If you have any comments or suggestions
on this manual, please contact us at:
www.nihonkohden.com
0614-007205B
Copyright Notice
The entire contents of this manual are copyrighted by Nihon Kohden. All rights are
reserved. No part of this document may be reproduced, stored, or transmitted in any
form or by any means (electronic, mechanical, photocopied, recorded, or otherwise)
without the prior written permission of Nihon Kohden.
GENERAL HANDLING PRECAUTIONS
This device is intended for use only by qualified medical personnel.
Use only Nihon Kohden approved products with this device. Use of
non-approved products or in a non-approved manner may affect the
performance specifications of the device. This includes, but is not limited
to, batteries, recording paper, pens, extension cables, electrode leads,
input boxes and AC power.
Please read these precautions thoroughly before attempting to operate the
instrument.
1. To safely and effectively use the instrument, its operation must be fully
understood.
2. When installing or storing the instrument, take the following precautions:
(1) Avoid moisture or contact with water, extreme atmospheric pressure,
excessive humidity and temperatures, poorly ventilated areas, and dust,
saline or sulphuric air.
(3) Avoid placing in an area where chemicals are stored or where there is
danger of gas leakage.
(4) Pay extra attention when the instrument is in combination with other
instruments to avoid misdiagnosis or other problems.
(5) Choose a room where a proper grounding facility is available.
3. Before Operation
(1) Check that the instrument is in perfect operating order.
(2) Check that the instrument is grounded properly.
(3) Check that all cords are connected properly.
(4) Pay extra attention when the instrument is in combination with other
instruments to avoid misdiagnosis or other problems.
(5) All circuitry used for direct patient connection must be doubly checked.
(6) Check that battery level is acceptable and battery condition is good when
using battery-operated models.
Operator’s Manual ZM-920PA/921PA/930PA/931PA i
4. During Operation
(1) Both the instrument and the patient must receive continual, careful
attention.
(2) Turn power off or remove electrodes and/or transducers when necessary to
assure the patient’s safety.
(3) Avoid direct contact between the instrument housing and the patient.
5. To Shutdown After Use
(1) Turn power off with all controls returned to their original positions.
(2) Remove the cords gently; do not use force to remove them.
(3) Clean the instrument together with all accessories for their next use.
6. The instrument must receive expert, professional attention for
maintenance and repairs. When the instrument is not functioning
properly, it should be clearly marked to avoid operation while it is out of
order.
personnel upon request from your Nihon Kohden representative.
9. When the instrument is used with an electrosurgical instrument, pay
careful attention to the application and/or location of electrodes and/or
transducers to avoid possible burn to the patient.
patient cables and/or transducers from the instrument to avoid possible
damage.
ii Operator’s Manual ZM-910PA/921PA/930PA/931PA
WARRANTY POLICY
Nihon Kohden Corporation (NKC) shall warrant its products against all defects in
materials and workmanship for one year from the date of delivery.
However, consumable materials such as recording paper, ink, stylus and battery
are excluded from the warranty.
NKC or its authorized agents will repair or replace any products which prove
to be defective during the warranty period, provided these products are used
as prescribed by the operating instructions given in the operator’s and service
manuals.
No other party is authorized to make any warranty or assume liability for NKC’s
products. NKC will not recognize any other warranty, either implied or in writing.
by someone other than NKC or its authorized agents without prior consent of
NKC may be cause for voiding this warranty.
Defective products or parts must be returned to NKC or its authorized agents,
along with an explanation of the failure. Shipping costs must be prepaid.
water or other casualty, improper installation or application, or on which the
originalidenticationmarkshavebeenremoved.
In the USA and Canada other warranty policies may apply.
CAUTION
United States law restricts this device to sale by or on the order of a
physician.
Equipment Authorization Requirement
Operation of this equipment requires the prior coordination with a frequency
coordinator designated by the FCC for the Wireless Medical Telemetry Service.
Operator’s Manual ZM-920PA/921PA/930PA/931PA iii
EMC RELATED CAUTION
This equipment and/or system complies with IEC 60601-2 International
Standard for electromagnetic compatibility for medical electrical equipment
and/or system. However, an electromagnetic environment that exceeds the
limits or levels stipulated in IEC 60601-1-2, can cause harmful interference to
the equipment and/or system or cause the equipment and/or system to fail to
perform its intended function or degrade its intended performance. Therefore,
during the operation of the equipment and/or system, if there is any
undesired deviation from its intended operational performance, you must
avoid, identify and resolve the adverse electromagnetic effect before
continuing to use the equipment and/or system.
The following describes some common interference sources and remedial
actions:
1. Strong electromagnetic interference from a nearby emitter source such as
an authorized radio station or cellular phone:
Install the equipment and/or system at another location. Keep the emitter
source such as cellular phone away from the equipment and/or system, or
turn off the cellular phone.
2. Radio-frequency interference from other equipment through the AC power
supply of the equipment and/or system:
Identify the cause of this interference and if possible remove this
interference source. If this is not possible, use a different power supply.
3. Effect of direct or indirect electrostatic discharge:
Make sure all users and patients in contact with the equipment and/or
system are free from direct or indirect electrostatic energy before using it. A
humid room can help lessen this problem.
4. Electromagnetic interference with any radio wave receiver such as radio or
television:
If the equipment and/or system interferes with any radio wave receiver,
locate the equipment and/or system as far as possible from the radio wave
receiver.
iv Operator’s Manual ZM-910PA/921PA/930PA/931PA
5. Interference of lightning:
When lightning occurs near the location where the equipment and/or
system is installed, it may induce an excessive voltage in the equipment
and/or system.
In such a case, disconnect the AC power cord from the equipment and/or
system and operate the equipment and/or system by battery power, or use
an uninterruptible power supply.
6. Use with other equipment:
When the equipment and/or system is adjacent to or stacked with other
equipment, the equipment and/or system may affect the other equipment.
Before use, check that the equipment and/or system operates normally
with the other equipment.
7. Use of unspecified accessory, transducer and/or cable:
When an unspecified accessory, transducer and/or cable is connected to
this equipment and/or system, it may cause increased electromagnetic
emission or decreased electromagnetic immunity. The specified
configuration of this equipment and/or system complies with the
electromagnetic requirements with the specified configuration. Only use
this equipment and/or system with the specified configuration.
8. Use of unspecified configuration:
When the equipment and/or system is used with the unspecified system
configuration different than the configuration of EMC testing, it may
cause increased electromagnetic emission or decreased electromagnetic
immunity.
Only use this equipment and/or system with the specified configuration.
9. Measurement with excessive sensitivity:
The equipment and/or system is designed to measure bioelectrical signals
with a specified sensitivity. If the equipment and/or system is used with
excessive sensitivity, artifact may appear by electromagnetic interference
and this may cause mis-diagnosis. When unexpected artifact appears,
inspect the surrounding electromagnetic conditions and remove this artifact
source.
If the above suggested remedial actions do not solve the problem, consult
your Nihon Kohden representative for additional suggestions.
Operator’s Manual ZM-920PA/921PA/930PA/931PA v
Conventions Used in this Manual and Instrument
Warnings, Cautions and Notes
Warnings, cautions and notes are used in this manual to alert or signal the reader
tospecicinformation.
WARNING
A warning alerts the user to the possible injury or death associated
with the use or misuse of the instrument.
CAUTION
A caution alerts the user to possible injury or problems with the
instrument associated with its use or misuse such as instrument
malfunction, instrument failure, damage to the instrument, or damage
to other property.
NOTE
A note provides specific information, in the form of recommendations,
prerequirements, alternative methods or supplemental information.
vi Operator’s Manual ZM-910PA/921PA/930PA/931PA
Explanations of the Symbols in this Manual and Instrument
The following symbols found in this manual/instrument bear the respective
descriptions as given.
SymbolDescriptionSymbolDescription
Power On
Attention, consult
operator’s manual
Power OffNurse call
Debrillationprooftype
BF applied part
Debrillationprooftype
CF applied part
Direct current
Replace battery
Check electrode
Operator’s Manual ZM-920PA/921PA/930PA/931PA vii
Introduction
The ZM-920PA/ZM-921PA/ZM-930PA/ZM-931PA transmitter transmits ECG and
other data from a patient to a Nihon Kohden monitor for continuous monitoring.
Available parameters and functions vary between the models. Read the operator’s
manual for the monitor before operation.
Model
ZM-920PA
ZM-921PA
ZM-930PA
ZM-931PA
The transmission frequency range is as follows.
ZM-920PA/930PA: 608.0250 MHz (channel number 9002) to 613.9750 MHz
ZM-921PA/931PA: 1395.0250 MHz (channel number E002) to 1399.9750 MHz
1427.0250 MHz (channel number E502) to 1431.9750 MHz
The transmitter channel can be changed by the QI-901PK Channel Writer. To
change the channel number, refer to the channel writer manual.
ParametersFunctions
The following information is indicated
• ECG
• Impedance respiration
• ECG
• Impedance respiration
• SpO
2
(channel number 9478)
(channel number E398)
(channel number E898)
by LED.
• Check ECG electrodes
• Replace batteries
The following information is indicated
on LCD.
• SpO2 value
• Pulse wave amplitude
• Replace batteries
The following information is indicated
by LED.
• Check ECG electrodes
Operator’s Manual ZM-920PA/921PA/930PA/931PA 1
WARNING
The following actions must be taken to properly receive the
transmitter signal of the correct patient on the receiving instrument.
Otherwise, there may be signal loss or signals may mix causing a
serious accident, such as monitoring a different patient.
• Assign a channel administrator in the hospital and only he or she
should manage channel assignment.
• The channel administrator must manage the channels in the facility
so that there is no signal interference.
• When the transmitter channel is changed, the channel
administrator must check that the channel on the receiving monitor
is also changed and the signal is properly received.
• The channel administrator must replace the channel number label
on the transmitter with the new one after changing the channel.
WARNING
Do not use the same transmitter on more than one patient at the
same time. Do not connect different sensors on different patients to
the same transmitter.
CAUTION
• Do not use the same channel for different patients. If the same
channel is used for two patients, the two patients’ data will be lost
due to mutual modulation interference, or another patient’s data
may appear on the receiving monitor screen.
• Do not use transmitters of adjacent channels in a hospital. If a
transmitter of an adjacent channel is used, radio waves from one
transmitter affect the receiver of the adjacent channel’s transmitter
and there may be interference.
2 Operator’s Manual ZM-920PA/921PA/930PA/931PA
NOTE
• To prevent interference between channels, assign a channel
administrator in the hospital and only he or she should manage channel
assignment.
• Use Nihon Kohden parts and accessories to assure maximum
performance from your instrument.
• For stable signal reception, it is recommended to use a diversity
antenna system on the receiving monitor. Otherwise, spike noise from
transient fading of electric field strength (for example, people moving)
may interfere with the transmitter signal and may be mistaken as an
arrhythmia on the receiving monitor.
• Do not diagnose a patient based on only part of the monitoring data on
the transmitter or only on the data acquired by the transmitter. Overall
judgement must be performed by a physician who understands the
features, limitations and characteristics of the transmitter by reading
this operator’s manual thoroughly and by reading the biomedical signals
acquired by other instruments.
• For details on the receiving monitor and upgrade information, contact
your Nihon Kohden representative.
Operator’s Manual ZM-920PA/921PA/930PA/931PA 3
Panel Description
Top Panel
For attaching a strap
Refer to the warnings below
Refer to the symbol page
ECG/RESP socket
Connects to the electrode lead for measuring ECG
and/or respiration by the impedance method.
ZM-930PA/931PA only
SpO2 socket
Connects to the SpO2 probe.
Refer to the symbol page
Refer to the warnings below
WARNING
When performing defibrillation, discharge as far as possible from
electrodes, patches and any gel, cream or medicine on the chest of
the patient. If there is a possibility that the defibrillator paddle could
touch these materials, remove them from the patient. If the
defibrillator paddle directly contacts these materials, the discharged
energy may cause skin burn to the patient.
WARNING
Before defibrillation, all persons must keep clear of the bed and must
not touch the patient or any equipment or cord connected to the
patient. Failure to follow this warning may cause electrical shock or
injury.
4 Operator’s Manual ZM-920PA/921PA/930PA/931PA
Front Panel
RALALL
Va Vb
USAch
E002
1395.025 MHz
ZM-920PA/921PA
Power switch
Turns transmitter power on or off.
Replace batteries LED
Lights when the batteries
need replacement.
CALL key
When this key is pressed,
a “peep” sounds at the
transmitter, and “CALL”
message appears at the
Check electrodes LED
Lights when the
electrode is detached
from the patient.
monitor. Depending
on the settings on
the monitor, an ECG
waveform is recorded
when this key is pressed.
Refer to the warning below.Battery case
Contains two 1.5 V dry cell batteries (AA).
WARNING
Close the battery case cover during operation. If the transmitter is
used with the battery case cover open, anyone who touches the
opened battery case may receive an electrical shock when
defibrillation is performed. Touching the opened battery case may
cause electrostatic discharge and intermittently interfere with the
waveform or data.
Only use your finger to press the CALL key. Do not press the key with
a sharp object. Otherwise the key may be damaged.
CAUTION
Operator’s Manual ZM-920PA/921PA/930PA/931PA 5
ZM-930PA/931PA
RALALL
Va Vb
USAch
E002
1395.025 MHz
RALALL
Va Vb
USAch
E002
1395.025 MHz
Battery replacement mark:
Appears when the batteries
are weak. Immediately
replace the batteries when
this appears.
For the descriptions except for the
LCD, refer to the “ZM-920PA/921PA”
section on the previous page.
LCD
LCD
%SpO2:
Displayed when the power is
turned on. This indication is not
displayed when SpO2 display is
turned off.
Pulse level bar graph:
Displays pulse level in
SpO2 data:
7 steps.
Displays SpO2 data.
When SpO2 is 41 to 100%, the value is displayed. (ex. )
When SpO2 is under 40%, “ ” is displayed.
When the detected pulse is too small to measure, “” is
displayed.
6 Operator’s Manual ZM-920PA/921PA/930PA/931PA
Important Safety Information
General
WARNING
Never use the transmitter in the presence of any flammable
anesthetic gas or high concentration oxygen atmosphere. Failure to
follow this warning may cause explosion or fire.
WARNING
Never use the transmitter in a hyperbaric oxygen chamber. Failure to
follow this warning may cause explosion or fire.
WARNING
Do not take this transmitter into the MRI test room. This transmitter is
not designed to be used during MRI tests.
WARNING
When performing MRI test, remove all electrodes from the patient
which are connected to this transmitter. Failure to follow this warning
may cause skin burn on the patient. For details, refer to the MRI
manual.
WARNING
When performing defibrillation, discharge as far as possible from
electrodes, patches and any gel, cream or medicine on the chest of
the patient. If there is a possibility that the defibrillator paddle could
touch these materials, remove them from the patient. If the
defibrillator paddle directly contacts these materials, the discharged
energy may cause skin burn to the patient.
Operator’s Manual ZM-920PA/921PA/930PA/931PA 7
WARNING
Before defibrillation, all persons must keep clear of the bed and must
not touch the patient or any equipment or cord connected to the
patient. Failure to follow this warning may cause electrical shock or
injury.
WARNING
When the transmitter is used with an electrosurgical unit (ESU),
firmly attach the entire area of the ESU return plate. Otherwise, the
current from the ESU flows into the electrodes of the transmitter,
causing electrical burn where the electrodes are attached. For
details, refer to the ESU manual.
WARNING
Close the battery case cover during operation. If the transmitter is
used with the battery case cover open, anyone who touches the
opened battery case may receive an electrical shock when
defibrillation is performed. Touching the opened battery case may
cause electrostatic discharge and intermittently interfere with the
waveform or data.
WARNING
If detergent or liquid spills into the transmitter, clean it and dry it
completely before use. If a wet transmitter is used, the patient or
operator may receive an electrical shock or injury.
WARNING
Do not use the same transmitter on more than one patient at the
same time. Do not connect different sensors on different patients to
the same transmitter.
8 Operator’s Manual ZM-920PA/921PA/930PA/931PA
CAUTION
Only use Nihon Kohden specified electrodes, electrode leads and
SpO2 probes. Otherwise, the maximum performance from the
transmitter cannot be guaranteed.
CAUTION
Do not reuse disposable parts and accessories.
CAUTION
Do not shake or swing the transmitter while holding the leads or
cables connected to the transmitter. The transmitter may come off
and injure someone or damage surrounding instruments.
CAUTION
Attach a strap to the transmitter to prevent the transmtiter from falling.
CAUTION
Turn off the power of mobile phones, small wireless devices and
other devices which produce strong electromagnetic interference
around a patient (except for devices allowed by the hospital
administrator). Radio waves from devices such as mobile phones or
small wireless devices may be mistaken as pulse waves and the
displayed data may be incorrect.
Operator’s Manual ZM-920PA/921PA/930PA/931PA 9
CAUTION
• Do not use the same channel for different patients. If the same
channel is used for two patients, the two patients’ data will be lost
due to mutual modulation interference, or another patient’s data
may appear on the receiving monitor screen.
• Do not use transmitters of adjacent channels in a hospital. If a
transmitter of an adjacent channel is used, radio waves from one
transmitter affect the receiver of the adjacent channel’s transmitter
and there may be interference.
Battery
WARNING
• Keep the batteries away from fire. They may explode.
• Keep the batteries away from patients.
• Never short-circuit the + and – terminals on the battery. It may
cause overheating and fire.
• Do not damage, disassemble, drop or give impact to the battery.
WARNING
If the battery is damaged and the substance inside the battery
contacts the eyes or skin, wash immediately and thoroughly with
water and see a physician. Never rub your eyes, because you may
lose your eyesight.
CAUTION
Battery replacement must be performed by the operator. When
replacing batteries of the transmitter currently used for a patient,
disconnect electrode leads from the transmitter before replacing
batteries or do not touch the patient during replacement.
10 Operator’s Manual ZM-920PA/921PA/930PA/931PA
CAUTION
The battery charger must be used outside the patient environment.
CAUTION
Refer to the battery and battery charger manuals for details on
handling the batteries.
Transmitter Channel Management
WARNING
The following actions must be taken to properly receive the
transmitter signal of the correct patient on the receiving instrument.
Otherwise, there may be signal loss or signals may mix causing a
serious accident, such as monitoring a different patient.
• Assign a channel administrator in the hospital and only he or she
should manage channel assignment.
• The channel administrator must manage the channels in the facility
so that there is no signal interference.
• When the transmitter channel is changed, the channel
administrator must check that the channel on the receiving monitor
is also changed and the signal is properly received.
• The channel administrator must replace the channel number label
on the transmitter with the new one after changing the channel.
Operator’s Manual ZM-920PA/921PA/930PA/931PA 11
For Patients Using Implantable Pacemaker
WARNING
Interaction Between Minute Ventilation Rate-Adaptive Pacemakers
and Cardiac Monitoring and Diagnostic Equipment
The bioelectric impedance measurement sensor of a minute
ventilation rate-adaptive implantable pacemaker may be affected by
transmitter which is connected to the same patient. If this occurs, the
pacemaker may pace at its maximum rate and the transmitter may
give incorrect data to the monitor. If this occurs, disconnect the
electrode leads from the patient or change the setting on the
pacemaker by referring to the pacemaker’s manual. For more details,
contact your pacemaker representative or Nihon Kohden
representative.
Output Signal
WARNING
Do not use the output signal from the receiving monitor as the
synchronization signal for other equipment such as IABP, MRI,
echocardiography or defibrillator. There may be time delay between
the monitor and the other equipment caused by waveform
transmission delay and spike noise may interfere on the output signal
and be mistaken as a trigger.
12 Operator’s Manual ZM-920PA/921PA/930PA/931PA
ECG Monitoring
CAUTION
Only use Nihon Kohden specified electrodes and electrode leads.
When other type of electrodes or electrode leads are used, the
“CHECK ELECTRODES” message may be displayed and monitoring
may stop.
CAUTION
When the “ELECTRODE OFF” or “CHECK ELECTRODE” message
is displayed on the receiving monitor, ECG is not monitored properly
and the ECG alarm does not function. Check the electrode, electrode
leads, and if necessary, replace with new ones.
SpO2 Monitoring
WARNING
SpO2 measurement may be incorrect in the following cases.
• When the patient’s carboxyhemoglobin or methemoglobin
increases abnormally.
• When dye is injected in the blood.
• When using an electrosurgical unit.
• During CPR.
• When measuring at a site with venous pulse.
• When there is body movement.
• When the pulse wave is small (insufficient peripheral circulation).
Operator’s Manual ZM-920PA/921PA/930PA/931PA 13
WARNING
Check the circulation condition by observing the skin color at the
measurement site and pulse waveform. Change the measurement
site every 8 hours for disposable probes and every 4 hours for
reusable probes (every 8 hours for TL-630T/TL-631T series probe).
The skin temperature may increase at the attached site by 2 or 3°C
(4 or 5°F) and cause a burn or pressure necrosis. When using the
probe on the following patients, take extreme care and change the
measurement site more frequently according to symptoms and
degree.
• Patient with a fever
• Patient with peripheral circulation insufficiency
• Neonate or low birth weight infant with delicate skin
• Patient who is receiving photodynamic therapy*
* Photodynamic therapy is a treatment to remove the affected tissue by using a
photosensitizing agent and exposing the tissue to light. This treatment has a side
a burn. This probe uses two light wavelengths in the range from 650 to 950 nm.
The maximum light intensity is less than 5.5 mW/sr.
WARNING
When not monitoring SpO2, disconnect the SpO2 cable from the
transmitter. Otherwise, noise from the probe sensor may interfere and
incorrect data is displayed on the screen.
14 Operator’s Manual ZM-920PA/921PA/930PA/931PA
WARNING
• When using the TL-201T finger probe, do not fasten the probe and
cable to the finger by wrapping with tape. This may cause burn,
congestion or pressure necrosis from poor blood circulation.
• When using probes other than the TL-201T finger probe, to avoid
poor circulation, do not wrap the tape too tight. Check the blood
circulation condition by observing the skin color and congestion at
the skin peripheral to the probe attachment site. Even for shortterm monitoring, there may be burn or pressure necrosis from
poor blood circulation, especially on neonates or low birth weight
infants whose skin is delicate. Accurate measurement cannot be
performed on a site with poor peripheral circulation.
CAUTION
While a patient is on medication which causes vasodilation, the pulse
waveform may change and in rare cases the SpO2 value might not be
displayed.
CAUTION
Normal external light does not affect monitoring but strong light such
as a surgical light or sunlight may affect monitoring. If affected, cover
the measuring site with a blanket.
CAUTION
If the attachment site is dirty with blood or bodily fluids, clean the
attachment site before attaching the probe. If there is nail polish on
the attachment site, remove the polish. Otherwise, the amount of
transmitted light decreases, and measured value may be incorrect or
measurement cannot be performed.
Operator’s Manual ZM-920PA/921PA/930PA/931PA 15
CAUTION
Do not use a probe which is deteriorated by aging. Accurate
measurement cannot be performed.
CAUTION
Do not use a damaged or disassembled probe. It causes incorrect
measurement and may injure the patient.
CAUTION
The disposable probe is not sterilized. Use the disposable probe only
for a single patient. Never reuse the disposable probe for another
patient because it causes cross infection.
CAUTION
If the skin gets irritated or redness appears on the skin from the
probe, change the attachment site or stop using the probe. Take
extreme care for the patients with delicate skin.
CAUTION
Do not pull or bend the probe cable, and do not put caster feet on the
probe cable. Do not immerse the probe cable in chemical solutions or
water. Failure to follow these instructions may cause cable
discontinuity, short circuit, skin burn on the patient and incorrect
measurement data. Replace any broken probe with a new one.
CAUTION
When the probe is attached on an appropriate site with sufficient
circulation and the error message confirming the probe attachment
repeatedly appears, the probe may be deteriorated. Replace it with a
new one.
16 Operator’s Manual ZM-920PA/921PA/930PA/931PA
CAUTION
When a message indicates a faulty probe, stop monitoring and
replace the probe with a new one.
CAUTION
When any of the following conditions occurs, the probe may be
broken. Replace it with a new one.
• The transmitter generates “pip” sounds.
• SpO
data is 85% and blinking.
2
CAUTION
Neonatal skin is delicate. Remove the probe and tape carefully and
slowly.
CAUTION
When removing a probe that is taped to the skin, do not pull the
probe cable because this can damage the cable.
CAUTION
When removing the probe from the attachment tape, do not pull the
sensor cable because this can damage the cable.
CAUTION
Do not immerse the disposable probe in detergents or water. If the
probe adhesive surface gets wet, adhesiveness becomes weak and
the probe cannot be attached to the skin.
CAUTION
Refer to the probe instruction manual for details.
Operator’s Manual ZM-920PA/921PA/930PA/931PA 17
Maintenance
CAUTION
If detergent or liquid spills into the transmitter, clean it and dry it
completely before use. If a wet transmitter is used, the transmitter
may malfunction or get damaged.
CAUTION
Never disassemble or repair the transmitter. Disassembly and repair
must be performed by qualified service personnel.
CAUTION
Before cleaning or disinfection, remove the batteries from the
transmitter. Failure to follow this instruction may result in electrical
shock or transmitter malfunction.
CAUTION
The transmitter cannot be sterilized. Sterilizing the transmitter may
damage it.
18 Operator’s Manual ZM-920PA/921PA/930PA/931PA
Preparation
Installing (Replacing) Batteries
Use two AA(R6) type alkaline dry cell batteries, manganese dry cell batteries,
NiCd rechargeable batteries or NiMH batteries.
Battery Lifetime
With new alkaline batteries, the transmitter can continuously measure for the
following number of days. Operation time depends on the thickness of the SpO2
probe attachment site.
Transmitter
ZM-920PA/930PA3 days4 days
ZM-921PA/931PA2.5 days3.5 days
The capacity of manganese, NiCd and NiMH batteries is less than that of
alkaline batteries, therefore the lifetime of the battery is shorter.
TypeLifetime
ManganeseAbout 1/2 of alkaline batteries
NiCdAbout 1/3 of alkaline batteries (when fully charged)
NiMHAbout 1/2 of alkaline batteries (when fully charged)
Recommended Batteries
Alkaline primary: Nihon Kohden Medipower (equivalent to Panasonic LR6 (G))
Operating time (Measuring parameters)
ECG, Resp, SpO
NOTE
2
ECG, Resp
CAUTION
Battery replacement must be performed by the operator. When
replacing batteries of the transmitter currently used for a patient,
disconnect electrode leads from the transmitter before replacing
batteries or do not touch the patient during replacement.
Operator’s Manual ZM-920PA/921PA/930PA/931PA 19
If electrode leads are attached to the patient and the person replacing the batteries
touches the patient, the patient leakage current over the amount allowed may
occur.
CAUTION
• Replace all batteries at the same time.
• Do not use different types of batteries together.
NOTE
Insert the batteries with the correct polarity (+ and −).
Procedure
1. Open the battery case cover.
1
1
3
3
2
2
2. Insert two dry cell batteries (LR6) into
the battery case observing the correct
polarity.
3. Close the cover and press it gently
until it clicks.
WARNING and CAUTION for Battery Handling
WARNING
• Keep the batteries away from fire. They may explode.
• Keep the batteries away from patients.
• Never short-circuit the + and – terminals on the battery. It may
cause overheating and fire.
• Do not damage, disassemble, drop or give impact to the battery.
20 Operator’s Manual ZM-920PA/921PA/930PA/931PA
WARNING
If the battery is damaged and the substance inside the battery
contacts the eyes or skin, wash immediately and thoroughly with
water and see a physician. Never rub your eyes, because you may
lose your eyesight.
CAUTION
Do not handle the batteries with wet hands.
CAUTION
When the transmitter is not in use, remove batteries. When batteries
are installed, battery power is consumed even if measurement is not
performed. Especially, when NiMH batteries remain in the transmitter
when the transmitter is not in use, the battery may become unusable
from overdischarge and leak liquid which will damage the transmitter.
CAUTION
Refer to the battery and battery charger manuals for details on
handling the batteries.
CAUTION
The battery charger must be used outside the patient environment.
NOTE
• When using rechargeable NiCd batteries or NiMH batteries, shallow
charging/discharging shortens battery capacity. For details, refer to the
battery operator’s manual.
• Remove the batteries from the transmitter before disposing of the
transmitter.
Operator’s Manual ZM-920PA/921PA/930PA/931PA 21
Loading...
+ 69 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.