This device is intended for use only by qualified medical personnel.
Use only Nihon Kohden approved products with this device. Use of
non-approved products or in a non-approved manner may affect the
performance specifications of the device. This includes, but is not
limited to, batteries, recording paper, pens, extension cables,
electrode leads, input boxes and AC power.
Please read these precautions thoroughly before attempting to operate
the instrument.
1.To safely and effectively use the instrument, its operation must be
fully understood.
2.When installing or storing the instrument, take the following
precautions:
(1) Avoid moisture or contact with water, extreme atmospheric
pressure, excessive humidity and temperatures, poorly ventilated
areas, and dust, saline or sulphuric air.
(2) Place the instrument on an even, level floor. Avoid vibration and
mechanical shock, even during transport.
(3) Avoid placing in an area where chemicals are stored or where there
is danger of gas leakage.
(4) The power line source to be applied to the instrument must
correspond in frequency and voltage to product specifications, and
have sufficient current capacity.
(5) Choose a room where a proper grounding facility is available.
3.Before Operation
(1) Check that the instrument is in perfect operating order.
(2) Check that the instrument is grounded properly.
(3) Check that all cords are connected properly.
(4) Pay extra attention when the instrument is in combination with
other instruments to avoid misdiagnosis or other problems.
Operator's Manual ZM-920PA/930PAi
(5) All circuitry used for direct patient connection must be doubly
checked.
(6) Check that battery level is acceptable and battery condition is good
when using battery-operated models.
4.During Operation
(1) Both the instrument and the patient must receive continual, careful
attention.
(2) Turn power off or remove electrodes and/or transducers when
necessary to assure the patient’s safety.
(3) Avoid direct contact between the instrument housing and the
patient.
5.To Shutdown After Use
(1) Turn power off with all controls returned to their original positions.
(2) Remove the cords gently; do not use force to remove them.
(3) Clean the instrument together with all accessories for their next
use.
6.The instrument must receive expert, professional attention for
maintenance and repairs. When the instrument is not functioning
properly, it should be clearly marked to avoid operation while it is
out of order.
7.The instrument must not be altered or modified in any way.
8.Maintenance and Inspection:
(1) The instrument and parts must undergo regular maintenance
inspection at least every 6 months.
(2) If stored for extended periods without being used, make sure prior
to operation that the instrument is in perfect operating condition.
(3) Technical information such as parts list, descriptions, calibration
instructions or other information is available for qualified user
technical personnel upon request from your Nihon Kohden
representative.
iiOperator's Manual ZM-920PA/930PA
9.When the instrument is used with an electrosurgical instrument,
pay careful attention to the application and/or location of electrodes
and/or transducers to avoid possible burn to the patient.
10. When the instrument is used with a defibrillator, make sure that the
instrument is protected against defibrillator discharge. If not,
remove patient cables and/or transducers from the instrument to
avoid possible damage.
WARRANTY POLICY
Nihon Kohden Corporation (NKC) shall warrant its products against all
defects in materials and workmanship for one year from the date of delivery.
However, consumable materials such as recording paper, ink, stylus and
battery are excluded from the warranty.
NKC or its authorized agents will repair or replace any products which
prove to be defective during the warranty period, provided these products
are used as prescribed by the operating instructions given in the operator’s
and service manuals.
No other party is authorized to make any warranty or assume liability for
NKC’s products. NKC will not recognize any other warranty, either implied
or in writing. In addition, service, technical modification or any other
product change performed by someone other than NKC or its authorized
agents without prior consent of NKC may be cause for voiding this
warranty.
Defective products or parts must be returned to NKC or its authorized
agents, along with an explanation of the failure. Shipping costs must be prepaid.
This warranty does not apply to products that have been modified,
disassembled, reinstalled or repaired without Nihon Kohden approval or
which have been subjected to neglect or accident, damage due to accident,
Operator's Manual ZM-920PA/930PAiii
fire, lightning, vandalism, water or other casualty, improper installation or
application, or on which the original identification marks have been removed.
In the USA and Canada other warranty policies may apply.
CAUTION
United States law restricts this device to sale by or on the order
of a physician.
Equipment Authorization Requirement
Operation of this equipment requires the prior coordination with a frequency
coordinator designated by the FCC for the Wireless Medical Telemetry
Service.
ivOperator's Manual ZM-920PA/930PA
EMC RELATED CAUTION
This equipment and/or system complies with the International
Standard IEC60601-1-2 for electromagnetic compatibility for
medical electrical equipment and/or system. However, an
electromagnetic environment that exceeds the limits or levels
stipulated in the IEC60601-1-2, can cause harmful interference to
the equipment and/or system or cause the equipment and/or
system to fail to perform its intended function or degrade its
intended performance. Therefore, during the operation of the
equipment and/or system, if there is any undesired deviation
from its intended operational performance, you must avoid,
identify and resolve the adverse electromagnetic effect before
continuing to use the equipment and/or system.
The following describes some common interference sources and
remedial actions:
1.Strong electromagnetic interference from a nearby emitter
source such as an authorized radio station or cellular phone:
Install the equipment and/or system at another location if it is
interfered with by an emitter source such as an authorized
radio station. Keep the emitter source such as cellular phone
away from the equipment and/or system.
2.Effect of direct or indirect electrostatic discharge:
Make sure all users and patients in contact with the equipment
and/or system are free from direct or indirect electrostatic
energy before using it. A humid room can help lessen this
problem.
3.Electromagnetic interference with any radio wave receiver
such as radio or television:
If the equipment and/or system interferes with any radio wave
receiver, locate the equipment and/or system as far as
possible from the radio wave receiver.
Operator's Manual ZM-920PA/930PAv
If the above suggested remedial actions do not solve the
problem, consult your Nihon Kohden representative for additional
suggestions.
viOperator's Manual ZM-920PA/930PA
Conventions Used in this Manual and Instrument
Warnings, Cautions and Notes
Warnings, cautions and notes are used in this manual to alert or signal the
reader to specific information.
WARNING
A warning alerts the user to the possible injury or death
associated with the use or misuse of the instrument.
CAUTION
A caution alerts the user to possible injury or problems with the
instrument associated with its use or misuse such as instrument
malfunction, instrument failure, damage to the instrument, or
damage to other property.
NOTE
A note provides specific information, in the form of
recommendations, prerequirements, alternative methods or
supplemental information.
Operator's Manual ZM-920PA/930PAvii
Explanations of the Symbols in this Manual and Instrument
The following symbols found in this manual/instrument bear the respective
descriptions as given.
SymbolDescriptionSymbolDescription
Power OnAttention, consult
operator’s manual
Power OffNurse call
Defibrillation proof
type BF applied part
Defibrillation proof
type CF applied part
Direct current
Replace battery
Check electrode
viiiOperator's Manual ZM-920PA/930PA
Introduction
Model Parameters Functions
ZM-920PA
• ECG
• Impedance respiration
The following information is
indicated by LED.
• Check ECG electrodes
• Replace batteries
ZM-930PA
•
ECG
• Impedance respiration
• SpO
2
The following information is
indicated on LCD.
• SpO
2
value
• Pulse wave amplitude
• Replace batteries
The following information is
indicated by LED.
• Check ECG electrodes
The ZM-920PA/930PA transmitter transmits ECG and other data from a patient to
a Nihon Kohden monitor for continuous monitoring. Available parameters and
functions vary between the models. Read the operator’s manual for the monitor
before operation.
The transmitter channel can be changed by the QI-901PK Channel Writer. To
change the channel number, refer to the channel writer manual.
The following actions must be taken to properly receive the
transmitter signal of the correct patient on the receiving instrument,
otherwise, there may be signal loss or signals may mix causing a
serious accident, such as monitoring a different patient.
••
• Assign a channel administrator in the hospital and only he or she
••
should manage channel assignments.
••
• The channel administrator must manage the channels in the
••
facility so that there is no signal interference.
••
• When the transmitter channel is changed, the channel
••
administrator must check that the channel on the receiving
monitor is also changed and the signal is properly received.
Operator's Manual ZM-920PA/930PA1
WARNING
••
• The channel administrator must replace the channel number label
••
on the transmitter with the new one after changing the channel.
CAUTION
• Do not use the same channel for different patients. If the same
channel is used for two patients, the two patients’ data will be lost
due to mutual modulation interference, or another patient’s data
may appear on the receiving monitor screen.
• Do not use transmitters of adjacent channels in a hospital. If a
transmitter of an adjacent channel is used, radio waves from one
transmitter affects the receiver of the adjacent channel’s
transmitter and there may be interference.
• Do not use the same transmitter on more than one patient at the
same time. Do not connect different sensors on different patients
to the same transmitter.
NOTE
• To prevent interference between channels, assign a channel
administrator in the hospital and only he or she should manage
channel assignment.
• Use Nihon Kohden parts and accessories to assure maximum
performance from your instrument.
• For stable signal reception, it is recommended to use a diversity
antenna system on the receiving monitor. Otherwise, spike noise
from transient fading of electric field strength (for example, people
moving) may interfere with the transmitter signal and may be
mistaken as an arrhythmia on the receiving monitor.
• For details on the receiving monitor and upgrade information,
contact your Nihon Kohden representative.
2Operator's Manual ZM-920PA/930PA
Panel Description
Top Panel
For attaching a strap
Refer to the warnings below
Refer to the symbol page
ECG/RESP socket
Connects to the electrode lead for measuring ECG
and/or respiration by the impedance method.
ZM-930PA only
socket
SpO
2
Connects to the SpO
Refer to the symbol page
Refer to the warnings below
probe.
2
WARNING
• Before defibrillation, check that the electrode leads and SpO
probe attached to the patient are properly connected to the
transmitter. Touching the metal parts of disconnected leads and
probes may cause electrical shock or injury by discharged energy.
• When performing defibrillation, discharge as far as possible from
electrodes, patches and any gel, cream or medicine on the chest
of the patient. If there is a possibility that the defibrillator paddle
could touch these materials, remove them from the patient. If the
defibrillator paddle directly contacts these materials, the
discharged energy may cause skin burn to the patient.
• Before defibrillation, all persons must keep clear of the bed and
must not touch the patient or any equipment connected to the
patient. Failure to follow this warning may cause electrical shock
or injury.
2
Operator's Manual ZM-920PA/930PA3
Front Panel
ZM-920PA
Power switch
Turns transmitter power on or off.
CALL key
When this key is pressed,
Replace batteries LED
Lights when the batteries
need replacement.
Check electrodes LED
Lights when the electrode is
detached from the patient.
Refer to the warning below.
RALALL
Va Vb
USAch
9002
608.025 MHz
Battery case
Contains two 1.5 V dry cell batteries (AA).
a "peep" sounds at the
transmitter, and "CALL"
message appears at the
monitor. Depending on the
settings on the monitor, an
ECG waveform is recorded
when this key is pressed.
Channel lable
WARNING
Close the battery case cover during operation. If the transmitter is
used with the battery case cover open, anyone who touches the
opened battery case may receive an electrical shock when
defibrillation is performed. Touching the opened battery case may
cause electrostatic discharge and intermittently interfere with the
waveform or data.
CAUTION
Only use your finger to press the CALL key. Do not press the key
with a sharp object, otherwise the key may be damaged.
4Operator's Manual ZM-920PA/930PA
ZM-930PA
RALALL
Va Vb
USAch
9002
608.025 MHz
Battery replacement mark:
Appears when the batteries
are weak. Immediately
replace the batteries when this
appears.
For the descriptions except for the LCD,
refer to the “ZM-920PA” section on the
previous page.
LCD
%SpO2:
Displayed when the power is turned
on. This indication is not displayed
when SpO
display is turned off.
2
Pulse level bar graph:
Displays pulse level in 7
steps.
SpO2 data:
Displays SpO
When SpO
When SpO
When the detected pulse is too small to measure, “
Operator's Manual ZM-920PA/930PA5
data.
2
is 41 to 100%, the value is displayed. (ex. )
2
is under 40%, “” is displayed.
2
” is displayed.
Important Safety Information
General
WARNING
• Never use the transmitter in the presence of any flammable
anesthetic gas or high concentration oxygen atmosphere. Failure
to follow this warning may cause explosion or fire.
• Never use the transmitter in a hyperbaric oxygen chamber.
Failure to follow this warning may cause explosion or fire.
• Never take this transmitter into an MRI test room.
• When performing MRI test, remove all electrodes from the patient
which are connected to this instrument. Failure to follow this
warning may cause skin burn on the patient. For details, refer to
the MRI manual.
• Before defibrillation, check that the electrode leads and SpO
probe attached to the patient are properly connected to the
transmitter. Touching the metal parts of disconnected leads and
probes may cause electrical shock or injury by discharged energy.
• When performing defibrillation, discharge as far as possible from
electrodes, patches and any gel, cream or medicine on the chest
of the patient. If there is a possibility that the defibrillator paddle
could touch these materials, remove them from the patient. If the
defibrillator paddle directly contacts these materials, the
discharged energy may cause skin burn to the patient.
• Before defibrillation, all persons must keep clear of the bed and
must not touch the patient or any equipment connected to the
patient. Failure to follow this warning may cause electrical shock
or injury.
• When the transmitter is used with an electrosurgical unit (ESU),
firmly attach the entire area of the ESU return plate. Otherwise,
the current from the ESU flows into the electrodes of the
transmitter, causing electrical burn where the electrodes are
attached. For details, refer to the ESU manual.
2
6Operator's Manual ZM-920PA/930PA
• Close the battery case cover during operation. If the transmitter is
used with the battery case cover open, anyone who touches the
opened battery case may receive an electrical shock when
defibrillation is performed. Touching the opened battery case may
cause electrostatic discharge and intermittently interfere with the
waveform or data.
CAUTION
• Only use Nihon Kohden specified electrodes, electrode leads and
SpO
probes. Otherwise, the maximum performance from the
2
instrument cannot be guaranteed.
• Do not reuse disposable items.
• Do not shake or swing the transmitter while holding the leads or
cables connected to the transmitter. The transmitter may come off
and injure someone or damage surrounding instruments.
• Attach a strap to the transmitter to prevent the transmitter from
falling.
• Turn off the power of mobile phones, small wireless devices and
other devices which produce strong electromagnetic interference
around a patient (except for devices allowed by the hospital
administrator). Radio waves from devices such as mobile phones
or small wireless devices may be mistaken as respiration or pulse
waves and the displayed data may be incorrect.
• Do not use the same channel for different patients. If the same
channel is used for two patients, the two patients’ data will be lost
due to mutual modulation interference, or another patient’s data
may appear on the receiving monitor screen.
• Do not use transmitters of adjacent channels in a hospital. If a
transmitter of an adjacent channel is used, radio waves from one
transmitter affects the receiver of the adjacent channel’s
transmitter and there may be interference.
Operator's Manual ZM-920PA/930PA7
Battery
WARNING
• Keep the battery pack away from fire. It may explode.
• Do not damage, disassemble, drop or give impact to the battery. If
the battery is damaged and substance inside the battery contacts
the skin or clothes, wash immediately and thoroughly with water.
• Never short-circuit the + and – terminals on the battery. It may
cause overheating and fire.
• Take care that the patient does not swallow batteries.
CAUTION
• Battery replacement must be performed by the operator. When
replacing batteries of the transmitter currently used for a patient,
disconnect electrode leads from the transmitter before replacing
batteries or do not touch the patient during replacement.
• The battery charger must be used outside the patient environment.
• Refer to the battery and battery charger manuals for details on
handling the batteries.
Transmitter Channel Management
WARNING
The following actions must be taken to properly receive the
transmitter signal of the correct patient on the receiving instrument,
otherwise, there may be signal loss or signals may mix causing a
serious accident, such as monitoring a different patient.
••
• Assign a channel administrator in the hospital and only he or she
••
should manage channel assignments.
••
• The channel administrator must manage the channels in the
••
facility so that there is no signal interference.
••
• When the transmitter channel is changed, the channel
••
administrator must check that the channel on the receiving
monitor is also changed and the signal is properly received.
8Operator's Manual ZM-920PA/930PA
Loading...
+ 49 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.