Installation Instructions

Contacting PerkinElmer

Features

This instruction shee t describes the installation of this accessory which is used with the FL 6500/8500 Fluorescence Spectrometer.
NOTE:
Read these instructions before you install this accessory.
Supplies, replacement parts, and accessories can be ordered directly from PerkinElmer, using the part numbers.
See our website: http://perkinelmer.com PerkinElmer's catalog service offers a full selection of high-quality supplies. To place an order for supplies and many replacement parts, request a free catalog, or ask for information: If you are located within the U.S., call toll free 1-800-762-4000, 8 a.m. to 8 p.m. EST. Your order will be
shipped promptly, usually within 24 hours. If you are located outside of the U.S., call your l oca l PerkinElmer sales or service office.
• Use for measuring various samples of powder, film, glass, and plate, etc.
• Easy installation and user convenient sampling
• Excellent durability
Figure 1 FL 6500/8500 Variable Angle Solid Sample Holder [P/N:N4201014]
PerkinElmer, 710 Bridgeport Avenue,
Shelton, CT 06484-4794, U.S.A
Produced in the USA.
Page 2
2

Dimensions and Specifications

Configuration of the Variable Angle Solid Sample Holder

Installation

Physical Characteristic Specification
Dimensions (mm) 130 x 267 x 95 (WDH)
Weight (Kg) 0.96
Angle (Degree) 0 ~ 80 (Depend on sample size)
Max. Length : 140
Sample Size (mm)
Max. Thickness : 13
09931488A
Figure 2 Variable Angle Solid Sample Holder Configuration
Figure 3 FL Precision Cell for Powder [P/N: N4201032]
1. Prepare the FL 6500/8500 Fluorescence Spectrometer to install this accessory.
2. Connect the power cord and the communication cable.
3. Loosen the accessory fixing bolt to take apart the existing cell holder.
Page 3
3
Figure 4 Loosening the Accessory Fixing Bolt
4. Pull out the cell holder by hand.
09931488A
Figure 5 Pulling Out the Cell Holder
5. After checking the pogo pin position of the sample compartment, attach the Variable Angle Solid Sample Holder to the pogo pin.
6. Tighten the accessory fixing bolt.
Figure 6 Tightening the Accessory Fixing Bolt
Page 4
4

Measurement

1. Double click on the Spectrum FL software and select the measurement mode.
2. Check the recognition of Accessory.
3. Set up the measurement parameters.
09931488A
NOTE:
For more detail of method, refer to Spectrum FL Software Users Guide.
4. Click Save to save the method after setting up the parameters.
5. Put the sample into the sample holder. Adjust the proper incident angle by rotating the holder.
6. After setting the angle, fix it with the hand bolt. Close the lid and select the Run icon.
7. Input the sample name and select OK.
8. Confirm the spectrum and results. Save or print the d ata.
Page 5
09931488A
5

Precision Cell Installation Procedure

Troubleshooting

When the sample does not fit in the Variable Angle Solid Sample Holder

When fluore s c e nc e is not meas ur e d or mea s ure d fluor escen c e i nt e ns it y is to o lo w

1. For powder sample, use the Precision Cell provided with the Variable Angle Solid Sample Holder.
2. Load the powder sample on the base of lower cover.
3. Place the screw top cover on the lower cover and tighten the precision cell by turning the screw cap clockwise.
1. Check the sample size and reduce the sample size if possible. The Variable Angle Solid Sample Holder is designed for the solid sample with maximum thickness of 13 mm and maximum diameter of 140 mm.
1. Check whether Variable Angle Solid Sample Holder is installed exactly on cell compartment.
Page 6
6
2. Check if incident angle is proper. If incident angle get out of recommended incident angle (30 ~70˚) or sample position get out of center, fluorescence intensity measured low because sample is not adequately struck by beam.
09931488A