PerkinElmer Photovac MicroFID User Manual

Page 1
Warning: Limitation of Liability
The ultimate responsibility of the consequences of use of toxic compounds rests with the user. PerkinElmer’s role is as a supplier of instrumentation to assist in the early detection of hazardous conditions involving such compounds.
MicroFID represents a major advance in this field and, as with all complex instruments, it is vitally important to ensure that MicroFID is maintained in accordance with PerkinElmer’s instructions and that proper calibration is regularly performed.
As with any complex device, MicroFID is subject to failure and, while PerkinElmer has taken, and continues to take, all possible precautions to (a) reduce the possibility of failure, and (b) warn the user in the event of failure, circumstances may occasionally occur in which there is a failure despite such precautions on PerkinElmer’s part. PerkinElmer regrets that it cannot accept liability for damages of any kind caused as a result of either failure of the user to follow instructions or of MicroFID to perform.
11
Page 2
Page 3
Photovac MicroFID
Portable Flame Ionization Detector
Part 396003 Rev. G
11
Page 4
Release History
Part Number Release Publication Date
396003 G February 2000 396003 F August 1997
User Assistance
PerkinElmer 761 Main Avenue Norwalk. CT 06859-0010 U.S.A.
Printed in Singapore
Notices
The information contained in this document is subject to change without notice. PerkinElmer makes no warranty of any kind with regard to the material, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in connection with furnishing, performance or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this publication may be reproduced in any form whatsoever or translated into any language without the prior, written permission of PerkinElmer, Inc.
Copyright © 2000 by PerkinElmer
Trademarks
Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are protected by law.
MicroFID is a trademark of PerkinElmer, Inc.
Page 5
Reference No. 600757 Rev G
33
Page 6
MicroFID Software Manual
Contents
Notices and Warnings.........................................................vii
Introduction.........................................................................15
Tutorial Session..................................................................38
Detailed Operation..............................................................63
Accessories.........................................................................85
Routine Maintenance........................................................107
Troubleshooting................................................................113
Technical Description.......................................................134
Appendices........................................................................142
Index...................................................................................156
Figures
Figure 1 MicroFID Layout..................................................29
Figure 2 MicroFID Refill Adapter......................................34
Figure 3 Refill Adapter 3-Way Valve.................................34
Figure 4 Default Display....................................................41
Figure 5 Keep MicroFID Level...........................................53
Figure 6 Printed Output for Interval Mode.......................56
Figure 7 Printed Output for Method 21 Mode .................58
Figure 8 Graphed Output for Interval Mode.....................61
Figure 9 Graphed Output for Method 21 Mode................62
Figure 10 Gas Bag Adapter...............................................67
Figure 11 Using the Gas Bag............................................68
Figure 12 Connecting MicroFID to a Serial Printer.........88
iv44
Page 7
Contents
Figure 13 Connection Description Dialog Box................91
Figure 14 Connect To Dialog Box.....................................92
Figure 15 Properties Dialog Box.......................................93
Figure 16 Properties Dialog Box.......................................94
Figure 17 Connecting Charcoal Filter..............................96
Figure 18 Connecting the Coalescing Filter....................97
Figure 19 Serial to Parallel Converter..............................99
Figure 20 Connecting MicroFID to the Serial to Parallel
Converter...........................................................................102
Figure 21 Bar Code Reader and Interface Module........103
Figure 22 Hydrogen Purge Outlet...................................111
Figure 23 Inlet Filter.........................................................112
Figure 24 Parallel Port Signals.......................................130
Figure 25 Connecting MicroFID Using the Null Modem
Cable...................................................................................132
Figure 26 Connecting MicroFID Using the Gender
Changer Cable...................................................................132
Figure 27 Flame Ionization Detector..............................136
Figure 28 Block Diagram.................................................137
Figure 29 Bar Code Reader Control Drawing 890047....154
v
55
Page 8
MicroFID Software Manual
Tables
Table 1 Instrument Status.................................................41
Table 2 TUTOR Displays....................................................44
Table 3 Check List for Field Operation............................82
Table 4 Additional Field Items..........................................83
Table 5 DIP Switch Settings for Baud Rate...................100
Table 6 DIP Switch Settings for Parity ..........................100
Table 7 DIP Switch Settings ...........................................101
Table 8 MicroFID I/O Port Pinouts .................................125
Table 9 Parallel Port Signals...........................................129
Table 10 Serial Port Signals ...........................................130
Table 11 Pin Definitions...................................................133
Table 12 Contact Information..........................................147
Table 13 MicroFID Response Factors............................152
vi66
Page 9
Notices and Warnings
77
Page 10
Page 11
Notices and Warnings
CONFORMS TO
UL STANDARD 913
Intrinsically safe/Sécurité intrinsèque for use in Class I, Division 1, Groups A, B, C, D, Hazardous Locations, Temperature Code T4, Exia
CERTIFIED TO CSA STANDARD CSA 22.2 No. 157
FCC Warning
This equipment has been tested and found to comply with the limits for a Class B Digital Device, pursuant to Subpart B, Class B of Part 15 of the FCC rules. These limits are designed to provide reasonable protection against harmful interference when the equipment is operated in a commercial environment. This equipment generates, uses and can radiate radio frequency energy and if not installed and used in accordance with the instruction manual, may cause harmful interference to radio communications. Operation of this equipment in a residential area is likely to cause harmful interference in which case the user will be required to correct the interference at his own expense.
MicroFID I/S & I/SC Notice
MICROFID I/S IS CLASSIFIED FOR USE IN CLASS I, DIVISION 1, GROUPS A, B, C, D HAZARDOUS LOCATIONS.
It has been listed by ETL® to comply with Underwriters Laboratories® Inc. UL® 913 Standard for Intrinsically Safe
Apparatus and Associated Apparatus for use in Class I, Division 1, Groups A, B, C, D Hazardous (Classified) Locations, Fourth
Edition.
ix
Page 12
MicroFID Software Manual
MICROFID I/SC IS CLASSIFIED FOR USE IN CLASS I, DIVISION 1, GROUPS A, B, C, D HAZARDOUS LOCATIONS. T4 (135ºC) RATING.
It has been listed by ETL to comply with CSA Standard 22.2 No.
157-92 Intrinsically Safe and Non-Incendive Equipment for Use in Hazardous Locations.
MICROFID I/S AND I/SC ARE NOT INTENDED TO DETECT COMBUSTIBLE LEVELS OF GASES. MICROFID I/S AND I/SC ARE CLASSIFIED FOR USE IN ATMOSPHERES CONTAINING COMBUSTIBLE LEVELS OF GASES.
These accessories are for use with MicroFID I/S and IS/C in a hazardous location:
MX395002 Shoulder Strap MX396001 Short Sample Probe MX396007 Carrying Case MX396011 Calibration Kit MX396018 Long Sample Probe MX396021 Charcoal Filters (2 pieces) MX396022 Charcoal Filters (10 pieces) MX396053 Headset MX396054 Bar Code Reader and Interface Module
(I/S)
Do not use any other accessories with MicroFID I/S or I/SC in a hazardous location.
Substitution of components may affect safety rating.
x
Page 13
CAUTION
Notices and Warnings
To reduce the risk of fire or injury to persons, read and follow these instructions:
1. All calibration, maintenance and servicing of this device, including battery charging, must be performed in a safe area away from hazardous locations.
2. For replacement battery pack use only Part No. MX396005 (I/S), MX396006 (I/SC).
3. Do not dispose of the battery pack in a fire. The cell may explode. Check with local codes for possible special disposal instructions.
4. Do not open or mutilate the battery pack. Released electrolyte is corrosive and may cause damage to the eyes or skin. It may be toxic if swallowed.
5. Exercise care in handling battery packs in order not to short the terminals with conducting materials such as rings, bracelets and keys. The battery or conductor may overheat and cause burns.
xi
6. Do not defeat proper polarity orientation between the battery pack and battery charger.
7. Charge the battery pack using the charger provided with or identified for use with this product only in accordance with the instructions and limitations specified in this manual. For battery charger use only Part No. MX396014 (110 Volts AC), MX396013 (220 Volts AC).
Page 14
MicroFID Software Manual
MICROFID I/SC EST CLASSIFIÉ POUR USAGE DANS LES EMPLACEMENTS DANGEREUX DE CLASSE I, DIVISION 1, GROUPES A, B, C, D. ÉVALUATION T4 (135ºC).
. Quatrième édition.
MicroFID ISC est conforme à la norme des Underwriters Laboratories Inc. UL 1203
MicroFID I/SC est conforme à la norme de CSA Standard 22.2 No. 157-92 - Intrinsically Safe and Non-Incendive Equipment for Use in Hazardous Locations.
MICROFID I/SC C’EST NE PAS INTENDER POUR DÉTECTER DES NIVEAUX DE COMBUSTION DES GAZ. CET APPAREIL EST CLASSIFIÉ POUR USAGE DANS DES ATMOSPHÈRES CONTENANT DES NIVEAUX DE COMBUSTION DES GAZ.
Les accessoires PerkinElmer suivants peuvent également être utilisés avec l’appareil dans un emplacement dangereux:
NOTICE
xii
MX395002 Bandoulière MX396001 Petite Gamme d’éc hantillons MX396007 Étui de Transport MX396011 Equipment de Calibration MX396018 Gamme d’échantillons MX396021 Filtres de Charbon (2 pièces) MX396022 Filtres de Charbon (10 pièces) MX396054 Lecteur de Codes à Barres et
Module d’I nterface (I/S)
MX396053 Casque telephonique
Ne pas utiliser d’autres accessoires avec cet appareil dans un emplacement dangereux.
Page 15
MISE EN
GARDE
Notices and Warnings
La substitution des composantes peut nuire à la sécurité d’e mploi.
Afin de réduire les risques d’incendie et les blessures, lire et suivre ces instructions:
1. Tout étalonnage, entretien et réparation de cet appareil, y compris le chargement de la pile, doit être effectué dans un endroit sûr, à l’ écart des zones dangereuses.
2. N’utiliser que la pièce PerkinElmer numéro MX396006 (I/SC) lorsqu’il faut remplacer le bloc-pile.
3. Ne pas jeter la pile dans un feu. La cellule pourrait exploser. Vérifier les codes locaux, qui peuvent comporter des instructions de mise au rebut particulières.
4. Ne pas ouvrir et ne pas abîmer le bloc-pile. L’acide sulfurique qui risque de s’en échapper est corrosif et peut causer des lésions aux yeux et à la peau. Cet acide est toxique lorsqu’avalé.
xiii
5. Manipuler le bloc-pile avec soin, afin de ne pas court­circuiter les bornes avec des matériels conducteurs tels qu’une bague, un bracelet ou des clés. La pile ou le conducteur pourraient surchauffer et causer des brûlures.
6. Ne pas modifier l’orientation de la polarité appropriée entre le bloc-pile et le chargeur.
7. Charger le bloc-pile fourni avec ou reconnu pour usage avec ce produit seulement conformément aux instructions et restrictions spécifiées dans ce manuel. Pour le chargeur, n’utiliser que la pièce numéro MX396014 (110 Volts AC), MX396013 (220 Volts AC).
Page 16
MicroFID Software Manual
CONFORMS TO COUNCIL DIRECTIVE 89/336/EEC COUNCIL DIRECTIVE 91/157/EEC
CE Directive
Council Directive 89/336/EEC - Electromagnetic Compatibility and the Applicable Standards. MicroFID complies with EN50082-01
Generic Immunity Standard for Electromagnetic Compatibility Part 1: Light Industrial Environment. MicroFID has been tested and conforms to the guidelines specified in EN50082-02 Generic Immunity Standard for Electromagnetic Compatibility Part 2: Industrial Environment
Council Directive 91/157/EEC - Recovery and Disposal of Spent Batteries
You must use the 220 V battery charger (Part No. MX396013) in order to comply with the requirements of the applicable Council Directives.
xiv
Page 17
Introduction
1
Page 18
Page 19
Introduction
About this Manual
This manual provides detailed instructions for the setup, operation and maintenance of the MicroFID Portable Flame Ionization Detector.
Before unpacking the instrument, please read Warnings and Safety Practices on page 18. This section describes possible hazards that may injure the user, damage the instrument or compromise its operation. Some general safety information is also provided.
To help you learn to use MicroFID quickly this manual is organized by tasks beginning with a tutorial in Chapter 2. More detailed operational instructions are provided in Chapter 3. Connecting accessories is covered in Chapter 4. Routine maintenance is covered in Chapter 5. Troubleshooting techniques are covered in Chapter 6. Chapter 7 provides a technical description of MicroFID.
The MicroFID manual uses a few conventions for key names on the keypad and for text that is shown on the display.
UPPERCASE Key names are denoted by uppercase text.
ARROW keys is the collective name for the UP ARROW and DOWN ARROW keys.
“ Display Text” Text that appears on the MicroFID display
is in quotation marks.
<Angle Brackets> Computer keyboard names are denoted by
angle brackets, e.g. <Ctrl>.
MD FID Text that must be typed in using the
computer keyboard is shown in italics.
17
Page 20
MicroFID Software Manual
In the text you will find various warnings and notes.
A warning indicates an operation that could cause personal injury if precautions are not followed.
WARNING
A caution indicates an operation that could cause damage to the instrument if precautions are not followed.
CAUTION
NOTE: A note indicates significant information.
Warnings and Safety Practices
18
Please read this section before operating MicroFID.
Approved Models of MicroFID
This manual provides operational information for all models of MicroFID. The MicroFID I/S and the MicroFID I/SC are intrinsically safe and approved for use in hazardous locations. Refer to the manual introduction for details of each approval.
Throughout the manual, notes are provided to inform you of the limitations of usage for the MicroFID I/S and MicroFID I/SC models.
Page 21
WARNING
Introduction
If the MicroFID you are using is not specifically identified as intrinsically safe, do not use it in a location were flammable concentrations of gases and vapors may exist.
Compressed Gases
Cylinders of compressed gas, such as hydrogen and calibration gas, must be handled with extreme care. When using the calibration gas bag adapter, take care not to kink or stress the tubing. For safety, the hydrogen and calibration gas cylinders must be secured before use.
Please observe the following handling procedures:
Do not mutilate cylinders.
Do not expose the cylinders to direct sunlight.
Do not heat the cylinders. The cylinders may rupture at high
temperatures.
Use only the specified regulator for the calibration gas. Confirm
regulator type and material with your specialty gas supplier. Use only the MicroFID refill adapter for the hydrogen (Part No. MX396004).
Always secure cylinders before removing the cylinder valve
protection cap
Do not drag or roll cylinders. Use a cylinder hand truck to move
large cylinders.
Wear safety glasses when working with compressed gases.
Store cylinders in an upright position.
Do not store cylinders in a hazardous location.
19
Page 22
MicroFID Software Manual
Store cylinders away from possible sources of ignition.
Keep regulators and related equipment in the same gas service.
Do not change service or adapt equipment without consulting your gas supplier.
20
Page 23
Introduction
Regulators for Compressed Gases
When connecting a regulator to a large cylinder:
Ensure cylinder valve and regulator connection match.
Ensure regulator construction materials are compatible with the
gas, and that the cylinder pressure gauge will withstand the cylinder pressure.
Never use the regulator as a shut-off valve. Close the cylinder
when it is not in use.
Do not subject the regulator to an inlet pressure greater than
recommended.
Do not move or detach the regulator when it is pressurized or
when it is in use.
Before connection, ensure the gas cylinder valve and the
regulator CGA connection are clean.
Turn the pressure control valve on the cylinder all the way out
(close the cylinder). Turn the regulator outlet to off. Open the gas cylinder valve slowly and check for leaks. Adjust the delivery pressure and then open the regulator outlet valve.
Hydrogen Gas
Hydrogen gas is a fire and explosion hazard when exposed to heat or flame. The lower explosive limit (LEL) is 4%. The lower explosive limit is the minimum concentration of gas or vapor in air that will ignite in the presence of a source of heat or sparks.
Refer to the Material Safety Data Sheet (MSDS) before handling this gas. You will receive the MSDS from your gas supplier when you order the gas.
Calibration Gas
21
Page 24
MicroFID Software Manual
Adequate ventilation must be provided when MicroFID is being calibrated.
If compound threshold limit values (TLV) are exceeded, you should use a gas bag for sampling and calibration.
To determine the TLV of the compounds contained in the calibration gas, refer to the Material Safety Data Sheet (MSDS) supplied with your calibration gas cylinder. See Calibration on page 65 for details of calibration using a gas bag.
Battery Pack Care
Do not leave battery packs uncharged for an extended period of time. This will result in damage to the battery packs and possible loss of logged data.
Leaving MicroFID for more than 3 months, without turning it on, may result in the loss of recorded data and setup parameters. If MicroFID is not used for long periods of time, turn the instrument on for a few hours every month to avoid loss of data. See Battery Charging on page 28 for instructions on charging the battery.
22
Please observe the following:
1. For replacement battery pack use only Part No. MX396005 or MX396006.
2. Do not dispose of the battery pack in a fire. The cell may explode. Check with local codes for possible special disposal instructions.
3. The battery pack is 57% lead (Pb) by weight. This battery pack must be disposed of properly. Check with local codes for special disposal instructions.
4. Do not open or mutilate the battery pack. Released electrolyte is corrosive and may cause damage to the eyes or skin. It may be toxic if swallowed.
5. Exercise care in handling battery packs in order not to short the terminals with conducting materials such as rings, bracelets
Page 25
Introduction
and keys. The battery or conduct or may overheat and cause burns.
6. Do not defeat proper polarity orientation between the battery pack and battery charger.
7. Charge the battery pack using the charger provided with or identified for use with this product only in accordance with the instructions and limitations specified in this manual. For battery charger use only Part No. MX396014 (110 V AC), MX369013 (220 V AC).
Excessive Heat and Cold
Do not expose the instrument to intense sunlight for prolonged periods.
Exposure to excessive heat may result in erroneous readings.
At low temperatures, water vapor, a by-product of the hydrogen flame, may condense at the exhaust port. At sub-zero temperatures the water vapor will freeze and obstruct the exhaust port. If the exhaust port becomes obstructed, pump operation will be inhibited. Flame out may also result.
Transporting MicroFID
When you transport MicroFID, you should empty the internal hydrogen cylinder and then refill it when you arrive at your destination.
If you are transporting MicroFID in the carry case or in any enclosed container, you must empty the hydrogen cylinder and fill it when you arrive at your destination
If you are traveling by passenger aircraft, you must empty the cylinder. You cannot transport MicroFID by passenger aircraft with hydrogen in the cylinder. Refer to Emptying the Hydrogen Cylinder on page 110 for more details.
Oxygen Concentration Limits
23
Page 26
MicroFID Software Manual
A minimum of 17% oxygen is required to start the hydrogen flame. The oxygen is supplied from the sample as it is drawn in by the pump. A minimum of 10% oxygen is required to maintain the hydrogen flame. An oxygen deficiency will reduce the height of the flame or cause the flame to be extinguished and may affect the displayed reading.
If you will be using the MicroFID in a highly contaminated area where it is possible that the oxygen content is below 10%, watch for indications of reduced flame height such as lowered detection limits or a flame out fault.
Flammable Gases
High concentrations of flammable gases (gases within their flammable range) can act as an additional fuel source. When this happens, the flame height may increase beyond the confines of the combustion chamber. The hydrogen supply will then be cut off and the flame will go out.
Flame out may also occur when the concentration of sample gas is so great that it causes an oxygen deficiency. This may occur when sampling enclosed or confined spaces where vapors and gases cannot escape. Watch for indications of increased flame height such as erratic readings or sudden high concentrations followed by a flame out fault.
24
Unpacking
The following accessories are included with your MicroFID:
1. Shoulder Strap
2. Sample Probe
3. MicroFID Instrument Manual
4. Hydrogen Refill Adapter
5. MicroFID Multi-Tool
Page 27
Introduction
6. Battery Charger with AC Line Cord
7. Replacement Sample Inlet Filters (10 pieces)
Ensure all of these accessories have been included with the instrument. If any items are missing or damaged, contact PerkinElmer immediately.
Support Equipment and Consumables
Hydrogen Fuel
You must obtain a tank of hydrogen from which you can fill the internal cylinder. When ordering hydrogen specify ultra high purity hydrogen, 99.999% pure. This grade of hydrogen is also referred to as Grade 5 or zero zero grade. The hydrogen must have less than 1 ppm hydrocarbon contamination.
You can obtain the hydrogen in various size cylinders. Regardless of the cylinder size, the cylinder must have a fill pressure of at least 1800 psig in order to completely fill the MicroFID internal cylinder. Specify a tank with no more than 2400 psig.
The hydrogen cylinder must also have CGA 350, male outlet.
The MicroFID refill adapter (Part No. MX396004) is shipped with the instrument. You cannot fill the MicroFID hydrogen cylinder without the refill adapter.
If the hydrogen supply tank is more than 6 months old it should be replaced with a new cylinder.
Calibration
For normal operation these items are required:
1. Calibration Kit (Part No. MX396011). The calibration kit contains a regulator, a gas sampling bag and a gas bag adapter.
2. Calibration gas containing the appropriate concentration of methane. The exact concentration of the calibration gas will
25
Page 28
MicroFID Software Manual
depend on your application. 500 ppm methane in air is available from PerkinElmer. (Part No. MX396028)
Some applications may require calibration gas at more than one concentration. When ordering calibration gas, specify methane in hydrocarbon free air. Air should have not more than 0.1 ppm total hydrocarbons (THC).
Method 21 protocol requires that commercial cylinders of calibration gas be analyzed and certified to be within +/- 2% accuracy and a shelf life must be specified on the cylinder. At the end of the shelf life, the cylinder must be replaced or re­analyzed.
If you will be using a portable cylinder, specify a regulator with a C-10, 5/8" - 18 UNF valve inlet and a 1/4” parallel, compression fitting outlet. A regulator, matching this specification, is included in the calibration kit.
If you will be using large tanks of gas, specify a single stage, high purity regulator with a CGA 590 connection at the inlet. The regulator should also have a 1/4” p arallel, compression fitting outlet with which to connect the regulator to the gas bag adapter. The delivery pressure must be adjustable to approximately 40 psig (275 kPa).
26
3. Zero air. Occasionally, clean, ambient air is suitable for calibration. If you are unsure about the quality of ambient air, you can connect a charcoal filter (Part No. MX396021 or MX396022) to MicroFID during the first step of calibration.
Alternatively you can use an additional gas sampling bag and a source of hydrocarbon free air. Air should have not more than
0.1 ppm total hydrocarbons (THC).
If you will be using large tanks of gas, specify a single stage, high purity regulator with a CGA 590 connection at the inlet. The regulator should also have a 1/4” p arallel, compression fitting with which to connect the regulator to the gas bag adapter. The delivery pressure must be adjustable to approximately 40 psig (275 kPa). You should obtain an
Page 29
Introduction
additional gas bag and gas bag adapter. (Part No. MX396017 ­gas bag, MX396010 - gas bag adapter)
Field Operation
For field operation a MicroFID Field Kit (Part No. MX396058) is available. The field kit includes a printer cable, a long sample probe, a carrying case and a calibration kit.
Refer to the check list in Preparing for Field Operation on page 82. Ensure you have all the necessary accessories and equipment before beginning field operation.
Printer
MicroFID may be used with a serial printer. The printer must have a print width of at least 65 characters and a proportionally spaced font. A serial printer with an RS-232 port, can be connected directly to MicroFID. See Printer on page 87 for details of connecting a printer to MicroFID.
If you are using a parallel printer, you will need the serial to parallel converter (Part No. MX380145). See Serial to Parallel Converter on page 98 for details.
NOTE: MicroFID is not classified for use in hazardous locations with a
printer.
Computer
MicroFID may also be connected to a computer. The computer must be 100% compatible with an IBM PC. You can use the printer cable to connect MicroFID to the computer.
You will also need terminal emulation software. If you are using Microsoft® Windows you do not need to purchase any separate software. If you are already using another type of communication or terminal emulation software package, it is not necessary to purchase separate software for MicroFID. See Computer on page 88 for details.
27
Page 30
MicroFID Software Manual
NOTE: MicroFID is not classified for use in hazardous locations with a
computer.
Battery Charging
Before beginning operation of MicroFID, the battery pack must be charged. You can connect the battery charger to the battery pack while the battery pack is still attached to MicroFID. You can also remove the battery pack and replace it with a fully charged, spare pack (Part No. MX396005 or MX396006) and charge the battery pack independently of the instrument.
Removing and Replacing the Battery Pack
NOTE: Do not remove or recharge the battery pack in a hazardous
location.
To remove the battery pack:
1. Stop the flow of hydrogen gas by turning the hydrogen shut­off valve fully clockwise. See Figure 1. If the instrument has been turned on, turn it off by pressing the On/Off switch twice.
28
2. Use the MicroFID multi-tool to loosen the two captive screws in the bottom of the battery pack.
3. A retainer at the rear of the instrument helps to secure the battery pack to the instrument. Free the battery pack from the instrument.
Page 31
Introduction
500
1000
1500
FULL
2500
psi
INTRINSICALLYSAFE/SECURITE,INTRINSÈQUE/Exia CLASSI,DIVISION1, GROUPSA, B, C,D ,T 4, HAZARDOUSLOCATIONS
IS-3000
I/O Connector
Shoulder Strap
Connector
On/Off Switch
Keypad
Shoulder Strap
Connector
Exhaust/Flame
Arrestor
Contents Gauge
Sample Inlet
Refill Connector
Battery Charger
Jack
Battery Pack
Hydrogen Shut-Off
Valve
Figure 1 MicroFID Layout
4. Connect the charged battery pack to the retainer at the rear of the instrument.
5. Use the MicroFID multi-tool to tighten the two captive screws into the bottom of the battery pack. Do not over­tighten the screws.
Charging the Battery Pack
To charge the battery pack:
1. Ensure the correct plug is installed on the line cord of the battery charger. If it is not correct for the wall outlet in your area, then it must be replaced. See Installing Alternate AC Plug on the Battery Charger on page 148.
NOTE: Do not charge the battery pack in a hazardous location. Use only a
MicroFID battery charger.
2. Plug the charger into the jack located on the front of the battery pack.
3. Plug the charger into an AC outlet.
The LED, on the battery pack, indicates the charge state. Red indicates the battery is being charged. Green indicates the battery is fully charged and ready for use.
29
Page 32
MicroFID Software Manual
It is normal for a fully charged battery to indicate it is charging (red light) when first plugged in. The LED will turn green as the battery charges.
4. When the battery pack is fully charged remove the charger, first from the wall outlet then from the battery pack.
5. If you removed the battery pack from the instrument to recharge it, you can replace the battery pack now or when necessary.
Charging a fully discharged battery pack will take approximately 8 hours. Leaving the charger connected to a charged battery pack will not harm the battery or the charger in any way. If a battery pack is to be left indefinitely, leave it connected to the charger so that it will be fully charged and ready for operation.
On average a fully charged battery pack will provide 15 hours of continuous operation. Battery life is shorter if the instrument is turned off and then on again repeatedly.
WARNING
30
Filling the Hydrogen Cylinder
The Refill Adapter
DO NOT attempt to fill the hydrogen cylinder without a refill adapter!
The refill adapter (Part No. MX396004) consists of an adapter fitting with a left handed thread for cylinder attachment, a three­way valve and a flexible hose. The three-way valve can be positioned to deliver gas from the cylinder to the port on the MicroFID. Alternatively, the valve can be turned to release gas held in the hose, after filling. A flow restrictor is also incorporated in line with the hose in order to limit the rate of gas flow.
Page 33
WARNING
Introduction
DO NOT MODIFY OR DISASSEMBLE THE REFILL ADAPTER! If you have any problems, the unit must be returned for repair or replacement.
31
Page 34
MicroFID Software Manual
Connecting the Refill Adapter to the Hydrogen Tank
NOTE: Read through this section before filling the internal cylinder.
Store the hydrogen tank in a well ventilated area, well away from heat or possible ignition sources.
Hydrogen gas is a fire and explosion hazard when exposed to heat or flame. The lower explosive limit is 4%.
WARNING
1. Connect the refill adapter to the tank of hydrogen. Hydrogen is usually supplied with a CGA 350 cylinder valve outlet. The refill adapter is supplied with the matching fitting for ease of connection.
NOTE: Do not force the connection.
Do not use Teflon tape with CGA fittings. In general, these fittings are designed for metal to metal sealing.
Do not use adapters to connect one CGA fitting to another type of CGA fitting. If the refill adapter does not match the outlet on your hydrogen tank contact PerkinElmer.
2. Tighten the refill adapter onto the tank with a wrench. Do not over tighten.
3. Hold the flexible hose and open the cylinder. Ensure there are no leaks. Use soap solution to leak test the connection.
32
Page 35
WARNING
WARNING
Introduction
DO NOT use an open flame to test for leaks!
Filling the Internal Hydrogen Cylinder
Do not fill the hydrogen cylinder in a hazardous location.
1. Turn MicroFID off.
2. Place MicroFID on a sturdy, flat surface within easy reach of the flexible filling hose.
3. Turn the refill adapter 3-way valve so that it points away from the cylinder. See Figure 3.
4. Hold the delivery hose and turn the 3-way valve on the refill adapter so that it points towards the cylinder. Open the cylinder valve. A steady flow of gas will be heard escaping from the end of the flexible hose. This serves to purge the hose of any impurities. Return the 3-way valve to point away from the cylinder.
33
Page 36
MicroFID Software Manual
Quick Connect
Flexible
Hose
Flow Restrictor
1/4" NPT Thread
Female Fitting
3-Way Valve
CGA 350
Adapter
Pneumatic
Silencer
Off/Vent
Hydrogen
Tank
Hydrogen
Tank
Deliver through
flexible hose
Figure 2
MicroFID Refill Adapter
34
5. Hold the knurled collar of the MicroFID refill connector and slide it in towards the instrument.
Figure 3
Refill Adapter 3-Way Valve
6. Insert the fitting at the end of the flexible hose into the port until it seats against the port. You will hear a click.
Page 37
WARNING
Introduction
7. Release the knurled collar and allow it to slide up to its original position.
8. Turn the 3-way valve on the refill adapter so that it once again points towards the cylinder and watch the pressure increase on the MicroFID Contents Gauge.
DO NOT leave the MicroFID unattended during the filling procedure! It is possible to overfill the cylinder and damage MicroFID.
9. The contents pressure MUST NOT exceed 2400 psig and the cylinder valve must be closed before this occurs.
10. When the contents gauge on MicroFID indicates Full (approximately 1600-1800 psig), turn the 3-way valve to point away from the cylinder. This will stop the flow of gas to MicroFID. High pressure gas trapped in the flexible hose will be vented. A check valve in the MicroFID prevents gas in the reservoir from escaping.
11. Close the cylinder valve on the tank of hydrogen.
12. AFTER relieving the gas pressure in the hose, disconnect it from MicroFID. Pull the knurled collar down and extract the fitting.
The internal cylinder is now full and will provide approximately 11 hours of operation.
If you have not used MicroFID for six months or longer, empty the hydrogen cylinder and refill it before beginning operation. See Emptying the Hydrogen Cylinder on page 110 for more details.
Overview
35
Page 38
MicroFID Software Manual
MicroFID measures the concentration of airborne organic, combustible gases and vapors and automatically displays and records these concentrations. It does not distinguish between individual pollutants. The reading displayed represents the total concentration of all ionizable chemicals present in the sample. MicroFID does not distinguish between individual pollutants.
MicroFID operates automatically. The user reads concentrations directly from the display. The display updates itself twice per second.
MicroFID has many datalogging options. You can select an averaging interval, you can log all data with no averaging or you can use the Method 21 mode. If you select an averaging interval, the minimum, maximum, and average concentrations measured in each period are automatically recorded in MicroFID’s datalogging memory, if the datalogger is turned on. If you are recording all data with no averaging you will only be able to log data for a short time before the datalogger becomes full. If you are averaging data you can log data for a longer period of time. The exact duration will depend on the averaging interval.
36
The Method 21 mode prompts you to locate a site and then to record a background and sample reading for the site. You can record up to 750 such entries. There is no averaging of data in Method 21 operation.
Recorded data can be played back from memory on MicroFID’s display, sent to a printer in either tabular or graphical format or sent to a computer. Data are recorded by date, time, and by an event name. Event names can be entered using the MicroFID keypad, an external bar code reader or a computer. Data are played back by specifying the start event name.
The keypad is used to set up and calibrate MicroFID, and allows you to manipulate the concentrations measured and recorded by MicroFID in various ways. MicroFID has 16 keys for direct numeric entry and for using MicroFID’ s functions.
All information entered from the keypad, the bar code reader or the computer and stored in MicroFID’s memory is retained when the
Page 39
Introduction
instrument is switched off. The clock and calendar continue to operate and do not need to be reset each time MicroFID is used.
37
Page 40
Tutorial Session
2
Page 41
Page 42
MicroFID Software Manual
Beginning Operation
To print or review logged data, you can operate MicroFID without the flame. You conserve the hydrogen this way.
If you choose to start the flame, MicroFID will attempt to ignite the flame once you have turned on the flow of hydrogen gas. If MicroFID has been unused for a while it is possible that the gas supply lines are filled with air. If the flame cannot be started, MicroFID will begin a 30 second purge cycle. During the purge cycle it will flush the gas supply lines with hydrogen. After the purge cycle it will attempt to light the flame again. If it fails again, another purge cycle will be performed and MicroFID will try a third time to ignite the flame.
To operate MicroFID:
1. Turn the instrument on by pressing the front of the On/Off switch. When the instrument is powered up, the version number and creation date of the instrument software are displayed. Press ENTER.
40
2. You will be prompted to start the flame. If you do not want to start the flame, use the ARROW keys to select “N o Flame Needed” and press ENTER. To start the flame use the ARROW keys to select Start Flame and press ENTER.
3. If you selected “ Start Flame”, MicroFID will prompt you to start the flow of hydrogen. Turn the shut-off valve counterclockwise to start the flow of hydrogen and press ENTER. See Figure 1 for the location of the shut-off valve.
4. The pump will start and MicroFID will then ignite the flame. You will hear a small pop when the flame has been ignited. Once the flame has been started the message “D etector flame has been started OK” will be displayed followed by the default display, shown in Figure 4.
Page 43
Tutorial Session
Instrument Status Detected Concentration
Time
Date
Event Name
Figure 4
Default Display
The default display provides the following information: instrument status, current detected concentration, Event name (if the datalogger is on), time, and date. If an event name is longer than three characters the bottom line of the display will scroll through the information.
The instrument status appears at the left of the upper line of the display and on the Print and Graph outputs. Each status has a priority assigned to it. If more than one status is in effect, then the status with the highest priority is displayed until the condition is corrected or until the option is turned off.
When the bar graph is selected, the instrument status is reduced to a one-letter code.
Status Code Priority Description
NoFlm N 1 Flame has not been ignited.
Check K 2 One of 4 faults is occurring. Press the
Over O 3 Detector electronics have been saturated.
TUTOR key for details.
Table 1
Instrument Status
41
Page 44
MicroFID Software Manual
Status Code Priority Description
Cal C 4 This status will never be observed on the
Alarm A 5 Detected concentration exceeds the set
LoBatt L 6 Battery pack power is low. Connect
Play P 7 The instrument is playing back previously
Endlog E 8 The datalogger is full. Sniff S 9 High Sensitivity operation. Ready R 10 Normal operation. Locate L 11 Locate site. Used for Method 21 operation BkGnd B 12 Record background reading. Used for
Sample S 13 Record sample reading. Used for Method 21
Max M 14 Displays maximum detected concentration.
display during normal operation as various calibration prompt messages are displayed while MicroFID is calibrating. If the instrument is turned off during calibration, Cal will appear on the display when MicroFID is turned on again indicating the last calibration was incomplete. “C al” status is also shown on printed and graphed output.
alarm level.
MicroFID to AC supply or replace the battery pack.
recorded data.
Method 21 operation
operation
42
Table 1
Instrument Status - continued
To turn the instrument off:
1. To turn the instrument off, press the front of the On/Off switch. You will then be prompted to turn of the flow of hydrogen.
Page 45
Tutorial Session
2. Stop the flow of hydrogen gas by turning the hydrogen shut-off valve fully clockwise.
3. Pressing the On/Off switch twice to power down the instrument.
Tutorial Session
TUTOR Key
To assist you in learning the key functions, MicroFID has a built-in tutorial session that displays a two-line description of the function of each key. Pressing the TUTOR key begins a tutorial session and pressing the EXIT key after any other key, ends the session. While in the tutorial session, key presses have no effect on MicroFID’s operation.
To begin a tutorial session:
1. Turn on MicroFID and select the “ No Flame” option.
2. Press the TUTOR key and begin a tutorial session.
3. Press each key and read the display. The tutorial display for each key is provided in Table 2.
If a fault is encountered during normal operation, press the TUTOR key to display a description of the fault. If more than one fault is occurring at the same time, press the TUTOR key to see each of the faults in turn. When all the faults have been viewed, press EXIT to return to the normal display.
43
Page 46
MicroFID Software Manual
Key Tutor
Description
BATT Shows battery V
normally 6-8.9V
DISPLAY Displays conc as
graph or numeral
LIGHT Shows detector
flame status
AUDIO Selects alarm or
tone or no audio
PRINT Prints table of
recorded data
GRAPH Prints graph of
recorded data
PLAY Replays recorded
data on display
CAL Calibrates with
zero & span gas
Table 2 TUTOR Displays
Key Tutor
Description
ALARM Shows set-point
for conc alarm
SETUP Sets date time
& options for keys
MAX Displays
highest conc measured
EVENT Sets options for
recording data
TUTOR Press a key then
read explanation
CLEAR Erases the last
number pressed
EXIT Exits from
tutorial session
ENTER Confirms
display then continues
44
Key Functions
Press a key to execute a function. If there are no options to the key function, the key acts immediately. The BATT key is one example. If there are options, the current option is displayed. You are prompted to display the other options by pressing the ARROW keys. Pressing ENTER confirms that the displayed value is correct and moves to the next option.
If the function requires numeric input, the current value is displayed. You can change the value using the numeric keys on the
Page 47
Tutorial Session
keypad. Pressing ENTER confirms that the displayed value is correct and moves to the next key option.
Some functions have multiple steps for options and/or numeric inputs. These are arranged so that the most frequently changed inputs are displayed first. Once the desired changes have been made you can bypass the rest of the steps by pressing EXIT.
Each key function is described in more detail in the following sections. Leave MicroFID on and try each key in turn.
BATT Key
Press the BATT key to open a window that displays the battery voltage. The voltage will be 9 volts when the instrument is fully charged. When the voltage drops below 7.2 volts, a “ LoBatt” status will be displayed and the battery pack must be recharged. See Battery Charging on page 28.
NOTE: Do not remove or recharge the battery pack in a hazardous
location.
An audio signal will be generated when “LoBatt” occurs. The signal will not clear until you connect the battery pack to an AC supply or replace the battery pack with a fully charged one.
NOTE: Do not leave the battery pack uncharged for an extended period of
time. This will result in damage to the battery pack.
DISPLAY Key
If a numerical display is shown, pressing DISPLAY will change it to a bar graph. If the bar graph is shown, pressing DISPLAY changes it to a numerical display. The top line of the bar graph display shows the instrument status and the concentration range. The bar graph range is selected with the SETUP key.
In the bar graph display, the instrument status becomes a one-letter code. See Table 1 for a description of the instrument status codes.
45
Page 48
MicroFID Software Manual
The bottom line of the display remains unchanged in bar graph mode and will continue to scroll and display data.
LIGHT Key
Press the LIGHT key to display the flame status. If the flame has been ignited and is burning, the status will be Detector flame is lit. If you are not analyzing, the flame status will be “Detector Flame is not lit”.
EXIT Key
The EXIT key terminates all MicroFID functions. When you press the EXIT key the display reverts to the default display. Most functions exit automatically if no key is pressed for 15 seconds.
When EXIT is pressed during printing, MicroFID stops sending information to the printer. The printer will continue to print until its buffer is empty.
46
CLEAR Key
CLEAR erases the last entry. If a value is entered in error, press CLEAR to erase the entry and then re-enter the correct number.
CLEAR used in conjunction with the MAX key clears the Max register.
MAX Key
Press the MAX key. The maximum concentration, the event during which it was encountered and the time and date of the occurrence will be displayed. Max is shown for 15 seconds and then the display reverts to the default display.
Page 49
Tutorial Session
Press MAX then press CLEAR twice to reset the Max register. “M ax Cleared” will be displayed with the current date and time. After 15 seconds the display reverts to the default display.
Recording of real time data is not interrupted when the MAX key is pressed or when the Max register is cleared.
AUDIO Key
If an “ Alarm” status, “LoBatt” status or a flame out fault occurs, an audio signal will be generated. Under these conditions you will hear the audio signal even if a headset is not connected.
If you have a “L oBatt” status, the audio signal will not clear until you connect the battery pack to an AC supply or replace the battery pack with a fully charged one.
If a flame out fault has occurred, the audio signal will not stop until the flame has been re-started or you turn MicroFID off.
In the case of an alarm condition, an audio signal will be generated when an alarm is encountered.
To turn off the audio signal on alarm:
1. Press the AUDIO key.
2. Select “A udio Off” with the ARROW keys and press ENTER.
ALARM Key
Press the ALARM key to display the current alarm level or to enter a new alarm level. The displayed alarm level is specific to the current Cal Memory. When the Cal Memory is changed, the alarm level associated with the newly loaded Cal Memory will be displayed and used during sampling.
To display the current alarm level or to enter a new alarm level:
47
Page 50
MicroFID Software Manual
1. Press CAL and select the desired Cal Memory. Press ENTER and then press EXIT. You do not want to calibrate MicroFID at this point.
2. Press ALARM to display the current alarm level. If this value is correct, press EXIT to return to the normal display.
3. If a new value is to be set, enter the new value, and press ENTER. This alarm level is specific to the current Cal Memory.
When an alarm condition is detected, the instrument status changes to “A larm” and an audio signal is generated. The “Alarm” status will not clear until the detected concentration falls below the set alarm level.
PLAY Key
The PLAY key plays back previously recorded data.
To play back previously recorded data:
1. Press PLAY. Two options are available. Pressing ENTER begins playback where it was last stopped. Press the SETUP (*) key to set the playback options.
2. Select the start Event. MicroFID will display a list of available events. Use the ARROW keys to scroll through the list. When the start event you want is displayed, press ENTER.
NOTE: An event name may not be available if it has been deleted.
3. If data has been averaged over an interval, select which value is to be displayed, either the Minimum, the Average, or the Maximum, and press ENTER. If you are playing back Method 21 data select Background, Sample or Difference to be displayed.
4. The data can be played back in either numerical or bar graph display. Press the DISPLAY key to select the desired option.
48
Page 51
Tutorial Session
5. Press the ARROW keys to play back the data. You can move forward or backward through the datalogger.
6. Press EXIT at any time to return to normal operation.
If the flame is on, MicroFID is measuring and recording real time concentrations when the instrument status is “P lay”. If, during playback, an instrument status with a priority higher than that of “P lay” is encountered in real time, it will be displayed. MicroFID will however, continue to play back recorded data.
The PLAY function provides a speed search to review data or to find the desired start and stop event names for printing or graphing.
SETUP Key
The setup key is used to enter analysis information. Press SETUP and step through the options. Press ENTER to accept the displayed data or enter a numerical value using the keypad and then press ENTER. If no values are entered the display reverts to the default display after 15 seconds.
To enter analysis information:
1. The first option is for the detector. If you have been using the instrument with no flame and you now need to start the flame, use this option. You can also use this option to start the flame after a flame out fault. If you do not want to change the status of the detector use the ARROW keys to select “No Change”. Press ENTER to move to the next option.
2. The next option is Repair. Use Repair to record the state of repair. You must define the repair codes and ensure all users are aware of the codes. This option is most useful for Method 21 monitoring.
For example, a code of 0 may represent no repair under way, 1 may represent first attempt at repair with 5 days, 2 may represent final repair attempt within 15 days, 3 may indicate irreparable damage, and a 4 may represent a state in which a repair can not be attempted until the next scheduled shut down.
49
Page 52
MicroFID Software Manual
3. Next, select the operating range. This option sets the full-scale range for the bar graph display, the graph output. Use the ARROW keys to select a Range of 20, 200, 2000, 10000, 50000 ppm. This does not affect MicroFID’s numeric display and datalogging.
4. Select the Interval next. You can average data over intervals of 1 second (s) 5 s, 10 s, 15 s, 30 s, 1 minute (m), 2 m, or 5 m. You may also select Method 21 mode.
If you select Method 21, the datalogger will be cleared. If you do not want to lose your previously recorded data, press EXIT. Then print or save the data to disk before changing the interval to Method 21. See PRINT Key on page 53, GRAPH Key Mode on page 58 and Computer on page 88 for more information on printing and saving logged data.
NOTE: The datalogger will be cleared when you change from Interval
mode to Method 21.
Method 21 mode first prompts you to locate a specific sampling site and then allows you to record both a background and sample entry in the datalogger. Method 21 mode will be covered in more detail on page 72.
50
5. The date option is next. Enter the current day, month and year. Press ENTER to confirm the entry. The date is retained when the instrument is turned off. You do not need to enter this information every time you use MicroFID.
6. Next enter the correct time. Press ENTER.
EVENT Key
Events may be used to identify a particular sample or sampling location in memory. An event name is used in Method 21 operation to identify a sampling location. Recorded data are played, printed and removed from the datalogger by specifying a start and stop event name.
Page 53
Tutorial Session
The EVENT key controls MicroFID’ s datalogger. Press the EVENT key to see the event options.
To display or add events:
1. Press the ARROW keys until the “In sert” option is displayed.
2. To insert an event, press ENTER. You will be prompted to enter an event name. You can enter a numerical name directly from the MicroFID keypad or you can connect MicroFID to a computer and enter the event name using the STAR 21 Software (Part No. MX396027). An event name may be up to 16 characters long. Refer to the Software Manual for the STAR 21 software for details of using the software with MicroFID.
3. Press the ARROW keys to move to the next option.
If you want to delete a range of events you must select a start and stop event. Use the ARROW keys to scroll through the list of events to select the start event and then to select the stop event.
NOTE: Deleted information cannot be recovered. You should play back
and print or download the contents of the datalogger before deleting any information.
4. You can delete the current event, all recorded events or a range of events. Press ENTER then use the ARROW keys to select the desired option.
5. You can turn the datalogger off by selecting Stop. MicroFID will continue to analyze, but no data will be stored. To turn the datalogger on again press the EVENT key and use the ARROW keys to select Start. MicroFID will begin logging data again.
6. You can find a specific event in the datalogger using the “Fin d” option. Use the ARROW keys to scroll through the list of events to select the desired event. When you press ENTER, the selected event will be displayed. Use this option in Method 21 operation to find an event for monitoring or re-monitoring.
51
Page 54
MicroFID Software Manual
7. The next option is “Scan” . You can use a bar code reader (Part No. MX396054) to scan a bar coded event name. See Bar Code Reader on page 103 for details of connecting the bar code reader to MicroFID.
8. To activate the bar code reader, use the ARROW keys to select “Scan ”. Press ENTER to activate the bar code reader. It will remain activated until you scan a code. If no bar code is scanned within 15 seconds the bar code reader will be deactivated. This is done to conserve battery life. “Scan ”, when used with the bar code reader has the same function as the “I nsert” option.
MicroFID can record continuously until the datalogger is full. When the datalogger is full the instrument status will change to “ Endlog”. The datalogger will stop logging data when it is full.
If it is necessary to retain a copy of recorded data, the data should be printed or stored in a computer when “E ndlog” is displayed. This will prevent loss of information.
52
CAL Key
The CAL key is used to calibrate MicroFID. Calibration has been automated for ease of operation. Before beginning calibration ensure you have a reliable source of zero air and calibration gas. A brief description of the functions under the CAL key will be provided here. For more information on calibrating MicroFID, see Calibration on page 65.
When you press the CAL key you will first be prompted to select a Cal Memory. Each Cal Memory stores a unique zero point, sensitivity setting, response factor and alarm level.
You will then be prompted to enter a response factor. Refer to Response Factors on page 149 for a list of response factors. If the compound is not in this list or you are measuring gas mixtures, enter a value of 1.00. The concentration detected by MicroFID will be multiplied by the response factor before it is displayed and logged.
Page 55
Tutorial Session
Next select Low Range or High Range operation. Use Low Range if you are sampling concentrations between 0.5 and 2000 ppm (methane equivalents). Use High Range if you are sampling concentrations between 10 and 50,000 ppm (methane equivalents).
You will now be prompted to connect a supply of zero air. You can use ambient air, a charcoal filter (Part No. MX396021, MX396022) to clean ambient air or, for best results, use a clean sampling bag filled with zero grade air.
After MicroFID has set its zero point, you then enter the concentration of the calibration gas and then connect the calibration gas. MicroFID sets its sensitivity for the specified Cal Memory. You must have a supply of calibration gas ready before calibrating MicroFID. See Calibration on page 25 for detailed calibration instructions.
When calibrating MicroFID, ensure the instrument is level. If MicroFID is tilted from side to side, gravity will affect the flame height and cause erroneous readings.
Figure 5 Keep MicroFID Level
53
Page 56
MicroFID Software Manual
PRINT Key
MicroFID has been designed for use with a serial printer. The printer must have a print width of at least 65 characters and a proportionally spaced font. If you are using a parallel printer, you will also need the serial to parallel converter (Part No. MX380145).
See Printer on page 87 for instructions on connecting and configuring a printer for use with MicroFID. See Serial to Parallel Converter on page 98 for instructions on connecting the serial to parallel converter to MicroFID and to the printer.
When MicroFID is printing data it is still measuring real time concentrations.
NOTE: MicroFID is not classified for use in hazardous locations with a
printer.
54
To print data:
1. Turn MicroFID off.
2. Use the MicroFID printer cable (Part No. MX396051) to connect the MicroFID I/O connector to the printer.
3. Turn MicroFID on and press the PRINT key and then the SETUP (*) key to select the desired setup options.
4. MicroFID will ask for the start and stop Events. Use the ARROW keys to scroll through the list of events to select the start event and then to select the stop event.
5. Enter an ID number if desired. Any number up to 16 digits long may be entered as an ID number.
The ID number can be used, for example, to differentiate between users or instruments. If more than one person is
Page 57
Tutorial Session
using a single instrument, each person can use an individual number for printed reports. If more than one MicroFID is being used, each instrument can have its own ID number on printed reports.
If an ID number is not required, leave this option blank and move to the next option by pressing ENTER.
6. Enter the baud rate. This value is specific to the printer being used and must be set correctly. Refer to the printer user’ s manual to determine the correct baud rate
7. Enter the parity. The parity is specific to the printer being used and must be set correctly. Most printers will have parity set to none.
8. When the setup is correct, ensure the printer is on-line and press ENTER.
MicroFID stores one set of readings (Min, Avg and Max) each interval. In each averaging interval the printed minimum is the minimum of all the stored readings in that interval. The printed Avg is the average of all the recorded average readings and the Max is the maximum of all the recorded maximum readings. In Method 21 operation, the background, sample and the difference are stored.
When you are using interval mode, the date and time are printed along with the event name, repair code and the minimum, average and maximum concentration of each interval. The date, time, event name, and repair code are only printed when they change.
55
Page 58
MicroFID Software Manual
Date Time Event REP Stat Min Avg Max
Nov 28,94 16:06 0.0 1.1
22.6
91.7
50.1
22.0
7.6
1.2
0.0
0.0
128.0
503.1
520.5
519.5
518.9
Photovac MicroFID Analysis Report
(All units are PPM)
ID: 100
16:07 0.0 11.6
Ready 1.7 22.5
Cal 8.0 16.0
1.2 3.7
0.0 0.4
0.0 0.0
16:08 0.0 0.0
112 HiSens 0.0 6.7
2.9 173.0
17.4 417.9
Cal 485.5 515.0
489.8 505.2
56
Figure 6 Printed Output for Interval Mode
When you are using Method 21 mode, the date and time are printed along with the event name, repair code and the background, difference and sample concentration of each interval. The date, time, event name, and repair code are only printed when they change. See Figure 7.
While the information is being printed, the display shows that printing is in progress. The keypad will not accept commands until the present print job has been completed.
Page 59
Tutorial Session
Pressing EXIT during printing stops the job. The printer will continue to print until its buffer is empty.
57
Page 60
MicroFID Software Manual
Date Time Event REP Stat BkGnd Diff Sample Nov 30,94 13:29 100 Ready 1.7 15.4 17.2
Dec 1, 94 10:39 109 0.2 0.0 0.2
Photovac MicroFID Analysis Report (All units are PPM)
ID:100781
13:35 101 1.7 15.8 17.6 13:48 102 0.8 0.0 0.8 14:13 103 1.6 8.2 9.8 14:53 104 3 Alarm 6.0 2464 2470 15:54 105 Ready 0.1 0.0 0.1 16:05 106 0.2 20.9 21.2 16:12 107 0.2 0.0 0.2 16:56 108 1.5 19.6 21.1
10:57 110 0.1 98.6 98.8 11:13 111 122.0 215.7 337.8 11:29 112 2 Alarm 55.8 2474 2530 11:35 113 2 13.9 809.3 823.2 12:00 114 Ready 0.2 154.5 154.8 12:21 115 0.2 20.8 21.0 12:32 116 0.2 0.0 0.2
Figure 7 Printed Output for Method 21 Mode
GRAPH Key Mode
Pressing the GRAPH key prints the recorded data in graphical format. When MicroFID is graphing recorded data it is also measuring real time concentrations.
NOTE: MicroFID is not classified for use in hazardous locations with a
printer.
To graph logged data:
1. Turn MicroFID off.
2. Use the MicroFID printer cable (Part No. MX396051)
to connect the MicroFID I/O connector to the printer.
58
Page 61
Tutorial Session
3. Turn MicroFID on and press the SETUP key to set the
range for the graph. Use the ARROW keys to select the 20, 200, 2000, 10,000 or 50,000 ppm range.
4. Press the GRAPH key and then the SETUP (*) key to
select the desired setup options.
5. MicroFID will ask for the start and stop events. Enter
the desired event and press ENTER. Use the ARROW keys to scroll through the list of events to select the start event and then to select the stop event.
6. Enter an ID number if desired. Any number up to 16
digits long may be entered as an ID number.
The ID number can be used, for example, to differentiate between users or instruments. If more than one person is using a single instrument, each person can use an individual number for printed reports. If more than one MicroFID is being used, each instrument can have its own ID number on printed reports.
If an ID number is not required, leave this option blank and move onto the next option by pressing ENTER.
7. Enter the baud rate. This value is specific to the printer
being used and must be set correctly. Refer to the printer user’ s manual to determine the correct baud rate.
8. Enter the parity. The parity is specific to the printer
being used and must be set correctly. Most printers will have parity set to none.
9. When the setup is correct, ensure the printer is on-line
and press ENTER.
MicroFID stores one set of readings (Min, Avg and Max) for each averaging interval. In each averaging interval the graphed minimum is the minimum of all the stored readings in that interval. The graphed Avg is the average of all the recorded average readings and the Max is the maximum of all the recorded maximum readings. In
59
Page 62
MicroFID Software Manual
Method 21 operation, the background, sample and the difference are stored.
In Interval mode the date and time are printed along with the event name and the repair code. The date, time, event name, and repair code are only printed when they change. See Figure 8.
A minus sign is used to draw the graph from the minimum to the average. Plus signs are used to graph the area between average and maximum.
In Method 21 mode the date and time are printed along with the event name and the repair code. The date, time, event name, and repair code are only printed when they change. See Figure 9.
A minus sign is used to draw the graph from the background to the difference. Plus signs are used to graph the area between difference and sample. Difference is calculated by subtracting the background from the sample.
60
Page 63
Tutorial Session
Photovac MicroFID Analysis Report
(All units are PPM)
ID:100
Date Time Event REP Stat [0---------------2000] Nov 29,94 13:36 10 Ready -
13:37 -
13:38 Cal ----+++
13:39 -
Ready -
13:40 -
13:41 --+++++
13:42 Cal -
13:43 11 HiSens ---------+++++++++++
-----
-----++++++++
-+
-+
---+++
-++
-----++
-
Figure 8 Graphed Output for Interval Mode
While the information is being printed, the display shows that printing is in progress. The keypad will not accept commands until the present print job has been completed.
Pressing EXIT during printing stops the job. The printer will continue to print until its buffer is empty.
61
Page 64
MicroFID Software Manual
Date Time Event REP Stat [0----------------- 2000] Nov 30,95 13:29 100 Ready -
Dec 1,94 10:39 109 -
Figure 9 Graphed Output for Method 21 Mode
Photovac MicroFID Analysis Report (All units are PPM)
ID: 100781
13:35 101 ­13:48 102 ­14:13 103 ­14:53 104 3 Alarm ------------------------+ 15:54 105 Ready ­16:05 106 ­16:12 107 ­16:56 108 -
10:57 110 -+ 11:13 111 --++ 11:29 112 2 Alarm -----------------------+ 11:35 113 2 ----------+
62
Page 65
Detailed Operation
3
Page 66
Page 67
Detailed Operation
Calibration
General Information
MicroFID must be calibrated in order to display concentration in ppm units equivalent to the calibration gas. First a supply of zero air, which contains no ionizable gases or vapors, is used to set MicroFID’s zero point. Then, calibration gas, containing a known concentration of an ionizable gas or vapor, is used to set the sensitivity.
Occasionally clean ambient air will be suitable as zero air. Due to MicroFID’s sensitivity, outdoor air is usually unsuitable for calibration. A charcoal filter (Part No. MX396021, MX396022) may be connected to the instrument to produce clean air from otherwise unsuitable ambient air. For best results, use a commercial source of zero grade air and a second sampling bag. Zero air should have not more than 0.1 ppm total hydrocarbons (THC).
Methane in air is recommended as span gas. The concentration of the calibration gas will depend on your application. When ordering calibration gas, specify methane in hydrocarbon free air. Balance air should have not more than 0.1 ppm total hydrocarbons (THC).
Method 21 protocol requires that commercial cylinders of calibration gas be analyzed and certified to be within +/- 2% accuracy and that a shelf life must be specified on the cylinder. At the end of the shelf life, the cylinder must be replaced or re­analyzed.
65
Page 68
MicroFID Software Manual
Preparing the Calibration Gas Bag and the Zero Air Bag
Use the calibration kit (Part No. MX396011) as follows:
Observe proper handling techniques for all gases! See Warnings and Safety Practices on page 18.
WARNING
1. Connect the regulator to the calibration gas cylinder.
If you are using a portable tank of calibration gas, connect the regulator supplied with the calibration kit.
If you are using a large cylinder of calibration gas, you must obtain a high purity regulator as specified in Support Equipment and Consumables on page 25. Methane in air is usually supplied with a standard CGA 590 cylinder valve outlet. Obtain a regulator with the matching fitting. Connect the regulator to the tank of calibration gas.
NOTE: Do not force the connection.
Do not use Teflon tape with CGA fittings. In general, these fittings are designed for metal to metal sealing.
Do not use adapters to connect one CGA fitting to another type of CGA fitting. If the regulator does not match the outlet on your calibration tank, contact your specialty gas supplier.
2 Tighten the regulator onto the tank with a wrench. Do not over-tighten.
3. Attach the knurled nut on the gas bag adapter to the regulator. Finger-tighten the fitting.
66
Page 69
Detailed Operation
1/8" to 3/16"
Union
1/8" Nut
To Regulator
Teflon Ferrules Required Here
Gas Bag
Tube Stub
1/8" Adapter
Tubing
Figure 10 Gas Bag Adapter
4. Loosen the knurled nut on the reducing union of the gas bag adapter.
NOTE: Do not remove the nut from the union, as the Teflon ferrules contained inside the nut may be lost. See Figure 10.
5. Insert the tube stub from the gas bag into the knurled
nut. Tighten the knurled nut and ensure the tube stub is secure. If the gas bag is not secure, ensure you have inserted the tube stub far enough into the knurled nut. Do not over tighten the fitting.
CAUTION
Over-tightening the Teflon ferrules will result in damage to the ferrules.
6. The union should be connected to the gas bag adapter.
If it is not, then tighten the nut on the adapter tube to the union.
7. Open the gas bag valve. Turn the knurled plastic knob
counter clockwise to loosen it, then use the knurled collar on the valve tube to gently push the valve tube toward the bag.
67
Page 70
MicroFID Software Manual
Locking Ring
Tube Stub
Gas Bag Valve
Injection Port
Septum Retainer
Replacing the Septum
Remove retainer and
replace silicone septum.
Place septum with
Teflons ide down.
Replace retainer
Do not overtighten.
Overtightening may result
in needle blocakge
Replace septum after
5-10 injections.To replace
septum, turn retainer counter-
clockwise. For replacement
septa order Teflon bonded to
Silcone, diameter = .375" (9.5 mm)
Figure 11 Using the Gas Bag
8. Turn the regulator knob counter clockwise about half a
turn to start the flow of gas. Fill the gas bag about half full and then close the regulator.
9. Disconnect the gas bag from the gas bag adapter and
empty the bag. Flush the bag a few times with the calibration gas and then fill it.
10. Close the gas bag valve. Gently pull the valve tube
away from the bag, and then turn the knurled plastic knob clockwise to tighten it against the valve tube.
11. Remove the knurled nut on the adapter tube from the
regulator.
12. Repeat this procedure, if necessary, to prepare a bag of
zero air.
NOTE: Do not use the same gas bag or gas bag adapter for the bag of zero
68
air. You will contaminate the bag of zero air.
Page 71
Detailed Operation
Calibrating MicroFID
To calibrate MicroFID:
1. Press CAL and select the desired Cal Memory. MicroFID has 10 Cal Memories and can be calibrated with 10 different span gases or response factors if desired. Only one Cal Memory can be used at a time. Each Cal Memory stores a different response factor, zero point, sensitivity, and alarm level
2. Enter the desired response factor and press ENTER. Refer to Response Factors on page 149 for a list of Response Factors. If the compound is not in this list or you measuring gas mixtures then enter a value of 1.00. The concentration detected by MicroFID will be multiplied by the response factor before it is displayed and logged.
3. Select Low Range or High Range and press ENTER. Use Low Range if you are sampling concentrations between 0.5 and 2000 ppm (methane equivalents). Use High Range if you are sampling concentrations between 10 and 50,000 ppm (methane equivalents)
4. Connect the supply of zero air. If you are using room air press ENTER.
5. If you are using a charcoal filter, connect the filter as outlined in Charcoal Filters on page 95. Press ENTER and MicroFID will set its zero point.
NOTE: The charcoal filter does not filter methane or ethane. If these
compounds are present, use a gas bag with a supply of commercial zero air.
6. If you are using a gas bag with zero air, connect the gas bag adapter to the inlet. Open the bag and press ENTER. MicroFID sets its zero point.
7. MicroFID then asks for the span gas concentration. Enter the known span gas concentration and then connect the gas bag adapter to the inlet. Open the bag.
69
Page 72
MicroFID Software Manual
8. Press ENTER and MicroFID sets its sensitivity.
NOTE: Readings may fluctuate slightly as the gas bag empties. Do not
allow the MicroFID to evacuate the bag completely.
9. When MicroFID’ s display reverts to normal, it is calibrated
10. Press the ALARM key and enter the alarm level for the
Programming the Cal Memories
MicroFID has 10 Cal Memories and can be calibrated with 10 different span gases or response factors if desired.
To program the Cal Memories:
1. Prepare the bags of calibration gas as outlined in Preparing
and ready for use. Remove the span gas bag from the inlet.
selected Cal Memory.
the Calibration Gas Bag and the Zero Air Bag on page 66. Use a different gas bag and gas bag adapter for each concentration and for each type of calibration gas.
You can use the same gas bag to zero all the Cal Memories; however, you must refill the bag for each Cal Memory.
2. Press CAL and select the desired Cal Memory (1 to 10) with the ARROW keys.
3. Enter the desired response factor and press ENTER. See Response Factors on page 149.
NOTE: It does not matter which Cal Memory is selected or which response
factor is entered, MicroFID’ s response is not specific to any one compound. The reading displayed represents the total concentration of all ionizable compounds in the sample.
4. Select the Low Range or High Range and press ENTER. The range will depend on the concentration of the calibration gas. Use Low range if you are sampling concentrations between 0.5 and 2000 ppm (methane equivalents). Use High Range if you are sampling
70
Page 73
Detailed Operation
concentrations between 10 and 50,000 ppm (methane equivalents).
5. You will now be prompted to connect a supply of zero air. You can use ambient air, a charcoal filter (Part No. MX396021, MX396022) to clean ambient air or, for best results, use a clean sampling bag filled with zero grade air.
6. After MicroFID has set its zero point, you then enter the concentration of the calibration gas and then connect the calibration gas. MicroFID sets its sensitivity for the specified Cal Memory. The span gas concentration, response factor and alarm level entered here are specific to the selected Cal Memory only.
When the calibration is completed it is automatically stored in the selected Cal Memory.
7. Press the ALARM key and enter the alarm level for the selected Cal Memory.
8. Repeat this procedure for each Cal Memory you need.
When calibrating MicroFID, ensure the instrument is level. If MicroFID is tilted side to side, gravity will affect the flame height and cause erroneous readings
Whenever the instrument is calibrated, MicroFID updates the selected Cal Memory only. Each Cal Memory must be calibrated at least once a day. Frequency of calibration will depend on ambient conditions and instrument response. If ambient conditions change or the response has drifted, a calibration must be performed for each Cal Memory to ensure reliable operation.
High Sensitivity Operation
MicroFID can be used as a high sensitivity leak detector. In high sensitivity operation, MicroFID does not read in ppm units equivalent to the calibration gas, but displays a reading proportional to the total concentration of ionizable gases and vapors detected.
71
Page 74
MicroFID Software Manual
During calibration, no span gas is required. MicroFID zeros its reading with zero air and then sets itself to maximum sensitivity.
To use the High Sensitivity option:
1. Press SETUP. Select the 0-20 ppm display range with the
2. Press CAL. Under Cal Memory, use the ARROW keys to
3. Press EXIT. Select the bar graph with the DISPLAY key.
4. Press CAL and calibrate MicroFID with zero air. You
As MicroFID samples air closer to the leak, the length of the shaded area on the display increases.
ARROW keys and press ENTER.
select High Sensitivity and press ENTER.
should use a supply of commercial zero grade air and a gas sampling bag when zeroing MicroFID for high sensitivity operation.
MicroFID’s 3 second response time and detection limit of 0.5 ppm methane permit fast detection of small leaks.
Method 21 Operation
Method 21 is a US EPA protocol for the determination of volatile organic compound (VOC) leaks in process equipment. MicroFID has a Method 21 mode that facilitates Method 21 measurements. The STAR 21 Software (Part No. MX396027) provides more comprehensive data management consistent with the Method 21 protocol. Contact PerkinElmer for more information.
You must be familiar with the Method 21 protocol to use MicroFID for Method 21 monitoring. You must use the specifications outlined in the Method 21 documentation for programming MicroFID’ s Cal Memories and monitoring the sites.
As part of Method 21 operation, you must setup MicroFID to monitor various locations. Since each location may contain different
72
Page 75
Detailed Operation
compounds and concentration ranges you will store a Cal Memory and the associated response factor and alarm level as part of each event. In this way you can sample numerous locations without having to re-calibrate MicroFID at each location.
You can enter the monitoring schedule in one of two methods. Regardless of the method you use to enter the monitoring schedule; you must calibrate the required Cal Memories before you enter the monitoring schedule.
Calibration
You must determine the number of calibration standards that will be required to perform Method 21 monitoring for your site. Refer to the Method 21 documentation for details.
To program your cal memories:
1. Prepare the bags of calibration gas as outlined in Preparing the Calibration Gas Bag and the Zero Air Bag on page 66. Use a different gas bag and gas bag adapter for each concentration and for each type of calibration gas.
You can use the same gas bag to zero all the Cal Memories; however, you must refill the bag for each Cal Memory.
2. Press CAL and select the desired Cal Memory (1 to 10) with the ARROW keys.
3. Enter the desired response factor and press ENTER. See Response Factors on page 149.
NOTE: It does not matter which Cal Memory is selected or which response
factor is entered, MicroFID’ s response is not specific to any one compound. The displayed reading represents the total concentration of all ionizable compounds in the sample.
4. Select the Low Range or High Range and press ENTER. The range will depend on the concentration of the calibration gas. Use Low Range if you are sampling concentrations between 0.5 and 2000 ppm (methane equivalents). Use High Range if you are sampling
73
Page 76
MicroFID Software Manual
5. Follow the displayed calibration instructions. When the
6. Press the ALARM key and enter the alarm level for the
Whenever the instrument is calibrated, MicroFID updates the selected Cal Memory only. Each Cal Memory must be calibrated at least once a day. Frequency of calibration will depend on ambient conditions and instrument response. If ambient conditions change or the response has drifted, a calibration must be performed for each Cal Memory to ensure reliable operation.
Programming the Monitoring Schedule in the Field
concentrations between 10 and 50,000 ppm (methane equivalents).
calibration is completed it is automatically stored in the selected Cal Memory. The span gas concentration and response factor entered here are specific to the selected Cal Memory only.
selected Cal Memory.
74
The easiest way of entering the Method 21 information is to go to each site and enter the required information. In this way, the schedule will be entered into the instrument in the exact manner in which it will be used.
To program the monitoring schedule:
1. Press the SETUP key. Press ENTER until “I nterval” is displayed. Use the ARROW keys to select Method 21 as the Interval. Press ENTER.
2. You will be prompted to clear the datalogger. If you
want to save the contents of the datalogger, press EXIT. You must print or save the data to disk. See PRINT Key on page 53, GRAPH Key Mode on page 58 or Computer on page 88 for details of printing or saving your data. Press ENTER to clear the datalogger.
3. Go to the first site to be monitored. You can enter the sites in any order; however, using the schedule after it has been
Page 77
Detailed Operation
stored is easier if the sites are stored in the order in which they will be monitored.
4. Press the EVENT key. Use the ARROW keys to select “In sert”. Press ENTER. Use the keypad to enter an event name up to 16 characters long. To obtain alphabetic characters you must use the STAR 21 software (Part No. MX396027).
If your site is bar coded, use the ARROW keys to select “S can”. Press ENTER to activate the bar code reader. The bar code reader will remain activated until you scan a code. If no bar code is scanned within 15 seconds the bar code reader will be deactivated. See Bar Code Reader on page 103 for details of connecting the bar code reader to MicroFID.
It is possible to enter an event name that already exists in the datalogger. When you select “In sert” or “Scan” you can enter the name of an event that already exists in the datalogger. MicroFID will not overwrite the existing event. It will insert an event with the same name. In this way you can create multiple copies of the same schedule in the datalogger. You can see the duplicate events when you print the contents of the datalogger.
5. Press the CAL key and select the calibration memory to be used for this site. Press ENTER and then press EXIT.
6. Press SETUP and enter a 2-digit repair code for this location. See SETUP Key on page 49 for a description of Repair code. This step is optional.
7. When you return to the default display, the instrument status will be “ Locate”. You should be at the site. If not, locate the site. If the event name is greater than 3 characters the bottom line of the display will scroll from right to left.
8. Once you have located the designated site, press ENTER. The instrument status will change to “BkGnd” . A background measurement must be made as specified in the Method 21 protocol. When you have an accurate background, press ENTER. MicroFID will record the maximum background concentration when you press the ENTER key.
75
Page 78
MicroFID Software Manual
9. The instrument status will now be “Sample ”. Take a sample
10. The instrument status will again be Locate. At this time,
11. Press EXIT at anytime to return to the beginning of the
12. To store the next location, go the next site and press the
13. Press the CAL key and select the calibration memory to be
14. Press SETUP and enter a repair code for this location.
measurement, as specified in the Method 21 protocol. When you have an accurate sample, press ENTER. MicroFID will record the maximum sample concentration when you press the ENTER key.
since there are no other locations stored in the datalogger the same event will be displayed.
current event and repeat the readings.
EVENT key. Select “Insert” or “Scan ” to enter an event name.
used for this site.
76
15. When you return to the default display, the instrument status will be “ Locate” . Record a background and sample reading for this site. You will now have two events recorded in the datalogger.
16. Repeat this procedure for all sites that must be monitored. You can record data for up to 250 sites.
17. Once all event entries have been recorded in the datalogger, you may want to download them to a computer to perform your own calculations.
Page 79
Detailed Operation
Programming the Monitoring Schedule before Field Operation
Use this method of programming the schedule if you are unable to go into the field. You will create a monitoring schedule with place­holders for the data.
To preprogram your monitoring schedule:
1. Press the SETUP key. Press ENTER until “ Interval” is displayed. Use the ARROW keys to select Method 21 as the Interval. Press ENTER.
2. You will be prompted to clear the datalogger. If you want to save the contents of the datalogger, press EXIT. You must print or save the data to disk. See PRINT Key on page 53, GRAPH Key Mode on page 58 or Computer on page 88 for details of printing or saving your data. Press ENTER to clear the datalogger.
3. You can enter the sites in any order; however, using the schedule after it has been stored is easier if the sites are stored in the order in which they will be monitored.
4. Press the EVENT key. Use the ARROW keys to select “In sert”. Press ENTER. Use the keypad to enter an event name up to 16 characters long. To obtain alphabetic characters you must use the STAR 21 software. (Part No. MX396027)
If your site is bar coded and you have a list of the bar codes, use the ARROW keys to select “Scan ”. Press ENTER to activate the bar code reader. The bar code reader will remain activated until you scan a code. If no bar code is scanned within 15 seconds the bar code reader will be deactivated. See Bar Code Reader on page 103 for details of connecting the bar code reader to MicroFID.
It is possible to enter an event name that already exists in the datalogger. When you select “In sert” or “Scan” you can enter the name of an event that already exists in the datalogger. MicroFID will not overwrite the existing event. It will insert an event with the same name. In this way you can create multiple copies of the same schedule in the
77
Page 80
MicroFID Software Manual
5. Press the CAL key and select the calibration memory to be
6. Press SETUP and enter a 2 digit repair code for this
7. When you return to the default display, the instrument
8. Press ENTER. The instrument status will change to
9. The instrument status will now be “Sample ”. Press ENTER
datalogger. You can see the duplicate events when you print the contents of the datalogger.
used for this site. Press ENTER and then press EXIT.
location. See SETUP Key on page 49 for a description of Repair code. This step is optional.
status will be “ Locate”. If the event name is greater than 3 characters the bottom line of the display will scroll from right to left.
“B kGnd”. Press ENTER to create a placeholder and to move to the next step.
to create a placeholder and to move to the next step. The instrument status will again be Locate. At this time, since there are no other locations stored in the datalogger the same event will be displayed.
78
10. Press EXIT at any time to return to the beginning of the current event.
11. To store the next location, press the EVENT key. Select “In sert” or “Scan” to enter an event name.
12. Press the CAL key and select the Cal Memory to be used for this site.
13. Press SETUP and enter a repair code for this location.
14. Press ENTER. The instrument status will change to “B kGnd”. Press ENTER to create a placeholder and to move to the next step.
Page 81
Detailed Operation
15. The instrument status will now be “Sample”. Press ENTER to create a placeholder and to move to the next step. The instrument status will again be “Locate”.
16. Repeat this procedure for all sites that must be monitored. You can record data for up to 250 sites.
Once all event entries have been recorded in the datalogger, you are ready to use the schedule for monitoring the sites. You may want to print the schedule to ensure all your sites have been entered correctly.
Using the Schedule
Once you have programmed the schedule you can use it at any time to perform Method 21 monitoring. Frequency of monitoring is determined from percentage leaker calculations. Refer to the Method 21 documentation for details.
To use the schedule:
1. Calibrate MicroFID as outlined in Calibration on page 65.
2. If the first event of the schedule is not displayed, you can find a specific event in the datalogger using the “F ind” option. Press SETUP and use the ARROW keys to scroll through the list of events. Select the desired event. When you press ENTER, the selected event will be displayed.
3. Locate the specified site and press ENTER.
4. The instrument status will change to “BkGnd ”. A background measurement must be made as specified in the Method 21 protocol. When you have an accurate background, press ENTER. MicroFID will record the displayed concentration when you press the ENTER key.
5. The instrument status will now be “Sample ”. Take a sample measurement, as specified in the Method 21 protocol. When you have an accurate sample, press ENTER. MicroFID will record the displayed concentration when you press the ENTER key.
79
Page 82
MicroFID Software Manual
6. The instrument status will again be “Locate”. If you
7. Press EXIT at any time to return to the beginning of the
8. To repeat a measurement recorded previously in the
If you change from Method 21 operation to a set averaging interval, you will lose your Method 21 schedule and associated programming. Print or save the data to disk before changing the interval. See PRINT Key on page 53, GRAPH Key Mode on page 58 or Computer on page 88 for details of printing or saving your data.
NOTE: The datalogger will be cleared when you change from Method 21 to
Interval mode.
recorded the sites in order, the next site will be displayed. You can move through the stored list. If the sites were not entered in order or the order of sites has changed, use the “F ind” option to locate the next site.
current event and repeat the readings.
datalogger, press the EVENT key and use the ARROW keys to select “ Find”. Select the event you need to repeat and press ENTER.
80
Response Factors for Gases and Vapors
In situations where only a single pure compound is present in air, MicroFID should be calibrated with a standard of that specific compound as span gas. MicroFID’s 10 Cal Memories can be used to store calibration information for 10 different span gases.
MicroFID’s reading will always be influenced by any other ionizable compounds present in the air sample. Even if MicroFID has been calibrated with a specific compound, its response is not specific and the presence of another ionizable impurity may render the numerical result invalid.
It is often impractical to carry a range of different standards into the field. Approximate results can be obtained by calibrating MicroFID
Page 83
Detailed Operation
with the recommended span gas and entering the appropriate response factor. The response factor is based on the ratio of the response of the specific compound to the response of the span gas. The response factor multiplies MicroFID’s reading then displays and records it (if the datalogger is on).
Response Factors on page 149 provides a list of response factors from which approximations can be made for guidance purposes. Data extrapolated from the use of response factors must be regarded as interim and approximate only. The response factors in the list should be used only for concentrations up to 500 ppm of the specific compound, as response factors change with concentration.
To use the response factors:
1. Press the CAL key and enter the response factor for the specific compound.
2. Calibrate MicroFID with zero air and 500 ppm methane as described in Calibration on page 65.
3. Expose MicroFID to the sample. The displayed reading is the approximate concentration of the specific compound.
The response factors on page 149 serve as a guide to concentrations measured by MicroFID.
Results are expected to be accurate to within +/-10 ppm or +/-25% of result, whichever is greater. Accuracy of response factors to other gases and vapors may differ from that stated.
NOTE: It does not matter which Cal Memory is selected or which response
factor is entered, MicroFID’ s response is not specific to any one compound. The displayed reading represents the total concentration of all ionizable compounds in the sample.
81
Page 84
MicroFID Software Manual
Preparing for Field Operation
Field Check List
The following items should be carried into the field to reduce or eliminate instrument down time. If you will be in the field for a single 8-10 hour day, you should include the following accessories:
Spare battery pack (Part No. MX396005 or MX396006)
Long sample probe (Part No. MX396018)
Calibration kit(s) (Part No. MX396011)
Tank(s) of calibration gas (Part No. MX396028)
Spare gas bag for zero air (Part No. MX396017)
Gas bag adapter for zero air (Part No. MX396010)
Supply of commercial zero air
Charcoal filters (Part No. MX396021, MX396022)
82
Shoulder strap (Part No. MX395002)
Spare inlet filters (Part No. MX396020 or MX396015)
Bar code reader and interface module (Part No. MX396054)
Spare 9 volt battery for bar code reader
Carrying case (Part No. MX396007)
Headset (Part No. MX396053)
MicroFID Instrument Manual (Part No. MX396003)
Table 3 Check List for Field Operation
Page 85
Detailed Operation
If you will be in the field for more than one day you should include the following additional items:
Battery chargers (Part No. MX396013 or MX396014). Take one charger for each battery pack.
Hydrogen and refill adapter (Part No. MX396004)
Printer cable (Part No. MX396051)
Computer and associated cables
Table 4 Additional Field Items
Operational Check List
Before beginning field work, set up and calibrate MicroFID for your particular application. Ensure the instrument is in working order before heading into the field.
To prepare MicroFID for field work:
1. Ideally, you should not transport MicroFID with the hydrogen cylinder filled. If it is possible to fill the internal hydrogen cylinder when you arrive at your destination then do so. If not, then you must fill the cylinder before you leave and observe all precautions for handling compressed, flammable gases. See Warnings and Safety Practices on page 18.
If you are travelling by passenger aircraft, you must make arrangements to fill the hydrogen cylinder when you arrive at your destination. You cannot transport MicroFID by passenger aircraft unless the internal hydrogen cylinder is empty. See Emptying the Hydrogen Cylinder on page 110 more information.
2. Press the BATT key and ensure the battery pack is fully charged. The voltage should be 9 volts.
3. Press the SETUP key and ensure the correct date and time are entered.
83
Page 86
MicroFID Software Manual
4. Program and calibrate all the Cal Memories you will be
5. Press the MAX key and clear the Max register if necessary.
6. If you are using an averaging interval, you may also want to
7. If you are performing Method 21 monitoring ensure you
using. See Calibration on page 65. After calibration is complete, sample the bag of calibration gas and the bag of zero air to ensure MicroFID has been calibrated correctly.
delete all events from the datalogger to avoid confusion between different days’ data and to avoid running out of space in the datalogger. See EVENT Key on page 50.
have programmed and calibrated all the Cal Memories. You must also program your monitoring schedule. See Method 21 Operation on page 72 for more information.
84
Page 87
Accessories
4
Page 88
Page 89
Accessories
Printer
MicroFID has been designed for use with a serial printer. The printer must have a print width of at least 65 characters and a proportionally spaced font. The printer cable (Part No. MX396051) has been configured for this application only. If the printer does not have an RS-232 serial interface you will need to use the serial to parallel converter (Part No. MX380145). See Serial to Parallel Converter on page 98 for details.
If you are unsure of the compatibility of the printer, refer to the printer user’ s manual or contact the printer’s manufacturer.
Ensure the printer is set to 8 data bits and 1 stop bit, parity has been set to none and that the appropriate baud rate has been selected.
NOTE: MicroFID is not classified for use in hazardous locations with a
printer.
To connect a printer:
1. Turn MicroFID off.
2. Connect the RS-232 printer cable (Part No. MX396051) to the I/O port on the back of the MicroFID and then to the serial port on the printer. See Figure 12.
3. Turn MicroFID on.
4. Determine the baud rate for your printer. Refer to the printer user’ s manual for details.
87
Page 90
MicroFID Software Manual
MicroFID
Printer Cable
MicroFID
Printer with 25 Pin,
Male,Serial, Port
Figure 12 Connecting MicroFID to a Serial Printer
5. Press the SETUP key and select the Baud Rate option. Use the ARROW keys to select the baud rate corresponding to the printer. The baud rate in the SETUP option must match the baud rate of the printer.
6. Use the PRINT and GRAPH keys as detailed on pages 53 and
58.
7. Turn MicroFID off before disconnecting the printer cable from the I/O port.
If this arrangement does not produce the desired results, see Printer Troubleshooting on page 124.
Computer
MicroFID will send information stored in its datalogger to a computer. This option may be used if you need to prepare reports based on MicroFID’s recorded data. This feature may also be used if you need the recorded data in a format that can be imported into a spreadsheet or database for further calculations.
NOTE: MicroFID is not classified for use in hazardous locations with
computers.
You may need to keep more data than can be logged by MicroFID. If this is the case you have two options: you may print the data before they are deleted from the datalogger and keep the printed reports, or you may store the data on disk for later use. You may
88
Page 91
Accessories
not, however, load data that has been stored on disk back into MicroFID.
In order to establish communication with MicroFID and send data stored in the datalogger to a computer, the computer must be set up to emulate a terminal. In order to do this you will need a communications software package that will allow you to transfer data from MicroFID to a computer. If you are using Microsoft Windows, you do not need to purchase any separate software. Instructions for downloading information to Windows’ HyperTerminal are provided below.
If you are already using another type of communication or terminal emulation software package it is not necessary to use HyperTerminal. Use your communications package to set up the computer to receive data with 8 data bits, 1 stop bit and no parity at the desired baud rate. Refer to the manual, provided with your software, for specific details.
Once the data have been transferred to HyperTerminal, it must be saved to a floppy disk or a hard disk. In order to manipulate the data you must import the data either as a text file into a spreadsheet, where it may be parsed and converted to numeric data, or into a text editor or word processor where it may be viewed and edited.
Refer to the computer setup options to determine the baud rate, as this value will depend on the computer. MicroFID’ s baud rate can be set from 1200 to 19200 to match the computer’s setting.
The instructions below will provide you with the most basic information for using MicroFID with communications software. Please refer to the software user’s manual for specific details of operation.
If you are using Microsoft Windows you do not need to purchase any separate software. You can use the HyperTerminal program that is shipped with Windows.
89
Page 92
MicroFID Software Manual
In order to use these instructions, you must be familiar with Microsoft Windows and it must be installed and running on your computer. These instructions are based on HyperTerminal Version
1.1.
NOTE: As newer versions of HyperTerminal and Windows are released,
the exact names and contents of dialog boxes may change. Use these instructions as a general guideline.
To initiate communications between MicroFID and Windows:
1. Turn MicroFID off.
NOTE: You must turn the instrument off before connecting or
disconnecting the printer cable.
2. MicroFID must be connected to a serial port. Use the printer cable (Part No. MX396051) to connect MicroFID to one of the computer’s serial ports. Remember which serial port you are using as you will need this information when you set up HyperTerminal. Normally you will use Com1 or Com2.
90
3. If the MicroFID printer cable plugs directly into the port on the computer without the gender changer or the 9 to 25 pin adapter, you are most probably connected to a parallel port. You will need at least one of the adapter cables to connect MicroFID to a serial port. To communicate with a parallel port, you will need to use the serial to parallel converter (Part No. MX380145).
4. Once all connections have been made, turn on the computer and the instrument.
5. Start Windows and then start the HyperTerminal. HyperTerminal will most likely be in the may be in the Accessories Program Group. Its location will depend on how you have set up Windows. Refer to your Windows documentation for details.
Page 93
Accessories
6. HyperTerminal may prompt you to connect a modem. You do not need a modem for this application, so press No, to close the pop up window.
7. The Connection Description dialog box will open. Enter the instrument name and select an icon. If you choose to save the session after you have completed downloading the data, this icon will then appear in the HyperTerminal group.
Figure 13 Connection Description Dialog Box
8. Press OK to accept the information.
9. The Connect To dialog box opens next. Select the serial port to which you connected MicroFID Press OK to move to the next dialog box.
91
Page 94
MicroFID Software Manual
10. The current properties of the selected serial port will be displayed. Enter 9600 as the baud rate (bits per second). This sets the baud rate at which the computer will receive data from MicroFID.
Figure 14 Connect To Dialog Box
NOTE: The baud rate setting of 9600 is being used here as an example. In
order for this setting to work, your computer serial port must be set to send/receive at 9600 baud. If your computer serial port is set to send/receive at a different baud rate, set the Speed command and MicroFID accordingly.
11. On MicroFID the number of data bits has been fixed at 8, stop bits has been fixed at 1. Parity has been set at None and the Flow control is Xon/Xoff. Enter these values and then press OK.
92
Page 95
Accessories
Figure 15 Properties Dialog Box
12. Select the correct Baud Rate, Data Bits, Stop Bits, Parity. Select Xon/Xoff for the Flow Control option. Leave the Parity Check and Carrier Detect options off. Select OK to close the dialog box and accept the settings.
13. Open the Transfer menu and select Capture Text. A “ Capture Text” dialog box will appear. Type in the desired path and filename for the data that are to be downloaded. Ensure the path is correct.
93
Page 96
MicroFID Software Manual
14. Ensure the file name has a .txt extension. The .txt extension will make it easier to use the downloaded data later.
15. Press Start. Select the Table Format option and then select OK. At the bottom of the screen the message, “Recei ving Data” is displayed.
16. On MicroFID, ensure the same baud rate is entered under the PRINT key. The number of Data Bits has been fixed at 8, the number of Stop Bits is fixed at one. Parity has been set to none. The settings in the Properties dialog box must match these settings.
Figure 16 Properties Dialog Box
94
17. Press the MicroFID PRINT or GRAPH key. Press SETUP (*) and set the start and stop run numbers. Press ENTER (on MicroFID). The logged data will begin to appear on the computer screen. It will be saved to the specified file as it appears on the screen.
18. To terminate communications select Stop from the bottom of the screen. This will close the text file.
The logged data can now be copied or cut to the Windows Clipboard and pasted into a text editor, such as Windows’ Notep ad, or into a spreadsheet for analysis.
Refer to the Microsoft Windows User’s Guide for detailed instruction on operation of HyperTerminal.
Page 97
Accessories
Chart Recorder
MicroFID’s output can be displayed as a 0 to 1 volt analog voltage on a chart recorder in real time.
NOTE: MicroFID is not classified for use in hazardous locations with a
chart recorder.
To connect MircoFID to a chart recorder:
1. Set the chart recorder to 1 volt full scale.
2. Turn MicroFID off and connect the analog output cable (Part No. MX396052) to the I/O port.
3. Connect the analog output cable to the chart recorder.
4. Turn MicroFID on.
The analog output has been fixed at 50000 ppm as full scale.
Gas Sampling Bag
A gas sampling bag is required for calibration of MicroFID. If you are unsure of the quality of ambient air you should use an additional gas bag filled with zero air. Connect the sample bag to the inlet fitting with the gas bag adapter. See Calibration on page 65 for details.
One sampling bag is included in the calibration kit. Additional gas bags are also available (Part No. MX396017).
Charcoal Filters
A charcoal filter (Part No. MX396021 or MX396022) may be connected to MicroFID during calibration to provide clean air to
95
Page 98
MicroFID Software Manual
1/4" Nut
Connect
to MicroFID
Charcoal Filter
Front Ferrule
Back Ferrule
MicroFID. The filter will remove hydrocarbon contamination from room air to provide zero grade air to MicroFID.
NOTE: The charcoal filter does not filter methane or ethane. If these
compounds are present, you should use a gas bag with a supply of commercial zero air.
To connect a charcoal filter:
Over-tightening the Teflon ferrules will result in damage to the ferrules.
CAUTION
1. Load the Teflon ferrules into the nut. The nut and ferrules are supplied with the filter. See Figure 17.
2. Connect the nut to MicroFID’ s inlet. Do not tighten the nut.
96
3. Remove the charcoal filter from its plastic bag and insert it into the nut. Finger-tighten the nut onto the inlet. If the filter is not secure, ensure you have inserted the tube stub far enough into nut. Do not over-tighten the fitting.
Figure 17 Connecting Charcoal Filter
Page 99
Accessories
2 C o n n e c t n u t t o M i c r o F I D i n l e t
3 I n s e r t f i l t e r i n t o n u t a n d t i g h t e n n u t
T e f l o n
F e r r u l e s
1 L o a d F e r r u l e s
F r o n t
B a c k
1 / 4 "
S s t l N u t
The charcoal filter will remove hydrocarbon contaminants for up to 4000 ppm hours. This means that the filter will be good for 1 hour removing 4000 ppm of hydrocarbon contaminants or will last for 4 hours removing 1000 ppm. The exact time will be determined by the operating environment. You will notice an increased hydrocarbon background when the filter requires replacement.
To replace the charcoal filter:
1. Loosen the nut on MicroFID’ s inlet.
2. Remove the used filter.
3. Insert the new charcoal filter into the nut and finger­tighten the nut onto the inlet. Do not over-tighten the fitting.
When the charcoal filter is not in use, place it in its plastic bag and store it in a clean, dry place.
Installing the Coalescing Filter
To install the Coalescing filter:
1. Load the Teflon® ferrules into the nut. The nut and ferrules are supplied with the filter.
2. Connect the nut to MicroFID’ s inlet. Do not tighten the nut.
Figure 18 Connecting the Coalescing Filter
97
Page 100
MicroFID Software Manual
Over-tightening the Teflon ferrules will result in damage to the ferrules.
CAUTION
The coalescing filter will remove particulates for up to 5 days of operation. The exact duration will be determined by the operating environment. MicroFID’ s inlet flow rate and sensitivity decrease when the filer is full.
3. Remove the filter from its plastic bag and insert it into the nut.
4. Finger-tighten the nut onto the inlet. If the filter is not secure, ensure you have inserted the tube stub far enough into nut. Do not over-tighten the fitting.
5. Calibrate all Cal Memories, that you are using, before continuing operation.
98
Serial to Parallel Converter
This device allows MicroFID, which is only capable of serial communication, to communicate with a parallel device. This will be most useful for printing, as most printers utilize parallel communication. Your printer must have a print width of at least 65 characters and a proportionally spaced font.
DIP Switch Settings
The DIP switch settings are very important and must be set correctly in order to correctly communicate with MicroFID. MicroFID will always be the serial device so the data flow direction will be serial to parallel.