Installation Qualification (IQ) and Operational Qualification (OQ)
Company Name:
Address:
Location, Room:
Serial Number or System Name:
Protocol Number:
Issue Date:
Date Tested:
Valid if tested within 1 year of Issue Date
Recertification Period
Template-Company
Template-Address
Template-SN
Template-Protocol
27 May 2009
Recommended at12 Months
Recertification Due Date:
Release History
Part Number Release Publication Date
09234069 D June 2009
Any comments about the documentation for this product should be addressed to:
User Assistance PerkinElmer Global Technical Support
PerkinElmer LAS (UK) Ltd M/S 215
Chalfont Road 710 Bridgeport Avenue
Seer Green Shelton
Beaconsfield Connecticut 06484-4794
Bucks HP9 2FX U.S.A.
United Kingdom
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Appendix C– Attachments List .................................................................................................. 28
Appendix D- Document History ................................................................................................. 29
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Template-Company IQ / OQ
Template-SN Page 3 of 29
Introduction
Objective
The objective of this document is to detail the environmental and facilities requirements necessary for proper
installation and operation of the Flexar Analytical / / FX-10 UHPLC Pump. The completed process demonstrates
that the equipment meets the vendor-developed standards of operation and safety, and performs the functions
specified by the manufacturer.
Scope and Responsibility
PerkinElmer is responsible for providing trained personnel, the IQ/OQ elements outlined in this plan and verifying
that these elements are fully executed and documented.
Template-Company is responsible for accepting the terms of this plan and providing the facility with all
environmental requirements for the proper installation of the Flexar Analytical / / FX-10 UHPLC Pump.
Template-Company is responsible for providing personnel and assistance to PerkinElmer for implementing the
IQ/OQ outlined in this document.
Warranty Period and Service
Full details of PerkinElmer’s instrument warranty have previously been provided with quotations, order
acknowledgements and invoices. PerkinElmer’s instrument warranty covers all parts and labor, but excludes
consumable materials. Exceptions may apply to instruments purchased used.
Contact your PerkinElmer service provider for a service plan which may be purchased to extend coverage
beyond warranty. PerkinElmer recommends contracting regular preventive maintenance. Over time, failure to
perform the recommended preventive maintenance may reduce the reliability of some systems.
Need to Re-Qualify
The instrument may need to qualified again following modifications made to the original configuration or if the
instrument is serviced or moved.
Notices
Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no warranty of any
kind with regard to this document, including, but not limited to, the implied warranties of merchantability
and fitness for a particular purpose.
PerkinElmer shall not be liable for incidental or consequential damages in connection with the furnishing or use
of this document.
Copyright Information
This document contains proprietary information that is protected by copyright.
All rights are reserved. No part of this publication may be reproduced in any form whatsoever or translated into
any language without the prior, written permission of PerkinElmer, Inc.
Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are
protected by law. PerkinElmer is a registered trademark of PerkinElmer, Inc.
All other trademarks and registered trademarks not owned by PerkinElmer, Inc. or its subsidiaries that are
depicted herein are the property of their respective owners.
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Template-Company IQ / OQ
Template-SN Page 4 of 29
Protocol Documentation
Documentation protocol may be superseded by customer SOP
Writing Instruments
All writing will be done with a black or blue ink pen.
Signatures
A signature of each party involved with the ownership, management, testing, and quality assurance of each
instrument will be required before testing will proceed and for final approval at completion.
On any page where a written entry is made, or data is gathered by a PerkinElmer representative, that page must
be signed and dated by the PerkinElmer representative.
Up to three signature entries are provided each for Customer and PerkinElmer on the Preliminary Report
Approval Page and Final Approval Page. If any signature lines are unused, draw a single diagonal line filling the
Signature area and then write initials and date next to the line.
All signatures will be the signer’s full legal name as legibly as can be written and dated.
Initials
At any point when initials are to be used, they will contain at least two of the initials and include a date.
Date
Dates will be given in the format of DD-MMM-YYYY. If a month or date is only one digit, a zero will be employed
to maintain the two-digit format for each entry.
Error Codes
When an error is corrected, a single horizontal line through the error will be used. The line will then be initialed
and dated along with an error code from the list found in the protocol.
Spaces
For columns where the protocol is asking whether something has been verified, a “Yes” will be written. If not, a
“No” will be written and an explanation will be stated in the Comments section. A check mark or “N” is
unacceptable. “NA” may be used where the verification was not applicable.
Pass/Fail
For areas where the protocol requires a “Pass/Fail”, the words “Pass” or “Fail” will be written in. If the instrument
passes, Pass will be entered. If the instrument fails, Fail will be entered.
Visual Inspections
All visual inspections will be filled out with a word that indicates the physical condition being observed. If no
damage is observed, use the word “None” if no damage is present. If damage is observed, write a description of
the damage observed. This will apply to any visual inspections.
Calibration/Expiration dates
MM/YY format will be used to denote calibration and reagent expiration dates. The instrument or reagent is valid
until the last day of the month noted.
Temperature
All temperatures will be given in the unit of Celsius (°C)
Blank Areas
If there are no comments for a “Comments” section, “None” is written and the rest of the lines are left blank.
If a section is not applicable, such as an alarm function, “NA” is sufficient.
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Template-Company IQ / OQ
Template-SN Page 5 of 29
Once an entry has been made (other than “None”) any remaining space in the Comments (either single or
multiple blank lines) requires a single diagonal line filling the area, along with initials and date next to the line.
EXCEPTION: DO NOT cross out blank lines on the “Error Code Abbreviation, Definition and Making Corrections”
page. These pages or sections are either copied for future documentation needs or provide space to maintain
records into the future.
Acceptance Criteria
All tests must include values for acceptance criteria and the results of that test must be included.
Precision and Accuracy Measurements
The following areas of precision and accuracy will be noted on a per-test basis.
All instrument precision values (when available) will be noted within the OQ document. All data entered into the
OQ document will reflect these values.
For example: If a precision value of four significant digits is supplied for a test by the manufacturer, then all four
digits will be entered with the last digit being rounded down if the fifth digit is below 5, rounded up if above 5, and
rounded up if fifth digit is 5.
Failure Mode Responses
If a test result is determined to be outside of the precision parameters set forth by manufacturer of instrument, it
will be discussed with the customer first and if agreed both by customer and PerkinElmer then it will be repeated
a second time to confirm the result. If the second result fails, the instrument owner will be contacted to discuss a
reasonable course of action. If however, the second test passes, a note will be made as to the reason for the
primary failure and the second test data will be used.
Test -> Fail -> Repeat -> PASS -> Report 2nd result
Deviation
Deviations are events that are a departure from the specific protocol procedures as they were approved.
Deviations to the plan are permitted if those deviations are clearly detailed, approved by all parties and entered in
Appendix - Deviations
Change Control
Change control procedures are in place to maintain the validation process after modifications are made to the
original configuration. Any changes to the hardware, firmware or software of a validated system must be clearly
specified and the validation confirmed and recorded.
1. A blank Change Control form is located in Appendix C. Consider making photocopies when blank.
2. If used, insert a completed Change Control Form and its associated data as an Attachment.
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Template-Company IQ / OQ
Template-SN Page 6 of 29
Error Code Abbreviation, Definition, and Making
Correction
The error code abbreviation and definition list below is provided for clients and reviewers to make necessary
entries and corrections after the document has been approved.
PerkinElmer technical staff and reviewers make all necessary efforts to ensure that this document and related
attachments are free of errors and mistakes. However, some typographical, formatting, and inadvertently
erroneous entries may exist within the document that can be corrected by client(s). Once an error is found, (1)
simply strike the erroneous word or entry with a black pen, (2) enter the correct word or entry above the
erroneous word or entry, (3) enter the code (for example; IE for incorrect entry), and (4) then initial and date.
Space is also provided to create new codes or to enter the codes used and recommended by your compliance
department.
CE Calculation Error NA Not Applicable
IE Incorrect Entry SE Spelling Error
IM Instrument Malfunction TE Transcription Error
IO Inadvertently Omitted TYP Typographical Error
IR Instrument Recalibrated WD Wrong Date Entered
LE Late Entry (of initials or date) WO Writeover
Additional Codes
Code Definition Initials Date
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Template-Company IQ / OQ
Template-SN Page 7 of 29
Approval Page
Preliminary Protocol Approval
This is to certify that the Installation Qualification and Operational Qualifications procedures for the
Flexar Analytical / / FX-10 UHPLC Pump Serial number Template-SN have been reviewed and both
PerkinElmer LAS and Template-Company have agreed to proceed with the procedures described in this
document.
Template-Company
Signature Name Title Date
PerkinElmer LAS
Signature Name Title Date
Final Approval
This is to certify that the Installation Qualification and Operational Qualifications procedures for the
Flexar Analytical / / FX-10 UHPLC Pump Serial number Template-SN have been performed and the
configuration installed meets [ ] does not meet [ ] the procedures and specifications described in
this document.
Template-Company
Signature Name Title Date
PerkinElmer LAS
Signature Name Title Date
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Template-Company IQ / OQ
Template-SN Page 8 of 29
This page intentionally blank
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Template-Company IQ / OQ
Template-SN Page 9 of 29
Part I – Installation Qualification
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
1.1 System Summary
1.1.1 Instrument Identification
Manufacturer
Model
Equipment Serial #
Firmware Version
Instrument Control Software Version
Model options, circle as applicable
PerkinElmer
Flexar Analytical / FX-10 UHPLC Pump
Template-SN
Isocratic / Binary / Quaternary / FX-10
1.1.2 Specifications
Refer to respective documents for specifications like Voltage requirements, Line Frequency, Power
Consumption, Temperature, Relative Humidity, etc.
Refer to respective documents for minimum PC requirements.
1.1.3 Location Requirements
Refer to respective documents for specific requirements for the location of the instrument. Requirements are
typically, space (actual size, bench space and weight requirements especially if heavy), environment, power
supply, gas supply, etc., and any other needs for the safe and consistent operation of the instrument.
1.1.4 Maintenance and Troubleshooting
Routine maintenance can be performed by a trained analyst using the instructions found in the respective
documents for maintenance and troubleshooting. PerkinElmer has trained service representatives, who perform
other planned maintenance service annually or as required. Departmental Standard Operating Procedures
(SOPs) shall also be consulted to implement and document the necessary repairs.
1.1.5 Hazards and Safety Precautions
Refer to respective documents for the safety summary outlines and explanation.
Comments:
PerkinElmer LAS Representative:
Reviewed By: Date:
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Date:
1.2 Documentation
1.2.1 Equipment Manual
Part No. Title
Flexar Analytical / FX-10 UHPLC Pump User’s Guide
1.2.2 Purchase Order Certification
The Flexar Analytical / FX-10 UHPLC Pump Template-SNis in compliance with the specifications of
the purchase order.
Purchase Order Number
Template-PO
1.2.3 PerkinElmer Service Engineer Training
Persons authorized by PerkinElmer LAS to perform validations have been properly trained. Training is
documented, and a certificate is issued by PerkinElmer LAS to that effect.
See Attachment #1 for PerkinElmer LAS Training Certificate.
1.3 Installation Qualification
Verify the following list and mark as appropriate:
Description Verified
1.3.1
1.3.2
The working environment is suitable with reference to the appropriate section in the
user’s manual
The correct basic equipment has been provided in accordance with the purchase
order
1.3.3
1.3.4
1.3.5
1.3.6
1.3.7
1.3.8
The packaging is undamaged
The instrument, PC and other supplied equipment are undamaged
Users Guide is identified
Manufacturer’s specifications included in manual
The power requirement settings of the instrument, PC etc., are compatible with those
of the working environment. Note: It is the responsibility of Template-Company to
ensure proper voltage / current / frequency of the electrical line supplied for
operating the Flexar Analytical / FX-10 UHPLC Pump.
Adequate room for ventilation and servicing the equipment
Comments:
PerkinElmer LAS Representative:
Reviewed By: Date:
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Date:
1.3.9
1.3.10
1.3.11
1.3.12
1.3.13
1.3.14
1.3.15
1.3.16
1.3.17
1.3.18
1.3.19
1.3.20
1.3.21
1.3.22
PC’s requirements are met with reference to the appropriate section in the user’s
guide.
Instrument is identified, connected and set up in accordance with reference to the
appropriate instructions in the user’s guide.
Adequate electricity supply available and unit is grounded
Software installed on the PC with reference to the appropriate instructions in the
user’s guide.
Necessary components/accessories installed with reference to the appropriate
instructions in the user’s guide.
Firmware verification
Software installed and appropriate panels displayed:
Instrument control software installed if applicable
Data Acquisition Software installed
Operating System (name and version)
Other Software installed if applicable (name and version)
Start-up diagnostics run to completion (instrument initialized)
Warranty and available services explained
System Familiarization and Orientation Completed
1.3.23
List Attendees:
1.4 Installation Qualification Verification
The Flexar Analytical / FX-10 UHPLC Pump, Serial Number Template-SN, has been
properly set up and installed, and all appropriate forms and documents supporting the
system have been filled out and are available.
Comments:
PerkinElmer LAS Representative:
Reviewed By: Date:
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
Date:
Verified:
Part II – Operational Qualification
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
2.1 Operating Procedure References
2.1.1 Standard Operating Procedures pertaining to the Flexar Analytical / FX-10
UHPLC Pump
It is responsibility of the customer to follow a SOP for use of the equipment to ensure consistent operations.
The customer SOP may be written after the Qualification and can be added as an attachment after
PerkinElmer completes this IQ/OQ.
2.2 Operational Qualification Instruction
The following instructions apply to the Operational Qualification Test Description
2.2.1
2.2.2
2.2.3
2.2.4
2.2.5
The following instructions apply to the Parameter Testing
2.2.6
2.2.7
2.2.8
2.2.9
2.2.10
2.2.11
Record Model (or Part Number) and Serial Number of any accessories or components not covered in
the OQ
Record the Serial Number and Calibration Date for each item of Test Equipment used.
Record the Batch/Lot Number and Expiration Date of each Standard, Reagent and Solvent used,
including those supplied by the customer.
Label all solutions prepared at customer site in accordance with the customer requirements (date,
content, expires by, etc)
Provide the Standard certificates as Attachments.
Conduct Parameter Testing and record test data, Pass/Fail and Comments.
Verify the completed OQ in Section 2.5
Sign, date and number attachments of all recordings or printouts made during the test procedure (ex.
settings, spectrograms).
Fill in All Appendices, as required.
Complete and sign off the Final Approval
Complete and affix an IQ/OQ Certification Sticker to the instrument.
Comments:
PerkinElmer LAS Representative:
Reviewed By: Date:
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
Date:
2.3 Operational Qualification Test Description
Configurations Covered
Accessories/Components not
covered
Estimated Certificate Testing Time
Pre-test Stabilization N/A
Testing Time 2.0 Hours
Materials Required
Documentation Part Number
Pump Users Manual (Not needed for
testing. Reference only)
Pump Service Manual (Not needed
for testing. Reference only
IQ/OQ Certification Sticker 09934512
Test Equipment Manufacturer Model
Restrictor Coil PerkinElmer 02540843 N/A N/A
Flexar Analytical/ FX-10 UHPLC Pump
Model Serial Number
Serial
Number
Calibration
Due Date
1000 PSI Back Pressure Regulator
Calibrated Scale or Balance
Calibrated Timing Device
Customer Supplied ManufacturerPart Number Batch/Lot Expiration
H2O
Acetone
All solutions prepared at customer site must be labeled in accordance with the customer requirements
Tests
Name of Test Description
PerkinElmer N2910381 N/A N/A
N/AN/A
Comments:
PerkinElmer LAS Representative:
Date:
Reviewed By: Date:
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
Pressure Pulsations
Pulsations (PSI) at 1 ml/min solvent flow and 1000 PSI backpressure
(approx)
Flow Accuracy
For Analytical pumps, solvent flow as measured at 1ml and 3ml per
minute.
For the FX-10 UHPLC pumps, solvent flow as measured at 0.3ml and
0.9ml per minute.
Compositional Accuracy
For Binary / Quaternary Analytical pumps, accuracy as measured by
relative UV absorbance (265 nm) at programmed proportions of 0.2%
Acetone/Water and Water alone.
PerkinElmer assumes no responsibility for test results.
2.4 Parameter Testing:
2.4.1 Pressure Pulsations Test
Test Conditions
Comments:
PerkinElmer LAS Representative:
Reviewed By: Date:
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
Date:
Delta
•The Pressure Pulsations Test is to verify that the output flow fluctuations are within
specification. The specification for this test to pass is 2 % peak-to-peak of the running
pressure at which the test is run.
Test Prerequisites
•Start the pump and let the pressure stabilize
Test Steps
•Connect the pump output to a restrictor coil or 1000 psi backpressure device. Set the
pump flow rate to 1 mL/min. Running pressure for this test must be more than 500 psi.
Adjust tubing length as required to obtain the required back pressure.
•Monitor the displayed pressure readout for 1 minute. Note the observed maximum and
minimum pressures.
•Using the numbers obtained determine:
1. The difference between maximum and minimum (Delta).
2. The value of the mid-point pressure (MP).
3. The 2% value of the midpoint pressure. This mid-point value is the pressure used
to compute what the 2 % peak-to-peak error band should be.
There is a minimum pressure requirement of 350 psi for the check valves to operate properly.
To ensure the valves are operating and seating during the test PerkinElmer runs the test with
the pressure approximately >500 psi. This is accomplished by using a length of 0.007” ID
tubing or a backpressure device that will create the necessary back pressure on the check
valves.
Test Conditions
The Composition Accuracy Test proves that the proportioning valves are operating correctly.
This is accomplished by varying the amounts of solvent entering the pump through the
proportioning valves. Using an acetone solution, which has an absorbance at 265 nm as one
solvent, and water, which does not absorb at 265 nm we look at the absorbance of the
resulting composition. By controlling or proportioning the amount of acetone in the
water/acetone composition we can change the absorbance of the mobile phase composition.
We do this by setting up a pump program to run through several different positions of the
proportioning valves and measure the absorbance of the resultant mobile phase.
Test Steps
•Connect pump output directly to the backpressure regulator or other device (length of
tubing for example) to create the necessary back pressure.
• Fill reservoirs A (and C if testing a quaternary model) with HPLC grade water.
• Pipette 1.0 mL of acetone into approximately 250 mL of HPLC-grade water. Dilute to 500
mL with HPLC grade water. Transfer the mobile phase to the labeled solvent reservoirs.
Place the 0.2% acetone solution into reservoirs B (and D if testing a quaternary model).
• Ensure that acetone solution is mixed well and degassed.
• Bypass the column and autosampler and connect pump output directly to the
backpressure regulator or other device (length of tubing for example) to create the
necessary back pressure.
• Connect the output of the back pressure device to the input of the detector flowcell.
• Equilibrate and insure the binary or quaternary valves are operating correctly by running
the pumps as follows:
•Binary pump - run 50/50 at 5mL/min for 10 minutes
Quaternary pump - run 25/25/25/25 at 5 mL/min for 10 minutes.
•Observe the P-counts on the pump (CNFG/CAL/TEST/PCNT). The P-count value should
fall typically in the range of 1300 + 50. Stability is important and the Range value should
be typically less than 12.
• Set the detector wavelength to 265 nm.
• Set the flow rate to 5 mL/minute from reservoir A and allow the pressure to stabilize.
Ensure the pump and baseline is stable at 5 mL/min.
•Create and run the method in Tables 1 and 2 as appropriate with TotalChrom® or
Chromera®, zeroing the detector at the end of step zero. Step 2 is the 100% reference.
Flexar Analytical / FX-10 UHPLC Pumps IQ/OQ Rev. D June 2009
•Record mV or absorbance values from the run for each step (AU1 through AU6, and if
applicable, for the quaternary pump, AU7 through AU10) in the AU column of Tables 1
and 2 below.
Table 1 - Binary and Quaternary (Reservoirs A and B)
STEP TIME FLOW
0 5.0 5.0 100 0 0.0
1 1.0 5.0 100 0 0.0
2 3.0 5.0 0 100 0.0 AU1 =
3 3.0 5.0 10 90 0.0 AU2 =
4 3.0 5.0 49 51 0.0 AU3 =
5 3.0 5.0 51 49 0.0 AU4 =
6 3.0 5.0 90 10 0.0 AU5 =
7 5.0 5.0 100 0 0.0 AU6 =
Table 2 - Quaternary (Reservoirs C and D)
STEP TIME FLOW
0 5.0 5.0 100 0 0.0
%A
(water)
%C
(water)
%B
(acetone)
%D
(acetone)
CURV
CURV
AU
(absorbance or mV)
AU
(absorbance or mV)
1 1.0 5.0 100 0 0.0
2 3.0 5.0 0 100 0.0 AU7 =
3 3.0 5.0 10 90 0.0 AU8 =
4 3.0 5.0 90 10 0.0 AU9 =
5 5.0 5.0 100 0 0.0 AU10 =
Comments:
PerkinElmer LAS Representative:
Date:
Reviewed By: Date:
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
2.4.7 Test Results
PARAMETER SPECIFICATION
Composition Accuracy
Within ± 1.0% absolute @ 5 mL/min
Channel A and B % Solvent B
49 / 51 Delta Range
Between 1.5% and 2.5%
Channel C and D % Solvent D
Comments:
10%
49%
51%
90%
10%
90%
TEST
RESULT
PASS/FAIL
2.5 Operational Qualification Verification
The Flexar Analytical / FX-10 UHPLC Pump, Serial Number Template-SN, operates
and performs according to the stated Operational Qualification, and all appropriate
forms and documents supporting the system have been filled out and are available.
Comments:
PerkinElmer LAS Representative:
Reviewed By: Date:
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
Date:
Verified:
Part III – Appendices
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
Appendix A– Deviations
Describe in detail any deviation from the standard IQ/OQ. The deviation must be approved by
Template-Company representative. If there is no deviation, no signature or date entry is required by
Template-Company representative (enter NA for signature and then date). If there are no deviations,
the PerkinElmer representative should enter “None” below and then sign and date.
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
Appendix B– Change Control
Note: Create copies of this page, leaving original blank. Read instructions below.
Change Control procedures maintain the validation processes after modifications are made to the original
configuration (by OEM, Template-Company, or PerkinElmer LAS).
The Change Control Form is located in Appendix B– Change Control and instructions are located in the
Introduction. Insert the completed Change Control Form and its data into this validation document as an
Attachment.
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
Appendix C– Attachments List
In the table below enter: each attachment; total pages for each attachment; company (OEM, Template-Company, or
PerkinElmer). Sign and date. Entries for up to 20 Attachments are available. Additionally, label each page of each
Attachment as follows:
Attachment Number (as per this table)
Page X of Y (total) Pages of the Attachment
Initials of Reviewer
Date
#
# Description
1 PerkinElmer Service Engineer Training Certificate PerkinElmer
2 Certificate of Conformity
3
4
5
Pgs
Company Signature Date
6
7
8
9
10
11
12
13
14
15
16
17
18
19
20
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
Appendix D- Document History
Revision Description of Change Page(s) Date
A First release May 2009
B Corrected math in instructions and typo in
20 May 2009
results
C Corrected footers. All June 2009
D Corrected footers All June 2009
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
Flexar Analytical / FX-10 UHPLC PumpIQ/OQ Rev. D June 2009
Appendix E- Spine Label for Instrument / System Log Book
Cut out the label below and insert into the spine of the Instrument / System Log Book, then destroy the remainder of this
page as it is no longer required.
Instrument / System Log Book
Template-Company
Model Serial Number Protocol ID
Flexar Analytical / FX-10
UHPLC Pump
Template-SN
Template-
Protocol
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.