![](/html/05/0574/0574aaab82b67436ef948ab35bbee6a8946afcc115d2e83647fc4a8d01f22287/bg1.png)
Sensors for Real-Time Analysis
FOXY Formulation FOSPOR Formulation HIOXY Formulation
(general-purpose coating) (high-sensitivity for low levels
of O2)
(for hydrocarbon
environments)
O2% range: 0-20% 0-<1% 0-20.9% at 1 ATM
DO range (ppm): 0-10 ppm 0-0.5 ppm 0-8 ppm
-20-+80 ºC for probes -20-+80 ºC for probes 0-60 ºC
0-60 ºC for patches 0-60 ºC for patches NA
O2 resolution: 100-500 ppm 10 ppm 100-500 ppm
DO resolution: 4-20 ppb at room temp. 0.4-2 ppb at room temp. 4-20 ppb at room temp.
O2 accuracy: 5% of reading 5% of reading 5% of reading
DO accuracy: 5% of reading 5% of reading 5% of reading
Min. detectable level: 100-500 ppm 10-100 ppm 100-500 ppm
< 1 second in gas < 1 second in gas < 1 second in gas
~30 seconds with overcoating
in gas
~30 seconds with overcoating
in gas
~45 seconds in pure water
~45 seconds in pure water ~45 seconds in pure water
Response time:
Temperature range:
advanced sampling options and detection technology
Ocean Optics optical sensors provide a viable alternative to traditional chemical sensing
devices and consist of transducer materials applied to the tips of optical fibers or to
substrates such as patches or cuvettes. These indicator materials change optical properties
in response to specific analytes in their immediate environment, with formulations
available for a variety of oxygen and pH sensing needs.
Thanks to our developments in sensor detection technology and sampling options, you
can enjoy lower system prices and greater application versatility than ever before.
NeoFox Phase Measurement System
A new option for the detection part of our fluorescence-based optical sensor systems helps reduce costs, improve
system stability and make calibration easier for most oxygen sensing setups. NeoFox is a benchtop device for
measurement of fluorescence lifetime, phase and intensity, and is especially useful for applications where sensitivity to
drift and system stability are important. What's more, NeoFox is about half of the cost of our previous phasemeasurement system, and includes a self-calibration feature for improved electronic stability.
System
Specifications
www.oceanoptics.com | info@oceanoptics.com | +1 727-733-2447
NeoFox Interface
Starting under $3,000
Complete O2 Sensing Systems including NeoFox, general purpose
probe, software and accessories.
![](/html/05/0574/0574aaab82b67436ef948ab35bbee6a8946afcc115d2e83647fc4a8d01f22287/bg2.png)
New Option for Monitoring O2 in Packaging
The award-winning RedEye™ Oxygen Sensor Patch is a self-adhesive patch with oxygen indicator that attaches directly
inside packages and containers for non-invasive measurements. RedEye sizes vary from a few millimeters to several
centimeters, giving you a custom fit; in fact, RedEye's sensing material is versatile enough to be applied to substrates
that meet most any specification. Also, different RedEye sensing formulations are available to optimize your results for
every sample environment. Use RedEye for monitoring oxygen levels in the headspace of bioreactors and blood bags
and for measuring oxygen gas and dissolved oxygen non-intrusively during bioprocesses and in pharmaceutical
packaging such as blister packs for pills.
New Probes Expand O2 Sampling Options
With the addition of new probes for small samples and for biological materials,
we now offer nearly a dozen oxygen sensing probe options – everything from
slender probes for fine spatial resolution to rugged stainless steel probes for process environments. Our newest
options include a 200 micron polyimide probe – our smallest probe yet – for very small sample sizes, and an angled
probe for biological samples. Optional silicone coatings (both standard and medical-grade) can be applied to probes
to exclude ambient light, improve chemical resistance and eliminate refractive index effects.
Phenol Red pH Test Kits
Use our Phenol Red pH Test Kits to determine the pH level in solutions. Simply add
3 mL solution to the test cuvette, cap it, and shake the cuvette to disperse the dye.
Measure the absorbance value of the reactive color to determine pH levels. CHEM-
TEST works with any Ocean Optics spectrometer system that is configured
for absorbance.
Coating Options for OEMs and Product Developers
Our optical-sensor coating technologies are available to OEMs and product developers designing new products for
chemical-sensing applications. With our oxygen and pH indicator materials, clients can develop and manufacture a variety of
custom optical sensor devices including fiber optic probes, cuvettes, Petri dishes, microscope slides and more. Also, RedEye patches
are designed especially with OEMs in mind, where high-volume production runs can lower costs and monitor sample conditions more
efficiently than existing methods. Additional services include sensor coating development and the licensing of Ocean Optics proprietary
oxygen and pH coating technologies. Contact an Applications Scientist for details.
Starting at $45
RedEye Evalution Kit
Special Volume Discounts Available