SR510 / SR530 FAST TIME CONSTANT MODIFICATION
The post-demodulator time constants of the SR510 may be removed or shortened by simply
changing one capacitor on the main circuit board. Removing this capacitor will reduce the time
constant to a few microseconds. Note that the SR510 does NOT use a square wave demodulator.
This means that there is always a 2f component to the output signal, which is equal in magnitude to
the desired DC output. Removing or shorting the time constant will allow this 2f component to
reach the output. In many servo type applications, this is not a problem since the servo loop has a
finite response time. In addition, changing the time constant will cause the noise measurement to
be wrong since it relies on the time constant to determine the noise bandwidth.
To remove the second pole of the output filter, simply select NONE for the post time constant. To
change the PRE time constant, the value of C414 needs to be changed. The standard value is 0.033
microfarads. This capacitor determines the PRE time constant from 1 ms to 300ms. To change
these time constants, simply scale the capacitance by the desired amount. For example, to reduce
the time constants by 10 (100uS to 30mS), C414 should be changed to 0.0033 microfarads.
Removing C414 altogether leaves a time constant of a few microseconds. Note that this change
only affects the PRE time constant positions on the front panel from 1ms to 300ms. The time
constant positions from 1s to 100s are set by C415 which is normally a 10uF capacitor. This
capacitor can be scaled also. For very short time constants however, it is recommended that only
C414 be changed and only the ms time constant positions be used.
To change C414, remove the top and bottom panels of the SR510. C414 is mounted at board
location F3 (near the front center). Carefully desolder C414 and either leave it open (no time
constant) or replace with the desired value. Note that the replacement should be made with a mylar
capacitor. C415 is located at G4 (the very large capacitor). Change C415 only if it is necessary to
have a wide selection of altered time constants. For the SR530 the y time constants must also be
changed. They are located on the quadrature oscillator board.
X Channel Y Channel
C414 - C415 C1114 - C1115
C416 - C417 C1116 - C1117
Note: For a time constant less than a few hundred microseconds, the output capacitor must be
removed. On the SR530 these capacitors are located on the X and Y outputs after U1204 and
U1205. Their 0.luF value gives an output time constant of about 220us. On the SR510 this
capacitor is C607 on the output of U513.