Any comments about the documentation for this product should be addressed to:
User Assistance
PerkinElmer, Inc.
710 Bridgeport Avenue
Shelton, CT 06484
Or emailed to: info@perkinelmer.com
Notices
The information contained in this manual is subject to change without notice.
Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no
warranty of any kind with regard to this manual, including, but not limited to, the implied
warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in
connection with furnishing, performance or use of this material.
Copyright Information
This manual contains proprietary information that is protected by copyright.
All rights are reserved. No part of this file may be reproduced in any form whatsoever or translated into
The Pyris installation and hardware guide contains information and warnings that must be
followed by the user to ensure safe operation and to maintain the instrument(s) in a safe
condition.
Possible hazards that could harm the user or result in damage to the instrument are clearly
stated at appropriate places throughout this manual.
The following safety conventions are used throughout:
We use the term WARNING to inform you about situations that could result in
personal injury if precautions are not followed.
Nous utilisons le terme ATTENTION pour vous informer de situations pouvant
entraîner des dommages graves à l'instrument ou à d'autres équipements.
We use the term CAUTION to inform you about situations that could result in
serious damage to the instrument or other equipment.
Nous utilisons le terme AVERTISSEMENT pour vous informer de situations
pouvant entraîner des blessures corporelles si les précautions ne sont pas
respectées.
Page 7
Symbols Used on the Instrument . 7
AVERTISSEMENT: Consulter les documents d’accompagnement.
AVERTISSEMENT: Surface chaude.
AVERTISSEMENT: Risque de choc électrique.
Symbols Used on t he Ins trume nt
The symbols that appear on the instruments discussed in this installation are as follows:
CAUTION - Risk of Danger
Sections in this manual must be consulted to determine the nature of the
potential hazard and any actions which hav e to be tak en.
If there is a hazard of heat or electric shock, the following warning symbols will appear on the
instrument:
This symbol indicates the danger of burns, if precautions are not followed,
due to the presence of heat in this area of the instrument. There may also be
hot surfaces that can be easily touched.
This symbol means CAUTION – Cold surface.
Ce symbole signifie ATTENTION – surface froide.
This symbol indicates the danger of electric shock, if precautions are not
followed, due to the presence of high voltage.
Additional graphic symbols used on the instrument are the following:
Indicates alternating current.
Indicates the primary protective grounding terminal.
Indicates the off position of the main power switch.
Indicates the on position of the main power switch.
Page 8
8 . TGA 8000 Installation and Hardware Guide
, à
Electrical Warnings
CAUTION: The protection provided by this equipment may be impaired if the
equipment is used in a manner not specified by PerkinElmer.
ATTENTION: La protection fournie par cet équipement risque d'être moins
efficace si l'équipement fait l'objet d'une utilisation différente de celle mentionnée
par PerkinElmer.
CAUTION: Connect the instrument to an AC line power outlet that has a
protective ground connection. To ensure satisfactory and safe operation of the
instrument, it is essential that the protective ground conductor (the green/yellow
lead) of the line power cord is connected to true electrical ground. Any
interruption of the protective ground conductor, inside or outside the instrument,
or disconnection of the protective ground terminal may impair the protection
provided by the instrument.
ATTENTION: Raccordez l'instrument à une prise secteur ayant une connexion
de terre de protection. Pour assurer un fonctionnement satisfaisant et sûr de
l'instrument, il est essentiel que le conducteur de terre de protection (le
conducteur vert / jaune) du cordon d'alimentation secteur soit connecté à une
vraie masse électrique. Toute interruption du conducteur de terre de protection
l'intérieur ou à l'extérieur de l'instrument, ou la déconnexion de la borne de terre
de protection peut nuire à la protection fournie par l'instrument.
CAUTION: Do not operate the instrument with any covers or parts removed.
ATTENTION: Ne pas utiliser l'instrument avec des couvercles ou des pièces
enlevés.
CAUTION: Do not attempt to make adjustments, replacements, or repairs to this
instrument except as described in this manual. Only a PerkinElmer service
representative should be permitted to service the instrument.
ATTENTION: N'essayez pas d'effectuer des réglages, des remplacements ou
des réparations sur cet appareil, sauf comme décrit dans ce manuel. Seul un
représentant de service PerkinElmer devrait être autorisé à réparer l'instrument.
CAUTION: Use only fuses with the required current rating and of the specified
type for replacement.
ATTENTION: N'utiliser que des fusibles avec le courant nominal requis et du
type spécifié pour le remplacem ent.
Page 9
Electrical Safety . 9
NOTE:
Electrical Safety
This analyzer conforms to IEC publication 61010-1 (Safety requirements for electrical
equipment for measurement, control and laboratory use) as it applies to IEC Class 1 (earthed)
appliances, and therefore meets the requirements of the Low Voltage Directive 2006/95/EC.
Electrical Safety Guidelines
Service on electrical components should be performed only by a qualified PerkinElmer
service engineer.
Be sure the power cord is the correct one for your laboratory. The TGA 8000 power cord
must be rated at or better than 100/120 VAC at 15 A or 150 AC at 10 A. The line cord used
must meet the National Safety Agency's guidelines for the particular country.
Fuses
Service on electrical components should be performed only by a qualified PerkinElmer service
engineer. The fuses in the power e ntry module can be changed by the user.
Page 10
10 . TGA 8000 Installation and Har d ware Guide
Safety Precautions for the TGA 8000
CAUTION: The protection provided by this equipment may be impaired if the
equipment is used in a manner not specified by PerkinElmer.
ATTENTION: La protection fournie par cet équipement risque d'être moins
efficace si l'équipement fait l'objet d'une utilisation différente de celle mentionnée
par PerkinElmer.
WARNING: Be sure that all instrument operators read and understand the
following precautions. It is advisable to post a copy of these precautions on or
near the instrument itself.
AVERTISSEMENT: Assurez-vous que tous les opérateurs de l'instrument lisent
et comprennent les précautions suivantes. Il est conseillé de poster une copie de
ces précautions sur ou près de l'instrument lui-même.
WARNING: The glass parts on the TGA 8000 are fragile and may be subject to
breakage if not handled carefully.
AVERTISSEMENT: Les pièces en verre du TGA 8000 sont fragiles et risquent
d'être cassées si elles ne sont pas manipulées avec soin.
The following precautions must be observed when using the TGA 8000 Thermogravimetric
Analyzer:
•Never turn the computer off until the following message appears: It is now safe to shut off your computer.
•Never press the Reset button on the computer if the Pyris software appears to
malfunction. Press the Ctrl–Alt–Del keys simultaneously and select the Task
Manager. From the Task Manager, close the Pyris software.
•Never remove the outer instrument cover without shutting down the instrument and
disconnecting its power cord from the power source.
•Only high-quality purge gases should be used. Minimum purity of 99.9% is
recommended. A high-quality filter dryer accessory is recommended for the removal
of any moisture from the purge gases.
•Always observe the proper startup and shutdown procedures with the TGA 8000 and
all related instruments.
•Always maintain a positive purge flow (20 – 100 cc/min) through the system gas
chamber at all times. An inert gas such as nitrogen or argon is recommended as a
balance purge.
•Do not raise the furnace when the Pyrex furnace tube has been removed. This will
cause severe damage to the furnace and other components of the analyzer.
•Prior to performing any experiment or calibration procedure, be sure that the
thermocouple currently installed is functioning properly. The thermocouple can be
checked by programming the TGA 8000 to 100°C and checking the temperature
displayed in the status panel. The temperature should be close to 100°C if the
instrument is calibrated.
Page 11
Safety Precauti ons for the TGA 80 00 . 11
Rated Power
Supply voltage
100 – 240VAC
Supply voltage frequency
50/60Hz
Glass Cartridge, 0.213 "Dia x 0.886" L
(5.40mm x 22.50)
Safe temperature range
10 – 40°C
45.8 cm (18-in)
Weight
26kg with autosampler (57lbs)
•To prolong the life of the standard furnace, it is recommended that the automatic
furnace cleaning procedure be performed routinely such as every 5 to 10 runs.
Pyrolysis products should be burned off after every analysis that involves pyrolysis of
a sample in the absence of air or oxygen.
•Always close the furnace door before attempting to run an analysis. When the TGA
8000 standard furnace is heating, keep the protective plastic shield in the down
position. This is for non-autosampler instruments. This will protect you from
accidentally touching the fur nace wh en it is hot.
•Do not attempt to move the TGA 8000 while it is on. Wait until the power is off and
the furnace cooling fan has stopped spinning before moving the instrument.
• Do not remove cables from the analyzer while it is on.
• Do not touch the opening on the furnace lift mechanism or insert objects into the opening.
• Do not use hydrogen gas with the TGA 8000 since it is highly explosive.
• When cleaning the instrument, consult PerkinElmer if there is any doubt about the
compatibility of decontamination or cleaning agents with parts of the equipment or
with material contained in it.
Specifications
Maximum power consumption
Fuses
Dimensions L x H x D
300W
6.3AT
45.8 cm (18-in) x 48.3 cm (19-in) x
Page 12
12 . TGA 8000 Installation and Har d ware Guide
Warnings and Labels
Warning Label on the Upper Ring
WARNING: Grounding circuit continuity is vital for the safe operation of
equipment. Never operate equipment with the grounding connector disconnected.
Disconnect supply cord before operating.
AVERTISSEMENT: La continuité du circuit de mise à la terre est essentielle au
bon fonctionnement de l'équipement. Ne jamais faire fonctionner l'équipement
avec le connecteur de mise à la terre débranché. Débrancher le cordon
d'alimentation avant de l'utiliser.
Page 13
Warning Label behind the Furnace
CAUTION – HOT SURFACE which indicates that the furnace at the center of the
ring may be hot.
ATTENTION - SURFACE CHAUD qui indique que le four au centre de l'anneau
peut être chaud.
Warnings and Labels . 13
Page 14
14 . TGA 8000 Installation and Har d ware Guide
Warning Labels on the Back of the TGA 8000
WARNING: Alwayskeep airflow unobstructed.
To maintain adequate ventilation, do not block the back of the instrument.
AVERTISSEMENT: Toujours garder l'écoulement d'air dégagé.
Pour maintenir une ventilation adéquate, ne pas bloquer le dos de l'instrument.
WARNING: For protection against fire hazard replace only with the same type
and rating of fuse.
AVERTISSEMENT: Pour la protection contre les risques d'incendie, remplacer
uniquement par le même type et la même classe de fusible.
WARNING: Grounding circuit continuity is vital for the safe operation of
equipment. Never operate equipment with the grounding connector disconnected.
Disconnect supply cord before operating.
AVERTISSEMENT: La continuité du circuit de mise à la terre est essentielle au
bon fonctionnement de l'équipement. Ne jamais faire fonctionner l'équipement
avec le connecteur de mise à la terre débranché. Débrancher le cordon
d'alimentation avant de l'utiliser.
WARNING: Toxic Gases – Fume Ventilation System
Without adequate ventilation, potentially toxic vapors can build up in the
laboratory. Your laboratory must have a reliable fume ventilation system before
you use this instrument.
AVERTISSEMENT: Gaz toxiques - Système de ventilation des fumées
Sans ventilation adéquate, des vapeurs potentiellement toxiques peuvent
s'accumuler en laboratoire. Votre laboratoire doit disposer d'un système fiable de
ventilation des fumées avant d'utiliser cet instrument.
Page 15
Environmental Requirements for the TGA 8000 . 15
Environmen tal R equi reme nts fo r t he TGA 800 0
The TGA 8000 has been designed for indoor use only and should not be operated in an explosive
environment. The TGA 8000 operates most efficiently under the following conditions:
•Ambient temperature range of 10°C – 35°C (50°F – 95°F). The instrument will remain
safe when operated between 5°C and 40°C (41°F to 104°F).
• Ambient relative humidity between 20% and 80% without condensation.
• Altitude no more than 2000 m.
• Clean area free of dust, smoke, vibration, and corrosive fumes.
• On a bench area out of direct sunlight.
• Away from heating or cooling units or ducts.
• The instrument must be positioned so the power switch on the rear panel can be
easily reached by the operator.
• There must be an adequate and stable power source for all system components.
• The TGA 8000 is a highly sensitive, precision laboratory instrument. The environment
in which it is used must be free of radio frequency interference sources which may
affect the performance of the instrument. If performance is affected by RF fields, the
instrument should be reoriented or relocated, or separated by a greater distance from
the interfering source. Consult PerkinElmer for additional assistance.
WARNING: If the heating of materials could lead to the liberation of hazardous
gases, the use of a fume extraction system will be required.
AVERTISSEMENT: Si le chauffage des matériaux peut entraîner la libération de
gaz dangereux, il sera nécessaire d'utiliser un système d'extraction des fumées.
WARNING: There may be possible hazards of explosion, implosion, or the
release of toxic or flammable gases arising from the materials being heated.
AVERTISSEMENT: Il peut y avoir des risques d'explosion, d'implosion ou de
dégagement de gaz toxiques ou inflammables provenant des matières chauffées.
WARNING: Do not mount the instrument of a surface of flammable material.
AVERTISSEMENT: Ne pas monter l'instrument d'une surface de matériau
inflammable.
CAUTION: Before using any cleaning or decontamination method, except that
recommended in this guide, the user should check with PerkinElmer Instruments
that the proposed method will not damage the instrument.
ATTENTION: Avant d'utiliser toute méthode de nettoyage ou de décontamination,
à l'exception de celle recommandée dans ce guide, l'utilisateur doit vérifier auprès
de PerkinElmer Instruments que la méthode proposée n'endommagera pas
l'instrument.
Page 16
16 . TGA 8000 Installation and Har d ware Guide
Hazardous Chemical Warnings
Before using any chemicals or solvents with the instrument, you should be thoroughly familiar
with all hazards and safe handling practices. Observe the manufacturer's recommendations
for use, storage, and disposal. These recommendations are normally provided in the material
safety data sheets (MSDS) supplied with the solvents.
We use the term CAUTION to inform you about situations that could result in
serious damage to the instrument or other equipment.
Nous utilisons le terme AVERTISSEMENT pour vous informer de situations
pouvant entraîner des blessures corporelles si les précautions ne sont pas
respectées.
WARNING: Some chemicals used with this instrument may be hazardous or
may become hazardous after completion of an analysis. The responsible body
(e.g., Lab Manager) must take the necessary precautions to ensure that the
surrounding workplace and instrument operators are not exposed to hazardous
levels of toxic substances as defined in the applicabl e Mater ial Saf et y Data Sheet s
or OSHA, ACGIH, or COSHH documents. Venting for fumes and disposal of waste
must be in accordance with all national, state, and local health and safety
regulations and laws.
AVERTISSEMENT: Certains produits chimiques utilisés avec cet instrument
peuvent être dangereux ou risquent de devenir dangereux une fois l'analyse
terminée. L'organisme responsable (p. Ex., Le responsable du laboratoire) doit
prendre les précautions nécessaires pour s'assurer que le lieu de travail
environnant et les exploitants d'instruments ne sont pas exposés à des niveaux
dangereux de substances toxiques tels que définis dans les fiches de données de
sécurité ou les documents OSHA, ACGIH ou COSHH. L'évacuation des fumées et
l'élimination des déchets doivent être conformes à toutes les réglementations et lois
nationales, nationales et locales en matière de santé et de sécurité.
Some definitions of terms used above are:
•Responsible body: "Individual or group responsible for the use and maintenance of
equipment, and for ensuring that operators are adequately trained" [per IEC 1010-1,
Amendment 2].
•Operator: "Person operating equipment for its intended purpose" [per IEC 1010-1,
Amendment 2].
• OSHA: Occupational Safety and Health Administration (United States)
• ACGIH: American Conference of Governmental Industrial Hygienists
• COSHH: Control of Substances Hazardous to Health (United Kingdom)
Installation Category
This instrument is able to withstand transient overvoltage according to Installation Category II
as defined in IEC 1010-1.
Page 17
Environmental Requirements for the TGA 8000 . 17
Pollution Degree 2
This product will operate safely in environments that contain nonconductive foreign matter up
to Pollution Degree 2 in EN/IEC 61010-1.
Normally only non-conductive POLLUTION occurs. Occasionally, however, a temporary
conductivity caused by condensation must be expected.
South Korea
Ingress Protection
The TGA 8000 is rated IP20 as defined in IEC 60529.
Storage Conditions
The TGA 8000 may be stored under the following conditions:
• ambient temperature from –20°C to +60°C (–4°F to +140°F)
• ambient relative humidity from 20% to 80% noncondensing
• altitude in the range of 0 – 12,000 m
Page 18
18 . TGA 8000 Installation and Har d ware Guide
WEEE Instructions for PerkinElmer Pro ducts
or
A label with a crossed-out wheeled bin symbol and a rectangular bar indicates that the
product is covered by the Waste Electrical and Electronic Equipment (WEEE) Directive and is
not to be disposed of as unsorted municipal waste. Any products marked with this symbol
must be collected separately, according to the regulatory guidelines in your area.
The objectives of this program are to preserve, protect and improve the quality of the
environment, protect human health, and utilize natural resources prudently Requirements for
waste collection, reuse, recycling, and recovery programs vary by regulatory authority at your
location. Contact your local responsible body (for example, your laboratory manager) or
authorized representative for information regarding applicable disposal regulations. Contact
PerkinElmer at the web site listed below for information specific to PerkinElmer products.
Web address:
http://www.perkinelmer.com/WEEE
For Customer Care telephone numbers select “Contact us” on the web page.
Products from other manufacturers may also form a part of your PerkinElmer system. These
other producers are directly responsible for the collection and processing of their own waste
products under the terms of the WEEE Directive. Please contact these producers directly
before discarding any of their products.
Consult the PerkinElmer web site (above) for producer names and web addresses.
Page 19
Prepare th e Lab orat ory
19
Page 20
20 . TGA 8000 Installation and Har d ware Guide
Overview
Thermogravimetry is a branch of thermal analysis that examines the mass change of a
sample as function of temperature or time as it is subjected to a controlled temperature
program in a controlled atmosphere. Not all thermal events cause a change in the sample
mass (for example, melting, crystallization, glass transition), but there are some important
exceptions such as desorption, a bsorption, oxidation, decomposition. Thermogravimetry
characterizes the decomposition and thermal stability of materials under a variety of
conditions and examines the kinetics of the physico-chemical processes occurring in the
sample.
The TGA 8000 is a computer-controlled analyzer that has been designed to take advantage
of the enhanced features of Pyris Software. It is made up of two major components: a
microbalance and a furnace element.
About the Microbalance
The microbalance of the TGA 8000 operates as a high gain electromechanical servo system that
permits detection of weight changes as small as 0.1 g, with a maximum capacity of 1300 mg.
The null balance design uses a servo-controlled torque motor to automaticall y compensate for
weight changes in the sample material. When a sample is placed in a sample pan and the pan is
on the hangdown wire, the beam that supports the sample pan deflects. A beam position
detector measures the deflection with an optical sensor and uses current to return the beam to
its original position. The amount of current necessa ry to maintain the system in the "null" state
is directly proportional to the weight change in the sample. The current is amplified and filtered
and then displayed in the display panel of the instrument in mg.
About the Furnace
The furnace allows operation of the TGA 8000 from subambient to 1200°C. The furnace uses
a microfurnace that has a low thermal mass for quick cooling and equilibration. Accurate
program temperature control reduces overshoot to 0.2°C at 100°C/minute which results in
stepwise analysis precision. The sample temperature is sensed by a chromel–alumel
thermocouple.
Features
Some of the features of the TGA 8000 are:
1. Faster furnace cool-down with the cooling fan above the furnace instead of underneath.
This creates positive upward ventilation of the furnace during cool-down.
2. iPad control.
3. The sample gas interface has been repositioned for easier access.
4. Controlled balance environment gives the TGA 8000 long-term ordinate stability. The
weighing balance is housed in a thermally-isolated enclosure that eliminates thermal
gradient effects.
5. An anti-convection iris prevents decomposition products and heat from the furnace
from backstreaming into the balance chamber.
6. The "quick release" furnace tube makes furnace tube cleanup easy. The furnace tube
is attached to the furnace base by a clamp.
7. Temperature control of the segmented furnace is optimized by the heater/sensor
furnace technology.
Page 21
Overview . 21
8. The removable quartz furnace chamber sleeve regulates furnace convection by
reducing the effective furnace volume. The reduced volume also enhances gas
switching time because there is less furnace volume to displace.
9. The antistatic device creates an ion stream which is an invisible curtain of charged
particles that surrounds the sample loading area. This minimizes static cling betwee n
the sample pan and the furnace wall dur ing sample loading and static drift during
analysis. Loading fine, powdered, or static-sensitive samples is no longer a problem.
The device can be controlled from the control pa nel in the software.
In addition to these features, there is an autosampler designed for both the R&D laboratory
and the quality control/quality assurance environment. The autosampler increases the speed
of analysis by being able to hold up to 48 samples ready for analysis. The autosampler
enables samples to be prepared and run during the same shift. Using the Pyris Player feature
of Pyris Software, you can create play lists that will run the samples in the autosampler tray
and analyze the data automatically.
The following sections describe requirements for your Thermal Analysis System. Make sure
your laboratory meets all of the requirements before you try to install the system. You should
step through the topics in the order presen ted be lo w. You can get from one topic to the next
by either returning to this topic or by clicking on a "continue" hotspot at the end of the topic.
Operating Variables in the TGA 8000
In thermogravimetric analysis, the object of an experiment is to record the change in weight
of a sample as function of temperature or time using a constant heating rate or a more
complex temperature program. Below are some techniques that can be used to obtain the
most accurate data in a minimum amount of time with a TGA 8000.
Sample Preparation
The TGA 8000 can analyze solid or liquid samples. Solid samples can be in the form of film,
crystal, or grains. In some cases, you may want to chop or grind the sample to create a large
surface area that is exposed to the purge gas atmosphere. In other cases, larger chunks or
sections of sample may be placed directly into the sample pan or crucible for analysis.
Sample Size
The size of sample that can be analyzed ranges from <1 mg to 400 – 500 mg. In most cases,
the typical sample size for TGA analysis is 2 – 50 mg. The type of transition or reaction that
you expect to occur in the sample should dictate the amount of sample used. For example, in
the case of polymer decomposition studies, sample sizes of 2 – 15 mg are recommended
since very large weight loss values are associat ed with this type of experiment. If there will
be a very small weight loss such as with wat er loss or solvent evaporation, larger samp le
sizes should be used. It is not uncommo n to use samples as large as 40 – 50 mg.
Temperature Range
The temperature used for your experiment will depend on the type of samples to be run as
well as the specific application performed. The standard furnace operates from subambient
to 1200°C with a temperature precision of ± 2°C. This furnace has broad applications in the
polymer, pharmaceutical, and organic chemical industries.
Because of the unique design of the TGA 8000, you can perform controlled heating or cooling
experiments as well as constant temperature experiments as a function of time. You can
even perform multiple-step experiments using any combination of heating, cooling, or
isothermal segments, thereby allowing the performance of virtually any thermogravimetric
Page 22
22 . TGA 8000 Installation and Har d ware Guide
NOTE:
application. In addition, there is also an AutoStepwise Scan available in the method program.
This scan heats the sample rapidly between weight loss regions, and slows down or stops
heating during rapid weight loss regions.
Scanning Rate
Scanning rates range from 0.1°C/min to 500°C/min for the standard furnace. The exact
scanning rate used will depend on the experiment and the end result you are trying to
achieve. For example, most TGA experiments are performed at heating rates of 5°C/min to
50°C/min. However, there may be times when you want to heat or cool rapidly to a selected
temperature and then hold isothermally or scan a controlled rate. In such cases, very fast
heating or cooling rates (i.e., 100°C/min or 200°C/min) are typically used to quickly increase
or decrease the sample temperature.
Sample Atmosphere
WARNING: Due to the highly explosive nature of hydrogen, it is recommended
that it not be used with this instrument.
AVERTISSEMENT: En raison de la nature très explosive de l'hydrogène, il est
recommandé de ne pas l'utiliser avec cet instrument.
The atmosphere to which the sample is exposed is carefully controlled by the selection of the
sample purge or sample gas and flow rate. Recommended sample gases are nitrogen, argon,
helium, carbon dioxide, air, oxyge n, or other inert or reactive gases, as long as they do not
react with the fur nace materials, thermocouple, or other analyzer components in which gases
come into contact. Analyses are done at normal pressure or at reduced pressures.
In addition, sample gases may be switched at any point during an analysis, for example, the
pyrolosis of a polymer occurs while in an inert atmosphere (Sample Gas A) then switching to
an O2 or Air to oxidize the remaining components ( Gas B), by using the internal gas
switching capabilities which accurately control the flow rates of both gasses.
Before starting installation, please re ad the Safety and Regulatory Information.
Cooling Device
Chiller: -20°C and above
Electrical Requir ements
Power Source: An independent power source should be provided for the system,
including the computer. The power source should not be associated with heavy-duty
equipment suc h as lar ge motors, or with possible sources of high-frequency
interference such as photocopying systems, discharge lamps, or radio transmitters.
The power supply should be fused at a max imu m of 20 A (12 0 V sy ste ms) o r 16 A
(200 240 V systems ).
Line Voltage: PerkinElmer analyzers and their associated instruments are designed
to operate within a line voltage range of 10% of the nameplate voltage or other such
voltage selected at installation to suit the particular country or region. (The range for
240 V systems is + 6%, –1 0%.) The supply must be smooth, clean, earthed and free
of transient voltages over 40 V.
The frequency range is +/-1% for 50 Hz and 60 Hz systems.
Instrument Maximum Power Requirements:
TGA 8000: 100 - 2 40 +/ - 10%, 50/60 Hz. +/- 1%
Page 23
Order Spares and Accessories . 23
NOTE:
Purge Gas and Pneumatic Supply Requirements
Minimally, the TGA 8000 requires a System P urge gas, also known as the balance purge. This
gas must be clean and dry. System gas is controlled by a mass flow controller resident in the
analyzer. Pressure, between 2-3bar is applied at the connector at the rear of the analyzer
labeled “System Gas”. Typically this is an inert, non-reactive gas i.e. N
such as Air or O
can also be used.
2
A Sample Purge gas is not a requirement to operate the TGA, however it is recommended.
There are two inputs, A and B for the sample gas which are also controlled by a separate mass
flow controller, allowing for gas switching exp eriments. Typically, “Sample Gas A” is the same
gas that is purging the “System” (Bal ance) and “Sample Gas B” would be Air or O2. Switching
gases at elevated temperatures facilitates the burning of the sample and aids in the cleaning of
the furnace and sam ple pans.
The recommended purge gas for all Thermal Analyzers at ambient temperatures is argon or
nitrogen with a minimum purity of 99.9%. Other gases, such as air or oxygen, may also be used.
The purge gas for any instrument must be dry. Use a size 1A cylinder equipped with a pressure
regulator that has a shutoff valve at the outlet. The shutoff valve should have 1/4-in. NPT male
threads on the outlet side for connection to the analyzer’s purge gas line.
CAUTION: Maximum pressure should not exceed that listed in the chart below.
ATTENTION: La pression maximale ne doit pas dépasser celle indiquée dans le
tableau ci-dessous.
, Ar, or He. Gases
2
Gas Pressure Flow
Purity
(Minimum)
System Gas:
Argon, nitrogen, helium, air or
oxygen
Sample Gas:
Argon, nitrogen, helium, air or
oxygen
Reactive Gas: Argon,
nitrogen, helium, air or oxygen
29- 43.5 psi (23 bar)
29- 43.5 psi (23 bar)
Unpressurized
0-200 cc/min 99.95
0-100 cc/min 99.95
0-100 cc/min
(Controlled with restrictor
at the source gas)
99.95
Install a Filter Dryer
The filter dryer is an optional accessor y that is used to remove moisture, carbon dioxide, and
other impurities from your purge gas or any other gas used with a particular analyzer. Refer
to Ordering Spares and Accessories, or contact your local PerkinElmer sa les representative
for assistance when ordering this accessory.
Space Requi rement s
Prepare a clean bench space large enough to accommodate all system components and
accessories. Click on the appropriate analyzer below to display the dimensions of a
component. For computer and printe r dimensions, refer to the manufacturers’ instruction s
that were included with them.
Before the instrument is installed, the area around, under and behind the instrument’s
planned location is to be cleared of any dirt and dust to prevent their entry into the
instrument’s interior, which could adversely affect performance.
Once all of the above requirements are met, you can install your analyzer.
Page 24
24 . TGA 8000 Installation and Har d ware Guide
Order Spares and Accessories
Supplies, accessories, and replacement parts can be ordered directly from PerkinElmer.
www.perkinelmer.com/supplies, PerkinElmer's web based catalog service, offers a full
selection of high-quality thermal analysis supplies.
To place an order, request a free catalog, or ask for information:
•If you are located within the U.S., call toll free 1-800-762-4000, 8 a.m. to 8 p.m.
EST. Your order will be shipped promptly, usually within 24 hours.
•If you are located outside of the U.S., call your local PerkinElmer sales office.
Page 25
Instal ling the TGA 8000
25
Page 26
26 . TGA 8000 Installation and Har d ware Guide
TGA 8000 Installation Summary
The installation procedure for the TGA 8000 comprises the following steps:
• Prepare the Laboratory
• Unpack the TGA 8000
• Set Up the TGA 8000 System Components
• Connect the TGA 8000 System Components
• Configure the TGA 8000
• Install the Furnace Hangdown Wire and Balance Tare Weight
• Calibrate the TGA 8000
• Install the Cooling Jacket Accessory or the Accupik Accessory (if applicable)
Instructions for installing a TGA 8000 with an autosampler are included where appropriate.
Page 27
Unpack the TGA 8000 . 27
Unpack the TGA 8000
Each component of the TGA 8000 system comes in its own box. The system components are:
• TGA 8000
• Personal Computer
• Printer (optional)
• Installation Kit containing the Spares Kit
CAUTION: Take care when unpacking the TGA. Some of the boxes packed with
it contain fragile parts that can break easily.
ATTENTION: Faites attention lors du déballage du TGA. Certaines des boîtes
emballées avec elle contiennent des pièces fragiles qui peuvent se casser
facilement.
The autosampler is used to automate the running of samp les. If you ordered a TGA with an
autosampler, it is shipped with the autosampler already installed.
When you unpack the box, check all the parts included against the shipping list. If any part is
damaged or missing, contact the shipping company and PerkinElmer immediately.
The TGA is shipped in a large box that contains the analyzer and a box that contains the
Install Kit (P/N N5320120). In the Install Kit is the Spares Kit (P/ N N5320026).
To unpack the TGA main box:
1. Place the instrument carton on the floor in front of the bench on which the
instrument will be placed. Open the box.
2. Remove the box from the packing, then remove the autosampler tray and furnace
door and set them aside.
Page 28
NOTE:
28 . TGA 8000 Installation and Har d ware Guide
The furnace shield is shipped in the inverted position on the furnace door.
3. Unscrew and remove the furnace shield from the furnace door, invert it then secure
it on the furnace door.
4. Remove the top foam piece that secures the instrument in the box.
5. Prepare to lift the instrument out of the carton.
WARNING: Since the instrument weighs 26 kg (57 lb), two people are needed to
safely do this. The bench should be nearby. DO NOT lift the unit by the top round
dome. This dome protects the balance mechanism and is not screwed down.
AVERTISSEMENT: Puisque l'appareil pèse 26 kg (57 lb), deux personnes sont
nécessaires pour le faire en toute sécurité. Le banc doit être à proximité. NE soulevez PAS l'appareil par le dôme supérieur. Ce dôme protège le mécanisme
de l'équilibre et n'est pas vissé.
6. One per son should put one hand underneath the bottom of the instrument and the
other hand beneath the front side. The other person should place one hand
underneath the top of the instrument and the other hand underneath the back side.
Lift the instrument simultaneously.
7. Place the analyzer carefully on the prepared bench.
8. Remove the tape on top of the dome.
Gently lift the dome cover STRAIGHT UP to ensure you clear the balance mechanism.
9. Remove the tape securing the balance and the zip tie securing t he furnace.
10. If there is an autosampler attached, cut the strap holding it in place as shown to the
right of the image below).
The procedure for unpacking the TGA is complete.
Page 29
Set Up TGA 8000 System Components . 29
Set Up TGA 8000 S ys tem Co mpo nen ts
Setting up the TGA 8000 system components consists of the following steps:
• Install the Purge Gases
• Level the TGA 8000
Install the Purge Gase s
The TGA 8000 has gas inlets at the rear of the analyzer, (Sample Gas A and B, System
(Balance) Gas, and Reactive Gas. Having two separate purges allows the flow rate of the
purge gas in the balance area to be greater than that in t he furnace tube area. This creates
positive pressure that helps prevent reactive gases from entering the balance area. Using
both a balance purge and a sample gas allows you to use reactive gases for various analyses.
This System gas supply is used for the a ntistatic device. Finally, there is an inlet on the rear
panel to connect Reactive gas that enters the sample area directly.
Overview of t he Purge Fl ow through the TGA 8000
The balance purge gas enters the back of the analyzer and then the cavity under the balance
cover. It then flows down the Pyrex ball joint into the furnace tube between the tube and the
liner. It exits the side arm of the furnace tube into the exhaust tubing.
CAUTION: The system purge flow rate should always be greater than the sample
gas flow rate to prevent decomposition products from entering the balance
chamber. The recommended flow rate is 20 – 100 ml/min for the balance purge
and 10 – 50 ml/min for the sample gas.
ATTENTION: Le débit de purge du système doit toujours être supérieur au débit
de gaz échantillon afin d'éviter que les produits de décomposition ne pénètrent
dans la chambre d'équilibrage. Le débit recommandé est de 20 - 100 ml / min
pour la purge de l'équilibre et de 10 - 50 ml / min pour le gaz échantillon.
CAUTION: The total flow rate of the balance purge gas, sample gas, and reactive
gas should be below 200 ml/min.
ATTENTION: Le débit total du gaz de purge d'équilibrage, du gaz échantillon et
du gaz réactif doit être inférieur à 200 ml / min.
The Sample Gas enters through connectors (labeled Sample Gas A or Sample Gas B) on the
back of the TGA 8000. It is applied as pressure 1 - 3 Bar (optimum 2 Ba r). The software
controls the gas flow controller to the sample area. Gas is directed to the furnace through
Teflon tubing from the flow controller to an inlet port on the bottom of the Balance mounting
plate. From here it flows through a chamber out of the plate and above the anti-connection
tube into the furnace area..
You can use the same gas source (house line or cylinder) for both balance system purge gas
and sample gas by using a serial pressure gauge se tup (P/N N5190462).
Connecting the System Purge and Sample Gas Lines to the Gas Supply
To connect the system purge and sample gases, using separate sources, you will need parts
included in the Spares kit (P/N N5320120).
Page 30
30 . TGA 8000 Installation and Har d ware Guide
NOTE:
1. Connect suitable regulators to the gas supplies that will be used to purge the system
(the customer is responsible for supplying the regulators). The regulators must be
equipped with shut off valves at the o utlet. The shutoff valves must have a 1/4-in.
NPT male thread on the outlet side.
2. Tape the threads of the outlet fitting on both shutoff valves with the thread sealant
tape provided (P/N 09908134).
3. Connect the end of a female connector (P/N 09903196) to the taped male end of the
shut off valve.
4. If you did not install a filter dryer into the purge gas or sample gas line, click here.
Click on the Back but ton to return here.
If you wish to install the GMD 8000, click here.
5. Place a 1/8-in. nut and ferrules over one end of the 1/8-in. Teflon tubing (P/N
02506483).
6. Connect the 1/8-in. Teflon tubing and nut to the end of the NPT male connector on
the gas supply.
7. Repeat steps 4 – 6 for the othe r gas line.
Connecting the System Purge Gas Line
1. Place a 1/8-in. nut and ferrules over the other e nd of the 1/8-in. Teflon tubing which
is connected to the gas cylinder that is to be used for the System purge.
2. Connect the end of the Teflon tubing to the fitting labeled Balance Purge located on
the back of the TGA 8000. Tighten the nut fingertight and then, using a wrench
tighten a quarter turn past fingertight.
If you are splitting the System Purge to supply also Sample Gas A, cut a short length of 1/8-
in Teflon tube and connect as described above by using the T-splitter. Connect the line to
System and the GSA to splitter.
Connect the Sample Gas Line
1. Place a 1/8-in. nut and ferrules over the other end of the 1/8-in. Teflon tubing which
is connected to the gas cylinder that is to be used for the sample gas.
2. Connect the end of the Teflon tubing to the fitting labeled Sample Gas located on the
back of the TGA 8000. Tigh te n the nut fingertight and then, a wrench, tighten a
quarter turn past fingertight.
Connecting the System Purge and Sample Gases Using a Common
Source
Instead of using two gas cylinders containing the same gas or two house lines of the same
gas, you can use one source of gas for both the system purge gas and the sample gas, for
example, compressed air or nitrogen. This can be done by installing gauges in series off of
the regulator on the source. These gauges are P/ N N5190462. Detailed instructions on how
to put two or more gauges together prior to installing them on the regulator are included with
the gauges. You should also install the connectors for the tubing to the outlets of the gauges
before attaching the gauges to the source. The cylinder or house line should be prepared
with a regulator as described above. Attach the series of gauges to the outlet connector on
the regulator with the arrows on the gauges, indicating the direction of the flow of gas,
pointing away from the source.
Page 31
Set Up TGA 8000 System Components . 31
Exhaust Gas Tubing
3. Connect the barbed fitting (P/N 09920026) to the exhaust port on the rear panel.
4. Insert the end of a piece of 0.25-inch Tygon tubing and to the barbed fitting
connector on the back of the instrument labeled EXHAUST. The tube should be
placed beneath a hood vent if you expect to have noxious exhaust fumes. Otherwise,
it can just hang from the instrument.
WARNING: If the heating of materials could lead to the liberation of hazardous
gases, the use of a fume extraction system will be required.
AVERTISSEMENT: Si le chauffage des matériaux peut entraîner la libération de
gaz dangereux, il sera nécessaire d'utiliser un système d'extraction des fumées.
Connecting Reactive Gas
The TGA 8000 has a reactive gas assembly with which a reactive gas can enter the sample
area. The reactive gas enter through the rear of the TGA
When using corrosive reactive gas, connect it to a connector on the rear of the balance
dome. The assembly is underneath the balance dome and goes down into the upper furnace
tube (ball joint).
Connect the TGA 8000 Components
The computer and optional pr inte r should already be conne cted together and the Pyris
software already installed. To complete the connection of the system components, the TGA
8000 has to be connected to the computer.
CAUTION: Make sure all components are turned off and unplugged before you begin.
ATTENTION: Assurez-vous que tous les composants sont éteints et débranchés
avant de commencer.
Page 32
32 . TGA 8000 Installation and Har d ware Guide
The image below shows the rear panel connections on the TGA 8000.
1. Install the Pyris So ftware
2. Connect one end of the USB cable to the back of the TGA 8000.
3. Connect the other end of the USB cable to the computer.
IComputer will recognize new hardware, install the drivers from the Prgram
Files\PerkinElmer\Pyris\Drivers folder.
4. From device manager determine which comport is assigned to the TGA8000
5. Connect the printer cable, if applicable, to an unused USB port on the computer.
6. Connect the analyzer, printer and computer p ower cords to AC outlets.
7. Once the system connections are made, switch on the system in the following order:
• Computer
• TGA 8000
• Printer
8. Turn on the purge gas supply, and the water for the TGA 8000 cooling system, if
applicable, or any other accessories for the cooling device you are using.
Configure the TGA 8000
Highlighting the TGA 8000 in Pyris Configuration dialog and clicking Edit displays the
Instrument Configuration dialog. This contains fields for Name, Port, Serial Number,
Accessories.
Before acquiring data, the software must be configured for the analyzer(s) connected to the
computer. Follow the steps below to configure the TGA:
1. In the Device Manager determine the COM Port assigned to the TGA 8000.
2. Select Pyris Configuration from the Pyris group under PerkinElmer Applications
on the Programs menu accessed from the Start button or select Configure Analyzer from the Pyris Manager Start button men u. The Configuration dialog box
appears on the screen.
Page 33
Set Up TGA 8000 System Components . 33
The following is a description of the controls on the TGA 8000 Configuration
dialog:
Name: If you are adding a TGA 8000, the system will display a default name
in this field. Type the name you wish to assign to the analyzer, using a
maximum of eight characters. The name identifies the analyzer in Pyris
Software; it also appears in the Instrument button on the Pyris Manager panel
and in the title bar of the Instrument Application.
Port: Displays the COM port to which the analyzer is attached and that you
selected in the Add Analyzer dialog box. Bes t on Port s 1,2,3,4
Serial Number: Displays the serial number of the analyzer.
Accessories: Lists the available accessories for the analyzer. For the TGA
8000, the following accessories are available:
• Cooling Jacket Accessory
• Accupik
• GMD 8000.
Firmware Version: Click the Firmware Version button to display the version
of firmware in the analyzer.
3. Click the Add Analyzer button. The Add Analyzer dialog box appears.
When a TGA 8000 is connected to the PC and has been added to the Pyris software
using the Add An a lyz er d i alo g, it shou ld be ident if ied in the P yris Configuration
dialog. If more than one TGA 8000 has been added, then the second one needs a
unique name when being connected.
4. From the list of available ports, select the port to which you connected the analyzer.
The TGA's Configuration dialog box is displayed.
5. The system detects the type of analyzer that is attached and displays the default
name and other information for a TGA in this dialog box. Select the accessory, if any,
that is connected to the TGA 8000 (for example, GMD 8000, Accupik).
Page 34
34 . TGA 8000 Installation and Har d ware Guide
6. Click OK to accept the entries.
7. Click OK again to exit the Configuration dialog box or select Add Analyzer to add
another analyzer to the configuration. Remember that the analyzer must be
connected to the communications port and powered on, otherwise it will not be
recognized by the Pyris software.
Once Pyris Software has been configured for the TGA, the application can be started from
Pyris Manager.
Starting the Pyris Manager
1.Select Pyris Manager from the Pyris group under PerkinElmer Applications in
All Programs on the Start menu. The Pyris Manager is displayed on the screen.
2. Select TGA 8000 from the Pyris Manager. The TGA application loads and the Control
Panel is displayed.
If you have an autosampler, you should now go to Autosampler Adjustment.
Configure the iOS App for the TGA 8000
This procedure will help guide you in configuring the TGA 8000 for use with an iPad/iPhone.
The interface is designed so the Pyris Light app communicates with the TGA8000 via the PC
which is connected to your company’s network. It is important that you co nfirm with your
network administrator that the PC has access to the network and that the iOS device can
connect to the network via WiFi. There is no direct co m munica ti o n between the iPhone/iPad
and the TGA 8000.
Requirements:
• iPhone (V6 and V6Plus) or iPad (V Air and V Air2). iOS version supported: iOS V8
• Lenovo M83 with Win7 (64bit) PKI supplied
• TGA 8000 with Pyris v12.1.1 connected to the PC via USB cable
• Pyris running and connected and communicating with the TGA 8000
• PC connection to the network
• Pyris Light App from iTunes
To set up an iOS connection follow this procedure:
1. Select Configure Analyzer from the Start Pyris button.
The TGA 8000 Configuration dialog appears.
Page 35
Set Up TGA 8000 System Components . 35
2. In the Remote Control section click the Enable checkbox and select the Maximum
number of devi c es the TGA 8000 can connect to. When complete click OK
3. Select Remote Control from the Pyris Tools menu.
4. In the Remote Control dialog the Local IP address and Port of the PC that the
app will connect with is displayed. Ensure the Start button is selected. The green
Status light confirms the port is open.
5. Open the Pyris Light
The app opens.
app on your iPad or iPhone.
Page 36
36 . TGA 8000 Installation and Har d ware Guide
6. Click the New Analyzer button
7. Enter the IP Address, Port Number, and Name of t he An al yz er (enter any
name) then click Next.
This displays the Settings screen which displays the General settings, Display
(graph) settings and Network conne ct ion se tt ings .
Page 37
Set Up TGA 8000 System Components . 37
8. Click Accept (these settings can be changed at any time). This displays the Main
Screen displaying a list of all Analyzers.
9. Select your analyzer and the connection will be made. This is confirmed when the
confirmation page appears.
Page 38
38 . TGA 8000 Installation and Har d ware Guide
10. Connection can also be confirmed by observing the Remo te C ont rol dialog in Pyris
which displays the Local IP address and Port of the iPad or iPhone. Connection
permissions can be controlled with all devices within this dialog.
• Start/Stop button – Starts and Stops connection with all devices.
• Refuse button – Blocks connection with this device.
Page 39
Initial Setup for t he Balance and Hangdown Wires . 39
Initial Setup for the Balance and Hangdown
Wires
CAUTION: Initial installation or movement of the instrument after installation calls
for the below steps to be performed.
ATTENTION: L'installation initiale ou le déplacement de l'instrum ent après
l'installation nécessite l'exécution des étapes ci-dessous.
• Leveling the TGA 8000
• Installing the hangdown wire and tare weight
• Adjusting the Convection Iris
• Adjusting the balance mechanism
Leveling the TGA 8000
Level the instrument using the spirit level provided.
1. Position the analyzer in the exact spot on the bench where you will be using it.
2. Remove the dome cover by gently lifting the cover straight up at least 8” after it is
released from the balance plate. Failure to do th is can result in damaging the
balance mechanism!
3. Observe the level on the balance plate near the flange of the anti-convection tub e.
c
4. Check t o see if the analyzer rocks on the bench or if one foot is not touching the
bench. Starting with that foot, turn the foot clockwise until it is touching the bench.
Once all the feet are touching the bench, make sure that the analyzer is steady.
5. Look at the level to see if the bubble is in the center. If it is not, slowly turn the
knobs of the adjustable feet until the bubble is centered in the level. The TGA 8000
is leveled when the bubble is centered inside the ring.
6. Once the analyzer appears to be level, make sure all of the feet are in contact with
the bench and that the analyzer does not rock.
You have now finished leveling the TGA8000. Final leveling will be complete after the
Autosampler calibration.
Page 40
40 . TGA 8000 Installation and Har d ware Guide
Install the Furnace Hangdown Wire
The TGA8000 comes with a kit that contains standard hangdown wires which is hung on the
platinum hangdown ribbon on the furnace side of the balance.
To install a hangdown wire and tare weig ht, fol lo w the procedur e be lo w:
1. Remove the dome cover by gently lifting the cover straight up at least 8” after it is
released from the balance plate. Failure to do th is can result in damaging the
balance mechanism!
2. Expand the iris ope ning by pushing the iris arm in a clockwise direction.
3. Using tweezers, pick up the appropriate tare weight for the TGA 8000 and carefully
hang it from the loop of the hangdown ribbon located on the right arm of the
balance.
4. Place the hangdown wire alignment tool (P/N N5376622) on the anti-convection cap.
CAUTION: The platinum hangdown ribbons are extremely fragile. Be careful not
to exert force on the loops of the hangdown ribbon or on the hangdown wire after
it is attached. Handle tweezers carefully when placing them near the platinum
hangdown ribbons, the hangdown wire, or stirrups as they may catch on one of
the components.
ATTENTION: Les rubans de platine sont extrêmement fragiles. Veillez à ne pas
exercer de force sur les boucles du ruban suspendu ou sur le fil pendulaire une
fois qu'il est fixé. Manipulez délicatement les pinces en les plaçant près des
rubans de pendage en platine, du fil pendulaire ou des étriers, car ils peuvent
s'accrocher à l'un des composants.
5. Using twe ezers, carefully pick up a hangdown wire and drop the wire through the
hangdown wire tub e. The wire will rest on the alignment tool.
6. Using twe ezers, grasp the hangdown wire with the tip facing the ribbon loop and
insert the tip through the loop; release the hangdown wire.
7. Remove the alignment tool and gently place a pan on the end of the hangdown wire.
Page 41
Initial Setup for t he Balance and Hangdown Wires . 41
NOTE:
Adjusting the Convection Iris
The iris is an anti-convection device. It prevents reaction gases from rising into the balance
chamber from the furnace.
CAUTION: If the iris is not adjusted properly then soot from the sample will
contaminate the balance mechanism. It is important that the hangdown wire does
not touch the iris when it is closed.
ATTENTION: Si l'iris n'est pas réglé correctement, la suie de l'échantillon
contaminera le mécanisme de l'équilibre. Il est important que le fil pendulaire ne
touche pas l'iris lorsqu'il est fermé.
1. Close the iris until it is approx 1mm clearence between the edge of the iris and the
hangdown wire all around. The wire should be centered in t he iris.
2. If it is not centered then the balance mechanism must be adjusted.
Adjusting the Balance Mechanism
If the hangdown wire is not centered in the iris then you will have to adjust the balance
mechanism position.
CAUTION: The balance mechanism is very fragile. Use extreme caution when
handling.
ATTENTION: Le mécanisme de l'équilibre est très fragile. Soyez extrêmement
prudent lors de la manipulation.
1. Ensure that the upper cooling fan is off; otherwise, it will cause the hangdown wire
to swing.
2. The adjustment scre ws are easily accessible (one shown below – red arrow). Loosen
the screws and ca refully move the balance to position the hangdown wire in the
center of the iris.
A small amount of rotation of the hangdown wir e (see Autosampler Calibration) is possible by
rotating the balance about the hangdown wire centerline.
3. Tighten the balance screws.
Page 42
42 . TGA 8000 Installation and Har d ware Guide
4. Adjust the iris again as per previous section.
5. Replace the balance cover
Autosampler Calibration
CAUTION: If the TGA 8000 is moved or installed for the first time, the Initial
Setup for the balance and hangdown wires must be performed prior to
Autosampler Calibration to ensure reliable operation of the autosampler.
ATTENTION: Si le TGA 8000 est déplacé ou installé pour la première fois, la
configuration initiale des fils d'équilibrage et de déconnexion doit être effectuée
avant l'étalonnage de l'échantillonneur automatique afin d'assurer un
fonctionnement fiable de l'échantillonneur automatique.
1. Ensure the hangdown wire is facing in the directi on of the sample tray. When a pan
is hung on the hangdown wire the stirr ups must be inline with the furnace center
and the large black column that supports the balance m echanism.
Page 43
Initial Setup for t he Balance and Hangdown Wires . 43
NOTE:
2. In the Pyris Software TGA 8000 screen (located at the top of the screen) click
Calibrate Autosampler from the Calibrate menu.
The Autosampler Calibration dialog box is displayed:
3. Calibr ate the tray height by clicking on the Tray Vertical button and follow the onscreen prompt. Make sure an empty samp le pan is present in position 1 in the
sample tray. Ensure the hook of the hangdown wire is:
• Aligned 2-3mm below the apex of sample pan handle (right image below)
• The hangdown wire is centered within the furnace tube (right image below))
• Aligned on the centerline of the apex. (left image below).
If the hangdown i s not on the centerline of the apex, the unit will have to be adjusted using
the leveling feet. Either raise or lower the foot nearest the furnace to ensure t he hangdown
wire is on center with the apex. If this procedure is done, re-check to ensure the iris is not
contacting the hangdown wire (see Adjusting the Convection Iris section).
Page 44
44 . TGA 8000 Installation and Har d ware Guide
4. Calibrate the sample pickup and drop-off position by clicking on the T ray Rotation
button and follow the on-screen prompt. For this calibration an empty sample pan
must be present on the hangdown wire before starting. During the calibration the
sample pan should hang on the centerline of the resting position on the tray to 12mm to the left the center as shown below.
5. To load an empty samp le pan click on the Autosampl er Control button.
The Autosampl er Co nt r o l dialog box is displayed:
Page 45
Initial Setup for t he Balance and Hangdown Wires . 45
6. Make sure an empty sample pan is present on position 1. Select carousel location as
1 and click Load Sample button.
Page 46
46 . TGA 8000 Installation and Har d ware Guide
Calibrate the TGA 8000
The TGA 8000 is calibrated at the factory and this calibration file is loaded from the CD upon
installation. There is also a default calibration which removes all of the calibration
adjustments and reverts to the hardware presets.
The TGA must be calibrated before obtaining accurate data. Calibration is performed using
the Pyris Software calibration routine which is accessed from the Calibrate menu through
the Method Editor window. There are four independent calibration routines accessed through
the three tabs on the Calibration menu item that should be performed for a TGA with or
without an autosampler:
• Weight Calibration
• Temperature Calibration
• Balance Drift Optimization
• Baseline Drift Optimization
In general, the calibrations should be performed in the above order; however, the weight
calibration will be done first and will likely only need to be done again if the TGA is moved or
re-leveled.
Before redoing the baseline calibration, restore the default baseline using the Restore
function in the Calibrate menu. The 7-hour Baseline DriftOptimization needs only to be
performed once, provided the full calibration was not set to defaults.
The Temperature and Baseline Calibration may be re-done to optimize the TGA for
optimum temperature and weight loss accuracy. For many routine separations this level of
calibration accuracy is unnecessary. It is wise t o Sa ve and Apply each calibration that is
performed under a new and descriptive file name so it can be recalled in the future using the
File Open menu item.
Performing a Weight Calibration
This calibration will be default depending on which Y Range you are using. Most people use
the Low Range which gives a dynamic range of +/- 160 mg which is appropriate for almost
all TGA analyses. This procedure will also wor k if you choose to calibrate the High Weight YRange which must be used if samples heavier than 160mg are to be analyzed. This
procedure needs to be modified if you calibrate the Ultrasensitive Y-Range in that you will
need to purchase and substitute an accurate 10mg reference standard. This may offer a
slight advantage when analyzing very small sam p les and optimizing the sensitivity for small
weight changes.
In order to perform the weight calibration, you will need the 100 mg calibration weight (P/N
09200501) in the Spares Kit (P/N N5320026).
1. Select Preferences from the Tools menu, then select the
TGA 8000 tab.
Page 47
Calibrate the TGA 8000 . 47
NOTE:
NOTE:
2. In the Y Data section select the range you plan to us e and click OK.
3. Select Calibrate Instrument from the Calibrate menu.
4. Select the Weight tab
in the Calibration menu.
Prior to calibrating the weight the display should look as below with the value in the
Measured(mg) bo x reading 100.000. If it indicates ano ther value you must restore the
default weight calibration using the Calibration Restore Weight Calibration menu item
5. Type a name in the Operator ID field (optional).
6. Enter the weight of the calibration weight in the Ref. Weight field.
Remove any pan from the hangdown wire because once you click Begin Calibration you
cannot remove it.
Page 48
48 . TGA 8000 Installation and Har d ware Guide
NOTE:
7. Click B egin Calibration.
A dialog box appears with instructions to remove anyt hing in the sample pan. For the
autosampler system, remove any sample from crucible “1”. The autosampler should
be in the Safe posit ion.
8. Follow the instructions in the dialog box, then click OK. The analyzer is tared for a
zero reading.
For the autosampler system, after accepting the reading, the crucible is unloaded
and the furnace is lowered. After approximately 30 seconds, the autosampler rotates
the tray so that position 1 is easily accessible .
9. When prompted, place and cent er the calibration weight in the sample pan and click
the OK button. The value of the calibr ation weight is read and displayed in the dialog
box. Click OK to accept the reading when the weight is stable.
For the autosampler, place the reference weight in crucible 1 and click OK. The value
of the calibration weight is read and displayed. When the signal settles, click OK to
accept the reading . The furnace goes to the Cool positi on and the crucible is
returned to the tray.
10. The just measured value is automatically displayed in the Measured field. Click the Save & Apply button to send the new calibration value to the analyzer and save the
calibration file.
11. Click the Close button to exit the Calibration window.
The weight calibration is complete.
Performing the Temperature Calibration
The temperature calibration is performed for a TGA 8000 with or without an autosampler.
Perform the procedure below to calibrate the TGA for temperature.
Prior to collecting the data to be used for temperature calibration as described below you
should open a calibration file which has the temperature calibration parameters set to their
default values, or you should go to the Restore menu item in Calibrate and restore the
temperature calibration.
For a temperature calibration use the magnetic standards of test metals Alumel, Nickel,
Perkalloy, and Iron supplied in the Spares kit (P/N N532002 6) . For calibrated analysis above
1000 °C substitute cobalt for one of the other standards; for example, check colbalt and
uncheck iron. The calibration runs should use the same conditions under which you would
run your samples and the scan rates for calibration should ideally bracket the rate(s) for
sample analyses.. (However, selecting rates above 50 °C/min for calibration is usually not
advisable.) Temperature calibration uses the Curie transition of the materials, that is, the
point at which the magnetic properties disappear.
1. Select Preferences from the Tools menu, then select the
TGA 8000 tab.
Page 49
Calibrate the TGA 8000 . 49
2. In the Y Data section set the range to Low Range +/-160mg, leave the Ordinate FilterON, and click OK. Precondition the TGA by running a conditioning method as
follows:
• Heat from 30 C 10 1080 C at 200 c/min.
• Cool from 1080 C to 30 C at 500 c/min.
3. While the conditioning method is running, cut 2-3mm lengths of test metals Alumel,
Nickel, Perkalloy and Iron, place them into a ceramic pan, and then cover them with
thin coating of alumina powder.
4. With a non-autosampler system, att ach the pan onto the hangdown wire, raise the
furnace, and observe the angle of the pan bale/stirrup.
With an autosampler system, place the ceram ic pan with samples into position 1 on
the autosampler tray. Click the Autosampler Control button in the Pyris software's
control panel. Click Load Sample “1”. The tray will move and hook the pan onto
the hangdown wire. Click the OK button to close the dialog box.
5. Raise the furnace.
6. Create a method for calibration, by selecting New from the File menu.
Page 50
50 . TGA 8000 Installation and Har d ware Guide
This method will collect data for all 4 calibra tion materials at 5 and 20 C/min, modify
the scan rate and ranges as needed. For accurate temperature calibration above
1000C you also need to run the cob alt Curie stand ard (using a 2mm by 2mm piece of
cobalt foil, run alone, not with other Curie standards) and using a similar
temperature program as above but heating between 1050 and 1200C.
7. Select the Sample Inf o t ab and fill in the Sample ID, Operator ID, Comment,
and File Name as required.
8. Tare the weight by clicking the Zero Weight button in the Control Panel. The
analyzer reads the weight of the sample which is entered into the Zero field of the
Enter Sample Weight box in the Method Editor.
9. Open the furnace door and carefully place the magnet (P/N N5370466) around the
furnace tube, then raise the furnace.
10. Rotate the magnet until the pan rotates from the magnetic field to its pre-magnet
angle. Make sure t he sample pan is not touching the sidewall of the furnace during
this procedure. If it is, swivel the magnet until the pan is centered.
If this is not possible, use smaller pieces of C urie standards.
11. Click the Sampl e Wei ght icon on the Control Panel.
This enters the we ight into the method. As the sample loses magnetism, the
apparent weight reading decreases.
12. Click the Start Method button in the Control Panel.
13. After the run, move the sample pan without disturbing the contents to a spot on the
autosampler carousel. It will be rerun later.
You are now ready to calculate the Onset Value for all four calibration standards.
Calculating the Onset Value
This is done in the Data Analysis window.
1. Click the Data Analysis ico n on the toolbar, the n open the data file in the Data
Analysis window.
Page 51
Calibrate the TGA 8000 . 51
Make sure that the curve is displayed with the following axes:
X = Temperature (change by selecting Rescale X from the Display menu and
then select Temperature in the dialog box).
Y = Weight (change by selecting Weight from the Curves menu).
2. In the Curves menu select the Weight curve.
3. In the Step Select dialog select step 2 and ste p 5 through 12. (These two curves
contain data for heating the Curie samples at 20 and 5 degrees per minute. (You can
delete the original weight data curve.
4. For the active curve, select from the Math menu, derivative, this will display the
derivative curve for the segment you are analyzing.
5. With the derivative curve active, Select Peak Area from the Calc menu.
Calulate the peak temperature at corresponds to each of the weight loses. This point
will be used to determine one of the points for each weight loss step.
6. Select Weight Loss curve so the derivative curve goes to the background.
7. Select Onset from the Calc menu.
8. Move the two red X marks on the curve. Position the left limit X at the inflection
point of the first derivative curve and place the right limit X past the inflection point
on the level part of the curve. Alternatively type in the values using the peak value
from the derivative calculation as the steepest slope and a temperature on the
trailing baseline.
9. Click the Calculate button in the Onset dialog box. T he Adjust Tangents box
appears.
10. Adjust the tangents of each point until they cross and each is tangential at the point
selected.
Perform this Onset calculation on each of the eight Curie temperatures (4 at 5 °C/min
and 4 at 20 or 25 °C/min.
11. Click the Calculate button in the Adjust Tangents dialog box. The onset
temperature is calculated for the sample and displayed on the screen (Onset X).
Page 52
52 . TGA 8000 Installation and Har d ware Guide
12. In the Method Editor, select Cali bra t e In s t r ument from the Calibrate menu.
13. Click the Temperature tab
reference material, the Expected Onset values, the Measured Onset values (four
temperatures at two rates).
14. Check the standards you plan to use and un che ck the ot her s. Always use a minimum
of 3.
15. Click Save and Apply to send the calibration values to the analyzer and save the
current calibration file (SN_weightcal_st.tg8c).
in the Calibration window. Enter the name of each
16. Click Close to exit the Calibration window.
The temperature calibration is now complete. For the case wher e you are calibrating to 1200
C you will be performing the onset calculation as above on the Cobalt analysis, inputting the
measured values into the table on the cobalt line, and selecting the cobalt checkbox instead
of one of the other standards.
Performing a Balance Drift Optimization
The Balance Drift Optimization takes approximately 7 hours.
1. From the Pyris menu, select Calibrate Instrument from the Calibrate menu.
2. Select the TG Drift tab in the Calibration menu.
Page 53
Calibrate the TGA 8000 . 53
3. On the TG Drift tab uncheck Perform Baseline Optimization and check Perform Balance Optimization (~7 hours) to perform the tests which compensate the
balance for temperature rise in the balance chamber.
4. For a TGA 8000 without an autosampler, place a clean empty sample pan in the
stirrup at the end of the hangdown wire. Place it there manually with tweezers while
the autosampler is in the Safe position.
For an autosampler system, place a pan in Position “1” on the autosampler tray and
use the Autosampl er Co ntrol dialog box to load a sample.
5. Click Begin Calibration.
An estimate of the amount of time needed to perform the calibration is displayed and
it begins to count down.
6. When the calibration is complete, click Save & Apply to send the new calibration
value to the analyzer and save the file (i.e. SN_weightcal_st_balanceopt.tg8c where
SN is the instrume nt serial number).
7. Click Close to exit the Calibration window.
Performing a Baseline Drift Optimization
The Baseline Drift optimization varies with scan rate and temperature range. Select a rate
and range applicable to the application.
1. From the Pyris menu, select Calibrate Instrument from the Calibrate menu.
2. Select the TG Drift tab in the Calibration menu.
Page 54
54 . TGA 8000 Installation and Har d ware Guide
3. On the TG Drift tab uncheck Perform Balance Optimization (~7 hours) and
check Perform Baseline Optimization.
4. For a TGA 8000 without an autosampler, place a clean empty sample pan in the
stirrup at the end of the hangdown wire. Place it there manually with tweezers while
the autosampler is in the Safe position.
For an autosampler system, use the Autosampler Co n tr ol dialog box to load a
sample.
5. Enter the minimum and maximum temperatures between which the baseline will be
determined (e.g., set minimum to 50 °C and maximum to 1150 °C).
6. Select a scanning Rate (°C/min) that you will be using for scanning analyses (e.g.,
20 °C/min).
7. Click B egin Calibration.
An estimate of the amount of time needed to perform the calibration is displayed and
it begins to count down.
8. When the calibration is complete, click Save & Apply to send the new calibration
value to the analyzer and save the file
(SN_weightcal_st_balanceopt_baselineopt20C.tg8c).
9. Click Close to exit the Calibration window.
The TGA 8000 calibration is now complete. This ends the installation of the TGA 8000.
Page 55
TGA 8000 Hardware
Maintenance
55
Page 56
56 . TGA 8000 Installation and Har d ware Guide
NOTE:
Maintenance Overview
The TGA 8000 needs little routine maintenance other than proper treatment as a sensitive
electronic device. Occasion ally, the furnace or furnace tube may become coated with sample
residue and require cleaning, or furnace components and other accessories, such as
hangdown wires and thermocouples, may need to be replaced.
The TGA 8000 case is painted metal. The exterior surfaces may be cleaned with a soft cloth,
dampened with a mild detergent and water solution.
Maintenance of the TGA 8000 includes the following procedures:
• Cleaning the Furnace
• Cleaning the Furnace Tube
• Cleaning the Sample Pan
• Furnace Maintenance
• Furnace Tube Maintenance
• Furnace Hangdown Wire Maintenance
• Replacing the Thermocouple
• Hangdown Ribbon Replacement
Page 57
Cleaning the Furnace . 57
vous que
NOTE:
Cool Furnac e
est une solution d'acide caustique
Cleaning the Furnace
An automatic procedure for cleaning the furnace is incl uded in the Pyris software. The
procedure involves lowering the furnace to expose it to the air and then heating it to
approximate ly 900 °C to burn off any materials coated onto the furnace surfaces.
WARNING: Since the furnace will be programmed to 900 °C, make certain that
the protective furnace door is closed. DO NOT touch the furnace during this
procedure.
AVERTISSEMENT: Comme le four sera programmé à 900 ° C, assurezla porte du four protecteur est fermée. NE PAS toucher le four pendant cette
procédure.
1. Start the Pyris software and make sure the Control Panel is displayed.
2. Select the Clean Furnace icon
3. The furnace will move to the lowered position and programmed to 900 °C. When the
cleaning procedure is complete, the furnace will remain in the lowered position and
cool down to 30 °C.
To abort the furnace cleaning procedure at any time, select the
from the Control Panel.
icon.
Cleaning the Furnace Tube
In order to clean the furnace tube, you must first remove the tube from the furnace. Clean the
furnace tube using one of the following methods:
1. Soak and scrub the furnace tube in a strong dete rgent (e.g., alcono x).
2. Soak the furnace tube in a strong solvent such as cleaning solution (e.g.,
concentrated H
WARNING: Concentrated H2SO4 and K2CrO7 is a very strong, caustic acid
solution. Wear protective gloves and safety glasses and perform this procedure in
a fume hood.
AVERTISSEMENT: H2SO4 concentré et K2CrO7
très forte. Porter des gants de protection et des lunettes de sécurité et effectuer
cette opération dans une hotte aspirante.
and K2CrO7).
2SO4
3. Dip the furnace tube in a dil ute hydrogen fluoride (HF) solution (1 0% solution, 5 min
maximum).
WARNING: Hydrogen fluoride is a dangerous skin irritant. Wear protective gloves
and safety glasses and perform this procedure in a fume hood.
AVERTISSEMENT: Le fluorure d'hydrogène est un irritant cutané dangereux.
Porter des gants de protection et des lunettes de sécurité et effectuer cette
opération dans une hotte aspirante.
4. Rotate the furnace tube over a Bunsen burner ( heat up slowly to av oid cracking).
5. After cleaning, reinstall the furnace tube.
Page 58
58 . TGA 8000 Installation and Har d ware Guide
TGA 8000 Furnace Maintenance
Standard furnace maintenance consists of the following procedures:
• Removing the Furnace
• Replacing the Furnace
• Adjusting the Furnace
Removing the Furnace
WARNING: Before performing this procedure, shut down the system and remove
the line power. Make sure the furnace has cooled sufficiently so that you do not
burn yourself when removing the furnace components.
To perform this procedure the furnace must be in the lowered position.
AVERTISSEMENT: Avant d'effectuer cette procédure, arrêtez le système et
retirez l'alimentation secteur. Assurez-vous que le four a suffisamment refroidi
pour ne pas vous brûler lors de la dépose des composants de la fournaise.
Pour exécuter cette procédure, le four doit être en position abaissée.
1. Disconnect the furnace power cord from its connector located behind the left panel
of the TGA. The furnace cable is screwed into the connector. To disconnect unscrew
and gently pull for ward on the connector.
Page 59
TGA 8000 Furnace Maint e n a nce . 59
2. Slide the instrument so the furnace is positioned off the edge of the bench
3. Loosen and remove the Clamp holding the Pyrex furnace tube to the base of the
furnace assembly.
4. Remove the Pyrex furnace tube. You do not have to disconnect the Tygon exhaust
tube from the furnace tube. Lay the furnace tube on the bench top.
5. Remove the inner Pyrex furnace lining.
6. Remove the platinum furnace shield by sliding it off the furnace. Do not deform the
shield.
7. Remove the plate (slotted ring) sitting on the spring.
8. Remove the spring
9. Unscrew the large knurled furnace nut to completely remove the furnace from the
mount
10. Remove the clip that is underneath the nut.
11. Lift the furnace assembly off of the mounting.
Page 60
60 . TGA 8000 Installation and Har d ware Guide
If you plan to use the thermocouple from the removed furnace on the new furnace, remove
it as described in “Replacing a Thermocouple.”
Replacing the Furnace
1. Insert thermocouple into furnace base following the thermocouple replacement
procedure
2. Place the furnace on the furnace base. Note that there is a locating pin on the
mounting assembly that fits into the hole on the bottom of the furnace to assist
alignment.
3. Place the furnace assembly onto the mounting assembly.
4. Place the knurled furnace locking nut over the furnace.
5. Place the clip underneath the knurled furnace locking nut.
6. Straighten the furnace and tighten the knurled nut.
7. Slide the platinum furnace shield over the furna ce.
8. Place the spring over the furnace and until it sits in the top of the metal furnace
base. Make sure that the top of the spring is straight.
Page 61
TGA 8000 Furnace Maint e n a nce . 61
9. Place the slotted plate on top of the spring. Make sure the plate is straight and the
slot aligns with the therm ocouple.
10. Place the quartz ins ert over the fur nace and seat it on the plate.
11. Place the Pyrex furnace tube over the furnace so that i t rests on the O-ring on the
top of the furnace tube base. Position the exhaust tube so that it is left of center
when the furnace i s out of the cooling pos ition.
12. Place the clamp around the base of the furnace tube and top of the furnace tube
base. The red handle should be placed s o that it is facing the cooling housing. The
exhaust tube should be to the right of the clamp. Note the position of the exhaust
tube and clamp when the furnace is in the Cooling position in the right-hand photo
below.
13. Plug in the furnace power cord to its connector on the left side of the TGA. There are
male and female guides on the connector that allow the connector to go on in only
one way.
After installing the standard furnace, you will have to adjust it with respect to the ball joint,
hangdown wire, and sample pan.
Page 62
62 . TGA 8000 Installation and Har d ware Guide
Adjusting t he Furnace
When your TGA 8000 with standard furnace and no autosampler is installed by a service
engineer, he performs a series of adjustments in order to optimize its performance. If you
need to remove the furnace and/or furnace tube, perhaps to clean it, you will have to perform
these steps to ensure proper operation of the instrument. These include:
• Adjusting the Height of the Furnace
• Adjusting the Balance Mechanism
• Adjusting the Convection Iris
Adjusting the Height of the Furnace
The height of the furnace is pre-set at the factory therefore no adjustme nt should be
required. However, if a thermocouple is replaced then the below procedure must be
followed:
1. The sampl e pan should still be on the hangdown wire. If not, place a pan on the end
of the hangdown w ire.
2. Click Ra is e Furnace button on the Pyris Control Panel.
3. Check t he position of the stirrup with respect to the tip of the thermocouple.
4. Adjust the height of the furnace using the furnace adjust height knob located below
the furnace adjustment ring. Turn the knob counterclockwise to raise the furnace
and clockwise the lower the furnace.
5. Raise the furnace using the knurl knob adjustment until the pan touches the tip of
the thermocouple, this is done by observing the weight status in the status panel.
When you see a weig ht change…
•Turn the furnace knob adjustment 3 full turns CCW (this will lower the
furnace ~3mm - each turn is 1 mm).
•Then turn the knob 2 full turns CW. The result will be the thermocouple will
be about 1 mm below the sample pan.
CAUTION: Make sure that the sample pan hangs in the center of the furnace. It is
important that the sample pan does not touch the side wall of the furnace or the tip
of the thermocouple.
ATTENTION: Assurez-vous que la casserole d'échantillonnage est suspendue au
centre du four. Il est important que le récipient d'échantillon ne touche pas la paroi
latérale du four ou la pointe du thermocouple.
Page 63
Adjusting the Furnace . 63
The TGA 8000 is no w ready to be calibrated.
Adjusting the Convection Iris
1. Lift the top cover of the analyzer up to access the balance dome.
2. Remove the balance dome b y turning the locking devices and pulling the arms out of
the groove in the dome. Lift the dome straight up and away from the balance. The
iris is now exposed.
3. Open the iris all the way by pushing the arm clockwise.
4. Loosen the screws on the two stands that hold the iris in place.
5. Move the iris around until the hangdown wire is centered in the iris.
6. When the hangdown wire i s centered, tighten the screws on the st ands. Do not
overtighten; the iris arm will not be able to move.
7. Close the iris as far as possible without it touching the hangdown wire by turning the
lever counterclockwise.
8. Replace the dome back onto the balance plate and lock in place.
9. Lower the access cover.
Page 64
64 . TGA 8000 Installation and Har d ware Guide
Furnace Tube Maintenance
Maintenance of the furnace tube co nsists of the following procedures:
• Removing the Furnace Tube
• Cleaning the Furnace Tube
• Replacing the Furnace Tube
Removing the Furnace Tube
1. Make sure that the furnace is in the lowered position.
2. Remove the exhaust line from the furnace tube side arm.
3. Remove the clamp around the base of the furnace tube.
4. Remove the glassware.
CAUTION: Do not pull the furnace tube very quickly or at an angle to avoid
breaking the thin stem on the furnace.
ATTENTION: Ne pas tirer le tube du four très rapidement ou à angle pour éviter
de briser la tige mince sur le four.
Replacing the Furnace Tube
This procedure is performed after removing the furnace tube.
1. Carefully place the tube over the furnace and place on the O-ring at the top of the
metal furnace base.
2. Position the furnace tube so that the exhaust tube connector is left of center while
the furnace is in the lowered position. The exhaust tube connector should be able to
go into the cool pos ition without hitting the side of the analyzer.
3. Place the clamp around the furnace tube bottom and the top of the furnace base and
tighten. Make sure that the red piece is just to the left of the exhaust tube.
4. Replace the exhaust tube to the furnace side arm.
Page 65
Furnace Tube Maintenance . 65
est une solution d'acide caustique
Note the positioning of the clamp and exhaust tube with respect to the autosampler when the
furnace is in the lowered position, and with respect to the wall of the cooling area with the
furnace in the cool position:
Cleaning the Furnace Tube
In order to clean the furnace tube, you must first remove the tube from the furnace. Clean the
furnace tube using one of the following methods:
1. Soak and scrub the furnace tube in a strong detergent (e.g., alconox).
2. Soak the furnace tube in a strong solvent such as cleaning solution (e.g.,
concentrated H
WARNING: Concentrated H2SO4 and K2CrO7 is a very strong, caustic acid
solution. Wear protective gloves and safety glasses and perform this procedure in
a fume hood.
AVERTISSEMENT: H2SO4 concentré et K2CrO7
très forte. Porter des gants de protection et des lunettes de sécurité et effectuer
cette opération dans une hotte aspirante.
3. Dip the furnace tube in a dil ute hydrogen fluoride (HF) solution (10% solution, 5 min
maximum).
and K2CrO7).
2SO4
WARNING: Hydrogen fluoride is a dangerous skin irritant. Wear protective gloves
and safety glasses and perform this procedure in a fume hood.
AVERTISSEMENT: Le fluorure d'hydrogène est un irritant cutané dangereux.
Porter des gants de protection et des lunettes de sécurité et effectuer cette
opération dans une hotte aspirante.
4. Rotate the furnace tube over a Bunsen burner (heat up slowly to av oid cracking).
5. After cleaning, reinstall the furnace tube.
Page 66
66 . TGA 8000 Installation and Har d ware Guide
Replacing a Thermocouple
To replace a thermocouple, follow the procedure below:
1. Disconnect the plastic thermocouple connector plug from the base of the furnace by
gently pulling up on the connector plug.
CAUTION: Do not bend the thermocouple severely as this may damage it. If the
thermocouple appears old, worn, or broken, replace it (P/N N5321002).
ATTENTION: Ne pliez pas le thermocouple avec précaution car cela pourrait
l'endommager. Si le thermocouple est vieux, usé ou cassé, remplacez-le (P/N
N5321002).
2. Remove the tip of the thermocouple from inside the furnace by pulling the
thermocouple straight down toward the bottom of the furnace.
3. Using the old thermocouple as a guide, carefully bend the new thermocouple in the
way the old thermocouple was bent
4. Take the new thermocouple and, while looking inside the furnace, feed the tip of the
new thermocouple through the thermocouple slot in the furnace stem until it is
approximately 1 or 2 mm below the half-height of the furnace wall.
Page 67
Replacing a Thermocouple . 67
Helpful Hint:
5. Plug the plastic thermocouple connector plug on the base of the thermocouple into
A new thermocouple is shipped in a plastic c ontainer with two caps, the cap is 10 mm
in height, place this into the furnace and raise or lower the thermocouple such that it
is just touching the bottom of the cap. Remove the cap.
the base of the furnace. There are two small pins in the furnace base where the
thermocouple connects. Be careful not to bend or break the thermocouple pins in the
furnace base.
Page 68
68 . TGA 8000 Installation and Har d ware Guide
Page 69
Furnace Hangdown Wire Maintenance . 69
platine, du fil pendulaire ou de l'étrier.
Furnace Hangdown Wire Maintenance
There are two hangdown wires associated with the furnace: a quartz (P/N N5370488) and a
nichrome (P/N N5370490) wire. These wires come in kits.
Maintenance of the furnace hangdown wire consists of the following:
• Removing the Furnace Hangdown Wire
• Installing the Furnace Hangdown Wire
WARNING: Before performing the procedures below, make sure that the furnace
has cooled sufficiently so that you do not accidentally burn yourself.
AVERTISSEMENT: Avant d'exécuter les procédures ci-dessous, assurez-vous
que le four a suffisamment refroidi pour ne pas vous brûler accidentellement.
Removing the Hangdown Wire
1. Lift the round cover straight up to gain access to the balance.
A platinum balance ribbon is connected to the left and right arms of the balance. The
ribbon has a loop at its end from which the hangdown wire is hung on the left side of
the balance. A tare weight, specific for the particular hangdown wire used, is hung
from the ribbon on the right side of the balance to offset the weight of the
hangdown wire, stirrup, and sample pan or crucible.
CAUTION: The platinum ribbon is extremely fragile. Care should be taken not to
exert force on the loop of the balance ribbon or on the hangdown wire after it is
attached. Handle tweezers with extra care when placed in the vicinity of the
platinum balance ribbon, the hangdown wire, or the stirrup. Tweezers can catch
any of the components and cause damage.
ATTENTION: Le ruban de platine est extrêmement fragile. Il faut veiller à ne pas
exercer de force sur la boucle du ruban de balance ou sur le fil de pendaison
après qu'il est attaché. Manipulez les pinces avec plus de soin lorsqu'elles sont
placées à proximité du ruban de balance en
Les pinces peuvent attraper n'importe lequel des composants et causer des
dommages.
2.To remove the hangdown wire from a non-autosampler TGA 8000:
i. Make sure that the anti-convection iris in the balance area is open all the way.
Place the hangdown wire guide on the end of the convection flange at the end
of the joint.
ii. Using tweezers, carefully remove the hangdown wire from the ribbon loop. Let
go of the hangdown wire.
To remove the hangdown wire from an autosampler system:
i. Make sure that the anticonvection iris is open all the way.
ii. Use tweezers to carefully remove the hangdown wire from the platinum ribbon
loop. Let the hangdown wire drop onto the wire guide.
iii. Grasp the hangdown wire with the twee zers whi le removing the guide.
iv. Carefully guide the hangdown wire down and out of the tube.
Page 70
70 . TGA 8000 Installation and Har d ware Guide
NOTE:
Installing the Hangdown Wire
1. Place the hangdown wire guide on the convection flange below the joint.
2. Make sure that the convection iris in the balance chamber is open.
3. Carefully remove a new hangdown wire from its container. Handle it with one hand.
4. From the balance chamber, drop the new hangdow n wire through the hangdown
wire tube. The end of the wire will rest on either the guide.
5. Check t he orientation of the hook at the end o f the hangdown wire. It should face
towards the autosampler. Use tweezers to grasp the hangdown wire near the top. Insert
the tip of the hangdown wire throug h the ribbon loop and release. Go to step 7.
6. For a no nautosampler system place an empty pan at the end of the hangdown wire.
Make sure that the proper tare weight is in place on the other side of the balance. If you
need to change the weight, carefully remove the present one using tweezers. Pick up the
new tare weight with tweezers and carefully hang it from the loop of the hangdown ribbon
on the right side of the balance mechanism.
CAUTION: The platinum hangdown ribbons are extremely fragile. Be careful not
to exert force on the loop of the hangdown ribbon or on the hangdown wire after it
is attached. Handle tweezers carefully when placing them near the platinum
hangdown ribbons, the hangdown wire, or stirrups as they may catch on one of
the components.
ATTENTION: Les rubans de platine sont extrêmement fragiles. Veillez à ne pas
exercer de force sur la boucle du ruban pendulaire ou sur le fil pendulaire une fois
qu'il est fixé. Manipulez délicatement les pinces en les plaçant près des rubans de
pendage en platine, du fil pendulaire ou des étriers, car ils peuvent s'accrocher à
l'un des composants.
7. Replace the balance cover over the balance.
After installing a new hangdown wire on either system, you may have to adjust the height of
the furnace with respect to the sample pan.
Page 71
Hangdown Ribbon Replacement . 71
Platinum Ribbon Insertion Tool
N5376741
Rubber Cement
09923480
Rubber Cement Application Tool
N5376742
Ribbons
N5370342
Hangdown Ribbon Replacement
The balance mechanism of the TGA 800 0 is designed with replaceable platinum hangdown
ribbons. If either the hangdown wire r ibbon or the tare weight ribbon break s, you can
replace it using t he TGA 8000 Hangdow n Ribbon Replacement Kit (P/N N53704 78) .
Owerview
The platinum wire suspension consists of an ultrathin platinum ribbon with a loop on one
end. The balance tare weight and the sample hangdown wire (quartz or nichrome) are
supported from the loop on their respective ribbons. A small glass sphere is affixed to the
other end of the ribbon. This sphere is used to s uspend the ribbon from the glass capillary
tube which is permanently mounted to the balance beam. The capillary tube is flared at the
top to accept the ball on the ribbon. This design provide s an inherently stable mount for the
ball so that, in a TGA 8000 without an autosampler , t he ball does not need to be
permanently bonded to the capillary t ube to achieve good balance performance. Should the
platinum ribbon break, it can be removed easil y; a pa ir of tweezers can easily grip the ball so
that the ribbon can be removed.
For an autosampler system, the orientation of the platinum wire loop is important. If affects
the position of the hook which is critical in sample pan loading and unloa di ng . The flared
tube design facilitates this alignment since the mount is stable. Once the loop is aligned, the
ball is bonded to the capillary tube using an easily removable rubber cement.
Parts Included in the Hangdown Ribbo n Replacement Kit (P/N N5370478)
Replacing the Platinum Ribbon
1. Turn off t he power to the analyzer.
2. Lift the balance dome straight up.
WARNING: Be careful not to touch components on the thermal controller PC
board behind the balance. They may be hot to the touch.
AVERTISSEMENT: Veillez à ne pas toucher les composants de la carte de circuit
imprimé du thermostat derrière la balance. Ils peuvent être chauds au toucher.
CAUTION: Be careful not to apply excessive force to any component of the
balance assembly. Permanent damage may result.
3.Locate the flared capillary tube mount and the ball support of the broken platinum
ATTENTION: Veillez à ne pas exercer de force excessive sur un composant de
l'ensemble de la balance. Un dommage permanent peut en résulter.
ribbon.
Page 72
72 . TGA 8000 Installation and Har d ware Guide
NOTE:
NOTE:
NOTE:
4. Apply a few drops of acetone to soften the rubber cement holding the ball to the
flared capillary tube.
WARNING: Acetone is extremely flammable. It has a flash point of 0°F. It can
cause central nervous system depression. Harmful if swallowed. Causes irritation
to eyes, skin, and respiratory tract. Avoid eye and skin contact. Avoid ingestion
and inhalation. Refer to the Material Safely Data Sheet for further information.
AVERTISSEMENT: L'acétone est extrêmement inflammable. Il a un point d'éclair
de 0°F. Il peut provoquer une dépression du système nerveux central. Nocif en
cas d'ingestion. Provoque une irri tat ion des yeux, de la peau et des voies
respiratoires. Évitez le contact avec les yeux et la peau. Éviter l'ingestion et
l'inhalation. Reportez-vous à la fiche de données de sécurité pour plus
d'informations.
5. Using of tweezers, grasp the short piece of platinum wire extending above the ball.
6. Lift the ball off of the flared mount. If there is resistance, apply more acetone until
the ball is free.
CAUTION: Do not apply excessive upward force! This may damage the balance.
ATTENTION: Ne pas appliquer de force excessive vers le haut! Cela peut
endommager l'équilibre.
If no platinum wire extends above the ball, a small piece of stiff wire (in the kit) may be
used to push up on t he ball from the bottom end of the capillary tube.
7. Once the ball is free from the capillary tube, clean off any residual rubber cement.
Use as many drops of acetone as needed. Acetone that remains trapped in the tube
may be "wicked" away with a tissue.
8. Using tweezers, grasp the new replacement ribbon by the glass ba ll so t ha t the loop
hangs downward.
9. Carefully feed the loop end of the ribbon into the flared end of the capillary. Some
gentle flexing action on the ribbon may be necessary to get it to feed through the
tube. Be careful not to kink the ribbon.
If the loop does not feed through the capillary tube properly, the platinum ribbon insertion
tool (P/N N5376741) may be used to pull the loop through the tube. Feed the hooked end of
the wire through the capilla ry tube from the botto m and capture t he platinum ribbon loop.
Carefully pull the platinum ribbon through the capillary tube.
This operation requires high dexterity with both hands. It is better accom p lishe d with two
people.
10. Suspend the proper tare weight from the loop in the new ribbon; go to step 15.
The following alignment steps are required ONLY when replacing the sample hangdown wire
side ribbon. (The tare weight ribbon may be fixed in any direction.)
11. Install a hangdown wire and pan.
12. Assess the position of the pan relative to the autosampler tray position. Decide in
which direction and by how much the ribbon loop needs to be rotated to achieve the
correct alignment.
13. If some correction needs to be made, use tweezers to grasp the short piece of
Page 73
Hangdown Ribbon Replacement . 73
NOTE:
NOTE:
platinum ribbon above the glass ball resting in the flared capillary tube.
14. Slightly lift and rotate the ribbon to correct the alignment. This may take a few tries
but it is extremely important to obtain the a lig nment shown below.
15. Carefully inspect (using a magnifier, if necessar y) t he position of the platinum ribbon
within the capillary tube below the glass ball. It must hang straight down without
touching the inner wall of the capillary tube. If it touches, the ball must be lifted
slightly to allow the ribbon to center itself within the tube.
If you replaced the tare weight ribbon, go to step 17.
16. After centering the ribbon in the capillary tube on the sample hangdown wire side,
check that the rotational alignment of the sample pan has not been disturbed. If so,
repeat step s 14 and 15.
17. Locate and open the tube of rubber cement (P/N 09923480). Apply a tiny amount of
cement to the end of the wire rubber cement applicator (P/N N5376742). Touch the
cement to the area between the glass ball and the flared capillary tube.
Take great care not to disturb the position of the ball.
Repeat this step until you have placed cement around the circumference of the glass
ball.
Be very careful no t to apply too much rubber cement in one area such that i t runs down the
capillary tube.
18. Allow 10 minutes for the glue to set.
Inspect the work to ensure that adequate cement was applied. You can test the
cement by gently trying to remove the ball from the flared capillary tube by pulling
on the ribbon abo ve the ball. The ball should not move.
Re-inspect the section of the ribbon within the capillary tube. It should hang freely.
19. Replace the dome on the balance plate assembly, taking care not to touch any
components.
Page 74
74 . TGA 8000 Installation and Har d ware Guide
Quartz Hangdown Wire Kit
N5320140
NiCr Hangdown Wire Kit
N5320131
Cooling Jacket Accessory
N5370553
Accupik Puncture Device (for an autosampler system)
Aluminum crucible liner
N5376732
Kit of 400 aluminum liners
N5370492
Stainless steel crucible liner
N5376749
Kit of 400 stainless steel crucible liners
N5370495
Platinum crucible liner
N5376702
Kit of 400 platinum crucible liners
N5370551
Tare Weights
Tare weight TGA 8000 194 mg
N5320128
Tare weight TGA 8000 272 mg
N5320129
Tare weight TGA 8000 226 mg
N5320130
Tare weight TGA 8000 375 mg
N5370563
Accessories
The following are accessories or optional item s for use with the TGA 8000:
N5320016
Page 75
Cooling Jacket Accessory . 75
Part Number
Description
Quantity
02506518
Tygon tubing 0.25-in. i.d.
10 feet
09903004
Hose clamps
4
09908739
Teflon tape
1 roll
09920073
Quick connect stem
2
09940294
Nylon tie wrap
6
N5370333
Glass tube
1
Cooling Jacket Accessory
The TGA Cooling Jacket Accessory allows cooling of the TGA 8000 and Pyris 1 TGA furnace
chamber. The cooling jacket is double-walled glassware through which cooling water or
antifreeze flows. For subambient operation, this accessory is to be used with a refrigerated
coolant circulator. The recommended cooling device from PerkinElmer is the Polyscience
Chiller (P/N N5370220/1). Note that a cooler that generates excess vibration or turbulence
may result in the poor performance of the TGA.
Parts Included
Parts included in the Cooling Jacket Acces s ory ki t (P/N N5370553) are:
Installation
To install the cooling jacket accessory, follow the steps below:
1. Switch your TGA on. If it is already switched on, switch it off and then on again after
a pause.
2. Press the Lower Furnace button on the Pyris software Control Panel to check that the
furnace assembly is in the Lowered position.
3. If your TGA has an autosampler, make sure it is in the Safe position.
4. Remove the Tygon exhaust tube from the furnace tube.
5. Remove the quartz insert
6. Loosen and remove the clamp holding the Pyrex furnace tube onto the base of the
furnace assembly.
7. Carefully lift the Pyrex furnace tube up and over the inner furnace sleeve and store
the tube in a safe place.
8. Place the cooling jacket over the furnace.
The Liquid In and the Liquid Out connectors should be facing toward the left and the
exhaust tube should be behind the furnace, as shown below.
Page 76
76 . TGA 8000 Installation and Har d ware Guide
9. Replace the clamp around the base of the jacket to secure it to the furnace base.
10. Install the quartz insert
11. Swing the red handle around and tighten the clamp.
Make sure that the red handle does not touch the lower liquid connector (Cooling
Liquid In).
12. Attach the exhaust tube to the connector in the back of the cooling jacket.
13. The TGA 8000 furnace plugs into the left side after you remove the left side panel.
Connecting to the Chiller
CAUTION: An appropriate fluid must be used with the chiller. The fluid should be
effective down to –20 °C. Any lower than that would impair the conveyance
power of the chiller's pump. Distilled water is recommended for operation at
temperatures between 15 °C and 90 °C. For operati on at –20 °C to 100 °C, 50%
ethylene glycol in water is recom mended.
ATTENTION: Un fluide approprié doit être utilisé avec le refroidisseur. Le fluide
doit être efficace jusqu'à -20 ° C. Tout ce qui est inférieur à celui-ci
compromettrait la puissance de transport de la pompe du refroidisseur. L'eau
distillée est recommandée pour des températures comprises entre 15 ° C et 90 °
C. Pour un fonctionnement à -20 ° C à 100 ° C, 50% d'éthylène glycol dans l'eau
est recommandé.
CAUTION: Ensure that the cooling air vents at the front and rear of the cooler are
not blocked. Maintain a minimum space of 15–20 cm.
ATTENTION: Assurez-vous que les évents d'air de refroidissement à l'avant et à
l'arrière du refroidisseur ne sont pas bloqués. Maintenir un espace minimal de
15-20.
Page 77
Cooling Jacket Accessory . 77
To connect to the chiller, follow the steps below:
1. Place the chiller on the floor to the left of the TGA.
2. Make sure that the chiller is switched off but that its power cord is plugged into a
power supply.
3. Screw one of the barbed hose couplings shipped with the chiller into the OUTLET
connection on the chiller unit, and t ighten with a suitable wrench.
4. Couplings suitable for several different hose sizes are provided with the chiller.
5. Push a ho se (compatible with the desired coola nt) onto the OUTLET coupling and
secure with a hose clamp.
6. Connect the hose from the chiller OUTLET to the lower connector on the cooling
jacket. Push the tube up onto the conne ctor and over the two ribs.
7. Screw the other hose coupling into the INLET connection on the chiller and tighten
with a suitable wrench.
8. Push a second piece of hose onto the INLET coupling and secure with a hose clamp.
9. Connect the hose from the chiller INLET to the upper connector on the cooling
jacket. Push the tube up onto the conne ctor and over the two ribs.
10. Place a hose clamp (P/N 09903004) around each hose to secure it to the jacket.
Position the clamp so that it is between the two ribs and that they do no t interfere
with the door or the autosampler operation.
11. Place the tie wraps (P/N 09940294) around the two coolant hoses in s trategic places
to help keep them out of the way when the furnace changes position.
Suggested location is around the TGA 8000 foot below the furnace. Do Not tig hten
the tie-wrap – keep it very loose so as to not kink the tubing. Allow for excess
tubing to allow furnace to be raised without stressing tubing. See the exam ple
below.
Page 78
78 . TGA 8000 Installation and Har d ware Guide
Setting up the Chiller (refer to Circulator User Manual
for more details)
1. Fill the reservoir wit h coolant.
2. Ensure that the cooling coils are completely covered. The maximum coolant level is
25 mm below the top of the reservoir.
3. Switch the chiller on at the mains supply and at the switch on the rear of the control
unit (NOT using the Power switch on t he front of the control unit).
4. When switching on the chiller for the first time, select the display language by rotating
the Select/Set control on the front panel, and then pressing to select the desired
option.
5. The chiller controller will continue with its start-up sequence and then display
"Standby" when ready.
6. Set the safety set temperature on the controller using a flat screwdriver to rotate the
control to the maximum temperature to which the bath should be heated.
7. Press the Power switch on the front panel.
8. The pump will begin operating.
Page 79
Cooling Jacket Accessory . 79
9. Top up the coolant reservoir to compensate for the fluid in the external circuit.
10. Rotate the Select/Set control on the front panel until the Pump/AutoTune menu is
displayed.
11. Press the Select/Set control until the pump speed bar is highlighted.
12. Rotate the Select/Set control to adjust the pump speed, and press the control to
accept the new setting.
CAUTION: The heat exchanger of the TGA will NOT accept pressure. Use the
lowest flow rate that provides sufficient cooling for your experiment to minimize
the pressure in the system.
ATTENTION: L'échangeur de chaleur du TGA n'accepte PAS de pression.
Utilisez le débit le plus bas qui offre un refroidissement suffisant pour votre
expérience afin de minimiser la pression dans le système.
CAUTION: It is essential to maintain a very constant flow rate of coolant. The
liquid should be free of air bubbles. Changes in the liquid flow rate or the
presence of air bubbles will reduce the quality of the measurement signal.
ATTENTION: Il est essentiel de maintenir un débit de fluide de refroidissement
très constant. Le liquide doit être exempt de bulles d'air. Des modifications du
débit de liquide ou de la présence de bulles d'air réduiront la qualité du signal de
mesure.
Setting the Coolant Temperature
When using a circulator, the temperature must be controlled to within 0.2 °C (set point ± 0.1 °C).
1. Press and release the Select/Set control on the front panel.
2. The set point temperature value is highlighted.
3. Turn the control to set the temperature to the nearest whole degree Celsius.
4. Press the control again.
5. The degree fraction figures are highlighted.
6. Turn the control to select the desired fraction of a degree.
7. Press the control to accept the value.
You can now switch on the TGA and begin sample runs.
Page 80
80 . TGA 8000 Installation and Har d ware Guide
Accupik Accessory
The TGA 8000 Autosampler Accupik is an accessory designed to help protect the validity of your
aqueous or volatile samples. Samples are se aled in volatile sample pans especially designed to fit
the TGA 8000 crucib les. The crucibles then are positioned in the autosampler tray for the run.
The Accupik is a mechanical device that attaches to the analyzer. From a command in the Pyris
software, it pierces the sealed sample pan just before the crucible in which it resides is loaded
from the autosam pler to the hangdown wire and then enters the furnace. The puncturing of the
sample pan provides an additional level of protection against volatilization of samples. Now you
can program volatile sample analysis for unattended overnight runs. The Accupik attaches to the
underside of the balance plate by inserting the standoff in one hole and using a screw in the
other hole to secure it.
Mounting the Accupik
The Accupik attaches to the underside of the balance plate by inserting t he standoff in one
hole and using a scr ew in the other hole to secure it.
To attach the Accupik to the balance plate:
1. Locate the two holes on the underside of the balance plate.
2. Align the round standoff on the Accupik with the rear hole in the bal ance plate.
3. Using the screw pr ovide d , secure the Accupik in place with a screwdriver.
Page 81
Alignment Procedure
NOTE:
Accupik Access ory . 81
You must have a crimped sample pan in a crucible in the autosampler in order to align the
Accupik properly. Use the Volatile Sample Pan Sealer Accesso ry (P/N N5370435) to crimp a
sample pan (P/N N5370550).
1. When the Accupik is installed it should have at least 5.0mm space between the
crucible handle and needle assembly.
2. From the Pyris Software T GA 8000 Control Panel, click on the Autosampler Control
Page 82
82 . TGA 8000 Installation and Har d ware Guide
button .
The Autosampl er Co nt r o l dialog box is displayed.
Note that the Piercing Position and Piercing Depth buttons are grayed out (inactive).
3. Click Change to set the Piercing Mode to ON.
The Piercing Position and Piercing Depth buttons become active.
4. Make sure that a crimped sample pan (use the Accupik volatile sample pan sealer to
crimp a volatile sample pan) is in a crucible and t ha t the crucible is in position 1 in
the autosampler tray.
Page 83
Accupik Access ory . 83
NOTE:
The Accupik needs to be adjusted so that when the pan is punctured, the needle is close to
the middle of the sample pan in the crucible but slightly to the right of the crucible's handle.
There should be approximately 0.5 – 1 mm of space between the needle guide and the
crucible’s handle.
5. Enter 1 in S elect Carousel Location, then click the Piercing Position button.
The autosampler will rotate so that position 1 i s underneath the Accupik needle.
6. Loosen the black thumbscrew adjustment knob to carefully move the stem so that
the needle assembly (contains the needle tip) is slightly above the sample in the
crucible.
The space between the needle tip and the top of the sample tray should be
approximately 1 mm.
7. Click the Piercing Depth button.
This moves the autosampler tray up and down to pierce the sample.
8. Watch the position of the knob as the needle punctures the sample pan.
When the vertical direction or height of the needle is adjusted properly.
9. Check the sample, if it is pierced (there is a hole) and the sample pan is not
deformed by the sleeve it is OK. If it is, the Accupik's height needs to be adjusted.
Page 84
84 . TGA 8000 Installation and Har d ware Guide
Needle Replacement/Adjustment Procedure
The needle should be replaced if it is old or contaminated. If a different size pan is used, you
may need to adjust the needle in the stem in order to maintain the 5-mm space between the
needle's tip and the top of the pan.
1. Remove the Accupik to replace the needle.
2. Using a screwdriver, loosen the screw that is attached to the metal shaft. Pull the old
needle out and insert a replacement needle (P/N N5376668).
3. Replace and tighten the screw with the screwdriver.
4. Replace the Accupik onto the TGA 8000.
5. Check the alignment and adjust the Accupik as necessary.
Page 85
TGA 8000 Autosampler
85
Page 86
86 . TGA 8000 Installation and Har d ware Guide
atmosphère explosiv.
le déchargement d'une casserole d'échantillons dans le carrousel.
TGA Autosampler Safety Precautions
WARNING: Be sure that all instrument operators read and understand the following
precautions. It is advisable to post a copy of these precautions on or near the TGA.
AVERTISSEMENT: Assurez-vous que tous les opérateurs de l'instrument lisent et
comprennent les précautions suivantes. Il est conseillé d'afficher une copie de ces
précautions sur ou près de la TGA.
The following precautions must be observed when using the TGA Autosample r. These are in
addition to the precautions that apply to the TGA.
WARNING: Your TGA Autosampler is either installed at the factory or is designed
to be installed by a PerkinElmer Service Engineer if purchased after the TGA. Do
not try to install or uninstall the autosampler without the help of a Service Engineer.
AVERTISSEMENT: Votre TGA Autosampler est soit installé en usine, soit est
conçu pour être installé par un ingénieur de service PerkinElmer si acheté après le
TGA. N'essayez pas d'installer ou de désinstaller l'échantillonneur automatique
sans l'aide d'un technicien de mainten anc e.
WARNING: Do not remove the gripper or covers. No user serviceable parts are
inside the autosampler. Refer all servicing to a qualified service engineer.
AVERTISSEMENT: Ne retirez pas la pince ou les couvercles. Aucune pièce
réparable par l'utilisateur ne se trouve à l'intérieur de l'échantillonneur automatique.
Confiez toutes les réparations à un technicien qualifié.
WARNING: Never operate the TGA autosampler in an explosive atmosphere.
AVERTISSEMENT: Ne jamais utiliser l'auto-échantillonneur TGA dans une
CAUTION: Make sure that all cables and tubes are clear of the autosampler's path
when going to the Load and the Safe positions.
ATTENTION: Assurez-vous que tous les câbles et tubes sont dégagés du trajet de
l'échantillonneur automatique lorsque vous passez aux positions de chargement et
de sécurité.
CAUTION: Never use the Autosampler Control dialog box to control the
autosampler while it is loading or unloading a sample pan to the carousel.
ATTENTION: N'utilisez jamais la boîte de dialogue Contrôle automatique des
échantillons pour contrôler l'échantillonneur automatique pendant le chargement ou
Page 87
How the TGA 8000 Autosampler Works . 87
NOTE
How the TGA 8000 Autosampler Works
The TGA 8000 autosampler is computer controlled and fully automated. The sample tray can
hold up to 48 crucibles. Carousel positions are labeled 1 through 48 for easy sample
identification. New samples can be added to the carousel to replace samples that have been
analyzed by simply pausing the play list and manua lly controlling the autosampler through
the Autosampler Control dialog box. Once the autosampler is in the Safe position (all the
way to the right), you can lift off the sample ring that holds the crucibles.
You can use the stainless steel sample pan liners (P/N N5376749) in the crucibles to keep
them from getting dirty. Simply throw away the liner after use.
Autosampler Work Cycle
When you begin a sample run by starting a play list, the autosampler performs a typical work
cycle, which includes the following steps:
: When running samples with an autosampler, the status of the Antistatic Device should be
Auto Enabled and the Upper Fan should be Off.
1. Before the start of the run, the furnace is in the C ooling position or the Lowered
position. The autosampler is in the Safe position.
2. Start the play list. Your play list may have an explicit Load Sample comma nd or, if
using a Sample Group, the entry Sample List implies Load Sample. Upon reaching
this line in the list, the autosampler tray turns to position the selected tray location
properly.
3. The autosampler swings to the Load position.
4. The tray turns so that the crucible handle engages the hook in the hangdown wire.
5. The antistatic device turns on.
6. The furnace rises almost all the way up. It stays at that position for about 15
seconds while the antistatic device deionizes the joint area.
7. The antistatic device turns off and the furnace finishes traveling up to engage the
joint.
8. The method runs.
9. Your play list may contain an explicit Return Sample command. If using a Sample
Group, the Sample List entry has an implied Return Sample command.
10. At the end of the run, the antistatic device comes back on as the furnace goes to the
Lowered position.
11. The antistatic device turns off as the furnace g oes to the Cooling position.
12. The autosampler swings to the Load position with the selected location underneath
the crucible.
13. The tray rises so the crucible goes into the selected location.
14. The tray turns so the crucible hook comes off of the hangdown wire.
Page 88
88 . TGA 8000 Installation and Har d ware Guide
Sample Handling - TGA 80 00 Autos ampler
The TGA 8000 measures the change in weight of a sample as a function of temperature or
time. Proper sample preparation and handling are important for obtaining optimal results.
Sample Preparation
The thermogravimetric analyzer analyzes solid samples in powder, crystal, or granular form.
Although quantitative accuracy will remain the same r egardless of sample shape, the
qualitative appearance of a run may be affected by the sample configuration. Proper sample
preparation that maximizes the contact surface between the crucible and the sample will
reduce any imbalance of the sample in the crucible and will result in maximum peak
sharpness and resolution. The best sample form for optimum performance is powder or fine
granules. Solid materials can be sliced into small pieces with a razor or knife.
Sample Pans
The sample tray holds up to 48 crucibles. You can also place an aluminum liner (P/N N5370492)
or a stainless steel liner (P/N N5370495) within the crucible to lengthen its usability. The liner can
be used for runs under 600 °C.
Sample Atmosphere
Once the sample is loaded onto the hangdown wire and the autosampler returns to the Safe
position, the sample is subject to the same atmosphere as a sample in a standard TGA 8000.
The furnace rises up around the sample and the sheath gas, if used, baths the furnace liner;
that is, the gas goes between the Pyrex furnace tube a nd the liner, not in the furnace.
Sample Loading
Before loading the samples into the crucibles, you will have to tare the crucibles, that is,
have the system weigh them so the weight of the crucible is not included in the data. Load
the empty crucible s into the sample ring. You can do this with the ring off of the autosampler
assembly. After filling the locations, carefully place the ring over the gripper and settle the
ring into position. There is a locating pin in the autosampler that goes into the hole in the
tray. You could al so keep the ring on the autosampler and use tweezers to load the crucibles
into the ring. Make sure that the auto sampler is in the Safe position. Be careful with the
tweezers around the hangdown wire .
You can tare all of the crucibles in the tray using the Pyris Player Tare All feature. Your play
list should contain a Sample Group. If you are creating a new play list, select Sample Group as the first entry. If you are using an existing play list, it should have a Sample Group
in it. In a new play list, with Sample Group highlighted, click the Tare All button to display
the Advanced Tare Options dialog box. The tare weights are automatically put into the
Sample List. If you chose Populate from Tray in the Advanced Tare Options dialog box,
the Sample List is filled with a line for each posit ion in the sample ring that contains a
crucible and the tare weight of the crucible is included.
Page 89
Sample Handling - T GA 8000 Autosampler . 89
Once you have tared the empty crucibles, remove the sample ring, remove each crucible, load
the sample into the crucible, and return the crucible to the same location in the sample ring.
Return the sample ring to the autosampler. Now you can have the system weigh all the samples
before running the play list, or you can have each sample weighed at the beginning of its run.
With Sample List highlighted, click the Weigh List button. The TGA Tare/Weigh System dialog
box appears. The system automatically starts t he p rogram to weigh each sample in the list. After
the last sample, if there was a missing sample e ncountered, a message is displayed:
The following samples could not be weighed. It is possible that the crucibles are
empty.
A list of missing samples follows. Click the dialog box's Close button to clear the box.
Click the Done button in the Tare/Weigh System window. The Sample List will now display
the weights of each sample:
For efficient use of the autosampler and to increase sample throughput, you can purchase a
second sample ring. This way you can prepare a second set of samples while the first set is
running.
Page 90
90 . TGA 8000 Installation and Har d ware Guide
Running a TGA 8000 Autosampler Playlist
Pyris Player is the backbone of Pyris Software automation. It was created with autosamplers
in mind. In addition to the standard play list items – Load Sample, Run Method, Return
Sample – there is a Sample Group. The Sample Group simplifies grouping like samples
together (as you would have in a sample tray of the autosampler). These like samples use
the same test method and data analysis. A Sample Group consists of a Sample List and Data
Analysis List. A specific method is selected for the sample in the Sample List. The Data
Analysis List allows you to access all necessary functions for data recall, curve manipulation,
optimization, and calculations for automatic data analysis.
Before starting a run, perform the steps below:
• Review the safety and warning notes for the TGA 8000 and the autosampler
• Verify that the system purge and sample gas, tubing are properly connected.
• Turn on the gases; adjust the pressures. The recommended flow rate for the system
purge is 20 – 100 ml/min. For the samp le gas, it should be between 10 – 50 ml/min.
For the furnace lift mechanism, use a pressure between 15 and 30 psi.
•Verify that the electrical and cable connections between the autosampler and the
TGA 8000 and other components of the system are properly connected.
Turn on the power t o the system components in the following order:
• Computer
• TGA 8000
• Printer
1. Prepare your samples.
2. Start Pyris Software and click the TGA button.
3. Click the Pyris Player button on the toolbar: . Either open an existing play list
or create a new one.
4. There are many ways to use the play list with the autosampler. A quick way to
create a play list using the Sample Group feature is given below.
5. Load empty sample pa ns into the locations you want to use in the sampl e tray.
Create a new play list that contains only a Sample Group. Click the Tare All button.
In the Advanced Tare Opt ions dialog box, select This Group Only; Populate from Tray.
.
Page 91
Running a TGA 8000 Autosampler Playlist . 91
6. While taring, the TGA Tare/Weigh System screen is displayed.
7. The software will tare the crucibles it finds and populate the Sample List in the
Sample Group for you.
Page 92
92 . TGA 8000 Installation and Har d ware Guide
8. Remove the sample tray from the autosampler and place the samples in the
crucibles. Carefully place the ring back on to the autosampler. You can also remove
each crucible from the tray individually and load the sample into the crucible and
return it to the tray.
Now you can have the system weigh each sample in the list. You can also choose to
have each sample weighed right before the run starts. To weigh all the samples
before starting the play list, highlight the Sample List line. A message informing you
that a method has not been selected is displayed. You can select the method after
the samples have been weighed. Click Weigh List. The TGA Tare/Weigh System
screen appears and the system begins to weigh the samples listed in the Sample
List. When complete, the weights are displayed, along with the tare weights, in each
sample line.
9. Next, you need to add some items to the Data Analysis List. If you have not entered
a method for the Sample List, however, you cannot fill in the Data Analysis List. To
enter a method, highlight the Sample List line. Type in the name of the method in
the Method Name field or click the Browse button and find and select the method
you want to use. Edit the method's program and initial state parameters by clicking
the Edit Meth o d button. When finished editing, close the Method Editor by clicking
Page 93
Running a TGA 8000 Autosampler Playlist . 93
NOTE:
NOTE:
the window's Close button in the upper-right-hand corner .
With Data Analysis List highlighted, click Add a step. Select Display Curve. Use
the default Use Cur rent Run. This means that the curve from the current run will
be displayed after the run is over. Add a Pause li ne after this line if you want to be
able to view the curve.
10. Once you add Display Cu rve to the Data Analysis List, other items become
available to add to the list.
11. Save the play list by selecting Save Player from the File menu.
12. Start playback of the play list by clicking on the Start at Top button or the Start at Current Item button on the Player toolbar.
If you select the Start at Current Step button, the focused item must be a main-level item,
that is, Prepare Sample, Data Analysis, or Sample Group. However, if you are in a Sample
Group, you can start a play list from a Sample line in the Sa mple List.
If you select Go To Load, Go To Temp, or Hold at Temp from the control panel while a play
list is running, the current sample run will end and the playback of the play list ends.
Page 94
94 . TGA 8000 Installation and Har d ware Guide
TGA 8000 Autosampler Maintenance
Observe the following guidelines when cleaning the autosampler:
1. The black area of the autosampler is anodized aluminum. To clean it, you can use a
cloth dampened with isopropyl alcohol.
2. The sample tray is clear anodized aluminum which can be cleaned with isopropyl
alcohol.
3. To keep the ceramic crucibles clean, you can use the disposable a luminum liners
(P/N N5370492) or stainless steel liners (P/N N5370495).