Thank you for purchasing ReliOn®HEM-412CREL Manual Inflation
Blood Pressure Monitor
Fill in for future reference.
DATE PURCHASED: __________________
SERIAL NUMBER: ____________________
Staple your purchase receipt here
Your new digital blood pressure monitor uses the oscillometric method
of blood pressure measurement. This means the monitor detects your
blood’s movement through your brachial artery and converts the
movements into a digital reading. An oscillometric monitor does not
need a stethoscope so the monitor is simple to use.
The HEM-412CREL comes with the following components:
• Main Unit• Arm Cuff
• Inflation Bulb
• Instruction Manual printed in English and Spanish
The HEM-412CREL Manual Inflation Blood Pressure Monitor is
intended for home use.
SAVE THESE INSTRUCTIONS
3
IMPORTANT SAFETY NOTES
To assure the correct use of the product basic safety measures should
always be followed including the precautions listed below:
þ Read all information in the instruction book and any other literature
included in the box before using the unit.
þ Contact your physician for specific information about your blood
pressure. Follow the instructions of your healthcare provider.
þ Operate the unit only as intended. Do not use for any other purpose.
þ The unit is intended for use in measuring blood pressure and pulse
rate in adult patient population. Do not use on infants.
þ Do not use a cellular phone near the unit. It may result in operational
failure.
þ Changes or modifications not approved by the manufacturer will void
the user warranty. Do not disassemble or attempt to repair the unit or
components.
4
BEFORE TAKING A MEASUREMENT
To ensure a reliable reading follow these recommendations:
1. Avoid eating, smoking, and exercising for 30 minutes before taking a
measurement. Rest for at least 15 minutes before taking the
measurement.
4. Remove tight-fitting clothing from your left arm.
5. Sit in a chair with your feet flat on the floor. Rest your left arm on a
table so that the cuff is at the same level as your heart.
6. Remain still and do not talk during the measurement.
7. Keep a record of your blood pressure and pulse readings for your
physician. A single measurement does not provide an accurate
indication of your true blood pressure. You need to take and record
several readings over a period of time. Try to measure your blood
pressure at the same time each day for consistency.
8. Wait 2-3 minutes between measurements. The wait time allows the
arteries to return to the condition prior to taking the blood pressure
measurement. You may need to increase the wait time depending on
your individual physiological characteristics.
5
Operating Instructions
KNOW YOUR UNIT
Main Unit:
Settings Button
Components:
Cuff Air Plug
Arm Cuff
Display
MEMORY Button
Battery Compartment
ON/OFF Button
Bulb Air Plug
Air Tube
Air Tube
6
Air Release Button
Inflation Bulb
Systolic Blood Pressure
Diastolic Blood Pressure
Memory Symbol
Average Value Symbol
Battery Low Symbol
UNIT DISPLAY
Heartbeat Symbol
Pulse Display
Deflation Symbol
Date/Time Display
Reinflation Symbol
7
BATTERY INSTALLATION
1. Press the ▼ indicator on the
battery cover and slide the cover
off in the direction of the arrow.
2. Install 4 “AAA” size batteries so
the + (positive) and - (negative)
polarities match the polarities of
the battery compartment as indicated.
3. Replace the battery cover.
BATTERY REPLACEMENT
Low Battery Indicator
When the Low Battery Indicator appears on the display screen remove
all the batteries. Replace with four new batteries at the same time. Long
life alkaline batteries are recommended.
NOTE: Measurement values stored in the memory will not be deleted
during battery replacement.
After replacing the batteries, reset the date and time.
Refer to page 9, Setting the Date and Time.
8
SETTING THE DATE AND TIME
Set the monitor to the current date and time before taking a measurement
for the first time. If the correct date and time are not set, the measurement
values stored in the memeory and the value will not be correct.
The Blood Pressure Monitor automatically stores up to 30
measurement values. The monitor also calculates an average reading
based on the three most recent measurements taken within 10 minutes
of the last reading.
NOTE:
• If only two readings are stored in the memory for the 10-minute
period, the average is based on the two readings.
• If one reading is stored in the memory for the 10-minute period,
this reading is displayed as the average.
1. TO START THE SETTING PRESS THE
ON/OFF BUTTON
Press the Date/Time Setting button to
adjust the date and time.
The year flashes on the display.
9
SETTING THE DATE AND TIME
2. SETTING THE YEAR
The year can be set between 2005 and 2030.
When the display reaches 2030, it will return
to 2005.
Press the MEMORY button to
advance by increments of one year.
Press the Date/Time Setting button to
set the current year.
The month flashes on the display.
NOTE:
• Press and hold the MEMORY button to
advance the date and time values faster.
3. SETTING THE MONTH
Press the MEMORY button to advance by increments of one month.
Press the Date/Time Setting button to set the
current month.
The day flashes on the display.
4. SETTING THE DAY
Press the MEMORY button to advance by increments of one day.
Press the Date/Time Setting button to set the
current day.
The hour flashes on the display.
10
SETTING THE DATE AND TIME
5. SETTING THE HOUR
The time is set using AM or PM.
Press the MEMORY button to advance by increments of one hour.
Press the Date/Time Setting button to set the current hour.
The minutes flash on the display.
6. SETTING THE MINUTE
Press the MEMORY button to advance by increments of one minute.
Press the Date/Time Setting button to set the current minute.
The year flashes on the display.
Press the START/STOP button to turn the
display off.
11
APPLYING THE ARM CUFF
1. Connect the air tube of the arm cuff and the inflation bulb to
their air jacks.
2. Remove tight-fitting clothing from your upper arm.
3. Sit in a chair with your feet flat on the floor. Place your arm on a
table so the cuff is level with your heart.
12
APPLYING THE ARM CUFF
4. Apply the cuff to your left upper arm so the blue arrow is centered
on the inside of your arm and aligned with your middle finger.
The air tube runs down the inside of your arm. The bottom of the
cuff should be approximately 1/2" above your elbow.
1/2 inch (1 - 2 cm)
Marker
Air Tube
5. Wrap the cuff firmly in place around your arm using the closure strip.
APPLYING THE CUFF ON THE RIGHT ARM
When taking a measurement using the right arm use this instruction
for Step 4.
NOTE: Be careful not to rest your arm on the air tube.
This will restrict the flow of air to the cuff.
13
TAKING A MEASUREMENT
1. Press the ON/OFF button
All display symbols appear on the screen.
NOTE: When the monitor completes the necessary preparations
before measurement, the Heart Symbol ( ) appears
next to the zero. Wait for the Heart Symbol ( )
before taking a measurement.
2. Begin inflating the cuff by rapidly squeezing the inflation bulb until
you have reached approximately 30-40 mmHg higher than your
normal systolic reading. If your normal systolic blood pressure is
unknown, inflate the cuff to 180 mmHg.
3. When the desired pressure has been achieved, release the
inflation bulb.
NOTE: Do not inflate the arm cuff more than necessary. The current
time continues to be displayed while the cuff is being inflated.
14
TAKING A MEASUREMENT
4. Inflation stops and the measurement is started.
As the cuff deflates, decreasing
numbers appear on the display. The
Heart Symbol ( ) flashes at
every heartbeat.
NOTE: If the re-inflation symbol
appears, squeeze the inflation
bulb to re-inflate the arm cuff.
5. When the Deflation Symbol ( ) appears on the display, press the
Air Release Button to release the air left in the cuff.
6. The measurement is complete, the Heart Symbol ( ) stops flashing
and your blood pressure and pulse rate are displayed.
Systolic Blood
Pressure
Diastolic Blood
Pressure
Pulse Display
Press the ON/OFF button to turn the monitor off.
7.
NOTE: The monitor will automatically turn off after five minutes.
15
USING THE MEMORY FUNCTION
Every time you complete a measurement the monitor stores the blood
pressure and pulse rate in the memory. The Blood Pressure Monitor
automatically stores up to 30 measurement values. When 30 sets of
measurement values are stored in the memory, the oldest record is
deleted to save the most recent measurement values.
The monitor also calculates an average reading based on the values of
the three most recent measurements taken.
NOTE:
• If only two readings are stored in the memory for the 10-minute
period, the average is based on the two readings.
• If one reading is stored in the memory for the 10-minute period,
this reading is displayed as the average.
If the date and time is not set, the measurement values will not be stored
in the memory correctly.
TO DISPLAY THE AVERAGE VALUE
Average Value Symbol
1.
Press the ON/OFF button to turn the power on. The average value
symbol will display on the screen with the average reading.
16
USING THE MEMORY FUNCTION
TO DISPLAY THE MEASUREMENT VALUES
The date and time is alternately displayed with the measurement values.
1. Press the MEMORY button while the Heart Symbol ( ) is
displayed. The most recent measurement values will be displayed.
NOTES:
• Press the button repeatedly to display the next values.
• Press and hold the button to display the values faster.
2. Press the ON/OFF button to turn the power off.
TO DELETE ALL VALUES STORED IN THE MEMORY
You cannot partially delete records stored in the memory. All records
will be deleted.
Press and hold the MEMORY button and ON/OFF button
1.
simultaneously for more than 2 seconds.
17
CARE AND MAINTENANCE
To keep your digital blood pressure monitor in the best condition and
protect the unit from damage follow the directions listed below:
Do not forcefully bend the arm cuff or air tube. Do not fold tightly.
Clean the monitor with a soft dry cloth. Do not use any abrasive or
volatile cleaners. Do not attempt to clean the cuff. Never immerse the
monitor or any components in water.
Store the monitor in a safe and dry location. Do not subject the monitor
to extreme hot or cold temperatures, humidity and direct sunlight.
Avoid subjecting the monitor to strong shocks, such as dropping the unit
on the floor.
Remove the batteries if the unit will not be used for three months or
longer. Always replace all the batteries with new ones at the same time.
Use the unit consistent with the instruction provided in this manual.
Use only authorized parts and accessories.
18
ERROR INDICATORS
SYMBOLCAUSECORRECTION
Cuff under-inflated.
Cuff over-inflated.
Batteries are worn.
TROUBLESHOOTING TIPS
PROBLEMCAUSES AND SOLUTIONS
No power.
No display appears on the unit.
Measurement values appear
too high or too low.
Remove the arm cuff. Read,
“Taking a Measurement” on
page 14.
Wait 2-3 minutes. Take another
measurement.
Replace the four batteries. Refer
to page 8 for battery installation.
Replace worn batteries.
Check the battery installation
for proper placement of the
battery polarities.
Blood pressure varies constantly.
Many factors including stress, time
of day, how you wrap the cuff, may
affect your blood pressure.
Review the sections “Before Taking
a Measurement” and “Taking a
Measurement.”
19
FCC STATEMENT
NOTE:
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for U.S.A. only)
This product has been tested and found to comply with the limits for a Class B digital
device, pursuant to part 15 of the FCC rules.
These limits are designed to provide reasonable protection against harmful interference
in a residential installation. The product generates, uses, and can radiate radio frequency
energy and, if not installed and used in accordance with the instructions, may cause
harmful interference to radio communications. However, there is no guarantee that
interference will not occur in a particular installation. If the product does cause harmful
interference to radio or television reception, which can be determined by turning the
product on and off, the user is encouraged to try to correct the interference by one or
more of the following measures:
• Reorient or relocate the receiving antenna
• Increase the separation between the product and the receiver.
• Connect the product into an outlet on a circuit different from that to which the
receiver is connected.
• Consult the dealer or an experienced radio/TV technician for help.
POTENTIAL FOR RADIO/TELEVISION INTERFERENCE (for Canada only)
This digital apparatus does not exceed the Class B limits for radio noise emissions from
digital apparatus as set out in the interference-causing equipment standard entitled
“Digital Apparatus”, ICES-003 of the Canadian Department of Communications.
Cet appareil numérique respecte les limites de bruits radioeléctriques applicables aux
appareils numériques de Clase B prescrites dans la norme sur le materiel brouilleur:
“Appareils Numériques”, ICES-003 édictée par le minister des communications.
Changes or modifications not expressly approved by the party responsible for compliance
could void the user’s authority to operate the equipment.
20
WARRANTY INFORMATION
LIMITED WARRANTY
Your HEM-412CREL ReliOn®Blood Pressure Monitor, excluding the monitor
cuff, is warranted to be free from defects in materials and workmanship appearing
within 5 years from the date of purchase, when used in accordance with the
instructions provided with the monitor. The monitor cuff is warranted to be free
from defects in materials and workmanship appearing within one year from the
date of purchase when the monitor is used in accordance with the instructions
provided with the monitor. The above warranties extend only to the original
retail purchaser.
We will, at our option, repair or replace without charge any monitor or monitor
cuff covered by the above warranties. Repair or replacement is our only
responsibility and your only remedy under the above warranties.
To obtain warranty service contact Omron Healthcare’s Customer Service by calling
1-800-634-4350 for the address of the repair location and the return shipping and
handling fee. Information for warranty service is available on our website at
www.omronhealthcare.com.
Enclose the Proof of Purchase. Include a letter, with your name, address, phone
number, and description of the specific problem. Pack the product carefully to
prevent damage in transit. Because of possible loss in transit, we recommend
insuring the product with return receipt requested.
ALL IMPLIED WARRANTIES, INCLUDING BUT NOT LIMITED TO
THE IMPLIED WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR PARTICULAR PURPOSE, ARE LIMITED TO THE DURATION OF
THE APPLICABLE WRITTEN WARRANTY ABOVE. Some provinces do
not allow limitations on how long an implied warranty lasts, so the above limitation
may not apply to you.
OMRON SHALL NOT BE LIABLE FOR LOSS OF USE OR ANY OTHER
INCIDENTAL, CONSEQUENTIAL OR INDIRECT COSTS, EXPENSES
OR DAMAGES. Some states do not allow the exclusion or limitation of incidental
or consequential damages, so the above exclusions may not apply to you.
This warranty gives you specific legal rights, and you may also have other rights
which may vary from state to state.
Visit our web site at:www.omronhealthcare.com
Call toll free:1-800-634-4350
FOR CUSTOMER SERVICE
SPECIFICATIONS
Model:ReliOn®HEM-412CREL
Display:LCD Digital Display
Measurement Range:Pressure: 0 to 299 mmHg, Pulse: 40-180/min.
Accuracy/Calibration:Pressure: ±3mmHg or 2% of reading
Pulse: ±5% of reading
Inflation:Manual by inflation bulb
Deflation: Automatic pressure release valve
Rapid Air Release:Manual release valve
Pressure Detection:Capacitive pressure sensor
Measurement Method:Oscillometric method
Pulse Wave Detection:Capacitance type pressure sensor
Power Source:4 “AAA” batteries
Battery Life:Approx. 600 uses when used twice a day
with new “AAA” standard batteries
Operating Temperature50ºF to 104ºF (10ºC to 40ºC) /30 to 85% RH
/Humidity:
Storage Temperature-4ºF to 140ºF (-20ºC to 60ºC) /10 to 95% RH
/Humidity:
Main Unit Weight:Approximately 12 3/4 oz. (360 g)
not including batteries
Main Unit Dimensions:Approx. 4 1/8" (w) x 3 1/8" (h) x 5 3/8" (l)
(106 mm x 80 mm x 136 mm)
Cuff Size:Approximately 5 3/4" (w) x 17 1/2" (l)
(146 mm x 446 mm)
Cuff tube 23 5/8" (l) (600 mm)
Cuff Circumference:Fits arm circumferences 8 5/8" to 12 5/8"
(220mm to 320mm)
Contents:Main Unit, Arm Cuff, Inflation Bulb, and
Instruction Manual
UPC Code:681131895361
NOTE: These specifications are subject to change without notice.