Copyright 2008-2014, PerkinElmer Health Sciences, Inc. All rights reserved.
V4.2
Preface2
Preface
Copyright
This manual is published by PerkinElmer Health Sciences, Inc., 68 Elm Street,
Hopkinton, MA 01748 USA. Copyright 2008-2014, PerkinElmer Health Sciences,
Inc and its parent, affiliated, and subsidiary companies. All rights reserved,
including but not limited to those rights to reproduce this publication or parts
thereof. Reproduction of this publication or parts thereof or the products it
describes by any means or in any form is expressly prohibited without the written
permission of PerkinElmer.
Trademarks
PerkinElmer, LabChip, the PerkinElmer logo, and the LabChip logo are registered
trademarks of PerkinElmer Health Sciences, Inc and/or its parent, affiliates, and/or
subsidiary companies (collectively “PerkinElmer”). Microsoft, Windows, and
Windows NT are either registered trademarks or trademarks of Microsoft
Corporation in the United States and/or other countries. All other trademarks and
registered trademarks are the property of their respective holders.
Content
Any errors or omission which may have occurred in this publication, despite the
utmost care taken in its production, will be corrected as soon as possible, but not
necessarily immediately upon detection. PerkinElmer provides this publication “As
Is” without warranty of any kind, either express or implied, including but not limited
to the implied warranties of merchantability or fitness for a particular purpose.
Some states or jurisdictions do not allow disclaimer of express or implied
warranties in certain transactions; therefore, this statement may not apply to you.
PerkinElmer reserves the right to revise this publication and to make changes from
time to time in the content hereof without obligation of PerkinElmer to notify any
person of such revision or changes. Further, PerkinElmer may make modifications
to the product described in this manual at any time without any obligation to notify
Proper Equipment Operation
any person of such modifications.
WARNINGS
•To reduce the risk of electric shock, do not remove the cover. No user
serviceable parts inside. Refer to qualified service personnel if help is
required.
•Use this product only in the manner described in this manual. If the equipment
is used in a manner not specified by the manufacturer, the protection provided
by the equipment may be impaired.
AVERTISSEMENTS
•Pour réduire le risque de choc électrique, ne pas retirer le couvercle. Ce
produit ne contient aucune pièce pouvant être réparée par l’utilisateur. Au
besoin, confier l’appareil à un réparateur qualifié.
•Ce produit ne doit être utilisé que comme décrit dans ce manuel. Si cet
appareil est utilisé d’une manière autre que celle spécifi ée par le fabricant, la
protection fournie par l’appareil peut être entravée.
V4.2LabChip GX User ManualPerkinElmer
Preface3
Contact Us
If you have a question about the LabChip GX that is not answered in this manual or
online Help, or if you need assistance with this product, contact the PerkinElmer
Technical Support Center:
For support outside the United States, contact your local PerkinElmer
representative.
Before you call, you should have the following information available for the
technical representative:
•Product serial number
•Software version (found by choosing About from the main Help menu)
•If applicable, the error number shown in the product software or in the log file.
Product Service and Customer Support Plans
PerkinElmer offers a full range of services to ensure your success. From our
original factory warranty through a comprehensive line of customer support plans,
PerkinElmer offers you Field Service Engineers and in-house Specialists who are
dedicated to supporting your hardware, software and application development
needs.
•Diagnostic servicing performed on-site by PerkinElmer field service engineers
or remotely via Technical Support
•Validation performed on-site by PerkinElmer field service engineers
•Extended use of the PerkinElmer Technical Support Center
•Software updates
•Parts, labor, and travel expense coverage
•Other customized services upon request
Training For Your Product
Contact the PerkinElmer Center for Training and Development for information
about the availability of training courses for your product:
Call:(508) 497-2634
Fax:
V4.2LabChip GX User ManualPerkinElmer
(508) 435-3439
FCC
CE
Preface4
This device complies with part 15 of the FCC (United States Federal
Communications Commission) Rules. Operation is subject to the following two
conditions:
•This device may not cause harmful interference, and
•This device must accept any interference received, including interference that
may cause undesired operation.
This device complies with all CE rules and requirements.
NOTE
Changes or modifications to this equipment not expressly approved by the party
responsible for compliance could void the user’s authority to operate the
equipment.
REMARQUE
Tout changement ou modification apporté à cet instrument non expresséme nt
approuvé par l’entité responsable de la conformité peut annuler l’autorisation
d’opérer l’appareil accordée à l’utilisateur.
Table of Symbols
Tab le 1 contains symbols that identify particularly important information and alert
you to the presence of hazards. These symbols may appear in this manual and/or
on the product it describes.
Table 1. Important Symbols
Symbol
Symbole
Description
Description
DANGER: An imminently hazardous situation, which, if not
avoided, will result in death or serious injury.
DANGER: Situation présentant un danger imminent qui, s’il
n’est pas éliminé, peut entraîner des blessures graves, voire la
mort.
WARNING: Caution, risk of danger. Refer to the User’s
documentation.
AVERTISSEMENT: Attention, danger potentiel. Se reporter à la
documentation de l’utilisateur.
NOTE: A cautionary statement; an operating tip or maintenance
suggestion; may result in instrument damage if not followed.
REMARQUE: Énoncé indiquant une précaution à prendre, un
conseil de fonctionnement ou une suggestion d’entretien; son
non-respect peut provoquer des dommages à l’instrument.
V4.2LabChip GX User ManualPerkinElmer
Table 1. Important Symbols (Continued)
Preface5
Symbol
Symbole
Description
Description
Hazardous voltage; risk of shock injury.
Tension dangereuse; risque de blessure par électrocution.
Crush hazard. Risk of body parts, hair, jewelry, or clothing
getting caught in a moving part.
Danger d’écrasement. Faire attention que les parties
corporelles, les cheveux, les bijoux ou les vêtements ne soient
pas pris dans une pièce mobile.
Risk of puncture injury.
Risque de blessure par piqûre.
Risk of eye injury; wear safety glasses.
Risque de lésion oculaire; porter des lunettes de sécurité.
Risk of fire.
Risque d’incendie.
Risk of poison.
Risque d’empoisonnement.
Risk of explosion.
Risque d’explosion.
Hazardous fumes.
Émanations dangereuses.
Laser light; avoid exposure. Risk of eye injury.
Rayonnement laser; éviter toute exposition. Risque de lésion
oculaire.
Lifting hazard. May result in injury.
Levage dangereux. Peut entraîner des blessures.
Protective ground symbol.
Symbole de terre de protection.
Ground symbol.
Symbole de terre.
Fuse.
Fusible.
Alternating current.
Courant alternatif.
On (supply).
Marche (alimentation).
V4.2LabChip GX User ManualPerkinElmer
Table 1. Important Symbols (Continued)
Preface6
Symbol
Symbole
Equipment
labels are color
coded:
Les étiquettes
de l’appareil
sont codées
couleur:
Description
Description
Off (supply).
Arrêt (alimentation).
CE compliance mark.
Marque de conformité CE.
Signifies that the unit has passed safety tests for grounding,
power line transience, and current leakage.
Signifie que l’appareil a réussi les tests de sécurité pour la mise
à la terre, le courant transitoire de ligne d’alimentation et la
perte de courant.
Input.
Entrée.
Output.
Sortie.
Yellow Caution, risk of danger
Red Stop
Blue Mandatory action
Green Safe condition or information
Jaune Attention, danger potentiel
Rouge Arrêter
Bleu Intervention obligatoire
Vert Condition sûre ou informations de sécurité
Helpful hints, additional information
Conseils utiles, informations supplémentaires
V4.2LabChip GX User ManualPerkinElmer
Instrument Safety
The following safety information about the LabChip GX is included
in this documentation. Read and review all safety information
before operating the LabChip GX.
•Required Training
•“Chemical Safety” on page 8
•“Laser Safety” on page 9
•“Electrical Safety” on page 10
•“Mechanical Safety” on page 11
Required Training
Ensure that all personnel involved with the operation of the
instrument have:
•Received instruction in general safety practices for laboratories.
•Received instruction in specific safety practices for the
instrument.
•Read and understood all related MSDSs.
Instrument Safety7
WARNING
Use this product only in the manner d escri bed in this ma nual. If the
equipment is used in a manner not specified by th e manufactur er,
the protection provided by the equipme nt may be impaired.
V4.2LabChip GX User ManualPerkinElmer
Chemical Safety
WARNING
Some chemicals used with the LabChip GX are potentially
hazardous and can cause ill ness.
•Read and understand the material safety data sheet (MSDS)
provided by the chemical manufacturer before you store,
handle, or work with any chemical or hazardous material.
•Minimize contact with and inhalation of chemicals and chemical
wastes. Wear appropriate personal protective equipment when
handling chemicals (e.g., safety glasses, gloves, or clothing).
For additional safety guidelines consult the MSDS.
•Do not leave chemical containers open. Use only with adequate
ventilation, including a fume hood, if necessary.
•Check regularly for chemical leaks or spills. If a leak or spill
occurs, follow the manufacturer’s cleanup procedures as
recommended on the MSDS.
•Dispose of waste in accordance with good laboratory practices
and local, state/provincial, or national environmental and health
regulations.
Instrument Safety8
•After emptying waste containers, seal them appropriately.
•Comply with all local, state/provincial, or national laws and
regulations related to chemical storage, handling, and disposal.
V4.2LabChip GX User ManualPerkinElmer
Laser Safety
WARNING
BRIGHT LIGHT HAZARD. LabChip GX Instruments contain Class 3B laser
diodes. The LabChip GX is classified as a Class 1 device because the lasers
are appropriately enclosed (embedded) and indicated with Warning labels.
Complies with 21 CFR 1040.10 except for deviations pursuant to Laser
Notice 50, dated June 24, 2007.
Complies with IEC 60825-1: 1993, A1: 1997, A2: 2001.
•Use of controls or adjustments or performance of procedures
other than those specified herein may result in hazard ous
radiation exposure.
•NEVER remove back, side, or front panels of the instrument
while the laser is powered. Panels (which, if removed, could
lead to laser exposure) are marked with the labels sh own be lo w:
•These panels are intended to be removed for service only by
qualified personnel; they are not intended to be removed during
operation or for maintenance by users. The only re movable
maintenance panel is the lower panel at the back of the
instrument, which can be removed to access the back of the
robot, if cleaning is necessary.
V4.2LabChip GX User ManualPerkinElmer
Electrical Safety
The LabChip GX is powered by a UL/CSA/VDE approved 100-240
VAC, 50/60 Hz input, 5, 15, 24 VDC output power supply.
Additionally, the LabChip GX High Voltage circuitry is currentlimited to non-hazardous levels. Users should observe the
following:
WARNING
Do not open the instrument enclosure. There are no user
serviceable parts inside.
The wall outlet or the power cable connector on the back of the
instrument should be accessible after the system's installation, to
enable trained service personnel to safely disconnect power from
the system during servicing.
The computer supplied with the LabChip GX instrument has internal
lithium batteries. Batteries should not be incinerated.
Instrument Safety10
WARNING
Danger of explosion if battery is incorrectly replaced. Replace only
with the same or equivalent type recommende d by the
manufacturer's instructions.
Power Cord Selection
United States and Canada
The LabChip GX instrument is shipped with a NEMA 5-15 / IEC 320
power cord. If the power cord needs to be replaced, substitute
power cords must be UL Listed, Type SJT or equivalent, minimum
No. 18 AWG, 3-conductor with ground conductor that for safety
considerations should never be disconnected or defeated. The
cord’s plug to the wall must be a three-pin grounding type connector
with a NEMA 5-15P (15A, 125V) plug configuration. The cord’s
connector at the unit must conform to requirements for an EN 60
320/IEC 320 Standard Sheet C13 connector.
The equipment is intended to be plugged into a standard NEMA 515R receptacle in the wall.
V4.2LabChip GX User ManualPerkinElmer
Instrument Safety11
International
All power cord sets must be approved by an acceptable, accredited
agency responsible for evaluation in the country where the power
cord set and system will be used.
The flexible cord must be <HAR> Type H05VV-F, 3-conductor,
minimum 0.75 - 1 mm
set fittings that is, the appliance coupler and wall plug, must bear
the certification mark of the agency responsible for evaluation in the
country where it will be used. The appliance coupler must meet the
mechanical configuration of an EN 60 320/IEC 320 Standard Sheet
C13 connector for mating with appliance inlet on the system.
Fuses
The LabChip GX instruments do not contain any user replaceable
fuses. Contact PerkinElmer Technical Support (see “Contact Us” on
page 3) if blown fuses are suspected.
Mechanical Safety
The LabChip GX instruments have a three axis robot that moves
quickly and can be a pinch hazard. Keep the front door of the
instrument closed when the robot is moving. Keep hands away from
the robot when not actually placing microplates in the instrument or
changing the ladder and buffer vials. Robot access areas are
marked with the following warning label:
This manual includes general instructions for using the LabChip GX
hardware and software. It includes general procedures for operating
the system, analyzing the data, using software security to comply
with 21 CFR Part 11 requirements, instrument maintenance, and
hardware and software troubleshooting.
DNA, gDNA, RNA, Protein, Glycan, and Protein Charge Variant
(CZE) Chip and Reagent Kits are available to run specific assays on
the LabChip GX. The Assay Kits include the reagents and
consumables required to run the specific assay. Protein, Glycan,
and Protein Charge Variant assays are only supported on LabChip
GX II instruments.
Assay User Guides
Assay User Guides provide information about the assay.
Instructions for preparing the chip, the plate, the ladder vial, and the
buffer vial are included in the LabChip GX/GXII Assay User Guide
for the specific assay that you are running. Detailed information
about the assays, including Specifications, Safety Warnings,
Preparation Procedures, Expected Results, Troubleshooting,
LabChip Kit Essential Practices, and Reordering Information is also
located in the LabChip GX/GXII Assay User Guide for the specific
assay that you are running.
The current version of the Assay User Guides can be accessed on
the PerkinElmer web site at:
www.perkinelmer.com/labchipguides.
Assay Quick Guides
Assay Quick Guides are included with each Assay Kit and include
instructions for preparing the chip to run an assay.
The current version of the Assay Quick Guides can be accessed on
the PerkinElmer web site at:
www.perkinelmer.com/labchipguides.
V4.2LabChip GX User ManualPerkinElmer
Principles of Operation
The LabChip GX assays are based on traditional gel
electrophoresis principles that have been transferred to a chip
format. The chip format dramatically reduces separation time and
provides automated sizing and quantitation information in a digital
format.
The chip contains an interconnected set of microchannels that join
the separation channel and buffer wells. One of the microchannels
is connected to a short capillary that extends from the bottom of the
chip at a 90-degree angle. The capillary sips sample from the wells
of a microplate during the assay.
Some of the channels in the chip are larger than others. The larger
channels contain buffer. During the chip preparation, the smaller
channels and some of the wells are filled with sieving gel and buffer.
Introduction19
Figure 1. DNA/RNA Chip and Protein Chip Schematic s
After the channels are filled, the chip functions as an integrated
electrical circuit. The circuit is driven by the 7 electrodes in the
electrode cartridge that contact solutions in the chip's wells when
the chip holder is closed. Each electrode is connected to an
independent power supply that provides maximum control and
flexibility.
The polymer filling the smaller channels in the chip is designed to
sieve DNA/RNA fragments or proteins by size as they are driven
through it by means of electrophoresis, similar to using agarose or
polyacrylamide gels. The sample and sieving buffers also contain a
fluorescent dye that gets brighter upon binding to double-stranded
DNA, RNA, or protein/SDS complex. (Protein, Glycan, and Protein
Charge Variant assays are only supported on LabChip GX II
instruments.)
V4.2LabChip GX User ManualPerkinElmer
Principles of Operation (Continued)
In the chip, each sample is sipped from the microplate by negative
pressure until a sufficient quantity is loaded in the chip. The sample
is then moved electrophoretically into the central channel. As the
fragments move down the central channel, they separate by size,
and then pass the laser, which excites the fluorescent dye bound to
the molecule. The LabChip GX software plots fluorescence intensity
versus time and produces electropherograms for each sample (see
Figure 2).
Introduction20
Figure 2. Electro pherogram
The data can be viewed in a gel-like format on the Gel Tab to
achieve the appearance of a slab gel. (The colors of the gel can be
changed.)
Figure 3. Gel View
V4.2LabChip GX User ManualPerkinElmer
Principles of Operation (Continued)
For DNA, RNA, Protein, and Glycan assays, quantitating the
concentration and accurately sizing each fragment are achieved by
comparing against a sizing ladder and running internal standards or
“markers” with each sample. Internal standards of known
concentration are mixed with the sample to aid in quantitation.
The amount of sample sipped into the chip depends on pH, salt
concentration, and buffer additives. The internal standards
normalize these factors so that the software can use the ratio of the
area of the curve of the standard to the unknown peak to determine
concentration directly. The internal standards lie slightly outside the
assay range so they do not interfere with analysis.
Capillary Zone Electrophoresis (CZE) is an electrophoretic
separation technique used to evaluate the charge heterogeneity of
proteins in a sample. For Protein Charge Variant assays, analytes
are separated based on their net charges, with molecules with a
higher net charge migrating faster than those with a lower net
charge. The data is analyzed to calculate the % Relative Amount of
the peaks.
Introduction21
Software Modes Available
The LabChip GX software has three modes available: Instrument,
Simulation, and Reviewer. The mode can be selected when the
LabChip GX software is installed.
Instrument Mode is used to control the LabChip GX instrument
and to review data generated by the LabChip GX.
•The 21 CFR Part 11 Compliance option is a purchased option
that can be used when the software is installed in Instrument
mode.
Simulation Mode is used to simulate running an instrument and to
review data generated by the LabChip GX.
Reviewer Mode is used to review data and develop assays, but
does not control an instrument or display the instrument controls.
If the LabChip GX software is installed in Instrument mode,
shortcuts can be created on the desktop to run the LabChip GX
software in Simulation or Reviewer mode (see “Creating Desktop
Shortcuts” on page 23).
V4.2LabChip GX User ManualPerkinElmer
Operation22
Operation
This section includes general instructions for using the LabChip GX
hardware and software to run an assay.
This section contains the following information:
•Opening the LabChip GX Software
•Creating Desktop Shortcuts
•Creating a New Assay
•Running an Assay
•Select the Auto Export Settings
•Monitoring the Run
•Stopping a Run
•Continuing a Stopped Run
•Saving Data Files
•Saving Workspace Files
•Selecting Wells using a Sample Names File
•Adding a New Plate
•Placing the Barcode on the Plate
•Calibrating the Optics
For assay-specific information, see the LabChip GX/GXII Assay User Guide for the specific assay that you are running. For
instructions on preparing the chip and plate, see the LabChip GX/GXII Assay Quick Guide for the specific assay. The current
version of the Assay User Guides and Quick Guides can be
accessed on the PerkinElmer web site at:
http://www.perkinelmer.com/labchipguides.
To run an assay in the LabChip GX:
1Open the LabChip GX software. (See “Opening the LabChip GX
Software” on page 23.)
2If necessary, create a new assay. (See “Creating a New Assay”
on page 24)
3Clean the electrodes and the O-Rings as directed in the
LabChip GX/GXII Assay User Guide.
4If necessary, calibrate the optics (see page 41) and run
Diagnostics (see page 291).
5Prepare the chip for the assay as directed in the LabChip
GX/GXII Assay Quick Guide.
V4.2LabChip GX User ManualPerkinElmer
6Prepare the plate for the assay as directed in the LabChip
GX/GXII Assay Quick Guide. (See “Placing the Barcode on the
Plate” on page 40 if using barcodes on the sample plates.)
7Run the assay. (See “Running an Assay” on page 25.)
Opening the LabChip GX Software
If the LabChip GX software was installed in Instrument mode, with
or without the 21 CFR Part 11 security option, the desktop icon
opens the software in Instrument mode. If desired, you can create
shortcuts for Reviewer mode or Simulation mode as described
below. If the LabChip GX software was installed in Reviewer mode,
the desktop icon opens the software in Reviewer mode.
To open the LabChip GX software:
1Double-click on the LabChip GX icon on the Windows desktop.
2If 21 CFR Part 11Security is installed, the Login Window opens.
Type a valid LabChip GX user name and password into the text
boxes and click the OK button. (For instructions on creating
LabChip GX user names, see “Adding New Users” on
page 118.)
Operation23
The LabChip GX Main Window opens.
3See “Creating a New Assay” on page 24 or “Running an Assay”
on page 25.
Creating Desktop Shortcuts
If the LabChip GX software was installed in Instrument mode, you
can create desktop shortcuts to run the LabChip GX software in
Reviewer mode or Simulation Mode, if desired.
To create new software shortcuts on the Windows desktop:
1Right-click on the LabChip GX icon on the Windows desktop
and click Copy.
2Right-click on the Windows desktop and click Paste. A copy of
the LabChip GX icon is created on the desktop.
3Rename the icon as desired.
4Right-click on the new icon and click Properties.
5On the Shortcut tab, enter the following in the Target text box:
V4.2LabChip GX User ManualPerkinElmer
•For Reviewer mode: "C:\Program Files\Caliper Life
Sciences\LabChip GX\LabChip.exe" /review
•For Simulation mode: "C:\Program Files\Caliper Life
Sciences\LabChip GX\LabChip.exe" /simulate
6Click OK in the Properties window.
Creating a New Assay
The LabChip GX software enables you to create a new assay by
opening and editing an existing assay and then saving the assay
with a new name.
To create a new assay:
1On the LabChip GX Main Window, select Tools Assay
Editor. The Select Assay To Edit window opens.
2Select the name of the Assay file that you want to edit and click
the OK button. The Assay Analysis Window opens and displays
the settings for the open assay.
Operation24
3Modify the settings as necessary for the new assay. (See
Modifying Analysis Parameters or Reanalyzing a Data File for
information on setting or changing analysis parameters.)
4Click the Save Assay button at the bottom of the Assay
Analysis window, specify the desired name for the new assay
file, and click the Save button.
V4.2LabChip GX User ManualPerkinElmer
Running an Assay
If a new chip has been inserted in the instrument, the chip is
automatically primed at the start of the first run. Since priming is a
lengthy process, you can start priming the chip before you are
ready to start the run. (To save time, the sample plate can be
prepared while the chip is priming.)
To prime a chip before starting the run:
1Select Instrument Sta rt P r ime to start priming for a specific
assay.
2Select the assay in the Choose Assay for Chip Prime window
and then click the OK button. (If the chip is not primed before
starting the run, the chip is primed automatically when you click
the Run button.)
To start an assay to read a plate:
1Click the Run button on the LabChip GX Main Window (see
page 143) to open the Start Run Window.
Operation25
If the maximum number of samples for the chip have already
been run or if the chip was last primed for a different assay, a
message displays instructions for resolving the problem.
2Click the Run tab in the Start Run window.
3Select the type of assay you want to run in the Assay Type
drop-down list. (Protein, Glycan, and Protein Charge Variant
assays are only supported on LabChip GX II instruments.)
4Type the operator's name in the Operator Name text box.
5Select the name of the plate in the Plate Name drop-down list.
To use a plate that is not listed in the Plate Name list, see
“Adding a New Plate” on page 38.
6To read the plate barcode, select the Use Barcode check box.
(The plate barcode can be used to name the data file if
selected.)
V4.2LabChip GX User ManualPerkinElmer
Running an Assay (Continued)
7On the Plate diagram, select the wells to be sampled.
•To select all wells and ladders on the plate, click the double-
arrow button in the lower-right corner of the plate.
•To select all rows on the plate, click the double-arrow button
at the top left corner of the plate.
•To select all columns on the plate, click the double-down-
arrow button at the top right corner of the plate.
•To select a single column, click the column number at the top
or bottom of the plate.
•To select or clear all wells in a quadrant, click the Q1, Q2,
Q3, or Q4 button.
•To select a single row, click the row letter on the left or right
side of the plate.
•Clear specific wells by clicking on a selected well.
•To use a Sample Names file to select the wells, see
“Selecting Wells using a Sample Names File” on page 35.
Operation26
8Select the desired sipping order for the selected wells, Row-
wise, Column-wise, or File Order.
9Click the Output Tab in the Start Run window.
10 To change the Data Path, either type the desired path or click
the Browse button and select the desired location for the data
files. Clicking the Default button restores the default data path.
NOTE
Data files should be saved to a local folder on the computer's hard
drive. Saving data files to a network drive may ca use loss of data if
the network connection is slow or interrupted.
11 If desired, select the Create Daily Sub-Directory check box to
create a new sub-directory for data files each day.
12 If the 21 CFR Part 11 option is installed and you want to save a
copy of the data files to a folder outside of the CDR or if the 21
CFR Part 11 option is not installed and you want to save a
second copy of the data files, see “Automatically Exporting
Copies of Data Files” on page 126.
V4.2LabChip GX User ManualPerkinElmer
Running an Assay (Continued)
13 To add the File Prefix, Computer Name, Project Name,
Barcode, Date, and/or Time to the data file name, select or
clear the desired check boxes or type the desired file prefix and
project name. The file name components are added to the data
file name in the order in which they are selected. Leave the text
box blank to omit the File Prefix or Project Name from the file
name.
14 To automatically export data tables, graphs, or gels, select the
Automatic Export check box, click the Auto Export Settings
button, and then select the desired Auto Export settings (see
page 29).
15 Click the Advanced Tab in the Start Run window.
16 To perform the assay multiple times on the same plate, select
the desired number of times to run the assay in the Plate
Cycles text box.
Operation27
17 To randomly sample a specific percent of the selected wells,
select the Random Selection check box and specify the
percent of wells to sample during the run.
18 To repeatedly run the selected wells and combine the data into
one data file, select the Sample Saver check box and select the
number of times to repeatedly run the selected wells in the
Repeats text box.
19 To use a file to supply the sample names, click the Browse
button next to Sample Name s File, select the name of the .csv
file that contains the sample names, and click the Open button.
The path and name of the file displays in the text box.
To use a Sample Names file to select the wells, see “Selecting
Wells using a Sample Names File” on page 35.
20 To use a file to supply the Expected Peaks, click the Browse
button next to Expected Peak s File , select the name of the
.gep file that contains the expected peaks, and click the Open
button. The path and name of the file displays in the text box.
(See “Using Expected Fragments/ Expected Proteins/ Expected
Glycans” on page 67 for more information.)
21 To use a file to supply the Excluded Peaks, click the Browse
button next to Excluded Pe aks Fi le, select the name of the
.gep file that contains the excluded peaks, and click the Open
button. The path and name of the file displays in the text box.
V4.2LabChip GX User ManualPerkinElmer
Operation28
22 Click the Start button to start the assay. A new workspace
opens to display the data.
See “Monitoring the Run” on page 30 for information about viewing
data during the run.
V4.2LabChip GX User ManualPerkinElmer
Operation29
Select the Auto Export Settings
The Auto Export settings specify which views to automatically
export at the end of each run and specifies the format for each view.
Click the Auto Export Settings button in the Output Tab on the
Start Run Window to open the Export Window.
To select the desired views to export:
1Select the check boxes next to the views to export. Selecting
Export All selects all check boxes.
2For each selected view, to change the location for the files, click
the Browse (...) button and select the desired location.
If the default path is selected, the path displays in blue. If the
path is not the default path, the path displays in purple. To set
the path to the default, right-click on the path and select Reset
to Default.
3If Raw Data is selected, click the AIA Format check box to
export in Chromatography Data Interchange Format or clear the
check box to export in CSV format. If CSV is selected:
•Select Include Size Data to align the data to the well's
ladder (for one file per well) or to the first well (for a single
data file) and include the size data in the exported data. (Not
available for CZE assays.)
•Select Export Single Table to export the data for all wells in
the plate to one CSV file. If not selected, the data from each
4If Gel is selected:
well is exported to a separate CSV file.
aSelect either Single File to include gels for all wells in the
run in the same image file, or select Separate Files to
export each gel to a separate image file.
bTo adjust the contrast minimum and maximum values for all
gels to a specific lane in the gel, choose the desired well in
the Contrast Lane drop-down list. (Only available after the
sample wells have been selected on the Run tab.)
cIf desired, change the height, in pixels, of the exported gel
graphics in the Height text box.
5If either Electropherogram or Gel is selected, choose the
desired format for the image files.
6Click OK to save the Export settings. The specified files are
exported at the end of the run.
V4.2LabChip GX User ManualPerkinElmer
Monitoring the Run
The following occurs after a run is started in the LabChip GX
software:
1The priming and warming steps are performed.
•The priming step fills the channels of the chip with reagent.
(Only performed if the chip holder has been opened since
the last run.)
•The warming step allows the heater plate located in the chip
holder to regulate chip temperature to 30°C.
•Data collection begins after the warming and priming steps
are completed.
2After the run begins, the Start button on the LabChip GX main
window changes to Stop.
3The Active Data Tab displays the electropherogram of the well
currently being read. Data is saved to a file with the name
shown above the plate diagram after each well is completed.
Operation30
4To view the results for individual wells as data is acquired or
after the run is finished, click a well in the Plate View or Plate
List, a sample name in the Well Table View, or a lane in the Gel
View. Data from the selected well displays in the Overlay
Electropherograms Tab and the Peak Table View.
For more information on data analysis, see “Data Analysis” on
page 42.
5To stop the run before it is complete, see Stopping a Run.
6When the assay is complete, Run Successful displays in the
Status line.
7If desired, remove the plate and/or remove the chip.
To save the analyzed data:
1Select File Sav e Workspace. The Save Workspace As
Window opens with the default workspace name as the name of
the data file.
2If desired, change the location and/or the name of the
workspace file and click the Save button. (Workspace files have
a .gxw file extension.)
To view or re-analyze the data, see Data Analysis.
V4.2LabChip GX User ManualPerkinElmer
Stopping a Run
If you need to stop the run before it is complete, click the Stop
button.
A message box confirms that you want to stop the run in progress.
Click Yes to stop the run. The Status line displays Run stopped by user.
Operation31
Data for any completed wells displays in the LabChip GX Main
Window.
To continue to read wells from an aborted run, see Continuing a
Stopped Run.
V4.2LabChip GX User ManualPerkinElmer
Continuing a Stopped Run
If a run is stopped before it is complete, you can restart the run to
finish the reading the plate. When you select the wells for the assay,
you should select only the wells that were not read so that the
reading starts with the well that was not completed.
To continue a run:
1Click Start in the upper-left corner of the main window to open
the Start Run Window.
2Select the wells that were not read in the previous run,
beginning with the well that was in progress when the run was
stopped.
3Select the same options for this run as were selected for the
stopped run.
4Click the Start button to begin.
Operation32
The run starts with the well that was not completed in the previous
run. A separate data file is created for the current run.
V4.2LabChip GX User ManualPerkinElmer
Saving Data Files
While running an assay, the raw time series data received from the
instrument is automatically saved to the data file (.gxd), one well at
a time as each well is completed. If a run is stopped before it is
complete, the data for the completed wells is saved in the data file.
The name of the data file is specified in the Output Tab on the Start
Run Window.
The analysis settings for a plate are saved at the end of the data file
(.gxd). If analysis settings are changed in the Assay Analysis
Window and the data file is saved, the new settings are added to
the end of the data file, but the previous settings are not
overwritten. This enables a plate to be restored to previous analysis
settings using the Restore Plate button in the Assay Analysis
window.
If the 21 CFR Part 11 option is installed, data files are saved in the
Centralized Data Repository (CDR). The CDR is a secure folder
that can only be accessed by the LabChip GX software. The
location of the CDR is specified in the CDR Utility Window.
Operation33
Updated analysis settings are saved when the plate data file is
saved. To save the data file, either:
•select Workspace Save Plate on the Menu Bar,
•right-click on the plate name in the Plate View and select Save Plate, or
•select File Save Workspace or File Save Workspace As
on the Menu Bar. (See Saving Workspace Files for more
information.)
If the 21 CFR Part 11 option is installed and Require Signature on
File Update is selected in the Set Policies tab on the User
Administration Window, the user must have signature rights to save
an updated data file. If the data file is locked, a user with Unlock
rights must unlock the data file before changes can be saved.
To save the data file when the 21 CFR Part 11 option is installed
and Require Signature on File Update is selected:
1The Perform Signature Window opens when you save a data
file.
2Select the name of the user who is signing the data file in the
Username drop-down list.
3Type a Comme nt that meets the requirements of the
compliance policies.
V4.2LabChip GX User ManualPerkinElmer
Saving Data Files (Continued)
4If appropriate, select the desired Approval State.
5If desired, select the Lock check box to lock the data file and
prevent it from being changed.
6Type the User Password for the signing user.
7Click the Sign button. The Signature Performed window
confirms that the signature has been performed.
8Click the OK button. The signature is recorded in the data file.
Note
Changing the analysis settings and sav ing the plate data file do es
not change the raw data from the run. Only the display of the data is
changed.
Operation34
If you change the analysis settings without saving the new settings,
and then try to close the workspace, exit the software, or acquire
new data, you are prompted to save the changes. Selecting Yes
opens the Save Workspace As Window.
Saving Workspace Files
When viewing plate data in the LabChip GX main window, you view
the selected data files in a Workspace. When a new run starts, a
new blank workspace opens, which contains the data file for the
plate in the run. After the run is complete, saving the workspace
saves the data file with the current analysis settings. (See Saving
Data Files for more information on the contents of the data files.)
The Data Files are not saved in the Workspace file. The workspace
file contains links to the revision of the data file that was last open in
the collection. If a Workspace file is moved to another folder or
computer, the data files must be moved with the Workspace file.
The data files must be in the same location relative to the
Workspace file as they were, otherwise you are prompted to find the
missing data files when you open the Workspace file. It is a good
practice to save the workspace files in the same location as the
data files that are included in the workspace.
Saving a Workspace file automatically saves any changes to the
data files that are open in the workspace, and the settings for each
collection in the workspace.
V4.2LabChip GX User ManualPerkinElmer
Operation35
If the 21 CFR Part 11 option is installed, saving Workspace files
does not require permission to save data files unless the data files
in the workspace have been modified.
Selecting Wells using a Sample Names File
The wells to be sampled during the run can be selected using a
Sample Names file. When using a Sample Names file to select the
wells, only wells with an entry in the Sample Names file are
selected for sampling. The format must match the file format
described in “Sample Name File Format (.CSV Format)” on
page 65.
The Sample Names file used to select the samples contains one
row for each well to be sampled, and does not contain rows for
wells that are not sampled.
Using the Sample Names file in Figure 4 to select the wells in the
run will select wells A1, A5, A11, A12, B1, B2, F5, and F6 for
sampling.
Figure 4. Sample Names File for Selecting Samples
Figure 5. Selecte d Wells on Run tab in Start Run Window
This Sample Names file cannot be created in the Sample Names
Editor window because the files created in the Sample Names
Editor window always contain one row for each well in the plate.
The Sample Names files for selecting wells are typically created by
a LIMS or can be created in Microsoft Excel.
V4.2LabChip GX User ManualPerkinElmer
Operation36
To select the wells using a Sample Names file:
1Create the sample names file with one row for each well to be
sampled.
2Click the Run button on the LabChip GX Main Window. The
Start Run Window opens.
3Click the Advanced Tab.
4Click the Browse button (...) next to the Sample Names text
box, select the sample names file, and click the Open button.
5Select the Use this file for sample selection check box.
Figure 6. Use This File For Sample Selection
6Click the Run Tab.
7Select the desired Sipping Order: Row-wise, Column-wise, or
File Order. (See page 250 for details.)
8Select any other run parameters as desired, and then start the
run.
Using a Barcode to Specify the Sample Names File
When using a LIMS, a Sample Names File can be specified by
using the plate barcode as the name of the file. The Sample Names
file is read at run-time, after reading the plate barcode.
To use the plate barcode as the name of the Sample Names file:
1Decide on a location and file name format for the Sample
Names files. For example, the Sample Names files can be
named “Sample Names” followed immediately by the barcode
on the plate.
2Create a temporary CSV file, using “%barcode%” as a
placeholder for the barcode string. For example, create a
temporary file named “Sample Names%barcode%.csv” and
save it in the location where the actual sample names files for
the runs will be saved.
3In the Run File Editor Window, select the desired run
parameters.
V4.2LabChip GX User ManualPerkinElmer
Operation37
4On the Advanced tab, click the Browse button next to Sample
Names File, and select the CSV file that contains the
%barcode% wildcard. The Use this file for Sample Selection
check box is automatically selected when the sample names file
contains the string %barcode%.
5On the Run tab, select the desired Sipping Order: Row-wise,
Column-wise, or File Order.
6Click the Export button and save the Run file.
7To start the run, click the Run button. The Start Run Window
opens.
8Click the Import button and select the saved Run file.
9When the run is started, the plate barcode is read and the
Sample Names file with the corresponding barcode is used to
select the wells to sample. In the example above, if the plate
barcode is 12345, then the file named “Sample
Names12345.csv” will be used to select the wells in the run.
V4.2LabChip GX User ManualPerkinElmer
Adding a New Plate
When using a plate other than the plates specified in the Start Run
Window, you must create a new plate in the Plate Information
Window. Use caution adding new plates. Entering wrong values in
the Add Plate Window can result in damaged chips and broken
sippers.
It is best to use the plate specifications provided by the plate
manufacturer. If the specifications are not available, measure the
plate with a caliper. Many plates have a large variation in the Z-axis
location of the well bottom.
Make sure the Sip Height has enough margin to accommodate this
variation:
•Minimum Sip Height is 2.5mm.
•Maximum Sip Height depends on the sample volume.
•Sip Height of 4mm is probably safe.
To add a new plate:
Operation38
1Select ToolsPlate Editor to open the Plate Information
Window.
2Click the Custom Plates tab.
3Click the Add Plate button. The Add Plate Window opens.
4Type the settings for the new plate. The diagram on the Add
Plate window shows the location for each measurement.
Acceptable plate parameters are:
•PlateHeight: 0 to 16mm
•WellDepth: 0 to 36mm
•SipHeight >= 2mm
•(WellDepth – SipHeight) <= PlateHeight
5Click the OK button.
6Perform a Punch Test from the Plate Information window to
verify the new plate settings are acceptable.
•With a piece of the tape provided by PerkinElmer, cover the
corner wells on the plate.
•Use an old, dead chip with a good sipper.
•In the Plate Editor, select the new plate name and click the
Verify Plat e button.
•Follow the onscreen instructions.
V4.2LabChip GX User ManualPerkinElmer
Adding a New Plate (Continued)
•The instrument will move to the specified number of wells in
each corner of the plate, punching holes in the tape.
•The Punch test is acceptable if the punched hole is >0.5mm
from the edge of the well.
Operation39
•Holes will never be perfectly centered.
Possible sources of errors:
•Sipper splay
•Plate tolerances
•Robot alignment error
•If the holes are too close to the edge (<0.5mm), adjust the XMargin or Y-Margin values in the Add Plate window.
7Select the new plate name in the Start Run Window when
starting the run.
V4.2LabChip GX User ManualPerkinElmer
Placing the Barcode on the Plate
Figure 7 shows the size limits for the barcode label and the location
on the microplate where the label should be placed. The barcode
must be located on the short (portrait) end of the microplate, closest
to well A1. If the barcode is not positioned properly, the barcode
reader will not be able to read the barcode.
Operation40
V4.2LabChip GX User ManualPerkinElmer
Figure 7. Ba rcode Label Po sition
Calibrating the Optics
The LabChip GX software provides an Optical Calibration function
using Test Chip C to calculate a correction factor for each individual
LabChip GX instrument. The correction factor is applied to all
samples run on the instrument to provide a common absolute
fluorescence across different instruments.
See “Maintenance and Service” on page 280 for recommendations
on optical calibration frequency. Optics Calibration should also be
run if the Optics Calibration Test in the Diagnostics fails.
To calibrate the optics:
1Insert Test Chip C in the instrument.
2Select Tools Calibrate Optics to open the Optics Calibration
Window. The laser and Test Chip warming step starts. Warm-up
takes 10 minutes.
3Wait while the scan is performed. The scan takes 5 minutes.
When the scan is complete, the correction factor is calculated
and displays in the New Correction Factor text box.
Operation41
4To apply the new calibration factor to the instrument, click the
Apply button. All future runs will be scaled by the new
calibration factor.
Walk-Away Operation
After the scan has started, you can leave the instrument. At the end
of the scan, the chip warming and laser are turned off after one
minute, but the correction factor displayed in the window can be
applied when the user returns to the instrument.
If the New Correction Factor is not within the factory-set limits, the
text box displays a red background and an error message displays.
Please contact PerkinElmer technical support (see page 3) to
determine the problem.
Failed to Find Focus Point
If the focus point cannot be found during a scan, retry the scan. If
the error still displays and another Test Chip C is available, try the
scan with a different Test Chip C. The laser and chip warming
remain on for one minute to allow you to retry the scan.
V4.2LabChip GX User ManualPerkinElmer
Data Analysis42
Data Analysis
After a run is complete, use the LabChip GX software to view and
analyze the plate data. The LabChip GX software can open multiple
data files in the same workspace to compare the data from different
plates. Analysis settings can be changed for single wells, entire
plates, or all plates in the workspace.
This section contains the following information:
•How the Software Analyzes DNA Data
•How the Software Analyzes Protein Data
•How the Software Analyzes RNA Data
•How the Software Analyzes Glycan Data
•How the Software Analyzes Protein Charge Variant Data
•How the Software Analyzes Genomic DNA Data
•Organizing, Retrieving, and Backing Up Data Files
The LabChip GX DNA Assay Analysiscalculates the size and
concentration of nucleic acid fragments. Results for each well are
calculated after all data for that well has been read.
The data analysis process for DNA assays consists of the following
steps:
1Raw data is read and stored by the system for each individual
well.
2The data is filtered (see “Data Filtering” on page 307) and the
resulting electropherograms of all wells are plotted.
3Peaks are identified (see “Understanding Peak Finding” on
page 108) for all wells and are tabulated by migration time. The
settings of the peak find algorithm can be changed and the data
can be reanalyzed after the run has finished. (See “Reanalyzing
a Data File” on page 95.) The peak find settings can be changed
for all wells or only for specific wells.
Data Analysis43
4A sizing ladder (see Figure 8), which is a mixture of DNA
fragments of different known sizes, is run first from the ladder
vial. The concentrations and sizes of the individual base pairs in
the ladder can be changed on the Analysis Tab.
Figure 8. Ladde r Graph - DNA
V4.2LabChip GX User ManualPerkinElmer
How the Software Analyzes DNA Data
(Continued)
5The ladder is analyzed (see “Ladder Analysis” on page 311) and
a standard curve of migration time versus DNA size is plotted
from the DNA ladder by interpolation between the individual
DNA fragment size/migration points. The standard curve derived
from the data of the ladder well should resemble Figure 9.
Data Analysis44
Figure 9. Standard Curve Window - DNA
6Two DNA fragments, the Lower Marker and Upper Marker, are
run with each of the samples, bracketing the DNA sizing range.
The Lower Marker and Upper Marker are internal standards
used to align the ladder data with data from the sample wells.
Figure 10 shows an example of assigned marker peaks in a
sample well.
Figure 10. Markers - DNA
V4.2LabChip GX User ManualPerkinElmer
How the Software Analyzes DNA Data
(Continued)
7The standard curve and the markers are used to calculate DNA
fragment sizes for each well from the migration times measured.
8To calculate the concentration of the individual DNA fragments
in all sample wells, the upper marker, in conjunction with a
calibration curve plotting assay-specific concentration against
base-pair size, is applied to the individual sample peaks in all
sample wells.
If a titer is being used, the ladder is not used for quantization.
Instead, peak sample areas are normalized using the upper
marker and then quantization is performed using the titer
standard curve to convert normalized area to concentration; see
“Titer” on page 328.
NOTE
Data Analysis45
The software allows you to define upper and lower markers.
Changing the markers causes quantitative changes in the
calibration procedure, however, and therefore in the entire data
evaluation.
How the Software Analyzes Protein Data
Protein assays are only supported on LabChip GX II instruments.
The LabChip GX Protein Assay Analysis consists of the following
steps:
1Raw data is read and stored by the system for each individual
well.
2The data is filtered (see “Data Filtering” on page 307) and the
resulting electropherograms of all wells are plotted.
3Peaks are identified (see “Understanding Peak Finding” on
page 108) for all wells and are tabulated by migration time. The
settings of the peak find algorithm can be changed and the data
can be reanalyzed after the run has finished. (See “Reanalyzing
a Data File” on page 95.) The peak find settings can be changed
for all or only certain wells.
V4.2LabChip GX User ManualPerkinElmer
Data Analysis46
How the Software Analyzes Protein Data
(Continued)
4A sizing ladder (see Figure 11), which is a mixture of protein
fragments of different known sizes, is run from the ladder vial
before and after every 12 samples. The concentration and sizes
of the individual proteins (in kDa) in the ladder can be changed
on the Analysis Tab.
Figure 11. Ladder Graph - Protein
5The ladder is analyzed (see “Ladder Analysis” on page 311) and
a standard curve of migration time versus mobility is plotted
from the ladder by interpolation between the individual protein
size/migration points. The standard curve derived from the data
of the ladder well should resemble Figure 12.
V4.2LabChip GX User ManualPerkinElmer
Data Analysis47
How the Software Analyzes Protein Data
(Continued)
Figure 12. Standard Curve - Protein
6A Lower Marker is the internal standard used to align the ladder
data with data from the sample wells. Figure 13 shows an
example of the assigned Lower Marker peak (marked LM) in a
sample well.
Figure 13. Marke rs - Protein
V4.2LabChip GX User ManualPerkinElmer
Data Analysis48
How the Software Analyzes Protein Data
(Continued)
7Before calculating either the protein size or concentration, the
sample data is processed relative to the two ladders that
bracket every 12 samples. First, all of the data is aligned to the
lower marker and then stretched, relative to the highest
molecular weight protein in the ladder.
8The concentration of the sample proteins is determined relative
to the bracketed ladders. There are seven proteins in the ladder
(not including the lower marker). The ladder concentration can
be changed on the Analysis Tab (see page 189).
9If a titer is being used, the ladder is not used for quantization.
Instead, peak sample areas are normalized using the lower
marker and then quantization is performed using the titer
standard curve to convert normalized area to concentration; see
“Titer” on page 328.
NOTES
•The seven proteins are assumed to represent th e average
staining behavior of proteins and are used to determine an
average peak area per g/mL o f protei n. Thi s factor is used to
convert sample peak areas into relative concentration. The total
relative concentration of all proteins is calcula ted by addition of
the relative concentration of the individ ual proteins and is
displayed in the Well Table View (see page 172).
•Since the conversion factor is determined for each ladder, the
factor can be linearly interpolated for each sample and thus
account for any assay drift.
•The software allows you to define the lower marker. Changing
the selection of the marker will lead to quantitative changes in
the calibration procedure, and theref ore in the entire data
evaluation.
V4.2LabChip GX User ManualPerkinElmer
How the Software Analyzes RNA Data
The LabChip GX RNA Assay Analysis determines the quality of the
RNA sample by measuring the relative amounts of know RNA
fragments relative to the total RNA present in the sample. Results
for each well are calculated after all data for the well has been read.
The data analysis process for RNA assays consists of the following
steps:
1Raw data is read and stored by the system for each individual
well.
2The data is filtered (see “Data Filtering” on page 307) and the
resulting electropherograms of all wells are plotted. The filtering
algorithm settings can be changed and the data can be
reanalyzed after the run is finished. (See “Reanalyzing a Data
File” on page 95.)
3A curve spline fit to the data is performed to generate a baseline
above which RNA fragment peaks are detected. This baseline is
displayed as a blue line on the electropherogram when Show
Peak Baselines is selected in the Graph View Properties (see
page 164).
Data Analysis49
4Peaks extending above the baseline are identified (see
“Understanding Peak Finding” on page 108) for all wells and are
tabulated by migration time. The settings of the peak find
algorithm can be changed and the data can be reanalyzed after
the run has finished. The peak find settings can be changed for
all or only certain wells.
5A sizing ladder (see Figure 14), which is a mixture of RNA
fragments of different known sizes, is run first from the ladder
vial. The concentrations and sizes of the individual nucleotides
in the ladder can be changed on the Analysis Tab.
V4.2LabChip GX User ManualPerkinElmer
How the Software Analyzes RNA Data
(Continued)
Data Analysis50
Figure 14. Ladder Graph - RNA
6The ladder is analyzed (see “Ladder Analysis” on page 311) and
a standard curve of migration time versus RNA size is plotted
from the RNA ladder by interpolation between individual RNA
fragment size/migration points. The standard curve derived from
the data of the ladder well should resemble Figure 15.
Figure 15. Standard Curve - RNA
V4.2LabChip GX User ManualPerkinElmer
How the Software Analyzes RNA Data
(Continued)
7A dye matching the lowest peak in the ladder is run with each of
the samples. This lower marker, labeled LM in the RNA sample
(see Figure 16) is used to align the ladder data with data from
the sample wells.
Data Analysis51
Figure 16. Markers - RNA
8The standard curve and the markers are used to calculate RNA
fragment sizes for each well from the migration times measured.
V4.2LabChip GX User ManualPerkinElmer
How the Software Analyzes RNA Data
(Continued)
9The Total RNA present is computed by finding the area under
the electropherogram trace. The baseline for this integration is a
straight line starting at the end of the lower marker and ending
at the baseline end time. The height of the baseline endpoints is
computed from an average of a five second window around the
baseline Start Time and End Time (shown on the Peak Find
Tab ). View the baseline by selecting Show Peak Baselines in
the Graph View Properties. Adjust the Start Time and End Time
by dragging the left (Start) and right (End) vertical dashed lines
to areas that more properly reflect the signal baseline. (Rightclick in the graph, select Set Scale, and change the X axis
Minimum and Maximum values if the start and end times are not
shown in the graph.)
Data Analysis52
Figure 17. Pea k Baseline - RNA
10 The Total RNA concentration in the sample is computed from
the ratio of the RNA area in the sample to the RNA area in the
ladder multiplied by the ladder concentration specified in the
assay.
If a titer is being used, the ladder is not used for quantization.
Instead, peak sample areas are normalized using the lower
marker and then quantization is performed using the titer
standard curve to convert normalized area to concentration; see
“Titer” on page 328.
V4.2LabChip GX User ManualPerkinElmer
How the Software Analyzes RNA Data
(Continued)
11 Assay-defined RNA fragments are identified from the peaks in
the peak table. Fragments are located by finding the largest
peak within a size range associated with the fragment. For
Eukaryote RNA assays, 5S, 18S and 28S fragments are
located. For Prokaryote assays, 5S, 16S and 23S fragments are
identified.
12 The following values are calculated for RNA assays:
Fragment_Area: Area of each rRNA Fragment.
%_of_Total_Area: Each Fragment area as a percent of total
area.
Corrected RNA Area: Total RNA Area corrected with Lower
Marker height.
RNA Concentration (ng/ul): Estimated Total RNA
Concentration in the sample.
rRNA Area Ratio [28S / 18S]: 28S area divided by 18S area.
rRNA Height Ratio [28S / 18S]: 28S height divided by 18S
height or 23S height divided by 16S height.
rRNA Fast Area Ratio: Region between 5S and 18S or 16S,
percent of total area.
RNA quality metrics: rRNA Area and Height Ratios (28S/18S
or 23S/16S) and Fast/Total RNA area ratio are computed.
Data Analysis53
Figure 18. rRNA Graph Analysis
13 Messenger RNA Assay: The RNA contamination ratio is
computed. This is the ratio of the area of all the fragments to
total RNA area.
V4.2LabChip GX User ManualPerkinElmer
Data Analysis54
How the Software Analyzes Glycan Data
Glycan assays are only supported on LabChip GX II instruments.
The LabChip GX Glycan Assay Analysis consists of the following
steps:
1Raw data is read and stored by the system for each individual
well.
2The data is filtered (see “Data Filtering” on page 307) and the
resulting electropherograms of all wells are plotted.
3Peaks are identified (see “Understanding Peak Finding” on
page 108) for all wells and are tabulated by migration time. The
settings of the peak find algorithm can be changed and the data
can be reanalyzed after the run has finished. (See “Reanalyzing
a Data File” on page 95.) The peak find settings can be changed
for all or only certain wells.
4A sizing ladder (see Figure 19), which consists of multiples of a
basic sugar molecule, is run from the ladder vial before and after
every 12 samples. The simplest molecule is assigned a size unit
of 1 CGU (Caliper Glucose Unit). The sequence of ladder peaks
is assigned integer size values, providing a migration time to
size ruler for sizing other sugar molecules.
Figure 19. Ladder Graph - Glyc an
5The ladder is analyzed (see “Ladder Analysis” on page 311) and
a standard curve of migration time versus mobility is plotted
from the ladder by interpolation between the individual
glycan/migration points. The standard curve derived from the
data of the ladder well should resemble Figure 20.
V4.2LabChip GX User ManualPerkinElmer
Data Analysis55
How the Software Analyzes Glycan Data
(Continued)
Figure 20. Standard Curve - Glycan
6A Lower Marker is the internal standard used to align the ladder
data with data from the sample wells. Figure 21 shows an
example of the assigned Lower Marker peak (marked LM) in a
sample well.
Figure 21. Markers - Glycan
V4.2LabChip GX User ManualPerkinElmer
Data Analysis56
How the Software Analyzes Glycan Data
(Continued)
7The standard curve and the markers are used to calculate
glycan sizes in each well from the migration times measured.
8The area under each peak is calculated.
9All the sample peak areas are added together and an area ratio
is calculated for the %Area for each peak.
NOTES
•In Glycan assays, the lower marker does not precede all ladder
peaks but appears between the 6th an d 7th ladder peak. The
concentration of the lower marker has been made sufficiently
large to avoid confusion with ladder peaks. Si nce peaks befo re
the lower marker are ignored, the ladder peaks used for the
sizing ruler range from 7 to 13 CGU.
•The software allows you to define the lower marker. Changing
the selection of the lower marker will lead to quantitative
changes in the calibration, and therefore in the entire data
evaluation.
V4.2LabChip GX User ManualPerkinElmer
Data Analysis57
How the Software Analyzes Protein Charge
Variant Dat a
Protein Charge Variant assays are only supported on LabChip GX II
instruments.
The LabChip GX Protein Charge Variant Assay Analysis consists of
the following steps:
1Raw data is read and stored by the system for each individual
well.
2The data is filtered (see “Data Filtering” on page 307) and the
resulting electropherograms of all wells are plotted.
3A baseline is generated for the data using the settings on the
Peak Find tab on the Assay Analysis Window. This baseline is
displayed on the electropherogram when Show Peak Baselines
is selected in the Graph View Properties (see page 164).
4Peaks are identified (see “Understanding Peak Finding” on
page 108) for all wells and are tabulated by migration time. The
settings of the peak find algorithm can be changed and the data
can be reanalyzed after the run has finished. (See “Reanalyzing
a Data File” on page 95.) The peak find settings can be changed
for all or only certain wells.
5The % Relative Amount for each peak is calculated as a percent
of total area.
V4.2LabChip GX User ManualPerkinElmer
Data Analysis58
How the Software Analyzes Genomic DNA Data
The Genomic DNA (gDNA) assay analyzes data similar to the other
DNA assays, except that the alignment of the sample data with the
ladder data is based solely on the Lower Marker. To improve the
accuracy of the alignment of the 12 samples between ladders,
gDNA assays use the bracketed alignment described in “How the
Software Analyzes Protein Data” on page 45.
For calculating concentration, the Genomic DNA assay uses a GQS
Smear, starting at 0.175 kb and extending to 300 kb, by default (see
Figure 22). The start size of the gDNA smear can be changed on
the Advanced Tab in the Assay Editor window. The calculation of
concentration is similar to that used for individual peaks, as
described in How the Software Analyzes DNA Data; however, the
GQS Smear is used in place of individual sample peaks, and the
lower marker is used in place of the upper marker to normalize
areas. The Total gDNA Concentration of each sample is reported in
the Well Table.
The Genomic DNA assay also reports a Genomic Quality Score
(GQS) for each sample in the Well Table. The GQS represents the
degree of degradation of a sample, with 5 corresponding to intact
gDNA and 0 corresponding to highly degraded gDNA. The GQS is
calculated using the size distribution of the sample.
Figure 22. Genomic DNA GQS Smear (in orange)
V4.2LabChip GX User ManualPerkinElmer
Data Analysis59
Organizing, Retrieving, and Backing Up Data
Files
As you work in the LabChip GX software, it's a good practice to
organize the LabChip GX files.
•Create a folder in which to save the data files. If desired, each
person can save data files to their own subfolder to organize the
data files.
•Save Workspace files in the same directory as the data files to
prevent missing data files in the workspaces.
•Review the files periodically, even if only one person uses the
LabChip GX software. Archive files you are no longer using but
want to save to a backup disk, and discard unneeded files.
Verify there is enough free space on the hard drive to save new
plate data files. If you are using the 21 CFR Part 11 option, see
“Software Security” on page 114 for backup options.
•Each user in the laboratory can specify a particular data file
name prefix to easily differentiate data files.
•A new folder can be created each day to store the data from all
runs. To automatically create the folders, select the Create Daily Sub-Directory check box on the Output Tab in the Start
Run Window.
Opening a New Workspace
A workspace displays data from one or more plate data files from
the same type of assay. Each workspace can contain one or more
Collections to display the data.
To open a new workspace:
1On the menu bar, select File New Workspace. If changes
have been made to an open workspace, you are prompted to
save any unsaved changes. A blank workspace opens in the
LabChip GX Main Window.
To view data, see:
•Opening a Data File
•Adding a Collection to a Workspace
•Selecting the Wells in a Collection
•Modifying Analysis Parameters
•Saving and Exporting Assays
V4.2LabChip GX User ManualPerkinElmer
Opening a Data File
Open a data file to view the data, to compare the data to other data
files in the same workspace, or to change analysis setting and view
the reanalyzed data. Data files generated by the LabChip GX
software have a .gxd file extension. Data files generated by the
LabChip HT software, which have a .cla file extension, can also be
opened in the LabChip GX software if the 21 CFR Part 11 option is
not installed.
To open a data file:
1Open a new workspace (see page 59) or a workspace that
already contains compatible data files.
2On the menu bar, select FileImport Data File. If the 21 CFR
Part 11 option is not installed, the Select a Data File Window
opens. If the 21 CFR Part 11 option is installed, the CDR
Manager Window opens.
Data Analysis60
3Select the name of the data file to open and click the Open
button. The selected data file is imported into the open
workspace. Use Ctrl+click or Shift+click to select multiple files.
•To select a .cla file, select CLA from the Files of Type drop-
down list in the Select a Data File window. The selected .cla
file is converted to a .gxd file, and the new .gxd file is
imported into the open workspace. The new .gxd data file is
saved in the same folder as the original .cla file. Any
changes are saved to the .gxd file. The .cla file cannot be
changed.
NOTE
After the .cla file is converted to a .gxd file, use the .gxd file in any
workspaces. Re-importing the same .cla file will over-write the
existing .gxd file and any analysis changes saved in the .gxd file will
be lost.
4If desired, right-click on the data file name in the Plate View or
Plate List and select Rename Plate to change the name of the
plate in the display. (If desired, select the Rename File to Match
check box to change the name of the data file.)
5Select the desired wells to view in each collection in the
workspace (see page 62).
6See “Modifying Analysis Parameters” on page 72 for information
about changing analysis parameters.
V4.2LabChip GX User ManualPerkinElmer
Adding a Collection to a Workspace
Collections are used to specify the wells selected for view in each
plate data file, the layout of the views in the Collection tab, and the
display properties for each view in each collection.
To add a new collection to an open workspace:
1On the menu bar, select Collection New Collection. The
New Collection Window opens.
2Select the desired option for creating the new collection:
•Template - Opens a new collection with the same settings
as a saved collection template file.
•Blank Collection - Opens a new template using the default
collection settings.
•Current Collection - Opens a new template based on the
current settings in the currently open collection.
3If no plates are open in the collection, choose the desired Assay
Type for the collection: DNA, Protein, RNA, Glycan, CZE, or
gDNA. (Protein, Glycan, and CZE are only supported on
LabChip GX II instruments.) Only data files from the selected
assay type can be imported into the workspace.
Data Analysis61
4If desired, type a new name for the collection in the Name text
box.
5Click the OK button to open the new collection.
V4.2LabChip GX User ManualPerkinElmer
Selecting the Wells in a Collection
Each collection can display different wells or the same wells in
different orders. To select the wells to display in a collection:
1Select the collection where the wells will be displayed.
2In the plate diagram in the Plate View, select the wells to
display:
Figure 23. Selecting Wells
Data Analysis62
•Select individual wells: click on the well.
•Select a block of wells: click on the well in one corner and
drag to the opposite corner well.
•Select an entire row: click the row letter at the left.
•Select an entire row and the corresponding ladder: click
the row letter at the right (if the plate sip order is row-wise).
•Select an entire column: click the column number at the
top.
•Select an entire column and any corresp onding ladd er:
click the column number at the bottom (if the plate sip order
is column-wise).
•Select all the wells by rows: click the button in the top left
corner of the microplate.
•Select all the wells by columns: click the button in the top
right corner of the microplate.
•Select all ladder wells and microplate wells: click the
button at the bottom right corner of the microplate.
•Select individual ladders: click on the ladder.
•Clear individual ladders: right-click on the ladder and
select Remove Ladder.
•Select all ladder wells: click the “L” icon at the bottom left
corner of the microplate.
•Clear individual wells: right-click on the well and select
Remove Well.
3Save the workspace to save the wells selected in the collection.
V4.2LabChip GX User ManualPerkinElmer
Using Sample Names Files
Sample Names files are used to import sample names into a data
file. The Sample Name file can also contain expected peaks. A
sample names file can be selected in the Advanced Tab on the Start
Run Window to assign the sample names and expected peaks as
the assay is running. Before the run, a Sample Name file can also
be used to select the wells and sipping order for the run (see
“Selecting Wells using a Sample Names File” on page 35).
After the run is complete, you can use the Sample Name Editor
Window to rename the samples or to import an existing sample
name file. Sample name files are saved as .CSV files and can be
edited with a spreadsheet program such as Microsoft Excel.
The Color, Name, and Display Property for an Expected Peak
cannot be set in the Sample Name Editor window. When new
Expected Peaks are created in the Sample Name Editor window,
default values are assigned for these properties.
Data Analysis63
Create a Sample Name File
Sample Names files created in the Sample Name Editor contain a
row for each well in the plate. To use the Sample Name file to select
the wells in the run, manually create the CSV file in Microsoft Excel
or use a LIMS to create the Sample Names file.
To create a Sample Names file to specify the sample names,
comments, and Expected Peaks:
1Select Tools Sample Name Editor on the LabChip GX Main
Window. The Sample Name Editor Window opens.
2Modify the sample names as desired in the Sample Name
Editor window.
3Enter any desired comments for the samples in the Sample
Comment column.
4For DNA, gDNA, Protein, or Glycan assays, if desired, click the
Edit Expected Peaks check box and enter the desired
Expected Peaks and Window (%) in the columns. Separate
each peak size or window value with a semi-colon (;). If only one
Window % value is entered, the same value applies to all
expected peaks. If multiple values are entered, separated by
semi-colons, each expected peak will use the corresponding
Window value. If more expected peaks are entered than
Window values, the last Window applies to all remaining
expected peaks in the list.
V4.2LabChip GX User ManualPerkinElmer
Data Analysis64
5Click the Export button, navigate to the desired location for the
file, type the desired name for the file, and click the Save button.
V4.2LabChip GX User ManualPerkinElmer
To import the sample names into a data file:
1Open the data file in an open workspace in the LabChip GX
Main Window.
2Select Tools Sample Name Editor on the LabChip GX Main
Window. The Sample Name Editor Window opens.
3Click the Import button, navigate to the file location, select the
name of the .csv file that contains the sample names, and click
the Open button.
If a Sample Names file is imported into the Sample Name Editor
window, rows are added for any wells that are not listed in the
original Sample Names file. All wells are included in the new
Sample Names file.
4Click the Apply button in the Sample Name Editor window to
apply the sample names to the active plate.
Sample Name File Format (.CSV Format)
Data Analysis65
The Sample Names created in the Sample Name Editor Window
can be exported to a .CSV file. A .CSV file can also be created in
Microsoft Excel or generated automatically with a LIMS to import
the Sample Names into the data file.
Figure 24 shows a Sample Names file open in Microsoft Excel.
Figure 24. Sample Nam e .CSV File
A Sample Name .CSV file must use the following format:
Column A: The well label (A1 to P24). The wells labels can be
entered in any order and do not need to cover the entire plate.
Column B: The desired sample name for each well.
Column C: A user comment to be added to the well properties.
Column D: A list of Expected Peak sizes for the well. This list is
separated by semi-colons (;).
V4.2LabChip GX User ManualPerkinElmer
Data Analysis66
Column E: The search window size for each Expected Peak in
column D as a % (10 = +/-10% of Expected Peak size). If all
Expected Peaks use the same window size, enter a single value;
otherwise enter unique values as a semicolon separated list. If
fewer windows than EPs are entered, the last window applies to all
remaining EPs in the list.
Column F… Any higher columns are ignored.
Applying Different Window Values to Expected Peaks of the
Same Size in Different Wells
Only one Expected Peak of a particular size can exist in the assay
analysis settings. If the same Expected Peak size with a different
Expected Peak window is entered in the Sample Name Editor
window, only the first Expected Peak is created. All other Expected
Peaks of the same size use the same window as the first Expected
Peak. To overcome this restriction, use Expected Peaks of slightly
different sizes; e.g. 200, 200.1, 200.2, etc. to apply different
windows to the same Expected Peak in different wells.
Sample Name File Import Errors
Any row in the .CSV file that does not match the expected format
and cannot be interpreted is ignored. If none of the rows can be
interpreted, usually because the first column does not contain the
well label, the following error message displays: “No rows found
matching format: label, name, comment, EP, window.”
V4.2LabChip GX User ManualPerkinElmer
Data Analysis67
Using Expected Fragments/ Expected Proteins/
Expected Glycans
You can track expected DNA fragments (EFs), proteins (EPs),
Glycans (EGs), or peaks (EPs) for the samples in a DNA, gDNA,
Protein, Glycan, or Protein Charge Variant assay, respectively.
(Protein, Glycan, and Protein Charge Variant assays are only
supported on LabChip GX II instruments.) You enter the EFs, EPs,
and EGs in the Assay Analysis Window.
Entering EFs, EPs, or EGs in the Assay Analysis Window
1Select Analysis Analysis Settings on the Menu Bar. The
Assay Analysis Window opens.
2Click on the Expected Fragments/Proteins/Glycans Tab.
3Click in the bottom (empty) row in the table.
4In the Size column, type the size of the expected fragment (bp
for DNA or kb for gDNA), protein (kDa), or glycan (CGU).
5For CZE assays, type the time of the expected peak, in
seconds, in the Time Column.
6In the Window (%) column, if desired, change the tolerance
value to allow for variations in the expected fragment/protein/
glycan size or peak time. This value is specified as a percent of
the expected size for that fragment/protein/glycan or time for a
peak. The default is 10% for DNA, protein, and CZE, and 2.5%
for Glycan.
Note: If there are multiple peaks in the tolerance range, the
largest peak is labeled as the expected peak, even if it is not the
exact size specified. To change the peak identified as an
expected peak, see “Forcing Expected Peaks ” on page 69.
7A default color is automatically assigned. To change the color,
click on the color block in the Color column and select the
desired color in the Color window.
8If desired, change the name shown in the Name column.
9If desired, change the Property Displayed in the Well Table.
This setting specifies the property that will be displayed in the
Expected Peak column for each expected peak listed in the Well
Table View.
V4.2LabChip GX User ManualPerkinElmer
10 To align the data to one or more specific peaks, select the For
Aligning check box. The data is realigned so the selected
peaks match their aligned size. Note: Incorrect alignment
settings can cause analysis errors.
11 To apply the expected peak only to specific wells, click in the
Apply to Wells column. The Select Wells window opens. Click
(or click and drag) to select the wells that you want to apply the
expected peaks to, and then click the OK button.
12 To apply the EFs, EPs, or EGs to the active plate, click the
Apply button.
13 To apply the EFs, EPs, or EGs to all plates in the workspace,
click the Apply Global button.
Exporting EFs, EPs, or EGs
1After the EFs, EPs, or EGs are entered in the Expected
Fragments/Proteins/Glycans Tab, click the Export button at the
bottom of the window. The Export Expected
Fragments/Proteins/Glycans Window opens.
Data Analysis68
2Navigate to the desired location for the file, type the desired
name for the file in the File Name text box, and then click the
Save button. A .GEP file is created to save the expected peak
settings.
Importing EFs, EPs, or EGs
After an Expected Peak file (*.GEP) has been exported, you can
import the settings into another plate or workspace.
1With the plate open in a workspace, in the Expected
Fragments/Proteins/Glycans Tab, click the Import button at the
bottom of the window. The Import Expected
Fragments/Proteins/Glycan Table window opens.
2Navigate to the location of the file, select the name of the file to
import, and then click the Open button. The expected peak
settings are imported into the tab.
3To apply the EFs, EPs, or EGs to the active plate, click the
Apply button.
4To apply the EFs, EPs, or EGs to all plates in the workspace,
click the Apply Global button.
V4.2LabChip GX User ManualPerkinElmer
Forcing Expected Peaks
If there are multiple peaks in the tolerance range, the largest peak
is labeled as the expected peak, even if it is not the exact size
specified. If a different nearby peak should have been selected as
the expected peak, you can specify which peak is labeled the
expected peak.
1In the Graph View, right-click on the peak that should be labeled
as the expected peak.
2On the shortcut menu, select Force Expected Fragment/Peak
and then select the desired fragment or peak from the menu.
To clear a forced peak and revert to the default expected peak,
right-click on the forced expected peak and select Clear Forced
EP.
Viewing the EFs/EPs/EGs in the Graph View
Data Analysis69
Expected Fragments, Expected Proteins, and Expected
Glycans are identified in the electropherogram by open triangles
over the peaks. The triangles are the same color as specified in the
Expected Fragments/Proteins/Glycans Tab.
To display the expected peak indicators in the Graph view:
1Click the Properties tab on the right side of the Graph view to
open the Graph View Properties.
2To view the size of all expected peaks, select Expected
Fragments, Expected Proteins, or Expected Glycan in one of
the Annotation list boxes.
Figure 25. Exp ected F ragments
V4.2LabChip GX User ManualPerkinElmer
Figure 26. Expected Peaks
Viewing the EFs/EPs/EGs in the Gel View
Expected Fragments/Proteins/Glycans are indicated in the Gel
View by colored lines. The color of the line matches the color
specified in the Expected Fragments/Proteins/Glycans Tab.
Data Analysis70
To display the expected peaks in the Gel View:
1Click the Properties tab on the right side of the Gel View to
open the Gel View Properties.
2Select the Show Expected Peaks check box.
To display the legend that identifies the colors and sizes of the
expected peaks:
1Click the Properties tab on the right side of the Gel View to
open the Gel View Properties.
2Select the Show Expected Peaks Legend check box.
Figure 27. Expe cted Peaks Legend
V4.2LabChip GX User ManualPerkinElmer
Viewing the EFs/EPs/EGs in the Well Table
A column is added to the Well Table for each Expected Peak
entered in the Expected Fragments/Proteins/Glycans Tab. The
column displays the property selected in the Property Displayed in
Well Table list in the Assay Analysis Window.
Viewing the EFs/EPs/EGs in the Peak Table
Expected Fragments, Expected Proteins, and Expected
Glycans are identified in the Peak Table with the peak name
displayed in the Type column.
Data Analysis71
Figure 28. DNA Assay Peak Table
V4.2LabChip GX User ManualPerkinElmer
Modifying Analysis Parameters
Some analysis parameters can be changed in the software to
modify the data evaluation for sample analysis. The following
procedures are included in this section:
•“Changing the Peak Find Parameters” on page 73
•“Adding a Peak” on page 74
•“Splitting a Peak” on page 75
•“Excluding a Peak” on page 74
•“Merging Two Peaks” on page 75
•“Adjusting the Peak Baseline” on page 76
•“Setting the Baseline for a Range of Peaks” on page 76
•“Selecting a Default Ladder” on page 77
•“Using the Default Ladder for Alignment” on page 79
•“Exporting the Default Ladder in an Assay” on page 79
•“Clearing the Default Ladder in a Plate” on page 79
•“Changing the Time Window for Analysis” on page 80
Data Analysis72
•Aligning or Unaligning the Marker Peaks
These settings can be changed after the run is complete or when
reanalyzing a previously saved data file.
V4.2LabChip GX User ManualPerkinElmer
Changing the Peak Find Parameters
After data filtering, the peak find algorithm locates the peaks and
calculates the local peak baselines. The algorithm begins by finding
all the peaks above the noise threshold to determine the baseline,
after which any peaks below the noise threshold are rejected. A
local baseline is calculated for each peak to allow for baseline drift.
The following Peak Find parameters can be changed:
•Min Peak Height
•Min Peak Width
•Slope Threshold
•Inflection Threshold
•Start Time
•End Time
•Filter Width
•Baseline Plateau
Data Analysis73
To change the Peak Find parameters for all wells:
1Select Analysis Analysis Settings to open the Assay
Analysis Window, and then click the Peak Find Tab.
2Change the parameters as necessary at the top of the window.
3Click the OK button to save the setting, reanalyze the data, and
close the window.
Click the Apply button to apply the changes and reanalyze the
data, but keep the Assay Analysis window open.
Click the Apply Global button the apply the settings to all plates
in the workspace and reanalyze the data, but keep the Assay
Analysis window open.
To change peak find settings for individual wells:
1Select Analysis Analysis Settings to open the Assay
Analysis Window, and then click the Peak Find Tab.
2In the Well drop-down list, select the well number that you want
to change the settings for.
3Change the settings at the bottom of the window under Well
Peak Find Settings to change the settings for the selected well.
4Click the OK button to save the setting, reanalyze the data, and
close the window.
Click the Apply button to apply the changes and reanalyze the
data, but keep the Assay Analysis window open.
V4.2LabChip GX User ManualPerkinElmer
Adding a Peak
You can manually add a peak in a region where a peak has not
been identified.
1In the Graph View, right-click at the top of the area where the
peak is to be added. The area must be outside any previously
identified peak and the cursor must be an up arrow.
2Select Add Manual Peak from the shortcut menu. A new peak
centered at the selected location is created.
3If necessary, adjust the Peak Baseline.
Excluding a Peak
You can exclude any peak or fragment from being used in the
analysis.
To exclude a peak:
Data Analysis74
1In the Peak Table View, right-click on the peak to be excluded.
2Select Exclude Peak from the shortcut menu. The Type for the
peak changes to X (excluded), and the value is not used in the
analysis.
3Right-click on an Excluded Peak in the peak table and select
Include Peak to include the peak in the data analysis.
OR
1In the Graph View, right-click near the top of the peak to be
excluded.
2Select Exclude Peak from the shortcut menu. The Type for the
peak changes to X (excluded), and the value is not used in the
analysis.
3Right-click on an Excluded Peak in the Graph view and select
Include Peak to include the peak in the data analysis.
V4.2LabChip GX User ManualPerkinElmer
Splitting a Peak
If two poorly resolved peaks have been identified as a single peak,
you can split the single peak into two separate peaks.
1In the Graph View, zoom in as necessary to view the single
2Position the cursor inside the peak at the horizontal location
Data Analysis75
peak.
where you want to split the peak. (The Split Peak option is not
available if the cursor is an up arrow. Move the cursor until the
peak is not selected.)
Figure 29. Splitting a Peak
3Right-click on the graph and select Split Pea k on the shortcut
menu. A new Manual Peak is created abutting the original peak.
Merging Two Peaks
If the analysis has defined two separate peaks, the two distinct
adjacent peaks can now be merged into one peak. This will include
the area of both peaks in the total concentration and %purity
calculations.
To merge two adjacent peaks:
1Exclude one of the peaks from the analysis, following the
procedure “Excluding a Peak” on page 74.
2Verify that Show Peak Baselines is selected in the Graph View
Properties.
3Click and drag the baseline of the remaining peak to include the
area under the excluded peak.
V4.2LabChip GX User ManualPerkinElmer
Adjusting the Peak Baseline
The Peak Baseline can be manually adjusted in the Graph View.
1In the Graph View, zoom in on the peak baseline. (If the peak
baselines are not displayed, select Show Peak Baselines in the
Graph View Properties to display the baseline for each peak.)
2Click the triangle at one end of the peak baseline and drag to
the desired location.
To reset the baseline to the original position, right-click near the
baseline end point and choose Reset to Defaults.
Setting the Baseline for a Range of Peaks
A straight line can be drawn under multiple peaks to create a single
baseline for all peaks above the straight line.
To draw a single baseline:
Data Analysis76
1In the Graph View, zoom in on the peaks. (If the peak baselines
are not displayed, select Show Peak Baselines in the Graph
View Properties to display the baseline for each peak.)
2Holding the SHIFT key, click at the point where the baseline will
begin and drag to the desired end point of the baseline.
3Release the mouse at the end of the baseline. The baselines for
peaks that are entirely contained within the baseline span are
adjusted to the new baseline.
To draw multiple baselines using Manual Baseline mode:
1In the Graph View, zoom in on the peaks. (If the peak baselines
are not displayed, select Show Peak Baselines in the Graph
View Properties to display the baseline for each peak.)
2Right-click on the graph, away from a peak, and select Manual
Baseline on the shortcut menu.
3Click and drag in the graph to draw manual baselines.
Note: Clicking and dragging in the graph does not zoom in while
Manual Baseline mode is on. Clicking and dragging in the Gel
View still zooms in when Manual Baseline mode is on.
4To turn off Manual Baseline mode, right-click on the graph, away
from a peak, and select Manual Baseline again.
V4.2LabChip GX User ManualPerkinElmer
Removing a Manual Baseline
To reset a manual baseline back to the original analyzed baseline:
1Right-click on one end of the manual baseline and select Reset
to Analyzed Base from the shortcut menu.
To reset all manual baselines:
1Right-click on one end of a manual baseline and select Reset
All Peak Bases.
2Click OK in the Reset All Peak Bases window.
Selecting a Default Ladder
For DNA, gDNA, and RNA assays, a ladder in an assay can be
defined as the default ladder. The default ladder can be used as the
ladder for wells on the plate, can be applied to all of the plates in
the open collection, or can be saved as the default ladder in a new
assay.
Data Analysis77
To set a specific ladder on a plate as the default ladder for the plate:
1With the plate data file open, select Analysis Analysis
Settings to open the Assay Analysis Window, and then click the
Peak Find Tab.
2Under Well Peak Find Settings, in the Well drop-down list,
select the ladder that will be defined as the default ladder.
3Click the Save as Default Ladder button. (The Save as Default
Ladder button only displays when a ladder is selected.)
4Click the Apply button to save the default ladder settings. The
Analysis Tab displays the ladder data in the Default Ladder table
and the Markers table (see Figure 30 on page 78).
V4.2LabChip GX User ManualPerkinElmer
Selecting a Default Ladder (Continued)
Figure 30. Default Ladder Settings
Data Analysis78
5The Ladder Peak Time and Area settings can be changed in the
Default Ladder table, but default ladder peaks cannot be added
or deleted and the ladder peak sizes cannot be changed.
6The Marker Time, Height and Area can be changed in the
Markers table.
7Click the Apply button at the bottom of the window to save the
changes to the active plate. Click the Apply Global button to
apply the default ladder settings to all open data files in the
collection. (This allows you to import the default ladder into a
plate that does not include any good ladders on the plate.)
8See “Using the Default Ladder for Alignment” on page 79 to
align wells in the plate using the default ladder.
V4.2LabChip GX User ManualPerkinElmer
Using the Default Ladder for Alignment
For DNA, gDNA, and RNA assays, after the default ladder is
selected for a data file, the default ladder must be assigned to the
desired wells in the plate.
1With the plate data file open, select Analysis Analysis
Settings to open the Assay Analysis Window, and then click the
Alignment Tab.
2Select the Align Well Groups to Specified Ladder option.
3For each group of wells under Align Well Group, select Default
Ladder in the To Ladder column.
4Click the Apply button to apply the changes to the active data
file or click the Apply Global button to apply the changes to all
data files in the current collection.
Exporting the Default Ladder in an Assay
Data Analysis79
The default ladder in a data file is included in the analysis settings
for a new assay when the settings are exported as an assay.
Exporting the assay enables you to use the default ladder in future
runs.
1After clicking the Apply or Apply Global button in the Assay
Analysis Window, the Export as Assay button displays.
2Click the Export as Assay button. The Export Assay Settings
from Plate to Assay File window opens.
3Select the desired location for the assay file, type the desired
name for the assay file, and click the Save button.
4Select the new assay in the Run Tab on the Start Run Window.
The new data file will include the default ladder.
Clearing the Default Ladder in a Plate
To delete the default ladder for a plate and return the alignment to
the ladder settings for the assay type:
1With the plate data file open, select Analysis Analysis
Settings to open the Assay Analysis Window, and then click the
Analysis Tab.
2Click the Clear Default Ladder button at the bottom of the
Default Ladder table.
V4.2LabChip GX User ManualPerkinElmer
Changing the Time Window for Analysis
The Start Ti me and End Time parameters in the Peak Find tab
define the time window within which peaks are found.
To change the Start Time and End Time parameters for all wells in
the open assay:
1Select Analysis Analysis Settings to open the Assay
Analysis Window, and then click the Peak Find Tab.
2Change the parameters as necessary.
3Click the OK button to save the setting, reanalyze the data, and
close the window.
Click the Apply button to apply the changes and reanalyze the
data, but keep the Assay Analysis window open.
Aligning or Unaligning the Marker Peaks
To perform data analysis for DNA, Protein, RNA, and Glycan
assays, the LabChip GX software aligns marker peaks included in
the sample wells with markers from the ladder.
•To view the unaligned data (spike rejected, filtered and baseline
corrected), choose Analysis Turn Off Analysis.
•To re-enable analysis, choose Analysis Turn On Analysis.
Data Analysis80
The default setting for this function is enabled. Turning off the
analysis displays the data without aligning the markers in the wells
and the ladders.
Figure 31. Data Before and After Alignment
V4.2LabChip GX User ManualPerkinElmer
Data Analysis81
Upper and Lower Marker Peaks for DNA Assays
For each DNA sample, the upper and lower marker peaks are
assigned first and then the data is aligned so that the well markers
match the ladder markers in time, allowing the size and
concentration of the sample peaks to be determined.
For DNA assays, the first peak is assigned to be the lower marker
and is then offset to match the lower marker in the ladder. The
upper marker is then assigned to the last peak in the sample well or
to the peak nearest the ladder’s upper marker. The Upper Marker
and Lower Marker are aligned to the ladder markers by resampling
the well data in a linear stretch or compression using a point-topoint fit.
If the sample marker peaks are either more than twice as far apart
or less than half as far apart as the ladder markers, they are
assumed to be the wrong peaks, and analysis of the well stops,
producing the error Marker peaks n ot detected.
In DNA assays, the height of marker peaks is assay dependent.
Ladder peaks are analyzed to calculate a marker peak threshold
that is used to locate the marker peaks in the sample wells. If the
marker peaks found using this calculated method fail to align with
those of a sample, the LabChip GX software will use the minimum
peak height threshold setting instead (if this value is lower than the
value for the marker peak). For example, the calculated threshold
might be too high to find the sample's markers if they happen to be
very small for some reason. Either no markers will be found or the
wrong peaks will be assumed to be markers and these may not
align with the ladder markers. Consequently, the software attempts
to use the minimum peak height threshold that, if it is set low
enough, will catch the real markers, allowing the sample to align.
If you get unexpected peaks in the ladder analysis or the markers
have been set incorrectly, you can manually exclude peaks or set a
peak to be used as a marker.
NOTES
•Excluding a peak or manually setting a peak to be an upper or
lower marker for a DNA assay can cause errors with analysis.
•You can move the boundary between the Peak Table and the
Graph view up or down to increase or decrease the size of the
Peak Table, making it possible to see all of the results at once.
Right-clicking in the Peak Table View of a well of a DNA assay
opens a shortcut menu with the following commands:
V4.2LabChip GX User ManualPerkinElmer
Data Analysis82
•Include Peak (only for peak type “?”)
•Exclude Peak
•Force Lower Marker
•Force Upper Marker
•Add Expected Peak
NOTE
You can also right-click on a peak in the Graph View to view the
same menu.
Lower Marker Peaks in Protein or RNA Assays
For each protein or RNA sample, the first peak in the sample is
designated as the Lower Marker. After all lower markers are
assigned, the data is aligned so that the well markers match the
ladder markers in time, allowing the size and concentration of the
sample peaks to be determined.
If there are unexpected peaks in the ladder analysis or if the marker
has been set incorrectly, you can manually exclude peaks or set a
peak to be used as a marker.
NOTES
•Excluding a peak or manually setting a peak to be the lower
marker can cause analysis errors.
•You can move the boundary between the Peak Table and the
Graph view up or down to increase or decrease the size of the
Peak Table, making it possible to see all of the results at once.
Right-clicking in the Peak Table View of a well of a Protein or RNA
assay opens a shortcut menu with the following commands:
•Include Peak (only for peak type “Xsys” or “?”)
•Exclude Peak
•Force Lower Marker
•Force Expected Peak
V4.2LabChip GX User ManualPerkinElmer
Data Analysis83
Lower Marker Peaks in Glycan Assays
For each Glycan sample, the lower marker is added to the samples
and ladders. The lower marker is identified because the
concentration is much higher than the sample or ladder. After all
lower markers are assigned, the data is aligned so that the well
markers match the ladder markers in time, allowing the size and
corrected area of the sample peaks to be determined.
If there are unexpected peaks in the ladder analysis or if the marker
has been set incorrectly, you can manually exclude peaks or set a
peak to be used as a marker.
NOTES
•Excluding a peak or manually setting a peak to be the lower
marker can cause analysis errors.
•You can move the boundary between the Peak Table and the
Graph view up or down to increase or decrease the size of the
Peak Table, making it possible to see all of the results at once.
Right-clicking in the Peak Table View of a well of a Glycan assay
opens a shortcut menu with the following commands:
•Include Peak (only for peak type “Xsys” or “?”)
•Exclude Peak
•Force Lower Marker
•Force Expected Peak
V4.2LabChip GX User ManualPerkinElmer
Saving and Exporting Assays
Assays are created by PerkinElmer and are included with the
LabChip GX software. The instrument and software settings used to
run each plate and to analyze the plate are contained in an assay
(.asy or .asyx) file, which is selected at the start of the run in the
Start Run Window. After each well is complete, the data is analyzed
using the analysis settings in the assay file. If the default analysis
settings often need to be modified for more optimal analysis, a new
assay file can be created with the modified settings to be used on
subsequent runs.
To save a modified assay:
1Run a plate with the original assay file.
2Modify the analysis settings for the plate as needed.
3Right-click on the plate title in the Plate View or Plate List, select
Export Assay, specify a name and location for the assay file,
and click the Save button.
-- OR -On the Assay Analysis Window, click the Export as Assay
button, specify a name and location for the assay file, and click
the Save button.
Data Analysis84
4When running the next plate, select the new assay file in the
Start Run window.
The procedure above can also be used to automatically include a
set of expected peaks in the assay.
V4.2LabChip GX User ManualPerkinElmer
Changing the View of the Data
The views in the LabChip GX Main Window can be customized to
display data according to the preferences of the user. These options
do not change the raw data but provide different means of
displaying the data.
To change the view in the main window, see:
•“Viewing Gels” on page 85
•“Viewing Zero Baselines ” on page 86
•“Adjust Pane Widths” on page 87
•“Show or Hide Views” on page 87
•“Zoom In and Zoom Out” on page 88
•“Viewing Graphs in the Overlay Electropherograms Tab” on
page 89
•“Viewing Graphs in the Electropherograms Tab” on page 90
•“Viewing Analysis Errors and Warnings” on page 92
Data Analysis85
•“Viewing Multiple Properties in the Well Table View” on page 93
•“Exporting Data Manually” on page 105
Viewing Gels
To compare the gels generated by the instrument, view the gels in
the Gel View.
Figure 32. Gel View
The color, width, and contrast of the gels can be changed using the
Gel View Properties.
To rearrange gels, click on the column header (well name) and drag
the well to the desired location. To hide a well, select the well and
then click the (X) button on the column header, or right-click on the
well in the Plate View or Plate List and select Remove Well.
V4.2LabChip GX User ManualPerkinElmer
Viewing Zero Baselines
All electropherograms produced with the instrument show some
amount of background fluorescence. By default, the LabChip GX
software enables the zero baseline function. To remove the zeroing,
select Analysis Analysis Settings to open the Assay Analysis
Window, click the Peak Find Tab, and select the None check box
under Baseline Algorithm.
Data Analysis86
Figure 33. Zero Bas eline On
Figure 34. Zero Basel ine Off
V4.2LabChip GX User ManualPerkinElmer
Adjust Pane Widths
The LabChip GX Main Window displays several different views of
the data files open in the workspace. You can change the height
and width of the views to make the views smaller or larger.
To adjust panes:
1Place the cursor over the edge of the pane that you want to
adjust. The cursor changes to a line with arrows on each end.
2Click and drag up, down, left, or right. The pane is resized after
you release the mouse button. The layout setting is saved as
part of the collection.
3To save the setting, save the workspace. To create new
collections with the same settings, the collection can be saved
as a collection template (see “Collection Menu” on page 146).
Show or Hide Views
Data Analysis87
The views displayed in the LabChip GX Main Window can be
hidden to maximize other views in the main window.
To hide a view:
1Select CollectionLayout on the LabChip GX Main window.
The Layout Options Window opens.
2Click on the location that is selected for the view to clear the
selection. The view is hidden.
3If a location (Left, Right, or Bottom) does not contain any views,
the pane closes and the remaining panes enlarge to fill the
space.
Note: The Gel View is always displayed and cannot be hidden.
To display a hidden view:
1Select CollectionLayout on the LabChip GX Main window.
The Layout Options Window opens.
2Click on the desired location for the hidden view (Left, Right, or
Bottom). If multiple views are displayed in the same location,
use the tabs at the top of the location to switch between views.
V4.2LabChip GX User ManualPerkinElmer
Zoom In and Zoom Out
You can zoom in and out on data displayed in the Gel View and the
Graph View. The Graph View and the Gel view both zoom to the
same levels when either view is zoomed in.
To z o om i n:
•Click and drag to enclose the region of interest. When you
release the mouse button, the selected area enlarges to fill the
view. In the Gel view, all lanes in the collection zoom to the
same level.
•You can continue zooming in until you reach the maximum
magnification (the graph will not zoom in any closer).
•Use the horizontal scroll bar under the graph to pan from left to
right.
To z o om o ut :
•Right-click in the Graph view or Gel view and select Unzoom to
go to the previous zoom setting or select Unzoom All to zoom
out to the default view.
Data Analysis88
For more information, see Viewing Graphs in the Overlay
Electropherograms Tab and Viewing Graphs in the
Electropherograms Tab.
V4.2LabChip GX User ManualPerkinElmer
Data Analysis89
Viewing Graphs in the Overlay Electropherograms Tab
If the Overlay Electropherograms Tab is not open, select Collection
Layout on the LabChip GX Main Window, select the location
where you want to display the Overlay Electropherograms tab, and
click the Apply button. The Overlay Electropherograms tab opens.
Figure 35. Overlay Electro pherograms Tab with Multiple
Graphs
Data from multiple wells can be overlaid in the same graph for
visual comparison. Click on one well, then hold down the CTRL key
and click on the additional wells to view in the graph. Each peak
graph is shown in a different color and line style with a legend at the
top of the window. You can remove wells from the overlay by CTRL
+ clicking the corresponding wells (the bounding box disappears).
Use the Overlay Offset text box on the Graph View Properties to
offset each of the graphs by the RFU value specified.
Figure 36. Ove rlay Offset
V4.2LabChip GX User ManualPerkinElmer
Data Analysis90
To add samples to the Overlay Electropherograms Tab, Ctrl +
click on the sample that you want to add in the Plate View or Plate
List, Gel View, Well Table View, or Peak Table View. Selected wells
are identified by: dashed outlines around the selected gel lanes in
the Gel view, light blue wells in the Plate view, and dark gray rows in
the Well Table view and Peak Table view.
To remove a specific sample from the graph, Ctrl + click on the
sample that you want to remove in the Plate View or Plate List, Gel
View, Well Table View, or Peak Table View.
T o display onl y one sample in the Overlay Electropherograms tab,
click on one sample in the Plate View or Plate List, Gel View, Well
Table View, or Peak Table View.
To view the point coordinates and size of the position of the
cursor, hold down the CTRL or Shift key and move the cursor over
the graph.
To view the point coordinates and slope of a point on the trace,
hold down the ALT key and the CTRL key and move the cursor over
the trace.
Viewing Graphs in the Electropherograms Tab
If the Electropherograms Tab is not open, select Collection
Layout on the LabChip GX Main Window, select the location where
you want to display the Electropherograms tab, and click the Apply
button. The Electropherograms tab opens.
Figure 37. Electro pherograms Tab with Multiple Graphs
V4.2LabChip GX User ManualPerkinElmer
Data Analysis91
Data from multiple wells can be displayed in the same tab for visual
comparison. A separate graph for each well in the collection is
displayed in the Electropherograms tab. The data file name and
well name display at the top of each graph. If all of the wells are not
visible, use the scroll bar on the tab to scroll through the wells in the
collection.
Zooming in or out on one graph zooms all of the graphs in the tab
to the same zoom level.
To only display the wells selected in the Gel View or Well Table
View, select the Graph Selected Gels Only check box in the Graph
View Properties.
To move the graphs in the tab, in the Gel View click and drag the
well to the desired position.
To change the number of graphs displayed in the tab, select the
desired number of Rows and Columns of graphs to display in the
Graph View Properties. Selecting 8 Rows by 12 Columns displays
an entire 96-well plate.
To view the point coordinates and size of the position of the
cursor, hold down the CTRL or Shift key and move the cursor over
the graph.
To view the point coordinates and slope of a point on the trace,
hold down the ALT key and the CTRL key and move the cursor over
the trace.
V4.2LabChip GX User ManualPerkinElmer
Viewing Analysis Errors and Warnings
Analysis errors and warnings are indicated by red or yellow
exclamation points in the wells in the Plate View or below the
headers in the Gel View. The details of the error or warning can be
displayed in the Well Table View. After the Analysis Error column is
added to the Well Table, the analysis error text can be displayed in
the Electropherograms Tab or the Overlay Electropherograms Tab.
Viewing Analysis Errors and Warnings in the Well Table
To view the details of an analysis error or warning, display the
Analysis Error in the Well Table View.
1Click the Well Table tab.
2Right-click in the table header. The Select Well Table Columns
Window opens.
3In the Available Columns list, select Analysis Error.
Data Analysis92
4Click the Right Arrow button to move Analysis Error to the
Selected Columns list.
5If desired, click and drag Analysis Error to the desired column
location. (The top column in the list is the first column in the
table.)
6Click the Save button.
Viewing Analysis Errors and Warnings on the Graphs
The text of analysis errors and warnings can be displayed on the
graphs in the Electropherograms Tab or the Overlay
Electropherograms Tab. The Analysis Error columns must be
displayed in the Well Table to be available for selection in the Graph
View Properties.
1Click the Electropherograms Tab or the Overlay
Electropherograms Tab.
2Click the Properties tab on the right side of the graph. The
Graph View Properties tab opens.
3Select Analysis Error in the Well Annotation drop-down list.
4If desired, click the Well Annotation Location button to change
the location.
5Click off of the Graph View Properties tab to close the tab.
V4.2LabChip GX User ManualPerkinElmer
Viewing Multiple Properties in the Well Table View
Properties for Expected Peaks and Smears display in the Well
Table View. When entering the settings in the Expected
Fragments/Proteins/Glycans Tab or the Smear Analysis Tab on the
Assay Analysis Window, the property to display in the well table is
specified in the Property Displayed in Well Table column.
To display multiple properties for the same Expected Peak or
Smear in the Well Table:
1In the Expected Fragments/Proteins/Glycans tab, or the Smear
Analysis tab, enter the properties for the expected peak or
smear, selecting one of the desired properties in the Property
Displayed in Well Table column.
2Click in the next row and type the same name as an existing EP
or smear in the Name field of the table. The row will
automatically update with the same entries as the original row.
The only field that can be changed is the Property Displayed in Well Tabl e column.
Data Analysis93
3Change the Property to the desired selection.
4Repeat until all desired properties have been added to the table.
5Click the Apply button to display the new columns in the Well
Table. Duplicate properties in the table will be removed when
the settings are applied to the plate.
V4.2LabChip GX User ManualPerkinElmer
Copying Information
The Edit menu and some of the right-click pop-up menus offer the
following choices for copying information from the LabChip GX
software for use with other applications, depending on the
selection:
•Copy Gel
•Copy Lane
•Copy Graph
•Copy Rows to Clipboard
•Copy Column to Clipboard
Choosing any of these commands places a copy of the selected
item on the computer's clipboard. You can then paste the item into a
word processing, graphics, or other program.
Choosing Copy Gel copies all of the wells displayed in the Gel View
with the labels as part of the graphic. To copy a gel, right-click in the
Gel View and choose Copy Gel.
Data Analysis94
Choosing Copy Lane copies the selected well in the Gel View with
the labels as part of the graphic. To copy a gel, right-click in the Gel
View and choose Copy Lane.
Choosing Copy Graph copies the graph displayed in the Graph
View. The size of the image that is placed on the clipboard when
copying a graph is the same size as the graph displayed in the
software.
Copying rows from the Peak Table View or the Well Table View
places ASCII information from the selected row or rows on the
clipboard. To copy a row or multiple rows, select the desired rows in
the table, right-click on one of the selected rows, and select Copy Rows to Clipboard.
Copying columns from the Well Table View is only available for
certain columns in the table. To copy a column, select a single cell
in the column, right-click on the cell, and select Copy Column to Clipboard.
V4.2LabChip GX User ManualPerkinElmer
Reanalyzing a Data File
Occasionally you may need to open and view or reanalyze a data
file that was run and saved previously. The raw data values are
saved in the plate data file, along with the original analysis settings
that were chosen for the run and any changed analysis settings.
This enables you to reanalyze the data with new settings or to view
previously saved settings.
The following analysis parameters can be changed:
•Plate peak find settings and Well peak find settings (see
“Changing the Peak Find Parameters” on page 73)
•Sample names and comments (see “Using Sample Names
Files” on page 63)
•Add a Peak (see “Adding a Peak” on page 74)
•Exclude peaks from analysis (see “Excluding a Peak” on
page 74)
•Reassign upper/lower markers (see “Upper and Lower Marker
Peaks for DNA Assays” on page 81 or “Lower Marker Peaks in
Protein or RNA Assays” on page 82)
Data Analysis95
•Alignment or no alignment with ladder peaks (see “Aligning or
Unaligning the Marker Peaks” on page 80)
•Assay - you can save the changed settings under a new assay
name, if desired (see “Saving and Exporting Assays” on
page 84)
To reanalyze a data file:
1Open the workspace that contains the plate data (see page 59).
2Change the analysis parameters (see list above) as needed.
3When you click the Apply, Apply Global, or OK buttons in the
Assay Analysis Window, the data is automatically re-analyzed
and the updated information is displayed.
4To view previous analysis settings, click the Restore Plate
button at the bottom of the Assay Analysis window and select
the version to view.
V4.2LabChip GX User ManualPerkinElmer
Printing Workspace Information
Choosing File Print opens the Print Windowto select the
information to print from the open workspace. The selected
information can be printed to a printer or to a PDF file. If the
workspace contains multiple collections, information from the active
(selected) collection is printed.
The header of each printout contains the workspace name, the
collection name, and the page number. The footer of each printout
contains information about the LabChip GX software, data file,
assay, and instrument. If the 21 CFR Part 11 option is installed, the
footer also includes the data file version, signed state, and lock
state.
The following information can be printed for the open workspace:
•Print All
•“Gel” on page 97
•“Electropherogram” on page 97
•“Overlay Electropherogram” on page 98
Data Analysis96
•“Well Table” on page 99
•“Peak Table” on page 100
•“Multiple Electropherogram” on page 101
See below for descriptions of each option.
Print All
This option prints the results of the assay in all of the available
formats. The page layout depends on the options selected in Print
Settings.
V4.2LabChip GX User ManualPerkinElmer
Data Analysis97
Gel
This option prints a gel image with the lanes marked by the sample
name.
Figure 38. Printed Gel
Electropherogram
Choosing this option prints an individual electropherogram of each
of the selected wells.
Figure 39. Pri nted Electrophe rogram
V4.2LabChip GX User ManualPerkinElmer
Data Analysis98
Overlay Electropherogram
Choosing this option prints one graph with the electropherograms
for all of the selected wells overlaid onto a single graph.
Figure 40. Pri nted Overlay Electrop herogram
V4.2LabChip GX User ManualPerkinElmer
Data Analysis99
Well Table
Choosing this option prints the columns displayed in the Well Table
View. To change the columns that are printed or the order of the
columns, change the columns in the Well Table View.
•Selecting the Add Border check box prints the table with a
border around each cell in the table.
•Selecting the Add Well Name as Header check box prints the
well name, well label, and sample name at the top of the page.
•Selecting both the Well Table and the Peak Table prints each
well on a separate page with the Well Table row at the top of the
page and the Peak Table for the well below it.
V4.2LabChip GX User ManualPerkinElmer
Figure 41. Prin ted Well Table
Data Analysis100
Peak Table
Choosing this option prints the columns displayed in the Peak Table
View. To change the columns that are printed or the order of the
columns, change the columns in the Peak Table View.
•Selecting the Exclude Marker check box removes all markers
from the printout.
•Selecting the Add Border check box prints the table with a
border around each cell in the table.
•Selecting the Add Well Name as Header check box prints a
well header (plate name, well name, and sample name) before
the peaks in each well.
•Selecting the Per Page or Per Well option for the Column
Header specifies whether the Peak Table column headers are
printed only at the top of each page (default), or if the headers
are also printed at the beginning of each new well table.
•Selecting both the Well Table and the Peak Table prints each
well on a separate page with the Well Table row at the top of the
page and the Peak Table for the well below it.
Figure 42. Pri nted Peak Table
V4.2LabChip GX User ManualPerkinElmer
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.