Clinitek, Clinitek Status, Multistix, Multistix PRO, Combistix, Uristix,
Labstix, Neostix, Clinitest and Chek-Stix are trademarks of
Siemens Healthcare Diagnostics.
Presept and Cidex are trademarks of Johnson & Johnson.
Theracide is a trademark of Lafayette Pharmaceuticals, Inc.
Amphyl is a trademark of Linden Corporation.
Kimwipes is a trademark of Kimberly-Clark.
US Pats 5,408,535; 5,477,326; 5,877,863; 6,239,445; D456,082; D489,816
Origin: UK
Siemens Healthcare Diagnostics Inc.
Tarrytown, NY 10591-5097 USA
Siemens Healthcare Diagnostics Ltd.
Sir William Siemens Sq.
Frimley, Camberley, GU16 8QD, UK
The information in this manual was correct at the time of printing.
How
ever, Siemens Healthcare Diagnostics continues to
and reserves the right to change specifications, equipment, and
maintenance procedures at any time without notice.
If this instrument is used in a manner differently than specified
in this manual, the protection provided by the equipment may
be impaired.
improve products
Table of Contents
A Brief Description ............................................................ v
Understanding the Symbols and Display Icons .............. vii
1 Unpacking & Set Up ................................................. 1-1
NOTE: Feature availability may
vary by geography due to regulatory
registration.
NOTE: Due to software changes,
some screens on the instrument
may appear slightly different from
those in this manual.
What does the analyzer do?
Your Clinitek Status
portable instrument for reading
Siemens Healthcare Diagnostics
urinalysis strips and Clinitest
immunoassay cassettes. No special
training is needed to use this
instrument. Several different Siemens
urinalysis strips (e.g., Multistix® 10 SG)
can be used with the analyzer as well
as the Clinitest hCG Pregnancy Test.
The analyzer can be set up to be
as simple or sophisticated as you
prefer. You may simply insert a
dipped urinalysis strip or a Clinitest
cassette into the analyzer and the
result will be reported (this is called
a Quick Test).
Or, you have the option to enter
an Operator Name, Patient Name
and Patient ID. This added
information will be reported along
with the test results (this is called a
Full Test).
The touch screen displays
instructions and prompts you
through operation of the analyzer. In
addition, you enter information
through the touch screen.
®
+
analyzer is a
®
Do I have to calibrate?
You do not have to do anything
to calibrate. The instrument
performs a system test each time it
is turned on. Then, each time a test
is run, the instrument automatically
calibrates. The white calibration bar
(on the test table) provides NIST
traceable calibration.
How does it work?
Testing starts (in the Quick Test
mode) when either the Strip Test
or Cassette Test is selected on the
main menu screen.
Touching the Strip Test key
prompts you to adjust the test table
to accept a urinalysis strip. Then
you touch the START key. You
have 8 seconds to dip the test strip,
blot the edge of the strip and place
it on the test strip table. The table is
partially pulled into the instrument
for calibration and then pulled
completely into the instrument to
read the test strip.
Touching the Cassette Test key
prompts you to adjust the test table
to accept a cassette. Then you
touch the START key. You have
8 seconds to draw the sample into
a pipette and dispense it into the
sample well of the cassette. The
analyzer automatically calibrates
and then pulls the table completely
into the instrument where the
cassette is read.
In the Full Test mode for either aStrip Test or Cassette Test, you
are prompted to enter an Operator
Name, Patient Name and/or Patient
ID prior to running a test.
v
A Brief Description
How do I get results?
Results will be displayed on the
touch screen and printed (if
desired). The results can also be
transferred to a computer using a
9-pin null modem serial cable and
the RS-232 serial port on the back
of the instrument.
The analyzer stores results from
950 patient tests. You are able to
recall past patient test results on
the analyzer using the RecallResults function.
How is the analyzer powered?
The analyzer can be plugged
into an electrical outlet for use on
the bench top, or it can be powered
by batteries and freely moved from
one testing site to another. The
batteries fit into an opening on the
bottom of the instrument.
the analyzer, use it for your daily
Siemens urinalysis and Clinitest
immunoassay testing and keep it in
good working condition.
As you read through the
Operator’s Manual, you will find
these symbols:
NOTES: contain useful tips
on using the analyzer.Notes
appear in italicized type.
CAUTION: should be
followed carefully to ensure your
analyzer operates correctly and
is not damaged. Cautions
appear in bold type.
In the Operator’s Manual, you will
notice some text is in bold/italic or
bold.
What about this Operator’s
Manual?
The Operator’s Manual contains
the directions you need to unpack
Bold text identifies a button
(touch sensitive area) on the
screen.
vi
Bold/italic text identifies screen names.
Understanding the Symbols and Display Icons
Instrument and Labeling Symbols
This section describes the symbols that appear on the exterior of the
Clinitek Status+ analyzer, the power supply provided with the instrument,
the carton in which the instrument was delivered and the supplies of
reagent strips and cassettes which you will use with the instrument.
Indicates that the input supply is direct current
Identifies that the instrument is type B equipment, which
provides a particular degree of protection against electric shock
Class 111
Identifies that the instrument is class 111 type equipment,
which is equipment for connection to SELV (Separated Extra
Low Voltage), i.e. a power supply
Double insulated product or transformer may also identify
class 2 equipment (power supply only)
Identifies that the instrument is listed by Underwriters
Laboratories as meeting U.S. and Canadian requirements for
safety
The CE mark identifies that the product complies with the
applicable directives of the European Union
Manufacturer
European authorized representative
Indicates a power on/off button
Caution, consult accompanying documents
In vitro diagnostic medical device
Consult instructions for use
vii
Understanding the Symbols and Display Icons
Indicates a serial port
This system contains certain toxic or hazardous substances or
elements. The environmental protection use period for this
system is 50 years. The system can be used safely during its
environmental protection use period. The system should be
recycled immediately after its environmental protection use
period has expired.
30
18
100
Temperature limitation (18° 30° C)
Contents sufficient for (n) tests (100)
Use by YYYY MM
Catalog number
Serial number
Batch code
Biohazard
Indicates that this equipment is classified as Waste Electrical
and Electronic Equipment under the European WEEE
Directive. It must be recycled or disposed of in accordance
with applicable local requirements.
Printed on recycled materials
viii
Indicates compliance with RESY packaging standards
Understanding the Symbols and Display Icons
Keep this way up
Fragile, handle with care
Keep dry
Keep away from sunlight and heat
VDE Testing and Certification Institute Germany
Manufacturer’s mark (FRIWO) and manufacturing location
(Hong Kong)
Manufacturer’s mark (FRIWO) and manufacturing location
(Geratebau, Germany)
Positive Temperature Coefficient (PTC) A thermistor device
used to protect the transformer from short circuits or overload.
This is an auto reset device.
Thermal cut out (TCO) This safety device disconnects the
supply voltage to the transformer at a specific temperature.
The operation temperature is stated below.
Ingress protection rating protected against the entry of solid
objects >1 mm but no protection from liquids.
This symbol indicates a risk of electric shock.
ix
Understanding the Symbols and Display Icons
Display Icons
There are seven icons which display in the top left of the display to show
the mode of the instrument. They also appear on the selection area for
each function. The icons are:
Instrument Set Up
This is displayed when the instrument is being set up to suit the
users’ requirements.
Strip Test (e.g., Multistix 10SG)
Shown when a test is being carried out using a reagent strip for
urinalysis, and when results are displayed following a strip test.
Cassette Test (e.g., Clinitest hCG)
Displayed when a cassette test is being carried out and when
results are shown after a cassette test.
Results recall
Used to show that results are being recalled from the
instrument’s memory.
Printer
This icon is displayed when results are being printed.
Data transfer to Personal Computer
Shows that data, including results, is being transferred to a PC.
Alert
Used when an error is being displayed.
There are two icons which may appear in the title bar.
Battery power
This has a maximum of four segments which show the level of
battery power. It will be shown in the top right corner of the title
bar when the instrument is battery powered.
Paper out
Appears in the top of the title bar when the printer paper/label
roll needs replacing.
x
Unpacking & Set Up
1
Unpacking
Unpacking
1 Carefully remove the contents
of the shipping carton. Check the
carton and instrument for visible
signs of damage; if seen,
immediately contact the carrier.
1
5
1
Clinitek Status®+ analyzer
Test table
2
Do not touch the white
calibration bar.
Test table insert
3
If you are using a reagent strip
that has 4 or fewer test pads,
e.g., Uristix® 4, you must use a short
test table insert. This has to be ordered
separately (for a list of suppliers see
Appendix A, Local Technical Support
Providers and Distributors).
2 Remove each of the wrappings
and check for the following items:
4
2
3
Power Supply
4
If the power cord is not the style
you need, contact your local
representative (for a list of contacts see
Appendix A, Local Technical Support
Providers and Distributors).
Paper Roll
5
Depending on the model you have
received, there may also be a Warranty
Registration Card, Unpacking/Setup Guide,
and/or Quick Reference Guide included.
1-1
Unpacking & Set Up
Set Up
1
Set Up
3
Analyzer Set Up
Place the instrument on a level
work surface where the
temperature and humidity are fairly
constant.
4
Plugging Analyzer In
Plug the appropriate end of the
power cord into the power inlet
socket located on the rear of the
Clinitek Status+ analyzer. Plug the
other end of the power cord into an
AC electrical wall outlet.
The best temperature for using
the instrument is between 22°C and
26°C (72°F and 79°F). Do not place
the analyzer outside or near
windows, ovens, hot plates, or
radiators.
Only use the power supply
adapter included with the unit.
5
Installing Batteries
(Optional)
Place the analyzer on its side and
remove the battery cover by
pressing down on the tab and
pulling out. Place the 6 new
alkaline AA-size batteries into the
analyzer. Replace the battery cover
and turn the instrument back onto
its base.
1-2
Unpacking & Set Up
1
Set Up
6
Inserting Test Strip Table
Insert the test strip table into the
analyzer by holding it by the end
opposite the white calibration bar
and with the white bar facing up.
Push the test table into the
analyzer, pushing it in just over
halfway.
Do not push the test table fully
into the analyzer as the test table
may become jammed and prevent
the use of the analyzer.
Do not touch the white
calibration bar.
7
Loading Test Table Insert
The test table insert adapts for use
with a Siemens Healthcare
Diagnostics urinalysis strip or
Clinitest immunoassay cassette.
One side is used for a strip test and
the other side is used for a cassette
test.
8
Interfacing to a Computer
The instrument can send results to
a computer via the serial port
located on the back of the analyzer.
This requires a 9-pin null modem
serial cable that can be purchased
separately at an electronics store or
from your Siemens Representative
(for a list of suppliers see Appendix
A, Local Technical Support
Providers and Distributors).
1-3
Unpacking & Set Up
1
Set Up
9
Interfacing to the
Clinitek Status Connector
The Clinitek Status connector
allows for Ethernet or wireless
network connectivity, Quality
Control, increased security, bar
code scanning, and additional
features with the Clinitek Status+
analyzer.
This connector provides standard
wired and wireless connectivity of
the Clinitek Status+ system to your
LAN, LIS, HIS, EMR, and allows
for centralized control of all
satellite Point of Care (POC)
Clinitek Status+ analyzers.
Refer to the Clinitek StatusConnect System Operator’s Guide.
Analyzer Software Upgrades
10
From time to time Siemens will add
new features and make improvements
to the Clinitek Status+ instrument
software.
These software updates will be
available on an electronic memory
card which is inserted into the
software update socket. This socket
is located under the printer cover
and is on the left-hand side of the
printer when you face the back of
the instrument.
Updating the software is a simple
procedure. Instructions for updating
the software on your instrument will
be supplied with the memory card.
1-4
Unpacking & Set Up
Set Up
1
11
Loading the Printer Paper or
Label Roll
1. Open the printer cover by pulling
up on the tab.
2. Open the paper roll compartment
cover by pressing down on its tab
and pulling out.
3. Lift the paper holding arm into
the open, upright position.
4. Place the new paper roll into the
printer paper compartment with the
paper unrolling from underneath
and toward the compartment wall.
5. Feed the paper up along wall
and through the printer. Once you
have approximately 4 inches (or
10 cm) of paper through the printer
then feed the edge of the paper
through the printer cover.
6. Push the paper holding arm
down in to the closed position.
paper holding arm
7. Close the printer and paper roll
covers by clicking them into
position.
The analyzer is set up to
automatically print the results (to
turn off the automatic print function
see Section 5, Instrument Set Up).
The analyzer uses ordinary
thermal paper as provided, or label
stock (for ordering information see
Appendix A, Local Technical
Support Providers and Distributors).
1-5
Unpacking & Set Up
Set Up
1
12
Warranty Registration
1. Lift the printer cover on the
instrument and the serial plate with
the instrument’s serial number will
be visible.
2. Write the serial number and
installation date on the Warranty
Registration Card. After the
instrument has been successfully
installed, complete the information
on the Warranty Registration Card
and return the card to your local
Siemens office (for a contact list
see Appendix A, Local Technical
Support Providers and
Distributors).
1-6
Unpacking & Set Up
1
Powering Up
Powering Up
Press the on/off button located
on the front of the instrument.
This is the first screen displayed.
The analyzer will run an automatic
system diagnostic test each time it
is turned on.
If this is the first time you have
turned on the analyzer, you will be
led through a Start-Up Wizard, a
quick set-up procedure. If you
require further instruction regarding
the Start-Up Wizard see Section 3,
page 3-1.
1-7
Unpacking & Set Up
Powering Down
1
Powering Down
1. Before turning the analyzer off,
always ensure that there is no strip
or cassette on the test table and
that the table and insert are clean.
2. Press the on/off button for
at least 2 seconds. The test table
will retract into the analyzer. If there
is no strip or cassette on the test
table, the door will close and the
analyzer will switch off.
If a strip or cassette is still on the
test table, the test table will be
pushed out and the analyzer will
turn off. The test table will remain
out. In order to retract the test table
into the analyzer, turn the analyzer
on, and then off (without a strip or
cassette on the test table).
Do not push the test table fully
into the analyzer as the test table
may become jammed and prevent
the use of the analyzer.
1-8
Interacting with the Touch Screen
2
Screens
Screens
The touch screen will guide you
through the operation of the
Clinitek Status®+ analyzer. The
screen will display messages,
instructions and options to which
you respond by touching the
appropriate area on the screen.
The first main screen you see is the
Select screen. It displays the time
and date, and indicates the
5 possible actions:
Instrument Set Up
Recall Results
QC Test
Cassette Test
Strip Test
Icon
Title Bar
Each screen that follows the Select
screen has an icon, title bar and
touch-sensitive active areas.
In some cases, the screen will also
display instructions, messages or
error messages.
The icon indicates the main section
in which you are working (1 of the
5 sections listed on the main
Select screen).
Touch Sensitive Area
2-1
Interacting with the Touch Screen
Screens
2
How to Touch the Screens
The screen needs to be touched
lightly in the touch-sensitive area to
activate a response.
Where to Touch the Screens
There are three types of areas that
respond to touching the screen.
• Round Buttons
• Boxed Areas
• Scroll Arrows
Round Buttons
These buttons typically appear on
screens that require a selection
among several items. The button
with a filled circle is the current
selection.
To change your selection, touch an
unfilled circle. The newly selected
circle (button) will now be
highlighted. You then touch the
Next button to move to the next
screen.
If a touched area does not
respond as expected, slide your
finger across the appropriate
selection area.
Use of anything hard or
pointed on the touch screen
may cause damage.
Round Buttons
In order to proceed, you will
always touch the box with a right
pointed arrow labeled Next.
In order to go back, you will
always touch the box with a left
pointed arrow. These “back” option
boxes vary in title.
2-2
Interacting with the Touch Screen
Screens
2
Boxed Areas
These are areas on the screen
enclosed in boxes. Simply touch
any area within the box to activate
that function.
The boxed areas vary in size. The
boxes located on the main Select
screen are examples of larger
areas. Smaller box selections
include boxes such as the
“Previous” and “Next” boxes found
at the bottom of the screen.
Boxed Area
Scroll Arrows
Press the up and down arrows on
the right side of the screen to scroll
through the list of information on
the left side of the screen. Once the
information on the left side of the
screen is highlighted, touch the
Select button to confirm your
selection and move to the next
screen.
Unavailable
Option
If there are double arrows on the
screen, these arrows (when
touched) will take you to the top or
bottom of the page.
When an option can be
selected it will be shown with a
thick black frame and will respond
when touched. If an option is not
available, it will be framed with a
thin black line and not respond
when touched.
2-3
Interacting with the Touch Screen
Keyboards
2
Keyboards
Press to switch to
Using the Alpha-Numeric
Keyboard
When the screen prompts you to
enter information for Operator,
Patient’s Name and/or Patient
Identification, a keyboard will
appear on the screen.
Depending on how your analyzer is
set up, either an alphabetic or
numeric keyboard will be displayed
first. The first keyboard displayed is
referred to as “keyboard priority.”
numeric keyboard
Data Entry Box
Back Space/
Erase key
If you require further instruction
regarding how to change keyboard
priority see Section 5, Instrument Set
Up.
To switch between the keyboards,
touch the 123 button to get to the
numeric keyboard. Touch the ABC
button to get to the alphabetic
keyboard.
To type in a name, number, birth
date, etc., touch the appropriate
button. Your selections will appear
in the data entry box.
If you switch between keyboards,
all values will be retained in the
data entry box on both keyboard
screens.
2-4
Return to
previous
screen
Press to switch to
alphabetic keyboard
Creates
blank
space
Confirm
data
entries
Interacting with the Touch Screen
2
Keyboards
The maximum number of
characters allowed is 32. An
audible tone will sound when you
have exceeded the maximum
number of characters.
Once you have finished entering
the information, touch Enter (from
either keyboard screen).
2-5
2-6
Start-Up Wizard
3
The first time your Clinitek Status®+
analyzer is turned on (following an
automatic system diagnostic test), it
will take you through a quick set up
procedure. This procedure will
allow you to select the basic
functions of the analyzer so you
can use the analyzer with your
choice of settings.
The Start-Up Wizard will allow you
to select the following settings:
Language
Date and time
Preferred test sequence
(e.g., Quick Test or Full
Test)
Type of urinalysis strip
Results format
If you require further instruction
regarding how to change the settings
see Section 5, Instrument Set Up.
Once you have selected your
choices, the analyzer will display a
Confirmation screen which allows
you to check that your preferences
are correct.
3-1
3-2
Testing
4
Quick Tests
Quick Tests
Urinalysis Strip Test
BIOHAZARD Wear personal protective
equipment. Use universal precautions.
Refer to Appendix H for recommended
precautions when working with
biohazardous materials.
Testing is started from the main
Select screen.
If you require more information
regarding use and storage of test
strips, please refer to the strip package
insert.
Touch Strip Test to conduct a
urinalysis strip test.
The next screen that appears is
Prepare Test.
If you would like the steps for
urinalysis testing to be shown on the
screen then touch Help.
Make sure the test table insert has
the reagent strip holder facing
upward.
Also, have the test strip, urine
sample and paper towel ready.
Touch the START button.
4-1
Testing
4
Quick Tests
The next screen that appears is
another Prepare Test. This screen
prompts you through the steps to
prepare the test strip.
A timer displays how much time
you have remaining to complete the
steps.
You have 8 seconds to complete
the following four steps:
1 Dip the reagent strip into the
urine sample, wetting all pads.
Immediately remove the strip from
the urine.
NOTE: Do not dip the automatic
identification band or color band
in the urine sample.
2 Drag the edge of the strip
against the side of the sample
container as you remove it.
4-2
Loading...
+ 114 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.