Safety and Preparation for Use III
Specifications V
Abridged Command List VI
GETTING STARTED
Your First Measurement 1-1
Analyzing a Sine Wave 1-2
Second Measurement Example 1-5
Amplifier Noise Level 1-6
Using Triggers and the Time Record 1-8
Using the Disk Drive 1-12
Using Data Tables 1-17
Using Limit Tables 1-20
Using Trace Math 1-24
Things to Watch Out For 1-28
ANALYZER BASICS
What is an FFT Spectrum Analyzer? 2-1
Frequency Spans 2-2
The Time Record 2-3
Measurement Basics 2-4
Display Type 2-5
Windowing 2-6
Averaging 2-7
Real Time Bandwidth and Overlap 2-8
Input Range 2-9
OPERATION
Front Panel3-1
Power On/Off 3-1
Video Display 3-1
Soft Keys 3-2
Keypad 3-2
Spin Knob 3-2
Disk Drive 3-2
BNC Connectors 3-2
Screen Display 3-3
Data Display 3-3
Single/Dual Trace Displays 3-3
Marker Display 3-5
Menu Display 3-5
Status Indicators 3-5
Keypad 3-7 Normal and Alternate Keys 3-7
Menu Keys 3-7
Entry Keys 3-8
START and PAUSE/CONT 3-8
MARKER 3-8
ACTIVE TRACE 3-9
AUTO RANGE 3-9
AUTOSCALE 3-9
SPAN UP/DOWN 3-9
MARKER ENTRY 3-9
MARKER MODE 3-9
MARKER REF 3-9
MARKER CENTER 3-9
MARKER MAX/MIN 3-9
PRINT 3-10
HELP 3-10
LOCAL 3-10
Rear Panel3-11
Power Entry Module 3-11
IEEE-488 Connector 3-11
RS232 Connector 3-11
Parallel Printer Connector 3-11
PC Keyboard Connector 3-11
MENUS
Frequency Menu 4-1
Measure Menu 4-3
Display Menu 4-15
Marker Mode Menu 4-17
Input Menu 4-19
Scale Menu 4-25
Analyze Menu 4-27
Average Menu 4-43
Plot Menu 4-47
Setup Menu 4-51
Store/Recall Menu 4-6
Default Settings 4-7
PROGRAMMING
GPIB Communications 5-1
RS232 Communications 5-1
Status Indicators and Queues 5-1
Command Syntax 5-1
Interface Ready and Status 5-2
Detailed Command List 5-3
Frequency Commands 5-4
Measurement Commands 5-5
Display and Marker Commands 5-6
Scale Commands 5-8
Input Commands 5-9
Analysis Commands 5-10
Data Table Commands 5-11
Limit Table Commands 5-12
Averaging Commands 5-13
Print and Plot Commands 5-14
Setup Commands 5-15
Store and Recall Commands 5-17
Trace Math Commands 5-18
Front Panel Control Commands 5-19
Data Transfer Commands 5-20
Interface Commands 5-22
Status Reporting Commands 5-23
4
2
i
TABLE OF CONTENTS
Status Byte Definitions 5-24
Serial Poll Status Byte 5-24
Standard Event Status Byte 5-24 FFT Status Byte 5-25
Error Status Byte 5-25
Program Examples Microsoft C, Nat'l Instruments GPIB 5-26
BASIC, Nat'l Instruments GPIB 5-29
TESTING
Introduction 6-1
Preset 6-1
Serial Number 6-1
Firmware Revision 6-1
General Installation 6-1
Necessary Equipment 6-3
If A Test Fails 6-3
Performance Tests
Self Tests 6-4
DC Offset 6-5
Common Mode Rejection 6-7
Amplitude Accuracy and Flatness 6-8
Amplitude Linearity 6-11
Anti-Alias Filter Attenuation 6-13
Frequency Accuracy 6-14
Phase Accuracy 6-15
Harmonic Distortion 6-17
Noise and Spurious Signals 6-19
Performance Test Record 6-21
CIRCUIT DESCRIPTION
Circuit Boards 7-1
Video Driver and CRT 7-1
CPU Board
Microprocessor System 7-3
Keypad Interface 7-3
Keyboard Interface 7-3
Spin Knob 7-4
Analog Input Board
Overview 7-7
Input Amplifier 7-7
Gain Stages and Attenuators 7-7
Anti-Alias Filter 7-7
A/D Converter 7-8
I/O Interface 7-8
Power 7-80
Parts Lists
CPU Board 7-9
Power Supply Board 7-13
DSP Logic Board 7-16 Analog Input Board 7-18
Chassis Assembly 7-26
Miscellaneous 7-28
Schematic Diagrams
CPU Board
Power Supply Board
DSP Logic Board
Analog Input Board
ii
SR760 FFT SPECTRUM ANALYZER
SAFETY AND PREPARATION FOR USE
WARNING
Dangerous voltages, capable of causing injury or death, are present in this instrument. Use extreme caution
whenever the instrument covers are removed. Do not remove the covers while the unit is plugged into a live
outlet.
CAUTION
This instrument may be damaged if operated with
the LINE VOLTAGE SELECTOR set for the wrong
AC line voltage or if the wrong fuse is installed.
LINE VOLTAGE SELECTION
The SR760 operates from a 100V, 120V, 220V, or
240V nominal AC power source having a line
frequency of 50 or 60 Hz. Before connecting the
power cord to a power source, verify that the LINE
VOLTAGE SELECTOR card, located in the rear
panel fuse holder, is set so that the correct AC
input voltage value is visible.
Conversion to other AC input voltages requires a
change in the fuse holder voltage card position
and fuse value. Disconnect the power cord, open
the fuse holder cover door and rotate the fuse-pull
lever to remove the fuse. Remove the small
printed circuit board and select the operating
voltage by orienting the printed circuit board so
that the desired voltage is visible when pushed
firmly into its slot. Rotate the fuse-pull lever back
into its normal position and insert the correct fuse
into the fuse holder.
LINE FUSE
Verify that the correct line fuse is installed before
connecting the line cord. For 100V/120V, use a 1
Amp fuse and for 220V/240V, use a 1/2 Amp fuse.
LINE CORD
The SR760 has a detachable, three-wire power
cord for connection to the power source and to a
protective ground. The exposed metal parts of the
instrument are connected to the outlet ground to
protect against electrical shock. Always use an
outlet which has a properly connected protective
ground.
SERVICE
Do not attempt to service or adjust this instrument
unless another person, capable of providing first
aid or resuscitation, is present.
Do not install substitute parts or perform any
unauthorized modifications to this instrument.
Contact the factory for instructions on how to
return the instrument for authorized service and
adjustment.
iii
SR760 FFT SPECTRUM ANALYZER
iv
SR760 FFT SPECTRUM ANALYZER
SPECIFICATIONS
FREQUENCY
Measurement Range 476 µHz to 100 kHz, baseband and zoomed.
Spans 191 mHz to 100 kHz in a binary sequence.
Center Frequency Anywhere within the measurement range subject to span and range limits.
Accuracy 25 ppm from 20°C to 40°C.
Resolution Span/400
Window Functions Blackman-Harris, Hanning, Flattop and Uniform.
Real-time Bandwidth 100 kHz
SIGNAL INPUT
Number of Channels 1
Input Single-ended or true differential
Input Impedance 1 MΩ, 15 pf
Coupling AC or DC
CMRR 90 dB at 1 kHz (Input Range < -6 dBV)
80 dB at 1 kHz (Input Range <14 dBV)
50 dB at 1 kHz (Input Range ≥ 14 dBV)
Noise 5 nVrms/√Hz at 1 kHz typical, 10 nVrms/√Hz max.
(-166 dBVrms/√Hz typ., -160 dBVrms/√Hz max.)
AMPLITUDE
Full Scale Input Range -60 dBV (1.0 mVpk) to +34 dBV (50 Vpk) in 2 dB steps.
Dynamic Range 90 dB typical
Harmonic Distortion No greater than -80 dB from DC to 100 kHz. (Input Range ≤ 0 dBV)
Spurious Input range ≥ -50 dBV: No greater than -85 dB below full scale below 200 Hz.
No greater than -90 dB below full scale to 100 kHz.
Input Sampling 16 bit A/D at 256 kHz
Accuracy ± 0.3 dB ± 0.02% of full scale (excluding windowing effects).
Averaging RMS, Vector and Peak Hold.
Linear and exponential averaging up to 64k scans.
TRIGGER INPUT
Modes Continuous, internal, external, or external TTL.
Internal Level: Adjustable to ±100% of input scale.
Positive or Negative slope.
Minimum Trigger Amplitude: 10% of input range.
External Level: ±5V in 40 mV steps. Positive or Negative slope.
Impedance: 10 kΩ Minimum Trigger Amplitude: 100 mV.
External TTL Requires TTL level to trigger (low<.7V, high>2V).
Post-Trigger Measurement record is delayed by 1 to 65,000 samples (1/512 to 127 time
records) after the trigger.
Delay resolution is 1 sample (1/512 of a record).
Pre-Trigger Measurement record starts up to 51.953 ms prior to the trigger.
Delay resolution is 3.9062 µs.
Phase Indeterminacy <2°
v
SR760 FFT SPECTRUM ANALYZER
DISPLAY FUNCTIONS
Display Real, imaginary, magnitude or phase spectrum.
Measurements Spectrum, power spectral density, time record and 1/3 octave.
Analysis Band, sideband, total harmonic distortion and trace math.
Graphic Expand Display expands up to 50x about any point in the display.
MARKER FUNCTIONS
Harmonic Marker Displays up to 400 harmonics of the fundamental.
Delta Marker Reads amplitude and frequency relative to defined reference.
Next Peak/Harmonic Locates nearest peak or harmonic to the left or right.
Data Tables Lists Y values of up to 200 user defined X points.
Limit Tables Automatically detects data exceeding up to 100 user defined upper and
lower limit trace segments.
GENERAL
Monitor Monochrome CRT. 640H by 480V resolution.
Adjustable brightness and screen position.
Interfaces IEEE-488, RS232 and Printer interfaces standard.
All instrument functions can be controlled through the IEEE-488 and RS232
interfaces. A PC keyboard input is provided for additional flexibility.
Hardcopy Screen dumps and table and setting listings to dot matrix and HP LaserJet
compatible printers. Data plots to HP-GL compatible plotters (via RS232 or
IEEE-488).
Disk 3.5 inch DOS compatible format, 720 kbyte capacity. Storage of data,
setups, data tables, and limit tables.
Power 60 Watts, 100/120/220/240 VAC, 50/60 Hz.
Dimensions 17"W x 6.25"H x 18.5"D
Weight 36 lbs.
Warranty One year parts and labor on materials and workmanship.
vi
SR760 FFT SPECTRUM ANALYZER
COMMAND LIST
VARIABLES g Trace0 (0), Trace1 (1), or Active Trace (-1)
i,j Integers
f Frequency (real)
x,y Real Numbers
s String
FREQUENCYpage
SPAN (?) {i} 5-4 Set (Query) the Frequency Span to 100 kHz (19) through 191 mHz (0).
STRF (?) {f} 5-4 Set (Query) the Start Frequency to f Hz.
CTRF (?) {f} 5-4 Set (Query) the Center Frequency to f Hz.
OTYP (?) {i} 5-4 Set (Query) the number of bands in Octave Analysis to 15 (0) or 30 (1).
OSTR (?) {i} 5-4 Set (Query) the Starting Band in Octave Analysis to -2 ≤ i ≤ 35.
WTNG (?) {i} 5-4 Set (Query) the Weighting in Octave Analysis to none (0) or A-weighting (1).
MEASUREMENTpage
MEAS (?) g {,i} 5-5 Set (Query) the Measurement Type to Spectrum (0), PSD (1), Time (2), or
DISP (?) g {,i} 5-5 Set (Query) the Display to LogMag (0),LinMag (1), Real (2), Imag (3), or Phase
UNIT (?) g {,i} 5-5 Set (Query) the Units to Vpk or deg (0),Vrms or rads (1), dBV (2), or dBVrms
VOEU (?) g {,i} 5-5 Set (Query) the Units to Volts (0), or EU (1).
EULB (?) g {,s} 5-5 Set (Query) the EU Label to string s.
EUVT (?) g {,x} 5-5 Set (Query) the EU Value to x EU/Volt.
WNDO (?) g {i} 5-5 Set (Query) the Window to Uniform (0), Flattop (1), Hanning (2), or BMH (3).
DISPLAY and MARKER page
ACTG (?) {i} 5-6 Set (Query) the Active Trace to trace0 (0) or trace1 (1).
FMTS (?) g {,i} 5-6 Set (Query) the Display Format to Single (0) or Dual (1) trace.
GRID (?) g {,i} 5-6 Set (Query) the Grid mode to Off (0), 8 (1), or 10 (2) divisions.
FILS (?) g {,i} 5-6 Set (Query) the Graph Style to Line (0) or Filled (1).
MRKR (?) g {,i} 5-6 Set (Query) the Marker to Off (0), On (1) or Track (2).
MRKW (?) g {,i} 5-6 Set (Query) the Marker Width to Norm (0), Wide (1), or Spot (2).
MRKM (?) g {,i} 5-6 Set (Query) the Marker Seeks mode to Max (0), Min (1), or Mean (2).
MRLK (?) {i} 5-6 Set (Query) the Linked Markers to Off (0) or On (1).
MBIN g,i 5-6 Move the marker region to bin i.
MRKX? 5-6 Query the Marker X position.
MRKY? 5-6 Query the Marker Y position.
MRPK 5-6 Move the Marker to the on screen max or min. Same as [MARKER MAX/MIN]
MRCN 5-6 Make the Marker X position the center of the span. Same as [MARKER
MRRF 5-6 Turns Marker Offset on and sets the offset equal to the marker position.
MROF (?) {i} 5-6 Set (Query) the Marker Offset to Off (0) or On (1).
MROX (?) {x} 5-6 Set (Query) the Marker Offset X value to x.
MROY (?) {x} 5-7 Set (Query) the Marker Offset Y value to x.
PKLF 5-7 Move the marker to the next peak to the left.
PKRT 5-7 Move the marker to the next peak to the right.
MSGS s 5-7 Display message s on the screen and sound an alarm.
SCALEpage
TREF (?) g {,x} 5-8 Set (Query) the Top Reference to x.
BREF (?) g {,x} 5-8 Set (Query) the Bottom Reference to x.
YDIV (?) g {,x} 5-8 Set (Query) the Vertical Scale (Y/Div) to x.
AUTS g 5-8 AutoScale graph g. Similar to the [AUTO SCALE] key.
EXPD (?) g {,i} 5-8 Set (Query) the Horizontal Expand to no expand (5), 128, 64, 30, 15, or 8 bins
ELFT (?) g {,i} 5-8 Set (Query) the Left Bin when expanded to bin i.
XAXS (?) g {,i} 5-8 Set (Query) the X Axis scaling to Linear (0) or Log (1).
description
description
Octave (3).
(4).
(3).
description
key.
CENTER] key.
description
(4-0).
vii
SR760 FFT SPECTRUM ANALYZER
INPUT page description
ISRC (?) {i} 5-9 Set (Query) the Input to A (0) or A-B (1).
IGND (?) {i} 5-9 Set (Query) the Input Grounding to Float (0) or Ground (1).
ICPL (?) {i} 5-9 Set (Query) the Input Coupling to AC (0) or DC (1).
IRNG (?) {i} 5-9 Set (Query) the Input Range to i dBV full scale. -60 ≤ i ≤ 34 and i is even.
ARNG (?) {i} 5-9 Set (Query) the Auto Range mode to Manual (0) or Auto (1).
AOFF 5-9 Perform Auto Offset calibration.
AOFM (?) {i} 5-9 Set (Query) the Auto Offset Mode to Off (0) or On (1).
TMOD (?) {i} 5-9 Set (Query) the Trigger Mode to Cont (0), Int (1), Ext (2), or Ext TTL(3).
TRLV (?) {x} 5-9 Set (Query) the Trigger Level to x percent. -100.0 ≤ x ≤ 99.22.
TDLY (?) {i} 5-9 Set (Query) the Trigger Delay to i samples. -13300≤ i ≤ 65000.
ARMM (?) {i} 5-9 Set (Query) the Arming Mode to Auto (0) or Manual (1).
ARMS 5-9 Manually arm the trigger.
ANALYSISpage
ANAM (?) g {,i} 5-10 Set (Query) the real time Analysis to None (0), Harmonic (1), Sideband (2), or
CALC? g,i 5-10 Query result i (0 or 1) of the latest real time analysis.
FUND (?) g {,f} 5-10 Set (Query) the Harmonic Fundamental to frequency f Hz.
NHRM (?) g {,i} 5-10 Set (Query) the Number of Harmonics to 0 ≤ i ≤ 400.
NHLT 5-10 Move the Marker or Center Frequency to the next harmonic to the left.
NHRT 5-10 Move the Marker or Center Frequency to the next harmonic to the right.
SBCA (?) g {,f} 5-10 Set (Query) the Sideband Carrier to frequency f Hz.
SBSE (?) g {,f} 5-10 Set (Query) the Sideband Separation to f Hz.
NSBS (?) g {,i} 5-10 Set (Query) the Number of Sidebands to 0 ≤ i ≤ 200.
BSTR (?) g {,f} 5-10 Set (Query) the Band Start to frequency f Hz.
BCTR (?) g {,f} 5-10 Set (Query) the Band Center to frequency f Hz.
BWTH (?) g {,f} 5-11 Set (Query) the Band Width to f Hz.
TABL 5-11 Turn on Data Table display for the active trace.
DTBL (?) g {,i}{,f} 5-11 Set (Query) Data Table line i to frequency f.
DINX (?) {i} 5-11 Set (Query) Data Table index to i.
DINS 5-11 Insert a new line in the data table.
DIDT 5-11 Delete a line from the data table.
DLTB 5-11 Delete the entire data table.
LIMT 5-12 Turn on Limit Table display for the active trace.
TSTS (?) {i} 5-12 Set (Query) the Limit Testing to Off (0) or On (1).
PASF? 5-12 Query the results of the latest limit test. Pass=0 and Fail=1.
LTBL (?) g {,i} {j,f1,f2,y1,y2} 5-12 Set (Query) Limit Table line i to Xbegin (f1), Xend (f2), Y1 and Y2.
LINX (?) {i} 5-12 Set (Query) Limit Table index to i.
LINS 5-12 Insert a new line in the limit table.
LIDT 5-12 Delete a line from the limit table.
LLTB 5-12 Delete the entire limit table.
LARM (?) {i} 5-12 Set (Query) the Audio Limit Fail Alarm to Off (0) or On (1).
AVERAGINGpage
AVGO (?) {i} 5-13 Set (Query) Averaging to Off (0) or On (1).
NAVG(?) {i} 5-13 Set (Query) the Number of Averagesto 2 ≤ i ≤ 32000.
AVGT (?) {i} 5-13 Set (Query) the Averaging Type to RMS (0), Vector (1), or Peak Hold (2).
AVGM (?) {i} 5-13 Set (Query) the Averaging Mode to Linear (0) or Exponential (1).
OVLP (?) {x} 5-13 Set (Query) the Overlap to x percent. 0 ≤ x ≤ 100.0.
PLOTpage
PLOT 5-14 Plot the entire graph (or graphs).
PTRC 5-14 Plot the trace (or traces) only.
PMRK 5-14 Plot the marker (or markers) only.
PTTL (?) {s} 5-14 Set (Query) the Plot Title to string s.
PSTL (?) {s} 5-14 Set (Query) the Plot Subtitle to string s.
PRSC 5-14 Print the screen. Same as the [PRINT] key.
PSET 5-14 Print the analyzer settings.
PLIM 5-14 Print the Limit Table of the active graph.
PDAT 5-14 Print the Data Table of the active graph.
SETUPpage
OUTP (?) {i} 5-15 Set (Query) the Output Interface to RS232 (0) or GPIB (1).
description
Band (3).
description
description
description
viii
SR760 FFT SPECTRUM ANALYZER
OVRM (?) {i} 5-15 Set (Query) the GPIB Overide Remote state to Off (0) or On (1).
KCLK (?) {i} 5-15 Set (Query) the Key Click to Off (0) or On (1).
ALRM (?) {i} 5-15 Set (Query) the Alarms to Off (0) or On (1).
THRS (?) {i} 5-15 Set (Query) the Hours to 0≤ i ≤ 23.
TMIN (?) {i} 5-15 Set (Query) the Minutes to 0 ≤ i ≤ 59.
TSEC (?) {i} 5-15 Set (Query) the Seconds to 0 ≤ i ≤ 59.
DMTH (?) {i} 5-15 Set (Query) the Month to 1 ≤ 1 ≤ 12.
DDAY (?) {i} 5-15 Set (Query) the Day to 1 ≤ 1 ≤ 31.
DYRS (?) {i} 5-15 Set (Query) the Year to 0 ≤ 1 ≤ 99.
PLTM (?) {i} 5-15 Set (Query) the Plotter Mode to RS232 (0) or GPIB (1).
PLTB (?) {i} 5-15 Set (Query) the Plotter Baud Rate to 300 (0), 1200 (1), 2400 (2), 4800 (3),
9600 (4).
PLTA (?) {i} 5-15 Set (Query) the Plotter GPIB Address to 0 ≤ i ≤ 30.
PLTS (?) {i} 5-15 Set (Query) the Plot Speed to Fast (0) or Slow (1).
PNTR (?) {i} 5-15 Set (Query) the Trace Pen Number to 1 ≤ i ≤ 6.
PNGD (?) {i} 5-15 Set (Query) the Grid Pen Number to 1 ≤ i ≤ 6.
PNAP (?) {i} 5-15 Set (Query) the Alphanumeric Pen Number to 1 ≤ i ≤ 6.
PNCR (?) {i} 5-16 Set (Query) the Cursor Pen Number to 1 ≤ i ≤ 6.
PRNT (?) {i} 5-16 Set (Query) the Printer Type to Epson (0) or HP (1).
STORE AND RECALL FILEpage
FNAM (?) {s} 5-17 Set (Query) the current File Name to string.
SVTR 5-17 Save the Active Trace Data to the file specified by FNAM.
SVST 5-17 Save the Settings to the file specified by FNAM.
RCTR 5-17 Recall the Trace Data from the file specified by FNAM to the active graph.
RCST 5-17 Recall the Settings from the file specified by FNAM.
MATH OPERATIONSpage
CSEL (?) {i} 5-18 Set (Query) the Operation to +, -, x, /, log, √ (0-5).
COPR 5-18 Start the calculation.
CARG (?) {i} 5-18 Set (Query) the Argument type to Constant (0), w (1), or Other Graph (2).
CONS (?) {x} 5-18 Set (Query) the Constant Argument to x.
CMRK 5-18 Set the Constant Argument to the Y value of the marker.
FRONT PANEL CONTROLSpage
STRT 5-19 Start data acquisition. Same as [START] key.
STCO 5-19 Pause or Continue data acquisition. Same as [PAUSE CONT] key.
PRSC 5-19 Print the screen. Same as [PRINT] key.
ACTG (?) {i} 5-19 Set (Query) the Active Trace to trace0 (0) or trace1 (1). Similar to [ACTIVE
ARNG (?) {i} 5-19 Set (Query) the Auto Range mode to Manual (0) or Auto (1). Similar to [AUTO
AUTS 5-19 AutoScale the graph. Same as the [AUTO SCALE] key.
DATA TRANSFERpage
SPEC? g {,i} 5-20 Query the Y value of bin 0 ≤ i ≤ 399.
BVAL? g, i 5-20 Query the X value of bin 0 ≤ i ≤ 399.
SPEB? g 5-20 Binary dump the entire trace g.
BDMP (?) g, {,i} 5-21 Set (Query) the auto binary dump mode for trace g.
INTERFACEpage
*RST 5-22 Reset the unit to its default configurations.
*IDN? 5-22 Read the SR760 device identification string.
LOCL(?) {i} 5-22 Set (Query) the Local/Remote state to LOCAL (0), REMOTE (1), or LOCAL
OVRM (?) {i} 5-22 Set (Query) the GPIB Overide Remote state to Off (0) or On (1).
STATUS page
*CLS 5-23 Clear all status bytes.
*ESE (?) {i} {,j} 5-23 Set (Query) the Standard Status Byte Enable Register to the decimal value i
*ESR? {i} 5-23 Query the Standard Status Byte. If i is included, only bit i is queried.
*SRE (?) {i} {,j} 5-23 Set (Query) the Serial Poll Enable Register to the decimal value i (0-255).
*STB? {i} -23 Query the Serial Poll Status Byte. If i is included, only bit i is queried.
description
description
description
TRACE] key.
RANGE] key.
description
description
LOCKOUT (2).
description
(0-255).
ix
SR760 FFT SPECTRUM ANALYZER
*PSC (?) {i} 5-23 Set (Query) the Power On Status Clear bit to Set (1) or Clear (0).
ERRE (?) {i} {,j} 5-23 Set (Query) the Error Status Enable Register to the decimal value i (0-255).
ERRS? {i} 5-23 Query the Error Status Byte. If i is included, only bit i is queried.
FFTE (?) {i} {,j} 5-23 Set (Query) the FFT Status Enable Register to the decimal value i (0-255).
FFTS? {i} 5-23 Query the FFT Status Byte. If i is included, only bit i is queried.
STATUS BYTE DEFINITIONS
SERIAL POLL STATUS BYTE (6-24)
name usage
bit
0 SCN No measurements in progress
1 IFC No command execution in progress
2 ERR Unmasked bit in error status byte set
3 FFT Unmasked bit in FFT status byte set
4 MAV The interface output buffer is non-empty
5 ESB Unmasked bit in standard status byte
set
6 SRQ SRQ (service request) has occurred
7 Unused
STANDARD EVENT STATUS BYTE (6-25)
name usage
bit
0 INP Set on input queue overflow
1 Limit Fail Set when a limit test fails
2 QRY Set on output queue overflow
3 Unused
4 EXE Set when command execution error
occurs
5 CMD Set when an illegal command is
received
6 URQ Set by any key press or knob rotation
7 PON Set by power-on
FFT STATUS BYTE (6-25)
name usage
bit
0 Triggered Set when a time record is triggered
1 Prn/Plt Set when a printout or plot is completed
2 NewData 0 Set when new data is available for trace 0
3 NewData 1 Set when new data is available for trace 1
4 Avg Set when a linear average is completed
5 AutoRng Set when auto range changes the range
6 High Voltage Set when high voltagedetected at input
7 Settle Set when settling is complete
ERROR STATUS BYTE (6-26)
name usage
bit
0 Prn/Plt Err Set when an printing or plotting error
occurs
1 Math Error Set when an internal math error occurs
2 RAM Error Set when RAM Memory test finds an error
3 Disk Error Set when a disk error occurs
4 ROM Error Set when ROM Memory test finds an error
5 A/D Error Set when A/D test finds an error
6 DSP Error Set when DSP test finds an error
7 Overload Set when the signal input overloads
x
YOUR FIRST MEASUREMENT
This sample measurement is designed to acquaint
the first time user with the SR760 Spectrum
Analyzer. Do not be concerned that your
measurement does not exactly agree with this
exercise. The focus of this measurement exercise
is to learn how to use the instrument.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function
is labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
Hardkeys
The keypad consists of five groups of hardkeys.
The ENTRY keys are used to enter numeric
parameters which have been highlighted by a
softkey. The MENU keys select a menu of
softkeys. Pressing a menu key will change the
menu boxes which are displayed next to the
softkeys. Each menu groups together similar
parameters and functions. The CONTROL keys
start and stop actual data acquisition, select the
marker and toggle the active trace the display.
These keys are not in a menu since they are used
frequently while displaying any menu. The
SYSTEM keys output the screen to a printer and
display help messages. These keys can also be
accessed from any menu. The MARKER keys
determine the marker mode and perform various
marker functions. The marker functions can be
accessed from any menu.
GETTING STARTED
Softkeys
The SR760 has a menu driven user interface. The
6 softkeys to the right of the video display have
different functions depending upon the information
displayed in the menu boxes at the right of the
video display. In general, the softkeys have two
uses. The first is to toggle a feature on and off or
to choose between settings. The second is to
highlight a parameter which is then changed using
the knob or numeric keypad. In both cases, the
softkey affects the parameter which is displayed
adjacent to it.
Knob
The knob is used to adjust parameters which have
been highlighted using the softkeys. Most numeric
entry fields may be adjusted with the knob. In
addition, functions such as display zooming and
scrolling use the knob as well. In these cases, the
knob function is selected by the softkeys. The
[MARKER] key, which can be pressed at any time,
will set the knob function to scrolling the marker.
Example Measurement
This measurement is designed to investigate the
spectrum of a 1 kHz sine wave. You will need a
function generator capable of providing a 1 kHz
sine wave at a level of 100 mV to 1 V, such as the
SRS DS345. The actual settings of the generator
are not important since you will be using the
SR760 to measure and analyze its output. Choose
a generator which has some distortion (at least 70 dBc) or use a square or triangle wave.
Specifically, you will measure the spectrum of the
sine wave, measure its frequency, and measure
its harmonic distortion.
1-1
GETTING STARTED
ANALYZING A SINE WAVE
1. Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the poweron tests are completed.
2. Turn on the generator, set the frequency to
1 kHz and the amplitude to approximately 1
Vrms.
Connect the generator's output to the A
input of the analyzer.
3. Press [AUTO RANGE]
4. Press the <Span> softkey to highlight the
span. Use the knob to adjust the span to
6.25 kHz.
You can also use the [SPAN UP] and
[SPAN DOWN] keys to adjust the span.
5. Press [MARKER MAX/MIN]
6. Press [MARKER]
Use the knob to move the marker around.
Take a look at some of the harmonics.
7. Let's measure the frequency exactly.
Decrease the span to 1.56 kHz using the
<Span> key and knob, the [SPAN DOWN]
key or by entering the span numerically.
Press [MARKER MAX/MIN]
Press [MARKER CENTER]
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The input impedance of the analyzer is 1 MΩ. The
generator may require a terminator. Many
generators have either a 50 Ω or 600 Ω output
impedance. Use the appropriate feedthrough
termination if necessary. In general, not using a
terminator means that the output amplitude will
not agree with the generator setting and the
distortion may be greater than normal.
Since the signal amplitude may not be set
accurately, let the analyzer automatically set its
input range to agree with the actual generator
signal. Note that the range readout at the bottom
of the screen is displayed in inverse when the
autoranging is on.
Set the span to display the 1 kHz signal and its
first few harmonics.
You can also use the numeric keypad to enter the
span. In this case, the span will be rounded to the
next largest allowable span.
This centers the marker region around the largest
data point on the graph. The marker should now
be on the 1 kHz signal. The marker readout above
the graph displays the frequency and amplitude of
the signal.
The [MARKER MAX/MIN] key can also be
configured to search for the minimum point on the
graph.
Pressing the [MARKER] key allows the knob to
adjust the marker position. The Span Menu box
becomes unhighlighted. A box is drawn around
the marker readout to indicate that the knob will
move the marker.
This isolates the 1 kHz fundamental frequency.
Move the marker to the peak at 1 kHz.
This sets the span center frequency to the marker
frequency. The signal will be at the center of the
1-2
8. Decrease the span to 97.5 Hz using the
<Span> key and knob, the [SPAN DOWN]
key or by entering the span numerically.
9. Press [MARKER MAX/MIN]
10. Press [AUTO SCALE]
11. Press [ANALYZE]
Press <Harmonic>
12. Press <Next Harmonic Right>
Use the <Next Harmonic Right> and
<Next Harmonic Left> keys to investigate
the harmonics of the signal.
13. Press [FREQ]
Press <Full Span>
Press [AUTO SCALE]
14. Press <Start Freq.>
Now adjust the span to 12.5 kHz using the
GETTING STARTED
span. Further adjustments to the span will keep
the center frequency fixed.
You may notice that the spectrum takes a while to
settle down at this last span. This is because the
frequency resolution is 1/400 of the span or
244 mHz. This resolution requires at least
4.096 seconds of time data. Note that the Settling
indicator at the lower right corner of the display
will stay on while the data settles.
This centers the marker more accurately. The
frequency of the signal can now be read with
244 mHz resolution.
This key adjusts the graph scale and top
reference to display the entire range of the data.
You can press this key at any time to optimize the
graph display.
Display the Analysis menu.
Select Harmonic analysis. The menu displays the
harmonic analysis menu. Notice that the
fundamental frequency (first menu box) has been
set to the frequency of the marker.
We used a narrow span to get an accurate
reading of the fundamental signal frequency. We
will use this measurement of the fundamental to
accurately locate the harmonics.
The harmonic measurement readout at the upper
left corner of the graph is under range because
the span is not wide enough to include any
harmonics.
This centers the span around the second
harmonic (approx. 2 kHz). You are now making
an accurate measurement of the 2nd harmonic
content of the signal.
With this narrow span, the harmonics should be
easily visible.
Let's have the analyzer measure the distortion for
us. First return to full span by displaying the
frequency menu and choosing full span.
Return the graph to a scale where the
fundamental is on screen.
This highlights the Start Frequency menu box. It
also fixes the start frequency when the span is
adjusted.
Reduce the span to resolve the first few
1-3
GETTING STARTED
<Span> key and knob, the [SPAN DOWN]
key or by entering the span numerically.
15. Press [ANALYZE]
Press <Harmonic>
16. Press <# Harmonics>
Press [1] [1] <Enter>
17. Now let's measure some harmonics using
the reference marker.
Press <Return>
Press <None>
Press [MARKER MAX/MIN]
Press [MARKER REF]
Press [MARKER]
Use the knob to measure the harmonic
levels relative to the fundamental.
18. Press [MARKER REF]
harmonics of the signal.
Display the Analysis menu.
Choose Harmonic analysis. (It should still be on
from before.) The fundamental frequency should
still be accurately set.
Highlight the number of harmonics menu box.
Enter 11 for the number of harmonics.
Notice that harmonic markers (little triangles)
appear on top of all of the harmonic peaks. These
indicate which data points are used in the
harmonic calculations.
The harmonic calculations are displayed in the
upper left corner of the graph. The top reading is
the harmonic level (absolute units) and the lower
reading is the distortion (harmonic level divided by
the fundamental level).
Return the menu display to the main Analysis
menu.
Choose No analysis. This turns off the harmonic
indicators and calculations.
This moves the marker to the fundamental peak.
This sets the marker reference or offset to the
frequency and amplitude of the fundamental. The
marker readout above the graph now reads
relative to this offset. This is indicated by the ∆ in
front of the marker readout. A small star shaped
symbol is located at the screen location of the
reference.
This allows the knob to move the marker.
The marker readout is now relative to the
reference or fundamental level.
Pressing [MARKER REF] again removes the
marker offset.
This concludes this measurement example. You
should have a good feeling for the basic operation
of the menus, knob and numeric entry, and
marker movement and measurements.
1-4
SECOND MEASUREMENT EXAMPLE
This sample measurement is designed to further
acquaint the user with the SR760 Spectrum
Analyzer. Do not be concerned that your
measurement does not exactly agree with this
exercise. The focus of this measurement exercise
is to learn how to use the instrument.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function
is labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
GETTING STARTED
The Measurement
This measurement is designed to investigate the
noise of an audio amplifier. You will need an audio
frequency amplifier such as the SRS SR560. You
will also need a function generator capable of
providing a 1 kHz sine wave at a level of 100 mV
to 1 V such as the SRS DS345.
Specifically, you will measure the output
signal/noise ratio of the amplifier and its input
noise level.
1-5
GETTING STARTED
MEASURING AMPLIFIER NOISE
1. Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the poweron tests are completed.
2. Turn on the generator, set the frequency to
1 kHz and amplitude to approximately 1
Vrms.
Connect the generator's output to the input
of the amplifier. Turn on the amplifier and
set its gain to at least 20 dB. Connect the
amplifier output to the A input of the
analyzer.
3. Press [AUTO RANGE]
4. Press [SPAN DOWN] until the span is
6.25 kHz
5. Press [AUTO SCALE]
6. Press [MAX/MIN]
7. Press [MARKER REF]
Use the knob to move the marker to a
region that is representative of the noise
floor.
7. Press [MARKER REF] again
8. Press {MEAS]
Press <Measure Menu>
Press <PSD>
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The input impedance of the analyzer is 1 MΩ. The
generator and/or amplifier may require a
terminator. Many instruments have either a 50 Ω
or 600 Ω output impedance. Use the appropriate
feedthrough termination if necessary. In general,
not using a terminator means that the output
amplitude will not agree with the instrument
setting and the distortion may be greater than
normal.
Since the signal amplitude may not be set
accurately, let the analyzer automatically set its
input range to the actual signal.
Set the span to display the 1 kHz signal and its
first few harmonics.
Set the graph scaling to display the entire range
of the data.
Move the marker to the signal peak (1 kHz). The
marker should read an amplitude equal to the
generator output times the amplifier gain.
This turns on the marker offset and sets the
reference marker to the current marker position.
From now on, the marker will now read relative to
the signal peak. A ∆ is displayed before the
marker readout to indicate that the reading is
relative. A small star symbol is located on the
graph at the marker offset position.
The marker is now providing a direct reading of
the signal to noise ratio. Remember, this is the
S/N for the generator/amplifier combination. It
may be that the amplifier is better than the
generator. To check this, turn off the generator. If
the noise floor is lower, then the generator is
determining the output S/N.
The [MARKER REF] key toggles the marker offset
on and off. We now want to turn the offset off.
Display the Measure menu.
Choose the Measurement type menu.
Select Power Spectral Density. The PSD
approximates the amplitude of the signal within a
1 Hz bandwidth located at each frequency bin.
1-6
9. Press [AVERAGE]
Press <Average Mode>
10. Press <Number of Averages>
Press [2] [0] <Enter>
11. Press <Averaging>
12. Press [MARKER]
Use the knob to move the marker to a
region representative of the noise floor.
13. Press [MEAS]
Press <Units Menu>
Press <Volts RMS>
14. Disconnect the generator output from the
amplifier. Leave the amplifier input
terminated.
GETTING STARTED
This allows measurements taken with different
linewidths (spans) to be compared.
To get a better measurement of noise, a little
averaging can help.
Display the Average menu.
Select Exponential averaging.
Highlight the Number of Averages menu box.
Enter 20 averages.
Turn averaging on. Notice how the noise floor
approaches a more stable value. We are using
RMS averaging to determine the actual noise
floor. See the section on Averaging for a
discussion of the different types of averaging.
The [MARKER] key allows the knob to move the
marker.
The Marker reading should be in dBV/√Hz. This is
the output noise amplitude at the marker
frequency, normalized to a 1 Hz bandwidth. To
generalize to other bandwidths, multiply by the
square root of the bandwidth. This approximation
only holds if the noise is Gaussian in nature.
Display the Measure menu.
Choose the Units menu.
Select Volts RMS as the display units.
The marker now reads in Volts RMS /√Hz. This is
a typical way of specifying amplifier input noise
levels.
Now we are measuring the amplifier's output
noise with a shorted input. If you take the noise
measurement and divide by the amplifier gain,
then you will have the amplifier's input noise at the
frequency of the marker reading.
An FFT is a convenient tool for measuring
amplifier noise spectra since the noise at many
frequencies can be determined in a single
measurement.
1-7
GETTING STARTED
USING TRIGGERS AND THE TIME RECORD
This sample measurement is designed to acquaint
the user with the triggering capabilities of the
SR760 Spectrum Analyzer. Do not be concerned
that your measurement does not exactly agree
with this exercise. The focus of this measurement
exercise is to learn how to use the instrument.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function
is labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
The Measurement
This measurement is designed to investigate the
trigger and time record. You will need a function
generator capable of providing a 100 µs wide
pulse at 250 Hz with an amplitude of 1 V. The
output should have a DC level of 0V.
Specifically, you will measure the output spectrum
when the signal is triggered. In addition, the trigger
delay will be used to delay the signal within the
time record.
Make sure that you have read "The Time Record"
in the Analyzer Basics section before trying this
exercise.
1-8
TRIGGERING THE ANALYZER
1. Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the poweron tests are completed.
2. Turn on the generator and choose a pulsed
output waveform. Set the frequency to
250 Hz, the pulse width to 100 µs and the
amplitude to approximately 1 V. Make sure
that the DC level of the output is near 0V.
Connect the generator's output to the A
input of the analyzer.
3. Press [INPUT]
Press <Coupling> to choose DC
Press <Input Range>
Press [4] <dBV>
4. Press [DISPLAY]
Press <Format> to choose Up/Dn
5. Press [MEAS]
Press <Measure Menu>
Press <Time Record>
Press <Return>
Press <Display Menu>
Press <Linear Mag.>
6. Press [INPUT]
Press <Trigger Menu>
Press <Trigger> to select Internal
Press <Trigger Level>
Press [.] [5] <Volts>
Press [AUTO SCALE]
GETTING STARTED
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section of this manual for a complete listing of the
settings.
The input impedance of the analyzer is 1 MΩ. The
generator may require a terminator. Many
generators have either a 50 Ω or 600 Ω output
impedance. Use the appropriate feedthrough
termination if necessary. In general, not using a
terminator only means that the output amplitude
will not agree with the generator setting and the
distortion may be greater than normal.
Let's choose DC coupling and an input range that
doesn't overload.
Set the input range to 4 dBV. Adjust the pulse
amplitude so that no overloads occur.
Show two traces.
We will show the time record on the upper trace.
Go to the Measure menu to choose Time Record.
Let's show the time record on a linear scale.
Now set up the trigger.
Trigger on the signal itself.
The input is a 1 V pulse so set the trigger level to
0.5 V.
The upper trace should display the pulse
waveform at the left edge. Auto scale will set the
display limits automatically. Remember that we
are displaying the magnitude of the signal. Any
negative portion of the signal will be folded back
1-9
GETTING STARTED
7. Press [MEAS]
Press <Window Menu>
Press <Uniform>
Press [ACTIVE TRACE]
Press [AUTO SCALE]
8. Press <Hanning>
9. Press [INPUT]
Press <Trigger Menu>
Press <Trigger Delay>
Press [2] [5] [6] <Samples>
10. Press [4] [7] [5] <Samples>
11. Press <Trigger> to select Continuous
around zero and appear as a positive magnitude.
Because the pulse is much shorter than the time
record, we need to use the Uniform window. The
other window functions taper to zero at the start
and end of the time record. Always be aware of
the effect windowing has on the FFT of thetime
record.
There should now be a spectrum on the lower
trace. Use [AUTO SCALE] to set the display.
The spectrum you see is the sinx/x envelope of a
rectangular pulse. The zeroes in the spectrum
occur at the harmonics of 1/pulse width (1/100µs
or 10 kHz).
Now choose the Hanning window. Notice how the
spectrum goes away. We can get the spectrum
back by delaying the time record relative to the
trigger so that the pulse is positioned in the center
of the time record.
Go back to the Trigger submenu.
Highlight the Trigger Delay menu box.
Enter 256 samples of delay. Because the pulse
repetition rate is 250 Hz, the period between
pulses is exactly equal to one time record. So
setting the delay to half of a time record will place
the pulse at the middle of the record.
Remember that the time record only displays the
first 400 points (out of 512) so that the middle of
the record is not the middle of the display trace.
The spectrum should reappear on the lower trace.
This is because windowing preserves the central
part of the time record.
Let's delay the signal some more. Now we've
delayed the time record by almost a full period.
The pulse is now near the end of the time record.
Notice how the spectrum is greatly attenuated.
This is the effect of the window function
attenuating the start of the timer record.
Now if we go to continuous triggering, the time
record becomes unstable. The spectrum is also
unstable because of the windowing. Some time
records place the pulse at the middle, some at the
ends.
1-10
12. Press [MEAS]
Press <Window>
Press <Uniform>
GETTING STARTED
If we set the window back to Uniform, we find that
the spectrum does not vary with the position of the
pulse within the time record.
1-11
GETTING STARTED
USING THE DISK DRIVE
The disk drive on the SR760 may be used to store
3 types of files.
1. Data File
This includes the data in the active trace,
the measurement and display type, the
units and the graph scaling. In addition,
the associated data and limit tables are
stored in this file as well. Data files may be
recalled into either trace0 or trace1.
2. ASCII Data File
This file saves the data in the active trace
in ASCII format. These files may not be
recalled to the display. This format is
convenient when transferring data to a PC
application.
3. Settings File
This files stores the analyzer settings.
Recalling this file will change the analyzer
setup to that stored in the file.
The disk drive uses double-sided, double density
(DS/DD) 3.5" disks. The disk capacity is 720k. The
SR760 uses the DOS format. A disk which was
formatted on a PC or PS2 (for 720k) may be used.
Files written by the SR760 may be copied or read
on a DOS computer.
Data files can store data in either binary or ascii
format. Binary format uses less disk space. Ascii
format allows trace data to be read by other
programs using a PC.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function
is labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
The Measurement
This measurement is designed to familiarize the
user with the disk drive. We will use a function
generator to provide an input signal so that there
is some data to save and recall. Use any function
generator capable of providing a 1 kHz sine wave
at a level of 100 mV to 1 V.
Specifically, you will save and recall a data file and
a settings file.
1-12
STORING AND RECALLING DATA
1. Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the poweron tests are completed.
2. Turn on the generator, set the frequency to
1 kHz and amplitude to approximately 1
Vrms.
Connect the generator's output to the A
input of the analyzer.
3. Press [AUTO RANGE]
4. Press [SPAN DOWN] until the span is
6.25 kHz
5. Press [AUTO SCALE]
6. Press [PAUSE CONT]
7. Put a blank double-sided, double density
(DS/DD)3.5" disk into the drive.
8. Press [STORE RECALL]
Press <Disk Utilities>
Press <Format Disk>
9. Press <Return>
Press <Save Data>
10. Press <File Name>
Press [ALT]
Press [D] [A] [T] [A] [1] <Enter>
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The input impedance of the analyzer is 1 MΩ. The
generator may require a terminator. Many
generators have either a 50 Ω or 600 Ω output
impedance. Use the appropriate feedthrough
termination if necessary. In general, not using a
terminator means that the output amplitude will
not agree with the generator setting and the
distortion may be greater than normal.
Since the signal amplitude may not be set
accurately, let the analyzer automatically set its
input range to actual signal.
Set the span to display the 1 kHz signal and its
first few harmonics.
Set the graph scaling to display the entire range
of the data.
Stop data acquisition. The graph on the screen is
the one we want to save. (You can actually save
graphs while the analyzer is running.)
Use a blank if disk if possible, otherwise any disk
that you don't mind formatting will do. Make sure
the write protect tab is off.
Let's format this disk.
Display the Store and Recall menu.
Choose Disk Utilities.
Make sure that the disk does not contain any
information that you want. Formatting the disk
takes about a minute.
Go back to the main Store and Recall menu.
Display the Save Data menu.
Now we need a file name.
[ALT] lets you enter the letter characters printed
below each key. The numbers and backspace
function as normal.
Enter a file name such as DATA1 (or any legal
DOS file name).
GETTING STARTED
1-13
GETTING STARTED
11. Press <Save Data>
12. Press <Catalog>
13. Press <File Name>
Press [ALT]
Press [D] [A] [T] [A] [2] <Enter>
Press <Save Data>
14. Press <Return>
Press [START]
Remove the input signal cable or turn off the
generator.
15. Press <Recall Data>
Press <Catalog>
16. Press [MARKER]
17. Press <Recall Data>
18. Press [DISPLAY]
Press <Format>
Press [ACTIVE TRACE]
19. Press [START]
This saves the active trace data to disk using the
file name specified above.
Display the disk catalog. This display lists all of
the files on the disk.
Save the data again using a new file name. This
way you can have multiple files in the disk
catalog.
Go back to the main Store and Recall menu.
Resume data acquisition. The graph should be
live again.
Now we have a spectrum which is different from
the one we just saved. Recalling the data from
disk will restore the graph to what it was.
Display the Recall Data menu.
Display the disk catalog. The 2 files which you just
saved should be listed.
Pressing the [MARKER] key allows the knob to
adjust the marker. When the disk catalog is
displayed, the marker highlights a file. Use the
knob to choose a file to recall.
This recalls the data file from disk and displays it
on the active graph. Data acquisition is stopped
so that the graph is not updated. The file name is
displayed below the graph.
The marker may be moved on the recalled graph
to read specific data points. The graph scaling
may also be changed.
Show the Display menu.
Choose the Up/Dn dual trace display format.
Make trace1 active (the lower graph). The active
graph has a highlighted label at its upper right.
This restarts data acquisition, but only for the
active trace (trace1). The recalled trace on graph
0 is still displayed. To restart data acquisition on
trace0, press [ACTIVE TRACE] to make trace0
active and then [START].
1-14
GETTING STARTED
STORING AND RECALLING SETTINGS
1. Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the poweron tests are completed.
2. Press [SPAN DOWN] a number of times to
change the span.
Press [INPUT]
Press <Coupling>
3. Press [STORE RECALL]
Press <Save Settings>
4. Press <File Name>
Press [ALT]
Press [T] [E] [S] [T] [1] <Enter>
5. Press <Save Settings>
6. Press [SPAN UP] a number of times to
change the span.
Press [INPUT]
Press <Coupling>
7. Press [STORE RECALL]
Press <Recall Settings>
Press <Catalog>
8. Press [MARKER]
9. Press <Recall Settings>
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
Change the analyzer setup so that we have a
non-default setup to save.
Show the Input menu.
Choose DC coupling.
Display the Store and Recall menu.
Choose the Save Settings menu.
Now we need a file name.
[ALT] lets you enter the letters printed below each
key. The numbers and backspace function as
normal.
Enter a file name such as TEST1 (or any legal
DOS file name).
Save the analyzer setup to disk using the file
name specified above.
Change the analyzer setup again.
Show the Input menu.
Choose AC coupling.
Now let's recall the analyzer setup that we just
saved.
Display the Store and Recall menu.
Choose the Recall Settings menu.
Display the disk catalog listing. Note that data files
have the type DAT and setting files have the type
SET.
Pressing the [MARKER] key allows the knob to
adjust the marker. When the disk catalog is
displayed, the marker highlights a file. Use the
knob to choose the file TEST1 to recall. (Or use
the <File Name> key to enter the file name.)
This recalls the settings from the file TEST1. The
analyzer settings are changed to those stored in
1-15
GETTING STARTED
TEST1. The span and input coupling should be
the same as those in effect when you created the
file.
1-16
USING DATA TABLES
A data table reports the Y values for user listed Xaxis values. For example, the entries could be a
set of harmonic frequencies which need to be
measured. The data table is a convenient way to
measure the data values at various points without
moving the marker around and manually recording
the answers. To generate a printed report of the
measurement, the data table may be printed using
the Plot menu.
Each trace has its own data table though only the
table associated with the active trace is on and
displayed at any time.
Data tables are saved along with the trace data
when data is saved to disk.
Data tables are not stored in non-volatile memory
and are not retained when the power is turned off.
Remember that the values in the table do not have
units associated with them. An X location of 10
kHz is stored as 10 k and a Y value of -20 dBV is
reported as simply -20. The Y values come directly
from the graph so it is important to use the proper
display units to get consistent data table readings.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function
is labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
GETTING STARTED
The Measurement
This measurement is designed to familiarize the
user with the data tables. We will use a function
generator to provide an input signal so that there
is some data to report. Use any function generator
capable of providing a 1 kHz sine wave at a level
of 100 mV to 1 V.
Specifically, you will generate a data table to
measure some harmonics as well as the noise
floor.
1-17
GETTING STARTED
DATA TABLES
1. Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the poweron tests are completed.
2. Turn on the generator, set the frequency to
1 kHz and amplitude to approximately 1
Vrms.
Connect the generator's output to the A
input of the analyzer.
3. Press [AUTO RANGE]
4. Press [SPAN DOWN] until the span is
6.25 kHz
5. Press [AUTO SCALE]
6. Press [ANALYZE]
Press <Data Table>
7. Press [MARKER MAX/MIN]
Press <X Value>
Press [MARKER ENTRY]
8. Press <Table Index>
Press [1] <Enter>
9. Press [MARKER]
Use the knob to locate the 2nd harmonic of
the signal.
Press <X Value>
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The input impedance of the analyzer is 1 MΩ. The
generator may require a terminator. Many
generators have either a 50 Ω or 600 Ω output
impedance. Use the appropriate feedthrough
termination if necessary. In general, not using a
terminator means that the output amplitude will
not agree with the generator setting and the
distortion may be greater than normal.
Since the signal amplitude may not be set
accurately, let the analyzer automatically set its
input range to actual signal.
Set the span to display the 1 kHz signal and its
first few harmonics.
Set the graph scaling to display the entire range
of the data.
Display the Analysis menu.
Select Data Table display. The display switches to
dual trace format with the spectrum on top and
the data table listed below.
This moves the marker to the peak of the
spectrum. This should center the marker on the 1
kHz fundamental frequency.
Highlight the X Value menu box.
This copies the marker X position into the X Value
menu box. The X value of data table line 0 is now
equal to the 1 kHz signal frequency. The Y value
of line 0 is updated each time the graph is
updated.
This highlights the Table Index menu box. Let's
add another line to the data table.
Entering an index or line number beyond the end
of the table adds a new line to the end.
Activate the marker.
We are going to enter the frequency of the 2nd
harmonic into the data table.
Highlight the X Value menu box.
1-18
Press [MARKER ENTRY]
10. Press <Table Index>
Press [2] <Enter>
11. Press <X Value>
Press [2] [.] [5] [4] <kHz>
12. Press <Insert Item>
Press <Delete Item>
13. Press [PLOT]
Press <Printing Menu>
14. Press [DISPLAY]
Press <Format>
GETTING STARTED
This copies the marker X location into the data
table. Line 1 now has the frequency of the 2nd
harmonic. Note how the Y values update with the
graph.
By now you probably realize that the [MARKER
ENTRY] key is pretty handy. In the Analysis
menu, many of the frequencies or X values may
be entered by copying the X location of the
marker into the highlighted menu field.
Let's add another line to the table.
And this time let's enter the X location
numerically.
Enter some frequency which is representative of
the noise floor of the signal.
We decided that we wanted another harmonic in
the table. This key inserts a new line before the
highlighted line.
We could enter an X value for this new line now.
But we changed our mind. Let's delete this line.
Display the Plot menu.
Display the Printing submenu.
If we have a printer attached, then the
<Print Data> function will print the data table, with
updated Y values.
Show the Display menu.
Choose the Single trace display format. This
removes the data table display and restores the
screen to a single trace display.
1-19
GETTING STARTED
USING LIMIT TABLES
A limit table lists the X,Y coordinates of the line
segments which define the trace test limits. When
trace data exceeds these limit segments, then the
test fails. The limit table is a convenient way to test
devices against a specification defined over a
range of frequencies. To generate a printed listing
of a limit table, use the Print Limits function in the
Plot menu.
Each trace has its own limit table though only the
table associated with the active trace is on and
displayed at any time.
Limit tables are saved along with the trace data
when data is saved to disk.
Limit tables are not stored in non-volatile memory
and are not retained when the power is turned off.
Remember that the values in the table do not have
units associated with them. An X location of 10
kHz is stored as 10 k and a Y value of -20 dBV is
simply -20. The limit test compares the data on the
graph (in the display units) to the Y values in the
table. It is important to use the correct units in the
display to get consistent limit table tests.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function
is labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
The Measurement
This measurement is designed to familiarize the
user with the limit tables. We will use a function
generator to provide an input signal. Use any
function generator capable of providing a 1 kHz
sine wave at a level of 100 mV to 1 V.
Specifically, you will generate a limit table to test
the signal level as well as the noise floor.
1-20
LIMIT TABLES
1. Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the poweron tests are completed.
2. Turn on the generator, set the frequency to
1 kHz and amplitude to approximately 1
Vrms.
Connect the generator's output to the A
input of the analyzer.
3. Press [AUTO RANGE]
4. Press [SPAN DOWN] until the span is
6.25 kHz
5. Press [AUTO SCALE]
6. Press [ANALYZE]
Press <Limit Table>
7. Press [MARKER MAX/MIN]
8. Press <X Values>
Press [9] [0] [0] <Hz>
Press <X Values> again
Press [1] [.] [1] <kHz>
9. Press <Y Values>
Press [-] [5] <Enter>
GETTING STARTED
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The input impedance of the analyzer is 1 MΩ. The
generator may require a terminator. Many
generators have either a 50 Ω or 600 Ω output
impedance. Use the appropriate feedthrough
termination if necessary. In general, not using a
terminator means that the output amplitude will
not agree with the generator setting and the
distortion may be greater than normal.
Since the signal amplitude may not be set
accurately, let the analyzer automatically set its
input range to agree with the actual generator
signal.
Set the span to display the 1 kHz signal and its
first few harmonics.
Set the graph scaling to display the entire range
of the data.
Display the Analysis menu.
Select Limit Table display. The display switches to
dual trace format with the spectrum on top and
the limit table listed below.
This moves the marker to the peak of the
spectrum and measures the fundamental
frequency.
Let's define an upper limit for the 1 kHz peak.
Highlight the upper X Value menu field.
Enter a frequency below the signal frequency.
Highlight the lower X Value menu field.
Enter a frequency higher than the signal
frequency.
As with data tables, it is also possible to copy the
marker X location into the X value fields. But this
time we want frequencies above and below the
peak so we entered them numerically.
Highlight the upper Y values menu field.
Enter a value somewhat less than the signal
peak.
1-21
GETTING STARTED
Press <Y Values>
Press [-] [5] <Enter>
10. Press <More>
Press <Audio Alarm>
Reduce the amplitude of the generator
output so that the peak falls below the limit
segment. The alarm should stop and the
PASS indicator should turn on.
Press <Audio Alarm>
Press <Return>
11. Press <Table Index>
Press [1] <Enter>
12. Press <X Values> until the upper field is
highlighted.
Press [2] [.] [2] <kHz>
Press <X Values>
Press [2] [.] [8] <kHz>
13. Press <YValues> until the upper field is
highlighted.
Highlight the lower Y values menu field.
Enter a value somewhat less than the signal
peak.
Notice that small line segment is drawn on the
display. This line starts at (Xbegin,Y1) and ends
at (Xend, Y2) and represents a limit segment. If
the data exceeds this limit (since it is an upper
limit), then the FAIL indicator will light at the
bottom of the screen. The FAIL indicator should
be on now.
Display the second limits menu.
Set the audio alarm on. Now whenever a trace is
taken that exceeds the limit, an alarm sounds.
Set the audio alarm off. You're probably ready to
turn off the alarms by now anyway.
Go back to the main limits menu.
Let's add another segment to this table.
Highlight the Table Index menu box.
Entering an index or line number beyond the end
of the table adds a new line to the end.
Notice how the new segment is a continuation of
the previous one. This makes building a
continuous limit much simpler. The starting point
of the new line equals the ending point of the
previous one. The new segment's length along
the X axis is the same as the previous segment's.
The only thing you need to edit is the value of Y2
and your new segment is finished.
But let's go on to define a noise floor limit.
Enter a segment which is between harmonics. In
this case, between 2.2 and 2.8 kHz. This is
representative of the noise floor.
Define an upper limit a little above the noise floor.
1-22
Press [-] [8] [0] <Enter>
Press <Y Values>
Press [-] [8] [0] <Enter>
14. Press <Limit Type>
15. Press <More>
Press <Testing>
16. Press [DISPLAY]
Press <Format>
GETTING STARTED
In this case, we define an upper noise limit of 80 dB. You should enter whatever is appropriate
for your display.
There should now be a horizontal segment above
the noise floor between 2 harmonics. The limit
test should still PASS.
This switches the noise limit from an upper limit to
a lower limit. Since the data will now be below the
lower limit, the test will FAIL.
Display the second limits menu.
Set limit testing to OFF. It is possible to display
the limit table without testing taking place. This is
helpful when a lot of the X values on the graph
have defined limits. The testing can slow down
the response of the analyzer noticeably. It is
simpler to define the limits with testing off.
Show the Display menu.
Choose the Single trace display format. This
removes the limit table display and restores the
screen to a single trace display. No testing occurs
when the limit table is not displayed.
1-23
GETTING STARTED
USING TRACE MATH
The Calculator submenu allows the user to
perform arithmetic calculations with the trace data.
Operations are performed on the entire trace,
regardless of graphical expansion.
Calculations treat the data as intrinsic values,
either Volts, Engineering Units or degrees. If a
graph is showing dB, then multiplying by 10 will
raise the graph by 20 dB and dividing by 10 will
lower the graph by 20 dB.
Performing a calculation on the active trace will set
the File Type to Calc to indicate that the trace is
not Live. This is shown by the "File=Calc"
message at the lower left of the graph. The
analyzer continues to run but the calculated trace
will not be updated. To return the trace to live
mode, activate the trace and press the [START]
key. The File Type will return to Live.
There are two types of front panel keys which will
be referenced in this section. Hardkeys are those
keys with labels printed on them. Their function is
determined by the label and does not change.
Hardkeys are referenced by brackets like this [HARDKEY]. The softkeys are the six gray keys
along the right edge of the screen. Their function
is labelled by a menu box displayed on the screen
next to the key. Softkey functions change
depending upon the situation. Softkeys will be
referenced as the <Soft Key> or simply the Soft
Key.
The Measurement
This measurement is designed to familiarize the
user with the trace math capabilities. We will use a
function generator to provide an input signal. Use
any function generator capable of providing a 1
kHz sine wave at a level of 100 mV to 1 V.
Specifically, you will ratio a spectrum with a
reference spectrum.
1-24
TRACE MATH
1. Turn the analyzer on while holding down the
[<-] (backspace) key. Wait until the poweron tests are completed.
2. Turn on the generator, set the frequency to
1 kHz and amplitude to approximately 1
Vrms.
Connect the generator's output to the A
input of the analyzer.
3. Press [AUTO RANGE]
4. Press [SPAN DOWN] until the span is
6.25 kHz
5. Press [AUTO SCALE]
6. Press [MEAS]
Press <Calculator Menu>
7. Press <Do Calc>
8. Press [DISPLAY]
Press <Format>
Press <Marker Width> twice to choose Spot
Marker.
Press [ACTIVE TRACE]
Press <Marker Width> twice to choose Spot
Marker.
GETTING STARTED
When the power is turned on with the backspace
key depressed, the analyzer returns to its default
settings. See the Default Settings list in the Menu
section for a complete listing of the settings.
The input impedance of the analyzer is 1 MΩ. The
generator may require a terminator. Many
generators have either a 50 Ω or 600 Ω output
impedance. Use the appropriate feedthrough
termination if necessary. In general, not using a
terminator means that the output amplitude will
not agree with the generator setting and the
distortion may be greater than normal.
Since the signal amplitude may not be set
accurately, let the analyzer automatically set its
input range to agree with the actual generator
signal.
Set the span to display the 1 kHz signal and its
first few harmonics.
Set the graph scaling to display the entire range
of the data.
Display the Measure menu.
Select the Calculator menu.
This operation defaults to adding zero to the trace
data. The default operation is +, the default
argument is the constant zero. We're doing this so
that the trace does not update. This is now the
graph we will use as the reference data.
Reference data normally comes from a disk file.
Recalling a stored file brings the data back to the
active graph but does not update it. See "Using
the Disk Drive" earlier in this section.
Bring up the Display menu.
Choose the Up/Dn dual trace format. The
reference graph will be the upper trace (Trace0)
and the live graph will be the lower trace (Trace1).
Make the marker on the upper graph a spot
marker.
Let's make the live graph the active trace.
Make the marker on the lower graph a spot
marker.
1-25
GETTING STARTED
9. Press [MARKER MODE]
Press <Linked Markers>
10. Press [MAX/MIN]
Adjust the generator amplitude either higher
or lower to make it different than the
reference peak amplitude.
11. Press [MEAS]
Press <Calculator Menu>
Press <Argument Type> twice to select
Other Graph.
Press <Operation> three times to select '/'
(divide).
12. Press <Do Calc>
Press [AUTO SCALE]
Display the Marker Mode menu.
Link the two markers together. Now when the
knob moves one marker, they both move
together. Since they are both spot markers, the
frequencies which they read on both graphs are
identical.
Move the markers to the signal peak (1 kHz).
The reference amplitude may be read from the
marker readout of the upper graph. The live
amplitude may be read from the marker of the
lower graph.
Now we have 2 traces which differ in amplitude.
Let's take the ratio.
Go back to the Calculator menu.
We will divide the active graph (Trace1 Live) by
the inactive graph (Trace0 reference).
Select the divide operation.
Do the calculation. Since the graphs are displayed
in dBV, the ratio of the peaks should simply be the
difference in their amplitudes expressed in dBV.
Remember, the calculations work on the
underlying data points (in Volts).
The active graph now displays the ratio of the two
traces in dB.
The marker on the lower graph should read the
difference between the two peak amplitudes (in
dB). Clearly, only the frequencies which
correspond to the signal and its harmonics have
much meaning in this ratio. One noise floor
divided by another noise floor is going to be pretty
noisy.
Move the marker to a harmonic by using the
reference trace as a guide. When the marker
aligns with a peak on the reference trace, it
selects the same point in the ratio trace since the
markers are linked and in spot mode.
A better way to read these harmonic ratios is
using the data table. A data table can display the
values of selected frequencies in easy to read
form. See "Using Data Table" earlier in this
section. The data table would be defined for
Trace1.
1-26
GETTING STARTED
Other operations which may be performed are +, , x, /, log, and square root. The second argument
may be a constant (scaling or offset), w (2πf to
differentiate or integrate the spectrum), or the
other graph (reference trace from disk).
1-27
GETTING STARTED
THINGS TO WATCH OUT FOR
If the analyzer is on but doesn't seem to be taking
data, there are a number of things to check.
1) Press the [START] key to make sure that
the indicator at the lower left of the screen
displays RUN instead of STOP.
2) Check if linear averaging is on. When the
analyzer finishes a linear average of N
spectra, the analyzer stops and the data is
no longer updated. Press [START] to take
another average.
3) Make sure the triggering mode is
CONTinuous. Otherwise, the analyzer
may be waiting for a trigger (as shown by
the Trg Wait indicator at the bottom of the
screen).
4) If the unit is being triggered, check that the
arming mode is set to AUTO. If the arming
mode is MANUAL, then the analyzer will
only trigger once and then wait for the
next manual arming command.
5) Check that the data is on scale by using
[AUTORANGE] and [AUTOSCALE].
6) Make sure that the analyzer is not in the
REMOTE state where the computer
interfaces have locked out the front panel.
Press the LOCAL key (the [HELP] key) to
restore local control.
If the analyzer still seems to function improperly,
turn the power off and turn it back on while holding
down the [<-] (backspace) key. This will reset the
analyzer into the default configuration. The
analyzer should power on running and taking
spectra.
1-28
ANALYZER BASICS
WHAT IS AN FFT SPECTRUM ANALYZER?
The SR760 FFT Spectrum Analyzer takes a time
varying input signal, like you would see on an
oscilloscope trace, and computes its frequency
spectrum.
Fourier's basic theorem states that any waveform
in the time domain can be represented by the
weighted sum of pure sine waves of all
frequencies. If the signal in the time domain (as
viewed on an oscilloscope) is periodic, then its
spectrum is probably dominated by a single
frequency component. What the spectrum
analyzer does is represent the time domain signal
by its component frequencies.
Why look at a signal's spectrum?
For one thing, some measurements which are
very hard in the time domain are very easy in the
frequency domain. Take harmonic distortion. It's
hard to quantify the distortion by looking at a good
sine wave output from a function generator on an
oscilloscope. When the same signal is displayed
on a spectrum analyzer, the harmonic frequencies
and amplitudes are displayed with amazing clarity.
Another example is noise analysis. Looking at an
amplifier's output noise on an oscilloscope
basically measures just the total noise amplitude.
On a spectrum analyzer, the noise as a function of
frequency is displayed. It may be that the amplifier
has a problem only over certain frequency ranges.
In the time domain it would be very hard to tell.
Many of these types of measurements used to be
done using analog spectrum analyzers. In simple
terms, an analog filter was used to isolate
frequencies of interest. The remaining signal
power was measured to determine the signal
strength in certain frequency bands. By tuning the
filters and repeating the measurements, a
reasonable spectrum could be obtained.
The FFT Analyzer
An FFT spectrum analyzer works in an entirely
different way. The input signal is digitized at a high
sampling rate, similar to a digitizing oscilloscope.
Nyquist's theorem says that as long as the
sampling rate is greater than twice the highest
frequency component of the signal, then the
sampled data will accurately represent the input
signal. In the SR760, sampling occurs at 256 kHz.
To make sure that Nyquist's theorem is satisfied,
the input signal passes through an analog filter
which attenuates all frequency components
above128 kHz by 90 dB. This is the anti-aliasing
filter. The resulting digital time record is then
mathematically transformed into a frequency
spectrum using an algorithm known as the Fast
Fourier Transform or FFT. The FFT is simply a
clever set of operations which implements
Fourier's basic theorem. The resulting spectrum
shows the frequency components of the input
signal.
Now here's the interesting part. The original digital
time record comes from discrete samples taken at
the sampling rate. The corresponding FFT yields a
spectrum with discrete frequency samples. In fact,
the spectrum has half as many frequency points
as there are time points. (Remember Nyquist's
theorem). Suppose that you take 1024 samples at
256 kHz. It takes 4 ms to take this time record.
The FFT of this record yields 512 frequency
points, but over what frequency range? The
highest frequency will be determined by the period
of 2 time samples or 128 kHz. The lowest
frequency is just the period of the entire record or
1/(4 ms) or 250 Hz. Everything below 250 Hz is
considered to be dc. The output spectrum thus
represents the frequency range from dc to 128
kHz with points every 250 Hz.
Advantages and limitations
The advantage of this technique is its speed. The
entire spectrum takes only 4 ms to measure. The
limitation of this measurement is its resolution.
Because the time record is only 4 ms long, the
frequency resolution is only 250 Hz. Suppose the
signal has a frequency component at 260 Hz. The
FFT spectrum will detect this signal but place part
of it in the 250 Hz point and part in the 500 Hz
point. One way to measure this signal accurately
is to take a time record that is 1/260 or 3.846 ms
long with 1024 evenly spaced samples. Then the
signal would land all in one frequency bin. But this
would require changing the sampling rate based
upon the signal (which you haven't measured yet).
Not a good solution. In fact, the way to measure
the signal accurately is to lengthen the time record
and change the span of the spectrum.
2-1
ANALYZER BASICS
FREQUENCY SPANS
Before we continue, let's clarify a couple of points
about our frequency span. We just described how
we arrived at a dc to 128 kHz frequency span
using a 4 ms time record. Because the signal
passes through an anti-aliasing filter at the input,
the entire frequency span is not useable. The filter
has a flat response from dc to 100 kHz and then
rolls off steeply from 100 kHz to 128 kHz. No filter
can make a 90 dB transition instantly. The range
between 100 kHz and 128 kHz is therefore not
useable and the actual displayed frequency span
stops at 100 kHz. There is also a frequency bin
labelled 0 Hz (or dc). This bin actually covers the
range from 0 Hz to 250 Hz (the lowest measurable
frequency) and contains the signal components
whose period is longer than the time record (not
only dc). So our final displayed spectrum contains
400 frequency bins. The first covers 0 - 250 Hz,
the second 250 - 500 Hz, and the 400th covers
99.75 - 100.0 kHz.
Spans less than 100 kHz
So the length of the time record determines the
frequency span and resolution of our spectrum.
What happens if we make the time record 8 ms or
twice as long? Well we ought to get 2048 time
points (sampling at 256 kHz) yielding a spectrum
from dc to 100 kHz with 125 Hz resolution
containing 800 points. But the SR760 places some
limitations on this. One is memory. If we keep
increasing the time record, then we would need to
store more and more points. Another limitation is
processing time. The time it takes to calculate an
FFT with more points increases more than linearly.
The net result is that the SR760 always takes
1024 point FFT's to yield 400 point spectra.
Here's how it's done. The analyzer digitally filters
the incoming data samples (at 256 kHz) to limit the
bandwidth. This is similar to the anti-aliasing filter
at the input except the digital filter's cutoff
frequency can be changed. In the case of the 8 ms
record, the filter reduces the bandwidth to 64 kHz
with a filter cutoff of 50 kHz (the filter rolls off
between 50 and 64 kHz). Remember that Nyquist
only requires samples at twice the frequency of
the highest signal frequency. Thus the digital filter
only has to output points at 128 kHz or half of the
input rate (256 kHz). The net result is the digital
filter outputs a time record of 1024 points
effectively sampled at 128 kHz to make up an 8
ms record. The FFT processor operates on a
constant number of points and the resulting FFT
will yield 400 points from dc to 50 kHz. The
resolution or linewidth is 125 Hz.
This process of doubling the time record and
halving the span can be repeated by using
multiple stages of digital filtering. The SR760 can
process spectra with a span of only 191 mHz with
a time record of 2098 seconds if you have the
patience. However, this filtering process only
yields baseband measurements (frequency spans
which start at dc).
Starting the span somewhere other than dc
Besides being able to choose the span and
resolution of the spectrum, we would also like the
span to be able to start at frequencies other than
dc. It would be nice to center a narrow span
around any frequency below 100 kHz. Using
digital filtering alone requires that every span start
at dc. What is needed is heterodyning.
Heterodyning is the process of multiplying the
incoming signal by a sine wave. The resulting
spectrum is shifted by the frequency of the sine
wave. If we incorporate heterodyning with our
digital filtering, we can shift any frequency span so
that it starts at dc. The resulting FFT yields a
spectrum offset by the heterodyne frequency.
When this spectrum is displayed, the frequencies
of the X axis are the frequencies of the actual
signal, not the heterodyned frequencies.
Heterodyning allows the analyzer to compute
zoomed spectra (spans which start at frequencies
other than dc). The digital filter processor can filter
and heterodyne the input in real time to provide
the appropriate filtered time record at all spans
and center frequencies. Because the digital signal
processors in the SR760 are so fast, you won't
notice any calculation time while taking spectra.
The longest it can take to acquire a spectrum is
the length of the time record itself. But more about
that later.
2-2
THE TIME RECORD
Now that we've described the process in simple
terms, let's complicate it a little bit. The SR760
actually uses 512 point complex time records.
Each point is a complex value (with real and
imaginary parts) so the record actually has 1024
data points in it. But how does a real point get to
be complex?
As we described in the previous section, the input
samples are digitally filtered and heterodyned to
produce a time record with the appropriate
bandwidth and a constant number of samples.
What we need to add to this is that the
heterodyning is a complex operation. This means
that the input points are multiplied by both sine
and cosine to yield a real and imaginary part.
So instead of using 1024 real points, we use 512
complex points. The time records have the same
duration so the complex record has half the
sampling rate of the real record. Thus at full span,
the real points would occur at 256 kHz and the
complex points at 128 kHz. You can think of the
complex record as two separate records, one real
and one imaginary, each with 64 kHz of
bandwidth. (1/2 of the sample rate). One covers 0
to +64 kHz and the other covers -64 kHz to 0 for a
total bandwidth of 128 kHz (the same bandwidth
as the real record). What a negative frequency
means is beyond this discussion but suffice to say
it works the same.
The time record display
What do you see when you display the time
record? Clearly the time record is not as simple as
the raw digitized data points you would see if this
were a digital oscilloscope.
The analyzer stores the 512 point complex time
record described above. Because the display is
designed for 400 point spectra, only the first 400
points of the time record are displayed. You can
use the trigger delay to "translate" the time record
to see the part not normally displayed.
The time record for every span has been digitally
filtered and heterodyned into a complex record.
You can display the magnitude, real or imaginary
part as well as the phase. Normally, the easiest
display to understand is Linear Magnitude.
ANALYZER BASICS
Remember that magnitudes are always positive.
The negative parts of the waveform will be folded
around zero so that they appear positive.
Because of the filtering and heterodyning, the time
waveform may not closely resemble the input
signal. For baseband measurements (when the
start frequency of the span is 0.0 Hz) the
waveform will resemble the signal waveform (with
folding if magnitude is displayed). The bandwidth
will be limited by the anti-alias filter and the digital
filtering. For zoomed measurements (when the
span start is not 0.0 Hz) the displayed waveform
will not closely resemble the input signal because
of the heterodyning.
Why use the time record?
The time display can be useful in determining
whether the time record is triggered properly. If the
analyzer is triggered, either internally by the signal
or externally with another pulse, and the signal
has a large component synchronous with the
trigger, then the time record should appear
stationary on the display. If the signal triggers
randomly, then the time display will jitter back and
forth.
Watch out for windowing!
The time display is not windowed. This means the
time record which is displayed will be multiplied by
the window function before the FFT is taken (see
"Windowing" later in this section). Most window
functions taper off to zero at the start and end of
the time record. If the transient signal occurs at the
start of the time record, the corresponding FFT
may not show anything because the window
function reduces the transient to zero.
Either use a Uniform window with transients, or
use the trigger delay to position the transient at the
center of the time record. (Remember that the
display only shows the first 400 points of the
record. The center is always at the 256th sample,
which is not at the center of the display.)
To repeat, the time record is not a snapshot of
the input signal. It is the output of the digital
filter and the input to the FFT processor.
2-3
ANALYZER BASICS
MEASUREMENT BASICS
Now that we know that the input to the FFT
processor is a complex time record, it should be
no surprise to find out that the resulting FFT
spectrum is also a complex quantity. This is
because each frequency component has a phase
relative to the start of the time record. If there is no
triggering, then the phase is random and we
generally look at the magnitude of the spectrum. If
we use a synchronous trigger then each frequency
component has a well defined phase.
Spectrum
The spectrum is the basic measurement of an FFT
analyzer. It is simply the complex FFT. Normally,
the magnitude of the spectrum is displayed. The
magnitude is the square root of the FFT times its
complex conjugate. (Square root of the sum of the
real part squared and the imaginary part squared).
The magnitude is a real quantity and represents
the total signal amplitude in each frequency bin,
independent of phase.
If there is phase information in the spectrum, i.e.
the time record is triggered in phase with some
component of the signal, then the real or
imaginary part or the phase may be displayed.
Remember, the phase is simply the arctangent of
the ratio of the imaginary and real parts of each
frequency component. The phase is always
relative to the start of the triggered time record.
Power Spectral Density or PSD
The PSD is simply the magnitude of the spectrum
normalized to a 1 Hz bandwidth. This
measurement approximates what the spectrum
would look like if each frequency component were
really a 1 Hz wide piece of the spectrum at each
frequency bin.
What good is this? When measuring broadband
signals such as noise, the amplitude of the
spectrum changes with the frequency span. This
is because the linewidth changes so the frequency
bins have a different noise bandwidth. The PSD,
on the other hand, normalizes all measurements
to a 1 Hz bandwidth and the noise spectrum
becomes independent of the span. This allows
measurements with different spans to be
compared. If the noise is Gaussian in nature, then
the amount of noise amplitude in other bandwidths
may be approximated by scaling the PSD
measurement by the square root of the bandwidth.
Thus the PSD is displayed in units of V/√Hz or
dBV/√Hz.
Since the PSD uses the magnitude of the
spectrum, the PSD is a real quantity. There is no
real or imaginary part or phase.
Octave Analysis
The magnitude of the normal spectrum measures
the amplitudes within equally divided frequency
bins. Octave analysis computes the spectral
amplitude within 1/3 octave bands. The start and
stop frequencies of each frequency bin are in the
ratio of 1/3 of an octave (2
analysis spectra will closely resemble data taken
with older analog type equipment commonly used
in acoustics and sound measurement.
To compute the amplitude of each band, the
normal FFT is taken. Those bins which fall within a
single band are rms summed together (square
root of the sum of the squared magnitudes). The
resulting amplitudes are real quantities and have
no phase information. They represent total signal
amplitude within each band.
We will have more about octave analysis later.
1/3
). The octave
2-4
DISPLAY TYPES
Spectrum
The most common measurement is the spectrum
and the most useful display is the Log Magnitude.
The Log Mag display graphs the magnitude of the
spectrum on a logarithmic scale using dBV as
units.
Why is the Log Mag display useful? Remember
that the SR760 has a dynamic range of 90 dB and
a display resolution of -114 dB below full scale.
Imagine what something 0.01% of full scale would
look like on a linear scale. If we wanted it to be 1
centimeter high on the graph, the top of the graph
would be 100 meters above the bottom. It turns
out that the log display is both easy to understand
and shows features which have very different
amplitudes clearly.
Of course the analyzer is capable of showing the
magnitude on a linear scale if you wish.
The real and imaginary parts are always displayed
on a linear scale. This avoids the problem of
taking the log of negative voltages.
The PSD and Octave analysis are real quantities
and thus may only be displayed as magnitudes. In
addition, the Octave analysis requires the display
to be Log Magnitude.
Phase
In general, phase measurements are only used
when the analyzer is triggered. The phase is
relative to the start of the time record.
ANALYZER BASICS
The phase is displayed in degrees or radians on a
linear scale from -180 (-π) to +180 (+π) degrees
(rads). There is no phase "unwrap".
The phase of a particular frequency bin is set to
zero if neither the real nor imaginary part of the
FFT is greater than 0.012% of full scale (-78 dB
below f.s.). This avoids the messy phase display
associated with the noise floor. (Remember, even
if a signal is small, its phase extends over the full
360 degrees.)
Watch Out For Phase Errors
The FFT can be thought of as 400 bandpass
filters, each centered on a frequency bin. The
signal within each filter shows up as the amplitude
of each bin. If a signal's frequency is between
bins, the filters act to attenuate the signal a little
bit. This results in a small amplitude error. The
phase error, on the other hand, can be quite large.
Because these filters are very steep and selective,
they introduce very large phase shifts for signals
not exactly on a frequency bin.
On full span, this is generally not a problem. The
bins are 250 Hz apart and most synthesized
sources have no problem generating a signal right
on a frequency bin. But when the span is
narrowed, the bins move much closer together
and it becomes very hard to place a signal exactly
on a frequency bin.
2-5
ANALYZER BASICS
WINDOWING
What is windowing? Let's go back to the time
record. What happens if a signal is not exactly
periodic within the time record? We said that its
amplitude is divided into multiple adjacent
frequency bins. This is true but it's actually a bit
worse than that. If the time record does not start
and stop with the same data value, the signal can
actually smear across the entire spectrum. This
smearing will also change wildly between records
because the amount of mismatch between the
starting value and ending value changes with each
record.
Windows are functions defined across the time
record which are periodic in the time record. They
start and stop at zero and are smooth functions in
between. When the time record is windowed, its
points are multiplied by the window function, time
bin by time bin, and the resulting time record is by
definition periodic. It may not be identical from
record to record, but it will be periodic (zero at
each end).
In the frequency domain
In the frequency domain, a window acts like a
filter. The amplitude of each frequency bin is
determined by centering this filter on each bin and
measuring how much of the signal falls within the
filter. If the filter is narrow, then only frequencies
near the bin will contribute to the bin. A narrow
filter is called a selective window - it selects a
small range of frequencies around each bin.
However, since the filter is narrow, it falls off from
center rapidly. This means that even frequencies
close to the bin may be attenuated somewhat. If
the filter is wide, then frequencies far from the bin
will contribute to the bin amplitude but those close
by will probably not be attenuated much.
The net result of windowing is to reduce the
amount of smearing in the spectrum from signals
not exactly periodic with the time record. The
different types of windows trade off selectivity,
amplitude accuracy, and noise floor.
The SR760 offers four types of window functions Uniform (none), Flattop, Hanning and BlackmanHarris (BMH).
Uniform
The uniform window is actually no window at all.
The time record is used with no weighting. A
signal will appear as narrow as a single bin if its
frequency is exactly equal to a frequency bin. (It is
exactly periodic within the time record). If its
frequency is between bins, it will affect every bin of
the spectrum. These two cases also have a great
deal of amplitude variation between them (up to
4 dB).
In general, this window is only useful when looking
at transients which do not fill the entire time
record.
Hanning
The Hanning window is the most commonly used
window. It has an amplitude variation of about
1.5 dB (for signals between bins) and provides
reasonable selectivity. Its filter rolloff is not
particularly steep. As a result, the Hanning window
can limit the performance of the analyzer when
looking at signals close together in frequency and
very different in amplitude.
Flattop
The Flattop window improves on the amplitude
accuracy of the Hanning window. Its between-bin
amplitude variation is about .02 dB. However, the
selectivity is a little worse. Unlike the Hanning, the
Flattop window has a wide pass band and very
steep rolloff on either side. Thus, signals appear
wide but do not leak across the whole spectrum.
BMH
The BMH window is a very good window to use
with this analyzer. It has better amplitude accuracy
(about 0.7 dB) than the Hanning, very good
selectivity and the fastest filter rolloff. The filter is
steep and narrow and reaches a lower attenuation
than the other windows. This allows signals close
together in frequency to be distinguished, even
when their amplitudes are very different.
If a measurement requires the full dynamic range
of the analyzer, then the BMH window is probably
the best one to use.
2-6
AVERAGING
The SR760 analyzer supports several types of
averaging. In general, averaging many spectra
together improves the accuracy and repeatability
of measurements.
RMS Averaging
RMS averaging computes the weighted mean of
the sum of the squared magnitudes (FFT times its
complex conjugate). The weighting is either linear
or exponential.
RMS averaging reduces fluctuations in the data
but does not reduce the actual noise floor. With a
sufficient number of averages, a very good
approximation of the actual random noise floor can
be displayed.
Since RMS averaging involves magnitudes only,
displaying the real or imaginary part or phase of
an RMS average has no meaning. The RMS
average has no complex information.
Vector Averaging
Vector averaging averages the complex FFT
spectrum. (The real part is averaged separately
from the imaginary part.) This can reduce the
noise floor for random signals since they are not
phase coherent from time record to time record.
Vector averaging requires a trigger. The signal of
interest must be both periodic and phase
synchronous with the trigger. Otherwise, the real
and imaginary parts of the signal will not add in
phase and instead will cancel randomly.
With vector averaging, the real and imaginary
parts as well as phase displays are correctly
averaged and displayed. This is because the
complex information is preserved.
Peak Hold
Peak Hold is not really averaging, rather the new
spectral magnitudes are compared to the previous
data, and if the new data is larger, then the new
data is stored. This is done on a frequency bin by
bin basis. The resulting display shows the peak
magnitudes which occurred in the previous group
of spectra.
ANALYZER BASICS
Peak Hold detects the peaks in the spectral
magnitudes and only applies to Spectrum, PSD,
and Octave Analysis measurements. However, the
peak magnitude values are stored in the original
complex form. If the real or imaginary part or
phase is being displayed for spectrum
measurements, the display shows the real or
imaginary part or phase of the complex peak
value.
Linear Averaging
Linear averaging combines N (number of
averages) spectra with equal weighting in either
RMS, Vector or Peak Hold fashion. When the
number of averages has been completed, the
analyzer stops and a beep is sounded. When
linear averaging is in progress, the number of
averages completed is continuously displayed
below the Averaging indicator at the bottom of the
screen.
Auto ranging is temporarily disabled when a linear
average is in progress. Be sure that you don't
change the input range manually either. Changing
the range during a linear average invalidates the
results.
Exponential Averaging
Exponential averaging weights new data more
than old data. Averaging takes place according to
the formula,
AverageN = (New Spectrum • 1/N) +
(Average
where N is the number of averages.
Exponential averages "grow" for approximately the
first 5N spectra until the steady state values are
reached. Once in steady state, further changes in
the spectra are detected only if they last
sufficiently long. Make sure that the number of
averages is not so large as to eliminate the
changes in the data that might be important.
) • (N-1)/N
N-1
2-7
ANALYZER BASICS
REAL TIME BANDWIDTH AND OVERLAP PROCESSING
What is real time bandwidth? Simply stated, it is
the frequency span whose corresponding time
record exceeds the time it takes to compute the
spectrum. At this span and below, it is possible to
compute the spectra for every time record with no
loss of data. The spectra are computed in "real
time". At larger spans, some data samples will be
lost while the FFT computations are in progress.
For all frequency spans, the SR760 can compute
the FFT in less time than it takes to acquire the
time record. Thus, the real time bandwidth of the
SR760 is 100 kHz. This includes the real time
digital filtering and heterodyning, the FFT
processing, and averaging calculations. The
SR760 employs two digital signal processors to
accomplish this. The first collects the input
samples, filters and heterodynes them, and stores
a time record. The second computes the FFT and
averages the spectra. Since both processors are
working simultaneously, no data is ever lost.
Averaging speed
How can you take advantage of this?Consider
averaging. Other analyzers typically have a real
time bandwidth of around 4 kHz. This means that
even though the time record at 100 kHz span is
only 4 ms, the "effective" time record is 25 times
longer due to processing overhead. An analyzer
with 4 kHz of real time bandwidth can only
process about 10 spectra a second. When
averaging is on, this usually slows down to about 5
spectra per second. At this rate it's going to take a
couple of minutes to do 500 averages.
The SR760, on the other hand, has a real time
bandwidth of 100 kHz. At a 100 kHz span, the
analyzer is capable of processing 250 spectra per
second. In fact, this is so fast, that the display can
not be updated for each new spectra. The display
only updates about 6 times a second. However,
when averaging is on, all of the computed spectra
will contribute to the average. The time it takes to
complete 500 averages is only a few seconds.
(Instead of a few minutes!)
Overlap
What about narrow spans where the time record is
long compared to the processing time? The
analyzer computes one FFT per time record and
can wait until the next time record is complete
before computing the next FFT. The update rate
would be no faster than one spectra per time
record. With narrow spans, this could be quite
slow.
And what is the processor doing while it waits?
Nothing. With overlap processing, the analyzer
does not wait for the next complete time record
before computing the next FFT. Instead it uses
data from the previous time record as well as data
from the current time record to compute the next
FFT. This speeds up the processing rate.
Remember, most window functions are zero at the
start and end of the time record. Thus, the points
at the ends of the time record do not contribute
much to the FFT. With overlap, these points are
"re-used" and appear as middle points in other
time records. This is why overlap effectively
speeds up averaging and smoothes out window
variations.
Typically, time records with 50% overlap provide
almost as much noise reduction as nonoverlapping time records when RMS averaging is
used. When RMS averaging narrow spans, this
can reduce the measurement time by 2.
Overlap percentage
The amount of overlap is specified as a
percentage of the time record. 0% is no overlap
and 99.8% is the maximum (511 out of 512
samples re-used). The maximum overlap is
determined by the amount of time it takes to
calculate an FFT and the length of the time record
and thus varies according to the span.
The SR760 always tries to use the maximum
amount of overlap possible. This keeps the display
updating as fast as possible. Whenever a new
frequency span is selected, the overlap is set to
the maximum possible value for that span. If less
overlap is desired, then use the Average menu to
enter a smaller value. On the widest spans (25, 50
and 100 kHz), no overlap is allowed.
Triggering
If the measurement is triggered, then overlap is
ignored. Time records start with the trigger. The
analyzer must be in continuous trigger mode to
use overlap processing.
2-8
INPUT RANGE
The input range on the SR760 varies from a
maximum of 34 dBV full scale to a minimum of 60 dBV full scale. A signal which exceeds the
current input range will cause the OvrLoad
message to appear at the bottom of the screen. A
signal which exceeds the maximum safe range will
turn on the HI V indicator.
The input range is displayed in dBV. The
maximum and minimum range equivalents are
tabulated below.
Max 34 dBVpk
31 dBVrms
50.1 Vpk
35.4 Vrms
Min -60 dBVpk
-63 dBVrms
1.0 mVpk
0.7 mVrms
ANALYZER BASICS
Manual Range
The input range can be specified in the Input
menu to be fixed at a certain value. Signals that
exceed the range will overload and become
distorted. Signals which fall to a small percentage
of the range will become hard to see.
Auto Range
The input range can be set to automatically correct
for signal overloads. When autoranging is on and
an overload occurs, the input range is adjusted so
that the signal no longer overloads. If the signal
decreases, the input range is not adjusted. You
must take care to ensure that the signal does not
fall dramatically after pushing the input range to a
very insensitive setting.
While the analyzer is performing linear averaging,
the input range is NOT changed even if the signal
overloads. The overload indicator will still light to
indicate an over range condition. Changing the
range during a linear average invalidates the
average.
2-9
ANALYZER BASICS
2-10
OPERATION
FRONT PANEL OVERVIEW
POWER BUTTON
The SR760 is turned on by pushing in the POWER
button. The video display may take a few seconds
to warm up and become visible. Adjust the
brightness until the screen is easily readable.The
model, firmware version and serial number of the
unit are displayed when the power is turned on.
A series of internal tests are performed at this
point. Each test is described as it is performed and
the results are represented graphically as OK or
NOT OK. The tests are described below.
RAM This test performs a read/write test to the
processor RAM. In addition, the
nonvolatile backup memory is tested. All
instrument settings are stored in
nonvolatile memory and are retained
when the power is turned off. If the
memory check passes, then the
instrument returns to the settings in effect
when the power was last turned off. If
there is a memory error, then the stored
settings are lost and the default settings
are used.
ROM This test checks the processor ROM.
CLR This test indicates whether the unit is
being reset. To reset the unit, hold down
the backspace [<-] key while the power is
turned on. The unit will use the default
settings. The default setup is listed in a
later chapter.
CLK This test checks the CMOS clock and
calendar for a valid date and time. If the
there is an error, the time will be reset to a
default time. Change the clock settings
using the SYSTEM SETUP menu.
DSP This test checks the digital signal
processors and fast memory.
A/DThis test checks the analog to digital
converter board.
VIDEO DISPLAY
The monochrome video display is the user
interface for data display and front panel
programming operations. The resolution of the
display is 640H by 480V. The brightness is
adjusted using the brightness control knob located
at the upper left corner. As with most video
displays, do not set the brightness higher than
necessary. The display may be adjusted left and
right using the Setup Screen function in the
SYSTEM SETUP menu.
3-1
OPERATION
horizontal axis. The graph is continuously updated
while the unit is in the RUN mode.
A complete description of the screen display
follows in the next section.
SOFT KEYS
The SR760 has a menu driven user interface. The
6 soft keys to the right of the video display have
different functions depending upon the information
displayed in the menu boxes at the right of the
video display. In general, the soft keys have two
uses. The first is to toggle a feature on and off or
to choose between settings. The second is to
highlight a parameter which is then changed using
the spin knob or numeric keypad. In both cases,
the soft keys affect the parameters which are
displayed adjacent to them.
KEYPAD
The keypad consists of five groups of keys. The
ENTRY keys are used to enter numeric
parameters which have been highlighted by a soft
key. The MENU keys select a menu of soft keys.
Pressing a menu key will change the menu boxes
which are displayed next to the soft keys. Each
menu presents a group of similar parameters and
functions. The CONTROL keys start and stop
actual data acquisition, select the marker and
toggle the active trace. These keys are not in a
menu since they are used frequently and while
displaying any menu. The SYSTEM keys print the
screen to a printer and display help messages.
Once again, these keys can be accessed from any
menu. The MARKER keys determine the marker
mode and perform various marker functions. The
marker functions can be accessed from any menu.
A complete description of the keys follows in the
next section.
SPIN KNOB
The spin knob is used to adjust parameters which
have been highlighted using the soft keys. Most
numeric entry fields may be adjusted using the
knob. In addition, functions such as display
zooming and scrolling use the knob as well. In
these cases, the knob function is selected by the
soft keys. The [MARKER] key, which can be
pressed at any time, will set the knob function to
scrolling the marker.
DISK DRIVE
The 3.5" disk drive is used to store data and
instrument settings. Double sided, double density
disks should be used. The disk capacity is 720k
bytes formatted. The disk format is DOS
compatible. Disks written by the SR760 may be
read by PC compatible computers equipped with a
3.5" drive and DOS 3.0 or higher.
Only use double sided double density (DS/DD)
disks. Do not use high density (DS/HD) disks.
Recently it has become difficult to find DS/DD
discs. However there is an easy way to convert a
high density DS/DD disc to a low density DS/DD
disc. To do this, proceed as follows:
1) Orient the high density disc so that the
label side is facing up and the metal
“slider” is facing down. Notice there are
two small rectangular holes near the top
edge of the floppy disc.
2) Use a single piece of sturdy opaque tape
to cover both sides of the hole in the
upper-left corner of the floppy disc (so
light cannot pass through the hole). Be
sure to stretch the tape tightly so it will not
snag when inserting the disc into the
SR760.
3) Format the disc in the SR760. Now the
disc is formatted as a DS/DD disc and can
be used in the SR760.
BNC CONNECTORS
TRIGGER
The rising or falling edge of the TRIGGER input
triggers a time record. The input impedance is 10
KΩ and the minimum pulse width is 10 ns. The
trigger level is adjustable from -5V to +5V with
either positive or negative slope. The minimum
pulse amplitude is 100 mV.
SIGNAL INPUTS
The input mode may be single-ended, A, or
differential, A-B. The A and B inputs are voltage
inputs with 1 MΩ, 15 pF input impedance. Their
connector shields are isolated from the chassis by
1 MΩ (float) or 50 Ω (ground). Do not apply more
than 50 V to either input. The shields should never
exceed 3V.
3-2
OPERATION
SCREEN DISPLAY
DATA DISPLAY
Data is graphed with signal on the Y axis and
frequency or time on the X axis. The physical size
of the graph remains constant while the vertical
and horizontal scales may be changed. The graph
area has a dotted grid for reference. There are 10
horizontal divisions and either 8 or 10 vertical
divisions. The frequency span consists of 400
frequency bins. The display normally shows all
400 bins. The X axis may be expanded and
translated to display less than 400 bins. This
expansion does not change the span or time
record, it merely changes the display of the data.
3-3
SINGLE and DUAL TRACE DISPLAYS
There are two data traces being acquired at all
times. The traces are labelled Trace0 and Trace1.
The traces may be different measurements, such
as spectrum and time record, or different displays,
such as magnitude and phase. When the two
traces are displaying live data, they have the same
signal input, frequency span, window function,
trigger, and averaging mode. If one of the traces is
a recalled file, then it can have a span and window
which differs from the live settings.
The display shown above is the SINGLE trace
format. The [ACTIVE TRACE] key toggles the
display between the two traces.
OPERATION
The dual trace or Up/Dn format is shown above.
The display format is selected in the DISPLAY
menu. Trace0 is always the upper trace. Each
trace is annotated the same way as the single
trace.
The left edge, center and right edge of the graph
are labelled directly below the graph. When
displaying spectral data with no horizontal
expansion, these values are the Start, Center and Stop frequencies of the frequency span in
use. When displaying time records, these values
are the Start, Middle and End of the time record. These times are always relative to the
start of the time record, they do not reflect any
trigger delay which may be programmed.
To expand a graph, use the SCALE menu. When
the display is expanded in the horizontal axis, the
labels reflect the displayed span and time, not the
actual acquisition span and time record. Expanded
traces have an EXPAND indicator below the right
hand edge of the graph as shown below.
The Top reference is the Y value of the upper
edge of the graph. The units can be Volts,
dBVolts, or EU (user defined engineering units).
The Vertical scale is shown as the number of dB,
Volts, or EU per division. This value is changed
whenever the vertical scale is adjusted.
The Window Function for the displayed data is
shown below the graph. In the case of a recalled
graph, this window is the one used to calculate the
recalled graph, not the window used for live
calculations.
The File Type refers to the source of the data
being displayed. Live means that the data is realtime, Calculated data is the result of Trace Math,
and a "filename" is data recalled from a disk file.
At the upper right, the measurement and display type and trace number, 0 or 1, are shown. The
measurement type can be Spectrum, PSD (power
spectral density), Time record, or Octave
analysis. The display types are Log Magnitude,
Linear Magnitude, linear Real part, linear
Imaginary part, and Phase.
On the active trace, the measurement type and
trace number are highlighted in inverse.
3-4
MARKER DISPLAY
The Marker Region is the graph region between
the two heavy vertical dashed lines. The marker
region may be set to 1 division (wide), 1/2 division
(norm), or a single vertical line (spot). The marker
region does not change with horizontal scaling.
The Marker is a small square which seeks the
minimum, maximum, or mean of the data within
the marker region. When seeking min or max, the
marker is located at the position of the data point
which is the min or max. This allows peaks and
valleys in the data to be easily read out. When
seeking the mean, the X position of the marker is
at the center of the marker region and the Y
position is the mean of the data within the region.
When a spot marker region is used, the marker is
confined to a single frequency or time bin.
The Marker Position displays the X position
(frequency or time) and the Y data of the marker.
Pressing the [MARKER] key will draw a box
around the marker information. When the marker
readout is surrounded by this box, the spin knob
adjusts the position of the marker region. The
marker region moves in increments of one
frequency or time bin.
MENU DISPLAY
The Soft Key menu boxes define the functions of
the 6 soft keys to the right of the screen. The
menu boxes are grouped into menus. Pressing
each of the ten Menu keys will display a different
menu of boxes. Related functions are grouped into
a single menu. In general, pressing a soft key
does one of two things. One is to toggle between 2
or 3 specific choices. An example is the Display
Format box illustrated on the previous page.
Pressing the first soft key toggles the display
between Single and Up/Dn. The second soft key
mode is to highlight an entry field and knob
function. An example would be the Start Freq.
Pressing the soft key will highlight the Start Freq.
value. The Start Freq. may then be adjusted with
the knob or entered as a value using the numeric
entry keys. Each menu is described at length in a
following section.
STATUS INDICATORS
In addition to the data display and menu boxes,
there are a number of status indicators which are
displayed at the bottom of the screen.
OPERATION
RUN/STOP/STOP-Invld
The RUN indicator is on whenever data is being
taken and spectra are being calculated. STOP
indicates that data is not being acquired and the
data display is not being updated. STOP-Invld
means that the data on the display may not match
the graph parameters or the analyzer settings. For
example, if the display is paused (using the
[PAUSE CONT] key or some other means), and
the span is changed, clearly the displayed data
does not reflect the new span. In this case, the
STOP-Invld indicator will turn on.
SETTLING
When changing between narrow frequency spans
with long acquisition times, the digital filter
requires some settling time before all of the data is
replaced with new data. This time is longer than
the record time. While this indicator is on, the filter
is still settling and the displayed spectrum may not
be accurate.
Input Range
The input range is always displayed. If the range is
set manually, the display is in normal characters. If
Auto Range is on, then inverse characters are
used.
OvrLoad
This indicator turns on if the input signal overloads
the analog amplifier or A/D converter.
No Avg./Avrging
This indicates whether averaging is in effect.
Averaging affects both traces if they are live. If
linear averaging is on, then the number below the
Avrging indicator is the number of averages
accumulated so far. If averaging is off or
exponential, then no number is displayed.
Trigger/Trg Wait
If triggering is on, then the Trigger indicator
flashes on whenever a time record is triggered.
TrgWait indicates that the unit is in triggered
mode and is waiting for a trigger to occur. Triggers
received while acquiring data from a previous
trigger are ignored.
Armed/Arm Wait
If triggering is on, the Armed indicator is on
whenever the unit is armed and awaiting a trigger.
Arm Wait means that the unit is in manual arming
mode and awaiting an arm command, either from
the front panel or via the computer interfaces.
3-5
OPERATION
GPIB/RS232
Flashes when there is activity on the computer
interfaces. This does not flash for printer or plotter
activity.
SRQ
This indicator is on whenever a GPIB Service
Request is generated by the SR760. SRQ stays
on until a serial poll is completed.
REM
This indicator is on when the front panel is locked
out by a computer interface. No front panel
adjustments may be made. To return the unit to
local control (if allowed), press the [HELP] key.
Pass/Fail
This indicates whether a trace passes or fails a
limit table test.
ERR
Flashes whenever there is a computer interface
error such as illegal command or out of range
parameter is received. This does not flash for a
printer or plotter error.
HV
The High Voltage indicator turns on whenever an
input greater than 50 V is detected. The analyzer
immediately switches in an attenuator to protect
the input circuitry. Any attempt to set the input
range to a setting which would remove this
attenuator will not be allowed until the input signal
is reduced to a safe level.
ALT
Indicates that the ALTERNATE keypad is in use.
The ALTERNATE keypad uses the alphabetic
legends printed below each key. To enter the ALT
mode, press the [ALT] key once. Pressing the
keys will now enter alphabetic characters into the
active entry field. The [0]...[9], [.], [-], [<-] and [ALT]
have the same function in the ALTERNATE
keypad. To return to the normal keypad, press the
[ALT] key again.
3-6
OPERATION
KEYPAD
NORMAL AND ALTERNATE KEYS
The normal key definitions are printed on each
key. In addition, each key also has an alternate
definition printed below it. The [ALT] key toggles
the keypad between the two definitions. The ALT
screen indicator is on when the alternate
definitions are in use. The [0]...[9], [.], [-], [<-] and
[ALT] keys have the same definition in both
modes. The alternate keys should only be used
when accessing files on the disk drive or labelling
plots.
MENUKEYS
All operating parameters of the SR760 are
grouped into ten function menus. The ten menu
keys select which menu of parameters is
displayed next to the six soft keys. The soft keys
then either toggle a parameter, highlight a
parameter entry field (for numeric entry or knob
adjustment), or display a submenu. The menus
are listed below.
[FREQ]Sets the frequency span and start
and center frequencies.
[MEAS]Displays submenus for selecting
the measurement type, displayed
quantity, units, window type, and
calculator.
[DISPLAY]Sets the display format, marker
on/off, and grid modes.
[INPUT]Configures the signal inputs, sets
the manual input range and
trigger setup.
[SCALE]Sets the graph scaling and
expansion and selects linear or
log X axis.
[ANALYZE]Turns on harmonic, sideband and
band analysis as well as data and
limit tables.
[AVERAGE]Turns averaging on and off and
selects the averaging type.
[PLOT]Plots the data graph on a plotter.
Also prints settings, data and limit
tables to the printer.
[SYSTEM SETUP]
Configures the computer
interfaces, sound, real time clock,
plotter, printer, and screen
3-7
OPERATION
location. The [TEST] submenu
tests the keypad, external
keyboard, knob, RS232 interface,
printer interface, disk drive, video
screen, and memory. The [INFO]
submenu displays various
information screens.
[STORE RECALL]
This menu stores and recalls data
and settings to and from the disk.
Also contains a disk utilities
submenu.
Detailed descriptions of each menu are provided
in a later chapter.
ENTRY KEYS
The numeric entry keys can be used to directly
enter parameter values. Parameters may be
entered only if their menu box is displayed and
their entry field is highlighted. For example, if the
FREQ menu is displayed, the fifth soft key is next
to the Start Freq. box. Pressing this soft key will
highlight the entry field displaying the start
frequency. The menu box will appear as below.
A new start frequency may now be entered using
the numeric keys. For example, to set the start to -
1.25 kHz, press [1] [.] [2] [5]. As soon as the [1] is
pressed, the entry parameter is displayed in the
upper left hand corner of the screen as shown
below.
Note that the frequency menu is
also replaced
menu. This menu shows the
available units for the active
entry field, in this case mHz, Hz,
or kHz.
The entry field displays the characters as the keys are pressed.
The '-' is the entry point. If an
error is made, the backspace [<] key will erase the last
character. Pressing the Escape
with a units
soft key will abort the entry operation and leave
the value unchanged. When the entry string is
correct, press the kHz units soft key to change the
start frequency to the new value.
Entries may be made in exponential form using
the [EXP] key. The entry above may be made by
pressing [1] [2] [5] [EXP] [1] and then the Hz units
soft key.
In general, whenever a parameter entry field is
highlighted, the knob may also be used to adjust
the value. If the knob is turned while making a
numeric entry but before a units key has been
pressed, the knob will adjust the marker position
instead.
Some entry fields allow only knob adjustment or
only numeric entry.
START and PAUSE/CONT
The [START] and [PAUSE/CONT] keys are used
to start, pause and continue data acquisition. If the
unit is in the RUN mode acquiring and displaying
data, as indicated by the RUN indicator, then the
[PAUSE/CONT] key will halt data acquisition. The
RUN indicator switches to STOP and no new
spectra will be taken and the display will not be
updated. If averaging is off, then either the
[START] key or the [PAUSE/CONT] key will
resume acquisition. If averaging is on, the
[START] key will reset the average and restart
acquisition. [PAUSE/CONT], on the other hand,
will continue the average where it was paused. In
the case of linear averaging when the average is
already completed, the [PAUSE/CONT] does
nothing since there is no average to continue.
MARKER
Pressing the [MARKER] key highlights the marker
information field by drawing a box around it. The
knob will now scroll the marker region. The
highlighted marker field appears below.
Any previously highlighted parameter field will
become non-highlighted. Pressing a soft key to
highlight a new parameter field will let the knob
adjust the new parameter while the marker
becomes unselected.
3-8
ACTIVE TRACE
Pressing [ACTIVE TRACE] toggles the active
trace. In the single trace display format, the graph
switches between Trace0 and Trace1. In the dual
screen display, [ACTIVE TRACE] switches which
trace is active as indicated by the highlighted trace
identification at the upper right of the graph. In
both cases, the active trace determines which
trace's parameters are displayed in the menus.
For example, activating Trace0 and then selecting
the Measure menu will allow you to select the
measurement for Trace0. Pressing [ACTIVE
TRACE] once allows you to select the
measurement for Trace1 using the same menu.
Only those parameters which are associated with
an individual trace have differing values between
the traces. Parameters such as input
configuration, frequency span and window
function are the same for any live trace.
AUTO RANGE
Pressing [AUTO RANGE] toggles the input
ranging mode between Manual and Auto. In
Manual mode, the input range is set within the
INPUT menu. When the mode is toggled to Auto,
the input range is stepped quickly from -60 dB
towards +30 dB until no overload is detected. Any
overload in the signal will cause the input scale to
change to remove the overload. If the signal
decreases, the input scale is not changed. The
range can be autoranged at any time by toggling
the mode from Auto to Manual and back to Auto.
Switching back to Manual ranging leaves the input
range at the current setting.
The Input Range indicator will be in inverse
characters if Auto Ranging is on.
AUTOSCALE
Pressing [AUTOSCALE] will automatically set the
vertical scale and translation to display the entire
range of the data. [AUTOSCALE] does not affect
the horizontal scaling.
[AUTOSCALE] may be pressed at any time during
or after data acquisition.
[AUTOSCALE] only operates on the data which is
displayed on the graph. If the graph is expanded,
data corresponding to frequency or time bins
which are not shown do not figure in the
autoscaling calculations.
SPAN UP and SPAN DOWN
OPERATION
The [SPAN UP] and [SPAN DOWN] keys
increment and decrement the frequency span by a
factor of 2. These keys provide a way of adjusting
the span when any menu is displayed. The span is
adjusted with either a fixed start or fixed center
frequency depending upon which frequency field
was most recently activated in the FREQ menu.
MARKER ENTRY
In the ANALYZE menu, pressing this key will enter
the marker frequency into the Fundamental
(Harmonic analysis), Carrier (sideband analysis),
Band Start and Center (band analysis) frequency
fields. This key also enters the marker frequency
into the X Value field of the Data and Limit Tables.
MARKER MODE
The [MODE] key in the MARKER section of the
keypad brings up a menu. This menu selects
linked cursors in the dual trace display and allows
marker offsets to be entered manually. The Peak
find functions are also in this menu.
MARKER REF
The [MARKER REF] key toggles the marker offset
or reference mode. Pressing this key once will turn
on the marker offset and set the X and Y offset to
the value of the current marker position.
Subsequent marker readings are relative to the
reference or offset values. The offset marker is
indicated by a ∆ (delta symbol) preceding the
marker readout above the graph as shown below.
The [MARKER REF] key may be used in any
menu.
The marker offset location on the graph is marked
by a small star shaped symbol.
MARKER CENTER
The [MARKER CENTER] key sets the span center
frequency to the marker frequency. If the span is
large so that this operation would require a span
which extends below 0 Hz or past 100 kHz, then
the span is decreased to the largest span which
allows the marker frequency to be the center.
MARKER MAX/MIN
Pressing [MARKER MAX/MIN] will center the
marker region around the maximum or minimum
3-9
OPERATION
data value on the screen. The Marker Seeks
mode in the DISPLAY menu chooses whether this
key finds the on-screen max or min. If the marker
seeks the mean, then the [MARKER MAX/MIN]
key finds the maximum on-screen point. The
marker will be positioned at the Min, Max, or Mean
of the data within the region, depending upon the
seeks mode. The [MARKER MAX/MIN] key only
searches the data which is on the screen. If the
max/min value occurs at more than one location,
then the one closest to the left edge is found.
PRINT
[PRINT] will print the currently displayed screen to
a printer attached to the rear panel parallel printer
port. The entire screen, including text and menus,
is printed. The time and date will also be printed.
The printer type needs to be configured in the
SYSTEM SETUP menu before using [PRINT]. A
"Printing in Progress" message will appear on the
screen while printing occurs. No other front panel
operations may be performed until printing is
completed. If no printer is attached or there is a
printer error, then the print operation is aborted
after about 10 seconds. A "Print Aborted!"
message will appear briefly on the screen.
HELP
[HELP] provides on screen help with any key or
soft key. Pressing [HELP] followed by any key will
display information about the function or use of
that key. [HELP] with a soft key will describe the
menu item next to the soft key. Pressing another
key will exit the help screen.
The [PRINT] key is the one key for which no help
is available. Pressing [PRINT] at any time will print
the screen, including the help screens.
LOCAL
When a host computer places the unit in the
REMOTE state, no keypad or knob input is
allowed. To return to front panel operation, press
the [HELP] key.
3-10
REAR PANEL
POWER ENTRY MODULE
The power entry module is used to fuse the AC
line, select the line voltage, and block high
frequency noise from entering or exiting the
instrument. Refer to the first page of this manual
for instructions on selecting the correct line
voltage and fuse.
IEEE-488 CONNECTOR
The 24 pin IEEE-488 connector allows a computer
to control the SR760 via the IEEE-488 (GPIB)
instrument bus. The address of the instrument is
set in the SETUP GPIB menu.
Also, a GPIB plotter with HPGL compatible
graphics may be connected to the IEEE-488 port.
In this case, the SR760 will control the plotter to
generate plots of the screen graph. Use the
SETUP PLOTTER menu to configure the SR760
for use with a GPIB plotter.
RS232 CONNECTOR
The RS232 interface connector is configured as a
DCE (transmit on pin 3, receive on pin 2). The
baud rate, parity, and word length are
programmed from the SETUP RS232 menu. To
connect the SR760 to a PC serial adapter, which
is usually a DTE, use a straight thru serial cable.
OPERATION
Also, a serial plotter with HPGL compatible
graphics may be connected to the RS232 port.
The SR760 will drive the plotter to generate plots
of the screen graph. Use the SETUP PLOTTER
menu to configure the SR760 for use with a serial
plotter.
PARALLEL PRINTER CONNECTOR
The [PRINT] key will print the screen to an Epson
compatible graphics printer or an HP LaserJet
compatible laser printer. Use a standard printer
cable to attach the printer to the printer port. Use
the SETUP PRINTER menu to choose the type of
printer.
PC KEYBOARD CONNECTOR
An IBM PC or XT compatible keyboard may be
attached to the keyboard connector. An AT
keyboard may be in its PC or 8088 mode. Typing
at the attached keyboard is the same as entering
numbers and letters from the front panel keypad.
Highlighted parameter entry fields will accept
characters from the keyboard. Typing 'E' or 'e' is
the same as [EXP]. In general, the keyboard is
only useful for alphabetic fields such as file names
or plot labels.
3-11
OPERATION
3-12
FREQUENCY MENU
Frequency
The Frequency menu is used to set the frequency span and location for the
measurement.
SpanPressing the Span key selects the frequency span as the active entry field. A
new span may be entered from the numeric keypad or the knob may be used
to adjust the span. The frequency span ranges from 191 mHz to 100 kHz in
factors of 2. A numerically entered span is rounded up to the next largest
allowable span.
If the new span is incompatible with the 0 to 100 kHz frequency range
because the start or center frequency is close to the limits of the range, then
the start or center frequency will be adjusted to accommodate the new span.
Changing the span will change the Linewidth (Span/400) and Acquisition
Time (400/Span).
LinewidthThe Linewidth key selects the linewidth as the active entry field. The
linewidth is defined as the span divided by 400. The linewidth ranges from
.477 mHz to 250 Hz in factors of 2. A numerically entered linewidth is
rounded up to the next largest allowable linewidth.
Changing the linewidth will change the Span (Linewidth*400) and Acquisition
Time (1/Linewidth). If the new span is incompatible with the 0 to 100 kHz
frequency range because the start or center frequency is close to the limits of
the range, then the start or center frequency will be adjusted to
accommodate the new span.
Acquisition TimeThe Acquisition Time key selects the acquisition time as the active entry
field. The acquisition time is defined as the reciprocal of the linewidth. The
acquisition time ranges from 2097.1 s to 4.00 ms in factors of 2. A
4-1
FREQUENCY MENU
numerically entered acquisition time is rounded down to the next fastest
allowable acquisition time.
Changing the acquisition time will change the Span (400/Acquisition Time)
and Linewidth (1/Acquisition Time). If the new span is incompatible with the 0
to 100 kHz frequency range because the start or center frequency is close to
the limits of the range, then the start or center frequency will be adjusted to
accommodate the new span.
Full SpanPressing this key immediately sets the Span to 100 kHz, Linewidth to 250
Hz, Acquisition Time to 4.00 ms, Start Frequency to 0.0 Hz, and Center
Frequency to 50.0 kHz.
Start FrequencyThe Start Frequency key selects the start frequency of the span as the active
entry field. The knob adjusts the start frequency in steps equal to the
linewidth. A numerically entered frequency is rounded to the nearest
frequency bin (exact multiple of the linewidth). If the new start frequency is
incompatible with the span because of the 0 to 100 kHz range limits, then the
start frequency will be set to the closest allowable value.
Center FrequencyThe Center Frequency key selects the center frequency of the span as the
active entry field. The knob adjusts the center frequency in steps equal to the
linewidth. A numerically entered frequency is rounded to the nearest
frequency bin (exact multiple of the linewidth). If the new center frequency is
incompatible with the span because of the 0 to 100 kHz range limits, then the
center frequency will be set to the closest allowable value.
Note:
Activating the Start or Center Frequency fields fixes the start or center
frequency for subsequent adjustments to the frequency span. Further
adjustments to the span leave the span start or center untouched, even
when the start or center frequency becomes de-activated as a menu choice.
The most recently activated of the Start or Center Frequency fields sets the
span adjustment mode.
Enlarging the frequency span may change the start and center frequencies.
This is because these frequencies are always exact frequency bins or
multiples of the linewidth. Larger spans have larger linewidths and thus the
start and stop frequencies may need to be rounded to the nearest allowable
bin of the new span.
4-2
MEASURE MENU
Measure
The Measure menu is used to select the measurement type, display type,
units and window function. The Measure menu also activates the calculator
for trace math.
Measure KeysEach Measure Key activates a sub menu. Each sub menu is described in
detail in the following pages.
4-3
MEASURE MENU
Measure
The Measure sub menu selects the type of measurement for the active trace.
SpectrumThe SR760 filters the input data in real time to provide a time record with the
desired frequency span and then performs an FFT on this record. Pressing
the Spectrum key displays this FFT on the active trace.
PSDThe PSD or Power Spectral Density is the magnitude of the spectrum (the
square root of [the FFT times its complex conjugate]) normalized to a
bandwidth of 1 Hz. This measurement approximates the amplitude within a 1
Hz bandwidth located at each frequency bin. The actual linewidth and
window function are compensated for in this calculation. This allows
measurements taken with different spans or windows to be compared.
Note:
PSD measurements are typically used to measure noise or noise density.
The data values are read out in Volts/√Hz or dBV/√Hz. When measuring
Gaussian noise sources, the noise in bandwidths other than 1 Hz may be
obtained by multiplying the reading by the square root of the desired
bandwidth. This is true only for Gaussian noise.
When measuring PSD, the Display may only be set to Log Magnitude or
Linear Magnitude.
4-4
MEASURE MENU
Time Record The Time Record is the minimum amount of filtered input data required to
generate an FFT with the desired span and linewidth. The SR760 filters the
input data in real time to provide a stream of data points with the correct
frequency span. The time record consists of 512 of these points, of which
only the first 400 are displayed.
When averaging is on, only spectra are averaged. The Time Record shows
the latest time record used to calculate a spectrum.
Note:
The SR760 is not a digital oscilloscope. The Time Record always shows
filtered data and does not resemble an oscilloscope trace of the same input.
The input data filter is a complex filter yielding complex outputs. Thus, the
time record has a real and imaginary part as well as phase associated with
each time bin.
Octave AnalysisOctave Analysis computes the spectral amplitude within 1/3 octave bands.
The analyzer computes a normal FFT, then calculates the rms sum of the
frequency components within each band. When Octave Analysis is on, only
the Log Magnitude may be displayed. Also, the display is always logarithmic
on the X axis, displaying evenly spaced octaves. The left and right most
bands are labelled on the graph by center frequency and band number. The
marker reads the center frequencies of the bands rounded to the nearest
even frequency. The actual band frequencies are exact according to the
ANSI standard.
Note:
When octave analysis is on for either trace, the FREQ menu will display the
band menu for both traces. This is because both traces must have the same
span. Thus, if one trace is measuring octave analysis, the other trace's span
is determined by the bands displayed in the octave analysis.
Furthermore, in order to perform 30 band analysis accurately, the SR760
must combine spectra taken with two different overlapping spans. This is
because the frequency range of 30 bands requires more than 400 linearly
spaced frequency points. When 30 band analysis is chosen, the analyzer
alternates between two different spans. If one trace is displaying 30 band
octaves, then the other trace will show spectra taken with alternating
frequency spans and is not very useful. In general, when using octave
analysis, only the trace showing octaves is meaningful.
Only one trace may be measuring octave analysis at a time. The other trace
must be measuring spectrum, PSD or time record.
4-5
MEASURE MENU
To choose the number of bands displayed, the starting band and the
weighting function, use the FREQ menu. The FREQ menu will display the
band selection menu shown at the right whenever octave analysis is on.
# BandsThe # Bands key toggles the octave analysis range between 15 and 30
bands. The bands are always 1/3 octave.
Starting BandPressing this key activates the Starting Band number entry field. The SR760
can display bands -2 through 49. The starting band can range from -2 to 49
minus the number of bands (15 or 30).
WeightingThe Weighting key toggles between no weighting and A weighting. A
weighting compensates for auditory sensitivity and can provide data
comparable to that derived from analog analysis equipment.
ReturnThe Return key will return to the main MEAS menu.
4-6
MEASURE MENU
Display
The Display sub menu allows the user to choose the displayed quantity for
the active trace.
Log Mag.This key displays the magnitude of the measurement on a logarithmic scale.
Only the active trace display is affected. Both the Time Record (as defined in
this analyzer) and the corresponding FFT are complex quantities. The
magnitude is the square root of the product of the measurement data and its
complex conjugate.
Linear Mag.This key displays the magnitude of the measurement on a linear scale. Only
the active trace display is affected. Both the Time Record (as defined in this
analyzer) and the corresponding FFT are complex quantities. The magnitude
is the square root of the product of the measurement data and its complex
conjugate.
Real PartThis key displays the real part of the measurement on a linear scale. Only
the active trace display is affected. Both the Time Record (as defined by the
SR760) and the corresponding Spectrum are complex quantities and thus
have a real part. PSD and Octave Analysis are not complex and only display
magnitudes.
Imag. PartThis key displays the imaginary part of the measurement on a linear scale.
Only the active trace display is affected. Both the Time Record (as defined
by the SR760) and the corresponding Spectrum are complex quantities and
thus have an imaginary part. PSD and Octave Analysis are not complex and
only display magnitudes.
4-7
MEASURE MENU
Phase This key displays the phase of the measurement on a linear scale. Only the
active trace display is affected. Both the Time Record (as defined by the
SR760) and the corresponding Spectrum are complex quantities and thus
have phase. PSD and Octave Analysis are not complex and only display
magnitudes.
ReturnThe Return key will return to the main MEAS menu.
4-8
MEASURE MENU
Units
The Units sub menu allows the user to choose the display units for the active
trace.
Volts Pk (EU PK)This key chooses units of Volts Peak or Engineering Units Peak for the
active trace.
Volts RMS (EU RMS)This key chooses units of Volts RMS or Engineering Units RMS for the active
trace.
dBV (dBEU)This key chooses units of dBVolts Peak or dBEngineering Units Peak for the
active trace.
dBVRMS (dBEURMS)This key chooses units of dBVolts RMS or dBEngineering Units RMS for the
active trace.
dB units are not available when displaying Real or Imaginary parts of the
spectrum. This is because the data values may be negative.
Volts/EUThis key chooses whether the fundamental unit is Volts or user defined
Engineering Units (EU). Choosing EU will activate the EU definition menu
shown below.
4-9
MEASURE MENU
EU Label Pressing this key activates the EU Label entry field. Use the ALT keys to
enter a name for the engineering units.
EU/Volt Pressing this key activates the EU scaling entry field. Enter the number of
engineering units per Volt.
ReturnThis key returns to the Units sub menu.
When Phase is being displayed on the active trace, the Units menu appears
as shown to the left. Phase values are always between -180 and +180
degrees. The analyzer does not "unwrap" phase.
The phase of a particular frequency bin is set to zero if neither the real nor
imaginary part of the FFT is greater than 0.012% of full scale (-78 dB below
f.s.). This avoids the messy phase display associated with the noise floor.
(Remember, even if a signal is small, its phase extends over the full 360
degrees.)
Degs This key chooses degrees for the Phase display.
Rads This key chooses radians for the Phase display.
Return The Return key will return to the main MEAS menu.
Note:
The choice of units does not affect the display scaling, whether linear or
logarithmic. The Marker and data readouts reflect the choice of units but the
graph remains unchanged.
4-10
MEASURE MENU
Window
The Window submenu allows the user to choose the window function. Both
traces use the same window function. A trace may be recalled from disk with
a window different than the "live" window. This is the only case where the
window on the graph is other than the "live" window shown in this menu.
UniformThis key selects no windowing (uniform or rectangular window function) of
the time record. This window provides high amplitude accuracy only for
frequencies exactly on a bin and poor frequency selectivity making it a poor
choice for continuous signals. It is primarily useful for analyzing impulses and
transients which are shorter than a time record.
FlattopThis key selects the Flattop window. This window has the least ripple and
thus the smallest amplitude errors for frequencies not exactly on a bin. It is
most useful for precise amplitude measurements.
HanningThis key selects the Hanning window. The Hanning window has a relatively
narrow mainlobe and low sidelobes providing low leakage (spectral
broadening) and good selectivity.
BMHThis key selects the Blackman-Harris window. This window has the
narrowest mainlobe and the fastest roll-off for the best selectivity. This
window is especially useful in measurements requiring the more than 70 dB
of dynamic range since it has the lowest leakage and broadening of the
skirts.
ReturnThe Return key will return to the main MEAS menu.
4-11
MEASURE MENU
Calculator
Calculations treat the data as intrinsic values, either Volts, EU or degrees. If
Performing a calculation on the active trace will set the File Type to Calc to
The Calculator sub menu allows the user to perform arithmetic calculations
with the trace data. Operations are performed on the entire trace, regardless
of graphical expansion.
a graph is showing dB, then multiplying by 10 will raise the graph by 20 dB
and dividing by 10 will lower the graph by 20 dB.
indicate that the trace is not Live. This is shown by the "File=Calc" message
at the lower left of the graph. The analyzer continues to run, but the
calculated trace will not be updated. To return the trace to live mode, activate
the trace and press the [START] key. The File Type will return to Live.
OperationThe Operation function selects the type of operation to be performed. The
add, subtract, multiply, and divide functions require a second argument
which may be a number, w (2πf), or the other trace. The log (base 10) and
square root function require no argument.
Do CalcPressing this key starts the actual calculation. The "Calculating" message
appears below the graph while calculations are in progress. The calculation
uses the operation specified by the Operation key and uses the argument
chosen by the Argument keys.
Note that many operations will require an AutoScale to display the result on
the graph.
4-12
MEASURE MENU
Argument Type The Argument Type function selects between a constant argument, w (2πf),
and a second data trace. A constant argument adds or subtracts a constant,
or multiplies or divides by a constant. Choosing w uses the argument
2π•frequency for each frequency bin. The other graph option uses the other
(inactive) trace as the argument. There is no attempt to check whether the
spans are the same, or even whether the measurement data are of the same
type. In this case, calculations are performed on a bin by bin basis, i.e. bin #1
of one trace is added to bin #1 of the other trace, bin #2 is added to bin #2,
etc. In the case of divide, the active trace is divided by the inactive trace. A
disk file may be used as one of the traces by recalling a file into one of the
graphs.
If the Argument type is a constant, then the Argument and Marker to
Argument functions are displayed.
ArgumentPressing the Argument key activates the constant argument entry field. Use
the keypad to enter a numerical argument. Integer (-3), real (-3.0), or floating
point (-0.3E+1) formats are all allowed.
Marker to Arg.The Marker to Argument will copy the data value of the marker to the
constant argument field above. This is convenient when subtracting a
baseline or normalizing to a data point.
Note:
This function takes the literal marker readout as shown above the graph and
copies it to the argument field. This is true even if the marker is reading in
dBV rather than Volts. The calculation will use the argument as if it were
Volts and result in meaningless data. Use linear units when using the Marker
to Arg. function to avoid this mistake.
ReturnThe Return key will return to the main MEAS menu.
4-13
MEASURE MENU
4-14
DISPLAY MENU
Display
The Display menu is used to change the graph parameters and marker type.
The settings for the active trace are displayed in this menu. Note that marker
movement is activated by the [MARKER] key and not by this menu.
FormatThe Format key toggles between single and dual trace screen formats. The
[ACTIVE TRACE] key toggles the active trace. If a single graph is displayed,
the [ACTIVE TRACE] switches the graph between Trace0 and Trace1. If two
graphs are displayed, [ACTIVE TRACE] selects either the upper or lower
trace as active. When the format is switched back to Single, the active trace
becomes the single displayed trace.
Marker On/Off/TrkThis function turns the marker on and off or selects tracking mode. This
function only affects the active trace. Each trace has its own marker. It is
sometimes desirable to turn off the marker before printing the screen. When
the marker is set to Track, the marker automatically seeks the maximum or
minimum point of the trace (according to the Marker Seeks selection).
Marker WidthThis function selects the width of the marker region defined by the vertical
dashed lines on the graph. Only the marker for the active trace is affected.
Normal width is 1/2 of a division, Wide is 1 division, and Spot is a single X
position on the screen (the marker is a single dashed line).
The marker region moves to the left and to the right a single bin at a time.
4-15
DISPLAY MENU
Marker Seeks The marker searches the data points within the marker region for the
maximum or minimum data value, or calculates the mean of the region. This
key toggles between Max, Min, and Mean and only affects the active trace's
marker.
When seeking minimum or maximum, the marker is located at the minimum
or maximum data point. This allows peaks and valleys in the data to be read
easily. When seeking mean, the X position of the marker is the center of the
marker region and the Y position is the mean of the data within the region.
Grid Div/ScrnThis function selects no grid or 8 or 10 vertical divisions per graph. The grid
is the set of dotted lines on the graph which mark each scale division. This
affects only the active trace graph.
Graph FillThe active trace can be selected to display the spectrum as the envelope of
the X values (line), or to fill the solid region below the trace (fill).
4-16
MARKER MODE MENU
Marker Mode
The Marker Mode menu is activated with the Mode key in the Marker area of
the keypad. This menu is used to manually enter a marker offset as well as
searching for peaks in the data.
Linked MarkersThis key links the markers on the two traces. When the dual trace display
format is used, linked markers means that the two markers are always at the
same location on the graph. This is true even if one of the traces is showing
the time record. This is strictly a graphical function. To move the markers,
activate either trace and use the [MARKER] key to move the markers with
the knob.
Marker OffsetThis function turns on the marker offset. When marker offset is on, a small
delta (∆) is displayed at the beginning of the marker readout above the
graph. The marker readout is now relative to the marker offset. The marker
offset location on the graph is marked with a small star shaped symbol.
The [MARKER REF] key toggles the marker offset on and off as well. When
the [MARKER REF] key turns on the offset, the X and Yoffsets are set to the
current marker position. Pressing the [MARKER REF] key again turns the
marker offset off.
X OffsetThis key activates the marker X Offset entry field. This is the offset of the
marker along the X axis. Only numeric entry is permitted. The X offset is
stored as unitless number. When displaying spectra, the offset is interpreted
as a frequency. The X offset does not have to be a frequency which is within
the current span.
4-17
MARKER MODE MENU
Y Offset This key activates the marker Y Offset entry field. This is the offset of the
marker along the Y axis. Only numeric entry is permitted. The offset is stored
as a unitless number. If the display units are changed, then the Y offset
needs to be changed. The Y offset does not have to be a value which is
currently within the vertical span of the graph.
Next Peak LeftThis function moves the marker to the next peak to the left of the current
marker position.
Next Peak RightThis function moves the marker to the next peak to the right of the current
marker position.
4-18
INPUT MENU
The Input menu is used to change the input configuration and input range. In
Input
addition, the Trigger and Arming submenus set the triggering mode, level
and delay.
Input SourceThe Input Source key selects the front end signal input configuration. The
input amplifier can be single-ended (A) or differential (A-B). In general, when
looking at very small signals, connect A to the signal source and B to the
signal source ground and use A-B. In this case, make sure that the two input
cables do not encompass any loop area (twist them together or run them
side by side).
GroundingThis key chooses the shield grounding configuration. The shields of the input
connectors (A and B) are not connected directly to the analyzer chassis
ground. In Float mode, the shields are connected by 1 MΩ to the chassis
ground. In Ground mode, the shields are connected by 50 Ω to the chassis
ground. In this mode, do not exceed 3V on the shields. The impedance
between the center conductor of each input and the shield is always 1 MΩ.
Note:
When the input source configuration is set to A-B, the grounding is
automatically set to Ground. This is because in the A-B case, the shields are
exactly that, shields, and do not carry signal.
CouplingThis key toggles the input coupling between AC and DC. The 3 dB
bandwidth of the AC setting is 0.16 Hz.
Input RangeThis key activates the Input Range entry field. The input range can only be
adjusted using the knob. The displayed value is the full scale signal input just
before overload. The input range limits are -60 dBV to +34 dBV in 2 dBV
4-19
INPUT MENU
steps. If the input ranging mode is Manual (as toggled by the [AUTO
RANGE] key and displayed by the Man.Rng. indicator), then this field sets
the input range. If the input ranging mode is AutoRng, then this field displays
the current range. Adjusting the input range automatically toggles back to
manual ranging mode.
Trigger MenuThis key displays the Trigger configuration menu as described on the
following page.
Auto OffsetThis function enables Auto Offset calibration. When Auto Offset is On, the
analyzer will periodically perform an auto offset calibration. Auto Offset is
always set to On when the analyzer is turned on. Setting Auto Offset to Off
defeats the periodic offset calibration. Turning Auto Offset On immediately
performs an offset calibration and then every few minutes for the first half
hour and then less often after that.
The offset calibration takes about a 10 seconds. After offset calibration, the
DC frequency bin of any baseband measurement will be minimized.
To calibrate the offsets, the inputs are internally grounded and the amplifier
offsets are measured and stored. When making very narrow span
measurements, the analyzer will need to re-settle after a calibration. In these
cases, it may be best to leave the Auto Offset Off and perform the calibration
only when necessary.
4-20
INPUT MENU
Trigger
The Trigger submenu is used to set the trigger mode, level and delay. The
Arming submenu selects the arming mode.
TriggerThe Trigger key selects the trigger mode and source.
Continuous is the same as free run. The analyzer takes time records
continuously.
Internal trigger means that time records are triggered by the input signal
itself. This is similar to an oscilloscope on internal trigger.
On External or External TTL, the time records are triggered by the external
trigger input on the front panel. Ext TTL triggers on a TTL level signal while
Ext trigger has a variable threshold.
Trigger LevelThis key activates the Trigger Level entry field. The level may be entered in
Volts or in percent of the input range (for Int trigger) or percent of 5V (for Ext
trigger). The knob only adjusts the percent value. A Voltage entry outside of
the limits of -100.0% to +99.22% of the applicable range will set the level to
the limit. The Voltage reading below the entry field displays the trigger level
in Volts.
Note:
Remember that the trigger requires a minimum 100 mV pulse amplitude to
successfully trigger. When using the internal trigger, this means that the
signal must exceed roughly 8% of the input range (-22 dB below the input
range).
4-21
INPUT MENU
Trigger Slope This key toggles the Trigger Slope between rising edge and falling edge.
Trigger Delay This key activates the Trigger Delay entry field. The delay may be entered
numerically or adjusted using the knob. The delay is set as a number of
samples rather than time.
The triggered time record does not have to start with the trigger event. The
time record can start before the trigger (negative delay values) or after the
trigger (positive delay values). A delay of 0 starts the time record with the
next sample following the trigger.
When the delay is positive, the delay is set with a resolution of one sample or
1/512 of the time record. This is equal to the acquisition time divided by 512
(7.8125 µs at full span). The positive limit of the delay is 65000 samples.
When the delay is negative, the delay resolution is one sample of the A/D
conversion or 3.9062 µs. The limit of the negative delay is 13300 (51.95 ms)
samples.
The time readout below the entry field is the equivalent delay in units of time.
Remember, changing the frequency span will change the positive delay
times because the time record and sample times change. The negative
delays are not affected by the span.
Arming MenuThis key displays the Arming configuration menu as described on the
following page.
Return This key returns to the main Input menu.
Note:
When the unit is successfully triggered, the Trigger indicator will light. No
indicator is present in the Continuous mode.
When the analyzer is in a triggered mode (not Continuous), and no triggers
are received, the display will not update even though the RUN indicator is on.
When no triggers have been received after a couple of seconds, the Trg Wait
indicator will turn on as a reminder that the unit is awaiting a trigger.
When a large trigger delay is used, the display may update slower since the
acquisition time for each record is the length of the time record plus the
trigger delay (which can be noticeably long).
The unit will trigger only if the trigger is armed. See the Arming menu.
4-22
INPUT MENU
Arming
Arming Mode This key selects the Arming Mode. Triggers are ignored unless the trigger is
Auto arming means that as soon as one triggered time record is processed,
With Manual arming, no time records are taken until the trigger is armed,
ArmThis key manually arms the trigger. This function only appears if the arming
The Arming submenu selects the arming mode.
armed. Arming allows a single triggered time record to be isolated even
when using a repetitive trigger source.
the trigger is immediately re-armed. Time records are basically taken as fast
as the trigger delay and actual time record length permit.
either from the front panel using the Arm Trigger softkey, or upon receipt of
an arming command from a computer interface. Once the trigger is armed,
the next trigger event will trigger a time record. The trigger is not
automatically re-armed but waits for an arming command or key.
Note:
If the trigger mode is continuous, the arming mode has no effect.
If the unit is in a triggered mode and arming is set to manual, then the Arm
Wait indicator will light if the unit is not armed after a few seconds. This is a
reminder that the unit is waiting. Once the unit is armed, the Armed indicator
will light. The Trg Wait and Trigger indicators will not turn on unless the unit
is armed.
mode is manual.
4-23
INPUT MENU
Return This key returns to the Trigger configuration submenu.
4-24
INPUT MENU
4-25
SCALE MENU
Scale
The Scale menu is used to change the graph X and Y scaling parameters for
the active trace.
Top ReferenceThis key activates the Top Reference entry field. The top reference is the Y
value of the top of the active trace graph. The top reference is expressed in
the same units as the display and marker as set in the Measure menu.
Bottom ReferenceThis key activates the Bottom Reference entry field. The bottom reference is
the Y value of the bottom of the active trace graph. The bottom reference is
expressed in the same units as the display and marker as set in the Measure
menu.
Y/DivThis key activates the Y/Division entry field. This value is the vertical scale of
the active trace graph. If the display is linear, then the vertical scale is
expressed in the displayed units. If the display is logarithmic, then the vertical
scale is always dB/division. Remember that the reference values and marker
readouts in this case are still shown in the previously selected units.
Using the knob adjusts the vertical scale in a 1-2-5-10 sequence. Knob
adjustments leave the marker at the center of the graph. The top and bottom
reference will be adjusted to make this happen. This gives the effect of
vertical zooming.
By using the numeric entry keys, almost any scale may be entered. In this
case, the top reference remains fixed while the bottom reference changes.
4-26
SCALE MENU
Auto Scale Pressing this key will automatically set the vertical scale and top and bottom
reference of the active trace to display the entire range of the trace.
Horizontal scaling is not affected. AutoScale only operates on the data which
is displayed on the graph. If the graph is expanded, data corresponding to
frequency or time bins which are not shown do not figure in the autoscaling
calculations.
This key is identical to the [AUTO SCALE] key.
Expand XThe Expand X key allows the active trace graph to be expanded and
translated in the X axis. This key activates knob control of translation (the left
icon) and expand (the right icon). Graph expansion is a convenient way of
examining closely spaced details of a spectrum without decreasing the span
and increasing the acquisition time.
Horizontal expansion displays 128, 64, 30, 15 or 8 bins across the graph.
Expansion is about the marker position unless the marker is too close the
edge of the span. In this case, expansion leaves one edge of the graph fixed.
Whenever a graph is expanded, the Expand at the bottom right of the graph
is on.
Horizontal translation is in increments of one bin.
No graphical expansion is allowed when the X-axis is logarithmic.
X-axisThis key selects the X-axis scaling for the active trace. Linear graphs are the
normal display for spectrum analyzers. The logarithmic graph is convenient
for certain types of filter or broadband noise measurements. The number of
frequency bins displayed is always 400 and they are linearly spaced. The log
axis merely displays these points differently. The first displayed point of a
baseband span (one that starts at DC) is the frequency of the first bin, not
DC or 0 Hz.
4-27
ANALYZE MENU
Analyze
The Analyze menu turns on real-time harmonic, sideband and band analysis
as well as Limit and Data tables for the active trace.
Note:
When real-time analysis functions or Limit tables are active, the trace update
rate may significantly slow down.
No harmonic, sideband or band analysis is available when measuring Time
Record or Octave Analysis, or when displaying Phase.
None This key turns off any real-time harmonic, sideband or band analysis.
Harmonic If no real-time analysis is on, then this key turns on harmonic analysis for the
active trace. The harmonic analysis submenu is displayed and the marker
frequency will be entered into the fundamental frequency entry field. If any
real-time analysis is already on, then this key simply displays the harmonic
analysis submenu without change. The harmonic submenu is described in
the following pages.
When harmonic analysis is on, the harmonic level (rms sum of the
magnitudes of the harmonic frequency bins) and Total Harmonic Distortion
or THD (the harmonic level divided by the fundamental magnitude) are
displayed in the upper left corner of the graph. Only those harmonics which
appear within the frequency span figure into the calculations of harmonic
power.
SidebandIf no real-time analysis is on, then this key turns on sideband analysis for the
active trace. The sideband analysis submenu is displayed and the marker
frequency will be entered into the carrier frequency entry field. If any real-
4-28
ANALYZE MENU
time analysis is already on, then this key simply displays the sideband
analysis submenu without change. The sideband submenu is described in
the following pages.
When sideband analysis is on, the sideband level (rms sum of the
magnitudes of the sideband frequency bins) and sideband level relative to
carrier (in dBc) are displayed in the upper left corner of the graph. Only those
sidebands which appear within the frequency span figure into the
calculations of sideband power.
BandIf no real-time analysis is on, then this key turns on band analysis for the
active trace. The band analysis submenu is displayed and the marker
frequency will be entered into the band center frequency entry field. The
band start will be adjusted consistent with the band width. If any real-time
analysis is already on, then this key simply displays the band analysis
submenu without change. The band submenu is described in the following
pages.
When band analysis is on, the band level (rms sum of the magnitudes of all
frequency bins within the defined band) is displayed in the upper left corner
of the graph. Only the portion of the band which is within the frequency span
contributes to the calculation of band level.
Data TableThis key turns on the data table for the active trace and displays the data
table submenu which is described in the following pages. The screen format
will switch to the dual trace mode and the inactive trace is replaced with the
data table window. The data table reports the Y values for user entered X
locations. For example, the entries could be a set of harmonic frequencies
which need to be monitored. To generate a report of the measurement, the
active trace's data table may be printed out using the Plot menu. Each trace
has its own data table though only the table associated with the active trace
is active and displayed at any time. To remove the data table display, change
the Format in the Display menu back to Single.
If no data table is entered (or the data table has been deleted), and harmonic
or sideband analysis is turned on, then this key not only activates the data
table display, but also enters the harmonic or sideband locations into the
table.
Limit TableThis key turns on the limit table for the active trace and displays the limit
table submenu which is described in the following pages. The screen format
will switch to the dual trace mode and the inactive trace is replaced with the
limit table window. The limit table lists the coordinates of the line segments
which define the trace limits. When trace data exceeds these limit segments,
then the Fail message appears and an audio alarm sounds. To generate a
listing of the active trace's limit table, use the Print Limits function in the Plot
menu. Each trace has its own limit table though only the table associated
with the active trace is active and displayed at any time. To remove the limit
table display, change the Format in the Display menu back to Single.
4-29
ANALYZE MENU
Harmonic
The Harmonic analysis submenu turns on real-time harmonic analysis and
sets the fundamental frequency and number of harmonics. Functions also
automatically move the span center frequency to adjacent harmonics.
FundamentalThis key activates the fundamental frequency entry field. If no real-time
analysis was on when the harmonic submenu was entered, then this field is
automatically filled with the marker frequency. The [Marker Entry] key copies
the marker frequency into this field even when it is not activated. When this
field is activated, knob adjustments and numeric entry are permitted. Note
that marker entries and knob adjustments are done with the resolution of the
current frequency span. If the actual fundamental is not exactly equal to a
frequency bin, then higher harmonic frequencies will be more and more
inaccurate. In this case, the frequency should be entered numerically with as
much precision as necessary.
The harmonic frequency bins on the graph are identified by a small triangle
marker located at the Y positions of each harmonic bin. This is helpful in
determining whether the fundamental frequency is accurate enough to
ensure that all harmonics are correctly identified.
# HarmonicsThis key activates the # of Harmonics entry field. Harmonics up to 400 may
be entered, though only those which are in the frequency span will enter into
the harmonic calculations and be identified on the graph. If 0 harmonics are
entered, the harmonic level readout will be zero.
Next Harmonic LeftThis function moves the marker to the next harmonic to the left of the current
marker position if it is on the graph. If it is beyond the edge of the graph, the
span center frequency is set to the frequency of the next harmonic (or as
close as the frequency range allows).
4-30
ANALYZE MENU
Next Harmonic Right This function moves the marker to the next harmonic to the right of the
current marker position if it is on the graph. If it is beyond the edge of the
graph, the span center frequency is set to the frequency of the next harmonic
(or as close as the frequency range allows).
ReturnThis key returns to the main Analyze menu.
4-31
ANALYZE MENU
Sideband
The Sideband analysis submenu turns on real-time sideband analysis and
sets the carrier and separation frequencies and number of number of
sidebands.
Sidebands are identified according to the formula
Sideband n = Carrier ± n•Separation
where n varies from 1 to the # of sidebands.
CarrierThis key activates the carrier frequency entry field. If no real-time analysis
was on when the sideband submenu was entered, then this field is
automatically filled with the marker frequency. The [Marker Entry] key copies
the marker frequency into this field even when it is not activated. When this
field is activated, knob adjustments and numeric entry are permitted. Note
that marker entries and knob adjustments are done with the resolution of the
current frequency span. If the actual carrier is not exactly equal to a
frequency bin, then higher order sideband frequencies will be more and more
inaccurate. In this case, enter the frequency numerically with as much
precision as necessary.
The sideband frequency bins on the graph are identified by small triangle
markers located at the Y positions of each sideband bin. This is helpful in
determining whether the carrier and separation frequencies are accurate
enough to ensure that all sidebands are correctly identified.
SeparationThis key activates the separation frequency entry field. When this field is
activated, the knob adjusts the separation with the resolution of the current
frequency span. This resolution may lead to the higher order sideband
frequencies being more and more inaccurate. In this case, enter the
frequency numerically with as much precision as necessary.
4-32
ANALYZE MENU
# Sidebands This key activates the # of Sidebands entry field. Up to 200 sidebands may
be entered, though only those which are in the frequency span will enter into
the sideband calculations and be identified on the graph. If 0 sidebands are
entered, the sideband level readout will be zero.
ReturnThis key returns to the main Analyze menu.
4-33
ANALYZE MENU
Band
The Band analysis submenu turns on real-time Band analysis and sets the
band location. A band is a range of frequencies defined by the band start,
center and width.
The frequency band is identified by a horizontal bar at the bottom of the
graph. The bar covers the region of the band.
Band StartThis key activates the band start entry field. If no real-time analysis was on
when the band submenu was entered, then this field is automatically filled
with the marker frequency minus half of the band width (the band center is
set to the marker frequency). When this field is activated, knob adjustments
and numeric entry are permitted. Note that knob adjustments are done with
the resolution of the current frequency span. For increased precision, enter
the frequency numerically.
Band CenterThis key activates the band center frequency entry field. If no real-time
analysis was on when the band submenu was entered, then this field is
automatically filled with the marker frequency. The [Marker Entry] key copies
the marker frequency into this field even when it is not activated. When this
field is activated, knob adjustments and numeric entry are permitted. Note
that marker entries and knob adjustments are done with the resolution of the
current frequency span. For increased precision, enter the frequency
numerically.
4-34
ANALYZE MENU
Band Width This key activates the bandwidth entry field. When this field is activated,
knob adjustments and numeric entry are permitted. Note that knob
adjustments are done with the resolution of the current frequency span. For
increased precision, enter the frequency numerically.
ReturnThis key returns to the main Analyze menu.
4-35
ANALYZE MENU
Data Table
The data table reports the Y values for user entered X locations. For
example, the entries could be a set of harmonic frequencies which need to
be monitored. To generate a report of the measurement, the active trace's
data table may be printed out using the Plot menu. Each trace has its own
data table though only the table associated with the active trace is active and
displayed at any time. To remove the data table display, change the Format
in the Display menu back to Single.
Note:
If no data table is entered (or the data table has been deleted), and harmonic
or sideband analysis is on, then entering this submenu not only activates the
data table display, it also enters the harmonic or sideband locations into the
table.
Data tables are saved along with the trace data when data is saved to disk.
Data Tables are not stored in non-volatile memory and are not retained when
the power is turned off. Save data to disk to preserve the data tables!
4-36
ANALYZE MENU
A sample data table display is shown above. The table entries are n (table
index or line number), X (user defined X values), and Y (measured data
values corresponding to the X values). The Y value is actually for the
frequency bin which is closest to the entered X value (within one bin
resolution). The Y values shown in the table are in the units of the display. X
values which are not in the frequency span have the message OVRG for
their Y values.
Table IndexThis key activates the Table Index entry field. Knob adjustment and numeric
entry are both permitted. When using the knob, scrolling past the last index
will add a new line. If an index greater than the last index is entered, then a
new line is added after the end of the table.
If the table is longer than what can be displayed in the window, then the table
index can be used to scroll the window. Entering an index will always display
that line in the window for viewing or editing.
X ValueThis key activates the X Value entry field. The X value of the highlighted line
may be entered using the numeric keypad. No knob adjustment is allowed.
When this field is active, the [MARKER ENTRY] key will copy the marker X
position into this field.
Insert ItemThis function inserts a new line before the highlighted line. The new line
becomes highlighted and is ready for editing.
Delete ItemThis function deletes the highlighted line and highlights the following line.
4-37
ANALYZE MENU
Delete Table This function deletes the entire table.
Return This key returns to the main Analyze menu.
4-38
ANALYZE MENU
Limit Table
The limit table lists the X,Y coordinates of the line segments which define the
trace test limits. When trace data exceeds these limit segments, then the test
fails. To generate a listing of the active trace's limit table use the Print Limits
function in the Plot menu. Each trace has its own limit table though only the
table associated with the active trace is active and displayed at any time. To
remove the limit table display, change the Format in the Display menu back
to Single.
A limit segment is defined as the line segment between the pair of points
(Xbegin,Y1) and (Xend,Y2) as shown below. The segment values between
the endpoints are calculated for the displayed frequency span.
A segment may be defined as either an upper or lower limit. Trace data
values which are greater than an upper or less than a lower limit cause the
test to fail.
Note:
Y values are entered without units. They are simply numbers. When the
display units are changed, the limit table is unaffected. and the limit tests will
compare the trace data in the new units with the old limit table values. Be
careful to use the limit tables only when the entered Y values match the
displayed units.
4-39
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.