This manual is published by PerkinElmer, Inc., 68 Elm Street, Hopkinton, MA
01748 USA. Copyright 2014 - 2019, PerkinElmer, Inc and its parent, affiliated, and
subsidiary companies. All rights reserved, including but not limited to those rights
to reproduce this publication or parts thereof. Reproduction of this publication or
parts thereof or the products it describes by any means or in any form is expressly
prohibited without the written permission of PerkinElmer.
LabChip and PerkinElmer are registered trademarks of PerkinElmer, Inc. Protein
Clear and ProteinEXact are trademarks of PerkinElmer, Inc. Microsoft and
Windows are either registered trademarks or trademarks of Microsoft Corporation
in the United States and/or other countries. All other trademarks are the property of
their respective holders.
Any errors or omission which may have occurred in this publication, despite the
utmost care taken in its production, will be corrected as soon as possible, but not
necessarily immediately upon detection. PerkinElmer provides this publication “As
Is” without warranty of any kind, either express or implied, including but not limited
to the implied warranties of merchantability or fitness for a particular purpose.
Some states or jurisdictions do not allow disclaimer of express or implied
warranties in certain transactions; therefore, this statement may not apply to you.
PerkinElmer reserves the right to revise this publication and to make changes from
time to time in the content hereof without obligation of PerkinElmer to notify any
person of such revision or changes. Further, PerkinElmer may make modifications
to the product described in this manual at any time without any obligation to notify
any person of such modifications.
Proper Equipment Operation
WARNINGS
•To reduce the risk of electric shock, do not remove the cover. No user
•Use this product only in the manner described in this manual. If the equipment
AVERTISSEMENTS
•Pour réduire le risque de choc électrique, ne pas retirer le couvercle. Ce
•Ce produit ne doit être utilisé que comme décrit dans ce manuel. Si cet
serviceable parts inside. Refer to qualified service personnel if help is
required.
is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.
produit ne contient aucune pièce pouvant être réparée par l’utilisateur. Au
besoin, confier l’appareil à un réparateur qualifié.
appareil est utilisé d’une manière autre qu e celle spécifi ée par le fabricant, la
protection fournie par l’appareil peut être entravée.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Contact Us
If you have a question about a product that is not answered in this manual or online
Help, or if you need assistance with this product, contact the PerkinElmer
Technical Support Center from 8:00 A.M. to 8:00 P.M., Eastern Time, Monday
through Friday:
Before you call, have the following information available for the technical
representative:
•Product serial number
•Software version and firmware version (found by touching the Info View button
in the lower right corner of the LabChip GX Touch software)
•If applicable, the error number shown in the LabChip GX Touch software or in
the log file.
Product Service and Customer Support Plans
Preface3
Training For Your Product
PerkinElmer offers a full range of services to ensure your success. From our
original factory warranty through a comprehensive line of customer support plans,
PerkinElmer offers you Field Service Engineers and in-house Specialists who are
dedicated to supporting your hardware, software and application development
needs.
•Diagnostic servicing performed on-site by PerkinElmer field service engineers
or remotely via Technical Support
•Validation performed on-site by PerkinElmer field service engineers
•Extended use of the PerkinElmer Technical Support Center
•Software updates
•Parts, labor, and travel expense coverage
•Other customized services upon request
Contact the PerkinElmer Center for Training and Development for information
about the availability of training courses for your product:
Phone: (US Toll Free): 800-762-4000
(Worldwide): +1 203-925-4602
Fax:+1 203-944-4904
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
FCC
CE
Preface4
This device complies with part 15 of the FCC (United States Federal
Communications Commission) Rules. Operation is subject to the following two
conditions:
•This device may not cause harmful interference, and
•This device must accept any interference received, including interference that
may cause undesired operation.
This device complies with applicable CE rules and requirements.
NOTE
Changes or modifications to this equipment not expressly approved by the party
responsible for compliance could void the user’s authority to operate the
equipment.
REMARQUE
Tout changement ou modification apporté à cet instrument non expresséme nt
approuvé par l’entité responsable de la conformité peut annuler l’autorisation
d’opérer l’appareil accordée à l’utilisateur.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Table of Symbols
Table 1 contains symbols that identify particularly important information and alert
you to the presence of hazards. These symbols may appear in this manual and/or
on the product it describes.
Preface5
Table 1. Important Symbols
Symbol
Symbole
Description
Description
DANGER: An imminently hazardous situation, which, if not
avoided, will result in death or serious injury.
DANGER: Situation présentant un danger imminent qui, s’il n’est
pas éliminé, peut entraîner des blessures graves, voire la mort.
WARNING: Caution. Refer to the User’s documentation. (ISO
7000-0434B)
AVERTISSEMENT: Attention. Se reporter à la documentation de
l’utilisateur.
NOTE: A cautionary statement; an operating tip or maintenance
suggestion; may result in instrument damage if not followed.
REMARQUE: Énoncé indiquant une précaution à prendre, un
conseil de fonctionnement ou une suggestion d’entretien; son nonrespect peut provoquer des dommages à l’instrument.
Hazardous voltage; risk of electric shock. (IEC 60417-6042)
Tension dangereuse; risque de blessure par électrocution.
Crush hazard. Risk of body parts, hair, jewelry, or clothing getting
caught in a moving part. (ISO 3864)
Danger d’écrasement. Faire attention que les parties corporelles,
les cheveux, les bijoux ou les vêtements ne soient pas pris dans
une pièce mobile.
Risk of puncture injury. (ISO 3864)
Risque de blessure par piqûre.
Risk of eye injury; wear safety glasses.
Risque de lésion oculaire; porter des lunettes de sécurité.
Risk of fire. (ISO 3864)
Risque d’incendie.
Risk of poison. (ISO 3864)
Risque d’empoisonnement.
Risk of explosion. (ISO 3864)
Risque d’explosion.
Hazardous fumes.
Émanations dangereuses.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Table 1. Important Symbols (Continued)
Preface6
Symbol
Symbole
Description
Description
Laser light; avoid exposure. Risk of eye injury. (ISO 3864)
Rayonnement laser; éviter toute exposition. Risque de lésion
oculaire.
Lifting hazard. May result in injury. (ISO 3864)
Levage dangereux. Peut entraîner des blessures.
Temperature limit. (ISO 7000-0632)
Limite de température.
Use by date. (ISO 7000-2607)
Utiliser par date.
Protective ground symbol. (IEC 60417-5019)
Symbole de terre de protection.
On (power). |(IEC 60417-5007)
Marche (alimentation).
Off (power). (IEC 60417-5008)
Arrêt (alimentation).
CE compliance mark.
Marque de conformité CE.
Batch Code (ISO 7000-2492)
Catalog number. (ISO 7000-2493)
Numéro de catalogue.
WEEE symbol (EN50419:2005). Do not dispose of as unsorted
municipal waste. See the PerkinElmer website
(www.perkinelmer.com) for more information.
Signifies that the unit has passed safety tests for grounding, power
line transience, and current leakage.
Signifie que l’appareil a réussi les tests de sécurité pour la mise à
la terre, le courant transitoire de ligne d’alimentation et la perte de
courant.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Table 1. Important Symbols (Continued)
Preface7
Symbol
Symbole
Equipment
labels are color
coded:
Les étiquettes
de l’appareil
sont codées
couleur:
Description
Description
Input.
Entrée.
Output.
Sortie.
Yel low Caution, risk of danger
Red Stop
Blue Mandatory action
Green Safe condition or information
Jaune Attention, danger potentiel
Rouge Arrêter
Bleu Intervention obligatoire
Ver t Condition sûre ou informations de sécurité
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Instrument Safety
The following safety information about the LabChip GX Touch/GXII
Touch is included in this documentation. Read and review all safety
information before operating the LabChip GX Touch/GXII Touch.
•Required Training
•Chemical Safety on page 9
•Laser Safety on page 10
•Electrical Safety on page 11
•Mechanical Safety on page 13
Required Training
Ensure that all personnel involved with the operation of the
instrument have:
•Received instruction in general safety practices for laboratories.
•Received instruction in specific safety practices for the
instrument.
•Read and understood all related MSDSs.
Instrument Safety8
WARNING
Use this product only in the manner d escri bed in this ma nual. If the
equipment is used in a manner not specified by th e manufactur er,
the protection provided by the equipme nt may be impaired .
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Chemical Safety
WARNING
Some chemicals used with the LabChip GX Touch/GXII Touch are
potentially hazardous and can cause ill ness.
•Read and understand the material safety data sheet (MSDS)
provided by the chemical manufacturer before you store,
handle, or work with any chemical or hazardous material.
•Minimize contact with and inhalation of chemicals and chemical
wastes. Wear appropriate personal protective equipment when
handling chemicals (e.g., safety glasses, gloves, or clothing).
For additional safety guidelines consult the MSDS.
•Do not leave chemical containers open. Use only with adequate
ventilation, including a fume hood, if necessary.
•Check regularly for chemical leaks or spills. If a leak or spill
occurs, follow the manufacturer’s cleanup procedures as
recommended on the MSDS.
•Dispose of waste in accordance with good laboratory practices
and local, state/provincial, or national environmental and health
regulations.
Instrument Safety9
•After emptying waste containers, seal them appropriately.
•Comply with all local, state/provincial, or national laws and
regulations related to chemical storage, handling, and disposal.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Laser Safety
WARNING
BRIGHT LIGHT HAZARD. LabChip GX /GXII Touch Instruments contain
Class 3B laser diodes. The LabChip GX Touch/GXII Touch are classified as a
Class 1 device because the lasers are appropriately enclosed (embedded)
and indicated with Warning labels.
Complies with 21 CFR 1040.10 except for deviations pursuant to Laser
Notice 50, dated June 24, 2007.
Complies with IEC 60825-1: 1993, A1: 1997, A2: 2001.
•Use of controls or adjustments or performance of procedures
other than those specified herein may result in hazard ous
radiation exposure.
•NEVER remove back, side, or front panels of the instrument
while the laser is powered. Panels (which, if removed, could
lead to laser exposure) are marked with the labels sh own be lo w:
•These panels are intended to be removed for se rvice only by
qualified personnel; they are not intended to be removed during
operation or for maintenance by users. The only re movable
maintenance panel is the lower panel at the back of the
instrument, which can be removed to access the back of the
robot, if cleaning is necessary.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Electrical Safety
The LabChip GX Touch/GXII Touch is powered by a UL/CSA/VDE
approved 100-240 VAC, 50/60 Hz input, 5, 15, 24 VDC output
power supply. Additionally, the LabChip GX Touch/GXII Touch High
Voltage circuitry is current-limited to non-hazardous levels. Users
should observe the following:
WARNING
Do not open the instrument enclosure. There are no user
serviceable parts inside.
The wall outlet or the power cable connector on the back of the
instrument should be accessible after the system's installation, to
enable trained service personnel to safely disconnect power from
the system during servicing.
The computer supplied with the LabChip GX Touch/GXII Touch
instrument has internal lithium batteries. Batteries should not be
incinerated.
Instrument Safety11
Power Cord Selection
WARNING
Danger of explosion if battery is incorrectly replaced. Replace only
with the same or equivalent type recommende d by the
manufacturer's instructions.
United States and Canada
The LabChip GX Touch/GXII Touch instrument is shipped with a
NEMA 5-15 / IEC 320 power cord. If the power cord needs to be
replaced, substitute power cords must be UL Listed, Type SJT or
equivalent, minimum No. 18 AWG, 3-conductor with ground
conductor that for safety considerations should never be
disconnected or defeated. The cord’s plug to the wall must be a
three-pin grounding type connector with a NEMA 5-15P (15A,
125V) plug configuration. The cord’s connector at the unit must
conform to requirements for an EN 60 320/IEC 320 Standard Sheet
C13 connector.
The equipment is intended to be plugged into a standard NEMA 515R receptacle in the wall.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Instrument Safety12
International
All power cord sets must be approved by an acceptable, accredited
agency responsible for evaluation in the country where the power
cord set and system will be used.
Fuses
The flexible cord must be <HAR> Type H05VV-F, 3-conductor,
minimum 0.75 - 1 mm
set fittings that is, the appliance coupler and wall plug, must bear
the certification mark of the agency responsible for evaluation in the
country where it will be used. The appliance coupler must meet the
mechanical configuration of an EN 60 320/IEC 320 Standard Sheet
C13 connector for mating with appliance inlet on the system.
The LabChip GX Touch/GXII Touch instruments contain two fuses.
See Changing the Fuses on page 142 for fuse replacement
instructions. Contact PerkinElmer Technical Support (see page 3) to
order replacement fuses.
2
conductor size (230 volt input). Power cord
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Mechanical Safety
The LabChip GX Touch/GXII Touch instruments have a three axis
robot that moves quickly and can be a pinch hazard. Keep the front
door of the instrument closed and keep hands away from the plate
loading position when the robot is moving. Keep hands away from
the robot when not actively placing microplates in the instrument or
changing the ladder and buffer tubes. Robot access areas are
marked with the warning label below:
Instrument Safety13
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Introduction18
Introduction
This manual includes general instructions for using hardware with
the LabChip GX Touch software. It includes general procedures for
operating the system, maintaining the instrument, troubleshooting
hardware and software problems, and using the LabChip GxP
option to provide compatibility with 21 CFR Part 11 requirements.
DNA, gDNA, RNA, Protein, Glycan, and Protein Charge Variant
(CZE) Chip and Reagent Kits are available to run specific assays on
the LabChip GX Touch/GXII Touch. The Assay Kits include the
reagents and consumables required to run the specific assay.
Protein, Glycan, and Protein Charge Variant assays are only
supported on LabChip GXII Touch instruments.
This section contains the following information:
•Usage on page 18
•Assay User and Quick Guides on page 19
•Principles of Operation on page 20
Usage
LabChip GX Touch software is for use with LabChip GX/GXII and
GX Touch/GXII Touch instruments. LabChip GX/GXII and GX
Touch/GXII Touch instruments are for research use only and not for
use in diagnostic procedures.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Assay User and Quick Guides
Assay User Guides
Assay User Guides provide information about a specific assay.
Instructions for preparing the chip, the plate, the ladder tube, and
the buffer tube are included in the LabChip GX Touch/GXII Touch Assay User Guide for the specific assay that you are running.
Detailed information about the assays, including Specifications,
Safety Warnings, Preparation Procedures, Expected Results,
Troubleshooting, LabChip Kit Essential Practices, and Reordering
Information is also located in the LabChip GX Touch/GXII Touch Assay User Guide for the specific assay that you are running.
The current version of the Assay User Guides are available on the
PerkinElmer web site at:
www.perkinelmer.com.
Assay Quick Guides
Assay Quick Guides are included with each Assay Kit and include
instructions for preparing the reagents and chip to run an assay.
Introduction19
The current version of the Assay Quick Guides are available on the
PerkinElmer web site at:
www.perkinelmer.com.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Principles of Operation
The LabChip GX Touch assays are based on traditional gel
electrophoresis principles that have been transferred to a chip
format. The chip format dramatically reduces separation time and
provides automated sizing and quantitation information in a digital
format.
The chip contains an interconnected set of microchannels that join
the separation channel and buffer wells. One of the microchannels
is connected to a short capillary that extends from the bottom of the
chip at a 90-degree angle. The capillary sips sample from the wells
of a microplate during the assay.
Some of the channels in the chip are larger than others. The larger
channels contain buffer. During the chip preparation, the smaller
channels and some of the wells are filled with sieving gel and buffer.
Introduction20
Figure 1. Examples of DNA/RNA Chip and Protein Chip
Schematics
After the channels are filled, the chip functions as an integrated
electrical circuit. The circuit is driven by the seven electrodes in the
electrode block that contact solutions in the chip wells when the
front door is closed. Each electrode is connected to an independent
power supply that provides maximum control and flexibility.
The polymer filling the smaller channels in the chip is designed to
sieve DNA/RNA fragments or proteins by size as they are driven
through it by means of electrophoresis, similar to using agarose or
polyacrylamide gels. The sample and sieving buffers also contain a
fluorescent dye that gets brighter upon binding to double-stranded
DNA, RNA, or protein/SDS complex.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Principles of Operation (Continued)
In the chip, each sample is sipped from the microplate by negative
pressure until a sufficient quantity is loaded in the chip. The sample
is then moved electrophoretically into the central channel. As the
fragments move down the central channel, they separate by size,
and then pass the laser, which excites the fluorescent dye bound to
the molecule. The software plots fluorescence intensity versus time
and produces electropherograms for each sample (see Figure 2).
Introduction21
Figure 2. Electropherogram
The data can be viewed in a gel-like format on the Gel Tab to
achieve the appearance of a slab gel. (The colors of the gel can be
changed.)
Figure 3. Gel View
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Principles of Operation (Continued)
For DNA, RNA, Protein, and Glycan assays, quantitating the
concentration and accurately sizing each fragment are achieved by
comparing against a sizing ladder and running internal standards or
“markers” with each sample. Internal standards of known
concentration are mixed with the sample to aid in quantitation.
The amount of sample sipped into the chip depends on pH, salt
concentration, and buffer additives. The internal standards
normalize these factors so that the software can use the ratio of the
area of the curve of the standard to the unknown peak to determine
concentration directly. The internal standards lie slightly outside the
assay range so they do not interfere with analysis.
Capillary Zone Electrophoresis (CZE) is an electrophoretic
separation technique used to evaluate the charge heterogeneity of
proteins in a sample. For Protein Charge Variant assays, analytes
are separated based on their net charges, with molecules with a
higher net charge migrating faster than those with a lower net
charge.
Introduction22
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Preparing the Instrument to Run Samples23
Preparing the Instrument to Run
Samples
This section includes general instructions for preparing the LabChip
GX Touch instrument to run samples.
For assay-specific information, see the LabChip GX Touch/GXII Touch Assay User Guide for the specific assay that you are
preparing to run. For instructions on preparing the chip and plate,
see the LabChip GX Touch/GXII Touch Assay Quick Guide for the
specific assay. The current version of the Assay User Guides and
Quick Guides are available on the PerkinElmer web site at:
www.perkinelmer.com.
To prepare the instrument to run samples:
1Open the LabChip GX Touch software (see page 24).
2Clean the electrodes and the O-Rings as directed in the
LabChip GX Touch/GXII Touch Assay User Guide.
3Purge the pressure lines (see page 24).
4If necessary, calibrate the optics (see page 139) and run
Diagnostics (see page 152).
5Prepare the chip for the assay as directed in the LabChip GX
Touch/GXII Touch Assay Quick Guide.
6Insert the chip into the instrument (see page 25).
7For Protein Clear HR chips, calibrate the instrument to run
Protein Clear HR Assays (see page 26).
8If desired, prime chips before the start of a run (not available for
Protein Clear HR chips). See page 27.
9Prepare the sample plate, ladder, and buffer for the assay as
directed in the LabChip GX Touch/GXII Touch Assay Quick Guide. If your instrument is equipped with a barcode reader, see
Placing the Barcode on the Plate on page 28 to use barcodes
on the sample plates.
10 If desired, define a new plate type to use a plate other than the
ones listed in the software (see page 29).
11 Load the Sample Plate, Ladder tube, and Buffer tube into the
instrument (see page 31).
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Preparing the Instrument to Run Samples24
Opening the LabChip GX Touch Software
To open the LabChip GX Touch software:
1Touch the LabChip GX Touch icon on the Windows desktop.
2If the LabChip GxP option is installed, the Unlock Application
Window opens. Type a valid LabChip GX Touch user name and
password into the text boxes and touch the Logon button.
LabChip GX Touch user names are created in the LabChip GX
Reviewer software. (For instructions on creating LabChip GX
Touch user names, see the LabChip GX Reviewer User Manual.)
3The LabChip GX Touch Main Window opens. The Navigation
Bar displays on the left side of the LabChip GX Touch Main
Window, and the Home Window displays in the center.
Purging the Pressure Lines
The LabChip GX Touch software provides a Purge Pressure Lines
function to improve instrument performance. A positive pressure is
applied to the lines and removes any potential liquid or debris out of
the lines through the chip interface.
Purge the pressure lines at the start of each day, before inserting a
chip.
To purge the pressure lines:
1If a chip is already inserted into the instrument, remove the chip
and close the chip door (see page 41).
2Touch the Purge Pressure Lines button on the Home Window.
The Status Window opens. The purge is complete when the Run
Status displays “Purge successfully completed” and the
instrument status returns to Idle.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Inserting the Chip
To insert the chip into the instrument:
1Prepare the chip and reagents for the assay as directed in the
Assay Quick Guide. Follow the instructions carefully to properly
prepare the chip.
2Touch the Home button on the Navigation Bar.
3Touch the Unload Chip button on the Home Window. The chip
door on the front of the instrument unlatches.
4Lift the chip door up.
5Place the chip into the cutout with the sipper through the hole in
the chip holder.
6Close the chip door and press down to latch the door.
7If the chip supports multiple assays, a dialog prompts you to
select the assay that the chip will be used for. Select the desired
assay and touch the OK button.
Preparing the Instrument to Run Samples25
8If a Protein Clear HR chip is inserted into the instrument,
calibrate the instrument to run Protein Clear HR assays (see
page 26).
9If any other chip is inserted into the instrument, the chip can be
primed while preparing the first plate (see page 27).
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Preparing the Instrument to Run Samples26
Calibrating the Instrument to Run Protein Clear
HR Assays
If a Protein Clear HR chip is inserted into the instrument, the user
must calibrate the instrument to run a Protein Clear HR assay
before starting the run. Calibration is performed by processing a
known protein sample, and uses the results to precisely set the
electrode currents for that assay to obtain consistent chip-to-chip
assay results.
To calibrate the instrument to run Protein Clear HR assays:
1Insert the Protein Clear HR chip (see page 25) OR
Open the LabChip GX Touch software (see page 24) with the
chip already inserted into the instrument.
2Touch the Prime and Calibrate button on the Home Window.
The Prime and Calibrate Window opens.
3Select the plate type from the Select Plate Type drop-down list.
4Select the assay from the Select Assay drop-down list. (To
select the location of the assay file, touch the Change Assay
Folder button and select the assay folder on the Change Assay
Folder Window.)
5Touch the Select Well Containing Standard Sample button to
select the well that contains the VeriMAb
6Select the desired well from the full-size plate diagram and
touch the Done button.
7Touch the Prime and Calibrate button on the Prime and
Calibrate Window.
•If the calibration is successful, a dialog box opens indicating
that the instrument is now calibrated for the Protein Clear
HR assay. Touch the OK button.
•If the calibration fails, a dialog box opens indicating that the
calibration process failed and a Protein Clear HR assay run
cannot be performed. Touch the OK button. Perform the
recommended actions on page 160 and retry the calibration.
•If the calibration is outside the % Purity target range or if
dips from the Egrams could not be eliminated, a dialog box
opens indicating that the calibration is marginal. Touch the
Yes button to run the chip despite the issue displayed on the
dialog box. Touch No to fix the issue. Perform the
recommended actions on page 161 and retry the calibration.
sample.
8To retry the calibration, select the Skip Prime Step check box
on the Prime and Calibrate Window. The chip is calibrated but
not primed.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Preparing the Instrument to Run Samples27
Priming the Chip Before the Run
If a new chip that is not a Protein Clear HR chip has been inserted
into the instrument, the chip is automatically primed at the start of
the first run.
Since priming is a lengthy process, the chip priming can be started
before starting the run. (To save time, the sample plate can be
prepared while the chip is priming.)
To prime a chip before starting the run:
1Insert the chip (see page 25).
2Load the buffer tube (see page 31). If running a test ladder, load
the ladder tube at the same time.
3Touch the Home button on the Navigation Bar.
4Touch the Prime button on the Home Window. The Prime
Window opens.
5On the Prime Window, select the assay for which to prime the
chip on the Assay drop-down list. (To select the location of the
assay file, touch the Change Assay Folder button and select
the assay folder on the Change Assay Folder Window.)
6If desired, select the Run Test Ladder after Prime check box to
run one ladder after the prime is complete.
7Touch the Prime button. The Prime Window displays the time
left until the prime is complete.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Preparing the Instrument to Run Samples28
Placing the Barcode on the Plate
For instruments equipped with a barcode reader, barcodes can be
placed on the plate before loading the plate into the instrument.
Figure 4 shows the size limits for the barcode label and the location
on the microplate where the label should be placed. The barcode
must be located on the short (portrait) end of the microplate, closest
to well A1. If the barcode is not positioned properly, the barcode
reader will not be able to read the barcode.
Figure 4. Barcode Label Position
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Preparing the Instrument to Run Samples29
Defining a New Plate Type
To use a plate other than the plates specified on the Setup Run Tab,
create a new plate on the Plate Editor Tab. Use caution adding new
plates. Entering wrong values on the Add Plate Window can result
in damaged chips and broken sippers.
It is best to use the plate specifications provided by the plate
manufacturer. If the specifications are not available, measure the
plate with a caliper. Many plates have a large variation in the Z-axis
location of the well bottom.
Make sure the Sip Height has enough margin to accommodate this
variation:
•Minimum Sip Height is 2.5mm.
•Maximum Sip Height depends on the sample volume.
•Sip Height of 4mm is probably safe.
To add a new plate:
1Touch the Tools button on the Navigation Bar.
2Touch the Plate Editor button to open the Plate Editor Tab.
3Touch the Custom Plates Tab.
4Touch the Add Plate button to open the Add Plate Window.
5Type the settings for the new plate. The diagram on the Add
Plate window shows the location for each measurement.
Acceptable plate parameters are:
•PlateHeight: 0 to 16mm
•WellDepth: 0 to 36mm
•SipHeight: 2 to 36mm
•(WellDepth – SipHeight) <= PlateHeight
6Touch the OK button.
7Perform a Punch Test from the Plate Editor Tab to verify the new
plate settings are acceptable:
•Cover the corner wells on the plate with a piece of the tape
provided by PerkinElmer.
•Use an old, expired chip with a good sipper.
•On the Plate Editor Tab, select the new plate name and
touch the Verify Plate button.
•Follow the onscreen instructions.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Preparing the Instrument to Run Samples30
Defining a New Plate Type (Continued)
•The instrument will move to the specified number of wells in
each corner of the plate, punching holes in the tape.
•The Punch test is acceptable if the punched hole is >0.5mm
from the edge of the well.
•Holes may not be perfectly centered.
Possible sources of errors:
•Sipper splay
•Plate tolerances
•Robot alignment error
•If all holes are too close to the edge (<0.5mm) in the same
direction, adjust the X-Margin or Y-Margin values on the Add
8Select the new plate name on the Setup Run Tab when starting
Plate window.
the run.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Preparing the Instrument to Run Samples31
Loading the Plate, Ladder Tube, and Buffer Tube
To load the Sample Plate, Ladder tube, and Buffer tube into the
instrument:
1Prepare the samples and load the samples into the plate as
instructed in the Assay Quick Guide. Follow the instructions
carefully to properly prepare the plate.
2Fill the Ladder and Buffer tubes as instructed in the Assay Quick
Guide.
3Touch the Home button on the Navigation Bar.
WARNING
Pinch Hazard. Keep hands away from plate carrier when carrier is
moving.
4Touch the Unload Plate button on the Home Window. The plate
carrier on the left side of the instrument moves out to the loading
position.
5Place the sample plate onto the plate carrier. Place the Ladder
tube and Buffer tube into the plate carrier.
6Touch the Load Plate button. The plate carrier retracts into the
instrument.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Running an Assay32
Running an Assay
This section includes general instructions for using the LabChip GX
Touch hardware and software to run an assay:
To run an assay:
1Complete the steps in the Preparing the Instrument to Run
Samples section.
2Define the parameters of the assay run (see page 33).
3Monitor the run (see page 37).
4If desired, edit the analysis parameters during the run (see
page 38).
5If necessary, stop the run before it completes (see page 39) and
continue the stopped run if desired (see page 40).
6Remove the Plate, Buffer Tube, and Ladder Tube (see page 41).
7Unload the chip (see page 41). Store the chip according to the
instructions in the Assay User Guide.
8If necessary, wash the chip (see page 42) immediately before
running the chip again.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Defining the Run Parameters
To define the run parameters:
1Touch the Home button and then touch the Run button.
If the maximum number of samples for the chip have already
been run, a message displays instructions for resolving the
problem.
The Select Wells Tab opens.
2Select the assay to run in the Select Assay drop-down list. If
the desired assay is not displayed, touch the Change Assay Folder button and select the assay folder in the Change Assay
Folder Window.
3Select the plate type in the Select Plate Type drop-down list. To
use a plate that is not listed in the Plate Name list, see Defining
a New Plate Type on page 29.
Running an Assay33
4To use an existing run file to specify the run settings, touch the
Import Run File button, navigate to the run file, select the
desired file, and touch the Open button. The settings specified
in the run file display in the Select Wells Tab and Setup Run
Tab . To use the settings in the run file, touch the Next button
twice and skip to step 26. For more information about run files,
see page 54.
5To sample the wells in the order in which the wells are selected
in the plate diagram, select the Selection Sip Order check box
before selecting the wells.
6On the Plate diagram, select the wells to be sampled.
•If desired, touch the Full Size button to enlarge the Plate
diagram to better view and select the desired wells.
•To select or clear all rows on the plate, touch the Select All Rows button at the top left corner of the plate.
•To select or clear all wells in a row, touch the row letter on
the left side of the plate. (Clear a selected row by touching
the row letter.)
•To select or clear a single column, touch the column number
at the top of the plate. (Clear a selected column by touching
the column number.)
•To select or clear all wells in a quadrant, touch the 1, 2, 3, or
4 button.
•To select or clear single wells, touch the well.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Running an Assay34
Defining the Run Parameters (Continued)
•To select or clear adjacent wells, touch and drag over the
desired wells.
•To use a Sample Names file to select the wells, see
Selecting Wells using a Sample Names File on page 57.
7If desired, select the Column-wise check box to sample the
plate in column order. If not selected, the plate is sampled in row
order.
8If desired, select the Advanced Settings check box.
•To repeatedly run the selected wells and combine the data
into one data file, touch the + or - button under Sample Sips
to select the number of times to sample the selected wells.
•To perform the assay multiple times on the same plate,
touch the + or - button under Plate Cycles to select number
of times to run the assay.
•To randomly sample a specific percent of the selected wells,
touch the + or - button under Random Select% to specify
the percent of wells to sample during the run.
9Touch the Next button . The Setup Run Tab opens.
10 If desired, type the name of the operator in the Operator text
box. (If the LabChip GxP option is installed, the name of the
current user is automatically displayed and cannot be edited.)
11 For ProteinEXact chips that have been primed before the run
(see page 27), select the Skip Warm and Skip Prime check
boxes to avoid repeating the priming and warming process.
12 To read the plate barcode (if your instrument is equipped with a
barcode reader), select the Read Plate Barcode check box.
(The plate barcode can be used as part of the data file name if
selected.)
13 To change the Data Path, touch the Browse button, select the
desired location for the data file in the Browse for Folder
window, and touch the OK button. If the LabChip GxP option is
installed, the data file must be located in the CDR. To restore
the default path, touch the Default button.
14 Select the Copy to check box to automatically create a copy of
the data files. Touch the Browse button, select the desired
location for the data files in the Browse for Folder window, and
touch the
Default button.
OK button. To restore the default path, touch the
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Running an Assay35
Defining the Run Parameters (Continued)
NOTE
If the LabChip GxP option is not installed, data files should be
saved to a local folder on the computer's hard drive. Saving data
files to a network drive may cause loss of data if the network
connection is slow or interrupted. If the LabChip GxP option is
installed and the network connection is interrupted, data files are
archived in a local folder and copied to the CDR when the ne twork
connection is restored. For more inf ormation ab out the GxP option,
see page 62.
15 If desired, select the Create Daily Sub-Directory check box to
create a new sub-directory for data files each day.
16 To automatically export data tables, graphs, or gels, select the
Auto Export check box. (See Selecting the Auto Export
Settings on page 46 for more information about selecting data to
export.)
17 To automatically print the exported data to a PDF file after the
run is complete, select the Auto Print check box.
18 To export all data at the end of the run rather than as each well
is completed, select the Defer Export to Plate Completion
check box.
19 To add the File Prefix, Project Name, Computer Name,
Barcode, Date, and/or Time to the data file name, select or
clear the desired check boxes or type the desired File Prefix and
Project Name. The file name components are added to the data
file name in the order in which they are selected. Leave the text
box blank to omit the File Prefix or Project Name from the file
name.
20 To use a Sample Names file to supply the sample names:
aSelect the Sample Names file check box.
bTouc h the Browse button next to Sample Names file.
cSelect the name of the CSV file that contains the sample
names, and touch the Open button. The path and name of
the file displays in the text box. (See Sample Names File on
page 177 for more information.)
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Running an Assay36
Defining the Run Parameters (Continued)
21 To use the Sample Names file to select the wells to read, select
the Use Sample Names File for Sample Selection check box.
(For details, see page 57).
22 To use a file to supply the Expected Peaks:
aSelect the Expected Peaks File check box.
bTouch the Browse button next to Expected Peaks File.
cSelect the name of the GEP file that contains the expected
peaks, and touch the Open button. The path and name of
the file displays in the text box. (See Expected Peak File on
page 170 for more information.)
23 To use a file to supply the Excluded Peaks:
aSelect the Excluded Peaks File check box.
bTouch the Browse button next to Excluded Peaks File.
cSelect the name of the GEP file that contains the excluded
peaks, and touch the Open button. The path and name of
the file displays in the text box. (See Excluded Peak File on
page 170 for more information.)
24 Touch the Next button . The Start Run Tab opens.
25 For ProteinEXact runs, type the Total Ladder Concentration of
the reagent into the text box.
26 Review the selections displayed. To change any selections, use
the Back and Next buttons to make the desired changes.
27 Touch the Start button to start the assay. The Status Window
displays the status of the run.
28 After the run has started, you can:
•Change the exported properties in the Well Table or Peak
Table (see page 47).
•Edit the sample names (see page 60).
•Edit the analysis settings (see page 38).
•View the raw data from the run (see page 49)
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Monitoring the Run
After a run is started in the LabChip GX Touch software:
1The priming and warming steps are performed.
•The priming step fills the channels of the chip with reagent.
(Only performed if the chip holder has been opened since
the last run and the chip has not been primed.)
•The warming step allows the heater plate in the chip holder
to regulate chip temperature to 30°C.
•Data collection begins after the priming and chip warming
steps are completed.
2For ProteinEXact chips, the calibration steps are performed.
3The Status Window displays the Chip and Run Status, the plate
diagram, and an electropherogram and gel of the well currently
being read.
4For ProteinEXact chips, the Status Window displays the
calibration status of the calibration step currently being
performed instead of the plate diagram (see page 37).
Running an Assay37
Figure 5. ProteinEXact Status Window
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Monitoring the Run (Continued)
NOTE
If the ProteinEXact calibration steps do not display, restart the
software and prime the chip.
5Data is saved to a GX Touch data file (*.gxd) with the name
shown in the Setup Run Tab after each well is completed.
6To view the results for individual wells as data is acquired or
after the run is finished, see page 49.
7If analysis settings are changed during the run (see page 38),
the analyzed data uses the current analysis settings if Perform
Sample Analysis is selected on the Select Wells Tab.
8To stop the run before it is complete, see page 39.
9When the assay is complete, Successfully Completed displays
in the Status line on the Status Window.
Running an Assay38
10 Remove the plate (see page 41) and/or remove the chip (see
page 41) as necessary.
Editing the Analysis Settings during a Run
Specific analysis settings can be adjusted in the LabChip GX Touch
software while a plate is running. If Auto Export is selected on the
Setup Run Tab, the Analysis Settings cannot be changed unless the
Defer Export to Run Completion check box is selected.
After a run is complete, the analysis settings can be changed to
view the data, but the analysis settings cannot be saved.
To adjust the analysis settings:
1After the run is started, touch the Views button and then the
Analysis Settings button. The Analysis Settings Tab opens.
2Change the Analysis Settings as desired. Press Enter or Tab, or
move to another text box, to enable the Apply and Cancel
buttons. (See Analysis Settings Tab for a description of each
analysis setting.)
3Click the Apply button to apply the settings to the current run.
The initial analysis settings are saved as Version 0. If changes
are made during the run, the analysis settings at the end of the
run are saved as Version 1. Use LabChip Reviewer to view the
analysis settings versions.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Running an Assay39
Stopping a Run before the Run is Complete
The run is completed automatically after reading the last selected
well. If you need to stop the run before it is complete, touch the
Stop button on the Navigation Bar. The run stops immediately. If a
well is in progress, the data for that well is not saved.
A message box confirms that you want to stop the run in progress.
Figure 6. Stop Run Message
Tou ch Yes to stop the run. The Status line on the Status Window
displays stopped by user as shown in Figure 7.
Data for any completed wells displays in the Views Window.
To continue to read wells from an aborted run, see Continuing a
Stopped Run.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Figure 7. Stopped by User
Continuing a Stopped Run
If a run is stopped before it is complete, you can start a new run to
read the remaining unread wells in the plate. When selecting the
wells for the assay, select only the wells that were not read so that
the run starts with the well that was not completed. The data is
saved in a new data file. It is not added to the data file from the
original run.
To continue a run:
1Touch the Run button on the Home Window.
2On the Select Wells Tab, select the wells that were not read in
the previous run, beginning with the well that was in progress
when the run was stopped.
3On the Setup Run Tab, select the same options for this run as
were selected for the stopped run.
4On the Start Run Tab, touch the Start button to begin.
Running an Assay40
The new run starts with the well that was not completed in the
previous run. A separate data file is created for the current run.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Running an Assay41
Removing the Plate, Buffer Tube, and Ladder
Tube
To remove the plate, Buffer tube, and Ladder tube from the
instrument when the run is complete:
1Touch the Home button on the Navigation Bar.
WARNING
Pinch Hazard. Keep hands away from plate carrier when carrier is
moving.
2Touch the Unload Plate button on the Home Window. Wait until
the plate carrier is fully extended.
3Remove the plate from the plate carrier.
4If necessary, clean the plate carrier with DI water or 70%-
isopropanol solution.
5Remove the Buffer and Ladder tubes if necessary.
WARNING
Pinch Hazard. Keep hands away from plate carrier when carrier is
moving.
6Touch the Load Plate button on the Home window to retract the
Removing the Chip
To remove the chip from the instrument:
1Touch the Home button on the Navigation Bar.
2Touch the Unload Chip button on the Home Window. The chip
3Lift the chip door up.
4Remove the chip from the chip holder.
plate carrier.
door on the front of the instrument unlatches.
5Close the chip door and press down to latch the door.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Washing the Chip
Some protein samples may have components which produce data
with extra peaks, spikes or other artifacts. When these artifacts are
present, washing protein chips on the LabChip GXII Touch
immediately before the next use can often restore data quality.
1Prepare the chip and buffer tube as instructed in the Assay User
Guide.
2Insert the prepared chip and buffer tube into the instrument (see
page 25).
3Touch the Home button on the Navigation Bar.
4Touch the Wash button on the Home Window.
5When the wash is complete, follow the instructions in the Assay
User Guide to run the chip.
Running an Assay42
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Exporting Data43
Exporting Data
The Peak Table, Well Table, Gel, Graph, or raw data can be
exported automatically at the end of a run or after each well is read.
Select the desired settings in the Export Settings Tab (see page 47)
and then select the Auto Export check box on the Export Setup
Tab . After the run has started, change the exported properties in the
Well Table or Peak Tables if desired (see page 47).
Data can also be exported after a run is complete (see page 48).
Peak Tables and Well Tables are exported to either CSV files or
XML files. CSV files can be imported into a spreadsheet program
such as Microsoft
Raw Data can be exported to a CSV file, an XML file, or to an AIA
file format (a CDF file) which is used by some graphical analysis
software tools.
Gel and Graph data is exported to the selected image format (i.e.,
BMP, GIF, JPEG, PNG, TIFF, WMF, or EMF).
®
Excel.
This section includes the following information for using the GX
Touch software to export Peak Tables, Well Tables, Gels, Graphs,
or raw data:
•Export Examples on page 44
•Selecting the Auto Export Settings on page 46
•Selecting the Exported Properties in the Well Table or Peak
Table on page 47
•Exporting Data in AIA Format for Use in Third-Party Analysis
Software on page 47
•Exporting a Data File after a Run is Complete on page 48
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Export Examples
This section shows export examples for the following data:
•Peak Table
•Raw Data
•Gel on page 45
Peak Table
Either a CSV file (ASCII text file in a comma-separated format) or
an XML file that contains the data in all columns in the Peak Table.
The peak table can be viewed in the LabChip GX Reviewer
software. Figure 8 is an example of part of a Peak Table CSV file
exported from a DNA assay (data truncated for this example):
Exporting Data44
Figure 8. Exported Peak Table
Excluded Peaks are not exported and are missing in the exported
file. For example, if peaks 3, 5, and 7 are excluded, when the data
is exported into Microsoft
included.
Raw Data
Either a CSV file (ASCII text file), an XML file, or an AIA Format file
that contains the signal data from the run as one file per well or
multiple wells in the same file. Exported data has been smoothed
using the polynomial filter.
Raw data can be exported in a Chromatography Data Interchange Format (formerly AIA format), which is used by some
graphical analysis software tools. The Include Size Data and Export
Single Table options are not available with Chromatography Data
Interchange Format.
In addition to exporting time and value information, you can choose
to export Size information. This information is determined based on
aligned data and is used to correlate the peaks from one row to
another.
®
Excel, peaks 3, 5, and 7 are not
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Export Examples (Continued)
Below is an example of part of a raw data CSV file, including size
data, exported from a DNA assay. In this example, the header is not
included and data is truncated.
DATA
Time,Value,Size
0.000,-1.097,0
0.017,-1.512,0
0.033,-1.913,0
...
12.933,19.774,14
12.950,19.836,16
12.967,20.165,19
12.983,20.761,21
13.000,21.384,23
13.017,21.498,25
…
Exporting Data45
Gel
Exports the gels in the selected image format. Options are available
to export all gels into the same image file or into a separate file for
each gel. The Height selector on the Export tab determines the
height, in pixels, of the exported graphic.
Figure 9 shows all gels exported to one image file.
Figure 9. Exported Gels
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Selecting the Auto Export Settings
The Auto export settings specify the data to automatically export at
the end of each run and specify the format for each type of data.
The settings in the export tab are only used if the Auto export
check box is selected in the Setup Run Tab.
To select the data to export:
1Touch the Tools button on the Navigation Bar and then touch the
Export Setup button to open the Export Setup Tab.
2Touch the Export Settings Tab.
3To export the files to a location other than the data folder, touch
the Browse button and select the desired location for the data
files. To reset the location to the default, touch the Default
button.
4Select the desired check boxes under Export Selections:
•Select the Export All check box to select all check boxes.
•To export all views of the same type to a single data file, the
Export All check box and select the Export as Single File
check box. If selected, the gels for all wells are included in a
single graphic file and the Raw Data for all wells is included
in a single AIA, CSV, or XML file.
•If Gel is selected, touch the + or - buttons next to Height
(Pixels) to specify the desired height of the exported gel
graphic.
•If Raw Data is selected, touch the AIA Format check box to
export in Chromatography Data Interchange Format or clear
the check box to export in CSV or XML format.
•Select Include Size Data to align the data to the well's
ladder (for one file per well) or to the first well (for a single
data file) and include the size data in the exported Raw data
file. (Not available for CZE assays or if the AIA Format check
box is selected.)
•If Peak Table, Well Table, or Raw Data is selected for
export, choose the desired format for the exported data files.
•If either Electropherogram or Gel is selected, choose the
desired format for the exported image files.
Exporting Data46
5To uc h Save to save the Export settings.
6Start the Run (see Running an Assay on page 32).
7After the run has started, change the exported properties in the
Well Table or Peak Table if desired (see page 47
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
).
Exporting Data47
Selecting the Exported Properties in the Well
Table or Peak Table
The properties that are included in the Exported Peak Table or Well
Table files can only be selected after the run is started.
NOTE
The desired properties should be selected during the W arm or Load
step to ensure that all properties are exported for all wells if Single
File is selected.
To change the exported data:
1Touch the Tools button on the Navigation Bar.
2Touch the Export Setup button to open the Export Setup Tab.
3To change the Peak Table or Well Table export properties:
aTouch the Peak Properties Tab or Well Properties Tab.
bMove the desired properties into the exported list by
selecting the properties and using the Left or Right arrow
buttons to move the properties into the desired list.
4The changes take effect with the next file that is exported.
Exporting Data in AIA Format for Use in ThirdParty Analysis Software
To export data in AIA format for use in third-party analysis software:
1Touch the Tools button on the Navigation Bar and then touch the
Export Setup button to open the Export Setup Tab.
2Touch the Export Settings Tab.
3To export the files to a location other than the data folder, touch
the Browse button and select the desired location for the data
files. To reset the location to the default, touch the Default
button.
4Select the Raw Data check box under Export Selections.
5Select the AIA Format check box under Raw Data Options.
6To uc h Save to save the Export settings.
7Start the Run (see Running an Assay on page 32).
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Exporting Data48
Exporting a Data File after a Run is Complete
To export a data file after a run is complete:
1If the data file is not open, open the data file (see page 53).
2Touch the Views button on the Navigation Bar.
3Touch the Select Wells button to open the Select Wells to View
Tab .
4Touch the Export Plate button to export a data file to using the
settings in the Export Settings Tab on the Views Window. If
Peak Table or Well Table are selected for export, the exported
table columns are selected in the Export Setup Tab on the Tool s
Window.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Viewing the Data49
Viewing the Data
Data can be viewed in the Status Window or in the Views Window.
View data for the current run while the run is in progress, or after
the run is complete.
The Status Window displays the raw data from the run.
The Egram/Gel Tab, EGram Tab, and Gel Tab on the Views Window
can display either analyzed or raw data.
This section includes the following procedures for viewing data in
the LabChip GX Touch software during a run:
•Selecting the Well Data to View on page 49
•Viewing Data in the Egram/Gel Tab on page 50
•Viewing Graphs in the Egram Tab on page 52
•Viewing Data in the Gel Tab on page 52
•Zoom In and Zoom Out on page 52
•Viewing a Data File after a Run is Complete on page 53
NOTE
After a run is complete, use the LabChip GX Reviewer software to
view and analyze the plate data. The LabChip GX Review er
software can open multiple data files in the same workspace to
compare data from different plates or different runs.
Selecting the Well Data to View
To select the well data to view in the graphs on the EGram/Gel tab
and the EGram tab:
1Touch the Views button on the Navigation Bar.
2Touch the Select Wells button to open the Select Wells to View
Tab .
3To view data from a single well:
aClear the Overlay check box.
bTouch the well in the plate diagram. Only one well can be
selected at a time.
4To view data from multiple wells overlaid on the same graph:
aSelect the Overlay check box.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Viewing the Data50
bTouch each well to view in the overlay graphs. Multiple wells
can be selected at the same time. Touch a well again to
deselect the well.
5To display the analyzed data on the Egram/Gel Tab, EGram Tab
and Gel Tab, select the Perform Sample Analysis check box.
To display the raw data only, clear the Perform Sample
Analysis check box.
6Use the gels in the Gel Tab or the Egram/Gel Tab to select the
wells if there are multiple cycles. Touching a well in the Select
Wells to View Tab selects the well data from the last plate cycle.
7To view the data in the Egram/Gel Tab, see page 50.
8To view the graphs in the EGram Tab, see page 52.
9To view the data in the Gel Tab, see page 52.
10 See Zoom In and Zoom Out on page 52 for details on zooming
graph and gel views.
Viewing Data in the Egram/Gel Tab
The Egram/Gel Tab displays a graph at the top of the tab and the
gels of all the well data at the bottom of the window. The graph at
the top of the window can display a single well or multiple wells
overlaid in the same graph. Gels for all wells in the run display at
the bottom of the window.
The wells selected on the Select Wells to View Tab display in the
Egram/Gel Tab.
The wells to view can also be selected using the gels on the bottom
of the Egram/Gel tab:
1Select or clear the Overlay check box.
2Touch a gel at the bottom of the tab to add the graph to the
overlay. If the overlay check box is selected, the most recently
added well displays at the top of the overlay graph. If the
Overlay check box is not selected, only one well can be
selected.
3To change the graph display options, select the desired options
in the Graph Settings on the EGram Tab.
4To change the gel display options, select the desired options in
the Gel Settings on the Gel Tab.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Viewing the Data51
Graph Overlay
The graph in the Egram/Gel Tab can display multiple wells overlaid
in the same graph for visual comparison. Each well is shown in a
different color and line style with a legend at the top of the window.
Figure 10 shows the electropherograms of two wells in the same
graph.
Figure 10. Graph with Multiple wells
To remove a specific sample
that you want to remove in the Select Wells to View Tab or
Egram/Gel Tab.
To display only one sample
box and touch one sample in the Select Wells to View Tab or
Egram/Gel Tab.
Overlay Offset
Use the Overlay Offset text box on the Graph Settings to offset
each of the graphs by the RFU value specified. Figure 11 shows the
electropherograms of two wells in the same graph, offset by 50
RFU.
from the graph, touch the sample
in the graph, clear the Overlay check
Figure 11. Overlay Offset
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Viewing Graphs in the Egram Tab
The EGram Tab displays a graph containing the data from one or
more completed wells. Data from multiple wells can be displayed in
the same graph for visual comparison. The data file name and well
name display at the top of the graph.
To navigate through the wells on a plate, touch the Left, Right, Up,
and Down buttons. The Select Wells to View Tab and the
Egram/Gel Tab update to display the current well.
Select the Scale to Sample Peaks check box to scale the view to
the minimum and maximum X values of the current sample peaks.
Marker and/or system peaks are ignored. (If Overlay is selected,
the view scales to the sample peaks in the most recently selected
well.)
Clear the Scale to Sample Peaks check box to scale the view to
the minimum and maximum X values of all peaks, including marker
peaks and system peaks.
Viewing the Data52
To change the graph view options, select the Graph Settings check
box and select the desired options in the Graph Settings.
Viewing Data in the Gel Tab
To compare the gels of all wells, view the gels in the Gel Tab.
The color, width, and contrast of the gels can be changed using the
Gel Settings.
Select the Overlay check box to select multiple gels. The selected
wells display in the graphs in the Egram/Gel Tab and the Gel Tab.
Zoom In and Zoom Out
You can zoom in and out on data displayed in the Egram/Gel Tab,
Gel Tab, and EGram Tab. The Graph and the Gel on the Egram/Gel
tab both zoom to the same levels when either view is zoomed in. All
tabs also zoom to the same levels.
To zoom in:
•Touch and drag on a graph or gel to enclose the region of
interest. The selected area enlarges to fill the view. In the Gel
tab, all lanes zoom to the same level. In a Gel, drag straight up
or down in a single gel lane to zoom in. Dragging horizontally
pans the view of the gels.
•You can continue zooming in until you reach the maximum
magnification (the graph will not zoom in any closer).
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Viewing the Data53
To zoom out:
•In the EGram tab, touch the Unzoom button to zoom out to the
previous zoom level. Touch the Unzoom All button to zoom out
to the default view.
•In the Gel tab or the Egram/Gel tab, double-tap a gel header to
zoom out to the previous zoom level.
•In the Egram/Gel tab or the EGram tab, double-tap the graph to
zoom out to the previous zoom level.
Viewing a Data File after a Run is Complete
To view a data file after a run is complete:
1Touch the Views button on the Navigation Bar.
2Touch the Select Wells button to open the Select Wells to View
Tab .
3Touch the Review Plate button to open a data file to view the
data. If a data file is already open, touch the OK button in the
Review Plate window to clear the current plate data. Select the
desired data file to open in the Select a Data File window and
then touch the Open button.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Creating a Run File54
Creating a Run File
A Run File contains all of the settings required to run a plate,
including all the tab settings on the Run Window. Importing a Run
File on the Select Wells Tab changes the settings on all tabs.
To create a run file if desired:
1If the run will include auto export data, touch Tools, touch the
Export Setup Tab, and then select the desired data to export.
See Selecting the Auto Export Settings on page 46 for details.
2Touch the Home button and then touch the Run button.
3Select the desired assay, plate and sampling options and select
the wells to sample on the Select Wells Tab. See Running an
Assay on page 32 for details.
4Touch the Next button to open the Setup Run Tab.
5Select the desired run options on the Setup Run Tab.
6Touch the Next button to open the Start Run Tab.
7Touch the Export button . The Save Run Parameters
to File window opens.
8Type the desired name for the run file, select the file location,
and touch the Save button to save the run file.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Using Sample Names Files55
Using Sample Names Files
If desired, Sample Names files can be used to select wells to read.
Sample Names files are also used to import sample names into a
data file as the data file is created.
This section includes general instructions for using Sample Names
files in the LabChip GX Touch software:
•Creating a Sample Names File on page 56
•Selecting Wells using a Sample Names File on page 57
•Using a Barcode to Specify the Sample Names File on page 59
•Editing Sample Names During a Run on page 60
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Creating a Sample Names File
Sample Names Files created in the LabChip GX Touch software can
be used to specify the desired sample names for each well in a
plate. The Sip order can also be specified to match the order of
wells in the Sample Names file. Sample Names Files created in the
LabChip GX Touch software do not contain Expected Peaks.
Existing Sample Names files can also be edited in the LabChip GX
Touch software by importing the existing file, editing the sample
names and User Comments, and then exporting the edited file.
To create a new Sample Names file:
1When a plate is not running, touch the Views button on the
Navigation Bar.
2Touch the Sample Names button to open the Sample Names
Tab .
3Touch the Blank 96 Wells or Blank 384 Wells button to open a
blank Sample Names file with 96 or 384 rows. The default
Sample Names for each well are the same as the well label.
Using Sample Names Files56
Or, to open an existing Sample Names file, click the Import
button and select the desired Sample Names file (*.csv).
4Double-tap a sample name and edit the sample name as
desired. The Well Labels cannot be changed and rows cannot
be added or deleted.
If desired, the wells can be displayed in Row Order or Column
Order using the Row Order and Column Order option buttons.
Select the By Quadrant check box to display the wells by
quadrants.
5If desired, type comments for the samples in the User
Comments column.
6To save the Sample Names file, touch the Export button and
specify the desired location and file name. The wells are saved
in the same order as they are displayed in the table.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Using Sample Names Files57
Selecting Wells using a Sample Names File
The wells to be sampled during the run can be selected using a
Sample Names file. When using a Sample Names file to select the
wells, only wells with an entry in the Sample Names file are
selected for sampling. Use the LabChip GX Touch software,
LabChip GX Reviewer software, Excel, or a LIMS to create the
Sample Names file. The format must match the file format
described in the LabChip GX Reviewer User Manual if the file is
created with Excel or a LIMS. Sample Names files that are created
in LabChip GX Touch software (see page 56) always contain all 96
or 384 wells on the plate. Existing Sample Names files containing
less than 96 or 384 wells can be edited but not created in the
LabChip GX Touch software.
The Sample Names file used to select the samples is a .csv file that
contains one row for each well to be sampled, and does not contain
rows for wells that are not sampled.
Using the Sample Names file in Figure 12 to select the wells in the
run will select wells A1, A5, A11, A12, B1, B2, F5, and F6 for
sampling. Figure 12 shows the wells selected in the Select Wells
Tab .
Figure 12. Sample Names file for Selecting Samples
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Using Sample Names Files58
Selecting Wells using a Sample Names File
(Continued)
Figure 13. Selected Wells on Run Window
This Sample Names file cannot be created in the Sample Names
Editor window in the LabChip GX Reviewer software because the
files created in the Sample Names Editor window always contain
one row for each well in the plate. The Sample Names files for
selecting wells are typically created by a LIMS or can be created in
Microsoft
To select the wells using a Sample Names file:
1Create the Sample Names file with one row for each well to be
2Touch the Home button and then touch the Run button.
3Select the desired run settings on the Select Wells Tab.
4Select at least one well in the plate diagram. It does not matter
5Touch the Next button or the Setup Run Tab.
6Select the Sample Names file check box.
7Touch the Browse button next to the Sample Names text box,
®
Excel and are saved as a *.csv file.
sampled.
which well is selected. The Next button is not enabled unless at
least one well is selected.
select the Sample Names file (*.csv), and touch the Open button.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Using Sample Names Files59
Selecting Wells using a Sample Names File
(Continued)
8Select the Use Sample Names file for Sample Selection
check box.
9Select the desired options on the Setup Run Tab.
10 Touch the Select Wells button.
11 Select or clear the desired sipping order options: Column-wise
and File Sip Order.
12 Select any other run parameters as desired, and then start the
run.
Using a Barcode to Specify the Sample Names
File
If your instrument is equipped with a barcode reader, the desired
Sample Names file can be specified by using the plate barcode in
the name of the Sample Names file. At run-time, the plate barcode
is read and then the barcode is used to read the corresponding
Sample Names file.
To use the plate barcode to specify the Sample Names file:
1Decide on a location and file name format for the Sample
Names files. For example, the Sample Names files can be
named “Sample Names” followed immediately by the barcode
on the plate.
2Create a temporary CSV file, using “%barcode%” as a
placeholder for the barcode string. For example, create a
temporary file named “Sample Names%barcode%.csv” and
save it in the location where the actual Sample Names files for
the runs will be saved.
3On the Select Wells Tab, select the desired run parameters.
4On the Setup Run Tab, touch the Browse button next to Sample
Names file, and select the CSV file that contains the
%barcode% wildcard. The Use Sample Names file for Sample
Selection check box is automatically selected when the Sample
Names file contains the string %barcode%.
5In the Select Wells tab, select the desired Sipping Order
options: Column-wise and/or File Order.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Using Sample Names Files60
Using a Barcode to Specify the Sample Names
File (Continued)
6Touch the Start Run button. The Start Run Tab opens.
7Touch the Export button and save the Run file. The Run File
can be imported to run multiple plates with the same settings.
8To start the run, touch the Start button.
9When the run is started, the plate barcode is read and the
Sample Names file with the corresponding barcode is used to
select the wells to sample. In the example above, if the plate
barcode is 12345, then the file named “Sample
Names12345.csv” will be used to select the wells in the run.
Editing Sample Names During a Run
The default Sample Names assigned during a run can be edited
after the run starts, but before the sample well read is complete.
(After the well has been read, the data is saved in the data file, and
the sample name can no longer be changed.) same names
assigned using a sample names file can also be edited before the
well has been read.
1After a run is started, touch the Views button on the Navigation
Bar.
2Touch the Sample Names button to open the Sample Names
Tab . If a sample names file was specified, the names from the
Sample Names File display. Otherwise, the well labels are used
as the default sample names.
3Edit the sample names as desired. (If a well has already been
read, the Sample Names and Comment fields are disabled for
the well and cannot be edited.)
4Add comments to the Comment fields as desired.
5Touch the Apply button to save the changes. You must touch
Apply before the well read is complete.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Saving Data Files61
Saving Data Files
While running an assay, the raw time series data received from the
instrument is automatically saved to the data file (.gxd), one well at
a time as each well is completed. The Analysis settings from the
assay are also saved in the data file. If a run is stopped before it is
complete, the data for all completed wells is saved in the data file.
The name and location of the data file is specified in the Setup Run
Tab .
This section includes general instructions for saving data files in the
LabChip GX Touch software:
•Saving Data Files with LabChip GxP Option Installed
•Organizing, Retrieving, and Backing Up Data Files
Saving Data Files with LabChip GxP Option
Installed
If the LabChip GxP option is installed, the raw time series data
received from the instrument is saved to the CDR, one well at a
time as each well is completed. If the connection to a remote CDR
server is lost after a run is started, the data is saved to a local folder
and copied to the CDR when the network connection is restored.
Organizing, Retrieving, and Backing Up Data
Files
As you work in the LabChip GX Touch software, it is a good practice
to organize the LabChip GX Touch data files.
•Create a folder in which to save the data files. If desired, each
user can save data files to their own subfolder to organize the
data files.
•Review the files periodically, even if only one person uses the
LabChip GX Touch software. If you are not using the LabChip
GxP option, archive files you are no longer using but want to
save by copying to a backup disk, and then delete unneeded
files. Verify there is enough free space on the hard drive to save
new plate data files. If you are using the LabChip GxP option,
see LabChip GxP Option on page 62 for backup options.
•Each user in the laboratory can specify a particular data file
prefix to easily differentiate data files.
A new folder can be created each day to store the data from all
runs. To automatically create the folders, select the Create Daily Sub-Directory check box on the Setup Run Tab.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
LabChip GxP Option62
LabChip GxP Option
A LabChip GxP option is available for the LabChip GX Touch
software. This option contains built-in technical controls and
features specifically designed to provide compatibility with 21 CFR
Part 11 requirements. Users are responsible for establishing
policies and standard operating procedures that complement the
capabilities provided by the software in order to ensure complete
compliance to the rule.
NOTE
Using the LabChip GxP compatibility option alone does not ensure
21 CFR Part 11 compliance. Laboratory processes and procedures
must comply with 21 CFR Part 11 regulations.
The LabChip GxP option ensures that assays, output data, analysis
settings, event data, and backup data files are not available for
editing or tampering. Data is stored in a secure folder on the local
computer or on a network server. To manage LabChip GX Touch
user accounts, Microsoft
with the LabChip GxP option.
®
SQL Server® 2014 Express is provided
The LabChip GxP option requires an internal USB Key (dongle)
installed inside the instrument cover. The USB Key is only to be
installed by trained PerkinElmer Service personnel.
The following procedures are included in this section:
•Locking and Unlocking the Software on page 63
•User Accounts on page 64
— Switching Users on page 64
•Central Data Repository (CDR) on page 65
— CDR Security Suggestions on page 65
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
LabChip GxP Option63
Locking and Unlocking the Software
The LabChip GX Touch software with the LabChip GxP option
installed allows you to lock the LabChip GX Touch software. The
software also locks after it has been left unattended for an amount
of time specified by the LabChip GX Touch Administrator.
Locking the LabChip GX Touch software prevents unauthorized
users from accessing the software while you are away from the
computer. After the software is locked, only the logged in user or a
LabChip GX Touch Administrator can unlock the software.
To lock the LabChip GX Touch software:
1On the Home Window, touch the User button in the lower right
corner. The Unlock Application Window opens on top of the
LabChip GX Touch Main Window and displays the User name of
the current user.
2To uc h th e Lock button. The Unlock Application Window opens
and the software is locked until unlocked by the current user or
an administrator.
Figure 14. Login Window
NOTE
If a run is in progress, the run continues and th e data is logge d
under the user name of the user that was logged in at the beginning
of the run, even if another user logs in during the run.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
LabChip GxP Option64
Locking and Unlocking the Software (Continued)
To unlock the LabChip GX Touch software:
1On the Unlock Application Window, type the password for the
logged in user in the Password text box and touch the Logon
button.
2If the password for the current user is not available, type a
LabChip GX Touch Administrator user name and password in
the Unlock Application Window and touch the Logon button.
The Administrator is logged into the LabChip GX Touch software
and the previous user is logged out.
User Accounts
Access to the LabChip GX Touch software is controlled by user
accounts when the LabChip GxP option is installed. Each user must
sign into the LabChip GX Touch software from the Unlock
Application Window. User accounts are managed in the LabChip
GX Reviewer software. See the LabChip GX Reviewer User Manual
for information on creating and managing user accounts.
Switching Users
The LabChip GX Touch software with the LabChip GxP option
installed allows you to switch users during operation.
To switch users:
1On the Unlock Application Window, touch the Switch User
button.
2Type the user name and password of the new user and touch
the Logon button.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Central Data Repository (CDR)
The Central Data Repository (CDR) is a protected folder located on
the local computer or on a network server. All data files (.gxd) are
saved in the CDR folder. The CDR folder is protected from changes
by unauthorized users. The CDR is only used when the LabChip
GxP option is installed.
The data files can be organized into virtual folders in the CDR using
the folder options in the Setup Run Tab. The folders are not actually
created in the CDR folder, but are displayed in the LabChip GX
Reviewer and LabChip GX Touch software to organize the data
files. The LabChip GX Reviewer software enables you to create
new folders, rename existing folder, and delete empty folders.
CDR Security Suggestions
To ensure proper security of data files, the LabChip GX Touch
Administrator should:
LabChip GxP Option65
1Change the administrator password. Make sure to keep a copy
of the password in a safe place. This password cannot be reset
if forgotten. To change the administrator password:
aLog into the LabChip GX Touch software as the
administrator.
bOn the Home Window, touch the User button in the lower
right corner. The Unlock Application Window opens.
cTouch the Change Password button. The Change
Password Window opens.
dType the current password in the Old Password text box.
eType the new password in both the New Password and
Confirm Password text boxes.
NOTE
Write down the administrator password and keep a copy in a safe
place. The administrator password cannot be reset or recovere d if
forgotten or lost.
fTouch the OK button.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Software Reference66
Software Reference
This section describes the windows in the LabChip GX Touch
software. Each topic describes the options and buttons on the
window, and how to open the window.
•CDR/Database Server Window on page 67
•Change Assay Folder Window on page 68
•Change Password Window on page 69
•LabChip GX Touch Main Window on page 70
— Navigation Bar on page 71
— Home Window on page 72
•Login Window on page 74
•Prime and Calibrate Window on page 76
•Prime Window on page 77
•Run Window on page 78
— Select Wells Tab on page 79
— Setup Run Tab on page 81
— Start Run Tab on page 84
•Status Window on page 86
•Tools Window on page 89
— Export Setup Tab on page 90
— Diagnostics Tab on page 95
— Optics Normalization Tab on page 97
— Plate Editor Tab on page 98
•Unlock Application Window on page 104
•Views Window on page 105
•Wash Window on page 124
— Software IQ Tab on page 100
— Factory Access Tab on page 102
— Error Message Display on page 103
— Select Wells to View Tab on page 106
— Egram/Gel Tab on page 108
— EGram Tab on page 109
— Gel Tab on page 113
— Run Info Tab on page 116
— EventLog Tab on page 119
— Sample Names Tab on page 120
— Analysis Settings Tab on page 122
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
CDR/Database Server Window
Use the CDR/Database Server Window to specify the location of
the CDR folder or to view the current location of the CDR folder
when the LabChip GxP option is installed.
Figure 15. CDR/Database Server Window
Software Reference67
The CDR/Database Server Window contains the following options:
Server Name / IP Address - The server name or IP address of the
computer where the CDR is installed.
Apply button - Connects to the CDR on the specified computer or
server. If the CDR cannot be contacted due to network problems or
because the CDR is not installed yet, an error message displays.
Cancel button - Closes the window without connecting to the
remote CDR Server. If the LabChip GX Touch software is not
connected to the CDR, assays cannot be run until the CDR is
connected.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Change Assay Folder Window
The Change Assay Folder window is used to select the Assay
Folder when priming a chip or running a plate.
To open the Change Assay Folder window, touch the Change Assay Folder button on the Prime Window or on the Select Wells
Tab .
Software Reference68
Figure 16. Change Assay Folder Window
The Change Assay Folder window contains the following options:
New Folder button - Enables you to create new assay folders.
Rename button - Enables you to rename the current assay folder.
Use the LabChip GX Reviewer software to create new assay
folders.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Change Password Window
Use the Change Password Window to change the password for the
current user. This window is only available if the LabChip GxP
option is installed with the LabChip GX Touch software.
To open the Change Password Window, touch the User button on
the Home Window and then touch the Change Password button.
Software Reference69
Figure 17. Change Password Window
The Change Password Window contains the following options:
Old Password - The current password for the current user.
New Password - The new password for the current user.
Confirm Password - The new password for the current user.
OK button - Saves the new password and closes the window.
Cancel button - Closes the window without saving changes to the
password.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
LabChip GX Touch Main Window
Use the LabChip GX Touch main window to access the Navigation
Bar and Home Window. The LabChip GX Touch Main Window
opens when the software is opened.
Software Reference70
Figure 18. LabChip GX Touch Main Window
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Navigation Bar
The Navigation Bar on the left side of the LabChip GX Touch Main
Window provides access to the main functions of the LabChip GX
Touch software. Touching a navigation button opens the
corresponding window.
Software Reference71
The Navigation bar contains the following buttons:
Home button - Opens the Home Window to start a prime,
calibration, pressure line purge, run, or wash operation or
to load/unload plates or chips.
Status button - Opens the Status Window to view the
current status of the instrument, chip, and run. For Protein
Clear HR and ProteinEXact assays, the status of
prime/calibration displays.
Views button - Opens the Views Window to view data from
the last completed run. After a new run is started or the
software is closed, data from a completed run is no longer
available in the LabChip GX Touch software. Use the
LabChip GX Reviewer software to view or analyze data
from previous runs.
Tools button - Opens the Tools Window to add plates, run
software and hardware diagnostics, select Auto export
options, calibrate the optics, or access PerkinElmer service
engineer software functions.
Exit button - Closes the LabChip GX Touch software.
Error button - Only displays if an error has occurred. Opens the
Current Events Tab to view the error details and to clear the error.
Warning button - Only displays if a Warning has occurred. Opens
the Current Events Tab to view the warning details and to clear the
warning.
Stop button - Stops the current operation in progress. If running a
plate, the run stops immediately and the current well does not finish
reading. Any data collected from the current well is discarded. If
priming or washing a chip, the prime or wash stops immediately.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Home Window
The Home Window displays to the right of the Navigation Bar and is
used to prime a chip, calibrate a chip, wash a chip, run a plate, or
purge the instrument’s pressure lines. Also used to load and unload
plates and chips in the instrument. The status of the instrument and
user information also display.
If the LabChip GxP option is installed, the Home Window displays
the CDR Server status and provides access to the Login window.
To open the Home window, touch the Home button on the
Navigation Bar.
Software Reference72
Figure 19. Home Window
The Home window contains the following buttons:
Prime - Opens the Prime Window to prime the chip (not available for
Protein Clear HR chips).
Prime and Calibrate - Opens the Prime and Calibrate Window to
calibrate the instrument to run Protein Clear HR assays (only for
Protein Clear HR chips).
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Home Window (Continued)
Run - Opens the Run Window to run an assay. See the Assay User
Guide for chip preparation instructions. While an assay is running,
the Exit button on the Navigation Bar changes to a Stop button to
stop the run in progress. Assays that have been stopped cannot
automatically be resumed. See Continuing a Stopped Run on
page 40 for details on restarting a stopped assay.
Wash - Opens the Wash Window and immediately begins washing
the chip. See the Assay User Guide for chip preparation instructions.
Unload/Load Plate - Ejects or retracts the microplate carrier to
access the microplate, buffer tube, and ladder tube.
Purge Pressure Lines - Forces any potential liquid or debris out of
the pressure lines to improve instrument performance.
Unload Chip - Unlatches the front door to access the chip.
Manually close the door to load the chip.
Software Reference73
Instrument Status - Displays the current status of the LabChip GX
Touch instrument:
•Calibrating - The GX Touch software is calibrating the
instrument to run a Protein Clear HR or ProteinEXact assay.
•Connecting - The GX Touch software is connecting to the
instrument.
•EjectPlate - The GX Touch is ejecting a plate carriage.
•Idle - The GX Touch is not active.
•Open Door - The GX Touch is opening the front door to access
the chip.
•Logon - A user has not yet logged into the software.
•Priming - The GX Touch is priming a chip.
•Purging - The GX Touch is purging the pressure lines.
•Retract Plate - The GX Touch is retracting a plate carriage.
•Startup - The GX Touch software is initiating communication
with the instrument.
•Running - The GX Touch is running an assay.
•Warming - The GX Touch is warming a chip.
•Washing - The GX Touch is washing a chip.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Home Window (Continued)
Server Status - Only displays if the LabChip GxP option is
installed. Displays the current status of the CDR/Database server
connection:
•Connected to server - The software is connected to the server.
Touch the button to open the CDR/Database Server Window.
•Mapping Server - The software is locating the server name or
IP address of the computer where the CDR is installed.
•Locked by Timeout - Displays if the software has been left
unattended for an amount of time specified by the LabChip GX
Touch Administrator.
•Locked by User - Displays if the software has been locked by
the user.
•Server is Not available - The software is not connected to the
server.
•Server not mapped - The software has not been mapped to a
CDR server.
Software Reference74
Login Window
User button - Only displays if the LabChip GxP option is installed.
Opens the Unlock Application Window to log into the software, lock
the software, switch users, or change the password of the current
user.
Information (i) button - Opens the InfoView window to access
information about LabChip GX Touch software, the LabChip GX
Touch online help, and the Log Zipper function.
Use the Login Window to log in to the LabChip GX Touch software,
lock the software, switch users, or change passwords when the
LabChip GxP option is installed. The LabChip GX Touch software
will not start until a valid user name and password are entered.
The Login Window opens when you start the LabChip GX Touch
software with the LabChip GxP option installed.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Login Window (Continued)
Figure 20. Login Window
The Login Window contains the following options:
Username text box - The user name to log into the LabChip GX
Touch software. Each user should have a unique user name.
Software Reference75
Password text box - The password assigned to the Username. All
passwords must be at least 5 characters long and must contain at
least one uppercase letter and at least one number. The User
Account Policies specify additional password requirements.
NOTE
The Switch User, Change Password, and Lock buttons are only
available if a user is already logged into the LabChip GX Touch
software.
Switch User button - Enables the User Name and Password text
boxes to enable a new user to log in. The current user is
automatically logged out.
Change Password Button - Opens the Change Password Window
to change the password for the current user.
Logon button - Logs in the user and closes the Login Window.
Lock button - Opens the Unlock Application Window and locks the
LabChip GX Touch software. Only the current user or an
administrator can unlock the software.
Cancel button - Closes the Login Window without logging the user
into the software.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Prime and Calibrate Window
Use the Prime and Calibrate window to calibrate the instrument to
run a Protein Clear HR assay. To open the Prime and Calibrate
window, touch the Prime and Calibrate button on the Home
Window.
Software Reference76
Figure 21. Prime and Calibrate Window
The Prime and Calibrate window contains the following options:
Change Assay Folder button - Touch to select the folder that
contains the desired assays.
Select Assay drop-down list - Touch to select the assay to run.
A1 button - Touch to enlarge the plate diagram to select the well
containing the VeriMAb sample.
Skip Prime step check box - Only displays if a Protein Clear HR
chip has already been calibrated. If selected, the chip is calibrated
without a second prime.
Select Plate Typedrop-down list - Select the desired plate type to
be used during the run.
Prime and Calibrate button - Starts the calibration process.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Prime Window
Use the Prime window to prime chips that do not support Protein
Clear HR assays. The chip must be primed for the same assay (or
assay family) as the assay that will run on the chip. To use the chip
for a different assay, wash and reprime the chip by selecting the
correct assay and using the appropriate reagents. Load the buffer
tube and the chip before opening the Prime window. To open the
Prime window, touch the Prime button on the Home Window.
Software Reference77
Figure 22. Prime Window
The Prime window contains the following options and buttons:
Change Assay Folder button - Touch to select the folder that
contains the desired assays.
Assay - Specifies the assay for which to prime the chip. The
selections available depend on the chip type and the assay type
that was selected when the chip was inserted.
Run Test Ladder after Prime check box - If selected, runs one
ladder well after priming is complete.
Prime button - Starts priming the chip and displays the status of
the priming step.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Run Window
Use the Run Window to define the assay run parameters. To open
the Run Window, touch the Run button on the Home Window.
The Run Window contains the following tabs:
•Select Wells Tab
•Setup Run Tab
•Start Run Tab
Software Reference78
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Select Wells Tab
Use the Select Wells tab to select the assay, plate type, wells to
read, and advanced run options. See the Assay User Guide for chip
and sample prep instructions. To open the Select Wells tab, touch the
Run button on the Home Window or touch the Select Wells tab on
top of the Run Window.
Software Reference79
Figure 23. Select Wells Tab
The Select Wells tab contains the following options and buttons:
Change Assay Folder button - Touch to select the folder that
contains the desired assays.
Select Assay drop-down list - Select the type of assay to run. The
assay types available depend on the chip that is loaded in the
instrument. The instrument automatically identifies the chip type
when the chip is loaded in the instrument.
Select Plate Type drop-down list - Choose the type of plate that
will be used in the run.
Import Run File button - Opens the Import Run Parameters from
XML File window to select a run file to import. See Creating a Run
File on page 54 for details on run files.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Select Wells Tab (Continued)
Plate Diagram - Select the wells to read during the run. Selected
wells are blue, unselected wells are white.
•To select all rows on the plate, touch the Select All Rows
button at the top left corner of the plate diagram.
•To select a row, touch the row letter on the left side of the plate.
•To select a column, touch the column number at the top of the
plate.
•Clear a single well by touching the selected well.
Full Size button - Enlarges the Plate Diagram to full screen to
enable easier selection of wells, especially when selecting wells on
384-well plates. Touch the Done button in the upper right corner to
return to the Select Wells tab.
Quadrants - Touch the 1, 2, 3, or 4 buttons to select or clear all
wells in each quadrant. Any wells that are already selected are not
changed unless the wells are in the specified quadrant.
Software Reference80
Column-wise check box - Specifies the order in which the wells
are sampled during the assay:
•If selected, the wells in each column are sampled from top to
bottom before proceeding to the next column (A1, B1, C1... A2,
B2, C2...). Ladders are sipped every 12 wells, so some columns
will include a ladder in the column.
•If cleared, the wells in each row are sampled from left to right
before proceeding to the next row (A1, A2, A3... B1, B2, B3...).
Ladders are sipped every 12 wells.
Selection Sip Order check box - If selected, the wells are sipped
in the order in which the wells are selected. Selecting a rectangular
region orders the wells from the start of the selection region to the
end of the selection region, either row-wise or column-wise. If not
selected, the wells are sipped in row or column order.
Advanced Settings - If selected, uses the Advanced Run Settings:
•Sample Sips - Specifies the number of times to run the selected
wells. Each well is sampled once, and then the entire run
repeats. The data from all wells is combined in one data file.
•Plate Cycles - Specifies how many times to repeat the assay on
the plate.
•Random Select% - Specifies the percent of wells to randomly
sample from the selected wells. For example, if 10 wells are
selected, and the Random Select% is set to 50%, five wells,
selected at random, will be sampled during the run.
Next button - Opens the Setup Run Tab.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Setup Run Tab
Use the Setup Run tab to specify the operator name, skip prime and
skip warm options (ProteinEXact assays only), barcode options,
data file options, auto export options, and auto print options. The
Sample Names file, Expected Peaks file, and Excluded Peaks file
are also selected on the Setup Run tab. To open the Setup Run tab,
touch the Next button on the Select Wells Tab or touch the Setup Run tab on top of the Run Window.
Software Reference81
Figure 24. Setup Run Tab
The Setup Run tab contains the following options:
Operator text box - Specifies the name of the operator running the
assay. The Operator Name is saved in the data file. If the LabChip
GxP option is not installed, the Operator text box is optional. If the
LabChip GxP option is installed, displays the first and last name of
the current user and cannot be edited.
Skip Prime check box - Only displays if a ProteinEXact chip has
been primed before the run. If selected, the priming process is not
repeated.
Skip Warm check box - Only displays if a ProteinEXact chip has
been warmed before the run. If selected, the warming process is
not repeated.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Setup Run Tab (Continued)
Read Plate Barcode check box - Only displays if the instrument is
equipped with a barcode reader. If selected, the LabChip GX Touch
reads the customer-applied barcode on the short (portrait) end of
the plate. The barcode is saved in the data file and can be used as
part of the data file name.
Data Path - Specifies the location where the data file will be saved.
To change the path, touch the Browse button. Create or select the
specific folder for the new path and touch the OK button. If the
LabChip GxP option is installed, data files are saved in folders in
the CDR. Touch the Default button to restore the default data path.
NOTE
If the LabChip GxP option is not installed, data files should be
saved to a local folder on the computer's hard drive. Saving data
files to a network drive may cause loss of data if the network
connection is slow or interrupted. If the LabChip GxP option is
installed and the network connection is interrupted, data files are
archived in a local folder and copied to the CDR when the ne twork
connection is restored.
Software Reference82
Copy To check box - If selected, the data file is copied to the
specified folder after the run is complete. To change the path, touch
the Browse button. Touch the Default button to restore the default
path.
Create Daily Sub-Directory check box - If selected, a new
directory is created each day in the specified Data Path, and all of
the data files from that day are saved in the directory. The directory
name is the current date, and the format is YYYY-MM-DD, where
YYYY is the year, MM is the month, and DD is the day.
Auto Export check box - If selected, data is automatically
exported at the end of each run. Select the type of data to export by
touching the Tool s button and choosing the desired settings in the
Export Setup Tab.
Auto Print check box - If selected, prints all exported data, except
Raw Data, into a single PDF at the end of the run.
Defer Export to Plate Completion check box - If selected, all data
from the run is exported at the end of the run instead of as each
well is completed.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Setup Run Tab (Continued)
Data File Name text box - Read-Only text box that displays the
selected format for the data file name. The file name components
are added to the data file name in the order in which they are
selected.
File Prefix text box - Specifies the text for the first characters of
the data file name. (Optional)
Project Name text box - Specifies a project name to use in the
data file name. (Optional)
Computer Name check box - If selected, adds the name of the
LabChip GX Touch computer to the data file name.
Barcode check box - Only displays if the instrument is equipped
with a barcode reader. If selected, the instrument reads the
customer-supplied barcode on the short (portrait) edge of the
microplate and includes the barcode in the data file name.
Software Reference83
Date check box - If selected, the current date is included in the
data file name. The date format is YYYY-MM-DD, where YYYY is
the year, MM is the month, and DD is the day.
Time check box - If selected, the time that the run was started is
included in the data file name. The time format is HH-MM-SS,
where HH is the hour (00 to 24), MM is the minutes, and SS is the
seconds.
Sample Names File - If selected, specifies the file that will supply
the Sample Names in the data file. Touch the Browse button next to
Sample Names File, select the name of the CSV file that contains
the sample names, and touch the Open button. The path and name
of the file displays in the text box. Use the LabChip GX Reviewer
software or Excel to create Sample Name files. See. See page 177
for more information about Sample Names files. To use the Sample
Names file to select the wells to sample, see page 57. To select the
sample names file using a plate barcode (if your instrument is
equipped with a barcode reader), see page 59.
Expected Peak File - If selected, specifies the file that will supply
the Expected Peaks for the data file. Touch the Browse button next
to Expected Peaks File, select the name of the GEP file that contains
the expected peaks, and touch the Open button. The path and
name of the file displays in the text box. Use the LabChip GX
Reviewer software to create the Expected Peak file. See page 170
for more information about Expected Peak files.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Setup Run Tab (Continued)
Excluded Peak File - If selected, specifies the file that will supply
the Excluded Peaks for the data file. Touch the Browse button next
to Excluded Peaks File, select the name of the GEP file that
contains the excluded peaks, and touch the Open button. The path
and name of the file displays in the text box. Use the LabChip GX
Reviewer software to create the Excluded Peak file. See page 170
for more information about Excluded Peak files.
The Setup Run tab contains the following buttons:
•Back - Opens the Select Wells Tab to change the selected wells
or sampling options.
•Next - Opens the Start Run Tab.
Start Run Tab
Use the Start Run tab to review the settings for the run, and start
the run if the settings are correct. To open the Start Run tab, touch
the Next button on the Setup Run Tab or touch the Start Run
button on top of the Run Window.
Software Reference84
Figure 25. Start Run Tab
The Start Run Tab contains:
Chip Status - Displays the Assay Type, Chip Reagent Expiration,
Chip Expiration date, and Chip Life.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Start Run Tab (Continued)
•Type - Displays the type of chip in the instrument, or the assay
type selected for the chip if the chip supports multiple assays.
•Chip Reagents - Displays the time and number of samples
(whichever comes first) until the chip reagents expire. Reagent
expiration depends on the chip type. The expiration time and
number of samples reset whenever the front door is opened.
•Chip Expiration - The expiration date of the chip. The
expiration date is stored on the RF ID tag on the chip.
•Chip Life - Displays the number of samples until the chip
expires. Chip Life depends on the chip type.
Run Parameters - Displays the settings selected on the Select
Wells Tab and the Setup Run Tab.
Plate Diagram - The wells selected to read are blue. Wells that are
not selected to read are white.
Software Reference85
Total Ladder Concentration text box - (ProteinEXact assays only)
Type the total ladder concentration of the reagent into the text box.
Back button - Opens the Setup Run Tab.
Export button - Opens the Save Run Parameters to File
window to export the current run settings to a run file (.xml).
Start button - Runs the plate using the current settings.
The Status Window opens to view the status of the run or priming
and calibration.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Status Window
Use the Status window to view the status of a prime, instrument
calibration, wash, or run while an operation is running or after the
operation is complete. To open the Status window, touch the Status
button on the Navigation Bar.
Software Reference86
Figure 26. Status Window
The Status window displays the following information:
Chip Status - Displays the current status of the chip. The LabChip
GX Touch automatically detects the chip type from an embedded
RF ID tag in each chip.
•Type - Displays the type of chip in the instrument, or the assay
type selected for the chip if the chip supports multiple assays.
•Chip Reagents - Displays the time and number of samples
(whichever comes first) until the chip reagents expire. Reagent
expiration depends on the type of chip inserted in the
instrument. The expiration time and number of samples reset
whenever the front door is opened.
Prime/Calibrate Status (Protein Clear HR and ProteinEXact chips
only) - Displays the status of the current or just completed prime
and calibration.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Status Window (Continued)
•Assay - The name of the assay for which the instrument will be
calibrated.
•Status - The current step of the operation in progress. For
completed operations, displays the status of the last operation.
•Step Time Left - The time remaining until the current step in the
priming/calibration completes.
•Run Time Left - The time remaining until the current run in the
priming/calibration completes.
•Maximum Calib. Time - The maximum amount of time
remaining until the priming/calibration completes.
Run Status - Displays the current status of the assay, prime, or
wash that is currently running or about the last completed
operation.
•Assay - The name of the assay that is running or that generated
the displayed data.
•Status - The current step of the operation in progress. For
completed operations, displays the status of the last operation.
•Step Time Left - The time remaining in the current step in the
run.
•Run Time Left - The time remaining until the assay is complete.
Software Reference87
•Barcode - Displays the barcode number if Use Barcode is
selected in the Setup Run Tab. Displays N/A if Use Barcode is
not selected in the Setup Run Tab.
•Current Cycle - Displays <CurrentPlateCycle>/
<TotalPlateCycle>, where CurrentPlateCycle is the number of
the plate cycle currently executing and TotalPlateCycle is the
total number of plate cycles specified in the Select Wells Tab.
CurrentPlateCycle displays 0 during the priming and warming
steps before reading the first well. Does not display if only one
cycle is selected.
Plate Diagram - Displays the wells selected for the run in progress
or the last completed run. Does not display for ProteinEXact chips.
•Blue wells - Well has been read.
•Red exclamation point in wells - An analysis error has
occurred in the well, such as no lower marker detected.
•Flashing green/orange - Well reading is in process.
•Green - Well is selected for reading but not read yet.
•Gray - Well is not selected to be read.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Status Window (Continued)
Calibration Steps (ProteinEXact chips only) - Displays the status
of the Area Variability Test, Laser Calibration, and Verification of
the calibration. See Figure 27.
•Flashing green/orange - Calibration step is in process.
•Green - Calibration step is successful.
•Gray - Calibration step has not yet begun.
•Red - Calibration step has failed.
Figure 27. Calibration Steps for ProteinEXact
Software Reference88
NOTE
If the ProteinEXact calibration steps do not display, restart the
software and prime the chip.
Realtime Electropherogram of Well - Displays the
electropherogram of the raw (unanalyzed) data of the well in
progress during a run or prime/calibrate, or displays the last
completed well. The RealTime Electropherogram does not show
any peak labels or marker indicators. The labels and markers are
identified after all data from the well is collected, and can be viewed
in the EGram Tab in the Views Window.
Gel - Displays the Gel of the well in progress during a run or
prime/calibrate, or the last completed well.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Tools Window
Use the Tools window specify the data export settings, view or add
plates, run the software IQ, run Diagnostics, and calibrate the
optics. To open the Tools window, touch the Tools button on the
Navigation Bar.
Software Reference89
The Tools window contains the following tabs:
•Export Setup Tab
•Diagnostics Tab
•Optics Normalization Tab
•Plate Editor Tab
•Software IQ Tab
•Factory Access Tab
•Error Message Display
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Figure 28. Tools Window
Export Setup Tab
Use the Export Setup tab to select the auto export options to
automatically export data at the end of a run. To automatically
export data, select the Auto Export check box on the Setup Run
Tab . To open the Export Setup tab, touch the Tools button on the
Navigation Bar and then touch the Export Setup button.
Peak Tables and Well Tables are exported to either CSV files or
XML files. CSV files can be imported into a spreadsheet program
such as Microsoft
an XML file, or to a Chromatography Data Interchange Format file
(formerly AIA), which is used by some graphical analysis software
tools.
Software Reference90
®
Excel®. Raw Data can be exported to a CSV file,
Figure 29. Setup Tab
The following tabs display on top of the Export Setup tab:
•Export Settings Tab
•Peak Properties Tab
•Well Properties Tab
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Export Settings Tab
The Export Settings tab contains the following options and buttons:
Export Folder
•Specifies the location to save the exported files. All files are
saved in the same location. Touch the Browse (...) button to
select the location for the exported files. Touch the Default
button to restore the default path.
Export Selections
•Export All - If selected, the Peak Table, Well Table, Raw Data,
Electropherogram, and Gel data are all exported.
•Export as Single File - If selected, the Raw Data, Peak Table,
Well Table, and Gel options each generate one file that contains
all the wells in the run. If not selected, the Raw Data, Peak
Table, Well Table, and Gel options each export one file per well
and the export occurs as soon as the well is completed (except
for assays which use Bracketed ladders, in which case the
export occurs when the bracket ladder well completes).
Software Reference91
•Peak Table - If selected, the data in the Peak Table is exported
to a CSV or XML file.
•Well Table - If selected, the data in the Well Table is exported to
a CSV or XML file.
•Raw Data - If selected, the raw (unanalyzed) data from the run
is exported to a CSV, XML or Chromatography Data Interchange
Format file.
•Electropherogram - If selected, a graph of each well is
exported in the selected image format. The file names are
<plate name>_Egram_<well number>.
•Gel - If selected, the gels of each well are exported in the
selected image format. The file names are <plate
name>_Gel_<well number>.
•Height (pixels) - If Gel is selected, specifies the height, in
pixels, of the exported gel graphic.
•File Naming Template - The template for naming the data files.
All exported data files use the same names. Type in the text box
or add wildcard variables.
•Wildcards drop-down list - Select predefined variables to add
to the name of the data files: %DATA% = datafile name,
%LABEL% = well label, %NAME% = well name, %TYPE% =
export type (gel, egram, etc.), %COMMENT% = user comments.
When a single file for the plate is exported, the %NAME% and
%LABEL% fields are blank, and any leading, trailing or duplicate
separators are removed.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Software Reference92
The %DATA%, %TYPE%, and at least one of %NAME% or
%LABEL% must be included in the template to ensure file
names are unique across export types. The template will appear
in Red letters if it fails to meet the minimum requirements and
will not be saved.
•Insert button - Inserts the text selected in the Wildcard dropdown list at the cursor position in the File Naming Template text
box. Each wildcard can only be inserted once in the template.
Raw Data Options
•If the AIA Format check box is selected, the raw data is
exported to a file in the Chromatography Data Interchange
Format (formerly AIA). (Include Size Data and Export as Single
File are not available.)
•If the AIA Format check box is NOT selected, the raw data is
exported to a CSV or XML file.
•If the Include Size Data check box is selected, the data is
aligned to the well's ladder (for one file per well) or to the first
well (for a single data file) and the size data is included in the
exported data. If not selected, the data is not aligned to a ladder.
(Not available for CZE assays.)
File Format for Peak Table, Well Table, and Raw Data
•Select the desired format for the exported data files. If CSV is
selected, select the desired symbol to use to separate the data
in the CSV Delimiter drop-down list.
File Format for Gel and Electropherogram
•Select the desired format for the exported image files.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Peak Properties Tab
Use the Peak Properties tab to select the columns to export in the
Peak Table if Peak Table is selected in the Export Settings Tab in
the Tools Window. The columns available for export do not display
until after the run is started.
Software Reference93
Figure 30. Peak Properties Tab
Not exported list box - Lists the columns in the Peak Table that
are not included in the Exported Peak Table. The columns available
depend on the type of assay that generated the data.
Exported list box - Lists the columns in the Peak Table that will be
included in the exported Peak Table.
Left and Right arrow buttons - Move the highlighted table column
to the Not exported or exported list boxes.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Well Properties Tab
Use the Well Properties tab to select the columns to export in the
Well Table if Well Table is selected in the Export Settings Tab in the
Tools Window. The columns available for export do not display until
after the run is started.
Software Reference94
Figure 31. Well Properties Tab
Not exported list box - Lists the columns in the Well Table that are
not included in the exported Well Table. The columns available
depend on the type of assay that generated the data.
Exported list box - Lists the columns in the Well Table that will be
included in the exported Well Table.
Left and Right arrow buttons - Move the highlighted table column
to the Not exported or exported list boxes.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Diagnostics Tab
Use the Diagnostics tab to run the diagnostics tests on the LabChip
GX Touch or GXII Touch instrument. All tests should be run
periodically to verify proper operation of the instrument. To verify a
particular function, select specific tests to run. To open the
Diagnostics Tab, touch the Tools button on the Navigation Bar, and
then touch the Diagnostics button
Software Reference95
The left side of the tab displays the tests to run on the instrument.
Expand each section to view all of the tests in each section. Tests
selected with a check mark will run. To skip a test, touch the check
box to clear the selection. Touch a section check box to select or
clear all of the tests in the section.
The right side of the tab displays the results of the tests. The icon
color indicates the status of each test:
•Blue - The test is in progress.
•Yellow - The test was skipped.
•Green - The test passed.
•Red - The test failed or was aborted.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Figure 32. Diagnostics Tab
Diagnostics Tab (Continued)
Run Tests button - Runs the selected tests (marked with a check
mark). After all tests are complete, the Test Report Generation
section on the right side of the tab displays the date and time when
the test report was created, the name of the test report (*.log), and
the location where the test report was saved. Test report files can
be opened with a text editor such as Windows Notepad.
Abort button - Stops the tests in progress. Only enabled when
tests are running.
Limits Reportbutton - Generates a report of test limits. After the
report is generated, the Limits Report Generation section at the
bottom of the right side of the tab displays the date and time when
the limit report was generated, the name of the limits report, and the
location where the limits report was saved.
Print button - Opens the Print Diagnostic Logs window to print
the selected report. The view of the report can be changed with the
Page Mode button, Two Page Mode button, Scroll Mode button,
and Zoom controls at the bottom of the right side of the tab.
Software Reference96
Print Summary button - Opens the Print Diagnostic Summary
window to print a summary of the selected test results.
The bottom of the window displays the status, an overall progress
bar for the entire set of selected tests and a separate test progress
bar for the test that is currently running.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Optics Normalization Tab
Use the Optics Normalization Tab to calibrate the instrument using
Test Chip C as the calibration standard.To open the Optics
Normalization Tab, touch the Tools button on the Navigation Bar
and then touch the Optics Normalization button.
Software Reference97
Figure 33. Optics Normalization Tab
The Optics Normalization Tab contains the following options and
buttons:
Reference Contrast - The Contrast value from Test Chip C that will
be compared to the scan results Contrast.
Instrument Gain Correction - The current correction factor for the
instrument.
Chip Temperature (Target 35) degC - The current temperature of
the chip. The chip is warmed to the target temperature to more
accurately calibrate the optics.
Background - The background measured on the current instrument
after focusing.
Contrast - The contrast measured on the current instrument.
New Correction Factor - The correction factor calculated from the
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Optics Normalization Tab (Continued)
Open Chip Door button - Opens the front door to insert Test Chip C.
Warm and Scan/Scan/Abort button:
•Warm and Scan - Warms up the chip and laser, focuses the
optics, and measures the optical contrast.
•Scan - Scans without the warmup (available only after a scan is
complete or aborted, while laser and chip are still warmed up).
•Abort - Stops the scan or warm currently in progress.
Apply button - Saves the New Correction Factor to the
instrument’s non-volatile flash memory.
Plate Editor Tab
Use the Plate Editor tab to view, add, or change plate names and
measurements. To open the Plate Editor tab, touch the Tools button
on the Navigation Bar and then touch the Plate Editor button.
Software Reference98
Figure 34. Plate Editor Tab
The following tabs display at the top of the Plate Editor tab:
•Predefined Plates Tab - Displays PerkinElmer-defined plates.
These plates are read-only and cannot be edited or deleted.
•Custom Plates Tab - Displays all user-created plates. Plates
can be added, deleted, and modified.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Plate Editor Tab (Continued)
The Plate Editor tab contains the following options and buttons:
Plate Information table - Displays the plate information for each plate
defined for use in the system. (See Add Plate Window for a description
of each column.)
Delete Selected Plate button - Deletes the selected plate from the
system. (Only available when a plate is selected on the Custom
Plates tab.)
Add Plate button - Opens the Add Plate Window to enter the
information for a new plate. (Only available when the Custom Plates
tab is selected.)
Modify Selected Plate - Opens the Modify Plate Window to edit the
information for the selected plate. The Modify Plate Window
contains the same options as the Add Plate Window. (Only available
when the Custom Plates tab is selected.)
Software Reference99
Verify Plate button - Opens the Verify Plate window to perform a
puncture test using the selected plate information. Follow the
instructions in the Verify Plate window to test the settings for the
selected plate.
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Add Plate Window
Use the Add Plate window to add new plates to the system. To open
the Add Plate window, touch the Add Plate button on the Custom Plates Tab on the Plate Editor Tab.
The Add Plate window contains the following settings:
Plate Name - Specifies the plate name that displays in the Start
Run Tab.
Software Reference100
Figure 35. Add Plate Window
Software IQ Tab
Plate Type - Specifies the number of wells in the plate, either 96 or
384.
X Margin - The distance from the outer edge of the plate to the
center of well A1 in the X direction.
Y Margin - The distance from the outer edge of the plate to the
center of well A1 in the Y direction.
Plate Height - The distance from the bottom of the plate to the top
of the plate.
Well Depth - The distance from the top of the plate to the bottom of
the well.
Sip Height - The distance from the bottom of the well to the bottom
of the sipper when the sipper is positioned to sip sample from the
well.
Use the Software IQ tab to perform the IQ test. The IQ test verifies
proper installation of the LabChip GX Touch software and verifies
no unauthorized changes have been made to the software. To open
the Software IQ tab, touch the Views button on the Navigation Bar
and then touch the Software IQ button.
The following tabs display at the top of the Software IQ tab:
•Run IQ Tab
•Detailed Result Tab
P/N CLS151164 Rev. CLabChip GX Touch/GXII Touch User ManualPerkinElmer, Inc.
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.