Siemens Ardoptix PT 140 AF4, Ardoptix PT 140 AF1, Ardoptix PT 143 AF4, Ardoptix PT 143 AF1 Operating Manual

Page 1
Two-colour Pyrometer
PT 14x
7MC3040-...
03/2017
Page 2
Page 3
Two-colour Pyrometer PT
Ident.-Nr.: 106 7393
PT14x_si_en.doc
Stand: 21.03.2017
830 hm / 810 het
Page 4
Copyright prohibits the reproduction or distribution of this instruction manual, including text, photographs or images contained herein, in whole or in part, for any purpose whatsoever, without prior consent of the au­thor. This applies to any form of mechanical or electronic reproduction as well as to electronic transmission in any form through any medium.
Please note:
Unless otherwise stated in this instruction manual, the instruments described herein are subject to change without prior notice, particularly modifications for the sake of technological advancement.
Siemens AG
Digital Factory Customer Services DF&PD DF CS SD OP ITM Telefon +49 (911) 750-9600 Fax +49 (911) 750-9609 Email: sirent.industry@siemens.com
Office/Testing
Warehouse/Counter service
Breslauer Str. 5
Gundelfinger Str. 20
90766 Fürth
90451 Nürnberg
Germany
Germany
Page 5
s
Operating manual
Contents
1 Product Numbers Ardoptix PT .................................................. 1
2 Miscellaneous ............................................................................. 2
2.1 Informationen about this manual ........................................................ 2
2.2 Explanation of symbols ....................................................................... 2
2.3 Liability and Warranty ......................................................................... 2
2.4 Copyright ............................................................................................ 3
3 Safety .......................................................................................... 3
3.1 Intended use ....................................................................................... 3
3.2 User’s responsibility ................................ ................................ ............ 4
3.3 Electromagnetic Compatibility ............................................................ 4
4 Scope of delivery ........................................................................ 4
5 General Description ................................................................... 4
5.1 Intended Use ...................................................................................... 4
5.2 Operating Controls and Display .......................................................... 4
6 Charging the batteries ............................................................... 5
7 Preparing the pyrometer for measurements ............................ 6
7.1 General Information ............................................................................ 6
7.2 Diopter compensation ......................................................................... 6
7.3 Brightness control to protect the eye .................................................. 6
7.4 Focusing the pyrometer ...................................................................... 6
7.5 Align the pyrometers ........................................................................... 7
7.6 Performing a measurement ................................ ................................ 7
7.7 Automatic switch-off ........................................................................... 8
8 Setting parameters at the pyrometer (basic configuration) .... 8
9 Menu ............................................................................................ 9
9.1 Measurement two-colour/ratio mode configuration layer C001.......... 9
9.2 Configuration data acquisition configuration layer C010 ................... 10
9.3 General function configuration layer C011 ....................................... 10
9.4 Displayed temperature readings configuration layer C020 ............... 11
10 Setting the Emissivity Ratio (two-colour/ratio mode) ............ 11
11 Initializing emissivity ratio constants ..................................... 12
11.1 Define number of emissivity ratio constants in array ........................ 12
11.1.1 Store and assign emissivity ratio constants .................................. 12
11.2 Select the emissivity ratio constant ................................................... 12
12 Determining and Setting Emissivity (Spectral mode) ............ 13
13 Initializing emissivity factors (Spectral mode) ....................... 13
13.1 Define number of emissivity constants in array ................................ 13
13.1.1 Store and assign emissivity constants .......................................... 14
13.2 Select the emissivity constant ........................................................... 14
14 Further configuration ............................................................... 14
15 Configuration – signal conditioning ....................................... 14
15.1.1 Transmission factor (spectral mode) ............................................ 14
15.1.2 Background Temperature Compensation (Spectral mode) .......... 15
15.1.3 Temperature offset using linear interpolation ............................... 15
15.1.4 Smoothing function ................................ ................................ ....... 16
15.2 Min/Max memory .............................................................................. 17
15.2.1 Min/Max Memory .......................................................................... 17
Page 6
s
Operating manual
15.2.2 Double Maximum Memory with hold time ..................................... 17
15.2.3 Double Maximum Memory with hold time "Combined" ................. 18
15.2.4 Automatic Temperature Detection (ATD) ..................................... 18
16 Setting Parameters at the device ............................................ 22
16.1 Configuration layers ................................................................ .......... 22
16.1.1 Temperature measurement using two-colour/ratio mode ............. 23
16.1.2 Temperature measurement lambda 1 .......................................... 24
16.1.3 Temperature measurement lambda 2 .......................................... 25
16.1.4 Configuration I/O .......................................................................... 27
16.1.5 General Functions ........................................................................ 28
16.1.6 Displayed temperature readings ................................................... 29
17 Software CellaView .................................................................. 29
18 PC Interface .............................................................................. 29
19 Remote configuration .............................................................. 30
19.1 Main Menu ........................................................................................ 31
19.2 View Current Configuration............................................................... 31
19.3 Submenus ........................................................................................ 32
19.3.1 Configure two-colour/ratio mode .................................................. 32
19.3.2 Lambda 1 ..................................................................................... 32
19.3.3 Lambda 2 ..................................................................................... 33
19.3.4 Quick access to emissivity / signal smoothing / operating mode .. 33
19.3.5 I/O Signal Configuration ............................................................... 33
19.3.6 Automatic temperature data output .............................................. 35
19.4 User Recalibration ............................................................................ 36
20 Maintenance ............................................................................. 38
20.1 Cleaning the pyrometer lens/protective glass ................................... 38
21 Accessories .............................................................................. 39
21.1 Supplementary lens .......................................................................... 39
22 Theory of Non-Contact Temperature Measurements ............ 40
22.1 Emissivity .......................................................................................... 40
22.2 Temperature measurement using spectral mode ............................. 41
22.3 Temperature measurement using two-colour/ratio mode ................. 41
22.4 Emissivity Coefficient Table – Spectral Mode ................................... 42
23 General technical data ............................................................. 43
24 Technical Data PT 140 (650 – 1700 °C) .................................... 43
24.1 Field of View Diagrams PT 140 (650 – 1700 °C) .............................. 43
25 Technical Data PT 140 (750 – 2400 °C) .................................... 45
25.1 Field of View Diagrams PT 140 (750 – 2400 °C) .............................. 45
26 Technical Data PT 143 (650 – 1700 °C) .................................... 46
26.1 Field of View Diagrams PT 143 (650 - 1700°C) ................................ 46
27 Technische Daten PT 143 (MB 750 - 2400°C).......................... 47
27.1 Field of View Diagrams PT 143 (MB 750 - 2400°C) ......................... 47
29 Dimensions ............................................................................... 48
30 Shipping, Packaging and Disposal ......................................... 49
30.1 Inspecting your shipment .................................................................. 49
30.2 Packaging ......................................................................................... 49
30.3 Disposal of used apparatus .............................................................. 49
31 Copyright .................................................................................. 50
Page 7
s
Operating manual
32 Default settings ........................................................................ 51
32.1 Temperature measurement (Configuration layer: C001)............. 51
32.2 Temperature measurement Lambda 1 /2(Configuration layer: C002) .. 52
32.3 Temperature measurement Lambda 2(Configuration layer: C003) ...... 53
32.4 General Functions (Configuration layer: C010) ........................... 53
32.5 General Functions (Configuration layer: C011) ........................... 54
Page 8
s
Operating manual
Page 9
s
Operating manual
1
1 Product Numbers Ardoptix PT
Pyrometer
Temperarute range
Product number
Ardoptix PT 140 AF1
MB: 650-1700 °C, D=80:1 (St.-Optic)
7MC3091-1AB40
Ardoptix PT 140 AF4
MB: 750-2400 °C, D=150:1 (St.-Optic)
7MC3091-1AB20
Ardoptix PT 143 AF1
MB: 650-1700 °C, DV=230:1, DH=45:1 (St.-Optic)
7MC3091-1AB43
Ardoptix PT 143 AF4
MB: 750-2400 °C, DV=350:1, DH=50:1 (St.-Otic)
7MC3091-1AE43
Page 10
s
Operating manual
2
2 Miscellaneous
2.1 Informationen about this manual
The Operating Manual shall enable the user to properly install the pyrometer and those accessories which are necessary.
Before starting installation, be sure to read and understand this entire manual, in particular the chapter on safety! The instructions contained in this manual, especially those concerning safety, as well as site-specific regulations governing UV radiation must be complied with at all times!
2.2 Explanation of symbols
Important safety-related references in this manual are marked with a symbol.
CAUTION! This symbol indicates important information which, if neglected, might result in pyrometer damage, malfunction or breakdown.
PLEASE NOTE !
This symbol points out guidelines which should be heeded for efficient and trouble-free operation.
2.3 Liability and Warranty
All information compiled in this manual is in accordance with applicable regulations. The statements made are based on state-of-the-art technol­ogy and reflect our extensive knowledge and many years of experience.
PLEASE NOTE !
Always carefully read this Operating Manual before beginning any work on or with the instrument, especially prior to installation and initial setup! The Manufacturer shall not be held liable for any damages or malfunctions arising from a disregard of the warnings and instructions contained herein.
This Operating Manual must be retained for future use. Please ensure that all persons who wish to operate the instrument have access to this manual.
Page 11
s
Operating manual
3
2.4 Copyright
This Operating Manual should be treated as confidential. It is solely in­tended for use by persons involved with the instrument. This manual may
not be made available to a third party without prior Manufacturer’s con-
sent. Please contact the Manufacturer if the need should arise.
PLEASE NOTE !
The data, texts, charts, drawings, images or other representations con­tained in this manual are copyright-protected and furthermore, subject to intellectual property rights. Violators will be prosecuted. Unauthorised use and copyright infringement will be subject to penalty by law.
Reproductions of any kind, in whole or in part, as well as the exploitation or disclosure of this manual’s content without the explicit written approval of the Manufacturer are expressly prohibited by law. Violations shall be subject to compensation claims by the Manufacturer. The right to claim additional indemnities remains reserved.
3 Safety
This chapter outlines all important safety aspects to be considered for optimum employee protection and to ensure safe and reliable operations.
3.1 Intended use
The pyrometer is solely intended for non-contact measurement of tem­peratures as described in this manual. Any other use is not intended. Operational safety can only be ensured when the instrument is used for its intended purpose.
CAUTION ! It is prohibited to use the pyrometer for any other purpose beyond what
is specified in this manual. Using the instrument in any other manner will be considered as improper.
The Manufacturer/Authorised Agent shall not be held liable for any dam­ages or loss resulting from such unintended or improper use; in this case the risk is solely borne by the user.
Page 12
s
Operating manual
4
3.2 User’s responsibility
The pyrometer may only be used when it is in perfect working condition.
3.3 Electromagnetic Compatibility
The devices comply with the essential safety requirements of the Elec­tromagnetic Compatibility Directive 2014/30/EU (EMC Act).
4 Scope of delivery
Make sure that all of the following components have been included with your shipment:
Pyrometer Carrying case Power supply Calibration certificate Software CellaView (Download Version) USB cable
5 General Description
5.1 Intended Use
The Ardoptix PT series provides efficient pyrometers for non-contact temperature measurements.
The two-colour pyrometer Ardoptix PT 14x measures the intensity of in­frared radiation at two different wavelengths. The ratio of these two in­tensities is proportional to the temperature. The Ardoptix PT 14x is used for temperature measurement from 650 °C to 2400 °C. These instru­ments serve a broad range of applications which include the iron and steel producing industry as well as the metal, glass, cement and chemi­cal industries.
5.2 Operating Controls and Display
The rear panel of the Ardoptix has a 4-digit display and 4 push-buttons. The display normally indicates the currently measured temperature. Dur­ing configuration (using the push-buttons), the display will show the pa­rameter you are setting. Whenever the display shows a parameter, the F1 Parameter LED (yellow) will light up. During a running measurement, the F2 Status LED (green) will light up.
Page 13
s
Operating manual
5
6 Charging the batteries
Connect the battery charging socket of the Ardoptix to the provided pow­er supply (9 V DC). The charge indicator lights as long as the batteries are being charged. A complete charging cycle takes approx. 15 hours.
Parameter LED
Status LED
Ocular
Mode key
Charge indicator
Charging socket
¼“ Tripod thread
Switch „ON“
USB connection
Parameter select / change
Display
Only use the power supply which comes with the pyrome­ter. Do not use any other power supply as this may cause irreparable damage to the device.
The charger is only for charging the battery. The rechargeable battery is only charged when the pyrom­eter is switched off.
Page 14
s
Operating manual
6
7 Preparing the pyrometer for measurements
7.1 General Information
The pyrometer’s field of view must remain unobstructed. Any obstruction may lead to measurement errors.
7.2 Diopter compensation
7.3 Brightness control to protect the eye
7.4 Focusing the pyrometer
To ensure precise temperature measurement, the pyrometer must be correctly focused on the target object
The pyrometer offers diopter compensa­tion to adjust the sighting device to the
user’s vision. Turn the ring until you see a sharp image of the target spot indica­tion in the viewfinder.
Instruments with a measuring range of > 2000 °C are equipped with a polarisation filter to control brightness and protect the
user’s eye.
The pyrometer is equipped with focusable optics. Turn the lens until the target object and the target marker (distinctly marked in the viewfinder) both appear in sharp focus simultaneously
Page 15
s
Operating manual
7
7.5 Align the pyrometers
Direct the Ardoptix to the object to be measured. The target marker in the through-the-lens sighting device indicates the size of the target. (see technical data/ course of target). When measuring with a two-colour py­rometer, the object to be measured is not required to fill the complete measurement area. The integrated traffic light function in the viewfinder signals an inadmissible partial illumination and the measurement is stopped.
7.6 Performing a measurement
Press the ON button to switch on the device. If necessary adjust the ratio correction. Focus the target marker at the diopter correction switch and adjust the pyrometer lens to the correct measuring distance.
Measurement is in progress as long as you press the ON button. The temperature (depending on the configured mode) is displayed on the rear panel when you release the ON button. As an option, the measure­ment readings can be transmitted via interface.
1) Traffic light indicates
2) Measurement area
3) Target object
2 3
1
The amount of thermal radiation emitted from an object de­pends on the specific radiation properties of the material and its surface. Non-contact temperature measurement requires that you determine the material constant (emissivity) of the target object prior to first-time use
Page 16
s
Operating manual
8
7.7 Automatic switch-off
In the default setting, the Ardoptix switches off automatically after 2 minutes if none of the control keys are pressed. Automatic switch-off can be configured from 1 – 60 minutes. The Ardoptix measures continuously when it is in "automatic disconnection deactivated" mode.
8 Setting parameters at the pyrometer (basic configura-
tion)
Use the buttons ▲▼ and the „MODE" button on the rear panel to access and configure parameters. With these buttons you can view and adjust all settings required for operating the pyrometer.
Menu structure:
1. Press the MODE button while in normal operating mode to switch to "configuration layer" mode.
2. Use ▲▼ to select the configuration layer for the parameter you wish to set.
3. Press MODE to confirm. Press ▲▼ to select the particular param­eter.
4. Press MODE to confirm. Press ▲▼ to adjust the parameter value.
5. Press MODE again to end. Press ▲▼ to select END.
Whenever a control key has not been pressed within 30 seconds, the py­rometer will automatically return to normal operating mode. The modified value is applied.
Normal
operating mode
(temperature)
Select
config. layer
C000-
C999
Select
parameter
EPS1
Adjust
parameter
99.9
MOD
MOD
MOD
▲▼
▲▼
▲▼
Key lock may have been activated at the terminal. When select­ing the configuration layer you will be prompted to enter an ac­cess code with
P000
. To obtain full access to parameter set-
tings, enter
P100
otherwise you will only be able to view pa-
rameters but not change them.
Page 17
s
Operating manual
9
9 Menu
9.1 Measurement two-colour/ratio mode configuration layer C001
In the default configuration, certain parameter settings will be hidden. If required, you can have them shown
Page 18
s
Operating manual
10
9.2 Configuration data acquisition configuration layer C010
9.3 General function configuration layer C011
Page 19
s
Operating manual
11
9.4 Displayed temperature readings configuration layer C020
10 Setting the Emissivity Ratio (two-colour/ratio mode)
When the radiation properties of the target object’s surface (emissivity)
differ at two different wavelengths, or when interferences in the field of viewsuch as dust or steamdo not weaken the signal to the same de­gree at each wavelength, the pyrometer can be adjusted by setting the ratio of these two emissivity coefficients.
During normal operating mode, the emissivity ratio can be set at the pyrometer using the ▲▼ buttons. When simultaneously pressing the MODE key, the display shows the current measuring temperature while the emissivity ratio coefficient continues to be adjusted in the background. This is an easy way to determine the emis­sivity ratio when the object temperature is known. The modified values are directly adopted.
Page 20
s
Operating manual
12
11 Initializing emissivity ratio constants
The Ardoptix lets you store up to 10 emissivity ratio factors. Before you perform a new measurement, simply use the ▲▼ control keys to select your preconfigured emissivity factor. When you select a stored emissivity ratio factor, the display will briefly show the preconfigured factor for that emissivity ratio factor.
11.1 Define number of emissivity ratio constants in array
Before you configure the pyrometer with emissivity ratio factors, you must define the size/length of the array. This is the number of emissivity ratio you would like the array to contain (10 maximum). Using parameter
E.TB.Q, enter the total number.
Measurement two-colour/ratio mode configuration layer C001
Parameter
Function
Explanation
E.TB.Q
Enter desired number of elements in array
If E.TB.1 = OFF, then use ▲▼ keys to manually set emis­sivity ratio
11.1.1 Store and assign emissivity ratio constants
Next, assign a value to each emissivity ratio constant in the index.
Measurement two-colour/ratio mode configuration layer C001
Parameter
Function
Explanation
E. 01
Emissivity ratio Memory position 1
z. B. 100.5 %
E. 02
Emissivity ratio Memory position 2
z. B. 101 %
E. 03
Emissivity ratio Memory position 3
z. B. 101.5 %
11.2 Select the emissivity ratio constant
The memory position can optionally be set using the menu.
Once you have configured the emissivity ratio parameter, the pyrometer will maintain this specific setting. The py­rometer will always operate with this value unless you change the setting.
Page 21
s
Operating manual
13
Measurement two-colour/ratio mode configuration layer C001
Parameter
Function
Explanation
E.IDX
Memory position
Shows current selected emissivity ratio con­stant, e.g. E. 02
12 Determining and Setting Emissivity (Spectral mode)
In pyrometry, the specific radiation characteristics of the measured object will influence the generated temperature data. In order to produce accu­rate and reliable temperature readings, the pyrometer must be config­ured according to the specific material constant (emissivity).
In spectral operation mode, emissivity can be set at the pyrometer using the ▲▼ buttons. This changed setting will apply immediately to the cur­rent temperature measurement
13 Initializing emissivity factors (Spectral mode)
The Ardoptix lets you store up to 10 emissivity factors for lambda 1 and lambda 2. Before you perform a new measurement, simply use the ▲▼ control keys to select your preconfigured emissivity factor. When you se­lect a stored emissivity factor, the display will briefly show the preconfig­ured factor for that emissivity factor.
13.1 Define number of emissivity constants in array
Before you configure the pyrometer with emissivity factors, you must de­fine the size/length of the array. This is the number of material constants you would like the array to contain (10 maximum). Using parameter
E.TB.1/ E.TB.2, enter the total number.
Once you have configured the emissivity parameter, the pyrometer will maintain this specific setting. The pyrome­ter will always operate with this value unless you change the setting.
To determine the correct setting of the emissivity of a compara­tive measurement is carried out by a contact measurement. In order to minimize measurement differences between the two different physical measuring methods, the comparison meas­urement is almost the same time and performs at the same measuring point as possible.
Page 22
s
Operating manual
14
Temperature measurement lambda 1 (Configuration layer: C002)
Parameter
Function
Explanation
E.TB.1
Enter desired number of elements in array
If E.TB.1 = OFF, then use ▲▼ keys to manually set emis­sivity
13.1.1 Store and assign emissivity constants
Next, assign a value to each material constant in the index (E. 01, E. 02 etc.).
Temperature measurement lambda 1 (Configuration layer: C002)
Parameter
Function
Explanation
E. 01
Emissivity Memory position 1
z. B. 75 %
E. 02
Emissivity Memory position 2
z. B. 60 %
E. 03
Emissivity Memory position 3
z. B. 50 %
13.2 Select the emissivity constant
The memory position can optionally be set using the menu.
Temperature measurement lambda 1 (Configuration layer: C002)
Parameter
Function
Explanation
E.IDX
Memory position
Shows current selected emissivity constant, e.g. E. 02
14 Further configuration 15 Configuration – signal conditioning
15.1.1 Transmission factor (spectral mode)
The pyrometer has to be adjusted for the transmission properties of any supplementary lens and/or protective window screwed onto the pyrome­ter. Set the pyrometer for the specific transmission factor (a percentage
value) which is indicated either in the product’s data sheet or on the lens
itself. This is parameter TAU.1 and can be accessed at configuration layer C002/C003. If you are not using any auxiliary lens or protective window, set the parameter to 100.0.
Parameters for lambda 2 are identical to lambda 1. See Chap­ter 16.1.3
Page 23
s
Operating manual
15
15.1.2 Background Temperature Compensation (Spectral mode)
Thermal radiation reflected from the surrounding will require that you make an additional correction when the reflected radiation is strong compared to the natural radiation of the target. This applies to object sur­faces which have very low emissivity or when the temperature of the ob­ject is lower than the temperature of its surroundings.
The background radiation reflected from the object consists of the follow­ing factors:
Background temperature Background size The capability of the background material to reflect infrared radia-
tion
To ensure that the measurement is not distorted and that the tempera­ture readings are completely accurate, you should activate "background temperature compensation" (switch on C002 or C003 / BAC.1). Now enter the temperature value of the ambient source of radiation (BAC.T) and its influence as a percentage (BAC.%). The influence in percent stands for the size and the capability of the material to emit infra­red radiation. You will have to determine these variables first.
15.1.3 Temperature offset using linear interpolation
When necessary, Ardoptix PT allows you to program an offset for the temperature reading reported. The offset correction can be configured
Source of extraneous radiation
Example: furnace 950 °C
Target object TO = 900 °C
Emissivity e < 1 Reflection s > 0
Furnace radiation T
Back
= 950 °C
Page 24
s
Operating manual
16
individually with a minimum of 2 and a maximum of 10 interpolating points (X/Y nodes). This information is stored in a user-defined tempera­ture-indexed lookup table and used for signal conditioning. For values lower than the 1st node and higher than the last node, the first and last linear segments are extrapolated. Enter the nodes in ascending order. Use the rear panel display to access this function via C001 / LIN.1.
15.1.4 Smoothing function
When the target object’s temperature is erratic, it makes sense to
smooth these temperature fluctuations in order to stabilize the signal. The greater the time constant t
98,
(user-definable), the lower the effect of these fluctuations on the yielded temperature reading. The pyrometer’s response time is proportional to the time constant. Set signal smoothing via parameter C001 / FIL.Q.
analogue output signal
time
output signal without smoothing
output signal with smoothing
X4
X3
X1
X2
Y1
Y2
Y3
Y4
Temperature before offset
Temperature after offset
Page 25
s
Operating manual
17
15.2 Min/Max memory
The pyrometer features a data memory to store minimum and maximum (peak) temperature readings. This feature can be configured in one of the following modes:
Memory off Store minimum value (single) Store maximum value (single) Store double max. value for cyclical processes ATD function (Only available in pyrometer with ATD function)
15.2.1 Min/Max Memory
In this modealso called peak pickerthe pyrometer determines the highest or lowest temperature reading and keeps the temperature read­ing. The value is stored until you press the ON button again. Additionally, you can define the smoothing filter setting.
15.2.2 Double Maximum Memory with hold time
In this operating mode the pyrometer continuously detects the maximum temperature reading. This value is stored for the duration of the config­ured hold time and is displayed on the rear panel. In mid-sequence – af­ter 50% lapse of the hold time – a second internal peak picker starts. When the hold time has expired without finding a new max value, the output signal decreases to the value of the second peak temperature (if it turned out being lower than the first peak).
This memory mode serves to detect the maximum temperature of ob­jects moving periodically past the pyrometer lens. The temperature read­ing is kept for the duration of the configured holding time when the py­rometer does not detect any hot object. The holding time should be con­figured to approximately 1.5 times as long as the cycle of the moving tar­gets. This ensures that a temperature measurement gap is avoided and temperature changes are detected quickly.
Page 26
s
Operating manual
18
target object in the pyrometer’s field of view
holdtime
time
temp. reading with double max. buffer
temp. measurement without double max. buffer
15.2.3 Double Maximum Memory with hold time "Combined"
The function of the double maximum memory "combined" is similar to that of the double maximum memory. However, the hold time starts when the spectral temperature is at its highest point. Displayed is the corresponding two-colour temperature. If the spectral temperature de­creases during the hold time, the corresponding two-colour temperature is only displayed when the hold time has elapsed. If the spectral temper­ature rises during the hold time, the corresponding two-colour tempera­ture is directly displayed.
15.2.4 Automatic Temperature Detection (ATD)
The ATD function allows a very easy and fully automatic capture of the temperature. As soon as the pyrometer is switched on and aimed at a hot object, the measurement starts automatically. The threshold value to detect a hot object is adjustable. Depending on the configuration, the measurement ends either after a selectable period or when there is no longer a hot object in the pyrometer's measurement area. At the end of the measurement the captured value is displayed until a new measuring object is detected. To eliminate unrealistic measurement differences, it is possible to check the determined measured value for plausibility in rela­tion to the previous measured value and to filter it out. The new meas­ured value can also be averaged with a weighting factor to eliminate large jumps between the measured values.
Page 27
s
Operating manual
19
Threshold limits
The start of a measuring cycle is determined automatically and is de­pendent on the following variables:
Parameter
Function
Limit 1 (LI. 1)
Before beginning the measurement, the temperature reading must have been lower than Limit 1 at least once. If Auto reset is activated (A.RST= ON), Limit 1 will be ignored
Limit 2 (LI. 2)
Limit 2 must be exceeded at least for the duration of the time delay (T.DEL).
Time Delay (T.DEL):
See Limit 2
When the conditions are fulfilled, the sampling time can begin (T.ACT).
Page 28
s
Operating manual
20
Parameter
Function
Sampling time
(T.ACT)
During the sampling time the temperature is detected and stored as a temperature val­ue.
Display and output of temperature reading [ANO] The parameter ANO (normal display mode) defines which temperature
value is saved during sampling.
Parameter
Function
Display mode (ANO)
T=0“ displays the lower temperature range
limit during the measurement.
T.HLD“ indicates the previous tempera-
ture reading during the current measure­ment.
The green status LED indicates that a measurement is currently in pro­cess.
Average weighting [F-PR] Plausibility check
When the sampling time has ended, an average value is calculated for recorded measuring cycles. The temperature reading is weighted with the previously saved average value and added.
Parameter
Function
Weighted average
(F-PR)
Factor for average weighting. If you enter 100%, averaging will be off.
The smaller you set the F-PR factor, the stronger the weighting will be. When the averaging function is activated (F-PR <100%) a plausibility
check will be performed. The difference in temperature between the cur­rent reading and the previously stored average is determined. If this de­viation is greater than the plausibility threshold TSP, the transmitted da­ta will be „0“ and the average value will remain unchanged. Enter the upper threshold and lower threshold for the permissible devia­tion separately.
The parameter T.ACT= 0 automatically recognizes the end of the discontinuous process (measured temperature < L2). The pyrometer then shows "Auto" instead of the time under parameter T.ACT.
Page 29
s
Operating manual
21
Parameter
Function
Plausibility (TSP_)
Permitted temperature difference for a valid measurement when the new value is lower than the stored measured value.
Plausibility (TSP~)
Permitted temperature difference for a valid measurement when the new value is higher than the stored measured value.
If a measuring cycle does not start during the period T.OUT the saved average will be deleted and reinitialized when the next cycle begins.
Parameter
Function
Timeout
(T.OUT)
Time cycle for deleting average value (in minutes)
At the end of the measuring cycle the pyrometer displays the averaged temperature value (or the invalid reading „----“ ).
Cut-off interval [T.DIS]
A cut-off interval (time lag) begins after the measuring time has ended. This cut-off interval must expire before the next measurement can start with the cycle starting conditions described above.
Parameter
Function
Cut-off interval
(T.DIS)
The interval between one completed sam­pling and the start of a new sampling.
Timeout [T.OUT] If a measuring cycle does not start during the period T.OUT, the saved
average will be deleted and reinitialized when the next cycle begins.
Parameter
Function
Timeout (T.OUT):
Time out for average function (in minutes)
Autoreset Function [A.RST]
Activate auto reset for the ATD function to run cyclically. Limit 1 will then be ignored.
Normally, for a new measuring cycle to start, the detected temperature must have dropped below Limit 1. If, however, you wish to measure a
Page 30
s
Operating manual
22
continuous process (continuous material flow), you must activate Auto Reset. Limit 1 will then be ignored. Sampling will continue (restarting au­tomatically and producing temperature readings cyclically) as long as Limit 2 is exceeded for the time period defined by T.DEL.
Parameter
Function
Autoreset (A.RST):
Autoreset on/off
Parameter Set Li2 check on tAct [CH.L.2]
When this parameter is set to ON (CH.L.2=on), the detected tempera­ture must exceed Limit 2 during the entire measuring time (TACT) in order for the pyrometer to generate a valid temperature reading. The measurement is discarded when the value falls below Limit 2. The display shows „----
Parameter
Function
Set Li2 check on tAct (CH.L.2)
on/off
16 Setting Parameters at the device
In addition to the configuration possibilities described, many parameters can be adjusted at the rear panel using push buttons. These settings can be accessed via configuration layers.
16.1 Configuration layers
The configuration layers are structured as follows:
C001 Temperature measurement two-colour/ratio mode C002 Temperature measurement lambda 1 C003 Temperature measurement lambda 2
C010 I/O configuration (LED / Buzzer)
C011 General function
C020 Display temperature readings
The following chart lists all parameters. In the factory default configura­tion, certain parameters and configuration layers will not be accessible by control key. This is meant to simplify pyrometer operation for the user. These parameters are indicated by. If required, remote access to these parameters can be enabled via PC interface or at configuration layer
C011. Open the user calibration menu (Menu mode: Full).
Some parameter settings will be hidden if their basic function has been deactivated. For example, you will not be able to adjust the smoothing
Page 31
s
Operating manual
23
time of the filter if the filter has been turned off or switched to automatic mode.
16.1.1 Temperature measurement using two-colour/ratio mode
(Configuration layer: C001)
Parameter
Function
Explanation
E.TB.Q
Array size
Use array with 1-10 entries or enter material constant directly into the pyrometer
EPS.Q
Ratio correction
E.IDX
Memory position
Customize the array; assign a value to each of the ratio correction in the array. Possible indexes depend on the size of the array
E. 01
Assign ratio correction constants
Customize the array; assign a value to each of the ratio correction in the array. Possible indexes depend on the size of the array
CH.R.Q
Plausibility check ratio mode
OFF off MIN deactivation when below limit MI.MA. deactivation when below or above limit
CH.R._
Relative limit min.
Relative lower limit [%] , two-colour temp. reading invalid (signal intensity)
CH.R.~
Relative limit max.
Relative upper limit [%] , two-colour temp. reading invalid (signal intensity)
CH.A.T
Absolute min. temp.
Absolute lower limit, two-colour temp. reading invalid
CH.A.%
Absolute minimum Emissivity
Absolute lower limit [%], two-colour temp. reading invalid
LIN.Q
Temperature offset using linear
OFF off 2-10 number of nodes used
L. X1
node x 1 - 10
Signal input (initial value) node n
L. Y1
node y 1 - 10
Signal output (resulting value) node n
FIL.Q
Smoothing filter
OFF ON
FIL.T
Smoothing time
Time in seconds t98
MEM.Q
Min/Max memory
OFF off MIN lowest (min.)temperature, single MAX highest (max.) temperature, single DBL.M double maximum DIS.M ATD function
MEM.T
Hold time for Min/Max
Hold time in sec.
FIL.M
Smoothing filter for min/max*
OFF OFF ON ON
FIL.T
Smoothing time*
Time t98 in sec.
T.DEL
time delay **
For ATD function, see Chap. 15.2.4
T.ACT
Sampling time **
For ATD function, see Chap. 15.2.4
T.DIS
cut-off interval **
For ATD function, see Chap. 15.2.4
T.OUT
Timeout**
For ATD function, see Chap. 15.2.4
LI. 1
Limit 1**
For ATD function, see Chap. 15.2.4
LI. 2
Limit 2**
For ATD function, see Chap. 15.2.4
F-PR
Average weighting **
For ATD function, see Chap. 15.2.4
TSP~
Plausibility threshold **
For ATD function, see Chap. 15.2.4
Page 32
s
Operating manual
24
TSP~
Plausibility threshold **
For ATD function, see Chap. 15.2.4
MEM.T
Hold time for Min/Max
Hold time in sec.
FIL.M
Smoothing filter for min/max*
OFF OFF ON ON
FIL.T
Smoothing time*
Time t98 in sec.
END
End
Exit menu
*
Only available with Min/Max and Double Max modes
**
Only available with ATD function
16.1.2 Temperature measurement lambda 1
(Configuration layer: C002) In the factory default configuration, all parameter settings for digital in­puts and outputs will be hidden.
Parameter
Function
Explanation
E.TB.1
Array size
Use array with 1-10 entries or enter material constant directly into the pyrometer
EPS.1
Emissivity factor L1
Enter material constant directly into the py­rometer
E.IDX
Memory position
Choose an entry from the material constants array
E. 01
Assign material constants
Customize the array; assign a value to each of the material constants in the array. Possible indexes depend on the size of the array
TAU.1
Transmittance factor L1
BAC.1
Ambient temperature compensation
BAC.T
Temp. of ambient source of radiation
BAC.%
Influence of ambient IR radiation
The reflected thermal radiation from the sur­roundings as a portion of the total IR radiation collected by the sensor in %
LIN.1
Temperature offset using linear
OFF off 2-10 number of nodes used
L. X1
node x 1 - 10
Signal input (initial value) node n
L. Y1
node y 1 - 10
Signal output (resulting value) node n
FIL.1
Smoothing filter
OFF smoothing not activated ON simple smoothing
FIL.T
Smoothing time
time t98 in sec for simple smoothing
MEM.1
Min/Max memory
OFF off MIN lowest (min.)temperature, single MAX highest (max.) temperature, single DBL.M double maximum DIS.M ATD function
MEM.T
Hold time for Min/Max
Hold time in sec.
FIL.M
Smoothing filter for min/max*
OFF OFF ON ON
FIL.T
Smoothing time*
Time t98 in sec.
Page 33
s
Operating manual
25
T.DEL
time delay **
For ATD function, see Chap. 15.2.4
T.ACT
Sampling time **
For ATD function, see Chap. 15.2.4
T.DIS
cut-off interval **
For ATD function, see Chap. 15.2.4
T.OUT
Timeout**
For ATD function, see Chap. 15.2.4
LI. 1
Limit 1**
For ATD function, see Chap. 15.2.4
LI. 2
Limit 2**
For ATD function, see Chap. 15.2.4
F-PR
Average weighting **
For ATD function, see Chap. 15.2.4
TSP~
Plausibility threshold **
For ATD function, see Chap. 15.2.4
TSP~
Plausibility threshold **
For ATD function, see Chap. 15.2.4
ANO
Mode of display **
T=0 show lower limit of temp. range dur-
ing running measurement
T.HLD Hold previous temp. reading during
running measurement
A.RST
Auto reset**
For ATD function, see Chap. 15.2.4
CH.L.2
Set Li2 check on tAct**
For ATD function, see Chap. 15.2.4
END
End
Exit menu
*
Only available with Min/Max and Double Max modes
**
Only available with ATD function
16.1.3 Temperature measurement lambda 2
(Configuration layer: C003) By default, some parameters are suppressed to facilitate operation.
Parameter
Function
Explanation
E.TB.
Array size
Use array with 1-10 entries or enter material constant directly into the pyrometer
EPS.2
Emissivity factor L2
Enter material constant directly into the py­rometer
E.IDX
Memory position
Choose an entry from the material constants array
E. 01
Assign material constants
Customize the array; assign a value to each of the material constants in the array. Possible indexes depend on the size of the array
TAU.2
Transmittance factor L2
BAC.2
Ambient temperature compensation
BAC.T
Temp. of ambient source of radiation
BAC.%
Influence of ambient IR radiation
The reflected thermal radiation from the sur­roundings as a portion of the total IR radiation collected by the sensor in %
LIN.2
Temperature offset using linear
OFF off 2-10 number of nodes used
Page 34
s
Operating manual
26
L. X1
node x 1 - 10
Signal input (initial value) node n
L. Y1
node y 1 - 10
Signal output (resulting value) node n
FIL.2
Smoothing filter
OFF smoothing not activated ON simple smoothing
FIL.T
Smoothing time
time t98 in sec for simple smoothing
MEM.1
Min/Max memory
OFF off MIN lowest (min.)temperature, single MAX highest (max.) temperature, single DBL.M double maximum DIS.M ATD function
MEM.T
Hold time for Min/Max
Hold time in sec.
FIL.M
Smoothing filter for min/max*
OFF OFF ON ON
FIL.T
Smoothing time*
Time t98 in sec.
T.DEL
time delay **
For ATD function, see Chap. 15.2.4
T.ACT
Sampling time **
For ATD function, see Chap. 15.2.4
T.DIS
cut-off interval **
For ATD function, see Chap. 15.2.4
T.OUT
Timeout**
For ATD function, see Chap. 15.2.4
LI. 1
Limit 1**
For ATD function, see Chap. 15.2.4
LI. 2
Limit 2**
For ATD function, see Chap. 15.2.4
F-PR
Average weighting **
For ATD function, see Chap. 15.2.4
TSP~
Plausibility threshold **
For ATD function, see Chap. 15.2.4
TSP~
Plausibility threshold **
For ATD function, see Chap. 15.2.4
ANO
Mode of display **
T=0 show lower limit of temp. range dur-
ing running measurement
T.HLD Hold previous temp. reading during
running measurement
A.RST
Auto reset**
For ATD function, see Chap. 15.2.4
CH.L.2
Set Li2 check on tAct**
For ATD function, see Chap. 15.2.4
END
End
exit menu
*
Only available with Min/Max and Double Max modes
**
Only available with ATD function
L1 stands for lambda 1, meaning the temperature reading from lambda 1 L2 stands for lambda 2, meaning the temperature reading from lambda 2
Page 35
s
Operating manual
27
16.1.4 Configuration I/O
(Configuration layer: C010) In the factory default configuration, all parameter settings for digital in­puts and outputs will be hidden.
Parameter
Function
Explanation
TYPE
Operating mode/ measuring technique
Q ratio mode (quotient) L1 Lambda 1
L2 Lambda 2
LED.
LED (green)
OFF OFF ON ON
LED.S
LED: define source
RDY Status LED indicates ‘ready’ L1 Lambda 1 L1.PR Lambda 1 without peak picker L2 Lambda 2 L2.PR. Lambda 2 without peak picker Q Two-colour/ratio mode Q. PR. Two-colour/ratio mode without peak
picker
TU Inner device temperature 1NTY. Signal intensity M.TR.1 Triggered by ATD function Lambda 1* M.TR.2 Triggered by ATD function Lambda 2* M.TR.Q Triggered by ATD function two-
colour/ratio mode*
DIRT Dirt Alert A.AC.1 Measuring time ATD Lambda 1* A.AC.2 Measuring time ATD Lambda 2 * A.AC.9 Measuring time ATD
two colour/ratio mode *
LED.F
LED function
LVL. Switch direction “Level” (LED acti-
vated if limit exceeded)
LVL.- Switch direction “Level” (LED / out-
put inverted)
RN6. Switch direction “Range” (LED acti-
vated if range exceeded)
RN6- Switch direction “Range” (LED / out-
put inverted)
LED.T
LED switching threshold
Temperature limit (only available at function “Level”)
LED.H
LED signal threshold
Hysteresis +/- relative to signal threshold (only available at function “Level”)
LED._
LED lower limit of range
Lower limit of range for switch signal (only available function “range”)
LED.~
LED upper limit of range
Upper limit of range for switch signal (only available function “range”)
LED.L
LED delay time
LED.M
LED hold time
BU2.
Buzzer
OFF OFF ON ON
BUZ.S
Buzzer: define source
RDY Status LED indicates ‘ready’ L1 Lambda 1 L1.PR Lambda 1 without peak picker L2 Lambda 2 L2.PR. Lambda 2 without peak picker Q Two-colour/ratio mode Q. PR. Two-colour/ratio mode without peak
picker
Page 36
s
Operating manual
28
TU Inner device temperature 1NTY. Signal intensity M.TR.1 Triggered by ATD function Lambda 1* M.TR.2 Triggered by ATD function Lambda 2* M.TR.Q Triggered by ATD function two-
colour/ratio mode*
DIRT Dirt Alert A.AC.1 Measuring time ATD Lambda 1* A.AC.2 Measuring time ATD Lambda 2 * A.AC.9 Measuring time ATD
two colour/ratio mode *
BU2.F
Buzzer function
LVL. Switch direction “Level” (LED acti-
vated if limit exceeded)
LVL.- Switch direction “Level” (LED / out-
put inverted)
RN6. Switch direction “Range” (LED acti-
vated if range exceeded)
RN6- Switch direction “Range” (LED /
output inverted)
BU2.T
Buzzer switching threshold
Temperature limit (only available at function “Level”)
BU2.H
Buzzer signal threshold
Hysteresis +/- relative to signal threshold (only available at function “Level”)
BU2._
Buzzer lower limit of range
Lower limit of range for switch signal (only available function “range”)
BU2.~
Buzzer upper limit of range
Upper limit of range for switch signal (only available function “range”)
BUZ.L
Buzzer delay time
BU2.M
Buzzer hold time
END
End
exit menu
*
Only available with ATD function
16.1.5 General Functions
(Configuration layer: C011)
Parameter
Function
Explanation
A.OFF
Auto switch-off
OFF automatic switch-off function deac-
tivated
1-60 time period in minutes for auto
switch-off delay
A.STR.
Auto temp. data ouput
OFF automatic temp. data output is off ON temp. data output at PC terminal
A.CYC.
Cycle for auto temp. data output
Select cycle time in s
ADDR.
Device address
Enter address of device for non-terminal mode
DISP.
Display control
"ON" Display panel indicates "on" IA1 Indicated temperature reading ac-
cording to operating mode
UNIT
temperature unit
°( degrees Celsius °F degrees Fahrenheit
MENU
Menu-Mode
NORM With marked parameter are
shown
FULL All parameters are shown
END
End
exit menu
Page 37
s
Operating manual
29
16.1.6 Displayed temperature readings
(Configuration layer: C002)
Parameter
Function
Explanation
Q.
Temp. reading Quotient
Shows current temperature reading in two­colour mode
L1.
Temp. reading Lambda1
Actual temperature reading lL 1
L2.
Temp. reading Lambda1
Actual temperature reading L 2
Q. PR.
Temp. reading Quoti­ent Pre
Shows current temperature reading in two­colour/ratio mode prior to peak picker
L1.PR.
Temp. reading Lamb­da1 Pre
Shows current temperature reading for L1 prior to peak picker
L2.PR.
Temp. reading Lamb­da 2 Pre
Shows current temperature reading for L2 prior to peak picker
1NTY.
Signal-Intensity
Calculated signal intensity
T.INT.
Inner temperature
Current inner temp. of device
END
Ende
Menü verlassen
17 Software CellaView
The software CellaView can be used for representation, analysis and ar-
chiving of measured values of your pyrometer.
You can download the CellaView software here:
http://www.industry.siemens.com/services/global/de/sirent/verkauf/Documents/software/cellaview.zip
18 PC Interface
The Ardoptix features a USB port to enable data communication to a PC for remote device configuration and temperature monitoring. Either use the CellaView software or use another standard terminal program.
The USB port is on the back of the pyrometer, covered by a protective cap. The pyrometer is supplied with a standard USB cable.
A Windowsoperating system released prior to Windows7 will not au­tomatically identify the Ardoptix as the peripheral device. The required driver is included in the CellaView.zip or use the link www.prolific.com.tw to download the PL 2303 driver.
When using the CellaView software no additional parameter must be set. For more information on CellaView, please refer to the software instruc­tion manual.
Page 38
s
Operating manual
30
If you are using a terminal program such as HyperTerminal, set parame­ters for the serial interface as follows:
When using a terminal software, e.g. Hyperterminal, enter the parame­ters manually as follows:
57600 Baud / 8 Data bits / odd parity / 1 stop bit / no handshake
After enabling the interface and selecting parameter settings, the pyrom­eter will automatically transmit data via serial interface (autoprint activat­ed).
19 Remote configuration
Pyrometer setup and temperature monitoring can be configured from a PC running a simple serial terminal such as HyperTerminal. Many key parameters can be easily accessed from a keyboard. Additional func­tions and settings can be programmed from cascading submenus, as shown in the menu navigation chart below:
To set the pyrometer to the terminal mode, simultaneously hold down the Ctrl key and press the E key twice in rapid succession. A help menu will appear on the screen.
Main Menu
Submenu 1
Submenu 2
Execute the
command
Execute the
command
Execute the
command
press
[RETURN] or [ESC]
press
[RETURN] or [ESC]
press
[RETURN] Select [KEY]
Select [KEY]
Backspace
Backspace
Enter
Enter
Enter
[ESC]
[ESC]
Serial communication starts approximately 2 seconds after the data terminal ready (DTR) control signal is enabled at the inter­face. Activate this signal in the terminal program or device con­figuration
Page 39
s
Operating manual
31
Direct commands have an assigned key. Example: E for epsilon (ratio correction). Submenu settings are shown in brackets. Exam­ple[Quotient]
19.1 Main Menu
After starting the terminal program or after entering „H“ the main menu will appear
-----------------------------------------------------------------­>H
-----------------------------------------------------------------­Mainmenu
-----------------------------------------------------------------­0: [QUOTIENT] E: Quick access EPSILON 1: [LAMBDA 1] A: Quick access FILTER 2: [LAMBDA 2] T: Quick access TYPE of measure C: [I/O] K: [CALIBRATION]
H: Show this help-site J: Show diagnosis W: Show ambient temperature Q: Show calibration data X: Show measure temperatures P: Show channel parameters
------------------------------------------------------------------
19.2 View Current Configuration
Command „P“ shows you how your Ardoptix PT is currently configured.
------------------------------------------------------------------
- PT 143 AF1 650-1700C - 00/00112 - Job - 21.10.13 -
- PT40SW101/0 QP 0,95/1,05um Version 01.02 10.07.13 -
------------------------------------------------------------------
Qu range .... 650.0 - 1700.0 C DISPLAY source ....... quotient
Qu epsilon ratio ...... 100.0 %
Qu check L2 rel.limit 10.00 % Qu abs.limit 650 C @ 50.00 %
Qu linearization .......... off
Qu filter .......... 0.10 s
Qu memory type ............ off GRN.LED source ... ready-signal
GRN.LED function level/signal
Unit .................. Celsius GRN.LED delay time ... 0.00 s
Terminal assigned to ...... USB GRN.LED hold time .... 0.00 s
Autoprint ......... on (cyclic) BUZZER source ............ off
Print cycle time ..... 0.1 s
Protocol address .......... 001
Display ........... temperature
Key lock .................. off
-----------------------------------------------------------------­>
At the top left, there is a list of data acquisition parameters for the two colour reading (Quotient). The right-side column shows LED/Buzzer­Configuration. At the bottom left you will find general settings.
Page 40
s
Operating manual
32
19.3 Submenus
19.3.1 Configure two-colour/ratio mode
Press key "0“ to access the data acquisition parameters for the quotient.
-----------------------------------------------------------------­Submenu QUOTIENT
------------------------------------------------------------------
Qu epsilon ratio ...... 100.0 %
Qu check L2 rel.limit 10.00 % Qu abs.limit 650 C @ 50.00 %
Qu linearization .......... off
Qu filter .......... 0.10 s
Qu memory type ............ off
C: [CONFIG EPSILON TABLE] E: Epsilon U: [Q-CHECK] L: [LINEARIZATION] F: Filter M: [MEMORY] P: Show parameter Q: Show calibration data O: Show signal intensity X: Show measure temperatures Y: Show premax measure temps. ESC: Back to MAIN-MENU
-----------------------------------------------------------------­>QUOTIENT >
19.3.2 Lambda 1
Press key „1“ to access all data acquisition parameters for Lambda 1.
-----------------------------------------------------------------­Submenu LAMBDA 1
------------------------------------------------------------------
L1 epsilon ............ 99.0 %
L1 transmission ....... 100.0 %
L1 backc. ................. off
L1 linearization .......... off
L1 filter .......... 0.10 s
L1 memory type ............ off
C: [CONFIG EPSILON TABLE] E: Epsilon T: Transmission B: Background-Compensation L: [LINEARIZATION] F: Filter M: [MEMORY] P: Show parameter Q: Show calibration data X: Show measure temperatures Y: Show premax measure temps. ESC: Back to MAIN-MENU
-----------------------------------------------------------------­>LAMBDA 1 >
Page 41
s
Operating manual
33
19.3.3 Lambda 2
Press key „2“ to access all data acquistion parameters for Lambda 2.
-----------------------------------------------------------------­Submenu LAMBDA 2
------------------------------------------------------------------
L2 epsilon ............ 99.0 %
L2 transmission ....... 100.0 %
L2 backc. ................. off
L2 linearization .......... off
L2 filter .......... 0.10 s
L2 memory type ............ off
C: [CONFIG EPSILON TABLE] E: Epsilon T: Transmission B: Background-Compensation L: [LINEARIZATION] F: Filter M: [MEMORY] P: Show parameter Q: Show calibration data X: Show measure temperatures Y: Show premax measure temps. ESC: Back to MAIN-MENU
-----------------------------------------------------------------­>LAMBDA 2 >
19.3.4 Quick access to emissivity / signal smoothing / operating
mode
Use keys "E", "T", „B“ und "F" to access and change the emissivity set-
ting, smoothing filter and measuring mode.
19.3.5 I/O Signal Configuration
Press „C“ to access and adjust all settings regarding LED and buzzer.
-----------------------------------------------------------------­Submenu I/O
-----------------------------------------------------------------­C: [STATUS LED CONTROL] D: [BUZZER CONTROL] M: [OPTIONS] ESC: Back to MAIN-MENU
-----------------------------------------------------------------­>I/O >
The LED and buzzer configuration are grouped in submenus where each can be accessed for further configuration.
LED-Control:
Page 42
s
Operating manual
34
-----------------------------------------------------------------­Submenu CONTROL LED
-----------------------------------------------------------------­GRN.LED source ... ready-signal GRN.LED function level/signal GRN.LED delay time ... 0.00 s
GRN.LED hold time .... 0.00 s
S: Set source F: Set function D: Set delay time O: Set hold time ESC: Back to MAIN-MENU
-----------------------------------------------------------------­>I/O >LED CONTROL >
Set status LED source: 0: Off 1: Ready-Signal 2: Lambda 1 3: Lambda 1 premax 4: Lambda 2 5: Lambda 2 premax 6: Quotient 7: Quotient premax 8: Signal intensity 9: Dirt Alert 10: Ambient Temperature 11: Lambda 1 ATD Trigger 12: Lambda 2 ATD Trigger 13: Quotient ATD Trigger 14: Lambda 1 ATD tAct 15: Lambda 2 ATD tAct 16: Quotient ATD tAct
-----------------------------------------------------------------­Your choice>
Buzzer-control:
-----------------------------------------------------------------­Submenu CONTROL 2
------------------------------------------------------------------
BUZZER source ............ off
S: Set source ESC: Back to MAIN-MENU
-----------------------------------------------------------------­>I/O >BUZZER CONTROL >S
Set buzzer control source: 0: Off 1: Ready-Signal 2: Lambda 1 3: Lambda 1 premax 4: Lambda 2 5: Lambda 2 premax 6: Quotient 7: Quotient premax 8: Signal intensity 9: Dirt Alert 10: Ambient Temperature 11: Lambda 1 ATD Trigger 12: Lambda 2 ATD Trigger
Page 43
s
Operating manual
35
13: Quotient ATD Trigger 14: Lambda 1 ATD tAct 15: Lambda 2 ATD tAct 16: Quotient ATD tAct
-----------------------------------------------------------------­Your choice>
In the submenu "Options" you can program a keylock. If keylock is acti­vated and you wish to change a parameter setting, enter Code P100 for full access. If you enter an incorrect code, you can view the configura­tion but you will not be able to change the setting.
-----------------------------------------------------------------­Submenu OPTIONS
------------------------------------------------------------------
Autoprint ......... on (cyclic)
Print cycle time ..... 0.1 s
Protocol address .......... 001
Display ........... temperature
Key lock .................. off
Unit .................. Celsius
A: Set autoprint function T: Set output cycle time P: Set protocol-address D: Set display function E: Set key lock F: Set unit Celsius/Fahrenheit R: Switch off pyrometer ESC: Back to MAIN-MENU
-----------------------------------------------------------------­>I/O >OPTIONS >
19.3.6 Automatic temperature data output
For continuous serial transmission of temperature data, you must acti­vate automatic temperature data output.
In the submenu “Options”, command A" activates (on) / deactivates (off) this function.
If ATD is activated, the temperature data will only be transmitted after successful completion of measurement.
If ATD is deactivated, press command "T" to select the cycle time in which the current temperature data are transmitted through the serial in­terface.
When automatic temperature data output is activated, the pyrometer will not display the parameter settings when it is switched on. Instead, the pyrometer will instantly begin transmitting data.
Page 44
s
Operating manual
36
Temperature data format
two-colour/ratio Lambda 1 Lambda 2 (one cycle):
Byte
Negative Temperature
Positive Temperature
Temperature exceeds measuring range
Temperature falls below measuring range
1
Space
Space
Space
Space
2
Minus symbol -
Space
Minus symbol -
Minus symbol -
3
Digit 1000
Digit 1000 O U
4
Digit 100
Digit 100 V N
5
Digit 10
Digit 10 E D
6
Digit 1
Digit 1 R E
7
Decimal point .
Decimal point .
Space
R
8
Decimal place
Decimal place
Space
Space
9
Space
Space
Minus symbol -
Minus symbol -
10
Unit C or F
Unit C or F
Space
Space
11
Tabulator
Tabulator
Tabulator
Tabulator
12
Space
Space
Space
Space
13
Minuszeichen -
Space
Minus symbol -
Minus symbol -
14
Digit 1000
Digit 1000 O U
15
Digit 100
Digit 100 V N
16
Digit 10
Digit 10 E D
17
Digit 1
Digit 1 R E
18
Decimal point .
Decimal point .
Space
R
19
Decimal place
Decimal place
Space
Space
20
Space
Space
Minus symbol -
Minus symbol -
21
Unit C or F
Unit C or F
Space
Space
22
Tabulator
Tabulator
Tabulator
Tabulator
23
Space
Space
Space
Space
24
Minuszeichen -
Space
Minus symbol -
Minus symbol -
25
Digit 1000
Digit 1000 O U
26
Digit 100
Digit 100 V N
27
Digit 10
Digit 10 E D
28
Digit 1
Digit 1 R E
29
Decimal point .
Decimal point .
Space
R
30
Decimal place
Decimal place
Space
Space
31
Space
Space
Minus symbol -
Minus symbol -
32
Unit C or F
Unit C or F
Space
Space
33
Carriage Return
Carriage Return
Carriage Return
Carriage Return
The cycle time in which the temperature reading is transmitted can be set via parameter ACYC (minimum cycle duration is 0.1 second).
19.4 User Recalibration
If ever required, the pyrometer can be recalibrated. Go to Submenu Cali­bration. Press command „K“ and then enter the password „100“ to ac- cess the calibration menu.
All signs/symbols are ASCII coded; preceding zeros will be transmitted
Page 45
s
Operating manual
37
------------------------------------------------------------------
-----------------------------------------------------------------­Submenu CALIBRATION
------------------------------------------------------------------
Name .... "Pyrometer PT Series"
Menu mode ............. default
1: [LAMBDA 1 CALIBRATION] A: Reset settings to factory default E: Set menu mode S: Set pyrometer name Z: End Calibration-Mode ESC: Back to MAIN-MENU
-----------------------------------------------------------------­>CALIBRATION >
You can reset all configurations you have made to your pyrometer and restore the factory default settings using command "A". This also applies to data acquisition parameters, LED and buzzer.
Press command "E" if you would like to view all parameters which are otherwise hidden. Because this list of menu items is quite long (Chapter
16), we only recommend this for advanced users. If you intend to carry out measurements of molten metal only, it makes sense to maintain the factory configuration. This will keep those parameters, which will not re­quire adjustment, hidden
Command "S" enables the user to enter a short word to designate the measurement location/task. This text can be accessed from the Main Menu by pressing "Q".
-----------------------------------------------------------------­Submenu LAMBDA 1
------------------------------------------------------------------
L1 range .... 0.0 - 1000.0 C
L1 User calibration ....... off
L1 User def. offset +0.00000 L1 User def. factor +1.00000
A: Set L1 - extended-range B: Set L1 User-Cal. On/Off ESC: Back to MAIN-MENU
-----------------------------------------------------------------­>CALIBRATION >LAMBDA 1 >
Use command "A" to reconfigure the pyrometer‘s measuring range. The range can be adjusted to be either larger or smaller than the factory­configured range. When customizing the temperature range, make sure that upper and lower temperatures you enter are compatible with the py­rometer’s actual working range.
Use key "B" for direct access to enable adjustments to the Ardoptix PT.
Page 46
s
Operating manual
38
If you make a mistake while making the adjustments, simply enter off­set=0.0 und factor=1.0, or set User Calibration to „off“.
20 Maintenance
20.1 Cleaning the pyrometer lens/protective glass
A false temperature reading will be generated when the lens/protective glass is dirty. Therefore check the lens periodically and clean it, if neces­sary.
Dust can be removed by simply blowing it away or by using a soft brush. A special lens cleaning cloth is ideal, but any soft, clean, lint-free cloth will be suitable.
If the lens is quite dirty, use a very mild liquid detergent and rinse careful­ly with clear water while holding the pyrometer down. Apply as little pres­sure as possible to avoid scratching the lens.
Make sure to turn off the pyrometer before removing or attaching the lens. Failure to do so may result in damage to the instrument!
T
actual
Actual temperature Influence of factor (slope) Influence of offset
T
target
User calibration Pyrometer PT
To recalibrate your Ardoptix, you will require a calibration fur­nace and a reference standard
Protect the pyrometer against high ambient temperatures, high air humidity, high voltage and strong electromagnetic fields. Never point the lens directly towards the sun
Page 47
s
Operating manual
39
21 Accessories
Description
Type
Item No.
Protective glass M46
70146
120314
USB Cable
VK 11/D
1009677
Power supply
1053975
Carrying case
PT 110/A
1052289
Supplementary lens
PZ 20/O-50
514744
Supplementary lens
PZ 20/O-63
514985
Supplementary lens
PZ 20/O-75
513840
Supplementary lens
PZ 20/O-120
514973
21.1 Supplementary lens
Pyrometer
Supplementary lens
Type
PZ 20/O-50
PZ 20/O-63
PZ 20/O-75
PZ 20/O-120
distance
[mm]
spot size
Ø in mm
distance
[mm]
spot size
Ø in mm
distance
[mm]
spot size
Ø in mm
distance
[mm]
spot size
Ø in mm
PT 140 AF 1
36-41
0.65-0.9
45 - 54
0.8 – 1,1
52 - 63
1.9 – 0.4
84 - 112
1.7 – 2.1
PT 140 AF 4
36-41
0.35-0.45
45 - 54
0.4 – 0.6
52 - 63
0.45 – 0.7
84 - 112
0.7 – 1.1
Pyrometer
Supplementary lens
Type
PZ 20/O-50
PZ 20/O-63
PZ 20/O-75
PZ 20/O-120
distance
[mm]
Mess-
fleck
in mm
distance
[mm]
spot size
in mm
distance
[mm]
spot size
in mm
distance
[mm]
spot size
in mm
PT 143 AF 1
h
36-41
1.2 – 1.6
45 - 54
1.4 – 1.2
52 - 63
1.6 – 2.4
84 - 112
2.4 – 3.8
v
0.2 – 0.3
0.4 – 0.4
0.5
0.5 – 0.8
PT 143 AF 4
h
35 - 41
1.0 – 1.4
45 - 54
1.3 – 1.8
52 - 63
1.4 – 2.1
84 - 112
2.1 – 3.4
v
0.2
0.2 – 0.3
0.3 – 0.4
0.4 – 0.6
Page 48
s
Operating manual
40
22 Theory of Non-Contact Temperature Measurements
All materials radiate thermal energy in all states of aggregation above absolute zero. This radiation is mainly caused by atomic or molecular oscillations. The intensity of this infrared radiation is a measure for the temperature of the radiating body. A pyrometer detects this radiation and displays it as a temperature value.
22.1 Emissivity
This coefficient lies between 0...100 %. A body with ideal radiation (black body) has a coefficient of 100 %. Materials with the same temperature but with lower radiation intensity have coefficients lower than 100 %. Ex­act temperature readings can only be obtained when the emissivity coef­ficient of the target is correctly adjusted on the pyrometer. The pyrometer then automatically compensates for the lower radiation.
For non-contact measurements the Ardoptix uses the intensity of the in- frared radiation at a certain wavelength. It is necessary to configure the Ardoptix to the respective emissivity coef- ficient of the measuring object. An incorrectly set emissivity coefficient leads to wrong temperature readings. Below is a table showing emissivi­ty coefficients for various materials.
Page 49
s
Operating manual
41
22.2 Temperature measurement using spectral mode
For non-contact measurements the Ardoptix uses the intensity of the in­frared radiation at one certain wavelength. It is necessary to configure the Ardoptix to the respective emissivity coef­ficient of the measuring object to get exact measuring results. An incor­rectly set emissivity coefficient leads to wrong temperature readings.
In pyrometry, the specific radiation characteristics of the measured object will influence the generated temperature data. In order to produce accu­rate and reliable temperature readings, the pyrometer must be config­ured according to the specific material constant (emissivity).
The annex includes a table showing emissivity coefficients for various materials.
22.3 Temperature measurement using two-colour/ratio mode
The quotient channel of the Ardoptix measures the infrared radiation at two wavelengths and calculates the ratio of these two radiation values. This quotient (or this ratio) of the two intensities is proportional to the temperature. The ratio remains constant even when the radiation intensi­ties decrease, e.g. with vapour and dust in the field of view, fogged-up lenses or changing surface properties (emissivity coefficient) of the measured object. Therefore, with a homogeneous weakening of the sig­nal, the two-colour pyrometer still provides a stable measuring signal. If weakening of the radiation is not homogeneous or depending on the wavelengths, the ratio of the two radiation intensities is not constant: the emissivity ratio changes. This emissivity ratio is the correction factor that is set at the pyrometer and that needs to be adapted to the meas­ured object when measuring with a two-colour pyrometer.
Resumé: The two-colour measuring method offers great advantages due to its in­sensivity to the described disturbances. The applications cover large segments of the iron and steel producing industry and the metal, cement and chemical industries.
Page 50
s
Operating manual
42
22.4 Emissivity Coefficient Table – Spectral Mode
List of emissivity coefficients of different materials in %
PT 140
Wavelength
0.8...1.1 m
"Black Body"
100
Aluminium, polished
15
Aluminium, blackened
25
Asbestos cement
70
Bronce, polished
3
Bronze, blackened
30
Chromium, polished
30
Iron, heavily scaled
95
Iron, rolling skin
90
Iron, liquid
30
Gold and silver
2
Graphite, blackened
90
Copper, oxidized
90
Brass, oxidized (tarnished)
70
Nickel
20
Porcelain, glazed
60
Porcelain, rough
85
Soot
95
Fireclay
50
Slag
85
Pottery, glazed
90
Bricks
90
Zinc
60
Page 51
s
Operating manual
43
23 General technical data
Digital output:
Cyclical data output, adjustable cycle time
Resolution Display:
1 K
Resolution USB:
0.1 K
Power supply
Rechargeable battery pack Mains adapter
Battery-life
Up to 12 hours of use (based on continuous operating at Ta = 23 °C
Permissible humidity
95% r.H. max. (non-condensing)
Ambient operating tempera­ture:
0 - 50 °C
Storage temperature:
-20 - 50 °C
Housing:
Aluminium
Protection rating:
IP 40 nach DIN 40050
Weight:
Approx.. 1.1 kg
Sighting
Through-the-lens sighting, par­allax free imaging, target spot indicator, diopter compensa­tion, polarizing filter
ATD function
automatically identifies the target and detects its tempera­ture
24 Technical Data PT 140 (650 – 1700 °C)
Measuring range:
650 ... 1700 °C
Sensors:
photo diode
Spectral sensitivity:
0.95/ 1.05 µm
Response time t98:
10 ms (T> 750 °C)
Measuring uncertainty :
1 % of range (at e =1.0 and T
A
= 23 °C)
Repeatability:
2 K
Temperature coefficient with reference to 23 °C:
0.05 %/K of measured value
24.1 Field of View Diagrams PT 140 (650 – 1700 °C)
PT 140
Lens
Focussing range
Distance to target size ratio
AF 1
PZ 20.01
0.4 m -
80:1
Page 52
s
Operating manual
44
1
2 3 4 5 6
7 8 9
10
13
25
38
50
63
75
88
100
113
125
D = 80 :1
Target distance [m]
Target diameter [mm]
0
PZ 20.01 (0.4 m - ∞)
Page 53
s
Operating manual
45
25 Technical Data PT 140 (750 – 2400 °C)
Measuring range:
750 … 2400 °C
Sensors:
photo diode
Spectral sensitivity:
0.95/ 1.05 µm
Response time t98: 10 ms (T> 950 °C)
Measuring uncertainty:
1 % of range (at e =1.0 and T
A
= 23 °C)
Repeatability:
2 K
Temperature coefficient with reference to 23 °C:
0.05 %/K of measured value
25.1 Field of View Diagrams PT 140 (750 – 2400 °C)
PT 140
Lens
Focussing range
Distance to target size ratio
AF 4
PZ 20.01
0.4 m -
150:1
1
2 3 4
5 6 7 8 9
10
6,7
13,3
20
26,7
33,3
40
46,7
53,3
60
66,7
D = 150 :1
Target distance [m]
Target diameter [mm]
0
PZ 20.01 (0.4 m - ∞)
2,7
Page 54
s
Operating manual
46
26 Technical Data PT 143 (650 – 1700 °C)
Measuring range:
650 ... 1700 °C
Sensors:
photo diode
Spectral sensitivity:
0.95/ 1.05 µm
Response time t98: 10 ms
Measuring uncertainty:
1,5 % of range and but at least 4 K (at e =1.0 and T
A
=
23 °C)
Repeatability:
3 K
Temperature coefficient with reference to 23 °C:
0.05 %/K of measured value
26.1 Field of View Diagrams PT 143 (650 - 1700°C)
PT 143
Lens
Focussing range
Distance to target size ratio
AF 1
PZ 20.01
0.4 m -
DV= 230:1 DH= 45:1
222
44
39
200
39
178
30
156
26
133
22
111
17
89
13
67 9 44 4 22 9 2
Target diameter [mm]
Target distance [m]
0,4
1 2 3 4 5 6 7 8 9
10
PZ 20.01 (0.4 m - ∞)
Page 55
s
Operating manual
47
27 Technische Daten PT 143 (MB 750 - 2400°C)
Measuring range:
750 ... 2400 °C
Sensors:
photo diode
Focussing range:
0.95/ 1.05 µm
Response time t98: 10 ms
Measuring uncertainty:
1.5 % of measured value (at e =1.0 and T
A
= 23 °C)
Repeatability:
3 K
Temperature coefficient:
0.05 %/K of measured value [°C] Deviation to TA= 23 °C
27.1 Field of View Diagrams PT 143 (MB 750 - 2400°C)
PT 143
Lens
Focussing range
Distance to target size ratio
AF 4
PZ 20.01
0.4 m -
DV= 350:1 DH= 50:1
200
29
26
180
23
160
23
140
17
120
14
100
11
80
9
60
6
40 3 20
8
1
Target distance [mm]
Target distance [m]
0,4 1 2 3 4
5 6 7 8 9
10
PZ 20.01 (0,4 m - ∞)
Page 56
s
Operating manual
48
29 Dimensions
Page 57
s
Operating manual
49
30 Shipping, Packaging and Disposal
30.1 Inspecting your shipment
Unpack and inspect the entire shipment immediately upon receipt to make sure it is complete and undamaged.
If the container/package shows visible signs of damage, please refuse the shipment. If this is not possible, accept the shipment on the condition that the freight carrier’s delivery record is noted with the extent of the damage in order to file a claim.
Should you discover a concealed loss or damage, report it to SIEMENS and to the freight carrier immediately. If the period for filing claims has expired, you will no longer be able to make any claims for compensation of damage or loss.
30.2 Packaging
The packages used by SIEMENS are made of carefully selected, envi­ronmentally compatible materials and are thus recyclable. We suggest you retain the packaging for possible future use; otherwise please en­sure that they are disposed of in an ecologically sound manner.
30.3 Disposal of used apparatus
Used electrical and electronic equipment often contain valuable compo­nents. The owner/user may either return such an instrument to the manufacturer for disposal, or he must dispose of it himself in a profes­sional and nonpolluting manner.
SIEMENS will not be held accountable for any inappropriate disposal carried out by the user/owner of SIEMENS instruments.
Page 58
s
Operating manual
50
31 Copyright
Portions of avr-libc are Copyright (c) 1999-2010 Werner Boellmann, Dean Camera, Pieter Conradie, Brian Dean, Keith Gudger, Wouter van Gulik, Bjoern Haase, Steinar Haugen, Peter Jansen, Reinhard Jessich, Magnus Johansson, Harald Kipp, Carlos Lamas, Cliff Lawson, Artur Lipowski, Marek Michalkiewicz, Todd C. Miller, Rich Neswold, Colin O’Flynn, Bob Paddock, Andrey Pashchenko, Reiner Patommel, Florin-Viorel Petrov, Alexander Popov, Michael Rickman, Theodore A. Roth, Juergen Schilling, Philip Soeberg, Anatoly Sokolov, Nils Kristian Strom, Michael Stumpf, Stefan Swanepoel, Helmut Wallner, Eric B. Weddington, Joerg Wunsch, Dmitry Xmelkov, Atmel Corporation, egnite Software GmbH, The Regents of the University of California. All rights reserved.
* Redistributions of source code must retain the above copyright notice, this list of conditions and the following disclaimer.
* Redistributions in binary form must reproduce the above copyright notice, this list of conditions and the following disclaimer in the documentation and/or other materials provided with the distribution.
* Neither the name of the copyright holders nor the names of contributors may be used to endorse or promote products derived from this software without specific prior written permission.
THIS SOFTWARE IS PROVIDED BY THE COPYRIGHT HOLDERS AND CONTRIBUTORS "AS IS"
AND ANY EXPRESS OR IMPLIED WARRANTIES, INCLUDING, BUT NOT LIMITED TO, THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS FOR A PARTICULAR PURPOSE ARE DISCLAIMED. IN NO EVENT SHALL THE COPYRIGHT OWNER OR CONTRIBUTORS BE LIABLE FOR ANY DIRECT, INDIRECT, INCIDENTAL, SPECIAL, EXEMPLARY, OR CONSEQUENTIAL DAMAGES (INCLUDING, BUT NOT LIMITED TO, PROCUREMENT OF SUBSTITUTE GOODS OR SERVICES; LOSS OF USE, DATA, OR PROFITS; OR BUSINESS INTERRUPTION) HOWEVER CAUSED AND ON ANY THEORY OF LIABILITY, WHETHER IN CONTRACT, STRICT LIABILITY, OR TORT (INCLUDING NEGLIGENCE OR OTHERWISE) ARISING IN ANY WAY OUT OF THE USE OF THIS SOFTWARE, EVEN IF ADVISED OF THE
POSSIBILITY OF SUCH DAMAGE.
Page 59
s
Operating manual
51
32 Default settings
32.1 Temperature measurement (Configuration layer: C001)
Function
Parameter configuration layer C001
Default
Customized confi-
guration
Array size
E.TB.1
OFF
1-10
OFF
Emissionsgrad Verhältnis
EPS.1
100
Memory posi­tion
E.IDX
1
Material constant 1
E. 01
100.0
Material constant 2
E. 02
100.0
Mode Q­Check
CHR.Q.
OFF
MIN
MI.MA
MIN
Relatives Limit Min.
CH.R._
CH.R._
10.0
Relatives Limit Max.
CH.R.~
100.0
Absolutes Minimum Temperatur
CH.A.T
MB*
Absolutes Minium Epsi­lon
CH.A.%
50
Smoothing filter
FIL.1
OFF
ON
ON
Smoothing time
FIL.T
0.10
Min/Max memory
MEM.1
OFF
MIN MAX
DBL.N
ATD
OFF
Smoothing time
MEM.T
1.00
Smoothing filter for min/max Smoothing time
FIL.M
FIL.T
FIL.M
FIL.T
OFF
0.10
T.DEL
1.0
T.ACT
3.0
T.DIS
0.0
T.OUT
1.0
LI. 1
MR*+50 °C
LI. 2
MR*+50 °C
F-PR
100.0
TSP~
200
TSP~
200
ANO
T.HLD.
A.RST
OFF
CH.L.2
OFF
END
* Start of measuring range
Page 60
s
Operating manual
52
32.2 Temperature measurement Lambda 1 /2(Configuration layer:
C002)
Function
Parameter configuration layer C002
Default
Customized confi-
guration
Array size
E.TB.1
OFF
1-10
OFF
Emissivity
EPS.1
99.6
Memory posi­tion
E.IDX
1
Material constant 1
E. 01
100.0
Material constant 2
E. 02
100.0
Transmission factor
TAU.
100.0
Smoothing filter
FIL.1
OFF
ON
Smoothing time
FIL.T
4.00
Min/Max memory
MEM.1
OFF
MIN MAX
DBL.N
ATD
OFF
Smoothing time
MEM.T
1.00
Smoothing filter for min/max Smoothing time
FIL.M
FIL.T
FIL.M
FIL.T
OFF
0.10
T.DEL
1.0
T.ACT
3.0
T.DIS
0.0
T.OUT
1.0
LI. 1
MR*+50 °C
LI. 2
MR*+50 °C
F-PR
100.0
TSP~
200
TSP~
200
ANO
T.HLD.
A.RST
OFF
CH.L.2
OFF
END
* Start of measuring range
Page 61
s
Operating manual
53
32.3 Temperature measurement Lambda 2(Configuration layer:
C003)
Function
Parameter Configuration layer C001
Default
Eigene
Einstellungen
Array size
E.TB.1
OFF
1-10
OFF
Emissivity
EPS.1
99.6
Memory posi­tion
E.IDX
1
Material constant 1
E. 01
100.0
Material constant 2
E. 02
100.0
Transmission factor
TAU.
100.0
Smoothing filter
FIL.1
OFF
ON
Smoothing time
FIL.T
4.00
Min/Max memory
MEM.1
OFF
MIN MAX
DBL.N
ATD
OFF
Smoothing time
MEM.T
1.00
Smoothing filter for min/max Smoothing time
FIL.M
FIL.T
FIL.M
FIL.T
OFF
0.10
T.DEL
1.0
T.ACT
3.0
T.DIS
0.0
T.OUT
1.0
LI. 1
MR*+50 °C
LI. 2
MR*+50 °C
F-PR
100.0
TSP~
200
TSP~
200
ANO
T.HLD.
A.RST
OFF
CH.L.2
OFF
End
END
* Start of measuring range
32.4 General Functions (Configuration layer: C010)
Function
Parameter
Default
Customized con­figuration
Operating mode
TYPE
Q Quotient
End
END
Page 62
s
Operating manual
54
32.5 General Functions (Configuration layer: C011)
Function
Parameter
Default
Customized configu­ration
Automatic Switch off
A.OFF
2
Temperature unit
UNIT
°(
End
END
Page 63
s
Operating manual
55
Page 64
Page 65
Page 66
Page 67
Page 68
Siemens AG
Digital Factory Customer Services DF&PD DF CS SD OP ITM Telefon +49 (911) 750-9600 Fax +49 (911) 750-9609 Email: sirent.industry@siemens.com
Office/Testing
Warehouse/Counter service
Breslauer Str. 5
Gundelfinger Str. 20
90766 Fürth
90451 Nürnberg
Germany
Germany
Subject to change without notice
Loading...