If you have any comments or suggestions on this
manual, please contact us at:
www.nihonkohden.com
Transmitter
ZM-540PA/ZM-541PA
0614-902754
Copyright Notice
The entire contents of this manual are copyrighted by Nihon Kohden. All rights are reserved. No part
of this document may be reproduced, stored, or transmitted in any form or by any means (electronic,
mechanical, photocopied, recorded, or otherwise) without the prior written permission of Nihon
Kohden.
GENERAL HANDLING PRECAUTIONS
This device is intended for use only by qualified medical personnel.
Use only Nihon Kohden approved products with this device. Use of non-approved
products or in a non-approved manner may affect the performance specifications
of the device. This includes, but is not limited to, batteries, recording paper, pens,
extension cables, electrode leads, input boxes and AC power.
Please read these precautions thoroughly before attempting to operate the instrument.
1. To safely and effectively use the instrument, its operation must be fully understood.
2. When installing or storing the instrument, take the following precautions:
(1) Avoid moisture or contact with water, extreme atmospheric pressure, excessive humidity and
temperatures, poorly ventilated areas, and dust, saline or sulphuric air.
(2) Place the instrument on an even, level floor. Avoid vibration and mechanical shock, even
during transport.
(3) Avoid placing in an area where chemicals are stored or where there is danger of gas leakage.
(4) The power line source to be applied to the instrument must correspond in frequency and
voltage to product specifications, and have sufficient current capacity.
(5) Choose a room where a proper grounding facility is available.
3. Before Operation
(1) Check that the instrument is in perfect operating order.
(2) Check that the instrument is grounded properly.
(3) Check that all cords are connected properly.
(4) Pay extra attention when the instrument is in combination with other instruments to avoid
misdiagnosis or other problems.
(5) All circuitry used for direct patient connection must be doubly checked.
(6) Check that battery level is acceptable and battery condition is good when using
batteryoperated models.
4. During Operation
(1) Both the instrument and the patient must receive continual, careful attention.
(2) Turn power off or remove electrodes and/or transducers when necessary to assure the
patient’s safety.
(3) Avoid direct contact between the instrument housing and the patient.
5. To Shutdown After Use
(1) Turn power off with all controls returned to their original positions.
(2) Remove the cords gently; do not use force to remove them.
(3) Clean the instrument together with all accessories for their next use.
Operator’s Manual ZM-540PA/541PA i
6. The instrument must receive expert, professional attention for maintenance and repairs.
When the instrument is not functioning properly, it should be clearly marked to avoid
operation while it is out of order.
7. The instrument must not be altered or modified in any way.
8. Maintenance and Inspection
(1) The instrument and parts must undergo regular maintenance inspection at least every 6
months.
(2) If stored for extended periods without being used, make sure prior to operation that the
instrument is in perfect operating condition.
(3) Technical information such as parts list, descriptions, calibration instructions or other
information is available for qualified user technical personnel upon request from your Nihon
Kohden representative.
9. When the instrument is used with an electrosurgical instrument, pay careful attention to
the application and/or location of electrodes and/or transducers to avoid possible burn to
the patient.
10. When the instrument is used with a defibrillator, make sure that the instrument is
protected against defibrillator discharge. If not, remove patient cables and/or transducers
from the instrument to avoid possible damage.
ii Operator’s Manual ZM-540PA/541PA
WARRANTY POLICY
Nihon Kohden Corporation (NKC) shall warrant its products against all defects in materials and
workmanship for one year from the date of delivery. However, consumable materials such as
recording paper, ink, stylus and battery are excluded from the warranty.
NKC or its authorized agents will repair or replace any products which prove to be defective during
the warranty period, provided these products are used as prescribed by the operating instructions
given in the operator’s and service manuals.
No other party is authorized to make any warranty or assume liability for NKC’s products.
NKC will not recognize any other warranty, either implied or in writing. In addition, service,
technical modification or any other product change performed by someone other than NKC or its
authorized agents without prior consent of NKC may be cause for voiding this warranty.
Defective products or parts must be returned to NKC or its authorized agents, along with an
explanation of the failure. Shipping costs must be pre-paid.
This warranty does not apply to products that have been modified, disassembled, reinstalled or
repaired without Nihon Kohden approval or which have been subjected to neglect or accident,
damage due to accident, fire, lightning, vandalism, water or other casualty, improper installation or
application, or on which the original identification marks have been removed.
In the USA and Canada other warranty policies may apply.
CAUTION
United States law restricts this product to sale by or on the order of a physician.
Operator’s Manual ZM-540PA/541PA iii
Equipment Authorization Requirement
This device complies with Part 15 of the FCC (Federal Communications Commission) Rules.
Operation is subject to the following two conditions:
(1) this device may not cause harmful interference, and
(2) this device must accept any interference received, including interference that may cause
undesired operation.
This device complies with Part 95 Subpart H of the FCC Rules to be used in wireless medical
telemetry service.
Operation of this equipment requires the prior coordination with a frequency coordinator designated
by FCC for the Wireless Medical Telemetry Service.
CAUTION
To comply with the FCC radio frequency (RF) exposure compliance requirements, no
change to the antenna or the device is permitted. Any change to the antenna or the
device could result in the device, exceeding the RF exposure requirements and void
user’s authority to operate this device.
NOTE
• Use this device only indoors.
• This device has been tested and complies with FCC radiation exposure limits set forth
for an uncontrolled environment. The RF transmission power from the antenna conforms
to the general public FCC RF Exposure Guidelines in Supplement C to OET65 limit of
Specific Absorption Rate (SAR) 1.6 W/kg. The maximum SAR value measured from this
device was extremely smaller than 1.6 W/kg. This device must not be located together
with or operated in conjunction with any other unspecified antenna or transmitter.
• The devices require registration and deployment by an authorized frequency coordinator.
The ASHE (American Society for Healthcare Engineering) has been designated by the
FCC to manage the WMTS frequencies. This device has frequency bands which may
not be used in some areas. For details, contact your Nihon Kohden representative. For
details on the guidelines, refer to the ASHE home page.
iv Operator’s Manual ZM-540PA/541PA
EMC RELATED CAUTION
This equipment and/or system complies with IEC 60601-2 International Standard for
electromagnetic compatibility for medical electrical equipment and/or system.
However, an electromagnetic environment that exceeds the limits or levels stipulated
in IEC 60601-1-2, can cause harmful interference to the equipment and/or system or
cause the equipment and/or system to fail to perform its intended function or degrade
its intended performance. Therefore, during the operation of the equipment and/or
system, if there is any undesired deviation from its intended operational performance,
you must avoid, identify and resolve the adverse electromagnetic effect before
continuing to use the equipment and/or system.
The following describes some common interference sources and remedial actions:
1. Strong electromagnetic interference from a nearby emitter source such as an
authorized radio station or cellular phone:
Install the equipment and/or system at another location. Keep the emitter source
such as cellular phone away from the equipment and/or system, or turn off the
cellular phone.
2. Radio-frequency interference from other equipment through the AC power supply of
the equipment and/or system:
Identify the cause of this interference and if possible remove this interference
source. If this is not possible, use a different power supply.
3. Effect of direct or indirect electrostatic discharge:
Make sure all users and patients in contact with the equipment and/or system are
free from direct or indirect electrostatic energy before using it. A humid room can
help lessen this problem.
4. Electromagnetic interference with any radio wave receiver such as radio or
television:
If the equipment and/or system interferes with any radio wave receiver, locate the
equipment and/or system as far as possible from the radio wave receiver.
5. Interference of lightning:
When lightning occurs near the location where the equipment and/or system is
installed, it may induce an excessive voltage in the equipment and/or system.
In such a case, disconnect the AC power cord from the equipment and/or
system and operate the equipment and/or system by battery power, or use an
uninterruptible power supply.
Operator’s Manual ZM-540PA/541PA v
6. Use with other equipment:
When the equipment and/or system is adjacent to or stacked with other equipment,
the equipment and/or system may affect the other equipment. Before use, check
that the equipment and/or system operates normally with the other equipment.
7. Use of unspecified accessory, transducer and/or cable:
When an unspecified accessory, transducer and/or cable is connected to this
equipment and/or system, it may cause increased electromagnetic emission
or decreased electromagnetic immunity. The specified configuration of this
equipment and/or system complies with the electromagnetic requirements with the
specified configuration. Only use this equipment and/or system with the specified
configuration.
8. Use of unspecified configuration:
When the equipment and/or system is used with the unspecified system
configuration different than the configuration of EMC testing, it may cause
increased electromagnetic emission or decreased electromagnetic immunity.
Only use this equipment and/or system with the specified configuration.
9. Measurement with excessive sensitivity:
The equipment and/or system is designed to measure bioelectrical signals with
a specified sensitivity. If the equipment and/or system is used with excessive
sensitivity, artifact may appear by electromagnetic interference and this may
cause mis-diagnosis. When unexpected artifact appears, inspect the surrounding
electromagnetic conditions and remove this artifact source.
If the above suggested remedial actions do not solve the problem, consult your Nihon
Kohden representative for additional suggestions.
vi Operator’s Manual ZM-540PA/541PA
Conventions Used in this Manual and Instrument
Warnings, Cautions and Notes
Warnings, cautions and notes are used in this manual to alert or signal the reader to specific
information.
WARNING
A warning alerts the user to the possible injury or death associated with the use or
misuse of the instrument.
CAUTION
A caution alerts the user to possible injury or problems with the instrument associated
with its use or misuse such as instrument malfunction, instrument failure, damage to
the instrument, or damage to other property.
NOTE
A note provides specific information, in the form of recommendations, prerequirements,
alternative methods or supplemental information.
Operator’s Manual ZM-540PA/541PA vii
Explanations of the Symbols in this Manual and Instrument
The following symbols found in this manual/instrument bear the respective descriptions as given.
On Panel
SymbolDescriptionSymbolDescription
Change screen
Attention, consult operator’s
manual
Defibrillation proof type BF
applied part
Defibrillation proof type CF
applied part
Moves cursor, scrolls dataSerial number
Direction for attaching battery
cover
Direction for inserting battery
Year of manufacture
RF transmitter
Non-ionizing radiation
Direct currentCSA mark
On LCD
SymbolDescriptionSymbolDescription
Full battery
Battery very weak
Replace battery
Battery 2/3 fullAlarm suspended
Low battery
NIBP cannot be measured
QRS/pulse sync mark
viii Operator’s Manual ZM-540PA/541PA
Intended Use
General
The ZM-540PA and ZM-541PA transmitter transmits ECG, respiration and pulse waveforms, SpO2
and NIBP data from a patient to a Nihon Kohden monitor for continuous monitoring. The front LCD
displays ECG (or pulse wave), numeric values of monitoring parameters, NIBP measuring mode and
interval, messages and battery condition.* It also displays the compressed waveform and numeric
data of the latest 10 minutes.
* Essential performance of this transmitter.
The difference between the ZM-540PA and ZM-541PA is the transmission frequency range.
ZM-540PA: 608.0250 MHz (channel number 9002) to 613.9750 MHz (channel number 9478)
ZM-541PA: 1395.0250 MHz (channel number E002) to 1399.9750 MHz (channel number E398)
1427.0250 MHz (channel number E502) to 1431.9750 MHz (channel number E898)
The transmitter channel can be changed with a QI-901PK channel writer.
Read the operator’s manual for the receiving monitor together with this manual before use.
WARNING
Do not use the same transmitter on more than one patient at the same time. Do not
connect different sensors on different patients to the same transmitter.
CAUTION
• Do not use the same channel for different patients. If the same channel is used
for two patients, the two patients’ data will be lost due to mutual modulation
interference, or another patient’s data may appear on the receiving monitor screen.
• Do not use two transmitters with adjacent channels in the same hospital. If
transmitters with adjacent channels are used, their radio waves interfere with each
other.
CAUTION
Be aware that signal loss and artifact may occur because of the multipath
cancellation* when using a transmitter.
* Multipath Cancellation (Standing Wave Interference)
When a radio wave reflects off a surface, there may be some points in the room where the
Operator’s Manual ZM-540PA/541PA 1
reflected and direct waves are exactly out of phase. At these points in the room, the reflected
and direct waves cancel each other out so that the signal strength is 0. This is called “multipath
cancellation” or “standing wave interference”. Locations where signal loss occurs are called “null
spots”. If the transmitter is moving or nearby people or objects are moving, null spots can occur
anytime and anywhere.
NOTE
• To prevent interference between channels, assign a channel administrator in the hospital
and only he or she should manage channel assignment.
• Use Nihon Kohden parts and accessories to assure maximum performance from your
instrument.
• For stable signal reception, it is recommended to use a diversity antenna system on the
receiving monitor. Otherwise, spike noise from transient fading of electric field strength
(for example, people moving) may interfere with the transmitter signal and may be
mistaken as an arrhythmia on the receiving monitor.
• For details on antennas and antenna construction, contact your Nihon Kohden
representative. You can also refer to the Telemetry System Installation Guide.
• Do not diagnose a patient based on only part of the monitoring data on the transmitter or
only on the data acquired by the transmitter. Overall judgment must be performed by a
physician who understands the features, limitations and characteristics of the transmitter
by reading this operator’s manual thoroughly and by reading the biomedical signals
acquired by other instruments.
Receiving Monitor
Any Nihon Kohden receiving monitor (central monitor with multiple patient receiver) can receive
signals from this transmitter as long as the protocol version and channel setting are the same on the
receiving monitor and transmitter.
NOTE
• For details on the receiving monitor and upgrade information, contact your Nihon Kohden
representative.
• The transmitter does not give any patient alarm, only a “no battery” alarm. Patient alarms
must be managed on the receiving monitor.
7 Lead/Scroll keys On the numeric and waveform screen, these keys change the ECG
lead.
On the waveform review screen, these keys scroll data.
On a SETUP screen, these keys move the cursor.
8 Battery case Contains three 1.5 V dry cell batteries (AA TYPE).
WARNING
Close the battery case cover during operation. If the transmitter is used with the
battery case cover open, anyone who touches the opened battery case may receive
an electrical shock when defibrillation is performed. Touching the opened battery case
may cause electrostatic discharge and intermittently interfere with the waveform or
data.
4 Operator’s Manual ZM-540PA/541PA
Rear Panel
Lock plate
Fastens the transmitter to an NIBP cuff.
Refer to the symbol
page.
Refer to the WARNING
below.
WARNING
This transmitter is not waterproof. If detergent or liquid spills into the transmitter, stop
using it and contact your Nihon Kohden representative. If a wet transmitter is used,
the patient or operator may receive an electrical shock or injury.
Operator’s Manual ZM-540PA/541PA 5
Top Panel
Refer to the WARNING on the
next page.
NIBP socket:
Connects the cuff
hose.
Refer to the symbol page.
Bottom Panel
Channel number label:
Indicates the channel number of the
transmitter. Attach the channel number label
to the panel of the receiving monitor.
Refer to the WARNING on the
next page.
Refer to the symbol
page.
ECG/impedance RESP socket:
Connects the electrode lead for measuring
ECG and/or respiration (impedance).
Refer to the symbol page.
Refer to the WARNING
on the next page.
6 Operator’s Manual ZM-540PA/541PA
SpO2 socket:
Connects the SpO2
probe.
WARNING
Before defibrillation, all persons must
keep clear of the bed and must not
touch the patient or any equipment or
cord connected to the patient. Failure
to follow this warning may cause
electrical shock or injury.
WARNING
When performing defibrillation,
discharge as far as possible from
electrodes, patches and any gel,
cream or medicine on the chest of the
patient. If there is a possibility that the
defibrillator paddle could touch these
materials, remove them from the
patient. If the defibrillator paddle
directly contacts these materials, the
discharged energy may cause skin
burn to the patient.
WARNING
When the transmitter is used with an
electrosurgical unit (ESU), firmly
attach the entire area of the ESU
return plate. Otherwise, the current
from the ESU flows into the electrodes
of the transmitter, causing electrical
burn where the electrodes are
attached. For details, refer to the ESU
manual.
CAUTION
Do not shake or swing the transmitter
while holding the leads or cables
connected to the transmitter. The
transmitter may come off and injure
someone or damage surrounding
instruments.
Operator’s Manual ZM-540PA/541PA 7
Important Safety Information
General
WARNING
Never use the transmitter in the
presence of any flammable anesthetic
gas or high concentration oxygen
atmosphere. Failure to follow this
warning may cause explosion or fire.
WARNING
Do not take this transmitter into the
MRI test room. This transmitter is not
designed to be used during MRI tests.
WARNING
Before defibrillation, all persons must
keep clear of the bed and must not
touch the patient or any equipment or
cord connected to the patient. Failure
to follow this warning may cause
electrical shock or injury.
WARNING
Never use the transmitter in a
hyperbaric oxygen chamber. Failure to
follow this warning may cause
explosion or fire.
WARNING
When performing MRI test, remove all
electrodes, probe and cuff from the
patient which are used with this
transmitter. Failure to follow this
warning may cause skin burn on the
patient. For details, refer to the MRI
manual.
WARNING
When performing defibrillation,
discharge as far as possible from
electrodes, patches and any gel,
cream or medicine on the chest of the
patient. If there is a possibility that the
defibrillator paddle could touch these
materials, remove them from the
patient. If the defibrillator paddle
directly contacts these materials, the
discharged energy may cause skin
burn to the patient.
8 Operator’s Manual ZM-540PA/541PA
WARNING
When the transmitter is used with an
electrosurgical unit (ESU), firmly
attach the entire area of the ESU
return plate. Otherwise, the current
from the ESU flows into the electrodes
of the transmitter, causing electrical
burn where the electrodes are
attached. For details, refer to the ESU
manual.
WARNING
This transmitter is not waterproof. If
detergent or liquid spills into the
transmitter, stop using it and contact
your Nihon Kohden representative. If a
wet transmitter is used, the patient or
operator may receive an electrical
shock or injury.
WARNING
Close the battery case cover during
operation. If the transmitter is used
with the battery case cover open,
anyone who touches the opened
battery case may receive an electrical
shock when defibrillation is performed.
Touching the opened battery case
may cause electrostatic discharge and
intermittently interfere with the
waveform or data.
CAUTION
Only use Nihon Kohden specified
electrodes, electrode leads, SpO
probes, and NIBP cuffs. Otherwise,
the maximum performance from the
transmitter cannot be guaranteed.
2
CAUTION
The measurement values and
displayed waveforms on the
transmitter and receiving monitor may
be different due to timing delay of the
display or difference in detection
settings.
WARNING
Do not use the same transmitter on
more than one patient at the same
time. Do not connect different sensors
on different patients to the same
transmitter.
CAUTION
Do not reuse disposable parts and
accessories.
CAUTION
Do not shake or swing the transmitter
while holding the leads or cables
connected to the transmitter. The
transmitter may come off and injure
someone or damage surrounding
instruments.
Operator’s Manual ZM-540PA/541PA 9
CAUTION
Be aware that signal loss and artifact
may occur because of the multipath
cancellation* when using a transmitter.
* Multipath Cancellation (Standing Wave
Interference)
When a radio wave reflects off a surface,
there may be some points in the room where
the reflected and direct waves are exactly
out of phase. At these points in the room, the
reflected and direct waves cancel each other
out so that the signal strength is 0. This is
called “multipath cancellation” or “standing
wave interference”. Locations where signal
loss occurs are called “null spots”. If the
transmitter is moving or nearby people or
objects are moving, null spots can occur
anytime and anywhere.
CAUTION
Turn off the power of mobile phones,
small wireless devices and other
devices which produce strong
electromagnetic interference around a
patient (except for devices allowed by
the hospital administrator). Radio
waves from devices such as mobile
phones or small wireless devices may
be mistaken as pulse waves and the
displayed data may be incorrect.
CAUTION
• Do not use the same channel
for different patients. If the same
channel is used for two patients, the
two patients’ data will be lost due to
mutual modulation interference, or
another patient’s data may appear
on the receiving monitor screen.
• Do not use two transmitters with
adjacent channels in the same
hospital. If transmitters with
adjacent channels are used, their
radio waves interfere with each
other.
Output Signal
WARNING
Do not use the output signal from the receiving monitor as the synchronization signal
for other equipment such as IABP, MRI, echocardiography or defibrillator. There may
be time delay between the monitor and the other equipment caused by waveform
transmission delay and spike noise may interfere on the output signal and be
mistaken as a trigger.
10 Operator’s Manual ZM-540PA/541PA
Battery
CAUTION
Battery replacement must be
performed by the operator. When
replacing the batteries of a transmitter
that is currently used for a patient,
disconnect the electrode leads from
the transmitter before replacing
batteries or do not touch the patient
during replacement.
Refer to the battery and battery
charger manuals for details on
handling the batteries.
CAUTION
Transmitter Channel Management
WARNING
The following actions must be taken to properly receive the transmitter signal of the
correct patient on the receiving instrument. Otherwise, there may be signal loss or
signals may mix causing a serious accident, such as monitoring a different patient.
• Assign a channel administrator in the hospital and only he or she should manage
channel assignment.
• The channel administrator must manage the channels in the facility so that there is
no signal interference.
• When the transmitter channel is changed, the channel administrator must check
that the channel on the receiving monitor is also changed and the signal is properly
received.
• The channel administrator must replace the channel number label on the
transmitter with the new one after changing the channel.
Operator’s Manual ZM-540PA/541PA 11
For Patients Using Implantable Pacemaker
WARNING
Interaction Between Minute Ventilation Rate-Adaptive Pacemakers and Cardiac
Monitoring and Diagnostic Equipment*
The bioelectric impedance measurement sensor of a minute ventilation rate-adaptive
implantable pacemaker may be affected by transmitter which is connected to the
same patient. If this occurs, the pacemaker may pace at its maximum rate and the
transmitter may give incorrect data to the monitor. If this occurs, disconnect the
electrode leads from the patient or change the setting on the pacemaker by referring
to the pacemaker’s manual. For more details, contact your pacemaker representative
or Nihon Kohden representative.
* Minute ventilation is sensed in rate-adaptive pacemakers by a technology known as bioelectric
impedance measurement (BIM). Many medical devices in addition to pacemakers use this
technology. When one of these devices is used on a patient with an active, minute ventilation rateadaptive pacemaker, the pacemaker may erroneously interpret the mixture of BIM signals created
in the patient, resulting in an elevated pacing rate.
For more information, see the FDA web site.
http://www.fda.gov/cdrh/safety.html
ECG Monitoring
CAUTION
Only use Nihon Kohden specified
electrodes and electrode leads. When
other electrodes or electrode leads
are used, the “CHECK
ELECTRODES” message may appear
and monitoring may stop.
12 Operator’s Manual ZM-540PA/541PA
When the “ELECTRODE OFF” or
“CHECK ELECTRODES” message is
displayed on the receiving monitor,
ECG is not monitored properly and
the ECG alarm does not function.
Check the electrode, electrode leads,
and if necessary, replace with new
ones.
CAUTION
SpO2 Monitoring
WARNING
SpO2 measurement may be incorrect
in the following cases.
• When the patient’s
carboxyhemoglobin or
methemoglobin increases
abnormally.
• When dye is injected in the blood.
• When using an electrosurgical unit.
• During CPR.
• When measuring at a site with
venous pulse.
• When there is body movement.
• When the pulse wave is small
(insufficient peripheral circulation).
WARNING
• When using the TL-201T finger
probe, do not fasten the probe and
cable to the finger by wrapping
with tape. This may cause burn,
congestion or pressure necrosis
from poor blood circulation.
• When using probes other than the
TL-201T finger probe, to avoid poor
circulation, do not wrap the tape too
tight. Check the blood circulation
condition by observing the skin
color and congestion at the skin
peripheral to the probe attachment
site. Even for short-term monitoring,
there may be burn or pressure
necrosis from poor blood circulation,
especially on neonates or low
birth weight infants whose skin is
delicate. Accurate measurement
cannot be performed on a site with
poor peripheral circulation.
WARNING
When not monitoring SpO2,
disconnect the SpO
transmitter. Otherwise, noise from the
probe sensor may interfere and
incorrect data is displayed on the
screen.
cable from the
2
WARNING
Check the circulation condition by
observing the skin color at the
measurement site and pulse
waveform. Change the measurement
site every 8 hours for disposable
probes and every 4 hours for reusable
probes (every 8 hours for TL-630T/TL631T series probe). The skin
temperature may increase at the
attached site by 2 or 3°C (4 or 5°F)
and cause a burn or pressure
necrosis. When using the probe on the
following patients, take extreme care
and change the measurement site
more frequently according to
symptoms and degree.
• Patient with a fever
• Patient with peripheral circulation
insufficiency
• Neonate or low birth weight infant
with delicate skin
Operator’s Manual ZM-540PA/541PA 13
WARNING
When monitoring SpO2 of a patient
who is receiving photodynamic
therapy, the light from the finger probe
sensor may cause a burn.
Photodynamic therapy uses a
photosensitizing agent that has a side
effect of photosensitivity.
CAUTION
Refer to the probe instruction manual
for details.
CAUTION
NIBP and SpO2 can be measured on
the same limb, but the SpO
monitoring might not be accurate
during NIBP measurement. Be careful
when reading the SpO
* Monitoring SpO2 during NIBP Measurement
When the SpO
probe is attached to the
2
same limb as the NIBP cuff, the blood
flow decreases during NIBP measurement
and pulse wave cannot be detected and
SpO
cannot be monitored properly. When
2
“INHIBIT SpO
DURING NIBP” on the
2
PARAMETER SETUP screen is set to ON
(factory default setting), SpO
paused during NIBP measurement to avoid
SpO
alarm occurrence. However, when
2
monitoring SpO
on the same limb as the
2
NIBP, be careful when reading SpO
2
values.*
2
monitoring is
2
values.
2
CAUTION
The disposable probe is not sterilized.
Use the disposable probe only for a
single patient. Never reuse the
disposable probe for another patient
because it causes cross infection.
CAUTION
While a patient is on medication which
causes vasodilation, the pulse
waveform may change and in rare
cases the SpO
value might not be
2
displayed.
CAUTION
Normal external light does not affect
monitoring but strong light such as a
surgical light or sunlight may affect
monitoring. If affected, cover the
measuring site with a blanket.
CAUTION
When a message indicates a faulty
probe, stop monitoring and replace
the probe with a new one.
14 Operator’s Manual ZM-540PA/541PA
CAUTION
Do not use a damaged or
disassembled probe. It causes
incorrect measurement and may
injure the patient.
CAUTION
Do not use a probe which is
deteriorated by aging. Accurate
measurement cannot be performed.
CAUTION
If the attachment site is dirty with
blood or bodily fluids, clean the
attachment site before attaching the
probe. If there is nail polish on the
attachment site, remove the polish.
Otherwise, the amount of transmitted
light decreases, and measured value
may be incorrect or measurement
cannot be performed.
CAUTION
Do not pull or bend the probe cable,
and do not put caster feet on the
probe cable. Do not immerse the
disposable probe cable in chemical
solutions or water. Failure to follow
these instructions may cause cable
discontinuity, short circuit, skin burn
on the patient and incorrect
measurement data. Replace any
broken probe with a new one.
CAUTION
If the skin gets irritated or redness
appears on the skin from the probe,
change the attachment site or stop
using the probe. Take extreme care for
the patients with delicate skin.
CAUTION
When the probe is attached on an
appropriate site with sufficient
circulation and the error message
confirming the probe attachment
repeatedly appears, the probe may be
deteriorated. Replace it with a new
one.
Operator’s Manual ZM-540PA/541PA 15
NIBP Monitoring
WARNING
Be careful when measuring NIBP on a
patient with known bleeding disorders
or coagulation. After NIBP
measurement, there may be dot
hemorrhage, or circulatory disorder by
thrombus where the cuff is attached.
WARNING
When performing NIBP
measurements in STAT mode or 5
minute intervals, periodically remove
the cuff from the patient for ventilation.
The skin temperature may increase at
the cuff attachment site by 2 or 3°C (4
or 5°F). When measuring a patient
with a fever or peripheral circulation
insufficiency, it may cause a burn.
CAUTION
Do not wrap the cuff on an arm or
thigh which is used for injection. NIBP
measurement on an arm or thigh
which is used for injection may cause
reflux of blood and stop injection.
WARNING
NIBP measurement may be incorrect
in the following cases.
• When using an electrosurgical unit
• When there is body movement
• When the pulse wave is small
(insufficient peripheral circulation)
• Too many arrhythmias
• When there is vibration
• When there is a rapid blood
pressure change
• During CPR
CAUTION
Do not wrap the cuff too tight. It may
cause poor blood circulation and
congestion. If the cuff is wrapped too
loosely, the NIBP value may increase.
CAUTION
Do not attach the cuff to the site
where there is injury or inflammation.
If the skin gets irritated or redness
appears on the skin from the cuff,
change the attachment site or stop
using the cuff. Take extreme care on
the patients with delicate skin.
16 Operator’s Manual ZM-540PA/541PA
When using an extension hose, check
that the extension hose is not bent or
squeezed. Otherwise, the cuff might
not inflate or deflate. If the cuff cannot
deflate, it may cause congestion on
the patient at the cuff attachment site.
CAUTION
CAUTION
When performing NIBP measurement
repeatedly, have a rest between
measurements to recover adequate
circulation.
Maintenance
CAUTION
This transmitter is not waterproof. If
detergent or liquid spills into the
transmitter, stop cleaning or
disinfecting it and contact your Nihon
Kohden representative. The
transmitter needs to be checked for
safety and function before use.
CAUTION
Never disassemble or repair the
transmitter. Disassembly and repair
must be performed by qualified
service personnel.
Operator’s Manual ZM-540PA/541PA 17
Preparation on Transmitter
Batteries
WARNING and CAUTION for Handling Batteries
CAUTION
Refer to the battery and battery
charger manuals for details on
handling the batteries.
CAUTION
Do not handle the batteries with wet
hands.
When the transmitter is not in use,
remove the batteries. When the
batteries are installed, battery power
is consumed even when
measurement is not performed. When
NiMH batteries are left in the
transmitter when it is not used, the
batteries may overdischarge and leak
liquid which makes the batteries
unusable and damages the
transmitter.
CAUTION
NOTE
Remove the batteries from the transmitter before disposing of it.
Battery Lifetime
Use three AA type alkaline dry cell batteries. NiMH batteries can also be used but the lifetime of
alkaline batteries is longer. The lifetime of NiMH batteries is about 1/2 of alkaline batteries (when
fully charged).
With new Nihon Kohden recommended alkaline batteries, the transmitter can continuously measure
ECG, respiration, SpO
temperature, NIBP is measured in auto mode at 60 minute intervals and SpO
index finger of a male patient with weight 60 kg. Operation time depends on the thickness of the
SpO
probe attachment site.
2
and NIBP for approximately 1 day. The measurement is performed at room
2
is measured on an
2
Recommended batteries
NiMH secondary: SANYO HR-3UF (W)
Battery charger: SANYO NC-M55
Alkaline primary: Nihon Kohden Medipower (equivalent to Panasonic LR6 (G))
18 Operator’s Manual ZM-540PA/541PA
Installing (Replacing) Batteries
CAUTION
Battery replacement must be performed by the operator. When replacing the batteries
of a transmitter that is currently used for a patient, disconnect the electrode leads
from the transmitter before replacing batteries or do not touch the patient during
replacement.
If electrode leads are attached to the patient and the person replacing batteries touches the patient
during battery replacement, excess patient leakage current may flow into the patient.
NOTE
• Replace all batteries at the same time.
• Do not use different types of batteries together.
• Insert the batteries with the correct polarity (+ and –).
• The capacity of rechargeable NiMH batteries is reduced if the batteries are recharged
before they are fully discharged. For details, refer to the battery operator’s manual.
1. Remove the battery case cover.
2. Insert three new or fully charged batteries into
the battery case observing the correct polarity.
Operator’s Manual ZM-540PA/541PA 19
3. Close the cover.
The transmitter is automatically turned on when the
batteries are installed and the cover is closed.
Situations Requiring Battery Replacement
Replace the batteries when any of the following occurs:
• The transmitter displays the “BATTERY WEAK” message or “
• The transmitter generates a constant alarm (continuous “peep” sound).
• The transmitter LCD does not display anything when the power is turned on.
• The receiving monitor displays a battery replacement message.
Battery Level Indication
The following icons on the LCD indicate the battery level. When “PROTOCOL” on the SYSTEM
SETUP screen is set to 57, the battery level indication is transmitted to the receiving monitor.
IndicationBattery LevelMessage on the Receiving Monitor
Fully charged batteries
” icon.
No indication
Batteries are 2/3 full.
Batteries are low. NIBP cannot
be measured. Replace batteries.
Batteries are very weak.
Dead batteries
There is no message on the monitor.
Message requiring battery replacement is
displayed.
No signal can be transmitted to the monitor.
There is no indication on the monitor.
Turning On the Transmitter
When the batteries are installed correctly, the power is turned on. A one second “peep” sounds and
the startup screen appears, then the check electrodes screen appears. (There is no “peep” sound
when there is no battery power.)
20 Operator’s Manual ZM-540PA/541PA
Loading...
+ 124 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.