PerkinElmer Frontier IR User Manual

Page 1
FRONTIER IR DUAL-RANGE SYSTEMS
MOLECULAR SPECTROSCOPY
User’s Guide
Page 2
Release History
Part Number Release Publication Date
L1050102 C July 2013
Any comments about the documentation for this product should be addressed to:
Or emailed to: info@perkinelmer.com
Notices
The information contained in this document is subject to change without notice.
Except as specifically set forth in its terms and conditions of sale, PerkinElmer makes no warranty of any kind with regard to this document, including, but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
PerkinElmer shall not be liable for errors contained herein for incidental consequential damages in connection with furnishing, performance or use of this material.
Copyright Information
This document contains proprietary information that is protected by copyright. All rights are reserved. No part of this publication may be reproduced in any form whatsoever or translated into any language without the prior, written permission of PerkinElmer, Inc.
Copyright © 2013 PerkinElmer, Inc.
Produced in the United Kingdom
Trademarks
Registered names, trademarks, etc. used in this document, even when not specifically marked as such, are protected by law. PerkinElmer is a registered trademark of PerkinElmer, Inc. Spectrum, AssureID and Frontier are trademarks of PerkinElmer, Inc.
Page 3
Contents
Introduction ............................................................................................... 5
About This Manual ............................................................................................ 6
Conventions Used in this Manual ........................................................................ 7
Notes, Cautions and Warnings ..................................................................... 7
Warnings and Safety Information ........................................................... 11
Safety Summary ...............................................................................................12
General Safety .................................................................................................13
Location and Ventilation ...................................................................................14
To allow for adequate cooling ....................................................................14
To allow any nitrogen used to disperse .......................................................14
Use of flammable solvents and samples ......................................................15
Electrical Safety ...............................................................................................16
Laser Safety Regulations...................................................................................17
Radiation Hazards and their Classification ..........................................................18
Laser radiation ..........................................................................................18
Radiation emitted by an NIR source ............................................................18
Labels .............................................................................................................19
Warning Signs on the Instrument ......................................................................21
EMC Compliance ..............................................................................................22
European EMC Directive .............................................................................22
FCC rules and regulations...........................................................................22
An Overview of the Frontier IR Dual-Range Spectrometers ................... 23
A Guided Tour of the Frontier IR Dual-Range Spectrometers ...............................24
Configuration changes ...............................................................................25
Optical system ...........................................................................................25
Top panel controls .....................................................................................26
The sample compartment ..........................................................................27
Internal accessories ...................................................................................28
External accessories ...................................................................................28
The Spotlight 400 Series Imaging System ...................................................29
Sample compartment windows ...................................................................30
Power switch and communications ports .....................................................30
Accessories for the Frontier IR Spectrometer .....................................................31
Frontier IR Sytems Upgrades ............................................................................33
Unpacking and Installation ...................................................................... 35
Requirements ..................................................................................................36
Electrical requirements ...............................................................................36
Environment ..............................................................................................36
Unpacking the spectrometer .............................................................................37
Opening the shipping container ..................................................................37
The Desiccant Indicator ....................................................................................39
Connecting up the Spectrometer .......................................................................40
Connecting to the PC .................................................................................40
Other connectors .......................................................................................41
Connecting the spectrometer to the electrical supply ...................................42
Installing the Instrument in the Software ...........................................................44
Installing the software ...............................................................................44
The Instrument Install Wizard ....................................................................44
Using the Spectrometer with Spectrum .................................................. 45
Basics of Software Control ................................................................................46
Starting Spectrum software ........................................................................46
Scanning samples ......................................................................................47
Changing the instrument range ..................................................................48
Changing the beampath .............................................................................49
Working with the instrument display and Go button .....................................49
Using the Spectrum on-screen Help system .................................................50
Page 4
Atmospheric (CO2/H2O) Suppression ................................................................. 51
What is atmospheric suppression? .............................................................. 51
What does atmospheric suppression do? .................................................... 51
AVI Correction ................................................................................................. 53
What is AVI correction? ............................................................................. 53
What does AVI correction do? .................................................................... 54
Look Ahead ..................................................................................................... 55
What is Look Ahead? ................................................................................. 55
What does Look Ahead do? ....................................................................... 55
Quality Checks ................................................................................................. 56
What are Quality Checks? .......................................................................... 56
What do Quality Checks do? ...................................................................... 56
Routine Maintenance ............................................................................... 57
Cleaning the spectrometer ............................................................................... 58
Cleaning the display .................................................................................. 58
Moving the Spectrometer ................................................................................. 59
Condensation ............................................................................................ 60
The Set Temperature utility ....................................................................... 60
Replacing the Frontier FT-IR/FT-FIR Sample Compartment Windows .................... 61
The Desiccant Indicator in Detail ...................................................................... 64
Changing the Desiccant.................................................................................... 65
Renewing the instrument desiccant ............................................................ 66
Installing rechargeable desiccant in the instrument ..................................... 67
Renewing the desiccant in the spare window compartment ......................... 68
Purging the Spectrometer ................................................................................ 70
Changing the External Fuse .............................................................................. 73
Cooling the MCT Detector (If Fitted) ................................................................. 75
Advanced Maintenance ............................................................................ 79
Opening the Main Cover ................................................................................... 80
Replacing a Source .......................................................................................... 83
Installing/Replacing Beam Port and Sample Compartment Windows ................... 85
Installing Filters in the Filter Wheel ................................................................... 87
Replacing the Laser and Power Supply .............................................................. 91
Appendices ............................................................................................... 95
Appendix 1: Changing the Sample Area Accessory ............................................. 96
Appendix 2: Instrument Self-Checks ................................................................. 98
Appendix 3: Instrument Performance Validation .............................................. 100
Appendix 4: Decontamination and Cleaning ..................................................... 101
Appendix 5: WEEE Instructions for PerkinElmer Products ................................. 102
Index ............................................................................................................ 103
Page 5

Introduction

Page 6
6 . Frontier IR Dual-Range Systems User's Guide

About This Manual

This manual contains the following sections:
Introduction
Warnings and Safety Information
An Overview of the Frontier-IR Dual-Range Spectrometers
Unpacking and Installation
Using the Spectrometer with Spectrum
Routine Maintenance
Advanced Maintenance
Appendices.
For further information on collecting, viewing and processing spectra using Spectrum software, refer to the help file that you can access from the Help menu, or by clicking Help on a dialog. The Help menu also includes links to on-screen tutorials.
NOTE: This manual shows details for using your instrument with the Spectrum software
package (version 10 or later). If you have Spectrum ES or AssureID software, please refer to the Administrator’s Guide for your software which can be found on the
IR & Raman Manuals CD
(L1050002), or refer to the on-screen Help.
Page 7
Introduction . 7

Conventions Used in this Manual

Normal text is used to provide information and instructions.
Bold text refers to text that is displayed on the screen.
UPPERCASE text, for example ENTER or ALT, refers to keys on the PC keyboard. '+' is used to show that you have to press two keys at the same time, for example, ALT+F.
All eight digit numbers are PerkinElmer part numbers unless stated otherwise.
The term “instrument” refers to a Frontier IR spectrometer, and any sampling accessory fitted. Frontier IR Dual-Range spectrometer refers to the Frontier FT-IR/FT-NIR and the Frontier FT-IR/FT-FIR spectrometers.

Notes, Cautions and Warnings

Three terms, in the following standard formats, are also used to highlight special circumstances and warnings.
NOTE: A note indicates additional, significant information that is provided with some
procedures.
Page 8
8 . Frontier IR Dual-Range Systems User's Guide
We use the term CAUTION to inform you about situations that could
or other equipment. Details
ou d'autre matériel. Les détails sur ces circonstances figurent
小心
注意
CAUTION
result in serious damage to the instrument about these circumstances are in a box like this one.
Caution (Achtung)
Bedeutet, daß die genannte Anleitung genau befolgt werden muß, um einen Geräteschaden zu vermeiden.
Caution (Bemærk)
Dette betyder, at den nævnte vejledning skal overholdes nøje for at undgå en beskadigelse af apparatet.
Caution (Advertencia)
Utilizamos el término CAUTION (ADVERTENCIA) para advertir sobre situaciones que pueden provocar averías graves en este equipo o en otros. En recuadros éste se proporciona información sobre este tipo de circunstancias.
Caution (Attention)
Nous utilisons le terme CAUTION (ATTENTION) pour signaler les situations susceptibles de provoquer de graves détériorations de l'instrument dans un encadré semblable à celui-ci.
Caution (Attenzione)
Con il termine CAUTION (ATTENZIONE) vengono segnalate situazioni che potrebbero arrecare gravi danni allo strumento o ad altra apparecchiatura. Troverete informazioni su tali circostanze in un riquadro come questo.
Caution (Opgelet)
Betekent dat de genoemde handleiding nauwkeurig moet worden opgevolgd, om beschadiging van het instrument te voorkomen.
Caution (Atenção)
Significa que a instrução referida tem de ser respeitada para evitar a
danificação do aparelho.
Caution (
我们使用“小心”这一术语来通知您有关可能会对
)
本仪器或其它设备造成严重损害的情况。 有关这些情况的详细信息可在此类方框中找到。
Caution (
分光器や他の機材等に深刻なダメージを与える恐れがある場合は、 この様なボックスの中に表示しています。
)
Page 9
WARNING
警告
警告
Introduction . 9
We use the term WARNING to inform you about situations that could result in personal injury to yourself or other persons. Details about these circumstances are in a box like this one.
Warning (Warnung)
Bedeutet, daß es bei Nichtbeachten der genannten Anweisung zu einer
Verletzung des Benutzers kommen kann.
Warning (Advarsel)
Betyder, at brugeren kan blive kvæstet, hvis anvisningen ikke overholdes.
Warning (Peligro)
Utilizamos el término WARNING (PELIGRO) para informarle sobre situaciones que pueden provocar daños personales a usted o a otras personas. En los recuadros como éste se proporciona información sobre este tipo de circunstancias.
Warning (Danger)
Nous utilisons la formule WARNING (DANGER) pour avertir des situations pouvant occasionner des dommages corporels à l'utilisateur ou à d'autres personnes. Les détails sur ces circonstances sont données dans un encadré semblable à celui-ci.
Warning (Pericolo)
Con il termine WARNING (PERICOLO) vengono segnalate situazioni che potrebbero provocare incidenti alle persone. Troverete informazioni su tali circostanze in un riquadro come questo.
Warning (Waarschuwing)
Betekent dat, wanneer de genoemde aanwijzing niet in acht wordt genomen, dit kan leiden tot verwondingen van de gebruiker.
Warning (Aviso)
Significa que a não observância da instrução referida poderá causar um ferimento ao usuário.
Warning (
我们使用“警告”这一术语来通知您有关可能会对您自己或他人造成人
)
身伤害的情况。 有关这些情况的详细信息可在此类方框中找到。
Warning (
使用者及びその他周辺に危害が及ぶ恐れがある場合は、 この様なボックスの中に注意事項が表示されています。
)
Page 10
10 . Frontier IR Dual-Range Systems User's Guide
Page 11
Warnings and Safety
Information
Page 12
12 . Frontier IR Dual-Range Systems User's Guide

Safety Summary

The Frontier IR spectrometers have been designed to comply with a wide variety of international standards governing the safety of laboratory equipment. In routine use, the instruments pose virtually no risk to you. If you take some simple, common-sense precautions, you can make sure that you maintain the continued safe operation of your instrument:
DO make sure that the instrument is properly connected to the electrical supply; in particular make sure that the ground (earth) is securely connected.
DO disconnect the electrical power cable before opening the main cover of the instrument.
DO keep the instrument dry. Avoid spilling liquid into the instrument. Clean all external spills immediately. If anything that is spilled enters the main body of the instrument, switch off the power and contact a PerkinElmer Service Engineer.
DO NOT stare into the internal laser beam under the instrument cover. The instrument contains a low power, visible (red) laser; momentary exposure to the beam is not dangerous, but deliberate, direct viewing of the beam along its axis could damage your eye.
DO NOT use a flammable gas to purge the instrument. The instrument contains a hot source, and a fire or explosion will result. Only use clean, dry, oil-free nitrogen or air to purge the instrument.
DO read the more detailed information on warnings and safety in the following pages to ensure the safe operation of the instrument.
Page 13
Warnings and Safety Information . 13

General Safety

The Frontier IR spectrometers have been designed and tested in accordance with PerkinElmer specifications and in accordance with the safety requirements of the International Electrotechnical Commission (IEC). The instruments conform to IEC publication 61010-1 (“Safety requirements for electrical equipment for measurement, control, and laboratory use”) as it applies to IEC Class 1 (earthed) appliances and therefore meets the requirements of EC low voltage Directive 2006/95/EC.
If the instrument is used in a manner not specified by the manufacturer, the protection provided by the instrument may be impaired. Only use the instrument indoors and under the following conditions:
Temperature 15 °C to 35 °C Relative Humidity 80% maximum (non-condensing)
If possible, avoid any adjustment, maintenance and repair of the opened, operating instrument. If any adjustment, maintenance and repair of the opened, operating instrument is necessary, this must only be done by a skilled person who is aware of the hazard involved.
Whenever it is likely that the instrument is unsafe, make it inoperative. The instrument may be unsafe if it:
Shows visible damage
Fails to perform the intended measurement
Has been subjected to prolonged storage in unfavorable conditions
Has been subjected to severe transport stresses.
If the equipment is used in a manner not specified herein the protection provided by the equipment may be impaired.
WARNING
The instrument has been designed to be safe under the following environmental conditions:
Indoor use
Altitude up to 2000 m (above mean sea level)
Ambient temperatures of 5 °C to 40 °C
A maximum ambient relative humidity of 80% for temperatures up to 31 °C, decreasing
linearly to 50% relative humidity at 40 °C
Mains supply fluctuations not exceeding ±10% of the nominal voltage.
Page 14
14 . Frontier IR Dual-Range Systems User's Guide

Location and Ventilation

Make sure that the switch at the electrical supply inlet on the rear of the
.
instrument is not obstructed
WARNING

To allow for adequate cooling

Do not site the instrument near to room heating equipment, for example, central heating radiators.
During operation, there should be a minimum gap of:
15 cm (6 inches) between any surface and the cooling louvers at the rear of the
instrument.
7 cm (3 inches) between the instrument and adjacent equipment.
45 cm (18 inches) between any surface and the top surface of the closed sample area
lid (to allow for the lid to be opened fully).

To allow any nitrogen used to disperse

Do not site the instrument in a poorly ventilated area if nitrogen will be used as a purge gas, or if liquid nitrogen will be used to cool a detector.
WARNING
The spectrometer includes a coupling that enables the instrument body, and another coupling that enables the sample compartment, to be purged using clean, dry, oil-free air or nitrogen. The recommended flow rate to each connector is 10 l/min, and both the instrument and the sample compartment vent to their surroundings.
If the instrument is fitted with an MCT detector, this is cooled using liquid nitrogen. A risk assessment should include personal protection protocols, and proper actions in the event of accidental spillage. 1 l (2 US pints) of liquid nitrogen evolves to 700 l (approximately 25 cubic feet) of nitrogen gas, and the cold vapor can pool at floor level.
Oxygen depletion in an enclosed space does not trigger a gasping reflex, and errors of judgment, confusion, or unconsciousness can occur in seconds and without warning.
Page 15
Warnings and Safety Information . 15
If flammable solvents or samples are spilled on the instrument and there

Use of flammable solvents and samples

The instrument contains a hot source and contact with flammable vapors may cause an explosion. When working with flammable solvents or samples, particularly during unattended operation with flow-cells, it is
WARNING
WARNING
recommended that the instrument optics area should be continuously purged with dry air or nitrogen to maintain a positive pressure and prevent flammable vapor from entering the instrument.
is any possibility that they have entered the interior (by coming into contact with cover gaskets, for example) then the instrument must be switched off immediately and disconnected from the power supply. The optics area should then be thoroughly purged with dry air or nitrogen, or the main cover should be opened to thoroughly ventilate the optics area before proceeding.
WARNING
WARNING
Flammable solvents or samples should not be stored on or near the instrument. Handling of such materials during preparation should be performed in a safe area away from the instrument such as a fume cabinet.
Do not use a flammable gas to purge the Frontier IR instruments. Use only clean, dry, oil-free nitrogen or air.
Page 16
16 . Frontier IR Dual-Range Systems User's Guide

Electrical Safety

Connect the instrument to a power supply line that includes a switch or other means of
disconnection from the electricity supply.
Only plug the instrument into an electricity-supply socket that is provided with a
protective earth connection.
When fuses need replacing, use only those with the required current rating and of the
specified type. Do not use makeshift fuses and do not short-circuit fuse holders.
When the instrument is connected to its electricity supply, terminals may be live and the
removal of covers other than those which can be removed by hand is likely to expose live parts.
Capacitors inside the instrument may still be charged even if the instrument has been
disconnected from all voltage sources.
The instrument must be disconnected from all voltage sources before it is opened for
any adjustment, replacement, maintenance or repair.
Any interruption of the protective earth conductor inside or outside the instrument or disconnection of the protective earth terminal can make the instrument dangerous.
WARNING
The instrument has an IEC Insulation Class I rating for external circuits – only connect other equipment that meets the requirements of IEC 61010-1, IEC 60950 or equivalent standards.
Page 17
Warnings and Safety Information . 17

Laser Safety Regulations

The Frontier IR spectrometers are Class 1 laser products as defined by IEC 60825-1. The optical module contains a Class 2 Helium Neon (HeNe) laser, which emits visible, continuous wave radiation at a wavelength of 633 nm and has a maximum output power of 1 mW. Some diffuse HeNe laser radiation, within Class 1 limits, emerges from:
The window in the left side of the sample compartment when an internal beam path is selected.
An external beam port when the port cover is removed, no accessory is fitted at the port, and the beam path to the port is selected.
Do not stare into any laser beam. Staring into a laser beam (intrabeam viewing) can cause permanent damage to your eyes.
WARNING
The laser is automatically shut down when the main cover of the instrument is raised.
Do not attempt to override or modify the interlock system.
Use of controls or adjustments or performance of procedures other than
WARNING
The instrument complies with the following laser safety regulations:
1. 21 CFR Part 1040.10 and 1040.11 except for deviations pursuant to Laser Notice No. 50 dated 24 June 2007. Administered by the Center for Devices and Radiological Health, U.S. Department of Health and Human Services.
2. European Standard EN 60825-1:2007 – “Safety of laser products – Part 1: Equipment classification and requirements”.
those specified herein may result in hazardous radiation exposure.
Page 18
18 . Frontier IR Dual-Range Systems User's Guide

Radiation Hazards and their Classification

Laser radiation

Indirect observation of the laser beam radiation in the optical path is not hazardous. Directly viewing the laser beam along its axis (allowing the laser beam radiation to pass into the eye) can be hazardous, depending upon the power of the beam, the length of time that the eye is exposed to the beam and the optical efficiency of the exposed eye. Direct viewing of a laser beam along its axis is termed
Protection of the eye during accidental, momentary intrabeam viewing of a Class 2 laser beam is normally given by the eye’s aversion response, including the blink reflex, which limits exposure of the eye to less than 0.25 seconds.
Class 1 levels of laser radiation are not considered to be hazardous.
intrabeam viewing

Radiation emitted by an NIR source

Your Frontier IR instrument can be fitted with an NIR source. The quartz halogen bulb, which produces the near infrared beam, emits ultraviolet, visible and infrared radiation. The majority of this radiation is in the infrared region. Do not stare into the beam produced by this bulb.
.
Measurements of the infrared radiation emitted from the sample area show that exposure limits recommended by the American Conference of Governmental Industrial Hygienists (ACGIH) and International Commission on Non-Ionizing Radiation Protection (ICNIRP) will not be exceeded during normal operation.
Page 19
Warnings and Safety Information . 19

Labels

When this label is attached to an instrument it means “Caution, risk of danger”. Refer to the manual to find out the nature of the potential hazard and any actions that have to be taken.
The product identification label is on the front of the instrument. Other labels are fixed to the Frontier IR spectrometers in the locations shown in Figure 1 and Figure 2:
Figure 1 Labels (rear of instrument)
NOTE: The label with a crossed-out wheeled bin symbol and a rectangular bar indicates that
the product is covered by the Waste Electrical and Electronic Equipment (WEEE) Directive. Refer to page 102.
Appendix 5: WEEE Instructions for PerkinElmer Products
on
Page 20
20 . Frontier IR Dual-Range Systems User's Guide
Serial number, part
number and date of manufacture
NOTE: There is a second copy of the CAUTION label inside the sample compartment cover.
Figure 2 Labels (sample compartment)
Page 21
Warnings and Safety Information . 21

Warning Signs on the Instrument

Caution, hot surface.
Caution, risk of electric shock.
Caution, laser radiation hazard.
Caution, risk of danger.
Refer to accompanying documents in all cases where this symbol is used to find out the nature of the potential hazard and any actions which have to be taken.
Page 22
22 . Frontier IR Dual-Range Systems User's Guide

EMC Compliance

European EMC Directive

The Frontier IR spectrometers have been designed and tested to meet the requirements of the EMC Directive 2004/108/EC.

FCC rules and regulations

These products are classified as digital devices used exclusively as industrial, commercial, or medical test equipment. They are exempt from the technical standards specified in Part 15 of the FCC Rules and Regulations, based on Section 15.103(c).
Page 23
An Overview of the
Frontier IR Dual-Range
Spectrometers
Page 24
24 . Frontier IR Dual-Range Systems User's Guide

A Guided Tour of the Frontier IR Dual-Range Spectrometers

PerkinElmer Frontier IR Dual-Range spectrometers are bench-top FT-IR instruments (Figure 3). Fully optimized performance is achieved by the provision of dual beamsplitters, dual sources (Frontier FT-IR/FT-NIR only) and dual detectors.
Figure 3 Frontier IR FT-IR/FT-NIR spectrometer
For the MIR configuration:
The optical system enables you to collect data over a total range of 7800 to 350 cm with a best resolution of 0.4 cm
1
.
1
A DTGS (deuterated triglycine sulfate) MIR detector or an MCT (mercury cadmium telluride) detector is fitted.
KBr windows are fitted in the sample compartment.
For the NIR configuration:
The optical system enables you to collect data over a total range of 14700 to 1250 cm with a best resolution of 0.4 cm
1
.
An NIR DTGS (deuterated triglycine sulfate) detector is fitted.
KBr windows are fitted in the sample compartment.
For the FIR configuration:
1
The optical system enables you to collect data over a total range of 700 to 30 cm a best resolution of 0.4 cm
1
.
with
An FIR DTGS (deuterated triglycine sulfate) detector is fitted.
1
A PE (polyethylene) window and a PE/D (polyethylene/diamond) window are supplied to be fitted in the sample compartment for operation in the FIR range.
The instrument can operate in ratio, single-beam, or interferogram mode.
Page 25
An Overview of the Frontier IR Dual-Range Spectrometers . 25
The instrument is connected to a PC, either point-to-point or over a network. The Spectrum software package supplied enables you to control the instrument and to manipulate the spectra that you collect.

Configuration changes

Spectrum software enables you to initiate an automated changeover between, for example, the MIR and NIR ranges, changing the source, detector, and beamsplitter. You can also individually select any combination of these elements.
NOTE: If you change between the MIR and FIR ranges, the software will prompt you to
change the sample compartment windows. For information about changing windows, see
Replacing the Frontier FT-IR/FT-FIR Sample Compartment Windows
When you change the source or detector, or swap between an external and internal accessory, the infrared beam is redirected using the appropriate rotating mirror.
Before you change to the MIR MCT detector, fit one of the attenuators supplied (Attenuator Set – L1160560) to the window to the right of the sample compartment to prevent the detector from overloading. If you have an open beam we recommend a 14%T attenuator. See
Windows Windows
on page 61 or
on page 85 for more information.
Replacing the Frontier FT-IR/FT-FIR Sample Compartment
Installing/Replacing Beam Port and Sample Compartment
on page 61.
NOTE: If you have a Frontier FT-IR/FT-FIR spectrometer with an MCT detector and the
instrument was in the MIR range when it was switched off, we recommended that a 6%T attenuator is fitted to the sample compartment window before switching on the instrument again, to prevent errors during the initialization of the FIR detector.
When you change the detector, allow it to equilibrate before scanning. The MCT detector (if fitted) will not perform correctly until it has been sufficiently cooled using liquid nitrogen.
When you select the alternate beamsplitter, there is a pause while the element is changed and initialized. The performance of the beamsplitter is temperature sensitive, so there is also an equilibration period.

Optical system

The optical system is under the main cover of the instrument. Usually the main cover of the instrument is closed, but to perform most maintenance tasks, the cover has to be open. When you do this, a safety interlock automatically switches off the power. Nevertheless, the instrument should be disconnected from the mains before opening the main cover for maintenance.
Consistent, reliable performance is achieved by having few adjustable parts, and by extensive insulation of the optical system from the effects of humidity and vibration.
Page 26
26 . Frontier IR Dual-Range Systems User's Guide
Display
Go button
Stability of the optical system
The entire optical system is purged and sealed at the factory. A supply of desiccant placed within the system removes any water vapor and carbon dioxide that may enter. A desiccant indicator is fitted in the top cover, which warns you when the desiccant needs changing (an internal humidity sensor is available as an option).
KBr windows (or polyethylene and polyethylene/diamond for the FIR range) separate the sample compartment from the purged optical system. You can purge the sample compartment with clean, dry, oil-free air or nitrogen. Either one removes water vapor; however, nitrogen is preferable because it also removes atmospheric carbon dioxide.
Do not use a flammable gas to purge the Frontier IR instruments. The instrument contains a hot source, and a fire or explosion will result. Only use clean, dry, oil-free nitrogen or air to purge the instrument.
WARNING
The optics are kinematically mounted to ensure accurate positioning and to make them rugged. The interferometer is enclosed and mounted on anti-vibration mounts to guard against air and bench borne disruptions. The interferometer uses very low-friction point bearings and a frictionless electromagnetic drive to ensure long life.

Top panel controls

The display on the front right of the top of your instrument (Figure 4) has two purposes:
To display messages generated by the instrument’s firmware, such as those that monitor initialization and diagnostics when the instrument is switched on.
To display prompts and other messages generated by Spectrum software.
The instrument’s Go button is used in concert with the displayed prompts and other messages generated by Spectrum software. It allows you to Scan or Halt data collection while you are away from your computer.
Figure 4 Frontier IR top panel controls
Page 27
An Overview of the Frontier IR Dual-Range Spectrometers . 27
Sample compartment lid
Baseplate

The sample compartment

The instrument has a large, purgeable sample compartment (Figure 5) located at the front of the instrument.
Figure 5 Frontier IR sample compartment
The windows in the sample compartment and external beam ports are composed of KBr (for MIR and NIR configurations) that, although coated, can be damaged by high levels of humidity.
CAUTION
If you spill a liquid in the sample compartment, wipe it up quickly. When working with water-based samples open to the air, either purge the sample compartment or leave the cover open.
A relative humidity higher than 80% can damage the windows of the
CAUTION
sample compartment. If you expect the humidity to exceed 80%, continually purge or desiccate the sample compartment.
Open the sample compartment by lifting the cover using the recess at the front.
The infrared beam enters the sample compartment through an aperture on the left. After passing through the sample, it enters the detector area through an aperture on the right side of the sample compartment.
Standard sampling accessories are mounted on a baseplate. The standard baseplate includes a sample holder located by a knurled screw.
Page 28
28 . Frontier IR Dual-Range Systems User's Guide

Internal accessories

A wide range of optional MIR and NIR accessories, such as the Universal ATR accessory, fit in the sample compartment (Figure 6).
Figure 6 Frontier IR with Universal ATR accessory
It is easy to remove an internal accessory and to replace it with another. See
Changing the Sample Area Accessory
The instrument identifies the type of accessory fitted using a coded connector on its baseplate that plugs into a socket on the rear wall of the sample compartment.
on page 96.
Appendix 1:

External accessories

The external beam port on the right side of the instrument is used, for example, with the optional External Near Infra Red Accessory (External NIRA). Spectrum software enables you to switch the instrument between this external accessory and another accessory fitted in the sample compartment.
Figure 7 Frontier IR with External NIRA accessory
NOTE: To use your instrument with an external accessory, you require the optional External
Beam Pack. If you did not purchase the External Beam Pack with your spectrometer, it is available as a field-fitted upgrade (L1280248).
For information about removing or refitting an external accessory, refer to its User’s Guide, which is distributed on the
IR & Raman Manuals CD
(L1050002) as a .pdf file.
Page 29
CAUTION
CAUTION
An Overview of the Frontier IR Dual-Range Spectrometers . 29
Do not attempt to remove the external accessory bracket mounted within the handhold on the right of the instrument or to transfer the bracket to another spectrometer.
The external accessory bracket is fitted by a PerkinElmer Service Engineer, along with any other internal components that may be required.
Do not attempt to lift the instrument, or to move it to another location, when an external accessory is attached.
The optical alignment of the accessory may be disturbed.
For information about moving the instrument, see
Moving the Spectrometer
on page 59.

The Spotlight 400 Series Imaging System

The external port on the left side of the instrument is used with the Spotlight 400 Imaging System (Figure 8) and SpectrumIMAGE software.
Figure 8 Spotlight 400 Series Imaging System
NOTE: To use your instrument with a microscope, you require the optional External Beam
Pack. If you did not purchase the External Beam Pack with your spectrometer, it is available as a field-fitted upgrade (L1280248).
Page 30
30 . Frontier IR Dual-Range Systems User's Guide
Desiccated
Compartment Spare window

Sample compartment windows

The Frontier FT-IR/FT-NIR instrument has KBr windows fitted in the sample compartment. It is possible to change the sample compartment windows if they are not suitable for your application. For information about changing windows, see
Sample Compartment Windows
The Frontier FT-IR/FT-FIR instrument has KBr windows fitted in the sample compartment for the MIR range, but a PE (polyethylene) window and a PE/D (polyethylene/diamond) window (supplied) should be fitted for the FIR range. For information about changing these windows,
Replacing the Frontier FT-IR/FT-FIR Sample Compartment Windows
see
on page 85.
Changing windows allows some CO2 and water vapor from the surroundings to enter the instrument.
CAUTION
The instrument desiccant packs should remove these contaminants within 30 minutes.
A desiccated compartment is provided for storing spare windows.
Installing/Replacing Beam Port and
on page 61.
NOTE: For long-term storage of spare windows we recommend that you use a desiccator.
desiccant indicator
compartment
Figure 9 Desiccated compartment for spare windows

Power switch and communications ports

The power switch, AC power cable connector, and communications ports are on the rear of the instrument. The power switch is marked I/O (on/off).
NOTE: It can take the instrument about two hours to equilibrate when switched on after
being switched off overnight. To save time, we suggest that you leave the instrument switched on at all times.
Page 31
An Overview of the Frontier IR Dual-Range Spectrometers . 31

Accessories for the Frontier IR Spectrometer

A wide range of optional accessories, such as the Universal ATR accessory (UATR), fit in the sample compartment. In addition, a number of external accessories are available, including the External NIRA. An external accessory can be fitted in conjunction with another internal accessory (Figure 10).
Figure 10 Frontier IR spectrometer with Universal ATR and External NIRA accessories
There is a range of specialized sampling accessories available for the Frontier IR instruments, as shown in Table 1.
Table 1 Sampling accessories for the Frontier IR instruments
Accessory Type MIR NIR FIR
Sample Shuttle Internal Yes Yes Yes
Horizontal ATR Internal Yes n/a n/a
Universal ATR Internal Yes n/a n/a
Diffuse Reflectance Internal Yes n/a Yes
NIRA Internal (limited utility) Yes n/a
External NIRA External (limited utility) Yes n/a
NIR Fiber Optic Probe Internal (limited utility) Yes n/a
NIR Tablet Autosampler* Internal n/a Yes n/a
Liquid Sipper* Internal Yes Yes n/a
TL 8000 EGA System (TG-IR) Yes Yes No
Page 32
32 . Frontier IR Dual-Range Systems User's Guide
Accessory Type MIR NIR FIR
General Purpose Optical Bench (GPOB) and External MCT
External Yes Yes n/a
Detector**
General Purpose Optical Bench (GPOB) and External LiTaO
3
External Yes Yes n/a
Detector**
*The NIR Tablet Autosampler accessory is supported by AssureID software and Spectrum software version 6.x. It is not supported in Spectrum software version 10.0 or later. ** For more details of the GPOB options, contact your PerkinElmer Sales Representative.
Further information about the use of these accessories can be found on the
Manuals CD
(L1050002).
IR & Raman
Page 33
An Overview of the Frontier IR Dual-Range Spectrometers . 33

Frontier IR Sytems Upgrades

The Frontier IR systems are designed to be flexible, upgradable instruments. The upgrades available for Frontier IR Dual-Range Systems are listed in Table 2.
Table 2 Frontier IR Dual-Range Systems Upgrades
Part Number Description
L1280248 External Pack Beam Upgrade – Microscope, GPOB, NIRA and
external beam input port ready The kit contains KBr windows, but other windows are available. Ask your PerkinElmer Customer Service Representative for details
L1280252 Frontier FT-IR/FT-NIR Narrow band MCT Detector Upgrade
(replaces standard MIR detector)
L1280253 Frontier FT-IR/FT-NIR Medium band MCT Detector Upgrade
(replaces standard MIR detector)
L1280254 Frontier FT-IR/FT-NIR Wide band MCT Detector Upgrade
(replaces standard MIR detector)
L1280255 Frontier FT-IR/FT-FIR Narrow band MCT Detector Upgrade
(replaces standard MIR detector)
L1280256 Frontier FT-IR/FT-FIR Medium band MCT Detector Upgrade
(replaces standard MIR detector)
L1280257 Frontier FT-IR/FT-FIR Wide band MCT Detector Upgrade
(replaces standard MIR detector)
L1200391 Electronic Humidity Sensor
Page 34
34 . Frontier IR Dual-Range Systems User's Guide
Page 35
Unpacking and
Installation
Page 36
36 . Frontier IR Dual-Range Systems User's Guide

Requirements

NOTE: Read the warnings and safety information at the start of this manual before you
install the instrument. They contain important information.

Electrical requirements

The Frontier IR instruments can operate on electricity supplies of 50 or 60 Hz and in the 100 V to 230 V range without any adjustment.
The nominal power consumption of the instrument is 120 VA.
The line supply must be within 10% of the nominal voltage.
If possible, do not connect the instrument to circuits that have heavy-duty equipment connected, such as large motors.
If possible, do not use photocopiers, discharge lamps, radio transmitters, and other equipment with large or frequent transient loads on the same supply circuit.
The primary fuse (2 AT, 250 V) is in the drawer on the rear of the instrument next to the mains inlet: the spare fuse is in the same drawer. The primary fuse is connected in the live line.

Environment

To obtain the best performance from your instrument:
Place the instrument in an environment that is relatively dust-free.
Make sure that the bench top is free from vibration or mechanical shocks.
Do not place the instrument or the PC near to room-heating equipment, such as
central-heating radiators.
Do not position the instrument in direct sunlight, as this may cause overheating.
Leave at least 15 cm (6 inches) from any surface and the cooling louvers at the rear of
the instrument.
Leave at least 7 cm (3 inches) from any vertical obstacle to the sides of the instrument, to permit an adequate flow of cooling air.
Make sure that there are no overhanging shelves, and no water pipes or faucets that could leak onto the instrument.
The area near the PC must be free of strong magnetic fields, direct sunlight, and heating or cooling units or ducts.
The instrument has been designed for indoor use and operates correctly under the following conditions:
Ambient temperature 15 °C to 35 °C
Ambient relative humidity 80% maximum (non-condensing).
Page 37

Unpacking the spectrometer

The Frontier IR spectrometer is a heavy precision instrument, so two people are required for safe handling.
WARNING
CAUTION

Opening the shipping container

The instrument weighs approximately 36 kg unpacked (40 kg packed) and has a lifting recess on either side. Consult the local codes of practice issued by safety advisors before attempting to lift it.
Take care not to injure yourself or others, or to drop the instrument.
Take great care when installing your Frontier IR spectrometer, and follow the procedures described in this manual. If you require assistance, contact your local PerkinElmer Service Engineer.
Unpacking and Installation . 37
Your instrument is packed inside the box in a silver bag that protects it from condensation. The beamsplitters are packed as a kit in a separate protective box.
1. First remove the manuals/software box, leads, and so on from the box, and check that all the following parts are present:
Part
Number
L1250432 Quick Change Window Kit (Frontier FT-IR/FT-FIR only) 1
L1202057 Polystyrene Calibration Film 1
L1250405 NIR Performance Validation Kit (Frontier FT-IR/FT-NIR only) 1
04974265 Quick Release Purge Coupling 2
L1200466 Ethernet Crossover Cable 1
04970839 2A, 250V Time Lag Fuse 2
L9003693 8.0 mm Hexagonal Wrench 1
N0171159 Desiccant Kit (Frontier FT-IR/FT-FIR only) 1
Description
L1050002 IR & Raman Manuals CD 1
LX108873
OR
LX108875
L1250230 Spectrum Configuration Disk 1
Spectrum Standard Software Kit
OR
Spectrum ES Software Kit
1
Page 38
38 . Frontier IR Dual-Range Systems User's Guide
Part
Number
L1240055 Cuvette Holder Assembly (Frontier FT-IR/FT-NIR only) 1
L1240056 Cell Holder Assembly and Disposable Cells (5)
(Frontier FT-IR/FT-NIR only)
L1160560 Attenuator Set: 1%T, 4%T, 6%T, 14%T and 32%T
(Frontier IR Systems with MCT MIR detector only)
L1250397
OR
L1250438
If any items are missing or damaged, contact your local PerkinElmer office.
2. Carefully remove the instrument from the shipping container, but not from the bag in which it was shipped.
Frontier FT-IR/FT-NIR Beamsplitter Kit (as applicable)
OR
Frontier FT-IR/FT-FIR Beamsplitter Kit (as applicable)
Description
1
1
1
The spectrometer and beamsplitters must be allowed to reach the
CAUTION
temperature of their surroundings. This means leaving them in their packaging overnight if moved from a cold area, and for at least four hours after removal from the shipping container.
3. When the spectrometer has been allowed to warm to the temperature of its surroundings, remove it from the bag and place it on the bench where it is to be used.
Ensure that you can reach the rear of the spectrometer to connect the cables.
Do not open the beamsplitter kit, or attempt to fit the beamsplitters yourself.
CAUTION
The beamsplitters must be fitted by a PerkinElmer Service Engineer.
NOTE: Any accessories will be shipped in separate boxes.
Page 39
Unpacking and Installation . 39
must
Pink
Blue
Blue

The Desiccant Indicator

The optical system of the spectrometer is purged at the factory. This protects the beam port windows (and, once fitted, the beamsplitters) from being damaged by humidity. Replaceable packs of desiccant keep the optics dry and free of CO
The top panel of the instrument includes a desiccant indicator, whose sectors change sequentially from blue to pink as the desiccant becomes exhausted (Figure 11). Change the desiccant in the instrument when the sector marked 10 becomes pink, but while the sectors marked 15 and 20 are still blue. These numbers correspond to the approximate % Relative Humidity in the instrument.
.
2
Refer to appropriate action to take if the sectors marked 15 and 20 are pink.
Figure 11 The desiccant indicator
The Desiccant Indicator in Detail
on page 61 for more information about the
If all three sectors of the desiccant indicator are pink, then you change the desiccant.
CAUTION
The instrument optics may be fogged. Do not switch the instrument either ON or OFF until all sectors are blue.
NOTE: A similar desiccant indicator is fitted in the desiccated compartment (Figure 34 on
page 68). This is used for storing sample compartment windows when not in use.
Page 40
40 . Frontier IR Dual-Range Systems User's Guide
instrument, or between a Local Area Network (LAN)
PC or network

Connecting up the Spectrometer

Connecting to the PC

NOTE: To control your instrument from a PC, use the crossover cable supplied.
To control your instrument over your network, use a standard Ethernet cable (not supplied).
1. Plug one end of the cable into the connector port on the right-hand side of the instrument (Figure 12).
Figure 12 PC Connection port
2. Plug the other end of the cable into the network connection on your PC (if you are using the crossover cable) or network hub (if you are using a standard Ethernet cable).
Connector details
Description Connector
Type
10/100 Base-T Ethernet connector. This is the standard interface between the PC and the
and the instrument. If connecting directly from the PC to the instrument use the Category 5 UTP Cross-Over cable supplied. If you are connecting to your network, use a standard Ethernet cable.
DO NOT USE THE CROSS-OVER CABLE TO CONNECT TO YOUR NETWORK.
Ethernet <5 V <100 mA
Voltage Maximum
Current
Page 41
Unpacking and Installation . 41
Sample area purge
Instrument purge
10101
Spt. Lt.
EXT.R

Other connectors

The communication ports for peripheral devices are shown in Figure 13.
Figure 13 Communication ports
Legends and icons that identify the function of each of the ports are printed on the hinge molding directly above the port.
Do not attempt to connect a monitor to either the EXT.R port or the
CAUTION
MICROSCOPE port or you will cause serious damage to the instrument
when it is switched on.
Connector details
Icon Description Connector
Type
10101
Serial port. Connects a PC to the instrument using an RS232 interface cable. This connector is only used for service diagnostics.
9-way
D-type
Voltages Maximum
Currents
+12 V <100 mA
12 V <100 mA
Spt.Lt
General Purpose Optical Bench (GPOB). Allows synchronization between the instrument and an external accessory.
Spotlight. Used to connect the instrument to a Spotlight Imaging System.
26-way
high density
D-type
11-way mixed
D-type
+5 V <50 mA
+5 V <100 mA
±12 V 50 mA
Page 42
42 . Frontier IR Dual-Range Systems User's Guide
Icon Description Connector
Type
EXT.R
Right external detector module. Connects to an external detector module. Usually the detector is located in an accessory placed on the right of the instrument.
Microscope external detector module. Connects to a PerkinElmer infrared microscope, placed on the left of the instrument.
15 way
high density
D-type
15 way
high density
D-type
Voltages Maximum
Currents
+12 V 0.65 A
12 V 0.65 A
+5 V 4 A
+12 V 0.65 A
−12 V 0.65 A
+5 V 4 A

Connecting the spectrometer to the electrical supply

The power cable for the electrical supply plugs into the rear of the instrument. It has a molded socket at one end. If it is necessary to fit a plug on the power cable, use the wire color code below:
Plug Pin Wire Color (100120 V) Wire Color (220240 V)
Ground (Earth) Green or Green/Yellow Green/Yellow
Live Black Brown
Neutral White Blue
To ensure safe and satisfactory operation of the instrument, it is essential that the green or green/yellow ground (earth) wire of the power cord is connected to a ground that complies with the regulations
WARNING
of the local electricity supply authority (or equivalent body); ground circuit continuity is essential for safe operation of the equipment.
Page 43
Unpacking and Installation . 43
Electricity supply inlet
Frontier IR instruments can operate on electricity supplies of 50 or 60 Hz and in the 100 V to 230 V range without any adjustment.
1. Fit the molded socket of the power cable into the electrical supply inlet (Figure 14) of the instrument.
Figure 14 Electrical supply inlet
2. Switch on the power at the supply.
3. Switch on the instrument at the switch above the supply inlet. I is on and O is off.
After a short diagnostic check, which takes about 2 minutes and is described in
Instrument Self-Checks
on page 98, your instrument will be ready to communicate with the
PC.
Appendix 2:
Page 44
44 . Frontier IR Dual-Range Systems User's Guide

Installing the Instrument in the Software

Before you can use the instrument it must be set up in the software.

Installing the software

NOTE: If you are supplying your own computer, make sure that it meets the minimum
requirements for hardware and software set out in the “PC Requirements” section of the
Administrator’s Guide
Manuals CD
To install Spectrum software, insert the supplied DVD and follow the instructions on the
screen. Details of the installation program are given in the
(part number L1050002) supplied with your instrument.

The Instrument Install Wizard

After installing your software you will need to install your instrument.
If you have Spectrum software or Spectrum ES software, select Add Instrument from
the Administration group on the Setup menu.
The Instrument Install Wizard starts.
for your software, which can be found on the
Administrator’s Guide
IR & Raman
.
OR
If you have AssureID software, select Configure Instruments from the Configure Instruments and Accessories group on the Tools menu.
The Instrument Install Wizard starts.
Details of the Instrument Install Wizard can be found in the software, which can be found on the
IR & Raman Manuals CD
Administrator’s Guide
supplied with your instrument.
for your
Page 45
Using the Spectrometer
with Spectrum
Page 46
46 . Frontier IR Dual-Range Systems User's Guide

Basics of Software Control

Starting Spectrum software

1. Switch on the power to the instrument using the switch on the rear of the instrument. The instrument will initialize, which will take approximately 2 minutes.
NOTE: It can take the instrument about two hours to equilibrate when switched on after
being switched off overnight. To save time, we suggest that you leave the instrument switched on at all times.
2. From the Start menu select Programs; the PerkinElmer Applications group; the Spectrum sub-group and then the Spectrum application.
OR
Double-click the shortcut icon The Spectrum start-up splash-screen is displayed, followed by a dialog that may require
your login details:
3. If required, enter your User name and Password, and then click OK.
If you already have an instrument set up in Spectrum or AssureID software for this user on this PC, the Instrument Connection dialog may be displayed, unless Auto Connect is set up for the instrument. If Auto Connect is enabled Spectrum will automatically connect to the instrument.
4. Select the Instrument you want to use.
OR
on the desktop.
If you want to work with data that has been collected previously, without connecting to an instrument, select work offline.
NOTE: Working offline can free a networked instrument for use by another user.
Spectrum starts.
Page 47
Using the Spectrometer with Spectrum . 47

Scanning samples

When you connect to your spectrometer the instrument settings will default to appropriate values for your instrument type and accessory.
The configurable Scan toolbars at the top of the workspace include the tools you need to collect a sample spectrum (Figure 15).
Figure 15 Frontier IR Dual-Range systems default scan toolbars
By default, the Measurement bar includes Scan, Halt, Background and Monitor buttons. You can also select Scanalyze and then choose one of the scan and then process options Scan and Compare, Scan and Search or Scan and Quant. All these commands can be selected from the Measurement menu.
NOTE: The Scan toolbars are not displayed if you have chosen to work offline.
To scan a sample:
1. Check and set the instrument parameters. Here you set the Start and End points of the scan range (by default in wavenumbers,
but the abscissa units can also be set to nanometers or microns on the Setup Instrument Basic tab) and the Accumulations required, either as a number of scans, or as a length of time. You can enter a unique Sample ID and Description for the sample.
If required, you can also change the wavelength range of the spectrometer using the Mode drop-down list. See information on configuring your Frontier Dual-Range spectrometer for different wavelength ranges.
2. If a background scan is required, the Scan button includes a small background flag. Clear the instrument beampath, or insert a suitable background material, and then click
to collect a background spectrum.
The background spectrum is displayed briefly, and then the Viewing Area is prepared for data collection from your sample.
We recommended that you run a background scan before every sample.
3. Place your sample in the instrument or accessory and then click your sample.
By default, during scanning the sample data is displayed on the Live tab in the Viewing Area.
The completed spectrum is displayed on the Graph tab, and added to the current Samples View in the Data Explorer. The results of the Quality Checks, selected by default on the Setup Instrument Advanced tab, are displayed in the results for the spectrum. If you have the Auto save option selected on the Setup Instrument Data Collection tab, your spectrum will be saved automatically.
Changing the instrument range
on page 48 for more
to begin scanning
Page 48
48 . Frontier IR Dual-Range Systems User's Guide
4. If, for any reason, you want to stop scanning your sample, click .
You can use the Sample Shuttle Accessory (L1200302) to set up an interleaved cycle scan so that you do not have to open the sample compartment between the background and sample scans. This helps to reduce changes in the concentration of carbon dioxide and water vapor in the sample compartment. Set the Scan Type to Interleaved on the Setup Instrument Basic tab. See the Spectrum on-screen Help for more information.

Changing the instrument range

You can change the instrument range from, for example MIR to NIR, using Spectrum software. Changing ranges requires the instrument to change the detector, beamsplitter and, for the Frontier FT-IR/FT-NIR only, source in concert.
NOTE: Depending on the beamsplitters and detectors fitted, the instrument may require
some time to equilibrate after changing ranges.
1. Select the Setup menu and then click Instrument.
The Setup Instrument tabs are displayed.
2. Select the Setup Instrument BeamPath tab. The Mode field in the settings table, shown here as MIR, displays the current range.
NOTE: The Mode option is also available in the Instrument Settings toolbar.
3. Select the range you wish to change to from the drop-down list. The options are MIR, NIR or Custom for the Frontier FT-IR/FT-NIR spectrometer, or
MIR, FIR or Custom for the Frontier FT-IR/FT-FIR spectrometer. If you select MIR, NIR or FIR, the Source, Beamsplitter, Detector, Window and
Optimum Scan Range fields will be updated. If you select Custom, the Source, Beamsplitter and Detector options can be selected independently. The Optimum Scan Range and Window fields will update for the combination you have selected.
NOTE: Some combinations may damage the instrument, and will not be allowed.
4. Click the Apply button or click Scan to change the range.
If you have a Frontier FT-IR/FT-FIR spectrometer, you will be prompted to change the windows. Change the windows (as described in
Sample Compartment Windows
While the instrument is changing range, you will not be able to collect data. The range change is monitored by messages in the Status Bar.
on page 61) and then click Scan.
Replacing the Frontier FT-IR/FT-FIR
Page 49
Using the Spectrometer with Spectrum . 49
Display
Go button

Working with the instrument display and Go button

Figure 16 Frontier IR top panel controls
The display on the front right of the top of your instrument is used to display prompts and other messages generated by Spectrum software.
Simply follow the prompts in the instrument display and use the Go button located on the instrument to complete a series of sample scans.

Changing the beampath

Hover over the icons on the Setup Instrument BeamPath tab to identify elements in the instrument beampath. As you click on each item, the appropriate row in the Settings Table is highlighted.
The accessory fitted in the sample compartment is identified by a graphic element in the beampath schematic. If an external accessory is fitted at the beam port on the right of the instrument, or a Spotlight Imaging System is fitted, the Setup Instrument Beampath tab includes additional elements that enable you to redirect the beampath to the accessory you want to use.
In Figure 17 the beampath is directed to the standard sample slide. In this example, the additional interface elements for the External NIRA (including its filter wheel), and the Spotlight Imager, are also shown.
Figure 17 Setup Instrument BeamPath tab with the beampath directed to the Slide Holder
Page 50
50 . Frontier IR Dual-Range Systems User's Guide
To direct the beam to the External NIRA, select Right external from the Beam
Location drop-down list.
Figure 18 Setup Instrument BeamPath tab with the beampath directed to the External NIRA (circled)

Using the Spectrum on-screen Help system

Use the Spectrum Help system to find further information about using Spectrum software to control, set up and adjust your instrument.
To open the Help file, select Contents from the Help menu. This menu also includes links to on-screen tutorials (Tutorials), and information about the software (About).
Page 51
Using the Spectrometer with Spectrum . 51

Atmospheric (CO2/H2O) Suppression

Atmospheric suppression can be selected in the Setup Instrument Advanced tab (Figure 19).
Figure 19 Setup Instrument Advanced tab in Spectrum

What is atmospheric suppression?

This is an atmospheric correction routine. This routine is more powerful than simple subtraction, overcoming the following issues:
Non-linearity due to resolution
The measured spectrum is temperature dependent
Lineshape and calibration are affected by J-stop and sample or accessory.

What does atmospheric suppression do?

When CO2/H2O is switched on, the software uses a single reference spectrum derived from high resolution data and our understanding of the instrument to model the lineshape, then finds the current real instrument parameters by least squares fitting to the measured spectrum.
Figure 20 Correction of an MIR spectrum at 4 cm
−1
resolution
Page 52
52 . Frontier IR Dual-Range Systems User's Guide
Figure 21 Atmospheric suppression
Page 53
Using the Spectrometer with Spectrum . 53

AVI Correction

AVI correction can be selected in the Setup Instrument Advanced tab in Spectrum (Figure 19).
NOTE: AVI correction is not available when in the FIR range.
AVI correction can only be performed if an AVI Calibration has been set up for the current configuration and resolution. Select AVI from the Adjustments Toolbox, available from the Setup Instrument Advanced Tab. Follow the instructions on-screen. For further information, refer to the Spectrum on-screen Help file.
NOTE: AVI correction uses a methane gas cell installed in the filter wheel.

What is AVI correction?

The objectives of Absolute Virtual Instrument (AVI) correction are:
Consistent performance over time and between instruments
Traceability for all measurements.
Although FT-IR spectrometers use a reference laser, the wavenumber calibration and lineshape are affected by differences in beam divergence and uniformity. This is true for all FT-IR spectrometers. Differences can occur between instruments, when using different sampling accessories and when components are changed. AVI allows calibration and lineshape to be maintained.
The Absolute Virtual Instrument is an instrument with theoretical performance, such that the result of measuring a known sample on such an instrument can be predicted. So, if we measure with a real instrument and calculate the software transform to match the theoretical result, we can apply this transform to future measurements.
The Absolute Virtual Instrument is defined by wavenumber calibration, instrument lineshape and ordinate accuracy.
Page 54
54 . Frontier IR Dual-Range Systems User's Guide

What does AVI correction do?

When AVI is switched on, the software measures the current instrument profile relative to an absolute standard (the methane cell) and an ideal lineshape function, and applies a correction (Figure 22). If you have an automated, internal filter wheel, this can provide correction for any sampling configuration.
−1
Figure 22 Spectra of methane at 4 cm AVI (bottom)
resolution as measured (top) and with
Page 55
Using the Spectrometer with Spectrum . 55

Look Ahead

Look Ahead can be selected in the Setup Instrument Advanced tab (Figure 19).

What is Look Ahead?

Look Ahead is a novel system where the spectrometer scans continuously and uses the properties of the measured spectrum to identify changes corresponding to sample removal, sample insertion, or sample change, automatically.
This information is used to identify the scans that are being collected, and to accumulate sample scans. This can decrease the overall scanning time required for those samples that require scan accumulation.

What does Look Ahead do?

If Look Ahead has been enabled, the spectrometer scans continuously. When you request a sample scan, the scan collected is compared to the scans that have already been collected. If identical scans are found, the number of scans requested is decreased by the number of scans found, so decreasing the overall scanning time.
Page 56
56 . Frontier IR Dual-Range Systems User's Guide

Quality Checks

Quality Checks can be selected in the Setup Instrument Advanced tab (Figure 19).

What are Quality Checks?

Quality checking is a tool for less experienced IR users that identifies possible problems in the spectrum being collected and suggests ways of improving the measurement.

What do Quality Checks do?

Simply select the Quality Checks that you want to perform from the list in the Setup Instrument Advanced tab and, if required, adjust the threshold range using the slider bars. To display more information about an individual test, double-click on its name in the list.
When you collect your spectrum, the selected tests are performed as the data is collected and a signal light (green
The results from the Quality Checks imply:
, amber , or red ) indicates the result.
Okay (
Caution (
order to improve the quality of the spectra you are collecting.
Warning ( cure before collecting further spectra.
) – the quality of the spectra is satisfactory.
) – the software has identified a problem that you may want to investigate in
) – the software has identified a serious problem that you should attempt to
Page 57

Routine Maintenance

Page 58
58 . Frontier IR Dual-Range Systems User's Guide

Cleaning the Spectrometer

Clean the outside of the instrument using a damp cloth. If necessary, a mild detergent may be used. Before you clean the entire instrument, always perform a patch test on an inconspicuous area.
Avoid spilling liquid into the instrument. Clean all external spills immediately. If anything that is spilled enters the main body of the spectrometer, switch off the power and contact a PerkinElmer Service Engineer.
Do not touch or attempt to clean any optical surface in the instrument,
CAUTION

Cleaning the display

Clean grease and dirt off the display using a soft damp cloth and a mild detergent.
because this will impair its performance and may easily damage the component.
Page 59

Moving the Spectrometer

Lifting handhold,
The spectrometer is a heavy precision instrument, so two people are required for safe handling.
Consult the local codes of practice issued by safety advisors before
WARNING
WARNING
attempting to lift it.
Take care not to injure yourself or others, or to drop the spectrometer.
Before moving the spectrometer, switch off the power supply, wait 60 seconds, and disconnect the power cable.
Routine Maintenance . 59
Do not attempt to lift the instrument, or to move it to another location, when an external accessory is attached. The optical alignment of the accessory may be disturbed.
CAUTION
The accessory is fitted to an external accessory bracket mounted within the handhold on the right of the instrument (Figure 23
). The two screws that attach the external accessory to this bracket are accessible from the right side of the accessory.
The spectrometer can be lifted using the shaped handholds on its sides, as shown in Figure 23. Two people are needed to lift it because its basic weight is approximately 36 kg.
(repeated on left of the instrument)
Figure 23 Lifting a Frontier IR spectrometer
Page 60
60 . Frontier IR Dual-Range Systems User's Guide

Condensation

Be aware that condensation caused by moving your spectrometer from a cooler environment to a warmer one can damage the windows of the sample compartment. To prevent this damage from occurring:
Make sure that the windows are protected by placing fresh bags of desiccant in the sample compartment.
Leave the spectrometer to reach the temperature of the surroundings before removing the desiccant.

The Set Temperature utility

A Frontier IR instrument is able to thermostat (that is, control the temperature of) its internal optical bench. By default, this capability is enabled by the PerkinElmer Service Engineer when the instrument is installed.
Usually, there is no need for adjustment. However, if the temperature range in your laboratory changes, or if you move the instrument to another location, it may be necessary to set a different temperature range to ensure optimal performance.
1. Browse to the C:\Program Files\PerkinElmer\ServiceIR or C:\Program Files (x86)\PerkinElmer\ServiceIR directory, and then double-click the Set Temperature shortcut.
The PerkinElmer Login dialog is displayed.
The ServiceIR folder contains a number of other utilities, many of which
CAUTION
are intended solely for use by a PerkinElmer Service Engineer. If used incorrectly, some utilities could damage the instrument.
2. Enter your login details. You must be a Spectrum Administrator. The Select Instrument dialog is displayed.
3. Select your instrument and then click OK.
The Set Temperature dialog is displayed.
4. Select the setting most appropriate for the range of temperature in your laboratory, and then click Apply.
The Disabled setting is not recommended.
The temperature range setting is stored in the instrument, and the utility closes.
NOTE: This setting applies until the Set Temperature Utility is used again. It is not reset
when the instrument power is recycled (that is, switched OFF and then ON).
Page 61
Routine Maintenance . 61
Seal
Window

Replacing the Frontier FT-IR/FT-FIR Sample Compartment Windows

NOTE: You can also change the sample compartment windows in the Frontier FT-IR/FT-NIR
spectrometer. For more information, see
Compartment Windows
The sample compartment windows in the Frontier FT-IR/FT-FIR spectrometer need to be changed when moving between the MIR and FIR ranges.
KBr windows are used in the MIR range. In the FIR range, the PE window is installed in the left side of the sample compartment and the PE/D window is installed in the right side of the sample compartment.
1. Open the sample compartment cover.
2. Using the 3 mm screwdriver (part number L9003317) supplied with the Quick Change Window Kit (part number L1250432), slacken the three screws securing the current window to the side of the sample compartment.
3. Remove the current window with its seal in place. Store the window in the desiccated spare window compartment (Figure 9).
on page 85.
Installing/Replacing Beam Port and Sample
Be careful not to get moisture on the optical surface of the window. Do
CAUTION
4. Ensure the seal is fitted to the replacement window and is correctly seated (Figure 24).
Figure 24 Seal fitted to window
not touch or breathe on the optical surface of the window.
Page 62
62 . Frontier IR Dual-Range Systems User's Guide
Screw
Notch and key
Notch and
Sample
wall
Sample compartment wall
Compartment Spare window
Screw
Screw
5. Fit the window to the sample compartment. Ensure the seal is fully seated and the key on the window lines up with the notch in the
main cover (Figure 25).
compartment
key
Figure 25 Window installed in sample compartment
6. Secure the window to the sample compartment wall by tightening the screws using the 3 mm screwdriver (Figure 25).
NOTE: Do not over tighten the screws.
7. Repeat steps 2–6 for the other sample compartment window.
8. Close the sample compartment cover.
To store the windows:
1. Open the spare window compartment cover (Figure 26).
desiccant indicator
Figure 26 Windows situated in the spare window compartment
Page 63
Routine Maintenance . 63
2. Place each window in the holder, with the seal in place, and tighten the screw using the screwdriver provided.
NOTE: Do not over tighten the screw.
3. Close the spare window compartment cover.
If you have an MCT detector for the MIR range, you will be supplied with an Attenuator kit (part number L1160560) containing 1%T, 4%T, 6%T, 14%T and 32%T attenuators. The attenuators attach to the KBr window magnetically.
Before you change to the MIR MCT detector in Spectrum software, fit one of the attenuators supplied to the window to the right of the sample compartment to prevent the detector from overloading (Figure 27). If you have an open beam, a 14%T attenuator is recommended.
Figure 27 KBr window with 14%T attenuator
NOTE: If you have a Frontier FT-IR/FT-FIR spectrometer with an MCT detector and the
instrument was in the MIR range when it was switched off, it is recommended that a 6%T attenuator is fitted to the sample compartment window before switching on the instrument again, to prevent errors during the initialization of the FIR detector.
Page 64
64 . Frontier IR Dual-Range Systems User's Guide
must
Desiccant change is recommended.
Pink
Blue
Blue

The Desiccant Indicator in Detail

The optical system of the spectrometer is purged at the factory. This protects the sample compartment windows, external port windows, or beamsplitters from being damaged by humidity. Replaceable packs of desiccant maintain the purge.
The top panel of the instrument includes a desiccant indicator (Figure 28), whose sectors change sequentially from blue to pink as the desiccant becomes exhausted. Change the desiccant packs in the instrument when the sector marked 10 becomes pink, but while the sectors marked 15 and 20 are still blue. These numbers correspond to the approximate % Relative Humidity in the instrument.
NOTE: A similar desiccant indicator is fitted in the desiccated compartment (Figure 34 on
page 68). This is used for storing sample compartment windows when not in use.
Figure 28 The desiccant indicator: when desiccant packs should be changed
If all three sectors of the desiccant indicator are pink, then you change the desiccant.
CAUTION
The instrument optics may be fogged. Do not switch the instrument either ON or OFF until all sectors are blue.
Sector Action Required
All sectors blue None.
The humidity levels in the instrument are below approximately 10% Relative Humidity.
Sector 10 pink
If the instrument has been switched OFF for an extended period, do not switch ON until you have changed the desiccant and all sectors are blue.
Sector 10 & 15 pink
Replace desiccant the immediately to avoid instrument damage.
Do not switch the instrument ON or OFF until you have changed the desiccant and all sectors are blue.
Page 65

Changing the Desiccant

Routine Maintenance . 65
Expect to change the desiccant
in the
spectrometer every six months.
Old, used desiccant releases moisture.
CAUTION
In regions experiencing high humidity levels we recommend that you change the desiccant more often.
The desiccant change interval is set in the Desiccant change due in (days) option on the Setup Instrument BeamPath tab in Spectrum software (Figure 29).
Figure 29 Setup Instrument BeamPath tab in Spectrum
If an instrument desiccant change is overdue, a warning will be displayed in the Status Bar and in the Setup Instrument BeamPath table.
The warning is displayed until you tell the software that the desiccant has been changed. To reset the desiccant change interval, select the current value, click Changed to clear any warning messages, and then enter the number of days before the next desiccant change is due.
Page 66
66 . Frontier IR Dual-Range Systems User's Guide
Screw
Screw
Sample cover

Renewing the instrument desiccant

Old, used desiccant releases moisture and can cause catastrophic failure
CAUTION
You can purchase disposable desiccant kits from PerkinElmer (part number N0171159). A kit contains two packs of desiccant, and three kits are required.
1. Inspect the humidity indicator card in the plastic bags in which the spare desiccant packs are packed. If the card indicates humidity in the bag, discard the desiccant pack.
2. Remove the sample area cover, if fitted, by opening the cover, pressing the clip and pulling the cover vertically to remove it (Figure 30).
of KBr optics.
Do not use damaged packs of desiccant. Make sure that the packs you use have not been left in contact with the air.
release clip
Figure 30 Removing the sample compartment cover
3. Undo the two captive screws securing the desiccant cover (Figure 31).
Figure 31 Captive screws securing the desiccant cover
Page 67
Routine Maintenance . 67
Desiccant
4. Open the cover and remove all the exhausted desiccant packs (Figure 32), noting how they are installed.
compartment (empty)
Figure 32 Desiccant removed
5. Place the first three packs of desiccant upright in the recess on the right of the desiccant holder then, one at a time, lay the three remaining packs flat in the upper part of the holder (Figure 33).
Figure 33 Stowing the desiccant packs
Ensure that when the packs have been installed they do not rise above the level of the black rubber purge seal.
6. Close the cover and tighten the screws.
7. Refit the sample area cover.

Installing rechargeable desiccant in the instrument

There is also a rechargeable desiccant kit available (part number L1250311).
To install rechargeable desiccant packs follow the procedure in
desiccant
The instrument requires two rechargeable desiccant packs; the packs are sized to fit into the left (70 mm × 78 mm × 33 mm) or right (102 mm × 78 mm × 84 mm) of the desiccant container.
The standard kit contains two pairs of rechargeable desiccant packs. This is so you can immediately replace the desiccant in the instrument with re-activated packs.
, starting on page 65.
Renewing the instrument
The desiccant packs can be re-activated by baking them in an oven at 250 °C for approximately 8 hours. They should be cooled in a dry atmosphere. For optimum results, re-activate the packs immediately before re-use.
Page 68
68 . Frontier IR Dual-Range Systems User's Guide
Thumb screw
Thumb screw

Renewing the desiccant in the spare window compartment

The windows in the sample compartment and the external ports can be exchanged for windows made from another material.
To avoid serious damage, spare windows must be kept in a dry atmosphere, such as in the desiccated compartment on top of the instrument.
CAUTION
Installing disposable desiccant in the window compartment
The desiccated compartment contains two disposable desiccant packs. You can purchase disposable desiccant kits from PerkinElmer (part number N0171159). A kit contains two packs of desiccant, and one kit is required.
1. Inspect the humidity indicator card in the plastic bags in which the spare desiccant packs are packed. If the card is pink (indicating humidity in the bag) discard the desiccant packs.
If more than one sector on the desiccant indicator inside the compartment is pink, change the desiccant.
2. Open the desiccated compartment, and then remove the two thumb screws securing the desiccant pack cover (Figure 34).
Figure 34 Desiccated compartment for spare windows
3. Lift off the desiccant pack cover and remove the desiccant packs noting how they were installed.
4. Install two new disposable desiccant packs, or the rechargeable desiccant pack, and then refit the desiccant pack cover.
5. Close the compartment. The desiccant indicator should be completely blue, indicating satisfactory storage
conditions, within approximately half an hour.
Page 69
Routine Maintenance . 69
Installing rechargeable desiccant in the window compartment
There is also a rechargeable desiccant kit available (part number L1251840).
To install rechargeable desiccant packs refer to the procedure for
desiccant in the window compartment
The desiccated compartment requires a single rechargeable desiccant pack that is sized to fit the desiccant container.
The rechargeable desiccant kit contains two rechargeable desiccant packs. This is so you can immediately replace the desiccant in the instrument with a re-activated pack.
Re-activate a desiccant pack by baking it in an oven at 250 °C for approximately eight hours. It should be cooled in a dry atmosphere. For optimum results, re-activate the pack immediately before re-use.
on page 68.
Installing disposable
Page 70
70 . Frontier IR Dual-Range Systems User's Guide

Purging the Spectrometer

Under most circumstances, you do not need to purge the optical system. However, after performing any maintenance tasks that involved opening the main cover, you may purge the optical system to remove water vapor and CO
that entered while the system was open.
2
Purge the sample compartment if you need to make sure that no residual water vapor or CO peaks are visible in the spectrum, particularly when operating in the FIR range.
There are two separate sets of connectors for purge gas lines, one for the sample compartment and one for the optical system. You can purge with either dry air or nitrogen. Both remove water vapor; however, nitrogen is preferable because it also removes atmospheric carbon dioxide.
3
A typical cylinder of dry nitrogen (or dry air) stores 6.26 m 2200 lbf/in
2
). Make sure that the gas is free of oil, water, or dirt particles larger than 25 µm
at 1.4 × 104 kPa (220 ft3 at
(0.001 inch).
Do not site the instrument in a poorly ventilated area if nitrogen will be used as a purge gas.
WARNING
Oxygen depletion in an enclosed space does not trigger a gasping reflex, and errors of judgment, confusion, or unconsciousness can occur in seconds and without warning.
Do not use a flammable gas to purge the instrument. The spectrometer contains a hot source, and a fire or explosion will result. Only use clean, dry, oil-free nitrogen or air to purge the instrument.
WARNING
2
Never connect the purge tubing directly to a gas cylinder or other high pressure supply; always use a pressure regulator and set the pressure to a maximum of 6.9 kPa (1 lbf/in
2
) before you start the flow.
WARNING
A length of clear plastic tubing (4 mm internal diameter) will be required to transfer the gas from the gas bottle to the instrument purge connectors.
Page 71
Routine Maintenance . 71
Instrument purge
Sample purge
Beveled end of connector
Knurled nut
Tubing
1. The tubing is connected to the instrument by a universal pipe fitting which is secured to the tubing with a nut. Two sets of fittings are supplied with the instrument. Assemble the fittings to the tubing as illustrated in Figure 35.
Figure 35 Purge fittings
2. Tighten the nut, ensuring it is not over tightened.
3. Push the connector on the end of the tubing into the purge connector of the compartment to be purged (Figure 36).
If the connector will not push in, press the clip at the bottom of the port and try again.
Figure 36 Connector fitted to purge port
4. Connect the free end of the tubing to the gas regulator.
NOTE: Ensure the regulator is set to a pressure not greater than 6.9 kPa (1 lbf/in2).
5. Open the gas flow valve on the regulator until a flow rate of 28 l/minute (1 ft3/minute) is achieved.
The instrument compartment is purged in 10 minutes. The sample compartment is purged in 2 minutes.
6. Once purging is complete turn off the gas supply.
Page 72
72 . Frontier IR Dual-Range Systems User's Guide
Locking clip
7. Disconnect the fitting from the instrument by pushing the clip located at the bottom of the purging port (Figure 37).
Figure 37 Disconnect the locking clip
Page 73
Routine Maintenance . 73
Fuse drawer

Changing the External Fuse

Very occasionally, a fuse may fail with age and will need to be changed.
NOTE: If fuses fail repeatedly, there is an electrical fault: disconnect the power supply and
contact your PerkinElmer Service Engineer.
There is one spare fuse in the fuse drawer. If you need more, order 2 A, 250 V Time Lag fuses (part number 04970839) from PerkinElmer. You must only replace the fuse with one of this type and rating. Do not use makeshift fuses and do not short-circuit fuse holders.
Before you begin this procedure, switch off the power to the spectrometer, wait for 60 seconds, and disconnect the power cable.
WARNING
The external fuse drawer is located at the rear of the instrument between the power switch and the power socket (Figure 38).
This will protect you from electrical shock.
Figure 38 Fuse drawer closed
Page 74
74 . Frontier IR Dual-Range Systems User's Guide
Spare fuse
Fuse
Fuse drawer
1. Lever out the fuse drawer so that it swings down over the power socket.
The fuse is in the right hand slot.
2. Remove the old fuse and discard it.
3. Fit the new fuse into the slot (Figure 39).
Figure 39 Fuse fitted; drawer open
4. Close the drawer. The new fuse is now installed.
5. Reconnect the power cable and switch on the instrument.
Page 75
Routine Maintenance . 75

Cooling the MCT Detector (If Fitted)

The instrument may be fitted with an MCT (mercury cadmium telluride) detector. When it is fitted in the instrument it must be cooled to 77 K before you collect spectra. The detector is mounted in a dewar that can be filled with liquid nitrogen. The detector is not enabled until it has been sufficiently cooled.
Do not site the instrument in a poorly ventilated area if liquid nitrogen will be used to cool a detector.
WARNING
WARNING
Oxygen depletion in an enclosed space does not trigger a gasping reflex, and errors of judgment, confusion, or unconsciousness can occur in seconds and without warning.
The extremely low temperature of liquid nitrogen can burn skin and eyes. Avoid exposure by wearing protective gloves and safety goggles whenever you work with it.
WARNING
WARNING
As liquid nitrogen cools the detector it evaporates so rapidly that the resulting pressures can send a funnel or detector cap suddenly and forcefully shooting upward from the top of the dewar.
Be sure to wait the specified time when filling the funnel and before replacing the plug. This enables the bubbling nitrogen to settle down and the pressure to dissipate. In addition to wearing safety goggles at all times, stand back from the instrument after each time you fill the funnel.
Page 76
76 . Frontier IR Dual-Range Systems User's Guide
MCT coolant cover
1. Remove the MCT coolant cover (Figure 40).
Figure 40 Location of MCT coolant cover
2. Remove the plug from the dewar.
3. Ensure that the funnel guide, which helps prevent overspill entering the instrument body rather than the dewar, is in place.
4. Make sure the supplied plastic funnel is dry.
Stand where you can see the inside of the funnel as you pour the
CAUTION
nitrogen in. Pour slowly, so that neither the funnel nor the dewar overflows. Take care not to overfill the funnel and splash liquid nitrogen onto the instrument covers.
5. Insert the plastic funnel into the funnel guide and slowly fill the detector dewar with liquid nitrogen.
NOTE: Make sure that the detector dewar is filled to the top with liquid nitrogen before
selecting the detector. Prevent radiation from reaching the detector by obstructing the beam path in the sample area during filling.
6. Stand back and wait two minutes. This nitrogen also vaporizes as the dewar continues to cool. The two-minute wait allows
the bubbling to settle down and the pressure of the vaporizing nitrogen to dissipate.
7. Continue to pour liquid nitrogen into the funnel, adding a little more each time the funnel empties.
The funnel takes longer to empty as the dewar fills. This happens after two or three more funnels of liquid nitrogen. Because the dewar has now cooled, the liquid nitrogen does not vaporize, but instead fills the dewar.
8. Remove the funnel and wait two minutes.
The liquid nitrogen settles down and bubbling slows.
Page 77
Routine Maintenance . 77
9. When the nitrogen stops bubbling, refit the plug. When the detector dewar has been filled, it will remain at its operating temperature for
approximately eight hours.
NOTE: Cooled detector dewars require pumping down after approximately 12 months
operation. When the boil-off rate of liquid nitrogen becomes excessive (that is, the liquid nitrogen level is low after 3–4 hours of operation), consult your PerkinElmer Service Engineer.
Page 78
78 . Frontier IR Dual-Range Systems User's Guide
Page 79

Advanced Maintenance

Page 80
80 . Frontier IR Dual-Range Systems User's Guide
Locking bolt
Locking bolt

Opening the Main Cover

To perform most maintenance tasks, you have to open the main cover of the spectrometer.
Switch off the mains power supply to the spectrometer, wait 60 seconds and disconnect the power cable before you open the cover of the spectrometer. This makes sure that you are safe from electrical shock
WARNING
CAUTION
The main cover is hinged at the back of the instrument and opens from the front. Figure 41 identifies the locking bolts, which are on the front of the instrument.
and laser radiation.
Do not open the main cover if the humidity is >80% relative humidity for an instrument fitted with KBr optics.
The humidity sensitive optics will be damaged if subjected to a relative humidity in excess of the specified level.
Figure 41 Main cover locking bolts
1. Switch off the power supply to the spectrometer.
2. Wait for 60 seconds.
3. Disconnect the power cable.
Page 81
Advanced Maintenance . 81
Sample cover release clip
Sampling accessory
4. Remove the sample compartment cover, if fitted; open the cover, press on the cover
retaining clip and then pull the cover vertically to remove it (Figure 42).
Figure 42 Removing the sample compartment cover
5. Remove the fitted sampling accessory, by pulling the release handle towards you and
sliding the accessory towards you (Figure 43).
release handle
Figure 43 Removing the sampling accessory
6. Undo the two locking bolts on the front of the instrument using the hexagonal wrench
supplied; insert the wrench, so that the free end is pointing horizontally away from the instrument and then rotate the wrench 180° in the direction indicated by the arrows in Figure 44.
Figure 44 Undoing the locking bolts
7. Lift the main cover from the front.
Page 82
82 . Frontier IR Dual-Range Systems User's Guide
MIR source
Laser
Laser power supply
Beamsplitters; within
carrier
MIR detector
NIR source
Power supply
NIR detector
MIR source
NIR source
Laser power supply
Power supply
Laser
NIR detector
MIR
Beamsplitters; within
carrier
8. Ensure the cover is fully open and retained by the stay.
9. Perform the necessary maintenance inside the instrument. Figure 45 illustrates the primary component parts of a Frontier IR Dual-Range
spectrometer.
interferometer and
interferometer and
detector
Figure 45 Frontier IR FT-IR/NIR components
NOTE: The Frontier FT-IR/FT-FIR spectrometer has one source only, in the MIR source
position. The FIR detector is in the NIR detector position.
NOTE: If you have an MCT detector fitted for the MIR range, the MCT detector is fitted
toward the rear of the instrument and the NIR/FIR detector is fitted at the front of the instrument.
10. Close the cover.
11. Lock the two locking bolts by inserting the wrench into each bolt in turn and rotating it opposite to the direction indicated by the arrows in Figure 44.
12. If required, reinstall the sampling accessory by sliding it on the ledge and pushing it into the sample compartment until it engages with the connector.
13. If required, refit the sample cover by inserting the hinge into the grooves and pushing the cover downwards until the clip engages.
14. Reconnect the power cable and switch the power back on.
15. If required change the desiccant (see purge the instrument (see
Purging the
Changing the Desiccant
Spectrometer on page 70).
on page 65), and/or
If the cover is left open for more than 1 hour, the desiccant should be replaced.
Page 83
Advanced Maintenance . 83
NIR source
MIR source
NIR source
MIR source

Replacing a Source

Switch off the mains power supply to the spectrometer, wait 60 seconds and disconnect the power cable before you open the cover of the spectrometer. This makes sure that you are safe from electrical shock
WARNING
WARNING
1. Open the main cover of the instrument as detailed on page 80 and identify the source (Figure 46).
and laser radiation.
Due to its high operating temperature allow sufficient time for the source to cool before attempting to remove it.
Figure 46 Source location for Frontier FT-IR/FT-NIR
NOTE: The Frontier FT-IR/FT-FIR spectrometer has one source only. It is located in the MIR
source position.
Caution, hot surface.
The source housing and source itself will be hot if recently switched on. Allow to cool before performing the following procedure.
Page 84
84 . Frontier IR Dual-Range Systems User's Guide
NIR source
MIR source
MIR source terminal
MIR source terminal
NIR source terminal
NIR source terminal
Source terminal
Source terminal
Source
The source element is brittle and can be broken if not handled properly. Be careful when removing and installing the source.
CAUTION
Do not touch the glass bulb as this will degrade the performance of the bulb and shorten its life.
2. Slacken the two terminal screws (Figure 47 or Figure 48) securing the wires leading to the source. Disconnect the wires from the connectors.
For the Frontier FT-IR/FT-NIR spectrometer, the connecting wires for the MIR source are white, and for the NIR source are red.
For the Frontier FT-IR/FT-FIR spectrometer, the connecting wires for the source are white.
3. By grasping the round metal source body, NOT the wires, remove the source from its housing by pulling, noting its installed position.
4. Install the new source in its housing, by pushing in until it clicks into place.
5. Reconnect the two wires to the terminals and secure by tightening the two terminal screws (Figure 47 or Figure 48).
Figure 47 Source terminal screws (from instrument left) for Frontier FT-IR/FT-NIR
Figure 48 Source terminal screws (from instrument left) for Frontier FT-IR/FT-FIR
6. Close and secure the main cover. Refer to steps 11 to 15 on page 82.
7. Allow the source temperature to equilibrate for approximately 2 hours before using the instrument.
Page 85
Advanced Maintenance . 85
Seal
Window

Installing/Replacing Beam Port and Sample Compartment Windows

The following procedure describes how to install a window in one of the external beam ports. It also applies to installing windows in the sample compartment of the Frontier FT-IR/FT-NIR spectrometer.
NOTE: For information about changing the Frontier FT-IR/FT-FIR sample compartment
windows, see page 61.
WARNING
1. Open the main cover of the instrument, as detailed on page 80.
Replacing the Frontier FT-IR/FT-FIR Sample Compartment Windows
on
Switch off the mains power supply to the spectrometer, wait 60 seconds, and disconnect the power cable before you open the cover of the spectrometer. This makes sure that you are safe from electrical shock and laser radiation.
Using the hexagonal wrench supplied with the new window kit, slacken and
remove the screws and washers securing the beam port cover or current window to the main cover.
Remove the screws carefully, and ensure that the washers do not fall into the instrument.
Remove the beam port cover or current window and seal. Retain for further use.
Be careful not to get moisture on the optical surface of the window. Do
CAUTION
2. Ensure the seal is fitted to the window and is correctly seated (Figure 49).
not touch or breathe on the optical surface of the window.
Figure 49 Seal fitted to window
Page 86
86 . Frontier IR Dual-Range Systems User's Guide
Main cover
Screw and washers
Screw and washers
Screw and washers
Screw and washers
Main cover
Notch and key
Notch and key
3. Fit the window to the main cover from the outside in. Ensure the seal is fully seated and the key on the window lines up with the notch in the
main cover.
Figure 50 Window installed in main cover
4. Secure the window to the main cover by fitting the screws and washers and tightening the screws using the hexagonal wrench (Figure 50).
5. Close and secure the main cover. Refer to steps 11 to 15 on page 82.
If you have an MCT detector for the MIR range, you will be supplied with an Attenuator kit (part number L1160560) containing 1%T, 4%T, 6%T, 14%T and 32%T attenuators. The attenuators attach to the window magnetically. You will require the ferritic window surround (part number C1201453) to attach the attenuators to the new windows.
Before you change to the MIR MCT detector in Spectrum software, fit one of the attenuators supplied to the window to the right of the sample compartment to prevent the detector from overloading. If you have an open beam, we recommend a 14%T attenuator.
Page 87
Advanced Maintenance . 87
Filter wheel
Filter wheel

Installing Filters in the Filter Wheel

The following procedure describes how to install a filter in the filter wheel. The same procedure should be followed when replacing a filter or the methane cell.
The filter wheel assembly is removed from its mounting and placed on the standard sample slide baseplate. It therefore may be necessary to remove the sample slide.
Switch off the mains power supply to the spectrometer, wait 60 seconds, and disconnect the power cable before you open the cover of the spectrometer. This makes sure that you are safe from electrical shock
WARNING
1. Set the Filter Wheel to Position 1 on the Setup Instrument BeamPath tab in Spectrum software.
Position 1 will be labelled as None 15000 – 0 cm drop-down list.
2. Open the main cover of the instrument as detailed on page 80 and identify the filter wheel (Figure 51).
and laser radiation.
-1
and be the first option in the
Figure 51 Filter wheel location
Page 88
88 . Frontier IR Dual-Range Systems User's Guide
Screw
Knurled screw
3. Install the standard sample holder baseplate, but remove the sample holder itself by unscrewing the knurled screw (Figure 52).
Figure 52 Knurled screw
4. To remove the filter wheel, loosen the set screw that secures it to its shaft using a
1.5 mm Allen key (Figure 53) and then gently ease the filter wheel off the shaft, rotating the filter wheel as required to clear the filter wheel from the other instrument components.
Take care not to remove the spacer ring from the spindle, as this may be lost.
NOTE: If the set screw is not accessible, you may not have moved the filter wheel to
Position 1 as described in Step 1.
Figure 53 Filter wheel securing screw
Avoid touching the filters.
CAUTION
Some of the filters are quite difficult to see.
Page 89
Advanced Maintenance . 89
Foam spacer
Filter holder
Filter
Opto flag
Position 7: for custom filter
Position 6: NIR polystyrene
Position 5: for custom filter
Position 4: Methane cell
Position 3:
Position 2:
Position 1: None Do not place filter here
5. Place the assembly face down on the sample holder baseplate.
6. Working clockwise around the filter wheel, identify the next available filter position (Figure 54).
Position 5 and Position 7 are available for your custom filters. There is no filter in Position 6 in the Frontier FT-IR/FT-FIR spectrometer.
NOTE: Do not change the positions of the standard filters, or place a filter in Position 1.
NG11 glass
MIR polystyrene
Figure 54 Filter wheel positions
7. Release the filter holder by turning it about 30 degrees counter-clockwise, and then remove the filter holder and foam spacer.
8. Place the filter in the hole, place the foam spacer on top of it, and then fit the filter holder.
Figure 55 illustrates the assembly sequence. Align the notches of the holder with the indents in the filter wheel. You must twist the
holder clockwise to lock it to the filter wheel.
Figure 55 Filter, foam spacer and filter holder
When fitted correctly, the small hole in the filter holder is close to the outer edge of the filter wheel.
Page 90
90 . Frontier IR Dual-Range Systems User's Guide
9. Refit the filter wheel to its spindle and secure it with the set screw.
Make sure that you tighten the set screw onto the flat of the shaft.
CAUTION
If the set screw is tightened onto a round portion of the shaft the filter wheel may loosen.
10. Refit the sample holder onto the baseplate with the knurled screw.
11. Close and secure the main cover. Refer to steps 11 to 15 on page 82.
12. Select the filter you would like to use in the beam path from the Filter Wheel drop-down list on the Setup Instrument BeamPath tab in Spectrum software.
NOTE: If you wish to run a background scan with a filter in place, you must set the Scan
Type to Background on the Setup Instrument Basic tab, rather than using the
Background toolbar button . See the Spectrum on-screen Help for more
information.
Page 91
Advanced Maintenance . 91
Laser
Power supply
Laser
Power supply
Replacing the Laser and Power Supply
Switch off the main power supply to the spectrometer, wait 60 seconds and disconnect the power cable before you open the cover of the spectrometer. This makes sure that you are safe from electrical shock
WARNING
and laser radiation.
Do not attempt to separate the Laser from the power supply. The Laser retains an electrical charge even when the main power is switched off.
WARNING
1. Open the main cover of the instrument as detailed on page 80 and identify the laser and the laser power supply.
Figure 56 Laser and power supply
Page 92
92 . Frontier IR Dual-Range Systems User's Guide
Instrument cover rear hinge
Connector
Cable clip
Screw
Screw
Cable tie
2. Snip, and remove, the cable tie surrounding the power supply, noting how it secures the laser cables (Figure 57).
The laser power supply fixings are accessible from the left of the instrument.
Figure 57 Laser power supply fixings (from left of instrument)
3. Slacken the two screws securing the laser power supply.
4. Unscrew and remove the cable clip that secures the thin black and red PSU laser cable at the rear of the instrument, underneath the laser assembly (Figure 58).
Figure 58 Laser board connector and cable clip (from top of instrument)
5. Disconnect the black and red PSU laser cable from the board.
WARNING
Be careful, the nearby source housing may be hot.
Page 93
Advanced Maintenance . 93
Laser bracket
Laser bracket Laser
Laser power supply
Laser spring retaining pin
Baffle retaining screw
Laser end stop
6. Unhook the spring at the rear of the laser, and remove the two screws on the retaining bracket at the front of the laser (Figure 59).
retaining screw
retaining screw
Figure 59 Laser and power supply installation
7. Remove the laser and power supply; noting the way the cabling is routed.
8. Install the new laser and power supply. Ensure the laser is pushed fully home to the end stop at the front of the housing. The laser must be installed so that the yellow laser warning label is facing directly
upwards.
9. Refit the laser retaining bracket and secure with the two securing screws.
10. Secure the spring on the retaining pin.
11. Reconnect the PSU laser cable to the board connector.
12. Secure the laser power supply with the two securing screws (Figure 57).
13. Refit the cable clip (Figure 58), so that it does not get trapped in the cover.
14. Secure the laser cabling to the laser power supply with a new cable tie.
15. Close and secure the main cover. Refer to steps 11 to 15 on page 82.
Page 94
94 . Frontier IR Dual-Range Systems User's Guide
Page 95

Appendices

Page 96
96 . Frontier IR Dual-Range Systems User's Guide
Release handle
Sample
release clip

Appendix 1: Changing the Sample Area Accessory

1. Remove the sample area cover if fitted by opening the cover, pressing the clip and pulling the cover vertically to remove (Figure 60).
cover
Figure 60 Removing the sample area cover
2. Remove the current sampling accessory by pulling the release handle (under the baseplate of the accessory) towards you and then sliding the accessory towards you (Figure 61).
Figure 61 Removing a sampling accessory
Page 97
Appendices . 97
3. Install the new accessory (Figure 62) by sliding it onto the ledge and pushing it into the sample compartment until it engages with the connector.
Figure 62 Installing a sampling accessory
Spectrum automatically recognizes which accessory is installed, and displays the accessory information on the Setup Instrument tabs.
Figure 63 Setup Instrument BeamPath tab with Universal ATR icon (circled)
Page 98
98 . Frontier IR Dual-Range Systems User's Guide

Appendix 2: Instrument Self-Checks

NOTE: It can take the instrument about two hours to equilibrate when switched on after
being switched off overnight. To save time, we suggest that you leave the spectrometer switched on at all times.
When you switch on the instrument, the instrument display provides this sequence of messages:
Initializing
>
Initializing Checking Program Integrity Checking Instrument Configuration Checking Source Checking Laser Checking Scan Checking Calibration
Initialization Complete.
When initialization is complete the last message fades, leaving the instrument identity message:
PerkinElmer Frontier FT-IR
When connected to a PC, the instrument display adds
When collecting data, the instrument display provides prompts generated by the PerkinElmer software.
Ready
.
Page 99
Appendices . 99

Initialization errors

If your instrument generates one of the following error messages, switch off the instrument, then switch the instrument on again. Consult your PerkinElmer service representative if the problem persists.
Error message Action
Program Integrity Failed The instrument’s firmware may be corrupt, so must be
reinstalled.
Instrument Configuration Failed Suggests that one or more internal devices is failing. If
the problem persists, note any associated error messages and consult your PerkinElmer service representative.
Scan Failed If the problem persists, note any associated error
messages and consult your PerkinElmer service representative.
Laser Failed Replace Laser
Source Failed Replace Source
Calibration Failed Too little energy may be reaching the detector. Make
Consult your PerkinElmer service representative.
The beamsplitters may be missing, the laser fringe pre­amplifier may have failed, or the laser itself may have failed.
Refer to page 91.
Refer to
sure the sample compartment is clear.
Replacing the Laser and Power Supply
Replacing a Source
on page 83.
on
Page 100
100 . Frontier IR Dual-Range Systems User's Guide

Appendix 3: Instrument Performance Validation

The Frontier FT-IR/FT-NIR spectrometer is supplied with NIR Performance Validation Kit (part number L1250405), which contains a polystyrene reference sample.
To validate the MIR performance of your instrument, use the optional MIR Performance Validation Kit (part number L1250404). This kit contains an MIR polystyrene reference sample and an NG11 Schott Glass reference sample.
Macros are supplied with Spectrum software (version 10 or later) that can be used with the Performance Validation kits to produce a report certifying your instrument.
NOTE: Instrument Performance Validation is not applicable for the FIR range.