PB-500 Electronic Balance
MODE key
COUNT key
TARE key
ON/OFF key
User Instructions
Keys
1

If no key is pressed within 120 seconds the balance
switches itself off.
Select unit of measure
Repeatedly press the MODE key to scroll through the unit choices: g, oz, dwt,
and ozt.
To zero the balance
Empty the scale and press the TAR E key.
Weighing
Place the item on the balance and the weight is displayed in the current unit of
measure.
Pushbutton tare
1. Zero the scale.
2. Place container to be tared on the balance and press TARE.
3. Place item(s) to be weighed in the container or place the container,
with the items already inside, on the balance.
Net weighed is displayed.
To remove a tare
Empty the balance and press TARE.
Tare is removed.
2

Counting
1. Be sure the balance is on and the display shows 0.0. Place the sample
size you want (25, 50, 75 or 100 items) on the balance.
2. Press and hold COUNT until …
100 is displayed. This is the default sample size.
3a. Press COUNT to accept this sample size …
OR
3b. Press MODE to change the sample size and press COUNT when the
sample size you want is displayed …
The display will show 100pcs (or 25pcs, 50pcs, 75pcs).
4. Replace items on the scale to continue counting.
5. Press MODE to return to normal mode.
Calibration
1. Be sure the balance is on and the display shows 0.0.
2. Press and hold MODE until …
CAL is displayed.
3. Press MODE …
CAL blinks then shows 500g blinking.
4. Place 500g of weight on the balance …
After few seconds, PASS is briefly displayed, then 500g.
5. The calibration is complete and you are ready to weigh.
3

© Salter Brecknell Weighing 2012. All rights reserved.
The information contained herein is the property of Salter Brecknell Weighing
and is supplied without liability for errors or omissions. No part may be
reproduced or used except as authorised by contract or other written
permission. The copyright and the foregoing restriction on reproduction and
use extend to all media in which the information may be embodied.
Brecknell
1000 Armstrong Dr., Fairmont, MN 56031
Toll free: 800-637-0529, Tel: 507-238-8702
Fax: 507-238-8271
Email: sales@brecknellscales.com
www.brecknellscales.com
Brecknell
Foundry Lane
Smethwick, West Midlands B66 2LP
Tel: +44 (0) 845 246 6717
Fax: +44 (0) 845 246 6718
Email: sales@brecknellscales.co.uk
www.brecknellscales.co.uk
4
April 2012 PB500_u_en_500913.fm AWT35-500913 Issue AA