Philips IntelliVue MX40 User manual

Page 1
IntelliVue MX40
Instructions for Use
Page 2
Notice
This document contains proprietary information, which is protected by copyright.
Copyright
Copyright © 2011 Koninklijke Philips Electronics N.V. All rights reserved. Reproduction in whole or in part is prohibited without the prior written consent of the copyright holder. Philips Medical Systems Nederland B.V. reserves the right to make changes in specifications and/or to discontinue any products at any time without notice or obligation and will not be liable for any consequences resulting from the use of this publication.
OxiCliq ® and OxiMax ® are registered trademarks of Nellcor Incorporated.
Duracell ® is a registered trademark of Procter & Gamble Incorporated.
STERRAD ® is a registered trademark of Advanced Sterilization Products.
GORE-TEX ® is a registered trademark of W.L. Gore & Assoc. Incorporated
Tone modulation is licensed under US patent 4,653,498 from Nellcor Puritan Bennett Incorporated.
Manufacturer
Philips Medical Systems 3000 Minuteman Road Andover, MA 01810-1099
(978) 687-1501
Printed in USA
Document number
453 564 242 441
Warranty
The information contained in this document is subject to change without notice. Philips Medical Systems makes no warranty of any kind with regard to this material, including, but not limited to, the implied warranties or merchantability and fitness for a particular purpose. Philips Medical Systems shall not be liable for errors contained herein or for incidental or consequential damages in connection with the furnishing, performance, or use of this material.
ii
Page 3
FCC
This device complies with Part 15 and/or Part 95 of the FCC Rules. Operation is subject to the following two conditions: (1) these devices may not cause harmful interference, and (2) these devices must accept any interference received, including interference that may cause undesired operation.
Changes and modifications not expressly approved by Philips Medical Systems can void your authority to operate this equipment under Federal Communications Commission's rules
Printing History
New editions of this document will incorporate all material updated since the previous edition. Update packages may be issued between editions and contain replacement and additional pages to be merged by a revision date at the bottom of the page. Note that pages which are rearranged due to changes on a previous page are not considered revised.
The documentation printing date and part number indicate its current edition. The printing date changes when a new edition is printed. (Minor corrections and updates which are incorporated at reprint do not cause the date to change.) The document part number changes when extensive technical changes are incorporated.
First Edition June 2011
Document Conventions
In this guide:
Warnings
Warning
A Warning alerts you to a potential serious outcome, adverse event or safety hazard. Failure to observe a warning may result in death or serious injury to the user or patient.
iii
Page 4
Cautions
Caution
A Caution alerts you to where special care is necessary for the safe and effective use of the product. Failure to observe a caution may result in minor or moderate personal injury or damage to the product or other property, and possibly in a remote risk of more serious injury.
Notes
A Note contains additional information on the product's usage.
iv
Page 5
Contents
1. Introducing the IntelliVue MX40 1-1
MX40 Features -------------------------------------------------------------------------- 1-2 MX40 Models ---------------------------------------------------------------------------- 1-3 MX40 Compatibility -------------------------------------------------------------------- 1-4
2. Product Safety 2-1
General Safety -------------------------------------------------------------------------- 2-2 Safety Symbols & Other Marks ----------------------------------------------------- 2-4
3. Basic Operation 3-1
Controls, Indicators and Connectors ---------------------------------------------- 3-2
MX40 Controls and Indicators -------------------------------------------------- 3-2
Operating and Navigating ------------------------------------------------------------ 3-8
Power-On Self Test --------------------------------------------------------------- 3-8 Navigating --------------------------------------------------------------------------- 3-8 Selecting Display Elements ----------------------------------------------------- 3-9 Locking the Display---------------------------------------------------------------- 3-9 Measurement Area ---------------------------------------------------------------- 3-9 Measurement Area Display Configurations -------------------------------- 3-10 Connecting/Disconnecting the Patient Cable ----------------------------- 3-10
Understanding Settings -------------------------------------------------------------- 3-12
Changing Measurement Settings --------------------------------------------- 3-12 ECG Settings at the MX40 ----------------------------------------------------- 3-12 Waveform Settings at the MX40 ---------------------------------------------- 3-13
Battery Information -------------------------------------------------------------------- 3-14
Battery Safety Information ------------------------------------------------------ 3-14 Lithium-ion Rechargeable Battery Care ------------------------------------ 3-15 Inserting/Removing Batteries -------------------------------------------------- 3-16 Inserting Batteries ---------------------------------------------------------------- 3-17 Removing the Batteries --------------------------------------------------------- 3-19 Battery Charge Status ----------------------------------------------------------- 3-20
Pouch Use------------------------------------------------------------------------------- 3-22
Securing the Pouch -------------------------------------------------------------- 3-22 Showering -------------------------------------------------------------------------- 3-24
Telemetry Mode Use ----------------------------------------------------------------- 3-26 Monitoring Mode Use----------------------------------------------------------------- 3-27 Briefing the Patient -------------------------------------------------------------------- 3-28
4. Alarms 4-1
Alarms Overview ----------------------------------------------------------------------- 4-2
Visual Alarm Indicators ----------------------------------------------------------- 4-3 Audible Alarm Indicators when in Monitoring Mode ---------------------- 4-5 Acknowledging Alarms ----------------------------------------------------------- 4-6
Contents - 1
Page 6
Pausing or Switching Off Alarms----------------------------------------------- 4-6 Alarm Limits ------------------------------------------------------------------------- 4-7 Reviewing Alarms ------------------------------------------------------------------ 4-8 Latching Alarms -------------------------------------------------------------------- 4-9 Alarm Latching Behavior --------------------------------------------------------- 4-9 Alarm Behavior at Power On ---------------------------------------------------- 4-9
Physiologic Alarms ------------------------------------------------------------------- 4-10 Technical Alarms (INOPs) --------------------------------------------------------- 4-14
5. ECG and Arrhythmia Monitoring 5-1
ECG Safety Information--------------------------------------------------------------- 5-2
For Paced Patients ---------------------------------------------------------------- 5-3
Measuring ECG ------------------------------------------------------------------------- 5-4 Connecting and Positioning ECG Electrodes ----------------------------------- 5-5 Selecting the Primary and Secondary ECG Leads ---------------------------- 5-6 Checking Paced Status --------------------------------------------------------------- 5-7 Understanding the ECG Display ---------------------------------------------------- 5-8 Monitoring Paced Patients ----------------------------------------------------------- 5-9
Avoiding Pace Pulse Repolarization Tails ----------------------------------- 5-9
Changing the Size of the ECG Wave ------------------------------------------- 5-10 Choosing EASI or Standard Lead Placement --------------------------------- 5-11 ECG Configuration ------------------------------------------------------------------- 5-12 ECG Leads Monitored --------------------------------------------------------------- 5-13 Reconstructed Leads ---------------------------------------------------------------- 5-15 3-Wire Placement -------------------------------------------------------------------- 5-16 5-Wire Placement (Standard Mode) --------------------------------------------- 5-17 5-Wire Placement (EASI Mode) -------------------------------------------------- 5-18 6-Wire Placement -------------------------------------------------------------------- 5-19
Selecting Positions of Va and Vb Chest Leads -------------------------- 5-19
Chest Electrode Placement -------------------------------------------------------- 5-20 Monitoring during Leads Off ------------------------------------------------------- 5-21
ECG Fallback --------------------------------------------------------------------- 5-21 Relearning ------------------------------------------------------------------------- 5-21
ST/AR Arrhythmia Monitoring ----------------------------------------------------- 5-23
ST/AR Arrhythmia Algorithm -------------------------------------------------- 5-23 How the ST/AR Algorithm Works -------------------------------------------- 5-23 ECG and Arrhythmia Alarm Overview -------------------------------------- 5-24 Using ECG Alarms --------------------------------------------------------------- 5-25 Learning ---------------------------------------------------------------------------- 5-27 Initiating Arrhythmia Relearning Manually --------------------------------- 5-29
ST/AR ST Analysis Algorithm ----------------------------------------------------- 5-30
Intended Use ---------------------------------------------------------------------- 5-30 The Measurement --------------------------------------------------------------- 5-31 Algorithm Processing ----------------------------------------------------------- 5-31 Displayed ST Data --------------------------------------------------------------- 5-32
Contents - 2
Page 7
EASI ST Analysis ----------------------------------------------------------------- 5-32 Turning ST Monitoring On/Off ------------------------------------------------- 5-32
QT Interval Monitoring --------------------------------------------------------------- 5-33
Intended Use ----------------------------------------------------------------------- 5-34 How the QT Analysis Algorithm Works ------------------------------------- 5-34
6. Monitoring Pulse Rate 6-1
Pulse Rate Measurement ------------------------------------------------------------ 6-2 Displaying the Pulse Rate Measurement at the MX40 ----------------------- 6-3
7. SpO2 Monitoring 7-1
SpO2 Safety Information -------------------------------------------------------------- 7-2
SpO2 Information for the User -------------------------------------------------- 7-4
Pulse Oximetry Measurement ------------------------------------------------------ 7-5
SpO2 Sensors ----------------------------------------------------------------------- 7-6 Selecting an SpO2 Sensor ------------------------------------------------------- 7-6 Sensor Application Safety Information --------------------------------------- 7-7 Applying the Sensor --------------------------------------------------------------- 7-8 Connecting SpO2 Cables -------------------------------------------------------- 7-8 Tone Modulation Indication ----------------------------------------------------- 7-8 Signal Quality Indicator ----------------------------------------------------------- 7-8 Measuring SpO2 -------------------------------------------------------------------- 7-9 Understanding SpO2 Alarms --------------------------------------------------- 7-10
8. Monitoring with other Assigned Devices 8-1
Assigning Devices ---------------------------------------------------------------------- 8-3
Device Assignment at the Information Center ----------------------------- 8-3 Device Assignment at the MX40 ----------------------------------------------- 8-3 Device Assignment at the Patient Monitor ---------------------------------- 8-4
Controls Available when Assigned to IntelliVue Cableless Measurements8-6 Controls Available when Assigned to IntelliVue Patient Monitors --------- 8-7 Networked Device Synchronized Alarm Settings ------------------------------ 8-8 MX40 Display when Wirelessly Connected to a Patient Monitor ---------- 8-9
9. Monitoring with the MX40 at the Information Center 9-1
MX40 Connection to the Information Center ------------------------------------ 9-2 MX40 Controls in the Patient Window -------------------------------------------- 9-3 Locating the MX40 (Find Device) -------------------------------------------------- 9-5 Viewing Device Location and Location History (optional) ------------------- 9-6 Using the Device Location Client (optional) ------------------------------------- 9-7 Patient Configurable Settings in Telemetry Setup ----------------------------- 9-8 Unit Configurable Settings ---------------------------------------------------------- 9-11
10. Operating with Information Center Release L or M 10-1
Display ----------------------------------------------------------------------------------- 10-2 Alarms ------------------------------------------------------------------------------------ 10-3
Contents - 3
Page 8
11. Trends (Optional) 11-1
Viewing Vital Trend Information -------------------------------------------------- 11-2
12. Maintenance 12-1
Cleaning --------------------------------------------------------------------------------- 12-2
Cleaning Materials for the MX40 --------------------------------------------- 12-2
Disposing of the MX40 -------------------------------------------------------------- 12-4 Label Assignment for Replacement MX40 ------------------------------------- 12-5
Re-assigning an Equipment Label ------------------------------------------ 12-5
Charging Lithium-ion Rechargeable Batteries -------------------------------- 12-7
Battery Power Indicators ------------------------------------------------------- 12-7 Battery Lifetime Management ------------------------------------------------ 12-8 Battery Disposal ------------------------------------------------------------------ 12-9
13. Safety Standards & Specifications 13-1
Regulatory Information -------------------------------------------------------------- 13-2
Software Hazard Prevention -------------------------------------------------- 13-2 AC Power Source ---------------------------------------------------------------- 13-2 Industrie Canada Compliance (Canada) ----------------------------------- 13-2 Safety Standards ----------------------------------------------------------------- 13-2 Intended Use Statement ------------------------------------------------------- 13-3 Indications for Use --------------------------------------------------------------- 13-3 Intended Uses of MX40 -------------------------------------------------------- 13-4 Authorized EU Representative ----------------------------------------------- 13-4 Patient Population --------------------------------------------------------------- 13-4 Rx ------------------------------------------------------------------------------------ 13-4 Essential Performance --------------------------------------------------------- 13-5
Electromagnetic Compatibility ----------------------------------------------------- 13-6
Reducing Electromagnetic Interference ------------------------------------ 13-7 Restrictions for Use-------------------------------------------------------------- 13-7 Electromagnetic Compatibility (EMC) Specifications ------------------- 13-7 Electromagnetic Emissions ---------------------------------------------------- 13-8 Electromagnetic Immunity ----------------------------------------------------- 13-8 Recommended Separation Distance --------------------------------------- 13-9
Battery Specifications -------------------------------------------------------------- 13-12 Lithium-ion Battery Charge Time ------------------------------------------------ 13-15 Physical Specifications ------------------------------------------------------------- 13-16 MX40 1.4 GHz Radio --------------------------------------------------------------- 13-17 MX40 2.4 GHz Radio --------------------------------------------------------------- 13-18 MX40 Short-Range Radio --------------------------------------------------------- 13-20 Environmental Specifications ----------------------------------------------------- 13-21 Measurement Specifications ----------------------------------------------------- 13-22
ECG -------------------------------------------------------------------------------- 13-22 ECG Performance Disclosure/Specifications ---------------------------- 13-23 FAST SpO2 ----------------------------------------------------------------------- 13-25 SpO2 Sensor Accuracy -------------------------------------------------------- 13-27
Contents - 4
Page 9
A. Accessories A-1
MX40 Accessories --------------------------------------------------------------------- A-2
Pouches ------------------------------------------------------------------------------ A-2 Miscellaneous ----------------------------------------------------------------------- A-2
ECG Accessories ----------------------------------------------------------------------- A-3
Electrodes --------------------------------------------------------------------------- A-3 Leadsets and Patient Cables --------------------------------------------------- A-4
SpO2 Accessories ---------------------------------------------------------------------- A-5
Philips/Nellcor Disposable Sensors ------------------------------------------- A-5 Philips Reusable Sensors ------------------------------------------------------- A-6 Adapter Cables --------------------------------------------------------------------- A-6
B. Default Settings B-1
Alarm Default Settings ---------------------------------------------------------------- B-2 ECG, Arrhythmia, ST and QT Default Settings --------------------------------- B-3 Configuration Default Settings at the MX40 ------------------------------------- B-5
C. Sales and Support Offices C-1
Contents - 5
Page 10
Contents - 6
Page 11
MX40 Features ........................................................................................... 1-2
MX40 Models ............................................................................................. 1-3
MX40 Compatibility .................................................................................. 1-4
1. Introducing the IntelliVue MX40
This section introduces the IntelliVue MX40 wearable patient monitor.
Introducing the IntelliVue MX40 1-1
Page 12
MX40 Features
Easy for clinicians to use and comfortable for patients to wear.
2.8" color, touch sensitive display. Smart, multi-measurement cable system available for use with reusable
and single-patient use supplies.
FAST SpO2 (continuous, or manual measurement). EASI or standard ECG selectable in one device. 6-lead with two V-leads for diagnosing multiple cardiac abnormalities,
including wide-QRS complex tachycardias and acute myocardial ischemia/infarction.
Local measurement trend/alarm history. Local alarming for measurements (requires IntelliVue Information
Center Release N or later).
Integrated Smart-hopping radio. Integrated Short-Range Radio (SRR). Communication with IntelliVue Patient Monitors and Cableless
Measurements via Short-Range Radio connection (MP5/MP5T/MP5SC, MP2 and X2 monitors only).
Powered by three AA batteries or rechargeable lithium-ion battery
pack.
Audio feedback for out-of-range and lost device. Battery gauge on device and at Information Center. Alarm suspend and resume from standby at device and Information
Center.
Pouch with clear front that closes securely.
Note Unlike a traditional bedside monitor which operates on AC power,
the MX40 is powered by battery and provides time-limited screen display and local alarming.
1-2 Introducing the IntelliVue MX40
Page 13
MX40 Models
The MX40 is available in three models (ECG only, ECG and FAST SpO2, or ECG and SpO2 Ready (for future upgrade).
Introducing the IntelliVue MX40 1-3
Page 14
MX40 Compatibility
The MX40 is compatible for use with IntelliVue Information Center Release N. Limited compatibility is offered when used with IntelliVue Information Center Release L or M. See the "Operating with Release L or M" chapter for more information.
The MX40 is compatible for use with IntelliVue Patient Monitors Release G or later when wirelessly connected.
The MX40 is compatible for use with IntelliVue Cableless Measurements Release A.1.
The MX40 is compatible for use with Access Point Controller 862147, Release B.00.19 and Access Point Controller 865346, Release C.00.XX.
The MX40 Patient Cable is compatible for use with IntelliVue Patient Monitor platforms MP2/X2, MP5/MP5T/MP5SC, MP20/30 with MMS or X2, MP40/50 with MMS or X2, MP60/70 with MMS or X2, MP80/90 with MMS or X2, and MX800/700/600 with MMS or X2.
1-4 Introducing the IntelliVue MX40
Page 15
General Safety ............................................................................................ 2-2
Safety Symbols & Other Marks ............................................................... 2-4
2. Product Safety
This section consolidates the general safety warnings associated with the IntelliVue MX40. These warnings are repeated throughout the book in context where relevant.
Safety symbols and other markings on the MX40 are also described here.
Product Safety 2-1
Page 16
General Safety
Warnings
The IntelliVue MX40 should not be used for primary monitoring in
applications where the momentary loss of the ECG is unacceptable at the Information Center.
For continued safe use of this equipment, it is necessary that the listed
instructions are followed. Instructions in this manual in no way supersede established medical procedures.
Do not touch the patient, or table, or instruments, during defibrillation.
The battery door must be closed during defibrillation. These steps protect the clinician from high defibrillator voltage.
This device is not to be used in the vicinity of electrosurgical units
because such use may interrupt or interfere with the transmission of signals from the MX40.
This equipment is not suitable for use in the presence of a flammable
anesthetic mixture with air, or with oxygen or nitrous oxide.
This equipment is not suitable for use in an MRI environment. Do not use patient cables with detachable lead wires that have exposed
male pins. Electrocution could result if these pins are plugged into AC power.
Do not use patient cables or accessory cables and sensors if prior visual
inspection reveals cable damage or the presence of liquid, lint or dust inside.
The system is not completely immune from radio interference although
it is designed to minimize interference. Sources of interference that may be a problem include failing fluorescent lights and construction equipment. See "Electromagnetic Compatibility p. 13-6". The product should not be used next to or stacked with other equipment. If you must stack the product, you must check that normal operation is possible in the necessary configuration before the product is used on patients.
Do not rely exclusively on the audible alarm system for patient
monitoring. Adjustment of alarm volume to a low level during patient monitoring may result in patient danger. Remember that the most reliable method of patient monitoring combines close personal surveillance with correct operation of monitoring equipment.
2-2 Product Safety
Page 17
If the MX40 enters a continuous "boot-up" cycle or the main display
does not appear or update, ensure that you are using a freshly charged lithium-ion battery or new disposable batteries. If the batteries are fresh and the device reboots or does not update, remove the device from service and contact your service personnel.
Place the MX40 in a pouch or over clothing, or both, during patient use.
The device should not touch the patient’s skin during use.
Patients should be instructed not to open the battery compartment
while the MX40 is in use.
Failure on the part of the responsible individual hospital or institution
employing the use of this equipment to implement satisfactory maintenance as needed may cause undue equipment failure and possible health hazards.
Because the coverage range of Access Points can sometimes overlap,
including different floor levels, the IntelliVue Device Location feature is not intended for use when attempting to locate a patient.
Short-range radio connections are subject to interruption due to
interference from other radio sources in the vicinity, including microwaves, bluetooth devices, and DECT phones. Outside the frequency band and 5% above and below, i.e. the exclusion band according to IEC 60601-1-2, the short-range radio connection is immune up to 3V/m in the frequency range from 80MHz to 2.5 GHz. Depending on the strength and duration of the interference, the interruption may occur for an extended period. Any interruption of the signal due to interference, moving out of range, or for other reasons is indicated with a Tele Disconnected INOP message on the IntelliVue Patient Monitor.
Caution
Philips recommends that when using a pouch to attach the MX40 to your patient that you consider your patient's condition and are careful about placement of the straps as the straps could present a strangulation hazard.
Product Safety 2-3
Page 18
Label
Definition
FCC ID: IC:
Federal Communications Commission (FCC) ID
Canadian ID
Federal Communications Commission (FCC) Grant of Equipment Authorization
CE Mark (MX40)
Compliance to Council Directive 93/42/EEC (Medical Device Directive) Class 2 Radio Equipment Identifier (1999/5/EC)
CE Mark (Rechargeable Lithium-ion Battery)
Compliance to Council Directive 2004/108/EC (EMC Directive)
Non-Ionizing Radiation Interference to electronic equipment
may occur in the vicinity of devices marked with this symbol.
Disposal Dispose of in accordance with the
local country’s requirements.
Follow operating instructions.
Safety Symbols & Other Marks
The table below describes the safety symbols and other markings present on the MX40 and the lithium-ion battery.
2-4 Product Safety
Page 19
Label
Definition
Prescription Device
Canadian and American standards compliance
Complies with applicable Canadian and American safety standards.
Defibrillation Proof Patient connections are protected
against defibrillation (DEFIBRILLATION-PROOF) and are a TYPE CF APPLIED PART.
Product Number
Serial Number
Used to identify the equipment during a call to the Philips Healthcare (Service).
MAC Address
Date of Manufacture
Battery Polarity
IPX Waterproof Rating
2D Barcode Underwriter's Laboratories Listed
Component
Service Identification Number
Used to identify the equipment during a call to Philips Healthcare (Service).
Product Safety 2-5
Page 20
Label
Definition
Attention! See Instructions for Use.
2-6 Product Safety
Page 21
Controls, Indicators and Connectors ...................................................... 3-2
Operating and Navigating ....................................................................... 3-8
Understanding Settings .......................................................................... 3-12
Battery Information ................................................................................. 3-14
Pouch Use ................................................................................................. 3-22
Telemetry Mode Use ............................................................................... 3-26
Monitoring Mode Use ............................................................................. 3-27
Briefing the Patient .................................................................................. 3-28
3. Basic Operation
This section gives you an overview of the IntelliVue MX40 and its functions. It tells you how to perform tasks that are common to all measurements, such as turning a measurement on and off, adjusting wave size and information in preparation for use.
Familiarize yourself with all instructions including warnings and cautions before starting to monitor patients. Read and keep the Instructions for Use that come with any accessories as these contain additional important information.
Basic Operation 3-1
Page 22
1. Patient Cable
2. Patient Information Area
3. Active Alarms Area
4. INOP Area
5. Measurement Area 1
6. Measurement Area 2
7. Waveform 1
8. Waveform 2
9. Radio/Network/Battery Status Area
10. Leads Off Status Area
11. Silence Alarms Button
12. SmartKeys Button
13. Main Screen Button
14. Multi-Function Button
Controls, Indicators and Connectors
This section describes the clinical controls of the IntelliVue MX40. These controls include buttons, display icons, visual and auditory indicators, ports, and safety labeling located on the front and back of the device.
MX40 Controls and Indicators
3-2 Basic Operation
Page 23
Button
Function
Depending on configuration at the Information Center:
generates a Nurse Call; Initiates a Delayed Recording; Both, or; None
Note the Multi-Function Button does not operate
when paired with an IntelliVue Patient Monitor via the short-range radio connection.
Button
Function
Initiates a local silence/acknowledgment of
all active alarms when enabled.
Silences the "Find Device" sound.
Note Alarms at the MX40 can be silenced
from the Information Center.
Button
Function
Displays the SmartKey Menu on the touch screen.
Multi-Function Button
Silence Alarm Button
SmartKeys Button
Basic Operation 3-3
Page 24
Button
Function
Activates the Touch Display if touched for two
seconds.
Cycles through the display screens if touched
repeatedly.
Resumes from Standby.
SmartKey
Function
Start SpO2
Note This
SmartKey is unavailable when SpO2 mode is continuous.
Starts a manual SpO2 measurement.
Delay Record
Starts a delayed recording at the Information Center.
Alarms
Review of up to 50 previous alarm conditions (entries are stored during power cycle). Pause Alarms for configured time period (if enabled at the Information Center).
Mode: Telemetry / Mode: Monitor
Toggles between modes. In Telemetry Mode, display and audio are off; in Monitor Mode, display and audio are always on.
Main Screen Button
SmartKeys
The following table lists the SmartKeys available on the display of the MX40.
Note—gray text on a SmartKey signifies that the item is unavailable.
3-4 Basic Operation
Page 25
SmartKey
Function
Standby
Puts the device into standby locally and at the Information Center. Displays purchased/enabled product options.
Add/Remove
Displays available monitors and IntelliVue Cableless Measurements to assign to via the short-range radio.
Print Reports
Prints the pre-configured report as designated at the Information Center.
Vitals Trends (Optional)
View up to 24 hours of tabular trend data.
Screen Setup
Determines time period that the display remains active after user interaction.
Lock/Unlock
Locks/Unlocks the display.
Op Mode
Selects either Monitoring, Demo, Config or Service modes.
The Alarm Area of the MX40 displays
physiological alarms and technical alarms.
A multiple alarm indicator (down arrow) is
displayed when multiple alarm conditions are present.
A check mark in front of the alarm text
signifies that the alarm has been acknowledged by touching the Silence Alarms button.
Alarm Indicators display in the Patient
Information Area in place of the time clock when alarm/INOP conditions are present but have not been acknowledged.
Touching the Alarms Area displays a list of
all active alarms.
The alarms paused icon communicates
whether the alarm system is on/off.
Local Alarm Audio is off when the alarm
volume symbol is present.
Alarms Area
Basic Operation 3-5
Page 26
The Patient Information Area displays the following information:
Bed Label Patient Name (up to 15 characters will display) Time
Touching the Patient Information Area displays the Patient Demogr. menu which lists the following:
Patient Name (Last, First, Middle) Lifetime ID Encounter ID Patient Category Paced Mode Height Weight Date of Birth Gender
Note If you use an alternative ID, it will display at the Information Center and on
printed reports. It will not display at the MX40.
1. Pacing algorithm is on.
2. Pacing algorithm is off.
Patient Information Area
Paced Status
3-6 Basic Operation
Page 27
The Lock symbol appears in the lower left of the display when the MX40 is in a locked state after five minutes of non-use. Locking the display provides additional protection against accidental patient access. The display is unlocked using the SmartKeys menu.
The status area of the MX40 displays short-range radio connection (optional) and system wireless connection status. You can also view battery strength for the type of battery used in the device, AA or rechargeable Li-on.
Display Lock
Status Area
Basic Operation 3-7
Page 28
Operating and Navigating
The principle method of operating your MX40 is via the Touch Display. Almost every element on the display is interactive. Display elements include measurement numerics, information fields, alarm fields, waveforms, SmartKeys and menus.
Power-On Self Test
Once battery power is supplied, the MX40 performs a power-on self test to check operational status prior to start-up. Should a failure be detected, an INOP tone will sound and if possible, the appropriate INOP message for the failure will be communicated to the Information Center and displayed locally.
A successful power-on self test will then transition the MX40 to the start-up screen. Selectable background colors can be configured and display on the screen for assistance with device identification. This can be helpful when devices are in a pooled use setting.
If the MX40 enters a continuous "boot-up" cycle or the main display does not appear or update, ensure that you are using a freshly charged lithium-ion battery or new disposable batteries. If the batteries are fresh and the device reboots or does not update, remove the device from service and contact your service personnel.
You must visually check that a waveform is present on the display. You can access further status information is by touching the status area on the display.
Navigating
Touching the Navigation Bar on the right of the display will scroll through additional display items. Solid downward arrows indicate there are additional elements that are not currently displayed. The arrows briefly illuminate when touched. Your selection from the menu also illuminates when touched.
3-8 Basic Operation
Page 29
Function
Display Locked/Active
Display Locked/Inactive
Display Unlocked/Active
Display Unlocked/Inactive
Display Touch
No
No
Yes
No
Main Screen Button
No
Yes
Yes
Yes SmartKeys Button
Yes
No
Yes
No
Silence Button
No
No
Yes
No
Selecting Display Elements
Touch a display element to get to the actions linked to that element. For example, touch the Patient Information element to call up the Patient Info window, or touch the HR numeric to call up the Setup ECG menu. Touch the ECG waveform to call up the wave selection menu.
Locking the Display
To provide additional protection against accidental patient access to the
MX40, the display can be locked using the Lock SmartKey. When Lock is selected, the SmartKey menu automatically changes to the Main Screen. When Unlock is selected, you must close the SmartKey menu to return to the Main Screen.
The display automatically locks when there is no interaction for five minutes.
Measurement Area
The measurement area of the MX40 display is optimized to show available parameter numerics, waveforms, and alarm limits. Each element is a touch object and when you select it, further controls and menus become available.
Basic Operation 3-9
Page 30
Measurement Area Display Configurations
The display of your MX40 is configured/can operate in one of four available orientations:
Portrait - One Waveform and four Numerics Portrait - Two Waveforms and two Numerics (IIC Release N only) Landscape - Two Waveforms and three Numerics (IIC Release N only) Portrait - Viewable Chest Diagram and two Numerics
Connecting/Disconnecting the Patient Cable
The patient cable is connected to the MX40 as shown in the illustration below.
When connecting to the MX40, there is a slight clicking sound that signifies that the cable is securely connected.
3-10 Basic Operation
Page 31
Disconnect the patient cable as shown below.
Caution
Never disconnect the patient cable by pulling on the leadwires, as this may damage wires over time.
Basic Operation 3-11
Page 32
Setting
Description
Alarm Limits
Heart Rate alarm limits can be viewed locally at the MX40. Limits set at the Information Center (Release N or later) are reflected at the MX40 when connected on the network.
Primary (used for arrhythmia analysis only)
I,II, III, aVR, aVL, aVF, V1-V9, MCL, V3R, V4R, V5R. Available waveforms are based on lead set type. Lead II is the default.
Secondary (used for arrhythmia analysis only)
I,II, III, aVR, aVL, aVF, V1-V9, MCL, V3R, V4R, V5R. Available waveforms are based on lead set type. Lead V is the default.
Paced Mode
Yes, No
Adjust Size
Set ECG gain to x1/2, x1, x2, x4
Arrhythmia
Initiate an Arrhythmia Relearn; View Arrhythmia Alarm Limits; Turn Arrhythmia Annotation On/Off.
Lead Placement
Set EASI, Standard
Understanding Settings
Each aspect of how the MX40 works and looks is defined by a setting. There are a number of different categories of settings, including:
Screen Settings - to define the selection and appearance of elements on
each individual display screen.
Measurement Settings - to define setting unique to each measurement,
e.g. high and low alarm limits.
Monitor Settings -including settings that affect more than one
measurement or display screen, for example alarm volume and alarm pause time.
You must be aware that, although many settings can be changed during use, permanent changes to settings can only be done in Configuration Mode. All settings are restored to their default setting when the patient is discharged or the MX40 is powered off.
Changing Measurement Settings
Each measurement has a setup menu in which you can adjust its settings. You enter the setup menu by selecting the measurement numeric.
ECG Settings at the MX40
3-12 Basic Operation
Page 33
Setting
Description
ECG
Set ECG On/Off
New Lead Setup
When IntelliVue Patient Monitor lead sets are in use, select 3-wire, or 5-wire.
Va Lead
Shows position of Va, or C1, electrodes. Choices are V1-V9, v3R, V4R, V5R.
Vb Lead
Shows position of Vb,or C2, electrodes. Choices are V1-V9, v3R, V4R, V5R.
Change Numeric
Selects parameter numeric to display in place of current HR numeric.
Setting
Description
Wave 1
Primary, Secondary, I, II, III, aVR, aVL, aVF, V1-V9, MCL, V3R, V4R, V5R. Available waveforms are based on patient cable type. Lead II is the default. If Primary or Secondary are selected, then the waveform displayed is the waveform configured as primary or secondary for arrhythmia analysis.
Wave 2
Primary, Secondary, I,II, III, aVR, aVL, aVF, V1-V9, MCL, V3R, V4R, V5R, Pleth (if SpO2 is available). Available waveforms are based on patient cable type. Lead V is the default.If Primary or Secondary are selected, then the waveform displayed is the waveform configured as primary or secondary for arrhythmia analysis.
Waveform Settings at the MX40
Primary or secondary waveform configuration changes made at the Information Center change the MX40.
Basic Operation 3-13
Page 34
Battery Information
Battery Safety Information
Warnings
The battery compartment door must be closed during defibrillation. Use the Philips Rechargeable Lithium-ion Battery or 3 Duracell Alkaline
batteries, size AA, MN 1500, 1.5V, to ensure specified performance and correct battery gauge reporting. Outdated, mismatched, or poor-quality batteries can give unacceptable performance (e.g., insufficient Battery-Low warning time). If you are using disposable batteries, the use of fresh high-quality alkaline batteries is strongly recommended.
Certain failure conditions, such as short circuits, can cause a battery to
overheat during use. High temperatures can cause burns to the patient and/or user. If the MX40 becomes hot to the touch, remove it from the patient and place it aside until it cools. Then remove the batteries and discard them. Have the MX40 checked by your service provider to identify the cause of overheating.
If you receive a TELE BATTERY LOW, TELE BATTERY EMPTY,
REPLACE BATTERY T, or TELE BATTERY TEMP alarm, the batteries must be promptly replaced. If these conditions are not corrected, they will result in a device shutdown and cessation of monitoring.
Disposable batteries should be removed from the MX40 at the end of
the battery’s useful life to prevent leakage.
If battery leakage should occur, use caution in removing the battery. The leaked substance may cause eye or skin irritation. Avoid contact with skin. Clean the battery compartment according to the instructions in the Maintenance section. Wash hands.
To eliminate the risk of electrical shock or burn, do not carry loose
batteries on your person, e.g. in clothing pockets.
Caution
Use of AA Lithium batteries or batteries with terminal voltage >1.6V may cause damage to the device.
3-14 Basic Operation
Page 35
Activity
When to Perform
Perform a visual inspection.
Before inserting a battery in the MX40.
Charge the battery.
Upon receipt, after use, or if a low battery state is indicated. To optimize performance, a fully (or almost fully) discharged battery should be charged as soon as possible.
Clean the battery
At each patient discharge, or in cases when the battery is exposed to contaminants.
Charge stored batteries to at least 40% of their capacity every six months.
When not in use for an extended period of time.
Decommission the battery
When any of the following INOPs are displayed on the MX40:
TELE SERVICE BATTERY TELE BATTERY TEMP
Lithium-ion Rechargeable Battery Care
Care of the rechargeable battery begins when you receive a new battery for use and continues throughout the life of the battery. The table below lists battery care activities and when they should be performed.
Rechargeable batteries are charged using the IntelliVue CL Charging Station. For information on charging station use, see Charging Li-ion Rechargeable Batteries p. 12-7 .
Note — The battery capacity of re-chargeable batteries degrades over time
and number of recharge cycles. Toward the end of its useful life, the battery capacity may be reduced by 25-30%. If this reduced battery life is unacceptable based on your use model, Philips recommends replacing the rechargeable battery sooner.
Lithium-ion Rechargeable Battery Handling Precautions
Lithium-ion batteries store a large amount of energy in a small package. Use caution when handling the batteries; misuse or abuse could cause bodily injury and/or equipment damage.
Do not short circuit - take care that the terminals do not contact metal
(e.g. coins) or other conductive materials during transport and storage.
Do not crush, drop or puncture - mechanical abuse can lead to internal
damage and internal short circuits that may not be visible externally.
Basic Operation 3-15
Page 36
Do not apply reverse polarity. Do not incinerate batteries or expose them to temperatures above 60
o
C
(140oF).
If a battery has been dropped or banged against a hard surface, whether damage is visible externally or not:
discontinue use. dispose of the battery in accordance with the disposal instructions.
Lithium-ion Rechargeable Battery Storage
When storing rechargeable batteries, make sure that the battery terminals do not come into contact with metallic objects or other conductive materials.
If batteries are stored for an extended period of time, they should be stored in a cool, dry place, ideally at 15oC (60oF), with a state of charge of 20% to 40%. Storing batteries in a cool place slows the aging process.
The batteries should not be stored at a temperature outside the range of
-20oC (-4oF) to 50oC (122oF).
Stored batteries should be should be charged to at least 40% of their capacity every 6 months.". They should be charged to full capacity prior to use.
Note — Storing batteries at temperatures above 38
periods of time could significantly reduce the batteries' life expectancy.
Inserting/Removing Batteries
Warning
Arrhythmia relearning is initiated whenever the MX40 is powered down
for one minute or longer. Be sure to check your patient’s arrhythmia
annotation for accuracy whenever relearn has occurred.
Caution
o
C (100oF) for extended
Remove the batteries before storing the MX40 for an extended period of time.
3-16 Basic Operation
Page 37
The battery compartment is located on the back of the MX40, accessible by opening the compartment door from the bottom. It accommodates three AA 1.5V Alkaline batteries or the Philips Rechargeable Lithium-ion battery. Only these batteries should be used.
Note Lithium-ion batteries should be fully charged prior to first use. Important Do not use other rechargeable batteries. Use of this type of
battery will adversely affect:
Battery gauge performance Battery low warnings Battery life performance
Inserting Batteries
Insert the rechargeable lithium-ion battery using the
following procedure:
Open the battery compartment by lifting up on both bottom sides of the compartment door.
1 Remove the AA battery tray if present. 2 Insert the battery pack so that the raised tab is aligned with the cutout
in the base of the battery compartment. Close the battery compartment door.
Basic Operation 3-17
Page 38
3 Close the battery compartment door. 4 Watch for the start-up screen on the front of the MX40 to illuminate
briefly.
Insert AA batteries into the MX40 using the following
procedure:
1 Open the battery compartment by lifting up on both bottom sides of the
compartment door.
2 Insert the AA battery tray if not already present. 3 Insert three AA 1.5V Alkaline batteries, matching the polarity with the
+indications inside the compartment.
Note—all batteries are inserted with the + polarity in the same direction.
3-18 Basic Operation
Page 39
4 Close the battery compartment door. 5 Watch for the start-up screen on the front of the MX40 to illuminate
briefly.
Removing the Batteries
Batteries should be removed when the MX40 is not in use or is being stored.
To remove the batteries, open the battery compartment door and push from the opening at the bottom of the compartment to pop the batteries out. Device settings (patient cable type, SpO2 mode, volume, etc.) are retained when the batteries are removed.
If you remove good AA batteries to turn off the MX40, keep them together as a set for later re-use so that all batteries will have the same level of power remaining.
Important— Do not "store" disposable AA batteries by leaving them in the
incorrect polarity position in the MX40.
Be careful not to short circuit the batteries. Batteries can get hot when shorted. Short circuits are caused when a piece of metal touches both the positive and negative terminals simultaneously. More than a momentary short circuit will generally reduce the battery life. In case of a short circuit, discard the batteries, or just the shorted one if the batteries are new.
Disposal of Batteries
When disposing of batteries, follow local laws for proper disposal. Dispose of batteries in approved containers. If local regulations require you to recycle batteries, recycle batteries in accordance with those regulations.
Basic Operation 3-19
Page 40
Approximate Battery Life Remaining
Approximate Time Remaining (ECG only)
Approximate Time Remaining (ECG & Spo2 Continuous)
Functionality Disabled
Battery Indicator LCD Segments
100%
~ 25 hours
~ 14 hours
None
5 Green
75%
< 19 hours
< 10.5 hours
None
4 Green
50%
< 13 hours
< 7 hours
None
3 Green
25%
< 6 hours
< 3.5 hours
None
2 Green
10%
< 3 hours
< 1.5 hours
None
1 Green
Low battery level to replace/charge battery level
< 30 minutes
< 30 minutes
SpO2 and short-range radio are disabled. Display is at half brightness
1 Red Red Battery Icon Audio Replace/charge battery level
< 10 minutes
< 10 minutes
Device shutdown
1 Red Red Battery Icon
Battery Charge Status
The battery charge indicator displays in the Status Area and communicates the remaining battery charge time when using both AA batteries or the rechargeable lithium-ion battery.
When the MX40 is initially powered-on, it takes approximately 25 seconds for the indicator to populate. During this time, the indicator displays a ? in the battery icon.
In order to guarantee overall device performance, certain functionality is disabled when the battery charge reaches critical levels. See the tables below for additional information about battery status.
Lithium-ion Rechargeable Battery Charge Status
3-20 Basic Operation
Page 41
Approximate Battery Life Remaining
Approximate Time Remaining (ECG only)
Approximate Time Remaining (ECG & Spo2 Continuous)
Functionality Disabled
Battery Indicator LCD Segments
100%
~ 24 hours
~ 9 hours
None
5 Green
75%
< 18 hours
< 7 hours
None
4 Green
50%
< 12 hours
< 5 hours
None
3 Green
25%
< 6 hours
< 2 hours
None
2 Green
10%
< 2 hours
< 1 hours
None
1 Green
Low battery level to replace/charge battery level
< 30 minutes
< 30 minutes
SpO2 and short-range radio are disabled. Display is at half brightness.
1 Red Red Battery Icon Audio Replace/charge battery level
< 10 minutes
< 10 minutes
Device shutdown
1 Red Red Battery Icon
AA Battery Charge Status
Basic Operation 3-21
Page 42
Pouch Use
The MX40 is not intended for direct contact with the patient’s skin. During normal use, the MX40 should be worn over clothing, in a pocket or, preferably, in a pouch. The Waterproof Carry Pouch with clear front is an appropriate means for holding the MX40. See Appendix A, "Accessories" for ordering information.
Securing the Pouch
1 See the Carry Pouch, Waterproof, Instructions for Use, P/N 453564267571,
for more information.
2 Insert the MX40 into the pouch with lead wires and SpO2 sensor cable, if
used, exiting from the side opening of the pouch. Pinch the velcro enclosures together to close the pouch around the cables.
3-22 Basic Operation
Page 43
3 Seal the pouch.
4 Secure the pouch on the patient with the ties around the patient’s
shoulder and under the arm.
5 Check that the patient is comfortable wearing the pouch with the MX40.
Caution
Philips recommends that when using a pouch to attach the MX40 to your patient, consider your patient's condition and be careful about placement of the straps, as the straps could present a strangulation hazard.
Basic Operation 3-23
Page 44
Showering
Warning
When the patient is showering, signal quality and leads off detection may be compromised due to significant movement. Appropriate clinical precautions must be taken.
Caution
Because the touchscreen display is sensitive to water impact, the display should be locked when showering.
The MX40 can be used to monitor a patient in the shower, but only when placed inside a Philips carrying pouch and secured on the patient as described above. The combination of the MX40 and pouch will withstand showering for up to 10 minutes.
Drying the MX40 after Showering
After showering, perform the following steps to continue monitoring:
1 Remove the battery. 2 Pat dry the patient cable connections at the electrodes. 3 Wipe the lead wires with care. 4 If wet, dry the outside of the MX40 with a non-lint producing cloth. 5 If wet, wipe dry the inside of the battery compartment. Dry the
batteries.
6 If wet, disconnect the patient cable and shake out any water. Dry the
connector pin area with a cotton swab.
7 Re-insert the battery.
Caution
The MX40 should not be used for monitoring if the battery compartment is wet. Remove the batteries and wipe the compartment dry before continued monitoring use.
3-24 Basic Operation
Page 45
Accidental Liquid Exposure
If the MX40 is accidentally immersed in liquid, no damage to the device and no electrical safety issues for the patient will result. Remove the device, dry it off, and follow the procedure for cleaning/sterilization under "Cleaning and Sterilization" as needed.
Basic Operation 3-25
Page 46
Telemetry Mode Use
To minimize patient disruption, the MX40 operates in Telemetry Mode when connected to the Information Center. In Telemetry Mode, the local volume is set to zero and the display is off. You can activate the display at any time by touching the Main Screen button for two seconds. All active alarms can be viewed when the display is on, however audible alarm indicators are not annunciated. Regardless of the display status, all measurement data is being sent to the Information Center. Telemetry Mode is only available when connected to the Information Center.
3-26 Basic Operation
Page 47
Monitoring Mode Use
You may find the use of Monitoring Mode helpful when spending extended time directly with your patient, e.g. during transport, showering, dressing change. The display is always on for easy viewing and should an alarm condition occur, it will be announced locally at the MX40 and at the Information Center if networked connected. If the MX40 is not network connected, the alarm is only announced locally.
To use Monitor Mode:
1 Press the SmartKeys Button. 2 Press the Mode: Telemetry / Mode: Monitor SmartKey and choose
Mode: Monitor.
Basic Operation 3-27
Page 48
Briefing the Patient
Warning
Patients should be instructed not to interact with the with display of the device and to not open the battery compartment while the MX40 is in use.
Note Pausing alarms at the Information Center activates the MX40
display. Patients should be notified that this is normal operation and not cause for any concern.
If the Multi-Function button has been configured to generate a Nurse Call alarm, recording at the Information Center, or both, instruct the patient to use the button when needed.
If desired, you can turn off patient use of the Multi-Function button at the Information Center. For more information see Patient Configurable Settings in Telemetry Setup p. 9-8.
3-28 Basic Operation
Page 49
Basic Operation 3-29
Page 50
Page 51
Alarms Overview ...................................................................................... 4-2
Physiologic Alarms ................................................................................. 4-10
Technical Alarms (INOPs) ..................................................................... 4-14
4. Alarms
The section provides alarm information that applies to all measurements. Measurement-specific alarm information is discussed in the sections on individual measurements.
Alarms 4-1
Page 52
Alarms Overview
The MX40 has two different types of alarms: physiological alarms and INOPs. For MX40 devices operating with IntelliVue Information Center Release L and M, physiological alarms are not available locally on the MX40. INOPs are displayed as described here.
For MX40 devices operating with IntelliVue Information Center Release N, physiological alarms are available locally on the MX40 regardless of network connection to the Information Center. Alarm settings are as configured by the Information Center. Changes to physiological alarm settings can only be made at the Information Center. Audible alarm indicators are annunciated only when operating in Monitor Mode or when the MX40 is not networked connected.
Physiological Alarms
Physiological alarms are red and yellow alarms. A red alarm indicates a high priority patient alarm such as a potentially life threatening situation (for example, asystole). A yellow alarm indicates a lower priority patient alarm (for example, a low SpO2 alarm limit violation). Additionally there are short yellow alarms, most of which are specific to arrhythmia-related patient conditions (for example, ventricular bigeminy).
INOPs
INOPs are technical alarms, they indicate that the monitor cannot measure or detect alarm conditions reliably. If an INOP interrupts monitoring and alarm detection (for example, LEADS OFF), the monitor places a question mark in place of the measurement numeric and an audible indicator tone will be sounded. INOPs without this audible indicator indicate that there may be a problem with the reliability of the data, but that monitoring is not interrupted.
Most INOPs are light blue, however there are a small number of INOPs which are always yellow or red to indicate a severity corresponding to red and yellow alarms. The following INOPs can also be configured as red or yellow INOPs to provide a severity indication:
ECG LEADS OFF REPLACE BATTERY (when using disposable batteries) TELE BATT EMPTY (when using the rechargeable battery pack)
All monitors in a unit should have the same severity configured for these INOPs.
4-2 Alarms
Page 53
The MX40 is designed to achieve visual alarm notification at a distance of up to one meter, which is consistent with its intended use model as a wearable monitor.
Alarms are indicated after the alarm delay time. This is made up of the system delay time plus the trigger delay time for the individual measurement. For more information see ECG Performance Disclosure/Specifications p. 13-23 .
If more than one alarm is active, the highest priority alarm is shown. A downward facing arrow symbol next to the alarm message informs you that more than one message is active. The monitor sounds an audible indicator for the highest priority alarm.
Visual Alarm Indicators
Warning
The MX40 display is inactive for a majority of the time because it is operating in Telemetry Mode. You must activate the screen to view any alarms locally. The alarm message text is displayed, however, the alarm volume setting is at zero.
Alarm Message
An alarm message text appears in the alarm status area at the top of the screen indicating the source of the alarm. If more than one alarm is present, there is a downward facing arrow symbol at the right side. The background color of the alarm message matches the alarm priority: red for red alarms, yellow for yellow alarms, light blue for standard INOPs, red for red INOPs and yellow for yellow INOPs. The asterisk symbols (*) beside the alarm message match the alarm priority: *** for red alarms, ** for yellow alarms, * for short yellow alarms. Standard INOPs do not have a symbol, red and yellow INOPs have exclamation marks beside the alarm message: !!! for red INOPs and !! for yellow INOPs.
Alarm limit violation messages are displayed in text form, for example ** SpO2 LOW.
Alarms 4-3
Page 54
INOP Color
On
Off
Yellow
1.0 seconds
1.0 seconds
Red
0.25 seconds
0.25 seconds
Alarm Indicator
An Alarm Indicator on the MX40 main display communicates alarm/INOP conditions that have not been acknowledged. The alarm indicator is divided into two sections and appears in the upper right hand corner normally occupied by the time display. The right section flashes for a physiological alarm, except for short yellow alarms where the indicator will light for approximately six seconds. The color is yellow or red corresponding to the highest priority alarm currently present.
An unacknowledged physiological alarm and INOP appears as (portrait view):
An acknowledged physiological alarm and INOP with an additional unacknowledged physiological alarm appears as (landscape view):
The left section lights continuously for a standard INOP and flashes for INOPs configured as red or yellow alarms as follows:
If only patient alarms are present, and no INOPs, the patient alarms will use both left and right sections to flash (for red and yellow alarms) or light for approximately six seconds (for short yellow alarms). If only INOPs are present, and no patient alarms, red and yellow INOPs will use both left and right sections to flash, but standard INOPs will always light continuously in the left section only.
Once all alarm/INOP conditions are acknowledged, the time display reappears.
Flashing Numeric
4-4 Alarms
The numeric of the measurement in alarm flashes.
Page 55
Audible Alarm Indicators when in Monitoring Mode
The audible alarm indicators configured for your monitor depend on which alarm standard applies in your hospital. Audible alarm indicator patterns are repeated until you acknowledge the alarm by switching it off or pausing it, or until the alarm condition ceases (if audible alarm indication is set to non-latching).
Warning
Do not rely exclusively on the audible alarm system for patient
monitoring. Adjustment of alarm volume to a low level or off during patient monitoring may result in patient danger. Remember that the most reliable method of patient monitoring combines close personal surveillance with correct operation of monitoring equipment.
No audible alarm indicators are available when the MX40's volume
setting is zero or when operating in Telemetry Mode. Audible alarm indicators become active as soon as the MX40 is no longer connected to the Information Center.
Traditional Audible Alarms (HP/Agilent/Philips/Carenet)
Red alarms and red INOPs: A high pitched sound is repeated once a
second.
Two-star yellow alarms and yellow INOPs: A lower pitched sound is
repeated every two seconds.
One-star yellow alarms (short yellow alarms): The audible indicator is
the same as for yellow alarms, but of shorter duration.
Standard INOPs: an INOP tone is repeated every two seconds.
ISO/IEC Standard Audible Alarms
Red alarms and red INOPs: A high pitched tone is repeated five times,
followed by a pause.
Two-star yellow alarms and yellow INOPs: A lower pitched tone is
repeated three times, followed by a pause.
One-star yellow alarms (short yellow alarms): The audible indicator is
the same as for yellow alarms, but of shorter duration.
Standard INOPs: a lower pitched tone is repeated twice, followed by a
pause.
Alarms 4-5
Page 56
Acknowledging Alarms
To acknowledge all active alarms and INOPs, touch the Silence Alarm button. This switches off the audible alarm indicators, if present, and alarm messages.
A check mark beside the alarm message indicates that the alarm has been acknowledged.
If the condition that triggered the alarm is still present after the alarm has been acknowledged, the alarm message stays on the screen with a check mark symbol beside it, except for NBP alarms and alarms from other intermittent measurements. When such an alarm is acknowledged the alarm message disappears.
If the alarm condition is no longer present, all alarm indicators stop and the alarm is reset.
Switching off the alarms for the measurement in alarm, or switching off the measurement itself, also stops alarm indication.
Pausing or Switching Off Alarms
If you want to temporarily prevent alarms from sounding, for example while you are moving a patient, you can pause alarms, if configured. Depending on your MX40 configuration, alarms are paused for one, two or three minutes.
Cautions
When operating with Information Center Release L or M, the alarm
pause time of the MX40 is not configurable. The alarm pause time for the MX40 is always two minutes.
When operating with Information Center Release L or M, if alarms are
paused at the Information Center, the "Alarms Paused" message is only displayed at the Information Center.
To Pause All Alarms
4-6 Alarms
Select the Alarms SmartKey and select Pause Alarms. A timer on the
display shows the remaining pause time.
Page 57
To Switch Individual Measurement Alarms On or Off
1 Select the measurement numeric to enter its setup menu. 2 Select Alarms to switch between on and off
The alarms off symbol is shown beside the measurement numeric.
While Alarms are Paused
In the alarm field, the MX40 displays the message ALARMS PAUSED
1:28 or ALARMS OFF, together with the alarms paused symbol or the alarms off symbol.
No alarms are sounded and no alarm messages are shown. INOP messages are shown but no INOP tones are sounded.
The only exceptions are the INOPs CUFF NOT DEFLATED, NBP CUFF OVERPRESS and INOPs relating to empty, missing and malfunctioning batteries.
These INOPs switch the alarms on, and the INOP tones are sounded, even if alarms are paused or off. You need to remove the INOP condition first before you can switch the alarm tones off again.
Warning
If connection to the Information Center is lost during an alarms paused period, upon reconnection, alarms remain paused at the Information Center for the set time. You can resume alarms from the Patient Window at the Information Center at any time.
Restarting Paused Alarms
To manually switch on alarm indication again after a pause, select Pause Alarms.
Alarm Limits
The alarm limits you set determine the conditions that trigger yellow and red limit alarms. For some measurements (for example, SpO2), where the value ranges from 100 to 0, setting the high alarm limit to 100 switches the high alarm off, or setting the low alarm limit to 0 switches it off. In these cases, the alarms off symbol is not displayed.
Alarms 4-7
Page 58
Warning
Be aware that the monitors in your care area may each have different alarm settings, to suit different patients. Always check that the alarm settings are appropriate for your patient before you start monitoring.
Viewing Individual Alarm Limits
You can see the alarm limits set for each measurement next to the measurement numeric on the main screen.
Reviewing Alarms
You can see which alarms and INOPs are currently active in the respective alarms and INOPs fields at the top of the screen.
To see the currently active alarms and INOPs listed in one place, touch the Alarms area.
All alarms and INOPs are erased from the Alarm Messages window when you discharge a patient, or if you change to Demonstration Mode.
Review Alarms Window
The Review Alarms window contains a list of the 50 most recent alarms and INOPs with date and time information.
The Review Alarms window also shows when alarms are paused or silenced.
4-8 Alarms
Note — Alarms that occur during an alarm suspend period will appear in
the Review Alarm window, however, they are not communicated to the Information Center.
Page 59
Red & Yellow Measurement Alarms
Non-latching Alarms
Visual and Audible Latching
Alarm has not been acknowledged.
Alarm condition still present.
Alarm tone on. Alarm message. Flashing numerics.
Alarm tone on. Alarm message. Flashing numerics.
Alarm condition no longer present.
All audible and visual alarm indicators automatically stop.
Alarm tone on. Alarm message . Flashing numerics.
Alarm has been acknowledged.
Alarm condition still present.
Alarm tone off. Alarm message with check mark. Flashing numerics. Audible alarm reminder (if configured)
Alarm tone off. Alarm message with check mark. Flashing numerics. Audible alarm reminder (if configured)
Alarm condition no longer present.
Audible and visual alarm indicators automatically stop.
Audible and visual alarm indicators automatically stop.
Latching Alarms
The alarm latching setting for your MX40 defines how the alarm indicators behave when you do not acknowledge them. When alarms are set to non-latching, their indicators end when the alarm condition ends. Switching alarm latching on means that visual and/or audible alarm indications are still displayed or announced by the monitor after the alarm condition ends. The indication lasts until you acknowledge the alarm by touching the Alarm Silence button.
Alarm Latching Behavior
Alarm Behavior at Power On
Alarms 4-9
If the MX40 is powered off for longer than one minute and then powered on again (or after a loss of power lasting longer than one minute, or when a patient is discharged), the device restores the alarm settings from the Information Center's configured default settings.
If power is lost for less than one minute, the alarm on/off condition prior to the power loss is restored.
Page 60
Alarm Text
Priority
Condition
Source
*** ASYSTOLE
Red
Asystole. No QRS for 4 consecutive seconds
ST/AR Basic & Enhanced Arrhythmia
*** EXTREME BRADY
Red
Extreme Bradycardia. Heart Rate (yyy) less than Extreme Brady limit (xxx)
ST/AR Basic & Enhanced Arrhythmia
Physiologic Alarms
Physiologic alarms indicate a life-threatening situation or a less urgent situation such as heart rate beyond limits.
Warning
Do not rely exclusively on the audible alarm system for patient
monitoring. Adjustment of alarm volume to a low level during patient monitoring may result in patient danger. Remember that the most reliable method of patient monitoring combines close personal surveillance with correct operation of monitoring equipment.
ST and QT related alarm messages appear at the Information Center
only.
Arrhythmia alarm chaining and customizing arrhythmia alarm settings are described in the ECG and Arrhythmia Monitoring chapter. There are two levels of arrhythmia analysis available: Basic and Enhanced. Enhanced analysis includes Basic alarms.
The MX40 provides physiological alarms based on the settings at the Information Center, Release N or later. Alarming is not active on the MX40 until it is configured via an active association with the Information Center.
In the following table, Red (***) alarms are listed alphabetically, followed by the Yellow (**) alarms, and the Yellow (*) alarms.
Note — The physiological alarm messages displayed on the MX40 use a
short text format. Alarm messages displayed at the Information Center use and extended text format, e.g. "*HR yyy<xxx".
4-10 Alarms
Page 61
Alarm Text
Priority
Condition
Source
*** DESAT
Red
Very Low SpO2 Saturation. SpO2 value below Desaturation limit
Note Desat limit is set 10 points
below low limit.
SpO2 ***EXTREME TACHY
Red
Extreme Tachycardia. Heart Rate (yyy) greater than Extreme Tachy limit
ST/AR Basic & Enhanced Arrhythmia
*** VENT FIB/TACH
Red
Ventricular Fibrillation. Fibrillatory wave (sinusoidal wave between 2-10 Hz) for 4 consecutive seconds
ST/AR Basic & Enhanced Arrhythmia
*** V-TACH
Red
Ventricular Tachycardia. Consecutive PVCs greater than or equal to V-Tach Run limit and Heart Rate greater than V-Tach limit (xxx)
ST/AR Basic & Enhanced Arrhythmia
*/**AFIB
Yellow
Atrial fibrillation waveform detected
ST/AR Enhanced Arrhythmia
**NBP High
Yellow
High limit has been exceeded for high pressure limit
NBP
**NBP Low
Yellow
Low limit has been exceeded for low pressure limit
NBP
** SpO2T HIGH
Yellow
High SpO2. SpO2 value (yyy) greater than high SpO2 limit (xxx).
SpO2
** SpO2T LOW
Yellow
Low SpO2. SpO2 value (yyy) less than low SpO2
limit (xxx).
SpO2
*/** HR High
Yellow
Heart Rate (yyy) greater than the upper Heart rate limit (xxx).
ST/AR Basic & Enhanced
Arrhythmia
*/** HR Low
Yellow
Heart Rate (yyy) lower than the lower Heart Rate limit (xxx).
ST/AR Basic & Enhanced
Arrhythmia
*/** IRREGULAR HR
Yellow
Consistently irregular rhythm (irregular R-R intervals).
ST/AR Enhanced
Arrhythmia
Alarms 4-11
Page 62
Alarm Text
Priority
Condition
Source
* MISSED BEAT
Yellow
No beat detected for 1.75 x average R-R interval for Heart Rate greater than 120, or no beat for 1 second with Heart Rate greater than 120 (non-paced patient only).
ST/AR Enhanced
Arrhythmia
* MULTIFORM PVCs
Yellow
The occurrence of two differently shaped Vs, each occurring at least twice within the last 300 beats as well as each occurring at least once within the last 60 beats.
ST/AR Enhanced
Arrhythmia
* NON-SUSTAIN VT
Yellow
A run of Vs having a ventricular Heart Rate greater than V-Tach limit but lasting for less than the V-Tach Run limit.
ST/AR Enhanced
Arrhythmia
*Nurse Call
Yellow
The patient has pressed the Multi-Function Button on the MX40.
* PACER NOT CAPTURE
Yellow
No QRS for 1.75 x the average R-R interval with Pace Pulse (paced patient only).
ST/AR Basic & Enhanced
Arrhythmia
* PACER NOT PACING
Yellow
No QRS and Pace Pulse for 1.75 x the average R-R interval (paced patient only).
ST/AR Basic & Enhanced
Arrhythmia
* PAIR PVCs
Yellow
Two consecutive PVCs between non-PVCs.
ST/AR Enhanced
Arrhythmia
* PAUSE
Yellow
No QRS detected for x seconds. Choices of >1.5 to 2.5 seconds.
ST/AR Enhanced
Arrhythmia
* PVCs /MIN HIGH
Yellow
PVCs within one minute exceed by the PVCs/min limit (xxx).
ST/AR Basic & Enhanced
Arrhythmia
4-12 Alarms
Page 63
Alarm Text
Priority
Condition
Source
* R-ON-T PVCs
Yellow
For Heart Rate less than 100, a PVC with R-R interval less than 1/3 the average interval followed by a compensatory pause of 1.25 x average R-R interval, or 2 such Vs without a compensatory pause occurring within 5 minutes of each other. (When Heart Rate is greater than 100, 1/3 R-R interval is too short for detection.)
ST/AR Enhanced
Arrhythmia
* RUN PVCs
Yellow
Run of PVCs greater than or equal to
2.
ST/AR Enhanced
Arrhythmia
* SVT
Yellow
Run of SVPBs greater than or equal to SVT Run limit and with SVT Heart Rate greater than the SVT Heart Rate limit.
ST/AR Enhanced Arrhythmia
* VENT BIGEMINY
Yellow
A dominant rhythm of N, V, N, V (where N= supraventricular beat, V=ventricular beat).
ST/AR Enhanced
Arrhythmia
* VENT RHYTHM
Yellow
A dominant rhythm of adjacent Vs greater than Vent Rhythm limit and ventricular Heart Rate less than V-Tach limit.
ST/AR Enhanced
Arrhythmia
* VENT TRIGEMINY
Yellow
A dominant rhythm of N, N, V, N, N, V (where N=supraventricular beat, V=ventricular beat).
ST/AR Enhanced
Arrhythmia
Alarms 4-13
Page 64
Alarm Text
Priority
Condition
What to do
BATTERY LOW T
Source - MX40
Soft
There is less than 15
minutes of monitoring time remaining (AA batteries).
Lithium-ion battery
level is < 10% or has <30 minutes remaining time.
Replace batteries
promptly to avoid shutdown and cessation of monitoring.
Insert a charged lithium-ion battery pack.
CANNOT ANALYZE ECG
Source - MX40 and Information Center
Hard
Arrhythmia algorithm cannot reliably analyze the ECG data on any monitored leads.
Assess the lead selections, initiate relearn, and validate analyzed rhythm. Check other INOPs for possible source of problem.
Technical Alarms (INOPs)
Technical Alarms, or INOPs (inoperative conditions), are sourced at the MX40, the ST/AR algorithm running at the Information Center, or the IntelliVue Patient Monitor. They identify inoperative conditions (that is conditions where the system is not operating properly and therefore cannot measure or detect alarm conditions reliably). There are four levels of Technical Alarms:
Severe - Monitoring and alarm generation are disabled. Visual alarm
indicator on the MX40. Audible tone at the Information Center. Must be acknowledged by a clinician.
Hard - Monitoring and alarm generation are disabled. Visual alarm
indicator on the MX40. Audible tone at the Information Center. If the hard INOP is "latched", the sound will be silenced, but the message will remain on the display until resolution of the offending condition.
Soft - Monitoring and alarms remain active. Visual alarm indicator on
the MX40 and at the Information Center. No audible tones are generated at the Information Center
Red/Yellow - Replace Battery and ECG Leads Off INOPs may be
configured to display as either Red or Yellow Technical Alarms.
Note - The ECG Leads Off INOP will initially display as a cyan
technical alarm until a valid ECG signal is obtained.
In the following table, technical alarms are listed alphabetically.
4-14 Alarms
Page 65
Alarm Text
Priority
Condition
What to do
CHECK PAIRING
Source - MX40
Yellow Technic al Alarm
There is a problem
with device pairing.
When the MX40 is
wirelessly paired with an X2 patient monitor (no label) docked with a larger networked MP series monitor, and the network connection is lost.
Check that the bedside
monitor is correctly paired.
Select the correct device
to be paired.
cl NBP Batt Low
Source - Cableless Measurement Device
Hard
CL NBP Pod weak battery condition.
Charge CL NBP Pod.
cl NBP Batt Empty
Source - Cableless Measurement Device
Severe
CL NBP Pod empty battery condition. Monitoring is not possible.
Replace CL NBP Pod. Recharge depleted CL
NBP Pod.
cl NBP DISCONNECT
Source - Cableless Measurement Device
Hard
CL NBP Pod is not connected with the MX40.
Resolve interference
condition.
Reduce range between
CL NBP Pod and MX40.
cl SpO2 Batt Low
Source - Cableless Measurement Device
Hard
CL SpO2 Pod weak battery condition.
Charge CL SpO2 Pod.
cl SpO2 Batt Empty
Source - Cableless Measurement Device
Severe
CL SpO2 Pod empty battery condition. Monitoring is not possible.
Replace CL SpO2 Pod. Recharge depleted SpO2
Pod.
cl SpO2 DISCONNECT
Source - Cableless Measurement Device
Hard
CL SpO2 Pod is not connected with the MX40.
Resolve interference
condition.
Reduce range between
CL SpO2 Pod and MX40.
!!!/!! CUFF NOT DEFLATED
Source - Cableless Measurement Device
Severe
Cuff pressure has exceeded the specified safety limit.
Remove cuff and tubing and expel air.
Alarms 4-15
Page 66
Alarm Text
Priority
Condition
What to do
!!!/!! CUFF OVERPRESS
Source - Cableless Measurement Device
Severe
Cuff pressure has increased above overpressure safety limits.
Remove cuff and tubing and expel air.
ECG/ARRH ALARM OFF
Source - MX40
Soft
ECG is turned off.
Turn on ECG.
ECG LEADS OFF Note This INOP may
also be configured to display as a Red or Yellow Technical Alarm.
Source - MX40
Red or Yellow or Hard Technic al Alarm
Multiple leads are off.
Re-attach ECG leads to patient..
<electrode> LEAD OFF
Source - MX40
Hard
Single lead is off. If primary lead is MCL,
lead will be identified as V/C in INOP text.
Re-attach ECG leads to patient.
LEADSET UNPLUGGED
Source - MX40
Hard
Patient cable has
been unplugged from the MX40.
Incompatible leadset
attached to patient cable.
Re-attach the patient
cable.
Replace the leadset.
LOCAL AUDIO OFF
Source - MX40
Note This is normal
operation in Telemetry Mode.
Soft
There is no alarm audio notification when operating in Telemetry Mode.
Change to Monitor Mode.
NBP INTERRUPTED
Source - Cableless Measurement Device
Hard
The preset maximum time for the total measurement has been exceeded.
Reduce patient movement and avoid interaction with the cuff and tubing.
NBP MEASURE FAILED
Source - Cableless Measurement Device
Hard
Measurement values cannot be derived.
Attach cuff to new location
on patient.
Replace cuff.
4-16 Alarms
Page 67
Alarm Text
Priority
Condition
What to do
NBP EQUIP MALF
Source - Cableless Measurement Device
Hard
Tubing may be
obstructed or kinked.
Hardware malfunction.
Check tubing. If condition persists,
contact Service.
NO ALARM DISPLAY
Source - MX40
Soft
When operating with Information Center Release L Or M, there is no local alarming at the MX40, networked or non-networked.
Condition is not present when operating with Information Center Release N or later (unless specifically configured to operate in this way).
NO CENTRAL MONITOR (appears at MX40 only)
Source - MX40
Hard
The MX40 is out of
range of the network.
Patient Sector at the
Information Center is in Standby.
Return the MX40 to the
coverage area.
Select Resume at the
Information Center.
NO HOST MONITORING
Source - MX40
Hard
The paired MX40/bedside monitor is out of short-range radio range or there is excessive radio interference.
Reduce the distance
between the devices.
Identify and remove
interference source.
NO SIGNAL (appears at the Information Center only)
Source - Information Center
Hard, Latched
The MX40 is outside
the coverage area, or
No batteries in the
MX40, or
The MX40 has failed.
Make sure that the MX40
is within the coverage area and has good batteries.
Replace the MX40 if
Power On Self Test fails. Put bed in Standby. Contact Service
REPLACE BATTERY T
Source - MX40
Note This INOP may
also be configured to display as a Red or Yellow Technical Alarm.
Red or Yellow or Hard Technic al Alarm, Latched
Dead battery. No
monitoring is
occurring. When operating
wirelessly, the patient
monitor is no longer
providing power to the
MX40, and battery
capacity is now
depleted.
Replace batteries.
Alarms 4-17
Page 68
Alarm Text
Priority
Condition
What to do
SpO2T EQUIP MALF
Source - MX40
Hard
Malfunction in the SpO2 equipment
Contact Service.
SpO2T ERRATIC
Source - MX40
Hard
Erratic SpO2 measurements, often due to a faulty sensor or invalid SpO2 measurements, or incorrect transducer position
Repeat measurement, reposition sensor on patient, or finally, replace sensor.
SpO2T EXTD UPDATE Numeric is replaced by a
-?-.
Source - MX40
Soft
The update period of displayed values is extended due to an NBP measurement on the same limb or an excessively noisy signal.
If NBP is not active, check the sensor placement. Reposition the sensor on patient, or replace sensor.
SpO2T LOW PERF
Source - Monitor
Soft
Accuracy may be reduced due to low perfusion. Data displayed with ?.
Increase perfusion. Change sensor site. Avoid site distal to BP cuff or intra-arterial line. Warm the site.
SpO2T INTERFERENCE
Source - MX40
Hard
Level of ambient light or level of electrical interference are so high that the SpO2 sensor cannot measure SpO2 and pulse rate.
Reduce ambient light to sensor or electrical noise sources.
SpO2T NO SENSOR
Note Silencing this
technical alarm turns off the SpO2 measurement on the MX40 only (not at the Information Center).
Source - MX40
Hard
No sensor attached to SpO2 device.
Attach SpO2 sensor.
%SpO2T NOISY SIGN
Source - MX40
Hard
Excessive patient movements or electrical interference are causing irregular pulse patterns
Reduce movement or electrical noise sources.
4-18 Alarms
Page 69
Alarm Text
Priority
Condition
What to do
SpO2T NO PULSE
Source - MX40
Note When paired
directly with an IntelliVue MP5 Patient Monitor, the INOP will display as SpO2T SENSOR OFF.
Hard
Pulse is too weak or
not detectable Sensor has fallen off
at patient.
Check connection to
patient. Change sensor site. Avoid
site distal to BP cuff or
intra-arterial line.
SpO2 POOR SIGNAL
Source - MX40
Soft
Although a measurement may be possible, its accuracy may be reduced due to poor signal quality.
Apply the sensor
according to the
manufacturer's
instructions. Relocate the sensor to a
different site on the
patient.
SpO2T SEARCHING
Source - MX40
Soft
The patient signal is analyzed, but a valid numeric is not available yet.
Wait for the measurement to complete.
SpO2T SENSOR OFF
Note The ability of
the algorithm to detect this condition depends on the sensor type in use.
Hard
The algorithm has determined that a sensor is connected, but not properly applied to the patient.
Apply the sensor
according to the
manufacturer's
instructions. If the condition persists,
relocate the sensor to a
different site on the
patient.
SpO2T SENSOR MALF
Source - MX40
Hard
Malfunction of the SpO2 sensor/adapter cable
Replace sensor.
SpO2 UNKN SENSOR
Source - MX40
Hard
The connected SpO2 sensor and/or adapter cable is not supported by the hardware version.
Use specified sensor and/or adapter cable.
SpO2T UPGRADE Source - MX40
Soft
SpO2 hardware is in upgrade process. Monitoring is not possible.
Wait for the upgrade process to complete.
TELE BATTERY LOW Source - MX40
Soft
Lithium-ion battery level is < 20% or has <30 remaining time.
Insert a charged lithium-ion battery pack.
Alarms 4-19
Page 70
Alarm Text
Priority
Condition
What to do
TELE BATT EMPTY
Note This INOP may
also be configured to display as a Red or Yellow Technical Alarm.
Source - MX40
Note For Information
Center Release L or M, this INOP will appear as "REPLACE BATTERY T".
Hard, Latched
Lithium-ion battery level is critically low. A 10-minute countdown begins. The MX40 will shut down if the condition is not cleared.
Insert a charged lithium-ion battery pack.
TELE BATTERY TEMP Source - MX40
Note For Information
Center Release L or M, this INOP will appear as "REPLACE BATTERY T"
Hard
The temperature of the lithium-ion battery is above 55o C or below -5o C.
Replace the lithium-ion battery.
TELE CHECK BATT Source - MX40
Soft
Lithium-ion battery has < 25 charge cycles remaining before reaching the charge cycle maximum limit.
Be aware that the Lithium-ion battery pack will soon need replacement.
TELE MALFUNCTION
Source - MX40
Hard
MX40 malfunction or self-test failure.
Contact Service to replace the MX40.
TRANSMITTER OFF
Source - MX40
Hard
RF Auto Shutoff after 10 minutes of all leads off and no SpO2 sensor connected.
Reattach ECG leads to
patient. Reattach SpO2 sensor.
TELE REMOVE BATT Source - MX40
Note For Information
Center Release L or M, this INOP will appear as "REPLACE BATTERY T".
Hard, Latched
The temperature of the lithium-ion battery is >60o C and the battery must be removed.
Replace the lithium-ion
battery. Dispose of old battery
properly.
TELE SERVICE BATT Source - MX40
Hard
The lithium-ion battery has exceeded the maximum charge cycle limit and reached the end of its useful life.
Replace the lithium-ion
battery. Dispose of old battery
properly.
4-20 Alarms
Page 71
Alarm Text
Priority
Condition
What to do
TELE WEAK SIGNAL
Source - MX40
Soft
Patient is at outer
range of the radio
coverage area. The MX40 is receiving
a weak signal with
high data loss from
the AP. Condition exists for
multiple devices in a
specific area
Return patient to the
coverage area. If patient is in close
proximity to AP, replace
the MX40. Contact
service. The AP covering the
specific area is suspect.
Contact Service
Alarms 4-21
Page 72
Page 73
ECG Safety Information ........................................................................... 5-2
Measuring ECG ......................................................................................... 5-4
Connecting and Positioning ECG Electrodes ........................................ 5-5
Selecting the Primary and Secondary ECG Leads ................................ 5-6
Checking Paced Status .............................................................................. 5-7
Understanding the ECG Display............................................................. 5-8
Monitoring Paced Patients ....................................................................... 5-9
Changing the Size of the ECG Wave .................................................... 5-10
Choosing EASI or Standard Lead Placement ...................................... 5-11
ECG Configuration .................................................................................. 5-12
ECG Leads Monitored ............................................................................ 5-13
Reconstructed Leads ............................................................................... 5-15
3-Wire Placement ..................................................................................... 5-16
5-Wire Placement (Standard Mode)...................................................... 5-17
5-Wire Placement (EASI Mode) ............................................................. 5-18
6-Wire Placement ..................................................................................... 5-19
Chest Electrode Placement ..................................................................... 5-20
Monitoring during Leads Off ................................................................ 5-21
ST/AR Arrhythmia Monitoring ............................................................. 5-23
ST/AR ST Analysis Algorithm ............................................................... 5-30
QT Interval Monitoring .......................................................................... 5-33
5. ECG and Arrhythmia Monitoring
This section covers the specifics of ECG measurement and the ST/AR Arrhythmia, ST, and QT algorithms used for arrhythmia monitoring.
ECG and Arrhythmia Monitoring 5-1
Page 74
ECG Safety Information
Warnings
Always confirm MX40 and Information Center observations with
clinical observation of the patient before administering interventions.
To avoid patient injury, assure that the patient cable is not positioned
where leads could become entangled around the patient, or cause choking, strangulation, or inhibit circulation in extremities.
Every lead must be secured to an electrode on the patient. Conductive
parts of electrodes must not contact earth or other conductive parts.
EASI derived 12-lead ECGs and their measurements are
approximations to conventional 12-lead ECGs. As the 12-lead ECG derived with EASI is not exactly identical to the 12-lead conventional ECG obtained from an electrocardiograph, it should not be used for diagnostic interpretations.
EASI lead placement is supported for adult patients only. Ensure that the patient cable is properly connected to the MX40. Do not mix and match electrodes of different types. In particular, do not
use electrodes of dissimilar metals. This helps ensure optimal signal quality.
Non-manufacturer supplied accessories and supplies can corrupt the
performance of the equipment. Use only AAMI EC-12 compliant electrodes with this device. Use of electrodes that are non-compliant may provide erroneous results.
During complete heart block or pacemaker failure (to pace or capture),
tall P-waves (greater than 1/5 of the average R-wave height) can be erroneously counted by the arrhythmia algorithm, resulting in missed detection of cardiac arrest.
Caution
To protect the MX40 from damage during defibrillation, to ensure
accurate ECG information, and to provide protection against signal noise and other interference, use only ECG electrodes and cables specified by Philips.
Philips recommends that you change the lead label only to reflect the
physical placement of electrodes. This will ensure a match between the monitored lead and the label, and prevent any possible confusion.
5-2 ECG and Arrhythmia Monitoring
Page 75
Note — When switching from EASI to standard monitoring, there is a
momentary loss of data.
For Paced Patients
Warnings
The output power of the MX40 and other sources of radio frequency
energy, when used in the proximity of a pacemaker, can be sufficient to interfere with pacemaker performance. Due to the shielding effects of the body, internal pacemakers are somewhat less vulnerable than external pacemakers. However, caution should be exercised when monitoring any paced patient.
In order to minimize the possibility of interference, position electrodes, electrode wires, and the MX40 as far away from the pacemaker as possible.
Consult the pacemaker manufacturer for information on the RF susceptibility of their products and the use of their products with the MX40. See the IntelliVue Information Center Instructions for Use for additional information on monitoring paced patients.
When an external pacemaker is being used on a patient, arrhythmia
monitoring is severely compromised due to the high energy level in the pacer pulse. This may result in the arrhythmia algorithm's failure to detect pacemaker non-capture or asystole.
Pacemakers that create fusion beats (pace pulse on top of the QRS
complex) cannot be detected by the monitor's QRS detector.
For paced patients who exhibit only intrinsic rhythm, the monitor can
erroneously count pace pulses as QRS complexes when the algorithm first encounters them, resulting in missed detection of cardiac arrest. The risk of missing cardiac arrest can be reduced by monitoring these patients with the low heart rate limit at or slightly above the basic/demand pacemaker rate. A low heart rate alarm notifies you when the patient begins pacing. Proper detection and classification of the paced rhythm can then be determined.
Note— During defibrillation, monitoring may be temporarily interrupted
or distorted. It may take several seconds for the ECG trace to reappear on the screen. After defibrillation, the device will continue to monitor as before; the device settings will not be affected.
ECG and Arrhythmia Monitoring 5-3
Page 76
Measuring ECG
The electrocardiogram (ECG) measures the electrical activity of the heart and displays it on the MX40 and the Information Center as a waveform and a numeric.
In order to compare measured ECG signals, the electrodes (or patient cables) are placed in standardized positions, forming "leads". To obtain ECG signals optimized for use in diagnosis and patient management in different care environments, different leadsets in varying lead placements are used. Both standard lead and EASI lead placements can be used with the MX40.
The Heart Rate calculation resides in the arrhythmia algorithm on the MX40 and at the Information Center. Arrhythmia analysis is always turned on for telemetry patients. Arrhythmia analysis is either Basic or Enhanced, depending on the product configuration.
5-4 ECG and Arrhythmia Monitoring
Page 77
Connecting and Positioning ECG Electrodes
Correct lead placement is always important for accurate diagnosis. Especially in the precordial leads, which are close the heart, QRS morphology can be greatly altered if an electrode is moved away from its correct location. Each electrode is color-coded. Use the placement diagrams available on the display of the MX40 and in this section for guidance. Additional lead placement information is available in the Online Help at the IntelliVue Information Center.
When placing electrodes on the patient, choose a flat, non-muscular site where the signal will not be impacted by either movement or bones.
Philips recommends that electrodes be changed every 24 hours.
In addition to correct positioning of the electrodes, optimal skin preparation prior to electrode placement will help ensure a clear signal for diagnosis.
1 Prepare the patient’s skin. Good electrode-to-skin contact is important
for a good ECG signal, as the skin is a poor conductor of electricity.
Select sites with intact skin, without impairment of any kind. Clip or shave hair from the site as necessary. Wash site with soap and water, leaving no soap residue.
Note Philips does not recommend using ether or pure alcohol,
because they dry the skin and increase the resistance.
Dry thoroughly. Use ECG skin preparation paper (abrasive) to remove dead skin
cells and to improve the conductivity of the electrode site.
2 Check electrodes for moist gel, and attach to the clips. If you are not
using pre-gelled electrodes, apply electrode gel to the electrodes before placement.
Note — Gel must be moist to provide a good signal.
3 Place the electrodes on the patient according to the lead placement you
have chosen (see the electrode placement diagrams following). Place the edge down, then "roll down" the rest of the pad. Press firmly around the adhesive edge toward the center.
4 Attach the patient cable to the MX40. An ECG waveform and numeric
appear on the monitor display.
ECG and Arrhythmia Monitoring 5-5
Page 78
Selecting the Primary and Secondary ECG Leads
The MX40 uses the primary and secondary lead selected at the Information Center to compute HR and to analyze and detect cardiac arrhythmias. They are also available for recordings and for display on the Information Center.
The secondary lead is used if your device is configured for multi-lead (instead of single-lead) arrhythmia analysis.
You should choose a lead as primary or secondary lead at the Information Center that has the following characteristics:
the QRS complex should be either completely above or below the
baseline and it should not be biphasic
the QRS complex should be tall and narrow the P-waves and T-waves should be less than 0.2 mV
5-6 ECG and Arrhythmia Monitoring
Page 79
Checking Paced Status
It is important to set the paced status correctly when you start monitoring ECG.
Note — Paced status is set at the Information Center and can only be
changed at the Information Center.
When Paced is set to Yes:
Pacer Algorithm is switched on. This means that pacemaker pulses are
not counted as extra QRS complexes.
The pacer spikes are shown in white. The paced symbol is displayed.
When Paced is set to No and your patient has a pacemaker, pace pulses may be counted as regular QRS complexes, which could prevent an asystole event from being detected.
Warning
Pace pulse rejection must be switched on for paced patients by setting
Paced to Yes. Switching pace pulse rejection off for paced patients may result in pace pulses being counted as regular QRS complexes, which could prevent an asystole event from being detected. At admission/discharge, always check that paced status is correct for the patient.
Some pace pulses can be difficult to reject. When this happens, the
pulses are counted as a QRS complex, and could result in an incorrect HR and failure to detect cardiac arrest or some arrhythmias. Make sure that pace pulses are detected correctly by checking the pace pulse markers on the display. Keep pacemaker patients under close observation.
ECG and Arrhythmia Monitoring 5-7
Page 80
1. Lead label of the displayed wave
2. 1 mV calibration bar
3. Pacer spikes
4. Current heart rate
5. Current heart rate alarm limits
6. EASI lead placement label (located here when active)
Understanding the ECG Display
Your display may be configured to look slightly different.
ECG HR numeric: This is the heart rate derived from the monitored ECG. Pacer Spikes: The pacer spikes are shown in white.
5-8 ECG and Arrhythmia Monitoring
Page 81
1. Normal Beats
2. Pace Pulses/Beats
1. Repolarization tail (note width)
Monitoring Paced Patients
An ECG optimized for monitoring a paced patient should look like this:
You should choose a lead as primary or secondary lead that has these characteristics:
the normal QRS complex should be either completely above or below
the baseline and it should not be biphasic. For paced patients, the QRS complexes should be at least twice the height of pace pulses.
the QRS complex should be tall and narrow the P-waves and the T-waves should be less than 0.2 mV.
Avoiding Pace Pulse Repolarization Tails
Some unipolar pacemakers display pace pulses with repolarization tails. These tails may be counted as QRSs in the event of cardiac arrest or other arrhythmias.
If you note a visible repolarization tail, choose a lead that decreases the size of the repolarization tail.
ECG and Arrhythmia Monitoring 5-9
Page 82
Changing the Size of the ECG Wave
If any of the displayed ECG waves is too small or clipped, you can change the size of the ECG waves on the screen.
Changing the adjustment factor only changes the visual appearance of the ECG wave on the MX40. It does not affect the ECG signal analyzed by the algorithm.
Comparing the wave size to the 1 mV calibration bar on the ECG wave segment can help you get an idea of the true ECG signal strength.
To change the size of the ECG waves on the screen by
a fixed adjustment factor:
1 Touch the HR parameter. 2 In the Setup ECG menu, scroll to the second page and select Adjust
Size.
3 Select the required adjustment factor from the list.
Size X1/2 to halve the wave size Size X1 to display the wave without zoom Size X2 to double the wave size Size X4 to multiply the wave size by four
5-10 ECG and Arrhythmia Monitoring
Page 83
Choosing EASI or Standard Lead Placement
Choose either standard lead placement or EASI lead placement:
1 In the Setup ECG menu, select Lead Placement to toggle between
Standard or EASI.
2 Select Standard or EASI. Note — When changing lead placement, the patient cable must be attached
to the MX40.
EASI is shown beside the 1 mV calibration bar on the ECG wave on the display, and EASI is marked on any recorder strips and printouts.
See the sections on Lead Placement for electrode placement diagrams.
ECG and Arrhythmia Monitoring 5-11
Page 84
Electrode Labels
Electrode Colors
AAMI
EASI
IEC
AAMI
IEC
RA I R
White
Red LA S L
Black
Yellow
LL A F
Red
Green
RL N N
Green
Black
V/Va
E
C/Ca
Brown
White
Vb Cb
Brown/White
White/Blue
ECG Configuration
The MX40 supports 3-, 5-, and 6-wire patient cables. The 5-wire patient cable can be used for either standard or EASI electrode configurations. The MX40 detects the patient cable type attached and automatically determines the ECG measurement and transmitted leads.
Note—The labels and colors of the ECG electrodes differ according to the
standards that apply for your hospital. The electrode placement references and illustrations in this chapter use the AAMI labels and colors. See the table below for additional label and color information.
Arrhythmia analysis resides in the arrhythmia algorithm on the MX40 and at the Information Center. Arrhythmia analysis is always turned on for telemetry patient. Arrhythmia analysis is either Basic or Enhanced (optional).
5-12 ECG and Arrhythmia Monitoring
Page 85
If you are using ...
these leads can be selected at the MX40 and the Information Center
3-wire
I, II, III Sourced (raw) waves are received as: Channel 1 = I, II, or III Factory Default is II.
5-wire (Standard mode)
I, II, III, aVR, aVL, aVF, MCL and V Sourced (raw) waves are received as: Channel 1 = II Channel 2 = III Channel 3 = MCL Factory Defaults are II, V, III.
5-wire (EASI mode)
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6
In EASI mode, the sourced (raw) waves are received as:
Channel 1 = Vector 1 (A-I) Channel 2 = Vector 2 (A-S) Channel 3 = Vector 3 (E-S) Factory Defaults are II, V2, III, V5. Arrhythmia monitoring is performed only on
the primary and secondary leads selected at the Information Center, although you can view and perform ST analysis on all 12 EASI derived leads.
ECG Leads Monitored
Depending on the patient cable connected to the MX40, a different set of viewable leads are available at the MX40 and the Information Center. The MX40 can source up to four raw ECG waves to the Information Center.
ECG and Arrhythmia Monitoring 5-13
Page 86
If you are using ...
these leads can be selected at the MX40 and the Information Center
6-wire
I, II, III, aVR, aVL, aVF, V1, V2, V3, V4, V5, V6, V7, V8, V9, V3R, V4R, V5R.
Sourced (raw) waves are received as: Channel 1 = II Channel 2 = III Channel 3 = Va Channel 4 = Vb Factory Defaults are II, Va = V2, III,Vb =
V5. The two chest leads, Va and Vb, can be
placed on the patient in any of the V lead positions (V1 through V9, V3R, V4R, V5R). Lead assignment is available at the Information Center. When unassigned, the chest leads use the defaults.
Note The lead label assigned to Vb
cannot be selected for Va even though Vb does not appear to be used.
When display of the pleth wave is enabled at the Information Center, the second chest lead (Vb) is not available for monitoring.
5-14 ECG and Arrhythmia Monitoring
Page 87
ECG Lead
Clinical Calculations in terms of electrodes
3-wire
5-wire Standard
6-wire
I I I
LA-RA II
(default)
II (default)
II (default)
LL-RA
III
III (default)
III (default)
LL-LA
-
MCL
V-LA, where V=C
-
aVR
aVR
RA-(LA+LL)/2
-
aVL
aVL
LA-(RA+LL)/2
-
aVF
aVF
LL-(LA+RA)/2
-
V (default)
V-(RA+LA+LL)/3, where V=C
Va
Va-(RA+LA+LL)/3, where Va=V2 (default) position
Vb
Vb-(RA+LA+LL)/3, where Vb =V5 (default) position
Reconstructed Leads
Reconstruction of leads from the sourced wave is defined by the calculations in the following table. EASI reconstructed leads are a linear combination of all three raw EASI leads
ECG and Arrhythmia Monitoring 5-15
Page 88
1. RA - directly below the clavicle and near the right shoulder
2. LA -directly below the clavicle and near the left shoulder
3. LL - on the left lower abdomen
3-Wire Placement
5-16 ECG and Arrhythmia Monitoring
Page 89
1. RA directly below the clavicle and near the right shoulder
2. LA directly below the clavicle and near the left shoulder
3. RL on the left lower abdomen
4. LL on the right lower abdomen
5. V on the chest, the position depends on your required lead selection. The typical position is V1, although this may vary according based on your hospital’s protocol.
V1 on the fourth intercostal space at the
right sternal border
V2 on the fourth intercostal space at the
left sternal border
V3 midway between the V2 and V4
electrode positions
V4 on the fifth intercostal space at the
left midclavicular line
V5 on the left anterior axillary line,
horizontal with the V4 electrode position
V6 on the left midaxillary line, horizontal
with the V4 electrode position
5-Wire Placement (Standard Mode)
ECG and Arrhythmia Monitoring 5-17
Page 90
1. E (V) on the lower sternum at the level of the fifth intercostal space
2. A (LL) on the left midaxillary line at the same level as the E electrode
3. S (LA) on the upper sternum
4. I (RA) on the right midaxillary line at the same level as the E electrode
5. N (Reference) can be anywhere, usually below the sixth rib on the right hip
Make sure that the S and E electrodes line up vertically on the sternum, and that the I, E and A electrodes align horizontally.
5-Wire Placement (EASI Mode)
5-18 ECG and Arrhythmia Monitoring
Page 91
6-Wire Placement
For a 6-lead placement use the positions from the 5-lead diagram above but with two chest leads. The two chest leads, Va and Vb, can be positioned at any two of the V1 to V6 positions shown in the chest electrode diagram below.
The default position of Va - the brown lead - is at the V2 position.
The default position for Vb - the brown/white lead - is at the V5 position.
The lead placement for the Va and Vb lead labels must be appropriate. If your unit uses other precordial leads for Va and Vb, they may be assigned in Unit Settings at the Information Center as defaults for your whole unit, or you may need to assign the new positions on a per-patient basis in the Patient Window at the Information Center.
Selecting Positions of Va and Vb Chest Leads
The two chest leads for the 6-lead placement can be positioned at any two of the V1 to V9 and V3R, V4R and V5R positions. Select the positions you have used in the Patient Window at the Information Center, so that the chest leads will be correctly labeled.
ECG and Arrhythmia Monitoring 5-19
Page 92
V1 on the fourth intercostal space at the right sternal border
V2 on the fourth intercostal space at the left sternal border
V3 midway between the V2 and V4 electrode positions
V4 on the fifth intercostal space at the left midclavicular line
V5 on the left anterior axillary line, horizontal with the V4 electrode position
V6 on the left midaxillary line, horizontal with the V4 electrode position
V3R-V6R on the right side of the chest in positions corresponding to those on the left
VE over the xiphoid process V7 on posterior chest at the left posterior
axillary line in the fifth intercostal space V7R on posterior chest at the right
posterior axillary line in the fifth intercostal space
Chest Electrode Placement
For accurate chest electrode placement and measurement, it is important to locate the fourth intercostal space.
To locate the fourth intercostal space:
1 Locate the second intercostal space by first palpating the Angle of Lewis
(the little bony protuberance where the body of the sternum joins the manubrium). This rise in the sternum is where the second rib is
5-20 ECG and Arrhythmia Monitoring
attached, and the space below this is the second intercostal space.
2 Palpate and count down the chest until you locate the fourth intercostal
space.
Page 93
Monitoring during Leads Off
ECG Fallback and Extended monitoring states are supported for the MX40 when the primary and/or secondary leads are in a "Leads Off" INOP condition. Both of these states are entered into after 10 seconds of "Leads Off" in an attempt to maintain monitoring and arrhythmia analysis.
ECG Fallback
ECG Fallback occurs when the primary lead is in "Leads Off" for 10 seconds and a secondary lead is available.
Multilead Analysis
If there is a "Leads Off" technical alarm in the primary lead for > 10 seconds, the active secondary lead becomes the primary lead. The arrhythmia algorithm switches the leads on the display, but relearn does not occur. When the "Leads Off" condition is corrected, the leads are switched back to their original state.
Single Lead Analysis
For single lead analysis, if there are two leads available, the secondary lead is made the primary lead until the "Leads Off" condition is corrected. The arrhythmia algorithm performs a relearn using the available lead.
Extended Monitoring
If both the primary and secondary leads are in a "Leads Off" condition, the ECG source on the MX40 will switch to any available lead. Relearning will occur in this condition.
Fallback for EASI
If one of the derived EASI leads is in a technical alarm condition, a flat line is displayed. After 10 seconds, the directly acquired EASI AS, EA, or AI lead, depending on which is available, is displayed. Arrhythmia relearn is performed with transition to or from EASI Fallback monitoring using the available lead(s).
Relearning
Whenever there is a "Leads Off" condition for more than 60 seconds, the arrhythmia algorithm performs a Relearn using the available leads.
ECG and Arrhythmia Monitoring 5-21
Page 94
Warning
Since Relearn happens automatically, if learning takes place during ventricular rhythm, the ectopics can be incorrectly learned as the normal QRS complex. This can result in missed detection of subsequent events of V-Tach and V-Fib. For this reason, you should:
1 Respond promptly to any technical alarm. 2 Ensure that the arrhythmia algorithm is labeling beats correctly.
5-22 ECG and Arrhythmia Monitoring
Page 95
ST/AR Arrhythmia Monitoring
ST/AR Arrhythmia Algorithm
Indications for Use
The ST/AR Arrhythmia Algorithm is indicated for use in instances where the clinician decides to monitor cardiac arrhythmias of adult and pediatric patients and/or the ST segment of adult patients to gain information for treatment, monitor the adequacy of treatment, or to exclude causes of symptoms.
How the ST/AR Algorithm Works
ST/AR multi-lead analysis is performed on the user-selected primary and secondary leads. If only one lead is available for multilead, ST/AR analysis is performed on the single available lead.
Arrhythmia analysis consists of several steps:
1 The ECG signal is pre-processed to filter out baseline wander, muscle
artifact, and signal irregularities. In addition, if the Paced status = Yes, pace pulses are detected then rejected from the processing to avoid seeing them as QRS beats.
2 Beat detection to locate the QRS complexes for further analysis. 3 Feature measurement such as R-wave height, width, and timing. 4 Beats classification. Templates are created and are matched to incoming
beats, and the appropriate beat label is determined.
5 Rhythm and alarm detection. Beat labels are used to produce the values
and events needed to generate rhythms and alarms.
Working in parallel with beat detection and classification, a separate detector examines continuously for ventricular fibrillation, asystole, and noise.
The quality of the ECG signal is important for accurate arrhythmia analysis. The section below provides guidelines for optimizing signals for arrhythmia analysis.
For additional information on the ST/AR Algorithm, refer to the Arrhythmia Monitoring ST/AR Algorithm Application Note, #453564115631.
ECG and Arrhythmia Monitoring 5-23
Page 96
Alarms with Basic Arrhythmia Option
Additional Alarms with Enhanced Arrhythmia Option
***Asystole
**Afib
***Ventricular Fibrillation/Tachycardia
**Supraventricular Tach ***Extreme Bradycardia
**Missed Beat
***Extreme Tachycardia
**Pause
Aberrantly-Conducted Beats
As P-waves are not analyzed, it is difficult and sometimes impossible for the monitor to distinguish between an aberrantly-conducted supraventricular beat and a ventricular beat. If the aberrant beat resembles a ventricular beat, it is classified as ventricular. You should always select a lead where the aberrantly-conducted beats have an R-wave that is as narrow as possible to minimize incorrect calls. Ventricular beats should look different from these 'normal beats'. Instead of trying to select two leads with a narrow R-wave, it may be easier to just select one lead and use single lead arrhythmia monitoring. Extra vigilance is required by the clinician for this type of patient.
Intermittent Bundle Branch Block
Bundle branch and the other fascicular blocks create a challenge for the arrhythmia algorithm. If the QRS during the block changes considerably from the learned normal, the blocked beat may be incorrectly classified as ventricular, causing false PVC alarms. You should always select a lead where the bundle branch block beats have an R-wave that is as narrow as possible to minimize incorrect calls. Ventricular beats should look different from these 'normal beats'. Instead of trying to select two leads with a narrow R-wave, it may be easier to just select one lead and use single lead arrhythmia monitoring. Extra vigilance is required by the clinician for this type of patient.
ECG and Arrhythmia Alarm Overview
The ECG and arrhythmia alarms available depend on the arrhythmia option enabled for your MX40.
Basic arrhythmia alarms are listed below along with the alarms provided with the optional Enhanced Arrhythmia.
To check your enabled settings, view the Standby screen.
5-24 ECG and Arrhythmia Monitoring
Page 97
Alarms with Basic Arrhythmia Option
Additional Alarms with Enhanced Arrhythmia Option
***Ventricular Tachycardia
**Irregular HR
**High heart rate
**Ventricular Rhythm
**Low heart rate
**Run PVCs High
**Pacer Not Capture (if Pacing set to On)
**Pair PVCs
**Pacer Not Pacing (If Pacing set to On)
**R-on-T PVCs
**PVCs/min HIGH (PVC > limit/min)
**Ventricular bigeminy **Ventricular trigeminy
**Non-sustain VT
**Multiform PVCs
1. Extreme Brady Limit
2. Low Limit
3. High Limit
4. Extreme Tachy Limit
5. Extreme Brady (D value)
6. Extreme Tachy (D value)
Note — The ST/AR Arrhythmia Algorithm operates and generates alarms
independently at both the MX40 and the Information Center. Therefore, you may see slight differences between the two, even though the alarm settings and limits are the same.
Using ECG Alarms
At the Information Center, ECG alarms can be switched on and off and the high and low alarm limits changed just like other measurement alarms, as described in the Alarms chapter. Special alarm features which apply only to ECG are described here.
Extreme Alarm Limits for Heart Rate
The extreme rate alarms, Extreme Tachy and Extreme Brady, are set at the Information Center by adding a set value (the D value) to the high and low alarm limits.
ECG and Arrhythmia Monitoring 5-25
Page 98
You need to know which value has been configured for your monitor. Changing the high and low alarm limits automatically changes the extreme alarm limits within the allowed range.
Arrhythmia Alarm Settings
Some arrhythmia alarms can be turned off at the Information Center. They are:
Non-Sustain VT, Vent Rhythm, Run PVCs, Pair PVCs, R-On-T PVC, V.Bigeminy, V.Trigeminy, Multif.PVCs, Pause, SVT, HR High, HR Low Irregular HR, Missed Beat, PVCs/min, Pacer Not Capture, Pacer Not
Pacing, Asystole and Afib.
It is also possible to turn all yellow arrhythmia alarms off at the Information Center.
Alarms that have been turned off at the Information Center will appear as off in the Arrhythmia menu of the MX40, but they are not accessible, nor can you change limits locally.
Yellow Arrhythmia Alarms
Yellow arrhythmia alarms are short yellow alarms specific to arrhythmia-related patient conditions. Depending on your Information Center configuration, they may be shown with one or two stars. The heart rate alarms (High HR and Low HR) can be configured as short yellow or standard yellow alarms. When they are standard yellow alarms they exist independently of the other arrhythmia alarms and no timeout periods apply.
Warning
When arrhythmia analysis is on, all yellow ECG and arrhythmia alarms are short yellow alarms (one-star). This means that the alarm tones (if volume is on) are active for six seconds only, after which the blinking numeric and the alarm message remain for up to three minutes. The only exception to this are the HR High and Low alarms which can be configured as standard yellow alarms. Red alarms behave as usual.
Viewing Arrhythmia Waves
To review arrhythmia beat labels:
1 Go to the Setup ECG menu.
5-26 ECG and Arrhythmia Monitoring
Page 99
2 Select Arrhythmia. 3 Change Annotate Arrhy from Off to On. Beat labels will be annotated
above the ECG wave and Delayed will appear beside it.
To return to the normal ECG primary lead display:
1 Select Annotate Arrhy. 2 Change to Off. 3 Exit from the Setup ECG menu.
Arrhythmia Beat Labels
Arrhythmia beat labels tell you how the monitor is classifying beats.
N = Normal V = Ventricular Ectopic S = Supra-ventricular Premature P = Paced ' = Pacer spike " = Biventricular Pacer Spike L = Learning patient's ECG A = Artifact (noisy episode) ? = Insufficient information to classify beats I = Inoperative condition (e.g., LEADS OFF) M = Pause or missed beat
Learning
The arrhythmia system’s goal is to learn the patient’s normal complexes so
it can differentiate abnormal beats. This "learning" process uses the 15 first valid beats (for example, free from noise) encountered during the learning phase.
While the system is learning the complex, the delayed arrhythmia wave displays the beat label "L".
ECG and Arrhythmia Monitoring 5-27
Page 100
Learning Phase
A learning phase involves the system learning the patient’s dominant complexes. During a learning phase:
Alarm timeout periods are cleared. Stored arrhythmia templates are cleared. Asystole, Vfib, and HR alarms (when there are enough beats to
compute the HR) are active.
All other alarms are not active.
Single Lead Analysis
If single lead analysis is selected, the arrhythmia system begins learning whenever:
ECG monitoring is initiated. The Relearn key is activated (see Initiating Arrhythmia Relearning
Manually p. 5-29).
The ECG Lead or Lead Label is changed manually, or when Fallback
occurs (see ECG Fallback p. 5-21).
A "LEADS OFF" INOP condition (that has been active for >60 seconds)
ends.
When the MX40 re-associates with the Information Center.
Multilead Analysis
If multilead analysis is selected, the arrhythmia system begins a learning on both leads whenever:
ECG monitoring is initiated. The Relearn key is activated (see Initiating Arrhythmia Relearning
Manually p. 5-29).
There has been a Leads Off INOP condition (that has been active for >60
seconds) for both leads, and the condition ends in either lead.
When the MX40 re-associates with the Information Center.
Multilead Analysis With Changes in One Lead
Since the arrhythmia system uses more than one lead for analysis, if there is a change in one lead, the system does a relearn only on the affected lead. This happens whenever:
An ECG lead or label is changed.
5-28 ECG and Arrhythmia Monitoring
Loading...