Industry leading measurement capability and reliability
■
Safety certified to IEC 61508 for SIL2 applications
■
Increased plant availability with predictive maintenance and easy troubleshooting
■
Reduced instrument count and process penetrations with a multivariable transmitter
$
%
&
Rosemount 5300 Level TransmitterOctober 2019
Taking guided wave radar benefits to the next level
Measurement principle
Low power, nano-second microwave pulses are guided down a probe submerged in the process media. When a microwave pulse
reaches a medium with a different dielectric constant, part of the energy is reflected back to the transmitter.
The transmitter uses the residual wave of the first reflection for measuring the interface level. Part of the wave, which was not
reflected at the upper product surface, continues until it is reflected at the lower product surface. The speed of this wave depends
fully on the dielectric constant of the upper product.
The time difference between the transmitted and the reflected pulse is converted into a distance, and the total level or interface
level is then calculated. The reflection intensity depends on the dielectric constant of the product: the higher dielectric constant
value, the stronger reflection.
Figure 1: Measurement Principle
A. Reference pulse
B. Level
C.
Interface level
Guided wave radar technology benefits
■
Highly accurate and reliable direct level measurement with no compensation needed for changing process conditions (such as
density, conductivity, viscosity, pH, temperature, and pressure)
■
No moving parts and no re-calibration result in minimized maintenance
■
Handles vapor, dust, turbulence, and foam well
■
Suitable for small tanks, difficult tank geometry, internal obstacles, and unaffected by the mechanical design of chambers
■
Top down installation minimizes risk for leakages
Contents
Taking guided wave radar benefits to the next level...........................................................................................................................2
Suitable for most liquid and solids level applications and liquid interface applications
■
Handles even the most challenging applications reliably, including process vessels, control, and safety systems
■
Easy retrofit in existing chambers or available as complete assembly with high quality Rosemount chambers
■
Dynamic Vapor Compensation assures accuracy also in saturated steam
■
Large coaxial probe optimized for interface applications were there is a need to measure level and interface level all the way up
to flange
Best performance and uptime
■
Unique Direct Switch Technology and Probe End Projection improve capability and reliability particularly in challenging
applications
■
Single lead probe for long measuring ranges, obstructions and low dielectrics ensures reliability in more applications, such as
viscous media
■
Signal processing algorithm makes it possible to distinguish between two liquids with a top layer down to 1 in. (2.5 cm).
■
Smart Galvanic Interface results in a more stable microwave and EMI performance with minimized effects from outside
disturbances
Robust design and increased safety
■
Heavy-duty unique hardware for extreme temperature and pressures with multiple layers of protection
■
EchoLogics® and smart software functions provide enhanced ability to keep track of the surface and detect a full vessel situation
■
Third party approved for overfill prevention and Safety Integrated System SIL3 suitability
■
Electronics and cable connections in separate compartments provides safer handling and improved moisture protection
■
Online device verification and reliable detection of high level conditions with the verification reflector
Easy installation and plant integration
■
Easy upgrade by matching existing tank connections and cut-to-fit probes
■
Long lengths of rigid probes for robust measurements become cost-effective and practical to ship, store and install with the
segmented probe option (code 4S)
■
Multivariable device reduces the number of process penetrations
■
Seamless system integration with HART®, FOUNDATION™ Fieldbus, Modbus®, or IEC 62591 (WirelessHART®) with the Emerson
Wireless 775 THUM™ Adapter
■
Pre-configured or easy configuration in Rosemount Radar Master with a five-step wizard, auto connect, and online help
■
Enhanced DD with step-by-step configuration and echo curve capability in tools such as AMS Device Manager, and handheld
communicator
■
DTM™ with echo curve capability for use in FDT®/DTM compatible configuration tools such as PACTware™, Yokogawa
FieldMate/PRM
Minimized maintenance reduces cost
■
Easy online troubleshooting with user friendly software, utilizing powerful echo curve and logging tools
■
Signal Quality Metrics diagnostics detect product build-up on probe to monitor turbulence, boiling, foam, and emulsions
■
Predictive maintenance with advanced diagnostics and Plantweb™ alerts
■
Modular design for reduced spare parts and easy replacement of the transmitter housing without opening the tank
Emerson.com/Rosemount3
Rosemount 5300 Level TransmitterOctober 2019
Ordering information
Rosemount 5301 and 5302 Level and/or Interface in Liquids
Rosemount 5301 and 5302 Guided Wave Radar Level Transmitters provide industry leading measurement
capabilities and reliability in liquids. Characteristics include:
■
Direct Switch Technology and Probe End Projection to handle low reflective media and long measuring
ranges
■
Wide range of probe styles, materials, and temperatures and pressures for application flexibility
■
HART 4-20 mA, FOUNDATION™ Fieldbus, Modbus, or IEC 62591 (WirelessHART®) with the THUM Adapter
(see Emerson Wireless 775 THUM™ Adapter for details)
■
Safety-certified to IEC 61508 (option code QT)
■
Advanced diagnostics (option code D01 or DA1)
■
Transmitter verification and high level supervision (option code HL1, HL2, or HL3)
Specification and selection of product materials, options, or components must be made by the purchaser of
the equipment. See Material selection for more information on material selection.
Table 1: Rosemount 5301 and 5302 Level and/or Interface in Liquids Ordering Information
The starred options (★) represent the most common options and should be selected for best delivery. The non-starred offerings are
subject to additional delivery lead time.
ModelProduct description
5301Guided Wave Radar Liquid Level or Interface Transmitter (interface available for fully submerged probe)★
5302Guided Wave Radar Liquid Level and Interface Transmitter★
Signal output
H4-20 mA with HART communication (default output from factory is HART 5, add option code HR7 for HART 7)
(see 4-20 mA HART® (output option code H) for details)
FFOUNDATION Fieldbus (see FOUNDATION™ Fieldbus (output option code F) for details)★
MRS-485 with Modbus communication (see Modbus® (output option code M) for details)★
URosemount 2410 Tank Hub Connectivity
Housing material
APolyurethane-covered Aluminum★
SStainless Steel, Grade CF8M (ASTM A743)
Conduit / cable threads
1½ - 14 NPT★
2M20 x 1.5 adapter★
★
42 pcs M20 x 1.5 adapter★
(1)(2)
G
(3)
E
(3)
M
Metal cable gland (½ - 14 NPT)★
M12, 4-pin, male connector (eurofast®)★
A size Mini, 4-pin, male connector (minifast®)★
4Emerson.com/Rosemount
October 2019Rosemount 5300 Level Transmitter
Table 1: Rosemount 5301 and 5302 Level and/or Interface in Liquids Ordering Information (continued)
Operating temperature and pressure (see Process temperature and pressure rating)
(4)
Probe type
SStandard:
- 15 to 580 psig @ 302 °F
1A, 2A, 3A, 3B, 3C, 4A, 4B,
4S, 5A, and 5B
(-1 to 40 bar @ 150 °C)
(5)
H
High Temperature / High Pressure:
5000 psi @ 100 °F and 2940 psi @ 752 °F
3A, 3B, 3V, 4A, 4B, 4S, 4U,
5A, and 5B
(345 bar @ 38 °C and 203 bar @ 400 °C)
(5)
P
High Pressure:
5000 psi @ 100 °F and 3320 psi @ 482 °F
Hazardous locations certifications (see Product certifications)
NANo Hazardous Locations Certifications★
(19)
E1
E3
E5
E6
E7
(19)
(19)
(19)
(19)
ATEX Flameproof★
China Flameproof★
USA Explosion-proof★
Canadian Explosion-proof★
IECEx Flameproof★
I1ATEX Intrinsic Safety★
(20)
IA
ATEX FISCO Intrinsic Safety★
Emerson.com/Rosemount9
Rosemount 5300 Level TransmitterOctober 2019
Table 1: Rosemount 5301 and 5302 Level and/or Interface in Liquids Ordering Information (continued)
I3China Intrinsic Safety★
(20)
IC
I5USA Intrinsic Safety and Non-Incendive★
(20)
IE
I6Canadian Intrinsic Safety★
(20)
IF
I7IECEx Intrinsic Safety★
(20)
IG
(19)
E2
(19)
EM
I2INMETRO Intrinsic Safety
(20)
IB
IMTechnical Regulations Customs Union (EAC) Intrinsic Safety
(20)
IN
(19)(21)
E4
(19)
EP
(19)
KA
(19)
KB
(19)
KC
(19)
KD
China FISCO Intrinsic Safety★
USA FISCO Intrinsic Safety★
Canadian FISCO Intrinsic Safety★
IECEx FISCO Intrinsic Safety★
INMETRO Flameproof
Technical Regulations Customs Union (EAC) Flameproof
INMETRO FISCO Intrinsic Safety
Technical Regulations Customs Union (EAC) FISCO Intrinsic Safety
Japan Flameproof
Republic of Korea Flameproof
ATEX, USA, Canadian Flameproof/Explosion-proof
ATEX, USA, IECEx Flameproof/Explosion-proof
ATEX, Canadian, IECEx Flameproof/Explosion-proof
USA, Canadian, IECEx Flameproof/Explosion-proof
KEATEX, USA, Canadian Intrinsic Safety
KFATEX, USA, IECEx Intrinsic Safety
KGATEX, Canadian, IECEx Intrinsic Safety
KHUSA, Canadian, IECEx Intrinsic Safety
(20)
KI
KJ
KK
KL
(20)
(20)
(20)
FISCO - ATEX, USA, Canadian Intrinsic Safety
FISCO - ATEX, USA, IECEX Intrinsic Safety
FISCO - ATEX, Canadian, IECEX Intrinsic Safety
FISCO - USA, Canadian, IECEX Intrinsic Safety
N1ATEX Type n
N7IECEx Type n
Options (include with selected model number)
Display
M1Integral digital display★
Communication
HR74–20 mA with digital signal based on HART 7 protocol★
10Emerson.com/Rosemount
October 2019Rosemount 5300 Level Transmitter
Table 1: Rosemount 5301 and 5302 Level and/or Interface in Liquids Ordering Information (continued)
Hydrostatic testing
(22)
P1
Hydrostatic testing★
Factory configuration
C1Factory configuration (Configuration Data Sheet required with order, available at Emerson.com/Rosemount)★
Alarm limit configuration
C4Namur alarm and saturation levels, high alarm★
C5Namur alarm and saturation levels, low alarm★
(23)
C8
Welding documentation
Low alarm (standard Rosemount alarm and saturation levels)★
(24)
Q66Welding Procedure Qualification Record Documentation
Q67Welder Performance Qualification Record
Q68Welding Procedure Specification
Special certifications
Q4Calibration Data Certification★
(25)
Q8
Material Traceability Certification consistent with ISO10474-3.1:2013 / EN10204-3.1:2004★
QGGOST Primary Verification Certificate (only for end-destination country Russia)
Safety certifications
QSPrior-use certificate of FMEDA Data. Only available with HART 4-20 mA output (output code H).★
QTSafety-certified to IEC 61508 with certificate of FMEDA data. Only available with HART 4-20 mA output (output
code H).
Country certification
J1Canadian Registration Number (CRN)★
(26)
J2
(27)
J7
(28)(29)
J8
ASME B31.1★
Indian Boiler Regulation (For witnessed Form III-C from factory, order certificate Q47 on separate line Item)
EN Boiler (European Boiler Approval in accordance with EN 12952-11 and EN 12953-9)★
Dye penetration test certificate
Q73Certificate of Liquid Penetrant Inspection★
Positive material identification certificate
Q76Positive Material Identification Certificate of Conformance★
Materials certification
(30)
N2
Marine / shipboard approvals
NACE® material recommendation per NACE MR0175/ISO 15156 and NACE MR0103/ISO 17945★
(31)
SBSAmerican Bureau of Shipping Type Approval
★
SDNDet Norske Veritas Germanischer Lloyd (DNV GL) Type Approval
SLLLloyd's Register Type Approval
SKRKorean Register Type Approval
Emerson.com/Rosemount11
Rosemount 5300 Level TransmitterOctober 2019
Table 1: Rosemount 5301 and 5302 Level and/or Interface in Liquids Ordering Information (continued)
SBVBureau Veritas Type Approval
Installation options
(32)
LS
BR316L Mounting Bracket for 1½-in. NPT Process Connection (RA) (see Figure 47)
Weight and anchoring options for flexible single probes (probe type 5A)
Long stud 9.8 in (250 mm) for flexible single lead probe to prevent contact with wall/nozzle. Standard stud
length is 3.9 in (100 mm) for probes 5A and 5B.
★
W3Heavy weight (recommended choice for most applications) Weight=2.43 lb (1.10 kg), Length=5.5 in. (140
mm), Diameter=1.5 in. (37.5 mm)
W2Short weight (when measuring close to the probe end) Weight=0.88 lb (0.40 kg. Length=2 in. (50 mm),
Diameter=1.5 in. (37.5 mm)
Weight assembly options for flexible single probes
WUWeight or chuck not mounted on the probe★
Transient protection
T1Transient Protection Terminal Block. Selectable with HART 4-20 mA output (output code H). Already included
in all FOUNDATION Fieldbus variations.
Diagnostic functionality
D01FOUNDATION Fieldbus Diagnostics Suite (includes Signal Quality Metrics diagnostics)★
DA1HART Diagnostics Suite (includes Signal Quality Metrics diagnostics)★
Cold temperature
(33)(34)(35)
BR5
(36)
-67 °F (-55 °C) cold temperature (pending, consult factory)
Verification reflectors (high level supervision)
(37)
HL1
Verification reflector for 3- to 6-in. pipe/chamber (High Level Supervision). See Verification reflector (option
code HL1, HL2, or HL3) for details.
(37)
HL2
Verification reflector for 8-in. pipe/chamber (High Level Supervision). See Verification reflector (option code
HL1, HL2, or HL3) for details.
(37)
HL3
Verification reflector for tanks and 10-in. or wider pipe/chamber (High Level Supervision). See Verification
reflector (option code HL1, HL2, or HL3) for details.
★
★
Overfill prevention
(38)
U1
WHG Overfill Approval★
Extended product warranty
(39)
WR3
WR5
(39)
3-year limited warranty★
5-year limited warranty★
Centering discs (see Table 33 for size recommendation)Outer diameter
(40)
S2
S3
S4
P2
P3
(40)
(40)
(41)
(41)
2-in. Centering disc1.8 in. (45 mm)★
3-in. Centering disc2.7 in. (68 mm)★
4-in. Centering disc3.6 in. (92 mm)★
2-in. Centering disc PTFE1.8 in. (45 mm)★
3-in. Centering disc PTFE2.7 in. (68 mm)★
12Emerson.com/Rosemount
October 2019Rosemount 5300 Level Transmitter
Table 1: Rosemount 5301 and 5302 Level and/or Interface in Liquids Ordering Information (continued)
(41)
P4
(40)
S6
(40)
S8
(41)
P6
(41)
P8
Remote housing mounting
4-in. Centering disc PTFE3.6 in. (92 mm)★
6-in. Centering disc5.55 in. (141 mm)
8-in. Centering disc7.40 in. (188 mm)
6-in. Centering disc PTFE5.55 in. (141 mm)
8-in. Centering disc PTFE7.40 in. (188 mm)
(42)
(see Figure 48)
B11 m / 3.2 ft. Remote Housing Mounting Cable and 316L Bracket
B22 m / 6.5 ft. Remote Housing Mounting Cable and 316L Bracket
B33 m / 9.8 ft. Remote Housing Mounting Cable and 316L Bracket
Reference reflectors for dynamic vapor compensation probes (required for probe type 3V and 4U)
(See Select reference reflector for reflector length guidelines)
R1Short reflector. Length=14 in. (350 mm)
R2Long reflector. Length=20 in. (500 mm)
Consolidate to chamber (see Rosemount chamber)
(43)
XC
Consolidate to Chamber
Engineered solutions (see Engineered solutions)
RxxxxEngineered Solutions beyond standard model codes. (Consult factory for details)
Example model string: 5301-H-A-1-S-1-V-1A-M-002-05-AA-I1-M1C1
E-002-05, means 2 ft. and 5 in. probe length. M-002-05, means 2.05 m.
Not available with explosion-proof or flameproof approvals.
(1)
Minimum temperature is -20 °C (-4 °F).
(2)
Not available with explosion-proof, flameproof, or type n approvals.
(3)
Process seal rating. Final rating depends on flange and O-ring selection. See “Temperature and pressure limits” on Process temperature and
(4)
pressure rating, Flange rating, Tri-Clamp® rating, and Plate design.
Requires option None for sealing (no O-ring).
(5)
For other materials, consult the factory.
(6)
Requires Operating Temperature and Pressure code H, P, or C.
(7)
Only available for Probe Type code 3C.
(8)
Consult factory.
(9)
Only available with Operating Temperature and Pressure code S.
(10)
Requires firmware version 2.L3 or later.
(11)
Only available with Operating Temperature and Pressure code H.
(12)
Not available with Remote housing code B1 or B2.
(13)
0.79 lb (0.36 kg) standard weight for flexible single lead probe. L=5.5 in. (140 mm). For PTFE covered probes: 2.2 lb (1 kg) standard weight for
(14)
flexible single lead probe. L=17.1 in. (434 mm).
Maximum probe length for Duplex 2205 probes is 105 ft (32 m).
(15)
Extra length for fastening is added in factory.
(16)
Requires model 5301.
(17)
Probe weight included if applicable. Give the total probe length in feet and inches or meters and centimeters, depending on selected probe length
(18)
unit. If tank height is unknown, please round up to an even length when ordering. Probes can be cut to exact length in field. Maximum allowable
length is determined by process conditions. See Total probe length for more probe length guidance.
Probes are intrinsically safe.
(19)
Requires FOUNDATION Fieldbus signal output (Ui parameter listed in Product certifications).
(20)
G ½ in. SST cable gland is included in delivery.
(21)
Available for tank connection with flange.
(22)
The standard alarm setting is high.
(23)
Weldings in accordance with EN/ISO standards.
(24)
Emerson.com/Rosemount13
Rosemount 5300 Level TransmitterOctober 2019
Certificate includes all pressure retaining wetted parts.
(25)
Design and manufacturing according to ASME B31.1. No code stamp available. Welding in accordance with ASME IX.
(26)
Only available with Material of construction code 1, Operating temperature and pressure code S, H, or P, Probe type 3A, 3B, 3V, 4U, 4A, 4B, 4S,
(27)
5A, or 5B, together with ASME flanges size 2-in, 3-in, or 4-in.
Only available with Signal output code H and Probe type code 3V or 4U.
(28)
Suitable for use as a level sensor part of a limiting device, in accordance with EN 12952-11 and EN 12953-9.
(29)
For Probe Type 3A, 3B, 3C, 4A, 4B, 4S, and 4U, and PTFE-coated 5A.
(30)
Transmitters with aluminum housing are not approved for open deck installations.
(31)
Not available with PTFE covered probes.
(32)
Consider any temperature limitations dependent on Material of construction, Hazardous locations certifications, and/or O-ring selection.
(33)
Not available with option code QS or U1.
(34)
Option code BR5 together with Remote housing options (B1, B2, B3) or Verification reflector options (HL1, HL2, HL3) is pending, consult factory.
(35)
For ambient temperatures between -67 °F (-55 °C) and -40 °F (-40 °C), the ambient temperature effect is ± 0.012 in. (0.3 mm) /°K or ± 45 ppm/°K
(36)
of measured value, whichever is greatest. Other performance specifications apply to ambient temperatures between –40 °F (–40 °C) and 185 °F
(85 °C).
Only available with HART 4-20 mA output (code H), standard operating temperature and pressure (code S), material of construction (code 1),
(37)
and flexible single lead probes (probe type 5A or 5B).
Cannot be combined with E2 (INMETRO Flameproof) or I2 (INMETRO Intrinsic Safety).
(38)
Rosemount extended warranties have a limited warranty of three or five years from date of shipment.
(39)
Available for SST, Alloy C-276, Alloy 400, and Duplex 2205 probes, type 2A, 4A, 4B, 4S, and 5A. Same disc material as probe material. For more
(40)
information, see Centering disc for pipe installations.
Available for probe types 2A, 4A, 4B, 4S, and 5A. Not available with Operating Temperature and Pressure code H or Material of Construction codes
(41)
7 and 8.
Not available with Marine / shipboard approvals.
(42)
Selecting the XC option code on the Rosemount 5300 Guided Wave Radar and a Rosemount chamber will result in matching, consolidating,
(43)
configuring, and shipping of the two products in one crate. Note that the flange bolts are only hand-tightened. Long rigid single lead probes (>8
ft./2.5 m) are ship separately in order to reduce transportation risk damage.
Rosemount 5303 Level for Solids
Rosemount 5303 Guided Wave Radar Level Transmitter provides industry leading measurement capabilities
and reliability on solids. Characteristics include:
■
Direct Switch Technology and Probe End Projection to handle low reflective media and long measuring
ranges
■
Measurement independent of dust, moisture and material fluctuations
■
HART 4-20 mA, FOUNDATION™ Fieldbus, Modbus, or IEC 62591 (WirelessHART®) with the THUM Adapter
(see Emerson Wireless 775 THUM™ Adapter for details)
■
Probes for high physical weight loads (probe type 6A and 6B)
■
Long stud available to prevent contact with nozzle (LS option)
Specification and selection of product materials, options, or components must be made by the purchaser of
the equipment. See Material selection for more information on material selection.
Table 2: Rosemount 5303 Level for Solids Ordering Information
The starred options (★) represent the most common options and should be selected for best delivery. The non-starred offerings are
subject to additional delivery lead time.
ModelProduct description
5303Guided Wave Solids Level Transmitter★
14Emerson.com/Rosemount
October 2019Rosemount 5300 Level Transmitter
Table 2: Rosemount 5303 Level for Solids Ordering Information (continued)
Signal output
H4-20 mA with HART communication (default output from factory is HART 5, add option code HR7 for HART
7) (see 4-20 mA HART® (output option code H) for details)
FFOUNDATION Fieldbus (see FOUNDATION™ Fieldbus (output option code F) for details)★
MRS-485 with Modbus communication (see Modbus® (output option code M) for details)★
Housing material
APolyurethane-covered Aluminum★
SStainless Steel, Grade CF8M (ASTM A743)
Conduit / cable threads
1½ - 14 NPT★
2M20 x 1.5 adapter★
42 pcs M20 x 1.5 adapter★
(1)(2)
G
(3)
E
(3)
M
Operating temperature and pressure (see Process temperature and
pressure rating)
SStandard:
Metal cable gland (½ - 14 NPT)★
M12, 4-pin, Male connector (eurofast)★
A size Mini, 4-pin, Male connector (minifast)★
(4)
Probe type
All★
- 15 to 580 psig @ 302 °F (-1 to 40 bar @ 150 °C)
Material of construction
(5)
: process connection / probeProbe type
★
1316/316L/ EN 1.4404All★
Sealing, O-ring material (consult factory for other O-ring materials)
VViton® Fluoroelastomer★
EEthylene Propylene (EPDM)★
KKalrez® 6375 Perfluoroelastomer★
BNitrile Butadiene (NBR)★
Probe typeProcess connectionProbe lengths
(6)
5A
5B
6A
6B
(7)
(8)
(8)
Flexible Single Lead with
weight, 4 mm
Flexible Single Lead with
chuck, 4 mm
Flexible Single Lead with
weight, 6 mm
Flexible Single Lead with
chuck, 6 mm
Flange / 1-in., 1½-in., 2-in.
Thread
Flange / 1-in., 1½-in., 2-in.
Thread
Flange / 1-in., 1½-in., 2-in.
Thread
Flange / 1-in., 1½-in., 2-in.
Thread
Min: 3 ft. 4 in. (1 m) Max: 115 ft. (35 m)★
Min: 3 ft. 4 in. (1 m) Max: 115 ft. (35 m)★
Min: 3 ft. 4 in. (1 m) Max: 164 ft. (50 m)★
Min: 3 ft. 4 in. (1 m) Max: 164 ft. (50 m)★
Probe length units
EEnglish (feet, in.)★
MMetric (meters, centimeters)★
Emerson.com/Rosemount15
Rosemount 5300 Level TransmitterOctober 2019
Table 2: Rosemount 5303 Level for Solids Ordering Information (continued)
(9)
(9)
Total probe length (feet/m)
XXX0-164 ft. or 0-50 m★
Total probe length (in./cm)
XX0-11 in. or 0-99 cm★
Process connection - size / type (consult the factory for other process connections)
ASME flanges
(10)
AA2 in. Class 150, RF (Raised Face Type)★
AB2 in. Class 300, RF (Raised Face Type)★
BA3 in. Class 150, RF (Raised Face Type)★
BB3 in. Class 300, RF (Raised Face Type)★
CA4 in. Class 150, RF (Raised Face Type)★
CB4 in. Class 300, RF (Raised Face Type)★
DA6 in. Class 150, RF (Raised Face Type)
EN 1092-1 flanges
(11)
HBDN50, PN40, Type A flat face★
IADN80, PN16, Type A flat face★
IBDN80, PN40, Type A flat face★
JADN100, PN16, Type A flat face★
JBDN100, PN40, Type A flat face★
HIDN50, PN40, Type E spigot face
HPDN50, PN16, Type C tongue face
HQDN50, PN40, Type C tongue face
IHDN80, PN16, Type E spigot face
IIDN80, PN40, Type E spigot face
IPDN80, PN16, Type C tongue face
IQDN80, PN40, Type C tongue face
JHDN100, PN16, Type E spigot face
JIDN100, PN40, Type E spigot face
JPDN100, PN16, Type C tongue face
JQDN100, PN40, Type C tongue face
KADN150, PN16, Type A flat face
JIS flanges
(11)
UA50A, 10K, RF (Raised Face Type)★
VA80A, 10K, RF (Raised Face Type)★
XA100A, 10K, RF (Raised Face Type)★
UB50A, 20K, RF (Raised Face Type)
16Emerson.com/Rosemount
October 2019Rosemount 5300 Level Transmitter
Table 2: Rosemount 5303 Level for Solids Ordering Information (continued)
VB80A, 20K, RF (Raised Face Type)
XB100A, 20K, RF (Raised Face Type)
YA150A, 10K, RF (Raised Face Type)
YB150A, 20K, RF (Raised Face Type)
ZA200A, 10K, RF (Raised Face Type)
ZB200A, 20K, RF (Raised Face Type)
Threaded connections
(10)
Probe type
RA1½-in. NPT threadAll★
RC2-in. NPT threadAll★
RB1-in. NPT threadAll
SA1½-in. BSP (G 1½-in.) threadAll
SB1-in. BSP (G 1-in.) threadAll
Hazardous locations certifications (see Product certifications)
NANo Hazardous Locations Certifications★
(12)
E1
E3
E5
E6
E7
(12)
(12)
(12)
(12)
ATEX Flameproof★
China Flameproof★
USA Explosion-proof★
Canadian Explosion-proof★
IECEx Flameproof★
I1ATEX Intrinsic Safety★
(13)
IA
ATEX FISCO Intrinsic Safety★
I3China Intrinsic Safety★
(13)
IC
China FISCO Intrinsic Safety★
I5USA Intrinsic Safety and Non-Incendive★
(13)
IE
USA FISCO Intrinsic Safety★
I6Canadian Intrinsic Safety★
(13)
IF
Canadian FISCO Intrinsic Safety★
I7IECEx Intrinsic Safety★
(13)
IG
E2
EM
(12)
(12)
IECEx FISCO Intrinsic Safety★
INMETRO Flameproof
Technical Regulations Customs Union (EAC) Flameproof
I2INMETRO Intrinsic Safety
(13)
IB
INMETRO FISCO Intrinsic Safety
IMTechnical Regulations Customs Union (EAC) Intrinsic Safety
(13)
IN
E4
(12)(14)
Technical Regulations Customs Union (EAC) FISCO Intrinsic Safety
Japan Flameproof
Emerson.com/Rosemount17
Rosemount 5300 Level TransmitterOctober 2019
Table 2: Rosemount 5303 Level for Solids Ordering Information (continued)
(12)
EP
(12)
KA
(12)
KB
(12)
KC
(12)
KD
KEATEX, USA, Canadian Intrinsic Safety
KFATEX, USA, IECEx Intrinsic Safety
KGATEX, Canadian, IECEx Intrinsic Safety
KHUSA, Canadian, IECEx Intrinsic Safety
(13)
KI
(13)
KJ
(13)
KK
(13)
KL
N1ATEX Type n
Republic of Korea Flameproof
ATEX, USA, Canadian Flameproof/Explosion-proof
ATEX, USA, IECEx Flameproof/Explosion-proof
ATEX, Canadian, IECEx Flameproof/Explosion-proof
USA, Canadian, IECEx Flameproof/Explosion-proof
FISCO - ATEX, USA, Canadian Intrinsic Safety
FISCO - ATEX, USA, IECEX Intrinsic Safety
FISCO - ATEX, Canadian, IECEX Intrinsic Safety
FISCO - USA, Canadian, IECEX Intrinsic Safety
N7IECEx Type n
Options (include with selected model number)
Display
M1Integral digital display★
Communication
HR74–20 mA with digital signal based on HART 7 protocol★
Hydrostatic testing
(15)
P1
Hydrostatic testing★
Factory configuration
C1Factory configuration (Configuration Data Sheet required with order, available at Emerson.com/
Rosemount)
Alarm limit configuration
C4Namur alarm and saturation levels, high alarm★
C5Namur alarm and saturation levels, low alarm★
C8Low alarm
Welding documentation
(16)
(standard Rosemount alarm and saturation levels)★
(17)
Q66Welding Procedure Qualification Record Documentation
Q67Welder Performance Qualification Record
★
Q68Welding Procedure Specification
Safety certifications
QSPrior-use certificate of FMEDA Data. Only available with HART 4-20 mA output (output code H).★
QTSafety-certified to IEC 61508 with certificate of FMEDA data. Only available with HART 4-20 mA output
★
(output code H).
18Emerson.com/Rosemount
October 2019Rosemount 5300 Level Transmitter
Table 2: Rosemount 5303 Level for Solids Ordering Information (continued)
Special certifications
Q4Calibration Data Certification★
(18)
Q8
Material Traceability Certification consistent with ISO10474-3.1:2013 / EN10204-3.1:2004★
QGGOST Primary Verification Certificate (only for end-destination country Russia)
Dye penetration test certificate
Q73Certificate of Liquid Penetrant Inspection★
Positive material identification certificate
Q76Positive Material Identification Certificate of Conformance★
Installation options
LSLong stud 9.8 in (250 mm) for flexible single lead probe to prevent contact with wall/nozzle. Standard stud
length is 3.9 in (100 mm) for probes 5A and 5B; 5.9 in. (150 mm) for probes 6A and 6B
BR316L Mounting Bracket for 1½-in. NPT Process Connection (RA) (see Figure 47)
Transient protection
T1Transient Protection Terminal Block. Selectable with HART 4-20 mA output (output code H). Already
included in all FOUNDATION Fieldbus variations.
Diagnostic functionality
D01FOUNDATION Fieldbus Diagnostics Suite (includes Signal Quality Metrics diagnostics)★
DA1HART Diagnostics Suite (includes Signal Quality Metrics diagnostics)★
Overfill prevention
(19)
U1
WHG Overfill Approval★
Extended product warranty
(20)
WR3
WR5
(20)
3-year limited warranty★
5-year limited warranty★
Remote housing mounting (see Figure 48)
B11m / 3.2 ft. Remote Housing Mounting Cable and 316L Bracket
B22m / 6.5 ft. Remote Housing Mounting Cable and 316L Bracket
B33m / 9.8 ft. Remote Housing Mounting Cable and 316L Bracket
★
★
Engineered solutions (see Engineered solutions)
RxxxxEngineered Solutions beyond standard model codes. (Consult factory for details)
Example model string: 5303-H-A-1-S-1-V-6A-M-025-50-AA-I1-M1C1.
E-025-05, means 25 ft. and 5 in. probe length. M-025-50, means 25.5 m.
Not available with explosion-proof or flameproof approvals.
(1)
Minimum temperature is -20 °C (-4 °F).
(2)
Not available with explosion-proof, flameproof, or type n approvals.
(3)
Process seal rating. Final rating depends on flange and O-ring selection. See “Temperature and pressure limits” on Process temperature and
(4)
pressure rating, Flange rating, Tri-Clamp® rating, and Plate design.
For other materials, consult the factory.
(5)
0.79 lb (0.36 kg) standard weight for flexible single lead probe. L=5.5 in. (140 mm).
(6)
Extra length for fastening is added in the factory.
(7)
1.2 lb (0.56 kg) standard weight for flexible single lead probe. L=5.5 in. (140 mm).
(8)
Emerson.com/Rosemount19
Rosemount 5300 Level TransmitterOctober 2019
Probe weight included if applicable. Give the total probe length in feet and inches or meters and centimeters, depending on selected probe length
(9)
unit. If tank height is unknown, please round up to an even length when ordering. Probes can be cut to exact length in field. Maximum allowable
length is determined by process conditions. See Total probe length for more probe length guidance.
Available in 316L. For other materials, consult the factory.
(10)
Available in 316L and EN 1.4404. For other materials consult the factory.
(11)
Probes are intrinsically safe.
(12)
Requires FOUNDATION Fieldbus signal output (Ui parameter listed in Product certifications).
(13)
G ½ in. SST cable gland is included in delivery.
(14)
Available for tank connection with flange.
(15)
The standard alarm setting is high.
(16)
Weldings in accordance with EN/ISO standards.
(17)
Certificate includes all pressure retaining wetted parts.
(18)
Cannot be combined with E2 (INMETRO Flameproof) or I2 (INMETRO Intrinsic Safety).
(19)
Rosemount extended warranties have a limited warranty of three or five years from date of shipment.
(20)
Accessories
Table 3: Accessories Ordering Information
The starred options (★) represent the most common options and should be selected for best delivery. The non-starred offerings are
subject to additional delivery lead time.
Weight kit
03300-7001-0002Weight kit flexible twin lead
03300-7001-0003Weight kit flexible 4 mm single lead
03300-7001-0004Weight kit flexible 6 mm single lead
Centering discs for rigid single lead probe (d=0.3 in./8 mm)
(1)(2)
Outer diameter
03300-1655-0001Kit, 2-in. Centering disc, SST1.8 in. (45 mm)★
03300-1655-0006Kit, 2-in. Centering disc, PTFE1.8 in. (45 mm)★
03300-1655-0002Kit, 3-in. Centering disc, SST2.7 in. (68 mm)★
03300-1655-0007Kit, 3-in. Centering disc, PTFE2.7 in. (68 mm)★
03300-1655-0003Kit, 4-in. Centering disc, SST3.6 in. (92 mm)★
03300-1655-0008Kit, 4-in. Centering disc, PTFE3.6 in. (92 mm)★
03300-1655-0004Kit, 6-in. Centering disc, SST5.55 in. (141 mm)
03300-1655-0009Kit, 6-in. Centering disc, PTFE5.55 in. (141 mm)
03300-1655-0005Kit, 8-in. Centering disc, SST7.40 in. (188 mm)
03300-1655-0010Kit, 8-in. Centering disc, PTFE7.40 in. (188 mm)
Centering discs for rigid single lead probe (d=0.5 in./13 mm)
(1)(2)
Outer diameter
03300-1655-0301Kit, 2-in. Centering disc, SST1.8 in. (45 mm)★
03300-1655-0306Kit, 2-in. Centering disc, PTFE1.8 in. (45 mm)★
03300-1655-0302Kit, 3-in. Centering disc, SST2.7 in. (68 mm)★
03300-1655-0307Kit, 3-in. Centering disc, PTFE2.7 in. (68 mm)★
03300-1655-0303Kit, 4-in. Centering disc, SST3.6 in. (92 mm)★
03300-1655-0308Kit, 4-in. Centering disc, PTFE3.6 in. (92 mm)★
03300-1655-0304Kit, 6-in. Centering disc, SST5.55 in. (141 mm)
20Emerson.com/Rosemount
October 2019Rosemount 5300 Level Transmitter
Table 3: Accessories Ordering Information (continued)
03300-1655-0309Kit, 6-in. Centering disc, PTFE5.55 in. (141 mm)
03300-1655-0305Kit, 8-in. Centering disc, SST7.40 in. (188 mm)
03300-1655-0310Kit, 8-in. Centering disc, PTFE7.40 in. (188 mm)
Centering discs for flexible single/twin lead probes
(1)(2)
Outer diameter
03300-1655-1001Kit, 2-in. Centering disc, SST1.8 in. (45 mm)★
03300-1655-1006Kit, 2-in. Centering disc, PTFE1.8 in. (45 mm)★
03300-1655-1002Kit, 3-in. Centering disc, SST2.7 in. (68 mm)★
03300-1655-1007Kit, 3-in. Centering disc, PTFE2.7 in. (68 mm)★
03300-1655-1003Kit, 4-in. Centering disc, SST3.6 in. (92 mm)★
03300-1655-1008Kit, 4-in. Centering disc, PTFE3.6 in. (92 mm)★
03300-1655-1004Kit, 6-in. Centering disc, SST5.55 in. (141 mm)
03300-1655-1009Kit, 6-in. Centering disc, PTFE5.55 in. (141 mm)
03300-1655-1005Kit, 8-in. Centering disc, SST,7.40 in. (188 mm)
03300-1655-1010Kit, 8-in. Centering disc, PTFE7.40 in. (188 mm)
Centering discs for mounting between segments (probe type 4S only)Outer diameter
03300-1656-10022-in. Centering disc (1 pc), PTFE, Segmented rigid single lead1.8 in. (45 mm)
03300-1656-10033-in. Centering disc (1 pc), PTFE, Segmented rigid single lead2.7 in. (68 mm)
03300-1656-10044-in. Centering disc (1 pc), PTFE, Segmented rigid single lead3.6 in. (92 mm)
03300-1656-10066-in. Centering disc (1 pc), PTFE, Segmented rigid single lead5.55 in. (141 mm)
03300-1656-10088-in. Centering disc (1 pc), PTFE, Segmented rigid single lead7.40 in. (188 mm)
03300-1656-30022-in. Centering disc (3 pcs), PTFE, Segmented rigid single lead1.8 in. (45 mm)
03300-1656-30033-in. Centering disc (3 pcs), PTFE, Segmented rigid single lead2.7 in. (68 mm)
03300-1656-30044-in. Centering disc (3 pcs), PTFE, Segmented rigid single lead3.6 in. (92 mm)
03300-1656-30066-in. Centering disc (3 pcs), PTFE, Segmented rigid single lead5.55 in. (141 mm)
03300-1656-30088-in. Centering disc (3 pcs), PTFE, Segmented rigid single lead7.40 in. (188 mm)
03300-1656-50022-in. Centering disc (5 pcs), PTFE, Segmented rigid single lead1.8 in. (45 mm)
03300-1656-50033-in. Centering disc (5 pcs), PTFE, Segmented rigid single lead2.7 in. (68 mm)
03300-1656-50044-in. Centering disc (5 pcs), PTFE, Segmented rigid single lead3.6 in. (92 mm)
03300-1656-50066-in. Centering disc (5 pcs), PTFE, Segmented rigid single lead5.55 in. (141 mm)
03300-1656-50088-in. Centering disc (5 pcs), PTFE, Segmented rigid single lead7.40 in. (188 mm)
Segmented rigid single lead probe spare part kit
03300-0050-000115.2 in. / 385 mm Segment for Top connection (1 pc)
03300-0050-000231.5 in. / 800 mm Segment (1 pc)
03300-0050-000331.5 in. / 800 mm Segment (3 pcs)
03300-0050-000431.5 in. / 800 mm Segment (5 pcs)
03300-0050-000531.5 in. / 800 mm Segment (12 pcs)
Emerson.com/Rosemount21
Rosemount 5300 Level TransmitterOctober 2019
Table 3: Accessories Ordering Information (continued)
Vented flanges
03300-1812-0092Fisher™ 249B, 259B
03300-1812-0093Fisher 249C
03300-1812-0091Masoneilan
Flushing connection rings
DP0002-2111-S62 in. ANSI, ¼ in. NPT connection
DP0002-3111-S63 in. ANSI, ¼ in. NPT connection
DP0002-4111-S64 in. ANSI/DN100, ¼ in. NPT connection
DP0002-5111-S6DN50, ¼ in. NPT connection
DP0002-8111-S6DN80, ¼ in. NPT connection
HART modem and cables
03300-7004-0001MACTek® VIATOR® HART Modem and cables (RS232 connection)★
03300-7004-0002MACTek VIATOR HART Modem and cables (USB connection)★
Remote housing mounting spare part kit
(3)(4)
(4)
™
03300-7006-00011 m / 3.2 ft. Remote Housing Mounting Cable and 316L Bracket
03300-7006-00022 m / 6.5 ft. Remote Housing Mounting Cable and 316L Bracket
03300-7006-00033 m / 9.8 ft. Remote Housing Mounting Cable and 316L Bracket
Verification reflector (high level supervision) spare part kit (requires Rosemount 5300 firmware version 2.H0 or later)
05300-7200-0001For 3- to 8-in. pipe/chamber (inner diameter)
05300-7200-0002For tanks or 10-in. pipe/chamber (inner diameter) or wider
If a centering disc is required for a flanged probe, the centering disc can be ordered with options Sx or Px in the model code. If a centering disc is
(1)
required for a threaded connection, or as a spare part, it should be ordered using the item numbers listed in this table. Refer to Table 33 for
centering disc size recommendation for different pipe schedules.
To order a centering disc in a different material consult the factory.
(2)
1-½ in. NPT threaded connection (RA) is required.
(3)
Not available with Country certification option code J1 (CRN).
(4)
22Emerson.com/Rosemount
October 2019Rosemount 5300 Level Transmitter
Specifications
Performance specifications
General
Reference conditions
Single Standard probe, 77 °F (25 °C) in water (DC=80) and ambient pressure in a 4-in. pipe using Trim Near Zone function.
Reference accuracy
± 0.12 in. (3 mm) or 0.03% of measured distance, whichever is greatest
For probes with spacers, the accuracy may deviate close to the spacers. Accuracy may be affected by remote housing.
Repeatability
± 0.04 in. (1 mm)
Ambient temperature effect
± 0.008 in. (0.2 mm) /°K or ± 30 ppm/°K of measured value, whichever is greatest
Electromagnetic interference effect
■
Shielded cable: ± 0.2 in. (5 mm)
■
Unshielded cable: ± 2 in. (50 mm)
For FOUNDATION™ Fieldbus units it may be required to ground the signal cable shield at the power supply and transmitter to achieve
optimum performance.
Thresholds may need to be adjusted, see section “Adjusting thresholds” of the Rosemount 5300 Reference Manual for general
guidelines on manual threshold settings.
Emission and Immunity: EMC directive 2014/30/EU, EN 61326-1:2013, and EN61326-3-1:2006.
NAMUR recommendations: NE21
CE-mark
Complies with applicable directives (EMC, ATEX).
Built-in lightning protection
EN 61326, IEC 61000-4-5, level 2kV (6kV with T1 terminal block)
(1) For the BR5 option code with ambient temperatures between -67 °F (-55 °C) and -40 °F (-40 °C), the ambient temperature effect is ±
0.012 in. (0.3 mm) /°K or ± 45 ppm/°K of measured value, whichever is greatest.
Deviation through electromagnetic interference according to EN 61326.
(2)
(3)
Namur NE21 not available with option code QT.
(3)
Emerson.com/Rosemount23
Rosemount 5300 Level TransmitterOctober 2019
Contamination/product build-up
■
Single lead probes are preferred when there is a risk of contamination (because build-up can result in the product bridging
across the two leads for twin versions; between the inner lead and outer pipe for the coaxial versions).
■
For viscous or sticky applications, PTFE probes are recommended. Periodic cleaning may also be required.
■
For viscous or sticky applications, it is not recommended to use centering discs mounted along the single lead probe.
■
Signal Quality Metrics (option code D01, or DA1) can be used to determine when to clean the probe. Transmitters equipped
with the Diagnostics Suite option can calculate Signal Quality Metrics.
Table 4: Maximum Recommended Viscosity and Contamination/Build-up
Probe typeMaximum viscosityContamination/build-up
Single lead8000 cP
Twin lead/Large coaxial1500 cPThin build-up allowed, but no bridging
Coaxial500 cPNot recommended
Consult your local Emerson representative in the case of agitation/turbulence and high viscous products.
(1)
Be cautious in HTHP viscous or crystallizing media applications where temperature at instrument connection is significantly lower than process
(2)
temperature with risk of coating in the upper part of probe that may reduce the measurement signal. Consider using HP or STD probes in such
applications.
(1)(2)
Build-up allowed
Measuring range
See Table 5 and Table 6 for each probe’s measuring range and minimum dielectric constant. Due to the measuring range
depending on the application and factors described below, the values are a guideline for clean liquids. For more information, ask
your local Emerson representative.
Note
For Remote Housing, see Table 7 for the maximum recommended measuring range for different Remote Housing lengths,
installation types, Dielectric Constants, and probe types.
Different parameters (factors) affect the echo and therefore the maximum measuring range differs depending on application
according to:
■
Disturbing objects close to the probe.
■
Media with higher dielectric constant (εr) gives better reflection and allows a longer measuring range.
■
Surface foam and particles in the tank atmosphere may affect measuring performance.
■
Heavy product build-up or contamination on the probe should be avoided since it can reduce measuring range and might cause
erroneous level readings.
Table 5: Maximum Measuring Range
Probe typeMaximum measuring range
Rigid single lead/segmented rigid single lead9 ft. 10 in. (3 m) for 8 mm probes (code 4A)
19 ft. 8 in. (6 m) for 13 mm probes (code 4B)
32 ft. 9 in. (10 m) for 13 mm probes (code 4S)
Flexible single lead164 ft. (50 m)
(1)
Coaxial19 ft. 8 in. (6 m)
Large coaxial19 ft. 8 in. (6 m)
Rigid twin lead9 ft. 10 in. (3 m)
Flexible twin lead164 ft. (50 m)
Maximum measuring range for Duplex 2205 probes type 5A and 5B is 105 ft. (32 m).
(1)
24Emerson.com/Rosemount
October 2019Rosemount 5300 Level Transmitter
Table 6: Minimum Dielectric Constant
Probe typeMinimum dielectric constant
StdHPHTHPC
Rigid single lead/
segmented rigid single
lead
Flexible single lead1.4, up to 49 ft. (15 m)
(1)(2)
1.4
(1.25 if installed in a metallic
bypass or stilling well)
(1)
1.8, up to 82 ft. (25 m)
2.0, up to 115 ft. (35 m)
(1)
(1)(3)
3, up to 138 ft. (42 m)
4, up to 151 ft. (46 m)
6, up to 164 ft. (50 m)
(1)(2)
1.6
(1.4 if installed in a metallic bypass or stilling well)
1.6, up to 49 ft. (15 m)
1.8, up to 82 ft. (25 m)
2.0, up to 115 ft. (35 m)
(1)
(1)
(1)(3)
3, up to 138 ft. (42 m)
4, up to 151 ft. (46 m)
6, up to 164 ft. (50 m)
Coaxial1.21.42.01.4
Large coaxial1.21.4N/A1.4
Rigid twin lead1.4N/AN/AN/A
Flexible twin lead1.4, up to 82 ft. (25 m)
2.0, up to 115 ft. (35 m)
2.5, up to 131 ft. (40 m)
(1)
N/AN/AN/A
(1)
(1)
3.5, up to 148 ft. (45 m)
6, up to 164 ft. (50 m)
Probe end projection software function will improve the minimum measurable dielectric constant. Consult the factory for details.
(1)
May be lower depending on installation.
(2)
Up to 49 ft. (15 m) for Duplex 2205 probes type 5A and 5B.
(3)
Table 7: Remote Housing Measuring Range, ft. (m)
DCRigid single
8 mm
Rigid single
13 mm/
segmented
Flexible
single
Coaxial/
Large
coaxial
rigid single
1 m remote
housing
(1)
Chamber / pipe
installations < 4 in.
(100 mm)
1.44 (1.25)19 (6)
210 (3)
(2)
19 (6)
8010 (3)19 (6)
(2)
(2)
(2)
33 (10)
33 (10)
33 (10)
(2)(3)
19 (6)10 (3)
(2)(3)
19 (6)10 (3)
(2)(3)
19 (6)10 (3)
Tank installations1.44 (1.25)4 (1.25)4 (1.25)19 (6)4 (1.25)4 (1.25)
24 (1.25)4 (1.25)4 (1.25)19 (6)4 (1.25)98 (30)
2 m remote
housing
(1)
Chamber / pipe
installations < 4 in.
(100 mm)
8010 (3)
(2)
1.49 (2.75)19 (6)
210 (3)
(2)
8010 (3)19 (6)33 (10)
10 (3)
19 (6)
(2)
(2)
(2)
159 (48.5)
(2)(3)
33 (10)
(2)(3)
33 (10)
(2)(3)
(2)
19 (6)10 (3)
19 (6)10 (3)
19 (6)10 (3)
19 (6)10 (3)
Tank installations1.49 (2.75)9 (2.75)9 (2.75)19 (6)9 (2.75)9 (2.75)
29 (2.75)9 (2.75)9 (2.75)19 (6)9 (2.75)98 (30)
8010 (3)
(2)
10 (3)
(2)
154 (47)
(2)
19 (6)10 (3)
Rigid twinFlexible
twin
(2)
(2)
(2)
(2)
(2)
(2)
(2)
(2)
33 (10)
33 (10)
33 (10)
159 (48.5)
33 (10)
33 (10)
33 (10)
154 (47)
(2)(3)
(2)(3)
(2)(3)
(2)
(2)
(2)(3)
(2)(3)
(2)(3)
(2)
(2)
Emerson.com/Rosemount25
1
35
7
9
11
0
16 (5)
33 (10)
49 (15)
66 (20)
82 (25)
98 (30)
115 (35)
A
B
C
Rosemount 5300 Level TransmitterOctober 2019
Table 7: Remote Housing Measuring Range, ft. (m) (continued)
DCRigid single
8 mm
Rigid single
13 mm/
segmented
Flexible
single
Coaxial/
Large
coaxial
Rigid twinFlexible
twin
rigid single
3 m remote
housing
Chamber / pipe
(1)
installations < 4 in.
(100 mm)
1.410 (3)19 (6)33 (10)
210 (3)19 (6)33 (10)
8010 (3)19 (6)33 (10)
Tank installations1.410 (3)14 (4.25)14 (4.25)19 (6)10 (3)
210 (3)14 (4.25)14 (4.25)19 (6)10 (3)
8010 (3)19 (6)
Validated for ambient temperature range -40 °F to 185 °F (-40 °C to 85 °C).
(1)
Accuracy may be affected up to + 1.2 in. (30 mm).
(2)
Required chamber/pipe size is 3 or 4 in. (75 -100 mm).
(3)
(2)
(2)(3)
(2)(3)
(2)(3)
149 (45.5)
19 (6)10 (3)
19 (6)10 (3)
19 (6)10 (3)
(2)
19 (6)10 (3)
(2)
(2)
(2)
(2)
(2)
(2)
(2)(3)
33 (10)
(2)(3)
33 (10)
(2)(3)
33 (10)
14 (4.25)
(2)
98 (30)
149 (45.5)
(2)
Interface measuring range
The maximum allowable upper product thickness/measuring range is primarily determined by the dielectric constants of the two
liquids.
Typical applications include interfaces between oil/oil-like and water/water-like liquids, with a low (<3) dielectric constant for the
upper product and a high (>20) dielectric constant for the lower product. For such applications, the maximum measuring range is
limited by the length of the coaxial, large coaxial, rigid twin, and rigid single lead probes.
For flexible probes, the maximum measuring range is reduced by the maximum upper product thickness, according to the diagram
below. However, characteristics may vary between the different applications. Maximum distance to the interface is 164 ft. (50 m)
minus the maximum upper product thickness.
Figure 2: Maximum Upper Product Thickness for the Flexible Single Lead Probe
A. Maximum upper product thickness, ft. (m)
B. Upper product dielectric constant
Lower product dielectric constant
C.
26Emerson.com/Rosemount
1
35
7
9
11
0
16 (5)
33 (10)
49 (15)
66 (20)
82 (25)
98 (30)
115 (35)
A
B
C
October 2019Rosemount 5300 Level Transmitter
Figure 3: Maximum Upper Product Thickness for the Flexible Twin Lead Probe
A. Maximum upper product thickness, ft. (m)
B. Upper product dielectric constant
Lower product dielectric constant
C.
Accuracy over measuring range
The measuring range depends on probe type, dielectric constant of the product and installation environment, and is limited by the
Blind Zones at the very top and bottom of the probe. In the Blind Zones, the accuracy exceeds ±1.18 in. (30 mm), and
measurements may not be possible. Measurements close to the Blind Zones will have reduced accuracy.
The following conditions will impact the Blind Zones:
■
If the single lead probes or twin probes are installed in a nozzle, the nozzle height shall be added to the specified Upper Blind
Zone.
■
The measuring range for the PTFE covered flexible single lead probe includes the weight when measuring on a high dielectric
media.
■
When using a metallic centering disc, the Lower Blind Zone is 8 in. (20 cm), including weight if applicable. When using a PTFE
centering disc, the Lower Blind Zone is not affected.
Figure 5 , Figure 7, and Figure 6 illustrate the accuracy over measuring range at reference condition with alternating probe types
and varying dielectric constant of the product.
Emerson.com/Rosemount27
$
%
%
&
C
AB
D
C
±0.12 in.
(3 mm)
3.5 in. (9 cm)
9.8 in. (25 cm)
4.7 in. (12 cm)
±1.18 in.
(30 mm)
±0.12 in.
(3 mm)
3.9 in. (10 cm)
9.8 in. (25 cm)
0.4 in. (1 cm)
±1.18 in.
(30
mm)
Rosemount 5300 Level TransmitterOctober 2019
Figure 4: Blind Zones
A. Upper Blind Zone
B. Reduced accuracy
Lower Blind Zone
C.
Note
Measurements may not be possible in the Blind Zones, and measurements close to the Blind Zones will have reduced accuracy.
Therefore, the 4-20 mA points should be configured outside these zones.
Figure 5: Accuracy Over Measuring Range for Single Lead Probes (Rigid/Segmented Rigid/Flexible)
A. Water (DC = 80)
B. Oil (DC = 2)
Accuracy
C.
D. Blind Zone
28Emerson.com/Rosemount
±0.12 in.
(3 mm)
3.2 in.(8 cm)
8.3 in.(21 cm)
4.7 in.(12 cm)
±1.18 in.
(30 mm)
±0.12 in.
(3 mm)
3.9 in. (10 cm)
9.8 in.(25 cm)
0.8 in.(2 cm)
±1.18 in.
(30 mm)
C
AB
D
C
±0.12 in.
(3 mm)
3.5 in. (9 cm)
7 in.(18 cm)
5.1 in.(13 cm)
±1.18 in.
(30 mm)
±0.12 in.
(3 mm)
2 in.(5 cm)
7.5 in.(19 cm)
0.8 in.(2 cm)
±1.18 in.
(30
mm)
C
AB
D
C
October 2019Rosemount 5300 Level Transmitter
Figure 6: Accuracy Over Measuring Range for Twin Lead Probes
A. Water (DC = 80)
B. Oil (DC = 2)
C.
Accuracy
D. Blind Zone
Figure 7: Accuracy Over Measuring Range for Coaxial Probe
A. Water (DC = 80)
B. Oil (DC = 2)
C.
D. Blind Zone
Accuracy
Emerson.com/Rosemount29
±0.12 in.
(3 mm)
3.9 in. (10 cm)
±1.18 in.
(30 mm)
±0.12 in.
(3 mm)
1.6 in.(4 cm)
±1.18 in.
(30
mm)
0.8 in.(2 cm)
5.1 in.(13 cm)
3.9 in.(10 cm)
C
AB
D
C
Rosemount 5300 Level TransmitterOctober 2019
Figure 8: Accuracy Over Measuring Range for Large Coaxial Probe
A. Water (DC = 80)
B. Oil (DC = 2)
C.
Accuracy
D. Blind Zone
Functional specifications
General
Field of application
Liquids and semi-liquids level and/or liquid/liquid interfaces or solids level
■
Model 5301, for liquid level or submerged interface measurements
■
Model 5302, for liquid level and interface measurements
■
Model 5303, for solid level measurements
Measurement principle
Time Domain Reflectometry (TDR)
Related information
Measurement principle
Microwave output power
Nominal 300 µW, Max. 45 mW
EMC
FCC part 15 subpart B and EMC Directive (2014/30/EU). Considered to be an unintentional radiator under the Part 15 rules.
Humidity
0 to 100% relative humidity
30Emerson.com/Rosemount