Rosemount FCL Instruction Manual

Instruction Manual
LIQ-MAN-56, Rev B July 2017
FCL with 56
ESSENTIAL INSTRUCTIONS
READ THIS PAGE BEFORE PROCEEDING!
Your purchase from Rosemount, Inc. has resulted in one of the finest instruments available for your particular application. These instruments have been designed, and tested to meet many national and international standards. Experience indicates that its performance is directly related to the quality of the installation and knowledge of the user in oper­ating and maintaining the instrument. To ensure their continued operation to the design specifica­tions, personnel should read this manual thoroughly before proceeding with installation, commissioning, operation, and maintenance of this instrument. If this equipment is used in a manner not specified by the manufacturer, the protection provided by it against hazards may be impaired.
Failure to follow the proper instructions may cause any one of the following situations to occur: Loss of life; personal injury; property dam­age; damage to this instrument; and warranty invalidation.
• Ensure that you have received the correct model and options from your purchase order. Verify that this manual covers your model and options. If not, call 1-800-854-8257 or 949-757-8500 to request correct manual.
For clarification of instructions, contact your Rosemount representative.
Follow all warnings, cautions, and instructions marked on and supplied with the product.
Use only qualified personnel to install, operate, update, program and maintain the product.
• Educate your personnel in the proper installation, operation, and maintenance of the product.
Install equipment as specified in the Installation section of this manual. Follow appropriate local and national codes. Only connect the product to electrical and pressure sources specified in this manual.
Use only factory documented components for repair. Tampering or unauthorized substitution of parts and procedures can affect the performance and cause unsafe operation of your process.
• All equipment doors must be closed and protec­tive covers must be in place unless qualified per­sonnel are performing maintenance.
• If this equipment is used in a manner not speci­fied by the manufacturer, the protection provided by it against hazards may be impaired.
W
ARNING
RISK OF ELECTRICAL SHOCK
Equipment protected throughout by double insulation.
Installation of cable connections and servicing of this product require access to shock hazard voltage levels.
Main power and relay contacts wired to separate power source must be disconnected before servicing.
Do not operate or energize instrument with case open!
Signal wiring connected in this box must be rated at least
240 V.
Non-metallic cable strain reliefs do not provide grounding between conduit connections! Use grounding type bushings and jumper wires.
Unused cable conduit entries must be securely sealed by non-flammable closures to provide enclosure integrity in compliance with personal safety and environmental protection requirements. Unused conduit openings must be sealed with NEMA 4X or IP65 conduit plugs to maintain the ingress protection rating (NEMA 4X).
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.
Operate only with front and rear panels fastened and in place over terminal area.
Safety and performance require that this instrument be connected and properly grounded through a three-wire power source.
Proper relay use and configuration is the responsibility of the
user.
CAUTION
This product generates, uses, and can radiate radio frequency energy and thus can cause radio communication interference. Improper installation, or operation, may increase such interfer­ence. As temporarily permitted by regulation, this unit has not been tested for compliance within the limits of Class A comput­ing devices, pursuant to Subpart J of Part 15, of FCC Rules, which are designed to provide reasonable protection against such interference. Operation of this equipment in a residential area may cause interference, in which case the user at his own expense, will be required to take whatever measures may be required to correct the interference.
W
ARNING
This product is not intended for use in th e light industrial, residential or commercial environments per the instruments certification to EN50081-2.
QUICK START
GUIDE
FOR FCL
TRANSMITTER
1. Refer to Section 2.0 for installation instructions and Section 3.0 for wiring instructions.
2. Once connections are secure and verified, apply power to the transmitter.
3. When the transmitter is powered up for the first time Quick Start screens appear.
4. The first quick start screen has two control boxes, one for language and the other for temperature units. a. The cursor, shown by dark blue backlighting, will be on the language control box. To change the language,
press the ENTER/MENU key. A list of available languages, shown two at a time, will appear. Using the up and down keys, scroll (see section 4.2) to display the choices. Press ENTER/MENU to select the desired language. Press the down key to move to the cursor to the temperature control box. To change units, press ENTER/MENU and scroll to either °F or °C. Press ENTER/MENU to store the selection.
b. To move to the next screen, use the navigation keys to move the cursor to NEXT and press
ENTER/MENU.
5. The next screen lists navigation rules. Press ENTER/MENU for the next screen.
6. The next step is to configure sensor 1. Sensor 1 is the free chlorine sensor. The screen has three control boxes. a. For measurement choose free chlorine. Do not choose pH-independent free chlorine. b. Choose the desired units, mg/L or ppm c. If you have an FCL-02 (with pH sensor), the third control box lets you choose between live/continuous
or manual pH correction. If you choose live/continuous (recommended), the transmitter will use the pH measured by the pH sensor to correct the chlorine reading for pH changes. If you choose manual (not recommended), a fourth control box will appear to let you enter the manual pH, and the transmitter
will use the entered value to correct the chlorine reading. d. If you have an FCL-01 (no pH sensor), enter the pH of the process liquid in the third control box. e. Move the cursor to NEXT and press ENTER/MENU. If you have an FCL-01, the display will change to
show some basic keypad operation guidelines. Press ENTER/MENU to show the main display. If you have an FCL-02, go to step 7.
7. The next step is to configure sensor 2. Sensor 2 is the pH sensor. The screen has two control boxes. a. For measurement choose pH. b. For pre-amplifier location choose transmitter. c. Move the cursor to NEXT and press ENTER/MENU. The display will change to show some basic keypad
operation guidelines. Press ENTER/MENU to show the main display.
8. The outputs, alarms, display configuration, and data logging are all assigned to default values. The default value for
data logging is disabled. To change the settings refer to Sections 4.5, 5.0, and 8.0.
About This Document
This manual contains instructions for installation and operation of the FCL-56 The following list provides notes concerning all revisions of this document.
Rev. Level
Date
Notes
5/11
This is the initial release of the product manual. The manual has been reformatted to reflect the Emerson documentation style and updated to reflect any changes in the product offering.
A 03/12
Update addresses - mail and web
B
07/17
Changed instances of analyzer to transmitter. Replaced Rosemount Analytical with Rosemount. Replaced Emerson Process Management with Emerson. Updated addresses and logos and reformatted back page.
MODEL FCL–56
TABLE OF CONTENTS
FCL-56
TABLE OF CONTENTS
Section Title Page
1.0 DESCRIPTION AND SPECIFICATIONS ................................................................ 1
1.1 Applications ............................................................................................................. 1
1.2 Features ................................................................................................................... 1
1.3 Specifications - General ........................................................................................... 2
1.4 Specifications - Sensor ............................................................................................ 2
1.5 Specifications - Transmitter .......................................................................................... 2
1.6 Ordering Information ................................................................................................ 3
2.0
INSTALLATION
.......................................................................................................
5
2.1
Unpacking and Inspection........................................................................................
5
2.2
Inst
allation ................................................................................................................
6
3.0 WIRING....................................................................................................................
9
3.1
Power, Alarm, and Output Wiring .............................................................................
9
3.2
Sensor Wiring .........................................................................................................
10
4.0 DISPLAY AND
OPERATION
...................................................................................
13
4.1
Display .....................................................................................................................
13
4.2
Keypad .....................................................................................................................
13
4.3
Operation .................................................................................................................
14
4.4
Hold..........................................................................................................................
16
4.5
Main Display ............................................................................................................
17
4.6
Security ....................................................................................................................
18
5.0 PROGRAMMING THE TRANSMITTER
19
5.1
Entering the Program Menus ...................................................................................
19
5.2
Outputs ....................................................................................................................
19
5.3
Relays ......................................................................................................................
20
5.4
Measurement ...........................................................................................................
22
5.5 Temperature .............................................................................................................
22
5.6
pH Diagnostic Setup ................................................................................................
23
5.7
Configuring Security Settings...................................................................................
23
5.8
Restoring Default Settings .......................................................................................
23
6.0 CALIBRATION ........................................................................................................
25
6.1
Introduction ..............................................................................................................
25
6.2
Entering the Calibration Menus................................................................................
25
6.3
Calibrating Temperature...........................................................................................
25
6.4
Calibrating the Free Chlorine Sensor ......................................................................
26
6.5
Calibrating the pH Sensor .......................................................................................
27
6.6
Calibrating the Analog Outputs ................................................................................
27
6.7
Reset........................................................................................................................
27
i
MODEL FCL–56
TABLE OF CONTENTS
TABLE OF CONTENTS CONT’D
Section Title Page
7.0 DIGITAL COMMUNICATIONS ............................................................................... 29
8.0 DATA AND EVENT LOGGING AND
RETRIEVAL
.................................................. 31
8.1 Overview .................................................................................................................. 31
8.2 Configuration............................................................................................................ 31
8.3 Downloading Data and Events................................................................................. 32
8.4 Viewing Events......................................................................................................... 32
8.5 Date and Time.......................................................................................................... 32
9.0 GRAPHICAL DISPLAY .......................................................................................... 33
9.1 Overview .................................................................................................................. 33
9.2 Configuration............................................................................................................ 33
10.0 MAINTENANCE ..................................................................................................... 35
10.1 Transmitter ................................................................................................................... 35
10.2 Chlorine Sensor ....................................................................................................... 36
10.3 pH Sensor ................................................................................................................ 37
10.4 Constant Head Flow Controller ................................................................................ 38
11.0 TROUBLESHOOTING ........................................................................................... 41
11.1 Overview .................................................................................................................. 41
11.2 Reading and Troubleshooting Fault and Warning Messages .................................. 41
11.3 Sensor Diagnostics .................................................................................................. 41
11.4 Troubleshooting Calibration Problems ..................................................................... 42
11.5 Other Troubleshooting – Chlorine ............................................................................ 42
11.6 Other Troubleshooting – pH ..................................................................................... 43
11.7 Other Troubleshooting – General............................................................................. 45
11.8 Simulating Inputs – Chlorine .................................................................................... 45
11.9 Simulating Inputs – pH ............................................................................................. 46
11.10 Simulating Inputs – Temperature ............................................................................. 47
LIST OF TABLES
Number Title Page
1.6 Ordering Information ............................................................................................... 3
1.6 Component Parts ..................................................................................................... 3
1.6 Accessories .............................................................................................................. 3
3.2 Sensor Wiring ........................................................................................................... 10
4.1 Display Abbreviations ............................................................................................... 13
10.1 Transmitter ................................................................................................................... 35
10.2 Spare Parts............................................................................................................... 37
10.3 Replacement Parts FCL-01 ...................................................................................... 39
10.4 Replacement Parts FCL-02 ...................................................................................... 40
11.5 Troubleshooting – Chlorine ...................................................................................... 42
ii
MODEL FCL–56
TABLE OF CONTENTS
LIST OF TABLES CONT’D
11.6
Other Troubleshooting – pH .....................................................................................
43
11.7
Other Troubleshooting – General .............................................................................
45
11.9
Simulating Inputs – pH .............................................................................................
46
11.10
Simulating Inputs Temperature.................................................................................
47
LIST OF
FIGURES
Number Title Page
2-1 Model FCL-01 .................................................................................................... 7
2-2 Model FCL-02 .................................................................................................... 7
3-1 Analog Output Connections ..................................................................................... 9
3-2 Alarm Relay Connections ......................................................................................... 10
3-3 Wiring Diagram for Free Chlorine Sensor ............................................................... 11
3-4 Wiring Diagram for 399VP-09 pH Sensor ............................................................... 11
3-5 Wiring Diagram for 3900VP-10 pH sensor (gray cable) .......................................... 11
3-6 Wiring Diagram for 3900VP-10 pH sensor (blue cable) .......................................... 11
4-1 Main Display .................................................................................................... 13
4-2 Transmitter Keypad .................................................................................................... 13
5-1 Menu Tree for Outputs Sub Menu ........................................................................... 19
5-2 Menu Tree for Relays Sub Menu ............................................................................. 20
5-3 Menu Tree for Measurements Sub Menu ................................................................. 22
5-4 Menu Tree for Temperature Sub Menu .................................................................... 22
5-5 Menu Tree for pH Diagnostic Sub Menu .................................................................. 23
5-6 Menu Tree for Security Sub Menu............................................................................ 23
10-1 Chlorine Sensor Parts .............................................................................................. 37
10-2 Replacement Parts for the Flow Controller Assembly used in Model FCL-01.......... 39
10-3 Replacement Parts for the Flow Controller Assembly used in Model FCL-02.......... 40
11-1 Simulating Chlorine .................................................................................................. 45
11-2 Simulating pH Inputs ................................................................................................ 46
11-3 Three-Wire RTD Configuration................................................................................. 47
11-4 Simulating RTD Inputs.............................................................................................. 47
iii
MODEL FCL– 56
This page left blank
intentionally
iv
MODEL FCL– 56
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
SECTION
1.0.
DESCRIPTION AND
SPECIFICATIONS
COMPLETE SYSTEM INCLUDES sensor, connecting cable, transmitter, and flow controller
CONTINUOUS pH CORRECTION eliminates expensive and messy reagents and trouble-
some sample conditioning systems
MEASURES FREE CHLORINE IN SAMPLES having pH as high as 9.5
1
VARIOPOL QUICK-DISCONNECT FITTINGS make replacing sensors easy
FEATURE-PACKED TRANSMITTER: four outputs, four fully-programmable relays, data
logger, and large full color display including graphics.
1
In some cases, the sensor can be used in samples having pH as great as 10.0. Consult the factory.
1.1 APPLICATIONS
The FCL free chlorine system is intended for the deter­mination of free chlorine in fresh water. Unlike free chlorine transmitters from other manufacturers, the FCL does not use expensive sample conditioning systems or messy reagents to control pH. Instead, the transmitter automatically compensates for changes in the pH of the sample. The FCL is not intended for the determi- nation of total chlorine or combined chlorine (like monochloramine). Nor, can the FCL be used for the determination of chlorine in seawater.
1.2 FEATURES
The FCL uses a membrane-covered amperometric sensor. A polarizing voltage applied to a platinum cathode behind the membrane reduces the chlorine diffusing through the membrane and keeps the con­centration of chlorine in the sensor equal to zero. The current generated by the cathode reaction is propor­tional to the rate of diffusion of chlorine through the membrane. Because the concentration of chlorine in the sensor is zero, th e diffusion rate and the current are proportional to the concentration of chlorine in the sample.
All amperometric free chlorine sensors respond to changes in pH. Although free chlorine is a mixture of hypochlorous acid and hypochlorite ion, hypochlorous acid alone is responsible for the sensor current. Because the relative amounts of hypochlorous acid and hypochlorite ion depend on pH, a pH change will cause the current and the apparent free chlorine concentration to change, even though the true con­centration remained constant. Most manufacturers solve the pH-dependence problem by treating the sample with acid, which lowers the pH and converts
hypochlorite ion into hypochlorous acid. The FCL avoids the expense and inconvenience of sample con­ditioning by measuring the pH and applying a correc­tion to the raw chlorine sensor signal. The correction is valid between pH 6.0 and 9.5. For samples having pH between 9.5 and 10.0, consult the factory.
The FCL is available in two options: Model FCL-01 with manual pH correction and Model FCL-02 with continuous pH correction. Choose the FCL-01 if the pH varies less than 0. 2 or if pH changes are predictable or seasonal. Choose the FCL-02 if the pH varies more than 0.2. To provide the continuous pH correction, the Model FCL-02 requires a separate pH sensor.
Maintenance is fast and easy. Replacing a membrane requires no special tools or fixtures. A screw cap holds the pre-tensioned membrane in place. Replacing the electrolyte solution takes only minutes.
The FCL includes the easy-to-use 56 transmitter. The transmitter features four fully programmable 4-20 mA out- puts and four fully programmable relays. The large, full color display allows the user to read sample pH and chlorine concentration at a glance. A data logger, graphical display, and HART digital communications are standard.
Valves, rotameters, and pressure regulators to control sample flow are things of the past with the FCL. A con­stant head overflow sampler ensures the correct sam­ple flow to each sensor. To eliminate wiring hassles, quick-disconnect Variopol cable is standard.
Stable free chlorine standards do not exist. The chlorine sensor must be calibrated using the results of a labora­tory test run on a grab sample.
1
MODEL FCL– 56
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.3 SPECIFICATIONS GENERAL
Sample requirements: Pressure: 3 to 65 psig (122 to 549 kPa abs)
A check valve in the inlet prevents the sensor flow cells from going dry if sample flow is lost. The check valve opens at 3 psig (122 kPa abs). If the check valve is removed, minimum pressure is 1 psig (108 kpa abs).
Temperature: 32 to 122°F (0 to 50°) Minimum Flow: 3 gal/hr (11 L/hr) Maximum flow: 80 gal/hr (303 L/hr); high flow causes
the overflow tube to back up
Sample Conductivity: >50 µS/cm at 25°C Process connection: 1/4-in OD tubing compression
fitting (can be removed and replaced with barbed fitting for soft tubing)
Drain connection: 3/4-in barbed fitting. Sample must
drain to open atmosphere.
Wetted parts:
Overflow sampler and flow cell: acrylic, polycar­bonate, Kynar®1, nylon, silicone
Chlorine sensor: Noryl®2, Viton®3, wood, silicone, polyethersulfone, polyester, and platinum
pH sensor (3900VP): stainless steel, glass, Teflon®4, polyphenylene sulfide, EPDM, and silicone
Response time: <80 sec to 95% of final reading for
inlet sample flow of 3 gph (11 L/hr)
Weight/shipping weight:
Model FCL-01: 10 lb/13 lb (4.5 kg/6.0 kg) Model FCL-02: 11 lb/14 lb (5.0 kg/6.5 kg)
[rounded to the nearest 1 lb. (0.5 kg)]
1.4 SPECIFICATIONS SENSOR
Free chlorine range: 0 to 10 ppm as Cl2. For higher
ranges, consult the factory.
pH correction range: 6.0 to 9.5. For samples having
pH between 9.5 and 10.0, consult the factory. If pH < 6.0, correction is not necessary. For manual pH correction, choose option -01. For continuous pH correction choose option -02.
Accuracy: Accuracy depends on the accuracy of the
chemical test used to calibrate the sensor.
Interferences: Monochloramine, permanganate,
peroxides.
Electrolyte volume: 25 mL (approx.) Electrolyte life: 3 months (approx.); for best results
replace electrolyte monthly.
1.5 SPECIFICATIONS TRANSMITTER
Case: Polycarbonate Display: Full color LCD, 3.75 x 2.20 in. (95 x 56 mm);
display can be customized by the user.
Languages: English, French, German, Italian, Spanish,
Portuguese, Chinese, Russian, and Polish.
Ambient Temperature and Humidity: 14 to 140°F (-
10 to 60°C); RH 5 to 95% (non-condensing). Between 23 and 131°F (-5 to 55°C) there is no visi­ble degradation in display response or performance.
Storage temperature: -4 to 140°F (-20 to 60°C) Power: 85 to 265 VAC, 47.5 to 65.0 Hz, 20 W RFI/EMI:
EN-61326 LVD: EN-6101-01 Outputs: Four 4-20 or 0-20 mA isolated current outputs;
assignable to measurement or temperature; fully scalable; maximum load 550 . HART digital signal is superimposed on output 1.
Alarms and Timers: Four relays, fully configurable as
a setpoint alarm, interval timer, TPC, bleed and feed timer, delay timer, date and time timer, and fault alarm.
Relays: Form C, SPDT, epoxy sealed. Relay Contact ratings:
5 A at 28 VDC or 300 VAC (resistive) 1/8 HP at 120/240 VAC
Control features: PID control (analog output) and time
proportional control or TPC (relays) are standard.
Data logger: Data automatically stored every 30 seconds
for 30 days; older data removed to make room for new data. The following data are automatically stored: Chlorine: date and time, ppm, temperature, raw sensor current pH: date and time, pH, temperature, mV, glass impedance, and reference impedance (if available)
Event logger: Stores up to 300 events with data and
time stamp: faults, warnings, calibration data, cali­bration results (pass or fail), power on/off cycles, and hold on/off. Alarm relay activation and deactivation can also be stored. Older events are automatically removed to make room for new events.
Data and event downloading: through USB port on
front panel.
Graphical display: Dual graphical display shows
measurement data on the y-axis and time on the x-axis. Y-axis is fully assignable and scalable. X- axis can be set to one hour, one day, seven days, or 30 days.
Digital communications: HART digital communica-
tions is standard.
1
Kynar is a registered trademark of Elf Atochem North America.
2
Noryl is a registered trademark of General Electric.
3
Viton is a registered trademark of E.I. duPont de Nemours & Co.
4
Teflon is a registered trademark of E.I. duPont de Nemours & Co.
2
MODEL FCL– 56
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
1.6 ORDERING INFORMATION
Model FCL Free Chlorine Measuring System. The FCL is a complete system for the determination of free chlo-
rine in aqueous samples. It consists of the sensor(s), transmitter, and constant head overflow device to control sam- ple flow. All components are mounted on a backplate. Model option -02 includes a pH sensor for continuous, automatic pH correction. Three replacement membranes and a 4-oz. bottle of electrolyte solution are shipped with the chlorine sensor.
FCL FREE CHLORINE MEASURING SYSTEM
CODE
pH CORRECTION (required selection)
01
Without pH sensor
02
With pH sensor
CODE
pH CORRECTION (required selection)
240
56-03-24-38-HT, 85–265 VAC, 47.5/65.0 Hz, chlorine only (option -01 only)
241
56-03-24-32-HT, 85–265 VAC, 47.5/65.0 Hz, chlorine and pH (option -02 only)
FCL-02 -241 EXAMPLE
COMPONENT PARTS
TRANSMITTER
DESCRIPTION
56-03-24-38-HT
56-03-24-38-HT, 85–265 VAC, 47.5/65.0 Hz, chlorine only
56-03-24-32-HT
56-03-24-32-HT, 85–265 VAC, 47.5/65.0 Hz, chlorine and pH
SENSOR MODEL
DESCRIPTION
499ACL-01-54-VP
Free chlorine sensor with Variopol connector
3900VP-02-10
pH sensor with Variopol connector
SENSOR CABLE
DESCRIPTION
23747-04
Interconnecting cable, Variopol for 499ACL sensor, 4 ft
23645-08
Interconnecting cable, Variopol for 3900VP sensor, 4 ft
ACCESSORIES
PART #
DESCRIPTION
9240048-00
Tag, stainless steel (specify marking)
3
MODEL FCL– 56
SECTION 1.0
DESCRIPTION AND SPECIFICATIONS
This page left blank
intentionally
4
MODEL FCL–56
SECTION 2.0
INST
ALLA
TION
SECTION
2.0.
INSTALLATION
2.1 UNPACKING AND INSPECTION
Inspect the shipping container. If it is damaged, contact the shipper immediately for instructions. Save the box. If there is no apparent damage, unpack the container. Be sure all items shown on the packing list are present. If items are missing, notify Rosemount immediately.
2.1.1 MODEL FCL-01 (free chlorine without continuous pH correction)
Model FCL-01 consists of the following items mounted on a back plate.
1. Model 56-03-24-38-HT transmitter with sensor cable attached.
2. Constant head flow controller with flow cell for chlorine sensor. The free chlorine sensor (Model 499ACL-01-54-VP), three membrane assemblies, and a bottle of electrolyte solution
are in a separate package.
2.1.2 MODEL FCL-02 (free chlorine with continuous pH correction)
Model FCL-02 consists of the following items mounted on a back plate.
1. Model 56-03-24-32-HT transmitter with sensor cables attached.
2. Constant head flow controller with flow cells for pH and chlorine sensors.
3. Ring clamp to hold pH buffer solution during calibration. The free chlorine sensor (Model 499ACL-01-54-VP), three membrane assemblies, and a bottle of electrolyte solution,
and the Model 3900VP-02-10 sensor, which replaces the Model 399VP-09 sensor, are in separate packages.
5
MODEL FCL–56
SECTION 2.0
INST
ALLA
TION
2.2 INSTALLATION
2.2.1 General Information
1. Although the system is suitable for outdoor use, do not install it in direct sunlight or in areas of extreme temperature.
CAUTION
The FCL free chlorine transmitter is NOT suitabl e for use in hazardous areas.
2. To keep the transmitter enclosure watertight, install plugs (provided) in the unused cable
openings.
3. Install the system in an area where vibrations and electromagnetic and radio frequency interference are min­imized or absent.
4. Be sure there is easy access to the transmitter and
sensors.
2.2.2 Sample Requirements
Be sure the sample meets the following requirements:
1. Temperature: 32 to 122ºF (0 to 50ºC)
2. Pressure: 3 to 65 psig (122 to 549 kPa abs)
3. Minimum flow: 3 gal/hr (11 L/hr)
2.2.3 Mounting, Inlet, and Drain Connections
The Model FCL is intended for wall mounting only. Refer to Figure 2-1 or 2-2 for details. The sensor(s) screw into the flow cell adapters as shown in the figures. For Model FCL-02 (free chlorine with continuous pH adjustment), the pH sensor must be installed as shown in Figure 2-2.
A 1/4-inch OD tubing compression fitting is provided for the sample inlet. If desired, the compression fitting can be removed and replaced with a barbed fitting. The fitting screws into a 1/4-inch FNPT check valve. The check valve prevents the flow cells from going dry if sample flow is lost.
The sample drains through a 3/4-inch barbed fitting. Attach a piece of soft tubing to the fitting and allow the waste to drain open atmosphere. Do not restrict the drain line.
Adjust the sample flow until the water level is even with the central overflow tube and excess water is flowing down the tube.
2.2.4 Electrical Connections
Refer to Section 3.1 for details.
2.2.5 Installing the Sensor(s)
The Model FCL is provided with sensor cables pre-wired to the transmitter. Connect the chlorine sensor (Model 499ACL-01-54-VP) to the cable labeled Chlorine. Connect the pH sensor (Model 3900VP-02-10 or older Model 399VP-09) to the cable labeled pH. The terminal end of the sensor is keyed to ensure proper mating with the cable receptacle. Once the key has slid into the mating slot, tighten the connection by turning the knurled ring clockwise.
The sensor(s) screw into the plastic fitting(s), which are held in the flow cell(s) by the union nut. Do not remove the protective cap on the sensor(s) until ready to put the sensor(s) in service.
MODEL FCL–56
SECTION 2.0
INST
ALLA
TION
6
MODEL FCL–56
SECTION 2.0
INST
ALLA
TION
INCH
MILLIMETER
FIGURE 2-1. Model FCL-01
INCH
MILLIMETER
FIGURE 2-2. Model FCL-02
7
MODEL FCL–56
SECTION 2.0
INST
ALLA
TION
This page left blank
intentionally
8
MODEL FCL– 56
SECTION 3.0
WIRING
SECTION
3.0.
WIRING
3.1 POWER, ALARM, AND OUTPUT WIRING
W
ARNING
RISK OF ELECTRICAL SHOCK
Electrical installation must be in accordance with the National Electrical Code (ANSI/NFPA-70) and/or any other applicable national or local codes.
3.1.1 Power
Wire AC mains power to the power supply board, which is mounted on the left hand side of the enclosure beneath the gray plastic cover. To remove the cover, grab it by the upper edges and pull straight out. The power connector is at the bottom of the board. See Figure 3-2. Bring the power cable through the conduit opening just below the con­nector. Unplug the connector from the board and wire the power cable to it. Lead connections are marked on the connector. (L is live or hot; N is neutral; the ground connection has the standard symbol.)
Replace the cover. The two tabs on the back edge of the cover fit into slots at the rear of the enclosure, and the three small slots in the front of the cover snap into the three tabs next to the relay terminal strip. See Figure 3-2. Once the tabs are lined up, push the cover to snap it in place.
AC power wiring should be 14 gauge or greater. Run the power wiring through the conduit opening nearest the power terminal. Provide a switch or breaker to disconnect the transmitter from the main power supply. Install the switch or breaker near the transmitter and label it as the disconnecting device for the transmitter.
3.1.2 Analog output wiring
Four analog current outputs are located on the main circuit board, which is attached to the inside of the enclosure door. Figure 3-1 shows the location of the terminals, the outputs they are assigned to, and the polarity.
For best EMI/RFI protection, use shielded output signal cable enclosed in earth-grounded metal conduit.
Keep output signal wiring sepa­rate from power wiring. Do not run signal and power or relay wiring in the same conduit or close together in a cable tray.
FIGURE 3-1. Analog output connections.
The analog outputs are on the main board near the hinged end of the enclosure door.
9
Loading...
+ 42 hidden pages