Clinitek, Clinitek Status, Multistix, Multistix PRO, Combistix, Uristix,
Labstix, Neostix, Clinitest and Chek-Stix are trademarks of
Siemens Healthcare Diagnostics.
Presept and Cidex are trademarks of Johnson & Johnson.
Theracide is a trademark of Lafayette Pharmaceuticals, Inc.
Amphyl is a trademark of Linden Corporation.
Kimwipes is a trademark of Kimberly-Clark.
US Pats 5,408,535; 5,477,326; 5,877,863; 6,239,445; D456,082; D489,816
Origin: UK
Siemens Healthcare Diagnostics Inc.
Tarrytown, NY 10591-5097 USA
Siemens Healthcare Diagnostics Ltd.
Sir William Siemens Sq.
Frimley, Camberley, GU16 8QD, UK
The information in this manual was correct at the time of printing.
How
ever, Siemens Healthcare Diagnostics continues to
and reserves the right to change specifications, equipment, and
maintenance procedures at any time without notice.
If this instrument is used in a manner differently than specified
in this manual, the protection provided by the equipment may
be impaired.
improve products
Table of Contents
A Brief Description ............................................................ v
Understanding the Symbols and Display Icons .............. vii
1 Unpacking & Set Up ................................................. 1-1
NOTE: Feature availability may
vary by geography due to regulatory
registration.
NOTE: Due to software changes,
some screens on the instrument
may appear slightly different from
those in this manual.
What does the analyzer do?
Your Clinitek Status
portable instrument for reading
Siemens Healthcare Diagnostics
urinalysis strips and Clinitest
immunoassay cassettes. No special
training is needed to use this
instrument. Several different Siemens
urinalysis strips (e.g., Multistix® 10 SG)
can be used with the analyzer as well
as the Clinitest hCG Pregnancy Test.
The analyzer can be set up to be
as simple or sophisticated as you
prefer. You may simply insert a
dipped urinalysis strip or a Clinitest
cassette into the analyzer and the
result will be reported (this is called
a Quick Test).
Or, you have the option to enter
an Operator Name, Patient Name
and Patient ID. This added
information will be reported along
with the test results (this is called a
Full Test).
The touch screen displays
instructions and prompts you
through operation of the analyzer. In
addition, you enter information
through the touch screen.
®
+
analyzer is a
®
Do I have to calibrate?
You do not have to do anything
to calibrate. The instrument
performs a system test each time it
is turned on. Then, each time a test
is run, the instrument automatically
calibrates. The white calibration bar
(on the test table) provides NIST
traceable calibration.
How does it work?
Testing starts (in the Quick Test
mode) when either the Strip Test
or Cassette Test is selected on the
main menu screen.
Touching the Strip Test key
prompts you to adjust the test table
to accept a urinalysis strip. Then
you touch the START key. You
have 8 seconds to dip the test strip,
blot the edge of the strip and place
it on the test strip table. The table is
partially pulled into the instrument
for calibration and then pulled
completely into the instrument to
read the test strip.
Touching the Cassette Test key
prompts you to adjust the test table
to accept a cassette. Then you
touch the START key. You have
8 seconds to draw the sample into
a pipette and dispense it into the
sample well of the cassette. The
analyzer automatically calibrates
and then pulls the table completely
into the instrument where the
cassette is read.
In the Full Test mode for either aStrip Test or Cassette Test, you
are prompted to enter an Operator
Name, Patient Name and/or Patient
ID prior to running a test.
v
A Brief Description
How do I get results?
Results will be displayed on the
touch screen and printed (if
desired). The results can also be
transferred to a computer using a
9-pin null modem serial cable and
the RS-232 serial port on the back
of the instrument.
The analyzer stores results from
950 patient tests. You are able to
recall past patient test results on
the analyzer using the RecallResults function.
How is the analyzer powered?
The analyzer can be plugged
into an electrical outlet for use on
the bench top, or it can be powered
by batteries and freely moved from
one testing site to another. The
batteries fit into an opening on the
bottom of the instrument.
the analyzer, use it for your daily
Siemens urinalysis and Clinitest
immunoassay testing and keep it in
good working condition.
As you read through the
Operator’s Manual, you will find
these symbols:
NOTES: contain useful tips
on using the analyzer.Notes
appear in italicized type.
CAUTION: should be
followed carefully to ensure your
analyzer operates correctly and
is not damaged. Cautions
appear in bold type.
In the Operator’s Manual, you will
notice some text is in bold/italic or
bold.
What about this Operator’s
Manual?
The Operator’s Manual contains
the directions you need to unpack
Bold text identifies a button
(touch sensitive area) on the
screen.
vi
Bold/italic text identifies screen names.
Understanding the Symbols and Display Icons
Instrument and Labeling Symbols
This section describes the symbols that appear on the exterior of the
Clinitek Status+ analyzer, the power supply provided with the instrument,
the carton in which the instrument was delivered and the supplies of
reagent strips and cassettes which you will use with the instrument.
Indicates that the input supply is direct current
Identifies that the instrument is type B equipment, which
provides a particular degree of protection against electric shock
Class 111
Identifies that the instrument is class 111 type equipment,
which is equipment for connection to SELV (Separated Extra
Low Voltage), i.e. a power supply
Double insulated product or transformer may also identify
class 2 equipment (power supply only)
Identifies that the instrument is listed by Underwriters
Laboratories as meeting U.S. and Canadian requirements for
safety
The CE mark identifies that the product complies with the
applicable directives of the European Union
Manufacturer
European authorized representative
Indicates a power on/off button
Caution, consult accompanying documents
In vitro diagnostic medical device
Consult instructions for use
vii
Understanding the Symbols and Display Icons
Indicates a serial port
This system contains certain toxic or hazardous substances or
elements. The environmental protection use period for this
system is 50 years. The system can be used safely during its
environmental protection use period. The system should be
recycled immediately after its environmental protection use
period has expired.
30
18
100
Temperature limitation (18° 30° C)
Contents sufficient for (n) tests (100)
Use by YYYY MM
Catalog number
Serial number
Batch code
Biohazard
Indicates that this equipment is classified as Waste Electrical
and Electronic Equipment under the European WEEE
Directive. It must be recycled or disposed of in accordance
with applicable local requirements.
Printed on recycled materials
viii
Indicates compliance with RESY packaging standards
Understanding the Symbols and Display Icons
Keep this way up
Fragile, handle with care
Keep dry
Keep away from sunlight and heat
VDE Testing and Certification Institute Germany
Manufacturer’s mark (FRIWO) and manufacturing location
(Hong Kong)
Manufacturer’s mark (FRIWO) and manufacturing location
(Geratebau, Germany)
Positive Temperature Coefficient (PTC) A thermistor device
used to protect the transformer from short circuits or overload.
This is an auto reset device.
Thermal cut out (TCO) This safety device disconnects the
supply voltage to the transformer at a specific temperature.
The operation temperature is stated below.
Ingress protection rating protected against the entry of solid
objects >1 mm but no protection from liquids.
This symbol indicates a risk of electric shock.
ix
Understanding the Symbols and Display Icons
Display Icons
There are seven icons which display in the top left of the display to show
the mode of the instrument. They also appear on the selection area for
each function. The icons are:
Instrument Set Up
This is displayed when the instrument is being set up to suit the
users’ requirements.
Strip Test (e.g., Multistix 10SG)
Shown when a test is being carried out using a reagent strip for
urinalysis, and when results are displayed following a strip test.
Cassette Test (e.g., Clinitest hCG)
Displayed when a cassette test is being carried out and when
results are shown after a cassette test.
Results recall
Used to show that results are being recalled from the
instrument’s memory.
Printer
This icon is displayed when results are being printed.
Data transfer to Personal Computer
Shows that data, including results, is being transferred to a PC.
Alert
Used when an error is being displayed.
There are two icons which may appear in the title bar.
Battery power
This has a maximum of four segments which show the level of
battery power. It will be shown in the top right corner of the title
bar when the instrument is battery powered.
Paper out
Appears in the top of the title bar when the printer paper/label
roll needs replacing.
x
Unpacking & Set Up
1
Unpacking
Unpacking
1 Carefully remove the contents
of the shipping carton. Check the
carton and instrument for visible
signs of damage; if seen,
immediately contact the carrier.
1
5
1
Clinitek Status®+ analyzer
Test table
2
Do not touch the white
calibration bar.
Test table insert
3
If you are using a reagent strip
that has 4 or fewer test pads,
e.g., Uristix® 4, you must use a short
test table insert. This has to be ordered
separately (for a list of suppliers see
Appendix A, Local Technical Support
Providers and Distributors).
2 Remove each of the wrappings
and check for the following items:
4
2
3
Power Supply
4
If the power cord is not the style
you need, contact your local
representative (for a list of contacts see
Appendix A, Local Technical Support
Providers and Distributors).
Paper Roll
5
Depending on the model you have
received, there may also be a Warranty
Registration Card, Unpacking/Setup Guide,
and/or Quick Reference Guide included.
1-1
Unpacking & Set Up
Set Up
1
Set Up
3
Analyzer Set Up
Place the instrument on a level
work surface where the
temperature and humidity are fairly
constant.
4
Plugging Analyzer In
Plug the appropriate end of the
power cord into the power inlet
socket located on the rear of the
Clinitek Status+ analyzer. Plug the
other end of the power cord into an
AC electrical wall outlet.
The best temperature for using
the instrument is between 22°C and
26°C (72°F and 79°F). Do not place
the analyzer outside or near
windows, ovens, hot plates, or
radiators.
Only use the power supply
adapter included with the unit.
5
Installing Batteries
(Optional)
Place the analyzer on its side and
remove the battery cover by
pressing down on the tab and
pulling out. Place the 6 new
alkaline AA-size batteries into the
analyzer. Replace the battery cover
and turn the instrument back onto
its base.
1-2
Unpacking & Set Up
1
Set Up
6
Inserting Test Strip Table
Insert the test strip table into the
analyzer by holding it by the end
opposite the white calibration bar
and with the white bar facing up.
Push the test table into the
analyzer, pushing it in just over
halfway.
Do not push the test table fully
into the analyzer as the test table
may become jammed and prevent
the use of the analyzer.
Do not touch the white
calibration bar.
7
Loading Test Table Insert
The test table insert adapts for use
with a Siemens Healthcare
Diagnostics urinalysis strip or
Clinitest immunoassay cassette.
One side is used for a strip test and
the other side is used for a cassette
test.
8
Interfacing to a Computer
The instrument can send results to
a computer via the serial port
located on the back of the analyzer.
This requires a 9-pin null modem
serial cable that can be purchased
separately at an electronics store or
from your Siemens Representative
(for a list of suppliers see Appendix
A, Local Technical Support
Providers and Distributors).
1-3
Unpacking & Set Up
1
Set Up
9
Interfacing to the
Clinitek Status Connector
The Clinitek Status connector
allows for Ethernet or wireless
network connectivity, Quality
Control, increased security, bar
code scanning, and additional
features with the Clinitek Status+
analyzer.
This connector provides standard
wired and wireless connectivity of
the Clinitek Status+ system to your
LAN, LIS, HIS, EMR, and allows
for centralized control of all
satellite Point of Care (POC)
Clinitek Status+ analyzers.
Refer to the Clinitek StatusConnect System Operator’s Guide.
Analyzer Software Upgrades
10
From time to time Siemens will add
new features and make improvements
to the Clinitek Status+ instrument
software.
These software updates will be
available on an electronic memory
card which is inserted into the
software update socket. This socket
is located under the printer cover
and is on the left-hand side of the
printer when you face the back of
the instrument.
Updating the software is a simple
procedure. Instructions for updating
the software on your instrument will
be supplied with the memory card.
1-4
Unpacking & Set Up
Set Up
1
11
Loading the Printer Paper or
Label Roll
1. Open the printer cover by pulling
up on the tab.
2. Open the paper roll compartment
cover by pressing down on its tab
and pulling out.
3. Lift the paper holding arm into
the open, upright position.
4. Place the new paper roll into the
printer paper compartment with the
paper unrolling from underneath
and toward the compartment wall.
5. Feed the paper up along wall
and through the printer. Once you
have approximately 4 inches (or
10 cm) of paper through the printer
then feed the edge of the paper
through the printer cover.
6. Push the paper holding arm
down in to the closed position.
paper holding arm
7. Close the printer and paper roll
covers by clicking them into
position.
The analyzer is set up to
automatically print the results (to
turn off the automatic print function
see Section 5, Instrument Set Up).
The analyzer uses ordinary
thermal paper as provided, or label
stock (for ordering information see
Appendix A, Local Technical
Support Providers and Distributors).
1-5
Unpacking & Set Up
Set Up
1
12
Warranty Registration
1. Lift the printer cover on the
instrument and the serial plate with
the instrument’s serial number will
be visible.
2. Write the serial number and
installation date on the Warranty
Registration Card. After the
instrument has been successfully
installed, complete the information
on the Warranty Registration Card
and return the card to your local
Siemens office (for a contact list
see Appendix A, Local Technical
Support Providers and
Distributors).
1-6
Unpacking & Set Up
1
Powering Up
Powering Up
Press the on/off button located
on the front of the instrument.
This is the first screen displayed.
The analyzer will run an automatic
system diagnostic test each time it
is turned on.
If this is the first time you have
turned on the analyzer, you will be
led through a Start-Up Wizard, a
quick set-up procedure. If you
require further instruction regarding
the Start-Up Wizard see Section 3,
page 3-1.
1-7
Unpacking & Set Up
Powering Down
1
Powering Down
1. Before turning the analyzer off,
always ensure that there is no strip
or cassette on the test table and
that the table and insert are clean.
2. Press the on/off button for
at least 2 seconds. The test table
will retract into the analyzer. If there
is no strip or cassette on the test
table, the door will close and the
analyzer will switch off.
If a strip or cassette is still on the
test table, the test table will be
pushed out and the analyzer will
turn off. The test table will remain
out. In order to retract the test table
into the analyzer, turn the analyzer
on, and then off (without a strip or
cassette on the test table).
Do not push the test table fully
into the analyzer as the test table
may become jammed and prevent
the use of the analyzer.
1-8
Interacting with the Touch Screen
2
Screens
Screens
The touch screen will guide you
through the operation of the
Clinitek Status®+ analyzer. The
screen will display messages,
instructions and options to which
you respond by touching the
appropriate area on the screen.
The first main screen you see is the
Select screen. It displays the time
and date, and indicates the
5 possible actions:
Instrument Set Up
Recall Results
QC Test
Cassette Test
Strip Test
Icon
Title Bar
Each screen that follows the Select
screen has an icon, title bar and
touch-sensitive active areas.
In some cases, the screen will also
display instructions, messages or
error messages.
The icon indicates the main section
in which you are working (1 of the
5 sections listed on the main
Select screen).
Touch Sensitive Area
2-1
Interacting with the Touch Screen
Screens
2
How to Touch the Screens
The screen needs to be touched
lightly in the touch-sensitive area to
activate a response.
Where to Touch the Screens
There are three types of areas that
respond to touching the screen.
• Round Buttons
• Boxed Areas
• Scroll Arrows
Round Buttons
These buttons typically appear on
screens that require a selection
among several items. The button
with a filled circle is the current
selection.
To change your selection, touch an
unfilled circle. The newly selected
circle (button) will now be
highlighted. You then touch the
Next button to move to the next
screen.
If a touched area does not
respond as expected, slide your
finger across the appropriate
selection area.
Use of anything hard or
pointed on the touch screen
may cause damage.
Round Buttons
In order to proceed, you will
always touch the box with a right
pointed arrow labeled Next.
In order to go back, you will
always touch the box with a left
pointed arrow. These “back” option
boxes vary in title.
2-2
Interacting with the Touch Screen
Screens
2
Boxed Areas
These are areas on the screen
enclosed in boxes. Simply touch
any area within the box to activate
that function.
The boxed areas vary in size. The
boxes located on the main Select
screen are examples of larger
areas. Smaller box selections
include boxes such as the
“Previous” and “Next” boxes found
at the bottom of the screen.
Boxed Area
Scroll Arrows
Press the up and down arrows on
the right side of the screen to scroll
through the list of information on
the left side of the screen. Once the
information on the left side of the
screen is highlighted, touch the
Select button to confirm your
selection and move to the next
screen.
Unavailable
Option
If there are double arrows on the
screen, these arrows (when
touched) will take you to the top or
bottom of the page.
When an option can be
selected it will be shown with a
thick black frame and will respond
when touched. If an option is not
available, it will be framed with a
thin black line and not respond
when touched.
2-3
Interacting with the Touch Screen
Keyboards
2
Keyboards
Press to switch to
Using the Alpha-Numeric
Keyboard
When the screen prompts you to
enter information for Operator,
Patient’s Name and/or Patient
Identification, a keyboard will
appear on the screen.
Depending on how your analyzer is
set up, either an alphabetic or
numeric keyboard will be displayed
first. The first keyboard displayed is
referred to as “keyboard priority.”
numeric keyboard
Data Entry Box
Back Space/
Erase key
If you require further instruction
regarding how to change keyboard
priority see Section 5, Instrument Set
Up.
To switch between the keyboards,
touch the 123 button to get to the
numeric keyboard. Touch the ABC
button to get to the alphabetic
keyboard.
To type in a name, number, birth
date, etc., touch the appropriate
button. Your selections will appear
in the data entry box.
If you switch between keyboards,
all values will be retained in the
data entry box on both keyboard
screens.
2-4
Return to
previous
screen
Press to switch to
alphabetic keyboard
Creates
blank
space
Confirm
data
entries
Interacting with the Touch Screen
2
Keyboards
The maximum number of
characters allowed is 32. An
audible tone will sound when you
have exceeded the maximum
number of characters.
Once you have finished entering
the information, touch Enter (from
either keyboard screen).
2-5
2-6
Start-Up Wizard
3
The first time your Clinitek Status®+
analyzer is turned on (following an
automatic system diagnostic test), it
will take you through a quick set up
procedure. This procedure will
allow you to select the basic
functions of the analyzer so you
can use the analyzer with your
choice of settings.
The Start-Up Wizard will allow you
to select the following settings:
Language
Date and time
Preferred test sequence
(e.g., Quick Test or Full
Test)
Type of urinalysis strip
Results format
If you require further instruction
regarding how to change the settings
see Section 5, Instrument Set Up.
Once you have selected your
choices, the analyzer will display a
Confirmation screen which allows
you to check that your preferences
are correct.
3-1
3-2
Testing
4
Quick Tests
Quick Tests
Urinalysis Strip Test
BIOHAZARD Wear personal protective
equipment. Use universal precautions.
Refer to Appendix H for recommended
precautions when working with
biohazardous materials.
Testing is started from the main
Select screen.
If you require more information
regarding use and storage of test
strips, please refer to the strip package
insert.
Touch Strip Test to conduct a
urinalysis strip test.
The next screen that appears is
Prepare Test.
If you would like the steps for
urinalysis testing to be shown on the
screen then touch Help.
Make sure the test table insert has
the reagent strip holder facing
upward.
Also, have the test strip, urine
sample and paper towel ready.
Touch the START button.
4-1
Testing
4
Quick Tests
The next screen that appears is
another Prepare Test. This screen
prompts you through the steps to
prepare the test strip.
A timer displays how much time
you have remaining to complete the
steps.
You have 8 seconds to complete
the following four steps:
1 Dip the reagent strip into the
urine sample, wetting all pads.
Immediately remove the strip from
the urine.
NOTE: Do not dip the automatic
identification band or color band
in the urine sample.
2 Drag the edge of the strip
against the side of the sample
container as you remove it.
4-2
Testing
Quick Tests
4
3 Blot by touching the edge of the
strip to the paper towel to remove
excess urine.
Do not lay the pads on the
paper towel or cover the pads by
the paper towel.
4 Place the reagent strip in the
channel of the table with the test
pads facing up. Slide strip to end of
the channel.
At the end of the 8 second
countdown, the test table and strip
will automatically be pulled into the
analyzer.
Do not push or pull the test
table.
The Clinitek Status®+ analyzer
will perform an automatic calibration
each time a test is run.
Be sure not to move or bump
the table while the instrument is
calibrating.
NOTE: A warning message
displays if you are not using a
Siemens reagent strip. Press OK to
continue. The Results may not
display if you are using a nonSiemens reagent strip. Repeat the
test using a Siemens reagent strip.
4-3
Testing
Quick Tests
4
The Analyzing screen will be
displayed when the calibration has
been completed and the analysis of
the strip has begun.
A timer will count down the time
remaining in analyzing the strip
results.
If the analyzer has been set up to
automatically print the results, then
the Printing screen will be
displayed until the print out has
been completed (otherwise the
Results screen will appear).
The date, time and test sequence
number will be printed along with
the test results. “Not Entered” will
be printed next to Color and Clarity.
If the results are positive, an
asterisk* will appear next to the
results (if “mark positive results” was
selected in Instrument Set Up).
4-4
Testing
4
Quick Tests
The next screen displayed is the
Results screen. The first page of
the test results are displayed on the
screen and the test table and strip
are automatically pushed out of the
analyzer.
To view the remaining test results,
touch More on the screen.
If you are using reagent strips with
a color strip or auto identification
band, you can view Sample
Interference notes about this test.
From the Results screen, touch
Notes.
If Sample Interference notes are
generated for this test, the
Interference notes screen displays.
Touch the up and down arrows to
scroll through the notes. Touch
Done to return to the main Results
screen.
4-5
Testing
4
Quick Tests
Up to 5 Sample Interference notes
display on the screen. Use the up
and down arrows to scroll through
the notes. If enabled, the notes
print with the test results.
If Sample Interference notes have
been disabled in the setup, the
NOTES button does not display.
NOTE: If you run a test with this
feature disabled, no notes will be
generated at the time of the actual
test. If you enable the sample
interface notes then recall the test
results, the analyzer generates
Sample Interference Notes for this
patient test.
If the analyzer has not been set up
to automatically print the test
results, touch Print to have the
results printed.
The results will automatically be
sent to the connected PC if this
option is set up in the analyzer.
If you require further instruction
regarding how to set up the analyzer
so the results are printed or sent to a
computer automatically see Section 5,
Instrument Set Up.
4-6
Testing
4
Quick Tests
From the test table, remove the
used urinalysis strip and dispose of
it according to your standard
laboratory procedures. Wipe the
table insert, if necessary.
Report the results to a laboratory
supervisor or physician.
Touch Done to complete the test
and return to main Select screen.
The results will be displayed on
screen for 2 minutes. After this time
the
elapses, the display will return to main
Select menu.
Touch Done to return the Strip Test
Prepare screen. You are ready to
start the next test. If testing is
complete, touch Back key to return
to the Select menu.
4-7
Testing
Quick Tests
4
Cassette Test
BIOHAZARD Wear personal protective
equipment. Use universal precautions.
Refer to Appendix H for recommended
precautions when working with
biohazardous materials.
Testing is started from the main
Select screen.
Please refer to the Clinitest
cassette test package insert for more
information regarding use and storage
of test cassettes.
Bring the test cassette and
patient sample to room
temperature 20
86°F) prior to testing.
Touch Cassette Test to conduct
the test.
The next screen that appears is
Test Type. Touch the Clinitest
hCG cassette button.
The next screen that appears is
Prepare Test screen.
If you would like the steps for
cassette testing to be shown on the
screen then touch Help.
°C to 30°C (68°F to
®
hCG
4-8
Testing
4
Quick Tests
Make sure the test table insert is in
position for a cassette test.
Remove the test cassette from the
foil package and place the cassette
on the test table.
Once you touch the START
button you have 8 seconds to
draw the urine sample into the
pipette and add the urine sample
into the well on the cassette.
Touch START button.
The next screen that appears is
another Prepare Test. This screen
prompts you through the steps to
prepare the cassette test.
A timer displays how much time
you have remaining to complete the
steps.
4-9
Testing
Quick Tests
4
You have 8 seconds to complete
the following two steps:
1 Draw the urine sample to the
line marked on the pipette
(approximately 0.2 mL).
2 Add entire contents of the
pipette into the sample well of the
test cassette.
At the end of the 8 second
countdown, the test table and
cassette will automatically be pulled
into the instrument.
Do not push or pull the test
table.
4-10
Testing
Quick Tests
4
The Clinitek Status+ analyzer will
perform an automatic calibration each
time a test is run.
Be sure not to move or bump
the
table while the instrument is
calibrating.
The Analyzing screen will be
displayed when the calibration has
been completed and the analysis of
the cassette has begun.
A timer will count down the time
remaining in analyzing the cassette
results.
The Clinitest hCG test results are
either negative, positive or borderline.
The analyzer takes approximately
5 minutes to confirm a negative
result. If the result is a clear positive,
the analyzer will report it sooner. If
the result is borderline, then you
should retest, with a new sample, in
48 to 72 hours. Please refer to the
Clinitest hCG cassette test package
insert for complete instructions for
use.
4-11
Testing
Quick Tests
4
If the analyzer has been set up to
automatically print the results, the
Printing screen will be displayed
until the print out has been
completed (otherwise the Results
screen will appear).
The date, time and test sequence
number will be printed along with
the test results.
The next screen displayed is the
Results screen. The test results
are displayed on the screen and
the test table and cassette are
pushed out of the analyzer.
The results will be printed automatically
if this option is set up in the
analyzer. If not, touch Print to print
the results on the analyzer’s printer.
If the result is positive, an
asterisk* will appear next to the result
(if “mark positive results” was
selected in Instrument Set Up).
The results will automatically be
sent to the connected PC if this
option is set up in the analyzer.
If you require further instruction
regarding how to set up the analyzer
so results are automatically printed or
sent to a computer see Section 5,
Instrument Set Up.
Remove the used cassette and
dispose of it according to your
standard laboratory procedures.
Report the results to a laboratory
supervisor or physician.
Touch Done to complete the test
and return to main Select screen.
4-12
The result will be displayed on
the screen for 2 minutes. After this
time elapses, the display will return to
main Select menu.
Testing
4
Full Tests
Full Tests
Urinalysis Strip Test
BIOHAZARD Wear personal protective
equipment. Use universal precautions.
Refer to Appendix H for recommended
precautions when working with
biohazardous materials.
A Full Strip Test allows you the
option to enter an Operator Name,
Patient Name and/or Patient ID
prior to inserting a strip.
The procedures to enter the
Operator and Patient data are
presented in this section.
The strip testing process is identical
to a Quick Strip Test.
If you require further instruction
regarding the procedures required for
running a Siemens Healthcare
Diagnostics urinalysis strip test see
Section 4, Quick Tests.
Testing is started from the main
Select screen.
Touch Strip Test to conduct a
Siemens urinalysis strip test.
4-13
Testing
4
Full Tests
The next screen that appears is
Operator Name.
There are two options under
Operator Name: Last Operator or
Enter New Operator Name.
Option 1: Last Operator
If this option is enabled, the last
operator that entered his/her Name
will be displayed on the screen in
the lower right side of the box. If
you are this operator, then touch
the Last Operator button to
proceed.
Option 2: Enter New Operator
Name
In order to enter the information for
a new operator, touch Enter NewOperator Name button. The next
screen that is displayed is EnterOperator Name.
Use the keyboards to enter
Operator Name using a maximum
of 13 characters. Touch Enter
when you have finished entering
the Name and to move to the next
screen.
If you require further instruction
regarding keyboard usage see
Section 2, Interacting with the Touch
Screen.
4-14
Testing
4
Full Tests
The next screen displayed is
Patient Information.
There are two options under Patient
Information: Recall Patient or
Enter New Patient.
Option 1: Recall Patient
In order to look up previous
patients, touch Recall Patient.
If previous patient identification has
been entered, a list of up to 950
patient results will appear on the
screen. Use the up and down arrow
buttons to scroll through the list of
patients. The most recently
performed test will be shown at the
top. Once the patient is highlighted,
touch Select button.
The next screen will be PrepareTest.
A total of 950 patient tests can
be stored in the analyzer. The tests
are listed in chronological order.
When the limit of 950 has been
reached, the oldest test will be
deleted from the analyzer. Deleted
information cannot be retrieved from
the analyzer.
4-15
Testing
Full Tests
4
Option 2: Enter New Patient
In order to enter the information for
a new patient, touch Enter NewPatient button. The next screen
displayed is Enter Patient Name.
Use the keyboards to enter Patient
Name using a maximum of 20
characters. Touch Enter when you
have finished entering the patient’s
name and to proceed to the next
screen.
If you require further instruction
regarding keyboard usage see
Section 2, Interacting with the Touch
Screen.
4-16
Testing
Full Tests
4
The next screen displayed is
Patient Identification. Use the
keyboards to enter Patient
Identification using a maximum of
13 characters. Touch Enter when
you have finished entering the
patient’s ID and to proceed to the
next screen.
If you require further instruction
regarding keyboard usage see
Section 2, Interacting with the Touch
Screen.
4-17
Testing
4
Full Tests
The next screen that appears is
Prepare Test.
If you require further instruction
regarding the procedures for running
a Siemens urinalysis strip test see
Section 4, Quick Tests.
While the strip is being analyzed, a
Select Appearance screen will be
displayed. The urine sample must
be visually observed and then the
appropriate color and clarity must
be selected.
If the urine sample is yellow and
clear, touch the Yellow and Clear
button.
If the urine sample is not yellow
and clear, touch the Other button
for more choices.
If you touched the Other button,
select the appropriate color by
touching the circle button that
corresponds to the correct
description.
NOTE: You can select only one
color for a urine sample.
4-18
Testing
4
Full Tests
Select the clarity by touching the
circle that corresponds to the
correct description. Then touch
Next.
There is a time indicator on the
Select Appearance screen that is
counting down the time remaining
in the analysis of the strip.
After color and clarity have been
entered the next screen displayed
will either be:
Analyzing – if the strip is still
being analyzed
Results – if analyzing the
strip has been completed
4-19
Testing
4
Full Tests
Entering the Strip Lot Number
and Expiration Date
To enter strip lot information for a
second strip test, perform the
following steps:
1. At the Select screen, touch StripTest.
The Strip screen displays.
2. To use the last strip number and
begin the test, touch Use LastLot.
To enter new strip data, touch
Enter new lot and expiration.
The Strip Lot screen displays.
3. Enter the strip lot number.
Use the alpha keyboard to enter
text.
To enter numeric text, touch 123.
4. Select Enter.
The Strip Expiration screen
displays.
5. Use the arrow keys to indicate
the strip expiration date.
6. Touch Enter.
The Prepare Test screen
displays.
7. Touch Start.
See above.
4-20
Testing
4
Full Tests
The test results displayed on the
screen and the printout will include
the following information:
Patient Name, ID or both
Type of strip used
Test date
Time
Operator
Test Number
Color
Clarity
Results
Sample Interference Notes
If the results are positive, an
asterisk* will appear next to the
results (if “mark positive results” was
selected in Instrument Set Up).
4-21
Testing
4
Full Tests
Cassette Test
BIOHAZARD Wear personal protective
equipment. Use universal precautions.
Refer to Appendix H for recommended
precautions when working with
biohazardous materials.
A Full Cassette Test allows you the
option to enter an Operator, Patient
Name and/or Patient ID prior to
inserting a cassette.
The procedures to enter the
Operator and Patient data are
presented in this section.
The cassette testing process is
identical to a Quick Cassette Test.
If you require further instruction
regarding the procedures for running
a Siemens Clinitest immunoassay test
see Section 4, Quick Tests.
Testing is started from the main
Select screen.
Touch Cassette Test to conduct an
hCG pregnancy test.
The next screen that appears is
Operator Name.
There are two options under
Operator Name: Last Operator or
Enter New Operator Name.
4-22
Testing
4
Full Tests
Option 1: Last Operator
If this option is enabled, the last
operator that entered his/her Name
will be displayed on the screen in
the lower right side of the box. If
you are this operator, then touch the
Last Operator button to proceed.
Option 2: Enter New Operator
Name
In order to enter the information for
a new operator, touch Enter NewOperator Name button. The next
screen that is displayed is EnterOperator Name.
Use the keyboards to enter
Operator Name using a maximum
of 13 characters. Touch Enter
when you have finished entering
the name and to move to the next
screen.
If you require further instruction
regarding keyboard usage see
Section 2, Interacting with the Touch
Screen.
The next screen displayed is
Patient Information.
There are two options under
Patient Information: Recall Patient
or Enter New Patient.
4-23
Testing
Full Tests
4
Option 1: Recall Patient
In order to look up previous
patients, touch Recall Patient.
If the previous patient identification
has been entered, a list of up to
950 patients will appear on the
screen. Use the up and down arrow
buttons to scroll through the list of
patients. The most recently
performed test will be shown at the
top. Once the patient is highlighted,
touch the Select button. The next
screen will be Test Type.
A total of 950 patient tests can
be stored in the analyzer. The tests
are listed in chronological order.
When the limit of 950 has been
reached, the oldest test will be
deleted from the analyzer. Deleted
information cannot be retrieved from
the analyzer.
4-24
Testing
Full Tests
4
Option 2: Enter New Patient
In order to enter the information for
a new patient, touch the Enter NewPatient button. The next screen
displayed is Enter Patient Name.
Use the keyboards to enter Patient
Name using a maximum of 20
characters. Touch Enter when you
have finished entering the patient’s
name and to move to the next
screen.
If you require further instruction
regarding keyboard usage see
Section 2, Interacting with the Touch
Screen.
4-25
Testing
4
Full Tests
The next screen displayed is Enter
Patient ID. Use the keyboards to
enter Patient Identification using a
maximum of 13 characters. Touch
Enter when you have finished
entering the patient’s ID and are
ready to proceed to the next
screen.
If you require further instruction
regarding keyboard usage see
Section 2, Interacting with the Touch
Screen.
4-26
Testing
4
Full Tests
The next screen that appears is
Test Type.
If you require further instruction
regarding the procedures for running a
Clinitest hCG cassette test see
Section 4, Quick Tests.
The test results displayed on the
screen and the printout will include
the following information:
Patient Name, ID or both
Type of Clinitest cassette test
Test date
Time
Operator
Test Number
Result
If the result is positive, an
asterisk* will appear next to the result
(if “mark positive results” was selected
in Instrument Set Up).
4-27
4-28
Instrument Set Up
5
Instrument Set Up
Instrument Set Up
Your Clinitek Status®+ analyzer
allows you to change settings to
suit your workplace requirements.
If no customizing is desired, the
analyzer will automatically be
programmed with default settings (to
view default settings please see
Appendix D: Instrument Default
Settings).
Touch Instrument Set Up in order
to initiate changing the settings.
If a password has been set, the
Enter Password screen will be
displayed. Enter the password into
the data entry box. The characters
you enter will be displayed as
asterisks.
Touch Enter to proceed to a list of
settings.
Settings
Use the up and down arrows to
scroll through the ChooseSettings screen.
Touch Select when you have
highlighted the setting you want to
change.
Touch Done to return to the main
Select screen.
5-1
Instrument Set Up
Language Settings
5
Language Settings
Use the up and down arrows to
scroll through the list of languages.
Touch Select when you have
highlighted the desired language.
Before changing the language an
alert screen will be displayed
requesting you to confirm your
decision.
Touch No to continue with same
language and retain current
settings. The display will return to
Choose Settings.
Touch Yes to change the language.
Changing the language will…
delete all results from the
memory.
change the instrument
settings to the defaults for
the new language (default
settings are listed in
Appendix D: Instrument
Default Settings).
5-2
Instrument Set Up
5
Password
Password
When a password has been set,
no changes can be made to the
Instrument Set Up until the
password has been entered.
Set Password
Enter a password into the data
entry box using the alphabetic and/
or numeric keyboards (maximum
12 characters).
Keep a record of the password to
be sure to have access to InstrumentSet Up when required.
Touch Enter to confirm the
password and return to ChooseSettings.
Touch Done to return to main
Select screen.
Remove Password
Once a password has been entered
the option will change from SetPassword to Remove Password.
Highlight the Remove Password
option and touch Select. The
option displayed will change to SetPassword and a password will no
longer be needed to access
Instrument Set Up.
Touch Done to return to the Select
Menu.
5-3
Instrument Set Up
5
Operator and Patient Information
Operator and Patient
Information
Selecting this option will allow you
to choose display options for data
entry.
The Input Settings screen allows
you to select 1 of 3 test sequences.
Quick Test
A test without any operator or
patient information required. The
test will be identified by a
sequential test number when the
results are displayed or printed.
Full Test
The operator will be prompted to
enter the following data during
testing:
Operator Name
Patient’s Name
Patient Identification
Sample Appearance (Color
and Clarity) of urine
sample, when testing with
a urinalysis strip.
Custom Set Up
This allows you to customize the
data to be entered when
conducting a test.
5-4
Instrument Set Up
5
Operator and Patient Information
On the Input Settings screen,
touch a round button in the box of
the preferred test sequence. Then
touch Next for test sequence
confirmation.
Touch Done on the Input Settings
screen to return to the ChooseSettings screen.
5-5
Instrument Set Up
5
Operator and Patient Information
Custom Settings
Touch the round button in the
Custom Set Up box to customize
selections for data entry. Touch
Next.
The next 5 screens present choices
for customizing your testing.
Choose settings by touching and
highlighting a round button for each
category.
Proceed through all 5 screens by
touching Next at the bottom of
each screen.
Screen 1 of 5
Operator Name – allows you to
choose whether Operator Name
will be required for running a test –
Enabled (on) or Disabled (off).
Keypad priority – allows you toselect either the Numeric or
Alphabetic keyboard to be
presented as the first keyboard
during data entry.
5-6
Instrument Set Up
5
Operator and Patient Information
Screen 2 of 5
Patient Name – allows you to
choose whether patient name will
be required for running a test –
Enabled (on) or Disabled (off).
Patient ID – allows you to choose
whether Patient ID will be required
for running a test – Enabled (on) or
Disabled (off).
A total of 950 patient records
can be stored in the analyzer. When
the limit of 950 has been reached,
the oldest test will be deleted from
the analyzer. Deleted information
cannot be retrieved from the
analyzer. If Patient’s Name and
Patient ID are both disabled, a
sequential test number will be shown
with the test results.
Screen 3 of 5
Choose which to show in
Results list – allows you to choosewhether Patient Name or Patient
ID will be included in the list of
results.
Last Operator’s Name displayed
– allows you to choose whether a
“Select Last Operator” option is
available to the user when you are
prompted to enter an Operator
Name during Strip or Cassette
Testing – Enabled (on) orDisabled (off).
5-7
Instrument Set Up
Operator and Patient Information
5
Screen 4 of 5
Choose which to record during a
strip test – allows you to select
which sample details you would like
to be recorded during a strip test –
Color and Clarity, Color only,
Clarity only or None (no sample
information).
Screen 5 of 5
Custom Field – allows you to
name the custom field in order to
customize data entry.
To name the custom field, touch the
Enter Custom Field box.
The next screen displayed is EnterCustom Field. Use the keyboards
to enter a custom field
(e.g., Physician Name). Touch
Enter to return to the Custom
Settings screen.
If you require further instruction
regarding keyboard usage see
Section 2, Interacting with the Touch
Screen.
5-8
Instrument Set Up
5
Operator and Patient Information
Touch the round button next to
Enabled (on) to highlight and
activate this custom setting.
Touch the round button next to
Disabled (off) to deactivate the
custom setting.
Touch Previous to scroll back
through custom setting screens to
review.
Touch Next to proceed to the InputSettings screen which will provide
a list of the selections made in
Custom Set Up.
Touch Done to confirm and return
to the Choose Settings screen.
5-9
Instrument Set Up
Operator and Patient Information
5
Managing Urine Colors
The following sections describe how
to customize and set urine color
choices and urine clarity for Siemens
strip tests. When you print patient
test results, you can include urine
color, clarity, or color and clarity in
the printout. Urine color and clarity
are optional; you can choose not to
print these parameters. Urine color
and clarity are available only in Full
Test or Custom mode.
Setting and Customizing Urine
Colors
You can choose from one of 10
instrument-provided colors and add
up to 4 customized colors to patient
test results.
To include instrument-provided
colors, perform the following steps:
1. At the Select screen, touchInstrument Set Up.
The Choose Settings screen
displays.
2. Use the arrow keys to select
Operator and Patient
information.
3. Touch Select.
The Input Settings screen
displays.
4. Touch Custom Set Up.
5. Touch Next.
The Custom Settings-Operator
screen 1 of 5 displays.
6. Touch Next 3 times.
The Custom Settings-SampleAppearance screen 4 of
5 displays.
5-10
Instrument Set Up
Operator and Patient Information
5
7. Touch Edit colors.
The Sample Appearance-Selectcolors screen 1 of 3 displays.
8. To choose colors, touch the
button for the color you want.
To remove a selected color,
touch that color button again.
9. Touch Next.
The Sample Appearance-Selectcolors screen 2 of 3 displays.
10. To choose colors, touch the
button for the color you want.
11.Touch Next.
The Sample Appearance-Selectcolors screen 3 of 3 displays.
12. Touch Next 3 times.
The Input Settings-Confirmation
screen displays.
13. Touch Done twice to return to
the Select screen.
Adding Customized Colors
To enter up to 4 custom colors,
perform the following steps:
1. At the Sample Appearance-Select colors screen 3 of 3,
touch Enter custom color 1
(2, 3, or 4) corresponding to
each custom color.
2. Enter the custom color.
Use the alpha keyboard to
enter text.
To enter numeric text, touch
123.
NOTE: The maximum number of
characters for each color is 10.
5-11
Instrument Set Up
Operator and Patient Information
5
3. Touch Enter.
The Sample Appearance-Selectcolors screen 3 of 3 displays.
CAUTION
Do not edit a custom color that
already exists because doing so
deletes all patient records stored on
the system.
If a custom color exists, the
Sample Appearance screen
displays.
Touch Yes, to edit that custom
color and delete all records.
Touch No, to return to the
Sample Appearance Select
Colors screen 3 of 3.
4. Touch Next 3 times.
The Input Settings-Confirmation
screen displays.
5. Touch Done twice to return to
the Select screen.
Managing Strip Lot Number and
Expiration Date
You can enter the strip lot number
and expiration date and associate
this information with each patient
record. Once entered, the
information is retained for the next
test, or you can enter a new lot
number and expiration date.
You can set the instrument to
prompt for new strip information or
use the information from the last
strip before each patient test.
5-12
Instrument Set Up
Intrument Settings
5
Setting Strip Information Prompt
To set the prompt for strip
information, perform the following
steps:
1. At the Select screen, touchInstrument Set Up.
The Choose Settings screen
displays.
2. Use the arrow keys to select
Instrument Settings.
3. Touch Enter.
The Instrument Settings screen
displays.
4. Use the arrow keys to select
Urinalysis Test Settings.
5. Touch Select.
The Urinalysis Test Settings
screen displays.
6. Touch Next.
The Urinalysis Test screen
displays.
7. To prompt for strip information
before each test, touch
Enabled.
To bypass a prompt to enter
strip information before each
test, touch Disabled.
8. Touch Done 3 times to return to
the Select screen.
5-13
Instrument Set Up
Date and Time Settings
5
Date and Time Settings
The date and time are displayed on
the Select screen and are recorded
with test results.
The format of the date and time are
displayed along with specific date
and time values.
If the date and time values are
incorrect, use the up and down
arrow buttons to adjust the date
and time to the correct values.
Touch AM or PM if you are in the
12-hour time format. Touch Set to
confirm your choices and move to
the next screen.
If the format is not correct for your
workplace, for example, you prefer
DD-MM-YY to MM-DD-YY or
24-hour clock to 12-hour clock,
touch Choose Format.
The next screen displayed is
Choose Format. Touch the round
button that corresponds to your
choice of a date format and a time
format. Touch Done to return to
Set Date & Time.
If the date and time are in the
format which you need, select
Done.
5-14
Instrument Set Up
5
Date and Time Settings
The AM and PM selections will
not be available if the time is in the
24 hour format.
The Set Date & Time display will
show the date and time in the
formats selected. Select Set to
confirm your choices and return to
the Choose Settings screen.
5-15
Instrument Set Up
Test Sequence Number
5
Test Sequence Number
The test sequence numbers run
from 0001 to 9999 and can be
reset. The screen shows the next
test number.
Touch the Reset to 0001 round
button if you would like to reset the
numbers to start at 0001 for the
next test.
Touch Done to confirm the reset or
to leave the screen if you have not
selected the option to reset the
number. You will return to the
Choose Settings screen.
5-16
Instrument Set Up
Instrument Settings
5
Instrument Settings
This section has a series of
screens which control the way in
which the Clinitek Status+ analyzer
displays information and operates.
There are 8 Instrument Setting items.
Results Format
System Settings
Display Contrast
Connectivity
Select Urinalysis Test
Authorized Operator
Printer Settings
QC Settings
Use the up and down arrows to highlight
a setting item and touch Select.
Results Format
This allows you to select the format
in which you would like the results
to be displayed and printed.
Choose Format
Screen 1 of 2
Units Selection – allows you to
select the format in which you
would like results by touching the
round button next to the
appropriate type of format.
NOTE: Nordic units are only
available in English and Swedish. If
you set the language to Chinese,
this screen does not display, as only
SI units are available in Chinese.
5-17
Instrument Set Up
Instrument Settings
5
Plus System – select Enabled
(on) if you wish to have results
shown in the Plus System. You will
also see in Appendix B that results
can be recorded in the Plus System
(which uses “+” symbols) instead
of clinical units such as mg/dL
(mmol/L). In some languages there
is no difference between the normal
system and the Plus System.
Screen 2 of 2
Mark Positive Results – allows
you to choose whether positive
results should be marked with an
asterisk (see Appendix B for results
to be shown as positive).
Touch the round button next to Yes
to mark and record positive results.
These will be marked on the
display, on a printout and when the
data is transferred to a host
computer.
Touch No and positive results will
not be marked.
Touch Done to enter your choice
and return to Instrument Settings.
5-18
Instrument Set Up
5
Instrument Settings
System Settings
This allows selections regarding the
printer, power and sound.
System Settings
Screen 1 of 2
Printer – allows you to select the
preferred printing option by
touching the round button.
Automatic – Results for
each test will be printed
automatically when the test
is complete.
Manual – Test results will
not be printed automatically.
They can be printed by
touching the Print selection
area on the Results
screen.
Off – Test results will not
be printed.
Power Save – allows you to enable
and disable this feature. If the
instrument is powered from an
electrical outlet and is not used for
5 minutes, you may choose to have
the test table retract and the
instrument power down. When it is
powered by batteries, Power Save
is always enabled and it will power
down after 5 minutes of non-use.
Touch the Enabled (on) to activate
or Disabled (off) if the Power Save
mode is not desired.
5-19
Instrument Set Up
5
Instrument Settings
Screen 2 of 2
Sound – allows you to adjust the
sound by highlighting and selecting
a level.
Sound on – The instrument
will use a range of audible
tones.
Sound off – No audible
tones will be made by the
instrument.
Key clicks only – Audible
clicks will be heard when
the user touches an active
button or key.
Display Contrast
This allows the contrast of the
display to be increased or
decreased to suit the work area in
which it is being operated. Use the
up and down arrows to sample the
contrast settings and touch Select
to set and return to the InstrumentSettings screen.
5-20
Instrument Set Up
5
Instrument Settings
Connectivity
You can connect the instrument to
a PC or host computer, or, if you
have a Clinitek Status connector, to
an LIS using an Ethernet or
wireless network. If you are using
the connector, refer to the
Clinitek Status Connect System
Operator's Guide for more
information.
Allow results to be sent to PC
If enabled, the system
automatically sends both new and
recalled patient tests to a host or
LIS.
To automatically send new and
recalled patient results to the LIS or
a PC directly connected to the
system, touch Enabled.To prevent
sending new and recalled patient
results to the LIS or PC, touch
Disabled.
NOTE: If there is no connector,
Siemens recommends setting the
connector to Disabled setting.
Setting the connector to the
Enabled setting without a connector
may prevent communication with
an external system.
5-21
Instrument Set Up
Instrument Settings
5
Store instrument serial number
in patient records
Touch round button of either Yes or
No depending on whether you
choose to include the serial number
of your Clinitek Status+ analyzer as
part of the patient results sent to
the computer.
If you have enabled the upload
capability, you will need to ensure
that the connectivity settings are
correct for the data transfer.
Touch Edit Connectivity Settings
and a series of 3 screens for
Connectivity Settings will allow
you to edit the settings.
Connectivity Settings
Screen 1 of 3
Baud rate – Touch a round button
to select the correct Baud rate and
Next to move to the next screen.
5-22
Instrument Set Up
Instrument Settings
5
Screen 2 of 3
Parity rate – Touch a round button
to select the correct Parity rate and
Next to move to the next screen.
Screen 3 of 3
Stop Bits – Touch a round button
to select the number of Stop Bits.
Select Urinalysis Test
This displays a list of some of the
Siemens Healthcare Diagnostics
urinalysis strips which can be used
with the Clinitek Status+ analyzer.
5-23
Instrument Set Up
Instrument Settings
5
Use the up and down arrows to
scroll through the list until the type
of strip you want to use is
highlighted. Touch Select to
confirm and return to the
Instrument Settings screen.
Do not use any reagent strip
product other than what is shown
on the display. Using the wrong
Reagent Strip will give you
incorrect results.
Some strips do not appear on
list (e.g., Clinitek
The analyzer will automatically identify
them through the color ID band on the
strip.
You do not need to select a type
of Clinitest
within Instrument Set Up.
®
Microalbumin 2).
®
immunoassay cassette
the
Touch Done repeatedly to return to
the main Select screen.
5-24
Instrument Set Up
Instrument Settings
5
Edit Reported Chemistries
This section describes how to
include or exclude urinalysis tests
from the reported results. Tests
include GLU, BIL, KET, SG, BLO,
pH, PRO, URO, NIT, LEU, ALB,
and CRE. The default is all tests
are reported and the button is
filled.
To set up tests for urinalysis,
perform the following steps:
1. At the Select Ready screen,
touch Instrument Set Up.
The Choose Settings screen
displays.
2. Use the arrow keys to select
Instrument Settings.
3. Touch Select.
The Instrument Settings screen
displays.
4. Use the arrow keys to select
Urinalysis Test Settings.
5. Touch Select.
The Urinalysis Test screen
displays.
6. Touch Edit reportedchemistries.
The Reported Chemistries
screen 1 of 2 displays.
7. To remove a selected test from
reported results, touch the
button for that test.
To include a test in reported
results, touch that test’s button
again.
8. Touch Next to advance to the
next screen.
The Reported Chemistries
screen 2 of 2 displays.
9. Touch the button for the test
you want to remove.
Touch the button again to
include that test.
10. Touch Done.
11.Touch Next.
12. Touch Done 3 times to return
to the Select Ready screen.
The reported results selection
applies to all strip types.
5-25
Instrument Set Up
5
Instrument Settings
Authorized Operator
This section describes how to set up
use of operator IDs and add, edit, or
delete the list of operator IDs. When
enabled, the system permits only
allowed operators to perform patient
tests, QC tests (when using the
connector), recall results, or modify
system settings. Operators gain
access by entering their ID.
The Clinitek Status+ analyzer
stores 700 operators.
NOTE: The Operator ID is never
printed or displayed with patient
results. If you wish to associate the
Operator’s Name with patient
results, enable Operator Name in
Custom Settings-Operator screen
1 of 5.
CAUTION
Once the Operator ID and Operator
Name settings are made, do not
change the Operator ID setting.
If you change the Operator
ID setting, all patient results are
erased.
2. Use the arrow keys to select
Instrument Settings.
3. Touch Select.
The Instrument Settings screen
displays.
Setting Operator IDs
To set up operator IDs, perform the
following steps:
1. At the Select Ready screen,
touch Instrument Set Up.
The Choose Settings screen
displays.
5-26
Instrument Set Up
5
Instrument Settings
4. Use the arrow keys to select
Authorized Operator.
5. Touch Select.
The Authorized operator screen
displays.
6. To permit access only by
authorized operators, touch
Enabled.
To allow all operators access to
the system, touch Disabled.
7. If you selected Enabled, seeAdding Operator IDs below to
add at least one operator.
If you selected Disabled, touch
Done 3 times to return to the
Select Ready screen.
CAUTION
If the instrument uses the operator
list sent by the LIS, do not
power down the system. If the
connector loses power, the
operator names are erased.
NOTE: The operator list sent by the
LIS overwrites an operator list
entered via the analyzer.
3. Touch Enter.
The Authorized Operator screen
displays indicating the Operator
ID and which functions the
operator can perform.
4. To edit this Operator ID, touch
Edit.
Adding Operator IDs
To add operator IDs, perform the
following steps:
1. At the Authorized operator
screen, touch Add operator.
2. Enter the new Operator ID.
Use the alpha keyboard to enter
text.
To enter numeric text, touch 123.
5-27
Instrument Set Up
Instrument Settings
5
5. To edit which functions this
Operator ID can access, touch
Edit.
The Authorized Operator-Operator access screen 1 of 2
displays.
6. To allow this operator to run
patient tests, touch Enabled.
To prevent patient tests, touch
Disabled.
7. To allow this operator to run
QC tests, touch Enabled.
To prevent QC tests, touch
Disabled.
8. Touch Next.
The Authorized Operator-Operator access screen 2 of
2 displays.
9. To allow this operator to recall
results, touch elect Enabled.
To prevent recall results, touch
Disabled.
10. To allow this operator to set up
the instrument, touch Enabled.
To prevent instrument setup,
touch Disabled.
11.Touch Done twice.
The Authorized Operator-Operators list screen displays.
12. Touch Exit.
13. Touch Done 3 times to return
to the Select Ready screen.
Viewing, Editing, Printing, and
Deleting Operator IDs
You can view, print, or delete the
entire operator list or edit individual
operators.
NOTE: If you delete the entire
operator list, ensure that
authorized operators is Disabled.
See Setting Operator IDs above.
5-28
Instrument Set Up
5
Instrument Settings
At the Authorized operator screen,
perform the following steps:
1. To delete the entire operators
list, touch Delete operators list.
The Delete operators list caution
screen displays.
2. To delete, touch Yes.
To keep the operators list,
touch No.
If you selected No, the
Authorized operator screen
displays.
If you selected Yes, go to Step 8.
3. To edit or view the operators list,
touch View operators list.
The Authorized Operator-Operators list screen displays.
4. Use the arrow keys to select the
operator you want to delete or
edit.
5. To delete that operator, touch
Delete entry.
To edit or delete that operator,
touch Select.
The Authorized operator screen
displays.
Refer to Adding Operator IDs
above, Step 6.
6. To print all operators, touch
Print.
NOTE: Enabling the instrument
password restricts access to
Instrument Setup to those who
know the password. If both
Operator ID and password are
enabled, the Operator ID has
priority.
NOTE: The system prints the first
100 operators listed alphabetically.
7. To return to the Authorizedoperator screen, touch Exit.
8. Touch Done 3 times to return to
the Select Ready screen.
5-29
Instrument Set Up
5
Instrument Settings
Printer Settings
This section describes how to
customize the printed test results.
Customizing the Printout
You can customize the test results
printout by including or excluding:
Operator name
Patient name
Patient ID
Instrument serial number
Urine color
Urine clarity
Up to 2 header lines of
customized alphanumeric text
To customize the printout, perform
the following steps:
1. At the Select Ready screen,
touch Instrument Set Up.
The Choose Settings screen
displays.
2. Use the arrow keys to select
Instrument Settings.
3. Touch Select.
The Instrument Settings screen
displays.
4. Use the arrow keys to select
Printer Settings.
5. Touch Select.
The Printer Settings-Includedin print-out screen 1 of 4
displays.
6. To select options, for example
Operator Name, Serial
Number, Patient Name, or
Patient ID to include in the
printout, touch the option
button.
5-30
Instrument Set Up
5
7. Touch Next.
8. To select options, for example,
9. Select Next.
10. To include a custom header in
11. To edit or create line 1 of a
12. Enter custom header text.
13. Touch Enter.
14. To edit or create line 2 of a
Instrument Settings
To remove a selected option,
touch that option button again.
The Printer Settings-Included inprint-out screen 2 of 4 displays.
Color, Clarity, or Custom
Information to include in the
printout, touch the option button.
To remove a selected option,
touch that option button again.
The Printer Settings-Set UpCustom Header screen 3 of 4
displays.
the printout, touch Enabled.
To exclude a custom header,
touch Disabled.
custom header, touch EnterLine 1.
The Custom Header screen
displays.
Use the alphabetic keyboard to
enter text.
To enter numeric text, touch 123.
The Printer Settings-Set UpCustom Header screen 3 of 4
displays.
custom header, touch Enter
Line 2.
NOTE: Each custom header line
accepts up to 24 alphanumeric
characters.
5-31
Instrument Set Up
5
Instrument Settings
15. Touch Next.
The Printer Settings screen 4
of 4 displays.
16. To print to the internal printer,
touch Internal printer.
To print to an external printer,
touch External printer.
17. If you selected Internal printer,
to print sample interference
notes, touch Enabled.
To disable printing sample
interference notes, touch
Disabled.
NOTE: If you select External
printer, sample interference notes
are automatically sent to the
printer.
18. Touch Done 3 times to return
to the Select Ready screen.
NOTE: To use an external printer,
you must connect and enable the
Clinitek Status connector.
Quality Control
For QC instructions, refer to the
Clinitek Status Connect System
Operator’s Guide.
5-32
Instrument Set Up
5
Restore Default Settings
Restore Default Settings
This option lists the analyzer’s
original settings.
Use the up and down arrows to
view the default settings.
Touch Restore in order to return
the analyzer to the settings listed.
Touch Yes to confirm your decision
or No to maintain the current
settings of your analyzer. You will
then return to the Restore screen.
Select Exit to return to ChooseSettings.
When the original settings are
restored, all results and patient
data will be deleted from the
memory.
5-33
Instrument Set Up
Diagnostics
5
Diagnostics
This option lists 7 diagnostic tests
which can be conducted.
only be conducted when instructed
by your local representative. The
representative will lead you through
the test procedures (see Appendix
A for Local Technical Support
Providers and Distributors).
5-34
Instrument Set Up
5
Sample Notes
Sample Interference Notes
To include Sample Interference
Notes, perform the following steps:
1. At the Select screen, touchInstrument Set Up.
The Choose Settings screen displays.
2. Use the arrow keys to select
Sample Notes.
3. Touch Select.
The Notes Settings screen displays.
4. To enable Sample Interference
Notes, touch Enabled.
To disable Sample Interference
Notes, touch Disabled.
5. Touch Done twice to return to
the Select screen.
5-35
Instrument Set Up
5
System Information
This screen has information about
your analyzer.
Touch System Configuration to
view details of the current system
configuration. This screen will show
the current settings for all the items
which can be changed within the
Input Settings and Instrument
Settings screens. Scroll through
the list using the up and down
arrows to view the details and print
the information if required.
If the printer paper roll needs
replacing, Print will be disabled and
you will be able to select Help to
view instructions on replacing the
printer paper (if you require further
instruction see page 1 5, Loading the
Printer Paper or Label Roll).
Select Done once to return to
System Information screen and a
second time to return to ChooseSettings screen.
5-36
Instrument Set Up
System Information
5
This completes the settings and
choices within Instrument Set Up.
Touch Done to return to the main
Select screen.
5-37
5-38
Recall Results
6
Patient Results
Patient Results
Recall results enables you to
search, view, and print patient test
results. These results are stored on
the analyzer. If you are using the
Clinitek Status connector, you
can also recall QC results.
At the Select Screen, touch RecallResults button.
The next screen displayed is
Select Test Results.
The test results are in chronological
order. The most recent test result is
displayed at the top of the screen
and is highlighted.
Use the up and down arrow keys to
scroll through the list of patient
tests and highlight the patient you
would like to recall.
To view details of a patient result,
touch the Select button.
The results will automatically be
sent to the connected computer if
this option is set up in the analyzer.
If you require further instruction
regarding how to set up the analyzer
so the results are automatically sent
to a computer see Section 5,
Instrument Set Up.
6-1
Recall Results
6
Patient Results
The first page of the patient’s
results is displayed on the screen.
If more than one page of results
exists for the patient, then the More
button will be present on the
screen.
If you would like to print the results,
touch the Print button. Any
information that was entered in
regard to the patient will be
included on the printout.
When you are finished viewing the
patient’s results, touch Done.
You will return to the Select TestResults screen. Press Exit to
return to the main Select screen.
6-2
Recall Results
6
Sending Data to a PC
Sending Data to a PC
To send data to a computer; if a
computer has been connected to the
analyzer, all results can be sent to the
computer by touching the Send alldata button.
If you are using a Clinitek Status
connector, refer to the
Clinitek Status Connect System
Operator's Guide.
6-3
6-4
Troubleshooting
7
General Information
Your Clinitek Status®+ analyzer will
operate properly if you follow the
directions for using and cleaning
the instrument.
Error Messages
Error messages will be displayed to
help you when the Clinitek Status+
analyzer detects something which
needs your attention. The format of
this advisory information depends
upon the importance of the
problem and the mode in which the
instrument is being used.
To correct an error, see the List of
Errors and Advisory Messages located
at the end of this section.
Errors which Disable the
Instrument
If the error is one which prevents
the instrument from being used, all
selection areas on the screen will
be disabled. Taking the corrective
action shown will remove the error
alert screen and allow you to use
the instrument.
Advisory Messages
Errors of less importance will be
presented via a message on the
main Select screen when this
screen is next displayed. When you
have taken corrective action, the
message will be removed from the
display. If more than one of this
class of error occurs, clearing one
message will enable the next to be
displayed in order of importance to
a user.
Results Alert
If an error occurs during testing and
the test cannot continue because of
the error, this will be presented via
the Results Alert screen. This will
provide details of the error and
show that the test has been
cancelled. The test table will be
extended so that the urinalysis strip
or Clinitest® cassette can be
removed.
Other Errors
There are certain errors which need
to be corrected to enable testing of
samples but do not prevent other
instrument functions from being
used. You will need to carry out the
corrective action to enable testing.
7-1
Troubleshooting
7
Battery Power Icon
The battery icon indicates the
power level of the battery. Power
can be reduced while testing
continues, with an advisory
message displayed on the main
Select screen. If battery level falls
too low to power the analyzer, all
selection areas on the display will
be disabled until the batteries are
replaced (if you require detailed
instructions regarding how to
change the batteries see Section 9,
Cleaning and Maintenance).
Paper-out Icon
A paper-out icon appears in the top
of the title bar when the printer
paper/label roll needs replacing.
An advisory message will be
displayed on the main Select
screen. Replace with new paper or
label roll as instructed in Section 1,
Loading the Printer Paper or Label
Roll.
Dashes may appear next to Color
and Clarity on test result printouts.
This occurs when the instrument is
powered by batteries. Color and
Clarity are selected in the
Instrument Set Up, but no
selections have been recorded on
the Select Appearance screens
before time-out.
The time-out on these screens is
designed to ensure that battery life
is preserved. The Color and Clarity
description may be added to the
printout in writing if needed.
Irregular or Slow Movement of
Test Table
If movement of the test table is
irregular or slow, this may be
caused by:
a) heavy buildup of dried urine on
the test table. Clean the test table
and insert as described in
Section 9, Periodic Cleaning of Test
Table.
b) low battery power. Replace the
batteries as described in Section 9,
Cleaning and Maintenance.
Dashes in Displays
Dashes are displayed in the
Results screens and on printouts
when no text has been entered for
a field enabled in Instrument SetUp.
7-2
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.