This instruction sheet describes the installation of this accessory which is used with the Lambda 365
Spectrophotometer.
NOTE:
Read these instructions before you install this accessory.
Contacting PerkinElmer
Supplies, replacement parts, and accessories can be ordered directly from PerkinElmer, using the part
numbers.
See our website:
http://perkinelmer.com
PerkinElmer's catalog service offers a full selection of high-quality supplies.
To place an order for supplies and many replacement parts, request a free catalog, or ask for
information:
If you are located within the U.S., call toll free 1-800-762-4000, 8 a.m. to 8 p.m. EST. Your order will be
shipped promptly, usually within 24 hours.
If you are located outside of the U.S., call your local PerkinElmer sales or service office.
reference) [P/N: N4101029] or Peltier Temp Unit (Multi) w/o Reference
Heating [P/N: N4101008]
PerkinElmer, 710 Bridgeport Avenue,
Shelton, CT 06484-4794, U.S.A
Produced in the USA.
09931286E
Dimensions and Specifications
Peltier Controller
Physical Characteristic Specification
Power 100-240 VAC, 50/60 Hz, 300W, Free Voltage
Temperature Range -5 to 100°C (Maximum Internal Temperature)
Maximum Ambient Operating Temperature 40°C
Dimensions
Weight 20.0 kg (44 lb)
Coolant Volume 2 L
N2 Gas Available
Liquid Cooling System
Magnetic Stirrer Control Available
Emergency Power Off Switch Available
310 (W) x 410 (D) x 275 (H) mm
(12.2 (W) x 16.1 (D) x 10.8 (H) in)
100 psi (6.9 bar)
6-Position Peltier Controlled Cell Changer (with and without heated reference)
Physical Characteristic Specification
The number of Sample Cells 6
The number of Reference Cell 1 (with or without heated reference)
Dimensions
135 (W) x 384 (D) x 143 (H) mm
(5.3 (W) x 15.1 (D) x 5.6 (H) in)
Weight
Temperature Accuracy ±0.15°C
Temperature Precision ±0.1°C
Temperature Stability ±0.1°C
Ramping Time (from 0 to 100°C) 5 minutes
With heated reference: 3.4 Kg (7.5 lb)
Without heated reference: 3.1 Kg (6.8 lb)
2
Safety Warnings
g
g
When this label is attached to an instrument it means refer to the
manual.
WARNING
WARNING
See the following figure for the location of warning labels on the back of the instrument:
Lorsque cette étiquette est attachée à un instrument, il est nécessaire
de voir le manuel.
There is risk of receiving a fatal electric shock if the fuses are replaced
with the power cord connected.
Il y a un risque d’électrocution si les fusibles sont remplacés tandis que
le cordon d'alimentation est encore branché.
09931286E
Warnin
Warning
FUSE
Avertissement
FUSE
For continued protection
against electrical shock,
replace with same type and
ratings of fuse.
Avertissement
Pour assurer la protection
contre le risque de choc
électrique, remplacez le fusible
seulement avec le même type
et la même valeur nominale.
Warnin
To prevent electrical shock,
do not open cover.
Refer all servicing to qualified
personnel.
Avertissement
Afin d’éviter tout risque
d’électrocution ne pas enlever ce
couvercle.
Faire appel au service Après-Vente.
3
09931286E
Configuration of Peltier Temp. Control Unit (Multi)
6-Position Peltier Controlled Cell Changer (with heated reference or without heated reference)
Peltier Controller
Power Cord, 3ea
Interface Cable (RS-232 and USB to RS-232), 1each
Spare Fuse (AC 250V T5AL), 2ea
Temperature Probe, 1ea
Magnetic Stirrer, 6ea
Coolant Hose, 1ea
Coolant, 2ea
Waste Basket for Coolant, 1ea
Waste Hose for Coolant, 1ea
Phillips round head screw with washer, 2ea
Description
6-Position Peltier Controlled Cell Changer (with heated reference or without
heated reference)
a. Interface connector between 6-Position Peltier Controlled Cell Changer and Peltier Controller
b. Coolant Inlet/Outlet Quick Coupler
c. N
Gas fitting
2
d. Interface slot for Temperature Probe
e. Cell Lifting knob for sample cells
f. Cell Lifting knob for reference cell
g. Phillips round headscrew with washer
Used to fix a cell holder, a base plate or a front plate for Lambda 365.
Spare screws (2ea) are enveloped with the accessory.
Figure 3 Phillips round headscrew with washer (M4 *12L)
h. Temperature Probe
Figure 4 Temperature probe
5
Peltier Controller
Front View of Peltier Controller
Figure 5 Front view
a. Manual touch pad with Display: Temperature Control Unit (Heating/Cooling Block).
09931286E
NOTE:
NOTE:
b. LED Indicators: Display the status of the operation of Coolant Circulation, Fan, Pump, etc.
(If there is any problem in the components, Red LED will flash with beep alarm.).
c. Manual ON/OFF button: As the Manual ON/OFF button is ON, the temperature can be
controlled manually by a) Manual touch pad with Display.
Manual operation is used only for the manufacturer so it should be turned off during the measurement.
d. N2 gas ON/OFF button: It is used for using N2 gas to remove the frost from the heating
and cooling.
e. Stirrer ON/OFF button: It is used for controlling of the stirring.
f. Stirrer Speed Controlling Knob: Control the stirring speed (rpm).
g. Emergency switch: Stop all operations in the emergency situation.
During the operation, Emergency switch should be turned off.
h. Coolant Inlet
6
Rear View of Peltier Controller
Figure 6 Rear view
a. RS-232 port
b. AC Power: Main Power ON/OFF
09931286E
NOTE:
c. Fuse: AC socket + Fuse Holder
d. Interface: Interface cable for connecting with 6-Position Peltier Controlled Cell Changer
e. Air Vent Manual button: It is used for removing the air in the coolant hose.
f. Buzzer: It makes an alarm sound. If there is any problem before the measurement, Buzzer
beeps 6 times at every 10 sec. If the malfunction occurs during the operation, Buzzer beeps
twice at every 10 sec.
g. Quick Coupler of Coolant lnlet/Outlet
h. Quick Coupler of Coolant Drain
Gas Inlet/Outlet ports
i. N
2
N2 gas is not always required [Tube for N2 gas is not supplied with the accessory].
j. Flow Gauge: Indicator of the flow rate of N
gas.
2
7
09931286E
Installation
CAUTION
ATTENTION
1. Prepare the Lambda 365 and Peltier Controller in a location that is compatible with the required
environmental conditions for the operation.
2. Open the lid of the coolant inlet on top of the Peltier Controller.
Make sure the instrument is turned off while installing this accessory.
Assurez-vous que l’instrument est éteint lors de l’installation de cet accessoire.
Figure 7 Lid of coolant inlet
3. Fill up the liquid coolant.
CAUTION
Liquid coolant should be used as refrigerant. Do not use any water (tab water,
DI, etc.). If water is used as refrigerant, the Peltier Controller’s components might
be corroded and its performance may be deteriorated.
Use any kind of liquid coolant of normal grade that is available. (e.g. liquid coolant
including Distilled Water 70 wt%, Propylene Glycol 27 wt%, Additives 3 wt%).
ATTENTION
Le liquide de refroidissement doit être utilisé comme réfrigérant. N'utilisez jamais
d'eau (eau du robinet, eau déminéralisée, etc.) L'eau risque en effet de corroder
les composants du contrôleur Peltier et de réduire les performances de ce
dernier.
Utilisez un liquide de refroidissement du commerce de qualité normale (par
exemple, un liquide de refroidissement à base d'eau distillée à 70 % M/V, de
propylène glycol à 27 % M/V et d'additifs à 3 % M/V).
NOTE:
Keep the amount of liquid coolant to the level where the indicator is located between the ‘L’(low) and
‘H’(high) mark in the scale on the left side of the Peltier Controller. When the coolant is running short for
the operation, LED of Low Water Level is changed red with alarm sounding.
4. Connect the power cord and communication cable of the Lambda 365. DO NOT turn on the power
of the instrument!
8
09931286E
5. Remove the two Phillips round head screws with washer (M4*12L) to disassemble the existing cell
holder and base plate.
Figure 8 Location of Phillips round head screws with washer
6. Pull out the cell holder and base plate by hand.
Figure 9 Pulling out the cell holder and base plate
7. Insert the 6-Position Peltier Controlled Cell Changer in the sample compartment.
Figure 10 Inserting the 6-Position Peltier Controlled Cell Changer
9
09931286E
8. Gently press the 6-Position Peltier Controlled Cell Changer to connect the communication port (male)
under the bottom of the 6-Position Peltier Controlled Cell Changer to the port (female) in the
sample compartment.
Figure 11. Connecting the communication ports
9. Fix and tighten the 6-Position Peltier Controlled Cell Changer with the Phillips round head screws.
or
Figure 12 Location of screws (with heated reference or without heated reference)
10
09931286E
10. Connect the accessory interface cable of the Peltier Controller to the interface connector of the 6-
Position Peltier Controlled Cell Changer.
Figure 13 Connecting the accessory interface cable of the Peltier Controller
11. Connect the coolant inlet/outlet and N2 gas tube between the Peltier Controller and the
6-Position Peltier Controlled Cell Changer.
NOTE:
Figure 14 Connecting the coolant inlet/outlet and N
gas fitting
2
N2 gas is not always required. If no gas is applied, user does not have to connect N2 gas tube.
12. Open the lid of Interface slot for Temperature Probe by hand.
Figure 15 Opening the lid of Interface slot
11
09931286E
13. Connect the Temperature Probe to the Interface slot for Temperature Probe.
Figure 16 Connecting the temperature probe
14. Stuff the probe line into the cell compartment and close the lid of the Interface slot for
Temperature Probe.
Figure 17 Stuffing the probe line into the cell compartment
15. Connect the Peltier Controller with the PC via the RS-232 or USB to RS-232 cable.
Figure 18 Connecting the Peltier Controller with the PC
12
16. Connect the power cord to the Peltier Controller.
Figure 19 Connecting the power cord
17. Turn on the AC power switch of the Lambda 365 and the Peltier Controller.
CAUTION
The air vent manual button should be off before turning on the main power of the
Peltier Controller and also during the operation since it could make a trouble for
the communication between Lambda 365 and Peltier Controller.
ATTENTION
Le bouton d'actionnement manuel de la ventilation doit être désactivé avant que
le contrôleur Peltier ne soit mis sous tension et pendant le fonctionnement de ce
dernier afin d'éviter qu'une erreur de communication entre le Lambda 365 et le
contrôleur Peltier ne se produise
09931286E
Figure 20 Turn on the AC power switch
18. Check that the Power LED is on as a blue light.
13
Figure 21 Location of Power LED
CAUTION
ATTENTION
If you hear the alarm sound during the operation, push the emergency
switch to stop all functions immediately and check the status of Peltier
Controller carefully. Then, turn the switch to the right again when
everything is solved properly and the functions will be on again.
Si l'alarme se déclenche en cours de fonctionnement, appuyez sur
l'interrupteur d'urgence pour arrêter immédiatement toutes les fonctions
et contrôlez soigneusement l'état du contrôleur Peltier. Ensuite, tournez à
nouveau l'interrupteur vers la droite une fois la situation rétablie afin de
restaurer les fonctions
09931286E
NOTE:
Figure 22 Location of Emergency Switch
Peltier controller can be compatible with various peltier cell holders (single or Multi). Wwhenever you
exchange the existing peltier cell holder to another one, you should perform the “Auto Tuning” setup
referring to chapter Peltier Controlled Auto Tuning Setup (p33).
19. Turn off the power after the experiment.
How to Drain the Coolant From the Peltier Controller
a. Push the Air Vent Manual button to be switched on the rear panel of Peltier Controller before
operation.
b. Connect the waste hose to the Coolant Drain port on the rear panel and prepare a waste
bucket.
c. Turn on the power of Peltier Controller, and then coolant will be drained automatically.
14
d. Once draining is complete, turn off the power of Peltier Controller.
Setting USB to RS-232 Cable
09931286E
NOTE:
When using the RS-232 cable, the COM Port is set automatically.
When using the USB to RS-232 cable, USB to RS-232 Driver has already been installed when installing
the Lambda 365 and UV Express software, user does not need to install it again.
When the communication by the USB to RS-232 converter is not established, change the port setting as
follows;
1. Select My Computer →Properties.
2. Select Device Manager.
15
09931286E
3. Select Ports (COM & LPT) to expand the listing. These are the devices currently connected to the
COM ports. USB Serial Port (COMx) is listed when the driver installation is completed
successfully.
4. Double click on the USB Serial Port (COMx) of the Ports (COM & LPT).
5. Select the Port Settings tab and select the Advanced… button.
16
6. Change the parameter values as shown below.
09931286E
7. Select OK after checking the changed parameter values.
8. If the Peltier controller fails to communicate with the PC, change the COM Port Number by the
following steps.
9. Open Advanced Setting for COMx window again by repeating the steps 1 to 6.
17
09931286E
10. Select the COM Port number list to expand it and change the COM port number to another one
which is not in use from COM 1 to COM 10.
11. Make sure that the changed COM Port Number is applied, and select OK.
12. After the port setting is changed, restart the computer.
18
09931286E
Measurement
Thermal Denaturation Mode
1. Install the 6-Position Peltier Controlled Cell Changer referring to the chapter Installation.
2. Close the sample compartment cover and turn on the instrument.
NOTE:
NOTE:
NOTE:
Before System Self Test is performed, the 6-Position Peltier Controlled Cell Changer should be installed.
Otherwise, the instrument can be damaged electrically and do not operate properly.
Start the System Self Test after warming up the system for at least 20 minutes.
3. Double-click on the UV Express folder and select Thermal Denaturation mode.
4. Select OK after finishing the System Self Test.
5. Select New to open a new window and select OK.
When the 6-Position Peltier Controlled Cell Changer is connected for the first time, you need to correct
the cell position. Calibrate the beam position of the 6-Position Peltier Controlled Cell Changer referring to
the chapter Calibration ofMulti-Cell Position (p30).
6. Open the method in main menu or mode icon on the left side of the main window. Set parameters
for Experiment.
19
09931286E
a. Y Unit: Select one of the display unit: Absorbance, Transmittance, Reflectance or Energy.
b. SBW (nm): Select bandwidth. There are five bandwidths selectable: 0.5, 1, 2, 5 or 20 nm.
c. Tm Method: Select a method for determining melting temperature. Options include: 1st
derivative and Average.
d. Equation Name:
Enter the name of the formula. %G-C is entered as default, of which equation is formulated in
Equation Expression.
e. Equation Expression
The default equation for the calculation of %G-C base pairs is:
%G-C=2.44 * (Tm - 81.5 - 16.66 * log(M) + 500/K)
Where M is the molarities of salt in mol/L, K is the DNA base pair length. %G-C will be
calculated with estimated T
and input values, Molarities (M) and length of DNA (K).
m
f. DNA Pair Length (K)
Enter the DNA base pair length. This value will be used calculating %G-C.
g. Calculation Start Temperature (°C)
Enter the start temperature for calculating the T
value.
m
h. Calculation End Temperature (°C)
Enter the end temperature for calculating the T
value.
m
i. Molarity (mol/L)
Enter the salt molarities of the solution [mol/L]. This value will be used in calculating %G-C.
7. After setting parameters for Experiment and Lamp, click the Accessory tab. And select Multi-
Cell Peltier.
20
09931286E
8.Select Connection Check, then Method Setup will be activated. Select Method Setup.
NOTE:
Check the communication between Computer and Peltier Controller referring to the chapter
Setting USB to RS-232 Cable (p15) if the Method Setup is not activated.
9. The Peltier window will be shown. Select the Multi Cell Peltier tab and set parameters according
to the experiment conditions.
▶Multi-Cell holder
a. Available Cells
Indicate cells that are available for measurement. In the case of 6-Position Peltier Controlled Cell
Changer, six cells are available.
21
09931286E
b. Using Cells
Show the cell positions which are selected for measurement. Remove the cells by and
switch positions by and keys.
c. Multi Zero
If checked, zero will be measured all the selected positions. If not, zero will be only measured at
st
cell position among the using cells.
the 1
NOTE:
NOTE:
To prevent inadequate baseline (zero) measurement because of the difference in transmittance among
cuvettes or any other reasons, Multi Baseline (Multi Zero) measurement is recommended.
d. Temperature (°C): Enter the returning temperature after the experiment is finished.
e. Monitor
Select temperature that used to monitor during measurement. Options include: Block, Probe1 or
Probe2.
f. Start (°C): Enter the start temperature for the measurement.
g. End (°C): Enter the end temperature for the measurement.
h. Interval (°C)
Enter the measurement interval temperature. For instance, if you enter 5°C, the sample will be
measured from start temperature to end temperature every 5°C.
i. Rate (°C/min): Enter the heating (or cooling) temperature rate in each temperature range.
The rate has to be set under or equal to interval value.
j. Hold (min)
Enter the holding time. For example, if you set 1 min, measurement will start one
minute passed after temperature reaches at the set point.
k. Error Range(°C/min)
It shows the temperature tolerance between the sampling and monitored temperature. The
measurement will start when the temperature tolerance reaches within the set Error Range.
Example: Set the starting temperature at 20°C and error range with 1°C then, the
measurement will start when the temperature reaches between 19°C and 21°C.
▶Temperature Display
l. Select which temperature will be displayed on the panel:Block, Probe 1 or Probe 2.
The selected temperature is only displayed in real time.
m. Offset:
This function is used to adjust temperatur e. It is only used for manufacturing process, so
do not modify the values.
22
09931286E
10. After completing the parameter setup, select Apply and OK. Then the LED for ON-LINE is turned
on, and it will start heating up or cooling down to the Start temperature in the Multicell Peltier
tab.
NOTE:
11. Insert blank solutions into the both reference and sample cell holder. Select Baseline and then
baseline will be measured when it reaches the set Start Temperature.
To monitor the probe temperature, the probes should be immersed in the sample, or to use the block
temperature for monitoring, the cell lid should be closed tightly.
a. Start Temp (°C): The starting temperature of experiment.
b. SP Temp (°C): The set temperature to go to the next measurement temp.
c. Block Temp (°C): The temperature of Cell Block.
d. Probe 1 Temp (°C): The temperature of Probe 1.
e. Probe 2 Temp (°C): The temperature of Probe 2.
f. Monitor: Selected temperature is being monitored.
23
09931286E
g. Error Range (°C): It shows the temperature tolerance between the sampling and monitored
temperature. The measurement will start when the temperature tolerance reaches within the set
Error Range.
h. Hold Time: It shows the set holding time.
i. Elapsed time (s): It shows the elapsed time to reach the sampling temperature.
j. Status: It shows the status of experimental progress.
12. Remove blank solution from sample cell holder and insert sample solution into the sample cell
holder. Select Sample, then the measurement will be started after the temperature reaches the
target temperature.
NOTE:
NOTE:
To pull out the cell easily, use the cell lifting knob.
Figure 23 Pull out the cell
13. After the measurement is finished, the results are displayed in the result window. Save or print the
results as required.
Other Measurement Modes
Peltier Temperature Control Unit is available in all methods, however, unlike Thermal Denaturation, the
measurement is only performed with isothermal state
.
NOTE:
1. Install the 6-Position Peliter Controlled Cell Changer referring to the chapter Installation.
2. Close the sample compartment cover and turn on the instrument.
When executing System Self Test, the 6-Position Peltier Controlled Cell Changer has to be installed.
Otherwise, the instrument can be damaged electrically and does not operate properly.
3. Double-click on the UV Express folder and select one of the modes except Thermal Denaturation.
In this case, open kinetics mode as example.
4. Select OK after finishing the System Self Test.
5. Select New to open a new window and select OK.
24
09931286E
NOTE:
When the 6-Position Peltier Controlled Cell Changer is connected for the first time, you need to correct
the cell position. Calibrate the beam position of the 6-Position Peltier Controlled Cell Changer referring to
the chapter Calibration ofMulti-Cell Position (p30).
6. Open the method in main menu or click the mode icon on the left side of the main window. Set
parameters for Experiment.
NOTE:
If you want to know about the meaning of each parameter of method tabs and other modes more
detail, please refer to X. Measurement Modes of the UV Express Software User Guide.
7. After setting parameters for Experiment, Lamp and Kinetics, click Accessory tab. Select Multi-Cell
Peltier.
25
8.Select Connection check, the Method Setup will be activated. Select Method Setup.
09931286E
NOTE:
Check the communication between Computer and Peltier Controller referring to the chapter Setting
USB to RS-232 Cable (p15) if the Method Setup is not activated.
9. The Peltier window will be shown. Select the Multi Cell Peltier tab and set parameters according to
the experiment conditions.
▶Multi-Cell holder
a. Available Cells
Indicate cells that are available for measurement. In the case of 6-Position Peltier Controlled Cell
Changer, six cells are available.
26
09931286E
b. Using Cells
Show the cell positions which are selected for measurement. Remove the cells by and
switch positions by and keys.
c. Multi Zero, Multi Baseline
If checked, zero (baseline) will be measured all the selected positions. If not, zero (baseline) will
st
be only measured at the 1
cell position among the using cells. Multi Baseline is only available at
the Scan and Scanning Kinetics mode.
NOTE:
To prevent inadequate baseline (zero) measurement because of the difference in transmittance among
cuvettes or any other reasons, Multi Baseline (Multi Zero) measurement is recommended.
d. Temperature (°C): Enter the returning temperature for the experiment.
e. Monitor
Select temperature that used to monitor during measurement. Options include: Block, Probe1 or
Probe2.
f. Error Range (°C/min)
It shows the temperature tolerance between the sampling and monitored temperature. The
measurement will start when the temperature tolerance reaches within the set Error Range.
Example: Set the starting temperature at 20°C and error range with 1°C then, the measurement
will start when the temperature is reached between 19°C and 21°C.
▶ Temperature Display
g. Select which temperature will be displayed on the panel:Block, Probe 1 or Probe 2. The only
selected temprature is displayed in real time.
h. Offset:
This function is used to adjust temperatur e. It is only used for manfacturing process, so
do not modify the values.
10. After completing the parameter setup, select Apply and OK. Then the LED of ON-LINE is turned on,
it will start heating up or cooling down to the set temperature in the Setup Multi Cell Peltier tab.
27
09931286E
11.Insert blank solutions into the both reference and sample cell holder. Select Baseline (or Zero) and
NOTE:
then baseline (or zero) will be measured when it reaches the set Temperature.
To monitor the probe temperature, the probes should be immersed in the sample, or to use the block
temperature for monitoring, the cell lid should be closed tightly.
a. Start Temp (°C): The starting temperature of experiment.
b. SP Temp (°C): The set temperature to go to the next measurement temp.
c. Block Temp (°C): The temperature of Cell Block.
d. Probe 1 Temp (°C): The temperature of Probe 1.
e. Probe 2 Temp (°C): The temperature of Probe 2.
f. Monitor: Selected temperature is being monitored.
g. Error Range (°C):
28
09931286E
It shows the temperature tolerance between the sampling and monitored temperature. The
measurement will start when the temperature tolerance reaches within the set Error
Range.
h. Hold Time: It shows the set holding time.
i. Elapsed Time (s): It shows the elapsed time to reach the sampling temperature.
j. Status: It shows the status of experimental progress.
12. Remove blank solution from sample cell holder and insert sample solution into the sample cell holder.
Select Sample, then the measurement will be started after the temperature reached the target
temperature.
NOTE:
To pull out the cell easily, use the cell lifting knob.
Figure 24 Pull out the cell
13. After the measurement is finished, the results are displayed in the result window. Save or print
results as required.
29
Calibration of Multi-Cell Position
09931286E
NOTE:
Calibrate the beam position of the 6-Position Peltier Controlled Cell Changer when the Multi-Cell is
installed for the first time or beam position is incorrect.
1. Execute one measurement mode of UV Express software (Scan, for example) and open the Method
Setup of the Multi-Cell Peltier.
①
②
③
2.Select the Calibration tab in the Peltier setup. The following dialog box will appear.
30
3. The Functions of the Multi-Cell Calibration are shown as follows.
Command Function
MultiCell Position It shows saved steps of each cell position of the Multi-Cell.
It is used for moving Multi-Cell position as clicking buttons
09931286E
NOTE:
Multicell Move
It is used for moving Multi-Cell position using buttons
by entered step.
Multicell Calibration Perform the Multi-Cell calibration.
Stop Stop operation during Multi-Cell calibration.
Save Result It is used to save the calibrated result.
Reset Multicell Move to ‘0’ step of Multi-Cell position.
.
Although eight positions (steps) are displayed, only the first six (6) positions are used for 6-Position
Peltier Controlled Cell Changer. The last two positions remain 0 steps.
4. Click Reset Multicell to format the Multi-Cell steps. Click OK.
5. Click MultiCell Calibration. Then the following dialog box will appear. Remove all samples from the
6-Position Peltier Controlled Cell Changer. [Empty the 6-Position Peltier Controlled Cell Changer].
Click OK.
31
09931286E
6. Then the Multi-Cell calibration will start. The current process of calibration will be shown in the main
window.
7. When calibration is finished, the following box appears. Click OK.
8. Click Save Result to save the result. If the following message box appears, click OK.
32
Peltier Controller Auto Tuning Setup
09931286E
NOTE:
Peltier controller can be compatible with various peltier cell holders (Single or Multi). Whenever you
exchange the existing peltier cell holder to another one, you should perform the “AUTO TUNING” set up
to minimize the temperature fluctuation at the target temperature.
1. Push the Manual button on the control panel of the peltier controller, and then the button light is on.
2. Set any SP temperature between 30~37°C using the button.
3. Press and hold the SET/ENT button until “G.AT” is shown.
4. Push the SET/ENT button again.
33
5. Push the button (OFF ON).
6. Press the SET/ENT button, and then the MAN LED will blink.
09931286E
7. The MAN LED will be off after a few minutes.
8. Push the Manual Button (Stop) and then the light is off.
34
09931286E
Troubleshooting
Power LED is not lit on
1. Check the connection of the power cord or the fuse. The fuse is located at the rear of the
instrument.
There is risk of receiving a fatal electric shock if the fuses are replaced
with the power cord connected.
WARNING
2. Turn off and unplug the instrument.
3. Locate the fuse cover on the left rear panel of the instrument.
4. Carefully open the compartment latch where the fuse is located.
Il y a un risque d’électrocution si les fusibles sont remplacés tandis que
le cordon d'alimentation est encore branché.
Location of latch on the
fuse compartment door
5. Disconnect the fuse.
6. Replace with a new T5AL fuse (AC 250V). One spare is contained in the power module.
7. Close the compartment door.
8. Plug in the instrument and turn on.
On-Line LED is not lit on
1. Check whether the communication cable is connected tightly.
2. Check whether the Air Vent Manual button is switched off.
3. Check whether the emergency switch is off.
4. When you use the USB to RS-232 cable, change the port setting referring to the chapter
USB to RS-232 Cable (p15).
Setting
35
FAN LED is lit on with an alarm sound
1. Fan needs to be replaced.
Low Water Level’ LED blinks with an alarm sound
1. Check the coolant level and if it is lower than L (low) mark, add more coolant.
2. Restart the Peltier Controller.
Flow Sensor LED is lit on with an alarm sound
09931286E
1. Check whether the tubing is bent or whether it is connected correctly, and then restart the Peltier
Controller.
2. Push the Air Vent Manual button on and check if coolant flows properly for about one minute after
the error occurs. If the Flow Sensor LED is continuously on with the alarm sounding, contact your
PerkinElmer service representative.
Pump LED is lit on with an alarm sound
1. Check whether the tubing is bent or whether it is connected correctly, and then restart the Peltier
Controller.
2. Push the Air Vent Manual button to be switched on and check if coolant flows properly for about
one minute after the error occurs. If the Pump LED is continuously on with the alarm sounding,
contact your PerkinElmer Service representative.
Connection is failed
1. Check the Interface connector is lined properly.
2. Check whether the MANUAL RUN/STOP button is pushed on. If yes, turn off the Manual button.
36
09931286E
6-Position Peltier Controlled Cell Changer does not move
1. Check if the 6-Position Peltier Controlled Cell Changerconnector is connected to the Lambda 365.
2. Turn on and off again the Lambda 365 with installed 6-Position Peltier Contr olled Cell Changer.
Intensity is too low
1. Recalibrate if the light beam does not reach the center of the cell holder’s hole.
2. If the intensity value is still low after the recalibration, the lamps of the light source may be
deteriorated and need replacing. Contact Perkin Elmer or the nearest representative.
When 6-Position Peltier Controlled Cell Changer does not reset
1. If reset is failed, the photo interrupt switch inside the 6-Position Peltier Controlled Cell Changerneeds
replacing. Contact Perkin Elmer or the nearest representative.
When 6-Position Peltier Controlled Cell Changer is not aligned precisely
1. Recalibrate the 6-Position Peliter Controlled Cell Changer if the light beam does not reach the center
of the cell holder’s hole.
37
09931286E
38
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.