Seca 434, 435 Service Manual

4340000009
30.08.03
4350000009
for seca
Description:
2. Display modul for bed scale mod. 984 / 985
434, 435
Content:
Operation 17-10-07-371 c Mounting instruction f. 2. Display 17-10-07-488 Replacement 30-34-00-710
Service Manual Number:
17-05-01-319-
Valid as of:
Spare parts 30-34-00-709
Manual number: 17-05-01-319-
Service Manual Replacement
seca 434, 435
1 Before starting any work on the scale, interrupt the power supply
2 Opening the display housing
2.1 Release 4 screws (32) on the bottom side of the display housing.
2.2 Turn the display housing over and carefully lift off the upper part.
2.3 Unsolder the connecting lug for the membrane keyboard (49) from the display electronics (50).
3 The individual display components can now be replaced.
observe the instuctions given when replacing the following parts:
3.1 Display module (50). The display module does not contain any individual data. It can be
exchanged for any type of display module! Resolder the cable conections one by one to the new board. Do not
confuse the cable conections.
3.2 Cover (33). the membrane keyboard must also be replaced.
4 Assemble in reverse order.