ReliOn Confirm Owner's Manual

5 (1)

Blood Glucose Monitoring System

User

Instruction

Manual

CAUTION:

Before using any product to test your blood sugar (blood glucose), read all instructions and practice the test. Do all quality control checks as directed and consult with a diabetes healthcare professional. These

recommendations apply to all blood glucose monitoring systems and are supported by the American Association of Diabetes Educators, the American Diabetes Association, the U.S. Food and Drug Administration, and the Health Industry Manufacturers Association.

TABLE OF CONTENTS

 

Introduction. . . . . . . . . . . . . . . . . . . . . . 5

Intended Use . . . . . . . . . . . . . . . . . . . . 6.

Understanding the ReliOn® Confirm Blood Glucose Meter . . . 7

Changing the Battery . . . . . . . . . . . . . . . . 10.

Meter Set-up . . . . . . . . . . . . . . . . . . . . . 13

Basic Rules . . . . . . . . . . . . . . . . . . . . 13

Audible Beep. . . . . . . . . . . . . . . . . . . 16

Year . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . 18

Month/Day . . . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . 19.

Hour/Minute. . . . . . . . . . . . . . . . . . . 20.

Test Averaging . . . . . . . . . . . . . . . . . . 21.

Ending Set-up . . . . . . . . . . . . . . . . . . . . . . . .

. . . . . . . . . . . . 22

Performing a Control Solution Test . . . . . . . . . . . .

. . . . . . . . . . . . 23

Troubleshooting Control Solution Test. . .

. . . . . . 30

3

Blood Glucose Testing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33 The Lancing Device. . . . . . . . . . . . . . . . . . . . 33 Getting a Drop of Blood . . . . . . . . . . . . . . . . . . 36 Performing a Blood Glucose Test. . . . . . . . . . . . . . . 38 Flagging Test Results. . . . . . . . . . . . . . . . . . . . 42

Alternate Site Testing (AST) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44 Understanding Test Results. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 53 Recalling Results - Entering Memory Mode. . . . . . . . . . . . 58 Maintenance. . . . . . . . . . . . . . . . . . . . . . . . 62 Cleaning. . . . . . . . . . . . . . . . . . . . . . . . . 62 Storage and Handling . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 63 Display Screen and Error Messages. . . . . . . . . . . . . . 64 Product Information . . . . . . . . . . . . . . . . . . . . . 69 Product Specifications. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 69 Warnings, Precautions & Limitations . . . . . . . . . . . . . . . . . . . . . . . . . . 70 Warranty. . . . . . . . . . . . . . . . . . . . . . . . . 73

4

INTRODUCTION

Thank you for choosing the ReliOn® Confirm Blood Glucose Monitoring System. This system measures blood glucose in fresh capillary whole blood. Results are shown as plasma values. This makes it easy to compare the ReliOn® Confirm blood glucose meter and lab results.

Please read this manual before using the system. If you have any questions, call Customer Service at 800.631.0076 (24 hours a day, 7 days a week). Contact your healthcare professional with questions if you are unable to reach Customer Service.

5

Intended Use

The ReliOn® Confirm Blood Glucose Monitoring System meter is intended for the quantitative measurement of glucose in fresh capillary whole blood samples drawn from the fingertips or palm. Testing is done outside the body (In Vitro diagnostic use). It is indicated for use at home (over the counter [OTC]) by persons with diabetes, or in clinical settings by healthcare professionals, as an aid to monitor the effectiveness of diabetes control.

The System Includes:

• Meter

• Carrying Case

• User Instruction Manual

Quick Reference Guide

Self-testing Logbook

• Lancing Device

• Lancets

Carrying Case

6

ReliOn Confirm Owner's Manual

UNDERSTANDING THE ReliOn® Confirm BLOOD GLUCOSE METER

Top

Bottom

Battery Compartment

Cover

Tab

Test Strip

Port

Connection Terminal

Up

Arrow

Down

Arrow

7

Explanation of the full display screen:

Memory Mode Symbol

Battery Symbol

Audible Beep Off

Apply Blood Symbol

 

Date

Unit of Measurement

 

Temperature Warning

Time Format

 

Delete Symbol

Control Solution

NOTE: Meters in the USA are programmed to show results in mg/dL. Meters used outside the USA are programmed to show results in either mg/dL or mmol/L. Results displayed in mmol/L will always have a decimal point. Results displayed in mg/dL will never have a decimal point.

8

ReliOn® Confirm/micro test strips

These test strips need only 0.3 µL (microliter) of blood. Apply blood to the sample application tip. The strip draws in the blood by capillary action. The blood glucose reacts with the enzyme on the test strip. This produces a current proportional to the blood glucose level. The meter detects the current and converts it into a blood glucose reading.

Test Strip Bottle

Code No.

Code(F-2, 3Noor. 4)

(F-2, 3 or 4)

F- 4

F-4

Expiration date

10

0018

Front

Back

 

Test Strip

 

Contact Bars

Insert this

Sample

way into

Application

meter

Tip (Apply blood

 

 

to end of test strip.

 

Capillary draw will

 

pull blood to

 

reaction site).

9

Changing the Battery

The ReliOn® Confirm blood glucose meter comes with one lithium battery CR2032 (or DL2032).

CAUTION: Replacing the battery will not delete meter settings or test results stored in the meter unless you:

Touch the metal portion on main unit with your fingers or any metal.

Take over 2 minutes to replace battery.

Load the battery incorrectly.

Load a used or weak battery.

NOTE:

To save battery power, the meter turns itself off 3 – 5 minutes after last use.

Replace batteries when low battery symbol appears. Otherwise inaccurate results could be obtained.

Discard used batteries as per local guidelines.

WARNING: Keep all batteries out of children’s reach. If a battery is swallowed, consult a doctor immediately.

10

To replace battery:

1. Open the battery cover on the back side of the meter by pushing the tab in the direction of the arrow and pulling the cover up.

2. The battery is held in place on the underside of the cover. Slide the battery out of the cover.

CAUTION: Do not touch the metal contacts inside the battery compartment with your finger or any metal.

3. Load a new battery into the battery cover. Place the “+” side down when loading the battery into the cover.

4. Make sure the battery is loaded in the proper position. Otherwise, the preprogrammed times will be reset.

11

5. Put the battery cover back into correct position and snap to close.

If the clock setting has been deleted, either “12:00” or “0:00” will blink when you put a test strip into the meter. The low battery symbol means you need to replace the battery as soon as possible. Discard used batteries as per local guidelines.

12

METER SET-UP

Before using the ReliOn® Confirm blood glucose meter for the first time, set the time and date. Do not perform a blood glucose test until you have setup the meter.

Meter set-up allows you to adjust the audible beep volume and to set time, date, and test averaging.

Basic rules for meter set-up:

Step 1: Enter meter set-up by first making sure the meter is turned off.

Up

Arrow

Down

Arrow

13

Step 2: Press the Up Arrow and Down Arrow buttons at the same time for 5 seconds. A beep sounds and the meter turns on. Next, a second beep sounds and “SET” shows on the screen.

Step 3: Use the Up Arrow to change the set-up screen. When you press the Up Arrow, the set-up screen changes in the order shown in the diagram on the following page. Press the Down Arrow button to change a setting within a set-up screen menu.

Note: If you need to go back to a setting, you must scroll through each set-up option. You cannot go backwards in set-up mode.

In set-up mode, the meter will turn off automatically after 3 minutes of non-use.

14

 

Setting

 

 

 

 

 

 

 

the

 

Setting

 

Setting the

 

Setting the

1

Audible

2

3

4

 

Beep

 

the Year

 

Month

 

Day

Up

Up

 

 

Up

 

Up

 

Arrow

Arrow

 

 

Arrow

 

Arrow

 

Gray Arrow

Return

 

 

 

 

8

Ending

 

Setting

 

Setting

7

Test

6

the

 

Set-up

Averaging

Minute

Up

Up

 

 

Up

 

Arrow

Arrow

 

Arrow

 

<Power OFF>

UP

Arrow

Setting the

5 Hour

UP

Arrow

15

Setting the Audible Beep

If the audible beep is turned “on”, the beep will sound when:

You insert a test strip into the meter.

You apply blood or control solution to a test strip.

The test has ended and the result is displayed.

The meter is turned on, but you do not use it for several minutes.

Meter will beep and automatically shut off.

You go into Memory Mode.

Make sure the meter is off. Press the Up Arrow and Down Arrow buttons at the same time for 5 seconds. A beep sounds and the meter

turns on. Next, a second beep sounds and “SET” shows on the screen.

Beep

Press the Up Arrow button. The screen will display “bUZ” and the current beep setting.

16

Press the Down Arrow button to cycle through the volume levels:

“OFF” (no sound) “on 1” (low)

“on 2”(middle) “on 3” (high)

“OFF” (no sound) …, etc.

A beep sounds at the set volume at each step. If “OFF” is set, the “audible beep mute (Off)” symbol lights up in the upper right of the screen. After you select the beep volume, press the Up Arrow button to confirm the setting. The screen changes to the year set-up screen.

[1]

Beep

p Bee

[1]

[1]

Beep

[1]

17

Setting the Year

In the year set-up screen, the last two digits of the currently set year appear at the top of the screen. “yER” appears in the middle of the screen. All four [1] digits of the currently set year appear at the

bottom of the screen. The last two digits will blink.

Press the Down Arrow to change the last

two digits of the year to the correct year. After you [2] set the year, press the Up Arrow button to confirm

the setting. The screen changes to the month set-up screen.

[3]

18

Setting the Month

Press the Down Arrow button to change the month. After you set the month, press the Up Arrow button to [1] confirm the setting. The screen changes to the day set-up screen.

Setting the Day

Press the Down Arrow button to change the day. After you set the day, press the Up Arrow button to confirm the setting. The screen changes to the hour set-up screen.

[2]

[3]

[4]

19

Setting the Hour

Press the Down Arrow button to change the hour. After you set the hour, press the Up Arrow button to confirm the setting. The screen changes to the minute set-up screen.

Setting the Minute

Press the Down Arrow button to change the minute. After you set the minute, press the Up Arrow button to confirm the setting. The screen changes to the test averaging set-up screen.

[1]

[2]

[3]

[4]

20

Setting Test Averaging

NOTE: “Lo” results are included as 20 mg/dL in the averages. “Hi” results are included as 600 mg/dL in the averages.

Press the Down Arrow button to cycle through the test averaging choices: “oFF” (not displayed), “14” (14-day average), or“30” (30-day average).“AVE” appears near the middle of the screen.

If you select either 14-day or 30-day averaging, your average test result will appear when you recall results from memory. See page 58.

After you set test averaging, press the

Up Arrow button to confirm the setting.

21

Ending Set-up

To make further changes to set-up settings, press the Down Arrow button while “End” is displayed. This will bring up the Audible Beep set-up screen. You can then move through the

set-up mode and adjust any settings that you wish

If you are done with set-up, press the Up Arrow button while “End” is displayed. This confirms all settings and turns the meter off.

Beep

<Power OFF>

The meter will turn off automatically 3 minutes after “End” is displayed.

22

PERFORM A CONTROL SOLUTION TEST

Use ReliOn® Confirm/micro CONTROL to check if:

the meter and test strips are working correctly as a system.

you are testing correctly.

To order control solution, call ReliOn® at 800.631.0076.

23

Loading...
+ 51 hidden pages