Nellcor OxiMax N-600 User manual

Page 1
N-600
Pulse Oximeter Operator’s Manual
Page 2
Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor, OxiCliq, Dura-Y, Max-Fast, SoftCare, and OxiMax are trademarks of Nellcor Puritan Bennett Inc.
Cet appareil ISM est conforme à la norme NMB-001 Canada.
To obtain information about a warranty, if any, contact Nellcor’s Customer Services Department, or your local representative.
Page 3

Contents

Contents
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . i
Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vii
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Warnings ........................................................................................... 1
Cautions ............................................................................................ 2
Notes ................................................................................................. 3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Intended Use for the N-600 ............................................................... 5
How to Use this Manual .................................................................... 6
Description of Controls, Indicators, and Symbols . . . . . . . . . . . . . . . . . . . . . . 7
Identification of Front Panel Buttons and Symbols ........................... 7
Identification of Rear Panel Components .......................................... 8
N-600 Symbols .................................................................................. 8
Description of Controls ...................................................................... 9
Description of Displays and Indicators ............................................ 10
Description of Audible Indicators ..................................................... 14
Setting up the N-600 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
List of Components ......................................................................... 19
Connecting the N-600 to AC Power ................................................ 19
Connecting an OxiMax Sensor to the N-600 ................................... 21
Battery Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
Operating the N-600 on Battery Power ........................................... 23
Low Battery Indicator ...................................................................... 24
OxiMax N-600 i
Page 4
Contents
Using the N-600 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .31
Introduction ..................................................................................... 31
Turning On the N-600 ..................................................................... 34
OxiMax Sensor Attached ...................................................... 36
No OxiMax Sensor Attached ................................................ 38
Turning the Backlight On or Off ...................................................... 39
Adjusting Screen Contrast .............................................................. 39
Selecting the Pleth View ................................................................. 40
Selecting the Blip View ................................................................... 41
Setting the Pulse Beep Volume ...................................................... 41
Setting the Alarm Volume ............................................................... 42
Setting the Date and Time .............................................................. 43
Setting Alarm Silence Duration ....................................................... 45
Disabling Audible Alarms ................................................................ 46
Selecting Standby Mode ................................................................. 47
Adult-Pediatric or Neonatal Settings ............................................... 48
Setting Patient Adult-Pediatric/Neonatal Mode .................... 49
Alarm Limit Changed Indicator ....................................................... 50
Setting Alarm Limits ........................................................................ 50
Setting SatSeconds Alarm Limit ..................................................... 52
Setting the N-600 Response Mode ................................................. 53
Selecting the Display Language ..................................................... 55
OxiMax Sensor Messages .............................................................. 56
OxiMax Sensor Adjust Condition Messages ........................ 57
OxiMax Sensor Adjust Messages ........................................ 58
N-600 Trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .59
N-600 Trend Data ........................................................................... 59
Trend Data Operation ........................................................... 61
Selecting the Trend Data Display Scale ......................................... 62
Reading the Trend Data Display ..................................................... 64
Dual Trend Data Display ................................................................. 65
SpO2 Trend Display ....................................................................... 65
Pulse Rate Trend Display ............................................................... 66
Histogram Trend Data Display ........................................................ 67
Pulse Amplitude Trend Data Display .............................................. 68
Clearing Trend Information ............................................................. 69
ii
Page 5
Contents
OxiMax Sensor Event Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 71
Setting In-Sensor Data Type ........................................................... 72
OxiMax Sensor Type ....................................................................... 74
OxiMax Sensor Data Type .............................................................. 75
OxiMax Sensor Event Record Data Available ................................. 75
OxiMax Sensor Event Record Not Available ................................... 77
OxiMax Sensor Event Record Graphical Data ................................ 77
Viewing and Printing OxiMax Sensor Event History Data ............... 79
OxiMax Sensor Tabular Event Data ................................................ 81
Viewing and Printing In-Sensor Tabular Event History Data ........... 82
Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 85
Printing N-600 Trend Information .................................................... 85
N-600 Trend Data in ASCII Mode ................................................... 88
Trend Data in Graph Mode ............................................................. 89
Real-Time Display/Printout Format ................................................. 89
Column Headings ................................................................. 91
Data Source .......................................................................... 91
Software Version .................................................................. 91
Alarm Limits .......................................................................... 92
Monitor Mode ........................................................................ 92
Response Mode .................................................................... 92
Data Column Headings ......................................................... 93
Time ...................................................................................... 93
Patient Data .......................................................................... 93
Operating Status ................................................................... 95
Using the Data Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 97
Overview ......................................................................................... 97
Connecting to the Data Port ............................................................ 97
Data Port Pinouts .................................................................. 98
Data Port Setup ............................................................................... 99
Using the Nurse Call Interface ...................................................... 101
Setting Nurse Call RS-232 Polarity ..................................... 102
Setting Nurse Call Relays Normally Open/Closed .............. 103
Calculating the Analog Voltage Output ......................................... 103
OxiMax N-600 iii
Page 6
Contents
OxiMax Sensors and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .105
OxiMax Sensor Event Record Data .............................................. 105
Selecting an OxiMax Sensor ........................................................ 105
OxiMax Sensor Features .............................................................. 109
Biocompatibility Testing ................................................................ 110
Optional Accessories .................................................................... 110
GCX Mounting Plate ........................................................... 111
GCX Vertical Wall Mount Arm ............................................ 112
GCX Roll Stand .................................................................. 113
Soft-Sided Carrying Case ................................................... 114
Performance Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 117
Performance Verification .............................................................. 117
N-600 Performance Considerations ............................................. 117
Dysfunctional Hemoglobins ................................................ 118
Anemia ............................................................................... 118
Saturation ........................................................................... 118
Pulse Rates ........................................................................ 118
OxiMax Sensor Performance Considerations ............................... 119
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .123
Error Codes .................................................................................. 123
Prompts and Error Messages ....................................................... 125
Corrective Action .......................................................................... 128
EMI (Electromagnetic Interference) .............................................. 131
Obtaining Technical Assistance .................................................... 133
OxiMax Sensor Message Setup ................................................... 133
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .135
Returning the N-600 ..................................................................... 135
Service .......................................................................................... 135
Periodic Safety Checks ................................................................. 136
Cleaning ........................................................................................ 136
Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .137
N-600 Menu Description ............................................................... 137
iv
Page 7
Contents
SatSeconds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 141
Describing SatSeconds ................................................................. 141
SatSeconds “Safety Net” ............................................................... 143
SatSeconds Display ...................................................................... 144
Factory Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 145
Neonate Default Settings .............................................................. 145
Adult Default Settings .................................................................... 146
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 149
Oximetry Overview ........................................................................ 149
Automatic Calibration .......................................................... 150
Functional versus Fractional Saturation ............................. 150
Measured versus Calculated Saturation ............................. 151
OxiMax Technology ...................................................................... 152
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 153
Performance .................................................................................. 153
Electrical ........................................................................................ 154
Environmental Conditions ............................................................. 156
Physical Characteristics ................................................................ 158
Compliance ................................................................................... 158
Manufacturer’s Declaration ........................................................... 160
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
OxiMax N-600 v
Page 8

Tables

Tables
Table 1: Audible Indicators .................................................... 14
Table 2: Low Battery and Critical Battery .............................. 26
Table 3: Parameter Ranges .................................................. 31
Table 4: Reading Trend Display ............................................ 62
Table 5: Data Port Pinouts .................................................... 96
Table 6: Analog Pinouts ...................................................... 101
Table 7: Nellcor Oximetry Sensor Models and
Patient Weights ..................................................... 105
Table 8: OxiMax Sensor Features ....................................... 107
Table 9: Error Codes ........................................................... 122
Table 10: Prompt/Error Messages ........................................ 124
Table 11: Neonate Factory Defaults ...................................... 143
Table 12: Adult Factory Defaults ........................................... 144
Table 13: Electromagnetic Emissions ................................... 157
Table 14: Electromagnetic Immunity,
RF Portable Equipment ......................................... 158
Table 15: Recommended Separation Distances ................... 161
Table 16: Cable Compliance ................................................. 162
vi
Page 9

Figures

Contents
Figure 1: Front Panel Buttons and Symbols ................................. 7
Figure 2: Rear Panel Components ............................................... 8
Figure 3: Trend Overwriting ........................................................ 59
Figure 4: ASCII Mode Printout ................................................... 86
Figure 5: Graph Mode Printout ................................................... 87
Figure 6: Real-Time Printout ...................................................... 88
Figure 7: Data Port Pin Layout ................................................... 97
Figure 8: GCX Mounting Plate .................................................. 109
Figure 9: GCX Poly-Mount Vertical Wall Mount ....................... 110
Figure 10: GCX Poly-Mount Horizontal Wall Mount ...................111
Figure 11: GCX Poly-mount Roll Stand ...................................... 112
Figure 12: GCX Utility Basket ..................................................... 113
Figure 13: Soft-Sided Carrying Case .......................................... 114
Figure 14: Alarm Response with SatSeconds ............................ 140
Figure 15: Oxyhemoglobin Dissociation Curve .......................... 149
OxiMax N-600 vii
Page 10
Page 11

Safety Information

Warnings

Warnings are identified by the WARNING symbol shown above.
Warnings alert the user to potential serious outcomes (death, injury, or adverse events) to the patient or user.
WARNING: The sensor extrapolates from the date and time provided by the N-600 when recording the sensor event record to the sensor. The accuracy of the date/time is the responsibility of the N-600. It is recommended that the N-600 user set the time/ date to the correct value before a sensor event record-enabled sensor is connected, and that this date/time not be changed while the sensor remains connected. Since a sensor with sensor event record data can be transported from one monitor to another, having discrepancies in the date/time between monitors and the sensor event record data will affect the order the sensor event record data appears. To eliminate this possible problem, all monitors within an institution should be set to the same time.
WARNING: Explosion hazard. Do not use the N-600 pulse oximeter in the presence of flammable anesthetics or gases.
WARNING: Chemicals from a broken LCD display panel are toxic when ingested. Use caution when handling a pulse oximeter with a broken display panel.
OxiMax N-600 1
Page 12
Safety Information
WARNING: Pulse oximetry readings and pulse signals can be affected by certain environmental conditions, OxiMax sensor application errors, and certain patient conditions. See the appropriate sections of this manual for specific safety information.
WARNING: The use of accessories, sensors, and cables other than those specified may result in increased emission and/or decreased immunity and inaccurate readings of the N-600 pulse oximeter.
WARNING: Failure to cover the OxiMax sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.

Cautions

2
Cautions are identified by the CAUTION symbol shown above.
Cautions alert the user to exercise care necessary for the safe and effective use of the N-600 pulse oximeter.
Page 13
Safety Information
Caution: When connecting the N-600 to any instrument, verify proper operation before clinical use. Both the N-600 and the instrument connected to it must be connected to a grounded outlet. Accessory equipment connected to the pulse oximeter’s data interface must be certified according to IEC Standard 950 for data-processing equipment or IEC Standard 60601-1 for electromedical equipment. All combinations of equipment must be in compliance with IEC Standard 60601-1-1 systems requirements. Anyone who connects additional equipment to the signal input port or signal output port (N-600 data port connector) configures a medical system and is therefore responsible for ensuring that the system complies with the requirements of system standard IEC Standard 60601-1-1 and the electromagnetic compatibility system standard IEC Standard 60601-1-2. The N-600 accuracy may degrade if it is connected to secondary I/O devices when the instrument is not connected to earth reference.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
Caution: Dispose of battery in accordance with local requirements and regulations.

Notes

Notes are identified by the Note symbol shown above.
Notes provide additional useful information.
OxiMax N-600 3
Page 14
Blank Page
Page 15

Introduction

WARNING: The N-600 is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.

Intended Use for the N-600

The OxiMax® N-600 pulse oximeter is indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO with neonatal, pediatric, and adult patients who are well or poorly perfused, in hospitals, hospital-type facilities, intra-hospital transport, and home environments. For prescription use only.
Note: Hospital use typically covers such areas as general care floors,
operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
2) and pulse rate. The N-600 is intended for use
Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.
Home Care use is defined as managed/used by a lay person (parent or other similar non-critical caregiver) in the home environment.
Use with any particular patient requires the selection of an appropriate oxygen OxiMax sensors as described in this manual.
OxiMax N-600 5
Page 16
Introduction

How to Use this Manual

All users should read this manual thoroughly. More experienced users of the N-600 will be able to go to the topics for the information they require.
The current copy of this manual is available on the internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
6
Page 17

Description of Controls, Indicators, and Symbols

Identification of Front Panel Buttons and Symbols

Adjust Down Button, page 10
1. SpO2 OxiMax Sensor Port, page 21
2.
AC Power Indicator, page 12 13. Neonate Mode Indicator, page 13
3.
On/Standby Button, page 9
4.
Low Battery Indicator, page 12 15. Fast Response Mode Indicator,
5. Waveform Display, page 11 16. Softkeys, page 10
6.
SatSeconds
7. %SpO
8. Pulse Rate Display, page 12
9.
10.
11.
2 Display, page 12
Alarm Silence Indicator, page 13 20. Pulse Search Indicator, page 13
Alarm Silence Button, page 10 Adjust Up Button,page 10
TM
Indicator, page 13
12.
Contrast Button, page 10
14.
page 13
17. Menu Bar, page 10
18.
Data In Sensor Indicator, page 13
19.
Interference Indicator, page 13
21. Speaker
Figure 1: Front Panel Buttons and Symbols
OxiMax N-600 7
Page 18
Description of Controls, Indicators, and Symbols

Identification of Rear Panel Components

1. Equipotential Terminal (Ground) 4. Fuse Holder
2. AC Power Connector, page 19 5. Supply Voltage Selector Switch, page 17
3. Data Port Connector, page 97
Figure 2: Rear Panel Components

N-600 Symbols

8
The symbols that are located on the rear panel of the N-600 are as follows:
See Instructions for Use
Fuse Replacement
Page 19
Equipotential Terminal (ground)
Date of Manufacture
Data Interface
There is one symbol located on the front panel of the N-600.
Type BF Applied Part - Not defibrillator proof.

Description of Controls

Description of Controls, Indicators, and Symbols
Note: A button press, except the On/Standby button, should result
in either a valid or an invalid key tone (refer to Table 1 on page 14). If the key pressed fails to emit a tone, contact qualified service personnel.
The
On/Standby button. Used to turn the N-600 on or off.
OxiMax N-600 9
Page 20
Description of Controls, Indicators, and Symbols
The Alarm Silence button. Used to silence current alarms for the alarm silence duration period. When an alarm has been silenced, pressing the button again reactivates, or “unsilences” the alarm. It is also used to view and adjust alarm silence duration and alarm volume. The
Alarm Silence button clears “SENSOR OFF,” “LOW
BATTERY,” and “SENSOR DISCONNECT” messages from the display.
The
Adjust Up button. Used to increase variable parameters of the
N-600.
The
Adjust Down button. Used to decrease variable parameters in
the N-600.
Contrast button. Used in conjunction with the Adjust Up and
The
Adjust Down buttons to lighten or darken the display screen.
The softkey buttons have multiple uses depending on the legend displayed above the button.

Description of Displays and Indicators

The type of display is user selectable. Refer to Selecting the Pleth View on page 40.
10
Page 21
Description of Controls, Indicators, and Symbols
The pleth display includes a “wiper bar” plethysmographic waveform, menu bar, and current measured %SpO SatSeconds are enabled, the pleth display includes the Indicator and SatSeconds setting. A decimal point after the %SpO
2 and pulse rate. If
SatSeconds
2 or
pulse rate indicate that the respective limits have been changed from the power on defaults (N-600 Trend Data on page 59).
The blip display includes a pulse amplitude blip bar, current measured %SpO
2 and pulse rate, and current upper and lower %SpO2 and pulse
rate limits. If SatSeconds are enabled, the blip display includes the
SatSeconds Indicator and SatSeconds setting. Decimal points after
the %SpO
2 or pulse rate indicate that the respective limits have been
changed from the power-on defaults.
There are various matrices within the N-600 algorithm. Some of these, are used to assess the severity of conditions presented to the N-600 in measuring SpO
2 and pulse rate on a patient. These
individual matrices or combinations of these matrices are used to drive the LED indicators on the N-600 front panel.
The advanced signal processing in the N-600 algorithm automatically extends the amount of data required for measuring SpO
2 and pulse
rate depending on the measurement conditions. During normal measurement conditions the averaging time is six seconds to seven seconds (approximately three seconds in Fast Mode). During conditions such as those caused by low perfusion, interference (e.g., external interference like ambient light), or a combination of the two, the N-600 algorithm automatically extends the amount of data required beyond the minimum as set by the response mode. See Setting the N-600 Response Mode on page 53. If the resulting dynamic averaging time exceeds 20 seconds, the indicator is lit solid and SpO
OxiMax N-600 11
2 and Pulse Rate will continue to be
Pulse Search
Page 22
Description of Controls, Indicators, and Symbols
updated every second. As these conditions extend, the amount of data required continues to increase. If the dynamic averaging time reaches 40 seconds, the pulse search indicator begins flashing, the SpO pulse rate displays flash zeros indicating a loss-of-pulse condition.
WARNING: Failure to cover the OxiMax sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.
2 and
The %SpO
2 Display. Shows the hemoglobin oxygen saturation level.
The display value flashes zeros during loss-of-pulse alarms and flashes the SpO
2 value when the SpO2 is outside the alarm limits.
During Pulse Search, the N-600 continues to update the display. If alarm limits have been changed from their power-on defaults, a decimal point (.) is displayed after the SpO
The
Pulse Amplitude Indicator (blip bar). Indicates pulse beat and
2 value (81.).
shows the relative pulse amplitude. As the detected pulse becomes stronger, more bars light with each pulse. This indicator is available only in the blip view.
The
Pulse Rate Display. Shows the pulse rate in beats per minute. It
flashes during loss-of-pulse alarms and when the pulse rate is outside of the alarm limits. During Pulse Search, the N-600 continues to update the display. Pulse rates outside of the pulse rate range (20 to 250 bpm) are displayed as the closest value within the range. If alarm limits have been changed from their power-on defaults, a decimal point (.) is displayed after the BPM value (112.).
The
AC Power Indicator. Lights continuously when the N-600 is
connected to AC power. It also indicates that the battery is charging. It is off when the N-600 is being powered by internal battery.
12
The
Low Battery Indicator. Lights continuously when 15 or fewer
minutes of battery capacity remain. Flashes when the battery capacity reaches critical condition.
Page 23
Description of Controls, Indicators, and Symbols
The Alarm Silence Indicator. Lights continuously when an audible alarm has been silenced. It flashes when the alarm silence duration has been set to Off.
The
Interference Indicator. The Interference Indicator lights
whenever the N-600 algorithm detects that the incoming signal
1
quality is degraded
.
An intermittently lit
Interference Indicator is common during
patient monitoring, and indicates the N-600 algorithm is dynamically adjusting the amount of data required for measuring SpO
2 and Pulse
Rate. When lit continuously, it indicates the N-600 algorithm has extended the amount of data required for measuring SpO
2 and Pulse
Rate and, consequently, fidelity in tracking rapid changes in these values may be reduced (see Description of Displays and Indicators on page 10).
The
Pulse Search Indicator. Lights continuously prior to initial
acquisition of a pulse signal and during prolonged and challenging monitoring conditions. The Pulse Search Indicator flashes during a loss-of-pulse signal.
The
Data In-Sensor Indicator. Lights to indicate that the attached
OxiMax sensor contains a patient sensor event record. The sensor event record information may be viewed or printed.
The
SatSeconds Indicator. Fills in clockwise as the SatSeconds
alarm management system detects a %SpO limit setting. Empties in counterclockwise direction when %SpO
2 reading outside of the
2
reading is within limits. When the indicator is full, a medium priority alarm will sound.
The
Fast Response Mode Indicator. The response mode setting
dictates the response time (two-seconds to four-seconds in fast mode and four-seconds to seven-seconds in normal mode) applied by the N- 600 algorithm in its calculation of SpO
1
Degradations can be caused by ambient light, electrical noise, electro-surgical interference, patient activity, or other causes.
OxiMax N-600 13
2. The N-600 algorithm
Page 24
Description of Controls, Indicators, and Symbols
calculation of pulse rate is unaffected by the response mode setting. The trending interval (two-seconds or four-seconds) is updated automatically by the N-600 to roughly correspond with the SpO calculation response time.
The
Neonate Alarm Limits Indicator. This symbol is displayed
when the alarm limits are set to neonate. No symbol is displayed when the N-600 is set to adult limits.

Description of Audible Indicators

Table 1 identifies the audible indicators of the N-600 indicators.
Table 1: Audible Indicators
Function Description
2
Alarm Silence Reminder
Confirmation Tone Three beeps sound to indicate that default
Invalid Button Press
Valid Button Press
High Priority Alarm
Medium Priority Alarm
Low Priority Alarm
Three beeps that sound approximately every three minutes when alarms are silenced with the alarm silence duration set to Off and the alarm silence reminder function is enabled.
settings have been saved or reset to factory defaults or trend data has been deleted.
A short, low button has been pressed that is not appropriate for the current state of the N-600.
A short, medium an appropriate button has been pressed.
A high loss
-of-pulse.
A medium SpO
2 or pulse rate limit violation.
A low
-pitched tone indicating that a
-pitched tone indicating that
-pitched, fast-pulsing tone indicating
-pitched, pulsing tone indicating an
-pitched, slow-pulsing tone indicating an
OxiMax sensor disconnect, low battery, or
N-600 failure.
14
Page 25
Description of Controls, Indicators, and Symbols
Table 1: Audible Indicators
Function Description
Power
-On Self-Tes t
Pass
A 1-second tone indicating that the N-600 has been turned on and has successfully completed the power
-on self-test.
Pulse Beep A single beep sounds for each detected pulse.
Volume Setting Tone A continuous tone that is used when adjusting
The pitch of the pulse beep signal changes with
-by-point rise or fall in the saturation
a point level.
the alarm volume.
OxiMax N-600 15
Page 26
Blank Page
Page 27

Setting up the N-600

WARNING: To ensure patient safety, do not place the pulse oximeter in any position that might cause it to fall on the patient.
WARNING: As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
WARNING: Ensure that the speaker is clear of any obstruction. Failure to do so could result in an inaudible alarm tone.
WARNING: Disconnect the N-600 and Nellcor OxiMax sensor from the patient during magnetic resonance imaging (MRI) scanning. Induced current could potentially cause burns.
WARNING: To ensure accurate performance and prevent device failure, do not subject the N-600 to extreme moisture, such as direct exposure to rain. Such exposure may cause inaccurate performance or device failure.
WARNING: Do not use an N-600 pulse oximeter, OxiMax sensor, cables, or connectors that appear damaged.
WARNING: Do not lift the pulse oximeter by the pulse oximetry cable or power cord because the cable or cord could disconnect from the pulse oximeter, causing the pulse oximeter to drop on the patient.
OxiMax N-600 17
Page 28
Setting up the N-600
WARNING: The N-600 is not defibrillator-proof. However, it may remain attached to the patient during defibrillation or while an electrosurgical unit is in use, but the readings may be inaccurate during the defibrillation and shortly thereafter.
WARNING: In the USA, do not connect the pulse oximeter to an electrical outlet controlled by a wall switch, because the pulse oximeter may be accidentally turned off.
WARNING: Use only the Nellcor pulse oximetry cable DOC-10 with the N-600 pulse oximeter. Use of another pulse oximetry cable will have an adverse effect on performance. Do not attach any cable that is intended for computer use to the OxiMax sensor port. Do not connect any device other than a Nellcor-approved OxiMax sensor to the OxiMax sensor connector.
18
WARNING: The N-600 should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the N-600 should be observed to verify normal operation in the configuration it is to be used.
Page 29

List of Components

Setting up the N-600
Quantity Item
1 N-600 Pulse Oximeter
1 Nellcor OxiMax Sensor or Assortment Pack
1
1 N-600 Operator’s Manual (applicable to
1 Power Cord (applicable to country of sale)
1 Quick Guide
DOC
country of sale) and/or Compact Disc

Connecting the N-600 to AC Power

WARNING: In the USA, do not connect the pulse oximeter to an electrical outlet controlled by a wall switch, because the pulse oximeter may be accidentally turned off.
Caution: The SUPPLY VOLTAGE SELECTOR switch must be set to the correct voltage (115 or 230) to avoid equipment damage and ensure battery charging.
-10 Pulse Oximetry Cable
OxiMax N-600 19
Page 30
Setting up the N-600
Caution: Use only the hospital-grade power cord provided by Nellcor.
1 — Power Connector 2 — Supply Voltage Selector
1. Set the SUPPLY VOLTAGE SELECTOR (2) switch to the
applicable voltage.
20
2. Plug the female connector end of the power cord into the N-600
POWER CONNECTOR (1) on the rear of the N-600.
3. Plug the male connector of the power cord into a properly
grounded AC outlet.
4. Verify that the N-600’s
Note: If the
AC Power Indicator is not lit, check:
AC Power Indicator is lit.
the power cord
the SUPPLY VOLTAGE SELECTOR switch
•the user-accessible fuses
the AC power outlet
Page 31

Connecting an OxiMax Sensor to the N-600

The OxiMax sensor type is displayed at the bottom of the display when an OxiMax sensor is connected to the N-600 or when the N-600 completes POST with an OxiMax sensor attached.
Caution: Use only Nellcor-approved OxiMax sensors and pulse oximetry cables.
Note: Physiological conditions, medical procedures, or external
agents that may interfere with the N-600’s ability to detect and display measurements include dysfunctional hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring agents, such as nail polish, dye, or pigmented cream.
Setting up the N-600
1 — SpO
1. Connect a DOC-10 pulse oximetry cable to the SpO
2 OxiMax Sensor Port
2 OxiMax
sensor port (1) of the N-600.
2. Connect a Nellcor OxiMax SpO
2 sensor to the other end of the
DOC-10 pulse oximetry cable.
OxiMax N-600 21
Page 32
Blank Page
Page 33

Battery Operation

WARNING: Dispose of battery in accordance with local requirements and regulations.

Operating the N-600 on Battery Power

The N-600 has an internal battery that can be used to power the N-600 during transport or when AC power is not available. A new, fully charged battery will provide at least two hours of monitoring time under the following conditions:
No audible alarms sound
No analog or serial output devices are attached to the N-600
The N-600 cannot operate with a fully discharged battery. Before attempting to turn on an N-600 whose battery charge has been completely depleted, first plug the N-600 into an AC outlet to allow the battery to charge for a few minutes. The N-600 may then be powered on.
To charge a low or dead battery, connect the N-600 to AC power. A full charge of a dead battery takes 14 hours while the N-600 is turned off. A full charge of a dead battery takes 18 hours while the N-600 is in operation (monitoring a patient).
When all of the following conditions are present for 15 minutes, the N-600 will automatically shut down:
N-600 is running on battery power
No buttons have been pressed
OxiMax N-600 23
Page 34
Battery Operation
Note: Whenever the N-600 is connected to AC power, the battery is

Low Battery Indicator

The Low Battery Indicator lights and a low priority alarm begins to sound when approximately 15 minutes of monitoring time is available on the existing battery charge. Refer to Table 2 for a description of the low and critical battery conditions.
No pulse has been detected (for example, when a patient is not connected to the OxiMax sensor or the OxiMax sensor is disconnected from the N-600)
No alarms are present (other than low battery or a non-correctable error)
being charged. Therefore, it is recommended that the N-600 remain connected to AC power when not in use. This will ensure a fully charged battery whenever it is needed.
24
If the N-600 is not on AC power, a low battery audible alarm can be canceled by pressing the Indicator and display screen message will continue to be displayed. Plugging the N-600 into AC power will silence the audible alarm, but the
Low Battery Indicator will stay lit as long as the battery is in the
low voltage condition. After the 15-minute period of low battery condition, a high priority alarm will sound for about 10 seconds before the N-600 shuts off.
If the N-600 backlight is turned off during a low battery condition, the backlight cannot be turned back on until the low battery condition is corrected.
It is recommended that qualified service personnel replace the internal battery every 24 months. Replaced batteries should be disposed of in accordance with local ordinances.
Alarm Silence button. The Low Battery
Page 35
Battery Operation
Caution: If the N-600 pulse oximeter is to be stored for a period of three months or longer, notify service personnel to remove the battery from the pulse oximeter prior to storage. Recharge the battery when the battery has not been charged for two or more months.
Caution: The pulse oximeter default settings will return to factory default setting if the battery becomes fully discharged or is replaced. Qualified service personnel will have to reset the institutional defaults, following the instructions in the N-600 Service Manual.
Note: If the AC voltage selector switch on the N-600 rear panel does
not match your AC voltage source, the N-600 may run on battery power, even though it is plugged into AC power, which will eventually result in a low priority alarm and a lighted
Battery
Indicator. Ensure that the switch setting matches your
Low
AC voltage.
Note: As the battery is used and recharged over a period of time, the
amount of time between the onset of the low battery alarm and the N-600 shut-off may become shorter
OxiMax N-600 25
Page 36
Battery Operation
Table 2: Low Battery and Critical Battery
State
1No No Yes
2No No No
Critical Battery
Low Battery
AC Operation
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY message
Audible alarm
Error code
Effect of key
Shutdown
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY message
2- normal
-on
-off
-off
-off
-none
Alarm Silence
-normal
-N/A
2-normal
-off
-off
-off
26
Audible alarm
Error code
Effect of key
Shutdown
-none
Alarm Silence
-normal
- N/A
-off
Page 37
Battery Operation
Table 2: Low Battery and Critical Battery (Continued)
State
3No Yes No
Critical Battery
Low Battery
AC Operation
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY message
Audible alarm
Error code
Effect of
SILENCE
silences audio alarm, second press cancels LOW BATTERY message. The LED stays on until Low Battery Condition is corrected.
Shutdown
2-normal
-off
-on
-on
-low priority
-logged
ALARM
key-First press
-N/A
OxiMax N-600 27
Page 38
Battery Operation
Table 2: Low Battery and Critical Battery (Continued)
State
4No Yes Yes
Critical Battery
Low Battery
AC Operation
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY message
Audible alarm
Error code
Effect of key until low battery condition is corrected.
Shutdown
Note: Connecting AC
2-normal
-on
-on
-off
-off
-logged
Alarm Silence
-N/A. The LED stays on
-N/A
functions the same as
Alarm Silence key
in state 3.
28
5 Not used
Page 39
Battery Operation
Table 2: Low Battery and Critical Battery (Continued)
State
6Yes Yes No
7 Yes Yes Yes
Critical Battery
Low Battery
AC Operation
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY
Audible alarm
Error code
Effect of
Shutdown
SpO
AC/Battery charge LED
LOW BATTERY LED
2-not displayed
(flashing)
message
-on
-displayed and
logged
Alarm Silence
key
-none
-after 10 seconds
2-not displayed
(flashing)
-off
-on
-high priority
-on
-on
LOW BATTERY
message
Audible alarm
Error code
logged
Effect of
key
Shutdown
OxiMax N-600 29
-on
-high priority
-displayed and
Alarm Silence
-N/A
-after 10 seconds
Page 40
Blank Page
Page 41

Using the N-600

Introduction

The parameters of the N-600 are preset to factory default settings. See Factory Defaults on page 145. The factory default parameters may be changed to institutional default parameters by following the procedures in the N-600 Service Manual.
Table 3 lists the parameters, ranges available, and the factory default setting. The parameters may be set on an individual basis, by the clinician, and these settings will remain in effect until the N-600 is turned off.
Table 3: Parameter Ranges
Parameter
%SpO
2 Upper
Alarm Limit
2 Lower
%SpO Alarm Limit
Pulse Rate Upper Alarm Limit
Pulse Rate Lower Alarm Limit
Alarm Silence Duration
Alarm Volume 1 to 10 7 7
Ranges/ Selections
Lower Alarm Limit plus 1 to 100%
20% to Upper Alarm Limit minus 1
Lower Alarm Limit plus 1 to 250 bpm
30 bpm to Upper Alarm Limit minus 1
Alarms 30, 60, 90, 120 seconds
Factory Adult Defaults
100% 95%
85% 80%
170 bpm 190 bpm
40 bpm 90 bpm
60 60
Factory Neonate Defaults
OxiMax N-600 31
Page 42
Using the N-600
Table 3: Parameter Ranges (Continued)
Parameter
Alarms
Data Port Baud Rate
Data Port Mode ASCII, GRAPH,
Default Display Format
Default Trend Display
Ranges/ Selections
Allow Off Ye s/ No
Off Reminder ­Ye s/ No
2400, 9600, 19200
OXINET, CLINICAL, AGILENT (HP Agilent), SPACELBS, MARQ (GE Marquette), DATEX (Datex­Ohmeda)
Pleth, Blip Pleth Pleth
Saturation, Pulse Rate, Dual, Histogram
-
Factory Adult Defaults
Ye s
Ye s
9600 9600
ASCII ASCII
Saturation Saturation
Factory Neonate Defaults
Ye s
Ye s
32
Display Contrast
Language English, Danish,
Limits Adult, Neonate Adult Neonate
Pulse Beep Vo l u m e
Response Mode
Low to high Medium Medium
English English Dutch, Finnish, French, German, Norwegian, Portuguese, Spanish, Italian, Swedish
0 to 10 4 4
Normal or Fast Normal Normal
Page 43
Table 3: Parameter Ranges (Continued)
Using the N-600
Parameter
RS-232 Level Nurse Call
Ranges/ Selections
Normally High, Normally Low
Polarity
SatSeconds Off, 10, 25, 50,
100
Sensor Event Date Format (SENSOR-R and
SpO
2,
SpO
2+Pulse Rate,
Default (default is factory default)
SENSOR-RW
Sensor
Yes, No Yes Yes Messages Enabled
Trend Display Dual, %SpO
2,
Pulse, Histogram,
Amplitude
Trend Scale 48, 36, 12,8, 4, 2,
1 hours, 30, 15
minutes, 40, 20
seconds
Factory Adult Defaults
Factory Neonate Defaults
Normally low Normally low
Off Off
Default Default
%SpO2 %SpO2
2 hours 2 hours
OxiMax N-600 33
Page 44
Using the N-600

Turning On the N-600

Before using the N-600 in a clinical setting, you must verify that the N-600 is working properly and is safe to use. Proper working condition will be verified each time the N-600 is turned on as described in the following procedure.
Caution: If any indicator or display element does not light when the pulse oximeter is turned on, do not use the pulse oximeter. Instead, contact qualified service personnel, your local Nellcor representative, or Nellcor’s Customer Services Department.
Note: Physiological conditions, medical procedures, or external
agents that may interfere with the N-600’s ability to detect and display measurements, include dysfunctional hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring agents such as nail polish, dye, or pigmented cream.
Note: The N-600 automatically starts the Power-On Self-Test
(POST), which tests the N-600 circuitry and functions.
Caution: During POST (immediately after power-up), confirm that all indicators light, all display segments turn on, and the pulse oximeter speaker sounds a one-second tone.
1. Turn on the N-600 by pressing the
2. Ensure that all of the front panel indicators illuminate.
On/standby button.
34
Page 45
Using the N-600
3. Once the display test portion of POST is complete, the N-600 software version is displayed for approximately five seconds.
Note: The software version shown above is only a sample. Check
your N-600 for the software version installed.
Software version numbers are often needed when calling Nellcor’s Customer Services Department or your local Nellcor representative for technical assistance. Write down the software version number and have it available prior to requesting technical assistance.
4. If the N-600 detects an internal problem during POST, an error tone sounds and the N-600 displays an error code (EEE) and the corresponding number. See Troubleshooting on page 123.
5. Upon successful completion of the POST, the N-600 sounds a one-second tone indicating that the N-600 has passed the test.
OxiMax N-600 35
Page 46
Using the N-600
WARNING: If you do not hear the POST pass tone, do not use the pulse oximeter.
WARNING: Ensure that the speaker is clear of any obstructions. Failure to do so could result in an inaudible alarm tone.
Note: In addition to serving as the POST pass verification, the POST

OxiMax Sensor Attached

When an OxiMax sensor is attached to the N-600, a “DATA TYPE: . . .” message is displayed briefly at the bottom of the N-600 display. For a sensor containing data, the message identifies the sensor data type. For a blank sensor, the message identifies the N-600’s current data type setting that will be used to write data to the sensor. The data type settings are SPO2 and SPO2+BPM.
pass tone also functions as an audible confirmation that the speaker is performing properly. If the speaker does not function, the alarm warning sounds cannot be heard.
36
Note: The type of data recorded is only displayed when data is
present in the OxiMax sensor.
The N-600 displays zeros in the %SpO
2 and Pulse Rate displays
while the N-600 is searching for a valid pulse. For optimal performance, allow the N-600 to search and lock onto a pulse for approximately 10 seconds in non-interference conditions.
Page 47
Using the N-600
When a valid pulse is detected, the N-600 enters the Monitoring Mode and displays patient parameters.
Look for movement of the blip bar or of the plethysmographic waveform indicating that the N-600 is displaying real-time data. Listen for the pulse beep tone. If the pulse beep tone does not sound with each pulse, it is an indication that the pulse beep volume is set to zero, the speaker is malfunctioning, or the signal is corrupted.
When an OxiMax sensor is attached to the N-600 and is applied to a patient, if the N-600 loses the pulse signal, the N-600 will display “---
& ---” (three dashes and three dashes) and remain in Pulse Search Mode for five seconds before displaying the poor signal condition screen. The poor signal condition screen is part of the N-600's Sensor Messages feature. For more information about OxiMax Sensor Messages, refer OxiMax Sensor Messages on page 56.
OxiMax N-600 37
Page 48
Using the N-600

No OxiMax Sensor Attached

Upon successful completion of the POST, the N-600 sounds a one-second tone indicating that the N-600 has passed POST.
or
38
The N-600 displays dashes (---) and the Pulse Search Indicator is not lit, indicating that the N-600 failed to detect an OxiMax sensor.
Page 49

Turning the Backlight On or Off

Note: When the backlight is off, any of the following conditions will
turn on the backlight:
pressing any of the softkeys
Using the N-600
pressing the
pressing the Alarm Silence button
•any alarm
LIGHT
With the N-600 in the normal monitoring mode, press the softkey.

Adjusting Screen Contrast

With the N-600 in the normal monitoring mode:
1. Press the
2. Press the
contrast is obtained.
Contrast button.
Adjust Up or Adjust Down button until the desired
Contrast button
LIGHT
3. Press the
OxiMax N-600 39
Contrast button.
Page 50
Using the N-600

Selecting the Pleth View

The pleth view displays the pleth waveform, %SpO2, and pulse rate data. Refer to Principles of Operation on page 149, for a description of the pleth waveform.
With the N-600 in the normal monitoring mode:
SETUP
VIEW
PLETH
1. Press the
2. Press the
3. Press the
SETUP softkey.
VIEW softkey.
PLETH softkey.
40
Page 51

Selecting the Blip View

Displays SpO2, pulse rate, blip bar, and limits in a larger format for easier viewing.
With the N-600 in the normal monitoring mode:
Using the N-600
SETUP
VIEW
BLIP
1. Press the
2. Press the
3. Press the
SETUP softkey.
VIEW softkey.
BLIP softkey.

Setting the Pulse Beep Volume

With the N-600 in the normal monitoring mode:
1. Press and hold the
Adjust Up/Adjust Down button to increase/
decrease pulse beep volume.
OxiMax N-600 41
Page 52
Using the N-600

Setting the Alarm Volume

The Alarm Volume display allows the user to adjust the volume of alarm tones.
With the N-600 in the normal monitoring mode:
1. Press the displayed and sounds on the N-600.
2. While continuing to press the hold the the volume.
Alarm Silence button until the alarm volume level is
Adjust Up/Adjust Down button to increase/decrease
Alarm Silence button, press and
42
Page 53

Setting the Date and Time

WARNING: The sensor extrapolates from the date and time provided by the N-600 when recording the sensor event record to the sensor. The accuracy of the date/time is the responsibility of the N-600. It is recommended that the N-600 user set the time/ date to the correct value before a sensor event record-enabled sensor is connected, and that this date/time not be changed while the sensor remains connected. Since a sensor with sensor event record data can be transported from one monitor to another, having discrepancies in the date/time between monitors and the sensor event record data will affect the order the sensor event record data appears. To eliminate this possible problem, all monitors within an institution should be set to the same time.
With the N-600 in the normal monitoring mode:
Using the N-600
SETUP
NEXT
CLOCK
SET
OxiMax N-600 43
1. Press the
2. Press the
3. Press the
4. Press the
SETUP softkey.
NEXT softkey.
CLOCK softkey.
SET softkey.
Page 54
Using the N-600
5. Press the SELECT softkey to select:
TIME HOURS : MINUTES : SECONDS (16:46:05)
DATE DAY - MONTH - YEAR (30-JAN-02)
EXIT
6. Use the selected value.
7. Press the
Adjust Up or Adjust Down buttons to change the
EXIT softkey.
44
Page 55

Setting Alarm Silence Duration

The Alarm Silence Duration display allows the user to adjust the alarm silence duration.
With the N-600 in the normal monitoring mode:
Using the N-600
1. Press the setting is displayed. Alarm silence durations that are available are OFF, 30, 60, 90, and 120 seconds.
2. Press and hold the button to increase the alarm silence duration setting.
3. Press and hold the button to decrease the alarm silence duration setting.
Note: Releasing the
alarm silence duration.
Alarm Silence button until the alarm silence duration
Alarm Silence button and the Adjust Up
Alarm Silence button and the Adjust Down
Adjust Up or Adjust Down button sets the
OxiMax N-600 45
Page 56
Using the N-600

Disabling Audible Alarms

Setting the alarm silence duration to OFF means that the N-600 will produce no audible alarms.
Note: The ability to set the alarm silence duration to OFF can be
enabled or disabled by qualified service personnel as described in the N-600 Service Manual. The current copy of the service manual is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_SuppProductManuals.html
WARNING: Do not silence the audible alarm function or decrease the audible alarm volume if patient safety could be compromised.
With the N-600 in the normal monitoring mode:
46
1. Press the setting is displayed.
Alarm Silence button until the alarm silence duration
Page 57
2. While pressing the Alarm Silence button, press and hold the
Adjust Up button until OFF is displayed. Release the buttons.

Selecting Standby Mode

The standby mode allows the N-600 to retain the alarm limit settings that are in effect while monitoring a patient. The N-600 must be powered by AC power to enter the standby mode.
Using the N-600
Usually the standby mode is used when a patient has to leave the N-600 for a period of time and will return to the same N-600.
To place a N-600 in the standby mode:
1. The N-600 should be monitoring a patient.
2. The N-600 alarm limits should be configured to the patient being monitored.
3. Disconnect the sensor from the N-600.
4. Press the alarms.
Alarm Silence button. This silences the audible
OxiMax N-600 47
Page 58
Using the N-600
5. Press the Alarm Silence button. This disables the alarm messages.
The N-600 is now in standby. To return to normal monitoring, connect the sensor to the N-600 and the patient.

Adult-Pediatric or Neonatal Settings

The clinician can set the N-600’s operating mode to adult-pediatric or neonatal by using the in the N-600 until the N-600 is turned off. The factory default power-on setting is for adult-pediatric patients. This default setting can be changed to neonatal by qualified service personnel using the procedures indicated in the N-600 Service Manual.
LIMITS softkey. The setting will only remain
48
Refer to Table on page 145, for neonate factory default limit settings. Refer to Table on page 145, for adult factory default limit settings.
WARNING: Each time the pulse oximeter is used, check alarm limits to ensure that they are appropriate for the patient being monitored.
Page 59

Setting Patient Adult-Pediatric/Neonatal Mode

With the N-600 in the normal monitoring mode:
Using the N-600
LIMITS
1. Press the
2. The N-600 will display the
LIMITS softkey.
Adult Limits or Neonate Limits
screen, depending on the patient setting being used.
ADULT or NEO
3. Press the
NEO or ADULT softkey to select ADULT LIMITS or
Neonate Limits as applicable for the patient being monitored.
OxiMax N-600 49
Page 60
Using the N-600

Alarm Limit Changed Indicator

Alarm limits that have been changed from the institutional or factory default settings are identified by a decimal point (.) after the displayed reading (%SpO
2 or BPM). The changed parameter is also identified
by a decimal point on the alarm limits screen.

Setting Alarm Limits

50
The Alarm Limit display allows the user to adjust the upper and lower saturation and pulse rate limits. It also allows the user to adjust the SatSeconds limit.
The Alarm Limit display is accessed by pressing the
LIMITS softkey
on the Main menu.
The Alarm Limit display includes the alarm limit table and current measured %SpO indicate whether the N-600 is in
2 and pulse rate. The title of the alarm limit table will
Adult or Neonate mode. If
SatSeconds are enabled, the Alarm Limit display also includes the
SatSeconds Indicator. Decimal points after the displayed %SpO2 or
pulse rate indicate that the respective limits have been changed from the power-on defaults.
With the N-600 in the normal monitoring mode:
Page 61
Using the N-600
LIMITS
ADULT or NEO
1. Press the LIMITS softkey. Current alarm limits are displayed.
or
2. Press the
ADULT or NEO softkey to select Adult-Pediatric or
Neonatal alarm limits screen.
SELECT
3. Press the
SELECT softkey as required to select the parameter to
be adjusted.
4. Use the
Adjust Up or Adjust Down buttons to increase or
decrease the selected limit parameter.
OxiMax N-600 51
Page 62
Using the N-600
5. Repeat steps 3, 4, and 5 as necessary to complete the alarm limits setup.
EXIT
6. To accept the changes, let the display time-out or press the softkey to exit the display and return to normal monitoring.
Note: Limit changes will only be in effect as long as the N-600
remains turned on. When the N-600 is turned off, the institutional or factory default limits will be restored into the N-600. When the N-600 is turned on, the institutional or factory default limits will be in effect. Factory or institutional defaults are selected by qualified service personnel following the procedure in the N-600 Service Manual.

Setting SatSeconds Alarm Limit

Refer to SatSeconds on page 141, for a description of the SatSeconds function.
With the N-600 in the normal monitoring mode:
LIMITS
1. Press the
LIMITS softkey. Current alarm limits are displayed.
EXIT
52
SELECT
2. Press the
SELECT softkey twice to select %SpO2 SAT-S.
Page 63
Using the N-600
3. Use the Adjust Up or Adjust Down buttons to select the limit. The choices are 10, 25, 50, or 100 seconds or OFF.
EXIT
4. Press the
EXIT softkey to save your choice.

Setting the N-600 Response Mode

The purpose of the response mode is to set the response time of the N-600 algorithm calculation of the SpO affect the N-600 algorithm’s calculation of pulse rate). The trending interval (two- or four-seconds) is updated automatically by the N-600 to roughly correspond with the SpO
The response mode programs the N-600 algorithm to display N-600 trend information at two-second intervals (Fast Mode) or four-second intervals (Normal Mode).
The response mode display screen includes the current SpO mode setting and the current measured %SpO in the fast mode, the screen displays the fast mode symbol.
With the N-600 in the normal monitoring mode:
2 (the response mode does not
2 calculation response time.
2 response
2 and pulse rate. When
SETUP
NEXT
NEXT
NEXT
MODE
OxiMax N-600 53
1. Press the
2. Press the
3. Press the
4. Press the
5. Press the
SETUP softkey.
NEXT softkey.
NEXT softkey.
NEXT softkey.
MODE softkey.
Page 64
Using the N-600
Note: When the N-600 is in the fast response mode the N-600 may
produce more SpO
2 and pulse rate alarms than the user is
accustomed to seeing.
EXIT
6. Use the response mode.
7. Press the
Adjust Up or Adjust Down buttons to select the desired
EXIT softkey.
54
Page 65

Selecting the Display Language

The N-600 can be programmed to display the information in various languages. The languages available are ENGLISH, DANSK (Danish), FRANCAIS (French), DEUTSCH (German), ITALIANO (Italian), ESPAÑOL (Spanish), NEDERLANDS (Dutch), NORSK (Norwegian), PORTUG (Portuguese), SUOMI (Finnish) and SVERIGE (Swedish).
With the N-600 in the normal monitoring mode:
Using the N-600
SETUP
NEXT
LANG
1. Press the
2. Press the
3. Press the
4. Use the language.
SETUP softkey.
NEXT softkey.
LANG softkey.
Adjust Up or Adjust Down buttons to select the desired
OxiMax N-600 55
Page 66
Using the N-600
EXIT
5. Press the EXIT softkey.
Note: The selected language will be displayed until the N-600 is
turned off. The selected language can be set as a default by qualified service personnel following the procedures in the N-600 Service Manual.

OxiMax Sensor Messages

OxiMax sensor messages consist of sensor adjust condition messages and sensor adjust messages which, when enabled, are displayed when the N-600 is not able to display saturation. When OxiMax sensor messages are displayed, it is an indication that the OxiMax sensor is functioning correctly, but the site to which the OxiMax sensor applies or the application method is not optimal for calculating %SpO Condition messages are followed by action messages. Up to three condition messages may be displayed on the “
CONDITION
condition display may be dismissed by using the exited, the OxiMax sensor message screen will not return until a new condition occurs.
2.
POOR SIGNAL
” display in priority order, highest on top. The
EXIT softkey. Once
56
If the
HELP softkey is pressed from the Condition message display,
the action messages are displayed. Action messages are linked to the sensor type; action messages will be displayed for the type of OxiMax sensor connected to the N-600. Up to five action messages may be displayed. Multiple screens may be required to display all of the messages. When multiple screens are required, navigation between screens can be accomplished through the
NEXT, BACK, and EXIT
softkeys.
Page 67
OxiMax sensor messages may be disabled. Refer to Troubleshooting on page 123 for selecting the OxiMax Sensor Messages, Enable/ Disable function.

OxiMax Sensor Adjust Condition Messages

Condition 1 — SENSOR OFF?
Condition 2 — SMALL PULSES
Condition 3 — WEAK SIGNAL
Using the N-600
Condition 4 — INTERFERENCE
Condition 5 — EXCESS INFRARED LIGHT
Condition 6 — INTERFERENCE
Condition 7 — HIGH PULSE AMPLITUDE
OxiMax N-600 57
Page 68
Using the N-600

OxiMax Sensor Adjust Messages

Message 1 — ALTERNATE SITE?
Message 2 — COVER SENSOR SITE?
Message 3 — EAR/FOREHEAD SENSOR?
Message 4 — NASAL/EAR SENSOR?
Message 5 — OxiMax ADHESIVE SENSOR
Message 6 — SECURE CABLE
Message 7 — HEADBAND
Message 8 — WARM SITE
Message 9 — BANDAGE ASSEMBLY
58
Message 10 — NAIL POLISH
Message 11 — SENSOR TOO TIGHT?
Message 12 — REPOSITION SENSOR
Message 13 — ISOLATE INTERFERENCE SOURCE
Message 14 — CLEAN SENSOR SITE
Page 69

N-600 Trend

N-600 Trend Data

The trend displays allow the user to view trend data. Two types of trend data can be viewed:
N-600 trend data can be viewed anytime patient trend is stored in the N-600. N-600 trend displays are accessed by pressing the softkey on the main menu and selecting the The N-600 trend submenu allows you to choose which trend data are displayed:
N-600 trend data stored in the N-600
Patient event data stored in the OxiMax sensor (single-patient-use OxiMax sensors only) and can be used with the sensor event record feature.
TREND
MONITR softkey option.
Saturation and pulse rate (Dual)
Saturation
•Pulse rate
Pulse amplitude
Histogram
The N-600 can graphically display trend data for SpO both. Trend data is stored at two- or four-second intervals. When the
2, pulse rate, or
TREND softkey is pressed, “READING TRENDS . . .” is
displayed at the bottom of the N-600 screen, indicating that the N-600 is formatting the trend data to be displayed.
OxiMax N-600 59
Page 70
N-600 Trend
The N-600 stores up to 48 hours of four-second trend data or 24 hours of two-second trend data. The amount of trend data displayed on the screen is determined by using the
ZOOM softkey. The settings
available are 20 and 40 seconds, 15 or 30 minutes, and 1, 2, 4, 8, 12, 24, 36, or 48 hours. All trend data are displayed in a graphical format except the 20- and 40-second trend displays, which are shown in tabular format.
The trend display is scrolled, that is, the data displayed can be moved throughout the 48 hours of trend data. Selecting the one-hour trend display allows you to view one hour of trend information. By using the scrolling feature, any one hour of trend data can be viewed over the 48 hours of trend information. The the display to the left and the
Adjust Up button scrolls the display to
Adjust Down button scrolls
the right.
When the data are displayed, the most recent readings are on the right side of the graph. The numbers below %SpO
2 indicate the highest and
lowest parameter values at the cursor position (vertical dotted line on the display). See Table 4 on page 64.
60
.
Trend data information may be retrieved through the N-600 data port or cleared using options available in a display menu.
Caution: N-600 trend data will be lost if the main battery fails or is removed.
Page 71

Trend Data Operation

While the N-600 is turned on, it stores the N-600 %SpO2 and pulse rate readings in memory every two seconds (fast response mode) or four seconds (normal response mode), regardless of whether the N-600 is monitoring a patient or not. The N-600 can store up to 48 hours of trend data in the normal response mode or 24 hours of trend data in the fast response mode. The 48/24 hours of stored trend data are available for downloading to a printer or a portable computer. Up to 50 alarm limit changes can be stored in the trend data. If more than 50 alarm limit changes occur during the 48/24 hours of trend data collection, then the additional alarm limit changes take the space reserved for trend data.
Caution: Change alarm limits only as needed.
Trend memory always contains the most recent 48 or 24 hours of data, with newly collected data overwriting the oldest data on a preemptive basis. The N-600 continues to record data points as long as the N-600 is turned on. Blank data points are collected when the N-600 is not connected to a patient. Blank data overwrites older patient data if the memory becomes full. Therefore, to retain old patient data, it is important to turn off your N-600 when you are not monitoring a patient. Download the trend memory before it fills up and overwrites the old data with new or blank data. Figure 3 illustrates when the new trend data starts to overwrite the old trend data.
N-600 Trend
Figure 3: Trend Overwriting
OxiMax N-600 61
Page 72
N-600 Trend

Selecting the Trend Data Display Scale

The trend scale is the amount of trend data displayed on the screen.
With the N-600 in the normal monitoring mode:
TREND
MONITR
VIEW
ZOOM
1. Press the
2. Press the
3. Press the
TREND softkey.
MONITR softkey.
VIEW softkey.
4. Press any of the trend softkeys ( select
HIST (histogram) or AMP (amplitude), press the NEXT
softkey and then the
5. Press the
ZOOM softkey. The Zoom menu is displayed.
HIST or AMP softkeys.
DUAL, SPO2, or PULSE). To
62
TIME
Pressing the
TIME softkey cycles the displayed trend time scale
through 48 hours, 36 hours, 12 hours, 8 hours, 4 hours, 2 hours, 1 hours, 30 minutes, 15 minutes, 40 seconds and 20 seconds.
Page 73
N-600 Trend
Note: The 20-second and 40-second trend displays are in tabular
format. The below display starts out in the normal response mode (left side of the display) and switches to the fast response mode.
SCALE
AUTO
BACK
Pressing the
SCALE softkey cycles the displayed trend amplitude
scale through ±5 points, ±10 points, ±15 points, ±20 points, ±25 points, ±30 points, ±35 points, ±40 points and ±50 points above and below the data point under the cursor. The saturation graphical N-600 trend display vertical scale default setting is from 10 to 100 if there is no data under the cursor. The pulse rate graphical N-600 trend display vertical scale is from 5 to 250 if there is no data under the cursor.
Pressing the
AUTO softkey presets the amplitude of the graphed
trend data. The maximum trend data point is rounded up to the nearest multiple of 10, this value is the top of the graph display. The minimum trend data point is rounded down to the next multiple of 10. Then 10 is subtracted from the rounded down number, this value is the bottom of the trend graph.
Pressing the
BACK softkey returns the N-600 to the Monitor menu.
OxiMax N-600 63
Page 74
N-600 Trend

Reading the Trend Data Display

Table 4 identifies the components of the trend data display.
Table 4: Reading Trend Display
Item Description
64
1 Amount of trend data displayed on the screen. Settings available
are 20 and 40 seconds, 15 and 30 minutes, 1, 2, 4, 8, 12, 24, 36, and 48 hours.
2 Date represented by the cursor (item 5).
3 No trend data recorded during this time.
4 Time represented by the cursor (item 5).
5
6 Highest and lowest reading at the cursor position.
7 Trend data that is being displayed (%SPO2, BPM, or PAU
- can be moved left or right using the Adjust Up (right)
Cursor
Adjust Down (left) buttons.
or
[pulse amplitude units]).
Page 75

Dual Trend Data Display

The dual trend data display displays both oxygen saturation (%SpO2) levels and pulse rate (bpm) trend data.
With the N-600 in the normal monitoring mode:
N-600 Trend
TREND
MONITR
VIEW
DUAL
1. Press the
2. Press the
3. Press the
4. Press the is displayed.
TREND softkey.
MONITR softkey.
VIEW softkey.
DUAL softkey. The dual trend (%SpO2 and Pulse Rate)

SpO2 Trend Display

With the N-600 in the normal monitoring mode:
TREND
MONITR
1. Press the
2. Press the
OxiMax N-600 65
TREND softkey.
MONITR softkey.
Page 76
N-600 Trend
VIEW
SPO2
3. Press the VIEW softkey.
4. Press the

Pulse Rate Trend Display

With the N-600 in the normal monitoring mode:
TREND
1. Press the TREND softkey.
SPO2 softkey. SpO2 trend data is displayed.
66
MONITR
VIEW
PULSE
2. Press the
3. Press the
4. Press the
MONITR softkey.
VIEW softkey.
PULSE softkey. The pulse trend data is displayed.
Page 77

Histogram Trend Data Display

The histogram displays trend data for the percent of oxygen blood saturation (SpO the trend data stored over the period of time indicated on the display. Refer to Selecting the Trend Data Display Scale on page 62, to set up the desired trend data scale.
Pulse amplitude cannot be displayed on the histogram display.
With the N-600 in the normal monitoring mode:
2) and pulse rate (bpm). The data displayed represents
N-600 Trend
TREND
MONTIR
VIEW
NEXT
HIST
1. Press the
2. Press the
3. Press the
4. Press the
5. Press the
TREND softkey.
MONITR softkey.
VIEW softkey.
NEXT softkey.
HIST softkey. The Histogram trend data is displayed.
OxiMax N-600 67
Page 78
N-600 Trend

Pulse Amplitude Trend Data Display

The pulse amplitude trend data display shows the amplitude of the patient’s pulse rate over the period of time indicated on the display. Refer to Selecting the Trend Data Display Scale on page 62, to setup the desired trend data scale.
With the N-600 in the normal monitoring mode:
TREND
MONITR
VIEW
NEXT
AMP
1. Press the
2. Press the
3. Press the
4. Press the
5. Press the data is displayed.
TREND softkey.
MONITR softkey.
VIEW softkey.
NEXT softkey.
AMP softkey. The pulse amplitude units (PAU) trend
68
The PAU reading (12 : 20) indicates the pulse amplitude units (upper and lower) at the cursor position (dashed line). The cursor is moved right or left using the
Adjust Up (right) and Adjust Down (left)
buttons.
Page 79

Clearing Trend Information

With the N-600 in the normal monitoring mode:
N-600 Trend
TREND
MONITR
NEXT
DELETE
YES
1. Press the TREND softkey.
2. Press the
3. Press the
4. Press the
Note: Press the
MONITR softkey.
NEXT softkey.
DELETE softkey.
NO softkey and then the EXIT softkey to leave this
function without deleting trend data.
5. Press the
YES softkey.
All the trend data is cleared and the N-600 sounds three beeps.
OxiMax N-600 69
Page 80
Blank Page
Page 81

OxiMax Sensor Event Record

WARNING: The sensor extrapolates from the date and time provided by the N-600 when recording the sensor event record to the sensor. The accuracy of the date/time is the responsibility of the N-600. It is recommended that the N-600 user set the time/ date to the correct value before a sensor event record-enabled sensor is connected, and that this date/time not be changed while the sensor remains connected. Since a sensor with sensor event record data can be transported from one monitor to another, having discrepancies in the date/time between monitors and the sensor event record data will affect the order the sensor event record data appears. To eliminate this possible problem, all monitors within an institution should be set to the same time.
The adhesive OxiMax sensors are capable of storing patient event data. A sensor event record allows alarm event history to travel with the patient on the sensor’s memory chip for quick assessment at every point of care where OxiMax monitors are used.
Patient (event) data is stored on the memory chip of adhesive OxiMax sensors (single-patient-use OxiMax sensors only). The event data is stored (recorded) with the limit/threshold settings that were active at the time of the event on the recording monitor. These events can be viewed on the next OxiMax sensor monitor when the patient moves to a new point of care.
An event occurs when the %SpO lower alarm limit for at least 15 seconds. The first OxiMax sensor event record event will be stored in the OxiMax sensor after the OxiMax sensor has been attached to a patient for five minutes and every five minutes thereafter. The maximum number of events that can be stored in an OxiMax sensor is 100.
OxiMax N-600 71
2 value exceeds either the upper or
Page 82
OxiMax Sensor Event Record
Event records can only be viewed after an OxiMax sensor containing patient data (event records) has been connected to an OxiMax monitor. Event records are designed to view patient events from prior areas of care or transport (history) while monitor trend should be used to view data or events from a patient currently being monitored. The monitor’s SENSOR EVENT RECORD Indicator will light when an OxiMax sensor containing event data is connected to the OxiMax monitor.
Patient event data is accessed by pressing the main menu and selecting the record can be viewed in graphical form ( table (
TABLE).
Note: Once the OxiMax sensor event record type is set up in the
OxiMax sensor and event data is stored in the OxiMax sensor,
the OxiMax sensor event record type cannot be reset. The monitor’s type set up can be changed at any time.
Recording and viewing of OxiMax sensor event record is only available on OxiMax compatable monitors. The OxiMax sensors may function on older technology monitors but the OxiMax sensor event record feature is not available.
Refer to the N-600 Service Manual for the procedure to disable the storage of sensor event record on an OxiMax sensor.

Setting In-Sensor Data Type

The In-Sensor Data Type display allows the user to set the type of trend data to be recorded in an OxiMax sensor. OxiMax sensors can be set to record either SpO
TREND softkey on the
SENSOR softkey option. Sensor event
GRAPH) or in a summary
2 or SpO2+BPM.
72
Note:The OxiMax sensor data type can only be set when an OxiMax
sensor is not connected to the N-600.
With the N-600 turned on and no cable attached to the SpO
2 OxiMax
sensor port:
Page 83
OxiMax Sensor Event Record
R
SETUP
SENSO
DATA
1. Press the SETUP softkey.
2. Press the
3. Press the
SENSOR softkey.
DATA softkey.
Note: OxiMax sensor data type settings are displayed on the N-600 as
shown in the figure below (in-sensor data type). If no sensor is connected, both sensor types and the full set of options for each are displayed. If a sensor is connected, only the sensor data type for that sensor is displayed.
Note: The SENSOR-R feature supports all the current OxiMax
sensors. The SENSOR-RW feature is only applicable to
sensors with a read/write chip installed.
SELECT
OxiMax
4. Use the SELECT softkey to toggle between SENSOR-R and
SENSOR-RW.
OxiMax N-600 73
Page 84
OxiMax Sensor Event Record
5. Use the Adjust Up or Adjust Down button to select the OxiMax sensor data type. are:
SENSOR-R and SENSOR-RW selections
EXIT
6. Press the

OxiMax Sensor Type

When an OxiMax sensor is connected to the N-600, a “SENSOR TYPE: …” message is displayed for four to six seconds at the bottom of the display. The message identifies the type (model) of OxiMax sensor connected to the N-600. Type is used in the determination of action messages in the OxiMax sensor message(s) function. This display is the first message displayed when an OxiMax sensor is connected to the N-600.
•SpO
2
•SpO2+BPM
•DEFAULT
EXIT softkey to set the OxiMax sensor type.
74
Page 85

OxiMax Sensor Data Type

When an OxiMax sensor with no previously recorded patient data is connected to the N-600, a “DATA TYPE: . . .” message is displayed briefly at the bottom of the display, this message is displayed after the OxiMax sensor type message. The message identifies the N-600's current data type setting that will be used to write data to the OxiMax sensor. The data type setting options are EVENT/SPO SPO
2+BPM.
The user can change the setting by referring to Setting In-Sensor Data Type on page 72. The OxiMax sensor event record type must be set prior to connecting the OxiMax sensor to the N-600.
OxiMax Sensor Event Record
2 and EVENT/

OxiMax Sensor Event Record Data Available

When an OxiMax sensor containing data (single-patient-use OxiMax sensors only) is connected to the N-600, the S Indicator on the N-600 front panel blinks at a medium priority flash
TREND
rate to indicate that the OxiMax sensor attached to the N-600 contains patient event data. The indicator blinks for approximately 60 seconds
SENSOR
OxiMax N-600 75
or until the OxiMax sensor is disconnected or until the sensor trend data is displayed by pressing
TREND, then SENSOR.
ensor Event Record
Page 86
OxiMax Sensor Event Record
A corresponding “DATA IN SENSOR” message is also displayed at the bottom of the display. After four to six seconds, if all the data has been read from the OxiMax sensor, the message is replaced with the main menu.
If data is still being read from the OxiMax sensor, after four to six seconds, the
READING TRENDS message with an ABORT option.
DATA IN SENSOR message is replaced with a
76
Selecting the
ABORT softkey stops the recording of additional data
in the OxiMax sensor and accessing or viewing the data that is in the OxiMax sensor.
Sensor event record can be viewed by accessing the
SENSOR
menu.
TREND/
The Sensor Event Record Indicator comes on steady when OxiMax sensor memory is full and stays on until the OxiMax sensor is disconnected.
Page 87

OxiMax Sensor Event Record Not Available

If the user selects the TREND/SENSOR option when a connected OxiMax sensor (single-patient-use OxiMax sensors only) does not contain data, because no events were recorded to the OxiMax sensor memory chip in the prior monitoring situation, a “NO DATA” message is displayed on the default trend or event graph.
A sample event display in which no data are available is shown below. The message will be cleared when the graph or summary is exited.
OxiMax Sensor Event Record

OxiMax Sensor Event Record Graphical Data

Graphical representations of patient event history is only available on single-patient-use OxiMax sensors. Graphed data points are the minimum or maximum %SpO throughout the duration of an event (%SpO alarm threshold for at least 15 seconds) and continuing every 30 seconds until the actual %SpO threshold.
The duration of an event is determined by the number of data points in the event. Each data point is stored at 30-second intervals.
Events end for one of four reasons:
The %SpO
OxiMax N-600 77
2 returns to or above the alarm limit
2 value for each 30-second interval
2 continuously below
2 value equals or exceeds the alarm
Page 88
OxiMax Sensor Event Record
Loss of pulse
•The OxiMax sensor is disconnected
•The OxiMax sensor is off the patient
The graph title shows the data type (
EVENT GRAPH) in the upper
left corner. The number of the displayed event and the total number of events recorded in the OxiMax sensor are shown to the right of the title (example, 2/2). The date and time of the displayed event are shown in the upper center and upper right corner.
The type of data displayed in the graph is indicated to the left of the vertical axis (%SpO
2). Below this is the range of values (min/max)
during the event. The duration of the event is shown below the range value. The vertical axis of the graph is labeled to show the magnitude scale of the graphed data. The horizontal axis is not labeled but automatically scales to accommodate the number of 30-second intervals during the event. The alarm threshold (lower than %SpO
2
alarm limit) is represented by a horizontal dotted line across the graph. The first data point is always the alarm threshold.
Events are displayed one at a time, one per graph. Graphs are displayed in chronological sequence with the most recent event shown first when accessing the graphical OxiMax sensor event display. The user can move between events by using the two left-most softkeys which are labeled with left- and right-facing arrow icons, respectively. At the beginning of an event sequence, event 1of 2 events, the left-arrow soft key is blank; at the end of a sequence, event 2 of 2 events, the right-arrow soft key is blank.
78
The
Adjust Up and Adjust Down buttons on the N-600 front panel
can also be used to move through events.
Page 89
OxiMax Sensor Event Record
R
The PRINT softkey allows the user to print the displayed event graph. The
BACK softkey takes the user back to the previous
TREND/SENSOR sub-menu level.

Viewing and Printing OxiMax Sensor Event History Data

With the N-600 in the normal monitoring mode, you must connect a printer, capable of printing graphs, to the N-600 data port connector to print OxiMax sensor event history data.
TREND
SENSO
GRAPH
The N-600 protocol must be set to
GRAPH to print the in-sensor
event history data. Refer to Printing N-600 Trend Information on page 85. To view and print in-sensor event history data:
1. Connect an OxiMax sensor containing patient data to the N-600.
2. Press the
3. Press the
4. Press the
TREND softkey.
SENSOR softkey.
GRAPH softkey.
Note: Use the left and right arrow softkeys to scroll through the pages
of the event graph.
OxiMax N-600 79
Page 90
OxiMax Sensor Event Record
PRINT
EXIT
5. Press the PRINT softkey to print the displayed screen.
6. Press the
A sequence of %SpO
EXIT softkey.
2 + BPM (saturation plus pulse rate)
“dual-view” event graphs are shown below. The dual-view graph is the same as a single graphical event history graph except the graphs are compressed horizontally to allow both %SpO
2 and pulse rate
graphs to be shown for the same event.
80
Page 91

OxiMax Sensor Tabular Event Data

The OxiMax sensor tabular event data is a listing of all events recorded on the OxiMax sensor’s memory chip.
OxiMax Sensor Event Record
The table title shows in the upper left corner. Below the table title is a six-column table with left-to-right column headings of event number (#), date ( %SPO
DATE), event start time (START), event duration (DUR),
2 minimum and maximum values during the event (%SPO2),
and pulse rate minimum and maximum values during the event (BPM).
Event data are listed in chronological order with the most recent event shown first, at the top of the list, when the tabular Event Summary display is first accessed. Four events can be displayed simultaneously; the table must be scrolled to view additional events. The user can move to the next screen view of the table, the next three events (the previously displayed bottom or top event is retained as the fourth event for context when a table is scrolled), using the two left-most softkeys which are labeled with left- and right-facing arrow
OxiMax N-600 81
Page 92
OxiMax Sensor Event Record
R
icons, respectively. At the beginning of an event sequence, Event 1 of 5 events, the left-arrow soft key is blank; at the end of a sequence, Event 5 of 5 events, the right-arrow soft key is blank, indicating you have reached the beginning or end of the table.
The
Adjust Up and Adjust Down buttons on the N-600 front panel
can be used to move through the Event Summary table line by line.
PRINT softkey allows the user to print the displayed event
The graph.
The
BACK softkey takes the user back to the previous TREND/
SENSOR
sub-menu level.

Viewing and Printing In-Sensor Tabular Event History Data

The N-600 should be in the normal monitoring mode.
To view and print in-sensor tabular event history data:
82
TREND
SENSO
TAB LE
1. Press the
2. Press the
3. Press the
TREND softkey.
SENSOR softkey.
TABLE softkey.
Page 93
OxiMax Sensor Event Record
PRINT
BACK
4. Press the PRINT softkey to print the data.
5. Press the
BACK softkey.
OxiMax N-600 83
Page 94
Blank Page
Page 95

Printing

Printing N-600 Trend Information

Trend information (N-600 and in-sensor event history) may be sent to a personal computer or to a serial printer.
Note: The protocol settings must be set to ASCII MODE for printing
text data or
GRAPH MODE for printing graphical data.
1 — Data Port Connector
With the N-600 in the normal monitoring mode:
1. Connect the serial printer to the N-600’s data port connector (1), using Nellcor printer cable part number 036341.
2. Turn on the printer.
SETUP
OxiMax N-600 85
3. Press the
SETUP softkey.
Page 96
Printing
R
NEXT
NEXT
COMM
4. Press the NEXT softkey.
5. Press the
6. Press the
7. Set the
Up
NEXT softkey.
COMM softkey.
BAUD rate to the appropriate number using the Adjust
button.
86
SELECT
EXIT
TREND
MONITR
SENSO
8. Press the
9. Set the graph printing using the
10. Press the
11. Press the
12. Press the
SELECT softkey to select PROTOCOL.
PROTOCOL to ASCII for text printing or GRAPH for
Adjust Up button.
EXIT softkey.
TREND softkey.
MONITR softkey for N-600 trend printing or press the
SENSOR softkey for in-sensor event history data printing.
Page 97
Printing
NEXT
PRINT
13. Press the NEXT softkey.
14. Press the
PRINT softkey.
ASCII printout:
GRAPH printout:
OxiMax N-600 87
Page 98
Printing

N-600 Trend Data in ASCII Mode

Refer to Printing N-600 Trend Information on page 85 for the procedure to print trend information.
The format of data displayed when a trend printout is shown in Figure 4. “
Readings are displayed in 2- or 4-second intervals depending on the response mode selected. The values on each row are an average of the response mode selected period.
At the end of the printout an “Output Complete” line indicates that the transmission was successful. If the “Output Complete” line is not present, a corruption of the data may have been detected and the data should be ignored.
TREND” is displayed in the top row.
88
Figure 4: ASCII Mode Printout
Once a trend printout has begun, it cannot be aborted without turning off the N-600 or the printer.
Page 99

Trend Data in Graph Mode

Refer to Printing N-600 Trend Information on page 85 for the procedure to print trend information. See Figure 5.
The graph mode disables all printout functions except trend data. Graph mode trend printouts are formatted for a Seiko DPU-414 and Okidata 320 serial printer.
Printing
Figure 5: Graph Mode Printout

Real-Time Display/Printout Format

Real-time data is continuously sent to the data port on the back of the N-600. Patient data can be obtained through the data port by connecting the N-600 data port to a PC or serial printer. When a real-time printout or display is being transmitted to a printer or PC, a new line of data is displayed every two seconds. Column headings are displayed or printed after every 25 lines, or if one of the values in the column heading changes. Readings are displayed at four-second intervals if the SpO two-second intervals when the SpO
Data cannot be obtained if the N-600 is operating on battery power.
Note: If the data output stops transmitting, turn the power off and
back on again or, if the N-600 is connected to a PC, send an XON (Ctrl-q) command to reset the N-600.
OxiMax N-600 89
2 response mode is set to normal and at
2 response mode is set to fast.
Page 100
Printing
An example of a real-time output is shown in Figure 6.
90
Figure 6: Real-Time Printout
Loading...