PerkinElmer OilExpress System User Manual

OilExpress System
User’s Gui de
Release History
Part Number Release Publication Date
L1050039 J December 2010
Any comments about the documentation for this product should be addressed to: User Assistance
PerkinElmer Ltd Chalfont Road Seer Green Beaconsfield Bucks HP9 2FX United Kingdom
Or emailed to: info@perkinelmer.com
Notices
The information contained in this document is subject to change without notice .
Except as specifically set forth in its terms and conditions of sale, PerkinEl me r mak es no warranty of any kind with regard to this document, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in connection with furnishing, performance or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this publication may be reproduced in any form whatsoever or translated into any language without the prior, written permission of PerkinElmer, Inc.
Copyright © 2010 PerkinElmer, Inc. Produced in the UK.
Trademarks
Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are protected by law.
PerkinElmer is a registered trademark of PerkinElmer, Inc. Frontier and Spectrum are trademarks of PerkinElmer, Inc.
Contents
Introduction ............................................................................................... 5
About this Manual ............................................................................................. 6
Using this Guide ......................................................................................... 6
Conventions ...................................................................................................... 7
Conventions Used in this Manual ................................................................. 7
Part Numbers ............................................................................................. 7
Notes, Warnings and Cautions ..................................................................... 7
The OilExpress System .....................................................................................10
A Guided Tour of the OilExpress System ............................................................11
Power Switch and Communications Ports ....................................................12
The Liquid Autosampler ....................................................................................13
X – Y – Z Arm ............................................................................................14
Pumps ......................................................................................................14
Sample Injection/Wash Port .......................................................................15
Sampling Probe/Needle ..............................................................................15
Sample Rack .............................................................................................16
Solvent Bottle ............................................................................................16
Overview of the Sampling Procedure ..........................................................17
Analyzer Measurement .....................................................................................20
Analyzer Sample Compartment ...................................................................20
2-Port Flowcell ..........................................................................................21
FT-IR Liquid Autosampler for Oils: Specifications ................................................22
Warnings and Safety Information ........................................................... 23
OilExpress System Safety Summary ...................................................................24
General Safety .................................................................................................25
Location and Ventilation .............................................................................26
Mechanical Hazards ...................................................................................26
Electrical Safety .........................................................................................26
Samples and Waste ...................................................................................27
Laser Safety Regulations...................................................................................28
Warning Labels ................................................................................................29
Autosampler Labels ...................................................................................29
Warning Signs on the Instrument ...............................................................30
EMC Compliance ..............................................................................................31
EC Directive ..............................................................................................31
FCC Rules and Regulations .........................................................................31
Use of Flammable Solvents and Samples ...........................................................32
Heptane ....................................................................................................33
Material Safety Data Sheets ..............................................................................35
Unpacking and Installation ...................................................................... 37
Overview .........................................................................................................38
Instrument Requirements .................................................................................39
Unpacking .......................................................................................................40
Unpacking the Autosampler ........................................................................40
Unpacking and Installing the Sample Shuttle ...............................................41
Positioning the OilExpress System .....................................................................42
Installing the Liquid Autosampler ......................................................................43
Electrical Connections ................................................................................43
Installing the Fluid Line Suppor t Arm ..........................................................44
Installing Racks (if ordered) .......................................................................45
Autosampler Tubing Connections ......................................................................46
Fluid Tubing System ..................................................................................46
Installing the Flowcell in the Sample Shuttle ................................................49
Waste Container ........................................................................................49
Installing the Syringe .................................................................................50
PC Software and Hardware Requirements ......................................................... 51
Software Requirements ............................................................................. 51
Hardware Requirements ............................................................................ 51
Upgrading from a Previous Version of OilExpress ............................................... 52
Upgrading from OilExpress v3.x ................................................................. 52
Upgrading from earlier versions ................................................................. 52
Installing the OilExpress Software ..................................................................... 54
Default Users for OilExpress ............................................................................. 58
Starting up OilExpress and Configuring your Instrument .................................... 59
Calibration ....................................................................................................... 60
Calibrating the Autosampler ....................................................................... 60
Calibrating the Pathlength of the Flowcell ................................................... 66
Installing the Arm Shield ........................................................................... 67
Background to Used Oil Analysis ............................................................. 69
Background to Used Oil Analysis ....................................................................... 70
Infrared Absorption ................................................................................... 70
Why Analyze Used Oils? ............................................................................ 70
Methods of Analysis .................................................................................. 71
Degradation Products ................................................................................ 72
Common Contaminants ............................................................................. 73
Analysis Methods ...................................................................................... 73
Sample Preparation ......................................................................................... 74
Filtering the Oil ......................................................................................... 74
Sample Presentation Requirements ............................................................ 74
Routine Maintenance and Troubleshooting ............................................. 75
Maintenance Schedule ..................................................................................... 76
Autosampler Maintenance Procedures ............................................................... 77
Flushing the Syringe Pump ........................................................................ 77
Cleaning the Syri nge Seal .......................................................................... 77
Cleaning the Sampling Probe ..................................................................... 77
Cleaning the Sample Rack ......................................................................... 77
Cleaning the Injection and Wash Port ......................................................... 77
Cleaning the Z-Rack .................................................................................. 77
Cleaning the Splas h Guard/Safety S hields ................................................... 78
Cleaning the Lead Screw and Lead Screw Encoder ...................................... 78
Cleaning X-Y-Z Mechanis m ........................................................................ 78
Inspecting the Insulation Block/ALIDUM Cable Asse mbly ............................. 79
Checking the Fluid Path (Tubing) ............................................................... 79
Replacing the Sample Probe/Needle ........................................................... 79
Replacing the Syring e Seal ........................................................................ 80
Replacing a Syringe ................................................................................... 80
Replacing the Rea gent Syringe Seal ........................................................... 80
Replacing the 3-port Valve ......................................................................... 82
Replacing the Insulation Block/ALIDUM Cable Assembly .............................. 82
Lubrication ............................................................................................... 84
Changing the Autosampler Fuses ............................................................... 84
Replacing the Tubing ................................................................................ 85
Replacing Reagent and Needle or Cell Tubing ............................................. 85
Replacing Waste Tubing ............................................................................ 86
Maintenance of the Flowcell ............................................................................. 87
Routine Cleaning of the Flowcell, Using Software Utilities ............................ 87
Deep Cleaning the Flowcell, Involving Disassembly ..................................... 88
Warranty Exclusions and Limitations ................................................................. 91
Spares and Consumables ................................................................................. 92
Troubleshooting .............................................................................................. 93
Appendices ............................................................................................... 97
Appendix 1: Decontamination ........................................................................... 98
Appendix 2: WEEE Instructions for PerkinElmer Products ................................... 99
Index ............................................................................................................ 100

Introduction

6 . OilExpress System User's Guide

About this Man ual

This manual contains the following sections :
Introduction
Warnings and Safety Information
Unpacking and Installation
Background to Used Oil Analysis
Routine Maintenance and Troubleshooting
Index

Using this Guide

This guide tells you how to install and maintain the OilExpress System; we recommend that you use it as follows:
1. Read
2. Read
3. Follow the procedures in
4. Read the
5. Read
You should use this guide in conjunction with the docume ntation supplied wit h your spectrometer.
Warnings and Safety Information
OilExpress System.
The OilExpress System
on page 10, to learn about the analyzer.
Unpacking and I nstallation
Maintenance Schedule
System requires.
Background to Used Oil Analysis
and the information that can be obtai ned by analyzing used oils.
on page 23 and this section before usi ng your
on page 37.
on page 76 to learn what maintenance the OilExpress
on page 69 to learn about sample preparation
Introduction . 7

Conventions

Conventions Used i n this Manua l

Normal text is used to provide information and instructions. Bold text refers to text that is displayed on the screen. UPPERCASE text refers to keys on the PC keyboard. ‘+’ is used to show that you have to
press two keys at the same time, for example, ALT+F.

Part Numb ers

All eight-digit numbers are PerkinElmer part numbers unless stated otherwise.

Notes, Warnings and Cautions

Three terms, in the following standard formats, are also used to highlight special circumstances and warnings.
NOTE: A note indicates additional, significant information that is provided with some
procedures.
8 . OilExpress System User's Guide
einen Geräteschaden zu vermeiden.
undgå en beskadigelse af apparatet.
este tipo de circunstancias.
We use the term CAUTION to inform you about situations that could
CAUTION
result in serious damage to the instrument or other equipment. Details about these circumstances are in a box like this one.
Caution (Achtung)
Bedeutet, daß die genannte Anleitung genau befolgt werden muß, um
Caution (Bemærk)
Dette betyder, at den nævnte vejledning skal overholdes nøje for at
Caution (Advertencia)
Utilizamos el término CAUTION (ADVERTENCIA) para advertir sobre situaciones que pueden provocar averías graves en este equipo o en otros. En los recuadros como éste se proporciona información sobre
Caution (Attention)
Nous utilisons le terme CAUTION (ATTENTION) pour signaler les situations susceptibles de provoquer de graves détériorations de l'instrument ou d'autre matériel. Les détails sur ces circonstances figurent dans un encadré semblable à celui-ci.
Caution (Attenzione)
Con il termine CAUTION (ATTENZIONE) vengono segnalate situazioni che potrebbero arrecare gravi danni allo strumento o ad altra apparecchiatura. Troverete informazioni su tali circostanze in un riquadro come questo.
Caution (Opgelet)
Betekent dat de genoemde handleiding nauwkeurig moet worden opgevolgd, om beschadiging van het instrument te voorkomen.
Caution (Atenção)
Significa que a instrução referida tem de ser respeitada para evitar a
danificação do aparelho.
Verletzung des Benutzers kommen kann.
overholdes.
este tipo de circunstancias.
dans un encadré semblable à celui-ci.
informazioni su tali circostanze in un riquadro come questo.
genomen, dit kan leiden tot verwondingen van de gebruiker.
WARNING
We use the term WARNING to inform you about situations that could result in personal injury to yourself or other persons. Details about these circumstances are in a box like this one.
Introduction . 9
Warning (Warnung)
Bedeutet, daß es bei Nichtbeachten der genannten Anweisung zu einer
Warning (Advarsel) Betyder, at brugeren kan blive kvæstet, hvis anvisningen ikke
Warning (Peligro)
Utilizamos el término WARNING (PELIGRO) para informarle sobre situaciones que pueden provocar daños personales a usted o a otras personas. En los recuadros como éste se proporciona información sobre
Warning (Danger)
Nous utilisons la formule WARNING (DANGER) pour avertir des situations pouvant occasionner des dommages corporels à l'utilisate u r ou à d'autres personnes. Les détails sur ces circonstances sont données
Warning (Pericolo)
Con il termine WARNING (PERICOLO) vengono segnalate situazioni che potrebbero provocare incidenti alle persone. Troverete
Warning (Waarschuwing)
Betekent dat, wanneer de genoemde aanwijzing niet in acht wordt
Warning (Aviso)
Significa que a não observância da instrução referida poderá causar um
ferimento ao usuário.
10 . OilExpress System User's Guide

The OilExpress Sys tem

The OilExpress System is optimized for the determination of lubricant degradation products. Based on infrared spectroscopy, the system is capable of measuring the following parameters in lubricants:
Soot loading
Anti-wear additive depletion
Glycol
Water
Nitration
Oxidation
Sulfonation
Fuel contamination
Introduction . 11

A Guided Tour of the OilExpress System

Figure 1 The OilExpress System
The PerkinElmer OilExpress System (Figure 1) is a bench-top instrument that provides all the following in one self-contained unit:
A spectrometer, fitted with a sample shuttle.
NOTE: The spectrometer may be a Frontier IR System, Spectrum 100 or 400 Series, or
Spectrum One instrument. For full details on how to install and use your spectrometer, refer to the documentation supplied with your instrument.
A Liquid Autosampler designed to analyze up to 60 samples an hour.
A water-resistant flowcell.
The system is controlled by dedicated oil analysis OilExpress software resident on a PC.
NOTE: See the help system available from the Help menu in the OilExpress software, which
includes a tutorial and full information on using the software.
12 . OilExpress System User's Guide

Power Switch and Communications Ports

The power switch, AC power cable connector, and communications ports are o n the rear of the spectrometer. The power switch is marked I/O (on/off).
The power switch, AC power cable connector, and communications ports are located on the left-hand side of the Autosampler (see Figure 2).
Power switch
Figure 2 Location of the Autosampler power switch
The spectrometer is connected to a PC, either point-to-point or over a network. The Liquid Autosampler module is connected to the same PC via an RS232 link.
Introduction . 13

The Liquid Autosampler

The OilExpress System uses a Liquid Autosampler to enable the rapid sampling of up to 60 oil samples per hour. Figure 3 shows a general view of the Liquid Autosampler.
The Liquid Autosampler provides:
Minimal sample carry-over by flus hing the needle and flowcell with solvent.
A 2-port flowcell that is designed for the analysis of viscous liquids.
Flexible sample presentation: a rack that holds 30 sample bottles, or test tube racks
with a capacity of 240.
Solvent bottle.
Two high-resolution syringe pumps to provide robust, reliable liquid handling.
Figure 3 The liquid autosampler
14 . OilExpress System User's Guide
X-frame
Arm
Insulation Block
Z- rack
Coax Cable
CCU-9000
ADRI-9 Board
ADRI-9 Board Cover
X-slide

X – Y – Z Arm

Figure 4 X – Y – Z arm mechanism
The X – Y – Z arm movements are controlled by stepper motors linked by drive belts.

Pumps

The Autosampler uses two syringe pumps. These pumps are located on the back panel of the auto stepper motor driven syringe pump (Figure 5). The pump consists of a syringe drive, a valve drive, and control electronics. The head of the syringe is attached to the valve assembly that controls the input and output of the fluid path. The plunger of the syringe is attached to the syringe drive that moves the plunger up and down.
Figure 5 The sample pump
Introduction . 15
Injection port

Sample Injection/Wash Port

The Sample Injection/Wash Port (Figure 6 and Figure 7) consists of a waste transfer and port. After completion of a liquid handling function, the excess sample is discarded in the waste port. The probe is then moved to the wash port for cleaning. Washing is accomplished by pumping wash fluid through the probe. The fluid then flows around the outside of the probe and overflows into the waste port. The waste fluid then travels by gravity out of the waste tubing to a waste reservoir located below the instrument.
Sample injection & wash ports
Waste exit port
Interconnect tubing port
Figure 6 Sample injection and wash ports (s ide v iew)
v2
v3
Wash station
Waste position
v1
Bolt
Figure 7 Sample injection and wash ports (top view)
The transfer port is used to transfer the sample to the flowcell located in the sample compartment of the spectrometer. The transfer port consists of a nylon fitting into which the sample probe is lowered.

Sampling Probe/Needle

The sample is drawn through the stainless steel needle and delivered to the flowcell via the transfer port. The needle is cleaned in the wash por t: the solvent flows down inside the needle, and then is forced up around the outside and to waste.
16 . OilExpress System User's Guide

Sample Rack

A 30-bottle sample rack is supplied as shown in Figure 8.
Figure 8 Autosampler sample rack

Solvent Bottle

In the event of a solvent spillage, isolate the power to the OilExpress System immediately and then clean up the solvent spill. The system can then be powered up but the spectrometer should be allowed to stabilize
WARNING
The recommended solvent for the OilExpress System is heptane. This should be aspirated from a supply adhering to laboratory safety standa r d s that advise on its safe storage, handling, use and disposal.
If using the solvent bottle supplied with the OilExpress System (Figure 9), the following is recommended:
The solvent bottle should be re-filled away from the OilExpress System, in a
well-ventilated area, and precautions should be taken against static discharge.
The bottle lid should be used at all times. The solvent bottle lid should be correctly
fitted to the bottle during operation of the OilExpress System.
The bottle should be stored to the rear of instrumentation at a location where it will not
be knocked accidentally.
for at least two hours before measurements are taken.
Introduction . 17
Solvent bottle lid
Solvent bottle
Figure 9 Solvent bottle

Overview of the Sampling Procedure

1. At the start of an ana lysis, the wash volume of solvent is drawn up by each pump.
2. The sample pump valve switches to the needle and the cell pump switches to the cell, and the solvent is dispensed.
3. The needle is lowered into the sample and the sample is aspirated by the drawing down of the sample pump.
The default volume aspirated is 3.5 ml.
18 . OilExpress System User's Guide
4. The needle then moves to the transfer port.
5. The sample pump dispenses the sample to the cell. Only 2.1 ml of sample is dispensed to the cell. To aid the transfer of the viscous liquid and prevent pressure build up, the cell pump is
simultaneously drawn back by the same volume.
6. The OilExpress software performs a sample det ection.
7. When the sample has been successfully detected the needle moves to the waste port to dispense the remaining sample.
8. The sample pump valve switches position so that the pump is connected to the solvent reservoir.
9. The pump aspirates the solvent.
10. The sample pump valve then switches back to the needle.
11. The needle moves and lowers into the wash port.
Introduction . 19
12. The sample pump dis penses the solvent to flush out the tubing and the needle. When the needle has been flushed the next sample can be aspirated.
13. When OilExpress has completed scanning the sample, the cell pump valve switches to the cell and the pump moves to push the sample back through the cell and back through the trans fer port where it then flows to the adjacent wash port.
14. The cell pump valve switches to the reservoir and aspirates solvent.
15. The cell pump valve then switches back to the cell and the pump moves up to flush the cell and the tubing.
Waste from this process flows back through the cell and the transfer port to waste.
20 . OilExpress System User's Guide

Analyzer Measurement

Analyzer Sample Compartment

DO NOT use the OilExpress System when the sample compartment lid is fitted. Flammable vapor may seep through instrument seals and gasket s and accumulate under the lid, where it could cause a fire or explosion
WARNING
CAUTION
hazard.
The sample compartment windows may be composed of a hygroscopic material, such as KBr; which, although coated, can be damaged by high levels of humidity. Refer to the documentation supplied with your spectrometer for details.
If you spill liquid in the sample compartment, wipe it up quickly.
Relative humidity higher than 80 % can damage the windows of the
CAUTION
The sample compartment (Figure 10) is located at the front of the instrument.
Figure 10 Sample compartment of a Frontier IR System
sample compartment.
The infrared beam enters the compartment through an aperture on the left. After passing through the sample, it enters the detector area through an aperture on the right side of the sample compartment.
A sample shuttle is installed in the sample compartment. The sample shuttle is used to move the flowcell of the Liquid Autosampler in and out of the beam.
Introduction . 21

2-Port Flowcell

The 2-port flowcell (Figure 11) is used in a sample shuttle so that a background spectrum can be collected regularly if sampling times are long, or to compensate for changing environmental co nditions. The sample shuttle moves the flowcell in and out of the beam between collecting a background and sample spectrum.
Flowcells are fitted with zinc selenide (ZnSe) windows for use with mid infrared spectrometers. Each flowcell is fitted with two windows.
Figure 11 The 2-port flowcell
Filling and emptying the flowcell
In the 2-port flowcell, the sample enters the flowcell at the bottom of the cell. When the cell is emptied, the sample is pushed back out of the bottom of the flowcell, up through the transfer port and evacuates to waste by gravity.
Rinsing the flowcell
When the flowcell is rinsed, the solvent enters the flowcell at the top of the cell, and leaves it at the bottom, then passes through the transfer/wash port to waste.
22 . OilExpress System User's Guide

FT-IR Liquid Autosampler for Oils: Specifications

Sample Viscosity Range 4–1000 cSt (centiStokes)
Sample Throughput Approximately 50 per hour for 200 cSt samples Sample Carry-over <0.1%
Wash Solvent Carry-over 1% typical
Capacity 30 × 4 oz bottles / 240 × 16 mm vials
Minimum Sample Vo lume 4 ml Flowcell – mid infrared Flow-through trans mission with ZnSe windows:
0.1 mm pathlength
Warnings and Saf et y
Information
24 . OilExpress System User's Guide

OilExpress System Safety Summary

The PerkinElmer OilExpress System has been designed to comply with a wide variety of international standards governing the safety of laboratory equipment. In routine use, the OilExpress System poses virtually no risk to you. If you take some simple, common sense precautions, you can make sure that you maintain the continued safe oper ation of the OilExpress System.
DO make sure that the OilExpress System is properly connected to the electrical supply; in particular make sure that the ground (earth) is securel y connected.
DO take care with solvents: heptane is highly flammable. Remove the solvent bottle and move it away from the OilExpress System before you fill it. Always keep the lid on the solvent bottle.
DO disconnect the electrical power cable before opening the main cover of the spectrometer. DO keep the OilExpress S ystem dry. Avoid spilling liquid into the OilExpress System. Clean all
external spills immediately. If anything that is spilled enters the main body of the OilExpress System, switch off the power and call a PerkinElmer Service Engineer.
DO NOT stare into the laser beam. The spectrometer contains a low power, visible (r ed) laser; momentary exposure to the beam is not dangerous, but deliberate, direct viewing of the beam along its axis could damage your eye.
DO NOT use the OilExpress System when the sample compartment lid is fitted. Flammable vapor under the lid could be a fire or explosion hazard.
DO NOT use a flammable gas to purge the spectrometer. The spectrometer contains a hot source, and a fire or explosion will result. Only use clean, dry, oil-free nitrogen or air to purge the instrument.
DO NOT place hands b eyond the line at the front of the Autosampler enclosure whilst operating the OilExpress Sy stem.
Read the more detailed information on warnings and safety in the following pages, and in the documentation supplied with your spectrometer, to ensure the safe operation of the instrument.
Warnings and Safety Information . 25

General Safety

The OilExpress System has been designed and tested in accordance with PerkinElmer specifications and in accordance with the safety requirements of the International Electrotechnical Commission (IEC). The spectrometer conforms to IEC publication 61010-1 (“Safety requirements for electrical equipment for measurement, control, and laboratory use”) as it applies to IEC Class 1 (earthed) appliances and therefore meets the requirements of EC low voltage directive 2006/95/EC.
If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired. Only use the OilExpress System indoors and under the following conditions:
Temperature: 15 °C to 35 °C Relative Humidity: 80% maximum (non–condensing)
If possible, avoid any adjustment, maintenance and re pair of the opened, operating instrument. If any adjustment, maintenance and repair of the opened, operating instrument is necessary, this must only be done by a skilled person who is aware of the hazard involved.
Whenever it is likely that the OilExpress System is unsafe make it inoperative. The OilExpress System may be unsafe if it:
Shows visible damage.
Fails to perform the intended measurement.
Has been subjected to prolonged storage in unfavorable conditions.
Has been subjected to severe transport stresses.
Use of controls or adjustments or performance o f p rocedures other than those specified herein may result in hazardous radiat ion exposure.
WARNING
The OilExpress System has been designed to be safe under the following environmental conditions:
Indoor use.
Altitude up to 2000 m (above mean sea level).
Ambient temperatures of 5 °C to 40 °C.
If the equipment is used in a manner not specified her ein the protection provided by the equipment may be impaired.
A maximum ambient relative humidity of 80% for temperatures up to 31°C, decreasing linearly to 50% relative humidity at 40 °C.
Mains supply fluctuations not exceeding ±10% of the nominal voltage.
26 . OilExpress System User's Guide

Location and Ventilation

To allow for adequate cooling, t he OilExpress System should not b e sited near to room heating equipment, for example, central heating radiators.
During operation, there should be a minimum gap of:
6 inches (15 cm) between any sur face and the cooling louvers at the rear of the spectrometer.
3 inches (7 cm) between the OilExpress System and adjacent equipment.
1 inch (2.5 cm) between any wall or obst ructing surface and the ventilation panel on
the back of the Auto sampler.
Make sure that the switches at the electrical supply inlet on the rear and
.
side of the OilExpress System are not obstructed
WARNING

Mechanical Hazards

Whenever the autosampler is in operation there is the risk of injury from moving mechanical parts. The autosampler is de signed for automatic hands-off operation only. Never reach into the autosampler workspace
WARNING
when the autosampler is in operation
.

Electrical Safety

Connect both units comprising the OilExpress System to a power supply line that includes a switch or other means of disconnection from the electricity supply.
Only plug the OilExpress System into an electricity supply socket that is provided with a protective earth connection.
When fuses need replacing, use only those with the requi red current rating and of the specified type. Do not use makeshift fuses and do not short-circuit fuse holders. On the Autosampler, replace both fuses if one blows as the second one will be weakened.
When the OilExpress System is connected to its electricity supply, terminals may be live and the removal of covers other than those that can be removed by hand is likely to expose live parts.
Capacitors inside the OilExpress System may still be c harged even if the instrument has been disconnected from all voltage sources.
Warnings and Safety Information . 27
The OilExpress System must be disconnected from all voltage sources before it is opened for any adjustment, replacement and maintenance, or cleaning up of any major liquid spills.
Do not touch any swi tches or outlets with wet hands.
Any interruption of the protective earth conductor inside or outside of any components comprising the OilExpress System or disconnection of the protective earth terminal can make the instrument dangerous.
WARNING
The OilExpress System has:
An IEC Pollution Degree 2 classification – usually only non-conductive atmospheric pollution of the equipment occurs; occasionally, however, a temporary conductivity caused by condensation must be expected.
An IEC Insulation class I rating for external circuits – only connect equipment that meets the requirements of IEC 61010-1, IEC 60950 or equivalent st andards.
The instrument is designed to be safe under transient overvoltages typically present on the MAINS supply.
NOTE: The normal level of transient overvoltages is impulse withstand (Overvoltage)
category II of IEC 60364-4-443.

Samples and Waste

Samples
When using the flowcell, to ensure that the sample will flow through the flowcell the maximum particle diameter must not exceed 50 µm.
Solvent
The recommended solvent for use in the OilExpress System is heptane. Observe the warnings in
Use of Flammable Solvents and Samples
Waste
DO position the waste container on a surface that is at l east 90 cm (36 inches) below the bench on which the OilExpress System is located because the flow of waste relies on gravity.
DO NOT immerse the ends of the tubes in the waste.
on page 32 when using hep t a ne.
Disposal of waste
Observe local and national regulations when disposing of waste from the OilExpress System .
28 . OilExpress System User's Guide

Laser S afe ty Reg ula tio ns

The OilExpress System is a CDRH Class I, BS EN 60825-1/IEC 60825-1 Class 1 laser product as defined by IEC 60825-1. The optical module contains a Class 2 Helium Neon (HeNe) laser, which emits visible, continuous wave radiation at a wavelength of 633 nm and has a maximum output power of 1 mW. Some diffuse HeNe laser radiation, within Class 1 limits, emerges from:
The window in the left hand side of the sample compartment when an internal beampath is selected.
An external beam port when the beam port cover is removed, no accessory is fitted at the port, and the beampath to the port is selected.
Do not stare into any laser beam. Staring into a laser beam (intrabeam viewing) can cause permanent damage to your eyes.
WARNING
The laser is automatically shut down when the main cover of the instrument is raised.
Do not attempt to override or modify the interlock system. Use of controls or adjustments or performance of procedures other tha n
WARNING
The instrument complies with the following laser safety regulations:
1. 21 CFR Part 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No.50 dated 24 June 2007. Administere d by the Center for Devices and Radiological Health, U.S. Department of Health and Human Services.
2. EN 60825-1:2007 – “Safety of laser products – Part 1: Equipme nt classification and requirements”.
For further information on the safe use of your spectrometer, refer to the documentation supplied with the i nstrument.
those specified herein may result in hazardous radiat ion exposure.
Warnings and Safety Information . 29

Warning Labels

When this label is attached to an instrument it means refer to the manual in order to find out the nature of the potential hazard and any actions which have to be taken.

Autosampler Labels

Labels are fixed to the Autosampler in the locations shown in Figure 12 & Figure 13.
Figure 12 Autosampler labels (front)
Figure 13 Autosampler labels (rear and left side)
30 . OilExpress System User's Guide

Warning Signs on the Instrument

Caution, hot surface
Caution, risk of electric shock
Caution, laser radiation hazard
Caution risk of danger. Refer to accompanying documents in all cases where this symbol is used
to find out the nature of the potential HAZARD and any actions which have to be taken.
Warnings and Safety Information . 31

EMC Compliance

EC Directive

The OilExpress System has be e n designed and tested to meet the requirements of the EMC Directive 2004/ 10 8/E C. The OilExpress System complies with the EMC standard EN 61326 (EMC standard for electrical equipment for measurement, control and laboratory use).

FCC Rules and Regulations

This product is classified as a digital device used exclusively as industrial, commercial, or medical test equipment. It is exempt from the technical standards specified in Part 15 of the FCC Rules and Regulations, based on Section 15.103 (c).
32 . OilExpress System User's Guide

Use of Flammable Solvents and Samples

The spectrometer contains a hot source and contact with flammable vapors may cause an explosion. When working with flammable solvents or samples, parti cularly during unattended operation with flow-cells, it is
WARNING
WARNING
WARNING
recommended that the instrument optics area should be continuously purged with dry air or nitrogen to maintain a positive pressure and prevent flammable vapor entering the instrument.
DO NOT use the OilExpress System when the sample compartment lid is fitted. Flammable vapor may seep through instrument seals and gasket s and accumulate under the lid, where it could cause a fire or explosion hazard.
If flammable solvents or samples are spilled on the instrument and there is any possibility that they have entered the interior (b y coming into contact with cover gaskets for example) then the instrument must be switched off immediately and disconnected from the power supply. The optics area should then be thoroug hly purged with dry air or nitrogen, or the main cover should be opened to thoroughly ventilate the optics area before proceeding.
WARNING
WARNING
Flammable solvents or samples should not be stored on or near the instrument. Handling of such materials during prepa r ation should be performed in a safe area away from the instrument such as a fume cabinet.
Some chemicals used with this instrument may be hazardous or may become hazardous after completion of an analysis. The responsible body (for example, the Laboratory Manager) must take the necessary precautions to make sure that the surrounding workplace is safe and that the instrument operators are not exposed to hazardous level s of toxic substances (chemical or biological) as defined in the applicable MSDS (Material Safety Data Sheets) or OSHA (Occupational Safety and Health Administration (United States)), ACGIH (American Conference of Governmental Industrial Hygienists (United States)), or COSHH (Control Of Substances Hazardous to Healt h (United Kingdom)) documents. Venting for fumes and disposal of waste must be in accordance with all national, state and local health and safety reg ula tions and laws.
WARNING
before
WARNING
Warnings and Safety Information . 33
We recommend that you use heptane as the rinse solvent in the OilExpress System. This highly flammable solvent presents no hazard during routine use of your OilExpress System, but:
DO clean up leakages and spillages immediat ely and thoroughly.
DO move the solvent bottle away from the OilExpress System you fill it.
DO always keep the lid on the solvent bottle.
DO NOT allow the syringe pumps to run dry; make sure that there is always sufficient solvent in the solvent bottle.
DO make sure that the waste tubing is not blocked and that it outlets to a suitable waste container which stands on a surface at least 90 cm below the bench on which the OilExpress System is located.
The end of the waste tubing should not be immersed in the waste material.
DO read and observe the information supplied with the solvent.

Heptane

Safety information for heptane is summarized below. See the Materials Safety Data Sheet (MSDS) for heptane, s upplied by your local Safety Officer, for full details.
Physical data
Description Clear, colorless liquid with a faint gasoline-like odor
Boiling point 98 °C
Specific gravity 0.6837
Solubility in water Insoluble Vapor pressure 40 mmHg at 20 °C
Vapor density 3.45
34 . OilExpress System User's Guide
Fire and explosion hazard
Heptane is flammable (USA definition)/highly flammable (European definition)
Flash point -4 °C
Explosive limits 1.05 - 6.7 % Ignition temperature 223 °C
Firefighting measures Extinguish fire with foam, dry powder, carbon dioxide or
vaporizing liquid
Storage and handling
When storing heptane, observe local and national regulations. Keep containers tightly closed. When handling heptane, wear safety goggles and protective gloves. Do not breathe the
vapor, and avoid contact with eyes, skin and clothing.
Warnings and Safety Information . 35

Material Safe ty Da ta Shee ts

Some of the materials described in these procedures are hazardous. Appropriate safety equipment and clothing should be used when handling them. See the Materials Sa fety Data Sheets (MSDS) supplied by your local Safety Officer for details.
WARNING
You can search for up-to-date copies of safety data sheets on materials, such as ZnSe, used in PerkinElmer products that are known to have safety issues from the Technical Resources section of the PerkinElmer website. The MSDS information is availab le in a range of languages, and includes data items required in specific national, supra-national and state jurisdictions.
To obtain a safety data sheet for a particular compound, follow the steps described below.
NOTE: To read MSDS .pdf files you will need Adobe Reader 5.0 or later. An installation of
this software is available on the
They must be disposed of with care, following your laboratory procedures.
Software Utilities CD
.
1. Launch your web browser and navigate to the PerkinElmer web site: www.perkinelmer.com If you are not redirected automatically you may have to select the home page
appropriate to your location.
2. Search for the term MSDS using the se arch box located at the top of the home page. The Search for Material Safety Data Sheets (MSDS) page is displayed.
3. Enter the key words for the compound, in the Product name box, and then click Go. A full list of all MSDS documents that refer to the compound is displayed.
4. Select the MSDS document you want to view.
36 . OilExpress System User's Guide
Unpacking and
Installation
38 . OilExpress System User's Guide

Overview

Installation of the OilExpress System and OilExpress software will be performed by a PerkinElmer Service Engineer. The installation proce d ure is summarized below.
Unpack the Autosampler
Unpack and install the sample shuttle in your spectrometer
Connect the Autosampler
Install OilExpress software
Configure the OilExpress system
Calibrate the OilExpress system
NOTE: For details of how to unpack and install your spectrometer, please refer to the
documentation s upplied with your instrument.
Unpacking and Installation . 39

Instrumen t Requi reme nts

NOTE: Read the warnings and safety information at the start of this manual before installing
the OilExpress System. They contain important information.

Electrical requirements

The OilExpress System can operate on electr icity supplies of 50 or 60 Hz and on voltage ranges of between 100 and 120 V or between 220 and 240 V.
The line supply must be within 10% of t he nominal voltage .
The power rating of the Autosampler is 300 W.
If possible, do not connect the OilExpress System to circuits that have heavy-duty
equipment, such as large motors, connected.
If possible, do not use photocopiers, discharge lamps, radio transmitters, and other equipment with large or frequent transient loads on the same supply circuit.

Environment

To obtain the best performance from your OilExpress System:
Place the OilExpress System in an environment that is relatively dust-free.
Make sure that the bench top is free from vibrations or mechanical shocks.
Do not place the OilExpre ss System or the PC near to room heating equipment, for
example, central-heating radiators.
Leave at least 7 cm (3 inches) from any vertical obstacle to the sides of the OilExpress System, to permit an adequate flow of cooling air.
Make sure that there are no overhanging shelves, and no water pipes or faucets that could leak onto the OilExpress System.
The area near the PC must be free of strong magnetic fields, direct sunlight, and heating or cooling units or ducts.
40 . OilExpress System User's Guide

Unpacking

Unpacking the Autosampler

The Autosampler is a heavy instrument with a basic weight of approximately 30 kg, so two people are required for safe handling.
WARNING
When you receive the instrument, look at the exterior of the container for possible
shipping damage. Check for the following items:
– Water damage or discoloration – Cuts or gashes – Collapsed corners
Consult the local codes of practice issued by safety advisors before attempting to lift it.
Take care not to injure yourself or others, or to drop the Autosampler.
– Crushed top or sides – Other physical damage
NOTE: If any of the above items are found, contact PerkinElmer and the shipping agent
immediately.
Before unpacking the Autosampler, make sure that there is sufficient space to lay out the instrument and accessories. This makes checking the contents of the box against the packing list easier. Follow the steps below to prevent damage to the instrument:
1. Remove the plastic shrink-wrap.
2. Cut the two (2) bands.
3. Lift off the top of the shipping carton.
4. Empty the small accessories from the end sections of the carton.
5. Gently lift the entire upper portion o f the inner container and set it aside.
6. Remove the styrofoam cushions on top of the instrument.
7. Remove instrument from the shipper and place it on a work surface.
Do not lift the Autosampler by its arm guard.
CAUTION
Unpacking and Installation . 41
8. Inspect instrument for any physical damage.
9. Open the box of accessories and place the contents on an open surface.
10. Check to make sure that all parts have been received.
NOTE: If any parts are missing from the packing list, contact PerkinElmer immediately.

Unpacking and Installing the Sample Shuttle

The sample shuttle accessory can be simply installed into the sample compartment of the spectrometer.
1. Unpack the sample shuttle from its box.
2. Rest the back of the sample shuttle accessory on the ledge in the sample area and slide it into position.
Push it firmly home to ensure that the multiway connector on the rear of the sample shuttle accessory mates properly with the spectrometer connector.
When the sample shuttle locks into place it aligns itse lf, which means that the slide holder may move. Be careful that your fingers, tie, etc are not trapped by the movement.
WARNING
Figure 14 Installing the sample shuttle accessory
NOTE: To remove the sample shuttle accessory, release the accessory by pulling the blue
handle under its baseplate, and then slide the accessory towards you and out of the sample compartment.
42 . OilExpress System User's Guide

Positioni ng the OilExpress System

The Autosampler is a heavy instrument with a basic weight of approximately 30 kg.
WARNING
1. Position the Liquid Autosampler on the bench as shown below (Figure 15). Leave a gap of 0.5–1 cm between the Autosampler and the spectrometer. Ensure that
you have space for the PC either to the right side of the spectrometer, or to the left side of the Autosampler.
Two people are required for safe handling. Consult the local codes of practice issued by safety advisors before
attempting to lift it. Take care not to injure yourself or others, or to drop the instruments.
Figure 15 The OilExpress System positioned for use
2. Align the Autosampler and the spectrometer so that the rear of the injection wash port and the sample shuttle are parallel.
The spectrometer protrudes approx imately 5 cm in front of the Autosampler when correctly alig ned.
Make sure that there is access to power and a waste container.
3. Place the PC in a suitable location.
4. Remove the Velcro straps and plastic cable ties around the Z-rack.
5. Remove any additional packaging and support from the Autosampler and make sure that the arm is able to move freely on all axes.
Unpacking and Installation . 43

Installing the Liquid Autosampler

Electrical Connections

Power cord
1. Make sure that the Autosampler is switched off.
2. Fit the molded socket of the mains cable into the plug on the side panel of the Autosampler (Figure 16).
3. Plug the power cord into the appropriate mains electrical supply.
Figure 16 The side panel of the autosampler
Cable to PC
1. Attach the male connector into the DB25 receptacle on the left side of the Autosampler (Figure 16), and tighten the screws.
2. Attach the female connector of the cable to the DB9 SERIAL COMM 1 port on the back of the PC, and tighten the screws.
If the computer has a DB25 connector, an adapter is required.
44 . OilExpress System User's Guide

Installing the Fluid Line Support Arm

Insert the fluid line support arm into the liquid connector bl ock on the right of the autosampler.
Figure 17 Fluid line support arm
NOTE: The Arm Shield is not installed until the remainder of the installation, including
software installation and calibration has been completed.
Unpacking and Installation . 45
Rack installs here
Rack installs here

Installing Racks (if ordered)

1. Remove and unwrap rack.
2. Place the sample rack on the pins as shown in Figure 18.
Figure 18 Installing the sample rack
Figure 19 Sample rack
46 . OilExpress System User's Guide

Autosample r Tubi ng Co nnec tion s

Fluid Tubing System

1. Remove the reagent tubing (L1200556) from the accessory bag. Route the ends with the fitting throug h the black grommet hole on the right panel of the autosampler. Install fitting on the right valve port of each of the syringe pum ps. Tighten the fitting fing er tight and then tighten an additional quarter turn with a wrench.
2. Remove the inte rconnect tubing (L1200585) from the accessory bag. Route one end of the tubing through t he black grommet hole on the right panel. Install the fitting on one end to the left valve port of the right syringe pump. Connect the fitting on the other end to the top of the 2-port flowcell. Tighten fittings as above.
3. Remove the needle tubing (L1200587) from the accessory bag. Install the end with the fitting into the left valve port of the left syringe pump. Insert the other end through the fluid line support arm (Figure 17), and then through the top of the Z-rack until it extends approximately one inch out from the lower end (Figure 21).
4. Remove the interconnect tubing (L 1200557) from the accessory bag. Install the fitting on one end to the bottom of the injection\wash port (Figure 6) and the other to the bottom of the 2-port flowcell. Tighte n the fitting finger-tight and then tighten an additional quarte r turn with a wrench.
From cell flush syringe
From transfer/wash port
Figure 20 Connecting tubing to the 2-port flowcell
Unpacking and Installation . 47
Cell
Needle
Solvent
Solvent
Z-rack
Figure 21 Tubing connections from the pumps to the 2-port flowcell
5. Remove the needle from the accessory bag. Loosen the slo tted screw in the insulation block, and then insert the shorter end of the needle through the sleeve in the insulation block as far as the collar. Retighten the screw.
Electrostatic discharge to the sampling probe may cause damage to the
CAUTION
ALIDUM liquid level detector (see
Cable Assembly
on page 82).
Inspecting the Insulation Block/ALIDUM
Z-rack
Probe Tubing
Allen screw
Slotted screw
Needle
Insulation block
Insulation block sleeve
Needle
Figure 22 Connecting the needle tubing to needle
48 . OilExpress System User's Guide
6. Push the needle tubing over the shorter end of the needle (to a minimum of 25 mm, 1 inch).
7. Loosen the Allen screw on the insulation block. Install the insulation block on the Z-rack with the insulation blo ck sleeve facing down and the insulatio n block (black coax) cable on the left side. Ensure that the insulation block cable is not twisted.
8. Push the insulatio n block onto the Z-rac k until firmly seated.
9. With an Allen wrench, tighten the insulation block set screw until it makes contact with the flat portion of the Z-rack. Do not overtighten the setscrew.
Connecting the tubing to the solvent bottle
1. Thread the tubing from top to bottom through the cap of the wash bottle as shown in Figure 23.
Solvent bottle lid
Solvent bottle
Figure 23 Connecting the sol v en t tu bing
2. Fill the solvent bottle with solvent.
3. Screw the cap on to the solvent bottle securely.
4. Position the bottle safely, behind the instrumentation at a location where it will not be accidentally knoc ke d.
Unpacking and Installation . 49

Installing the Flowcell in the Sample Shuttle

DO NOT use the OilExpress System when the sample compartment lid is fitted. Flammable vapor may seep through instrument seals and gasket s and accumulate under the lid, where it could cause a fire or explosion
WARNING
1. Unpack the flowcell.
2. Slide the flowcell into the sample shuttle (Figure 24).
3. Make sure that the tubing cannot obstruct the optical b eampath.
hazard.
Figure 24 Sample shuttle accessory and flowcell fitted

Waste Container

We recommend that you use a waste container that is larger than the solvent bottle. Then, if you empty the waste container whenever you refill the solvent bottle, the waste container will not overflow.
The waste drains into the container by gravity, so the cont ainer should stand on a surface that is at least 90 cm below the bench on which the OilExpress System is located.
Fitting waste tubing
1. Attach the waste tubing to the hose connector at the rear of the transfer/wash port.
2. Insert the waste tubing into the hole on the rear right-hand side of the base plate of the Autosampler.
3. Adjust the routing o f the tubing so that it is not kinked.
4. Place the other end of the waste tubing in the waste container. Ensure that the end will not be immersed in the waste.
Make sure that the ends of the tubing do not become immersed in the
CAUTION
waste or the operation of the analyzer will be impaired.
50 . OilExpress System User's Guide

Installing the Syringe

The syringes cannot be installed until the OilExpress software has been installed.
1. Switch on the Autosampler.
2. Start the AutoSampler Utilities. Select Start, All Programs , Pe rkinElme r Application s, OilExpress and then click
AutoSampler Utilities.
3. Log in as an Administrator. The User Name is Administrator and the Password is Administrator.
4. Select Pump Diagnostics from the File menu.
5. Click Change Syringe.
6. Assemble the syringe (L1200552) by placing the plunger in syringe barrel and putting the syringe clamp over the Luer lock fitting.
7. Unscrew the pin from the plunger sha ft.
8. Mount the syringe plunger on the plung er shaft. Line up the barrel with the Luer lock fitting. Screw the barrel onto the valve fitting while pushing upward slightly.
When mounting the syringe on the val ve, you must assist the Luer lock
CAUTION
9. Replace the plunger pin.
10. Rotate the syringe clamp clockwise then tighten the clamp screw.
11. Replace the plunger screw at the base of the syringe plunger.
threads by pushi ng upward. Otherwise they may become stripped.
Unpacking and Installation . 51

PC Software and Hardware Requirements

This section details the minimum hardware and software requirements for the PC. To ensure successful installation of the software, please check these requirements before starting the installation.
Only if all of these criteria are met should you start to install the OilExpress software (see
Installing the OilExpress Software

Software Requirements

The minimum software requirements for the OilExpress System are:
Windows 7 (32-bit), Windows Vista or Windows XP with Servi ce Pack 2 must be installed, and you must be logged on at Administra tor level.
Internet Explorer 5.01 or later must be installed.
Spectrum software must be installed, as described in the documentatio n supplied with
your spectrometer on the
on page 54).
IR & Raman Manuals CD
(L1050002).

Hardware Requirements

The PC you install the software on must meet the following minimum specification:
®
Intel
At least 256 MB of Random Access Memory (RAM).
The capability of displaying at least High Color (16 bit) at 1024 × 768 SVGA.
10 GB Hard disk with at least 1 GB free space as an NTFS drive.
CD-ROM drive.
Ethernet network connection.
A keyboard and PS/2
A serial port, for autosampler connection.
Pentium 4 processor with 500 MHz or great er clock speed, or equivalent.
®
-style mouse.
52 . OilExpress System User's Guide

Upgrading from a Previous Version of OilExpress

Upgrading from OilExpress v3.x

If you are upgrading OilExpress from version 3.x, simply uninstall the previous version of OilExpress before installing the latest version.
All your OilExpress data is automatically retained.

Upgrading from earlier versions

If you are upgrading OilExpress from a version e arlier than 3.0, and you want to retain the databases from the previous installation, you must copy some files from the old installation into the new instal lation.
NOTE: The security da tabase cannot be retained. After completing the upgrade you must
recreate all users.
1. Copy the files specified in the upgrade notes that follow this procedure.
2. Uninstall OilExpress.
3. If required, uninstall Spectrum, and then install the latest version. Follow any applicable guidance provided in the
(L1050095), which is provided on t he
4. Log in to Spectrum and install the spectrometer.
5. Install the latest version of OilExpress (see page 54).
6. Restore any OilExpress database files you copied at step 1.
IR & Raman Manuals CD
Spectrum Administrator's Guide
(L1050002).
Installing the OilExpress Software
Upgrading from OilExpress v2.2
Before uninstalling OilExpress v2.2, make copies of these database files:
C:\Documents and Settings\All Users\Application Data \PerkinElmer\OilExpress\Local Repository\*
C:\Documents and Settings\All Users\Application Data \PerkinElmer\OilExpress\Fuel Calibrations\*
C:\Documents and settings\All users\Application Data\PerkinElmer\OilExpress\Local Resultstore\resultstore.mdb
on
(where * means copy the entire contents of the folder).
Unpacking and Installation . 53
After uninstalling OilExpress v2.2, copy the contents of the Local Repository and Fuel Calibrations folders back to their original locations.
After installing the latest version of OilExpress (see page 54), migrate the resultstore.mdb file:
1. Start the Database Tools utility. Select Start, All Programs , PerkinElmer, OilExpress and then click Database Tools.
2. Select Result Stores and then click Migrate Database.
3. Open the resultstore.mdb file you copied from the previous installation.
4. Enter a short name and a file name for the migrated database.
5. Select the new database from the list and click Set Active Database.
Installing the OilExpress Software
on
Upgrading from OilExpress v1.0 or v2.0
NOTE: Make copies of these database files before uninstalling OilExpress.
Copy the batchstore.mdb, reference spectra.mdb and repository.mdb files into folder C:\Documents and Settings\ All Users\Application Data\PerkinElmer\OilExpress \Local Repository.
Copy the resultstore.mdb file into folde r C:\Documents and Settings\ All Users\Application Data\PerkinElmer\OilExpress\Local Resultstore.
Copy the fuel calibration files into folder C:\Documents and Sett ings\ All Users\Application Data\PerkinElmer\OilExpress\Fuel Calibrations.
For OilExpress version 1.0, these files can be found in:
C:\Program Files\PerkinElmer\OilExpress\Local Repository C:\Program Files\PerkinElmer\OilExpress\FuelCalibrations C:\Program Files\PerkinElmer\OilExpress\Local Resultstore.
After installing the latest version of OilExpress (see page 54), migrate the resultstore.mdb file:
1. Start the Database Tools utility. Select Start, All Programs , PerkinElmer, OilExpress and then click Database Tools.
2. Select Result Stores and then click Migrate Database.
3. Open the resultstore.mdb file you copied from the previous installation.
4. Enter a short name and a file name for the migrated database.
5. Select the new database from the list and click Set Active Database.
Installing the OilExpress Software
on
54 . OilExpress System User's Guide

Installing the OilExpre ss Software

The Liquid AutoSampler software CD contains an Installation Wizard to help you install OilExpress on your PC.
NOTE: The Spectrum software must be installed
documentation s upplied with your spectrometer, on the (L1050002) for details.
Once the Spectrum software has been installed, det ails of how to administer it can be found in the the Manuals CD.
To install OilExpress:
1. Place your Liquid AutoSampler software CD into your CD-ROM drive.
2. If the CD does not auto-run, select Run from the Start menu. The Run dialog is displayed.
3. Enter d:\Setup.exe and clic k OK. Replace d:\ with the drive letter for your CD-ROM drive. After preparing to install, the InstallShield Wizard Welcome dialog is displayed.
Spectrum Administrator's Guide
before
OilExpress. Consult the
IR & Raman Manuals CD
(L1050095), which is also provided on
Unpacking and Installation . 55
4. Click Next. The License Agreement dialog is displayed.
5. Read the agreement, select I accept the terms in the license agreement, and then click Next.
The Product Selection dialog is displayed.
56 . OilExpress System User's Guide
6. Select the OilExpress option and enter your license key; then click Next. The Instrument Selection dialog is displayed.
7. Select PerkinElmer FT-IR with Autosampler and click Next. Your software can now be installed.
Unpacking and Installation . 57
8. Click Install. The installation will start.
When the installation is complete, the following dialog will be displayed:
9. Click Finish.
58 . OilExpress System User's Guide

Default Users fo r OilExpress

The default passwords for the default user groups in OilExpress are the same as the name of the group, as follows:
Login Name Password
Administrator administrator
Advanced Analyst advanced analyst Analyst analyst
Supervisor supervisor
You should immediately change these passwords to stop any unauthorized access to the software.
For further information see
Setup Users and Groups
in the on-screen Help.
Unpacking and Installation . 59

Starting up OilExpres s and Configu ring your Instrument

The first time you start the OilExpress system, if you have not already installed instruments using the Spectrum software, the Instrument Configuration dialog is displayed.
Use the Add/Edit button to add one or more instr uments. See the on-screen help for details.
NOTE: Subsequently, if you want other instruments to be made available, you can use the
menu option System Settings > Instruments Settings > Configure Instrument.
Click OK.
NOTE: If you make more than one instrument available for use wi th OilExpress, or a single
instrument capable of operating in both nea r- and mid-infrared, you are prompted to select which instrument and mode is to be used each time you start OilExpress.
OilExpress should only be used with a spectromet er operating in mid-infrared.
60 . OilExpress System User's Guide

Calibration

Before using the OilExpress System two calibration tasks must be performed:
Calibrate the Autosampler
Calibrate the flowcell

Calibrating the Autosampler

The Autosampler must be calibrated to correct for any differences in the X-Y-Z zero alignment and any differences in the injection/wash port positioning.
1. From the Start menu select: Programs > PerkinElmer > Applications > OilExpress > Autosampler Utilities.
The login screen (PerkinElmer or Windows, as appropriate) is displayed.
2. Enter a User name and Password. The Autosampler Utilities dialog is displayed.
3. From the File menu select WorkTable Editor. The following dialog is displayed.
Unpacking and Installation . 61
4. Right-mouse click on the Waste Port, and from the drop-down menu displayed select Properties.
The following dialog is displayed.
5. Select the Calibrate tab.
62 . OilExpress System User's Guide
6. Click Position Evaluation. The following dialog is displayed.
7. Separate the three dialogs on your desktop as shown below.
8. On the Moving Arm Dialog click Open Com. The Autosampler initializes.
Unpacking and Installation . 63
9. In the Anchor Point section of the User-defined Rack Properties dialog, click Go.
10. Click the down button close as possible to the top of the injection/wash port without actually touching it.
on the Moving Arm Dialog to lower the needle so that it is as
Do not constantly press as this will cause the needle to accelerate.
CAUTION
Take care not to ram the needle into the injection/wash port as this may cause the needle to bend.
11. Use the left and right arrow buttons anchor point.
to move the needle so that it is at the
64 . OilExpress System User's Guide
Needle position
Needle position
v1 v2 v3
The anchor point depends on the version of your wash st ation. Examine the injection/wash port fitted, then use the appropriate needle position anchor point, as shown in Figure 25.
anchor point
anchor point
10 mm
Needle position anchor point
Figure 25 Injection wash port anchor positions
12. Once the needle has been aligned with the anchor point, click the Anchor Point Insert button
.
The new numbers appear in the X= and Y= fields.
13. Record the Anchor Point X and Y values in your labora tory notebook as the numbers can be used in the calibration of other worktables.
14. Click OK.
Unpacking and Installation . 65
15. Right-mouse click within the worktable area of the Autosampler Utilities dialog and from the menu select Save Worktable.
NOTE: The worktable changes are not automatically saved. If you do not select Save
Worktable, the changes will be lost.
NOTE: The procedure described above must be repeated for all worktables used by the
OilExpress software as anchor points are not app lied universally. Follow steps 16 to 20 to calibrate all additional worktables.
16. Right-mouse click within the worktable area of the Autosampler Utilities dialog and from the menu select Load Existing Worktable.
17. Select the required worktable, PerkinElmer BOTTLES or PerkinElmer VIALS.
18. Repeat steps 4 to 7.
19. Enter the values you previously record ed in your laboratory notebook into the X= and Y= fields of the User-defined Rack Properties dialog.
66 . OilExpress System User's Guide
20. Repeat steps 14 and 15.
21. If the syringes have not yet been installed, comple te the hardware installation on page 50.

Calibrating the Pathlength of the Flowcell

The OilExpress software calculates the cell pathlength by using a Partial Least Squares multi­component calibration on heptane. A default calibrat ion file is provided and stored as c:\quant\pls1\methods\cellpath.md, where c:\ is the drive to whic h the OilExpress software has been installed. Heptane must be used as the solvent within the OilExpress System for accurate cell calibration.
1. Log in to OilExpress.
2. Select System Settings > Instrument Settings > System Checks Setup. The Check Scheduling Properties dialog is displayed.
3. Select Pathlength and click OK.
4. Click the System Checks thumbnail OilExpress dialog.
The System Checks dialog is displayed.
on the left-hand panel in the main
Unpacking and Installation . 67
5. Click Start. The Autosampler pumps are primed when the system is initialized to ensure that the
cell is filled with heptane. (To prime the pumps manually, refer to in the online help). The check will be performed.
NOTE: If additional system checks are selected, these will be performed in addition to the
pathlength check in the order that they appear on the screen.
On completion of the pathlength check, Passed or Failed will appear in the dialog. If the pathlength falls between the limits expected for the ZnSe cell, then the pathlength will be automatically stored for future use in OilExpress analyses.
If the pathlength check fails, checks the following and then repeat the steps described above:
The correct solvent was used
The solvent was delivered to the cell correctly and there were no leaks
The beam of the spectrometer is not obstructed
Autosampler Control
The cell is in the correct position within the sample shuttle
The limits within the pathlength setup are correct

Installing the Arm Shield

1. Remove the clear rectangular arm shield from the accessory box.
2. Remove the three screws (Figure 26).
3. Position the arm shield under the arm and adjust the shie ld so that the holes in the shield are aligned with the screw holes on the side of the arm cover.
68 . OilExpress System User's Guide
4. Insert the socket head screws through t he holes and tighte n by turning the screw clockwise.
CAUTION
Do not over tighten the screws as this may crack the plastic shield.
screw
screw
Figure 26 Installing the arm shield
screw
Background to Used Oil
Analysis
70 . OilExpress System User's Guide

Background to Used Oil Analysis

Infrared Absorption

The chemical bonds within organic molecules are in a state of continual vibration, with bonds stretching and contracting as well as be nding relative to each other. When an infrared beam falls on a molecule, waves of specific frequencies are absorbed from the beam by the molecule. The actual frequencies absorbed depend on the types of bonds present in the molecule structure. Chemical bonds within a molecule are said to exhibit “characteristic infrared absorptions”.
A record of the frequencies at which absorption takes place for an organic compound is called its infrared spectrum. An infrared spectrum of a compound will reveal information about molecular structure as the existence of specific groups of atoms may be confirmed from the presence of their characteristic absorpt ions.

Why Analyze Used Oils?

Infrared analysis of used oils provides critical information regarding the state of the lubr icant itself and the engine from which it comes.
Used oil analysis provides a great deal of information in two areas:
1. The physical condition of the lubricating oil, such as its viscosity, neutralization number, or level of contamination.
2. The condition of the component or system from which the sample was taken.
The presence of chemical degradation products due to oxidation, nitration, sulfate formation, ester breakdown, and anti-wear additive depletion, and contaminants such as soot, water, ethylene glycol and unburned fuel is used to measure the degradation of the oil.
Analyzing used oil samples and comparing the resulting spectrum (showing characteristic absorptions of the degradation products, oil and contaminants) with an unused o il spectrum contained in a “new oil” library accurately monitors the chemical degradation of engine lubricants.
Background to Used Oil Analysis . 71
4000.0 3600 3200 2800 2400 2000 1800 1600 1400 1200 1000 800 500.0
0.0
10
20
30
40
50
61.2
cm-1
%T
4000.0 3600 3200 2800 2400 2000 1800 1600 1400 1200 1000 800 500.0
0.0
10
20
30
40
50
61.2
cm-1
%T

Methods of Analysis

The OilExpress System is an automated system that uses infrared (IR) analysis to monitor oil quality parameters: chemical degradat ion products and common contaminants of lubricating oils. The OilExpress System provides a range of Analysis Methods for measuring these parameters, including standard methods for Mineral Oils, Synthetic Oils, and the Joint Oil Analysis Program (JOAP). Within OilExpress, methods can be added or deleted, and the properties of existing methods customized.
Used oil samples are complex mixtures of a large number of different chemicals and include compounds derived from the original formulation of base oil and its additives, oil degradation products and oil contaminants. As a result of this a us ed oil spectrum is complex; because of this complexity, the spectrum of a used oil alone is of limited value and must be compared against a spectrum of the unused oil to be of any significant analytical value.
Figure 27 Spectrum of a used oil sample
Figure 28 Spectrum of an unused oil sample
72 . OilExpress System User's Guide
4000.0 3600 3200 2800 2400 2000 1800 1600 1400 1200 1000 800 500.0
-0.100
-0.08
-0.06
-0.04
-0.02
0.00
0.02
0.04
0.06
0.08
0.100
cm-1
A
Hydroxy
Oxidation
Nitration
Oxidation/ Sulfate
Antiwear Depletion
Fuel
Sulfate
Figure 27 and Figure 28 illustrate the comparison of a used oil sample and a sample of the same oil unused. From these spectra it would be fair to assume that minimal degradation has occurred.
This situation changes dramatically when a differential spectrum is viewed. Figure 29 shows the baseline corrected difference spectrum of the same two samples,
obtained by subtracting one spectrum from the other and then performing a baseline correction.
Figure 29 Baseline corrected difference spectrum
This data can now be manipulated mathematically to provide information about the differences between the new and used oil, and may be considered to be a spectrum of the degradation products that exist in the used oil.
In both cases, the trends observed in the results from a series of samples taken at regular time intervals from the same engine have a much higher diagnostic value than the results from a lone sample.

Degradation Products

Oxidation
Oil exposed to oxygen from the air at elevated temperature will oxidize to a variety of compounds, the majority of which are carbonyl compounds, including carboxylic acids. Carboxylic acids contribute to the acidity o f the oil, depleting the basic additives present in the oil and contributing to corrosio n. Oxidation can also increase the viscosity of the oil. The degree of oxidation is a good indicator of oil degradation. A rapid increase in oxidation may indicate an engine overheating or a depletion of the anti-oxidant additive due to an over-extended oil change period.
Nitration
Nitrogen oxides, produced fr om the oxidation of atmospheric nitrogen during the combustion process, react with the oil. Nitration increases the viscosity of the oil and is the major cause of the build-up of va rnish or lacquer.
Background to Used Oil Analysis . 73
A high nitration value, also known as NOX or nitro-oxidation, indicates incorrect fu el/air ratio, incorrect spark timing, excessive loads, low operating temperature or piston ring blow-by.
Sulfate
Sulfur oxides are produced by the combustion of sulfur compounds present in the fuel. These oxides react with water, also produced by the combustion process, to form sulfuric acid. The sulfuric acid is neutralized by the oil’s basic add itives, forming inorganic sulfates. A rapid increase in the sulfate value may indicate the use of a high sulfur content fuel, poor combustion, over-cooling or the rapid depletion of anti-wear additive.
Ester breakdown
Synthetic lubrication oils usually contain a high prop ortion of synthetic polyol esters. These esters are susceptible to breakdown in the presence of water and acids (hydrolysis). Ester breakdown contributes to the acidity of the oil and can result in the formation of crystals of the base polyol, leading to clogging of filters. Ester breakdown may indicate the presence of water from condensation caused by low operation temperatures or from a coolant leak.

Common Contaminants

Soot
Soot particles result from the incomplete combustion of fuel and, since they are too small to be removed by the filter, remain suspended in the oil. Soot builds up continuously until it reaches an unacceptable level, that level depends on the type of engine and lubricant. Diesel oils tolerate higher soot levels than gasoline oils. The rate of soot build up depends on engine design, type of fuel and operating conditions. A high soot value may indicate poor combustion due to incorrect fuel/air ratio, a clogged air filter or an over-extended oil change period.
Water and glycol
The presence of water and glycol indicate a leak from the cooling system. It is essential to detect such a problem early to avoid a serious failure. The presence of water alone does not necessarily indicate a coolant problem, but may result from condensation due to a low operating temperature.
Unburned fuel
The fuel measurement depends not only on the level of contamination, but also on the aromatic hydrocarbon content of the fuel. The amount of aromatic content in fuels vary in location and season, therefore a calibration must be done on the fuel being used. The presence of fuel indicates poor combustio n due to incorrect fuel/air ratio, clogged air filter or an ignition problem. It may also indicate an over-extended oil change period.

Analysis Methods

The OilExpress System ships with a set of default analysis methods that will appear in the drop-down menu. The methods vary in the analytes they determine, the peaks used and the warning limits on the results. For furt her information about the default analysis methods see the on-screen Help.
74 . OilExpress System User's Guide

Sample Preparation

Filtering the Oil

It is important that all oil samples are filtered to prevent large particles blocking the flowcell, however, care should be taken not to remove any soot particles. A steel mesh filter (90 microns) is recommended.

Sample Presentation Requirements

Carefully record details of the oil contained in each bottle. Place the sample bottles in the sample rack, ensuring that the lids have been removed.
The OilExpress System bottle rack accepts standard 4 oz oil sampling bottles. If a smaller diameter bottle is used ensure that it is centered in the sample holder.
NOTE: For information on how to use the OilExpress System, see the help system available
from the Help menu in the OilExpress software, which includes a tutoria l and full information on using the software .
Routine Mai nten ance and
Troubleshooting
76 . OilExpress System User's Guide

Maintenance Schedule

Maintenance must be performed on a regular basis in or der to maintain the accuracy and precision of the OilExpress System. The schedule below outlines the proper intervals to check or replace components of the instrument.
NOTE: For information on maintaining your spectrometer, refer to the documentation
supplied with the i nstrument.
Item Daily Weekly 6 Months
ALIDUM Cable - X (p.79) R (p.82)
Sampling Probe C/X (p.77) C/X (p.77) R (p.80)
Z-Rack - C (p.77) ­Splash Guard - C (p.78) -
Transfer/Wash Port - C (p.77) -
Syringe C (p.77) C (p.77) S (p.80)
Syringe Seal C (p.77) - R (p.80) Sample Rack C (p.77)
Sampling Probe C (p.77) - R (p.79)
X/Y/Z Mechanism - - X (p.78)
X/Y/Z Lubrication - - S (p.84) Tubing - - S (p.85)
Key:
C = Clean this item - = No action necessary R = Replace component S = Special case X = Perform service
Routine Maintenance and Troubleshooting . 77

Autosampler Maintenance Procedures

Flushing the Syringe Pump

1. Do not allow the pump to run dry for more than a few cycles.
2. Inspect the pump for leaks, and correct any problems immediately.
3. Wipe up all spills on and around the pump immediately.

Cleaning the Syringe Seal

1. Clean the seal with heptane.
2. Check for cracks in the Teflon coating.

Cleaning the Sampling Probe

Electrostatic discharge to the sampling probe may damage the ALIDUM.
CAUTION
1. Turn off power to the Autosampler.
2. Clean the needle by wiping gently with a lint-free cloth dampened with heptane.

Cleaning the Sample Rack

We recommend that you clean the sample rack daily with detergent or degreasant to remove dust and oil.

Cleaning the Injection and Wash Port

Residues from samples may be deposited in the injection/wash-port. We recommend that you remove the lid occasionally and rinse it with a suitable detergent if necessary.

Cleaning the Z-Rack

The Z-Rack must be cleaned weekly.
NOTE: Do not use alcohol or solvents when cleaning the Z-Rack.
1. Turn off the power to the Autosampler.
2. With a lint-free cloth, wipe the Z-Rack thoroughly. If necessary, use a toothbrush to remove dust or dirt from the teeth of the
Z-Rack.
78 . OilExpress System User's Guide

Cleaning the Splash Guard/Safety Shields

The Splash Guard/Safety Shields must be cleaned weekly.
1. Turn off the power to the Autosampler.
2. With a lint-free cloth, wipe the splashguards thoroughly.

Cleaning the Lead Screw and Lead Screw Encoder

If the pump is used in a dusty environment, the lead screw encoder should be cleaned periodically with air. The lead screw should be lubricated:
If the pump is making a "screeching" noise
If the syringe is stalling frequently
NOTE: Only PerkinElmer lubricant (part number L1200559) should be used for lubricating
the lead screw. Lubrication is usually only needed after 1 million syringe strokes (or every 6 months). If the pump appears to need lubrication more frequently, suspect another problem.

Cleaning X-Y-Z Mechanism

The X-Y-Z mechanism should be cleaned every six months. However, if the instrument is operated in a dusty or humid environment, it must be cleaned every three months. Follow the instructions to prevent damage to the instrument.
Use only isopropyl alcohol and a lint-free cloth to clean the Autosampler.
CAUTION
1. Turn off the power to the Autosampler.
2. Clean the Z-Rack.
3. Clean the needle.
4. Wipe the arm using a lint-free cloth dampened with isopropyl alcohol to remove any residual dust.
5. Wipe the inside of the flex cable channel using a lint-free cloth dampened with isopropyl alcohol.
Other cleaning agents may affect the performance of the instrument. Never clean the X-Y-Z axis guide rails or Z-Rack with al c ohol or solvent. Serious damage to the instrument may occur.
CAUTION
Do not wipe the X-axis guide rails. The guide rails are lubricated with a grease that does not require removal unless found to be "extremely" dirty.
Routine Maintenance and Troubleshooting . 79
6. Wipe the inside of the X-frame with a lint-free cloth dampened with isopropyl alcohol, ensuring that the lubricant on the X-axis guide rails is not removed.
7. Wipe the square shaft pinion located underneath the arm using a lint-free cloth dampened with isopropyl alcohol.
When cleaning the square shaft pini on, ensure no alcohol enters the
CAUTION
Z-bearing or is wiped on the Y-axis guide rails.

Inspecting the Insulation Block/ALIDUM Cable Assembly

The insulation block/cable assembly must be inspected weekly and replaced every six months. Examine t he cable where it connects to the insulation block. If there is any residue, liquid, or if the cable is cracked at this connection, replace the assembly.
NOTE: Instructions on how to replace the insulation block cable can be found in
the Insulation Block/ALIDUM Cable Assembly
on page 82.
Replacing

Checking the Fluid Path (Tubing)

Check that all the tubing is clean and free of crimps. Refer to page 85 for procedures to change tubing.
Replacing the Tubing
on

Replacing the Sample Probe/Needle

1. Turn the power off to the Autosampler.
2. Loosen the sampling probe set screw using a flat screwdriver.
3. Loosen the insulation block set screw using an Allen wrench.
4. Gently pull downward on the insulation block until the system tubing is approximately 1 inch below the Z-Rack.
Electrostatic discharge to the needle may damage the
CAUTION
ALIDUM detector.
5. While holding the needle, remove the system tubing.
6. Remove the needle from the bottom of the insulation block.
7. To reinstall the needle, refer to
NOTE: If the needle has been bent, it should be replaced.
Sampling Probe/Needle
on page 15.
80 . OilExpress System User's Guide
.

Replacing the Syringe Seal

The syringe seal will need to be replaced periodicall y. How often will depend o n the duty cycle of the pump, the type of fluids being run through the system, the size of the syringe and how well the instrument is maintained. If the syringe seal becomes worn and is not replaced, the following problems may occur:
Poor precision and accuracy.
Variable or moving air gaps.
Fluid leaks from the bottom of the syringe.
The tip of the plunger breaks through the seal and scra tches the inside of the syringe
barrel. If this happens, the entire syringe will need to be replaced.

Replacing a Syringe

1. Refer to
2. Remove the plunger pin at the base of the syringe plunger.
3. Release the syringe clamp and carefully unscrew the syringe barrel from the Luer lock fitting (about 1½ turns) while pulling downward slightly. Then slide the syringe plunger from the plunger shaft.
4. To install a syringe, place the plunger over the plung er shaft. Line up the barr el with the Luer lock fitting. Screw the barrel onto the fitting while pushing up ward slightly.
Installing the Syringe
on page 50.
You must assist the Luer lock threads by pushing upward, otherwise they
CAUTION
5. Replace the plunger screw at the base of the syringe plunger.
may become stripped

Replacing the Reagent Syringe Seal

1. Refer to
2. Remove the plunger pin at the base of the syringe plunger.
Installing the Syringe
on page 50.
3. Release the syringe clamp and carefully unscrew the syringe barrel from the Luer lock fitting (about 1½ turns) while pulling downward slightly. Then slide the syringe plunger from the plunger shaft.
4. Remove the syringe plu nger.
Routine Maintenance and Troubleshooting . 81
O-ring
5. Remove the seal from the plunger tip using a pair of pliers and gripping the seal approximately one third o f the way down (Figure 30).
O-ring
Narrow lip Sharp raised edge
Plunger
Figure 30 Plunger and O-ring configuration
The 10.0 ml syringe has an O-ring beneath the seal. Be careful not to
CAUTION
damage the O-ring. If it is necessary to replace the O-ring, simply slip the new one over the narrow lip on the plunger. It may be necessary to use needle nose pliers to assist in re moving the O-ring from the plunger.
6. Wet the plunger tip and the O-ring with distilled or de-ionized water.
7. Place the seal on a flat surface with the open end facing up. Press the plunger tip firmly into the hole until it snaps into position.
8. Lay the plunger on a flat tabletop. Position the plunger so that the seal (from the O-ring up) hangs over the edge (Figure 31). Slowly roll the plunger along the table edge, pressing firmly on the portion of the seal below the O-ring. Rotate the plunger three complete turns. This is necessary for the sharp, raised edge of the plunger to bite into the seal for a secure fit.
Table
Sharp raised edge
Figure 31 Seal installation
9. Wet the seal and replace the plunger in the barrel.
10. Replace the syringe.
82 . OilExpress System User's Guide
Front view
Flat, stem

Replacing the 3-port Valve

1. Remove the syringe as described in
2. Power down the Analyzer.
3. Remove the aspirate tubing, reagent tubing and syringe.
4. Remove the valve by loosening the two screws on the valve.
5. Install the new valve by aligning the "D" coupler on the valve stem with the "D" hole in the encoder.
6. Rotate the valve body so that the Luer fitting is oriented toward the bottom of the pump.
7. Gently push the valve in place ensuring the locating pins on the frame side of the valve body fit in the holes on the XL 3000 frame and replace the screws.
8. Replace the syringe.
9. Power up the Autosampler.
Replacing a Syringe
on page 80.
Right port
Syringe port
Figure 32 CAVRO 3-port valve

Replacing the Insulation Block/ALIDUM Cable Assembly

1. Turn off the power to the Autosampler.
2. Remove two (2) screws on the back of the top cover.
3. Remove the top cover of the instrument by lifting both sides.
4. Remove the sampling probe.
5. Remove the plastic cable strap underneath the arm, which is holding the insulation block cable.
NOTE: Use a small wire cutter to remove the plastic cable binder. A new cable binder is
included with each spare replacement insulation block.
Routine Maintenance and Troubleshooting . 83
Gol d Coax Connect or
6. Push the insulatio n block cable through the back end of the arm until the gold connector is visible.
Figure 33 Disconnecting the insulation block cable
7. Disconnect the in sulation block cab le from the connector.
8. Carefully pull the insulation block cable back out through the back of the arm.
9. Remove the new insulation block cable and the plastic cable strap from the packaging.
10. Feed the gold connector of the insulation block cable through the back of the arm.
11. Attach the gold connector to the matching connector on the ALIDUM cable. Be sure that the connectors are firmly seated.
12. Adjust the ALIDUM cable until all of the slack is removed from behind the arm.
Make sure the insulation block cable is not twisted or kinked.
CAUTION
13. Fasten the insulat ion block cable to the arm using the new plastic cable binder.
14. Reinstall the sampling probe.
15. Replace the top cover and retaining screws.
84 . OilExpress System User's Guide

Lubrication

The Autosampler does not require lubrication under normal operating conditions. If the lubricant applied at the factory should be wiped away or if the X/Y axis guide rails become dirty, cleaning and lubrication is necessary. Follow the instructions below when applying lubricants.
1. Turn off the power to the Autosampler.
2. Move the arm to the far right side of the X-frame.
3. Clean the X-axis guide rails using a lint-free cloth and isopropyl al cohol.
4. While moving the arm to the left side of the X-frame, clean the three roller bearings mounted to the base of the X-slide.
5. Wipe the section of the X-axis guide rail that was covered by the arm.
6. Using a fingertip, apply a thin film of lubricant to both X-axis guide rails.
7. If the Y-axis guide rails are dirty, follow steps 2 through 6.

Changing the Autosampler Fuses

Two 5A fuses are fitted to the Autosampler.
1. Switch off the Autosampler and disconnect the mains electrical supply.
2. Disconnect the mains electrical supply lead from the side of the Autosampler.
3. Insert a small screwdriver into the slot above the fuse drawer and lever the fuse drawer out.
4. Remove both fuses from the fuse drawer.
NOTE: If a fuse blows, it is recommended that both fuses be replaced.
5. Fit the replacement fuses.
6. Push the fuse drawer back into the Autosampler, and switch on the Autosampler.
For continued protection against risk of fire, only rep lace fuses which meet the following specified types and current rating.
WARNING
The following fuse must be used when operating the Autosampler:
Name Location Type
Main Fuse Power Entry Module T5.0 A/250 V × 2
Routine Maintenance and Troubleshooting . 85

Replacing the Tubing

It is important that all tubing is kept clean and free of crimps. Tubing that has become dirty, blocked or crimped can result in poor accuracy and precision, loss of air gap, or syringe stalls. Replace the tubing if necessary. Frequency of replacement will be dependent on duty cycle, reagents and maintenance.
Switch off the Autosampler before replacing any tubing.
WARNING

Replacing Reagent and Needle or Cell Tubing

Removing the old reagent tubing
1. Unscrew the connector from the right hand side of the va lve (reagent tubing).
2. Pull the tubing clear of the Autosampler.
Installing the new reagent tubing
1. Route the end of the tubing with the fitti ng through the black grommet hole on the right panel.
2. Install fitting on the right valve port of the syringe pump.
3. Finger-tighten the fitting.
Removing the old needle tubing
1. Unscrew the connector from the left hand side of the left hand syringe pump.
2. Remove the insulation block from the Z-drive.
3. Pull the tubing from the needle end.
4. Withdraw the tubing from the Z-drive and tubing support.
Installing the new needle tubing
1. Route the end of the tubing with the fitti ng through the black grommet hole on the right panel.
2. Install the end of the tubing with the fitting into the left valve port of the left syringe pump.
3. Route the other end of the tubing through the tubing support and down the Z-drive.
4. Push the end of the tubing over the end o f the needle and replace the insulating block onto the Z-drive.
86 . OilExpress System User's Guide

Replacing Waste Tubing

Removing the old waste tubing
Before removing the waste tubing, be sure the fluidic system/wash
CAUTION
Disconnect the waste tubing from the wash station and remove the tubing from the
Autosampler.
Installing the new waste tubing
1. Attach the waste tubing to the hose connector at the rear of the transfer/wash port.
2. Insert the waste tubing into the hole on the rear right-hand side of the base plate of the Autosampler.
3. Adjust the routing o f the tubing so that it is not kinked.
station is dry.
4. Place the other end of the waste tubing in the waste container. Ensure that the end will not be immersed in the waste.
Routine Maintenance and Troubleshooting . 87

Maintenance of the Flowcell

In normal use of the OilExpress System, the cleaning utilities provided by the software ensure that the flowcell is routinely kept clean and free of contaminants.
If, however, the flowcell does become contaminated, it may be carefully dismantled for deep cleaning using lens tissues and heptane, then re-assembled. Conta mination of the flowcell may be detected during calibration (see
Troubleshooting

Routine Cleaning of the Flowcell, Using Software Utilities

The procedure described below may provide an effective method for the routine cleaning of your flowcell without the need to disassemble it.
The frequency with which you need to carry out routine cleaning will depend upon the number and nature of the samples you work with. In particular, if you work with dirty or sooty samples, you may find that you need to clean your flowcell more frequently than will otherwise be the case.
on page 93).
1. With heptane in the solvent bottle, select Autosampler Control from the View menu and then click Wash.
2. Repeat the heptane wash 10 times.
3. Place a mixture of toluene with 10% isopropyl alcohol in a sample bottle, then select Autosampler Control from the View menu and use the Aspirate function.
4. Aspirate the solution into the cell, allow it to settle for a few seconds, and then wash it out with heptane, using the Wash function.
5. Repeat steps 3 and 4 a total of 10 times, then allow the solution to settle in the cell for 30 minutes before washing it out.
6. Repeat steps 3 and 4 a further 10 times.
7. Carry out 15 heptane washes.
8. Disconnect the tubing from the flowcell and dry it, either with air or by attaching i t to a vacuum pump.
9. Finally, carry out a Throughput check us ing System Settings > Instrument Settings > System Checks Setup.
88 . OilExpress System User's Guide

Deep Cleaning the Flowcell, Involving Disassembly

Wear gloves when dismantling the cell as the zinc selenide (ZnSe) windows are toxic.
WARNING
NOTE: A torque driver with a range of at least 10 – 25 cNm and a 3 mm hexagonal bit, for
1. Using a 3 mm hexagonal wrench, undo and remove the three hexagonal screws.
2. Remove the parts of the flowcell in the following order:
example part numbers L9003631 and L9003632, must be used to reassemble the flow cell. It is not recommended to start disassem b ling unless you have such a torque driver to-hand.
– The front of the cell – The PTFE gasket – The front window – The PTFE spacer – The former – The back window (this window has no holes in it) – The neoprene gasket
Routine Maintenance and Troubleshooting . 89
Figure 34 Dismantling or assembling the flowcell
3. When the flowcell is dismantled, use a lens tissue and a suitable solvent, for example heptane, to clean the flowcell.
Do not use acid to clean the windows, as toxic fumes will be given off. Take care not to crease the wedged PTFE spacer during cleaning as this
WARNING
4. Refit the parts of the cell in the following order:
– The neoprene gasket – The back window – The former – The PTFE spacer (with the thin edge of the wedge towards the top of the cell) – The front window – The PTFE gasket
may reduce its effectiveness at suppressing interference fringes, and may also alter the pathlength.
5. Align the two holes in the PTFE gasket with the two holes at the ends of the channels in the front window.
90 . OilExpress System User's Guide
Screw #1
Screw #2
Screw #3
Screw #4
6. Lay the front of the cell on top of the gasket.
7. Align the hole in the front of the cell with the holes in the gasket and front window.
The windows in the flowcell are brittle and easily broken. Be careful that
CAUTION
you do not tighten the screws holding the flowcell together too tightly.
8. Insert and tighten the four screws using a hexagonal wrench until the screws just touch the front of the cell.
9. Use a torque driver to tighten the screws, in the sequence shown, until a torque value of 10 cNm is reached.
NOTE: The sequence in which you tighten the screws is important. The sequence s ho wn
below is designed to prevent unequal pressure being applied to the flowcell, and thereby help prevent any damage being caused.
Figure 35 Flowcell securing sequence
10. Repeat step 9 until a torque value of 16 cNm is reached; and then again, until you reach a final torque value of 24 cNm.
11. Recalibrate the flowcell pathlength after dismantling and reassembling it.
NOTE: It is essential that the pathlength of the flowcell is recalculated after the cell has
been dismantled. Failing to carry out this procedure will seriously affect the results obtained using the OilExpress System.
Routine Maintenance and Troubleshooting . 91

Warranty Exclusions and Limitations

The following consumable items are excluded from the warranty agreement. See
Consumables
Flowcell Windows
Tubing
Fuses
Syringes
Syringe piston noses
on page 92 for details of part numbers a nd suppliers.
Spares and
Any attempts to perform installation or maintenance operations that are not detailed in this manual are at the user’s own risk.
WARNING
If a user-attempted service results in a visit by a PerkinElmer Service Engineer, the visit will not be covered by the instrument warranty.
Damage caused by the following is not covered by the warranty:
Not maintaining the solvent level.
Failure to flush the system adequately after use.
Blockage of tubing, the transfer/wash port, needle or flowcell.
Failure to observe the precautions described in this manual.
92 . OilExpress System User's Guide

Spares and Consumables

Item PerkinElmer Part Number
Matched pair of ZnSe Windows for flow cell L1191081
Replacement Gasket and Spacer Set for flowcell L1191082 Sample Rack (4 oz) L1200551
Syringe (10 ml) L1200552
Seal Kit (10 ml) for syringe plunger L1200553
Valve 3-port L1200554 Needle L1200555
Pump to Solvent Tubing (2) L1200556
Injection/Wash Port to Cell Tube L1200557
Lubrication Kit L1200559 Waste Tubing L1200560
Plunger Screw L1200571
Syringe Clamp L1200574
RS232 Cable L1200576 Wash Station Mounting Plate L1200577
Protective Shield L1200579
Consumables Kit L1200581
Injection Port L1200586 Solvent Bottle and Cap (drilled) L1195019
Syringe Pump to Cell Tubing L1200585
Syringe Pump to Needle Tubing L1200587
Pack of 6, 16-position Racks for Test tube Vials L1200589 Torque Driver 10 – 120 cNm L9003631
Hex Bit 3.0 mm (for use with L9003631) L9003632
Routine Maintenance and Troubleshooting . 93
The oil may not be
Ensure that the portion Loss of air gap due to tubing
Replace the defective tubing
Lead Screw Encoder may be
Clean Lead Screw Encoder with
transfer/wash port.
the Autosampler
on page 60).

Troubleshooting

NOTE: Troubleshooting as described in this chapter should be performed only by a qualified
service engineer.
Problem Possib l e C ause Corrective Action
During Oil Analysis
The results obtaine d for one or more properties are unacceptable. The cell pathlength may not
The flowcell may be dir ty or
Syringe Pump and Tubing
Syringe pump is not pumping or stalls.
Flush out the fluid system, (see
The oil type entered during 'Prepare Run' is incorrect.
be correct.
have a build-up of material on the windows.
representative of the whole sample.
Syringe pump or tubing is blocked.
Ensure the correct oil type has been entered.
Calibrate the flowcell (see
Calibrating the Pathlength of the Flowcell
Clean the flowcell (see
Maintenance of the Flowcell
page 87) and record a spectrum of the empty cell to ensure that it is clean. Ensure the pathlength is calculated correctly.
sampled was collected roughly halfway through the filtration process.
Flush out the syrin ge pump (see
Flushing the Syringe Pump
page 77).
on page 66).
on
on
Checking the Fluid Path (Tubing)
on page 79).
Needle and Wash Port
Cannot align needle. Call for a PerkinElmer Service Needle dirty or blocked. Residues from samples may
Sample spraying above
blocked, kinked or split.
dirty.
(see
Replacing the Tubing
page 85). air (see
Cleaning the Lead
Screw and Lead Screw Encoder
on page 78).
Engineer. Flush the needle using the
be deposited in the needle. The needle is not aligned. Align the needle (see
The waste tubing is i n the waste, causing a back­pressure.
Autosampler controls in the OilExpress software.
Empty the waste container and/or raise the tubing out of the waste (see
Fluid Path (Tubing)
Checking the
on page 79).
on
Calibrating
94 . OilExpress System User's Guide
Solvent is not being drawn
The solvent bottle is empty.
Re-fill the solvent bottle and
A vacuum has built up in the
Make sure that there is a gap
The nut on the flange of the
Refit the tubing and hand-
Sampler
Sampling rate is slow.
Air is leaking into the
Check all the connections to
Air bubbles in the sample.
Connections are loose or
Check all the connections to
Sample/Solvent
through.
solvent bottle, sucking the
sides of the bottle inwards. Sample/solvent is not being drawn through.
Solvent is leaking out. Connections are loose. Check all the connections to The tubing is split or leaking. Change the tubing (see
During data collection
The message not responding is displayed. The Autosampler is not
The Autosampler cable is
The tubing is split or leaking. Examine it and change it if
tube is not fitted correctly.
The Autosampler is switched
off.
connected to the PC.
connected to the wrong
COMM port on the PC.
empty the waste container at the same time.
around the tubing through the lid of the wash bottle.
necessary (see
on page 85).
Tubing
valves and pumps.
Replacing the Tubing
page 85). tighten the nut at the end of the
tubing.
Make sure that the Autosampler is switched on.
Make sure that the Autosampler is connected to the PC (see
Cable to PC
Make sure that the Autosampler is connected to the PC using the correct COMM port (default is COMM 1) (see page 43).
Replacing the
on page 43).
Cable to PC
on
on
Sampling rate is getting slower.
No spectrum was obtained during the run.
The cell is not installed in
The screws in the flowcell
system.
The waste tubing is i n the
waste, causing a back-
pressure.
The sample was not present
or the cell is not filling
correctly.
the sample shuttle.
leaking.
are loose.
valves and pumps, the tube-end fittings, and the connection to the top of the needle. Empty the waste container and/or raise the tubing out of the waste. Check all connections.
Install the cell in the fr ont of the sample shuttle (see page 49).
valves, flowcell and pumps, the tube-end fittings, and the connection to the top of the needle. Tighten the screws to correct torque (see
Deep Cleaning the
Figure 24
on
Flowcell, Involving Disassembly
on page 88).
During data collection (cont.)
Sample flow is poor.
Air is leaking into the
Check all the connections to The tubing is worn.
Change the tubing (see
Waste is not draining
The waste tubing is
Empty the waste container pumps.
Engineer.
Mechanical movement
Check movement mechanism for
CCU-9000 defective.
Call for a PerkinElmer Service
Routine Maintenance and Troubleshooting . 95
system.
Bubbles are present in the flowcell. Spectra are weak. The energy reaching the
properly.
General Autosampler mechanical failures:
No movement on arm or
LED on Autosampler does not come on.
Connections to the flowcell are reversed.
detector is low because the tubing in the sample compartment is in the beampath.
immersed in the waste.
Flex cable defective. Call for a PerkinElmer Service Main fuses blown. Replace fuses. Call customer
blocked.
valves and pumps, and the connection to the top of the needle.
Replacing the Tubing
page 85). Make sure that the connections are correct. Move the tubing out of the beampath.
and/or raise the tubing out of the waste.
support (see Autosampler Fuses on page 84).
obstructions.
Changing the
on
Autosampler X, Y or Z-axis failures.
If any of these problems persist, or if there are any other problems that you cannot correct, contact your local PerkinElmer office or agent .
Call for a PerkinElmer Service
Engineer. Engineer.
96 . OilExpress System User's Guide

Appendices

98 . OilExpress System User's Guide

Appendix 1: Decontamination

Before using any cleaning or decontamination methods except those specified by PerkinElmer, users should check with PerkinElmer that the proposed method will not damage the equipment.
Customers wishing to return instrumentation and/or associated materials to PerkinElmer for repair, mai nt enance, warranty or trad e -i n purposes are advised that all returned goods must be certified as clean and free from contamination.
The customer’s responsible body is required to follow the "Equipment Decontamination Procedure" and complete the “Certificate of Decontamination”. These documents are available on the PerkinElmer public website:
http://las.perkinelmer.com/OneSource/decontamination.htm
Alternatively, if you do not have access to the internet contact Customer Care: Customer Care USA: 1-800-762-4000 (inside the USA)
(8:30 a.m. – 7 p.m. EST ) (+1) 203-925-4602 (outside the USA)
Customer Care Canada : 800-561-4646 Customer Care EU: 0800 40 858 (Brussels)
080 0 90 66 42 (Monza)
If you are located outside of these regions, please call your local PerkinElmer sales office for more information.
Appendices . 99

Appendix 2: WEEE Instructions for P erkinElmer Products

or
A label with a crossed-out wheeled bin symbol and a rectangular bar indicates that the product is covered by the Waste Electrical and Electronic Equi pment (WEEE) Directive and is not to be disposed of as unsorted municipal waste. Any products marked with this symbol must be collected separately, and in accordance with the regulatory guidelines in your area.
The objectives of this program are to preserve, protect and improve the quality of the environment, protect human health, and utilize natural resources prudently and rationally. Specific treatment of WEEE is indispensable in order to avoid the dispersion of pollutants into the recycled material or waste stream. Such treatment is the most effective means of protecting the customer’s environment.
The requirements for waste collection, reuse, recycling, and recovery programs are set by the regulatory authority in your location. Contact your local responsible person (such as your laboratory manager) or authorized representative for information regarding applicable disposal regulations. Contact PerkinElmer at the web site listed below for information specific to PerkinElmer products.
Web address:
http://las.perkinelmer.com/OneSource/Environmental-directives.htm
Customer Care USA: 1-800-762-4000 (inside the USA )
(+1) 203-925-4602 (outside the USA)
Customer Care EU: 0800 40 858 (Brussels)
0800 90 66 42 (Monza)
Products from other manufacturers may also form a part of your PerkinElmer system. These other manufacturers are directly responsible for the collection and processing of their own waste products under the terms of the WEEE Directive. Please contact these manufacturers directly before discarding any of their products.
Consult the PerkinElmer web site (above) for manufacturer’s names and web addresses.
Loading...