ReliOn HEM-780REL User Manual

780
INSTRUCTION MANUAL
Automatic Blood Pressure Monitor
with Easy Wrap Cuff
Model HEM-780REL
ESPAÑOL
ENGLISH
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 1
CONTENTS
Before Using the Monitor
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3
Important Safety Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .4
Before Taking a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . .5
Operating Instructions
Know Your Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .6
Unit Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .7
Irregular Heartbeat Symbol . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8
Using the AC Adapter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9
Setting the Date and Time . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10
Applying the Arm Cuff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13
Taking a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16
Using the Memory Function . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
Care and Maintenance
Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .20
Error Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .21
Troubleshooting Tips . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22
Quick Reference Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23
FCC Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25
Warranty Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .26
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .27
2
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 2
INTRODUCTION
Thank you for purchasing the ReliOn®HEM-780REL Automatic Blood Pressure Monitor with the Easy Wrap Cuff.
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
Staple your purchase receipt here
Your new digital blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through your brachial artery and converts the movements into a digital reading. An oscillometric monitor does not need a stethoscope so the monitor is simple to use.
®
The ReliOn
• Monitor • Easy Wrap Cuff
• AC Adapter • 4 “AA” Batteries
• Storage Case
• Instruction Manual printed in English and Spanish The ReliOn
Monitor is intended for home use.
HEM-780REL comes with the following components:
®
HEM-780REL Automatic Inflation Blood Pressure
SAVE THESE INSTRUCTIONS
3
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 3
IMPORTANT SAFETY NOTES
To assure the correct use of the product basic safety measures should always be followed including the precautions listed below:
þ Read all information in the instruction book and any other literature
included in the box before using the unit.
þ Contact your physician for specific information about your blood
pressure. Follow the instructions of your healthcare provider.
þ Operate the unit only as intended. Do not use for any other purpose.
þ The unit is intended for use in measuring blood pressure and pulse
rate in the adult patient population. Do not use on infants or persons who cannot express their intentions.
þ Do not use a cellular phone near the unit. It may result in
operational failure.
þ Do not plug or unplug the adapter power cord with wet hands.
þ Changes or modifications not approved by the manufacturer will void
the user warranty. Do not disassemble or attempt to repair the unit or components.
4
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 4
BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
1. Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes, before taking a measurement. Rest for at least 15 minutes before taking the measurement.
2. Stress raises blood pressure. Avoid taking measurements during stressful times.
3. Measurements should be taken in a quiet place.
4. Remove tight-fitting clothing from your left arm.
5. Sit in a chair with your feet flat on the floor. Rest your left arm on a table so that the cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Keep a record of your blood pressure and pulse readings for your physician. A single measurement does not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time. Try to measure your blood pressure at the same time each day for consistency.
8. Wait 2-3 minutes between measurements. The wait time allows the arteries to return to the condition prior to taking the blood pressure measurement. You may need to increase the wait time depending on your individual physiological characteristics.
5
780
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 5
KNOW YOUR UNIT
Main Unit:
Memory Buttons
Date/Time Setting Button
Components:
Grip
Arm Cuff
Air Plug
(Connects to Air Jack)
Display
Air Tube
Start/Stop Button
Battery Compartment
Air Jack
AC Adapter Jack
Storage Case
4 ''AA” Batteries
AC Adapter
6
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 6
UNIT DISPLAY
Systolic Blood Pressure
Diastolic Blood Pressure
Memory Symbol Average Value Symbol
Movement Error Symbol
Low Battery Symbol
Irregular Heartbeat Symbol
Heartbeat Symbol
Pulse Display Deflation Symbol
Date/Time Display
IRREGULAR HEARTBEAT SYMBOL
IRREGULAR HEARTBEAT SYMBOL
When the monitor detects an irregular rhythm two or more times during the measurement, the Irregular Heartbeat Symbol ( ) will appear on the display with the measurement values.
Pulse
Blood
pressure
Pulse
Blood
pressure
Normal Heartbeat
Irregular Heartbeat
Short Long
An irregular heartbeat rhythm is defined as a rhythm that varies by less than 25% from the average rhythm or 25% more than the average rhythm detected while the monitor is measuring the systolic and diastolic blood pressure.
If the irregular heartbeat symbol ( ) displays with your measurement results we recommend you consult your physician. Follow the directions of your doctor.
7
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 7
BATTERY INSTALLATION
1. Press the and slide the cover off in the direction of the arrow.
2. Install 4 “AA” size batteries so the + (positive) and - (negative) polarities match the polarities of the battery compartment as indicated.
3. Replace the battery cover.
BATTERY REPLACEMENT
When the Low Battery Symbol appears on the display screen remove all the batteries. Replace with four new batteries at the same time. Long-life alkaline batteries are recommended.
NOTE:
indicator on the battery cover
Low Battery Symbol
• Measurement values stored in the memory will not be deleted during battery replacement.
• After replacing the batteries, reset the date and time. Refer to page 10, Setting the Date and Time.
• Dispose of batteries according to applicable local regulations.
8
1
2
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 8
USING THE AC ADAPTER
Use only the authorized Omron AC Adapter, (Omron Model Number HEM-ADPT1), with this monitor. Use of any other adapter may cause damage to the monitor and will void the user warranty.
To connect the AC Adapter
Insert the AC Adapter Plug into the AC Adapter Jack on the back side
of the monitor as illustrated.
Plug the AC Adapter into a 120V AC outlet (60 cycles).
To disconnect the AC Adapter remove the AC Adapter from the electrical outlet first and then remove the AC Adapter Plug from the monitor.
NOTE:
We recommend batteries are installed and used with the AC
Adapter. If batteries are not installed the Date and Time must be set each time the AC Adapter is disconnected from the electrical outlet.
• When the AC Adapter is used, the battery life is extended.
9
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 9
SETTING THE DATE AND TIME
Set the monitor to the current date and time before taking a measurement for the first time. If the correct date and time are not set, the measurement values stored in the memory and the average value will not be correct.
The Blood Pressure Monitor automatically stores up to 90 measurement values with the date and time. The monitor also calculates an average reading based on the last three measurement values.
1. TO START THE SETTING
Press the Date/Time Setting button to adjust the date and time.
The year flashes on the display.
2. SETTING THE YEAR
The year can be set between 2004 and 2030. When the display reaches 2030, it will return to 2004.
Press the MEMORY up button to advance by increments of one year.
Press the Date/Time Setting button to set the current year.
The month flashes on the display.
NOTES: • Press the MEMORY down button to decrease the day NOTE: Press the down memory button to decrease the day and
and time values by increments of one.
time values by increments of one.
• Press and hold the MEMORY up button to increase or memory
down button to decrease the date and time values faster.
Press and holdthe memory down button to increase
10
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 10
SETTING THE DATE AND TIME
3. SETTING THE MONTH
Press the MEMORY up button to advance by increments of one month.
Press the Date/Time Setting button to set the current month.
The day flashes on the display.
SETTING THE DAY
4.
Press the MEMORY up button to advance by increments of one day.
Press the Date/Time Setting button to set the current day.
The hour flashes on the display.
5.
SETTING THE HOUR
The time is set using AM or PM.
Press the MEMORY up button to advance by increments of one hour.
Press the Date/Time Setting button to set the current hour.
The minutes flash on the display.
11
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 11
SETTING THE DATE AND TIME
6. SETTING THE MINUTE
Press the MEMORY up button to advance by increments of one minute.
Press the Date/Time Setting button to set the current minute.
The year flashes on the display.
7. Press the START/STOP button to turn the display off.
12
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 12
APPLYING THE ARM CUFF
1. Make sure the air plug is securely inserted in the main unit.
2. Remove tight-fitting clothing from your upper arm.
3. Sit in a chair with your feet flat on the floor. Place your arm on a table so the cuff is level with your heart.
13
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 13
APPLYING THE ARM CUFF
4. Hold the grip on the cuff securely
Grip
with your hand.
Closure Strip
5. Turn the palm of your hand upward.
6. Apply the cuff to your left upper arm so the blue strip is on the inside of your arm and aligned with your middle finger. The air tube runs down the inside of your arm. The bottom of the cuff should be approximately 1/2" above your elbow.
Blue Strip
1/2 inch (1 - 2 cm)
14
Air Tube
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 14
APPLYING THE ARM CUFF
7. Wrap the cuff firmly in place around your arm using the closure strip.
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm use the instruction below.
Apply the cuff to your right upper arm so the grip is centered on the inside of your inner arm. The bottom of the cuff should be approximately 1/2'' above your elbow.
Grip
NOTE: Be careful not to rest your arm on the air tube. This will restrict
the flow of air to the cuff
15
R
EL
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 15
TAKING A MEASUREMENT
The monitor automatically determines your ideal inflation level. In rare circumstances when a higher inflation may be necessary, the monitor automatically re-inflates the cuff up to 30 mmHg higher than the initial inflation and restarts the measurement.
Press the START/STOP button.
1. All display symbols appear on the screen. The cuff starts to inflate
automatically.
NOTE: To stop the inflation or measurement, push the
START/STOP button. The monitor will stop inflating, start deflating, and will turn off.
2. Inflation stops automatically and the measurement is started. As the cuff deflates, decreasing
numbers appear on the display. The Heartbeat Symbol ( ) flashes at every heartbeat.
16
780RELN IM Final 8-29-05.qxd 8/29/05 5:24 PM Page 16
TAKING A MEASUREMENT
You3blood pressure and pulse rate are dis-
3. When the measurement is complete, the arm played.
cuff completely deflates. Your blood pressure and pulse rate are displayed.
Press the START/STOP button to
4. turn the monitor off.
NOTE: The monitor will automatically
turn off after five minutes.
INSTRUCTIONS FOR SPECIAL CONDITIONS
If your systolic pressure is known to be more than 220 mmHg, press and hold the START/STOP button until the monitor inflates 30 to 40 mmHg higher than your suspected systolic pressure.
NOTES:
• Do not apply more pressure than necessary.
• The monitor will not inflate above 299 mmHg.
17