BX-URA2
BX-RFA
U-LH100HGAPO
U-LH100HG
Power Supply Unit
U-25ND6-2
U-25ND25-2
U-25ND50-2
U-RSL6
U-RSL6EM
BX-RFSS
U-EXBABG
U-EXBAUB
U-EXBAUG
INSTRUCTIONS
REFLECTED
FLUORESCENCE
SYSTEM
This instruction manual is for the Olympus Reflected Fluorescence System. To ensure the safety,
obtain optimum performance and to familiarize yourself fully with the use of this system, we recommend that you study this manual thoroughly before operating the microscope. Retain this instruction manual in an easily accessible place near the work desk for future reference.
This publication is printed on 100% recycled paper
A X 7 6 2 5
CONTENTS
Correct assembly and adjustments are critical for the reflected fluorescence system to exhibit its full performance. If you are
going to assemble the reflected fluorescence system yourself, please carefully read section 9, “ASSEMBLY” (pages 30 to 35).
IMPORTANT — Be sure to read this section for safe use of the equipment. —
I. REFLECTED FLUORESCENCE OBSERVATION
1 NOMENCLATURE
REFLECTED FLUORESCENCE OBSERVATION PROCEDURE
2
3 USING THE CONTROLS
1General Precautions for Observation....................................................................................................................... 8
2Selecting the Fluorescence Mirror Unit....................................................................................................... 8-10
3Objectives for Various Observation Modes........................................................................................ 10-11
4Turning the Power Supply Unit On............................................................................................................................. 11
5Centering the Field Iris Diaphragm.......................................................................................................................... 12
6Centering the Aperture Iris Diaphragm.............................................................................................................. 13
1-3
4-5
6-7
8-15
7Centering the Mercury Burner................................................................................................................................ 14-15
8Mounting the ND Filters ............................................................................................................................................................ 15
4 SIMULTANEOUS FLUORESCENCE OBSERVATIONS
1
Simultaneous Reflected Fluorescence and Phase Contrast Observations
Simultaneous Reflected Fluorescence and Transmitted Light Nomarski
«UIS2 (UIS) Series for Reflected Light Observation»
6 TROUBLESHOOTING GUIDE
■
PROPER SELECTION OF THE POWER SUPPLY CORD .........................................................................
...... 40-42
43-44
45
46-47
IMPORTANT
This system employs a UIS2/UIS (Universal Infinity System) optical design, and should be used only with
UIS2/UIS microscopes, eyepieces, objectives and condensers for the BX2 series. (Some of the modules
designed for the BX series and objectives/eyepieces for the UIS series are also usable. For details,
please consult Olympus or the catalogues.) Less than optimum performance may result if inappropriate
accessories are used.
The use of a universal reflected fluorescence illuminator has enabled the installation of necessary fluorescence mirror units.
By combining the microscopy techniques as shown below, this system can efficiently be used to find fluorescence emission in any area of cells:
In addition, the following observations are also by installing a general reflected light observation unit (BX-URA2 only):
1. Reflected brightfield/darkfield observations
2. Reflected Nomarski DIC observation
3. Reflected simplified polarized light observation
This manual describes the instructions for I. Reflected Fluorescence Observations in the first half and those for II. Reflected
Light Observations in the second half.
Please find the pages giving you the appropriate instructions for your observation.
SAFETY PRECAUTIONS
1. This system is composed of precision instruments. Handle it with care and avoid subjecting it to sudden or severe
impact.
2. The ultrahigh-pressure mercury burner used should be the USH-103OL DC burner (mfd. by USHIO, Inc.) or the HBO103W/
2 burner (mfd. by OSRAM) that Olympus supplies.
3. Make sure that a mercury burner is attached and that cables are plugged in firmly.
4. The inside of the lamp housing is very hot and hazardous during lighting and for about 10 minutes after turning off. Do not
open the lamp housing in this period. (Page 11)
5. Do not apply excessive force to the stoppers which are provided for some functions. Otherwise, the stopper or equipment
may be damaged.
6. Do not attempt to open or disassemble the power supply unit because it includes high voltage parts inside.
7. Always use the power cord provided by Olympus. If no power cord is provided, please select the proper power cord by
referring to the section “PROPER SELECTION OF THE POWER SUPPLY CORD” at the end of this instruction manual. If the
proper power cord is not used, product safety and performance cannot be guaranteed.
Before plugging the power cord to the power outlet, make sure that the main switch of the power supply unit is set to
“
” (OFF).
8. To ensure safety, be sure to ground the power supply unit. Otherwise, Olympus can no longer warrant the electrical safety
performance of the system.
9. Before opening the lamp housing for replacement of the burner or any other internal part, set the main switch to “
(OFF), then unplug the lamp housing connection cable from the power supply unit, and wait for more than 10 minutes
until the lamp housing cools down.
10
.
The top panel of the lamp housing becomes very hot during operation. To prevent fire hazard, do not block the ventilation
through the top panel.
”
1
Safety Symbols
The following symbols are found on the microscope. Study the meaning of the symbols and always use the equipment
in the safest possible manner.
SymbolExplanation
Indicates the presence of high voltage (1 kV or more). Take caution to guard against electric
shock.
Indicates that the surface becomes hot, and should not be touched with bare hands.
Before use, carefully read the instruction manual. Improper use could result in personal injury to
the user and/or damage to the equipment.
l
Warning indications
Warning indications are placed at parts where special precaution is required when handling and using the System.
Always heed the warnings.
1. This manual pertains only to the reflected fluorescence system. Before using this system together with the BX2 microscope and associated options, make sure that you have carefully read and understood their manuals, and understand
how the system should be operated together.
2. The reflected fluorescence system is composed of precision instruments. Handle it with care and avoid subjecting it to
sudden or severe impact.
3. Do not use the system where it is subjected to direct sunlight, high temperature and humidity, dust or vibrations.
4. To allow heat from the unit to dissipate well, reserve a distance of at least 10 cm between the lamp housing and power
supply unit.
5. The power cord can also be used to cut the power supply in case of emergency. To make this possible, the power supply
unit should be installed so that the power cord connector (on the rear of the power supply unit) or the power outlet is
easily accessible for unplugging in case of emergency.
2
2 Maintenance and Storage
1. To clean the lenses and other glass components, simply blow dirty away using a commercially available blower and wipe
gently using a piece of cleaning paper (or clean gauze).
If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available absolute
alcohol.
!Since the absolute alcohol is highly flammable, it must be handled carefully.
Be sure to keep it away from open flames or potential sources of electrical sparks --- for example, electrical
equipment that is being switched on or off.
Also remember to always use it only in a well-ventilated room.
2. With any part of the system other than glass components gets dirty, do not use organic solvents but wipe it with a clean
cloth. If the part is extremely dirty, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.
3. Do not disassemble any part of the system. This could result in malfunctions or reduced performance.
4. The mercury burner has a service life period of 300 hours (USH-103OL, HBO103W/2). When the hour counter on the
power supply unit indicates this value, set the main switch to “
replacing the mercury burner (Page 33). Unlike electric bulbs, the mercury burner seals high-pressure gas inside. If it
continues to be used after the service life has expired, the glass tube may eventually explode due to accumulated
distortion.
5. When not using the microscope, be sure set the main switch to “
cooled down sufficiently, cover the microscope with the dust cover for storage.
6. When disposing of the microscope, check the regulations and rules of your local government and be sure to observe
them.
” (OFF) and wait for more than 10 minutes before
” (OFF). After confirming that the lamp housing has
3 Caution
If the system is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the
system equipment may also be damaged. Always use the system as outlined in this instruction manual.
The following symbols are used to set off text in this instruction manual.
! : Indicates that failure to follow the instructions in the warning could result in bodily harm to the
user and/or damage to equipment (including objects in the vicinity of the equipment).
# : Indicates that failure to follow the instructions could result in damage to equipment.
} : Indicates commentary (for ease of operation and maintenance).
Fluorescence Mirror Units
U-MWU2, etc., total 24 models
Indicator sheets
Power Supply Unit
(for 100 W mercury burner)
Burner ON LED
}Up to six fluorescence mirror units can be mounted on the BX-RFA or BX-
URA2.
#Each filter unit includes a dichroic mirror, barrier filter and excitation
filter that have been combined according to the excitation method. It is
basically not recommended to open a fluorescence mirror unit.
}It is recommended that you use the U-MF2 dummy filter unit (which does
not contain a filter) when making your original fluorescence unit. (Page 32)
Blank indicator sheets provided with the illuminator can be used to write
the names of original fluorescence mirror units.
Hour counter
Main switch
I : ON
: OFF
ND Filters
U-25ND6-2, U-25ND25-2, U-25ND50-2
Input
receptacle
Output
connector
Centering Target
U-CST
5
REFLECTED FLUORESCENCE OBSERVATION PROCEDURE
}If you need simultaneous observation of reflected fluorescence observation with the phase contrast observation or trans-
}Engage the shutter if you interrupt observation for a short time.
‡ Collector lens focusing knob
Š Field iris diaphragm knob
‰ Aperture iris diaphragm knob
‹ Shutter knob
(P. 14)
(P. 12)
(P. 13)
(P. 12)
6
|
†
‡
‹
ƒ
@
²
…
} Make a photocopy of the observation procedure pages and post it near your microscope.
‰
Š
³
7
USING THE CONTROLS
1 General Precautions for Observation
1. Verify that the power supply voltage and frequency match the requirements inscribed on the Rating plate.
2. Make sure that the power cord and connecting cables are plugged in securely.
3. If you perform only transmitted light phase contrast or transmitted light DIC observations, leave one cube position on the
turret empty. This allows for transmission of white light.
The turret must always be set to one of the click position. If it is deviated from a click position, the cover may be deformed
by heat.
4. Enlarge the field iris diaphragm so it just circumscribes the field of view. If decentered, center it using the Allen screwdriver.
5. Always use immersion for oil immersion objectives.
6. If you use an objective with correction collar such as the UPlanSApo40X, UPlanFLN60X, UPlanApo40X or PlanApo40X,
you can correct variations in cover glass thickness by adjusting the correction collar.
Correction procedure
If the cover glass thickness is known, match the correction collar to the cover glass thickness using the collar scale
provided. If the thickness is not known, turn the collection collar and adjust the fine adjustment knob to where the
image is as sharp as possible.
7. Engage the shutter if you interrupt observation for a short time.
(Turning the mercury burner ON and OFF repeatedly will significantly shorten the life span of the burner.)
8. Color fading of specimens
This system features high excitation light intensity to ensure bright observation of dark fluorescence specimens.
In consequence, after long period of observations using high-power objectives, the colors of specimens will fade quicker
than usual, causing the view (contrast) of fluorescent images to deteriorate.
In such a case, slightly reduce the excitation light intensity to slow color fading down and improve the fluorescence
images.
To reduce the excitation light intensity, use ND filters or aperture iris diaphragm as far as the observation is not affected or
use the shutter to limit the exposure of specimen to more than necessary light.
Commercially-marketed color fading protection agent (DABCO, etc.) can also delay fading of specimen colors. The use of
fading protection agent is recommended especially when you perform high-magnification observations frequently.
#Remember that the fading protection agents cannot be used with certain kinds of specimens.
2 Selecting the Fluorescence Mirror Unit
Select the fluorescence mirror unit which matches the fluorochrome in use.
#Never mount or use the U-MBF3 brightfield mirror unit together with a with a mirror unit for fluorescence. The U-
MBF3 brightness is excessive and injury to the eyes could occur. If this type of mirror unit is to be used together
with a mirror unit for fluorescence, use the U-MBFL3 mirror unit equipped with a built-in ND filter or add a 3% ND
filter to the U-MBF3.
}Use according to the excitation wevelength:
Olympus has prepared some sets of fluorescence mirror unit combined with appropriate filters which are variable depending on wavelengths.
The wide-band (W) set is normally used. There may be cases, however, where superwide-band (SW) or Narrow-band (N)
sets are recommendable.
@Extremely weak fluorescence brightness
(B- and G-excitation only):
² Specimens emitting strong autofluorescence:Use the narrow band (N).
Use the super-wide band (SW).
}With the SWB, strong autofluorescence may reduce
image contrast.
}The fluorescence bright is somewhat reduced.
8
Dichroic Mirror and Filter Configurations of Fluorescence Mirror Units
¦: Recommended combination.
¦* : Slightly inferior in U-excitation.
––: Not usable, or applicable objective is not available.
¦** : A phase contrast (Ph) objective is necessary for phase contrast observation.
UIS Series
Objective
UPlanApo4X
10X
10X O
10X W
20X
20XO3
40X
40XO Ι 3
60X
60XW3
100XOΙ 3
PlanApo40X
60XO3
100XO3
UPlanFI4X
10X
20X
40X
60XO Ι 3
100XO, OΙ3
UApo
20X3/340
20XW3/340
40X3/340
40X
40XW3/340
OΙ 3/340
Reflected light fluorescence
U, V, BVB, IB, G, IY
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
–
¦
–
¦*
¦*
¦*
¦*
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦*
¦*
¦*
¦*
¦
¦
¦
¦
¦
¦
¦
Phase contrast
difference
–
¦**
–
–
¦**
–
–
¦**
–
–
¦**
–
¦**
–
–
¦**
¦**
¦**
¦**
¦**
–
–
–
–
–
Transmitted
light DIC
–
¦
¦
–
¦
¦
¦
¦
–
¦
¦
–
¦
–
–
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦
¦: Recommended combination.
¦* : Usable, but image be dark depending on NA.
––: Not usable, or applicable objective is not available.
¦** : A phase contrast (Ph) objective is necessary for phase contrast observation. The Ph objective is not
available for the UPlanFI100XOI3.
4 Turning the Power Supply Unit On
Set the main switch to “ I ” (ON). The arc will stabilize in 5 to 10 minutes after ignition.
}The discharge type mercury burner may not be ignited from the beginning on rare occasions due to its characteristics.
In this case, set the main switch to “ ” (OFF), wait for 5 to 10 seconds, then set it again to “ I ” (ON).
#To extend the mercury burner life, do not turn the mercury burner off for 15 minutes after ignition.
#The mercury burner cannot be reignited until the mercury vapor has cooled down and liquefied. Before re-igniting
a mercury burner, wait for about 10 minutes after the last time it was turned off.
}For the shake of safety, the power supply to the lamp housing is shut down if the lamp housing is opened while the burner
is on. If this happens, set the main switch to “ ” (OFF), wait for more than 10 minutes, then set it again to “ I ” (ON). Do not
open the lamp housing until it has cooled down enough.
#To reset the hour counter, hold its reset button till “000.0” is displayed.
11
@
²
Fig. 1
³
5 Centering the Field Iris Diaphragm
1. Close the light path by sliding the shutter knob @ to position marked {.
2. Engage the B or IB mirror unit in the light path by rotating the turret.
(If these mirror units are not available, engage another fluorescence mirror unit in the light path.)
3. Open the light path by sliding the shutter knob to position marked \.
4 Engage the 10X objective in the light path, place the specimen on the
stage and bring the image into approximate focus.
5. Pull out the field iris diaphragm knob ² to minimize the field iris diameter.
6. Fit the Allen wrench provided with the microscope frame in the two field
iris centering screws ³ and adjust so that the iris image comes at the
center of the field of view.
7. While pushing in the field iris diaphragm knob ², enlarge the field iris
diaphragm until the field iris image inscribes the field of view. If eccentricity is found after this, try centering again.
8. Enlarge the iris diaphragm until the iris image becomes almost the same
size as (i.e. circumscribes) the field of view.
Effects of Field Iris Diaphragm
The field iris diaphragm restricts the diameter of the beam of light entering the objective and thus excludes extraneous light, improving image
contrast. The field iris diaphragm also functions to prevent color fading of
fluorescent light in other part than the observed region.
To exclude extra light, set the field iris diaphragm knob ² on the fluorescence illuminator according to the objective power, so that the image of
the field iris diaphragm just circumscribes the field of view.
(Fig. 1)
12
Loading...
+ 36 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.