Nihon Kohden ZM 940PA User Manual

Page 1
TRANSMITTER
ZM-940PA
0614-009881
Page 2
Model: ZM-940PA
Manual code no.: 0614-009881
Reader Comment Card
We welcome your comments about this manual. Your comments and suggestions help us
improve our manuals. Please circle the number for each of the following statements
Fax or send your completed comment card to:
Fax: +81 (3) 5996-8100
International Div., Sales Promotion Section, Nihon Kohden Corp., 1-31-4, Nishiochiai
Shinjuku-ku, Tokyo 161-8560, Japan
Strongly Agree 1 Disagree 4
Agree 2 Strongly Disagree 5
Neutral 3
cutting line
This manual is organized. 12345
I can find the information I want. 12345
The information is accurate. 12345
I can understand the instructions. 12345
The illustrations are appropriate and helpful. 12345
The manual length is appropriate. 12345
Comments:
Thank you for your cooperation. We appreciate it very much.
Name:
Occupation/Position:
Hospital/Company:
Address:
Phone:
Page 3
Contents
GENERAL HANDLING PRECAUTIONS.................................................................... i
WARRANTY POLICY ............................................................................................... iii
Equipment Authorization Requirement ................................................................... iii
EMC RELATED CAUTION ....................................................................................... iv
Conventions Used in this Manual and Instrument ................................................. vii
Warnings, Cautions and Notes ......................................................................... vii
Explanations of the Symbols in this Manual and Instrument ........................... viii
Introduction ....................................................................................................................1
Panel Description .......................................................................................................... 2
Front Panel ............................................................................................................... 2
Rear Panel ............................................................................................................... 3
Top Panel ................................................................................................................. 4
Bottom Panel ............................................................................................................ 4
LCD .......................................................................................................................... 6
Note on Parameter Settings .......................................................................................... 8
Important Safety Information ......................................................................................... 9
General .................................................................................................................... 9
Output Signal ........................................................................................................... 9
Preparation .................................................................................................................. 10
Installing (Replacing) Batteries ............................................................................. 10
WARNING and CAUTION for Battery Handling ............................................... 10
Battery Lifetime ................................................................................................. 10
Installing (Replacing) Batteries ........................................................................ 11
Situations Requiring Battery Replacement ..................................................... 12
Battery Condition Indication ............................................................................. 12
Turning the Transmitter On/Off ............................................................................... 13
Turning On the Power ....................................................................................... 13
Turning Off the Power ....................................................................................... 13
Check Items Before Use .................................................................................. 13
Check Items After Power On ............................................................................ 1 4
Check Items After Use ...................................................................................... 14
Changing the Transmitter Channel ............................................................................. 15
Changing Parameter Setup Settings .......................................................................... 16
Parameter Setup Setting List ................................................................................. 16
Displaying the PARAMETER SETUP Screen ....................................................... 17
Changing Settings ................................................................................................. 18
Operator's Manual ZM-940PA i
Page 4
SELECTABLE INTERVALS .............................................................................. 1 8
INITIAL INTERVAL ............................................................................................ 18
INITIAL CUFF PRESS ...................................................................................... 18
NIBP MODE AFTER STAT ................................................................................ 19
START/FINISH SOUND .................................................................................... 19
OLD NIBP DATA/AFTER .................................................................................. 19
INHIBIT SpO
DURING NIBP ........................................................................... 20
2
2ND PARAMETER ........................................................................................... 20
LEADS OFF DISPLAY ...................................................................................... 21
ECG ELECTRODE ........................................................................................... 21
Changing System Setup Settings ............................................................................... 22
System Setup Setting List ...................................................................................... 22
Displaying the SYSTEM SETUP Screen ............................................................... 22
Changing Settings ................................................................................................. 23
CHANNEL/TYPE .............................................................................................. 23
PRESSURE UNIT ............................................................................................ 23
LANGUAGE ...................................................................................................... 24
BRIGHTNESS .................................................................................................. 24
SYSTEM INITIALIZE ......................................................................................... 24
Attaching NIBP Cuff, Electrodes and SpO
Probe to the Patient ................................ 25
2
Attachment Examples ............................................................................................ 25
Attaching the NIBP Cuff ......................................................................................... 26
Selecting the NIBP Cuff .................................................................................... 26
Connecting the NIBP Cuff to the Transmitter .................................................... 29
Attaching the NIBP Cuff to the Patient ............................................................. 3 1
Attaching Electrodes .............................................................................................. 35
Selecting Electrode Lead ................................................................................. 35
Connecting the Electrode Lead to the Transmitter ........................................... 36
Selecting the Electrode Position ...................................................................... 36
Attaching Electrodes to the Patient and Connecting the Electrode Leads to
Disposable Electrodes ..................................................................................... 40
Electrode Position for Respiration Monitoring ................................................. 41
Attaching the SpO
Selecting the SpO Connecting the SpO
Probe ...................................................................................... 43
2
Probe ................................................................................ 43
2
Probe to the Transmitter ................................................ 46
2
Attaching the Probe to the Patient ................................................................... 4 6
Locking the Keys on the Transmitter ........................................................................... 48
Monitoring .................................................................................................................... 49
NIBP Monitoring ..................................................................................................... 49
Selecting the Initial Cuff Inflation Pressure ...................................................... 49
Selecting the Measurement Mode and Interval ............................................... 49
Measuring NIBP ............................................................................................... 50
ii Operator's Manual ZM-940PA
Page 5
Monitoring SpO2 during NIBP Measurement ................................................... 53
ECG and Respiration Monitoring .......................................................................... 53
Electrode Detachment ..................................................................................... 55
SpO
Monitoring ..................................................................................................... 56
2
SpO
and PR Display Order ............................................................................. 57
2
Monitoring SpO
during NIBP Measurement ................................................... 57
2
Display and Message List ........................................................................................... 59
Battery Indication ................................................................................................... 59
ECG/Respiration .................................................................................................... 59
SpO
....................................................................................................................... 60
2
NIBP ....................................................................................................................... 61
Troubleshooting........................................................................................................... 62
Transmitter ............................................................................................................. 62
ECG/Respiration .................................................................................................... 63
SpO
....................................................................................................................... 63
2
NIBP ....................................................................................................................... 64
Maintenance ................................................................................................................ 66
1. External Check ............................................................................................ 66
2. Transmitter Channel..................................................................................... 66
3. LCD Display ................................................................................................. 68
4. Key Operation .............................................................................................. 69
5. NIBP Cuff for Attaching Transmitter to Patient Arm ...................................... 70
Maintenance Check Sheet .................................................................................... 7 1
Repair Parts Availability Policy.................................................................................... 72
Lifetime and Disposal .................................................................................................. 7 3
Disposing of Used Batteries .................................................................................. 73
Battery Lifetime ................................................................................................. 73
Disposal ........................................................................................................... 73
Disposing of Electrodes, SpO
Probes and NIBP Cuffs ........................................ 73
2
Cleaning, Disinfection and Sterilization ...................................................................... 74
Transmitter and Electrode Leads........................................................................... 74
Cleaning ........................................................................................................... 74
Disinfection....................................................................................................... 74
SpO
Probe ............................................................................................................ 75
2
YP-943P/944P NIBP Cuffs ..................................................................................... 75
Cleaning ........................................................................................................... 75
Disinfection....................................................................................................... 75
Specifications .............................................................................................................. 76
Measuring Parameters..................................................................................... 76
Transmitting Data ............................................................................................. 76
Displayed Data................................................................................................. 76
ECG Measurement........................................................................................... 76
Operator's Manual ZM-940PA iii
Page 6
Respiration Measurement ............................................................................... 76
SpO
Measurement .......................................................................................... 76
2
NIBP Measurement .......................................................................................... 77
Pulse Rate ........................................................................................................ 77
Transmitter ........................................................................................................ 77
Power Requirements ........................................................................................ 77
Dimension and Weight..................................................................................... 77
Environment ..................................................................................................... 78
Safety Standards .............................................................................................. 78
Electromagnetic Compatibility ......................................................................... 79
Electromagnetic Emissions .............................................................................. 79
Electromagnetic Immunity ................................................................................ 80
Recommended Separation Distance between Portable and Mobile RF
Communications Equipment............................................................................ 82
System Composition for EMC Test ................................................................... 82
Standard Accessories ................................................................................................. 83
Options ........................................................................................................................ 84
Transmitter ........................................................................................................ 84
ECG/RESP ....................................................................................................... 84
NIBP ................................................................................................................. 84
SpO
................................................................................................................. 85
2
Transmission Frequencies .......................................................................................... 86
iv Operator's Manual ZM-940PA
Page 7
GENERAL HANDLING PRECAUTIONS
This device is intended for use only by qualified medical personnel. Use only Nihon Kohden approved products with this device. Use of non-approved products or in a non-approved manner may affect the performance specifications of the device. This includes, but is not limited to, batteries, recording paper, pens, extension cables, electrode leads, input boxes and AC power.
Please read these precautions thoroughly before attempting to operate the instrument.
1. To safely and effectively use the instrument, its operation must be fully understood.
2. When installing or storing the instrument, take the following precautions:
(1) Avoid moisture or contact with water, extreme atmospheric pressure, excessive
humidity and temperatures, poorly ventilated areas, and dust, saline or sulphuric air.
(2) Place the instrument on an even, level floor. Avoid vibration and mechanical shock,
even during transport.
(3) Avoid placing in an area where chemicals are stored or where there is danger of gas
leakage.
(4) The power line source to be applied to the instrument must correspond in frequency
and voltage to product specifications, and have sufficient current capacity.
(5) Choose a room where a proper grounding facility is available.
3. Before Operation
(1) Check that the instrument is in perfect operating order. (2) Check that the instrument is grounded properly. (3) Check that all cords are connected properly. (4) Pay extra attention when the instrument is in combination with other instruments to
avoid misdiagnosis or other problems. (5) All circuitry used for direct patient connection must be doubly checked. (6) Check that battery level is acceptable and battery condition is good when using battery-
operated models.
4. During Operation
(1) Both the instrument and the patient must receive continual, careful attention. (2) Turn power off or remove electrodes and/or transducers when necessary to assure the
patient’s safety. (3) Avoid direct contact between the instrument housing and the patient.
5. To Shutdown After Use
(1) Turn power off with all controls returned to their original positions. (2) Remove the cords gently; do not use force to remove them.
Operator's Manual ZM-940PA i
Page 8
(3) Clean the instrument together with all accessories for their next use.
6. The instrument must receive expert, professional attention for maintenance and repairs. When the instrument is not functioning properly, it should be clearly marked to avoid operation while it is out of order.
7. The instrument must not be altered or modified in any way.
8. Maintenance and Inspection:
(1) The instrument and parts must undergo regular maintenance inspection at least every 6
months.
(2) If stored for extended periods without being used, make sure prior to operation that the
instrument is in perfect operating condition.
(3) Technical information such as parts list, descriptions, calibration instructions or other
information is available for qualified user technical personnel upon request from your Nihon Kohden distributor.
9. When the instrument is used with an electrosurgical instrument, pay careful attention to the application and/or location of electrodes and/or transducers to avoid possible burn to the patient.
10. When the instrument is used with a defibrillator, make sure that the instrument is protected against defibrillator discharge. If not, remove patient cables and/or transducers from the instrument to avoid possible damage.
ii Operator's Manual ZM-940PA
Page 9
WARRANTY POLICY
Nihon Kohden Corporation (NKC) shall warrant its products against all defects in materials and workmanship for one year from the date of delivery. However, consumable materials such as recording paper, ink, stylus and battery are excluded from the warranty.
NKC or its authorized agents will repair or replace any products which prove to be defective during the warranty period, provided these products are used as prescribed by the operating instructions given in the operator’s and service manuals.
No other party is authorized to make any warranty or assume liability for NKC’s products. NKC will not recognize any other warranty, either implied or in writing. In addition, service, technical modification or any other product change performed by someone other than NKC or its authorized agents without prior consent of NKC may be cause for voiding this warranty.
Defective products or parts must be returned to NKC or its authorized agents, along with an explanation of the failure. Shipping costs must be pre-paid.
This warranty does not apply to products that have been modified, disassembled, reinstalled or repaired without Nihon Kohden approval or which have been subjected to neglect or accident, damage due to accident, fire, lightning, vandalism, water or other casualty, improper installation or application, or on which the original identification marks have been removed.
In the USA and Canada other warranty policies may apply.
CAUTION
United States law restricts this device to sale by or on the order of a physician.
Equipment Authorization Requirement
Operation of this equipment requires the prior coordination with a frequency coordinator designated by FCC for the Wireless Medical Telemetry Service.
Operator's Manual ZM-940PA iii
Page 10
EMC RELATED CAUTION
This equipment and/or system complies with IEC 60601-2 International Standard for electromagnetic compatibility for medical electrical equipment and/or system. However, an electromagnetic environment that exceeds the limits or levels stipulated in IEC 60601-1-2, can cause harmful interference to the equipment and/or system or cause the equipment and/or system to fail to perform its intended function or degrade its intended performance. Therefore, during the operation of the equipment and/or system, if there is any undesired deviation from its intended operational performance, you must avoid, identify and resolve the adverse electromagnetic effect before continuing to use the equipment and/or system.
The following describes some common interference sources and remedial actions:
1.Strong electromagnetic interference from a nearby emitter source such as an
authorized radio station or cellular phone: Install the equipment and/or system at another location. Keep the emitter source such as cellular phone away from the equipment and/or system, or turn off the cellular phone.
2.Radio-frequency interference from other equipment through the AC power
supply of the equipment and/or system: Identify the cause of this interference and if possible remove this interference source. If this is not possible, use a different power supply.
3.Effect of direct or indirect electrostatic discharge:
Make sure all users and patients in contact with the equipment and/or system are free from direct or indirect electrostatic energy before using it. A humid room can help lessen this problem.
4.Electromagnetic interference with any radio wave receiver such as radio or
television: If the equipment and/or system interferes with any radio wave receiver, locate the equipment and/or system as far as possible from the radio wave receiver.
5.Interference of lightning
When lightning occurs near the location where the equipment and/or system is installed, it may induce an excessive voltage in the equipment and/or system. In such a case, disconnect the AC power cord from the equipment and/or system and operate the equipment and/or system by battery power, or use an uninterruptible power supply.
iv Operator's Manual ZM-940PA
Page 11
6.Use with other equipment When the equipment and/or system is adjacent to or stacked with other equipment, the equipment and/or system may affect the other equipment. Before use, check that the equipment and/or system operates normally with the other equipment.
7.Use of unspecified accessory, transducer and/or cable When an unspecified accessory, transducer and/or cable is connected to this equipment and/or system, it may cause increased electromagnetic emission or decreased electromagnetic immunity. The specified configuration of this equipment and/or system complies with the electromagnetic requirements with the specified configuration. Only use this equipment and/or system with the specified configuration.
8.Use of unspecified configuration When the equipment and/or system is used with the unspecified system configuration different than the configuration of EMC testing, it may cause increased electromagnetic emission or decreased electromagnetic immunity. Only use this equipment and/or system with the specified configuration.
9.Measurement with excessive sensitivity The equipment and/or system is designed to measure bioelectrical signals with a specified sensitivity. If the equipment and/or system is used with excessive sensitivity, artifact may appear by electromagnetic interference and this may cause mis-diagnosis. When unexpected artifact appears, inspect the surrounding electromagnetic conditions and remove this artifact source.
If the above suggested remedial actions do not solve the problem, consult your Nihon Kohden distributor or representative for additional suggestions.
This equipment complies with International Standard IEC 60601-1-2 (1993) which requires CISPR11, Group 1, Class B. Class B EQUIPMENT is equipment suitable for use in domestic establishments and in establishments directly connected to a low voltage power supply network which supplies buildings used for domestic purposes.
Operator's Manual ZM-940PA v
Page 12
This page is intentionally left blank.
vi Operator's Manual ZM-940PA
Page 13
Conventions Used in this Manual and Instrument
Warnings, Cautions and Notes
Warnings, cautions and notes are used in this manual to alert or signal the reader to specific information.
WARNING
A warning alerts the user to the possible injury or death associated with the use or misuse of the instrument.
CAUTION
A caution alerts the user to possible injury or problems with the instrument associated with its use or misuse such as instrument malfunction, instrument failure, damage to the instrument, or damage to other property.
NOTE
A note provides specific information, in the form of recommendations, prerequirements, alternative methods or supplemental information.
Operator's Manual ZM-940PA vii
Page 14
Explanations of the Symbols in this Manual and Instrument
The following symbols found in this manual/instrument bear the respective descriptions as given.
On Panel
Symbol Description Symbol Description
Defibrillation proof type BF applied part
Defibrillation proof type CF applied part
Attention, consult operator’s manual
Direction for attaching battery cover
Inside Battery Case
Direct current
Serial number
Year of manufacture
CSA mark
Symbol Description Symbol Description
Battery position
On LCD
Attention, consult operator’s manual
Symbol Description Symbol Description
Full battery
Replace battery NIBP cannot be measured
Replace battery Check electrode
viii Operator's Manual ZM-940PA
Page 15
Introduction
The ZM-940PA transmitter transmits ECG, respiration, SpO2, NIBP and pulse waveform from a patient to a Nihon Kohden monitor for continuous monitoring. The transmitter can change channels when connected to the QI-901PK channel writer. The front LCD displays SpO NIBP, pulse rate, pulse waveform amplitude, electrode condition mark, battery condition and NIBP measuring mode and interval.
Read the operator’s manual for the receiving monitor together with this manual before operation.
CAUTION
• Do not use the same channel for different patients. Otherwise, two patients’ data will be lost due to mutual modulation interference, or another patient’s data may appear on the receiving monitor screen.
• Do not use transmitters of adjacent channels in a hospital. Otherwise, radio waves from one transmitter affect the receiver of the adjacent channel’s transmitter and there may be interference.
• Do not use the same transmitter on more than one patient at the same time. Do not connect different sensors on different patients to the same transmitter.
NOTE
• To prevent interference between channels, assign a channel administrator in the hospital and only he or she should manage channel assignment.
• Use Nihon Kohden parts and accessories to assure maximum performance from your instrument.
• For stable signal reception, it is recommended to use a diversity antenna system on the receiving monitor. Otherwise, spike noise from transient fading of electric field strength (for example, people moving) may interfere with the transmitter signal and may be mistaken as an arrhythmia on the receiving monitor.
• NIBP cannot be measured on a neonate. (ECG, respiration and SpO monitored on a neonate.)
can be
2
%,
2
Operator's Manual ZM-940PA 1
Page 16
Panel Description
Front Panel
LCD:
Displays measuring data, settings and other information.
NIBP START/STOP key:
Starts/stops NIBP measurement in selected mode.
NIBP INTERVAL key:
Selects NIBP measurement mode.
Refer to the WARNING below. (This symbol is attached to the rear of the battery case cover.)
Battery case:
Contains three 1.5 V AA batteries.
WARNING
Close the battery case cover during operation. If the transmitter is used with the battery case cover open, the patient may get an electrical shock when defibrillation is performed, and electrostatic discharge by the patient may intermittently interfere with the waveform or data.
2 Operator's Manual ZM-940PA
Page 17
Rear Panel
Lock plate:
Fastens the transmitter to an NIBP cuff.
Refer to the WARNING below.
WARNING
This transmitter is not waterproof. If detergent or liquid spills into the transmitter, stop using it and contact your Nihon Kohden distributor. If a wet transmitter is used, the patient or anyone in contact with the transmitter may receive an electric shock or patient leakage current over the allowed amount may flow.
Operator's Manual ZM-940PA 3
Page 18
Top Panel
Refer to the WARNING on the next page.
Refer to the WARNING on the next page.
NIBP socket:
Connects the cuff hose.
Refer to the symbol page.
Bottom Panel
Channel number label:
Indicates the channel number of the transmitter. Attach the channel number label to the panel of the monitor.
Refer to the symbol page.
ECG/impedance RESP socket:
Connects the electrode lead for measuring ECG and/or respiration by the impedance method.
Refer to the symbol page.
SpO
socket:
2
Connects the SpO
2
probe.
Refer to the WARNING on the next page.
4 Operator's Manual ZM-940PA
Page 19
WARNING
• Before performing defibrillation, check that the electrode leads and SpO probe attached to the patient are properly connected to the transmitter. Touching the metal parts of disconnected leads and probes causes serious electrical shock or injury by discharged energy.
• When performing defibrillation, all persons must keep clear of the bed and must not touch the patient, any equipment connected to the patient or the metal parts of leads and probes connected to the patient. Failure to follow this warning may result in serious electrical burn, shock or other injury.
• When performing defibrillation, discharge as far as possible from electrodes and medicine on the chest of the patient. If there is a possibility that the defibrillator paddle could touch electrodes and medicine, remove electrodes and medicine from the patient. If the defibrillator contacts these materials, the discharged energy may cause serious electrical burn on the patient.
• When using this transmitter with an ESU, the ESU return plate and the electrodes for monitoring must be firmly attached to the patient. If the return plate is not attached correctly, it may burn the patient’s skin where the electrodes are attached. Refer to the instruction manual for the ESU.
2
CAUTION
Do not shake or swing the transmitter holding the leads/cables connected to the transmitter. The transmitter may come off and injure someone or damage surrounding instruments.
Operator's Manual ZM-940PA 5
Page 20
LCD
1234
6
5
10
LOCK KEYS
11
9
8
No. Name Description
1 NIBP SYS Displays NIBP systolic value. 2 NIBP DIA Displays NIBP diastolic value. 3 NIBP MEAN Displays NIBP mean value.
“CUFF” is displayed with the cuff inflation pressure during measurement.
4 Check electrode mark Appears when an electrode or electrode lead becomes
detached during ECG measurement.
5 Battery replacement mark Appears when the batteries are weak. For details, refer to
the “Battery Condition Indication” section.
6 Message display area Displays messages.
When ECG is monitored with 6 electrodes and an electrode or electrode lead is detached, “Check electrode” is indicated as below, depending on the PARAMETER SETUP setting. Refer to the “Changing Parameter Setup Settings” and “ECG and Respiration Monitoring” sections.
7
LEADS OFF DISPLAY set to CHAR ECG ELECTRODE set to AHA
LEADS OFF DISPLAY set to IMAGE
6 Operator's Manual ZM-940PA
LEADS OFF DISPLAY set to CHAR ECG ELECTRODE set to IEC
Page 21
No. Name Description
7 NIBP measurement mode Displays NIBP measurement mode. When set to auto mode,
the measurement interval is displayed.
8 NIBP interval bar graph In auto NIBP measurement, the remaining time from the last
measurement to the next measurement is displayed as a bar graph.
9 Pulse level bar graph Displays pulse level in 7 steps.
10 %SpO
2
11 PR Displays pulse rate when NIBP or SpO
Displays SpO2 data.
the SpO
probe is attached to the patient, the real time pulse
2
rate is displayed. When the SpO
is measured. When
2
probe is not attached to the
2
patient, the pulse rate at the end of NIBP measurement is displayed.
Operator's Manual ZM-940PA 7
Page 22
Notes on Parameter Settings
When monitoring NIBP and SpO2, the following setting must be set as indicated in the table to properly transmit the monitoring data to the receiving monitor. Otherwise, SpO monitored properly during NIBP measurement.
Some receiving monitors require the software to be upgraded. For details, contact your Nihon Kohden distributor.
SpO2 probe attachment site INHIBIT SpO2 DURING NIBP setting
Probe attached to the same limb as the cuff ON Probe attached to the limb without cuff* OFF
* When the SpO2 probe is attached to the same limb as the NIBP cuff and the cuff is
inflated, the SpO
value becomes unstable and SpO2 or PR alarm may occur.
2
cannot be
2
8 Operator's Manual ZM-940PA
Page 23
Important Safety Information
General
WARNING
• Never use this transmitter in the presence of any flammable anesthetic gas or high concentration oxygen atmosphere. Failure to follow this warning may cause explosion or fire.
• Never use this transmitter in a high-pressure oxygen medical care tank. Failure to follow this warning may cause explosion or fire.
• Do not take this transmitter into the MRI test room. This transmitter is not designed to be used during MRI tests.
Output Signal
WARNING
Do not use the output signal from the receiving monitor as the synchronization signal for other equipment such as IABP, MRI, echocardiography or defibrillator because there may be time delay between the monitor and the other equipment caused by waveform transmission delay and spike noise may interfere on the output signal and be mistaken as a trigger.
Operator's Manual ZM-940PA 9
Page 24
Preparation
Installing (Replacing) Batteries
WARNING and CAUTION for Battery Handling
WARNING
• Do not dispose of the battery in fire because it may explode.
• Do not use a disassembled or damaged battery. The contents of the battery are harmful and the battery may catch fire.
• If the contents of the battery contact the skin or clothes, immediately wash it thoroughly with water.
• Never short-circuit the + and – terminals. The battery may overheat and catch fire.
• Take care that the patient does not swallow batteries.
CAUTION
• Refer to the battery and battery charger manuals for details on handling the batteries.
• Do not handle the batteries with wet hands.
• When the transmitter is not in use, remove batteries. When batteries are installed, battery power is consumed even if measurement is not performed. Especially, when NiMH batteries remain in the transmitter when the transmitter is not in use, the battery may become unusable from overdischarge and leak liquid which will damage the transmitter.
• The battery charger must be used outside the patient environment.
Battery Lifetime
Use three AA type alkaline dry cell batteries. NiMH rechargeable batteries can also be used.
Type Lifetime (Measuring parameters)
ECG, SpO2, NIBP ECG, SpO
2
ECG only
NiMH secondary 2 days 2 days 2.5 days Alkaline primary 1 day 2.5 days 3 days
The above data is when new batteries are used at room temperature, NIBP is measured in auto mode at 60 minute intervals and SpO
is measured on an index finger of a male patient with
2
weight 60 kg. Operation time depends on the thickness of SpO
10 Operator's Manual ZM-940PA
probe attachment site.
2
Page 25
NOTE
When using rechargeable NiMH batteries, shallow charging/discharging shortens battery capacity. For details, refer to the battery operator’s manual.
Installing (Replacing) Batteries
CAUTION
Battery replacement must be performed by medical staff. When replacing batteries of the transmitter currently used for a patient, disconnect electrode leads from the transmitter before replacing batteries or do not touch the patient during replacement.
If electrode leads are attached to the patient and a person replacing batteries touches the patient during battery replacement, patient leakage current over the allowed amount may flow.
CAUTION
• Replace all batteries at the same time.
• Do not use different types of batteries together.
NOTE
Insert the batteries with the correct polarity (+ and –).
Procedure
1. Remove the battery case cover.
2. Insert three new or fully charged batteries into the battery case observing the correct polarity.
Operator's Manual ZM-940PA 11
Page 26
3. Close the cover.
NOTE
Remove the batteries before disposing of the transmitter.
Situations Requiring Battery Replacement
Replace the batteries when any of the following occurs.
• The transmitter LCD displays the “
• The transmitter generates a constant alarm (continuous “peep” sound).
• The transmitter LCD does not display anything when the power is turned on.
• The monitor displays the battery replacement message on the screen.
Battery Condition Indication
The battery condition is indicated as follows.
Indication Condition Receiving Monitor
Fully charged battery
Batteries are low. Replace batteries. Batteries are low. NIBP cannot be measured. Replace batteries.
No
indication
Dead batteries
” or “ ”mark.
Batteries are full. There is no indication on the monitor.
Message requiring battery replacement is displayed.
No signal can be transmitted to the monitor. There is no indication on the monitor.
12 Operator's Manual ZM-940PA
Page 27
Turning the Transmitter On/Off
Turning On the Power
Peep
Check Items Before Use
Before turning on the transmitter power, check the following to confirm that the transmitter can be used in normal and safe condition.
When the batteries are installed correctly, the power is turned on. A one second “peep” sounds and the startup screen appears. (There is no “peep” sound when there is no battery power.)
NOTE
Replace the batteries when the LCD displays the “
Turning Off the Power
To turn off the power, remove batteries.
” or “ ”mark.
Appearance
• There are no damaged or dirty parts on the outside of the transmitter (LCD, keys, sockets, battery case cover, battery case, lock plate, etc.).
• The transmitter is completely dry.
• The electrode lead, SpO
• There are no damaged or dirty parts on the disposable SpO
probe and NIBP cuff are not broken.
2
probe, disposable electrodes or
2
disposable NIBP cuff.
Batteries
• The battery polarity is correct.
• The battery case spring is firmly fixed and the battery is not loose.
• The battery case cover is firmly closed.
Channel Setting
• The transmitter channel corresponds to those of the receiving monitor.
• There is no transmitter in the surrounding area with the same channel.
Operator's Manual ZM-940PA 13
Page 28
Check Items After Power On
After turning on the power, check the following.
Power On
• The transmitter generates about a one second “peep” sound and the startup screen appears.
• The transmitter does not generate a continuous “peep” sound.
• The transmitter does not give excessive heat.
• The transmitter LCD displays a “
” mark.
• The transmitter does not interfere with the operation of other medical instruments in use.
Basic Operation
• The “signal loss” message is not displayed on the receiving monitor when the transmitter is inside the receiving range of the monitor.
• The battery replacement message is not displayed on the monitor.
• The keys on the transmitter function properly.
• The LCD brightness is appropriate. To adjust brightness, refer to the “Changing System Setup Settings” section.
Check Items After Use
To use the transmitter in safe and optimum condition for next time, check the following.
Before Turning Power Off
• Temporarily changed settings are changed back to the previous settings.
• There was no malfunction on the transmitter.
Storage
• ECG electrode leads, SpO
probe and NIBP cuff are cleaned and disinfected.
2
• When the transmitter gets wet, liquid is wiped off and the transmitter is thoroughly dried.
• There are enough consumables, such as disposable electrodes.
• The transmitter power is turned off by removing batteries from the transmitter.
• Dead batteries are disposed of properly.
14 Operator's Manual ZM-940PA
Page 29
Changing the Transmitter Channel
The channel of the transmitter can be changed. The optional QI-901PK Channel Writer is required.
WARNING
The following action must be taken to properly receive the transmitter signal of the correct patient on the receiving instrument. Otherwise, there may be signal loss or signals may mix causing a serious accident, such as monitoring a different patient.
• Assign a channel administrator in the hospital and only he or she should manage channel assignment on his or her responsibility.
• The channel administrator must manage the channels in the facility so that there is no signal interference.
• When the transmitter channel is changed, the channel administrator must check that the channel on the receiving monitor is also changed and the signal is properly received.
• The channel administrator must replace the channel number label on the transmitter with the new one after changing the channel.
NOTE
The software version of the QI-901PK channel writer must be 02-01 or later to change the channel on the ZM-940PA transmitter.
To check the transmitter channel, refer to “CHANNEL” in the “Changing System Setup Settings” section.
Operator's Manual ZM-940PA 15
Page 30
Changing Parameter Setup Settings
The initial settings on the PARAMETER SETUP screen must be changed before monitoring. Changing these settings during monitoring interrupts monitoring.
Parameter Setup Setting List
The factory default settings are underlined.
Setting Item Description Settings
SELECTABLE INTERVALS
INITIAL INTERVAL
INITIAL CUFF PRESS
NIBP MODE AFTER STAT START/FINISH SOUND
OLD NIBP DATA AFTER
INHIBIT SpO
2
DURING NIBP 2ND PARAMETER Set SpO2 and PR display order. SpO2, PR
LEADS OFF DISPLAY
ECG ELECTRODE
Select the NIBP measurement modes for the mode selection.
Select the initial NIBP measurement mode at power on.
Select the NIBP cuff inflation pressure.
Select the NIBP measurement mode after completing STAT measurement. Turn ON or OFF the sound for NIBP measurement start/finish. Select whether to hide or dim the NIBP data after measurement and how long to wait after measurement to dim or hide it. Turn SpO
2
monitoring on or off during
NIBP measurement.
Select the mode for displaying electrode off. This setting is only available when ECG is monitored with 6 electrodes. Select the electrode lead type. This setting is only available when CHAR is selected for LEADS OFF DISPLAY.
STAT, 5, 10, 15, 30, 60, 120, 240 MAN., 5 min, 10 min, 15 min, 30 min, 60 min, 120 min, 240 min 120 mmHg, 150 mmHg, 180 mmHg, 210 mmHg, 240 mmHg MAN., 5 min, 10 min, 15 min, 30 min
ON, OFF/ON, OFF
DATA: HIDE, DIM AFTER: 5 min, 10 min, 30 min
ON, OFF
CHAR, IMAGE
IEC, AHA
16 Operator's Manual ZM-940PA
Page 31
Displaying the PARAMETER SETUP Screen
1. Remove one battery.
2. While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery.
The SETUP screen appears.
3. Press the NIBP START/STOP key to enter the PARAMETER SETUP screen.
When the cursor is moved to “EXIT” by pressing the NIBP INTERVAL key and the NIBP START/STOP key is pressed, the startup screen appears, then the monitoring screen appears.
4. To select or change a setting, press the NIBP START/STOP key.
To move the cursor, press the NIBP INTERVAL key.
Selects or changes setting
Moves cursor
When the cursor is moved to “RETURN” by pressing the NIBP INTERVAL key and the NIBP START/STOP key is pressed, the SETUP screen appears.
Operator's Manual ZM-940PA 17
Page 32
Changing Settings
SELECTABLE INTERVALS
During monitoring, when the NIBP INTERVAL key is pressed, the measurement mode changes according to the modes selected in this item. MANUAL mode is already selected for the mode selection.
1. Press the NIBP INTERVAL key to move the cursor to the desired mode.
2. Press the NIBP START/STOP key to select or unselect the mode. Selectable modes are: STAT, 5, 10, 15, 30, 60, 120 and 240 min.
INITIAL INTERVAL
Select the initial NIBP measurement mode at power on.
1. Press the NIBP INTERVAL key to move the cursor to “INITIAL INTERVAL”.
2. Press the NIBP START/STOP key to select the mode. Selectable modes are the modes selected for “SELECTABLE INTERVALS” and “MAN.” (MANUAL).
INITIAL CUFF PRESS
Select the NIBP cuff inflation pressure.
1. Press the NIBP INTERVAL key to move the cursor to “INITIAL CUFF PRESS”.
2. Press the NIBP START/STOP key to select the inflation pressure from 120, 150, 180, 210 and 240 mmHg.
18 Operator's Manual ZM-940PA
Page 33
NIBP MODE AFTER STAT
Select the NIBP measurement mode after completing the STAT measurement.
1. Press the NIBP INTERVAL key to move the cursor to “NIBP MODE AFTER STAT”.
2. Press the NIBP START/STOP key to select the mode. The selected mode is automatically selected for “SELECTABLE INTERVALS” as well.
START/FINISH SOUND
Turn on or off the sound for NIBP measurement start and finish.
Start sound
End sound
1. Press the NIBP INTERVAL key to move the cursor to “START/FINISH SOUND”.
2. Press the NIBP START/STOP key to turn ON or OFF.
OLD NIBP DATA/AFTER
Select whether to dim or hide the NIBP data after measurement and how long to wait after NIBP measurement to dim or hide it.
1. Press the NIBP INTERVAL key to move the cursor to “OLD NIBP DATA/AFTER”.
2. Press the NIBP START/STOP key to select the setting.
DATA: DIM NIBP data is dimmed after the “AFTER” interval.
HIDE NIBP data is hidden after the “AFTER” interval. “– – –” is displayed on the
screen.
AFTER: Select the interval after NIBP measurement to dim or hide.
Dimmed Hidden
Operator's Manual ZM-940PA 19
Page 34
INHIBIT SpO2 DURING NIBP
Set whether or not to monitor SpO
during NIBP measurement.
2
When the SpO OFF, the pulse may become unstable and SpO set this setting to ON so that SpO
When the SpO
probe is attached to the same limb as the NIBP cuff and this setting is set to
2
or PR alarm may occur. It is recommended to
2
is not measured during NIBP measurement.
2
probe is attached to the other limb from the NIBP cuff, this setting can be set to
2
OFF.
NOTE
When this “INHIBIT SpO2 DURING NIBP” is set to OFF, refer to the “Monitoring SpO
during NIBP Measurement” section.
2
1. Press the NIBP INTERVAL key to move the
ON: Stops SpO OFF: SpO
monitoring during NIBP measurement.
2
is monitored during NIBP measurement.
2
2ND PARAMETER
Set the display order of SpO
and PR.
2
cursor to “INHIBIT SpO
2. Press the NIBP START/STOP key to select “ON” or “OFF”.
DURING NIBP”.
2
1. Press the NIBP INTERVAL key to move the cursor to “2ND PARAMETER”.
2. Press the NIBP START/STOP key to select “SpO
” or “PR”.
2
When set to SpO
20 Operator's Manual ZM-940PA
2
When set to PR
Page 35
LEADS OFF DISPLAY
Select the mode for displaying electrode off. This setting is only available when ECG is monitored with 6 electrodes.
1. Press the NIBP INTERVAL key to move the cursor to “LEADS OFF DISPLAY”.
2. Press the NIBP START/STOP key to select “CHAR” or “IMAGE”.
When set to CHAR
When set to IMAGE
ECG ELECTRODE
Select the electrode lead type. This setting is only available when “CHAR” is selected for LEADS OFF DISPLAY.
1. Press the NIBP INTERVAL key to move the cursor to “ECG ELECTRODE”.
2. Press the NIBP START/STOP key to select “IEC” or “AHA”.
AHA: RA, LA, LL, Va, Vb IEC: R, L, F, Ca, Cb
Operator's Manual ZM-940PA 21
Page 36
Changing System Setup Settings
NOTE
Changing System Setup settings must be done only by a qualified personnel.
System Setup Setting List
The factory default settings are underlined.
Setting Item Description Settings
CHANNEL Displays the transmitter channel. PRESSURE UNIT Select the units for NIBP. mmHg, kPa LANGUAGE Select the language for screen display. JPN, ENG BRIGHTNESS Select the LCD brightness. 1, 2, 3, 4 SYSTEM INITIALIZE
Displaying the SYSTEM SETUP Screen
1. Remove one battery.
2. While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery. The SETUP screen appears.
Initializes all settings to the factory default settings.
3. Press the NIBP INTERVAL key to move the cursor to “SYSTEM SETUP”.
4. Press the NIBP START/STOP key to enter the SYSTEM SETUP screen.
When the cursor is moved to “EXIT” by pressing the NIBP INTERVAL key and the NIBP START/STOP key is pressed, the startup screen appears, then the monitoring screen appears.
22 Operator's Manual ZM-940PA
Page 37
5. To select or change a setting, press the NIBP START/STOP key. To move the cursor, press the NIBP INTERVAL key.
Selects or changes setting
Moves cursor
When the cursor is moved to “RETURN” by pressing the NIBP INTERVAL key and the NIBP START/STOP key is pressed, the SETUP screen appears.
Changing Settings
CHANNEL
The channel of this transmitter is displayed.
Channel of this transmitter
PRESSURE UNIT
Select the unit for NIBP.
1. Press the NIBP INTERVAL key to move the cursor to “PRESSURE UNIT”.
2. Press the NIBP START/STOP key to select “mmHg” or “kPa”.
Operator's Manual ZM-940PA 23
Page 38
LANGUAGE
Select the language for screen display.
1. Press the NIBP INTERVAL key to move the cursor to “LANGUAGE”.
2. Press the NIBP START/STOP key to select the language.
BRIGHTNESS
Select the LCD brightness.
1. Press the NIBP INTERVAL key to move the cursor to “BRIGHTNESS”.
2. Press the NIBP START/STOP key to select the LCD brightness from 1 to 4.
1234
Light .............................. Dark
SYSTEM INITIALIZE
Do the following procedure to initialize the settings to the factory default settings.
1. Press the NIBP INTERVAL key to move the cursor to “SYSTEM INITIALIZE”.
2. Press the NIBP START/STOP key. The “EXECUTE” message appears.
3. Press the NIBP START/STOP key to initialize the settings to the factory default settings.
24 Operator's Manual ZM-940PA
Page 39
Attaching NIBP Cuff, Electrodes and SpO2 Probe
to the Patient
The transmitter can be attached to an arm of the patient or placed on the bedside. The required length of the electrode leads and SpO attached to the patient.
Monitoring SpO2 during NIBP Measurement When the SpO
probe is attached to the same limb as the NIBP cuff, the blood
2
flow decreases during NIBP measurement and pulse wave cannot be detected and SpO
cannot be monitored properly. When “INHIBIT SpO2 DURING NIBP” on
2
the PARAMETER SETUP screen is set to ON (factory default setting), SpO monitoring is paused during NIBP measurement to avoid SpO2 alarm occurrence. However, when monitoring SpO careful when reading SpO
probe cable depends on how the transmitter is to be
2
NOTE
on the same limb as NIBP, be
2
values.
2
2
When monitoring SpO
is important, attach the probe to the limb to which the
2
NIBP cuff or catheter is not attached.
Attachment Examples
When transmitter is attached on an arm
When placing the transmitter on a bedside, place it on a stable and flat place. If the transmitter falls off, it may be damaged.
Operator's Manual ZM-940PA 25
When transmitter is placed on a bedside
NOTE
Page 40
Attaching the NIBP Cuff
Selecting the NIBP Cuff
Select the NIBP cuff appropriate for the patient.
NOTE
NIBP cannot be measured on neonates using this transmitter.
Reusable Cuffs
When attaching the transmitter to the patient arm, a special NIBP cuff is required. An optional YN-990P extension hose (1.5 m) is available to extend the length between the NIBP socket on the transmitter and NIBP cuff (e.g. when not attaching the transmitter to the patient arm and placing the transmitter on a bedside).
Reusable cuff Model Width (cm) Air hose length (cm)
For adult
Standard YP-943P 13 15
Large YP-944P 15 15
Air hose
Width
When not attaching the transmitter to the patient arm, the following cuffs can be used. To use these cuffs, an optional YN-990P extension hose (1.5 m) is required.
Reusable cuff Model Width (cm) Air hose length (cm)
For infants YP-960T 5
For children
For adults
26 Operator's Manual ZM-940PA
Small YP-961T 7
Standard YP-962T 10
Standard YP-963T 13
Large YP-964T 15
Width
Air hose
15
Page 41
Disposable Cuffs
CAUTION
The disposable cuffs are not sterilized. If necessary, sterilize the disposable cuffs using glutaraldehyde solution by following the instructions for the glutaraldehyde.
When not attaching the transmitter to the patient arm, the following disposable cuffs can be used. To use these cuffs, an optional YN-990P extension hose (1.5 m) is required.
Reusable cuff Model Width (cm) Air hose length (cm)
For infants YP-910P 6
For children YP-912P 9
For adults
Small YP-913P 12
Standard YP-914P 14
Large YP-915P 16
Width
Air hose
20
Extension Hose
CAUTION
When using an extension hose, check that the extension hose is not bent or squeezed. Otherwise, the cuff may not inflate or deflate. If the cuff cannot deflate, it may cause congestion on the patient at the cuff attachment site.
YN-990P extension hose, 150 cm
Operator's Manual ZM-940PA 27
Page 42
Reference for selecting a cuff
The AHA (American Heart Association) recommends that the cuff width be 40% of the
circumference of the upper arm. Refer to the following graph and select the cuff which suits the
patient’s arm.
NOTE
• If a range of arm circumference appropriate for the cuff is prescribed, use a cuff within that range.
• To obtain accurate measured values, select a wide cuff which can be attached to the upper arm. Measuring with a very narrow cuff may result in measured values higher than the actual values.
• The YP-943P NIBP cuff is for standard size adult. Do not use this cuff when it does not fit the patient.
Cuff Width and Arm Circumference
Reusable Cuffs
Cuff width (cm)
15
10
Children small YP-961T
5
Infants YP-960T
Adults large YP-944P (cuff for transmitter)
YP-964T
Adults standard YP-943P (cuff for transmitter)
YP-963T
Children standard YP-962T
Disposable Cuffs
Cuff width (cm)
0
20
15
10
5
0
10 20 30
Children standard YP-912P
Infants YP-910P
10 20 30
40
Adults large YP-915P
Adults standard YP-914P
Adults small YP-913P
40
50
Arm circumference (cm)
50 60
Arm circumference (cm)
60
28 Operator's Manual ZM-940PA
Page 43
Connecting the NIBP Cuff to the Transmitter
When Using YP-943P/944P NIBP Cuff
To attach the YP-943P/944P NIBP cuff to the transmitter, the lock plate is required.
YP-943P/944P NIBP cuff
Lock plate
Front cover
Belt for the strap
Lock plate pocket
Top tab
For attaching the NIBP cuff to the transmitter
Air hose
Front cover open
NOTE
Do not roll up or put weight on the cuff with the lock plate attached to it. Otherwise, the lock plate may break.
D ring
Belt
Bottom tab
Operator's Manual ZM-940PA 29
Page 44
1. Remove the lock plate from the transmitter.
2. Insert the lock plate into the lock plate pocket on the NIBP cuff.
3. Attach the transmitter to the lock plate by inserting the tabs on the lock plate into the slots on the transmitter.
4. Cover the transmitter with the front cover of the NIBP cuff.
5. Connect the air hose to the NIBP socket on the transmitter. Turn the cuff connector joint until it clicks.
2
1
30 Operator's Manual ZM-940PA
Page 45
When Using YP-960T series or YP-910P series NIBP Cuff
To use these NIBP cuffs, an optional YN-990P extension hose (1.5 m) is required.
NOTE
Connect the joints properly. If there is an air leak, NIBP cannot be measured properly.
1. Connect the NIBP cuff to the extension hose.
2. Connect the other end of the extension hose to the NIBP socket on the transmitter. Turn the joint clockwise until it clicks.
To disconnect the cuff from the transmitter, turn the hose joint counterclockwise.
Attaching the NIBP Cuff to the Patient
WARNING
Be careful when measuring NIBP on a patient with known bleeding disorders or coagulation. After NIBP measurement, there may be dot hemorrhage, or circulatory disorder by thrombus where the cuff was attached.
CAUTION
• Do not wrap the cuff on an arm or thigh which is used for injection. NIBP measurement on an arm or thigh which is used for injection may cause a reflux of blood and stop injection.
• Do not wrap the cuff too tight. It may cause poor blood circulation and congestion. If the cuff is wrapped too loosely, the NIBP value may be increased.
• If the skin gets irritated or redness appears on the skin from the cuff, change the attachment site or stop using the cuff.
Operator's Manual ZM-940PA 31
Page 46
• NIBP and SpO2 can be measured on the same limb, but the SpO2 monitoring may not be accurate during NIBP measurement. Be careful when reading the SpO
values.*
2
• Do not reuse disposable cuffs.
* Monitoring SpO2 during NIBP Measurement
When the SpO during NIBP measurement and pulse wave cannot be detected and SpO properly. When “INHIBIT SpO set to ON (factory default setting), SpO avoid SpO NIBP, be careful when reading SpO
probe is attached to the same limb as the NIBP cuff, the blood flow decreases
2
cannot be monitored
2
DURING NIBP” on the PARAMETER SETUP screen is
2
monitoring is paused during NIBP measurement to
2
alarm occurrence. However, when monitoring SpO2 on the same limb as the
2
values.
2
NOTE
• Measuring NIBP at a site other than the upper arm gives different values from those measured at the upper arm. When making diagnosis based on the NIBP values, measure NIBP on an upper arm.
• To accurately detect the pulsatile flow of the artery, the cuff should be wrapped around a bare upper arm.
• Do not use an abnormal cuff. The cuff deteriorates from use and cleaning. Before use, check the cuff and confirm that there is no flaw, crack or hole in it. Be careful not to damage the inflation bag. If the inflation bag has a hole or a flaw, it may burst during use. Dispose of an abnormal cuff and replace it with a new one.
• Refer to the NIBP cuff manual for details.
Cuff Position
Place the cuffed upper arm (brachium) at the same height as
When placing the transmitter on a bed, make sure that the hose is not bent.
the patient’s heart. If the cuff is not at the same level as the heart, the weight of the blood affects the blood pressure reading. The pressure difference per unit height is 0.7 mmHg/cm. The blood pressure reading decreases when the arm is higher than the heart and increases when lower.
Heart
The best measuring condition is when the patient is lying on his/her back with arms and legs relaxed. If the cuff position cannot be on the same level as the heart, the displayed blood pressure reading must be mathematically adjusted.
32 Operator's Manual ZM-940PA
Page 47
Using the YP-943P/944P NIBP Cuff
Belt
1. Attach the NIBP cuff to the transmitter. Refer to the “Connecting the NIBP Cuff to the Transmitter” section.
2. Insert the end of the cuff into the belt and then through the D ring as shown at left.
D ring
End of cuff
3. Fold back the cuff at the D ring and fasten it using the velcro tape.
Make sure that the cuff is not attached on a joint.
NOTE
The cuff must not wrap around the elbow.
Operator's Manual ZM-940PA 33
Page 48
Attaching the Strap to the Transmitter
NOTE
• Use the strap to prevent the transmitter from falling.
• Do not attach the clip to hard objects such as thick cloth or zipper. It will break the clip.
Attach a strap provided with the transmitter to the NIBP cuff and patient clothes.
To open the clip, firmly pull out the tab in direction of the arrow.
To adjust the strap length, push down the tab on the adjuster and slide.
Belt for the strap on the NIBP cuff
1. Adjust the length of the strap.
2. Clip one end of the strap to the belt for the strap on the NIBP cuff.
3. Clip the other end of the strap to the patient’s clothes as shown left.
34 Operator's Manual ZM-940PA
Page 49
Using the YP-960T series Reusable Cuffs or YP-910P series Disposable Cuffs
ARTERY
1. Put the cuff on the upper arm so that the
mark of “ARTERY ” aligns with the artery
of the patient.
2. Wrap the cuff so that “INDEX ” comes
within the “ RANGE ”.
RANGE
If “Index
”, change the cuff size.
” is not within the “ RANGE
Attaching Electrodes
Selecting Electrode Lead
CAUTION
Use only Nihon Kohden specified electrode leads. With electrodes and electrode leads other than specified ones, the “CHECK ELECTRODE” message appears and monitoring may stop.
The following electrode leads can be used on the transmitter (option).
BR-903PA, 3 electrodes, clip type
Operator's Manual ZM-940PA 35
BR-913PA, 3 electrodes, snap type
BR-906PA, 6 electrodes, clip type
BR-916PA, 6 electrodes, snap type
Page 50
Connecting the Electrode Lead to the Transmitter
Connect the electrode lead to the ECG/RESP socket on the transmitter.
When the transmitter is attached on an arm
CAUTION
Hold the connector of the electrode lead when connecting/disconnecting the electrode lead. If you disconnect the electrode lead by pulling the lead, it damages the electrode lead.
Selecting the Electrode Position
Follow the physician’s instructions for electrode placement when available. For ECG monitoring, electrodes are attached only on the chest to allow patient movement and obtain continuous stable ECG. Following leads are examples. When also monitoring respiration, refer to the “Electrode Position for Respiration Monitoring” section.
NOTE
The optimum electrode positions for ECG measurement of a patient are not always optimum for respiration measurement of the patient. Select positions suitable for both ECG and respiration measurements, or positions which have priority for one measurement.
36 Operator's Manual ZM-940PA
Page 51
Six Electrodes
The 6-electrode method with lead II and lead V5 is effective for monitoring myocardial ischemia. You can improve monitoring accuracy considerably by adding lead V4 to this combination. Va and Vb can be at any position of the standard 12 leads V1 to V6, but V4 and V5 are most appropriate for myocardial ischemic monitoring.
RA/R
Va/Ca
RL/RF
Electrode Position
LA/L
Vb/Cb
LL/F
Symbol Lead Color
AHA IEC
AHA IEC
Left infraclavicular fossa LA L Black Yellow
Right infraclavicular fossa RA R White Red
Below lowest rib on the left anterior axillary line LL F Red Green
Right anterior axillary line at the same level as LL/F Fifth intercostal space on the left midclavicular line. (V4 position of standard 12 leads) Left anterior axillary line at the same level as Va. (V5 position of standard 12 leads)
RL RF Green Black
Va Ca
Vb Cb
Brown-
blue
Brown-
orange
White-
brown
White-
black
Operator's Manual ZM-940PA 37
Page 52
Lead Position
Standard limb leads
Lead I
Lead II Lead III
RA
LA
LL
N (RL)
RA
LL
N (RL)
Monopolar limb leads
aV
R
lead aVL lead aVF lead
RA
LA
RA
LL
N (RL)
Monopolar chest leads
V1 to V6 leads
LA
RA
LL
N (RL)
N (RL)
to
LA
LL
N (RL)
N (RL)
RA
RA
LALA
LL
LA
LL
38 Operator's Manual ZM-940PA
Page 53
Three Electrodes
• Lead MII, which is similar to standard lead II, used when ECG measurement has priority
Electrode Position
Symbol Lead Color
AHA IEC AHA IEC
Left infraclavicular fossa LA L Black Yellow Right infraclavicular fossa RA R White Red Below lowest rib on the left anterior axillary line
• Lead MI, which is similar to standard lead I Change F/LL and L/LA of the lead
LL F Red Green
• Lead MIII, which is similar to standard lead III. Change R/RA and L/LA of the lead MII.
MII.
If the electrode position shown above is not available due to chest surgery, attach the electrodes to the root of the limbs or below the clavicles for stable ECG monitoring.
Operator's Manual ZM-940PA 39
Page 54
Attaching Electrodes to the Patient and Connecting the Electrode Leads to Disposable Electrodes
Prepare the Patient Skin
Shave off excessive body hair. To reduce skin impedance, clean the electrode site with cream or with a cotton pad moistened with alcohol. Thoroughly dry the skin with a clean cotton pad.
NOTE
• For a patient with frequent body movement, rub the sites with Skinpure skin preparation gel. However, do not use Skinpure skin preparation gel on sensitive skin.
• Do not place electrodes on a wound or on an inflamed, wrinkled or uneven skin surface.
Attaching Electrodes to the Patient
CAUTION
Do not reuse disposable products.
NOTE
• To maintain good contact between the electrode and skin, check that the paste of the disposable electrode is not dry.
• When contact between the disposable electrode and skin becomes poor, replace electrodes with new ones immediately. Otherwise, contact impedance between the skin and the electrode increases and the correct ECG cannot be obtained.
Refer to the electrode operator’s manual for details.
1. Carefully remove the backing paper from the electrode. Avoid touching the adhesive surface.
2. Place the electrode on the previously cleaned skin. Pay attention to the electrode lead color and symbol.
3. Clip the electrode lead to the electrode.
4. Fasten the electrode lead wire with surgical tape with an extra length of wire between the tape and the electrode. This lessens the movement of electrode leads by body movement and helps stable monitoring.
40 Operator's Manual ZM-940PA
Page 55
Electrode Position for Respiration Monitoring
Place the R/RA and F/LL electrodes so that the lungs are between the electrodes.
NOTE
The optimum electrode positions for ECG measurement of a patient are not always optimum for respiration measurement of the patient. Select positions suitable for both ECG and respiration measurements, or positions which have priority for one measurement.
Electrode Position Examples
NOTE
The following examples are when monitoring with 3 electrodes. ECG cannot be monitored correctly when electrodes are attached as the following examples when monitoring with 6 electrodes.
Position 1 In this position, respiration measurement is available; however, there is a difference in amplitude between different patients.
R or RA F or LL
Right infraclavicular fossa
Fifth intercostal space on the left midclavicular line, V4
Position 2 In this position, the waveform amplitude is usually large and the ECG lead is similar to Lead MII. This position can be generally recommended.
R or RA F or LL
Right infraclavicular fossa
Operator's Manual ZM-940PA 41
Fifth intercostal space on the left midaxillary line, V6
Page 56
Position 3 In this position, the respiration waveform is optimum, but the ECG lead is unusual.
R or RA F or LL
Right midaxillary at the horizontal level of V4
Fifth intercostal space on the left midaxillary line, V6
Position 4 In this position, the respiration measurement is influenced by the impedance variation of the abdomen, so the cardiac pulse wave included in the respiration wave is reduced. Note that the waveform is inverted in phase compared with the chest movement (the waveform goes down during inspiration). It is difficult to measure the ECG at the same time.
R or RA F or LL
Lowest rib on the right anterior axillary line
Lowest rib on the left anterior axillary line
42 Operator's Manual ZM-940PA
Page 57
Attaching the SpO2 Probe
Selecting the SpO2 Probe
Select an appropriate probe for the patient.
CAUTION
• Use Nihon Kohden specified SpO2 probe to assure maximum performance from your instrument.
• Do not use damaged or disassembled probe. It causes incorrect measurement and may hurt the patient.
Reusable Probes
When using a TL-201T finger probe, choose the appropriate cable length for attachment.
Probe Cable Length Patient Attachment site
Finger probe TL-201T
0.6 m
1.6 m
Adult or children 20 kg or more
Finger
Attachment tape
Adult or Infant
Finger or toeMulti-site probe TL-220T
3 kg or more
Neonate
Instep and sole
3 kg or less
Operator's Manual ZM-940PA 43
Page 58
Disposable Probes
CAUTION
• Use the disposable probe only for one patient. Never reuse the disposable probe for another patient because it causes cross infection.
• Disposable probes are not sterilized.
Probe Patient Attachment site
TL-251T Adult
30 kg or more
Finger or toe
TL-252T Child
3 to 40 kg
TL-253T Neonate
3 kg or less
Multi-site Y probe TL-260T
Attachment tape
Low birth weight infant 1 kg or less Neonate or Child 3 kg or more Neonate 3 kg or less Adult 50 kg or more
40 mm
Neonate 3 kg or less
Cable length TL-051S: 80 cm
TL-052S: 160 cm
Child or Adult 15 to 50 kg
Finger or toe
Instep and sole
Instep and sole
Attachment tape S
Finger or toe Attachment
tape S Instep and sole
Attachment
tape L FingerTL-051S/052S
Instep and sole
FingerTL-061S/062S
35 mm
Infant
Toe
3 to 15 kg
Cable length TL-061S: 80 cm
TL-062S: 160 cm
44 Operator's Manual ZM-940PA
Page 59
Model Patient Attachment site
TL-271T Adult
30 kg or more
TL-272T Child
10 to 50 kg
TL-273T
Neonate 3 kg or less
Adult 40 kg or more
TL-274T Infant
3 to 20 kg
Finger or toe
Instep
Finger or toe
Operator's Manual ZM-940PA 45
Page 60
Connecting the SpO
Connect the probe to the SpO
Probe to the Transmitter
2
socket on the transmitter.
2
When the transmitter is attached on an arm
CAUTION
Hold the connector when connecting/disconnecting the probe. If you disconnect the SpO
probe by pulling the cable, it damages the cable.
2
Attaching the Probe to the Patient
Attach the probe to the patient by referring to the probe’s manual. Make sure that the light emitter and photo detector of the probe face each other at the attachment site.
WARNING
• When using a TL-201T finger probe, do not fasten it to a finger by wrapping the probe to the site with some tape. It may cause poor blood circulation, congestion, pressure necrosis or burn.
• When using probes other than a TL-201T finger probe, do not wrap the tape too tight. Check the blood circulation condition by observing the skin color and congestion at the skin peripheral to the probe attachment site. Even for short­term monitoring, there may be burn or pressure necrosis from poor blood circulation, especially on neonates or low birth weight infants whose skin is delicate. Accurate measurement cannot be performed on a site with poor peripheral circulation.
46 Operator's Manual ZM-940PA
Page 61
• Check the circulation condition by observing the skin color of the measuring site and the pulse waveform. Change the measuring site every 8 hours for disposable probes and every 4 hours for reusable probes. The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree.
• A patient with a fever
• A patient with a peripheral circulation insufficiency
• Neonate or low birth weight infant with delicate skin
CAUTION
• If the attachment site is dirty with blood, clean the attachment site before attaching the probe. If there is nail polish on the attachment site, remove the polish. Otherwise, the amount of transmitted light decreases and the measured data may be incorrect or measurement cannot be performed.
• If the skin gets irritated or redness appears on the skin from the probe, change the attachment site or stop using the probe.
• When the probe is attached on an appropriate site with sufficient circulation, but the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one.
• Do not use a probe that is deteriorated by aging. Accurate measurement cannot be performed.
• When using probes other than a TL-201T finger probe on a neonate, be careful when removing the adhesive tape from neonatal skin.
• When removing a probe taped to the skin, do not pull the cable. Otherwise the cable may break.
• When removing the probe from the attachment tape, do not pull the sensor cable. Otherwise the cable may get damaged.
• Before using the TL-260T multi-site Y probe, be sure to attach the probe to the sponge attachment tape S or L. Do not use the probe without the sponge attachment tape attached. It causes incorrect measurement and may damage the attachment site on the skin.
• When fixing the TL-260T multi-site Y probe with the sponge attachment tape, confirm that the adhesive part of the tape is not on the skin. The adhesive may cause oversensitive symptoms on the skin such as redness or itch. If the adhesive touches the skin, remove it carefully and slowly because neonatal skin is very delicate.
• Do not use a dirty sponge attachment tape. The measurement value may be incorrect.
• Refer to the probe instruction manual for details.
Operator's Manual ZM-940PA 47
Page 62
Locking the Keys on the Transmitter
To prevent the patient from pressing the keys on the transmitter during monitoring, you can lock the NIBP START/STOP and NIBP INTERVAL keys.
1. Press the NIBP INTERVAL key for about 3 seconds.
2. A “pip, pip” sounds and the “LOCK KEYS” message is displayed on the LCD.
Pip, pip
LOCK KEYS
When the NIBP START/STOP key or NIBP INTERVAL key is pressed while the keys are locked, the “PRESS INT. KEY 3S TO UNLOCK” message appears.
To unlock the keys:
1. Press the NIBP INTERVAL key for about 3 seconds.
2. A “pip, pip” sounds and the keys are unlocked. The “UNLOCK KEYS” message appears and the keys are unlocked.
48 Operator's Manual ZM-940PA
Page 63
Monitoring
When preparation is done, monitoring starts.
NIBP Oscillometric Method
NIBP is measured from the change in amplitude pattern of pulsatile oscillation in cuff pressure as the cuff pressure is reduced from above systolic to below diastolic pressure. The occlusive­oscillometry method uses this to determine the systolic, diastolic and mean arterial pressure.
NIBP Monitoring
Selecting the Initial Cuff Inflation Pressure
The initial cuff inflation pressure can be changed on the PARAMETER SETUP screen. The default setting is 180 mmHg. To change the setting, refer to the “Changing Parameter Setup Settings” section.
Selecting the Measurement Mode and Interval
Measurement Modes
There are three measurement modes: manual, auto and STAT. The selected mode or interval is displayed at the lower right of the screen.
The measurement mode and interval can be changed by pressing the NIBP INTERVAL key. When the key is pressed, the measurement mode changes according to the modes selected at “SELECTABLE INTERVALS” on the PARAMETER SETUP screen. MANUAL mode is already selected for the mode selection.
To select the modes for the mode selection, refer to the “Changing Parameter Setup Settings” section.
Manual Measurement
In Manual mode, a single NIBP measurement is performed when the NIBP START/STOP key is pressed.
STAT (Continuous) Measurement
In STAT mode, measurement is continuously repeated for 15 minutes after the NIBP START/ STOP key is pressed.
When the STAT measurement for 15 minutes is completed, the measurement mode automatically changes to the Manual mode or Auto mode of selected interval depending on the “NIBP MODE
Operator's Manual ZM-940PA 49
Page 64
AFTER STAT” setting on the PARAMETER SETUP screen. The default setting is Manual mode. Refer to the “Changing Parameter Setup Settings” section.
The STAT measurement completes within 15 minutes. When more than 12 minutes elapse from the start of measurement, there will be no more measurement performed and the measurement mode changes to the mode selected for “NIBP MODE AFTER STAT” on the PARAMETER SETUP screen.
Auto Measurement
In Auto mode, measurement is performed automatically at the preset time intervals.
In Auto mode, a single measurement can be performed by pressing the NIBP START/STOP key between auto measurements.
Measuring NIBP
WARNING
• Be careful when measuring NIBP on a patient with known bleeding disorders
or coagulation. After NIBP measurement, there may be dot hemorrhage, or circulatory disorder by thrombus where cuff is attached.
• NIBP measurement may be incorrect in the following cases.
• When using an electrical surgery unit.
• When there is body movement.
• When the pulse wave is small (insufficient peripheral circulation).
• Too many arrhythmia.
• When there is vibration.
• When there is a rapid blood pressure change.
• During CPR.
• When performing NIBP measurements in STAT mode or 5 minute intervals,
periodically remove the cuff from the patient for ventilation. Otherwise, the skin temperature may increase at the cuff attachment site by 2 or 3°C (4 or 5°F). When measuring a patient with a fever or peripheral circulation insufficiency, it may cause a burn.
CAUTION
When performing NIBP measurement repeatedly, have a rest between measurements to recover adequate circulation.
50 Operator's Manual ZM-940PA
Page 65
NOTE
• When measuring patients who are conscious, help the patient to relax. Measurement may not be accurate if the patient’s arm is tense or if the patient talks.
• The data for measurement on a leg tends to be higher than measurement on the arm. When making diagnosis based on the NIBP values, measure NIBP on an upper arm.
• Do not apply pressure to the cuff or air hose. NIBP may not be measured correctly because of noise or NIBP measurement may stop due to the NIBP safety circuit.
• When the transmitter is attached to the patient arm and the NIBP measurement is performed when moving, tell the patient to relax and keep quiet. Otherwise, measurement may be stopped or remeasurement is repeated due to body movement.
• If there is an abnormal noise generated during measurement, stop using the transmitter and contact your Nihon Kohden distributor.
• Do not measure NIBP of a patient on whom an IABP is being used. Measurement may be incorrect due to the mixing of the patient’s own pulse and IABP pulse.
• NIBP cannot be measured on a neonate using this transmitter.
1. Select the measurement mode by pressing the NIBP INTERVAL key.
2. Press the NIBP START/STOP key to perform measurement.
Operator's Manual ZM-940PA 51
Page 66
The cuff is inflated and the inflation pressure is displayed on the screen.
Inflation pressure
In manual mode: Measurement is performed once. In STAT mode: Measurement is performed repeatedly for 15 minutes. In auto mode: The first measurement is performed when the NIBP START/STOP key
is pressed. The second measurement is performed when the current time in the transmitter reaches the selected time interval.
To stop measurement during measurement, press the NIBP START/STOP key again.
In STAT mode, after completing the STAT measurement, the measurement mode changes to the mode set for “NIBP MODE AFTER STAT” on the PARAMETER SETUP screen.
In auto mode, to stop measurement in auto mode, change the mode to manual. To cancel one measurement, press the NIBP START/STOP key during measurement.
After the measurement is complete, the measured data is displayed on the screen and is transmitted to the monitor.
When SpO2 is not monitored, the pulse rate at the end of NIBP measurement is displayed.
During auto mode measurement, the measurement mode can be changed. During the interval, press the NIBP INTERVAL key to change the mode. When “MANUAL” is displayed for more than one second, the measurement in auto mode is stopped.
A buzzer can be set to sound at the start and end of NIBP measurement. Refer to the “Changing Parameter Setup Settings” section.
52 Operator's Manual ZM-940PA
Page 67
Data Display After NIBP Measurement
When the time set at “OLD NIBP DATA” on the PARAMETER SETUP screen elapses after the last measurement, the NIBP data is dimmed or hidden. Whether to dim or hide the old data can also be selected at “OLD NIBP DATA”. Refer to the “Changing Parameter Setup Settings” section.
Monitoring SpO
When the SpO during NIBP measurement and pulse wave cannot be detected and SpO properly. When “INHIBIT SpO to ON (factory default setting), SpO SpO
alarm occurrence. However, when monitoring SpO2 on the same limb as the NIBP, be
2
careful when reading SpO
during NIBP Measurement
2
probe is attached to the same limb as the NIBP cuff, the blood flow decreases
2
cannot be monitored
2
DURING NIBP” on the PARAMETER SETUP screen is set
2
monitoring is paused during NIBP measurement to avoid
2
values.
2
ECG and Respiration Monitoring
When the electrodes are attached and the ECG leads are connected to the electrodes, heart rate, ECG, respiration rate and respiration waveform appear on the monitor.
When 6 leads are used on this transmitter, up to 8 lead (I, II, III, aVR, aVL, aVF, Va and Vb) of ECG waveforms can be displayed on the receiving monitor. The heart rate is also measured. When 3 leads are used, one channel ECG waveform of lead II can be displayed on the receiving monitor. Refer to the operator’s manual of the monitor for details.
Respiration is monitored by measuring changes in impedance between the RA and LL ECG electrodes. This transmitter sends the changes in impedance to the monitor as a respiration waveform. The monitor displays the respiration waveform and calculates respiration rate. Refer to the operator’s manual of the monitor for details.
WARNING
Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac monitoring and Diagnostic Equipment* The bioelectric impedance measurement sensor of a minute ventilation rate­adaptive implantable pacemaker may be affected by the transmitter which is connected to the same patient. If this occurs, the pacemaker may pace at its maximum rate and the transmitter may give incorrect data to the monitor. If this occurs, disconnect the electrode leads from the patient or change the setting on
Operator's Manual ZM-940PA 53
Page 68
the pacemaker by referring to the pacemaker’s manual. For more details, contact your pacemaker distributor or Nihon Kohden distributor.
* Minute ventilation is sensed in rate-adaptive pacemakers by a technology known as bioelectric
impedance measurement (BIM). Many medical devices in addition to pacemakers use this technology. When one of these devices is used on a patient with an active, minute ventilation rate-adaptive pacemaker, the pacemaker may erroneously interpret the mixture of BIM signals created in the patient, resulting in an elevated pacing rate.
For more information, see the FDA web site. http://www.fda.gov/cdrh/safety.html
WARNING
When using this transmitter with an ESU, the ESU return plate and the electrodes for monitoring must be firmly attached to the patient. If the return plate is not attached correctly, it may burn the patient’s skin where the electrodes are attached. Refer to the instruction manual for the ESU.
CAUTION
Turn off the power of cell telephones, small wireless devices and other devices which produce strong electromagnetic interference. Otherwise, the waveforms and measurements are affected by such interference and the displayed data may be incorrect.
NOTE
• Noise generated from an electrosurgery unit may interfere on an ECG waveform, but will not damage it.
• If an electric blanket is used and incorrect heart rate is displayed on the monitor, turn off the pacing spike detection on the monitor.
• Turn the pacing spike detection to ON on the monitor when monitoring a pacemaker patient. Pacing pulse is detected by the transmitter and transmitted to the monitor. If the pacing spike detection is turned OFF, QRS and pacemaker spike may not be distinguished and pacemaker failure may not be recognized.
54 Operator's Manual ZM-940PA
Page 69
Electrode Detachment
In the following conditions, the check electrode indication is displayed on the LCD of the transmitter and the “CHECK ELECTRODE” message is displayed on the monitor.
• Electrode is detached from skin.
• Electrode lead is disconnected from the electrode.
• Polarization voltage between the electrode and skin is excessively high.
In these cases, check the cause and if necessary, replace electrodes with new ones.
CAUTION
When the “ELECTRODE OFF” or “CHECK ELECTRODE” message is displayed on the receiving monitor, check electrodes and electrode leads and remove the cause. While the “ELECTRODE OFF” or “CHECK ELECTRODE” message is being displayed, there is no ECG monitoring and no alarms.
Check Electrode Indication on the Transmitter when Monitoring with 3 Electrodes
The “
” mark is displayed.
Check electrode symbol
Check Electrode Indication on the Transmitter when Monitoring with 6 Electrodes
The “
” mark and either the detached lead or detached electrode position is indicated,
depending on the LEADS OFF DISPLAY setting on the PARAMETER SETUP screen.
Check electrode symbol
Check electrode indication
When LEADS OFF DISPLAY is set to CHAR, the detached lead is indicated
Operator's Manual ZM-940PA 55
When LEADS OFF DISPLAY is set to IMAGE, the detached electrode position is indicated with X
Page 70
SpO2 Monitoring
When monitoring starts, SpO2 and pulse waveform are sent to the monitor and SpO2, pulse rate and pulse level bar graph are displayed on the transmitter LCD.
WARNING
• Measurement may be incorrect in the following cases.
• When the patient’s carboxyhemoglobin or methemoglobin increases
abnormally.
• When dye is injected in the blood.
• When using an electrical surgery unit.
• During CPR.
• When there is body movement.
• When there is vibration.
• When measuring at a site where there are venous pulses.
• When the pulse wave is small (insufficient peripheral circulation).
• When using an IABP (intra-aortic balloon pump).
• Check the circulation condition by observing the skin color of the measuring site and pulse waveform. Change the measuring site every 8 hours for disposable probes and every 4 hours for reusable probes. The skin temperature may increase at the attached site by 2 or 3°C (4 or 5°F) and cause a burn or pressure necrosis. When using the probe on the following patients, take extreme care and change the measurement site more frequently according to symptoms and degree.
• A patient with a fever
• A patient with peripheral circulation insufficiency
• Neonate or low birth weight infant with delicate skin
• When not monitoring SpO Otherwise, noise may interfere from the probe sensor and incorrect data is displayed on the screen.
, disconnect the SpO2 cable from the transmitter.
2
CAUTION
• Turn off the power of cell telephones, small wireless devices and other devices which produce strong electromagnetic interference. Otherwise, the waveforms and measurements are affected by such interference and the displayed data may be incorrect.
• Normally external light does not affect monitoring, however, strong light such as an operating lamp or sunlight may affect monitoring. If affected, cover the measuring site with a blanket.
56 Operator's Manual ZM-940PA
Page 71
• Do not pull or bend the probe cable, and do not run over the probe cable with caster feet. Do not immerse the probe cable in detergents or water. Failure to follow these cautions may cause cable discontinuity, short circuit, skin burn on the patient and incorrect measurement data. Replace any broken probe with a new one.
• When the probe is attached on an appropriate site with sufficient circulation and the error message confirming the probe attachment repeatedly appears, the probe may be deteriorated. Replace it with a new one.
• When the probe failure message appears on the screen, replace it with a new one. Otherwise SpO
data may not be accurate.
2
• While a patient is on medication which causes vasodilation, the pulse waveform may change and in rare cases the SpO
value may not be displayed.
2
NOTE
In order to maintain sufficient blood circulation, keep the measurement site warm by covering it with a blanket or something similar. Warming the site is effective, especially for a patient with a small pulse amplitude.
SpO
and PR Display Order
2
You can select the display order for SpO “Changing Parameter Setup Settings” section.
and PR (pulse rate) on the LCD. Refer to the
2
Monitoring SpO
When the SpO during NIBP measurement and pulse wave cannot be detected and SpO properly. When “INHIBIT SpO to ON (factory default setting), SpO SpO
alarm occurrence. However, when monitoring SpO2 on the same limb as NIBP, be careful
2
when reading SpO
When monitoring SpO
during NIBP Measurement
2
probe is attached to the same limb as the NIBP cuff, the blood flow decreases
2
cannot be monitored
2
DURING NIBP” on the PARAMETER SETUP screen is set
2
monitoring is paused during NIBP measurement to avoid
2
values.
2
is important, attach the probe to the limb to which the NIBP cuff or
2
catheter is not attached.
When SpO2 monitoring is paused during NIBP measurement, the SpO2 value just before the start of NIBP measurement and an When NIBP measurement is not completed after 30 seconds, “– – –” is displayed for the SpO
mark are displayed on the transmitter for 30 seconds.
2
value. The same data also appears on the monitor screen.
Operator's Manual ZM-940PA 57
Page 72
NOTE
• When continuous SpO2 monitoring is necessary, attach the probe to the limb to which the NIBP cuff is not attached and set “INHIBIT SpO the PARAMETER SETUP screen to OFF.
• When the probe is attached to the same limb as the NIBP cuff, set the sync source to a parameter other than SpO
• When monitoring SpO
during STAT NIBP measurement, attach the probe to the
2
on the receiving monitor.
2
limb to which the NIBP cuff is not attached.
DURING NIBP” on
2
58 Operator's Manual ZM-940PA
Page 73
Display and Message List
Battery Indication
Indication Cause Countermeasure
Fully charged battery
Batteries are low.
No
indication
Batteries are low. NIBP cannot be measured.
Dead batteries
Replace batteries.
ECG/Respiration
Indication Cause Countermeasure
Electrode lead is disconnected from the electrode. Electrode lead is disconnected from the transmitter.
Electrode lead discontinuity.
Electrode is not firmly attached to the skin.
Polarization voltage is abnormally high.
When monitoring ECG with 6 electrodes, the electrode or lead detached position is indicated by either lead or electrode position. This is set at LEADS OFF DISPLAY on the PARAMETER SETUP screen. Refer to the “Changing Parameter Setup Settings” section.
LEADS OFF DISPLAY set to CHAR ECG ELECTRODE set to AHA
LEADS OFF DISPLAY set to CHAR ECG ELECTRODE set to IEC
Firmly connect the electrode lead to the electrode. Firmly connect the electrode lead to the transmitter. Replace the electrode lead with a new one.
Replace the electrode with a new one.
LEADS OFF DISPLAY set to IMAGE
Operator's Manual ZM-940PA 59
Page 74
SpO
2
Message Cause Countermeasure
During NIBP measurement
Detecting body movement
SpO2 CHECK PROBE
SpO2 CHECK PROBE SITE
SpO2 DETECTING PULSE
SpO2 monitoring is paused for NIBP measurement.
Considerable body movement.
The probe is not attached to the patient properly.
The probe is not attached to the patient properly. The probe is not attached at the appropriate site.
Wait for NIBP measurement to finish. When the message is displayed frequently, check the patient condition and, if necessary, change the attachment site. Attach the probe to the patient properly. Attach the probe to a site 6 to 14 mm thick.
Probe is expired. Replace the probe with a new one.
Searching for the correct pulse wave.
Wait until the pulse wave is
detected. The SpO2 value cannot be obtained because the waveform is unstable. The probe is not attached to
Attach the probe to the patient
properly.
the patient properly. SpO
measurement site is
SpO2 LIGHT INTERFERENCE
SpO2 PROBE FAI LU RE
SpO2 WEAK PULSE
2
under fluorescent light, surgical light, sunlight, etc.
Probe is expired. Replace the probe with a new one.
Probe is damaged or short­circuited.
Poor peripheral circulation.
The probe is attached too tightly and is obstructing the blood circulation.
Cover the measurement site with a
blanket or cloth.
Replace the probe with a new one.
Check the patient condition and
change the attachment site.
Check the probe attachment
condition and if necessary,
reattach the probe.
60 Operator's Manual ZM-940PA
Page 75
NIBP
Message Cause Countermeasure
The cuff and extension hose are not
NIBP AIR LEAK
NIBP CANNOT DETECT PULSE
NIBP CUFF OCCLUSION
NIBP HIGH CUFF PRESS
NIBP INFLATION PRESS LOW
NIBP MEAS TIME­OUT
NIBP MODULE FAI LU RE
NIBP REMEASURING
NIBP SAFETY CIRCUIT RUNNING (When this message is displayed, measurement cannot be performed for 40 seconds.)
NIBP SYS OUT OF RANGE
NIBP WEAK PULSE
NIBP ZEROING
properly connected. The cuff hose (or extension hose) is not properly connected to the NIBP socket. The cuff or extension hose is damaged.
The patient’s pulse wave is small. Measure by palpation or auscultation.
The cuff is not wrapped on the patient properly.
Transmitter malfunction.
Enormous pressure was applied by the pressure of the cuff.
Insufficient cuff inflation pressure.
The measuring time exceeded the specified time due to arrhythmia, body movement, vibration or, cuff or air hose being squeezed.
Module malfunction.
NIBP is being remeasured due to arrhythmia, body movement, vibration or, cuff or air hose being squeezed.
Measurement stopped by the safety circuit.
The maximum blood pressure cannot be measured even when the cuff inflation pressure exceeded 280 mmHg when using adult cuff. The patient’s pulse wave is too small.
The cuff is wrapped too loosely. Wrap the cuff properly.
The cuff size is not appropriate. Use the appropriate cuff.
NIBP zero balance is being adjusted.
Connect them properly.
Replace with a new one.
Wrap the cuff on the patient properly.
Immediately remove the cuff from the patient and contact your Nihon Kohden distributor.
Remove the cause.
Wait for the remeasurement to be performed with increased cuff inflation pressure.
Remove the cause if the cause is body movement, vibration or squeezing of cuff or hose.
Contact your Nihon Kohden distributor. If the message still appears after remeasurement, remove the cause if the cause is body movement, vibration or squeezing of cuff or hose. Check that the hose is not bent or squeezed. Wait 40 seconds, then perform remeasurement. If the message still appears, contact your Nihon Kohden distributor.
Measure by palpation or auscultation.
Measure by palpation or auscultation.
Do not touch the cuff during zeroing. Wait for the message to disappear.
Operator's Manual ZM-940PA 61
Page 76
p
Troubleshooting
If the problem still remains after checking the following, contact your Nihon Kohden distributor.
Transmitter
Problem Cause Countermeasure
displayed on the LCD after turning the power on. LCD is difficult to see (too dark or too light).
Nothing is displayed on the monitor after turning the transmitter power on.
Signal receiving condition is
oor.
Batteries are not installed correctly. The battery polarity is wrong. Batteries are completely discharged. LCD brightness is not appropriate.
The channel of the transmitter and monitor does not match. The software version of the multiple patient receiver is old.
Another transmitter of the same channel is used nearby.
Signals are mixing. Follow the instructions of your channel
The transmitter is damaged.
Install the batteries correctly.Nothing is
Replace the batteries with new ones.
Change the LCD brightness on the SYSTEM SETUP screen. Refer to the “Changing System Setup Settings” section. Set the correct channel on the monitor.
Upgrade the multiple patient receiver software to receive signal from the transmitter. The software version must be 01-09 or later. Turn the transmitter power off. If the monitor still receives a signal, there is a high probability that another transmitter of the same channel is used nearby. Follow the instruction of your channel administrator and use another transmitter of a different channel.
administrator and use another transmitter of a different channel. Contact your Nihon Kohden distributor.
62 Operator's Manual ZM-940PA
Page 77
ECG/Respiration
Problem Cause Countermeasure
The heart rate is unstable.
The “CHECK ELECTRODE” message appears on the receiving monitor.
ECG baseline is thick. (Hum is overlapping)
Respiration waveform measurement is unstable.
Pacing detection setting on the monitor is not correct.
Turn off the pacing detection setting on the monitor. When monitoring a pacemaker
patient, turn on pacing detection. Electrode lead is disconnected from the electrode.
Firmly connect the electrode lead to
the electrode. Electrode lead discontinuity Replace the electrode lead with a new
one. Electrode is not firmly
Replace the electrode with a new one. attached to the skin. Polarization voltage is abnormally high. The gel on the electrode is
Use Nihon Kohden specified
electrodes.
Replace the electrode with a new one. dried out. The gel on the electrode is coming off. Electric blanket is used. Cover the blanket with a shield cover. Hum filter is set to OFF on the
Set the filter to ON. monitor The gel on the electrode is
Replace the electrode with a new one. dried out. The gel on the electrode is coming off.
SpO
2
Problem Cause Countermeasure
SpO2 data is unstable and not reliable.
The probe size is not appropriate for the patient. Probe attachment condition is poor. Probe is partly
Use the appropriate probe for the patient.
Firmly attach the probe according to the
procedure in the probe operator’s manual. detached from the skin. External light gets in. Measurement site is dirty.
Remove dirt and nail polish. Patient is wearing nail polish. Probe is attached to the same limb that is used for NIBP measurement.
When the probe and cuff are attached to
the same limb, set “INHIBIT SpO
2
DURING NIBP” setting on the
PARAMETER SETUP screen to ON.
Attach the probe to the opposite limb.
Avoid a site where blood circulation
condition changes greatly.
Operator's Manual ZM-940PA 63
Page 78
NIBP
Problem Cause Countermeasure
Cuff inflation pressure is less than 10 mmHg.
The cuff does not inflate when the NIBP START/STOP key is pressed.
Abnormal measurement results are displayed.
The cuff is suddenly deflated during inflation.
Auto mode measurement does not start even when the time interval has passed. The cuff suddenly inflates.
Cannot connect cuff to the air hose.
The cuff hose is not connected to the NIBP socket properly. The cuff is not wrapped around the arm or is wrapped too loosely. The cuff hose is not connected to the NIBP socket. The cuff hose or extension hose may be folded or squeezed when the cuff pressure display on the screen increases quickly but the actual cuff does not inflate.
The cuff size is not correct. Select the cuff which fits the
The cuff is not wrapped around the arm correctly.
NIBP data is not correct because of body movement.
Vibration on the cuff.
The NIBP START/STOP key is pressed during inflation.
The NIBP INTERVAL key is pressed and the measurement mode is changed.
The measurement mode is set to auto mode.
Unspecified cuff is used. Use a cuff specified by Nihon
Connect the cuff hose to the socket properly. Wrap the cuff around the upper arm.
Connect the cuff hose to the socket firmly. Check the cuff hose and air hose.
patient’s limb circumference. Wrap the cuff around the upper arm, not too tightly or too loosely.
Prevent the patient from moving during measurement. Check that nothing is touching the cuff during measurement. Change the measuring site.
Check the measurement mode and interval.
Check the time interval. If necessary, stop measurement.
Kohden.
64 Operator's Manual ZM-940PA
Page 79
Problem Cause Countermeasure
Cannot measure NIBP.
Vibration on the cuff.
The cuff hose or extension
Check that nothing is touching the cuff during measurement.
Remove the cause. hose is bent or squeezed. The cuff has worn out. Use a new cuff.
Measuring over a long
occurs.
period of time at short intervals.
Thrombus occurs. Measuring on a patient
with known bleeding
Increase the measuring interval.Blood congestion
Do not measure NIBP over a long
time.
Do not perform NIBP measurement
on such a patient. disorders or coagulation.
NIBP data on the screen is --- or dark.
The time set for “OLD NIBP DATA” on the PARAMETER SETUP
When NIBP is measured again, the
data is displayed in normal
brightness. screen elapsed from the last measurement.
Three loud pip sounds indicting NIBP measurement
The cuff is not deflated enough to start another measurement.
Wait 30 seconds and measure again.
cannot be started.
Operator's Manual ZM-940PA 65
Page 80
Maintenance
To use the instrument in safe and optimum condition, perform maintenance check once every six months.
CAUTION
Do not disassemble the transmitter when performing maintenance and inspection. Do not repair the transmitter. When there is any problem with the transmitter after maintenance and inspection, contact your Nihon Kohden distributor.
A maintenance check sheet is provided at the end of this section. Make a copy of this check sheet before performing maintenance check.
1. External Check
• There are no damaged or dirty parts on the outside of the transmitter.
• The battery case cover is not damaged, the spring is firmly fixed and the battery case cover can be closed firmly.
• NIBP socket is not damaged.
• Keys are not damaged.
• Electrode leads are not damaged.
• There is no blood or chemicals on the transmitter.
2. Transmitter Channel
Check that the channel of the transmitter and the label match.
1. Check that the channel number label attached to the transmitter is not torn or removed.
Channel label
66 Operator's Manual ZM-940PA
Page 81
2. Remove one battery.
3. While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery. The SETUP screen appears.
4. Press the NIBP INTERVAL key to move the cursor to “SYSTEM SETUP”.
5. Press the NIBP START/STOP key to enter the SYSTEM SETUP screen. The channel of this transmitter is displayed.
Channel
6. Check that the channel displayed on the LCD
matches the label on the transmitter.
Operator's Manual ZM-940PA 67
Page 82
3. LCD Display
Check that there are no dots missing on the LCD.
1. Remove one battery.
2. While pressing the NIBP START/STOP and NIBP INTERVAL keys, install the battery. The SETUP screen appears.
3. Press the NIBP INTERVAL key twice to move the cursor to “MANUAL CHECK”.
4. Press the NIBP START/STOP key to enter the MANUAL CHECK screen.
5. Check that the cursor is on “LCD TEST” and press the NIBP START/ STOP key.
68 Operator's Manual ZM-940PA
Page 83
6. Every time the NIBP INTERVAL key is pressed, the screen changes as below. Check that there are no dots missing.
When the NIBP START/STOP key is pressed, the screen returns to the MANUAL CHECK screen.
4. Key Operation
NIBP START/STOP Key
1. Attach the NIBP cuff to your upper arm.
2. Press the NIBP START/STOP key. Check that the cuff inflates and deflates properly.
3. Press the NIBP START/STOP key again. During inflation, press the NIBP START/STOP key to check that the cuff deflates properly.
NIBP INTERVAL Key
1. Press the NIBP INTERVAL key and check that the NIBP measuring mode can be changed.
2. Select any interval and press the NIBP START/STOP key to perform auto measurement. Check that the NIBP is measured at the selected interval.
Operator's Manual ZM-940PA 69
Page 84
5. NIBP Cuff for Attaching Transmitter to Patient Arm
The NIBP cuff is a consumable. Check the following and when necessary, replace it with a new one.
Appearance
• There are no dirty parts.
• There are no broken stitches on the cuff.
• The label on the cuff is readable.
• The velcro tape on the cuff is not removed and there are no broken stitches.
• The lock plate is not damaged and functions properly.
Inflation bag
• The inflation bag is not torn or damaged.
• There is no water inside the inflation bag.
• The connector on the inflation bag is not damaged.
70 Operator's Manual ZM-940PA
Page 85
Hospital/Organization:
Service Personnel:
Maintenance Check Sheet
Instrument Name:
Instrument Model:
Instrument Serial Number:
Hardware Revision Number:
Software Revision Number:
1. External Check OK No
2. Transmitter Channel OK No
3. LCD Display OK No
4. Key Operation OK No
5. NIBP Cuff for Attaching Transmitter to Patient Arm OK No
Overall Judgement
OK
Transmitter
ZM-940PA
Can be used but needs maintenance
Maintenance required. Cannot be used.
Operator's Manual ZM-940PA 71
Page 86
Repair Parts Availability Policy
Nihon Kohden Corporation (NKC) shall stock repair parts (parts necessary to maintain the performance of the instrument) for a period of 8 years from the date of delivery. In that period NKC or its authorized agents will repair the instrument. This period may be shorter than 8 years if the board or part necessary for the faulty section is not available.
72 Operator's Manual ZM-940PA
Page 87
Lifetime and Disposal
Disposing of Used Batteries
Battery Lifetime
Replace the batteries when the battery replacement indication appears on the transmitter. When using rechargeable batteries, recharge them.
Type Lifetime (Measuring parameters)
ECG, SpO2, NIBP ECG, SpO
NiMH secondary 2 days 2 days 2.5 days Alkaline primary 1 day 2.5 days 3 days
The above data is when new batteries are used at room temperature, NIBP is measured in auto mode at 60 minute intervals and SpO weight 60 kg.
Operation time depends on the thickness of SpO2 probe attachment site.
Disposal
is measured on an index finger of a male patient with
2
NOTE
Remove the batteries before disposing of the transmitter.
Before disposing of the batteries, check with your local solid waste officials for details in your area for proper disposal. It may be illegal to dispose of these batteries in the municipal waste stream.
2
ECG only
Disposing of Electrodes, SpO2 Probes and NIBP Cuffs
Refer to the manual of each item.
Operator's Manual ZM-940PA 73
Page 88
Cleaning, Disinfection and Sterilization
Transmitter and Electrode Leads
CAUTION
• This transmitter is not waterproof. If detergent or liquid spills into the transmitter, stop cleaning/disinfecting/sterilizing it and contact your Nihon Kohden distributor. The transmitter needs to be checked for safety and function before use.
• Before cleaning or disinfecting the transmitter, remove the batteries from the transmitter.
• The transmitter cannot be sterilized.
Cleaning
Wipe the transmitter and electrode leads with a soft cloth moistened with disinfecting alcohol or neutral detergent diluted with water. After cleaning, dry them completely.
Disinfection
CAUTION
• Do not immerse the electrode lead connector in liquid.
• Do not disinfect with hypochlorous acid.
• Use the recommended concentration.
Wipe the outside surface of the transmitter and electrode lead with a non-abrasive cloth moistened with any of the disinfectants listed below. Use the recommended concentration.
Disinfectant Concentration (%) Glutaraldehyde solution 2.0 Hydrochloric alkyl diaminoethylglycine 0.5 Benzalkonium chloride 0.2 Benzethonium chloride solution 0.2 Chlorohexidine gluconate solution 0.5
74 Operator's Manual ZM-940PA
Page 89
SpO2 Probe
Refer to the probe manual.
YP-943P/944P NIBP Cuffs
CAUTION
• Do not autoclave.
• Use only glutaraldehyde solution.
• Never allow liquid to get inside the rubber cuff.
• Do not sterilize or disinfect the cuff with ultraviolet light or ozone.
Cleaning
To clean the cuff, remove the lock plate and carefully pull out the inflation bag from the cloth cover. Cloth cover: Wash with neutral detergent and water. Thoroughly dry it. When washing in a
washing machine, put it in a net.
Inflation bag: Wipe with a soft cloth or cotton moistened with disinfecting alcohol.
Thoroughly dry it.
Disinfection
To disinfect the cuff, use glutaraldehyde solution. Use the recommended concentration of the disinfectant. Refer to the disinfectant manual for details. After disinfection, clean the cuff as described above.
Operator's Manual ZM-940PA 75
Page 90
Specifications
Measuring Parameters
Measuring waveforms: ECG, Respiration in impedance method, pulse Measuring numeric data: SpO
Transmitting Data
Waveform data: ECG, respiration, pulse wave Numeric data: SpO Status information: Battery replacement, channel ID, type of transmitter, check
Displayed Data
SpO2, NIBP, pulse rate, pulse wave bar graph, check electrode, battery replacement, NIBP measurement mode and status information
ECG Measurement
Channels: 4 Input range: ±5 mV or more DC offset: ±500 mV or more Input impedance: 5 M or more (5 Hz) Pacing pulse detection: ANSI/AAMI EC13
, NIBP, pulse rate
2
and NIBP
2
electrodes, abnormal polarization voltage, pacing data, SpO status
Based upon Pacemaker pulse rejection Capability
2
Respiration Measurement
Measuring method: Impedance method Impedance range: 0 to 2 k or less
SpO
Measurement
2
Display range: Depends on the receiving monitor Measuring range: 0 to 100%, in 1% steps Minimum display range: 1%
Measuring accuracy (When the measuring accuracy of the SpO
±1 (80% SpO
±2 (50% SpO
100%)
2
< 80%)
2
probe is not considered):
2
Less than 50% is not specified.
(When considering the measuring accuracy of the SpO
±2 (80% SpO
±3 (70% SpO
probe):
2
100%)
2
< 80%)
2
Less than 70% is not specified.
76 Operator's Manual ZM-940PA
Page 91
NIBP Measurement
Displayed items: Systolic, diastolic, mean Cuff pressure display range: 0 to 300 mmHg Measurement modes: Manual, STAT, auto at 5, 10, 15, 30, 60, 120 or 240 minute
interval
Pulse Rate
Measuring range: 30 to 200 beats/minute ±8 beats/min (NIBP)
30 to 250 beats/min ±3% ±1 beat/min (SpO
)
2
Transmitter
FCC regulation: FCC part 95 Subpart H
Wireless Medical Telemetry Service (WMTS) Field strength limits: <200 mV/m (at 3 m) Undesired emission: below 960 MHz: 200 µV/m (at 3 m)
above 960 MHz: 500 µV/m (at 3 m) Antenna: Internal Transmission channel: indicated on the transmitter Transmission frequency range: 608.0250 to 613.9750 MHz Channel spacing: 50 kHz (25 kHz when interleave) Type of emission: F1D Occupied bandwidth: <20 kHz Effective radiated power: 1.0 mW (conducted)
Power Requirements
Operating voltage: 3.2 to 4.8 V Battery type: Three AA type NiMH secondary batteries
Three AA type alkaline dry cell primary batteries Battery lifetime:
Type Lifetime (Measuring parameters)
ECG, SpO2, NIBP ECG, SpO
2
ECG only
NiMH secondary 2 days 2 days 2.5 days Alkaline primary 1 day 2.5 days 3 days
The above data is when new batteries are used at room temperature, NIBP is measured in auto mode at 60 minute intervals and SpO
is measured on an index
2
finger of a male patient with weight 60 kg. Operation time depends on the thickness of SpO
probe attachment site.
2
Dimension and Weight
Dimension: 114 W × 103 H × 58 D (mm) Weight: 280 g ±30 g (excluding batteries, NIBP cuff and other
accessories)
Operator's Manual ZM-940PA 77
Page 92
Environment
Operating environment
Operating temperature: 5 to 40°C, 41 to 104°F
When using NIBP cuff, 10 to 40°C, 50 to 104°F Operating humidity: 30 to 85% (non-condensing) Operating atmospheric pressure: 70 to 106 kPa
Storage environment
Storage temperature: 20 to 65°C, 4 to 149°F Storage humidity: 10 to 95% Storage atmospheric pressure: 70 to 106 kPa
Safety Standards
Safety standard: CAN/CSA-C22.2 No. 601-1 M90:1990
CAN/CSA-C22.2 No. 601-1. 1S1-94:1994 CAN/CSA-C22.2 No. 601-1. 1B-90:R2002 CAN/CSA-C22.2 No. 60601-2-49-04:2004 IEC 60601-1:1988 IEC 60601-1 Amendment1:1991 IEC 60601-1 Amendment2:1995 IEC 60601-1-2:2001 IEC 60601-2-27:1994 IEC 60601-2-30:1999
IEC 60601-2-49:2001 According to the type of protection against electrical shock: INTERNALLY POWERED EQUIPMENT According to the degree of protection against electrical shock:
ECG and impedance method respiration: DEFIBRILLATION-PROOF TYPE CF APPLIED
PA RT
SpO
and NIBP: DEFIBRILLATION-PROOF TYPE BF APPLIED
2
PA RT According to the degree of protection against harmful ingress of water: IPX0 (Ordinary equipment) According to the degree of safety of application in the presence of a FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE: Equipment not suitable for use in the presence of
FLAMMABLE ANAESTHETIC MIXTURE WITH AIR, OR WITH OXYGEN OR NITROUS OXIDE
According to the mode of operation: CONTINUOUS OPERATION
78 Operator's Manual ZM-940PA
Page 93
Electromagnetic Compatibility
IEC 60601-1-2 (2001) Emissions: CISPR11 Group1,Class B
Electromagnetic Emissions
This Model ZM-940PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-940PA should assure that it is used in such an environment.
Emissions test Compliance Electromagnetic environment guidance
RF emissions CISPR 11
Group 1 The ZM-940PA uses RF energy only for its
internal function. Therefore, its RF emissions are very low and are not likely to cause any
interference in nearby electronic equipment. RF emissions CISPR 11 Harmonic emissions
Class B
Not applicable
The ZM-940PA is suitable for use in all
establishments, including domestic establishments.
IEC 61000-3-2 Voltage fluctuations/
Not applicable flicker emissions IEC 61000-3-3
Operator's Manual ZM-940PA 79
Page 94
Electromagnetic Immunity
This Model ZM-940PA is intended for use in the electromagnetic environment specified below. The customer or the user of the ZM-940PA should assure that it is used in such an environment.
Immunity test
Electrostatic discharge (ESD) IEC 61000-4-2
IEC 60601 test
level
6 kV contact
±
8 kV air
±
Compliance
level
6 kV contact
±
8 kV air
±
Electromagnetic
environment -Guidance
Floors should be wood, concrete or ceramic tile. If floors are covered with synthetic material, the relative humidity should be at least
30%. Electrical fast transient/ burst IEC 61000-4-4
2 kV for power
±
supply lines
1 kV for
±
Not applicable
input/output lines Surge IEC 61000-4-5
Voltage dips, short
1 kV differential
±
mode
2 kV common mode
±
T
<5% U
(>95% dip in
U
T) for 0.5 cycle
Not applicable
Not applicable
interruptions and voltage variations on
T
40% U
(60% dip in
T
U
) for 5 cycles power supply input lines IEC 61000-4-11
Power frequency (50/60 Hz) magnetic field IEC 61000-4-8
T
(30% dip in
70% U
T
U
) for 25 cycles
T
(>95% dip in
<5% U
T
U
) for 5 s
3 A/m 3 A/m Power frequency magnetic fields
should be at levels characteristic of a typical location in a typical commercial or hospital environment.
NOTE UT is the AC mains voltage prior to application of the test level
80 Operator's Manual ZM-940PA
Page 95
Immunity test
IEC 60601
test level
Compliance
level
Electromagnetic environment -
guidance
Portable and mobile RF communications equipment should be used no closer to any part of the ZM-940PA, including cables, than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter.
Recommended separation distance
Conducted RF
3 Vrms
3 Vrms
d = 1.2√P
IEC 61000-4-6
Radiated RF IEC 61000-4-3
3 V/m 80 MHz to 2.5 GHz
3 V/m 80 MHz to 2.5 GHz
d = 1.2√P 80 MHz to 800 MHz
d = 2.3√P 800 MHz to 2.5 GHz
where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as deter mined by an electromagnetic site
1
, should be less than the compliance
survey* level in each frequency range*
2
.
Interference may occur in the vicinity of equipment marked with the following symbol:
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies. NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
*1Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
2
*
absorption and reflection from structures, objects and people.
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the ZM-940PA is used exceeds the applicable RF compliance level above, the ZM-940PA should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orienting or relocating the ZM-940PA. Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Operator's Manual ZM-940PA 81
Page 96
Recommended Separation Distances between Portable and Mobile RF Communications Equipment
The ZM-940PA is intended for use in an electromagnetic environment in which radiated RF disturbances are controlled. The customer or the user of the ZM-940PA can help prevent electromagnetic interference by maintaining a minimum distance between portable and mobile RF communications equipment (transmitters) and the ZM-940PA as recommended below, according to the maximum output power of the communications.
Rated maximum output power of transmitter (W)
0.01 0.12 0.12 0.23
0.1 0.38 0.38 0.73 1 1.2 1.2 2.3
10 3.8 3.8 7.3
100 12 12 23 For transmitters rated at a maximum output power not listed above, the recommended separation distance d in meters (m) can be estimated using the equation applicable to the frequency of the transmitter, where P is the maximum output power rating of the transmitter in watts (W) according to the transmitter manufacturer. NOTE 1 At 80 MHz and 800 MHz, the separation distance for the higher frequency range
applies.
NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
Separation distance according to frequency of transmitter (m)
150 kHz to 80 MHz
d = 1.2√P
80 MHz to 800 MHz
d = 1.2√P
800 MHz to 2.5 GHz
d = 2.3√P
System Composition for EMC Test
The ZM-940PA bedside monitor is tested to comply with IEC 60601-1-2 (2001) with the following composition. If any part which is not specified by Nihon Kohden is used, the EMC specifications may not comply.
Units Cable length
ZM-940PA transmitter --­YP-943P NIBP cuff 0.15 m
BR-906P ECG electrode lead 0.8 m TL-201T finger probe 1.6 m
82 Operator's Manual ZM-940PA
Page 97
Standard Accessories
1
2
No. Name Model Q’ty Supply Code No.
1 NIBP cuff for adult, standard YP-943P 1 S938B 2 Strap --- 1 Y236
The following parts are available for replacement.
34
Lock plate is a standard accessory of the YP-943P/944P NIBP cuff.
No. Name Model Q’ty Supply Code No.
3 Battery case cover 1 6144-012004 4 Lock plate 1 6113-049585
Operator's Manual ZM-940PA 83
Page 98
Options
CAUTION
Use only Nihon Kohden electrodes, electrode leads, SpO2 probes and NIBP cuffs to assure maximum performance from your instrument.
Transmitter
Channel writer, QI-901PK
ECG/RESP
Name Application Model Q’ty Supply Code No.
Electrode lead
3 electrodes,clip type, lead length 80 cm 3 electrodes, snap type, lead length 80 cm 6 electrodes,clip type, lead length 80 cm 6 electrodes,snap type, lead length 80 cm
BR-903PA 1 K911A
BR-913PA 1 K910B
BR-906PA 1 K912A
BR-916PA 1 K915A
NIBP
(cm)
Air Hose
Length
(cm)
15
15
20
Model Q’ty
YP-944P*
YP-964T
YP-915P
Name
Cuff for adult, for attaching transmitter to patient arm Cuff for infants 5 YP-960T S943A
Cuff for children
Cuff for adult
Disposable cuff for infants 6 YP-910P --­Disposable cuff for children 9 YP-912P ---
Disposable cuff for adults
Extension hose 150 YN-990P 1 S903
* The lock plate is provided with these NIBP cuffs.
84 Operator's Manual ZM-940PA
Standard 13 YP-943P* S938B
Large 15
Small 7 YP-961T S943B Standard 10 YP-962T S943C Standard 13 YP-963T S944B Large 15
Small 12 YP-913P --­Standard 14 YP-914P --­Large 16
Width
20
Supply
Code
No.
1
S938C
1
S944C
---
Page 99
SpO
2
Name
Finger probe (reusable)
Cable length
0.6 m P225H
1.6 m
Multi-site probe (reusable) TL-220T
Model/
Code No.
TL-201T
Q’ty
1
Supply
Code No.
P225F
P225G SpO2 probe (for adult, disposable) TL-251T P201A SpO2 probe (for child, disposable) TL-252T P201B SpO2 probe (for neonate, disposable)
TL-253T P201C
5 Multi-site Y probe (for low birth weight infant/child/ neonate, disposable)
1.6 m
TL-260T
P205A
SpO2 probe (for adult, disposable) TL-271T P203A SpO2 probe (for child, disposable) TL-272T P203B SpO2 probe (for neonate/adult, disposable) SpO2 probe (for child/infant, disposable) SpO2 probe(for adult/neonate, disposable) SpO2 probe (for child/infant, disposable)
TL-273T P203C
TL-274T
0.8 m TL-051S P228A
1.6 m TL-052S P228B
0.8 m TL-061S P229A
1.6 m TL-062S
24
P203D
5
P229B COTTONY tape 340703 20 P259 Foam tape for TL-051S/052S/ 061S/062S Attachment tape for TL-220T/ 251T/252T/253T
---
---
Attachment tape S for TL-260T P260A Attachment tape L for TL-260T
4 × 25 package
3 × 30 package
24
P260
P263
P260B
Operator's Manual ZM-940PA 85
Page 100
Transmission Frequencies
USA
BAND
Transmission
frequency
(MHz)
608.0250 9002 608.7750 9062 609.5250 9122
608.0500 9004 608.8000 9064 609.5500 9124
608.0750 9006 608.8250 9066 609.5750 9126
608.1000 9008 608.8500 9068 609.6000 9128
608.1250 9010 608.8750 9070 609.6250 9130
608.1500 9012 608.9000 9072 609.6500 9132
608.1750 9014 608.9250 9074 609.6750 9134
608.2000 9016 608.9500 9076 609.7000 9136
608.2250 9018 608.9750 9078 609.7250 9138
608.2500 9020 609.0000 9080 609.7500 9140
608.2750 9022 609.0250 9082 609.7750 9142
608.3000 9024 609.0500 9084 609.8000 9144
608.3250 9026 609.0750 9086 609.8250 9146
608.3500 9028 609.1000 9088 609.8500 9148
608.3750 9030 609.1250 9090 609.8750 9150
608.4000 9032 609.1500 9092 609.9000 9152
608.4250 9034 609.1750 9094 609.9250 9154
608.4500 9036 609.2000 9096 609.9500 9156
608.4750 9038 609.2250 9098 609.9750 9158
608.5000 9040 609.2500 9100 610.0000 9160
608.5250 9042 609.2750 9102 610.0250 9162
608.5500 9044 609.3000 9104 610.0500 9164
608.5750 9046 609.3250 9106 610.0750 9166
608.6000 9048 609.3500 9108 610.1000 9168
608.6250 9050 609.3750 9110 610.1250 9170
608.6500 9052 609.4000 9112 610.1500 9172
608.6750 9054 609.4250 9114 610.1750 9174
608.7000 9056 609.4500 9116 610.2000 9176
608.7250 9058 609.4750 9118 610.2250 9178
608.7500 9060 609.5000 9120 610.2500 9180
CHANNEL
SPACING
C
25 kHz step
Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
C
25 kHz step
Channel No.
USA
BAND
Transmission
frequency
(MHz)
CHANNEL
SPACING
25 kHz step
Channel No.
C
86 Operator's Manual ZM-940PA
Loading...