With Rosemount 3051S High Pressure Solutions, operations can be optimized in these critical areas: production, quality, energy efficiency, and safety and environment. By leveraging the power of the scalable Rosemount 3051S across the entire operation, you’ll be able to minimize process variability, gain greater process insight, reduce maintenance and downtime, and meet regulatory demands. What’s more, it’s easy to use, ensuring the full potential of the measurement investment is realized.
Rosemount 3051S High Pressure

Rosemount 3051S SuperModule™ Platform

The most advanced pressure, flow, and level measurements
The all-welded hermetic design delivers the industry's highest field reliability
Patented electronics within the SuperModule
SIL 3 capable: IEC61508 certified by an accredited third party agency for use in safety instrumented
systems up to SIL 3 (minimum requirement of single use [1oo1] for SIL2 and redundant use [1oo2] for SIL 3)

Rosemount 3051S High Pressure Solutions

Rosemount 3051S High Static Differential Pressure Transmitter
Coned and threaded direct connection enables operation in applications with static pressures up to
15,000 psi (1,034 bar)
Coplanar platform enables integrated seal system solutions
Calibrated Differential Pressure spans from 5 inH
Dual-capacitance Saturn
Available with Alloy C-276 process isolators
sensor technology corrects for overpressure and line pressure effects
O to 150 psi (12,4 mbar to 10,4 bar)
2
April 2018
Contents
Rosemount 3051S High Static Differential Pressure
Transmitter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
2
Product certifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . .17
Dimensional drawings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Emerson.com/Rosemount
April 2018
Rosemount 3051S High Pressure

Advanced functionality

WirelessHART® (IEC 62591) capabilities
Available on coplanar, in-line, and level transmitters
Quickly deploy new pressure, level and flow measurements in 70 percent less time
Eliminate wiring design and construction complexities to lower costs by 40 to 60 percent
Extended range antenna capabilities provide access to remote locations
Delivering over a decade of maintenance free performance with 15-year stability and 10-year
power module life
Advanced diagnostic capabilities
Provides diagnostic coverage from the process to the transmitter to the host
Prevent on-scale failures by diagnosing electrical loop issues with power advisory diagnostics
Statistical process monitoring detects abnormal process conditions enabling more productive
and safer operations
Extend diagnostic coverage to Safety Instrumented Systems (SIS) with IEC 61508 SIL 2/3 capable
rating
Additional functionality
Remote display and interface allows for direct mounting to process to eliminate impulse lines
and enables access to the transmitter's interface from 100 ft. away
Optional differential pressure + temperature measurement option available on the Rosemount
3051SHP reduces overall installation cost with a 2-in-1 DP and process temperature measurement
Emerson.com/Rosemount
3
Rosemount 3051S High Pressure
April 2018

Rosemount 3051S High Static Differential Pressure Transmitter

Rosemount 3051S High Static Differential Pressure Transmitters bring reliability based on the industry leading Rosemount 3051S SuperModule design to installations with high static pressure conditions. The lightweight coplanar design results in a more accurate measurement while reducing transmitter weight over traditional high static pressure transmitters. Capabilities include:
Accurate and reliable measurements up to static lines pressure of 15,000 psi (1.034 bar)
4–20 mA HART
Safety certification (options code QT)
Advanced diagnostics (option code DA2)
Differential pressure and temperature (measurement type option code 7)
Specification and selection of product materials, options, or components must be made by the purchaser of the equipment. See page 15 for more information on material selection.
Table 1. Rosemount 3051S High Static Differential Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
®
, WirelessHART, FOUNDATION™ Fieldbus Protocols
Additional information Specifications: page 10 Certifications: page 17 Dimensional drawings: page 22
Rosemount model
3051SHP High Static Differential Pressure Transmitter
Performance class
1 Ultra: 0.055% span accuracy, 15-yr limited warranty
2 Classic: 0.055% span accuracy
(1)
Connection type
C Coplanar
Measurement type
D Differential pressure
7 Differential pressure and temperature
Differential pressure range
6 –250 to 250 inH2O (–623 to 623 mbar)
7 –700 to 700 inH2O (–1,74 to 1,74 bar)
8 –150 to 150 psi (–10,34 to 10,34 bar)
Static pressure range
A None
Maximum static line pressure
3 15,000 psi (1,034 bar)
4
Emerson.com/Rosemount
April 2018
Rosemount 3051S High Pressure
Table 1. Rosemount 3051S High Static Differential Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Temperature input
N None
R RTD input (type Pt 100, –328 to 1562 °F [–200 to 850 °C])
Isolating diaphragm
(2)(3)
3 Alloy C-276
Process connection
(4)(5)
B12
Assemble to two Rosemount 1199 Seals
H11 Coned and threaded, compatible with autoclave type F-250-C
Transmitter output
A 4–20 mA with digital signal based on HART Protocol
(6)
F
(7)
X
Housing style Material
1A Plantweb housing Aluminum
1B Plantweb housing Aluminum M20 x 1.5
1C Plantweb housing Aluminum G1/2
1J Plantweb housing
1K Plantweb housing SST M20 x 1.5
1L Plantweb housing SST G1/2
2A Junction Box housing Aluminum
2B Junction Box housing Aluminum M20 x 1.5
FOUNDATION Fieldbus Protocol
Wireless (requires wireless options and wireless Plantweb™ housing)
Stainless steel (SST)
Conduit entry size
1
/2–14 NPT
1
/2–14 NPT
1
/2–14 NPT
2C Junction Box housing Aluminum G1/2
2E Junction Box housing with output for remote display and interface Aluminum
1
/2–14 NPT
2F Junction Box housing with output for remote display and interface Aluminum M20 x 1.5
2G Junction Box housing with output for remote display and interface Aluminum G1/2
2J Junction Box housing SST
2M Junction Box housing with output for remote display and interface SST
(8)
5A
5J
7J
(8)
(8)
Wireless Plantweb housing Aluminum
Wireless Plantweb housing SST
Quick connect (A size mini, 4-pin male termination)
SST N/A
1
/2–14 NPT
1
/2–14 NPT
1
/2–14 NPT
1
/2–14 NPT
Emerson.com/Rosemount
5
Rosemount 3051S High Pressure
Table 1. Rosemount 3051S High Static Differential Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
April 2018
Wireless options (requires option code X and wireless Plantweb housing)
Update rate
WA User-configurable update rate
Operating frequency and protocol
3 2.4 GHz DSSS, IEC 62591 (WirelessHART Protocol)
Omni-directional wireless antenna
WK External antenna
WJ Remote antenna
WM Extended range, external antenna
WN High-gain, remote antenna
SmartPower
(8)
1 Adapter for black power module (I.S. power module sold separately)
Other options (Include with selected model number)
Extended product warranty
WR3 Three-year limited warranty
WR5 Five-year limited warranty
Plantweb control functionality
A01 FOUNDATION Fieldbus advanced control function block suite
Plantweb diagnostic functionality
(9)
D01
DA2
(10)(11)
FOUNDATION Fieldbus diagnostics suite
Advanced HART diagnostics suite
RTD cable (RTD sensor must be ordered separately)
C12 RTD Input with 12 ft. (3,66 m) of shielded cable
C13 RTD Input with 24 ft. (7,32 m) of shielded cable
C14 RTD Input with 75 ft. (22,86 m) of shielded cable
C22 RTD Input with 12 ft. (3,66 m) of armored shielded cable
C23 RTD Input with 24 ft. (7,32 m) of armored shielded cable
(9)
C24 RTD Input with 75 ft. (22,86 m) of armored shielded cable
C32 RTD Input with 12 ft. (3,66 m) of ATEX/IECEx flameproof cable
C33 RTD Input with 24 ft. (7,32 m) of ATEX/IECEx flameproof cable
C34 RTD Input with 75 ft. (22,86 m) of ATEX/IECEx flameproof cable
Mounting bracket
B4 Coplanar flange bracket, all 316 SST, 2-in. pipe and panel/ bracket
6
Emerson.com/Rosemount
April 2018
Rosemount 3051S High Pressure
Table 1. Rosemount 3051S High Static Differential Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
Software configuration
C1 Custom software configuration (requires Configuration Data Sheet)
Alarm level
(11)(12)
C4 NAMUR alarm and saturation levels, high alarm
C5 NAMUR alarm and saturation levels, low alarm
C6
Custom alarm and saturation signal levels, high alarm (requires C1 and Configuration Data Sheet)
C7 Custom alarm and saturation signal levels, low alarm (requires C1 and Configuration Data Sheet)
C8
Hardware adjustments
Low alarm (standard Rosemount alarm and saturation levels)
(11)(12)(13)
D1 Hardware adjustments (zero, span, alarm, security)
Ground screw
(12)
D4 External ground screw assembly
Conduit plug
(14)
DO 316 SST conduit plug
Customer specified bar code tag
D6 Bar code tag
Product certifications
(15)
E1 ATEX Fla mep roof
E2 INMETRO Flameproof
E5 US Explosion-proof, Dust Ignition-proof
(16)
E6
Canada Explosion-proof, Dust Ignition-proof, Division 2
E7 IECEx Flameproof, Dust Ignition-proof
EM Technical Regulation Customs Union (EAC) Flameproof and Dust
I1 ATEX Intrinsic Safety
I5 US Intrinsically Safe; Nonincendive
I6 Canada Intrinsically Safe
I7 IECEx Intrinsic Safety
IA ATEX FISCO Intrinsic Safety (FOUNDATION Fieldbus Protocol only)
IE US FISCO Intrinsically Safe (FOUNDATION Fieldbus Protocol only)
IF Canada FISCO Intrinsically Safe (FOUNDATION Fieldbus Protocol only)
IG IECEx FISCO Intrinsic Safety (FOUNDATION Fieldbus Protocol only)
IM Technical Regulation Customs Union (EAC) Intrinsic Safety
K1 ATEX Flameproof, Intrinsic Safety, Type n, Dust
K5 US Explosion-proof, Dust Ignition-proof, Intrinsically Safe, Division 2
Emerson.com/Rosemount
7
Rosemount 3051S High Pressure
April 2018
Table 1. Rosemount 3051S High Static Differential Pressure Transmitter Ordering Information
The starred offerings () represent the most common options and should be selected for best delivery. The non-starred offerings are subject to additional delivery lead time.
K6
(17)
Canada Explosion-proof, Dust Ignition-proof, Intrinsically Safe, Division 2
K7 IECEx Flameproof, Dust Ignition-proof, Intrinsic Safety, Type n
KA
KB
(17)
(17)
ATEX and Canada Flameproof, Intrinsically Safe, Division 2
US and Canada Explosion-proof, Dust Ignition-proof, Intrinsically Safe, Division 2
KC US and ATEX Explosion-proof, Intrinsically Safe, Division 2
(17)
KD
US, Canada, and ATEX Explosion-proof, Intrinsically Safe
KG US, Canada, ATEX and IECEx FISCO Intrinsic Safety
KM Technical Regulations Customs Union (EAC) Flame-proof, Intrinsic Safety
N1 ATEX Type n
N7 IECEx Type n
ND ATEX Dus t
Display type
(18)
M5 Plantweb LCD display
(11)(17)(19)
M7
M8
M9
(11)(20)
(11)(20)
Remote mount LCD display and interface, Plantweb housing, no cable, SST bracket
Remote mount LCD display and interface, Plantweb housing, 50 ft. (15 m) cable, SST bracket
Remote mount LCD display and interface, Plantweb housing, 100 ft. (31 m) cable, SST bracket
Pressure testing
P1 Hydrostatic testing with certificate
Calibration certification
Q4 Calibration certificate
Material traceability certification
Q8 Material traceability certification per EN 10204 3.1B
Quality certification for safety
(11)(13)
QS
(21)
QT
Transient protection
Prior-use certificate of FMEDA data
Safety-certified to IEC 61508 with certificate of FMEDA data
(20)(22)
T1 Transient terminal block
Toolkit performance reports
QZ Remote seal system performance calculation report
NACE® certificate
(20)
Q15 Certificate of compliance to NACE MR0175/ISO 15156 for wetted materials
Q25 Certificate of compliance to NACE MR0103 for wetted materials
Typical model number: 3051SHP 1CD6A3N3 H11A1A DA2 B4 M5
8
Emerson.com/Rosemount
April 2018
1. For detailed specifications see “Specifications” on page 10.
2. Materials of construction comply with metallurgical requirements highlighted within NACE MR0175/ISO 15156 for sour oil field production environments. Environmental limits apply to certain materials. Consult latest standard for details. Selected materials also conform to NACE MR0103 for sour refining environments. Order with Q15 or Q25 to receive a NACE certificate.
3. Isolator diaphragm selection will dict ate materials of construction for wetted parts.
4. Consult factory for available Rosemount 1199 high pressure remote seal options.
5. “Assemble to” items are specified separately and require a completed model number.
6. Requires Plantweb housing.
7. Only intrinsically safe approval codes apply.
8. Long-life power module must be shipped separately, order power module 701PBKKF.
9. Only available with output code F.
10. Only available with output code A.
11. Requires Plantweb housing and output code A. Includes hardware adjustments as standard.
12. This assembly is included with approval options EP, KP, E1, N1, K1, ND, E4, E7, N7, K7, E2, E3, KA, KC, KD, IA, IB, IE, IF, IG, KG, K2, N3, EM, and KM. It is also included with transient protection T1.
13. Not available with housing style codes 00, 01, 2E, 2F, 2G, 2M, 5A, 5J, or 7J.
14. Transmitter is shipped with 316 SST conduit plug (uninstalled) in place of standard carbon steel conduit plug.
15. Valid when SuperModule platform and housing have equivalent approvals.
16. Not available with M20 or G
17. Not available with output code F, option code DA2, or option code QT.
18. Not available with Housing code 7J.
19. See the Rosemount 3051S Reference Manual
20. NACE compliant wetted materials are identified by Footnote 2.
21. Not available with output code F or X. Not available with housing code 7J.
22. Not available with housing code 00, 5A, 5J, or 7J.
1