Omron M3W User Manual

Manufacturer
EU-representative
Production facility Site de production Herstellungswerk Stabilimento di produzione
Subsidiary
Succursale Niederlassung Consociata
OMRON HEALTHCARE Co., Ltd.
53, Kunotsubo, Terado-cho, Muko, Kyoto, 617-0002 JAPAN
OMRON HEALTHCARE EUROPE B.V.
Scorpius 33, 2132 LR Hoofddorp, THE NETHERLANDS www.omron-healthcare.com
Planta de producción Productiefaciliteit
Производственное подразделение
Empresa filial Dochteronderneming Филиал
OMRON HEALTHCARE MANUFACTURING VIETNAM CO., LTD.
Binh Duong Province, VIETNAM
OMRON HEALTHCARE UK LTD.
Opal Drive Fox Milne, Milton Keynes, MK15 0DG, U.K.
OMRON MEDIZINTECHNIK HANDELSGESELLSCHAFT mbH
John-Deere-Str. 81a, 68163 Mannheim, GERMANY www.omron-medizintechnik.de
OMRON SANTÉ FRANCE SAS
14, rue de Lisbonne, 93561 Rosny-sous-Bois Cedex, FRANCE
Uniquement pour le marché français:
OMRON Service Après Vente N° Vert 0 800 91 43 14 consommateurs@omron-sante.fr www.omron-sante.fr/contact
Digital Automatic Blood Pressure Monitor
Instruction Manual
English Français Deutsch Italiano Español Nederlands Русский
IM-HEM-7202-E(V)-02-01/2012
5334060-6B
Check following components!
Vérifier les composants suivants ! Prüfen Sie folgende Teile des Lieferumfangs! Controllare i componenti indicati di seguito! ¡Compruebe los siguientes componentes!
Eng
lis
h
Français
D
Deu
ig
ital A
tsc h
ut
om
a t
ic B
Italia
lo
o d
no
P r
ess
u re
M
Moni
o
d
Esp
el
to
M
r
3W
o l
I n
s
t r u
ct
Ned
i o
erland
n
M
a n
s
u
a l
Рус
ский
IM
­H E
M
-7 2
02
-E (V )
-0
2
­0
1/ 20
12
5334060
-6 B
Controleer de volgende onderdelen! Проверьте следующие компоненты!
Contents
21
F
c
L
d
J
b
a
Thank you for purchasing the OMRON M3W Digital Automatic Blood Pressure Monitor. The OMRON M3W is a compact, fully automatic blood pressure monitor, operating on the
oscillometric principle. It measures your blood pressure and pulse rate simply and quickly. For comfortable controlled inflation without the need of pressure pre-setting or re-inflation the devices uses its advanced “IntelliSense” technology.
Intended Use This product is designed to measure the blood pressure and pulse rate of people within the range of the designated arm cuff, following the instructions in this instruction manual. It is mainly designed for general household use. Please read the Important Safety Information in this instruction manual before using the unit.
Important Safety Information........................ 1
1. Overview ...................................................3
2. Preparation ...............................................4
2.1 Installing/Replacing the Batteries...........4
2.2 Setting the Buzzer/ Date and Time ........5
3. Using the Unit...........................................6
3.1 Applying the Arm Cuff ............................6
3.2 How to Sit Correctly ...............................7
3.3 Taking a Reading...................................8
3.4 Using the Memory Function .................12
Please read this instruction manual thoroughly before using the unit. Please keep for future reference. For specific information about your own blood pressure, CONSULT YOUR DOCTOR.
4. Troubleshooting and Maintenance....... 14
4.1 The Icons and Error Messages ........... 14
4.2 Troubleshooting................................... 16
4.3 Maintenance........................................ 18
4.4 Storage................................................19
5. Optional Parts.........................................20
6. Technical Data ........................................21
7. Some Useful Information about Blood Pressure.... 24
F E
A
B
C
D
I J K
L
G
H
M
N O P Q
R S T
U
V
W X
Y Z AA
Important Safety Information
Consult your doctor prior to using in pregnancy or if diagnosed with arrhythmia or arteriosclerosis. Please read this section carefully before using the unit.
Warning:
• Indicates a potentially hazardous situation which, if not avoided, could result in death or serious injury.
(General Usage)
• Always consult your doctor. Self-diagnosis of measurement results and self-treatment are dangerous.
• People with severe blood flow problems, or blood disorders, should consult a doctor before using the unit, as cuff inflation can cause internal bleeding.
(AC Adapter Usage)
• Never plug in or unplug the power cord from the electric outlet with wet hands.
(Battery Usage)
• If battery fluid should get in your eyes, immediately rinse with plenty of clean water. Consult a doctor immediately.
Caution:
• Indicates a potentially hazardous situation which, if not avoided, may result in minor or moderate injury to the user or patient or damage to the equipment or other property.
EN
(General Usage)
• Do not leave the unit unattended with infants or persons who cannot express their consent.
• Do not use the unit for any purpose other than measuring blood pressure.
• Do not disassemble the unit or arm cuff.
• Use only the approved arm cuff for this unit. Use of other arm cuffs may result in incorrect measurement results.
• Make sure that the air tube is not wrapped around other parts of your body when taking measurements. This could result in injury when the air pressure in the air tube is increased.
• Do not leave the cuff wrapped on the arm if taking measurements during the night. This could result in injury.
• Do not inflate the arm cuff over 299 mmHg.
• Do not use a mobile phone or other devices that emit electromagnetic fields, near the unit. This may result in incorrect operation of the unit.
• Do not operate the unit in a moving vehicle (car, airplane).
• To inflate the cuff manually, refer to Section 3.3. If the cuff is over inflated, it can cause internal bleeding.
1
Important Safety Information
(AC Adapter Usage)
• Use only the original AC adapter (optional) designed for this unit. Use of unsupported adapters may damage and/or may be hazardous to the unit.
• Plug the AC adapter into the appropriate voltage outlet. Do not use a multiple-tap.
• Do not use the AC adapter if the unit or the power cord is damaged. Turn off the power and unplug the power cord immediately.
(Battery Usage)
• If battery fluid should get on your skin or clothing, immediately rinse with plenty of clean water.
• Use only four “AA” alkaline or manganese batteries with this unit. Do not use other types of batteries.
• Do not insert the batteries with their polarities incorrectly aligned.
• Replace old batteries with new ones immediately. Replace all four batteries at the same time.
• Remove the batteries if the unit will not be used for three months or more.
• Do not use new and used batteries together.
General Precautions
• Do not apply strong shocks and vibrations to or drop the unit and arm cuff.
2
• Do not take measurements after bathing, drinking alcohol, smoking, exercising or eating.
• Do not forcibly bend the arm cuff or bend the air tube excessively.
• When removing the air tube, pull on the air plug at the connection with the main unit not the tube itself.
• Do not inflate the arm cuff when it is not wrapped around your arm.
• Do not wash the arm cuff or immerse it in water.
• Read and follow the “Important information regarding Electro Magnetic Compatibility (EMC)” in the Technical Data Section.
• Read and follow the “Correct Disposal of This Product” in the Technical Data Section when disposing of the device and any used accessories or optional parts.
1.
Open the rear cover page to read following:
A
B
CDE
F
G
H
I
J
KLM
N
O
PQR
S
T
U
V
W
X
Y
Z
AA
Overview
The alphabet in the rear cover page correspond to those in the body page.
Main unit
Display O/I START ( ) button User A/B button Up/Down ( ) buttons Buzzer and Date/Time setting ( ) button Air Jack Battery compartment AC adapter jack (for optional AC adapter)
Arm cuff
Arm cuff (Wide Range Cuff: arm
Display
Systolic blood pressure Diastolic blood pressure Heartbeat symbol
(Flashes during measurement)
Average value symbol Deflation symbol Memory symbol Irregular heartbeat symbol Movement error symbol Blood pressure level indicator Pulse display and Memory
number Buzzer symbol
EN
circumference 22-42cm) Marker Air tube Air plug
Cuff wrapping guide USER ID symbol (A/B) Battery low symbol Date/Time display
3
2.
2.1
Preparation
Installing/Replacing the Batteries
1. Remove the battery cover.
2. Insert four “AA” batteries as
indicated in the battery compartment and then replace the battery cover.
Notes:
• If the battery low symbol ( ) appears on the display, turn the unit off then replace all batteries at the same time.
• The measurement values continue to be stored in memory even after the batteries are replaced.
4
Disposal of used batteries should be carried out in accordance with the national regulations for the disposal of batteries.
2. Preparation
2.2
1)
2)
1)
2)
Setting the Buzzer/ Date and Time
1. Press the button.
2. Set the Buzzer.
Push or buttons to select On or Off.
Push the button to confirm.
On
Off
3. Set the monitor to the correct date and time before
taking a measurement for the first time.
Push or buttons to change.
EN
Push the button to confirm.
Next setting appears.
Year
Month
Day
Hour
4. Press the O/I START button to store the setting.
Notes:
• If the batteries have been removed for 30 seconds or more, the date and time setting will need to be reset.
• If the date and time are not set, “-:--” appears during or after measurement.
Minute
5
3.
3.1
Open both the front and rear covers to read following:
L
F
a
J
b
Using the Unit
The alphabet and number in the cover page correspond to those in the body page.
Applying the Arm Cuff
Remove tight-fitting clothing or tight rolled up sleeve from your upper arm. Do not place the cuff over thick clothes.
1. Insert the air plug into the air jack.
2. Put your arm through the cuff loop.
3. Position the arm correctly.
The bottom edge of the cuff should be 1 to 2 cm above the elbow. Marker (arrow under tube) is centred on the middle of your inner arm. Close the fabric fastener FIRMLY.
Notes:
• When you take a measurement on the right arm, air tube will be at the side of your elbow. Be careful not to rest your arm on the air tube. ---
• The blood pressure can differ between the right arm and the left arm, and therefore also the measured blood pressure values can be different. OMRON recommends to always use the same arm for measurement. If the values between the two arms differ substantially, please check with your physician which arm to use for your measurement.
6
3. Using the Unit
3.2
c
d
How to Sit Correctly
To take a measurement, you need to be relaxed and comfortably seated, under comfortable room temperature. No eating, smoking or exercising 30 minutes before taking a measurement.
• Sit on a chair with your feet flat on the floor.
• Sit upright with your back straight. ---
• The cuff should be at the same level as your heart. ---
EN
Cuff wrapping guide
The Cuff Wrapping Guide is a unique feature that indicates if the cuff is not wrapped tightly enough around the arm. Even when the is displayed, a blood pressure reading will be taken.
Note:
This reading is NOT reliable due to the incorrect wrapping of the cuff. Please wrap the cuff again, taking care to wrap it correctly and take the measurement again. When the is displayed, the cuff is correctly wrapped tightly enough on the arm and the reading is accurate and reliable.
7
3. Using the Unit
3.3
Taking a Reading
Notes:
• To cancel a measurement, press the O/I START button to turn off the unit and to release the air in the arm cuff.
• Remain still while taking a measurement.
The monitor is designed to take measurements and store the measurement values in the memory for two people using USER ID A and USER ID B.
1. Press the O/I START button.
2. Select your USER ID (A or B).
8
3. Press the O/I START button again.
The cuff will start to inflate automatically. The buzzer will sound when measurement is completed.
3. Using the Unit
If your systolic pressure is more than 220 mmHg
After the cuff starts to inflate, press and hold the O/I START button until the monitor inflates 30 to 40 mmHg higher than your expected systolic pressure.
Notes:
• The monitor will not inflate above 299 mmHg.
• Do not apply more pressure than necessary.
START
INFLATE
DEFLATE
END
COMPLETE
Notes:
• During measurement, the buzzer (if set to “on”) will beep in rhythm with your heartbeat.
• Wait 2-3 minutes before taking another blood pressure measurement. Waiting between readings allows the arteries to return to the condition prior to taking the blood pressure measurement.
Warning: Self-diagnosis of measured results and treatment are dangerous. Please follow the instructions of your doctor.
EN
9
3. Using the Unit
4. Remove the arm cuff.
5. Press the O/I START button to turn off the monitor.
The monitor automatically stores the measurement in its memory. It will automatically turn off after five minutes.
Important:
• Recent research suggests that the following values can be used as a guide to
high blood pressure for measurements taken at home.
Systolic Blood Pressure Above 135 mmHg
Diastolic Blood Pressure Above 85 mmHg
This criteria is for home blood pressure measurement. For professional office blood pressure measurement criteria, please refer to Chapter 7.
10
3. Using the Unit
• Your blood pressure monitor includes an irregular heartbeat feature. Irregular heartbeats
can influence the results of the measurement. The irregular heartbeat algorithm automatically determines if the measurement is usable or needs to be repeated. If the measurement results are affected by irregular heartbeats but the result is valid, the result is shown together with the irregular heartbeat symbol ( ). If the irregular heartbeats cause the measurement to be invalid, no result is shown. If the irregular heartbeat symbol ( ) is shown after you have taken a measurement, repeat the measurement. If the irregular heartbeat symbol ( ) is shown frequently, please make your doctor aware of it.
• If you move during measurement, the movement error symbol ( ) will appear on the
display. Keep still and repeat the measurement.
EN
11
3. Using the Unit
3.4
Using the Memory Function
The monitor automatically stores the result up to 60 sets for each user (A and B). It can also calculate an average reading based on the measurements from the last three readings taken within 10 minutes. If there are only two readings in memory for that period, the average will be based on two readings. If there is one reading in memory for that period, the average will be based on one reading.
Notes:
• If the memory is full, the monitor will delete the oldest readings.
• When viewing the reading taken without setting the date and time, “-:--” is displayed instead of the date
and time.
To View the Average Value
1. Press the or button.
2. Select your USER ID (A or B).
12
Note: If there are no measurements results stored in the memory, the
screen to the right is displayed.
3. Using the Unit
To View the Readings Stored in Memory
1. Press the button, while the average value is displayed.
The Memory number appears for a second before the pulse rate is displayed. The newest set is numbered “1”.
Alternating date/ time display
2. Press the or button to view the readings stored in
memory.
: To the older readings
: To the more recent readings
To Delete All the Values Stored in Memory
The values stored in the memory are deleted by USER ID. When the memory symbol ( ) appears, first press
EN
button. Then while holding it down, press
the the O/I START button simultaneously for about 2-3
seconds.
Note: You cannot partially delete values stored in the memory.
All values for the user you select will be deleted.
13
4.
4.1
Blink
Lit
Troubleshooting and Maintenance
The Icons and Error Messages
Error Display Cause Remedy
Remove the arm cuff. Wait 2-3 minutes and then
Irregular or weak pulses are detected.
Movement during measurement.
Cuff is not applied correctly.
The batteries are low.
The batteries are exhausted.
take another measurement. Repeat the steps in section 3.3. If this error
continues to appear, contact your doctor.
Carefully read and repeat the steps in section 3.3.
Apply the arm cuff correctly. Refer to section 3.1.
You should replace them with new ones ahead of time.
Refer to section 2.1.
You should replace them with new ones at once. Refer to section 2.1.
14
4. Troubleshooting and Maintenance
Error Display Cause Remedy
Cuff is under inflated.
Movement during measurement.
Air plug disconnected.
Arm cuff not applied correctly.
Clothing is interfering with the arm cuff.
Air is leaking from the arm cuff.
The arm cuff was inflated above 299 mmHg when inflating the cuff manually.
Carefully read and repeat the steps listed under section 3.3.
Repeat measurement. Remain still and do not talk during measurement.
Refer to section 3.3.
Insert the air plug securely. Refer to section 3.1.
Apply the arm cuff correctly. Refer to section 3.1.
Remove any clothing interfering with the arm cuff. Refer to section 3.1.
Replace cuff with new one. Refer to Chapter 5.
Do not inflate the arm cuff above 299 mmHg. Refer to section 3.3.
EN
Device error. Contact your OMRON retail outlet or distributor.
Note:
The irregular heartbeat symbol ( ) may also be displayed with error messages.
15
4. Troubleshooting and Maintenance
4.2
Troubleshooting
Problem Cause Remedy
The reading is extremely low (or high).
Arm cuff pressure does not rise.
Arm cuff not applied correctly.
Movement or talking during measurement.
Clothing is interfering with the arm cuff.
The air tube is not securely connected into the main unit.
Air is leaking from the arm cuff.
Apply the arm cuff correctly. Refer to section 3.1.
Remain still and do not talk during measurement.
Refer to section 3.3.
Remove any clothing interfering with the arm cuff.
Refer to section 3.2.
Make sure that the air tube is connected securely.
Refer to section 3.1.
Replace the arm cuff with a new one.
Refer to Chapter 5.
16
Arm cuff deflates too soon. The arm cuff is loose.
Cannot measure or readings are too low or too high.
The arm cuff has not been inflated sufficiently.
Apply the cuff correctly so that it is firmly wrapped around the arm.
Refer to section 3.1.
Inflate the cuff so that it is 30 to 40 mmHg above your previous measurement result.
Refer to section 3.3.