The information contained in this document is subject to change without notice.
Except as specifically set forth in the terms and conditions of sale, PerkinElmer makes no warranty of any kind with regard to
this document, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein or for incidental consequential damages in connection with the furnishing,
performance, or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright. All rights are reserved. No parts of this document may
be reproduced in any form whatsoever or translated into any language without the prior written permission of
Perkin Elmer, Inc.
Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are protected by
law
PerkinElmer is a trademark of Perkin Elmer, Inc.
FIAS is a trademark of Perkin Elmer, Inc.
GemCone is a trademark of Perkin Elmer, Inc.
GemTip is a trademark of Perkin Elmer, Inc.
Hewlett-Packard and HP LaserJet are trademarks of Hewlett Packard Corporation.
Meinhard is a registered trademark of J.E. Meinhard Associates, Inc.
MS-DOS, Windows, and Microsoft are registered trademarks of Microsoft Corporation.
Ryton is a registered trademark of Phillips Petroleum Company.
Teflon is a registered trademark of E.I. duPont deNemours & Co.
Tygon is a registered trademark of Norton Company.
Viton is a registered trademark of E.I. duPont deNemours & Co.
WinLab is a trademark of Perkin Elmer, Inc.
PerkinElmer
710 Bridgeport Avenue
Shelton, Connecticut 06484-4794 USA
Tel: (800) 762-4000 or (203) 762-4000
Supplies, accessories, and replacement parts can be ordered directly from
PerkinElmer using the eight-digit part numbers provided in this manual. To place an
order for supplies and many replacement parts, request a free catalog, or ask for
information visit our website.
www.perkinelmer.com/supplies
The most up-to-date information on part numbers, product brochures, spare parts and
application notes are located in the PerkinElmer website.
•If you are located within the U.S., call toll-free: (800) 762-4000, Monday Friday, 8:30 a.m. to 7 p.m. EST. Your order will be shipped promptly, usually
within 24 hours.
•If you are located outside the U.S., call your PerkinElmer sales office.
11
Safety and Regulatory Information
!
Warning
Safety Information in the Manual
Safety information is contained in Chapter 1 of this manual. Before setting up and
operating this instrument, carefully read the safety precautions described in
this chapter and observe them at all times.
The protection provided by this equipment may be impaired if
Caution
the equipment is used in a manner not specified by PerkinElmer.
This manual contains important information regarding potential hazards that may
arise during the operation of the instrument. It is essential that this information is
read and thoroughly understood by all potential users of the instrument.
The instrument should be used according to the instructions provided in this
manual. If used otherwise, the protection provided by the instrument may be
impaired.
In this manual, the following graphic symbols and special text formats are used to
set apart important safety information.
12
Caution
A warning indicates an operation that could cause personal injury if
precautions are not followed..
A caution indicates an operation that could cause instrument damage
if precautions are not followed.
FCC Compliance (United States)
This product is classified as Industrial, Scientific and Medical (ISM) equipment and
has been tested and found to comply with the limits specified for non-consumer
equipment, pursuant to Part 18 of the FCC Rules.
These limits are designed to provide reasonable protection against harmful
interference in a non-residential installation.This equipment generates, uses, and can
radiate radio frequency energy and, if not installed and used in accordance with the
instructions, may cause harmful interference to radio communications. However,
there is no guarantee that interference will not occur in a particular installation.
If this equipment does cause harmful interference to radio or television reception,
which, can be determined by turning the equipment on and off, the user is
encouraged to try to correct the interference by one or more of the following
measures:
•reorient or relocate the receiving antenna
•increase the separation between the equipment and the receiver
•connect the equipment into an outlet on a circuit different from that to which
the receiver is connected
•consult the dealer or an experienced RFI technician for assistance.
Europe
All information concerning EMC standards is in the Declaration of Conformity, and
these standards may change as the European Union adds new requirements.
This instrument has been designed and manufactured, having regard to the state of
the art, to ensure that
a) the electromagnetic disturbance generated does not exceed the level
above which radio and telecommunications equipment or other
equipment cannot operate as intended;
b) it has a level of immunity to the electromagnetic disturbance to be
expected in its intended use which allows it to operate without
unacceptable degradation of its intended use.
13
Environment
!
Warning
Operating Conditions
Explosive Atmosphere
This instrument is not designed for operation in an explosiv e atmosphere.
Pollution Degree
This equipment will operate safely in environments that contain non-conductive
foreign matter up to Pollution Degree 2 in EN/IEC 61010–1.
Recommended operating conditions:
• Indoors.
• The location must be free of smoke, dust, corrosive fumes, direct sunlight and
• Temperature +15 °C to +35 °C (+59 °F to +95 °F) with a maximum rate of
• Relative humidity 20% to 80%, without condensation.
• Altitude in the range 0 m to 2,000 m (sea level to 6,562 feet) with a maximum
excessive vibration.
change of 2.8 °C (5 °F) per hour. For optimum instrument performance, the
room temperature should be controlled at 20° ± 2 °C. For optimum instrument
performance, the relative humidity should be between 35% and 50%.
rate of change of 2.8 °C (5 °F) per hour.
Safe operating conditions:
• Indoors
• Temperature +5 °C to +40 °C (+41 °F to +104 °F).
• Relative humidity 20% to 80%, without condensation.
• Altitude in the range 0 m to 2,000 m (sea level to 6,562 feet).
Storage conditions:
• Ambient temperature –20 °C to +60 °C (–4 °F to +140 °F).
• Relative humidity 20% to 80%, without condensation.
• Altitude in the range 0 m to 12,000 m (sea level to 39,370 feet).
14
When you remove the instrument from storage, before unpacking or putting it into
!
Warning
operation, allow it to stand for at least a day under the approved operating conditions.
Symbols Used on the Instrument
The three different types of warning symbols that appear on the instrument are
shown below:
This symbol indicates Caution, risk of electric shock.
AVERTISSEMENT: Risque de choc électrique.
This symbol indicates Caution, risk of danger.
Documentation must be consulted to determine the nature of the
potential hazard and any actions which have to be taken.
AVERTISSEMENT: Risque de Danger: Réfère à la documentation
de déterminer la nature de l’hasard potentiel et aucunes actions qui
sont nécessaire de le réparer.
This symbol indicates Caution, hot surface.
AVERTISSEMENT: La surface est chaude.
The following graphic symbols are also found on the instrument:
Indicates the OFF position of the main power switch.
15
Indicates the ON position of the main power switch.
Indicates alternating current.
16
Warning Labels on the Instrument
1
2
Labels on the front of the instrument
LabelLocationInstructions for Safety
Label #1 in
Figure A.
Label #2 in
Figure A.
UV Radiation. Risk of eye damage.
Rayonnement UV. Risque de blessure des yeux.
The plasma may emit UV radiation which can damage your eyes. Do not gaze
into the emitted light. Always wear UV-absorbing safety glasses when looking at
the light from the plasma.
Rayonnement UV. Risque de blessure des yeux. La plasma peut émettre des UV
qui peut blesser les yeux. Ne regardez pas fixement à la lumière emis. Tourjours
porter des lunettes de protection absorbant les UV pour regarder le plasma.
Hot gases are vented through the chimney into the venting system.
Les gaz chauds passent par la cheminée dans l’installation de ventilation.
Do not touch this area of the instrument when the plasma is on.
The plasma generates high intensity ultraviolet radiation.
Le plasma crée des rayons ultraviolets a haute intensité.
Do not view the plasma through the chimney.
Figure A.Location of warning label on the front of the instrument.
17
Warning labels on the back and side of the instrument
.
Air
GermanLuft
FrenchAir
DanishLuft
FinnishIlma
Greek
Vmax = 30 V AC, Imax = 0.5 A AC
Vmax = 30 V DC, Imax= 2 A DC
Do not exceed stated voltage and current.
Label #2 in
Figure B.
Label #3 in
Figure B.
Label #4 in
Figure B.
Label #5 in
Figure B.
Air / N
2
For the shear gas, use air or nitrogen.
Warning - Grounding circu it continuity is v it al for the safe ope ration
of this instrument. Grounding is accomplished by the use of an
IEC309 instrument power plug furnished by the factory and it is not
to be removed during or after installation.
Attention: La continuité de circuit par mettre a la terre est nécessaire pour
l’opération de cet instrument sans risque. L'usine fournit la fiche
d’instrument IEC309 qui mits à la terre l’instrument. N’enlève pas la fiche
pendant ou après l’installation.
Warning: All field wiring must have insulation suitable for at least 300 V.
Avertissement: Le cablage a pied d’ouvre doit etre certifiés pour un
minimum de 300 V.
Warning: Disconnect supply before servicing.
Avertissement: Couper l’alimentation avant l’entretien.
18
Label #6 in
Figure B.
CAUTION: This unit contains protective circuitry. Contact PerkinElmer
qualified personnel before performing any AC line tests.
Attention: Cet appareil continent un circuit protecteur. Contact le personnel
qualifié de PerkinElmer d’effectuer un quelconque controle de circuit en
courant alternatif.
The instrument has AC line surge suppressing components which require
disconnection before testing. For safety reasons, testing and servicing
should be carried out only by a PerkinElmer service engineer or similarly
authorized and trained person.
19
Figure B.Location of warning labels on the back and side of the instrument.
1
2
3
45
6
Caution: Do not restrict air intake or exhaust.
Attention: N’obstruez pas l’arrivée ou l’évacuation d’air.
T o provide adequate space for ventilation, allow at least 30 cm (12 inches)
of free space around the instrument. Do not obstruct the two air filters.
20
Warning labels in the torch compartment
.
MarkingLocationInstructions for Safety
Label #1 in
Figure C
Label #2 in
Figure C
WARNING – Do not defeat interlocks.
Avertissement: Ne pas desengager les enclenchements de securite.
The instrument has safety interlocks to protect the operator from exposure
to radio frequency and ultraviolet radiation. Do not attempt to defeat the
safety interlocks.
Risk of hot surfaces.
Risque de surfaces chaudes.
Wait until the torch and surrounding surfaces have cooled to room
temperature before you touch them.
21
22
1
2
Figure C.Location of warning labels in the sample compartment.
Using this Guide
This manual is your hardware guide to the Optima 8000 instrument. Main topics
include:
•safety practices
•preparing your laboratory
•a description of the system
•installation
•maintenance
•troubleshooting
•error messages
Documents to Help You
This manual contains hardware information for Optima 8000 instrument. The
following documentation is also provided for Optima 8000 instrument.
•Winlab32 Software Guide on the WinLab CD. This guide contains basic
information on the ICP WinLab software, step-by-step procedures, and
comprehensive chapters that cover each of the software windows and parameters
for reference.
•Winlab32 Software Installation and Administration Guide on the WinLab CD.
This guide contains step-by-step procedures for software installation and
administration information on the ICP WinLab software.
23
How this Guide is Organized
Thoroughly read the Safety chapter before using the instrument. Consult the
appropriate reference materials listed at the end of the Safety chapter on page 46.
Read Chapter...To find out about...
1 Safety PracticesImportant safety information.
2 Preparing Your LaboratoryThe preparation of your laboratory
3 System DescriptionThe components of the instrument, how
4 InstallationThis chapter includes information for
needed for the Optima 8000 inst rument.
Environmental, electrical, space,
exhaust, gases, and cooling water
requirements are reviewed.
the echelle spectrometer works and
instrument specifications.
installing or re-installing the system
components (autosampler, computer,
and printer) and is provided for your
reference should you ever need to move
the system.
24
5 MaintenanceMaintenance and cleaning procedures
for the various components of your
system, particularly for the sample
introduction area.
6 TroubleshootingPerformance checks and
troubleshooting information for both
hardware and software problems.
7 Error MessagesExplanations of software and system
error messages.
Using the Index
Use the index to look up information on the different hardware components. If you
need to find information on cleaning the nebulizer, for example, look up “nebulizer”
in the index. Also, use the index to look up information on the following topics:
•installation
•maintenance
•troubleshooting
A Word on Troubleshooting
We recommend that you consult the troubleshooting information included in this
manual before you call a PerkinElmer service engineer. The Troubleshooting
chapter contains suggestions to help you determine systematically whether
instrument problems are due to improper analytical techniques, improper selection
of instrument parameters, or malfunction of the instrument.
25
26
Safety Practices
1
28
Safety Practices
Introduction
This chapter describes general practices designed to aid you in safely operating the
spectrometer and accessories.
This advice is intended to supplement, not supersede, the normal safety codes in
the user's country. The information provided does not cover every safety procedure
that should be practiced. Ultimately, maintenance of a safe laboratory environment
is the responsibility of the analyst and the analyst's organization.
Please consult all manuals supplied with the spectrometer and accessories before
you start working with the instrument. Carefully read the safety information in this
chapter and in the other manuals supplied. When setting up the instrument or
performing analyses or maintenance procedures, strictly follow the instructions
provided.
Laboratory Guidelines
This section describes some general laboratory safety guidelines. For additional
information, we recommend The CRC Handbook of Laboratory Safety (1) and
Prudent Practices for Handling Hazardous Chemicals in Laboratories (2).
General
Caution
Note
The protection provided by this equipment may be impaired if the
equipment is used in a manner not specified by PerkinElmer.
This equipment requires no specified inspection or preventive
maintenance to ensure the continuous functioning of its safety features
Always wear safety glasses when using the instrument. Safety glasses with side
shields will provide an extra margin of safety as well as mechanical protection for
your eyes.
For safety reasons and to avoid contaminating samples, be sure that the instrument
and work area are kept scrupulously clean. This is especially important when
working with toxic elements or when measuring trace amounts of any element.
Clean up spilled chemicals immediately and dispose of them properly.
Smoking is a source of significant contamination as well as a potential route for
!
Warning
ingesting harmful chemicals.
Food should not be stored, handled, or consumed in the work area.
Environmental Conditions
The instrument should be used indoors in a laboratory having the environmental
conditions as recommended in Environmental Conditions on page 49.
This instrument is not designed for operation in an explosive
environment.
Pollution Degree
This product will operate safely in environments that contain nonconductive
foreign matter up to Pollution Degree 2 in EN/IEC 61010-1
Laboratory Guidelines
29
Handling of the Instrument
Contact a PerkinElmer service engineer for assistance in installing or
evaluating the system after moving. The Optima 8000 weighs 146 kg (321
pounds), or 201 kg (442 pounds) with the shipping container. If the instrument
needs to be moved please contact PerkinElmer Service.
Laboratory Ventilation
Toxic combustion products, metal vapor, and ozone can be generated by the
system, depending upon the type of analyses. Combustion products vary with the
type of sample analyzed. If the sample compartment door is opened, a maximum
of 0.9 liters of Argon may escape during the first minute. After that, 0.1 liters/
minute of Argon will continue to escape while the door is open.
An exhaust venting system is always required to remove those gases which are
generated during the normal operation of the instrument.
30
!
Warning
Safety Practices
An efficient ventilation system must be provided for the instrument. Detailed
specifications for a recommended venting system are described in the section titled
Exhaust Vent Requirements on page 50.
Instrument Safety Practices
This section describes potential hazards and recommended safety practices. You
should thoroughly review this information.
This equipment requires no specified inspection or preventive maintenance
to ensure the continuous functioning of its safety features.
Do not position the instrument so that it is difficult to operate the main power
switch which is the disconnecting device.
The protection provided by this equipment may be impaired if
Caution
Safety Interlocks
the equipment is used in a manner not specified by PerkinElmer.
The instrument has safety interlocks to protect the operator from radio frequency
(RF) radiation and ultraviolet radiation, and to prevent access to high voltage areas.
Do not attempt to defeat the safety interlocks. This would place the
operator's safety at risk.
The following interlocks must be satisfied in order to ignite the plasma. If any of
these interlocks is interrupted while the plasma is on, the plasma will automatically
be shut down.
•The front door on the torch compartment must be closed before the plasma can
ignite.
•Argon pressures for the torch must be correct.
•Emergency Off Switch must be released.
•Cooling water must be flowing to the plasma induction plates and oscillator.
•Quick change torch must be installed.
•The shear gas must be flowing.
Radio Frequency Radiation
!
Warning
!
Warning
The instrument generates high levels of Radio Frequency (RF) energy, which is
potentially hazardous if allowed to escape. The instrument is designed to contain
the RF energy within the shielded enclosures of the sample compartment and the
RF power supply. Safety interlocks prevent you from operating the system without
all covers, doors, and shields in place.
Ultraviolet Radiation
The plasma generates high intensity ultraviolet radiation. A safety interlock is used
to automatically shut off the plasma if the sample compartment door is opened or
the torch is removed. The sample compartment has a viewing window for safely
viewing the plasma.
Directly viewing the plasma (without protection) may cause
permanent impairment of eyesight.
Instrument Safety Practices
31
Electrical Safety
The instrument has been designed to protect the operator from potential electrical
hazards. This section describes some recommended electrical safety practices.
Lethal voltages are present at certain areas within the instrument.
Installation and internal maintenance of the instrument should only
be performed by a PerkinElmer service engineer or similarly
authorized and trained person.
When the instrument is connected to line power, opening the instrument
covers is likely to expose live parts. Even when the power switch is off,
high voltages can still be present. Power supplies inside the instrument
may still be charged even if the instrument has been disconnected from
all voltage sources.
32
!
Warning
!
Warning
!
Warning
!
Warning
!
Warning
Safety Practices
Grounding circuit continuity is vital for safe operation of the
equipment. Never operate equipment with grounding conductor
disconnected.
Connect the instrument to an AC line power outlet that has a
protective ground connection. To ensure satisfactory and safe operation
of the instrument, it is essential that the protective ground conductor (the
green/yellow lead) of the line power cord is connected to true electrical
ground. Any interruption of the protective ground conductor, inside or
outside the instrument, or disconnection of the protective ground terminal
may impair the protection provided by the instrument.
Grounding circuit continuity is vital for safe operation of the
equipment. Grounding is accomplished by use of an IEC60309
instrument power plug furnished by the factory and not removed during
or after installation.
The instrument must be correctly connected to a suitable electrical supply. The
supply must have a correctly installed protective conductor (earth ground) and
must be installed or checked by a qualified electrician before connecting the
instrument.
Any interruption of the protective conductor (earth ground) inside or
outside the instrument or disconnection of the protective conductor
terminal is likely to make the instrument dangerous. Intentional
interruption is prohibited.
Do not operate the instrument with any covers or parts removed.
Disconnect the power cord from the AC line power wherever the cover is
removed.
Instrument Safety Practices
When working with the instrument:
•Connect the instrument to a correctly installed line power outlet that has a
protective conductor connection (earth ground).
•Do not operate the instrument with any covers or internal parts removed unless
required to do so.
•Never leave the instrument unattended with the cover off and connected to AC
power!
•Do not attempt to make internal adjustments or replacements except as directed
in the manuals.
•Disconnect the instrument from all voltage sources before opening it for any
adjustment, replacement, maintenance, or repair. If afterwards, the opened
instrument must be operated for further adjustment, maintenance, or repair,
this must only be done by a qualified person who is aware of the hazards
involved.
•Whenever it is possible that the instrument is no longer electrically safe for use,
make the instrument inoperative and secure it against any unauthorized or
unintentional operation. The electrical safety of the instrument is likely to be
impaired if, for example, the instrument shows visible damage; has been
subjected to prolonged storage under unfavorable conditions; or has been
subjected to severe stress during transportation.
33
High Temperatures
The torch components remain hot for some time after the plasma has been shut off.
Allow sufficient time for these items to cool to room temperature before you handle
them.
Hot Exhaust Gases
When the plasma is on, hot gases are vented through the chimney above the sample
compartment. An efficient ventilation system must be provided, as referenced in
the Exhaust Vent Requirements on page 50. In addition, avoid this area of the
instrument to prevent burns.
34
!
Warning
Safety Practices
Safe Use of Drain Systems
For safe operation of the system, the pumped drain system should be properly
installed.
Never place the vessel in an enclosed cabinet. Doing so could result in
a build-up of hazardous gases, which may result in a serious explosion or
fire.
Never use a glass drain vessel. A glass drain vessel may break and spill
flammable, toxic, or corrosive liquids.
Recommended safety practices for drain systems are given below.
•Place the drain vessel in an area that is visible to the operators, who can observe
the level of collected effluent and empty the vessel when necessary.
•Check the condition of the drain tubing regularly to monitor deterioration.
Organic solvents deteriorate the tubing more quickly than aqueous solutions.
When the tubing becomes brittle or cracked, replace it.
•Empty the drain bottle regularly when using organic solvents. Before
switching from organic to aqueous solutions, always empty the drain bottle.
Waste Disposal Procedures
•Carefully monitor the collection of effluent in the drain vessel and empty the
drain vessel frequently. When switching between organic and aqueous
solutions, flush the drain tube thoroughly and empty and flush out the drain
vessel.
•Drain vessels may contain flammable, acidic, caustic, or organic solutions, and
small amounts of the elements analyzed . The collected effluent may have to be
disposed of as hazardous waste.
•The responsible body is responsible for the correct collection and disposal of
waste materials. This includes the necessity for a suitably sized waste container
of appropriately resistant material for the collection of organic solvent waste
and provision for the removal into an appropriate exhaust system of any gases
or vapors which may be produced in hazardous concentrations.
WEEE Instructions for PerkinElmer Products
!
Warning
Dispose of waste in accordance with the regulations applicable to
your locality, state, and/or country.
WEEE Instructions for PerkinElmer Products
A label with a crossed-out wheeled bin symbol and a rectangular bar indicates that
the product is covered by the Waste Electrical and Electronic Equipment (WEEE)
Directive and is not to be disposed of as unsorted municipal waste. Any products
marked with this symbol must be collected separately, according to the regulatory
guidelines in your area.
35
The objectives of this program are to preserve, protect and improve the quality of
the environment, protect human health, and utilize natural resources prudently and
rationally. Specific treatment of WEEE is indispensable in order to avoid the
dispersion of pollutants into the recycled material or waste stream. Such treatment
is the most effective means of protecting the customer's environment.
Requirements for waste collection reuse, recycling, and recovery programs vary by
regulatory authority at your location. Contact your local responsible body (e.g.,
your laboratory manager) or authorized representative for information regarding
applicable disposal regulations. Contact PerkinElmer at the web site listed below
for information specific to PerkinElmer products.
Web address:
http://las.perkinelmer.com/OneSource/Environmental-directives.htm
36
!
Warning
Safety Practices
For Customer Care telephone numbers select "Contact us" on the web page.
Products from other manufacturers may also form a part of your PerkinElmer
system. These other producers are directly responsible for the collection and
processing of their own waste products under the terms of the WEEE Directive.
Please contact these producers directly before discarding any of their products.
Consult the PerkinElmer web site (above) for producer names and web addresses.
Safe Handling of Gas Cylinders
The permanent installation of gas supplies is the responsibility of the user
Note
and should conform to local safety and building codes.
Gases commonly used with ICP instruments include argon and nitrogen. The major
hazard associated with these gases is suffocation. This can occur if the gas is
allowed to escape in an enclosed area and displaces the oxygen in the air. These
gases are neither explosive nor combustible.
Contact the gas supplier for a material safety data sheet (MSDS)
containing detailed information on the potential hazards associated
with the gas.
Carefully use, store, and handle compressed gases in cylinders. Gas
cylinders can be hazardous if they are mishandled.
If liquid argon or nitrogen is used, the gas cylinder must be fitted with an overpressure regulator, which will vent the cylinder as necessary to prevent it from
becoming a safety hazard.
Listed below are some general safety practices for the proper identification,
storage, and handling of gas cylinders. Consult Referen ces 3 and 4 on page 46 for
more detailed information and additional guidelines.
Identification of Gas Cylinders
•Legibly mark cylinders to identify their contents. Use the chemical name or
commercially accepted name for the gas.
Safe Handling of Gas Cylinders
37
Storing Cylinders
•Cylinders should be stored in accordance with the regulations and standards
applicable to the customer’s locality, state, and country.
•When cylinders are stored indoors in storage rooms, the storage room should
be well ventilated and dry. Ensure that the ventilation is adequate to prevent
the formation of dangerous accumulations of gas. This is particularly important
in small or confined areas.
•Do not store cylinders near elevators, gangways, or in locations where heavy
moving objects may strike or fall against them.
•Use and store cylinders away from exits and exit routes.
•Locate cylinders away from heat sources, including heat lamps. Compressed
gas cylinders should not be subjected to temperatures above 52 °C (126 °F).
•Do not allow ignition sources in the storage area and keep cylinders away from
readily ignitable substances such as gasoline or waste, or combustibles in bulk,
including oil.
•Store cylinders standing upright, fastened securely to an immovable bulkhead
or permanent wall.
•When storing cylinders outdoors, they should be stored above ground on a
suitable floor and protected against temperature extremes (including the direct
rays of the sun).
Handling Cylinders
•If it becomes necessary to move cylinders, do so with a suitable hand truck
after ensuring that the container cap is secured and the cylinder properly
fastened to the hand truck.
•Use only gas-pressure regulators, tubing, and hose connectors approved by an
appropriate regulatory agency.
•Arrange gas hoses where they will not be damaged or stepped on, and where
objects will not be dropped on them.
•Do not refill gas cylinders.
38
!
Warning
Safety Practices
•Check the condition of pipes, hoses, and connectors regularly. Perform gas
leak tests at all joints and seals of the gas system regularly, using an approved
gas leak detection solution.
•When the equipment is turned off, close all gas cylinder valves tightly at the
cylinder. Bleed the remainder of the line before turning the exhaust vent off.
Hazardous Chemicals Warnings
Before using any chemicals or solvents with the instrument, the customer should
be thoroughly familiar with all hazards and safety handling practices. Observe the
manufacturer’s recommendations for use, storage and disposal. These
recommendations are normally supplied in the material safety data sheets (MSDS)
supplied with the solvents.
Some chemicals used with this instrument may be hazardous or may
become hazardous after completing an analysis. The responsible body
(e.g. Lab Manager) must take the necessary precautions to ensure that the
surrounding workplace and the instrument operators are not exposed to
hazardous levels of toxic substances (chemical or biological) as defined
in the applicable Material Safety Data Sheets (MSDS) or OSHA, ACGIH,
or COSHH documents. Venting for fumes and disposal of waste must be
in accordance with all national, state, and local health and safety
regulations and laws.
Some definitions of terms used in "Hazardous Chemicals Warnings" above are
given next.
OSHA: Occupational Safety and Health Administration (United States)
ACGIH: American Conference of Governmental Industrial Hygienists
COSHH: Control of Substances Hazardous to Health (United Kingdom)
Responsible body: "Individual or group responsible for the use and maintenance
of equipment, and for ensuring that operators are adequately trained." [per EN/IEC
61010-1].
Operator: "Person operating equipment for its intended purpose." [per EN/IEC
!
Warning
61010-1].
Safe Use of Chemical Reagents
This section provides some general safety practices that should be observed when
working with any chemicals.
Give careful attention to the hazards associated with the chemical
reagents being used. Refer to the safety data sheets provided by the
manufacturer, for example, Material Safety Data Sheets (MSDS) in
the USA (for other names, see Reference 5).
Protective Equipment
•Wear appropriate eye protection at all times while handling chemicals. Use
safety glasses with side shields, goggles, or full-face shields, according to the
types of chemicals being handled.
•Wear suitable protective clothing, including gloves specifically designed to
resist the chemicals being handled.
Safe Handling of Gas Cylinders
39
Use, Storage, and Disposal
Review the following information to ensure the safe use, storage, and disposal of
chemicals.
•Use, store, and dispose of chemicals in accordance with the manufacturer's
recommendations and regulations applicable to the locality, state, and/or
country.
•When preparing chemical solutions, always work in a fume hood that is
suitable for the chemicals you are using.
•Conduct sample preparation away from the instrument to minimize corrosion
and contamination.
•Clean up spills immediately using the appropriate equipment and supplies such
as spill cleanup kits.
•Do not put open containers of solvent near the instrument.
40
!
Warning
Safety Practices
•Store solvents in an approved cabinet (with the appropriate ventilation, as
required) away from the instrument.
•The responsible body has the responsibility for carrying out appropriate
decontamination if hazardous material is spilt on or inside the instrument.
Sample Preparation
Sample preparation for ICP spectroscopy may require the handling of organic or
corrosive solutions. Dilution of samples and adding dilute acid to a liquid sample
is generally less hazardous than putting a solid sample into solution. This section
will deal specifically with the potential hazards associated with dissolving solid
samples, although the information can be applied to any sample handling situation.
Also, refer to the general precautions for handling chemicals described at the
beginning of this chapter.
Reference 6 contains general information on sample preparation procedures as well
as an appendix on laboratory safety. Reference 7 contains information on
laboratory safety and sample preparation methods prescribed by the United States
Environmental Protection Agency (EPA). Reference 8 contains information on
microwave sample preparation, which is not covered in this section.
Solid samples may be put into solution by dissolving the sample in an organic
solvent or acidic solution; digesting the sample; or reacting the sample with a
fusion mixture. Each of these techniques has certain hazards associated with it.
Always wear appropriate eye protection while preparing samples.
Use safety glasses with side shields, goggles, or full-face shields,
depending on the chemicals you are handling.
Dissolving Solid Samples in Organics
•Always work in a fume hood so that flammable and/or toxic solvents do not
concentrate in the work area.
•Wear protective clothing and gloves. Some solvents are readily absorbed
through the skin.
When selecting a solvent, consider the following:
•Is the solvent compatible with the equipment?
Aliphatic hydrocarbons, ketones and esters, alcohols, and xylene, are the most
frequently used solvents. Make sure that all equipment that will come into
contact with the solvent is compatible with the solvent, e.g., tubing, sample
cups, storage and waste containers, etc.
•Is the solvent toxic?
Avoid solvents known to be health hazards, such as benzene or methyl isobu tyl
ketone (MIBK).
Some solvents, such as the halogenated hydrocarbons, generate toxic gases
when heated.
Consult the data sheets provided by the manufacturer for information on health
hazards (see Reference 5 on page 46).
Dissolving Solid Samples Using Acids
Sample Preparation
41
Dissolving solid samples in aqueous solutions often requires the use of
concentrated acids.
Using Acids
When using acids, use the following precautions:
•Always work in an acid-resistant fume hood.
•Wear protective clothing, including gloves specifically designed to resist the
acid being used.
•Attempt to dissolve the sample in a dilute acid solution before using
concentrated acid.
•Add concentrated acid to a sample cautiously. Dissolving a sample in
concentrated acid may provoke a vigorous reaction.
42
!
Warning
!
Warning
Safety Practices
Acid Digestions
Acid digestions, either at atmospheric pressure or at increased pressure, require
special care. Spattering and foaming of the sample/acid mixture may expose the
user to a hazard, as well as compromise the sample integrity. A digested sample
containing concentrated acid will react violently with water.
Perchloric acid and hydrofluoric acid are particularly hazardous to work with.
Perchloric Acid
Perchloric acid (HClO
presents severe fire and explosion hazards.
Before using perchloric acid, you should be thoroughly familiar with
its hazards and safe handling practices. Observe the manufacturer's
recommendations for use, storage, and disposal.
•Use hoods, ducts, and other devices for removing vapors specifically designed
to accommodate this kind of fume. There is a severe explosion hazard if a
normal hood is used, or if the hood is not properly used and maintained.
•Use goggles and face shields. Wear protective clothing and pol yvinyl chloride
gloves. Do not use rubber gloves.
•Additional hazards and precautions are given in References 1, 2, 9, 10 and 11.
Hydrofluoric Acid
Hydrofluoric acid (HF) is also used for digestions. It is toxic and extremely
corrosive. Hydrofluoric acid will readily burn skin, and if the fumes are inhaled,
lung tissue. Burns may not be immediately painful or visible. Contact with eyes
could result in blindness.
) is extremely corrosive and a powerful oxidizing agent. It
4
Before using hydrofluoric acid, you should be thoroughly familiar
with its hazards and safe handling practices. Observe the
manufacturer's recommendations for use, storage, and disposal.
Sample Preparation
!
Warning
•Always wear suitable protective equipment, including goggles, a face shield,
acid-resistant gloves, and protective clothing when using hydrofluoric acid.
•Do not breathe HF vapors. Always work in a fume hood when using
hydrofluoric acid.
•Do not use a glass beaker. Hydrofluoric acid attacks glass.
•Observe the additional hazards and precautions outlined in References 1, 2, 9,
10 and 11 on page 46.
Performing Digestions
When carrying out sample digestions:
•Always work in a hood suitable for the type(s) of chemicals you are using.
•Add very small quantities of the acid dropwise, while observing the magnitude
of the reaction.
•Apply heat to a digestion solution slowly in case a further vigorous reaction is
initiated.
43
•Cool the completed digest before transferring it and diluting it. Add water
cautiously.
Digestions at Elevated Pressure
Acid digestions in a pressure digestion apparatus require special care at several
points, in addition to the precautions described above.
When using high pressure digestion vessels, consult the
manufacturer's instructions and recommendations, particularly
regarding organic material and possible explosive reactions.
Never use perchloric acid in a pressure digestion.
•Use only the manufacturer's recommended amounts of sample and acid for the
type of sample you are preparing.
•Be sure the apparatus is in good condition and provides safe release of pressure
in the case of excess pressure buildup.
44
Safety Practices
•Cool the apparatus to room temperature before attempting to open it.
•Open the apparatus in a fume hood to vent the entrapped fumes safely.
Fusion Mixtures
Dissolving a sample by reacting with a fusion-flux mixture is generally chosen
when other techniques for sample dissolution fail. The fusion-flux mixture should
be carefully selected. Knowledge of fusion-flux mixture/crucible compatibility is
essential. Heating the sample-flux mixture may initiate a vigorous or explosive
reaction.
Caution
The fusion matrix will deposit on and dissolve the quartz torch over time.
It is recommended to use the unslotted (0 slots) torch because the
slotted torches will dissolve much faster.
•Heat the mixture slowly and intermittently until the system is characterized.
•If you use a muffle furnace rather than a flame for heating, make a trial using
the standard amount of flux but with a smaller amount of sample.
•Most reacted fusion mixtures are dissolved in acidic solutions. Observe the
precautions described previously for handling acids.
Decontamination and Cleaning
Before using any cleaning or decontamination methods except those specified by
PerkinElmer, users should check with PerkinElmer that the proposed method will
not damage the equipment.
Decontamination
Customers wishing to return instrumentation and/or associated materials to
PerkinElmer for repair, maintenance, warranty or trade-in purposes are advised
that all returned goods must be certified as clean and free from contamination.
The customer's responsible body is required to follow the "Equipment Decontamination Procedure" and complete the "Certificate of Decontamination". These
documents are available on the PerkinElmer public website.
If you do not have access to the internet and are located in the U.S., call toll free at
1-800-762-4000 or (+1) 203-925-4602, 8:30 a.m. - 7 p.m. EST and speak to
Customer Support.
In Canada, call toll free 800-561-4646 and speak to Customer Support.
If you are located outside of the United States or Canada, please call your local
PerkinElmer sales office for more information.
Cleaning the Instrument
Exterior surfaces may be cleaned with a soft cloth, dampened with a mild detergent
and water solution. Do not use abrasive cleaners or solvents.
Decontamination and Cleaning
45
46
Safety Practices
References
1. Furr, K., ed., CRC Handbook of Laboratory Safety, 5th ed., The Chemical
Rubber Co. Press, 2000.
2. National Research Council, Prudent Practices for Handling Hazardous Chemicals in Laboratories, National Academy Press, Washington, D.C., USA,
1981.
3. Compressed Gas Association (USA), "Safe Handling of Compressed Gases in
Containers," pamphlet no. P-1, 2008.
4. Compressed Gas Association (USA), "The Inert Gases – Argon, Nitrogen and
Helium," pamphlet no. P-9, 2008.
5. Data sheets provided by chemical manufacturers, for example:
The items listed below need to be considered when preparing the laboratory for the
Optima 8000:
•Environmental Conditions
•Exhaust Vent Requirements
•Laboratory Space Requirements
•Cooling Water Requirements
•Electrical Requirements
•Pneumatic Requirements
Environmental Conditions
The laboratory in which the Optima 8000 Spectrometer system is located must meet
the following conditions:
•corrosive-free environment.
•The instrument will operate with a laboratory temperature between 15 and 35 °C
(59 - 95 °F). For optimum instrument performance, the room temperature should
be controlled at 20
•The heat dissipated directly into the laboratory when the Optima 8000 is
properly vented is approximately 6600 BTU/ho ur (2200 W).
•Relative humidity between 20% and 80%, non-condensing. For optimum
instrument performance, the relative humidity should be between 35% and 50%.
•Dust levels not above 36 000 000 particles, 0.5 micron or larger, per cubic meter
of air. The environment should be relatively dust-free to avoid sample and
instrument contamination problems.
•Free of excessive vibration.
•Altitude: in the range -400 to 2,000 m (sea level to 6,562 feet)
± 2 °C.
Preparing Your Laboratory
The Optima 8000 has been designed for indoor use. Do not use the instrument in an
area where explosion hazards may exist.
Pollution Degree:
This product will operate safely in environments that contain nonconductive foreign
matter up to Pollution Degree 2 in EN/IEC 61010-1.
49
Preparing Your Laboratory
!
Warning
Exhaust Vent Requirements
The Optima 8000 requires an exhaust vent to remove combustion fumes and vapors
from the torch compartment. Exhaust venting is important for the following reasons:
•It protects laboratory personnel from toxic vapors that may be produced by some
samples.
•It improves the stability of the ICP torch by removing the effects of room drafts
and laboratory atmosphere.
•It helps to protect the instrument from corrosive vapors that may originate from
the sample(s).
•It removes dissipated heat produced from the torch and power supply.
Warning: Toxic fumes
The use of ICP-OES instruments without adequate ventilation to outside air may
constitute a health hazard. For exam p l e, the co mb u stio n of halo ge n at ed
hydrocarbons produces toxic vapors. Extreme care should be taken that exhaust
gases are vented properly.
50
Note
The maximum temperature for the ICP torch vent system is 200 °C (392
withstand these temperatures, stainless-steel must be used.
ICP torch vent required minimum flow rate is 5600 liters/min (200 cubic feet/min)
at the end of the venting hood.
Local electrical codes do not allow PerkinElmer Service Engineers to install the
blower and vent assembly.
The blower capacity depends on the duct length and number of elbows or bends used
to install the system. If an excessively long duct system or a system with many bends
is used, a stronger blower may be necessary to provide sufficient exhaust volume.
Alternatively, smooth stainless-steel may be used instead of flexible stainless-steel
where flexibility is not required to reduce system friction loss or “drag.” If smooth
stainless steel is used, there must be a way to move the vent hood out of the way for
servicing. A length of smooth stainless-steel ducting has 20-30% less friction loss
F). To
Preparing Your Laboratory
than a comparable length of flexible ducting. When smooth stainless-steel is used,
elbows must be used to turn corners. These elbows should turn at a center line radius
of 150 mm with a maximum bend angle of 45 degrees to reduce friction losses, and
the number of elbows should be minimized.
Additional recommendations on the venting system include:
•Make sure duct casing is installed using fireproof construction. Route ducts
away from sprinkler heads.
•Locate the blower as close to the discharge outlet as possible. All joints on the
discharge side should be airtight, especially if toxic vapors are being carried.
•Equip the outlet end of the system with a back draft damp er an d tak e the
necessary precautions to keep the exhaust outlet away from open windows or
inlet vents and to extend it above the roof of the building for proper dispersal of
the exhaust.
•Equip the exhaust end of the system with an exhaust stack to improve the overall
efficiency of the system.
•Make sure the length of the duct that enters int o the blower is a straight length at
least ten times the duct diameter. An elbow entrance into the blower inlet causes
a loss in efficiency.
•Provide make-up air in the same quantity as is exhausted by the system. An
"airtight" lab will cause an efficiency loss in the exhaust system.
•Ensure that the system is drawing properly by using an air flow meter.
•Equip the blower with a pilot light located near the instrument to indicate to the
operator when the blower is on.
The venting system for the ICP torch should be positioned over the Torch
Compartment Chimney located on top of the sample compartment. For proper
instrument venting, order PerkinElmer Venting Kit (Part No. 03030447 Kit for 115
V or Part No. 03030448 Kit for 230 V) by contacting your PerkinElmer sales
representative. The minimum distance from the top of the instrument’s Torch
Compartment Chimney to the bottom of the extraction vent should be 7.5-12.5 cm.
(3-5 in.). Gas, water, and air lines of 6 m (20 ft.) are also provided. Figure 2-1 shows
the location of the ICP torch exhaust vent.
51
Preparing Your Laboratory
1
2
3
4
Figure 2-1Location of the Torch Compartment Chimney and Vent.
ItemDescription
1Recommended Vent
Width: 30.6 cm (12 in.)
Depth: 15.3 cm (6 in.)
Height: 30.6 cm (12 in.)
2Torch Compartment Chimney
328 cm (11 in.) from the instrument edge to the chimney
center
47.5 to 12.5 cm (3 to 5 in.)
52
Laboratory Space Requirements
The Optima 8000 system includes the Optima 8000 instrument, a PolyScience
Recirculating Chiller (or equivalent), a computer, and a printer. The Optima 8000
instrument itself includes optics, electronics, a plasma torch, and an RF power
supply, all housed in a self-contained unit. The minimum door width must be 81 cm
(32 in.).
Optima 8000 Instrument
The Optima 8000 is 132-cm long (52-in.), 76-cm high (3 0-in.), and 81-cm deep (32in.). The bottom of the vent must be a minimum of 75-cm (3-in.) from the top of the
torch compartment chimney. Figure 2-2 illustrates the dimensions of the instrument.
The Optima 8000 weighs 146 kg (321 pounds), or 201 kg (442 pounds) with the
shipping container.
The Optima 8000 may be placed on a bench. The main power cable length is 2.4 m
(8.0 ft.).
To provide access for servicing and space for ventilation, allow at least 61 cm (24
in.) of space between the instrument and the laboratory walls.
Preparing Your Laboratory
Computer and Printer
Refer to the manuals supplied with your computer and printer for dimensions.
53
Preparing Your Laboratory
2
3
1
Figure 2-2Outside Dimensions of the Optima 8000 Instrument.
ItemDescription
176 cm (30 in.)
281 cm (32 in.)
3132 cm (52 in.)
PolyScience Chiller
®
The PolyScience
WhisperCoolTM (or equivalent) is usually located on the floor, to
the right side of the instrument. Its dimensions are 36.5-cm (14.4 in.) wide, 67.3-cm
(26.5 in.) deep and 61.0-cm (24 in.) high. It weighs 80 kg (178 lbs.). The Chiller has
air intakes and exhausts that need to be unobstructed. The power is 2850 watts. It
requires a minimum of 45.7 cm (18 in.) on all sides for adequate ventilation. The
coolant hoses restrict the distance that the Chiller can be placed away from the
instrument (hose length = 3.6 meters or approximately 12 feet).
54
If a longer hose distance is required for the chiller location, it may be
Caution
necessary to increase the hose inner diameter to accommodate the
specification of no greater than 15 psig at the instrument water
outlet. The hose inner diameter should also be sized large en ough to
accommodate the specification of 45 to 80 psig at the water inlet.
The hose also needs to be insulated sufficiently to maintain a
temperature at the instrument water inlet between 15 °C and 25 °C.
Cooling Water Requirements
A water supply is required to dissipate heat from the plasma induction plates and the
oscillator.
Recirculating System
A recirculating system (Chiller) must be used. The requirements for the chiller are:
Preparing Your Laboratory
•Cooling Capacity at 20 °C is 2000 watts
o
Temperature Stability ±1
C
Pump Rate 1 US gal/min. at 55 psi max (45 psi min).
Coolant: distilled water
A PolyScience Recirculating Chiller meets these requirements and is recommended
for the instrument. The PolyScience
®
WhisperCoolTM chiller is available through
PerkinElmer in the following two configurations:
•230V, 60 Hz (Part No. N0772046)
•240V, 50 Hz (Part No. N0772045)
For detailed information on PolyScience
®
WhisperCoolTM chiller electrical
requirements see the following section, Electrical Requirements.
®
The PolyScience
WhisperCoolTM chiller comes with its own Instruction Manuals.
The headquarters for PolyScience Instruments is located at:
•PolyScience Instruments, Inc.
6600 West Touhy Ave
Niles, Il 60714
Grounding circuit continuity is vital for safe operation of this instrument.
Grounding is accomplished by use of an IEC60309 instrument power plug
furnished by the factory and not to be removed during or after installation.
•Optima 8000 requires an AC line voltage of 200 - 230 VAC +/-10% (180 - 253
VAC) 50/60 Hz. +/-1% under full instrument load, that has a correctly wired
protective earthing system (ground connection) and a separate circuit breaker.
Maximum power consumption is 2800 VA. The line power supply must
conform with local safety regulations and be checked by a qualified electrician
before you connect the instrument to line power.
•The line power supply should be free of line transients in excess of 50 V peak.
If the electrical supply voltage produces large AC line voltage fluctuations, a
qualified electrician should install a voltage regulator between the electrical
outlet and the instrument.
56
•The Optima 8000 instrument is supplied with a 2.5 m (98 in.) line power cord
that supplies both the spectrometer and the RF Generator. The line power supply
point must be within 2.5 meters of the rear of the spectrometer.
•Connect the spectrometer, computer, printer, and any accessories to the same
phase of the line power supply and the same protective earth.
Note
PerkinElmer instruments will normally operate wi th in ± 10% of the specified voltage
and within
unstable, fluctuates in frequency or is subject to surges, additional control of the
incoming power by the user may be required.
The Optima 8000 must not have a Ground Fault Circuit Interruptor (GFCI) protected
outlet. The instrument will trip the interruptor if this type of outlet protection is used.
1% of the specified frequency, unless otherwise noted. If th e power line is
The Optima 8000 is equipped with an IEC60309 250 V 16/20A 2 pole plus
protective earth plug (PerkinElmer Part No. 09997530) that inserts i nto an equivalent
IEC60309 series receptacle (Perkin Elmer part number 09290304 or 09290305
Preparing Your Laboratory
surface mount version) both of which are contained in the N0770425 conduit box kit
that ships with the instrument.
As an alternative you can use the following receptacle part numbers which can be
directly ordered from Hubble. For the US/Canada (20A service) use PerkinElmer
Part Number 09997529 Hubble Part No. C320R6SVL or C320R6W. For Europe
(16A service) use Hubble Part No. C316R6S. Also, Conduit Box-Cast Device Back
Box Hubble Part No. SP 20301 is available for the above Hubble versions.
Computer and Printer
Refer to the guides supplied with your computer and printer for electrical
requirements. The computer and printer must share a common earth ground with the
Optima 8000.
PolyScience Chiller
®
The specific electrical requirements for the PolyScience
WhisperCoolTM Chiller
are printed on a serial number label located on the back of the unit. The voltage of
the power source must meet the specified voltage
10%. In addition, an adequate
ground connection must be provided.
For 60 Hz installations, the PolyScience 6106PE V, 60 Hz, 15A unit is supplied
with a 15-ampere, 250-volt Hubbell #4570-C Twist-Lock power plug (NEMA L615P configuration). A Hubbell #4560 or equivalent receptacle (NEMA L6 -15R
configuration) is also required, and is supplied with the instrument.
(NEMA: The National Electrical Manufacturer’s Association)
For 50 Hz installations, to accommodate most countries, the PolyScience 220/240
V, 50 Hz, 15A unit is supplied with two detachable line cords.
57
Preparing Your Laboratory
Pneumatic Requirements
Argon
Liquid or gaseous argon can be used with the Optima 8000 system. The use of liquid
or gaseous argon tanks is determined primarily by the usage rate. Liquid argon is
usually less expensive per unit volume to purchase, but cannot be stored for extended
periods. If liquid argon is used, the tank should be fitted with an over-pressure
regulator for safety reasons. The over-pressure regulator vents the tank as necessary
to keep the argon cool enough to remain in its liquid state, thus preventing the
cylinder from exploding due to pressure build-up. A tank of liquid argon containing
160 liters will typically last for 80 hours of continuous running time.
Gaseous argon tanks do not require venting and consequently can be stored for
longer periods without loss. A tank of gaseous argon will last 5 to 6 hours of running
time. The normal argon usage is 9-20 liters/min. with a maximum of 25 liters/min
(0.04 - 1.0 cu. ft/min). The argon flow for the system (including the ICP) may vary
between 1 and 25 L/min.
CautionGas delivery lines from th e argon tank should be contaminant-free and
not made of plastic. Teflon delivery lines are acceptable.
Available argon pressure should be between 550 to 825 kPa (5.5 to 8.25 bar or 80 to
120 psig ).
Argon can be purchased from local suppliers. The argon for use with ICP systems
should be 99.996% pure.
Purge Gas
Nitrogen or argon can be used to purge the Optima 8000 spectrometer optics. The
purge gas high flow is 5 L/min. The low gas flow is 1.0 L/min.
CautionGas delivery lines from the purge gas tank should be contaminant-free and not
made of plastic. Teflon delivery lines are acceptable.
The available pressure should be between 275 and 825 kPa (2.75 to 8.25 bar or 40 to
120 psig).
The purge gas should be 99.999% pure and is available from local suppliers.
58
Shear Gas
Regulator
Preparing Your Laboratory
Air or nitrogen can be used to shear the plasma for the Optima 8000. The Optima
8000 RF generator typically consumes the shear gas at a rate of 25 L/min.
The flow rate should be approximately 25 L/min (1.0 cu. ft/min).
Available pressure should be between 550 and 825 (5.5 to 8.25 bar or 80 to 1 20 psig).
A pressure regulator for use with either argon or nitrogen is available from
PerkinElmer as Part No. 0303-0284. To connect the regulator to the instrument gas
controls, use the 1/4-in. Swagelok connector. The gas tube provided has 1/4-in.
Swagelock fittings.
59
Preparing Your Laboratory
60
System Description
3
Introduction
The Optima 8000 instrument consists of three major components: the spectrometer,
the ICP Source and the sample introduction system. Each component is further
divided into the different modules described below. This modular system design
facilitates system access, testing and servicing.
NOTE: All illustrations in this guide are based on the Optima 8000 instrument.
The spectrometer for the Optima 8000 consists of the following modules:
The ICP source for the Optima 8000 consists of three modules:
System Description
•Optics module
•Spectrometer electronics module
•Spectrometer pneumatics module
•RF generator module
•High voltage power supply module
•Plasma pneumatics module
The sample introduction system for the 8000 which comprises:
•A Quick-Change Adjustable Torch Module
•Spray chamber and nebulizer
•Peristaltic pump for the sample and drain.
The Optima 8000 has a number of software and firmware enhancements.
63
System Description
1
2
3
4
Figure 3-1.The Optima 8000.
64
ItemDescription
1Sample Introduction Compartment with the Quick Change
Adjustable Torch Module (inside)
2Peristaltic Pump
3Torch Compartment
4Spectrometer
Spectrometer
Introduction
The core of the optical system comprises a dual Echelle monochromator with a dual,
backside-illuminated, cooled, CCD detector. The system is specifically designed for
ICP-OES. Computer controlled transfer optics direct the radiation from the plasma
into the monochromator. The optics housing is sealed and continuously purged with
high purity nitrogen.
Signals at the required analytical wavelengths are measured using a scanning CCD
(charged coupled device) based technology, with simultaneous measurement of the
background emission and a neon spectrum for active wavelength correction.
Transfer Optics
The computer controlled transfer optics are used to select either radial or axial
viewing of the plasma and to direct the radiation from the plasma onto the entrance
slit of the monochromator. The exact viewing position, horizontal and vertical
position, can be selected in the software.
System Description
An automatic shutter closes between measurement cycles to reduce the exposure of
the optics to excess UV radiation.
65
System Description
88
1
10
5
3
6
7
8
4
2
9
12
17
19
14
15
16
13
11
18
Figure 3-2Schematic diagram of the optical system.
ItemDescription
1Radial View
2Transfer Optics Mirrors
3Echelle Grating
66
System Description
ItemDescription
4Detector
5Parabolic Collimators
6Prism (internally reflective face)
7Entrance Slit
8Intermediate Slit (This is where the neon reference beam enters)
9Torch
10Axial View
11Transfer Optics Mirrors
12Echelle Grating
13Detector
14Parabolic Collimators
15Prism (i internally reflective face)
16Entrance Slit
17Intermediate Slit (This is where the neon reference beam enters)
18This mirror moves out of the optical path for Axial viewing.
19Torch
Monochromator
In the dual echelle monochromator, the prism monochromator acts as a preselection
system to select the required wavelength range to pass on to the echelle
monochromator. Optimally positioned slits and baffles result in very low stray light
levels reaching the echelle monochromator. The prism and echelle dispersion
systems use Littrow configurations, designed to eliminate astigmatism, with
identical, 300 mm focal length, 10° off-axis, parabolic, collimating and focusing
mirrors.
The echelle grating has 79 lines/mm with a blaze angle of 63.4°. The grating is used
in the higher orders, where the high efficiency and high dispersion allow a relatively
short focal length resulting in a compact optical system.
67
System Description
Wavelength selection is achieved by simultaneous rotation of the prism and grating.
Since the maximum rotation required for either element is not more than ± 2°, the
average wavelength selection time is less than 2 seconds. To further optimize the
analysis time, the wavelengths required are sorted to minimize the change-over time
between any two consecutive wavele ngths.
The dual monochromator system enables relatively high slits to be used with no loss
of image quality, which contributes to the high optical throughput. In addition, part
of the slit height is used for the simultaneous measurement of a neon reference
spectrum for wavelength correction.
Two slit width settings are available, optimized for UV and visible wavelengths. The
system automatically selects the correct slit for each analytical measurement.
68
System Description
1
2
4
3
5
6
Figure 3-3Monochromator
ItemDescription
1Echelle Grating
2Detector
3Intermediate Slit
4Fiber optic for the neon reference beam.
5Entrance Slit
6Prism
Detector
The detector is a two-dimensional CCD device containing approximately 25,600
pixels. The photosensitive area is separated into two differently sized arrays that are
used for separate reference and analytical measurements. The analytical signal is
measured in the larger, lower array.
69
System Description
1
3
5
7
2
4
6
8
9
The rear of the actual detector area is thinned to a few micrometers to allow
illumination from the rear. This prevents the absorption of radiation by components
other than the detector pixels and maximizes the quantum efficiency without the use
of a fluorescent coating.
Figure 3-4Detector
ItemDescription
1Output
2Register for the reference measurement.
32 mm
4CCD Array
70
53.5 mm
System Description
ItemDescription
6CCD Array
7Output
8Register for the reference measurement.
93 mm
At 240 nm the array covers a wavelength range of approximately 0.52 nm, and at 850
nm, approximately 25 nm. Thus the emission line for the analyte of interest and
emission on each side of the analytical line fall simultaneously on the array. This
allows simultaneous measurement of the analyte and background signals.
To improve performance and reduce noise levels, the CCD detector is cooled
between –7 and –8 °C with an integrated Peltier cooler. The entire CCD is
hermetically sealed and the housing filled with dry nitrogen.
How the CCD works
Photons of radiation from the analyte emission strike the photosensitive area of the
detector where photoelectrons are produced in each pixel of the detector. The
electrons are moved into the register where they accumulate as electric charge. The
charge is allowed to accumulate for the period selected for the integration time. At
the end of the integration time, the charge is transferred out of the register to the
signal processing electronics. The pixels in each vertical row are binned into the
register for the respective array. This use of an effective pixel length of 3 mm results
in the low noise performance typical of CCD devices.
71
System Description
1
2
3
Figure 3-5Storage of photoelectrons during integration.
ItemDescription
1Photons
2Photosensitive Register
3Register
Wavelength Correction
The optical system uses a two-component, active wavelength stabilizing system. The
overall stability is controlled by temperature and pressure sensors in the optics
housing.
Any residual deviations are compensated for by measuring a neon reference
spectrum simultaneously with each measurement of an analytical emission line. The
output from a neon discharge lamp is collected by an optical fi ber and projected onto
the top half of the intermediate slit in the monochromator, which is the entrance to
the echelle monochromator. The neon spectrum passes through the echelle system
with the analytical radiation and illuminates the top array of the detector. The neon
spectrum acts as a wavelength scale to enable active wavelength correction.
72
ICP Source
RF Generator
The Optima 8000 uses a 40-MHz free-running solid state RF generator. The RF
power from the solid state oscillator is used to ionize the argon in the torch and excite
the atoms of the liquid sample so that they emit energy at their atomic wavelength in
the form of photons. The photons from the torch are detected optically and measured
electronically in the spectrometer section of the Optima 8000. It is displayed in terms
of wavelength and intensity, which are converted to sample concentration.
The RF generator provides a power output of 750 to 1500 watts. The power output
levels are computer-controlled and may be adjusted in one-watt steps for different
sample matrices.
The RF generator is designed with RF Power Control (RFPC), using a power control
loop which maintains the plasma setting regardless of line voltage fluctuations and
changes in the plasma.
System Description
73
System Description
1
2
3
4
7
8
56
Figure 3-6Block diagram illustrating the RF Power Control (RFPC)
ItemDescription
1AC Power Input
240 MHz RF Power Generator
3RF Power to Plasma Induction Plates
4Plasma
5Control Feedback Signal
74
ItemDescription
6Power Measurement Signal
7RF Power Control Loop
8Microprocessor
RF Control Electronics
The RF generator uses solid-state circuits. The solid state RF generator is designed
to significantly increase reliability and reduce the need for recalibration.
The RF generator also monitors plasma conditions. If the plasma is unstable, the
system automatically shuts it off.
To ensure operator safety and to protect the instrument from damage, the system
includes extensive use of RF shielding and safety interlocks. Proper RF shielding
and filtering are provided so that the system complies with regulations regarding
radio frequency radiation.
System Description
Sample Introduction System
Sample and Torch Compartments
The sample compartment has a side door to provide easy access to the sample
introduction module. The torch compartment door has a window so that the operator
can safely view the plasma, and has safety interlocks that shut down the plasma if the
door is opened during operation. The torch compartment has extensive shielding to
prevent exposure to radio frequency radiation.
Quick-Change Adjustable Torch Module
The Quick-Change Adjustable Torch Module is the quartz torch, injector, spray
chamber, and nebulizer/end cap – all in one assembly. This module can be quickly
removed from the sample compartment. This adjustable mount allows the torch an
adjustment of up to 8 mm.
In addition, adapters are available for special applications, such as the use of an
ultrasonic nebulizer or an external spray chamber.
75
System Description
Torch
The torch has a standard alumina injector with a 2.0 mm inner diameter. Other
injectors available include alumina injectors in different sizes for optimum sample
flow rate, and quartz injectors for different sample types.
Spray Chamber
Depending on the instrument configuration ordered the instrument may include a
Scott-type spray chamber or GemTip Cross-Flow pneumatic nebulizer and a
cyclonic spray chamber. Other combinations are also be available. A Ryton, doublepass Scott-type spray chamber and the GemTip Cross-Flow pneumatic nebulizer are
one of the standard configurations. This rugged combination provides the best results
for a variety of elements and sample types.
Nebulizers
PerkinElmer offers these nebulizers for a wide variety of applications:
Description/Part No.Uses/Advantages
GemTip Cross-Flow
Nebulizer end cap
N0770546
GemCone (Conespray)
Nebulizer
High Solids: N0690670
Low Flow: N0690671
Concentric Glass
Nebulizer
(MEINHARD)
MEINHARD Type A
00472020
76
Good general purpose nebulizer for the analysis of strong mineral acids
(including HF) and samples with less than 5% diss ol ved solids. Uses
GemTips made of sapphire and ruby in a Ryton end cap for maximum
chemical resistance.
High Solids GemCone is for samples with high dissolved solids (up to
20%). Low Flow GemCone permits lower nebulizer gas flow rates, useful
for spectral lines with high excitation energies and for providing a more
robust plasma. GemCone nebulizers require an end cap (Part No.
N0680343) for use with the Scott spray chambers.
Provides excellent sensitivity and precision for aqueous solutions and
samples with few dissolved solids (less than 1%). Self-aspirating. Not to be
used with solutions containing hydrofluoric acid. PerkinElmer offers three
types, Meinhard A, C, and K, which are described below. All require an end
cap (Part No. N0680343) for use with the Scott spray chambers.
The MEINHARD nebulizer will aspirate liquid sample naturally. The
nebulizer may also be operated with externally pumped sample, provided
that the sample flow is not significantly below the natural aspiration rate.
Used for samples containing high dissolved solids (up to 20%).
Optimized for samples that require reduced nebulizer gas flow rates, such as
organic-based samples.
For samples with low analyte and low matrix concentrations. Typically
improves detection limits by a factor of 10 over conventional pneumatic
nebulizers.
The Mira Mist Nebulizer is recommended for aqueous solutions. Organics
will wet the nebulizer tip and performance will deteriorate.
Figure 3-7GemTip Cross-Flow nebulizer and end cap N0680503 (also shown in
cross-sectional view).
77
System Description
78
Figure 3-8GemCone nebulizer and end cap N0680343 (also shown in cross-sec-
tional view).
System Description
Figure 3-9Concentric glass nebulizer and end cap N0680343 (also shown in
cross-sectional view).
Nebulizer Heater (optional)
The nebulizer heater allows you to set the temperature of the spray chamber
enclosure to provide a constant temperature for the spray chamber. The nebulizer
heater ensures maximum long-term stability even as room temperature varies.
The use of the nebulizer heater also ensures that the sample aerosol is uniformly
maintained with respect to temperature and aerosol droplet size. The heater can be
used with the cyclonic and the Scott-type spray chamber. The nebulizer heater
temperature can be controlled via the software and the spray chamber door must be
closed for the heater to work properly. The heater should always be on for inorganic
analyses.
For organic, high solids or salt analyses the spray chamber should be as cool as
possible, therefore we recommend that you do not use the heater to achieve the best
performance. We also recommend that you leave the sample compartment door open
when running organic analyses. Also do not use the nebulizer heater when using an
ultrasonic nebulizer.
79
System Description
Peristaltic Pump
The peristaltic pump is fully computer-controlled. As an added feature, the pump
speed can be programmed to run at a fast speed for the read or rinse cycle. The
TubingSaver mode is a feature for extending the life span of pump tubing.
Autosampler
PerkinElmer offers autosamplers for automated sample handling. These
autosamplers consist of a sample table, a sample tray, and a motorized sampling arm
with an attached probe. Different sample trays are available for each autosampler,
covering a variety of sample volume and total sample capacity requirements. All
sample trays have one location for a 150-mL wash vessel.
Switches and Controls
Main On/Off Switch
The Main Instrument switch is used to turn on the spectrometer (and is normally left
on). Once the plasma has been ignited, you should wait one half hour for the system
to stabilize before running samples.
Interlocks
80
Interlocks are designed to ensure operator safety and protect the instrument from
damage. The main system interlocks are described below.
The following interlocks must be satisfied in order to ignite the plasma. If any of
these interlocks is interrupted while the plasma is on, the plasma will automatically
be shut down. Before you can ignite the plasma:
•The torch compartment door must be closed;
•the Emergency Off (EMO) switch on front of instrument must be released;
•the argon pressure for the torch must be correct;
•the cooling water must be flowing to the plasma induction plates and oscillator;
•the torch must be installed;
•the shear gas pressure must be correct.
EMO Switch
1
System Description
The EMO switch, Emergency Off Switch, is the illuminated red switch on the front of
the instrument. If it is blinking slowly the instrument is in the middle of an ignition
cycle.It blinks rapidly after the Emergency Off Switch is depressed. This is an
indication that the switch has been depressed and to remind you to push the switch
again to release it. This switch shuts off the plasma in an emergency by
disconnecting the main voltage circuitry in the RF generator. To restart, you must
release the switch by pressing the switch again. Reset the RF generator using the
software Reset button. (The spectrometer stays on.)
ItemDescription
Software Controls
Figure 3-10Location of Red Emergency Plasma Off Switch.
1Emergency Off Switch
Many of the hardware settings are controlled by the software.
•RF Power: Power levels can be adjusted in 1-watt increments.
•Plasma and auxiliary argon flow rates: Flow rates can be automated during the
analysis with specific flow rates for each element if desired. Plasma argon is
adjustable in 1 L/min increments. Auxiliary argon is adjustable in 0.1 L/min
increments.
81
System Description
•Nebulizer argon flow rate: The flow rate is automatically controlled using a
•Pump rate: By specifying the desired flow rate (adjustable in 0.1 mL/min
•Nitrogen purge rate: A high or normal flow rate can be specified.
•Dark current measurement: A shutter under software-control can be closed to
•TubingSaver: A feature for extending the life span of pump tubing for the
•The area of the plasma viewed by the optical system can be adjusted horizontally
mass flow controller in 0.01 L/min increments.
increments) and the tubing diameter, the software calculates the pump speed.
block light from the plasma from reaching the detector in order to measure dark
current.
peristaltic pump.
and vertically using a computer-controlled moveable transfer optic. The
software also has a built-in optimization function for plasma viewing.
82
System Description
Connections to Electrical, Gas, and Cooling Water
Supplies
See the following figure.
MarkingFunction
O I
IEEE 488.2
N
2
Air / N
2
Shear Gas
Ar
H2O IN
(Red Tie)
H2O OUT
(White Tie)
Main power switch.
Connections for remote control of accessories. Do not exceed the stated
voltage and current:
Umax = 30 V AC, Imax = 0.5 A AC
Umax = 30 V DC, Imax= 2 A DC
IEEE-488 cable 0999-1355 to connect instrument to the computer.
Line power cord, permanently attached with an IEC 309 connector. Connect
to 200-240 VAC, 15 A, IEC-309 line power outlet.
Purge gas inlet for nitrogen (or argon). Use hose N0 69 27 5
Shear gas inlet for air or nitrogen. Use Air hose N0770348
included with Air Dryer Filter and regulator assembly.
Argon inlet for torch. Use Argon hose N0690274.
Cooling water inlet. Use hose N0770341 from the Chiller.
Cooling water outlet. Use hose N0770342 to the Chiller.
83
System Description
1
10
13
11
12
2
6
4
5
7
3
9
8
Figure 3-11Electrical, gas, and cooling water connections on the sp ec trome te r
side of the instrument.
ItemDescription
1Gas Connections
2Ar Supply
3Shear Gas Supply
84
ItemDescription
System Description
4N
5Coolant IN
6Coolant OUT
7IEEE Computer Connection
8Terminal strip (Part No. 09987900) for chiller remote operation
9Power Switch
10Instrument to 200-240 VAC, 16 AMP, IEC-309 Outlet
11White Tie Wrap (Part No. N0770342)
12Red Tie Wrap (Part No. N0770341)
13Chiller
Supply
2
System Initialization
When you switch on the spectrometer:
1. The Peltier cooling system for the detector starts to cool the detector.
2. The system sets up all the motors for the optics at the default positions for the
axial viewing mode.
3. The system has been continuously monitoring the detector temperature. When
the temperature is stable at about -8 °C, the system continues with the
initialization.
4. The system performs a dark current measurement; this takes about 30 seconds.
5. The system switches on the neon reference lamp and measures the intensity. If
the energy is too low, the system displays an error message. When the measured
intensity is above the lower limit, the system is ready for use.
The spectrometer initialization is completed when the spectrometer sends a "system
ready" message to the computer and can be viewed under the diagnostics spectrometer window. Once the plasma has been ignited, you should wait one half
hour for the system to stabilize before running samples..
85
System Description
Technical Data
General
Principle
Inductively coupled plasma optical emission spectrometer.
Computer controlled, using a special application program running
under a graphical user interface.
Power
requirements
Electrical
protection
Safety
standards
EMC
standards
Environmental
requirements
200 to 240 V AC ~, 50/60 Hz
Power consumption 2800 VA (maximum)
Insulation: Class I
Pollution degree: 2
See the Declaration of Conformity.
See the Declaration of Conformity.
Recommended temperature: +15 °C to +35 °C (59 °F to 95 °F)
20–80% relative humidity; non-condensing
Altitude in the range 0 m to 2,000 m.
Dimensions
Width: 1320 mm, (52 inches)
Height: 760 mm, (30 inches)
Depth: 810 mm, (32 inches)
Mass
(weight)
141 kg, (310 lb.) Spectrometer alone, not including the computer,
autosampler, and chiller.
ICP Plasma Generator System
86
Frequency: 40 MHz, free-running
Output Power Stability: <0.1%
Power Output: 750 to 1500 watts, computer-controllable in 1 watt increments. RF
generator is located in the left side of the instrument for efficient use of laboratory
space.
RF Shielding: Meets all FCC certification requirements for RF emissions.
Plasma Induction Plates
Cooling Water: System requires a flow of 1 gal/min at 310 to 550 kPa at a
temperature between 15 °C and 25 °C. A recirculating cooling system is required
(PolyScience
Automatic Ignition: Plasma ignition is computer controlled and totally automated.
The plasma can be turned on at a set time, warming up the system prior to an
analysis, and can be turned off automatically after an analysis.
Safety Interlocks: System checks water flow, shear gas flow, argon pressures,
emergency plasma off switch, torch compartment door interlocks, torch installed and
plasma stability. The status of these interlocks is constantly monitored and text
information is displayed on the computer screen. If any interlock is interrupted, the
plasma is shut down automatically.
Gas Flow Controls
Plasma Argon Flow: Closed-loop flow control using a proportional valve and a
measured pressure across a known resistor. Computer-controlled to regulate the flow
automatically within the range of 0 to 20.0 liter/minute in 1.0 liter/minute
increments.The flow system is interlocked to prevent ignition without plasma gas
flowing.
System Description
®
WhisperCoolTM chiller or equivalent is recommended).
Auxiliary Argon Flow: Closed-loop flow control using a proportional valve and a
measured pressure across a known resistor. Computer-controlled to regulate the flow
automatically within the range of 0 to 2.0 liter/minute in 0.1 liter/minute increments.
Nebulizer Argon Flow: Computer-controlled, using a mass flow controller, and is
variable between 0 and 2.00 liter/minute in 0.01 liter/minute increments.
Shear Gas: A compressed-air shear gas (18-20 liters/minute) is used to push the
plasma plume out of the optical path, minimizing the impact of self-absorption in the
cooler plasma plume.
Sample Introduction System
Torch: Demountable design using one-piece quartz tubing for plasma and auxiliary
gas flow. The standard torch is supplied with a 2.0-mm alumina in jector for full
corrosion resistance to all acids, including hydrofluoric and aqua regia. A 0.8-mm
injector is available as an option for analysis of very volatile organic solutions. As
options, 3.0-mm, 1.6-mm and 1.2-mm quartz injectors are available.
87
System Description
Spray Chamber: Scott-type designed to minimize pulsations from the peristaltic
pump and constructed of Ryton for complete corrosion resistance to most acids,
including HF, and all organic solvents normally used in ICP analyses.
Nebulizer: Cross-flow design with chemically resistant GemTips manufactured
from corrosion-resistant (sapphire tips in a PEEK body) material. The system can
routinely handle 50% (v/v) solutions of HCl, HNO
and 30% (v/v) NaOH. Up to 20% NaCl can be aspirated by the nebulizer for 1 hour
without clogging. The system is fully compatible for use with other nebulizers such
as ultrasonics, concentrics (MEINHARD) and GemCone types.
Peristaltic Pump: A three-channel, variable speed, computer-controlled pump.
Speed is variable from 0.2 to 5 mL/minute in 0.1 mL/minute increments using 0.76
mm (0.030 inch) tubing.
Torch Mount: In Quick-Change Torch Module both horizontal and vertical viewing
of the plasma are optimized by computer-controlled movement of the first transfer
mirror.
User Plasma Viewing: The full plasma is viewed through a UV-blocking, lowtransmittance window located in the sample compartment door.
Transfer Optics: Computer-controlled toroidal mirrors. Adjustable viewing
position: ± 15mm along the plasma; ± 10 mm across the plasma.
Detector: Dual, backside-illuminated, cooled, CCD detector. Cooling to between 7 and -8 °C with integrated Peltier cooler. Detector area: approximately 3 x 5 mm
split into a reference and an analyte array with separate read-out registers. Reference
array: 64 x 192 pixels of 18 x 32 m. Analyte array: 64 x 192 pixels of 18 x 56 m.
Read-out noise: 30 electrons. Dark current: 150 electrons/pixel/second. Read-out
time: 64 s. Full well capacity: 1.1 million electrons. Charge transfer efficiency:
0.99995.
Wavelength correction: Temperature and pressure sensors for the optics housing.
Spectrum from a neon discharge lamp is collected with each analytical
wavelength measurement and acts as a wavelength scale to enable active
wavelength correction.
88
Installation
4
Installation Summary
Contact a PerkinElmer service engineer for assistance in installing or
evaluating the system after moving. The Optima 8000 weighs 141 kg (310
pounds), or approx. 200 kg (440 pounds) with the shipping container. If the
instrument needs to be moved please contact PerkinElmer Service.
A PerkinElmer service engineer should install your system for the first time or help
you in moving the system. This chapter is provided for your reference should you
need information on moving the system, or reinstalling accessories. In conjunction
with this material, be sure to consult the Safety Practices chapter beginning on page
27and Preparing Your Laboratory chapter beginning on page 47. The following
sections are included in this chapter:
The installation of the system is divided into the following steps:
•Moving the Instrument
•Connecting the Gases and Cooling Water
Installation
•Connecting the System Components
•Installing the Quick-Change Adjustable Torch Module
•Installing and Setting Up the Autosampler
•Setting the Torch Viewing Position
•Switching on the System
•Setting Instrument Parameters
91
Installation
Setting Up the Computer and Printer
To install the ICP WinLab software, computer, and printer, refer to Winlab32
Software Installation and Administration Guide (Part No. 09936397).
Moving the Instrument
Caution
f moving the Optima 8000 will subject the instrument to any freezing temperatures
I
you must contact a PerkinElmer service engineer to assist you in the move. The
PerkinElmer service engineer will flush all traces of cooling water from the RF
generator to prevent freezing of RF generator components.
If you must move the instrument any great distance, especially if the instrument may
be subject to vibration or jolts contact a PerkinElmer service engineer to assist you in moving your system.
If you are only moving the instrument within the lab or to a laboratory nearby you
may use the following procedure.
•Remove the Quick-Change Torch module.
•Remove the chiller.
•Remove any accessories, for example an autosampler.
•Disconnect all gas and water lines.
•Remove the computer and printer.
•Attach the handles.
•Lift the instrument onto a movable table or put it in the wooden platform
originally shipped with the instrument. See the Unpacking Instructions (Part
No. 09931013).
92
•Lift into position at the new location. Make sure that the new location
complies with the laboratory requirements; see Laboratory Space Requirements on page 53.
•Remove the handles.
•At the new location reconnect the Quick-Change Torch module, the chiller,
any accessories, gas and water lines, the computer and the printer; see the
procedures later in this chapter Installing the Quick-Change Adjustable Torch Module on page 107.
Connecting the Gases and Cooling Water
After the instrument has been moved into its position, it can be connected to the
various services in the laboratory.
Connecting the Gases
Installation
Note
CautionThe pneumatic tubing kinks easily.Install it so that it is less likely to twist, fold and
All gas hose connections use 1/4 inch Swagelok fittings.
Use two wrenches when making Swagelok connections. Use a space collar (Part No.
09920125, supplied in the hose kit) at each Swagelok connection. Tighten the
Swagelok nut until you can no longer turn the knurled space collar. Only if you cannot locate a space collar, tighten the nut until it is finger-tight, then turn it an
additional 1/4 turn with a wrench. Pre-swaged fittings should be turned 1/8 turn past
finger-tight. Do not over-tighten fittings. Check for leaks with all the connections
made.
kink.
Connect the gases to the instrument as described in the following procedures.
Connecting the Nitrogen Supply
1. Connect the instrument to the nitrogen supply using the nitrogen hose (Part No.
N0690275, uses 1/4-in. Swagelok fitting) using a 1/4-in. space collar (Part No.
09920125) on both ends of the tubing. Refer to Figure 4-1.
2. Set the nitrogen supply regulator between 275 kPa and 825 kPa (2.75 to 8.2 5 bar
or 40-120 psig).
3. Check for leaks using a proprietary leak testing fluid for high purity gas lines.
93
Installation
1
23
654
7
Figure 4-1.Pneumatic, water and shear gas connections.
ItemDescription
1Air Supply
2Air/ N
3N
4Out (water Hose with White Tie Wrap)
5Cooling Water
6In (Water Hose with Red Tie Wrap)
94
7Cooling minimum pressure no maximum
Shear Gas and RF Cooling Gas Supply
2
Supply
2
Installation
Connecting the Argon Supply
Use the following procedure to connect the argon supply.
CautionGas delivery lines from the argon tank must be contaminant-free and not made of
plastic, although PTFE lines are acceptable.
1. Connect the instrument to the argon supply using the argon hose (Part No.
N0690274, uses a 1/4-in. Swagelok fitting) using a 1/4-in. space collar (Part No.
09920125) on both ends of the tubing.
2. Set the argon supply regulator between 550 kPa and 825 k Pa (5.5 to 8.25 b ar or
80-120 psig).
3. Check for leaks using a proprietary leak testing fluid for high purity gas lines.
Connecting the Shear Gas Supply
The shear gas used is typically compressed air; however, nitrogen may also be used.
The shear gas must be clean and dry, so the instrument includes an air dryer filter
assembly (Part No. N0775325) and is included in the Installation kit (Part No.
N0770432). For more information on the requirements for the shear gas supply, refer
to Pneumatic Requirements on page 58.
Note
An air hose (Part No. N0770348) with 1/4-in. Swagelok fittings at each end, is
supplied in the hose kit that is shipped with the instrument. A second air hose is also
included. The following procedure describes how to connect the PerkinElmer air
compressor, filter and regulator.
All If you are using house air, it must be clean and dry. The instrument
includes an air dryer assembly (Part No. N0775325). Refer to Connecting the Shear Gas Supply later in this chapter.
1. Install the Air Dryer Filter as described in the instructions that are included with
the filter.
2. Locate the air hose (Part No. N0770348, uses 1/4-in. Swagelok fittings at each
end). Connect the air hose from the air compressor to the AIR IN fitting on the
95
Installation
!
Warning
!
Warning
Air Dryer Filter using a 1/4-in. space collar at both ends. Make sure the air flow
is in the proper direction, as indicated by arrows on top of the filters.
3. Locate the second air hose (Part No. N0770348, uses 1/4-in. Swagelok fittings
at each end) that is shipped with the instrument in the hose kit. Connect the air
hose from the AIR OUT fitting on the Air Dryer Filter using a 1/4-in. space
collar at both ends.
Setting the Shear GasPressure
Do not set the air compressor pressure higher than 690 kPa (100 psig). The
maximum pressure and temperature that the filter bowls can withstand is 1035
kPa (150 psig) at 50 °C
(125 °F). At a higher pressure, the filter bowls can be blown off and cause injury.
The following procedure describes how to set the shear gas pressure when using the
PerkinElmer Air Dryer Filter.
1. Make sure the shut-off valve on the Air Dryer Filter is closed (knob turned fully
clockwise), then set the air pressure on the air compressor to between 550 kPa
and 825 kPa (5.5 to 8.25 bar or 80-120 psig).
96
2. Check that the metal covers are in position on the filter bowls and the filter bowl
holding rings are properly locked.
Before opening the shut-off valve, always check to make sure the filter bowls
are properly secured. Injury can result if the bo wls ar e blo wn off.
3. Open the shut-off valve fully (turn counterclockwise). Partial opening may
defeat the action of the water separator portion of the system.
4. Close the shut-off valve fully (knob turned fully clockwise), then set the air
pressure on the air compressor to between 550 kPa (5.5 bar and 80 psig) and 8 25
kPA (8.25 bar and 120 psig).
5. Connect the air hose from the Air Dryer Filter to the SHEAR GAS SUPPLY
fitting on the side of the spectrometer.
6. Open the shut-off valve fully (turn counterclockwise).
Connecting the PolyScience Chiller
The PolyScience® WhisperCoolTM Chiller must be used to supply cooling water to
the instrument. The PolyScience
instruction manual. You should be familiar with this manual before proceeding.
®
WhisperCoolTM Chiller comes with a complete
Installation
Note
A qualified electrician must install the single wall receptacle for the
PolyScience® WhisperCoolTM chiller.
®
The PolyScience
WhisperCoolTM is available through PerkinElmer in the
following two configurations:
208/230V, 60 Hz
220/240V, 50 Hz
Connecting the Chiller to the Instrument
Teflon tape (PTFE tape) should be used for all pipe-thread (NPT) fittings.
Use of Teflon tape can prevent leaks.
Connect the instrument to the chiller inlet using coolant drain hose. The direction of
the flow through the system can be controlled by the way the hoses are connected to
the chiller. The "INLET" port will draw liquid into the chiller; the “OUTLET" port
will pump liquid out. See the following figure for the location of the cables.
1. Locate the instrument return/Output coolant line (Part No. N0770342 and
connect to the chiller "INLET" fitting. The instrument water outputs have a
fitting and tubing to connect them to the return or drain. Use 3/8-in. space collars
(Part No. 09920584) at the instrument. The following figure shows the water
supply connections at the instrument.
2. Locate the instrument Input coolant lines (Part No. N0770341, uses 3/8-in
Swagelock fittings) and connect to the chiller “OUTLET” fitting. Use a 3/8-in.
space collars (Part No. 09920584) at the instrument. The following figure shows
the chiller connections. Figure 4-3 on page 101shows the water supply
connections at the instrument.
3. Make sure that the hoses and fittings are tight and that there are no bends or
crimps in the hoses.
97
Installation
Remote Chiller Connections
The chiller can be turned on or off via the software. Your service engineer will
connect the remote cable (Part No. N0770175) from the chiller to the Optima 8000
so that the chiller can be operated remotely. See the following figure for the location
of the cable connections.
98
Installation
11
10
1
2
7
4
5
9
6
3
8
Figure 4-2Remote chiller connections.
ItemDescription
1Terminal Strip Part No. 09987900 (For Chiller Remote)
2Coolant Out
33/8 in. Space Collars (Part No. 09920584)
4Remote Cable (Part No. N0770175)
99
Installation
ItemDescription
5Coolant Drain Hose (Part No. N0770342) White Tie Wrap
6From Optima Out
7Coolant In
8To Optima In
9Coolant Supply Hose (Part No. N0770341) Red Tie Wrap
10Line Cord Plug
11Chiller
Remote Chiller Operations
The remote chiller will automatically be on when the Optima 8000 Spectrometer is
warming up. The warming up period includes just being turned on, coming out of
Sleep mode or Standby mode.
If the instrument has just been turned on, starting WinLab32 will automatically begin
warm up. Initiating WinLab 32 will have no effect on the chiller if the chiller is
already on.
100
Note
If the Optima 8000 Spectrometer is in Sleep or Standby mode and the instrument is
still ignited, the chiller will remain on. If the instrument is not ignited in Sleep or
Standby mode the chiller will turn itself off.
The plasma should never be ignited if the chiller is off.
Installation
1
2
4
3
7
6
5
Figure 4-3Water flow diagram with the chiller
ItemDescription
1Terminal Strip Part No. 09987900 (For Chiller Remote)
23/8 in. Space Collars (Part No. 09920584)
3Remote Cable (Part No. N0770175)
4Coolant Out (Part No. N0770342) White Tie Wrap
101
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.