ReliOn 7100REL User Manual

4 (1)

INSTRUCTION MANUAL

Automatic Blood Pressure Monitor

Model 7100REL

ENGLISH ESPAÑOL

TABLE OF CONTENTS

Before Using the Unit

 

Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

3

Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4

Operating the Device . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

4

Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

6

Before Taking a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

7

Operating Instructions

Know Your Unit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .8 Unit Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .9 Display Symbols . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Heartbeat Symbol ( ) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .10 Battery Installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .11 Applying the Arm Cuff . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .13 Taking a Measurement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .16 Using the Memory function . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18

Care and Maintenance

Care and Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .19 Error Indicators and Troubleshooting Tips . . . . . . . . . . . . . . . . . . . . . .20 FCC Statement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .22 Warranty . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .23 Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .24 Notes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .25

2

Before Using the Unit

INTRODUCTION

Thank you for purchasing the ReliOn® 7100REL Automatic Blood Pressure Monitor.

Fill in for future reference.

DATE PURCHASED: _______________

SERIAL NUMBER: _________________

Staple your purchase receipt here

Your new digital blood pressure monitor uses the oscillometric method of blood pressure measurement. This means the monitor detects your blood’s movement through your brachial artery and converts the movements into a digital reading.

The 7100REL comes with the following components:

• Monitor

• Standard Adult Arm Cuff

• 4 “AAA” Batteries

• Instruction Manual printed in English and Spanish

 

The ReliOn® 7100REL Automatic Blood Pressure Monitor is intended for home use.

SAVE THESE INSTRUCTIONS

3

SAFETY INFORMATION

To assure the correct use of the product basic safety measures should always be followed including the warnings and cautions listed in this instruction manual.

SAFETY SYMBOLS USED IN THIS INSTRUCTION MANUAL

Indicates a potentially hazardous situation

WARNING which, if not avoided, could result in death or serious injury.

Indicates a potentially hazardous situation

which, if not avoided, may result in minor or CAUTION moderate injury to the user or patient or

damage to the equipment or other property.

OPERATING THE DEVICE

Contact your physician for specific information about your blood pressure. Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider.

The monitor is not intended to be a diagnostic device.

If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Contact a physician immediately.

Read all of the information in the instruction manual and any other literature in the box before operating the unit.

This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions.

4

SAFETY INFORMATION

OPERATING THE DEVICE (continued)

Read carefully “Taking a Measurement” No.3 of the instruction manual if your systolic pressure is known to be higher.

Inflating to a higher pressure than necessary may result in bruising where the cuff is applied.

Operate the device only as intended. Do not use the device for any other purpose.

Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.

Do not use a cellular phone near the device. It may result in an operational failure.

Do not inflate the arm cuff when it is not wrapped around your arm.

Use only authorized parts and accessories. Parts and accessories not approved for use with the device may damage the unit.

Use only 1.5V “AAA” alkaline batteries with this device. Do not use any other type of battery. This may damage the unit.

5

SAFETY INFORMATION

CARE AND MAINTENANCE

Do not subject the monitor to strong shocks, such as dropping the unit on the floor.

Do not submerge the device or any of the components in water.

Store the device and the components in a clean, safe location.

Changes or modification not approved by the manufacturer will void the user warranty. Do not disassemble or attempt to repair the unit or components.

6

BEFORE TAKING A MEASUREMENT

To ensure a reliable reading follow these recommendations:

1.Avoid eating, drinking alcohol, smoking, exercising, and bathing for 30 minutes before taking a measurement. Rest for at least 15 minutes before taking the measurement.

2.Stress raises blood pressure. Avoid taking measurements during stressful times.

3.Measurements should be taken in a quiet place.

4.Remove tight-fitting clothing from your arm.

5.Sit on a chair with your feet flat on the floor. Rest your arm on a table so that the cuff is at the same level as your heart.

6.Remain still and do not talk during the measurement.

7.Wait 2-3 minutes between measurements. The wait time allows the arteries to return to the condition prior to taking the blood pressure measurement. You may need to increase the wait time depending on your individual physiological characteristics.

8.Keep a record of your blood pressure and pulse readings for your physician. A single measurement does not provide an accurate indication of your true blood pressure. You need to take and record several readings over a period of time. Try to measure your blood pressure at about the same time each day for consistency.

7

ReliOn 7100REL User Manual

Operating Instructions

KNOW YOUR UNIT

Main Unit:

Display

Battery Compartment

On/Off Start Button

Air Jack

 

Components:

Air Tube

Air Plug

(Connects to Air Jack)

Arm Cuff

Four “AAA” Alkaline

(LR03) Batteries

INSTRUCTION MANUAL

Automatic Blood Pressure Monitor

Model 7100REL

ESPAÑOL

ENGLISH

Instruction Manual

8

UNIT DISPLAY

Systolic Blood Pressure

Diastolic Blood Pressure

Heartbeat Symbol

Memory Symbol

Low Battery Indicator

Pulse Display

Deflation Symbol

9

DISPLAY SYMBOLS

The American Heart Association (AHA) recommends the following guideline for upper limit of normal for home blood pressure readings:

AHA Home Guideline for Upper Limit of Normal BP

Systolic Blood Pressure

135 mmHg

 

 

Diastolic Blood Pressure

85 mmHg

 

 

This is a general guideline because blood pressure depends on age and morbidity. The AHA recommends a lower home blood pressure goal for certain patients, including diabetic patients, pregnant women, and patients with renal failure1.

WARNING

The monitor is not intended to be a diagnostic device.

Contact your physician for specific information about your blood pressure. Self-diagnosis and treatment using measured results may be dangerous. Follow the instructions of your physician or licensed healthcare provider.

Heartbeat Symbol ( )

TAKING A MEASUREMENT

The Heartbeat Symbol flashes on the display during the measurement. The Heartbeat Symbol flashes at every heartbeat.

1 American Heart Association Scientific Statement 2004

10

BATTERY INSTALLATION

1.Turn the main unit upside down.

2.Press the indicator on the battery cover and slide the cover off in the direction of the arrow.

3.Install 4 “AAA” size batteries so the + (positive) and - (negative) polarities match the polarities of the battery compartment as indicated.

4.Replace the battery cover. Slide the battery cover as indicated until it clicks into place.

Battery Cover

11

BATTERY INSTALLATION

BATTERY REPLACEMENT

Low Battery Indicator

When the Low Battery Indicator appears on the display screen turn the monitor off and remove all the batteries. Replace with four new batteries at the same time.

WARNING

If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Contact a physician immediately.

NOTE: The batteries supplied with the device may have a shorter life.

CAUTION

Dispose of the device, components and optional accessories according to applicable local regulations. Unlawful disposal may cause environmental pollution.

12

APPLYING THE ARM CUFF

CAUTION

Read all of the information in the instruction manual and any other literature in the box before operating the unit.

CAUTION

This device is intended for use in measuring blood pressure and pulse rate in the adult population. Do not use this device on infants or persons who cannot express their intentions.

APPLYING THE CUFF ON THE LEFT ARM

1. Make sure the air plug is securely inserted in the main unit.

NOTE: If the cuff is not assembled, lay the cuff flat with the metal D ring facing down and on the right side. Pass the left edge through the metal D ring to form a loop. The smooth cloth should be inside of the loop.

2. Remove tight-fitting clothing from your upper left arm.

13

APPLYING THE ARM CUFF

3.Apply the cuff to your left upper arm so the blue arrow is centered on the inside of your arm and aligned with your middle finger. The air tube runs down the inside of your arm. The bottom of the cuff should be approximately 1/2" above your elbow.

1/2 inch (1 - 2 cm)

Blue Arrow

Air Tube

 

4. Wrap the cuff firmly in place around your arm using the closure strip.

14

APPLYING THE ARM CUFF

APPLYING THE CUFF ON THE RIGHT ARM

When taking a measurement using the right arm use this instruction for Step 4 on page 14.

Apply the cuff to your right upper arm so the blue arrow and the air tube runs along the side of your arm. The bottom of the cuff should be approximately 1/2'' above your elbow.

NOTE: Be careful not to rest your arm on the air tube. This will restrict the flow of air to the cuff.

15

TAKING A MEASUREMENT

CAUTION

Operate the device only as intended. Do not use the device for any other purpose.

1.Sit on a chair with your feet flat on the floor. Place your left arm on a table so the cuff is level with your heart.

Remain still and do not move your arm until the entire measurement is complete.

2. Press the ON/OFF START button.

All symbols appear on the display.

CAUTION

Inflating to a higher pressure than necessary may result in bruising where the cuff is applied.

16