Frequency to the next harmonic to the left.
NHRT 5-10 Move the Marker or Center Frequency to the next harmonic to the right.
SBCA (?) g {,f} 5-10 Set (Query) the Sideband Carrier to frequency f Hz.
SBSE (?) g {,f} 5-10 Set (Query) the Sideband Separation to f Hz.
NSBS (?) g {,i} 5-10 Set (Query) the Number of Sidebands to 0 ≤ i ≤ 200.
BSTR (?) g {,f} 5-10 Set (Query) the Band Start to frequency f Hz.
BCTR (?) g {,f} 5-11 Set (Query) the Band Center to frequency f Hz.
BWTH (?) g {,f} 5-11 Set (Query) the Band Width to f Hz.
TABL 5-11 Turn on Data Table display for the active trace.
DTBL (?) g {,i}{,f} 5-11 Set (Query) Data Table line i to frequency f.
DINX (?) {i} 5-11 Set (Query) Data Table index to i.
DINS 5-11 Insert a new line in the data table.
DIDT 5-11 Delete a line from the data table.
DLTB 5-11 Delete the entire data table.
LIMT 5-12 Turn on Limit Table display for the active trace.
TSTS (?) {i} 5-12 Set (Query) the Limit Testing to Off (0) or On (1).
PASF? 5-12 Query the results of the latest limit test. Pass=0 and Fail=1.
LTBL (?) g {,i} {j,f1,f2,y1,y2} 5-12 Set (Query) Limit Table line i to Xbegin (f1), Xend (f2), Y1 and Y2.
LINX (?) {i} 5-12 Set (Query) Limit Table index to i.
LINS 5-12 Insert a new line in the limit table.
LIDT 5-12 Delete a line from the limit table.
LLTB 5-12 Delete the entire limit table.
LARM (?) {i} 5-12 Set (Query) the Audio Limit Fail Alarm to Off (0) or On (1).
AVERAGINGpage
AVGO (?) {i} 5-13 Set (Query) Averaging to Off (0) or On (1).
NAVG(?) {i} 5-13 Set (Query) the Number of Averages to 2 ≤ i ≤ 32000.
AVGT (?) {i} 5-13 Set (Query) the Averaging Type to RMS (0), Vector (1), or Peak Hold (2).
AVGM (?) {i} 5-13 Set (Query) the Averaging Mode to Linear (0) or Exponential (1).
OVLP (?) {x} 5-13 Set (Query) the Overlap to x percent. 0 ≤ x ≤ 100.0.
SOURCEpage
STYP (?) {i} 5-14 Set (Query) Source Type to Off (0), SIne (1), 2-Tone (2), Noise (3) or
Chirp (4).
SLVL (?) i {,x} 5-14 Set (Query) the Level of sine, tone 1, tone 2, noise, chirp (i=0..4) to x mV.
SFRQ (?) i {,f} 5-14 Set (Query) the Frequency of sine, tone 1, tone 2 (i=0..2) to f Hz..
NTYP (?) {i} 5-14 Set (Query) the Noise Type to white (0) or pink (1).
SCAL (?) {i} 5-14 Set (Query) Source Cal off (0) or on (1) (noise and chirp only).
APHS 5-14 Do Auto Phase (chirp source only).
PLOT AND PRINTpage
PLOT 5-14 Plot the entire graph (or graphs).
PTRC 5-14 Plot the trace (or traces) only.
description
Band (3).
description
description
description
ix
SR770 FFT SPECTRUM ANALYZER
PMRK 5-14 Plot the marker (or markers) only.
PTTL (?) {s} 6-14 Set (Query) the Plot Title to string s.
PSTL (?) {s} 6-14 Set (Query) the Plot Subtitle to string s.
PRSC 6-14 Print the screen. Same as the [PRINT] key.
PSET 6-14 Print the analyzer settings.
PLIM 6-14 Print the Limit Table of the active graph.
PDAT 6-14 Print the Data Table of the active graph.
SETUPpage
OUTP (?) {i} 5-15 Set (Query) the Output Interface to RS232 (0) or GPIB (1).
OVRM (?) {i} 5-15 Set (Query) the GPIB Overide Remote state to Off (0) or On (1).
KCLK (?) {i} 5-15 Set (Query) the Key Click to Off (0) or On (1).
ALRM (?) {i} 5-15 Set (Query) the Alarms to Off (0) or On (1).
THRS (?) {i} 5-15 Set (Query) the Hours to 0≤ i ≤ 23.
TMIN (?) {i} 5-15 Set (Query) the Minutes to 0 ≤ i ≤ 59.
TSEC (?) {i} 5-15 Set (Query) the Seconds to 0 ≤ i ≤ 59.
DMTH (?) {i} 5-15 Set (Query) the Month to 1 ≤ 1 ≤ 12.
DDAY (?) {i} 5-15 Set (Query) the Day to 1 ≤ 1 ≤ 31.
DYRS (?) {i} 5-15 Set (Query) the Year to 0 ≤ 1 ≤ 99.
PLTM (?) {i} 5-15 Set (Query) the Plotter Mode to RS232 (0) or GPIB (1).
PLTB (?) {i} 5-15 Set (Query) the Plotter Baud Rate to 300 (0), 1200 (1), 2400 (2), 4800 (3),
PLTA (?) {i} 5-15 Set (Query) the Plotter GPIB Address to 0 ≤ i ≤ 30.
PLTS (?) {i} 5-15 Set (Query) the Plot Speed to Fast (0) or Slow (1).
PNTR (?) {i} 5-15 Set (Query) the Trace Pen Number to 1 ≤ i ≤ 6.
PNGD (?) {i} 5-16 Set (Query) the Grid Pen Number to 1 ≤ i ≤ 6.
PNAP (?) {i} 5-16 Set (Query) the Alphanumeric Pen Number to 1 ≤ i ≤ 6.
PNCR (?) {i} 5-17 Set (Query) the Cursor Pen Number to 1 ≤ i ≤ 6.
PRNT (?) {i} 5-17 Set (Query) the Printer Type to Epson (0) or HP (1).
STORE AND RECALL FILEpage
FNAM (?) {s} 5-18 Set (Query) the current File Name to string.
SVTR 5-18 Save the Active Trace Data to the file specified by FNAM.
SVTA 5-18 Save the Active Trace Data to the file specified by FNAM as an ASCII file.
SVST 5-18 Save the Settings to the file specified by FNAM.
RCTR 5-18 Recall the Trace Data from the file specified by FNAM to the active graph.
RCST 5-18 Recall the Settings from the file specified by FNAM.
MATH OPERATIONSpage
CSEL (?) {i} 5-19 Set (Query) the Operation to +, -, x, /, log, √ (0-5).
COPR 5-19 Start the calculation.
CARG (?) {i} 5-19 Set (Query) the Argument type to Constant (0), w (1), or Other Graph (2).
CONS (?) {x} 5-19 Set (Query) the Constant Argument to x.
CMRK 5-19 Set the Constant Argument to the Y value of the marker.
FRONT PANEL CONTROLSpage
STRT 5-20 Start data acquisition. Same as [START] key.
STCO 5-20 Pause or Continue data acquisition. Same as [PAUSE CONT] key.
PRSC 5-20 Print the screen. Same as [PRINT] key.
ACTG (?) {i} 5-20 Set (Query) the Active Trace to trace0 (0) or trace1 (1). Similar to [ACTIVE
ARNG (?) {i} 5-20 Set (Query) the Auto Range mode to Manual (0) or Auto (1). Similar to [AUTO
AUTS 5-20 AutoScale the graph. Same as the [AUTO SCALE] key.
DATA TRANSFERpage
SPEC? g {,i} 5-21 Query the Y value of bin 0 ≤ i ≤ 399.
BVAL? g, i 5-21 Query the X value of bin 0 ≤ i ≤ 399.
SPEB? g 5-21 Binary dump the entire trace g.
BDMP (?) g, {,i} 5-22 Set (Query) the auto binary dump mode for trace g.
INTERFACEpage
*RST 5-23 Reset the unit to its default configurations.
description
9600 (4).
description
description
description
TRACE] key.
RANGE] key.
description
description
x
SR770 FFT SPECTRUM ANALYZER
*IDN? 5-23 Read the SR770 device identification string.
LOCL(?) {i} 5-23 Set (Query) the Local/Remote state to LOCAL (0), REMOTE (1), or LOCAL
LOCKOUT (2).
OVRM (?) {i} 5-23 Set (Query) the GPIB Overide Remote state to Off (0) or On (1).
STATUS page
*CLS 5-24 Clear all status bytes.
*ESE (?) {i} {,j} 5-24 Set (Query) the Standard Status Byte Enable Register to the decimal value i
*ESR? {i} 5-24 Query the Standard Status Byte. If i is included, only bit i is queried.
*SRE (?) {i} {,j} 5-24 Set (Query) the Serial Poll Enable Register to the decimal value i (0-255).
*STB? {i} 5-24 Query the Serial Poll Status Byte. If i is included, only bit i is queried.
*PSC (?) {i} 5-24 Set (Query) the Power On Status Clear bit to Set (1) or Clear (0).
ERRE (?) {i} {,j} 5-24 Set (Query) the Error Status Enable Register to the decimal value i (0-255).
ERRS? {i} 5-24 Query the Error Status Byte. If i is included, only bit i is queried.
FFTE (?) {i} {,j} 5-24 Set (Query) the FFT Status Enable Register to the decimal value i (0-255).
FFTS? {i} 5-24 Query the FFT Status Byte. If i is included, only bit i is queried.
STATUS BYTE DEFINITIONS
SERIAL POLL STATUS BYTE (5-25)
name usage
bit
0 SCN No measurements in progress
1 IFC No command execution in progress
2 ERR Unmasked bit in error status byte set
3 FFT Unmasked bit in FFT status byte set
4 MAV The interface output buffer is non-empty
5 ESB Unmasked bit in standard status byte
set
6 SRQ SRQ (service request) has occurred
7 Unused
STANDARD EVENT STATUS BYTE (5-26)
name usage
bit
0 INP Set on input queue overflow
1 Limit Fail Set when a limit test fails
2 QRY Set on output queue overflow
3 Unused
4 EXE Set when command execution error
occurs
5 CMD Set when an illegal command is
received
6 URQ Set by any key press or knob rotation
7 PON Set by power-on
description
(0-255).
FFT STATUS BYTE (5-27)
name usage
bit
0 Triggered Set when a time record is triggered
1 Prn/Plt Set when a printout or plot is completed
2 NewData 0 Set when new data is available for trace 0
3 NewData 1 Set when new data is available for trace 1
4 Avg Set when a linear average is completed
5 AutoRng Set when auto range changes the range
6 High Voltage Set when high voltagedetected at input
7 Settle Set when settling is complete
ERROR STATUS BYTE (5-27)
name usage
bit
0 Prn/Plt Err Set when an printing or plotting error
occurs
1 Math Error Set when an internal math error occurs
2 RAM Error Set when RAM Memory test finds an error
3 Disk Error Set when a disk error occurs
4 ROM Error Set when ROM Memory test finds an error
5 A/D Error Set when A/D test finds an error
6 DSP Error Set when DSP test finds an error
7 Overload Set when the signal input overloads
xi
SR770 FFT SPECTRUM ANALYZER
xii
YOUR FIRST MEASUREMENT
GETTING STARTED
This sample measurement is designed to acquaint
the first time user with the SR770 Network
Analyzer. Do not be concerned that your
measurement does not exactly agree with this
exercise. The focus of this measurement exercise
is to learn how to use the instrument.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function is
labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
Hardkeys
The keypad consists of five groups of hardkeys. The
ENTRY keys are used to enter numeric parameters
which have been highlighted by a softkey. The
MENU keys select a menu of softkeys. Pressing a
menu key will change the menu boxes which are
displayed next to the softkeys. Each menu groups
together similar parameters and functions. The
CONTROL keys start and stop actual data
acquisition, select the marker and toggle the active
trace the display. These keys are not in a menu
since they are used frequently while displaying any
menu. The SYSTEM keys output the screen to a
printer and display help messages. These keys can
also be accessed from any menu. The MARKER
keys determine the marker mode and perform
various marker functions. The marker functions can
be accessed from any menu.
The SR770 has a menu driven user interface. The 6
softkeys to the right of the video display have
different functions depending upon the information
displayed in the menu boxes at the right of the
video display. In general, the softkeys have two
uses. The first is to toggle a feature on and off or to
choose between settings. The second is to highlight
a parameter which is then changed using the knob
or numeric keypad. In both cases, the softkey
affects the parameter which is displayed adjacent to
it.
Knob
The knob is used to adjust parameters which have
been highlighted using the softkeys. Most numeric
entry fields may be adjusted with the knob. In
addition, functions such as display zooming and
scrolling use the knob as well. In these cases, the
knob function is selected by the softkeys. The
[MARKER] key, which can be pressed at any time,
will set the knob function to scrolling the marker.
Example Measurement
This measurement is designed to investigate the
spectrum of a 1 kHz sine wave. You will need a
function generator capable of providing a 1 kHz sine
wave at a level of 100 mV to 1 V, such as the SRS
DS345. The actual settings of the generator are not
important since you will be using the SR770 to
measure and analyze its output. Choose a
generator which has some distortion (at least -70
dBc) or use a square or triangle wave.
Specifically, you will measure the spectrum of the
sine wave, measure its frequency, and measure its
harmonic distortion.
Softkeys
1-1
GETTING STARTED
ANALYZING A SINE WAVE
1.Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the power-on
tests are completed.
2.Turn on the generator, set the frequency to
1 kHz and the amplitude to approximately 1
Vrms.
Connect the generator's output to the A input
of the analyzer.
3.Press [AUTO RANGE]
4.Press the <Span> softkey to highlight the
span. Use the knob to adjust the span to
6.25 kHz.
You can also use the [SPAN UP] and
[SPAN DOWN] keys to adjust the span.
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The input impedance of the analyzer is 1 M. The
generator may require a terminator. Many
generators have either a 50 or 600 output
impedance. Use the appropriate feedthrough
termination if necessary. In general, not using a
terminator means that the output amplitude will not
agree with the generator setting and the distortion
may be greater than normal.
Since the signal amplitude may not be set
accurately, let the analyzer automatically set its
input range to agree with the actual generator
signal. Note that the range readout at the bottom of
the screen is displayed in inverse when the
autoranging is on.
Set the span to display the 1 kHz signal and its
first few harmonics.
You can also use the numeric keypad to enter the
span. In this case, the span will be rounded to the
next largest allowable span.
5.Press [MARKER MAX/MIN]
6.Use the knob to move the marker around.
Take a look at some of the harmonics.
7. Let's measure the frequency exactly.
This centers the marker region around the largest
data point on the graph. The marker should now be
on the 1 kHz signal. The marker readout above the
graph displays the frequency and amplitude of the
signal.
The [MARKER MAX/MIN] key can also be
configured to search for the minimum point on the
graph.
Pressing the [MARKER MAX/MIN] key also selects
the knob to adjust the marker position. The Span
Menu box becomes unhighlighted. A box is drawn
around the marker readout to indicate that the knob
will move the marker.
1-2
GETTING STARTED
Decrease the span to 1.56 kHz using the
<Span> key and knob, the [SPAN DOWN]
key or by entering the span numerically.
Press [MARKER MAX/MIN]
Press [MARKER CENTER]
8.Decrease the span to 97.5 Hz using the
<Span> key and knob, the [SPAN DOWN]
key or by entering the span numerically.
9.Press [MARKER MAX/MIN]
This isolates the 1 kHz fundamental frequency.
Move the marker to the peak at 1 kHz.
This sets the span center frequency to the marker
frequency. The signal will be at the center of the
span. Further adjustments to the span will keep the
center frequency fixed.
You may notice that the spectrum takes a while to
settle down at this last span. This is because the
frequency resolution is 1/400 of the span or
244 mHz. This resolution requires at least
4.096 seconds of time data. Note that the Settling
indicator at the lower left corner of the display will
stay on while the data settles.
This centers the marker more accurately. The
frequency of the signal can now be read with
244 mHz resolution.
10. Press [AUTO SCALE]
11. Press [ANALYZE]
Press <Harmonic>
12. Press <Next Harmonic Right>
This key adjusts the graph scale and top reference
to display the entire range of the data. You can
press this key at any time to optimize the graph
display.
Display the Analysis menu.
Select Harmonic analysis. The menu displays the
harmonic analysis menu. Notice that the
fundamental frequency (first menu box) has been
set to the frequency of the marker.
We used a narrow span to get an accurate reading
of the fundamental signal frequency. We will use
this measurement of the fundamental to accurately
locate the harmonics.
The harmonic measurement readout at the upper
left corner of the graph is under range because the
span is not wide enough to include any harmonics.
This centers the span around the second harmonic
(approx. 2 kHz). You are now making an accurate
measurement of the 2nd harmonic content of the
signal.
1-3
GETTING STARTED
Use the <Next Harmonic Right> and
<Next Harmonic Left> keys to investigate the
harmonics of the signal.
Now adjust the span to 12.5 kHz using the
<Span> key and knob, the [SPAN DOWN]
key or by entering the span numerically.
With this narrow span, the harmonics should be
easily visible.
Let's have the analyzer measure the distortion for
us. First return to full span by displaying the
frequency menu and choosing full span.
Return the graph to a scale where the fundamental
is on screen.
This highlights the Start Frequency menu box. It
also fixes the start frequency when the span is
adjusted.
Reduce the span to resolve the first few harmonics
of the signal.
15. Press [ANALYZE]
Press <Harmonic>
16. Press <# Harmonics>
Press [1] [1] <Enter>
17. Now let's measure some harmonics using the
reference marker.
Press <Return>
Display the Analysis menu.
Choose Harmonic analysis. (It should still be on
from before.) The fundamental frequency should still
be accurately set.
Highlight the number of harmonics menu box.
Enter 11 for the number of harmonics.
Notice that harmonic markers (little triangles)
appear on top of all of the harmonic peaks. These
indicate which data points are used in the
harmonic calculations.
The harmonic calculations are displayed in the
upper left corner of the graph. The top reading is the
harmonic level (absolute units) and the lower
reading is the distortion (harmonic level divided by
the fundamental level).
Return the menu display to the main Analysis
menu.
Press <None>
Choose No analysis. This turns off the harmonic
indicators and calculations.
1-4
Press [MARKER MAX/MIN]
Press [MARKER REF]
Use the knob to measure the harmonic levels
relative to the fundamental.
18. Press [MARKER REF]
GETTING STARTED
This moves the marker to the fundamental peak.
This sets the marker reference or offset to the
frequency and amplitude of the fundamental. The
marker readout above the graph now reads relative
to this offset. This is indicated by the in front of
the marker readout. A small star shaped symbol is
located at the screen location of the reference.
The [MARKER REF] key also allows the knob to
move the marker.
The marker readout is now relative to the reference
or fundamental level.
Pressing [MARKER REF] again removes the
marker offset.
This concludes this measurement example. You
should have a good feeling for the basic operation of
the menus, knob and numeric entry, and marker
movement and measurements.
1-5
GETTING STARTED
SECOND MEASUREMENT EXAMPLE
This sample measurement is designed to further
acquaint the user with the SR770 Network
Analyzer. Do not be concerned that your
measurement does not exactly agree with this
exercise. The focus of this measurement exercise
is to learn how to use the instrument.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function is
labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
The Measurement
This measurement is designed to investigate the
noise of an audio amplifier. You will need an audio
frequency amplifier such as the SRS SR560. We
will use the SR770's source to provide the test
signal.
Specifically, you will measure the output
signal/noise ratio of the amplifier and its input noise
level.
1-6
MEASURING AMPLIFIER NOISE
GETTING STARTED
1.Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the power-on
tests are completed.
2.Turn on the amplifier and set its gain to at
least 20 dB.
Connect the amplifier output to the A input of
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The input impedance of the analyzer is 1 M. The
amplifier output may require a terminator. Many
instruments have either a 50 or 600 output
impedance. Use the appropriate feedthrough
termination if necessary. In general, not using a
terminator means that the output amplitude will not
agree with the instrument setting and the distortion
may be greater than normal.
Display the Source menu.
Turn on the Sine output.
Display the sine configuration menu.
Highlight the Sine Output Level. Leave the
frequency at 1 kHz.
Enter 100 mV (pk).
Connect the SR770 Source output to the
amplifier input.
3.Press [AUTO RANGE]
Press [FREQ]
Press [SPAN DOWN] until the span is
6.25 kHz
Press [AUTO SCALE]
4.Press [MARKER MAX/MIN]
Press [MARKER REF]
This should be a small enough signal for the
amplifier to handle. If not, simply lower the sine
output level to a suitable level.
Since the amplifier output amplitude may not be set
accurately, let the analyzer automatically set its
input range to the actual signal.
Display the Frequency menu.
Set the span to display the 1 kHz signal and its
first few harmonics.
Set the graph scaling to display the entire range of
the data.
Move the marker to the signal peak (1 kHz). The
marker should read an amplitude equal to the
source output level times the amplifier gain.
This turns on the marker offset and sets the
reference marker to the current marker position.
From now on, the marker will now read relative to
1-7
GETTING STARTED
the signal peak. A is displayed before the marker
readout to indicate that the reading is relative. A
small star symbol is located on the graph at the
marker offset position.
Use the knob to move the marker to a region
that is representative of the noise floor.
The [MARKER REF] key automatically allows the
knob to adjust the marker position. The marker is
now providing a direct reading of the signal to noise
ratio. Remember, this is the S/N for the
source/amplifier combination.
The [MARKER REF] key toggles the marker offset
on and off. We now want to turn the offset off.
Display the Measure menu.
Choose the Measurement type menu.
Select Power Spectral Density. The PSD
approximates the amplitude of the signal within a 1
Hz bandwidth located at each frequency bin. This
allows measurements taken with different
linewidths (spans) to be compared.
To get a better measurement of noise, a little
averaging can help.
Display the Average menu.
Select Exponential averaging.
Highlight the Number of Averages menu box.
Enter 20 averages.
Turn averaging on. Notice how the noise floor
approaches a more stable value. We are using
RMS averaging to determine the actual noise floor.
See the section on Averaging for a discussion of
the different types of averaging.
7.Press [MARKER]
Use the knob to move the marker to a region
representative of the noise floor.
The [MARKER] key allows the knob to move the
marker.
The Marker reading should be in dBV/Hz. This is
the output noise amplitude at the marker frequency,
normalized to a 1 Hz bandwidth. To generalize to
other bandwidths, multiply by the square root of the
bandwidth. This approximation only holds if the
noise is Gaussian in nature.
1-8
Press [MEAS]
Press <Units Menu>
Press <Volts RMS>
8.Disconnect the source output from the
amplifier. Leave the amplifier input terminated
(with 50 Ohms).
GETTING STARTED
Display the Measure menu.
Choose the Units menu.
Select Volts RMS as the display units.
The marker now reads in Volts RMS /Hz. This is
a typical way of specifying amplifier input noise
levels.
Now we are measuring the amplifier's output noise
with a shorted input. If you take the noise
measurement and divide by the amplifier gain, then
you will have the amplifier's input noise at the
frequency of the marker reading.
An FFT is a convenient tool for measuring amplifier
noise spectra since the noise at many frequencies
can be determined in a single measurement.
1-9
GETTING STARTED
USING TRIGGERS AND THE TIME RECORD
This sample measurement is designed to acquaint
the user with the triggering capabilities of the
SR770 Network Analyzer. Do not be concerned that
your measurement does not exactly agree with this
exercise. The focus of this measurement exercise
is to learn how to use the instrument.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function is
labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
The Measurement
This measurement is designed to investigate the
trigger and time record. You will need a function
generator capable of providing a 100 µs wide pulse
at 250 Hz with an amplitude of 1 V. The output
should have a DC level of 0V.
Specifically, you will measure the output spectrum
when the signal is triggered. In addition, the trigger
delay will be used to delay the signal within the
time record.
Make sure that you have read "The Time Record" in
the Analyzer Basics section before trying this
exercise.
1-10
TRIGGERING THE ANALYZER
GETTING STARTED
1.Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the power-on
tests are completed.
2.Turn on the generator and choose a pulsed
output waveform. Set the frequency to
250 Hz, the pulse width to 100 µs and the
amplitude to approximately 1 V. Make sure
that the DC level of the output is near 0V.
Connect the generator's output to the A input
of the analyzer.
3.Press [INPUT]
Press <Coupling> to choose DC
Press <Input Range>
Press [4] <dBV>
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section of this manual for a complete listing of the
settings.
The input impedance of the analyzer is 1 M. The
generator may require a terminator. Many
generators have either a 50 or 600 output
impedance. Use the appropriate feedthrough
termination if necessary. In general, not using a
terminator only means that the output amplitude
will not agree with the generator setting and the
distortion may be greater than normal.
Let's choose DC coupling and an input range that
doesn't overload.
Set the input range to 4 dBV. Adjust the pulse
amplitude so that no overloads occur.
The upper trace should display the pulse waveform
at the left edge. Auto scale will set the display
limits automatically. Remember that we are
displaying the magnitude of the signal. Any
negative portion of the signal will be folded back
around zero and appear as a positive magnitude.
Because the pulse is much shorter than the time
record, we need to use the Uniform window. The
other window functions taper to zero at the start
and end of the time record. Always be aware of the
effect windowing has on the FFT of the time record.
There should now be a spectrum on the lower
trace. Use [AUTO SCALE] to set the display.
The spectrum you see is the sinx/x envelope of a
rectangular pulse. The zeroes in the spectrum
occur at the harmonics of 1/pulse width (1/100µs or
10 kHz).
Now choose the Hanning window. Notice how the
spectrum goes away. We can get the spectrum
back by delaying the time record relative to the
trigger so that the pulse is positioned in the center
of the time record.
Highlight the Trigger Delay menu box.
Enter 256 samples of delay. Because the pulse
repetition rate is 250 Hz, the period between pulses
is exactly equal to one time record. So setting the
delay to half of a time record will place the pulse at
the middle of the record.
Remember that the time record only displays the
first 400 points (out of 512) so that the middle of the
record is not the middle of the display trace.
The spectrum should reappear on the lower trace.
This is because windowing preserves the central
part of the time record.
1-12
10. Press [4] [7] [5] <Samples>
11. Press <Trigger> to select Continuous
12. Press [MEAS]
Press <Window>
Press <Uniform>
GETTING STARTED
Let's delay the signal some more. Now we've
delayed the time record by almost a full period. The
pulse is now near the end of the time record.
Notice how the spectrum is greatly attenuated. This
is the effect of the window function attenuating the
start of the timer record.
Now if we go to continuous triggering, the time
record becomes unstable. The spectrum is also
unstable because of the windowing. Some time
records place the pulse at the middle, some at the
ends.
If we set the window back to Uniform, we find that
the spectrum does not vary with the position of the
pulse within the time record.
1-13
GETTING STARTED
USING THE DISK DRIVE
The disk drive on the SR770 may be used to store 3
types of files.
1.Data File
This includes the data in the active trace,
the measurement and display type, the
units and the graph scaling. In addition, the
associated data and limit tables are stored
in this file as well. Data files may be
recalled into either trace0 or trace1.
2.ASCII Data File
This file saves the data in the active trace in
ASCII format. These files may not be
recalled to the display. This format is
convenient when transferring data to a PC
application.
3.Settings File
This files stores the analyzer settings.
Recalling this file will change the analyzer
setup to that stored in the file.
The Measurement
This measurement is designed to familiarize the
user with the disk drive. We will use the SR770
source to provide a test signal so that there is some
data to save and recall.
Specifically, you will save and recall a data file and
a settings file.
The disk drive uses double-sided, high density
(DS/HD) 3.5" disks. The disk capacity is 1.44M.
The SR770 uses the DOS format. A disk which was
formatted on a PC or PS2 may also be used. Files
written by the SR770 may be copied or read on a
DOS computer.
Data files can store data in either binary or ascii
format. Binary format uses less disk space. Ascii
format allows trace data to be read by other
programs using a PC.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function is
labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
1-14
STORING AND RECALLING DATA
GETTING STARTED
1.Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the power-on
tests are completed.
2.Connect the SR770 Source output to the A
input.
Press [SOURCE]
Press <Sine>
3.Press [AUTO RANGE]
4.Press [SPAN DOWN] until the span is
6.25 kHz
5.Press [AUTO SCALE]
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
Display the Source menu.
Turn on the Sine output. Leave the sine frequency
and level at the default settings (1 kHz and 1 Vpk).
Let the analyzer automatically set its input range.
Set the span to display the 1 kHz signal.
Set the graph scaling to display the entire range of
the data.
6.Press [PAUSE CONT]
7.Put a blank double-sided, high density
(DS/HD)3.5" disk into the drive.
8.Press [STORE RECALL]
Press <Disk Utilities>
Press <Format Disk>
9.Press <Return>
Press <Save Data>
Stop data acquisition. The graph on the screen is
the one we want to save. (You can actually save
graphs while the analyzer is running.)
Use a blank if disk if possible, otherwise any disk
that you don't mind formatting will do. Make sure
the write protect tab is off.
Let's format this disk.
Display the Store and Recall menu.
Choose Disk Utilities.
Make sure that the disk does not contain any
information that you want. Formatting the disk
takes about a minute.
Go back to the main Store and Recall menu.
Display the Save Data menu.
Now we need a file name.
[ALT] lets you enter the letter characters printed
below each key. The numbers and backspace
function as normal.
Enter a file name such as DATA1 (or any legal
DOS file name).
This saves the active trace data to disk using the
file name specified above.
Display the disk catalog. This display lists all of the
files on the disk.
Save the data again using a new file name. This
way you can have multiple files in the disk catalog.
14. Press <Return>
Press [START]
Remove the input signal cable.
15. Press <Recall Data>
Press <Catalog>
16. Press [MARKER]
17. Press <Recall Data>
Go back to the main Store and Recall menu.
Resume data acquisition. The graph should be live
again.
Now we have a spectrum which is different from the
one we just saved. Recalling the data from disk will
restore the graph to what it was.
Display the Recall Data menu.
Display the disk catalog. The 2 files which you just
saved should be listed.
Pressing the [MARKER] key allows the knob to
adjust the marker. When the disk catalog is
displayed, the marker highlights a file. Use the
knob to choose a file to recall.
This recalls the data file from disk and displays it
on the active graph. Data acquisition is stopped so
that the graph is not updated. The file name is
1-16
Loading...
+ 237 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.