In the course of 2006 the Roche OMNI S system was relaunched under the
Roche Diagnostics professional IVD user brand cobas®.
Systems with a serial number of 5001 or above are cobas b 221 systems.
Systems with a serial number up to 5000 are Roche OMNI S systems.
Every effort has been made to ensure that all the information contained in this
manual is correct at the time of printing. However, Roche Diagnostics GmbH reserves
the right to make any changes necessary without notice as part of ongoing product
development.
Any customer modification to the instrument will render the warranty or service
agreement null and void.
Software updates are done by Roche Service representatives.
The contents of this document may not be reproduced in any form or communicated
to any third party without the prior written consent of Roche Diagnostics.
While every effort is made to ensure its correctness, Roche Diagnostics assumes no
responsibility for errors or omissions which may appear in this document.
Subject to change without notice.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.01
Brands
Contact addresses
cobas b 221 system
COBAS, COBAS B, LIFE NEEDS ANSWERS, ROCHE OMNI, AUTOQC,
ROCHE MICROSAMPLER, COMBITROL and AUTO-TROL are trademarks of
Roche.
Edition
Manufacturer
Roche Diagnostics GmbH
D-68298 Mannheim / Germany
www.roche.com
Revision 10.0, April 2009
First edition: May 2003
REF/No. 03261395001
Roche DiagnosticsApril 2009
2Instructions for Use · Revision 10.0
cobas b 221 system
Table of contents
Revision History 1
Edition notice 1
Copyright 1
Brands 2
Contact addresses 2
Edition 2
Tab l e o f co n ten t s 3
Preface 5
How to use this manual 5
Where to find information 5
Conventions used in this manual 5
Introduction and specificationsPart A
1 Safety information
Important information A-5
Operating safety information A-6
2 General descriptions
Introduction A-9
General notes A-11
Measurement and calibration procedure A-13
Measurement evaluation A-14
Safety instructions for specific dangers A-14
Handling solutions A-15
Handling electrodes A-15
General notes on the use of the MSS cassette A-16
System description A-18
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.03
TroubleshootingPart D
11 Troubleshooting
Troubleshooting - general D-5
System stops D-5
Module stops D-12
System warnings D-16
Status messages of measuring and calibration values
D-20
Status messages on the measurement report D-39
Barcode D-40
AppendixPart E
12 List of consumables
Order information E-5
Glossary E-9
cobas b 221 system
IndexPart F
Index F-3
Roche DiagnosticsApril 2009
4Instructions for Use · Revision 10.0
cobas b 221 system
Preface
The cobas b 221 system is an analyzer with integrated AutoQC drawer option.
This manual has detailed descriptions of cobas b 221 system features and general
operational concepts, specification functions and use of controls, operating
techniques, emergency procedures, product labeling and maintenance procedures.
How to use this manual
o
Keep this manual in a safe place to ensure that it is not damaged and remains available for use.
o
This Instructions for Use should be easily accessible at all times.
To help you find information quickly, there is a table of contents at the beginning of
the book and each chapter. In addition, a complete index can be found at the end.
Where to find information
In addition to the Instructions for Use, the following documents are also provided to
assist in finding desired information quickly:
ocobas b 221 system Reference Manual
ocobas b 221 system Short Instruction
Conventions used in this manual
Visual cues are used to help locate and interpret information in this manual quickly.
This section explains formatting conventions used in this manual.
SymbolsHelping to locate and interpret information in this manual the following symbols are
used:
SymbolUsed for
aProcedural step
oList item
e
hCall up of screen
Cross-reference
Note
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.05
SymbolUsed for
Caution
All sections / passages that are marked with this symbol describe procedures
and/or indicate conditions or dangers that could damage or lead to a
malfunction in the cobas b 221 system, and which therefore should never be
attempted and contain information that must be observed to avoid potential
injuries (to patients, users and third parties).
Risk of infection
All sections and parts of texts that are marked with this symbol describe
procedures that may involve risk of infection.
cobas b 221 system
IVD symbolsThe symbols are used in accordance with DIN EN 980
SymbolDescription
Conformité Européenne:
This product complies with the requirements in the guideline for
In Vitro Diagnostic 98/79/EC.
Lot designation
Use by...
The product should not be used after expiry of the specified date.
If a day is not indicated, apply the last day of the respective month.
Temperature limitation
The conditions necessary to preserve the product's shelf life before
opening.
In Vitro Diagnostic Medical Device
Manufacturer
(according to In Vitro Diagnostic guidelines 98/79/EG)
Catalogue number
(a)
and DIN EN ISO 780
(b)
.
Serial number (model plate)
Caution, consult accompanying documents
Please consult instructions for use
(a) DIN EN 980: Medical devices - Symbols to be used with medica l device labels, labelling and information
to be supplied (Part 1: General requirements)
(b) DIN EN ISO 780: Packaging - Pictorial marking for the handling of goods
Roche DiagnosticsApril 2009
6Instructions for Use · Revision 10.0
cobas b 221 system
SymbolDescription
Biological risk!
(according to the standard IEC/EN 61010-2-101)
Biological risk!
(according to the standard DIN EN ISO 980)
Do not use if package damaged
Do not reuse
Fragile. Handle with care
Handle with care
(a)
(Instrument)
(b)
(Consumables)
Valid only for Roche MICROSAMPLER:
Method of sterilization using ethylene oxide
Valid only for BS2 Blood Sampler:
Method of sterilization using irradiation
(a) IEC/EN 61010-2-101: Safety requirements for electrical equipment for measurement, control, and
laboratory use - (Part 2-101: Particular requirements for in vitro diagnostic (IVD) medical equipment).
(b) DIN EN ISO 980: Medical devices - Symbols to be used with medical device labels, labelling and
information to be supplied (Part 1: General requirements).
Other symbolsThe following symbols are listed as additional information:
SymbolDescription
Electrodes:
This date indicates the limit of the maximum storage time of
an electrode. The electrode must be installed in the instrument
no later than the imprinted date.
If the installation takes place on the imprinted date, it still falls
within the specifications. The calculation of the “Install
before” date is based on the production date of the elctrode.
Danger symbol: "Irritant" (on the label and the packaging of
S2 Fluid Pack)
Rating: Although not corrosive, momentary, longer-lasting, or
repeated contact with skin or mucous membrane may result in
inflammation. Danger of sensitization during contact with
skin (when classified with R 43).
Caution: Avoid contact with eyes and skin, do not inhale
vapors.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.07
SymbolDescription
Invisible Laser Radiation
Avoid direct radiation to eyes!
Laser Class 3R according to EN 60825-1
P0 ≤ 5 mW
λ = 635 - 850 nm
Store upright
"Grüner Punkt" (in Germany)
Protective gloves, protective goggles and suitable protective
clothing must be worn.
AbbreviationsThe following abbreviations are used:
cobas b 221 system
AbbreviationDefinition
A
ANSIAmerican National Standards Institute
AQCAutomatic Quality Control
B
BGBlood gas
BUNAbbr. for blood urea nitrogen
C
CLIAClinical Laboratory Improvement Amendments
CLSIClinical and Laboratory Standards Institute
condConductivity
CSACanadian Standards Association
D
dBADecibel weighted against the A-frequency response curve. This curve
approximates the audible range of the human ear.
DILDiluent
DNSDomain Name Server
E
ECEuropean community
e.g.exempli gratia – for example
ENEuropean standard
F
FMSFluid mixing system
H
HctHematrocrit
HIVHuman immunodeficiency virus
HWHardware
Roche DiagnosticsApril 2009
8Instructions for Use · Revision 10.0
cobas b 221 system
AbbreviationDefinition
I
i.e.id est – that is to say
ISEIon selective electrode
IVDIn vitro Diagnostic Directive
L
LCDLiquid cristal display
LISLaboratory Information System
LJLevey Jennings
M
MACMedia Access Control
MCMeasuring chamber
MSDSMaterial safety data sheet
MSSMetabolite sensitive sensor
MVMean value
P
PPPeristaltic pump
Q
QCQuality control
R
RConReference contact
REFReference solution
S
SIPSample inlet path
SDCSample distributor cartridge
S1S1 Rinse Solution
S2S2 Fluid Pack
S3S3 Fluid Pack
SConSensor contact
SDStandard deviation
SO
T
2
Oxygen saturation
T&DTurn & dock
tHbTotal hemoglobin
U
ULUnderwriters Laboratories Inc.
V
VDE Association of German Electrical Engineers (Verband Deutscher
Elektrotechniker)
e
For writing the measuring, calculated and input values see Chapter 9 Softwaremodi >
Parameter on page B-75!
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.09
cobas b 221 system
Roche DiagnosticsApril 2009
10Instructions for Use · Revision 10.0
The information provided in this chapter is essential for the safe, trouble-free
operation of the instrument and must be read and understood by the user.
In this chapter
Important information ...................................................................................................5
Operating safety information .........................................................................................6
Chapter
1
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-3
1 Safety informationcobas b 221 system
Contents
Roche DiagnosticsApril 2009
A-4Instructions for Use · Revision 10.0
cobas b 221 system1 Safety information
Important information
Important information
These Instructions for Use contain vital warnings and safety information.
This instrument is intended to be used only for the specialized purpose described in
the instructions. The most important prerequisites for use, operation, and safety are
explained to ensure smooth operation. No warranty or liability claims will be covered
if the machine is used in ways other than those described or if the necessary
prerequisites and safety measures are not observed.
The instrument may be operated only by persons whose qualifications enable them to
comply with the safety measures that are necessary during operation of the
instrument.
Suitable protective equipment, like laboratory clothing, protective gloves, protective
goggles and if necessary mouth protectors, must be worn to prevent direct contact
with biological working materials. In addition, a face mask is required if there is a
risk.
Adjustments and maintenance performed with covers removed and power connected
may be attempted only by a qualified technician who is aware of the associated
dangers.
Instrument repairs are to be performed only by the manufacturer or qualified service
personnel.
Only accessories and supplies either delivered by or approved by Roche are to be used
with the instrument. These items are manufactured especially for use with this
instrument and meet the highest quality requirements.
Operation of the instrument with solutions whose composition is not consistent with
that of the original solutions can negatively affect the long-term measurement
accuracy. Deviations in the composition of the solutions can also decrease the service
life of the electrodes.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each exchange
of solutions and packs and after startup of the instrument.
Additionally complete a quality control test on one level between two automatic 2P
calibrations. The level have to be alternated (low, normal, high).
Since the measurements of the instrument depend not only on the correct
characteristic function, but also on a series of marginal conditions (e.g. pre-analysis),
results obtained from the instrument should be submitted for an expert opinion
before taking additional measures based on the supplied measurements.
Caution (refer to accompanying documents)!
Please refer to safety-related notes in the manual accompanying this instrument.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-5
1 Safety informationcobas b 221 system
Operating safety information
Operating safety information
The instrument has been constructed and tested according to the following European
Standards:
It was delivered from the factory in flawless condition with regards to safety features.
In order to preserve this condition and ensure safe operation, the user must respect
the notices and warnings that are contained in these Instructions for Use.
oThis equipment is a Class I laser product, and it complies with FDA Radiation
Performance Standards, 21 CFR Subchapter J (only valid for
cobas b 221<1> system, cobas b 221<3> system and cobas b 221<5> system with
tHb/SO
oThis instrument is classified under the protection class I according to
IEC /EN 61010-1.
oThe instrument meets the conditions for overvoltage category II.
oThe instrument meets the conditions for contamination level 2.
oDo not operate the instrument in an explosive environment or in the vicinity of
explosive anesthetic mixtures containing oxygen or nitrous oxide.
oIf objects or liquids enter the internal areas of the instrument, remove the
instrument from its power supply and allow an expert to check it thoroughly
before using it again.
oThe instrument is suitable for long-term operation indoors.
module).
2
o
The power cord must be plugged into a grounded power receptacle. When using an extension
cord, make sure it is properly grounded.
o
Any rupture of the ground lead inside or outside the instrument or a loose ground connection
may result in hazardous operating conditions for the operating personnel. Intentional
disconnection of the grounding is not permitted.
o
The instrument is not suitable for operation with a direct current power supply. Use only the
original power plug delivered with the cobas b 221 system.
o
The use of controls or adjustments or performance of procedures other than those specified
herein may result in hazardous radiation exposure.
Roche DiagnosticsApril 2009
A-6Instructions for Use · Revision 10.0
cobas b 221 system2 General descriptions
Contents
General descriptions
This chapter contains a general description of the instrument, as well as
precautionary measures against special dangers and the proper handling of sensors,
solutions and the MSS cassette.
Warning and identification labels (incl. nameplate) .......................................24
Roche DiagnosticsApril 2009
A-8Instructions for Use · Revision 10.0
cobas b 221 system2 General descriptions
Introduction
Introduction
Figure A-1cobas b 221 system
The cobas b 221 system is an analyzer with integrated AutoQC drawer option.
Depending on combination and configuration, the following parameters can be
measured in whole blood, serum, plasma, acetate and bicarbonate containing dialysis
solutions and QC materials:
opH
oBlood gas BG (PO
oElectrolyte ISE (Na
, PCO2)
2
+
, K+, Cl–, Ca2+)
oHematocrit (Hct)
oMetabolite MSS
Urea/BUN - only cobas b 221<6> system
oTo t al h em og lo b in ( tH b)
oOxygen saturation (SO
oHemoglobin derivative COOX (O
)
2
Hb, HHb, COHb, MetHb)
2
oBilirubin (neonatal)
The following configurations are available:
ocobas b 221<1> system
(a)
BG, pH, tHb/SO
2
ocobas b 221<2> system BG, pH, COOX, Bili
ocobas b 221<3> system
(a)
BG, pH, ISE, Hct, tHb/SO2
ocobas b 221<4> system BG, pH, ISE, Hct, COOX, Bili
ocobas b 221<5> system
(a)
BG, pH, ISE, Hct, MSS, tHb/SO
2
ocobas b 221<6> system BG, pH, ISE, Hct, MSS, COOX, Bili
(a) are no longer manufactured or offered.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-9
2 General descriptionscobas b 221 system
Introduction
During the measurement or calibration or other processes, it is possible to conduct
database operations, perform certain settings or call up general information at the
same time.
e
For details see Chapter 9 Software modes
The individual, mutually independent software modes are defined as follows:
oSetupInstrument settings
oDatabaseData about patients, measurements, calibrations, QC, and
the instrument
oInfo
Roche DiagnosticsApril 2009
A-10Instructions for Use · Revision 10.0
cobas b 221 system2 General descriptions
General notes
General notes
Application area
The instrument has been tested for measuring parameters in whole blood, serum,
plasma and dialysis solutions (electrolytes only) and the validity of measurements was
tested accordingly.
In order to achieve accurate measurements of recommended aqueous control
solutions (with regards to deviations from biological samples), choose the proper
components and make the corresponding corrections in the QC measurement mode.
The accuracy of measurement values of undefined aqueous solutions cannot be
guaranteed (e.g. due to the possibility of interfering components and/or missing or
insufficient buffer systems, and/or differences in ionic strength and diffusion
potential when compared to biological samples).
Operating instructions
The cobas b 221 system should be switched on at all times!
If the instrument is switched off for an extended period of time (more than 24 hours),
a shutdown must be performed.
e
For additional information, see Chapter 3 Installation and shutdown, section Installation
on page A-27 and Shutdown on page A-48.
Prevent any other liquids from entering the instrument except samples and
QC material at the fill port.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each exchange
of solutions and packs and after startup of the instrument.
Additionally complete a quality control test on one level between two automatic
2P calibrations. The level have to be alternated (low, normal, high).
e
For additional information, see Chapter 7 Quality control.
With Software V 6.0 onwards, using cobas bge link, the instrument can be monitored
from one location, any disturbances can be remedied and the analytical quality
monitored.
cobas bge link is a remote monitoring and remote maintenance software for Roche
Point-of-Care analyzers.
e
see Figure A-2 on page A-12!
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-11
2 General descriptionscobas b 221 system
General notes
B
A
A "Screen sharing" SymbolB "Screen sharing" active
Figure A-2
Confirm the message with [OK] either on the instrument or on the PC. The "screen
sharing" symbol is added in the status line.
To avoid multiple operation of the instrument, the message "Screen sharing active" is
displayed with a yellow background in the error and message window of the
instrument.
As long as the "screen sharing" symbol is displayed in the status line, the service connection is
active. In order to prevent multiple operation of the instrument, no buttons on the screen should be
pressed!
Important buttons on the screen
ButtonsDescription
"Analyzer" active / inactive
"Database" active / inactive
"Setup" active / inactive
"Info" active / inactive
e
For additional information, see Chapter 9 Software modes, section Buttons on page B-76.
Roche DiagnosticsApril 2009
A-12Instructions for Use · Revision 10.0
cobas b 221 system2 General descriptions
Measurement and calibration procedure
Measurement and calibration procedure
Measurement procedure
PO2: Use of the Clark measurement principle: measurement of current generated by
the reduction of oxygen.
PCO
: Use of the Severinghouse principle: potentiometric measurement of the pH
2
change in the electrode caused by CO
pH- , Na
+
-,K+-, Ca2+- und Cl- electrodes are potentiometric electrodes. Special
glasses are used as the sensitive element for pH and Na
membranes contain special neutral carriers. A special ion exchanger is used for
chloride membranes. Calculation of these variables also requires the use of a reference
electrode—a permanently contacted chloride electrode in the cobas b 221 system.
Glucose, lactate: Glucose oxidizes to form gluconolacton using atmospheric oxygen
and the glucose-oxidase (GOD) enzyme, lactate oxidizes to form pyruvate using the
lactate oxidase enzyme.
The generated H
is determined amperometrically by using manganese dioxide/
2O2
carbon electrode at 350 mV.
.
2
+
. The potassium and calcium
Calibration procedure
Oxygen (O
PCO
, pH, ISE:are calibrated using two solutions mixed under different conditions, thereby avoiding
2
):Ambient air and a zero point solution are used to calibrate oxygen.
2
MSS:The calibration is carried out with four (Glu, Lac) or five solutions (Urea/BUN)
Urea: Urea is broken into ammonia and carbon dioxide through urease. Ammonia
and carbon dioxide react through hydrolysis with physiological pH to form ammonia
or bicarbonate ions. The ammonia ions can be determined using a potentiometrical
ammonia ion-selective electrode. This measurement requires a reference electrode
such as those used in ion-selective electrodes.
tHb/SO
: Light absorption in whole blood is measured at four different wavelengths,
2
the sample is subjected to light radiation and the dispersed light is also evaluated.
COOX: The hemoglobin derivatives and the total bilirubin (= neonatal) are
determined spectrophotometrically based on the Lambert-Beer law.
Hematocrit: Measurement of the sample's conductivity in the ISE measuring
chamber.
tHb and SO2 was calibrated when the instrument was manufactured.
the gas supply which is required by other instruments.
whose weighing concentrations form the basis for measured value determination.
COOX:Determining the hemoglobin derivatives and the total bilirubin (= neonatal) are
carried out spectral-photometrically using a cuvette.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-13
2 General descriptionscobas b 221 system
Measurement evaluation
Measurement evaluation
The validity of the test results from the cobas b 221 system must be carefully
examined by a clinical-medical specialist who will take the patient's clinical condition
into consideration before any clinical decisions are reached based on the test results.
In order to ensure the quality of the measurement results, complete a quality control
test on 3 levels (low, normal, high) after each electrode exchange, after each exchange
of solutions and packs and after startup of the instrument.
Additionally complete a quality control test on one level between two automatic 2P
calibrations. The level have to be alternated (low, normal, high).
e
For detailed information, see Chapter 7 Quality control.
Safety instructions for specific dangers
Handling samples
While handling samples, all necessary regulations concerning hygiene must be
observed. Dangerous pathogenic agents could be present.
e
For more detailed information, see Chapter 6 Measurement
Disposal of waste water, bottles, packs, electrodes and the instrument
Dispose of waste water, bottles, packs, electrodes and the instrument according to local and/or labor
regulations (biologically contaminated—hazardous waste!).
Decontamination
The purpose of this decontamination is to minimize risk when handling items that
were in contact with biological samples.
Roche recommends following a decontamination procedure in addition to
regulations specific to the laboratory.
These decontamination procedures should be performed periodically to minimize the
risk of infections.
Always wear gloves!
e
For more detailed information about decontamination, see Chapter 10 Maintenance
Roche DiagnosticsApril 2009
A-14Instructions for Use · Revision 10.0
cobas b 221 system2 General descriptions
Handling solutions
Handling solutions
Store the cobas b 221 system wash/calibrating solutions according to the specified
packaging requirements. The temperature of the solutions should be adapted to the
ambient temperature before use.
The shelf life of the solutions is limited.
Please read the bottle label and the packaging for the correct storage temperature and
the maximum shelf life.
DO NOT FREEZE!
If frozen, the solution's concentration may change and cause calibration errors!
Do not use damaged fluid packs (S2 and S3)! Do not mix the individual components!
e
For "Storage specifications", see Chapter 4 Specifications.
Handling electrodes
Store the electrodes according to the packaging specifications.
The shelf life of the electrodes is limited.
Please read the label and the packaging for the correct storage temperature and the
maximum shelf life.
CAUTION! Installation note for the PCO2 electrode
Insert the electrode into the measuring chamber within 5 minutes of opening the ALU-PE
packaging.
A special protective gas atmosphere designed to condition the PCO2 electrode during storage is
found inside the ALU-PE packaging.
This gas atmosphere ensures immediate potential stability during insertion of the electrode into the
measuring chamber and immediate readiness for measuring the first 2 point calibration.
If more than 5 minutes elapse after opening the ALU-PE packaging, the level of gas conditioning
could be lost and the time required for the first-time calibration could be increased.
e
For "Storage specifications", see Chapter 4 Specifications.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-15
2 General descriptionscobas b 221 system
General notes on the use of the MSS cassette
General notes on the use of the MSS cassette
For instrument versions with MSS module only!
Attention:
MSS cassette may only be brought into contact with liquids in the cobas b 221 system while
electrodes are changed!
Replace the MSS cassette within 28 days of installation!
After initial contact with liquids, the MSS cassette may no longer be removed from the instrument.
It may lead to the destruction of the enzyme sensors.
Storage:
At 2 – 8 °C, maximum of 2 weeks at room temperature.
MSS cassette removed from the measuring chamber
Once an MSS cassette is exposed to liquid, it must not be allowed to dry out under
any circumstances since this would destroy the enzymes. The enzymes are equipped
with a special protectant prior to shipping for transportation purposes. This
protectant is washed out inside the instrument during the warm-up phase and MSS
polarization.
Incompatible substances
The following substances may not be introduced into the MSS measuring chamber
under any circumstances since they would immediately destroy the MSS sensors or
severely impact their functionality.
oDeproteinizer (NaOCl)
oO
oCleaning solution
oNa electrode conditioning solution
oRinse additive
oSolutions containing heavy metals (Ag, Hg, Au, etc., e.g. Thiomersal)
oCleaning solutions containing detergent (e.g. washing material or liquid
oAll solutions for disinfections (e.g. high-percentage alcohol, glutaric dialdehyde,
oSolutions with pH values that deviate greatly from neutral
zero point solution
2
detergents)
cresol, etc.)
(e.g. pH value of < 6.0 and > 9.0)
The use of anticoagulants other than those approved by Roche Diagnostics
(approved: heparin salts), such as EDTA, citrate, NH4 heparin and glycolysis
inhibitor such as NaF and oxalate can lead to erroneous results.
Roche DiagnosticsApril 2009
A-16Instructions for Use · Revision 10.0
cobas b 221 system2 General descriptions
General notes on the use of the MSS cassette
Inserting the MSS cassette
Hold the MSS cassette only at the designated handle and avoid touching the contacts.
e
For a detailed description see Chapter 10 Maintenance, section Changing the MSS cassette
(cobas b 221<5> system and cobas b 221<6> system only) on page C-32.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-17
2 General descriptionscobas b 221 system
System description
System description
Visual identification
For example:cobas b 221<6> system
P
A
O
N
B
C
D
D
E
A Screen/PC unit
B Reverse side
C Docking mechanism
D AutoQC drawer
E Barcode scanner
F W Waste container
Figure A-3cobas b 221<6> system
G S1 Rinse Solution
H S2 Fluid Pack
IS3 Fluid Pack
JBottle compartement cover
K Bottle compartement
L COOX module (tHb/SO
FGHI
module)
2
M
L
K
J
M Input unit
N Measuring chambers
O Printer
P Pumps
Roche DiagnosticsApril 2009
A-18Instructions for Use · Revision 10.0
cobas b 221 system2 General descriptions
System description
Screen/PC unit
The screen/PC unit serves as the graphical user interface.
All information (results, error messages, alarms, warnings, etc.) is displayed on the
screen. The screen consists of a color LCD that is covered with a touch-sensitive film
("touch screen").
As sharp objects can damage the touch-sensitive film, only touch the film using suitable pins and/or
with your fingers.
The screen/PC unit also contains a diskette drive.
Printer
Low-noise thermoprinter with integrated paper cutter (manually activated using the
"Cut" key) and optional winder.
The "Feed" key feeds in the paper.
Measuring chamber
tHb/SO2 module
With an installed winder, the "Automatic Cut" function is deactivated.
Underneath the top cover are the BG and, depending on the configuration, ISE
measuring chamber with the electrodes, the MSS measuring chamber with the MSS
cassette and the tHb/SO
or COOX module.
2
The electrodes are flow-through electrodes with a visible sample channel.
Figure A-4tHb/SO2 module
The tHb/SO2 module is an optical sensor module for determining the level of total
hemoglobin (tHb) and oxygen saturation (SO
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-19
) in whole blood.
2
2 General descriptionscobas b 221 system
System description
COOX module
The COOX module consists of the hemolyzer and the COOX measuring chamber.
The measurement is based on the principle of spectral photometry.
Pumps
Depending on the configuration, up to three peristaltic pumps transport the sample
and the operating fluids inside the instrument.
Input unit
The sample insertion as well as the aspiration of solutions is carried out via input unit
which consists of the following:
oT&D module:
oT&D disk
oT&D tubing set with wash-water jet
oPlug control
oFill port
oSample drip tray
Bottle compartment
Behind the bottle compartment cover are the S1 Rinse Solution bottle, the
S2 Fluid Pack, the W Waste Container and, depending on the configuration,
S3 Fluid Pack (cobas b 221<5> system and cobas b 221<6> system only).
Roche DiagnosticsApril 2009
A-20Instructions for Use · Revision 10.0
cobas b 221 system2 General descriptions
System description
Reverse side
E
Power supply
A
B
A Power supply
B Main power switch and connector
C Warning and identification labels
Figure A-5Reverse side
D Air filter
E Interfaces
This unit also contains the main power switch and the connector.
D
C
B
A
A Power supply
B Main power switch OFF
C Main power switch ON
Figure A-6Power supply
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-21
C
2 General descriptionscobas b 221 system
System description
Interfaces
Only data processing units manufactured according to the standards IEC 950
(UL1950) may be attached to the interface connections!
ABCD EF
A Power supply
B Service connector
C RS 232
Figure A-7Interfaces (without USB)
ABCDEF
A Power supply
B Service connector
C Ext. keyboard/barcode scanner
Figure A-8Interfaces (with USB)
D Ext. keyboard/barcode scanner
E RS 232
F 10BaseT
D RS 232
E USB
F 10BaseT
oVar ia nt 2 :
2x RS 232 interfaces (COM 1 and COM 2) (SN < 1500)
e
see Figure A-7 on page A-22
oVar ia nt 1 :
1x RS 232 interface (COM 1) and 1x USB (SN > 1500)
e
see Figure A-8 on page A-22
o1x 10BaseT Ethernet (RJ45)
oExt. keyboard / barcode scanner: PS/2 DIN - 6 pin female connector
o1 service connector
oPower (power supply is connected)
No reverse compatibility from Variant 2 to Variant 1 possible.
Roche DiagnosticsApril 2009
A-22Instructions for Use · Revision 10.0
cobas b 221 system2 General descriptions
System description
Barcode scanner
Figure A-9Barcode scanner
oScanning of electrode data (type, lot, expiration date)
oScanning of patient or user identity
oScanning of QC data (QC material, lot, basis, expiration date, target values, etc.)
oScanning of desired alphanumeric code
Press the button on the underside to activate the scanner! A beeping sound and a brief illumination
of the LED on the upper side indicate the successful scanning of the barcode.
For more detailed information, please see enclosed manual of the PS2 hand-held scanner
(included in scope of delivery).
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-23
2 General descriptionscobas b 221 system
System description
Warning and identification labels (incl. nameplate)
Figure A-10cobas b 221<1> system, cobas b 221<3> system and cobas b 221<5> system (with tHb/SO2 module)
Figure A-11cobas b 221<2> system, cobas b 221<4> system and cobas b 221<6> system
Roche DiagnosticsApril 2009
A-24Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Contents
Installation and shutdown
In this chapter, the software-guided installation and shutdown of the instrument are
described step by step. The sequence of the steps described must be strictly followed.
Roche DiagnosticsApril 2009
A-26Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
Installation
Location
For best results, a suitable, level location that is not subject to direct sunlight is
required for the instrument.
When installing an instrument that was stored in a cool room or was transported at
low temperatures, be aware that condensation may have formed and could cause
disturbances to the instrument. The instrument must be climatized at room
temperature for at least one hour before beginning operation.
The following conditions must be fulfilled:
oAmbient temperature: 15 °C to 31 °C
oAmbient air pressure: 797 - 526 mmHg (106.225 - 70.13 kPa)
From approx. 3000 m above sea level or air pressure < 526 mmHg (70.13 kPa), the specifications
for parameter PO
of the clinical decisions.
After successful installation, the parameter must be permanently deactivated.
are no longer fulfilled and the parameter must no longer be used for evaluation
2
e
See section 23. Checking the barometer value on page A-47
oAvoid direct sunlight, vibration and strong electromagnetic fields (electric
oA stable and level work surface (max. 1° incline with bottles installed)
oRelative humidity: 20 to 85%
oAt least 10 cm free space around the instrument for air circulation and electrical
connections
oCorrect voltage: 100 to 240 VAC (±10%)
After setting up the cobas b 221 system at a location that meets the necessary
conditions, the following steps must be performed to ensure the instrument is ready
for operation:
oFirst check the instrument and the accessories for completeness and damage. The
completeness of the delivery can be checked through comparison with the delivery
packing slip.
If anything is missing, inform the Roche representative immediately.
If the delivery has suffered damage despite careful packing, inform the transportation
company immediately. Retain the packing material and products as evidence for the
damage claim.
Handle the instrument only at the specified holding points — risk of injury!
Take care when lifting - weight of the instrument without wash/calibrating solutions and AutoQC
is approx. 45 kg!
e
See illustration on the outer packaging and in Chapter 4 Specifications, section Holding
points on page A-92!
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-27
3 Installation and shutdowncobas b 221 system
Installation
Accessories
The following parts are delivered as standard equipment with the cobas b 221 system:
o1 barcode scanner
o2 Power cords (US and European version)
o1 roll printer paper
o2 pcs fill port
o1 sample inlet path (glass tube)
o5 system disks
o1 RCon (reference contact)
o1 shutdown kit
o1 dummy electrode
o1 dummy MSS cassette
o2 SCon (sensor contact)
o1 13 mm wrench (for screen/PC unit)
1 Phillips screwdriver
o3 pump tubes
Not shown in Figure A-12 on page A-29:
o
1 screen/PC unit
o
1 power supply
o
1 fill port
o
2 system disks
Roche DiagnosticsApril 2009
A-28Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
A
M
B
C
D
L
E
F
G
K
J
I
H
A 1 Roll printer paper
B 1 Dummy electrode
C 1 Dummy MSS cassette
D RCon (reference contact)
E SCon (sensor contact)
F 1 Sample inlet path (glass tube)
Figure A-12Accessories
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-29
G 2 Power cords (US and European version)
H System disks (total of 5 pcs.)
I1 13 mm wrench (for screen/PC unit);
1 Phillips screwdriver
J1 Barcode scanner
K 1 Fill port
L 3 Pump tubes
M 1 Shutdown kit
3 Installation and shutdowncobas b 221 system
Installation
Installation
1. Screen/PC unit
Ensure that the printed serial number on the rear of the screen/PC unit is the same as the unit serial
number on the nameplate!
1
Unscrew the fixing nut from the screen.
2
Place the screen/PC unit on the swivel arm.
3
At the base of the swivel arm, place the brake packet and lock nut on the shaft and
tighten using the 13 mm wrench provided in the accessories.
A
B
C
D
A Screen/PC unit
B Swivel arm
Figure A-13Swivel arm of the Screen/PC unit
4
Connect the cable to the screen and push it into the cable routing bar.
C Fixing nut
D Brake packet
Roche DiagnosticsApril 2009
A-30Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
2. Power supply
1
Place the power supply, including the two adapter connectors, on the holder and
position them.
A
A ScrewB Holder
Figure A-14Power supply
2
Tighten the screw.
3. Attach power cord and barcode scanner
1
Connect the power cord.
2
Connect the barcode scanner, and, if necessary, the network connection to the
appropriate port on the rear side of the cobas b 221 system.
4. Switch on
oSwitch the instrument on and wait until the program has completely loaded and
started. Before starting the installation, you must set the language, in which the
unit is to be operated, the date and the time.
B
B
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-31
3 Installation and shutdowncobas b 221 system
Installation
5. Installation
When carrying out the installation, follow the on-screen instructions.
Installation must be carried out completely and may not be interrupted.
Observe the listed sequence while performing the actions.
If the automatic first installation is unsuccessful, you must carry out the installation process
manually. To do this, press the following buttons:
[System] > [Utilities] > [Installation]
Processing the actions
ManualThe corresponding line of the list box contains an instruction which must be
performed manually. Then press [Confirm action].
AutomaticIf there is an automatic sequence for any action, you can start this by clicking
[Start process].
If an action has been completed successfully (manually or automatically),
this symbol is displayed.
6. Select language
7. Set the date and time
1
Press the following buttons:
h Setup > Instrument > Language
If the current language is "English": [Instrument] > [Language]
2
Select the language.
oPress the following buttons:
h Setup > Times & Intervals > Act. time / date
Figure A-15Act. time / date
Roche DiagnosticsApril 2009
A-32Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
8. Cal. intervals & timing
oPress the following button:
h Setup > Times & intervals > Cal. intervals & timing
System calibrationEvery 8, 12 or 24 hours.
2P calibrationEvery 4, 8 or 12 hours.
1P calibrationAll 30 or 60 minutes (USA: only every 30 minutes).
9. Set valves for FMS
Figure A-16Cal. intervals
Use this function to enter the automatic calibration times and intervals for system,
1 point and 2 point calibrations.
The time scale uses markers to show the selected interval for the 2P calibration
and the start time for the system calibration.
Intervals:
Enter the [Start time] of a system calibration to which all calibrations are oriented.
(a)
tubing exchange
oPress [Start process]. This action is performed automatically.
Valve V19 is pushed in to prevent the tube from being pinched while the aluminum part is
tightened! Valve VM is pushed out.
(a) "Fluid Mixing System" - Mixing of calibration solution A and B in a certain ratio
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-33
3 Installation and shutdowncobas b 221 system
Installation
10. Fix screws at V19 (bottle compartment)
1
Open the bottle compartment cover and the docking mechanism "S3".
2
Tighten the screws on valve V19 (approx. 2-3 rotations).
e
see Figure A-17!
Use the delivered screwdriver!
AA
A Screws on valve V19
Figure A-17Valve V19 and VM
3
To return to the installation window, close the docking mechanism and the bottle
compartment cover.
11. Insert right FMS tube at VM (bottle compartment)
1
Open the bottle compartment cover and the docking mechanism "S3".
2
Slide the tube under the tube clip of valve VM.
AB
A VMB V19
Figure A-18Valve VM
3
Close docking mechanism and bottle compartment cover.
Roche DiagnosticsApril 2009
A-34Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
12. Insert fill port and sample inlet path (glass tube)
1
Pull out the sample drip tray.
2
Remove the T&D cover and the unit cover.
3
Insert the fill port started from the 6 o’clock position as shown below.
4
Push the fill port straight onto the insert needle.
Do not bend the insert needle during this process!
A
A Needle
Figure A-19Insert needle
5
Rotate the fill port 90° clockwise and upwards until it snaps into place.
Figure A-20
6
Open the T&D lock.
e
see Figure A-21 on page A-36, A
7
Insert the glass tube into the guides, fasten it and check it for a correct position.
e
see Figure A-21 on page A-36, C
e
see Figure A-21 on page A-36, D
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-35
3 Installation and shutdowncobas b 221 system
A
Installation
A
A T&D lock
B Glass tube
Figure A-21Glass tube
B
C
C Insert the glass tube into the guides
D Fasten and check for correct position
C
D
8
Close the T&D lock again. Check the correct positioning of the sample inlet path
to the bypass nipple (see below)!
A Bypass nipple
Figure A-22T&D lock
9
Close the T&D cover.
10
Insert the sample drip tray.
Roche DiagnosticsApril 2009
A-36Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
13. Insert printer paper
The printer paper is heat sensitive on one side only. Observe the correct insertion of the thermal
paper roll.
A
B
A Printer coverB Paper lid
Figure A-23Printer
1
Open the printer cover and the paper lid.
2
Cut the start of the paper so that it is straight.
3
Place the paper roll into the holder.
4
Make sure that the printer lever is in the "down" position (see below).
A
A Printer lever "down" position
Figure A-24Printer lever
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-37
3 Installation and shutdowncobas b 221 system
Installation
5
Insert the beginning of the paper according to the instructions on the inside of the
paper lid (see below).
A
B
A Paper lidB Printer lever
Figure A-25Insert printer paper - without take-up unit
Figure A-26Insert printer paper - with take-up unit (optional)
6
The paper is automatically pulled into the printer.
7
Close paper lid.
With take-up unit (optional)
1
Press the paper feed button until the paper is long enough.
2
Insert the beginning of the paper in the take-up unit according to the instructions
on the inside of the paper lid.
e
see Figure A-26 on page A-38
Press the take-up unit (rods) fully onto the holder and rotate until the paper is taut on the rods and
paper lid, so that the entire roll of paper can be taken up. During operation, the paper should be
tautened now and then by turning the take-up roller.
3
Close printer cover.
With an installed take-up unit, the "Automatic Cut" function is deactivated.
Roche DiagnosticsApril 2009
A-38Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
14. Insert peristaltic pump tubes
1
Open the peristaltic pump's clear plastic cover (tension lever).
2
Push the linear bracket (white plastic part) upwards (see below).
A
B
C
A Te ns i on l e ver
B Pump head
C Linear bracket
Figure A-27Peristaltic pump
3
Place the tubing set around the corresponding rolling wheel (see below/A). Check
that the tubing set is correctly orientated (the grip end must be pointing upwards,
see below/B).
4
Close the clear plastic cover (tension lever). The tubing holder is then pressed into
the sealer (see below/B).
A Place the tubing setB Close the tension lever
Figure A-28Peristaltic pump
AutoQC module (option)
The installation with an AutoQC module (optional) must be performed by a Roche Diagnostics
Service Representative!
15. Go to AutoQC service position
oPress [Start process]. This action is performed automatically.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-39
3 Installation and shutdowncobas b 221 system
Installation
16. Open the AutoQC drawer and remove the AutoQC valve clamp
1
Pull out the AutoQC drawer.
2
Pull the key of the AutoQC valve up and out (see below).
A
A AutoQC valve clamp
Figure A-29AutoQC valve clamp
3
Close the AutoQC drawer.
17. Go to AutoQC home position
oPress [Start process]. This action is performed automatically.
18. Open AutoQC drawer and insert ampoule holder
1
Pull the AutoQC drawer out again.
A without ampoule holderB with ampoule holder
Figure A-30AutoQC drawer
2
Insert the AutoQC ampoule holder.
3
Close the AutoQC drawer.
19. Open the measuring chamber cover and insert the sensors
a BG / ISE measuring chamber
1
Open the measuring chamber cover (push the right edge of the MC cover to the
left with a finger and open up the MC cover).
In each case, open only the relevant measuring chamber.
Keep the bottle compartment cover closed.
Roche DiagnosticsApril 2009
A-40Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
The following screen appears:
Figure A-31Changing of electrodes
2
Open the locking lever.
e
see Figure A-34 on page A-42
3
Follow the instructions on the screen.
Check the internal electrolyte of the electrodes for possible air bubbles (see below).
If there are air bubbles between the contact pin and the membrane, there will not be effective
electrical conduction. Result: calibration and measurement errors!
4
Remove any air bubbles.
Remove air bubbles by holding the electrode vertically and by tapping lightly with
a fingernail against the electrode body (see below).
A
A Free of air bubbles!
Figure A-32Electrode
5
Insert the electrodes, beginning at the right and proceeding left according to the
color code.
6
Push all electrodes slightly to the right so that they are lined up together without
gaps.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-41
3 Installation and shutdowncobas b 221 system
Installation
a Insertion of the reference electrode
1
Insert the reference electrode.
Figure A-33Reference electrode
2
Insert the reference tube into the upper tube guide channel of the left locking lever
and into the tube holder of the cover hinge. Close the locking lever (see below).
A
A Locking lever
Figure A-34Insertion of the reference electrode
3
Connect the white connector on the end of the tube to the measuring chamber
cassette (see below).
A
B
A ConnectorB Measuring chamber cassette
Figure A-35Insertion of the reference electrode 2
4
Scan the barcodes located on the inner packaging of each electrode or enter the
barcodes manually with the help of the keyboard.
5
Close the measuring chamber cover.
Roche DiagnosticsApril 2009
A-42Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
a MSS measuring chamber (for instrument versions with MSS module only)
Hold the MSS cassette only at the designated handle and avoid touching the contacts.
1
Open the cover of the MSS measuring chamber (apply force to the right edge of
the MC cover with a finger to push it to the left and open up the MC cover).
Keep the bottle compartment cover closed!
2
Open the contact clip and the locking lever.
3
Depending on the MSS parameter configuration, insert the MSS reference
electrode (Ref + dummy) (see Figure A-36/A) or the reference contact (RCon)
(see Figure A-36/B) and the MSS cassette, close the contact clip and the locking
lever.
A
A Ref + dummy (for Glu/Lac/Urea)B RCon (Glu or Glu/Lac)
D
C
C Locking lever
D Contact clip
Figure A-36MSS measuring chamber
4
Read in the barcode of the packaging.
5
Close the measuring chamber cover.
6
Close the top cover.
7
Prepare a syringe or capillary with whole blood for polarization. Having
B
completed the installation process, the unit requests a blood sample.
The blood should have a volume of at least 150 μL, contain heparin as an anticoagulant, and be
stored for less than 24 hours.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-43
3 Installation and shutdowncobas b 221 system
Installation
20. Open bottle compartment cover and insert Waste container & packs
A
A
B
A Ruber sealingsBcobasb221<5> system and
cobas b 221<6> system only
Figure A-37Waste container & packs
1
Open the bottle compartment cover.
2
Open the corresponding docking mechanism.
3
Insert an empty waste water bottle and a S1 Rinse Solution bottle.
Remove packs’ rubber sealings.
4
Push the two packs into the appropriate location in accordance with the labeling
on the docking mechanisms until the packs lock.
Using the transponder attached to the bottle/packs, the instrument automatically
recognizes the corresponding bottle or packs.
A
Acobasb221<5> system and cobas b 221<6> system only
Figure A-38Changing of bottles and packs
Roche DiagnosticsApril 2009
A-44Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
A
Acobasb221<5> system and cobas b 221<6> system only
Figure A-39Bottle compartment
5
Close the docking mechanism and the bottle compartment cover.
To avoid splashing the S1 Rinse Solution, deaerate the bottle at about 3000 m above sea level or
higher before inserting it.
6
Place the bottle tool on the screw cap of the S1 Rinse Solution (see below).
A Bottle toolB Screw cap with placed bottle tool
Figure A-40Screw cap
7
Press the grips together and press the transparent disk downward (see below/A).
8
Rotate the transparent disk clockwise and stop when you notice a resistance after a
short distance (see below/B).
AB
Figure A-41Open bottle
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-45
3 Installation and shutdowncobas b 221 system
Installation
21. Complete installation
1
Press the [Complete installation] button.
Automatic sequences take place and the unit warms up.
2
Installation is complete.
If a power failure occurs during installation, the installation starts anew with the next restart.
Actions which were performed successfully are discarded.
22. Perform MSS polarization (cobas b 221<5> system and cobas b 221<6> system only)
1
Prepare a syringe or capillary with whole blood for polarization.
The blood should have a volume of at least 150 μL, contain heparin as an anticoagulant, and be
stored for less than 24 hours.
Figure A-42MSS polarization
2
The blood sample is inserted via fill port similar to a measurement.
e
see Chapter 6 Measurement
3
The MSS cassette is subsequently exposed to liquid, polarized and heated.
4
A system calibration is carried out.
5
If, after inserting the cassette, the automatic polarization was not successful and
the MSS sensors are not calibrated, you must manually polarize the MSS cassette.
To do this, press the following buttons:
h System > Utilities > MSS polarization
6
Follow the instructions on the screen!
Roche DiagnosticsApril 2009
A-46Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Installation
23. Checking the barometer value
h System > Component test > Control sensors > Baro sensor
1
If the barometer value deviates by more than ± 4 mbar from the value indicated by
a precision barometer, it will be necessary for Technical support to calibrate the
barometer!
A wrong barometer value leads to wrong PO2 measurement results.
Important:
From approx. 3000 m above sea level or air pressure < 526 mmHg (70.13 kPa), the specifications
24. Quality control
for parameter PO
of the clinical decisions. The parameter PO2 must be permanentely deactivated.
2
To deactivate the parameter PO2 press the following buttons:
h Setup > Parameter > Miscellaneous settings > Activated / deactivated for calibrations
1
Define the material and if an AutoQC drawer (option) is available insert the mats
before performing a quality control measurement.
are no longer fulfilled and the parameter must no longer be used for evaluation
2
e
For details, see Chapter 7 Quality control
2
Perform quality control tests for all 3 levels (low, normal, high). Make sure that
the results agree with the target values.
e
See Chapter 7 Quality control
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-47
3 Installation and shutdowncobas b 221 system
Shutdown
Shutdown
Less than 24 hours
If the cobas b 221 system is not used for a short period of time only (< 24 hours), then
activate the following function, starting with the top level of the analyzer mode:
h System > Utilities > Shutdown PC
This function allows for switching off the touch screen/PC unit and is completed with
manually switching off the instrument.
Follow the instructions on the screen!
MSS sensors (Glu / Lac / Urea/BUN) are destroyed during this operation.
If the instrument is turned on again, a new MSS cassette must be inserted.
e
See section 19. Open the measuring chamber cover and insert the sensors on page A-40.
Longer than 24 hours
If the cobas b 221 system will be shut down for longer than 24 hours, perform the
following procedure.
Before performing a shutdown, Roche Diagnostics recommends decontaminating all surfaces and
tube paths.
e
see Chapter 10 Maintenance, Abschnitt Decontamination on page C-5!
Activate the following function, starting with the top level of the analyzer mode:
h System > Utilities > Put out of operation
All solutions and electrodes have to be removed during the shutdown procedure.
The procedure ends in switching off the instrument.
Follow the instructions on the screen.
Observe the listed sequence while performing the actions.
Processing the actions:
Manual:The corresponding line of the list box contains an instruction which must be
performed manually. Then press [Confirm action].
Automatic:If there is an automatic sequence for any action, you can start this by clicking [Start
process].
Roche DiagnosticsApril 2009
A-48Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Shutdown
Upon successful completion, this symbol is displayed.
1. Open bottle compartment cover and only remove bottle S1 and packs
(depending on the configuration S2 and S3).
1
Open bottle compartment cover and docking mechanism and remove bottle S1
and the packs (S2 and S3).
Do not remove the waste container!
2
Close docking mechanism and bottle compartment cover.
2. Fill the shutdown kit with distilled water
oFill the shutdown kit about halfway with distilled water.
Figure A-43Shutdown kit
3. Insert shutdown kit into space S2
1
Open bottle compartment cover and docking mechanism S2 and insert the
shutdown kit into space S2.
2
Close docking mechanism and bottle compartment cover.
3
Perform "Washing of the tubes".
4. Remove shutdown kit from space S2
1
Open bottle compartment cover and docking mechanism S2 and remove the
shutdown kit.
2
Close docking mechanism and bottle compartment cover.
3
Perform "Emptying of the tubes".
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-49
3 Installation and shutdowncobas b 221 system
Shutdown
5. Insert shutdown kit into space S3 (cobas b 221<5> system and cobas b 221<6> system only)
1
Open bottle compartment cover and docking mechanism S3 and insert the
shutdown kit into space S3.
2
Close docking mechanism and bottle compartment cover.
3
Perform "Washing of the tubes".
6. Remove shutdown kit from space S3 (cobas b 221<5> system and cobas b 221<6> system only)
1
Open bottle compartment cover and docking mechanism S3 and remove the
shutdown kit.
2
Close docking mechanism and bottle compartment cover.
3
Perform "Emptying of the tubes".
7. Remove Waste container
1
Open bottle compartment cover and docking mechanism W.
2
Remove the waste water container (W Waste Container).
3
Close docking mechanism and bottle compartment cover.
8. Open the measuring chamber cover and remove the sensors
1
Remove the top cover and open all measuring chamber covers.
2
Open the measuring chamber cover (push the right edge of the MC cover to the
left with a finger and open up the MC cover).
3
Open the locking levers and the contact clip (MSS measuring chamber).
4
Sequentially remove the electrodes and the MSS cassette from the measuring
chambers.
5
Close the locking lever, the contact clip and all the measuring chamber covers.
9. Remove the peristaltic pump tubes
1
Open the peristaltic pump's clear plastic cover (tension lever) (see below).
A
B
C
A Te ns i on l e ver
B Pump head
C Linear bracket
Figure A-44Peristaltic pump
2
Push the linear bracket (white plastic part) upwards (see below/A).
Roche DiagnosticsApril 2009
A-50Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Shutdown
3
Remove the complete tubing set (tubing holder and tubing) of the corresponding
pump (see below/B)
A Push the linear bracket upwardsB Remove the tubing set
Figure A-45Peristaltic pump
4
Close the tension lever.
10. Remove the printer paper
1
Open the printer cover and the paper lid.
A
B
A Printer coverB Paper lid
Figure A-46Printer cover / paper lid
2
Move the printer lever upwards (see below/A).
A Printer lever "upwards"B Printer lever "down"
Figure A-47Printer lever
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-51
3 Installation and shutdowncobas b 221 system
Shutdown
3
Remove the printer paper.
4
Move the printer lever down again (see above/B).
5
Close the paper lid and the printer cover.
11. Open T&D
oPress [Start process]. This action is performed automatically.
The T&D disk turns to position 1.
12. Remove fill port and sample inlet path (glass tube)
1
Remove the sample drip tray.
2
Remove the T&D cover.
3
Open the T&D lock and remove the sample inlet path (glass tube).
A
B
A T&D lockB Sample inlet path (glass tube)
Figure A-48T&D lock & sample inlet path
4
Turn the fill port downward by 90° and pull it straight off of the needle.
Do not bend the needle!
Roche DiagnosticsApril 2009
A-52Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Shutdown
A Fill portB Needle
Figure A-49Fill port
5
Close the T&D lock again.
6
Close the T&D cover.
13. Set valves for FMS tubing exchange
oPress [Start process]. This action is performed automatically.
Both valves are pushed out.
14. Release screws at V19 (bottle compartment)
1
Open the bottle compartment cover and the docking mechanism S3.
2
Loosen the screws (A) of the aluminum part of valve V19 (approx. 2-3 turns).
A
B
A
A
V 19
A Screws
Figure A-50Valve V19
3
Close the docking mechanism S3 and the bottle compartment cover.
15. Remove right FMS tube at VM (bottle compartment)
1
Open the bottle compartment cover and the docking mechanism S3.
2
Slide the tube out under the tube clip of valve VM.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-53
3 Installation and shutdowncobas b 221 system
Shutdown
B
A
A VMB Tube clip
Figure A-51Valve VM
3
Pressure is removed from the tubes.
4
Close the docking mechanism and the bottle compartment cover.
If available (option):
16. Go to AutoQC home position
oPress [Start process]. This action is performed automatically.
17. Open the AutoQC drawer and remove the ampoule holder
1
Pull out the AutoQC drawer.
2
Remove the AutoQC ampoule holder.
3
Remove the already opened ampoules from the mats and dispose of them
according to the local guidelines.
If individual ampoules remain in the white ampoule holder after removing the mats, note that
these open ampoules may break on removal with the attendant risk of injury.
Before inserting a new mat remove them all carefully!
Always wear gloves! CAUTION: Danger of spilling!
4
Leave the full ampoules in the mats and store them in a refrigerator in accordance
with their storage temperature (see packaging insert).
5
Close the AutoQC drawer.
18. Go to AutoQC service position
oPress [Start process]. This action is performed automatically.
19. Open AutoQC drawer and insert the AutoQC valve clamp
1
Pull out the AutoQC drawer.
2
Insert the clamp of the AutoQC valve (see below).
Roche DiagnosticsApril 2009
A-54Instructions for Use · Revision 10.0
cobas b 221 system3 Installation and shutdown
Shutdown
A
A AutoQC valve clamp
Figure A-52AutoQC valve clamp
3
Close the AutoQC drawer.
20. Go to AutoQC home position
oPress [Start process]. This action is performed automatically.
21. Complete shutdown
1
Press the [Complete shutdown] button.
Shut down is complete. The following screen appears:
Figure A-53Shutdown
2
Press the [Shutdown PC] button. Follow the instructions on the screen.
The PC is booted down.
3
Turn off the device.
4
Close top cover.
Remove the transport, power cable, scanner and, if available, network connectors.
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-55
3 Installation and shutdowncobas b 221 system
Shutdown
Roche DiagnosticsApril 2009
A-56Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Contents
Specifications
In this chapter, the performance data, as well as product and environmental data are
described.
In this chapter
Performance data ..........................................................................................................59
(a) Due to the current specifications, clinically significant deviations in the range < 3mmol/L can occur compared to other glucose measuring systems.
Especially in the neonatal field, we therefore recommend carrying out a comparative blood measurement relative to a known reference system or to
adapt the correlation table (refer to the Reference Manual, chapter "Setup" section "Correlation"). For any questions concerning this matter, contact
the local Roche organization.
B/Q0 - 800 mmHg
B/Q4 – 200 mmHg
B/Q/S0.5 – 40 mmol/L9.01 - 720.8 mg/dL
B/Q/S0.2 – 20 mmol/L1.8016 - 180.16 mg/dL
B
Q
A
D
S
PF
(a) with approximate physiological ion matrix and buffer capacity
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-59
Whole blood
Aqueous QC material
(a)
Dialysis solutions containing acetate
Dialysis solutions containing bicarbonate
Serum or plasma
Pleural fluid (can be measured in serum/plasma mode)
4 Specificationscobas b 221 system
Performance data
Reproducibility
"Within-Run (Swr)" and "Total Precision (ST)" was determined from 2 runs per day
with 2 replicates per run for 20 days on four cobas b 221 systems.
The mean value is the measured value of the corresponding parameter for which S
and S
are representative resp. have been determined.
T
ParameterUnit
pHpH units
PCO
2
PO
2
Sodiummmol/L
Potassiummmol/L
Chloridemmol/L
ionized Calciummmol/L
Hct%
Lactate (cobas b 221<5> system, cobas b 221<6> system only)
Glucose (cobas b 221<5> system, cobas b 221<6> system only)
Urea (cobas b 221<6> system only)mmol/L
tHb (tHb module)g/dL
(tHb module)%
SO
2
tHb (COOX)g/dL
(COOX)%
SO
2
Hb%
O
2
COHb%
MetHb%
HHb%
Bilirubin (neonatal)mg/dL
Ta b le A - 2Units of the parameters
mmHg
mmHg
mmol/L
mmol/L
wr
Material: acetat - standard solution (Level 1), NIST Traceable, n=80
ParameterMeanS
wr
Sodium140.0 0.5600 0.400.7405 0.53
Potassium2.02 0.0165 0.820.0290 1.44
Chloride-----
ionized Calcium1.622 0.0155 0.960.0205 1.26
Ta b l e A - 3Acetat - standard solution (Level 1), NIST Traceable, n=80
Roche DiagnosticsApril 2009
A-60Instructions for Use · Revision 10.0
(CV%)S
T
(CV %)
cobas b 221 system4 Specifications
Performance data
Material: acetat - standard solution (Level 2), NIST Traceable, n=80
ParameterMeanS
wr
(CV%)S
T
(CV %)
Sodium140.1 0.5107 0.360.7747 0.55
Potassium4.00 0.0171 0.430.0273 0.68
Chloride-----
ionized Calcium1.166 0.0077 0.660.0141 1.21
Ta b l e A - 4Acetat - standard solution (Level 2), NIST Traceable, n=80
Material: tonometered human whole blood, 20 different probands, n=80
ParameterMeanS
wr
pH7.441 0.0042 0.06--
PCO
PO
2
2
18.3 0.3331 1.820.6262 3.42
137.9 0.9371 0.682.3258 1.69
Sodium139.5 0.4878 0.35--
Potassium4.58 0.0260 0.57 --
Chloride108.4 0.4310 0.40--
ionized Calcium1.181 0.0079 0.67--
Hct43.3 0.3203 0.74 --
Lactate11.5 0.1769 1.54 --
Glucose1.8 0.0648 3.51--
Urea4.8 0.0529 1.11--
tHb (tHb module)15.4 0.1461 0.95--
(tHb module)96.6 0.3744 0.39--
SO
2
tHb (COOX)14.1 0.0773 0.55--
(COOX)99.9 0.0613 0.06--
SO
2
Hb97.9 0.0684 0.07 --
O
2
COHb1.4 0.0377 2.79--
MetHb0.7 0.0287 4.10--
HHb0.1 0.0601 ---
Ta b l e A - 5Tonometered human whole blood, 20 different probands, n=80
(CV%)S
T
(CV %)
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-61
4 Specificationscobas b 221 system
Performance data
Material: tonometered human whole blood, 20 different probands, n=80
ParameterMeanSwr (CV%)ST (CV %)
pH7.129 0.0049 0.07--
PCO
2
PO
2
Sodium142.3 0.7126 0.50--
Potassium4.32 0.0392 0.91 --
Chloride105.2 0.5184 0.49--
ionized Calcium1.301 0.0136 1.05--
Hct40.4 0.2795 0.69 --
Lactate8.7 0.2021 2.33--
Glucose2.3 0.0977 4.31--
Urea4.9 0.0583 1.18--
tHb (tHb module)15.9 0.1315 0.83--
(tHb module)55.0 0.8839 1.61--
SO
2
tHb (COOX)14.1 0.1691 1.20--
(COOX)67.8 0.2479 0.37--
SO
2
Hb66.9 0.3437 0.51 --
O
2
COHb1.6 0.0549 3.53--
MetHb0.4 0.0504 12.14--
HHb31.5 0.3121 0.99 --
Ta b l e A - 6Tonometered human whole blood, 20 different probands, n=80
79.5 1.2629 1.591.9644 2.47
40.1 0.3297 0.820.5976 1.49
Material: human plasma, n=80
ParameterMeanSwr (CV%)ST (CV %)
pH7.670 0.0097 0.130.0549 0.72
PCO
2
PO
2
Sodium140.9 0.7783 0.550.9920 0.70
Potassium3.99 0.0514 1.290.0603 1.51
Chloride106.0 0.4967 0.470.7877 0.74
ionized Calcium1.155 0.0174 1.510.0339 2.94
Hct-----
Lactate2.3 0.0349 1.520.1150 5.00
Glucose5.7 0.0818 1.440.1695 2.97
Urea4.8 0.0873 1.810.1005 2.08
Ta b l e A - 7Human plasma, n=80
-----
-----
Roche DiagnosticsApril 2009
A-62Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Performance data
Material: serum, n=80
ParameterMeanSwr (CV%)ST (CV %)
pH7.731 0.0120 0.150.0334 0.43
PCO
2
PO
2
Sodium140.2 0.3226 0.230.6567 0.47
Potassium4.18 0.0149 0.360.0330 0.79
Chloride105.2 0.4310 0.410.6871 0.65
ionized Calcium1.098 0.0092 0.840.0323 2.94
Hct-----
Lactate2.3 0.0353 1.530.0989 4.30
Glucose5.1 0.0737 1.450.1834 3.62
Urea5.2 0.0451 0.860.1197 2.29
Ta b l e A - 8Serum, n=80
Material: bicarbonate, n=80
-----
-----
ParameterMeanS
wr
(CV%)S
T
(CV %)
Sodium137.9 0.7201 0.521.0185 0.74
Potassium2.00 0.0224 1.120.0301 1.51
Chloride-----
ionized Calcium1.605 0.0091 0.570.0167 1.04
Ta b l e A - 9Bicarbonate, n=80
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-63
4 Specificationscobas b 221 system
Performance data
Material: AUTOTROL PLUS B Level 1, n=40
ParameterMeanS
wr
(CV%)S
T
(CV %)
pH7.182 0.0039 0.050.0060 0.08
PCO
PO
2
2
65.8 0.8109 1.231.7861 2.72
55.4 3.6232 6.534.5447 8.20
Sodium121.2 0.6188 0.511.1226 0.93
Potassium2.97 0.0161 0.540.0283 0.95
Chloride84.2 0.4971 0.591.6465 1.96
ionized Calcium1.557 0.0089 0.570.0153 0.98
Hct51.8 0.9534 1.841.1250 2.17
Lactate9.2 0.0821 0.890.4539 4.92
Glucose5.4 0.0612 1.120.1299 2.38
Urea23.5 0.3307 1.410.6664 2.84
tHb (tHb module)-----
(tHb module) -----
SO
2
tHb (COOX)7.8 0.0317 0.410.0599 0.77
(COOX)72.1 0.0690 0.100.1941 0.27
SO
2
Hb46.8 0.0844 0.180.2383 0.51
O
2
COHb23.0 0.0371 0.160.1043 0.45
MetHb12.0 0.0180 0.150.0513 0.43
HHb18.1 0.0294 0.160.0830 0.46
Bili6.10.02870.470.04770.78
Ta b l e A - 1 0AUTOTROL PLUS B Level 1, n=40
Material: AUTOTROL PLUS B Level 2, n=40
ParameterMeanS
wr
pH7.411 0.0031 0.040.0047 0.06
PCO
PO
2
2
41.0 0.4626 1.130.7116 1.74
93.2 2.9752 3.195.0160 5.38
Sodium139.6 0.3827 0.270.7718 0.55
Potassium4.76 0.0131 0.270.0250 0.53
Chloride101.0 0.3290 0.330.9795 0.97
ionized Calcium1.154 0.0064 0.550.0138 1.20
Hct38.6 0.2840 0.740.6195 1.60
Lactate1.9 0.0135 0.700.0798 4.12
Glucose2.4 0.0197 0.810.1172 4.83
Urea7.3 0.0538 0.740.1939 2.67
tHb (tHb module)-----
(tHb module) -----
SO
2
tHb (COOX)12.1 0.0715 0.590.1182 0.98
(COOX)89.6 0.1442 0.160.1507 0.17
SO
2
Hb74.3 0.2843 0.380.3011 0.41
O
2
Ta b l e A - 1 1AUTOTROL PLUS B Level 2, n=40
(CV%)S
T
(CV %)
Roche DiagnosticsApril 2009
A-64Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Performance data
ParameterMeanS
wr
(CV%)S
T
(CV %)
COHb11.1 0.1265 1.140.1306 1.18
MetHb6.0 0.0577 0.960.0671 1.12
HHb8.6 0.1001 1.170.1041 1.21
Bili12.40.08570.690.11880.96
Ta b l e A - 1 1AUTOTROL PLUS B Level 2, n=40
Material: AUTOTROL PLUS B Level 3, n=40
ParameterMeanS
wr
pH7.571 0.0027 0.040.0050 0.07
PCO
PO
2
2
20.3 0.3114 1.530.5568 2.74
144.2 5.3745 3.736.5040 4.51
Sodium158.9 0.5680 0.360.8495 0.53
Potassium6.97 0.0343 0.490.0514 0.74
Chloride119.0 0.4810 0.401.0305 0.87
ionized Calcium0.546 0.0041 0.760.0078 1.43
Hct26.9 0.4193 1.560.4298 1.60
Lactate0.8 0.0103 1.290.0562 7.02
Glucose21.0 0.1298 0.620.4006 1.91
Urea2.1 0.0202 0.940.0757 3.53
tHb (tHb module)-----
(tHb module) -----
SO
2
tHb (COOX)20.4 0.1940 0.950.2357 1.15
(COOX)97.5 0.1396 0.140.1400 0.14
SO
2
Hb92.5 0.3581 0.390.3617 0.39
O
2
COHb3.3 0.1564 4.750.1565 4.75
MetHb1.9 0.0773 4.130.0809 4.32
HHb2.4 0.1244 5.230.1249 5.25
Bili21.60.16210.750.16900.78
Ta b l e A - 1 2AUTOTROL PLUS B Level 3, n=40
(CV%)S
T
(CV %)
Material: AUTOTROL PLUS B Level 4B, n=40
ParameterMeanS
wr
pH7.418 0.0014 0.020.0050 0.07
PCO
PO
2
2
41.3 0.2720 0.660.6088 1.48
96.4 5.0118 5.208.9120 9.24
Sodium140.6 0.3242 0.230.5710 0.41
Potassium4.77 0.0135 0.280.0220 0.46
Chloride101.6 0.3679 0.360.9279 0.91
ionized Calcium1.104 0.0048 0.430.0092 0.83
Hct36.7 0.3883 1.060.5049 1.38
Lactate5.6 0.0304 0.540.1607 2.85
Ta b l e A - 1 3AUTOTROL PLUS B Level 4B, n=40
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-65
(CV%)S
T
(CV %)
4 Specificationscobas b 221 system
Performance data
ParameterMeanS
wr
(CV%)S
T
(CV %)
Glucose1.4 0.0204 1.450.1026 7.26
Urea13.1 0.2512 1.910.7169 5.46
tHb (tHb module)-----
(tHb module) -----
SO
2
tHb (COOX)6.4 0.0265 0.410.1241 1.93
(COOX)62.7 0.2002 0.320.2514 0.40
SO
2
Hb36.5 0.1973 0.540.2474 0.68
O
2
COHb27.6 0.0869 0.320.1091 0.40
MetHb14.2 0.0414 0.290.0518 0.36
HHb21.7 0.0690 0.320.0866 0.40
Bili4.20.01460.340.07071.67
Ta b l e A - 1 3AUTOTROL PLUS B Level 4B, n=40
Material: AUTOTROL PLUS B Level 5B, n=40
ParameterMeanS
wr
pH7.412 0.0033 0.040.0061 0.08
PCO
PO
2
2
41.4 0.3787 0.910.7924 1.91
94.7 3.1077 3.283.2578 3.44
Sodium139.4 0.7465 0.540.8404 0.60
Potassium4.76 0.0365 0.770.0339 0.71
Chloride102.1 0.7601 0.741.3617 1.33
ionized Calcium1.119 0.0062 0.560.0103 0.92
Hct38.0 1.4404 3.791.4027 3.69
Lactate12.9 0.1628 1.260.5348 4.14
Glucose25.4 0.1913 0.750.5098 2.00
Urea26.4 0.4122 1.562.5774 9.75
tHb (tHb module)-----
(tHb module) -----
SO
2
tHb (COOX)23.0 0.2175 0.940.3139 1.36
(COOX)98.1 0.1568 0.160.1744 0.18
SO
2
Hb94.2 0.4087 0.430.4554 0.48
O
2
COHb2.5 0.0852 3.370.1053 4.16
MetHb1.4 0.0397 2.770.0519 3.62
HHb1.8 0.0675 3.830.0837 4.75
Bili24.10.26291.090.27281.13
Ta b l e A - 1 4AUTOTROL PLUS B Level 5B, n=40
(CV%)S
T
(CV %)
Roche DiagnosticsApril 2009
A-66Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Performance data
Material: AUTOTROL TS+ Level 1, n=40
ParameterMeanS
wr
(CV%)S
T
(CV %)
pH7.172 0.0037 0.050.0054 0.08
62.3 0.8276 1.331.5073 2.42
PCO
2
PO
2
50.6 4.2084 8.325.4278 10.74
Sodium121.9 0.8432 0.691.0952 0.90
Potassium2.98 0.0333 1.120.0352 1.18
Chloride84.8 0.5243 0.620.9029 1.06
ionized Calcium1.591 0.0170 1.070.0217 1.36
Hct56.8 1.5547 2.741.5912 2.80
Lactate9.3 0.0710 0.760.4680 5.01
Glucose5.5 0.0564 1.020.1729 3.13
tHb (tHb module)18.7 0.0256 0.140.0440 0.24
(tHb module)100.0 0.0112 0.010.0112 0.01
SO
2
Ta b l e A - 1 5AUTOTROL TS+ Level 1, n=40
Material: AUTOTROL TS+ Level 2, n=40
ParameterMeanS
wr
pH7.406 0.0022 0.030.0043 0.06
43.1 0.3800 0.880.5716 1.33
PCO
2
PO
2
92.5 2.8967 3.133.5631 3.85
Sodium136.9 0.5024 0.370.7705 0.56
Potassium4.70 0.0303 0.640.0414 0.88
Chloride99.4 0.3950 0.400.5617 0.56
ionized Calcium1.146 0.0141 1.230.0178 1.56
Hct41.6 0.3827 0.920.7501 1.80
Lactate1.9 0.0163 0.860.0672 3.53
Glucose2.5 0.0204 0.830.1044 4.23
tHb (tHb module)14.3 0.0773 0.540.0794 0.56
(tHb module)93.3 0.1776 0.190.1855 0.20
SO
2
Ta b l e A - 1 6AUTOTROL TS+ Level 2, n=40
(CV%)S
T
(CV %)
Material: AUTOTROL TS+ Level 3, n=40
ParameterMeanS
wr
pH7.567 0.0023 0.030.0047 0.06
22.9 0.2652 1.160.4763 2.08
PCO
2
PO
2
141.8 3.4549 2.444.0597 2.86
Sodium156.3 0.8170 0.521.0626 0.68
Potassium7.03 0.0510 0.720.0628 0.89
Chloride120.3 0.4829 0.400.6012 0.50
ionized Calcium0.599 0.0100 1.670.0128 2.13
Ta b l e A - 17AUTOTROL TS+ Level 3, n=40
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-67
(CV%)S
T
(CV %)
4 Specificationscobas b 221 system
Performance data
ParameterMeanS
wr
(CV%)S
T
(CV %)
Hct22.9 0.5431 2.370.5799 2.53
Lactate0.8 0.0316 4.020.0446 5.68
Glucose21.3 0.6006 2.820.7883 3.70
tHb (tHb module)8.3 0.0122 0.150.0319 0.39
(tHb module)93.3 0.0355 0.040.0450 0.05
SO
2
Ta b l e A - 17AUTOTROL TS+ Level 3, n=40
Material: AUTOTROL TS+ Level 4A, n=40
ParameterMeanS
wr
pH6.880 0.0054 0.080.0076 0.11
87.5 1.3229 1.512.4825 2.84
PCO
2
PO
2
22.7 3.6828 16.235.1660 22.77
Sodium88.0 0.5162 0.590.8391 0.95
Potassium8.94 0.0584 0.650.1029 1.15
Chloride67.8 0.5941 0.881.3054 1.93
ionized Calcium2.543 0.0272 1.070.0452 1.78
Hct75.8 0.8202 1.081.0158 1.34
Lactate-----
Glucose-----
Urea-----
tHb (tHb module)11.0 0.0206 0.190.0281 0.26
(tHb module)88.2 0.0303 0.030.0450 0.05
SO
2
Ta b l e A - 1 8AUTOTROL TS+ Level 4A, n=40
(CV%)S
T
(CV %)
Material: AUTOTROL TS+ Level 5A, n=40
ParameterMeanS
wr
pH7.730 0.0042 0.050.0061 0.08
9.2 1.3769 15.031.4213 15.51
PCO
2
PO
2
253.6 6.0686 2.398.7795 3.46
Sodium174.5 0.8890 0.511.2891 0.74
Potassium2.00 0.0250 1.250.0346 1.73
Chloride130.3 0.7821 0.601.1922 0.91
ionized Calcium0.403 0.0065 1.610.0109 2.70
Hct22.0 0.6997 3.180.7713 3.50
Lactate-----
Glucose-----
Urea-----
tHb (tHb module)15.8 0.0196 0.120.0316 0.20
(tHb module)95.8 0.0469 0.050.0596 0.06
SO
2
Ta b l e A - 1 9AUTOTROL TS+ Level 5A, n=40
(CV%)S
T
(CV %)
Roche DiagnosticsApril 2009
A-68Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Performance data
Material: MSS Level 1, NIST Traceable, n=80
ParameterMeanS
wr
(CV%)S
T
(CV %)
Lactate9.4 0.0670 0.710.2626 2.78
Glucose5.7 0.0337 0.600.1231 2.18
Urea4.9 0.0391 0.800.1837 3.74
Ta b l e A - 2 0MSS Level 1, NIST Traceable, n=80
Material: MSS Level 2, NIST Traceable, n=80
ParameterMeanS
wr
Lactate1.9 0.0188 0.960.0497 2.55
Glucose2.6 0.0267 1.050.0972 3.81
Urea14.5 0.2263 1.560.4100 2.83
Ta b l e A - 2 1MSS Level 2, NIST Traceable, n=80
(CV%)S
T
(CV %)
Material: human whole blood incl. bilirubin Level 1, n=40
ParameterMeanS
wr
Bili8.20.12021.470.61987.56
Ta b l e A - 2 2Human whole blood incl. bilirubin Level 1, n=40
(CV%)S
T
(CV %)
Material: human whole blood incl. bilirubin Level 2, n=40
ParameterMeanS
wr
(CV%)S
T
(CV %)
Bili24.10.11710.490.96634.01
Ta b l e A - 2 3Human whole blood incl. bilirubin Level 2, n=40
Material: human whole blood incl. bilirubin Level 3, n=40
ParameterMeanS
wr
Bili44.00.16230.372.15094.89
Ta b l e A - 2 4Human whole blood incl. bilirubin Level 3, n=40
(CV%)S
T
(CV %)
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-69
4 Specificationscobas b 221 system
Performance data
Linearity
Tonometered whole bloodWhole blood was tonometered at 37 °C to various level of gravimetrically prepared
gases with CO
manufacturer. Expected and observed values for PCO
760 mmHg.
Aqueous SolutionsExpected values for the aqueous solutions are based on weighted samples.
NIST standardsNIST standards are precise serums with accredited target values.
HematocritMeasurement results of the hemofuge, which is representing the Golden Standard for
hematocrit measurements, are used as expected values for hematocrit results.
and O2 concentrations certified to ± 0.03% absolute by the
2
and PO2 were corrected to
2
Human whole blood incl.
Parameter: PO
bilirubin
(mmHg)
2
Expected bilirubin values for human whole blood incl. bilirubin are based on
weighted samples.
Material: tonometered whole blood
Number of instruments: 4 cobas b 221 systems
Measurements per measuring point and instrument: 5
Expected valueMeanSwr Recovery
55.3955.660.4860100.5
83.8383.450.498299.5
103.55103.160.903499.6
216.97218.541.9437100.7
Ta b le A - 25Parameter PO2 (mmHg)
Correlation
Slope0.9904 - 1.0097
Intercept± 0.857
Correlation coefficient0.9998
Roche DiagnosticsApril 2009
A-70Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Performance data
Parameter: PCO2 (mmHg)
Material: Tonometered whole blood
Number of instruments: 4 cobas b 221 systems
Measurements per measuring point and instrument: 5
Expected valueMeanSwr Recovery
14.9013.780.114192.5
39.7437.780.391195.1
119.43117.091.350598.0
Ta b le A - 26Parameter PCO2 (mmHg)
Correlation
Slope0.9898 - 1.0103
Intercept± 1.225
Correlation coefficient0.9999
Parameter: pH (pH units)
Material: Tonometered whole blood
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 5
Expected valueMeanSwr Recovery
7.527.520.0050100
7.327.320.0042100
6.986.990.0066100.1
Ta b le A - 27Parameter pH (pH units)
Correlation
Slope0.9825 - 1.0178
Intercept± 0.133
Correlation coefficient0.9998
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-71
4 Specificationscobas b 221 system
Performance data
Parameter: Hct (%)
Material: human whole blood, traceable to golden standard (micro centrifuge)
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 5
Expected valueMeanSwr Recovery
11.0011.720.4146106.5
24.0023.600.180498.3
36.0036.511.0171101.4
48.0049.731.0046103.6
68.0068.160.2210100.2
78.0077.800.392599.7
Ta b le A - 28Parameter Hct (%)
Correlation
Slope0.997 - 1.003
Intercept± 0.620
Correlation coefficient0.999
Parameter: sodium (mmol/L)
Material: aqueous solution, NIST traceable
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
19.8521.250.4979107.1
91.5292.440.3018101.0
153.49154.830.3808100.9
205.66208.300.5619101.3
258.42262.681.6465101.6
Ta b le A - 29Parameter Sodium (mmol/L)
Correlation
Slope0.988 - 1.012
Intercept± 0.365
Correlation coefficient0.9999
Roche DiagnosticsApril 2009
A-72Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Performance data
Parameter: potassium (mmol/L)
Material: aqueous solution, NIST traceable
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
0.230.260.0159115.0
3.123.120.0114100.0
5.115.130.0144100.4
9.9610.160.0378102.0
14.7115.190.0624103.3
19.3620.150.0757104.1
Ta b le A - 30Parameter Potassium (mmol/L)
Correlation
Slope0.960 - 1.042
Intercept± 0.109
Correlation coefficient0.9999
Parameter: ionized Calcium (mmol/L)
Material: aqueous solution, NIST traceable
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
0.100.090.004292.0
0.800.760.006794.7
1.251.190.005595.2
2.502.390.012295.7
4.003.860.022596.5
6.005.840.034797.4
Ta b le A - 31Parameter ionized Calcium (mmol/L)
Correlation
Slope0.975 - 1.026
Intercept± 0.024
Correlation coefficient0.9999
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-73
4 Specificationscobas b 221 system
Performance data
Parameter: chloride (mmol/L)
Material: aqueous solution, NIST traceable
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
24.8625.840.5081103.9
93.2093.080.224199.9
149.34146.850.354198.3
194.98190.160.611097.5
239.86232.041.072196.7
Ta b le A - 32Parameter Chloride (mmol/L)
Correlation
Slope0.959 - 1.043
Intercept± 2.908
Correlation coefficient0.9999
Parameter: pH (pH units)
Material: aqueous solution, NIST traceable
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
6.206.240.0022100.7
6.876.890.0024100.2
7.387.380.0023100.1
7.707.670.002399.7
8.007.970.003599.7
Ta b le A - 33Parameter pH (pH units)
Correlation
Slope0.960 - 1.042
Intercept± 0.293
Correlation coefficient1.0000
Roche DiagnosticsApril 2009
A-74Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Performance data
Parameter: CO2 (mmHg)
Material: tonometered aqueous solution
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
10.0011.360.1389113.6
20.0020.590.1962103.0
60.0057.570.655795.9
120.00114.241.552195.2
180.00175.372.435897.4
Ta b le A - 34Parameter: CO2 (mmHg)
Correlation
Slope0.961 - 1.041
Intercept± 0.865
Correlation coefficient0.9994
Parameter: O
(mmHg)
2
Material: tonometerd aqueous solution
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
600.00550.248.259491.7
300.00278.073.713192.7
140.00140.250.5353100.2
60.0060.290.2923100.5
10.0011.710.4329117.1
Ta b le A - 35Parameter O2 (mmHg)
Correlation
Slope0.908 - 1.101
Intercept± 6.609
Correlation coefficient0.9995
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-75
4 Specificationscobas b 221 system
Performance data
Parameter: glucose (mmol/L)
Material: aqueous solution, NIST traceable
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
0.800.800.0309100.5
10.0010.910.1274109.1
20.0020.210.3101101.0
30.0029.970.437799.9
40.0038.120.883395.3
Ta b le A - 36Parameter Glucose (mmol/L)
Correlation
Slope0.919 - 1.088
Intercept± 1.773
Correlation coefficient0.998
Parameter: lactate (mmol/L)
Material: aqueous solution, NIST traceable
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
0.500.450.006489.8
5.005.000.042099.9
10.0010.110.0873101.1
15.0014.840.092098.9
20.0019.070.281895.3
Ta b le A - 37Parameter Lactate (mmol/L)
Correlation
Slope0.961 - 1.041
Intercept± 0.191
Correlation coefficient0.9989
Roche DiagnosticsApril 2009
A-76Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Performance data
Parameter: urea (mmol/L)
Material: aqueous solution, NIST traceable
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
0.60.830.0145138.8
7.507.580.0921101.0
15.0014.840.232898.9
22.5022.130.321198.4
30.0029.620.509498.7
Ta b le A - 38Parameter Urea (mmol/L)
Correlation
Slope0.979 - 1.021
Intercept± 0.198
Correlation coefficient0.9991
Parameter: glucose (mmol/L)
Material: NIST 965
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
5.685.560.122197.9
11.1011.010.225099.2
16.3616.690.3826102.1
Ta b le A - 39Parameter Glucose (mmol/L)
Correlation
Slope0.9591 - 1.0426
Intercept± 0.4273
Correlation coefficient0.9991
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-77
4 Specificationscobas b 221 system
Performance data
Parameter: sodium (mmol/L)
Material: NIST 956a
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
121.40122.040.4136100.5
141.00141.370.2483100.3
160.90160.290.312799.6
Ta b le A - 40Parameter Sodium (mmol/L)
Correlation
Slope0.9719 - 1.0289
Intercept± 4.0475
Correlation coefficient0.9999
Parameter: potassium (mmol/L)
Material: NIST 956a
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
6.016.040.0202100.6
3.994.000.0103100.3
2.031.910.020994.1
Ta b le A - 41Parameter Potassium (mmol/L)
Correlation
Slope0.9629 - 1.0385
Intercept± 0.1788
Correlation coefficient0.9999
Roche DiagnosticsApril 2009
A-78Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Performance data
Parameter: sodium (mmol/L)
Material: NIST 909b
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
120.76119.960.366299.3
141.00144.310.4298102.3
Ta b le A - 42Parameter Sodium (mmol/L)
Correlation
Slope0.8311 - 1.2032
Intercept± 25.3383
Correlation coefficient0.9997
Parameter: potassium (mmol/L)
Parameter: chloride (mmol/L)
Material: NIST 909b
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
3.423.290.016296.1
6.286.560.0273104.4
Ta b le A - 43Parameter Potassium (mmol/L)
Correlation
Slope0.8738 - 1.1444
Intercept± 0.6284
Correlation coefficient1.0000
Material: NIST 909b
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
89.1188.590.667499.4
119.43115.960.976397.1
Ta b le A - 44Parameter Chloride (mmol/L)
Correlation
Slope0.9032 - 1.1072
Intercept± 8.1053
Correlation coefficient0.9990
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-79
4 Specificationscobas b 221 system
Performance data
Parameter: glucose (mmol/L)
Material: NIST 909b
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
5.405.060.029493.7
15.0012.170.123981.1
Ta b le A - 45Parameter Glucose (mmol/L)
Correlation
Slope0.7429 - 1.3461
Intercept± 1.0482
Correlation coefficient0.9997
Parameter: urea (mmol/L)
Material: NIST 909b
Number of instruments: 4 cobas b 221 systems
Measurements per level and instrument: 15
Expected valueMeanSwr Recovery
5.515.400.024898.0
Ta b le A - 46Parameter Urea (mmol/L)
Parameter: tHb (g/dL), SO
(%)
2
Material: Tonometered whole blood
Number of instruments: 4 cobas b 221<5> systems
ParameterSlopeinterceptCoefficientRangen
tHb0.9892 - 1.0109± 0.08330.99046-18 [g/dl]250
SO
2
Ta b l e A - 47Parameter tHb (g/dL), SO2 (%)
0.99999 - 1.00001± 0.8560.987451.7-100 [%]382
Roche DiagnosticsApril 2009
A-80Instructions for Use · Revision 10.0
cobas b 221 system4 Specifications
Performance data
Parameter: bilirubin (mg/dL)
Material: human whole blood incl. bilirubin
Number of instruments: 2 cobas b 221 systems
Measurements per level and instrument: 3
Expected valueMeanSwr Recovery
6.006.860.0928114.33
14.0014.550.1417103.93
28.0026.280.190193.86
44.0041.520.092094.36
Ta b le A - 48Parameter Bilirubin (mg/dL)
Correlation
Slope0.9038 - 1.1064
Intercept± 1.514
Correlation coefficient0.9996
Roche DiagnosticsApril 2009
Instructions for Use · Revision 10.0A-81
4 Specificationscobas b 221 system
Performance data
Correlation to other methods
pH
Comparison instrumentSlope and interceptBiasCorr. coeff. [r]No. of samples
(a) typical sample volume for Hct ≤ 45%
(b) typical sample volume for 45% < Hct ≤ 75%: if a sample with high Hct is expected, the sample volume for high Hct is recommended.
(c) The sample volume limitation is the maximum volume of sample which is aspirated from the container.
or COOX112128148
2
or COOX172186210
2
The volume limitation by the sample sensor depends on INSTALLED modules, regardless whether
they are activated or deactivated!
The actual required sample volume depends on the used sample container.