INTRODUCTION
Thank you for purchasing the ReliOn®HEM-780REL
Automatic Blood Pressure Monitor with the Easy Wrap Cuff.
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
Staple your purchase receipt here
Your new digital blood pressure monitor uses the oscillometric method
of blood pressure measurement. This means the monitor detects your
blood’s movement through your brachial artery and converts the
movements into a digital reading. An oscillometric monitor does not
need a stethoscope so the monitor is simple to use.
®
The ReliOn
• Monitor• Easy Wrap Cuff
• AC Adapter• 4 “AA” Batteries
• Storage Case
• Instruction Manual printed in English and Spanish
The ReliOn
Monitor is intended for home use.
HEM-780REL comes with the following components:
®
HEM-780REL Automatic Inflation Blood Pressure
SAVE THESE INSTRUCTIONS
3
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 3
IMPORTANT SAFETY NOTES
To assure the correct use of the product basic safety measures should
always be followed including the precautions listed below:
þ Read all information in the instruction book and any other literature
included in the box before using the unit.
þ Contact your physician for specific information about your blood
pressure. Follow the instructions of your healthcare provider.
þ Operate the unit only as intended. Do not use for any other purpose.
þ The unit is intended for use in measuring blood pressure and pulse
rate in the adult patient population. Do not use on infants or persons
who cannot express their intentions.
þ Do not use a cellular phone near the unit. It may result in
operational failure.
þ Do not plug or unplug the adapter power cord with wet hands.
þ Changes or modifications not approved by the manufacturer will void
the user warranty. Do not disassemble or attempt to repair the unit or
components.
4
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 4
BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing
for 30 minutes, before taking a measurement. Rest for at least 15
minutes before taking the measurement.
4. Remove tight-fitting clothing from your left arm.
5. Sit in a chair with your feet flat on the floor. Rest your left arm on
a table so that the cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Keep a record of your blood pressure and pulse readings for your
physician. A single measurement does not provide an accurate
indication of your true blood pressure. You need to take and record
several readings over a period of time. Try to measure your blood
pressure at the same time each day for consistency.
8. Wait 2-3 minutes between measurements. The wait time allows the
arteries to return to the condition prior to taking the blood pressure
measurement. You may need to increase the wait time depending on
your individual physiological characteristics.
5
780
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 5
KNOW YOUR UNIT
Main Unit:
MemoryButtons
Date/Time Setting Button
Components:
Grip
Arm Cuff
Air Plug
(Connects to Air Jack)
Display
Air Tube
Start/Stop
Button
Battery Compartment
Air Jack
AC Adapter Jack
Storage Case
4 ''AA” Batteries
AC Adapter
6
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 6
UNIT DISPLAY
Systolic Blood Pressure
Diastolic Blood Pressure
Memory Symbol
Average Value Symbol
Movement Error Symbol
Low Battery Symbol
Irregular Heartbeat Symbol
Heartbeat Symbol
Pulse Display
Deflation Symbol
Date/Time Display
IRREGULAR HEARTBEAT SYMBOL
IRREGULAR HEARTBEAT SYMBOL
When the monitor detects an irregular rhythm two or more times during
the measurement, the Irregular Heartbeat Symbol ( )
will appear on the display with the measurement values.
Pulse
Blood
pressure
Pulse
Blood
pressure
Normal Heartbeat
Irregular Heartbeat
Short Long
An irregular heartbeat rhythm is defined
as a rhythm that varies by less than 25%
from the average rhythm or 25% more
than the average rhythm detected while
the monitor is measuring the systolic
and diastolic blood pressure.
If the irregular heartbeat symbol ( )
displays with your measurement results we
recommend you consult your physician.
Follow the directions of your doctor.
7
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 7
BATTERY INSTALLATION
1. Press the
and slide the cover off in the direction of
the arrow.
2. Install 4 “AA” size batteries so the
+ (positive) and - (negative) polarities
match the polarities of the battery
compartment as indicated.
3. Replace the battery cover.
BATTERY REPLACEMENT
When the Low Battery Symbol appears on the display screen remove
all the batteries. Replace with four new batteries at the same time.
Long-life alkaline batteries are recommended.
NOTE:
▼ indicator on the battery cover
Low Battery Symbol
• Measurement values stored in the memory will not be deleted
during battery replacement.
• After replacing the batteries, reset the date and time.
Refer to page 10, Setting the Date and Time.
• Dispose of batteries according to applicable local regulations.
8
1
2
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 8
USING THE AC ADAPTER
Use only the authorized Omron AC Adapter, (Omron Model Number
HEM-ADPT1), with this monitor. Use of any other adapter
may cause damage to the monitor and will void the user warranty.
To connect the AC Adapter
① Insert the AC Adapter Plug into the AC Adapter Jack on the back side
of the monitor as illustrated.
② Plug the AC Adapter into a 120V AC outlet (60 cycles).
To disconnect the AC Adapter remove the AC Adapter from the electrical
outlet first and then remove the AC Adapter Plug from the monitor.
NOTE:
We recommend batteries are installed and used with the AC
•
Adapter. If batteries are not installed the Date and Time must
be set each time the AC Adapter is disconnected from
the electrical outlet.
• When the AC Adapter is used, the battery life is extended.
9
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 9
SETTING THE DATE AND TIME
Set the monitor to the current date and time before taking a measurement
for the first time. If the correct date and time are not set, the measurement
values stored in the memory and the average value will not be correct.
The Blood Pressure Monitor automatically stores up to 90 measurement
values with the date and time. The monitor also calculates an average
reading based on the last three measurement values.
1. TO START THE SETTING
Press the Date/Time Setting
button to adjust the date
and time.
The year flashes on the display.
2. SETTING THE YEAR
The year can be set between 2004 and 2030.
When the display reaches 2030, it will return
to 2004.
Press theMEMORY up button to advance
by increments of one year.
Press the Date/Time Setting button to set
the current year.
The month flashes on the display.
NOTES: • Press theMEMORY down button to decrease the day NOTE:•Press the down memory button to decrease the day and
and time values by increments of one.
time values by increments of one.
• Press and hold the MEMORY up button to increase or memory
down button to decrease the date and time values faster.
•Press and holdthe memory down button to increase
10
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 10
SETTING THE DATE AND TIME
3. SETTING THE MONTH
Press the MEMORY up button to advance by increments
of one month.
Press the Date/Time Setting button to set the current month.
The day flashes on the display.
SETTING THE DAY
4.
Press theMEMORY up button to
advance
by increments of one day.
Press the Date/Time Setting button to set
the current day.
The hour flashes on the display.
5.
SETTING THE HOUR
The time is set using AM or PM.
Press the MEMORY up button to advance by increments
of one hour.
Press the Date/Time Setting button to set the
current hour.
The minutes flash on the display.
11
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 11
SETTING THE DATE AND TIME
6. SETTING THE MINUTE
Press the MEMORY up button to advance by increments
of one minute.
Press the Date/Time Setting button to set the current minute.
The year flashes on the display.
7. Press the START/STOP button to turn the display off.
12
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 12
APPLYING THE ARM CUFF
1. Make sure the air plug is securely inserted in the main unit.
2. Remove tight-fitting clothing from your upper arm.
3. Sit in a chair with your feet flat on the floor. Place your arm on a
table so the cuff is level with your heart.
13
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 13
APPLYING THE ARM CUFF
4. Hold the grip on the cuff securely
Grip
with your hand.
Closure Strip
5. Turn the palm of your hand upward.
6. Apply the cuff to your left upper arm so the blue strip is on the inside
of your arm and aligned with your middle finger. The air tube runs
down the inside of your arm. The bottom of the cuff should be
approximately 1/2" above your elbow.
Blue Strip
1/2 inch (1 - 2 cm)
14
Air Tube
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 14
APPLYING THE ARM CUFF
7. Wrap the cuff firmly in place around your arm using the closure strip.
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm use the
instruction below.
Apply the cuff to your right upper arm so the grip is centered on
the inside of your inner arm. The bottom of the cuff should be
approximately 1/2'' above your elbow.
Grip
NOTE: Be careful not to rest your arm on the air tube. This will restrict
the flow of air to the cuff
15
R
EL
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 15
TAKING A MEASUREMENT
The monitor automatically determines your ideal inflation level. In rare
circumstances when a higher inflation may be necessary, the monitor
automatically re-inflates the cuff up to 30 mmHg higher than the initial
inflation and restarts the measurement.
Press the START/STOP button.
1.
All display symbols appear on the screen. The cuff starts to inflate
automatically.
NOTE: To stop the inflation or measurement, push the
START/STOP button. The monitor will stop inflating,
start deflating, and will turn off.
2. Inflation stops automatically and the measurement is started.
As the cuff deflates, decreasing
numbers appear on the display.
The Heartbeat Symbol ()
flashes at every heartbeat.
16
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 16
TAKING A MEASUREMENT
You3blood pressure and pulse rate are dis-
3. When the measurement is complete, the arm
played.
cuff completely deflates. Your blood pressure
and pulse rate are displayed.
Press the START/STOP button to
4.
turn the monitor off.
NOTE: The monitor will automatically
turn off after five minutes.
INSTRUCTIONS FOR SPECIAL CONDITIONS
If your systolic pressure is known
to be more than 220 mmHg,
press and hold the START/STOP
button until the monitor inflates
30 to 40 mmHg higher than your
suspected systolic pressure.
NOTES:
• Do not apply more pressure than necessary.
• The monitor will not inflate above 299 mmHg.
17