Rosemount 3095MFA Operating Manual

4.5 (2)

Reference Manual

00809-0100-4809, Rev AA

August 2002

The Annubar® Flowmeter Series

Model 3095MFA

Mass ProBar

Model 3051SFA

ProBar®

Model 485

Annubar® Primary

www.rosemount.com

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

The Annubar Flowmeter Series

NOTICE

Read this manual before working with the product. For personal and system safety, and for optimum product performance, make sure you thoroughly understand the contents before installing, using, or maintaining this product.

The United States has two toll-free assistance numbers and one International number.

Customer Central

1-800-999-9307 (7:00 a.m. to 7:00 P.M. CST)

International

1-(952) 906-8888

National Response Center

1-800-654-7768 (24 hours a day) Equipment service needs

The products described in this document are NOT designed for nuclear-qualified applications. Using non-nuclear qualified products in applications that require nuclear-qualified hardware or products may cause inaccurate readings.

For information on Rosemount nuclear-qualified products, contact your local Rosemount

Sales Representative.

This device is intended for use in temperature monitoring applications and should not be used in control and safety applications.

May be protected by one or more of the following U.S. Patent Nos.4,559,836; 4,717,159; 5,710,370; 5,773,726; 4,633, 713; and various foreign patents. Other foreign patents issued and pending.

www.rosemount.com

Reference Manual

00809-0100-4809, Rev AA

June 2002

Annubar Flowmeter Series

Table of Contents

SECTION 1

Introduction

SECTION 2

Installation

SECTION 3

Commissioning

Using This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1

Receiving and Inspection. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

Returning the Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

Considerations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2

Environmental. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-3

Process Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-4

Electrical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-5

Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1

Installation Flowchart and Checklist . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2

Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4

Tools and Supplies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4

Mounting Brackets . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4

Bolt Installation Guidelines . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4

Instrument Manifolds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5

Straight Run Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-6

Integral (Direct) Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-8

Remote Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-9

Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12

Pak-Lok Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-12

Flanged Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-18

Flange-Lok Model. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-23

Threaded Flo-Tap Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-28

Flanged Flo-Tap Model . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-34

Connect the Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-40

Wiring Diagrams . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-40

Equipment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-41

Field Wiring (Power and Signal) . . . . . . . . . . . . . . . . . . . . . . . . . . 2-42

Grounding. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-42

Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1 Commissioning on the Bench (Model 3051SFA Only) . . . . . . . . . . . . 3-2 Set the Switches. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2 Commissioning The Annubar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5 Direct Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-5 Remote Mount . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-10 Setting the Loop to Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14 Model 275 HART Communicator. . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14 Connections and Hardware . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14 Model 3051SFA Probar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-14 Updating the Model 275 HART Communication Software . . . . . . 3-15 HART Menu Tree . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-15 Fast Key Sequences . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17

www.rosemount.com

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

June 2002

SECTION 4

Operation and

Maintenance

APPENDIX A

Specifications and

Reference Data

Review Configuration Data. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-17

Check Output . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18

Basic Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-18

Detailed Setup . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20

Diagnostics and Service . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-20

Calibration . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-21

Advanced Functions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30

Model 3095MFA Mass Probar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-30

EA Software/ HART Communicator Comparison . . . . . . . . . . . . . 3-30

Engineering Assistant (EA) Software . . . . . . . . . . . . . . . . . . . . . . 3-31

Model 275 HART Communicator . . . . . . . . . . . . . . . . . . . . . . . . . 3-31

Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1

Model 3095MFA Mass Probar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2

Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2

Disassembly . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3

Remove the Flowmeter from Service . . . . . . . . . . . . . . . . . . . . . . . 4-3

Terminal Block . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3

Electronics Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4

RTD Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5

Replacing a RTD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5

Model 3051SFA ProBar Flowmeter . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1

Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-1

Functional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-2

Installation Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4

Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-4

Model 3095MFA Mass ProBar Flowmeter. . . . . . . . . . . . . . . . . . . . . . A-7

Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7

Functional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-7

Installation Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9

Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-9

Model 485 Annubar Primary . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12

Performance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12

Functional . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-12

Installation Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-13

Physical . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-13

Dimensional Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-16

Model 3051SFA Pak–Lok Probar . . . . . . . . . . . . . . . . . . . . . . . . . A-16

Model 3051SFA Flange–Lok Probar . . . . . . . . . . . . . . . . . . . . . . A-17

Model 3051SFA Flange Probar . . . . . . . . . . . . . . . . . . . . . . . . . . A-18

Model 3051SFA Flange Flo–Tap Probar . . . . . . . . . . . . . . . . . . . A-19

Model 3051SFA Threaded Flo–Tap Probar . . . . . . . . . . . . . . . . . A-20

Model 3095MFA Pak–Lok Mass ProBar . . . . . . . . . . . . . . . . . . . A-21

Model 3095MFA Flange–Lok Mass ProBar . . . . . . . . . . . . . . . . . A-22

Model 3095MFA Flange Mass ProBar . . . . . . . . . . . . . . . . . . . . . A-23

Model 3095MFA Flange Flo–Tap Mass ProBar . . . . . . . . . . . . . . A-24

Model 3095MFA Threaded Flo–Tap Mass ProBar. . . . . . . . . . . . A-25

Model 485 Pak–Lok Annubar . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-26

Model 485 Flange–Lok Annubar . . . . . . . . . . . . . . . . . . . . . . . . . A-27

Model 485 Flange Annubar . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-28

Model 485 Flange Flo–Tap Annubar . . . . . . . . . . . . . . . . . . . . . . A-29

TOC-ii

Reference Manual

00809-0100-4809, Rev AA

June 2002

Annubar Flowmeter Series

APPENDIX B

Approvals

Model 485 Threaded Flo–Tap Annubar . . . . . . . . . . . . . . . . . . . . A-30

Mounting. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-31

Ordering Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . A-32

Hazardous Locations Installations . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1

Hazardous Locations Certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1

Models 3051SFA . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-1

Model 3095MFA Mass ProBar . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-2

Installation Drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . B-4

Model 3051SFA ProBar Flowmeter . . . . . . . . . . . . . . . . . . . . . . . . B-4

Model 3095MFA Mass ProBar Flowmeter . . . . . . . . . . . . . . . . . . . B-4

TOC-iii

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

June 2002

TOC-iv

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

Section 1

Introduction

USING THIS MANUAL

Using This Manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-1 Receiving and Inspection . . . . . . . . . . . . . . . . . . . . . . . . . page 1-2 Returning the Product . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-2 Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 1-2

This product manual provides installation, configuration, calibration, troubleshooting, and maintenance instructions for the Annubar Flowmeter Series.

Section 2: Installation

Installation flowchart and checklist

Setting the Failure Model Alarm and Write Protect switches

Orienting, mounting, and installing the flowmeter

Connecting the Wiring

Commissioning the flowmeter according to the application

Section 3: Commissioning

Using the Model 275 HART® Communicator

Configuring the flowmeter using the Model 275 HART Communicator

Calibrating the flowmeter

Section 4: Operation and Maintenance

Troubleshooting information

Disassembly

RTD maintenance

Model 275 HART Communicator diagnostic messages

Appendix A: Specifications and Reference Data

Specifications

Dimensional drawings

Appendix B: Approvals

Approvals certifications

Installation drawings

www.rosemount.com

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

RECEIVING

AND INSPECTION

RETURNING THE PRODUCT

CONSIDERATIONS

Limitations

Flowmeters are available in different models and with different options, so it is important to inspect and verify that the appropriate model was delivered before installation.

Upon receipt of the shipment, check the packing list against the material received and the purchase order. All items are tagged with a model number, serial number, and customer tag number. Report any damage to the carrier.

To expedite the return process, call the Rosemount National Response Center toll-free at 800-654-7768. This center, available 24 hours a day, will assist you with any needed information or materials.

The center will ask for the following information:

Product model

Serial numbers

The last process material to which the product was exposed

The center will provide

A Return Material Authorization (RMA) number

Instructions and procedures that are necessary to return goods that were exposed to hazardous substances

NOTE

If a hazardous substance is identified, a Material Safety Data Sheet (MSDS), required by law to be available to people exposed to specific hazardous substances, must be included with the returned materials.

Information in this manual applies to circular pipes only. Consult Rosemount Customer Central for instructions regarding use in square or rectangular ducts.

Structural

Structural limitations are printed on the sensor tag. Exceeding structural limitations may cause sensor failure.

Functional

The most accurate and repeatable flow measurement occurs in the following conditions:

The structural limit differential pressure, as printed on the sensor tag, is not exceeded.

The instrument is not used for two-phase flow or for steam service below saturation temperature.

Install the flowmeter in the correct location within the piping branch to prevent measurement inaccuracies caused by flow disturbances.

1-2

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

The flowmeter can be installation with a maximum misalignment of 3 degrees (see Figure 1-1). Misalignment beyond 3 degrees will cause flow measurement errors.

Figure 1-1. Permissible

Misalignment

3° max.

28-490000-940A01A

3° max.

3° max.

Environmental

Mount the flowmeter in a location with minimal ambient temperature changes.

 

Appendix A: Specifications and Reference Data lists the temperature

 

operating limits. Mount to avoid vibration, mechanical shock, and external

 

contact with corrosive materials.

 

Access Requirements

 

Consider the need to access the flowmeter when choosing an installation

 

location and orientation.

 

Process Flange Orientation

 

Orient the process flanges on a remote mounted flowmeter so that process

 

connections can be made. For safety reasons, orient the drain/vent valves so

 

that process fluid is directed away from technicians when the valves are used.

 

In addition, consider the possible need for a testing or calibration input.

 

Housing Rotation

 

The electronics housing may be rotated up to 180 degrees (left or right) to

 

improve field access to the two compartments or to better view the optional

 

LCD meter. To rotate the housing, release the housing rotation set screw and

 

turn the housing up to 180 degrees.

1-3

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

NOTE

Rotating the housing more than 180 degrees will damage the sensor module and void the warranty.

Electronics Housing

Terminal Side

The circuit compartment should not routinely need to be opened when the unit is in service. Wiring connections are made through the conduit openings on the top side of the housing. The field terminal side is marked on the electronics housing. Mount the flowmeter so that the terminal side is accessible. A 0.75-in. (19 mm) clearance is required for cover removal. Use a conduit plug on the unused side of the conduit opening. A 3-in. (76 mm) clearance is required for cover removal if a meter is installed.

Exterior

The integral span and zero push-buttons are located under the certifications plate on the top of the ProBar. The plate will be blank if no certifications are ordered.

Cover Installations

Always install the electronics housing covers metal-to-metal to ensure a proper seal.

Figure 1-2. Electronics Housing

Rosemount Model 3051S Transmitter

Rosemount Model 3095 Transmitter

 

 

 

 

Process Considerations

3051S_COPLANAAR_3051A 01F, 3095-3095G05C

The process connections on the transmitter flange are 1/4–18 NPT. Flange adapter unions with 1/2–14 NPT connections are available as options. These are Class 2 threads; use the plant-approved lubricant or sealant when making the process connections. The process connections on the transmitter flange are on 21/8–in. (54 mm) centers to allow direct mounting to a threeor five-valve manifold. By rotating one or both of the flange adapters, connection centers of 2–, 21/8–, or 21/4–in. (51, 54, or 57 mm) may be obtained.

1-4

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

Failure to install proper flange adapter O-rings can cause process leaks, which can result in death or serious injury. There are two styles of Rosemount flange adapters, each requiring a unique O-ring, as shown below. Each flange adapter is distinguished by its unique groove.

 

MODEL 3051/2024/3001/3095

 

Flange Adapter

 

O-ring

 

Unique O-ring

 

Grooves

 

Use only the O-ring designed to seal with the corresponding flange adapter.

 

Refer to the factory for the correct part numbers of the flange adapters and

 

O-rings designed for the flowmeter.

 

Teflon® (PTFE) O-rings tend to cold flow when compressed, which aids in

 

their sealing capabilities. Whenever flanges or adapters are removed, visually

 

inspect the Teflon (PTFE) O-rings. Replace them if there are any signs of

 

damage. If the O-rings are replaced, the flange bolts may need to be

 

retorqued after installation to compensate for cold flow.

Electrical

The signal terminals are located in a compartment of the electronics housing.

 

Connections for the Model 275 HART Communicator are located below the

 

signal terminals. The Model 272 Field Calibrator can be connected at the

 

signal terminals to provide temporary power to the electronics for calibration

 

or diagnostic purposes. Otherwise, the calibrator may be attached to the test

 

connections on the terminal block of the electronics for indication purposes.

 

Power Supply

 

The dc power supply should provide power with less than 2% ripple. The total

 

resistance load is the sum of the resistance of the signal leads and the load

 

resistance of the controller, indicator, and related pieces. Note that the

 

resistance of intrinsic safety barriers, if used, must be included.

 

NOTE

 

A loop resistance between 250-1100 ohms is required to communicate with a

 

personal computer. With 250 ohms of loop resistance, a power supply voltage

 

of at least 16.5 V dc is required.(1)

 

If a single power supply is used to power more than one Model 3095MFA

 

Mass ProBar, the power supply used, and circuitry common to the Mass

 

ProBars, should not have more than 20 ohms of impedance at 1200 Hz.

(1)Quick troubleshooting check: There must be at least 11.0 V DC across the Model 3095MFA Mass ProBar electronics terminals.

1-5

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

1-6

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

Section 2

Installation

SAFETY MESSAGES

Safety Messages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-1 Installation Flowchart and Checklist . . . . . . . . . . . . . . . . page 2-2 Mounting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-4 Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-12 Connect the Wiring . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . page 2-40

Instructions and procedures in this section may require special precautions to ensure the safety of the personnel performing the operations. Please refer to the following safety messages before performing any operation in this section.

Explosions could result in death or serious injury:

Do not remove the transmitter cover in explosive atmospheres when the circuit is live.

Before connecting a Model 275 HART Communicator in an explosive atmosphere, make sure the instruments in the loop are installed in accordance with intrinsically safe or non-incendive field wiring practices.

Verify that the operating atmosphere of the transmitter is consistent with the appropriate hazardous locations certifications.

Both transmitter covers must be fully engaged to meet explosion-proof requirements.

Failure to follow these installation guidelines could result in death or serious injury:

Make sure only qualified personnel perform the installation.

www.rosemount.com

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

INSTALLATION FLOWCHART AND CHECKLIST

Figure 2-1. Installation Chart

Figure 2-1 is an installation flowchart that provides guidance through the installation process. Following the figure, an installation checklist has been provided to verify that all critical steps have been taken in the installation process. The checklist numbers are indicated in the flowchart.

Start.

Unpack Instrument

Review Product

Manual.

Verify proper location.

Hazardous

 

Review Appendix B.

Location?

 

 

 

 

 

 

 

 

 

 

 

 

 

Bench

Configure write-protect and

Configure?

failure alarm

 

Connect the bench power supply

 

Connect the instrument to a PC

 

Perform bench configuration tasks

 

(Optional) Perform bench

Verify model

calibration tasks

 

Remote

 

Mounted

Install hardware

Electronics?

 

Install flowmeter

 

 

Install electronics

Wire

 

Remote

Commission

Mounted

Electronics?

 

Commission

 

Finish.

 

2-2

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

The following list is a summary of the steps required to complete a flowmeter installation. If this a new installation, begin with step 1. If the mounting is already in place, verify that the hole size and the fittings match the recommended specifications (see Table 2-3 on page 2-13) and begin with step 5.

1.Determine where the flowmeter is to be placed within the piping system.

2.Establish the proper orientation as determined by the intended application.

3.Review Appendix B: Approvals and determine if the flowmeter is located in a hazardous location.

4.Confirm the configuration.

5.Drill the correct sized hole into the pipe.

For instruments equipped with opposite-side support, drill a second hole 180° from the first hole.

6.Weld the mounting and clean the burrs and welds.

7.Measure the pipe’s internal diameter (ID), preferably at 1 x ID from the hole (upstream or downstream).

NOTE

To maintain published flowmeter accuracy, provide the pipe ID when purchasing the flowmeter.

8.Check the fit-up of the instrument assembly to the pipe.

9.Install the flowmeter.

10.Wire the instrument.

11.Supply power to the flowmeter.

12.Perform a trim for mounting effects.

13.Check for leaks.

14.Commission the instrument

2-3

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

MOUNTING

Tools and Supplies

Mounting Brackets

Bolt Installation

Guidelines

Tools required include the following:

Open end or combination wrenches (spanners) to fit the pipe fittings and bolts: 9/16-in., 5/8-in., 7/8-in.

Adjustable wrench: 15-in. (1½-in. jaw).

Nut driver: 3/8-in. for vent/drain valves (or 3/8-in. wrench).

Phillip’s screwdriver: #1.

Standard screwdrivers: ¼-in., and 1/8-in. wide.

Pipe wrench: 14-in.

Wire cutters/strippers

7/16-in. box wrench (required for the ferry head bolt design)

Supplies required include the following:

½-in. tubing (recommended) or ½-in. pipe to hook up the electronics to the sensor probe. The length required depends upon the distance between the electronics and the sensor.

Fittings including (but not limited to)

Two tube or pipe tees (for steam or high temperature liquid) and

Six tube/pipe fittings (for tube)

Pipe compound or Teflon (PTFE) tape (where local piping codes allow).

Optional mounting brackets available with the instrument facilitate mounting to a panel, wall, or 2-in. (50.8 mm) pipe. The bracket option for use with the Coplanar flange is 316 SST with 316 SST bolts. See “Mounting” on page A-31 for bracket dimensions.

When installing the transmitter to one of the mounting brackets, torque the bolts to 125 in-lb (169 n-m).

The following guidelines have been established to ensure a tight flange, adapter, or manifold seal. Only use bolts supplied with the instrument or sold by the factory.

The instrument is shipped with the coplanar flange installed with four 1.75-in. (44.5 mm) flange bolts. The following bolts also are supplied to facilitate other mounting configurations:

Four 2.25-in. (57.2 mm) manifold/flange bolts for mounting the coplanar flange on a three-valve manifold. In this configuration, the 1.75-in. (44.5 mm) bolts may be used to mount the flange adapters to the process connection side of the manifold.

(Optional) If flange adapters are ordered, four 2.88-in. (73.2 mm) flange/adapter bolts for mounting the flange adapters to the coplanar flange.

2-4

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

Figure 2-2. Coplanar Mounting Bolts and Bolting Configurations for Coplanar Flange.

Stainless steel bolts supplied by Rosemount Inc. are coated with a lubricant to ease installation. Carbon steel bolts do not require lubrication. Do not apply additional lubricant when installing either type of bolt. Bolts supplied by Rosemount Inc. are identified by the following head markings:

Carbon Steel Head

 

B7M

 

 

 

 

Markings (CS)

 

 

 

 

 

 

 

 

 

 

 

Stainless Steel Head

316

316

B8M

STM

 

SW

Markings (SST)

R

 

316

316

316

 

 

Transmitter with 3-Valve

 

Transmitter with Optional

Manifold, Manifold/Flange

Transmitter with

Flange Adapters and

Bolts, Flange Adapters

Flange Bolts

Flange/Adapter Bolts

and Flange/Adapter bolts

 

 

 

Instrument Manifolds

 

 

 

 

3095B29A

 

 

2.25 (57)

4

3095E05A,

Flange/adapter bolts (4)

2.88 -in. (73 mm)

 

 

3095D05A,-3095

1.75 (44) 4

2.88 (73) 4

 

 

 

Description

Size in. (mm)

 

 

 

Flange bolts (4)

1.75 -in. (44 mm)

1.75 (44) 4

 

Manifold/flange bolts (4)

2.25 -in. (57 mm)

 

 

 

Figure 2-3 on page 2-6 identifies the valves on a 5-valve and a 3-valve manifold. Table 2-1 on page 2-6 explains the purpose of these valves.

An instrument manifold is recommended for all installations. A manifold allows an operator to equalize the pressures prior to the zero calibration of the electronics as well as to isolate the electronics from the rest of the system without disconnecting the impulse piping. Although a 3-valve manifold can be used, a 5-valve manifold is recommended.

5-valve manifolds provide a positive method of indicating a partially closed or faulty equalizer valve. A closed faulty equalizer valve will block the DP signal and create errors that may not be detectable otherwise. The labels for each valve will be used to identify the proper valve in the procedures to follow.

NOTE

Some recently-designed instrument manifolds have a single valve actuator, but cannot perform all of the functions available on standard 5-valve units. Check with the manufacturer to verify the functions that a particular manifold can perform. In place of a manifold, individual valves may be arranged to provide the necessary isolation and equalization functions.

2-5

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

Figure 2-3. Valve Identification

for 5-valve and 3-Valve 5-Valve Manifold 3-Valve Manifold

Manifolds

To PH

To PL

 

To PH

To PL

 

MV

 

 

ME

MH

ML

MH

 

ML

 

2

 

 

2

MEH

MEL

 

 

 

DVH

DVL

DVH

DVL

 

 

 

1

 

1

8900_8900_35A

Table 2-1. Description of

Impulse Valves and

 

 

Name Description

Purpose

Components

 

 

Manifold and Impulse Pipe Valves

 

PH

Primary Sensor – High Pressure

PL

Primary Sensor – Low Pressure

DVH

Drain/Vent Valve – High Pressure

DVL

Drain/Vent Valve – Low Pressure

MH

Manifold – High Pressure

ML

Manifold – Low Pressure

MEH

Manifold Equalizer – High Pressure

MEL

Manifold Equalizer – Low Pressure

Isolates the flowmeter sensor from the impulse piping system

Drains (for gas service) or vents (for liquid or steam service) the DP electronics chambers

Isolates high side or low side pressure from the process.

Allows high and low pressure side access to the vent valve, or for isolating the process fluid

ME

Manifold Equalizer

Allows high and low side pressure to equalize

MV

Manifold Vent Valve

Vents process fluid

Components

 

1

Electronics

Reads Differential Pressure Isolates and

2

Manifold

equalizes electronics.

3

Vent Chambers

Collects gases in liquid applications.

4

Condensate Chamber

Collects condensate in gas applications.

Straight Run

Requirements

Use the following to aid in determining the straight run requirements

NOTE

For gas service, multiply values from Table 2-2 on page 2-7 by 1.5.

If longer lengths of straight run are available, position the mounting such that 80% of the run is upstream and 20% is downstream.

The information contained in this manual is applicable to circular pipes only. Consult the factory for instructions regarding use in square or rectangular ducts.

Straightening vanes may be used to reduce the required straight run length.

Row 5 in Table 2-2 is to be used if a “through type” valve will remain open. Row 6 in Table 2-2 applies to gate, globe, plug, and other throttling valves that are partially opened, as well as control valves.

2-6

Rosemount 3095MFA Operating Manual

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

Table 2-2. Straight Run

Requirements

Figure 2-4. Mounting

Configuration

 

 

Upstream dimension

 

 

Downstream

 

Without vanes

With vanes

 

Dimensions

1.

In plane A Out of plane A

A’

C

C’

B

 

 

 

 

 

 

-0573B

8

10

4

 

 

 

 

 

 

1295

8

4

4

4

 

2.

 

 

 

 

 

 

0573C

11

16

4

 

 

 

 

 

 

1295-

8

4

4

4

 

 

 

 

 

 

3.

 

 

 

 

 

 

-0573D

23

28

4

 

 

 

 

 

 

1295

8

4

4

4

 

4.

 

 

 

 

 

 

-0573E

12

12

4

8

4

4

4

1295

 

 

 

 

 

 

5.

 

 

 

 

 

 

0573F

18

18

4

 

 

 

 

 

 

1295-

8

4

4

4

 

 

 

 

 

 

6.

 

 

 

 

 

 

-0573G

30

30

4

 

 

 

 

 

 

1295

8

4

4

4

 

Integral Mount

 

 

Remote Mount

 

 

 

Electronics

 

 

 

 

Electronics

 

 

 

 

Mounting

Mounting

Configuration

Configuration

 

Sensor

Sensor

28-490000-945A01A, 946A01A

2-7

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

Integral (Direct) Mount

NOTE

The integral mount flowmeter is usually shipped with the electronics bolted directly to the sensor. If this is not so, contact Rosemount Customer Central for more information.

Horizontal Pipes

Liquid or Steam Applications

Due to the possibility of air getting trapped in the probe, the sensor should be located according to Figure 2-5 for liquid or steam applications. The area between 0° and 30° angle should not be used unless full bleeding of air from the probe is possible.

For liquid applications, mount the side drain/vent valve upward to allow the gases to vent.

In steam applications, fill the lines with water to prevent the steam from contacting the electronics. Condensate chambers are not required because the volumetric displacement of the electronics is negligible.

Air and Gas Applications

Figure 2-5 illustrates the recommended location of the flowmeter in air or gas applications. The sensor should be located on the upper half of the pipe, at least 30° above the horizontal line.

For air and gas applications, mount the drain/vent valve downward to allow liquid to drain.

Figure 2-5. Horizontal Pipe

Applications

Liquid or Steam Applications

Air or Gas Applications

 

30

30

 

Recommended

909A01A

degrees

degrees

 

 

 

 

 

120 degrees

30

Zone 120

909A02A,

 

30

 

 

 

degrees

 

 

Recommended

degrees

degrees

-

 

Zone

 

 

490000-9

 

 

 

 

Vertical Pipes

Liquid, Steam, Air, and Gas Applications

Figure 2-6 illustrates the recommended location of the flowmeter in liquid, air, or gas applications.

The sensor can be installed in any position around the circumference of the pipe, provided the vents are positioned properly for bleeding or venting. Vertical pipe installations require more frequent bleeding or venting, depending on the location.

2-8

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

Figure 2-6. Vertical Pipe

Liquid or Steam Applications

Air or Gas Applications

Applications

 

 

 

 

Flow

 

360

 

 

degrees

 

Flow

360 degrees

Remote Mount

Instrument head connections differ between horizontal and vertical pipes. Consult your specification head code number to confirm the proper pipe orientation.

Valves and Fittings

Throughout the remote mounting process:

Use only valves, fittings, and pipe thread sealant compounds that are rated for the service pipeline design pressure and temperature as specified in Appendix A: Specifications and Reference Data.

Verify that all connections are tight and that all instrument valves are fully closed.

Verify that the sensor probe is properly oriented for the intended type of service: liquid, gas or steam (see Figures “Integral (Direct) Mount” on page 2-8 and “Remote Mount” on page 2-9).

Impulse Piping

Impulse piping connects remote mounted electronics to the sensor. Temperatures in excess of 250 °F (121 °C) at the electronics will damage electronics components; impulse piping allows service flow temperatures to decrease to a point where the electronics is no longer vulnerable.

The following restrictions and recommendations apply to impulse piping location.

Piping used to connect the sensor probe and electronics must be rated for continuous operation at the pipeline-designed pressure and temperature

Impulse piping that runs horizontally must slope at least 1–in. per foot (83mm/m).

It must slope downwards (toward the electronics) for liquid and steam applications.

It must slope up (away from the electronics) for gas applications.

For applications where the pipeline temperature is below 250 °F

(121 °C), the impulse piping should be as short as possible to minimize flow temperature changes. Insulation may be required.

2-9

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

For applications where pipeline temperature is above 250 °F (121 °C), the impulse piping should have a minimum length of 1-ft. (0.30 m) for every 100 °F (38 °C) over 250 °F (121 °C), which is the maximum operating electronics temperature. Impulse piping must be uninsulated to reduce fluid temperature. All threaded connections should be checked after the system comes up to temperature, because connections may be loosened by the expansion and contraction caused by temperature changes.

A minimum of 1/2-in. (12mm) outer diameter (OD) stainless steel tubing with a wall thickness of at least 0.035-in. is recommended.

Outdoor installations for liquid, saturated gas, or steam service may require insulation and heat tracing to prevent freezing.

For installations where the electronics are more than 6-ft. (1.8m) from the sensor probe, the high and low impulse piping must be run together to maintain equal temperature. They must be supported to prevent sagging and vibration.

Threaded pipe fittings are not recommended because they create voids where air can become entrapped and have more possibilities for leakage.

Run impulse piping in protected areas or against walls or ceilings. If the impulse piping is run across the floor, ensure that it is protected with coverings or kick plates. Do not locate the impulse piping near high temperature piping or equipment.

Use an appropriate pipe sealing compound rated for the service

temperature on all threaded connections. When making threaded connections between stainless steel fittings, Loctite® PST® Sealant is recommended.

Figure 2-7. Liquid Service

Horizontal

Vertical

29-490000-941A01A, 942A01A

2-10

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

Figure 2-8. Gas Service

Horizontal

Vertical

29-490000-9

Figure 2-9. Steam Service

Horizontal

Vertical

29-490000-980A02A, 981A01A

2-11

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

INSTALLATION

This manual contains the horizontal and vertical installation procedures for

 

the Pak-Lok, Flanged, Flange-Lok, and Threaded Flow-Tap Annubar models.

Pak-Lok Model

Figure 2-10 identifies the components of the Pak-Lok assembly.

Figure 2-10. Components

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Transmitter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Coplanar Flange with Drain

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Vents

Direct Mount Electronics

 

 

 

 

Connection with Valves

 

 

 

O-Rings (2)

Nuts

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Compression Plate

 

 

 

 

Follower

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Packing Rings (3)

 

 

 

 

 

 

 

 

Retaining Ring

 

 

 

 

 

 

Studs

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Pak-Lok Body

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Opposites Side Support

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Annubar Sensor

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

(optional)

 

 

 

 

28-49000-956A, 900A

Step 1: Set the Switches

Refer to “Mounting” on page 2-4 for more information

Step 2: Determine the Proper Orientation

Please refer to “Mounting” on page 2-4 for straight run requirements and orientation information.

Step 3: Drill a Hole into the Pipe

Follow the steps below to drill the hole in the pipe.

1.Depressurize and drain the pipe.

2.From the previous steps, select the location to drill the hole.

3.Determine the diameter of the hole to be drilled according to the specifications in Table 2-3 and drill the hole. Do not torch cut the hole.

2-12

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

Table 2-3. Drill Hole into Pipe

 

 

 

Sensor Size / Hole Diameter Chart

 

 

Sensor Diameter

 

 

T1

3/4-in.

+ 1/32-in (1 mm)

 

(19 mm)

– 0.00

 

 

 

T2

15/16-in.

+ 1/16-in. (1 mm)

 

(34 mm)

– 0.00

 

 

 

T3

21/2-in.

+ 1/16-in. (1 mm)

 

(64 mm)

– 0.00

 

 

 

 

 

 

Note: Drill the hole 180 degrees from the first hole for opposite-side support models.

Drill the appropriate diameter hole through the pipe wall.

8900-8900_15A

4.If opposite-side support coupling is supplied, a second identically sized hole must be drilled opposite the first hole so that the sensor can pass completely through the pipe. (To determine a opposite-side support model, measure the distance from the tip of the first slot or hole. If the distance is greater than 1-in. (25.4 mm), it is the opposite-side model.) To drill the second hole, follow these steps:

a.Measure the pipe circumference with a pipe tape, soft wire, or string (for the most accurate measurement the pipe tape needs to be perpendicular to the axis of flow).

b.Divide the measured circumference by two to determine the location of the second hole.

c.Rewrap the pipe tape, soft wire, or string from the center of the first hole. Then, using the number calculated in the preceding step, mark the center of what will become the second hole.

d.Using the diameter determined from Table 2-3, drill the hole into the pipe with a hole saw or drill. Do not torch cut the hole.

5.Deburr the drilled hole(s) on the inside of the pipe.

Step 4: Weld the Mounting Hardware

1.Center the Pak-Lok body over the mounting hole, gap 1/16-in. (1.5 mm) and place four 1/4-in. (6-mm) tack welds at 90° increments. Check alignment of the Pak-Lok body both parallel and perpendicular to the axis of flow. If alignment of mounting is within tolerances (see Figure 2-11), finish weld per local codes. If alignment is outside of specified tolerance make adjustments prior to finish weld.

2-13

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

Figure 2-11. Alignment

 

Tack

LMH

Welds

 

28-490000_906A04A

2.If opposite side support is being used, center the fitting for the opposite

side support over the opposite side hole, gap 1/16-in. (1.5 mm) and place

four 1/4-in. (6 mm) tack welds at 90° increments. Insert the sensor into the mounting hardware. Verify that the tip of the bar is centered in the opposite side fitting and verify that the plug will fit around bar. If the bar is centered in the fitting and plug fits around the bar, finish weld per local codes. If the alignment of the bar does not allow enough clearance to insert the opposite side plug, make the necessary adjustments prior to making the finish weld.

NOTE

To avoid serious burns, allow the mounting hardware to cool before continuing.

Step 5: Insert into the Pipe

After the mounting hardware has cooled, use the following steps for installation.

1.Thread studs into the Pak-Lok body.

2.To ensure that the flowmeter contacts the opposite side wall, mark the

tip of the sensor with a marker. (Do not mark if the sensor was ordered with special-cleaned option code P2.)

3.Rotating the flowmeter back and forth, insert the flowmeter into the Pak-Lok body until the sensor tip contacts the pipe wall (or support plug).

4.Remove the flowmeter.

5.Verify that the sensor tip made contact with the pipe wall by removing the pipe and ensuring that some of the marker has been rubbed off. For special-cleaned bars, look for wear marks on the tip. If the tip did not touch the wall, verify pipe dimensions and the height of mounting body from the OD of the pipe and re-insert.

6.Re-insert the flowmeter into the Pak-Lok body and install the first packing ring on the sensor between the lock ring and the packing follower. Do not damage the split packing rings.

7.Push the packing ring into the Pak-Lok body and against the weld lock ring. Repeat this process for the two remaining rings, alternating the location of the packing ring split by 180°.

2-14

Reference Manual

00809-0100-4809, Rev AA

August 2002

Annubar Flowmeter Series

Figure 2-12. Packing Ring Detail

Compression

 

 

Plate

 

Retaining

 

Ring

 

 

Follower

Packing

Ring

28-490000_942A01A

8.Tighten the nuts onto the studs:

Place the included split-ring lock washer between each of the nuts and the compression plate. Give each nut one half (1/2) turn in succession until the split-ring lock washer is flat between the nut and the compression plate. Inspect the unit for leakage; if any exists, tighten the nuts in one-quarter (1/4) turn increments until there is no leakage.

NOTE

On sensor size (1), failure to use the split-ring lock washers, improper washer orientation, or over-tightening the nuts may result in damage to the flowmeter.

Figure 2-13. Split-Ring Lock

 

Washer Orientation

Stud

 

Nut

 

Split ring

 

lock washer

 

Compression

 

Plate

Before Tightening

After Tightening

Stud

Nut

Split ring lock washer

Compression Plate

28-490000_943A01A

NOTE

Pak-Lok sealing mechanisms generate significant force at the point where the sensor contacts the opposite pipe wall. Caution needs to be exercised on thin-walled piping (ANSI Schedule 10 and below) to avoid damage to the pipe.

Step 6: Mount the Transmitter

Direct Mount Head

With Valves

Place Teflon (PTFE) O-rings into grooves on the face of head.

Align the high side of the transmitter to the high side of the probe (“Hi” is stamped on the side of the head) and install.

Tighten the nuts in a cross pattern to 400 in•lb (45 N•m).

2-15

Annubar Flowmeter Series

Reference Manual

00809-0100-4809, Rev AA

August 2002

Without Valves

Place Teflon (PTFE) O-rings into grooves on the face of head.

Orient the equalizer valve or valves so they are easily accessible. Install manifold with the smooth face mating to the face of the head. Tighten in cross pattern to a torque of 400 in•lb (45 N•m).

Place Teflon (PTFE) O-rings into grooves on the face of the manifold.

Align the high side of the transmitter to the high side of the probe (“Hi” is stamped on the side of the head) and install.

Tighten the nuts in a cross pattern to 400 in•lb (45 N•m).

Remote Mount Head – temperatures below 250 °F (121 °C)

See “Remote Mount” on page 2-9 for more information.

Liquid Applications

Secure the electronics below the sensor to ensure that air will not be introduced into the impulse piping or the electronics.

Gas Applications

Secure the electronics above the sensor to prevent condensable liquids from collecting in the impulse piping and the DP cell.

28_490000_931A01A, 932A01A

Remote Mount Head – temperature above 250 °F (121 °C)

Liquid or Steam Applications

The electronics must be mounted below the process piping. Route the impulse piping down to the electronics and fill the system with cool water through the two tee fittings.

2-16

Loading...
+ 160 hidden pages