Any comments about the documentation for this product should be addressed to:
User Assistance
PerkinElmer Ltd
Chalfont Road
Seer Green
Beaconsfield
BUCKS
HP9 2FX
United Kingdom
Or emailed to: info@perkinelmer.com
Notices
The information contained in this document is subject to change without noti ce .
Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no
warranty of any kind with regard to this document, including, but not limited to, the
implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in
connection with furnishing, performance or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright.
All rights are reserved. No part of this publication may be reproduced in any form whatsoever or
translated into any language without the prior, written permission of PerkinElmer, Inc.
Appendix 2: Customer Care Contacts ....................................................... 87
Index ..................................................................................................... 89
Page 5
Introduction
Page 6
6 . Lambda XLS/XLS+ User's Guide
About this Manual
This user’s guide describes the installation and operating procedures for your
Lambda XLS/XLS+ UV/Vis spectrometer.
The manual contains the following sections:
• Introduction;
• Safety Information;
• Installing your Instrument;
• Using your Instrument;
• Instrument Methods ;
• Accessories;
• Maintenance;
• Appendices.
Page 7
Introduction . 7
personal injury
serious damage to the instrument
Conventions Used in this Manual
Normal text is used to provide information and instructions.
Bold text refers to text that is displayed on the screen.
UPPERCASE text, for example ENTER or ALT, refers to keys on the PC keyboard. The
symbol “+” is used to show that you have to press two keys at the same time, for example,
ALT+F.
All eight digit numbers are PerkinElmer part numbers unless stated otherwise.
Unless stated otherwise, in this manual “instrument” refers to the Lambda XLS or
Lambda XLS+ instrument.
Notes, cautions and warnings
Three terms, in the following standard formats, are also used to highlight special
circumstances and warnings.
NOTE: A note indicates additional, significant information that is provided with so me
procedures.
We use the term WARNING to inf orm you about situations that could
to yourself or other persons. Details about
WARNING
result in
these circumstances are in a box like this one.
We use the term CAUTION to inform you about situations that co uld
CAUTION
result in
or other equipment.
Details about these circumstances are in a box like this one.
Page 8
8 . Lambda XLS/XLS+ User's Guide
Page 9
Safety Information
Page 10
10 . Lambda XLS/XLS+ User's Guide
the sample chamber during sample measurements. Under normal use,
removed for cleaning only when
General Safety
The Lambda Bio/XLS instruments conform to the requirements of the Low Voltage Directive
2006/95/EC, EMC Directive 2004/108/EC and the IVD Directive 98/79/EC.
Standards to which conformity is declared:
•EN 61010-1 Safety requirements for electrical equipment for measurement, control and
laboratory use.
• EN 61010-2-101 Particular requirements for IVD medical equipment.
• BS EN 591 Instruction for use for in vitro diagnostic instruments for professional use.
• BS EN 13612 Performance evalua tion of in vitro diagnostic medical devices.
• EN 61326-2.3 Electromagnetic compatibili ty – generic emission standard Electr ical
equipment for measurement, control and laboratory use.
•EN 61326 Electromagnetic compat ibility – generic emission standard Electrical
equipment for measurement, control and laboratory use. Classified as Class B for
conducted and radiated emissions.
The Lambda Bio/XLS complies with the following environmental legislation regarding the
restriction of the use of certain hazardous substances:
• Europe: EU directive 2002/ 95/EC (RoHS Directive).
• China: Administration on the Control of Pollution Caused by Electronic Informatio n
Products No 39.
This appliance is marked according to the European directive 2002/96/EC on Waste Electrical
and Electronic Equipment (WEEE). See the
more information.
There are a number of warning label s and symbols on your instrument. These are there to
inform you where potential dange r exists or particular caution is required. Before
commencing installation, please take tim e to familiarize yourself with these symbols and their
meaning.
This instrument is subject to the following hazards :
Lambda Bio/XLS Safety Manual
(L6050014) for
The UV source contained within the unit generates a light beam that passes
through
this beam is contained within the unit and no UV radiation is emitted from the
system. However, care should be taken that no reflective items are placed in
the beam during sample measurements so that the beam is reflected out of
the system. Also, the sample holder should be
the power is disconnected from the instrument. Prolonged exposure to the
beam may cause permanent eye damage.
Page 11
Safety Information . 11
•High voltages exis t inside the unit. Repair and maintenance should only be carried out
by individuals trained specifically to work on these instruments.
There are no bio-hazardous materials within the unit; however, this unit could
be used with bio-hazardous samples. Given the general-purpose nature of the
instruments it is not possible to define exact decontamination procedures for
the many types of samples being tested. These should be established by the
user, or their Health and Safety representative, before use of the instrument.
Guidance can be found in document
from Occupationally Acquired Infections: Approved Guidelines
the US Clinical and Laboratory Standards Institute. The sample compartment
may be removed and cleaned by soaking in a disinfectant appropriate for the
bio-hazard for at least 30 minutes. The compartment should then be rinsed in
distilled water and allowed to dry. Similarly, the outer cover can be wiped over
with an appropriate disinfectant with a wipe or soft cloth. Any internal spills
should be wiped up using absorbent material through the sample compartment
aperture, and then similarly decontaminated using disinfectant soaked wipes.
M29-A3 Protection of Laboratory Workers
published by
Decontamination Certificates
If the instrument needs to be returned, a decontamination certificate is required.
Decontamination certificates can be found at the following web address or be provide d by
Customer Care.
http://www.perkinelmer.com/LambdaBioXLSSupport
It is the responsibility of the user to ensure a safe working environment is provided to the
engineer or staff by appropriate risk assessments and, if appropriate, decontamination
procedures before work commences.
Chemicals
Use, store, and dispose of chemicals that y ou require for your analyses in accordance wi th
the manufacturer’s recommendations and local safety regulations.
Toxic Fumes
If you are working with volatile solvents or toxic substances, then you must provide an
efficient laboratory ventilation system to remove vapors that may be produced when you are
performing analyses.
Page 12
12 . Lambda XLS/XLS+ User's Guide
Waste Disposal
Waste containers may contain corrosive or organic solutions and small amounts of the
substances that were analyzed. If these materials are toxic, you may have to treat the
collected effluent as hazardous waste. Refer to your local safety regulations for proper
disposal procedures.
Personal Protective Equipment
No Personal Protective Equipment (PPE) is required for operation of the equipment, although
the samples under test may require the use of such PPE and should be cove red by a risk
assessment made by the user or the appropriative Health and Safety representative.
Intended User
This instrument is intended for use by individuals trained in and familiar with the use of
spectrophotometers and their associated hazards as detailed above. In the event of a
malfunction or hazard occurring, the user responsible shall disconnect the unit from power,
and isolate the instrument for decontamination by an appropriate means if biohazardous
material is spilled on, in or around the instrument.
Page 13
Safety Information . 13
e protection
Unpacking and Installation
•This equipment is designed to be installed in accordance with the Canadian Electrical
Code (CEC) Part 1, CSA C22.1 and CSA C22.2 No. 0.
•Inspect the instrument for any signs of damage caused in transit. If any damage is
discovered, access the Troubleshooting Guide on the PerkinElmer website
http://www.perkinelmer.com/LambdaBioXLSSupport and follow the process there. If
you do not have access to the webs ite, contact Customer Care (see
Customer Care Contacts
•Ensure your proposed installation site conform s to the environmental conditions for safe
operation:
Indoor use only.
Temperature range 5 °C to 35 °C.
Maximum relative humidity of 80% up to 31 °C, decreasing linearly to 50% at 40 °C.
NOTE: If you use the instrument in a room subjected to extremes of temperature change
during the day, it may be necessary to recalibrate (by switching off and then on
again) once thermal equilibrium has been established (2–3 hours).
on page 87).
Appendix 2:
•If the instrument has just been unpacked or has been stored in a cold environment, it
should be allowed to come to thermal equilibrium for 2–3 hours in the laborator y before
turning the instrument on. This will prevent ca lib r ation failure as a result of internal
condensation.
•The instrument must be placed on a stable, level bench or table that can take its weight
(approximately 4.5 kg) with enough space around the instrument so that air can
circulate freely.
•This equipment must be connected to the supply outlet with the power cord and power
supply provided. It can be used on 100–240 V, 50–60 Hz supplies. The power cord or
power supply outlet should be is fused at 3 A, 250 V.
•The instrument should be positione d so that the power cable can be readily
disconnected from the unit in the event of a hazard or malfunction occurring.
• Site the instrument in an atmosphere free from dust and corrosive fumes.
• Please read through the instrument User’s Guide suppl ied on the Lambda Bio/XLS
Documentation CD (L6050018) prior to use.
•Switch on the instrument via the ON/OFF button on the keypad after it has been
plugged in. The instrument will perform a serie s of self-diagnostic checks.
If the equipment i s used in a manner not specified herein, then t h
provided by the equipment may be impaired and instrument warranty
withdrawn.
WARNING
Page 14
14 . Lambda XLS/XLS+ User's Guide
Page 15
Installing yo ur Instrument
Page 16
16 . Lambda XLS/XLS+ User's Guide
Unpacking
Take great care when installing your instrument, and follow the
CAUTION
1. Check that you have the items listed in the following table.
Part Number Description
L7110189/L7110192 Lambda XLS/XLS+ (if applicable)
L7110190/L7110193 Lambda XLS/XLS+ with printer (if applicable)
L7110191/L7110194 Lambda XLS/XLS+ with SD (if applicable)
L6050014 Lambda Bio/XLS Safety Manual
procedures described in this manual.
L6050018 Lambda Bio/XLS Documentation CD
L7111025 Cuvette dust co ver
L7110231 Lambda Bio/XLS Print Utility and Cable (if applicable)
L7110270 UV macro cuvettes (8 pack)
N/A Power cable
2. Examine the components for any sig ns of damage in shipment:
•Check the enclosure of the instrument for damage, and make sure that the
terminals are not damaged.
•Make sure that the compartment is free of dust or other foreign matter.
NOTE: In the event of damage, or missi ng parts, acce ss the Troubleshooting Guide on the
PerkinElmer website at http://www.perkinelmer.com/LambdaBioXLSSupport and
follow the process there. If you do not have access to the website, contact Customer
Care (see
Appendix 2: Customer Care Contacts
on page 87).
Page 17
Installing your Instrument . 17
Electrical Hazard
USB cable connection
Line power in
Connecting to the Line Power Supply
To prevent potential injury to yourself and damage to the instrument,
WARNING
Connect to the supply outlet with power supply provided and the power cord that is
appropriate for your region.
You must use the power cord and power supply supplied with the instrument.
NOTE: If the instrument has just been unpacked or has been stored in a cold environment,
it should be allowed to come to thermal equilibrium in the laboratory for 2–3 hour s
before turning the instrument on. This will prevent calibration failure as a result of
internal condensation.
make any electrical connections in the system before connecting to the
line power supply.
The instrument automatically adjusts to the correct operating voltage.
Figure 1 Rear connections on the Lambda Bio/XLS instrument
The ON/OFF switch is located at the top left-hand of the keypad panel.
Page 18
18 . Lambda XLS/XLS+ User's Guide
Page 19
Using your Instrument
Page 20
20 . Lambda XLS/XLS+ User's Guide
Arrow keys
Alphanumeric keys
Cancel/Back/Stop
Run/OK/Next/Scan
0A/100%T (scan reference)
Display panel
ON/OFF
View Options
Overview of the Instrument
The Lambda XLS/XLS+ instrument is a simple-to-use UV/Visible spectrometer with a CCD
array detector (1024 pixels).
To access the folders and functions on your Lambda XLS/XLS+ instrument, navigate around
the display using the alphanumeric keys and the navigation arrows on the keypad.
Figure 2 Keypad and display
Key Action
Display panel Displays folders and menu options that guide you through taking
measurements and viewing your results.
Turns the instrument on or off.
Use the four arrows to navigate around the display and select the required
setting from the active (highlighted) option.
The Options ke y displays the options for that method or folder. See
on page 25 for more information. Options unique to a specific method are
described in the relevant section.
The 0A/100%T key sets a reference solution measurement to 0.000 A or
100%T at the current wavelength. When in scan mode, it performs a
reference scan.
Options
Page 21
Using your Instrument . 21
Alphanumeric
keys
The Cancel/Ba c k/Stop key Cancels a selection and returns to the previous
folder, goes Back to previous screen in a method or Stops a measurement.
The OK/Next/Run/Scan key, enters or confirms a selection, moves to the
Next screen or takes a measurement.
Use these to enter Method Settings, make Options menu selections and to
write text descriptions, where appropriate. Use repeated presses to cycle
through lower case, number and upper case. Leave for 1 second before
entering the next character.
Use the
key to backspace. Press 1 to enter a space.
Page 22
22 . Lambda XLS/XLS+ User's Guide
Switching on the Instrument
NOTE: If the instrument has just been unpacked or has been stored in a cold environment, it
should be allowed to come to thermal equilibrium for 2–3 hours in the laboratory
before being switched on. This will prevent ca lib r ation failure as a result of internal
condensation.
Do not turn power to the instrument on and off quickly as this may
CAUTION
Switch on the instrument via the ON/OFF on the keypad, as shown in Figure 2.
The instrument will perform a series of diagn ost ic che c ks.
NOTE: If during the calibration the instrument disp lays an error message it will prompt you
to press OK to restart the calibration procedure. However, after three attempts
you will still be able to view the default home page. You should NOT use the
instrument under these circumstances. Instead, make a note of the error message
and refer to the Troubleshooting Guide on the PerkinElmer website to address the
problem.
After switch on and calibration, the default home page is displayed.
To open a method or folder, press the appropriate number on the keypad.
damage the power supply.
Wait at least 30 seconds before restarting the instrument.
Figure 3 Lambda Bio home page
NOTE: If you have an SD Memory Card installed in your instrument that contains methods,
the SD Memory Card folder will be added to the home page. You can view and open
the first 9 methods saved to this folder. Further methods can be saved to the SD
Memory Card folder; see
these methods on your PC. If the SD Memory Card is empty it will not be displayed
on the home page.
SD Memory Card
on page 29 for information on how to view
Page 23
Using your Instrument . 23
Press OK to save
Press Cancel to
Using the Software
The home page contains a list of folders and functions or methods. Each item in the list is
numbered and is ope ned by using the associated number on the keypad. Opening a folder
takes you to another screen of further options. Opening a method takes you to a Method
Settings screen where you can set the required parameters before using the method to take
measurements.
the selected Method
Settings and go on
to the next screen
Figure 4 Features of a typical Method Settings screen
Navigation
Move between settings using the up and do wn arrows.
Entering Method Settings
Enter Method Settings by using the keypad numbers.
OR
If the box contains the symbol
choose an option from the next screen.
OR
If the box contains arrow symbols, use the le ft
required setting.
, either type in a value or press options , and
erase selection s
and return to the
previous screen
and right arrows to select the
Page 24
24 . Lambda XLS/XLS+ User's Guide
Taking measurements
After the method settings have been entered, press OK to start measurements. The
results screen for the method is displayed.
Figure 5 A typical results screen
1. Insert the Reference sample in the sample holder, and press OA/100%T
the reference measurement.
2. Insert your first sample and press
3. Repeat Step 2 for each sample.
.
to take
Results
The results are displayed on screen.
NOTE: Results shown as ---- indicate that the concentration is out of range, for example, if a
negative absorbance value is obtained.
Figure 6 The Single Wavelength me th od results screen
Page 25
Using your Instrument . 25
Options
Within each method you are able to select various options that define the way results are
treated. These options may be specific to the method, allow you to print the results or view
method settings.
Press Options
The options are:
Option Text Function
1 Method Settings Returns to the Method Settings screen for the method.
2 Print Prints the results.
3–6 N/A Specific to the method.
7 Sample Number Defines the sample number you wis h to start taking
to view the Options menu.
measurements from.
8 Save Method Allows you to save the Method Settings as a new method to
the Methods or Favorites folders or, if installed, to an S D
Memory Card. You can save up to 9 methods in each folder.
NOTE: Methods from more than one type of instrument can
be saved to the same SD Memory Card. The
Methods are saved to the c ard in a default directory
that depends on the instrument type: \<Instrument
type>\Methods. For example, \Lambda
XLS\Methods
9 Auto-Print Toggles Auto-Print on and off. The default is off.
Press or , or wait 20 seconds to exit the method options.
You can use the numeric keys as a shortcut to the option required without needing to
enter the Options menu.
Page 26
26 . Lambda XLS/XLS+ User's Guide
If you are not using a previously stored method, it is advisable to check that these options
have been appropriately set for your experiment before running your samples.
NOTE: Setting the History method setting to On (see
the instrument to store its last settings. If the History method setting is Off, all
method settings and options will return to their default s when you leave t hat method.
Preferences
on page 36) will cause
Page 27
Using your Instrument . 27
Methods and Favorites
The Methods folder allows you to store user-modified methods, based on the default
instrument methods, which can be saved using the Options menu. You can save up to 9
methods in each folder. To retrieve previously stored methods, press the appropriate keypad
number to enter the Methods folder. You can also export (backup) methods to an SD card or
import (restore) methods from an SD Memory Card.
Figure 7 Methods folder
Folder Options
Press to display the available Options in the Methods folder, which you can select
using the keypad numbers:
Option Text Function
1 Folder Names Allows you to rename the Methods folders.
2 Lock Folder Allows you to lock the folder.
3 Unlock Folder Allows you to unlock the folder.
4 SD Memory
Card
Allows you to access options for the SD card. (This option is
only available if you have an SD installed in the instrument.)
Page 28
28 . Lambda XLS/XLS+ User's Guide
Folder Names
1. Open the Methods folder.
2. Press
3. Press 1 to select Folder Names.
4. Select the Folder you wish to rename using the left and right arrows, and then press
the down arrow.
5. Enter a New Name for the folder using the alphanumeric keypad.
6. Press OK
to display the available Options.
to rename the folder.
Lock Folder
1. Open the Methods folder.
2. Press
3. Press 2 to select Lock Folder.
4. Select the Folder you wish to lock using the left and right arrows, and then press the
down arrow.
5. Enter a Pass Code for the folder using the keypad numbers or left and right arrows.
to display the available Options.
6. Press Lock
to lock the folder.
Unlock Folder
1. Open the Methods folder.
2. Press
3. Press 3 to select Unlock Folder.
4. Select the Folder you wish to unlock using the left and right arrows, and then press the
down arrow.
5. Enter the Pass Code for the folder using the keypad numbers or left and right arrows.
6. Press Unlock
to display the available Options.
to unlock the folder.
Page 29
Using your Instrument . 29
SD Memory Card
If an SD Memory Card is installed in your instrument, then you can access the SD Memory
Card folder option. The options Backup Folder or Backup All Folders allow you to copy a
methods folder or all of the methods folders on your instrument, respectively, to the SD
Memory Card. Methods from more than one instrument can be saved to the same card. The
methods are saved to a default directory \<Serial Num ber >\BACKUP
For example, \1000\BACKUP\METHODS\M et hods 1
The SD Memory Card can be plugged into a card reader attached to your PC. Methods can be
renamed (up to a maximum of 24 characters) and moved between folders.
The options Restore Folder or Restore All Folders restore the contents of the folder or
folders, respectively, to the content of the SD Memory Card BACKUP directory. All methods in
the folder(s) will be overwritten if you select Restore Folder or Restore All Folders. If you
have renamed an instrument folder, the folder name will be overwritten by those in the
BACKUP directory of the SD Memory Card.
NOTE: If you rename a folder on the SD Memory Card and then try to restore the contents,
your instrument will not recognize the folder and it will not be restored.
The Restore All Folders function can also be used if you wish to set multiple instruments of
the same type to have the same methods and folder structure. Navigate in Windows Explorer
to the \<Serial Number>\BACKUP folder that contains the methods you would like to use and
then rename the folder \<Instrument Type>\BACKUP. You can then restore e ach ins trum ent
in the normal way.
To Backup or Restore your folders:
1. Open the Methods folder.
2. Press
3. Press 4 to select SD Memory Card.
The SD Memory Card screen is displayed.
to display the available Options.
4. Select the mode of Operation using the left and right arrows, and then press the down
arrow.
The options are Backup Folder, Backup All Folders, Restore Folder or Restore All Folders.
Page 30
30 . Lambda XLS/XLS+ User's Guide
5. Select the Folder using the left and rig ht arrows.
6. Press OK
NOTE: The LED next to the card will be lit while methods are saved to the card. The card
should not be removed from the instrument while the LED is lit as this could corrupt
your data.
.
Method Options
Press to display the available Options in the Methods folder, which you can select
using the keypad numbers:
Option Text Function
1 Delete Method Allows you to delete the Method.
2 Lock Method Allows you to lock the method.
3 Unlock Method Allows you to unlock the method.
Delete Method
1. Open the appropriate methods folder.
2. Press
3. Press 1 to select Delete Method.
4. Select the method to be deleted using the left and right arrows.
5. Press Delete
to display the available Options.
to delete the method.
Lock Method
1. Open the appropriate methods folder.
2. Press
3. Press 2 to select Lock Method.
4. Select the method to be locked using the left and right arrows, and then press the down
arrow.
5. Select a Pass Code using the keypad numbers or left and right arrows.
to display the available Options.
6. Press Lock
to lock the method.
Page 31
Unlock Method
1. Open the appropriate methods folder.
Using your Instrument . 31
2. Press
3. Press 3 to select Unlock Method.
4. Select the method to be unlocked using the left and right arrows, and then press the
down arrow.
5. Enter a Pass Code using the keypad numbers or le ft and right arrows.
6. Press Unlock
to display the available Options.
to unlock the method.
Page 32
32 . Lambda XLS/XLS+ User's Guide
Lambda XLS/XLS+ Instrument Methods
The pre-defined methods available on your Lambda XLS/XLS+ instrument are:
• Single Wavelength
• Multi Wavelength
• Spectrum
• Concentration
• Standard Curve
• Kinetics
• Absorbance Ratio.
Figure 8 Lambda XLS home page
NOTE: If you have an SD Memory Card installed in your instrument that contains methods,
the SD Memory Card folder will be added to the home page.
Page 33
Using your Instrument . 33
Utilities
You can use the Utilities folder to modify the instrument setup.
From the home page, press the appropriate keypad number to enter the Utilities folder.
The Utilities screen is displayed.
Figure 9 The Utilities screen
The Utilities are:
1 Date and Time Set correct time and date.
2 Regional Select preferred language and number format.
3 Printer Printer/output options.
4 Preferences Select screen layout (themes) and history.
5 Contrast Adjust screen contrast and brightness.
6 About Serial number and software version.
7 Games Spectro Blocks/Sudoku.
Page 34
34 . Lambda XLS/XLS+ User's Guide
Date and Time
1. Press 1 to select Date and Time.
The Date and Time screen is displayed.
2. Using the keypad numbers or left and right arrows, enter the Day, Month, Year, Hour
and Minute.
3. Press OK
folder.
Seconds are zeroed when you press OK.
to store the settings to store the settings and return to the Utilities
Regional
1. Press 2 to select Regional.
The Regional screen is displayed.
2. Using the left and right arrows, select a Language and then press the down arrow.
The options are Chinese, English, French, German, Italian, Japanese or Spanish.
Page 35
Using your Instrument . 35
3. Select the Number Format (the type of decimal separator) using the left and right arrows.
The options are “,” o r “.”.
4. Press OK
to store the settings and return to the Utilities folder.
Printer
1. Press 3 to select Printer.
The Printer screen is displayed.
2. Select whether Auto-Print is On or Off using the left and right arrows, and then press
the down arrow.
When Auto-Print is On, the results are automatically printed after a measurement is taken
or when you exit the application, depending on the method. The appro pri at e icon for yo ur
Printer option will be displayed in the top right corner of the method results screen.
When Auto-Print is Off, printing has to be initiated manually.
3. Select how the data are sent.
The options are Built-in (the internal printer), Computer (USB) (to a computer via USB
port), SD Memory Card or Computer (Bluetooth) (to a computer via Bluetooth).
The options available depend on what is installed in your instrument.
NOTE: If SD Memory Card is selected, data is saved as .pvc files in a folder that depends
upon the instrument serial number: \<Serial number >\PVC. The method type is
saved in the file name. For information on viewing your results and savi ng your data
into other formats, see “Lambda Bio XLS Report Utility Viewer” in the
Lambda Bio/XLS Print Utility User’s Guide
NOTE: If an SD Memory Card is installed in your instrument, the LED next to the card will be
lit while data is being printed to the card. The card should not be removed from the
instrument while the LED is lit as this could corrupt your data. For some methods you
have to exit the method to end printing. See the “Receiving Pri nt Data” in the
Lambda Bio/XLS Print Utility User’s Guide
4. Press OK to store the settings and return to the Utilities folder.
(L6050015).
(L6050015) for more information.
Page 36
36 . Lambda XLS/XLS+ User's Guide
Preferences
1. Press 4 to select Preferences.
The Preferences screen is displayed.
2. Select whether Games are turned available using the left and right arrow keys.
The options are Yes or No.
3. Define the Theme (the on-screen layout of folders).
The options are either a Grid format or a List.
4. Select the History mode using the left and right arrow keys.
If Off is selected the instrument will not remember your method settings unless you
restore them as a new method. The default is Off.
5. Select whether to use an Auto Standby mode after defined periods.
The options are 1 hour, 2 hours, At Night or Off.
6. Press OK
to store the settings and return to the Utilities folder.
Page 37
Using your Instrument . 37
Contrast
Ambient temperature can affect the display. The contrast function can optimize the display
for local conditions.
The procedure is as follows:
1. Press 5 to select Contrast.
The Contrast screen is displayed.
2. Adjust the Brightness using the left and right arrows, and then press the down arrow.
3. Adjust the Contrast using the left and right arrows.
4. Press OK
to store the settings and return to the Utilities folder.
About
The About screen displays the instrument serial number and the software version.
1. Press 6 to select About.
The About screen for your instrument is displayed.
2. Press OK
to close the window and return to the Utilities folder.
Page 38
38 . Lambda XLS/XLS+ User's Guide
Sample Handling
•The light beam is directed from RIGHT to LEFT through the cell chamber, as indicated
by the white arrows on the instrument enclosure. Make sure that the cell is inserted in
the correct alignment.
•The cell holder supplied with the instrument accepts standard 10–40mm (Lambda XLS)
or 10 mm (Lambda XLS+) pathlength quartz, glass or plastic cells.
•The optical height is 15 mm, and the minimum volume that can be used is
approximately 70 µl in a micro cell.
NOTE: When using calibration standard filters, insert such that the flat surface is facing
away from the spring end of the cell holder.
NOTE: A cover is provided to keep the sample compartment free from dust and debris.
When the instrument is in use, the d ust cover should be kept in a safe place, nearby.
Page 39
Instrumen t Methods
Page 40
40 . Lambda XLS/XLS+ User's Guide
Single Wavelength
The Single Wavelength method makes absorbance (A) and percentage transmission (%T)
measurements on samples by measuring the amount of light that has p assed through a
sample relative to a reference (this can be air).
The procedure is as follows:
1. From the home page, press 1 to open the Standard Methods folder.
The Standard Methods screen is displayed.
2. Press 1 to open the Single Wavelength method.
The Single Wavelength – Method Settings screen is displayed.
3. Set the Wavelength by using the keypad numbers or the left and right arrows, and
then press the down arrow.
4. Select the Mode, using the left and right arrows.
The options are Absorbance or %Transmission.
Page 41
5. Press OK to start measurements.
The Single Wavelength results screen is displayed.
Instrument Methods . 41
6. Insert the reference sample and press
This reference measurement will be used for all subsequent samples until repeated.
7. Insert your first sample and press
The result at the selected wavelength is displayed on screen. Use the left and right
arrows to move the cursor and display the va lue at the cursor position (±20 nm from
the set wavelength).
.
.
8. Repeat Step 7 for each sample.
Page 42
42 . Lambda XLS/XLS+ User's Guide
Press to display the available Options, which you can select using the keypad
numbers:
1 Method Settings Returns to the Method Settings screen.
2 Print Prints the result using the printer selected in the Utilities folder.
3 A/%T Toggles between Absorbance and %Transmission mode.
4 Print Graph Prints the graph (this option is grayed out if no data are available).
7 Sample Number Adds a prefix to the sample number and resets the incrementing
number to the desired value.
8 Save Method Use the left and right arrows to select the folder you want to store the
method in (Favorites/Methods 1–9/SD Mem ory Card), then press the
down arrow and enter a name for the method. Press Save
the method.
9 Auto-Print Toggles Auto-Print on and off.
Press or , or wait 20 seconds to exit the method options.
Press
measured.
to return to the Standard Methods folder after all your samples have been
to save
Page 43
Instrument Methods . 43
Multi Wavelength
The Multi Wavelength method can provide absorbance readings at up to 5 wavelengths.
The procedure is as follows:
1. From the home page, press 1 to open the Standard Methods folder.
The Standard Methods screen is displayed.
2. Press 2 to open the Multi Wavelength method.
The Multi Wavele ngth – Method Settings screen is displayed.
3. Select the number of Wavelengths and then press the down arrow.
The options are between 2 and 5.
4. Enter the first wavelength (λ1) using either the number keys or the left and right
arrows, and then press the down arrow.
5. Repeat Step 4 for each of the wavelengths.
Page 44
44 . Lambda XLS/XLS+ User's Guide
6. Press OK to start measurements.
The Multi Wavele ngth results screen is displayed.
7. Insert the Reference sample and press
This reference measurement will be used for all subsequent samples until repeated.
8. Insert your first sample and press
A spectrum covering the range of wavelengths is di splayed, with cursors marking the
chosen wavelengths. A table of the absorbance values at those wavelengths is also
displayed.
.
.
9. Repeat Step 8 for each sample.
Page 45
Instrument Methods . 45
Press to display the available Options, which you can select using the keypad
numbers:
1 Method Settings Returns to the Method Settings screen.
2 Print Prints the data on the results screen (the absorbance table) us ing the
printer selected in the Utilities folder.
4 Print Graph Prints the graph (this option is grayed out if no data are available).
7 Sample Number Adds a prefix to the sample number and resets the incrementing
number to the desired value.
8 Save Method Use the left and right arrows to select the folder you want to store
the method in (Favorites/Methods 1–9/SD Memory Card), then press
the down arrow and enter a name for the method. Press Save
to save the method.
9 Auto-Print Toggles Auto-Print on and off.
Press or , or wait 20 seconds to exit the method options.
Press
measured.
to return to the Standard Methods folder after all your samples have been
Page 46
46 . Lambda XLS/XLS+ User's Guide
Spectrum
The Spectrum method enables you to collect an absorption spectrum, and determine peak
heights and positions.
The procedure is as follows:
1. From the home page, press 1 to open the Standard Methods folder.
The Standard Methods screen is displayed.
2. Press 3 to open the Spectrum method.
The Spectrum – Method Settings screen is displayed.
3. Set the Start Wavelength by using the keypad numbers or the left and right arrows,
and then press the down arrow.
4. Set the End Wavelength by using the keypad numbers or the left and right arrows,
and then press the down arrow.
Page 47
Instrument Methods . 47
5. Select the Mode using the left and right arrows, and then press the down arrow.
The options are Absorbance or %Transmission.
6. Press OK
The Spectrum results screen is displayed.
7. Insert the Reference sample and press
This reference measurement will be used for all subsequent samples until repeated.
8. Insert your first sample and press
A spectrum and a table of the Absorbance/%Transmission values at each peak are
displayed.
Use the left and rig ht arrows to move the cursor along the graph. When it reaches a
peak, the peak height and width are displayed at the top of the screen. To zoom in on
the wavelength scale, use the up arrow key. (This automatically scales the
Absorbance/%Transmission axis, depending upon the Graph Scale option, and this is
retained for subsequent measurements.) To zoom out again, use the down arrow key.
to start measurements.
.
.
9. Repeat Step 8 for each sample.
Page 48
48 . Lambda XLS/XLS+ User's Guide
Press to cancel the selections; pre ss to accept them.
Press to display the available Options, which you can se lect using the keypad
numbers:
1 Method Settings Returns to the Method Settings screen.
2 Print Prints the result using the printer selected in the Utilities folder.
3 A/%T Toggles between Absorbance and %Transmission mode
4 Peak Detection Displays the Peak Detection settings screen.
Auto-detect Peaks : Tur ns on and off the automatic peak
detection.
The following options determine how peaks are detected:
•Min Peak Height: Minimum height the peak has to be above
the higher of the two adjacent minima for the peak to be
detected.
•Min Peak Width: Minimum width of the peak as determined
by the difference in wavelength between the higher of the two
adjacent minima and the opposing intersection of that higher
minimum level and the peak profile.
•Peak Detect on Zoom: Determines if peaks are reassessed
and tabulated when you zoom into a region of the spectrum.
If No is selected, the peak detection remains as determined
before zooming.
•Sort Peaks By: Determines the sequence that peaks are
reported by. The options are Peak Width, Wavelength and
Peak Height.
•Indicate Peaks: Switches display of peak cursors on and off.
The peak cursors show vertical dashed lines displaying the
measured peak height and horizontal dashed lines showing
the peak width.
Page 49
Instrument Methods . 49
5 Add Peak/Delete
Peak
Add Peak manually adds a peak position to the peak table in the
results screen at the position set by the cursor. The entry in the
peak table is then displayed in inverse shading to discriminate
between User Defined Peaks and Auto-Detect peaks. If the cursor is
positioned over a user-defined peak, the legend User Defined Peak is displayed at the top of the scan. The option then changes
to Delete Peak to enable you to remove the peak.
NOTE: Saving a method at t his stage will save these user-defined
wavelengths, so that each time the method is run the
Absorbance values at these wavelengths are reported.
Page 50
50 . Lambda XLS/XLS+ User's Guide
and display the required graph.
6 Graph Scale Displays the Graph Scale settings screen. This enables you to set
up a defined graph by defining the limi ts in either or both of the
y
axes.
and
Zoom Mode: This sets up the operation of the Zoom keys (up and
down arrows). The mode x axis zooms the display around the
cursor point x axis, the mod e y axis zooms the display around the
cursor point y axis, and x & y axes expands both axes about the
x
cursor point. With
be permitted within the set limits.
or y axis limits set to On, zooming out will only
x
x axis limits/y axis limits: Setting to On activates the start and
finish points of the desired graph to user-defined, specific
wavelengths and/or absorbance values.
Press
7 Sample Number Adds a prefix to the sample number and resets the incrementing
number to the desired value.
8 Save Method Use the left and right arrows to select the folder you want to store
the method in (Favorites/Methods 1–9/SD Memory Card), then
press the down arrow and enter a name for the method. Press
Save
9 Auto-print Toggles Auto-Print on and off.
Press or , or wait 20 seco nds to exit the method options.
Press
measured.
to return to the Standard Methods folder after all your samples have been
to cancel the selections; press to accept them
to save the method.
Page 51
Instrument Methods . 51
Concentration
The Concentrati on method makes simple concentration measurements on samples by
measuring the amount of light that has passed through a sample relative to a reference (this
can be air). The concentration is determined by measuring the absorbance at a specific
wavelength and then multiplying by a factor. If the factor is known in advance it can be
entered into the method. If the facto r is not known, a standard of known concentration can
be measured and the factor then calculated by the instrument.
The procedure is as follows:
1. From the home page, press 1 to open the Standard Methods folder.
The Standard Methods screen is displayed.
2. Press 4 to open the Co ncentration method.
The Concentration – Method Settings screen is displayed.
3. Enter the Wavelength by using the keypad numbers or the left and right arrows, and
then press the down arrow.
Page 52
52 . Lambda XLS/XLS+ User's Guide
4. Select the Mode using the left and right arrows, and then press the down arrow.
The options are Enter Factor (where a known factor i s entered), Measure Standard
(where the factor is calculated from a calibration sample) or Enter Neg. Factor (where
a known negative factor is entered).
5. If you selected Enter Factor or Enter Neg. Factor, enter the Factor (or Negative Factor) using the keypad numbers, and then press the down arrow.
The available range is from 0.01 to 9999.
OR
If you selected Measure Standard, enter the Concentration using the keypad
numbers and then press the down arrow.
The available range is from 0.001 to 9999.
6. Enter the Units and then press the down arrow.
You can enter text up to 8 characters long using the keypad, or press Options
display the Units screen and access a list of pre-defined units .
In the Units scree n, use the left and right arrows to select from the available options.
The options are µg/ml, µg/µl, pmol/µl, mg/dl, mmol/l, µmol/l, g/l, mg/l, µg/l, U/l, %,
ppm, ppb, conc or none.
The Units screen also enables you to select the number of displayed decimal points
(DP). The options are Auto, 0, 1 and 2.
NOTE: The result will always be fixed to 5 significant figures, regardless of how ma ny
decimal points are selected (so 98768.2 is displayed as 98768, even with 1 decimal
point selected).
to
Press OK to save your options and return to the Method Settings screen.
You can still edit the units.
Page 53
7. Press OK to start measurements.
The Concentration results screen is displayed.
Instrument Methods . 53
8. Insert the Reference sample and press
This reference measurement will be used for all subsequent samples until repeated.
9. If you selected Measure Standar d, insert your standard and press OK
The Run Standard screen is displayed.
OR
If you selected Factor or Negative Factor, go to Step 11.
.
.
Page 54
54 . Lambda XLS/XLS+ User's Guide
10. Press Run to run the standard.
The Factor is calculated and displayed.
11. Insert your first sample and press
The Concentration of the sam p le is displayed.
12. Repeat Step 11 for each sample.
.
Page 55
Instrument Methods . 55
Press to display the available Options, which you can select using the keypad
numbers:
1 Method Settings Returns to the Method Settings screen.
2 Print Prints the result using the pri nter selected in the Utilities
folder.
3 Show Graph Toggles the graph on and off. The wave length range of the
graph displayed is ±20 nm from the selected wavelength,
which is marked by the cursor.
4 Run Standard Returns to the Run Sta ndard screen.
7 Sample Number Adds a prefix to the sample number and resets the
incrementing number to the desired value.
8 Save Method Use the left and right arrows to select the folder you want to
store the method in (Favorites/Methods 1–9/SD Memory
Card), then press the down arrow and enter a name for the
method. Press Save
9 Auto-Print Toggles Auto-Print on and off.
Press or , or wait 20 seconds to exit the method options.
Press
measured.
to return to the Standard Methods folder after all your samples have been
to save the method.
Page 56
56 . Lambda XLS/XLS+ User's Guide
Standard Curve
This method enables the construction of a multi-point calibration curve from standards of
known concentration that can then be used to quantify unknown samples. This instrument
has the advantage of being able to store this curve as a method, using up to nine standards.
If you wish to have a zero concentration standard, include this in the number of standards to
be entered and enter 0.00 for the Concentration, then use a reagent blank when required to
enter the zero standard.
The procedure is as follows:
1. From the home page, press 1 to open the Standard Methods folder.
The Standard Methods screen is displayed.
2. Press 5 to open the Standard Curve method.
The Standard Curve – Method Settings screen is displayed.
3. Enter the required Wavelength using the keypad numbers or left and right arrows, and
then press the down arrow.
Page 57
Instrument Methods . 57
4. Enter the number of Standards to be use d in the standards curve using the keypad
numbers or left and right arrows, and then press the down arrow.
The options are 1–9.
5. Enter the Units and then press the down arrow.
You can enter text up to 8 characters long using the keypad, or press Options
display the Units screen and access a list of pre-defined units.
In the Units scree n, use the left and right arrows to select from the available options.
The options are µg/ml, µg/µl, pmol/µl, mg/dl, mmol/l, µmol/l, g/l, mg/l, µg/l, U/l, %,
ppm, ppb, conc or none.
The Units screen also enables you to select the number of displayed decimal points
(DP). The options are Auto, 0, 1 and 2.
NOTE: The result will always be fixed to 5 significant figures, regardless of how ma ny
decimal points are selected (so 98768.2 is displayed as 98768, even with 1 decimal
point selected).
to
Press OK to save your options and return to the Method Settings screen.
You can still edit the units.
6. Select the type of Curve Fit using the left and right arrows, and then press the down
arrow.
The options are Zero Regression (thi s forces the straight line through the origin),
Regression (straight-line regression), Interpolation (linear interpolation) or
Cubic Spline.
7. Select the Calibration mode using the left and right arro ws.
The options are Standards (to measure prepared standards) or Manual (entering
absorbance values by keypad) or New Standards (to replace any standards that have
already been collected).
8. If you selected Standards, press the down arrow and enter the number of Replicates.
This determines the number of standards to be measured and averaged at each
standard concentr ation point.
The options are Off (1), 2 or 3.
NOTE: If you selected Manual in Step 7, the option Replicates is not available.
Page 58
58 . Lambda XLS/XLS+ User's Guide
9. Press Next.
The Standard Curve – Standards screen is displayed.
10. Enter the concentration values for each standard using the keypad numbers and the up
and down arrows to move be tween the different standard boxes.
The available range is from 0.001 to 9999.
11. Press Next
The Standard Curve – Calibration screen is displayed.
NOTE: If any dupli c ate or non-monotonic entries are entered on the Standards screen, when
Next is pressed the instrument will beep and highlight the incorrect entry.
.
Page 59
Instrument Methods . 59
12. If you selected Standards, insert the reference sample and then press .
This reference measurement will be used for all subsequent samples until repeated.
OR
If you selected Manual, enter the known absorbance value for each concentration using
the keypad. Go to Step 15.
The available range is from 0.001 to 9999.
13. Insert your first standard and press
If there are any replicates, press Replicates
measure each replicate.
14. Repeat Step 13 for all your standards.
A graph will display the results and the curve fit (for example, zero regression) a s the
measurements are made.
.
to view them and then press to
NOTE: You can use the up and down arrows to select a standard to be repeated. Press
to clear the previous reading before measuring.
Page 60
60 . Lambda XLS/XLS+ User's Guide
If you selected Replicates, then when all replicates have been measured press Next
to return to the Standards screen.
15. Press OK
The Standard Curve results screen is displayed.
16. If you selected Manual, insert the reference sample and then press
This reference measurement will be used for all subsequent samples until repeated.
OR
to accept the calibration.
.
If you selected Standards, go to Step 17.
Page 61
Instrument Methods . 61
17. Insert your first sample and then press .
The absorbance of the sample is measured, and its concentra tion is calculated and
displayed.
18. Repeat Step 17 for each sample.
Press
keypad numbers:
1 Method Settings Returns to the Method Settings screen.
2 Print Prints the data on the results scre en using the printer selected in
3 Show Graph Toggles the calibration graph on and off. The cursors display the
7 Sample Number Adds a prefix to the sample number and resets the incrementing
8 Save Method Use the left and right arrows to select the folder you want to store
9 Auto-print Toggles Auto-Print on and off.
to display the available method options, which you can select us ing the
the Utilities folder.
absorbance and corresponding concentration values of the last
measured sample.
number to the desired value.
the method in (Favorites/Methods 1–9/SD Memory Card), then
press the down arrow and enter a name for the method. Press
Save
to save the method.
Press or , or wait 20 seconds to exit the method options.
Press
measured.
to return to the Standard Methods folder after all your samples have been
Page 62
62 . Lambda XLS/XLS+ User's Guide
Kinetics
Kinetics studies, where the change in absorbance needs to be followed as a function of time
at a fixed wavelength, can readily be performed using the Kinetics method.
Reagent test kits are routinely used for the enzymatic determination of compounds in food,
beverage and clinical laboratories by measuring NAD/NADH conversion at 340 n m. The
change in absorbance over a specified time period can be used to provide useful information
when an appropriate factor, defined in the reagent -kit protocol, is applied. Reaction rate and
enzyme activity can be calculated if the fa c tor used takes account of the absorbance
difference per unit time, as opposed to only the absorbance difference.
For this reason, the change in absorbance per minute (∆A/min), concentration (∆A/min x
factor) and correlation coefficient (calculated from a best fit of the data points) are displayed.
They may not be relevant for simple kinetics experiments.
The procedure to define a new method is as follows:
1. From the home page, press 1 to open the Standard Methods folder.
The Standard Methods screen is displayed.
Page 63
Instrument Methods . 63
2. Press 6 to open the Kinetics method.
The Kinetics – Metho d Set ti ngs 1 screen is displayed.
Enter all numerical values using the keypad numbers or the left and right arrows.
3. Set the Wavelength by using the keypad numbers or left and right arrows, and then
press the down arrow.
The default value is 340 nm.
4. Enter the Delay Time in seconds before measurements are taken and then press the
down arrow.
This can be a maximum of 600 seconds (10 minutes).
5. Enter the time in minutes over which measurements are taken (Duration) and then
press the down arrow.
This can be a maximum of 60 minutes.
6. Enter the Interval time in seconds between measurements using the left and right
arrows, and then press the down arrow.
The options are 1, 2, 5, 10, 15, 20, 30 or 60 seconds.
Page 64
64 . Lambda XLS/XLS+ User's Guide
7. Press Next.
The Kinetics – Method Setti ngs 2 scr ee n is displayed.
8. Select the measurement Mode using the left and right arrows, and then press the down
arrow.
The options are Delta A (the change in absorbance over the measurement duration, or
selected period), Final A (the absorbance at the end of the measurement duration, or
selected time) and Slope (the rate of change of absorbance over the measurement
duration or selected period).
9. Enter the Units and then press the down arrow.
You can enter a text string up to 8 characters long. To access a list of pre-defined uni ts
on the Units screen, press Options
µg/µl, pmol/µl, mg/dl, mmol/l, µmol/l, g/l, mg/l, µg/l, U/l, %, ppm, ppb, conc or
none). These units can also be edited once OK is pressed.
The Units screen also enables you to select the number of displayed decimal points
(DP). The options are Auto, 0, 1 and 2.
NOTE: The result will always be fixed to 5 significant figures, regardless of how many
decimal points are selected (so 98768.2 is displayed as 98768, even with 1 decimal
point selected).
and then use the left/right arrows (µg/ml,
Press OK
You can still edit the units.
to save your options and return to the Method Setti ngs screen.
Page 65
Instrument Methods . 65
10. Set the Factor by which the absorbance is multiplied to give the concentration using
the left and right arrows.
The available range is from 0.01 to 9999.
11. Press OK
to start measurements.
The Kinetics results screen is displayed.
12. Insert the reference sample and press
.
This reference measurement will be used for all subsequent samples until repeated.
13. Insert your first sample and then press
to start the run.
The running Time is displayed at the bottom of the screen, and absorbance data are
plotted on the graph as the measurement proceed.
The table below the graph displays A
the absorbance value at tn (finish of calculation); dA, the change in absorbance; the
A
n,
Slope and the regression parameter (R
the absorbance value at t0 (start of calculation);
0,
2
) of the calculated slope. The concentration is a
calculated using the parameters selected in Method Settings 2 (dA, Final A or Slope) –
the name of the Result table column depends on the units chosen in Step 9.
Use the left and right arrows to move the cursor and display the time and absorbance
value at measured data points.
Use the up and down arrows to zoom in or out.
Page 66
66 . Lambda XLS/XLS+ User's Guide
Press to display the available method options, which you can select using the
keypad numbers:
1 Method Settings Returns to the Method Settings screen.
2 Print Prints the data on the results scre en using the printer selected in
the Utilities folder.
3 Print Data Prints all the dat a.
4 Set t0 at cursor Sets the t0 position (the starting point for the slope and dA
calculations) at the current cursor positio n. The value is retained
for subsequent samples.
5 Set tn At Cursor Sets the tn position (the finishing point for t he slope and dA
calculations) at the current cursor position. The value is retained
for subsequent samples.
6 Show Slope Toggles the calculated slope line on and off.
NOTE: If any data points enclosed by t0 and tn are beyond the
range of the instrument (that is, >2.5 A or <−0.3 A) then
this option is grayed out.
7 Sample Number Adds a prefix to the sample number and resets the incrementing
number to the desired value.
8 Save Method Use the left and right arrows to select the folder you want to store
the method in (Favorites/Methods 1–9/SD Memory Card), then
press the down arrow and enter a name for the method. Press
Save
9 Auto-print Toggles Auto-Print on and off.
Press or , or wait 20 seconds to exit the method options.
Press
measured.
to return to the Standard Methods folder after all your samples have been
to save the method.
Page 67
Instrument Methods . 67
Absorbance Ratio
The Absorbance Ratio method makes simple absorbance ratio measurements on samples, by
measuring the amount of light that has passed through a sample relative to a blank (this can
be air) at two wavelengths.
The procedure is as follows:
1. From the home page, press 1 to open the Standard Methods folder.
The Standard Methods screen is displayed.
2. Press 7 to select the Absorbance Ratio method.
The Absorbance Ratio – Wavelengths screen is displayed.
3. Enter Wavelength 1 by using the keypad numbers or the left and right arrows, and
then press the down arrow.
4. Enter Wavelength 2 as above and then press the down arrow.
5. Select whether a Background correction is applied to both wavelengths 1 and 2 using
the left and right arrows.
Page 68
68 . Lambda XLS/XLS+ User's Guide
6. If background correction is On, enter Wavelength 3, from which the background
correction will be obtained.
7. Press Next
The Absorbance Ratio – Method Settings screen is displayed.
8. Select the Pathlength using the left and right ar rows, and then press the down arrow.
The options are 5 mm or 10 mm.
9. If the Dilution Factor is known, enter a dilution factor by using the keypad numbers,
press the down arrow and then go to Step 13.
The available range is from 1.00 to 9999.
.
OR
If the dilution factor is unknown, press Options
screen.
10. Enter the Volume of the sample using the keypad numbers and then press the down
arrow.
The available range is from 0.01 to 9999.
to display the Dilution Factor
Page 69
11. Enter the volume of Diluent by using the keypad numbers.
The available range is from 0.01 to 9999.
Instrument Methods . 69
12. Press OK
13. Select the Units of measurement using the left and right arrows, and then press the
down arrow.
The options are µg/ml, ng/µl, µg/µl.
14. Enter the Factor using the keypad numbers and then press the down arrow.
The available range is from 0.001 to 9999 .
15. Press OK
The Absorbance Ratio results screen is displayed.
to calculate the dilution factor and return to the Method Settings screen.
to start measurements.
16. Insert the reference sample and then press
This reference measurement will be used for all subsequent samples until repeated.
17. Insert your first sample and press
The absorbance at each wavelength is measured. If Background was selected, the
background absorbance is then subtracted from the absorbance at Wavelength 1 and
Wavelength 2. The ratio between Wavelength 1 and Wavelength 2 is calculated.
.
.
Page 70
70 . Lambda XLS/XLS+ User's Guide
18. Repeat Step 17 for each sample.
Press
keypad numbers:
1 Method Settings Returns to the Method Settings screen.
2 Print Prints using the printer selected in the Utilities folder.
3 Show Graph Toggles the graph on and off. Th e graph shows a spectrum across
7 Sample Number Adds a prefix to the sample number and resets the incrementing
8 Save Method Use the left and right arrows to select the folder you want to store
9 Auto-Print Toggles Auto-Print on and off.
Press or , or wait 20 seconds to exit the method options.
to display the available method options, which you ca n select usi ng the
the range of the measured wavelengths, which are marked by
cursors.
number to the desired value.
the method in (Favorites/Methods 1–9/SD Memory Card), then
press the down arrow and enter a name for the method. Press
Save
to save the method.
Press
measured.
to return to the Standard Methods folder after all your samples have been
Page 71
Accessories
Page 72
72 . Lambda XLS/XLS+ User's Guide
Accessories and Consumables
Supplies, accessories, and replacement parts can be ordered directly from PerkinElmer.
See the PerkinElmer website at http://www.perkinelmer.com for more information.
Part Number Accessory/Consumable
L7110230 Lambda Bio/XLS Printer Module
L7110231 Lambda Bio/XLS Print Utility and Cable
L7110232 Lambda Bio/XLS Spare Printer Paper (20 Rolls)
L7110233 Lambda Bio/XLS Bluetooth Accessory
A range of cells and microcells are also availab le from PerkinElmer that are suitable for use
with the Lambda XLS/XLS+ instrument.
Page 73
Accessories . 73
Cap head screws
Printer Installation and Configuration
The Printer Accessory (L7110230) is an optional accessory that allows you to print your data
as you run your samples, directly from your instrument.
Printer installation
1. Turn off the line power supply and disconnect the power cable.
2. Turn the instrument over and remove the two cap head screws using the 3 mm
hexagonal wrench provided .
3. Turn the instrument back over and lift the ac c essory cover vertically upwards to
remove. Remove the tie-wrap from the cable.
4. Invert the instrument and replace the cap head screws.
Page 74
74 . Lambda XLS/XLS+ User's Guide
Locating bosses
Accessory cable
Printer
5. Plug the accessory cable into the printer.
6. Lower the printer onto the locating b osses and push down firmly.
7. Reconnect the power cable.
Printer configuration
1. Switch the instr ument on and then open the Utilities folder.
2. Press 3 to open the Printer folder.
3. Select Built-in as the Printer option using the arrow keys.
4. Press OK
to save the settings.
Page 75
Loading or changing the printer paper
Platen lock released
Platen locked
1. Lift off the paper cover.
2. Lock the platen and turn the knob to feed the paper.
3. Feed in the paper.
Sometimes it helps if the platen lock is released.
Accessories . 75
4. Replace the cover.
Page 76
76 . Lambda XLS/XLS+ User's Guide
Cap head screws
Bluetooth Accessory Installation
1. Remove the power cable from the instrument.
2. Turn the instrument over and remove the two cap head screws using the 3 mm
hexagonal wrench provided .
3. Turn the instrument back over and lift the ac c essory cover vertically upwards to remove
it.
4. Remove the tie-wrap from the cable.
Page 77
Accessories . 77
Bluetooth accessory slots
5. Plug the accessory cable into the Bluetooth module.
6. Plug the two lugs on the Bluetooth module into the innermost slots in the base of the
case.
7. Lower the accesso ry cover vertically downwards onto the instrument, engaging the PCB
in the innermost slots in the accessory cover.
8. Invert the instrument and replace the two cap head screws.
9. Connect the power cable.
Page 78
78 . Lambda XLS/XLS+ User's Guide
10. Switch the instrument on and open the Utilities folder.
11. Press 3 to open the Printer folder.
12. Select Computer (Bluetooth) as the Printer option using the arrow keys.
13. Press OK
to save the settings.
Page 79
Accessories . 79
Lambda Bio/XLS Print Utility
There are two software applications comprising the Lambda Bio/XLS Print Utility.
The Lambda Bio/XLS Report Utility is an application running separately under Windows XP
that enables you to connect via a USB cable to a PC onto which the software has been
installed. You can connect to several instruments simultaneously. Once your data is on the
PC it can be printed directly to any printer that is connected to that PC or saved. The Report
Utility can save data as an Enhanced Metafile (.emf) graphics file, a comma delimited (.csv)
data file, a tab delimited (.txt) data file or as an Excel file, or in print utility (.pvc) format for
viewing using the Lambda Bio/XLS Report Viewer.
NOTE: An internal print er is also available for the instrume nt. This is either supplied
pre-installed in the instrument or as an optional accessory (L7110230).
If you have this option, please see the
on the
Lambda Bio/XLS User Manuals CD
Lambda Bio/XLS Print Utility User’s Guide
(L6050018) for more information.
(L6050015)
Page 80
80 . Lambda XLS/XLS+ User's Guide
Page 81
Maintenance
Page 82
82 . Lambda XLS/XLS+ User's Guide
Unauthorized Adjustments and Servicing
General Maintenance
Do not attempt to make adjustments, replacements or repairs to this
WARNING
CAUTION
instrument. There are no user-serviceable parts in this instrument. If
you have any problems, access the Troub leshooting Gui de on the
PerkinElmer website and work through the process there:
http://www.perkinelmer.com/LambdaBioXLSSupport
If you do not have access the Troubleshooting Guide, or have any
further problems, contact Customer Care (see Appendix 2: Customer
Care on page 87
Before using any cleaning or decontamination me thods except those
specified here, users should check with Customer Care that the
proposed method will not damage the instrument.
for more information).
Cleaning the instrument
External Cleaning
Switch off the instrument and disconnect the power cord before cleaning your instrument.
Clean the outside of the instrument using a damp cloth. A mild liquid detergent may be used,
if necessary. Always test an inconspicuous area of the instrument before you clean the entire
instrument.
Clean all spilled materials from the affected area immediately and wipe it dry with lint-free
paper or cloth.
Removing the cell holder for cleaning
To remove the sample cell holder for cleaning:
1. Loosen the thumbscrews on the sample-holder cover on the base of the instrument by
turning them anticlockwise.
The thumbscrews remain attached to the plate.
Page 83
2. Lower the sample holder out of the instrument.
Thumbscrews
Figure 10 Position of cell holder thumbscrews
Maintenance . 83
Page 84
84 . Lambda XLS/XLS+ User's Guide
Storage and Shipping
This unit should be stored or shipped in its original packaging, which should be kept for this
purpose. The unit should be kept in i ts packaging in a dry, non condensing environm ent, free
from dust or dirt, in the range −10 to 50 °C.
Should you need to return your instrument to PerkinElmer, a decontamination certificate is
required before the instrument is shipped. Decontamination certificates can be obtained from
PerkinElmer. The Troubleshooting Guide, which will help you work thr ough the returns
process, can be accessed at:
http://www.perkinelmer.com/LambdaBioXLSSupport
If you do not have access to the Troubleshooting Guide, you can also contact Customer Care
Appendix 2: Customer Care
(see
It is the responsibility of the user to carry out appropriate risk assessments and, if applicable,
decontaminatio n procedures be fo re returning the instrument.
on page 87 for more information).
Page 85
Appendices
Page 86
86 . Lambda XLS/XLS+ User's Guide
Appendix 1: Specifications
Specifications are measured after the instrument has warmed up at a constant ambient
temperature and are typical of a production unit.
Specification Lambda XLS Lambda XLS+
Wavelength range 190–1100 nm
Wavelength scanning range 200–950 nm
Monochromator Flat grating
Wavelength calibration Automatic when the instrument is switched on
Spectral bandwidt h 5 nm 3 nm
Wavelength accuracy ±2 nm
Wavelength reproducibility ±1 nm
Light sources Pulsed xenon lamp
Detector 1024 element CCD array
Absorbance range −0.300 to 2.500 A (0 to 199%T)
Photometric linearity ±0.005 A or 1% of the reading, whichever is the greater at 546 nm
Photometric reproducibility ±0.003 A (0–0.5 A), ±0.007 A (0.5–1.0 A)
Stray light <0.5% at 220 nm and 340 nm using NaNO2
Zero stability ±0.01 A/hour after 20 minutes warm up at 340 nm
Noise 0.005 peak-to-peak 0.002 rms
Digital output USB port standard, Bluetooth option
Dimensions 340 × 310 × 170 mm (340 × 420 × 170 mm with printer attached)
Weight Approximately 4.5 kg
Power input 100–250 V, 50/60 Hz, Max 30 VA
Page 87
Appendices . 87
Appendix 2: Customer Care Contacts
NOTE: If you experience any problems with your instrument, please refer to the
Troubleshooting Guide at http://www.perkinelmer.com/LambdaBioXLSSupport before
contacting Customer Care.
Customer Care USA: 1 00 762 4000 (In the USA)
(+1) 203 925 4602 (Outside USA)
Customer Care EU: 0800 40 858 (Belgium)
0805 111 333 (France)
0800 0 00 66 79 (Germany)
0800 90 66 42 (Italy)
800 099 164 (Spain)
0800 896 046 (UK)
Customer Care Brazil: 11 38 68 6200
Customer Care Canada: 800 561 4646
Customer Care China: 10 5820 8166
Customer Care India: 022 6760 1700
Customer Care Japan: 45 339 5889
Customer Care Singapore: 67799 539
Page 88
88 . Lambda XLS/XLS+ User's Guide
Page 89
Index
Appendices . 89
A
Absorbance Ratio ................................ 67