Olympus U-LH100HGAPO, U-LH100HG, U-25ND6-2, U-25ND25-2, U-25ND50-2 User manual

...
BX-URA2 BX-RFA U-LH100HGAPO U-LH100HG Power Supply Unit U-25ND6-2 U-25ND25-2 U-25ND50-2 U-RSL6 U-RSL6EM BX-RFSS U-EXBABG U-EXBAUB U-EXBAUG
INSTRUCTIONS
REFLECTED
FLUORESCENCE
SYSTEM
This instruction manual is for the Olympus Reflected Fluorescence System. To ensure the safety, obtain optimum performance and to familiarize yourself fully with the use of this system, we recom­mend that you study this manual thoroughly before operating the microscope. Retain this instruc­tion manual in an easily accessible place near the work desk for future reference.
This publication is printed on 100% recycled paper
A X 7 6 2 5
CONTENTS
Correct assembly and adjustments are critical for the reflected fluorescence system to exhibit its full performance. If you are going to assemble the reflected fluorescence system yourself, please carefully read section 9, “ASSEMBLY” (pages 30 to 35).
IMPORTANT — Be sure to read this section for safe use of the equipment. —
I. REFLECTED FLUORESCENCE OBSERVATION
1 NOMENCLATURE
REFLECTED FLUORESCENCE OBSERVATION PROCEDURE
2
3 USING THE CONTROLS
1 General Precautions for Observation....................................................................................................................... 8
2 Selecting the Fluorescence Mirror Unit....................................................................................................... 8-10
3 Objectives for Various Observation Modes........................................................................................ 10-11
4 Turning the Power Supply Unit On............................................................................................................................. 11
5 Centering the Field Iris Diaphragm.......................................................................................................................... 12
6 Centering the Aperture Iris Diaphragm.............................................................................................................. 13
1-3
4-5
6-7
8-15
7 Centering the Mercury Burner................................................................................................................................ 14-15
8 Mounting the ND Filters ............................................................................................................................................................ 15
4 SIMULTANEOUS FLUORESCENCE OBSERVATIONS
1
Simultaneous Reflected Fluorescence and Phase Contrast Observations Simultaneous Reflected Fluorescence and Transmitted Light Nomarski
2
Differential Interference Contrast (DIC) Observations .................................................................. 16
5 TROUBLESHOOTING GUIDE
6 SPECTRAL CHARACTERISTICS OF FILTERS
7 SPECIFICATIONS
16
.............. 16
17
18-22
23
8 OPTIONAL MODULES
1 6-Position Filter Slider U-RSL6........................................................................................................................... 24-25
2 6-Position Barrier Filter Slider U-RSL6EM .................................................................................................. 26
3 Rectangle Field Stop BX-RFSS (for exclusive use with the BX-RFA).................... 27
4
Exciter Balancers U-EXBABG/EXBAUB/EXBAUG (for exclusive use with the BX-RFA) ......
24-29
28-29
9 ASSEMBLY
9-1 Assembly Diagram ............................................................................................................................................................................ 30
9-2 Detailed Assembly Procedures ............................................................................................................................. 31-35
— See this section for the replacement of the light bulb. —
II. REFLECTED OBSERVATIONS (BX-URA2 Only)
1 CONFIGURATION OF REFLECTED OBSERVATION SYSTEM
2 ASSEMBLY
FIELD IRIS AND APERTURE IRIS DIAPHRAGM ADJUSTMENTS
3
4 OBSERVATIONS
4-1 Reflected Light Brightfield/Darkfield Observations ........................................................................... 39
30-35
36
37
37-38
39-42
Reflected Light Nomarski Differential Interference Contrast (DIC) Observation
4-2
4-3 Reflected Light Simple Polarized Light Observation....................................................................... 42
5
OPTICAL CHARACTERISTICS
«UIS2 (UIS) Series for Reflected Light Observation»
6 TROUBLESHOOTING GUIDE
PROPER SELECTION OF THE POWER SUPPLY CORD .........................................................................
...... 40-42
43-44
45
46-47
IMPORTANT
This system employs a UIS2/UIS (Universal Infinity System) optical design, and should be used only with UIS2/UIS microscopes, eyepieces, objectives and condensers for the BX2 series. (Some of the modules designed for the BX series and objectives/eyepieces for the UIS series are also usable. For details, please consult Olympus or the catalogues.) Less than optimum performance may result if inappropriate accessories are used.
The use of a universal reflected fluorescence illuminator has enabled the installation of necessary fluorescence mirror units. By combining the microscopy techniques as shown below, this system can efficiently be used to find fluorescence emis­sion in any area of cells:
1. Reflected fluorescence observation + Transmitted light phase contrast observation
2. Reflected fluorescence observation + Transmitted Nomarski Differential Interference Contrast (DIC) observation
3. Reflected fluorescence observation + Transmitted Light Observation
In addition, the following observations are also by installing a general reflected light observation unit (BX-URA2 only):
1. Reflected brightfield/darkfield observations
2. Reflected Nomarski DIC observation
3. Reflected simplified polarized light observation This manual describes the instructions for I. Reflected Fluorescence Observations in the first half and those for II. Reflected Light Observations in the second half. Please find the pages giving you the appropriate instructions for your observation.
SAFETY PRECAUTIONS
1. This system is composed of precision instruments. Handle it with care and avoid subjecting it to sudden or severe impact.
2. The ultrahigh-pressure mercury burner used should be the USH-103OL DC burner (mfd. by USHIO, Inc.) or the HBO103W/ 2 burner (mfd. by OSRAM) that Olympus supplies.
3. Make sure that a mercury burner is attached and that cables are plugged in firmly.
4. The inside of the lamp housing is very hot and hazardous during lighting and for about 10 minutes after turning off. Do not open the lamp housing in this period. (Page 11)
5. Do not apply excessive force to the stoppers which are provided for some functions. Otherwise, the stopper or equipment may be damaged.
6. Do not attempt to open or disassemble the power supply unit because it includes high voltage parts inside.
7. Always use the power cord provided by Olympus. If no power cord is provided, please select the proper power cord by referring to the section “PROPER SELECTION OF THE POWER SUPPLY CORD” at the end of this instruction manual. If the proper power cord is not used, product safety and performance cannot be guaranteed. Before plugging the power cord to the power outlet, make sure that the main switch of the power supply unit is set to “
” (OFF).
8. To ensure safety, be sure to ground the power supply unit. Otherwise, Olympus can no longer warrant the electrical safety performance of the system.
9. Before opening the lamp housing for replacement of the burner or any other internal part, set the main switch to “ (OFF), then unplug the lamp housing connection cable from the power supply unit, and wait for more than 10 minutes until the lamp housing cools down.
10
.
The top panel of the lamp housing becomes very hot during operation. To prevent fire hazard, do not block the ventilation through the top panel.
1
Safety Symbols
The following symbols are found on the microscope. Study the meaning of the symbols and always use the equipment in the safest possible manner.
Symbol Explanation
Indicates the presence of high voltage (1 kV or more). Take caution to guard against electric shock.
Indicates that the surface becomes hot, and should not be touched with bare hands.
Before use, carefully read the instruction manual. Improper use could result in personal injury to the user and/or damage to the equipment.
l
Warning indications
Warning indications are placed at parts where special precaution is required when handling and using the System. Always heed the warnings.
Warning indication position:
Indicates that the main switch is ON.
Indicates that the main switch is OFF.
· Mercury burner lamp housing (U-LH100HG, U-LH100HGAPO
· Power supply unit for 100 W mercury burner
· ND filters (U-25ND6, U-25ND25, U-25ND50)
[Warning against
high temperature]
[Warning against
high voltage]
1 Getting Ready
1. This manual pertains only to the reflected fluorescence system. Before using this system together with the BX2 micro­scope and associated options, make sure that you have carefully read and understood their manuals, and understand how the system should be operated together.
2. The reflected fluorescence system is composed of precision instruments. Handle it with care and avoid subjecting it to sudden or severe impact.
3. Do not use the system where it is subjected to direct sunlight, high temperature and humidity, dust or vibrations.
4. To allow heat from the unit to dissipate well, reserve a distance of at least 10 cm between the lamp housing and power supply unit.
5. The power cord can also be used to cut the power supply in case of emergency. To make this possible, the power supply unit should be installed so that the power cord connector (on the rear of the power supply unit) or the power outlet is easily accessible for unplugging in case of emergency.
2
2 Maintenance and Storage
1. To clean the lenses and other glass components, simply blow dirty away using a commercially available blower and wipe gently using a piece of cleaning paper (or clean gauze). If a lens is stained with fingerprints or oil smudges, wipe it gauze slightly moistened with commercially available absolute alcohol.
!Since the absolute alcohol is highly flammable, it must be handled carefully.
Be sure to keep it away from open flames or potential sources of electrical sparks --- for example, electrical equipment that is being switched on or off. Also remember to always use it only in a well-ventilated room.
2. With any part of the system other than glass components gets dirty, do not use organic solvents but wipe it with a clean cloth. If the part is extremely dirty, use a lint-free, soft cloth slightly moistened with a diluted neutral detergent.
3. Do not disassemble any part of the system. This could result in malfunctions or reduced performance.
4. The mercury burner has a service life period of 300 hours (USH-103OL, HBO103W/2). When the hour counter on the power supply unit indicates this value, set the main switch to “ replacing the mercury burner (Page 33). Unlike electric bulbs, the mercury burner seals high-pressure gas inside. If it continues to be used after the service life has expired, the glass tube may eventually explode due to accumulated distortion.
5. When not using the microscope, be sure set the main switch to “ cooled down sufficiently, cover the microscope with the dust cover for storage.
6. When disposing of the microscope, check the regulations and rules of your local government and be sure to observe them.
” (OFF) and wait for more than 10 minutes before
” (OFF). After confirming that the lamp housing has
3 Caution
If the system is used in a manner not specified by this manual, the safety of the user may be imperiled. In addition, the system equipment may also be damaged. Always use the system as outlined in this instruction manual.
The following symbols are used to set off text in this instruction manual. ! : Indicates that failure to follow the instructions in the warning could result in bodily harm to the
user and/or damage to equipment (including objects in the vicinity of the equipment). # : Indicates that failure to follow the instructions could result in damage to equipment. } : Indicates commentary (for ease of operation and maintenance).
3
I. REFLECTED FLUORESCENCE OBSERVATION
NOMENCLATURE
Reflected Illuminator BX-URA2 Fluorescence Illuminator BX-RFA
Note
Mirror unit turret
Mirror unit inscription pocket (Page 31)
The diagram shows the BX-RFA. Parts marked * are not provided on the BX-URA2.
100 W Mercury Apo Lamp Housing U-LH100HGAPO 100 W Mercury Lamp Housing U-LH100HG
Field iris diaphragm centering screws (Page 12)
x2 screws.
Collector lens focusing knob (Page 14)
Burner centering knobs
(Page 14)
Mirror focusing screw
(Page 15)
On the rear of the
lamp housing.
Aperture iris diaphragm centering screws (Page 13)
x2 screws.
Shutter knob (Page 12) \: Shutter OUT
{: Shutter IN
Field iris diaphragm knob (Page 12)
6-position filter inlet (Page 24)
Analyzer/6-position barrier filter slider inlet (Page 26)
3
3
2
1
Aperture iris diaphragm knob
6
5
4
(Page 13)
4
ND filter/* exciter balancer inlet (Pages 15 & 28)
* 6-Position filter slider inlet (Page 24)
Fluorescence Mirror Units U-MWU2, etc., total 24 models
Indicator sheets
Power Supply Unit (for 100 W mercury burner)
Burner ON LED
}Up to six fluorescence mirror units can be mounted on the BX-RFA or BX-
URA2.
#Each filter unit includes a dichroic mirror, barrier filter and excitation
filter that have been combined according to the excitation method. It is basically not recommended to open a fluorescence mirror unit.
}It is recommended that you use the U-MF2 dummy filter unit (which does
not contain a filter) when making your original fluorescence unit. (Page 32) Blank indicator sheets provided with the illuminator can be used to write the names of original fluorescence mirror units.
Hour counter
Main switch I : ON
: OFF
ND Filters U-25ND6-2, U-25ND25-2, U-25ND50-2
Input receptacle
Output connector
Centering Target U-CST
5
REFLECTED FLUORESCENCE OBSERVATION PROCEDURE
}If you need simultaneous observation of reflected fluorescence observation with the phase contrast observation or trans-
mitted light Nomarski Differential Interference Contrast (DIC) observation, please read Chapter 4, “SIMULTANEOUS FLUO­RESCENCE OBSERVATION”. (Page 16)
(Controls Used)
Preparation
· Attach the fluorescence mirror unit and objective matching the observation method. (Pages 8 to 11)
· Center the mercury burner. (Page 14 or 15)
Set the main switch to “ I ” (ON) and wait for
the arc to stabilize.
Place the specimen on the stage.
Engage the fluorescence mirror unit
matching the specimen in the light path.
Engage the objective in the light path and
focus on the specimen.
Engage an ND filter in the light path as required.
@ Main switch
² Specimen holder ³ X-/Y-axis knobs
| Mirror unit turret
ƒ Revolving nosepiece … Coarse/fine adjustment knobs
† ND filters
(Page)
(P. 11)
(P. 15)
Adjust so that the entire field is
uniform and brightest.
Adjust the field iris diaphragm.
Adjust the aperture iris diaphragm.
Start observation.
}Engage the shutter if you interrupt observation for a short time.
‡ Collector lens focusing knob
Š Field iris diaphragm knob
‰ Aperture iris diaphragm knob
‹ Shutter knob
(P. 14)
(P. 12)
(P. 13)
(P. 12)
6
|
ƒ
@
²
} Make a photocopy of the observation procedure pages and post it near your microscope.
Š
³
7
USING THE CONTROLS
1 General Precautions for Observation
1. Verify that the power supply voltage and frequency match the requirements inscribed on the Rating plate.
2. Make sure that the power cord and connecting cables are plugged in securely.
3. If you perform only transmitted light phase contrast or transmitted light DIC observations, leave one cube position on the turret empty. This allows for transmission of white light. The turret must always be set to one of the click position. If it is deviated from a click position, the cover may be deformed by heat.
4. Enlarge the field iris diaphragm so it just circumscribes the field of view. If decentered, center it using the Allen screwdriver.
5. Always use immersion for oil immersion objectives.
6. If you use an objective with correction collar such as the UPlanSApo40X, UPlanFLN60X, UPlanApo40X or PlanApo40X, you can correct variations in cover glass thickness by adjusting the correction collar.
Correction procedure
If the cover glass thickness is known, match the correction collar to the cover glass thickness using the collar scale provided. If the thickness is not known, turn the collection collar and adjust the fine adjustment knob to where the image is as sharp as possible.
7. Engage the shutter if you interrupt observation for a short time. (Turning the mercury burner ON and OFF repeatedly will significantly shorten the life span of the burner.)
8. Color fading of specimens This system features high excitation light intensity to ensure bright observation of dark fluorescence specimens. In consequence, after long period of observations using high-power objectives, the colors of specimens will fade quicker than usual, causing the view (contrast) of fluorescent images to deteriorate. In such a case, slightly reduce the excitation light intensity to slow color fading down and improve the fluorescence images. To reduce the excitation light intensity, use ND filters or aperture iris diaphragm as far as the observation is not affected or use the shutter to limit the exposure of specimen to more than necessary light. Commercially-marketed color fading protection agent (DABCO, etc.) can also delay fading of specimen colors. The use of fading protection agent is recommended especially when you perform high-magnification observations frequently.
#Remember that the fading protection agents cannot be used with certain kinds of specimens.
2 Selecting the Fluorescence Mirror Unit
Select the fluorescence mirror unit which matches the fluorochrome in use.
#Never mount or use the U-MBF3 brightfield mirror unit together with a with a mirror unit for fluorescence. The U-
MBF3 brightness is excessive and injury to the eyes could occur. If this type of mirror unit is to be used together with a mirror unit for fluorescence, use the U-MBFL3 mirror unit equipped with a built-in ND filter or add a 3% ND filter to the U-MBF3.
}Use according to the excitation wevelength:
Olympus has prepared some sets of fluorescence mirror unit combined with appropriate filters which are variable de­pending on wavelengths. The wide-band (W) set is normally used. There may be cases, however, where superwide-band (SW) or Narrow-band (N) sets are recommendable.
@Extremely weak fluorescence brightness
(B- and G-excitation only):
² Specimens emitting strong autofluorescence: Use the narrow band (N).
Use the super-wide band (SW). }With the SWB, strong autofluorescence may reduce
image contrast.
}The fluorescence bright is somewhat reduced.
8
Dichroic Mirror and Filter Configurations of Fluorescence Mirror Units
Excitation
Method
U
V
BV
B
IB
G
IG IY
Mirror Unit Dichroic Mirror Excitation Filter Barrier Filter Fluorochromes
U-MWU2
U-MNU2
U-MNV2
U-MWBV2
U-MNBV2 BP420-440
U-MWB2 U-MNB2
U-MSWB2
U-MWIB3
U-MNIB3
U-MWG2
U-MNG2
U-MSWG2
U-MWIG3 DM570 BP530-550 BA575IF
U-MWIY2 DM600
DM400
DM455 BP400-410 BA455
DM455
DM500
DM505
DM570
BP330-385
BP360-370
BP400-440
BP460-490 BP470-490
BP420-480
BP460-495 BP470-495
BP510-550
BP530-550
BP480-550
BP545-580 BA610IF
BA420
BA475
BA520IF
BA510IF
BA590
· Autofluorescence observation
· DAPI: DNA staining
· Hoechest 33258, 33342: Chromosome
· Catecholamine
· Serotonin
· Tetracyline: Bones, teeth
· Quinacrine, quinacrine mustard: Chromosome
· Thioflavine S: Lymphocyte
· Acriflavine: Nucleic acid
· ECFP
· FITC: Fluorescent antibody
· Acridine orange: DNA, RNA
· Auramine: Tubercle bacillus
· EGFP, S65T, RSGFP
· Rhodamine, TRITC: Florescent antibody
· Propidium iodide: DNA
· RFP
Texas Red: Fluorescent antibody
Color Separation Filter Combinations
U U-MNUA2
U-MWIBA3
IB
U-MNIBA3
U-MWIGA3
G
U-MNIGA3
Mirror Unit Name Meaning
DM400 BP360-370 BA420-460
DM505
DM570
U- MN I B A 2
Mirror unit
Universal
For observing only the U-excitation stain, when using U-excitation stain together with FITC.
BP460-495
BA510-550
BP470-495
BP530-550
BA575-625
BP540-550
Model number (2 or 3) For color separation Excitation (U, V, BV, B, IB, G, IG or IY) Bandwidth (SW: Superwide band. W: Wide band. N: Narrow band.)
For observing only the B-excitation stain, when using B-excitation stain with TRITC or Texas Red.
For observing only the G-excitation stain, when using G-excitation stain together with Cy5.
9
Exclusively for Fluorescent Proteins
Excitation
Method
CFP U-MCFPHQ DM450HQ BP425-445HQ BA460-510HQ For ECFP GFP U-MGFPHQ DM485HQ BP460-480HQ BA495-540HQ For EGFP YFP U-MYFPHQ DM505HQ BP490-500HQ BA515-560HQ For EYFP RFP U-MRFPHQ DM565HQ BP535-555HQ BA570-625HQ For RFP
Mirror Unit Name Meaning
Mirror Unit Dichroic Mirror Excitation Filter Barrier Filter Fluorochromes
U- MC F P HQ
High Quality CFP/GFP/YFP/RFP
Mirror unit
Universal
3 Objectives for Various Observation Modes
UIS2 Series
Objective
UPlanSApo 4X
10X 20X 20X O 40X 60X W 60X O
100X O
PlanApoN 60X O UPlanFLN 4X
10X 20X 40X 40X O 60X 60X OΙ
100X O2 100X OΙ2
Reflected light
fluorescence
¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦
¦* ¦
¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦
Phase contrast
difference
– – – – – – – –
–¦ –
¦** ¦** ¦** – – ¦** ¦**
Transmitted light DIC
¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦
– ¦ ¦ ¦ ¦
– ¦ ¦ ¦
10
¦ : Recommended combination. ¦* : Slightly inferior in U-excitation. –– : Not usable, or applicable objective is not available. ¦** : A phase contrast (Ph) objective is necessary for phase contrast observation.
UIS Series
Objective
UPlanApo 4X
10X 10X O 10X W
20X
20X O3 40X 40X O Ι 3 60X 60X W3
100X OΙ 3
PlanApo 40X
60X O3
100X O3
UPlanFI 4X
10X 20X 40X 60X O Ι 3
100X O, OΙ3
UApo
20X 3/340
20X W3/340
40X 3/340
40X
40X W3/340
OΙ 3/340
Reflected light fluorescence
U, V, BV B, IB, G, IY
¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦
¦
¦* ¦* ¦* ¦* ¦ ¦
¦ ¦ ¦ ¦ ¦
¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦ ¦
¦ ¦ ¦
¦* ¦* ¦* ¦* ¦ ¦
¦ ¦ ¦ ¦ ¦
Phase contrast
difference
– ¦** – – ¦** – – ¦** – – ¦**
– ¦**
– ¦** ¦** ¦** ¦** ¦**
– – – – –
Transmitted
light DIC
– ¦ ¦
– ¦ ¦ ¦ ¦
– ¦ ¦
¦
– ¦ ¦ ¦ ¦ ¦
¦ ¦ ¦ ¦ ¦
¦ : Recommended combination. ¦* : Usable, but image be dark depending on NA. –– : Not usable, or applicable objective is not available. ¦** : A phase contrast (Ph) objective is necessary for phase contrast observation. The Ph objective is not
available for the UPlanFI100XOI3.
4 Turning the Power Supply Unit On
Set the main switch to “ I ” (ON). The arc will stabilize in 5 to 10 minutes after ignition.
}The discharge type mercury burner may not be ignited from the beginning on rare occasions due to its characteristics.
In this case, set the main switch to “ ” (OFF), wait for 5 to 10 seconds, then set it again to “ I ” (ON).
#To extend the mercury burner life, do not turn the mercury burner off for 15 minutes after ignition. #The mercury burner cannot be reignited until the mercury vapor has cooled down and liquefied. Before re-igniting
a mercury burner, wait for about 10 minutes after the last time it was turned off.
}For the shake of safety, the power supply to the lamp housing is shut down if the lamp housing is opened while the burner
is on. If this happens, set the main switch to “ ” (OFF), wait for more than 10 minutes, then set it again to “ I ” (ON). Do not open the lamp housing until it has cooled down enough.
#To reset the hour counter, hold its reset button till “000.0” is displayed.
11
@
²
Fig. 1
³
5 Centering the Field Iris Diaphragm
1. Close the light path by sliding the shutter knob @ to position marked {.
2. Engage the B or IB mirror unit in the light path by rotating the turret. (If these mirror units are not available, engage another fluorescence mir­ror unit in the light path.)
3. Open the light path by sliding the shutter knob to position marked \.
4 Engage the 10X objective in the light path, place the specimen on the
stage and bring the image into approximate focus.
5. Pull out the field iris diaphragm knob ² to minimize the field iris diameter.
6. Fit the Allen wrench provided with the microscope frame in the two field iris centering screws ³ and adjust so that the iris image comes at the center of the field of view.
7. While pushing in the field iris diaphragm knob ², enlarge the field iris diaphragm until the field iris image inscribes the field of view. If eccentric­ity is found after this, try centering again.
8. Enlarge the iris diaphragm until the iris image becomes almost the same size as (i.e. circumscribes) the field of view.
Effects of Field Iris Diaphragm
The field iris diaphragm restricts the diameter of the beam of light enter­ing the objective and thus excludes extraneous light, improving image contrast. The field iris diaphragm also functions to prevent color fading of fluorescent light in other part than the observed region. To exclude extra light, set the field iris diaphragm knob ² on the fluores­cence illuminator according to the objective power, so that the image of the field iris diaphragm just circumscribes the field of view.
(Fig. 1)
12
Loading...
+ 36 hidden pages