Nellcor Portable Bedside User manual

Portable Bedside Capnograph/Pulse Oximeter
Service Manual
CI05225E
Notice: Purchase of this instrument confers no express or implied license under any Oridion Medical patent or Nellcor Puritan Bennett patent to use the instrument with any accessory that is not manufactured or licensed by Oridion Medical 1987 Ltd.
Capnography is covered by one or more of the following United States patents: 6,491,643; 6,656,127; 6,168,958; 5,300,859 and 4,755,675 and their foreign equivalents
この製品は、次の日本における特許、1980816と海外での同等のものによって保護されて 居ります。
Microstream Smart BiteBloc™ and NIV Line™ are trademarks of Oridion Medical 1987 Ltd.
Nellcor Puritan Bennett Incorporated is an affiliate of Tyco Healthcare.
The following are trademarks of Nellcor Puritan Bennett Incorporated: Oxiband OxiCliq
Pulse Oximetry is covered by one or more of the following US patents and foreign equivalents: 4,802,486; 4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,485,847; 5,743,263; 5,865,736; 6,035,223; 6,298,252; 6,463,310; 6,591,123; 6,675,031; 6,708,049; 6,801,797; Re 35,122.
®
and FilterLine® are registered trademarks and Smart CapnoLine™, CapnoLine™,
®
; Dura-Y®; and OXIMAX®.
.
®
; Durasensor®;
TABLE OF CONTENTS
List of Figures List of Tables
Section 1: Introduction........................................................................ 1-1
Section 2: Required Equipment.......................................................... 2-1
Section 3: Cleaning ........................................................................... 3-1
Section 4: Service Mode...................................................................... 4-1
Section 5: Troubleshooting................................................................. 5-1
Section 6: Flow Calibration Check ..................................................... 6-1
Section 7: Disassembly Guide............................................................ 7-1
1.1 Manual Overview ............................................................... 1-1
1.2 Warnings, Cautions, and Notes ......................................... 1-1
1.2.1 Warning.............................................................. 1-1
1.2.2 Caution...............................................................1-1
1.2.3 Note.................................................................... 1-1
1.3 Product Description ...........................................................1-2
2.1 Required Equipment .......................................................... 2-1
3.1 Cleaning ........................................................................... 3-1
3.1.1 Procedure........................................................... 3-1
4.1 Introduction ........................................................................ 4-1
4.2 Service Level 1 .................................................................. 4-1
4.2.1 Information Screen............................................. 4-1
4.2.2 Event Messages Screen .................................... 4-1
4.2.3 Interface Test Screen......................................... 4-2
4.2.4 Accessing Service Mode and Service Level 1 ... 4-3
4.3 Service Level 2 .................................................................. 4-5
4.3.1 Default Alarm Limits ........................................... 4-7
4.3.2 Default Settings.................................................. 4-9
4.3.3 Flow Calibration ............................................... 4-10
5.1 Introduction ........................................................................ 5-1
5.2 Who Should Perform Repairs ............................................ 5-1
5.3 Repair Level Supported ..................................................... 5-1
5.4 How to Use This Section ...................................................5-1
5.5 Obtaining Replacement Parts............................................ 5-1
5.6 Troubleshooting Guide ......................................................5-2
6.1 Introduction ........................................................................ 6-1
6.2 Flow Rate Check ...............................................................6-1
6.3 Flow Calibration Process ................................................... 6-2
7.1 Introduction ........................................................................ 7-1
7.2. Opening the Monitor Case.................................................7-2
7.3 Replacing the CO
7.4 Replacing the SpO
2 Board .................................................. 7-4
2 Board................................................. 7-5
7.5 Replacing the Pump ..........................................................7-6
7.6 Replacing the Flow System ...............................................7-7
7.6.1 Replacing the lines to the pump.........................7-8
7.6.2 Replacing the lines to the solenoid ....................7-9
7.7 Replacing LCD and LED Display..................................... 7-11
7.7.1 Replacing the LCD ........................................... 7-11
7.7.2 Replacing the LED Display .............................. 7-11
iii
Table of Contents
7.8 Replacing Housing Components ..................................... 7-12
7.8.1 ON-OFF Button and Gas Outlet....................... 7-12
7.8.2 Keypads ...........................................................7-12
7.8.3 Front Cover ......................................................7-13
7.8.4 SpO
Connector Latch, Rear Cover and
2
SpO
Insulation Plate .......................................7-13
2
7.9 Updating Software Version .............................................. 7-14
Section 8: Electrical Safety Tests.......................................................8-1
8.1 Electrical Safety Tests ....................................................... 8-1
Section 9: Periodic Maintenance ........................................................ 9-1
9.1 Periodic Maintenance ........................................................ 9-1
Section 10: Performance Verification............................................... 10-1
10.1 Flow System Leak Check ................................................ 10-1
10.1.1 Flow System Leak Check Procedure ............... 10-1
10.2 Performance Verification Procedure................................ 10-2
Section 11: Packing For Shipment ................................................... 11-1
11.1 General Instructions......................................................... 11-1
11.2 Packing Monitor in Original Carton .................................. 11-1
11.3 Packing in a Different Carton........................................... 11-1
Section 12: Specifications................................................................. 12-1
12.1 Physical ......................................................................... 12-1
12.1.1 Size .................................................................. 12-1
12.1.2 Weight.............................................................. 12-1
12.1.3 Noise Emission ................................................ 12-1
12.2 Environmental..................................................................12-1
12.2.1 Temperature.....................................................12-1
12.2.2 Relative Humidity ............................................. 12-1
12.2.3 Pressure and Altitude....................................... 12-1
12.3 Safety Standards ............................................................. 12-1
12.4 Performance .................................................................... 12-2
12.4.1 Capnograph ..................................................... 12-2
12.4.2 Pulse Oximeter................................................. 12-3
12.5 Power Specifications ....................................................... 12-4
12.5.1 External Power Source .................................... 12-4
12.5.2 Internal Power Source...................................... 12-4
12.6 Components and User interface ...................................... 12-4
12.6.1 Displays............................................................ 12-4
12.6.2 Controls and Indicators .................................... 12-4
12.6.3 Connections ..................................................... 12-4
Section 13: Spare Parts..................................................................... 13-1
13.1 Spare Parts List ............................................................... 13-1
Glossary of Terms
iv
Table of Contents
LIST OF FIGURES
LIST OF TABLES
1-1 Handheld Capnograph/Pulse Oximeter............................. 1-2
1-2 LCD Graphic Display and 7 Segment Digital Displays ......1-2
7-1 Removing the Battery Pack ............................................... 7-2
7-2 Removing Mounting Screws.............................................. 7-3
7-3 Separating Rear and Front Covers.................................... 7-3
7-4 Replacing CO 7-5 Replacing SpO
2 Board and Housing Components.............. 7-4
2 Board....................................................... 7-5
7-6 Replacing the Pump .......................................................... 7-6
7-7 Flow System ......................................................................7-7
7-8 Main Line ........................................................................... 7-8
7-9 Exhaust Line...................................................................... 7-8
7-10 Lines to Solenoid (Zero line, Input line 1, Input line 2........ 7-9
7-11 Zero line........................................................................... 7-10
7-12 Replacing LCD and LED Display.....................................7-11
7-13 Replacing SpO
2 Connector Latch.................................... 7-13
7-14 Replacing EPROM........................................................... 7-14
10-1 Leak Test Jig Connection ................................................10-1
10-2 Initialization Screen.......................................................... 10-2
10-3 Self-Test Screen .............................................................. 10-2
10-4 Measuring Mode .............................................................. 10-2
10-5 Quick Guide ..................................................................... 10-3
10-6 Connecting the monitor to Printer/PC with
Communication Adapter Kit ............................................. 10-9
11-1 Packing the Monitor in Original Packing .......................... 11-2
4-1 Accessing Service Mode and Changing Parameter
Settings................................................................... 4-3
4-2 Accessing Service Level 2................................................. 4-5
4-3 Changing Default Alarm Limits .......................................... 4-7
4-4 Changing Default Settings................................................. 4-9
4-5 Accessing Flow Calibration.............................................. 4-10
5-1 Troubleshooting Guide ...................................................... 5-2
5-2 Advisory Messages Guide.................................................5-4
5-3 Event Messages ................................................................ 5-5
6-1 Flow Calibration................................................................. 6-2
10-1 CO
2 Calibration Check..................................................... 10-4
10-2 Calibration Process.......................................................... 10-6
10-3 Calibration Process -Troubleshooting.............................. 10-7
10-4 Plethysmograph Waveform ............................................. 10-8
10-5 Accessing Monitor’s Print Functions.............................. 10-10
13-1 Spare Parts List ............................................................... 13-1
v
Table of Contents
SECTION 1: INTRODUCTION
1.1 Manual Overview
1.2 Warnings, Cautions, and Notes
1.3 Product Description
1.1 MANUAL OVERVIEW
This manual contains information for servicing the handheld capnograph/ pulse oximeter (the monitor). Only authorized service personnel should service this product. The Handheld Capnograph/Pulse Oximeter Operator’s Manual (the operator’s manual) is an integral part of the service procedures. Before servicing the monitor, read the operator’s manual carefully for a thorough understanding of how to operate the unit.
Only use the tools and test equipment as specified in this manual. Only use original spare parts available from your local distributor.
Warning: Incorrect procedures may harm the patient, or damage the monitor.
1.2 WARNINGS, CAUTIONS, AND NOTES
1.2.1 Warning
1.2.2 Caution
1.2.3 Note
A warning precedes an action that may result in injury or death to the patient or user. Warnings are boxed and highlighted in boldface type.
A caution precedes an action that may result in damage to, or malfunction of, the monitor. Cautions are highlighted in boldface type.
A note gives information that requires special attention.
1-1
Section 1: Introduction
1.3 PRODUCT DESCRIPTION
The handheld capnograph/pulse oximeter is used to continuously monitor end-tidal carbon dioxide (EtCO
2), and pulse rate. The monitor is for attended monitoring only and
(SpO must be used in the continuous presence of a qualified healthcare provider. The monitor can be used on adult, pediatric, and infant/neonatal patients. It is intended for use where continuous, noninvasive monitoring of these parameters is desired, including hospital and mobile use (when protected from excessive moisture such as direct rainfall). The monitor operates on AC power or a rechargeable Nickel Metal Hydride battery pack.
2), respiratory rate, oxygen saturation
1-2
Figure 1-1: Handheld Capnograph/Pulse Oximeter
The monitor is operated using a four-key keypad. Measurements are displayed on the front panel on the 7-segment LED digital displays (LED) and on the LCD graphic display (LCD) as shown in Figure 1-2. Refer to the operator’s manual for complete operating instructions.
Figure 1-2: LCD Graphic Display and 7-Segment Digital Displays
SECTION 2: REQUIRED EQUIPMENT
2.1 Required Equipment
2.1 REQUIRED EQUIPMENT
You will need the following equipment to disassemble, replace parts, check, adjust, or calibrate the monitor.
For disassembly and replacement:
• Phillips-head screwdriver (medium)
• Flat-head screwdriver (small)
• IC Extractor for PLCC socket
• Threadlocker glue e.g. Loctite
For checks, adjustments and calibration:
®
222 super screw lock
• Calibration Gas (5% CO
2 in air)
• FilterLine or FilterLine H and a connecting means
• DS-100A - O
XIMAX Durasensor
®
, Nellcor
• Leak Test Jig
• Vacuum manometer with a range of 0-500 mBar e.g. MPL-3200-DM, Digital Manometer MicroPneumatic Logic, Inc.
• Communication Adapter Kit (only necessary to verify print functions)
• Seiko DPU-414 printer (only necessary to verify print functions)
• External power supply
• Handheld Capnograph/Pulse Oximeter Operator’s Manual
• Flow Meter (range 0-50 ml/min) e.g. Top Track Mass Flow Meter, model: 822-13-OV1-PV1-V1, Sierra Instruments, Inc.
Warning: Observe ESD (electrostatic discharge) precautions when handling, adjusting or performing any procedure with the monitor’s internal components.
2-1
SECTION 3: CLEANING
3.1 Cleaning
3.1 CLEANING
Always clean the outer surface of the monitor before servicing.
Warning: Always wear latex or surgical gloves when cleaning or servicing the monitor after hospital use.
Warning: Do not allow liquids to drip inside the housing.
Warning: Do not clean internal components.
Warning: Do not use preparations based on: phenols, halogen producing compounds, strong organic acids, or oxygen producing compounds.
3.1.1 Procedure
1. Unplug the unit from AC power.
2. With a damp cloth, gently wipe the outer case of the monitor with a
disinfectant based on aldehyde, alcohol or quaternary ammonium producing compounds.
3-1
SECTION 4: SERVICE MODE
4.1 Introduction
4.2 Service Level 1
4.3 Service Level 2
4.1 INTRODUCTION
The monitor supports a service mode built into the instrument’s firmware. To access the service mode, refer to Table 4-1: "Accessing Service Mode and Changing Parameter Settings".
The architecture of the monitor’s service mode divides the service mode into two levels as described in the sections below.
• Service Level 1
• Service Level 2
4.2 SERVICE LEVEL 1
The Service Level 1 provides information and tools on three service screens to test and calibrate specific functions without disassembling the monitor.
• Information Screen
• Event Messages Screen
• Interface Test Screen
From these display screens the service technician can check the operating hours counter and event messages, test the Human Interface functions (HMI), set the light detection threshold, adjust the LCD contrast middle point and do a calibration check and a calibration process, if necessary. Refer to Table 4-1: "Accessing Service Mode and Changing Parameter Settings".
4.2.1 Information Screen
The information screen displays the monitor’s main board serial number, the number of operating hours, and the SpO versions.
4.2.2 Event Messages Screen
The monitor automatically performs a series of built-in tests (BITs) at start-up and during operation to check sub-systems and modules for errors. The event messages screen lists the last five event messages, the time each occurred relative to the unit’s operating hours, and a recommended service action. There are three event messages and each has a specific response as follows:
2 and CO2 boards’ software
Event Message: Action:
2 board Replace CO2 board
CO
2
board Replace SpO2 board
SpO Flow Replace CO
2 board
4-1
Section 4: Service Mode
Note: To exit the service mode and save event messages, turn off the
4.2.3 Interface Test Screen
The interface test screen shows four parameters (listed below) that provide tools for the user to test, adjust, and calibrate the unit’s interfacing systems.
•HMI Test
• Light Detector
• Contrast Cal. (Calibration)
monitor. Event messages will be erased when exiting the service mode by a long press of .
•CO
2 Calib. (Calibration)
4.2.3.1 HMI Test
When the HMI Test is turned on, the monitor performs a self-test of the following sub-systems:
• Buzzer (sounds for one second)
• LCD graphic display (screen shows a checkerboard pattern)
• 7-segment digital display (8s scroll left to right)
• Alarm bar (lights red, yellow, off)
• Backlight (turns on and off)
• Alarm silence indicator (light turns on and off)
When the HMI test is completed the status field returns automatically to
OFF.
4.2.3.2 Light Detector
The Light Detector parameter allows the service technician to set the threshold value of the phototransistor. The default value is 1 and the range is from 1 to 5 with 1 as the most sensitive (backlight on at all times) and 5 as the least sensitive (backlight off). The threshold value should be set according to the lighting conditions of the working environment. A higher threshold value saves on battery power consumption.
4-2
4.2.3.3 Contrast Cal. (calibration)
The Contrast Cal. gives the service technician the option to change the middle point default value of the contrast intensity of the LCD. Refer to Table 4-1: "Accessing Service Mode and Changing Parameter Settings".
4.2.3.4 CO
The CO
2 Calib. (calibration)
2 Calib. allows the service technician to perform a calibration
check and calibration process to ensure the accuracy of the monitor. Section 10: Performance Verification describes the calibration check procedure.
4.2.4 Accessing Service Mode and Service Level 1.
Table 4-1: "Accessing Service Mode and Changing Parameter Settings" describes the process to access the service mode, change parameter settings and how to change Contrast Cal. settings.
Table 4-1: Accessing Service Mode and
Changing Parameter Settings
Objective Action Response
Section 4: Service Mode
To access Service Mode During self-test,
long press simultaneously
and
To access Event Messages screen
To access Interface Test screen

long press

long press

Self-test occurs and the Information Screen appears:
To change the parameter setting or value
To move to the next parameter

short press

short press

4-3
Section 4: Service Mode
Table 4-1: Accessing Service Mode and
Changing Parameter Settings (Continued)
Objective Action Response
To change Contrast

short press

Cal. setting to set

To start Contrast Cal. short press “Contrast Cal.” flashes.

To adjust Contrast Cal. middle point

press

Screen lightens or darkens

until desired setting is reached. (Pressing on the left lightens the screen, and pressing on the right darkens the screen.)

To accept new setting short press

To reset and return to measuring mode (automatically erasing the event messages memory)
To exit service mode from any service screen without erasing the event messages memory

long press

Note: Reset time is approximately one minute.
Turn off the
monitor.
4-4
4.3 SERVICE LEVEL 2
Section 4: Service Mode
The Service Level 2 gives the service technician tools on three service screens:
• Default Alarm Limits
• Default Settings
• Flow Calibration
To access this service level, the service technician must enter a valid password (1627), refer to Table 4-2: "Accessing Service Level 2".
Note: The password should not be disclosed to avoid unauthorized setting
of parameters.
Table 4-2: Accessing Service Level 2
Objective Action Response
To access Service Level 2
Enter Password: 1627
Enter the 1st digit short press (right)

Move to the next digit short press

After entering the service mode

long press (x3)

(x1)

Pressing on the right arrow scrolls the numbers up, the left arrow scrolls the numbers down.

Password Screen appears
Enter the 2nd digit short press (right)
x6
Repeat the same procedure as above for the next two digits (2, 7).
4-5
Section 4: Service Mode
Table 4-2: Accessing Service Level 2 (Continued)
Objective Action Response
To accept the last digit and access the Default Alarm Limits menu
In case you entered a wrong digit:
To restart password validation process
To return to the Interface Test Screen

short press

long press

long press

To exit service mode and save event messages
Turn off the
monitor.
4-6
4.3.1 Default Alarm Limits
The service technician can change the default alarm limits for either patient mode (Adult or Neonatal) as described in Table 4-3: "Changing Default Alarm Limits".
The monitor has the following alarms with adjustable level settings:
Section 4: Service Mode
•EtCO
•FiCO
•SpO
2 high and low levels (mmHg)
2 high level (mmHg)
2 high and low levels (%)
• Pulse rate high and low levels (bpm)
Warning: Changing default settings will permanently change the user’s start-up value and should be verified with the responsible clinical personnel.
Table 4-3: Changing Default Alarm Limits
Objective Action Response
To access the Default

short press Alarm Limits menu after entering last password digit

To change the patient
short press mode
To access a new
short press parameter

To change the value short press

4-7
Section 4: Service Mode
Table 4-3: Changing Default Alarm Limits (Continued)
Objective
To reset to measuring mode (automatically erasing event messages)
To exit service mode and save event messages

Action

long press

then long press

Turn off the
monitor.
Response
4-8
4.3.2 Default Settings
Section 4: Service Mode
The service technician can change two default settings (refer to Table 4-4: "Changing Default Settings"):
•3 Min Alert
When the 3 Min Alert is ON, the unit will beep every three minutes to remind the user that an alarm (or all alarms) are set to OFF.
• BTPS (body temperature, pressure, saturation)
When BTPS is ON, the CO
2 value is corrected for this factor
Pulse Tone Pulse Tone default is ON. When changed to OFF, the Pulse Tone default is silent (as indicated on screen).
.
Table 4-4: Changing Default Settings
Objective Action Response
To access the Default Settings menu (after accessing the Default Alarm Limits menu)

To change the setting short press

To access BTPS parameter (and save the changed setting for 3 Min Alert)
To reset and return to measuring mode (automatically erasing event messages memory)

long press

short press

long press

To exit service mode and save event message
Turn off the
monitor.
4-9
Section 4: Service Mode
4.3.3 Flow Calibration
The Flow Calibration screen allows the service technician to perform flow calibration after flow adjustment and/or pump or flow system replacement, refer to Table 4-5: "Accessing Flow Calibration"
Note: When changing the pump voltage setting, refer to
Flow Calibration Check
Table 4-5: Accessing Flow Calibration
Objective Action Response
To access the Flow Calibration screen
To change the parameter setting or value
To move to the next parameter (and save the changed setting)
At any screen after the password long press simultaneously
for change guidelines.
and
short press
short press
Section 6:
4-10
SECTION 5: TROUBLESHOOTING
5.1 Introduction
5.2 Who Should Perform Repairs
5.3 Repair Level Supported
5.4 How to Use This Section
5.5 Obtaining Replacement Parts
5.6 Troubleshooting Guide
5.1 INTRODUCTION
This section provides information for troubleshooting the monitor and isolating a failure on the unit.
5.2 WHO SHOULD PERFORM REPAIRS
Only qualified service personnel should remove and replace components of the monitor. If your facility does not have qualified service personnel, contact your local distributor.
5.3 REPAIR LEVEL SUPPORTED
The monitor has the following replaceable components: CO2 board, SpO2 board, Flow System, Pump Assembly, Housing Components, LCD graphic display (LCD), 7-segment LED digital displays (LED display) and Software Assembly.
The procedures for disassembling and replacing the mentioned parts are described in Section 7: Disassembly Guide.
5.4 HOW TO USE THIS SECTION
Failures of the Housing Components; Front Cover, Rear Cover, Gas Outlet, External SpO are determined by visually inspecting these components for cracks or deformations, and checking for mechanical failures.
Refer to Table 5-1: “Troubleshooting Guide”to isolate failures of the boards, Flow System, Pump Assembly, LCD and LED displays. Once a failure has been isolated, refer to Section 7: Disassembly Guide for instructions for removing and replacing a component of the monitor.
5.5 OBTAINING REPLACEMENT PARTS
Your local distributor provides technical assistance information and replacement parts. Refer to parts by the part name listed in Section 13: Spare Parts.
2 Connector Latch, ON/OFF button and Keypads,
5-1
Section 5: Troubleshooting
5.6 TROUBLESHOOTING GUIDE
If you encounter a problem that cannot be resolved through a visual inspection, refer to Table 5-1: “Troubleshooting Guide” which provides a list of symptoms, probable causes, and recommended actions to take to correct the problem. It is recommended that corrective actions be performed in the order presented. For a symptom that is not listed in Table 5-1: “Troubleshooting Guide”, contact your local distributor.
If a Caution or Advisory message is displayed on the LCD, refer to Table 5-1: “Troubleshooting Guide”for the indicated failure and corrective action. Caution messages appear on the display screen followed by !! and Advisory messages are followed by !.
Check in the Service Mode for any Error Code and refer to Table 5-3: “Event Messages”for the indicated failure and corrective action.
Symptom Probable Cause Corrective Action
Table 5-1: Troubleshooting Guide
The unit does not turn on when the ON/OFF button is switched
ON
when the unit is operated with batteries.
The unit does not turn on when the ON/OFF button is switched
ON
when the monitor is connected to mains power.
The battery pack is missing or is installed incorrectly.
The battery pack’s charge is low.
The battery pack’s contacts are defective.
The contacts for the battery pack on the CO board are defective or a CO
2 board component
has failed.
The ON/OFF button has a mechanical defect.
Check if there is a battery pack in the monitor, if not, place one in the monitor. Be sure the battery pack is installed correctly.
Install a fully charged battery pack; charge the empty battery pack.
Install a new battery pack and check the monitor functions.
Connect the monitor to mains line power
2
and if the monitor functions, replace the CO
2 board.
Inspect the button for cracks, deformations or other damage. Replace with a new button.
5-2
A CO
2 board component
has failed.
Replace with a new CO
2 board.
Section 5: Troubleshooting
Table 5-1: Troubleshooting Guide (Continued)
Symptom Probable Cause Corrective Action
The unit does not turn on when the ON/OFF button is switched
ON
when the monitor is connected to main lines power.
One or more buttons on the upper or lower keypad do not work.
The wall socket is not receiving power or is defective.
Check the wall socket for power supply or mechanical defects. If necessary, use a different wall socket.
The AC adapter is defective.
A CO
2 board component
is defective.
Replace with new AC adapter.
Replace with a new CO
2 board.
The keypad is defective. Inspect the keypad for
cracks, deformations or other damage. Replace with a new keypad.
A CO
2 board component
has failed.
Replace with a new CO
2 board.
No information is displayed on the LCD.
LCD Backlight does not come on when the unit is placed in a dark environment.
The contrast default value is set too low or high.
An LCD component has failed.
A CO
2 board component
has failed.
The Power Management option is set at
LOW.
An LCD component has failed.
A CO
2 board component
has failed.
Check the contrast value in the Service Mode using the
Contrast Cal.
option and correct accordingly.
Replace with a new LCD.
Replace with a new CO
2 board.
Refer to the monitor’s operator’s manual to change the Power Management option to
NORMAL or
HIGH.
Replace with a new LCD.
Replace with a new CO
2 board.
5-3
Section 5: Troubleshooting
Table 5-1: Troubleshooting Guide (Continued)
Symptom Probable Cause Corrective Action
LED display
does not light when the unit is functioning.
Beeper does not beep.
The Power Management option is set at
LOW.
An
LED display has
failed.
A CO
2 board component
has failed.
The Pulse Tone option on the Alarm Silence menu has been disabled.
One or more of the Alarm Silence options has been turned
A CO
2 component has
OFF.
failed.
Refer to the monitor’s operator’s manual to change the Power Management option to
NORMAL or
HIGH.
Replace with new LED Display.
Replace with a new CO
2 board.
Turn the Pulse Tone ON. Refer to the monitor’s operator’s manual.
Turn the respective alarm
ON. Refer to
the monitor’s operator’s manual.
Replace with a new CO
2 board.
Table 5-2: Advisory Messages Guide
Message Probable Cause Corrective Action
Check Unit
!
A CO2 board component has failed.
Check Event Messages screen in Service Mode.
Check CO
Check SpO2 !
2 !
An SpO2 board component has failed.
Replace with a new SpO
2 board.
5-4
Section 5: Troubleshooting
Table 5-3: Event Messages
Event Message Probable Cause Corrective Action
CO
2 board A CO2 board component
failed.
2 board Replace with a new
SpO
Replace with a new CO
2 board.
SpO
2 board.
Flow - Loose connections
- Kinks or twists in tubing
- Flow system blocked
- Pump Malfunction - Replace pump
- A CO
2 board
component failed
- Inspect all tubing ends for loose connections and ensure tight connections
- Straighten kinks and undo any twists in the tubing.
- Check flow rate and if necessary change Flow System.
- Replace with a new CO
2 board
5-5
SECTION 6: FLOW CALIBRATION CHECK
6.1 Introduction
6.2 Flow Rate Check
6.3 Flow Calibration Process
6.1 INTRODUCTION
Perform a Flow Calibration Check after replacing the Flow System and/or Pump. Flow Calibration Check includes Flow Rate Check and Flow Calibration Process.
6.2 FLOW RATE CHECK
To perform the Flow Rate Check, follow the steps below:
1. Connect the FilterLine to the Handheld Capnograph/Pulse Oximeter.
2. Turn on the monitor.
3. Connect the other end of the FilterLine to the Flow meter gas outlet.
4. Access the Flow Calibration screen of the Service Mode, refer to Table 6-1: "Flow Calibration".
5. Check that the Flow Meter reading is 50 ±5 ml/min. (at sea level)
If the Flow Meter reading is 50 ±5 ml/min:
5a Perform Flow Calibration Process, refer to section 6.3 "Flow
Calibration Process".
If the Flow Meter reading is not 50 ±5 ml/min:
5b Change the pump voltage value until the flow rate displayed
in the flow meter is 50 ±5 ml/min, refer to Section 4: Service Mode, Table 4-5: "Accessing Flow Calibration".

5c Perform Flow Rate Check

5d Perform Flow Calibration Process, section 6.3 "Flow
Calibration Process".
If you cannot set the flow rate to 50 ±5 ml/min after adjusting the pump voltage value:
5e Change the Flow System, refer to section 7.6 "Replacing the
Flow System".

5f Perform Flow Rate Check

5g Perform Flow Calibration Process.

If after changing the Flow System you still cannot set the flow rate display in the flow meter to 50 ±5 ml/min:

5h Change the Pump, refer to section 7.5 "Replacing the Pump".

5i Perform Flow Rate Check

5j Perform Flow Calibration Process.

6-1
Section 6: Flow Calibration Check
6.3 FLOW CALIBRATION PROCESS
To perform the Flow Calibration Process, refer to Table 6-1: "Flow Calibration".
Objective Action Response
Table 6-1: Flow Calibration
To access Flow Calibration screen
To change the parameter setting or value
To move to the next parameter (and save the changed setting)
To select Flow Cal. short press
Simultaneously
long press
and

short press

short press

6-2
To start Calibration short press

short press

Wait until either a Pass or Fail message is displayed.
Section 6: Flow Calibration Check
Table 6-1: Flow Calibration (Continued)
Objective Action Response
If Fail is displayed Replace Flow
System and/or Pump
Pump volt.
(refer to Section 5: Troubleshooting)
and repeat above procedure
If Pass is displayed Exit Service Mode
2.0
To reset and return to measuring mode
To exit service mode and save event messages
long press
Turn off the monitor
Pump volt.
2.0
6-3
SECTION 7: DISASSEMBLY GUIDE
7.1 Introduction
7.2 Opening the Monitor Case
7.3 Replacing the CO
7.4 Replacing the SpO
7.5 Replacing the Pump
7.6 Replacing the Flow System
7.7 Replacing the LCD and LED Display
7.8 Replacing the Housing Components
7.9 Updating the Software Version
7.1 INTRODUCTION
The monitor can be disassembled and the following components can be replaced:
2 Board
2 Board
•CO
•SpO
2 Board
2 Board
Pump Assembly
Flow System
Housing components
•EPROM
LCD graphic display (LCD)
7-segment LED digital displays (LED Display)
Caution: Observe ESD (electrostatic discharge) precautions when disassembling and reassembling the monitor and when handling any of the components of the monitor.
Use the following procedures to disassemble the monitor and replace parts as needed. Reassemble the monitor in reverse order. The manufacturer recommends that you follow the disassembly procedure in the order presented.
Note: After replacing the Flow System and/or Pump, perform a flow
calibration check as described in Section 6: Flow Calibration
Check.
Note: After reassembling any part of the monitor, be sure to complete
the performance verification as described in Section 10: Performance Verification.
7-1
Section 7: Disassembly Guide
7.2 OPENING THE MONITOR CASE
1. Be sure the monitor is disconnected from mains power and remove the battery pack from the monitor; press the release button at the bottom of the monitor and pull out the battery pack as shown in Figure 7-1: Removing the Battery Pack.
7-2

Figure 7-1: Removing the Battery Pack

Section 7: Disassembly Guide
2. Place the monitor on a clean, dry, nonabrasive surface with the back of the monitor facing up.
3. Remove the two Bottom Mounting and two Back Mounting Screws from the rear panel of the monitor as shown in Figure 7-2: Removing Mounting Screws.
.

Figure 7-2: Removing Mounting Screws

4. Carefully lift the rear case cover up and remove the 14-pin flat cable from the CO
2 Board as shown in Figure 7-3: Separating
Rear and Front Covers.

Figure 7-3: Separating Rear and Front Covers

7-3
Section 7: Disassembly Guide
7.3 REPLACING THE CO2 BOARD
Refer to Figure 7-4: Replacing CO2 Board and Housing Components
1. Open the monitor case as described in section 7.2 "Opening the Monitor Case".
2. Remove the two CO
2 Board Mounting Screws.
3. Remove the CO2 Connector Screw from the CO2 Input Connector.
4. Remove the ON-OFF Button and save for reassembly.
5. Pull out and remove the Gas Outlet and save for reassembly.
6. Remove the CO
2 Board and replace with a new CO2 Board and
reassemble in reverse order.
Note: When replacing the keypads, be sure the icons are facing the correct
direction. (Refer to Figure 1-1: Handheld Capnograph/Pulse Oximeter.)
7-4

Figure 7-4: Replacing CO2 Board and Housing Components

7.4 REPLACING THE SPO2 BOARD
(Refer to Figure 7-5: Replacing SpO2 Board.)
1. Open the monitor case as described in section 7.2 "Opening the Monitor Case".
Section 7: Disassembly Guide
2. Pull out the Internal SpO
3. Remove the three SpO
2 Connector Lock.
2 Board Mounting Screws and Washers.
4. Remove the SpO2 Board with attached cables and replace with a new SpO Internal SpO
2 Board. Be sure to reattach the cables and use the new
2 Connector Lock. Be sure the SpO2 Insulation
Plate is placed correctly over the three screw mounts.
Caution: When replacing the SpO
2 Mounting Screws, be sure you
use the plastic washers supplied with the original unit.
Caution: When replacing the SpO
2 Board, carefully hand-tighten
the Mounting Screws. Overtightening can strip the screw-holes.

Figure 7-5: Replacing SpO2 Board

7-5
Section 7: Disassembly Guide
7.5 REPLACING THE PUMP
Refer to Figure 7-6: Replacing the Pump.
1 Open the monitor case described in section 7.2 "Opening the
2 Remove the Damping cover from the Pump.
3 Locate the two tubings, Main line and Exhaust line, connected to
4 Locate the purple Main line restrictor and the green Exhaust line
5 Disconnect both tubings from their restrictors.
Monitor Case".
the Pump, refer to Figure 7-7: Flow System.
restrictor, refer to Figure 7-7: Flow System.
Note: Both restrictors must remain with their flow system lines attached to the CO
2 Board.
6 Disconnect the Pump connector from it’s CO
7 Remove the CO
2 Board from the Front Cover, refer to section 7.3
2 Board socket.
"Replacing the CO2 Board".
8 Unscrew and remove the Pump mounting screws.
9 Remove the Lower damping and Damping holder.
10 Remove the Pump and the Upper damping.
11 Place the new Pump.
12 Reassemble the unit, using the new assembly parts, in reverse
order.
Note: Before replacing the pump mounting screws, first dip them into
®
Loctite
glue or any other threadlocker glue. Carefully twist the
screws into the Lower damping, gently pressing against the
2 Board.
CO
7-6

Figure 7-6: Replacing the Pump

7.6 REPLACING THE FLOW SYSTEM
1. Open the monitor case as described in section 7.2 "Opening the
Monitor Case".
The Flow System has 5 lines, refer to Figure 7-7: Flow System.
Two lines go from the pump:
•Main line
•Exhaust line
Three lines go from the solenoid:
Zero line
Input line 1
Input line 2 (not viewed in Figure 7-7)
Section 7: Disassembly Guide

Figure 7-7: Flow System

7-7
Section 7: Disassembly Guide
7.6.1 Replacing the lines to the pump
7.6.1.1 Replacing the Main line
(Refer to Figure 7-8: Main line and Figure 7-9: Exhaust line)

1. Locate the purple restrictor.

2. Carefully disconnect the Main line from the tubing connected to the Pump, leaving the purple restrictor on the Main line.

3. Remove the tubing from the Sensor.

4. Remove the tubing from the Pressure sensor.

5. Reassemble the new Main line in reverse order.

Note: In order to fit the Main line on the board without twists and
kinks, coil
the tubing around the Pressure sensor.

Figure 7-8: Main line

7.6.1.2 Replacing the Exhaust line
(Refer to Figure 7-9: Exhaust line)
1. Locate the green restrictor.
2. Carefully disconnect the Exhaust line, at the green restrictor,
from the tubing connected to the pump. Leave the green restrictor on the Exhaust line.
3. Reassemble the new Exhaust line in reverse order.
7-8

Figure 7-9: Exhaust line

7.6.2 Replacing the lines to the solenoid
Refer to: Figure 7-10: Lines to Solenoid (Zero line, Input line 1, Input line 2)
7.6.2.1 Removing the lines to the solenoid
1. Carefully disconnect the Zero line from the solenoid.
Warning: The scrubber on the Zero line is filled with a lithium based compound. Follow local governing ordinances for disposal.
2. Remove Input line 1 from the solenoid connector and the other end from the FilterLine input connector.
3. Remove Input line 2 from the solenoid connector and the other end from the sensor connector.
Section 7: Disassembly Guide

Figure 7-10: Lines to Solenoid (Zero line, Input line 1, Input line 2)

7-9
Section 7: Disassembly Guide
7.6.2.2 Reassembly of the lines to Solenoid

1. Connect the new Input line1 to the line 1 solenoid connector.

2. Connect the new Input line 2 to the line 2 solenoid connector.

3. Bring the Input line 2 tubing over Input line 1, and connect to
4. Lead the Input line 1 under the FRS cables (connected to the
5. Connect the Zero line to the Zero line solenoid connector. Place
Refer to Figure 7-10: Lines to Solenoid (Zero line, Input line 1,
Input line 2) and Figure 7-11: Zero line
the sensor connector.
FilterLine input connector, not viewed in Figure 7-10) and connect to the FilterLine input connector.
the scrubber close to the pump under the coiled Main line tubing (not viewed in Figure 7-11: Zero line) around the pressure sensor.
Figure 7-11: Zero
line
7-10
7.7 REPLACING LCD AND LED DISPLAY
(Refer to Figure 7-12: Replacing LCD and LED display)

7.7.1 Replacing the LCD

1. Remove the CO2 Board as described in Figure 7-4: Replacing CO2
Board and Housing Components.
2. Unscrew and remove the four LCD mounting screws.
3. Carefully remove the LCD and replace with a new LCD.

7.7.2 Replacing the LED Display

1. Remove the CO2 Board as described in section 7.3 "Replacing the CO2 Board".
2. Carefully remove the faulty LED Display from its socket and replace with a new LED Display. Note the correct orientation of the LED Display when replacing it.
Section 7: Disassembly Guide

Figure 7-12: Replacing LCD and LED display

7-11
Section 7: Disassembly Guide
7.8 REPLACING HOUSING COMPONENTS
The Housing kit includes the following replaceable components:
Front Cover
Rear Cover
ON-OFF Button
Upper Keypad (Alarm Silence/Alarm Silence Menu Button and Contrast/Value Change Button)
Lower Keypad (Event/Home Button and Next/Menu Button)
•Gas Outlet
•SpO
External SpO
Internal SpO
2 Insulation Plate
2 Connector Latch
2 Connector Lock
•Screws
•Washers
Note: When reassembling the monitor after replacing any part, be sure
to follow the disassembly steps in reverse order.

7.8.1 ON-OFF Button and Gas Outlet

(Refer to Figure 7-4: Replacing CO2 Board and Housing Components)

7.8.2 Keypads

(Refer to Figure 7-4: Replacing CO2 Board and Housing Components)
1. Open the monitor as described in section 7.2 "Opening the
Monitor Case".
2. Remove the Gas Outlet from the tubing and replace with a new
Gas Outlet.
2
3. Remove the ON-OFF Button from the CO
board as described in
section 7.3 "Replacing the CO2 Board", and replace with a new button.
1. Open the monitor as described in section 7.2 "Opening the
Monitor Case".
2. Remove the CO
2 Board as described in section 7.3 "Replacing the
CO2 Board".
3. Remove the old keypad. When replacing the new keypad, be sure
the icons are facing the correct direction. (Refer to Figure 1-1: Handheld Capnograph/Pulse Oximeter.)
7-12

7.8.3 Front Cover

(Refer to Figure 7-4: Replacing CO2 Board and Housing Components)
Section 7: Disassembly Guide
1. Open the monitor as described in section 7.2 "Opening the Monitor Case".
2. Remove the CO
2 Board as described in section 7.3 "Replacing the
CO2 Board".
3. Remove the keypad and ON-OFF Button. Place them in the new Front Cover. When replacing the keypad, be sure the icons are facing the correct direction. (Refer to Figure 1-1: Handheld Capnograph/Pulse Oximeter.)
4. Place the CO
2 board in the new cover.

7.8.4 SpO2 Connector Latch, Rear Cover and SpO2 Insulation Plate

(Refer to Figure 7-5: Replacing SpO2 Board, Figure 7-12: Replacing LCD and LED
display and Figure 7-13: Replacing SpO2 Connector Latch)
1. The External SpO
2 Connector Latch can be removed without
opening the monitor casing. Carefully pull the latch out of its hinges by squeezing the hinge and pulling the SpO Latch to the side. Replace with a new SpO
2. To replace the Rear Cover, open the monitor as described in section 7.2 "Opening the Monitor Case".
3. Remove the SpO
2 board and cables as described in section 7.4
"Replacing the SpO2 Board".
4. Remove the SpO
2 Insulation Plate and replace with a new plate
(if needed). Be sure to place the plate over the three screw mounts.
2 Connector
2 Connector Latch.
5. Place the SpO new Internal SpO
2 board and cables in the new back cover using a
2 Connector Lock.

Figure 7-13: Replacing SpO2 Connector Latch

7-13
Section 7: Disassembly Guide
7.9 UPDATING SOFTWARE VERSION
Warning: Observe ESD (electrostatic discharge) precautions when disassembling and reassembling the monitor and when handling any of the components of the monitor.
1. Open the monitor as described in section 7.2 "Opening the Monitor Case".
2. Remove the EPROM from its socket (on the CO
2 board) using an
IC extractor. (Refer to Figure 7-14: Replacing EPROM.)
3. Insert the new EPROM.
Note: When replacing the EPROM, be sure it is positioned correctly.
Note: When operating the monitor for the first time after updating the
software version, the warm-up and the self-test periods will be longer than the usual.
7-14

Figure 7-14: Replacing EPROM

SECTION 8: ELECTRICAL SAFETY TESTS
8.1 Electrical Safety Tests
8.1 ELECTRICAL SAFETY TESTS
Do not return the monitor to the user until the conditions are met in the following “Warning” statement.
Warning: After servicing the monitor, always verify product performance per the procedures in Section 10: Performance Verification and perform any safety tests required by local regulatory bodies or mandated by your institution.
8-1
SECTION 9: PERIODIC MAINTENANCE
9.1 Periodic Maintenance
9.1 PERIODIC MAINTENANCE
Periodic maintenance is recommended according to operating hours (refer to section 4.2 "Service Level 1"):
The Pump and Flow System should be replaced every 7,000 operating hours.
For replacing the pump, refer to section 7.5 "Replacing the Pump".
For replacing the flow system, refer to section 7.6 "Replacing the Flow System".
The monitor should be returned to the manufacturer for periodic maintenance every 21,000 operating hours.
9-1
SECTION 10: PERFORMANCE VERIFICATION
10.1 Flow System Leak Check
10.2 Performance Verification Procedure
The performance of the monitor must be verified after all service procedures. Perform all procedures in this section to verify the correct functioning of the monitor.
10.1 FLOW SYSTEM LEAK CHECK
Perform the Flow System Leak Check after all service procedures described in this manual.
10.1.1 Flow System Leak Check Procedure
1. Connect the vacuum manometer to the Leak Test Jig as shown in Figure 10-1: Leak Test Jig Connection.
Figure 10-1: Leak Test Jig Connection
2. Open the Leak Test Jig clamp and empty the syringe’s contents.
3. Connect the open-ended tubing (as shown in Figure 10-1: Leak Test Jig Connection) of the Leak Test Jig to the Gas Outlet of the monitor.
4. Connect the other open-ended tubing of the Leak Test Jig to the
input connector.
CO
2
5. Pull the syringe valve out until the pressure drops (as displayed on the vacuum manometer) to approximately -350 mBar (or equivalent) and close the clamp.
6. Wait for pressure to drop to 300 mBar.
7. Wait 30 seconds.
8. The change in reading should be less than 20 mBar (or equivalent).
10-1
Section 10: Performance Verification
9. If the reading drops more than 20 mBar, there is a leak in the flow system.
Open the unit.
Check for loose fittings and if any are found, reconnect them.
Repeat the above process (steps 1-8).
If the reading continues to show more than 20 mBar, replace the Flow System, refer to section 7.6 "Replacing the Flow System".
10.2 PERFORMANCE VERIFICATION PROCEDURE
Verify the performance of the monitor using the following procedure.
1. Connect the monitor to main power using the AC adapter.
2. Connect the FilterLine and the DS-100A - Durasensor monitor.
3. Turn ON the monitor.
4. The initialization screen appears. (Refer to Figure 10-2: Initialization Screen.)
5. Verify that an audio tone sounds. This verifies proper operation of the Alarm interfacing. Check that the LEDs on the alarm bar turn red, yellow and then off. The
Figure 10-2:
Initialization Screen
LED display shows 8s moving from left to right.
6. As the monitor continues to test its internal subsystems, the Self-Test Screen appears. (Refer to Figure 10-3: Self-Test Screen.).
Figure 10-3: Self-
Test Screen
7. After the Self-test, the monitor automatically goes to Measuring Mode. The message CO
2
Warmup appears on the Silent
Advisory Message area. (Refer to Figure 10-4: Measuring Mode.)
Figure 10-4: Measuring
Note: If BTPS is on, the message will
read BTPS ON-Ready
®
to the
Mode
10-2
8. After CO
2 Warmup, verify that the message changes to Ready.
9. Disconnect the FilterLine and verify that the message FilterLine is displayed on the Silent Advisory Message Area and verify that the pump stops.
Section 10: Performance Verification
10. Reconnect the FilterLine and verify that the message FilterLine disappears and the pump is operating.
11. Disconnect the DS-100A - O
XIMAX Durasensor
®
and verify that the
message
2 Sensor is displayed in the Silent Advisory Message area.
SpO
12. Reconnect the DS-100A - O
XIMAX Durasensor
®
and verify that the
message
2 Sensor disappears.
SpO
13. Verify that all buttons on the unit work properly. Refer to the monitor’s Quick Guide for their applications. (See Figure 10-5: Quick Guide.)
Figure 10-5: Quick Guide
14. Verify the CO
2 Calibration Check process functions. Follow the
procedure as described below. When completed, return to the Measuring Mode screen.
Note: Calibration must be performed annually or every 4000 operating
hours , whichever comes first, regardless of the CO
Calibration
2
Check results
10-3
Section 10: Performance Verification
CO2 Calibration Check
Caution: Do not check CO
2 values from the measuring mode if the
BTPS setting is ON (factory default). This mode corrects the CO value for BTPS. The calibration check mode disables this correction automatically.
In order to ensure accuracy, a CO
2 calibration check needs to be done once
a year. Calibration gas and a FilterLine are needed for this procedure. Start the process from the Interface Test screen (refer to Table 4-1: "Accessing Service Mode and Changing Parameter Settings") as follows in “Table 10-1: CO
Note: Connect the FilterLine to the monitor before starting CO
2 Calibration Check”.
2
Calibration Check. Use the calibration kit (5% gas in air and connecting means) approved by Oridion Medical 1987 Ltd.
Note: If the CO
2 Calibration Check is performed on a battery powered
device, ensure that the battery is fully charged prior to calibration.
Table 10-1: CO2 Calibration Check
Objective Action Result
Access CO Calib.
2
short press
3 times
2
Change option to Start.
Prepare for Cal. Check.
short press
No action
required
No action
required
Connect the FilterLine to the calibration gas.
10-4
Section 10: Performance Verification
Table 10-1: CO
2 Calibration Check (Continued)
Objective Action Result
Check the measured values shown in Vol% in the EtCO
2 digital
Press the gas valve for 15 seconds until the readings stabilize
display.
Calibration is not required if the measured value is the same as the concentration of the calibration gas (±0.3 Vol%) e.g. Concentration of calibration gas used is 5% (CO
2); the
measured value should be between 4.7%-5.3%, therefore, calibration is not required.
To reset and
long press return to measuring mode (automatically erasing event messages)
To exit service mode and save
Turn off the
monitor
event messages
If the reading is outside these limits, calibration is required. Refer to Table 10-2: "Calibration Process" for the calibration process procedure.
10-5
Section 10: Performance Verification
Calibration Process
If calibration is required, perform the Calibration Process (after performing CO Process".
Objective Action Result
2 Calibration Check) as shown in Table 10-2: "Calibration
Table 10-2: Calibration Process
To start calibration.
To start calibration process after calibration check.
To activate calibration process.
Adjust the display value to the gas concentration you are using.
Press the gas
valve and
long press
No action
required.
long press
short press
10-6
Monitor performs an internal calculation.
Calibration succeeded.
Return to calibration check mode.
Continue pressing the gas valve until the message
Disconnect Gas appears.
Disconnect Gas
from the
FilterLine.
No action
required.
long press
Section 10: Performance Verification
Table 10-2: Calibration Process (Continued)
Objective Action Result
To reset and return to measuring mode
To exit service mode and save event messages
long press
Turn off the
monitor.
Calibration Errors
Table 10-3: "Calibration Process - Troubleshooting" describes calibration problems that can occur, the message displayed and the appropriate corrective action.
Table 10-3: Calibration Process - Troubleshooting
Message Possible Causes Action
• Gas concentration does not match the concentration value selected in the first calibration screen.
• Problems with gas sampling
•Flow system problems (purging, blockage or no FilterLine)
Check setup, correct probable causes, and recalibrate unit as above.
•CO
2 Sensor problem
• Calibration process exceeds 3 minutes
•Contact your local distributor.
•Restart. If problem persists, contact your local distributor.
10-7
Section 10: Performance Verification
15. Advance to the Plethysmograph Display Screen as described in Table 10-4: "Plethysmograph Waveform" below.
To View Action Screen
CO
2 waveform
(Measuring Mode)
Plethysmograph short press (x4)
Table 10-4: Plethysmograph Waveform
No action
required.
16. Connect the DS-100A - O the SpO
2 value appears on the LED digital display and the Pulse
XIMAX Durasensor
®
to your finger. Verify
rate appears on the LCD graphic display.
17. Disconnect the monitor from the AC power supply. Verify the unit continues to work from its battery pack and the battery icon appears on the lower right corner of the display.
18. Reconnect the monitor to main line power with the AC adapter. Verify that the unit functions, and the plug icon appears.
Note: If the battery pack has a low charge level, the battery icon will appear instead of the plug. The battery icon will fill indicating the battery charging process.
10-8
Section 10: Performance Verification
To verify printing functions:
19. Adjust the printer to the following communication settings: Input Serial Data Length 8 bits Parity Settings None Baud 9600 bps
20. Using the Communication Adapter Kit, connect the printer to the monitor. (Refer to Figure 10-6: Connecting the monitor to Printer/ PC with Communication Adapter Kit)
Figure 10-6: Connecting the monitor to Printer/PC
with Communication Adapter Kit
21. Turn the printer ON and select on-line mode.
22. Access the monitor’s print functions as described in Table 10-5: "Accessing Monitor’s Print Functions":
10-9
Section 10: Performance Verification
Objective Action Result
Table 10-5: Accessing Monitor’s Print Functions
To access the
long press (x2) Instrument Setup menu (first access the Alarm Limits menu from any measuring display)
To select Print short press (x4)
Be sure the Print
short press option shows Screen; if it doesn’t, change to the Screen option.
To return to
long press
Measuring Mode
To print long press
simultaneously
and
(printout)
10-10
SECTION 11: PACKING FOR SHIPMENT
11.1 General Instructions
11.2 Packing Handheld Capnograph in Original Carton

11.3 Packing in a Different Carton

11.1 GENERAL INSTRUCTIONS
To ship a monitor or one of its components for any reason, follow the instructions in this section. Failure to follow the instructions in this section may result in loss or damage not covered by any applicable manufacturer warranty.
Pack the monitor or component(s) carefully. If available, use the original carton and packing materials and follow the instructions in “Packing monitor in Original Carton.” If the original shipping carton and material are not available, use other suitable shipping materials and container and follow the instructions in “Packing in a Different Carton.”
Prior to shipping the monitor or a component, contact your local representative for a Returned Material Authorization (RMA) number. Mark the shipping carton and any shipping forms with the RMA and the monitor’s serial number.
Caution: Observe ESD (electrostatic discharge) precautions when packing any monitor components.
11.2 PACKING MONITOR IN ORIGINAL CARTON
If the original carton and packing material are available, repack the monitor as follows. (Refer to Figure 11-1: Packing the Monitor in Original Packing.)
1. Place the unit in a plastic bag. Place it into the corresponding space, in the original foam padding, with the front panel facing up.
2. Place the foam padding cover over the top of the unit.
3. Seal the carton with packing tape.
4. Label the carton with the correct shipping address, return address and RMA number.
11.3 PACKING IN A DIFFERENT CARTON
If the original carton and packing material are not available or if shipping any monitor component not in its original carton:
1. Place the monitor in a plastic bag or component in a plastic anti­static bag.
2. Locate a corrugated cardboard shipping carton (approximately the size of the monitor or component) with at least 200 pounds per square inch (psi) bursting strength.
3. Fill the bottom of the carton with at least 2 inches of packing material.
11-1
Section 11: Packing For Shipment
4. Place the bagged monitor or component on the layer of packing
5. Seal the carton with packing tape.
6. Label carton with shipping address, return address and RMA
material and fill the box completely with packing material such that there is at least 2 inches of packing material around all sides of the item.
number.
Figure 11-1: Packing the Monitor in Original Packing
11-2
SECTION 12: SPECIFICATIONS
12.1 Physical
12.2 Environmental
12.3 Safety Standards
12.4 Performance
12.5 Power Specifications
12.6 Components and User Interface
12.1 PHYSICAL
12.1.1 Size
206 mm H x 88 mm W x 52.5 mm D (8.11”H x 3.46” W x 2.06”D)
12.1.2 Weight
850 grams (1.87 lb.) (including battery pack)
12.1.3 Noise Emission
maximum 45 dB(a)
12.2 ENVIRONMENTAL
12.2.1 Temperature
Operating
o
C to 45oC (32oF to 113oF)
0
Storage
o
C to 70oC (-31oF to 158oF)
-35
12.2.2 Relative Humidity
10 to 95% (noncondensing)
12.2.3 Pressure and Altitude (for operating and storage)
Pressure
430 mmHg to 795 mmHg
Altitude
-380m to 4,570m (-1,250 ft. to 15,000 ft.)
12.3 SAFETY STANDARDS
The monitor was designed to comply with EN60601-1/1990, A1/1993, A2/ 1995, UL2601-1, CSA C22.2 No. 601.1-M90, ISO 10993-1 (Sensors)
12-1
Section 12: Specifications
12.4 PERFORMANCE
12.4.1 Capnograph
Sampling Rate
50 ±7.5 ml/min.
CO
0-99 mmHg (0-13.2 kPa and 0-13.0 Vol%) at sea level
Accuracy
From power-up until steady state is reached, the CO2 reading accuracy is: 0 - 38 mmHg: (+ 4 mmHg) 39 - 99 mmHg: (+ 12% of reading)
The CO2 reading reaches its steady state accuracy 20 minutes after power up. 0 - 38 mmHg: (+ 2 mmHg) 39 - 99 mmHg: (+5% of reading + 0.08% for every 1 mmHg above 40mmHg)
Equivalent values for kPa and Vol%Frequency Response:
EtCO accuracy for respiration rate over 60 bpm, use the neonatal mode.) From 81 to 150 bpm accuracy is ±12%, if the EtCO in neonatal mode.
2 Range
2 accuracy is maintained up to 80 breaths/min. (for maintaining
2 is higher than 18.8 mmHg
Respiration Rate
0-150 breaths/min.
Warm-up Time
30 seconds (typical)
Rise Time
190 msec for neonatal mode with low dead space endotracheal tube adapter; 240 msec for adult mode with Microstream Airway Adapter
System Response Time
2.45 seconds (typical), 2.9 seconds maximum (includes delay time and rise
time)
Ambient Pressure
Compensated internally - automatic
Alarms
2 high, EtCO2 low, FiCO2 high, No Breath.
EtCO
12-2
12.4.2 Pulse Oximeter
12.4.2.1 Display Ranges
SpO2 Saturation Range
0-100% SpO
2
Saturation Accuracy
(% SpO
2, ±1 SD)
Adults 70-100%±2 digits
0 - 69% Unspecified
Neonates 70-100%±3 digits
0 - 69% Unspecified
Pulse Rate Range
20-250 beats per minute (bpm)
Pulse Rate Accuracy
± 3 bpm
Note: Pulse rate accuracy is expressed as ±3 bpm across the
display range. This variation equals ±1 standard deviation (1SD), which encompasses 68% of the population.
Section 12: Specifications
Alarms
2 high, SpO2 low, Pulse high, Pulse low
SpO
12-3
Section 12: Specifications
12.5 POWER SPECIFICATIONS
12.5.1 External Power Source
12V DC Medical Grade Adapter
12.5.2 Internal Power Source
Ni-MH Rechargeable Battery Pack 7.2V 2100 mA/h (intended for continuous operation)
Operating Time (fully charged)
Between 3 to 6 hours depending on power management. This spec reflects the performance of a new battery; age and usage will degrade capacity.
Note: If the battery pack is stored for 6 months or longer, you must charge
and discharge it (leave the unit on, not connected to AC power, until battery is empty) three times before use in order to ensure full capacity.
Recharging Period
Approximately 4.5 hours internal recharging
Charger Type
Internal
12.6 COMPONENTS AND USER INTERFACE
12.6.1 Displays
LCD graphic display
(128 x 64 dots) with LED backlight dimension 75 mm x 53 mm.
Two numeric fields
3 digits each, using 7-segment LED graphic display dimension 22mm x 14 mm.
Two-color alarm bar
yellow, red
12.6.2 Controls and Indicators
Front Panel
ON/OFF switch; Alarm Silence/Alarm Menu button; Contrast/Value Change button; Event/Home button; Next/Menu button.
12.6.3 Connections
Front Panel
2 Input connector
CO
Top Panel
2 Input socket
SpO
12-4
Rear Panel
Clamp connector, battery release button
Side Panel
Power Supply/Communication Adapter Port, Gas output connector
SECTION 13: SPARE PARTS
13.1 Spare Parts List
13.1 SPARE PARTS LIST
Spare parts are shown below in Table 13-1. For spare part numbers contact your local distributor.
Item Description
Table 13-1: Spare Parts List
1
CO
2 board - includes electronic board, sensor, pump
and scrubber, solenoid, LCD, 7-segment display, flow system, CO
2 input connector, software EPROM and
screws.
2 Software Assembly - includes EPROM chip and last
updated software version.
3SpO
4SpO
Board
2
Kit - includes inter-board connection cable,SpO2
2
input cable, SpO
insulation plate and internal
2
connector lock
5 Housing Kit - includes front cover, rear cover,
ON/OFF button, keypads, gas outlet, screws, external SpO
connector latch, SpO2 insulation plate and
2
internal SpO
connector lock
2
6 7-Segment Digital Display (x2)
7 LCD Assembly - Graphic Display Screen and PC
driver
8 Calibration Gas (5% CO
2 in air) and FilterLine
9 Leak Test Jig - vacuum test accessories
10 Communication Adapter Kit
11 Battery Pack
12 Pump Assembly
13 Flow System
13-1
GLOSSARY OF TERMS
BIT: Built in Tests.
BTPS: Body Temperature, Pressure and Saturation
Compensation.
Where Body Temperature is assumed 37°C with
vapor pressure 47mmHg
The Calculations are made according to:
P
CO2
= F
x (Pb - 47)
CO2
Where:
F
is the Fractional concentration of CO2 in
CO2
Dry gas, F
= % CO2/100
CO2
Pb = the ambient pressure
P
= the partial pressure of CO2 at BTPS
CO2
EPROM: Erasable Programmable Read Only Memory.
ESD: Electrostatic Discharge.
EtCO
: End Tidal Carbon Dioxide: amount of CO2 present
2
at the end of the exhalation.
FiCO
: Fractional Inspired Carbon Dioxide: amount of
2
present during inhalation.
CO
2
FRS: FilterLine Recognition Safeguard.
LCD: Liquid Crystal Display.
HMI: Human Machine Interface.
Respiration Rate: Number of respiratory cycles (inhalation and
exhalation) per minute.
7-segments LED: Graphic display of seven segments of light emitter
diodes.
Solenoid: 2-way electro-pneumatic valve controlling the
directional flow either from the patient or from the
Scrubber (during Autozero).
CO
2
Scrubber: Lithium compound CO2 absorber.
Loading...