Frequency Range
Max. Freq. Resolution
Sine 3.1 MHz 1 µHz
Square 3.1 MHz 1 µHz
Ramp 10 kHz 1 µHz
Triangle 10 kHz 1 µHz
Noise 3.5 MHz (Gaussian weighting)
Output
Source impedance 50 Ω
Grounding Output may float up to ±40 V
(AC + DC)
Amplitude
Range 50 mVpp to 10 Vpp (50 Ω),
100 mVpp to 20 Vpp (Hi-Z)
Resolution 3 digits (DC offset = 0V)
Offset ±5 VDC (50 Ω), ±10 VDC (Hi-Z)
Offset resolution 3 digits
Accuracy 0.1 dB (sine output)
Sine Wave
Spurious response < −65 dBc to 1 MHz
< −55 dBc to 3.1 MHz
Harmonic distortion
DC to 100 kHz < −60 dBc
100 kHz to 1 MHz < −50 dBc
1 MHz to 3 MHz < −40 dBc
Phase noise < −60 dBc (30 kHz band centered
on carrier)
Square Wave
Rise/fall time 15 ns ± 5 ns (10 % to 90 %)
Asymmetry <3 ns + 1 % of period
Overshoot <5 % (full-scale output)
Ramps and Triangles
Rise/fall time 100 ns
Linearity ±0.1 % of full scale
Settling time 200 ns (0.5 % of final value)
FSK Modulation
Modes Internal, External
Max rate 50 kHz, internal
External FSK TTL input, 1 MHz (max.)
DS335 Specifications
Stanford Research Systems phone: (408)744-9040
www.thinkSRS.com
Sweeps
Type Linear and logarithmic (phase
continuous)
Span Linear (full frequency range),
log (6 decades)
Sweep rate 0.01 Hz to 1 kHz
Timebase Accuracy
Standard ±5 ppm (20 °C to 30 °C)
Optional TCXO, 2 ppm stability,
2 ppm aging (20 °C to 50 °C)
General
Interfaces Optional RS-232 and GPIB. All
instrument functions are
controllable over the interfaces.
Non-volatile memory Up to nine sets of instrument
settings may be stored and recalled.
Dimensions 8.5" × 3.5" × 13" (WHD)
Weight 8 lbs.
Power 22 W, 100/120/220/240 VAC,
50/60 Hz
Warranty One year parts and labor on defects
in materials and workmanship
Ordering Information
DS335 3 MHz function generator $1095
Option 01 GPIB and RS-232 interfaces $495
Option 02 2 ppm TCXO timebase $350
O335RMD Double rack mount kit $85
O335RMS Single rack mount kit $85
DS335 rear panel (with opt. 01)