Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor,
OxiCliq, Dura-Y, Max-Fast, SoftCare, and OxiMax are trademarks of Nellcor Puritan Bennett Inc.
Cet appareil ISM est conforme à la norme NMB-001 Canada.
To obtain information about a warranty, if any, contact Nellcor’s Customer Services Department,
or your local representative.
Purchase of this instrument confers no express or implied license under any Nellcor Puritan
Bennett patent to use the instrument with any sensor that is not manufactured or licensed by
Nellcor Puritan Bennett.
Warnings are identified by the WARNING symbol shown above.
Warnings alert the user to potential serious outcomes (death, injury, or
adverse events) to the patient or user.
WARNING: The sensor extrapolates from the date and time
provided by the N-600 when recording the sensor event record to
the sensor. The accuracy of the date/time is the responsibility of
the N-600. It is recommended that the N-600 user set the time/
date to the correct value before a sensor event record-enabled
sensor is connected, and that this date/time not be changed while
the sensor remains connected. Since a sensor with sensor event
record data can be transported from one monitor to another,
having discrepancies in the date/time between monitors and the
sensor event record data will affect the order the sensor event
record data appears. To eliminate this possible problem, all
monitors within an institution should be set to the same time.
WARNING: Explosion hazard. Do not use the N-600 pulse
oximeter in the presence of flammable anesthetics or gases.
WARNING: Chemicals from a broken LCD display panel are
toxic when ingested. Use caution when handling a pulse oximeter
with a broken display panel.
OxiMax N-6001
Page 12
Safety Information
WARNING: Pulse oximetry readings and pulse signals can be
affected by certain environmental conditions, OxiMax sensor
application errors, and certain patient conditions. See the
appropriate sections of this manual for specific safety
information.
WARNING: The use of accessories, sensors, and cables other
than those specified may result in increased emission and/or
decreased immunity and inaccurate readings of the N-600 pulse
oximeter.
WARNING: Failure to cover the OxiMax sensor site with opaque
material in high ambient light conditions may result in inaccurate
measurements.
Cautions
2
Cautions are identified by the CAUTION symbol shown above.
Cautions alert the user to exercise care necessary for the safe and
effective use of the N-600 pulse oximeter.
Page 13
Safety Information
Caution: When connecting the N-600 to any instrument, verify
proper operation before clinical use. Both the N-600 and the
instrument connected to it must be connected to a grounded
outlet. Accessory equipment connected to the pulse oximeter’s
data interface must be certified according to IEC Standard 950
for data-processing equipment or IEC Standard 60601-1 for
electromedical equipment. All combinations of equipment must
be in compliance with IEC Standard 60601-1-1 systems
requirements. Anyone who connects additional equipment to the
signal input port or signal output port (N-600 data port
connector) configures a medical system and is therefore
responsible for ensuring that the system complies with the
requirements of system standard IEC Standard 60601-1-1 and
the electromagnetic compatibility system standard IEC Standard
60601-1-2. The N-600 accuracy may degrade if it is connected to
secondary I/O devices when the instrument is not connected to
earth reference.
Caution: Federal law (U.S.A.) restricts this device to sale by or on
the order of a physician.
Caution: Dispose of battery in accordance with local
requirements and regulations.
Notes
Notes are identified by the Note symbol shown above.
Notes provide additional useful information.
OxiMax N-6003
Page 14
Blank Page
Page 15
Introduction
WARNING: The N-600 is intended only as an adjunct in patient
assessment. It must be used in conjunction with clinical signs and
symptoms.
Intended Use for the N-600
The OxiMax® N-600 pulse oximeter is indicated for the continuous
non-invasive monitoring of functional oxygen saturation of arterial
hemoglobin (SpO
with neonatal, pediatric, and adult patients who are well or poorly
perfused, in hospitals, hospital-type facilities, intra-hospital transport,
and home environments. For prescription use only.
Note: Hospital use typically covers such areas as general care floors,
operating rooms, special procedure areas, intensive and critical
care areas, within the hospital plus hospital-type facilities.
Hospital-type facilities include physician office based
facilities, sleep labs, skilled nursing facilities, surgicenters, and
sub-acute centers.
2) and pulse rate. The N-600 is intended for use
Intra-hospital transport includes transport of a patient within
the hospital or hospital-type facility.
Home Care use is defined as managed/used by a lay person
(parent or other similar non-critical caregiver) in the home
environment.
Use with any particular patient requires the selection of an
appropriate oxygen OxiMax sensors as described in this
manual.
OxiMax N-6005
Page 16
Introduction
How to Use this Manual
All users should read this manual thoroughly. More experienced users
of the N-600 will be able to go to the topics for the information they
require.
The current copy of this manual is available on the internet at:
Alarm Silence Button, page 10
Adjust Up Button,page 10
TM
Indicator, page 13
12.
Contrast Button, page 10
14.
page 13
17. Menu Bar, page 10
18.
Data In Sensor Indicator, page 13
19.
Interference Indicator, page 13
21. Speaker
Figure 1: Front Panel Buttons and Symbols
OxiMax N-6007
Page 18
Description of Controls, Indicators, and Symbols
Identification of Rear Panel Components
1. Equipotential Terminal (Ground) 4. Fuse Holder
2. AC Power Connector, page 195. Supply Voltage Selector Switch, page 17
3. Data Port Connector, page 97
Figure 2: Rear Panel Components
N-600 Symbols
8
The symbols that are located on the rear panel of the N-600 are as
follows:
See Instructions for Use
Fuse Replacement
Page 19
Equipotential Terminal (ground)
Date of Manufacture
Data Interface
There is one symbol located on the front panel of the N-600.
Type BF Applied Part - Not defibrillator proof.
Description of Controls
Description of Controls, Indicators, and Symbols
Note: A button press, except the On/Standby button, should result
in either a valid or an invalid key tone (refer to Table 1 on
page 14). If the key pressed fails to emit a tone, contact
qualified service personnel.
The
On/Standby button. Used to turn the N-600 on or off.
OxiMax N-6009
Page 20
Description of Controls, Indicators, and Symbols
The Alarm Silence button. Used to silence current alarms for the
alarm silence duration period. When an alarm has been silenced,
pressing the button again reactivates, or “unsilences” the alarm. It is
also used to view and adjust alarm silence duration and alarm volume.
The
Alarm Silence button clears “SENSOR OFF,” “LOW
BATTERY,” and “SENSOR DISCONNECT” messages from the
display.
The
Adjust Up button. Used to increase variable parameters of the
N-600.
The
Adjust Down button. Used to decrease variable parameters in
the N-600.
Contrast button. Used in conjunction with the Adjust Up and
The
Adjust Down buttons to lighten or darken the display screen.
The softkey buttons have multiple uses depending on the legend
displayed above the button.
Description of Displays and Indicators
The type of display is user selectable. Refer to Selecting the Pleth
View on page 40.
10
Page 21
Description of Controls, Indicators, and Symbols
The pleth display includes a “wiper bar” plethysmographic
waveform, menu bar, and current measured %SpO
SatSeconds are enabled, the pleth display includes the
Indicator and SatSeconds setting. A decimal point after the %SpO
2 and pulse rate. If
SatSeconds
2 or
pulse rate indicate that the respective limits have been changed from
the power on defaults (N-600 Trend Data on page 59).
The blip display includes a pulse amplitude blip bar, current measured
%SpO
2 and pulse rate, and current upper and lower %SpO2 and pulse
rate limits. If SatSeconds are enabled, the blip display includes the
SatSeconds Indicator and SatSeconds setting. Decimal points after
the %SpO
2 or pulse rate indicate that the respective limits have been
changed from the power-on defaults.
There are various matrices within the N-600 algorithm. Some of
these, are used to assess the severity of conditions presented to the
N-600 in measuring SpO
2 and pulse rate on a patient. These
individual matrices or combinations of these matrices are used to
drive the LED indicators on the N-600 front panel.
The advanced signal processing in the N-600 algorithm automatically
extends the amount of data required for measuring SpO
2 and pulse
rate depending on the measurement conditions. During normal
measurement conditions the averaging time is six seconds to seven
seconds (approximately three seconds in Fast Mode). During
conditions such as those caused by low perfusion, interference (e.g.,
external interference like ambient light), or a combination of the two,
the N-600 algorithm automatically extends the amount of data
required beyond the minimum as set by the response mode. See
Setting the N-600 Response Mode on page 53. If the resulting
dynamic averaging time exceeds 20 seconds, the
indicator is lit solid and SpO
OxiMax N-60011
2 and Pulse Rate will continue to be
Pulse Search
Page 22
Description of Controls, Indicators, and Symbols
updated every second. As these conditions extend, the amount of data
required continues to increase. If the dynamic averaging time reaches
40 seconds, the pulse search indicator begins flashing, the SpO
pulse rate displays flash zeros indicating a loss-of-pulse condition.
WARNING: Failure to cover the OxiMax sensor site with opaque
material in high ambient light conditions may result in inaccurate
measurements.
2 and
The %SpO
2 Display. Shows the hemoglobin oxygen saturation level.
The display value flashes zeros during loss-of-pulse alarms and
flashes the SpO
2 value when the SpO2 is outside the alarm limits.
During Pulse Search, the N-600 continues to update the display. If
alarm limits have been changed from their power-on defaults, a
decimal point (.) is displayed after the SpO
The
Pulse Amplitude Indicator (blip bar). Indicates pulse beat and
2 value (81.).
shows the relative pulse amplitude. As the detected pulse becomes
stronger, more bars light with each pulse. This indicator is available
only in the blip view.
The
Pulse Rate Display. Shows the pulse rate in beats per minute. It
flashes during loss-of-pulse alarms and when the pulse rate is outside
of the alarm limits. During Pulse Search, the N-600 continues to
update the display. Pulse rates outside of the pulse rate range (20 to
250 bpm) are displayed as the closest value within the range. If alarm
limits have been changed from their power-on defaults, a decimal
point (.) is displayed after the BPM value (112.).
The
AC Power Indicator. Lights continuously when the N-600 is
connected to AC power. It also indicates that the battery is charging.
It is off when the N-600 is being powered by internal battery.
12
The
Low Battery Indicator. Lights continuously when 15 or fewer
minutes of battery capacity remain. Flashes when the battery capacity
reaches critical condition.
Page 23
Description of Controls, Indicators, and Symbols
The Alarm Silence Indicator. Lights continuously when an audible
alarm has been silenced. It flashes when the alarm silence duration
has been set to Off.
The
Interference Indicator. The Interference Indicator lights
whenever the N-600 algorithm detects that the incoming signal
1
quality is degraded
.
An intermittently lit
Interference Indicator is common during
patient monitoring, and indicates the N-600 algorithm is dynamically
adjusting the amount of data required for measuring SpO
2 and Pulse
Rate. When lit continuously, it indicates the N-600 algorithm has
extended the amount of data required for measuring SpO
2 and Pulse
Rate and, consequently, fidelity in tracking rapid changes in these
values may be reduced (see Description of Displays and Indicators
on page 10).
The
Pulse Search Indicator. Lights continuously prior to initial
acquisition of a pulse signal and during prolonged and challenging
monitoring conditions. The Pulse Search Indicator flashes during a
loss-of-pulse signal.
The
Data In-Sensor Indicator. Lights to indicate that the attached
OxiMax sensor contains a patient sensor event record. The sensor
event record information may be viewed or printed.
The
SatSeconds Indicator. Fills in clockwise as the SatSeconds
alarm management system detects a %SpO
limit setting. Empties in counterclockwise direction when %SpO
2 reading outside of the
2
reading is within limits. When the indicator is full, a medium priority
alarm will sound.
The
Fast Response Mode Indicator. The response mode setting
dictates the response time (two-seconds to four-seconds in fast mode
and four-seconds to seven-seconds in normal mode) applied by the N-600 algorithm in its calculation of SpO
1
Degradations can be caused by ambient light, electrical noise, electro-surgical
interference, patient activity, or other causes.
OxiMax N-60013
2. The N-600 algorithm
Page 24
Description of Controls, Indicators, and Symbols
calculation of pulse rate is unaffected by the response mode setting.
The trending interval (two-seconds or four-seconds) is updated
automatically by the N-600 to roughly correspond with the SpO
calculation response time.
The
Neonate Alarm Limits Indicator. This symbol is displayed
when the alarm limits are set to neonate. No symbol is displayed
when the N-600 is set to adult limits.
Description of Audible Indicators
Table 1 identifies the audible indicators of the N-600 indicators.
Table 1: Audible Indicators
FunctionDescription
2
Alarm Silence
Reminder
Confirmation ToneThree beeps sound to indicate that default
Invalid Button Press
Valid Button Press
High Priority Alarm
Medium Priority
Alarm
Low Priority Alarm
Three beeps that sound approximately every
three minutes when alarms are silenced with
the alarm silence duration set to Off and the
alarm silence reminder function is enabled.
settings have been saved or reset to factory
defaults or trend data has been deleted.
A short, low
button has been pressed that is not appropriate
for the current state of the N-600.
A short, medium
an appropriate button has been pressed.
A high
loss
-of-pulse.
A medium
SpO
2 or pulse rate limit violation.
A low
-pitched tone indicating that a
-pitched tone indicating that
-pitched, fast-pulsing tone indicating
-pitched, pulsing tone indicating an
-pitched, slow-pulsing tone indicating an
OxiMax sensor disconnect, low battery, or
N-600 failure.
14
Page 25
Description of Controls, Indicators, and Symbols
Table 1: Audible Indicators
FunctionDescription
Power
-On Self-Tes t
Pass
A 1-second tone indicating that the N-600 has
been turned on and has successfully completed
the power
-on self-test.
Pulse BeepA single beep sounds for each detected pulse.
Volume Setting ToneA continuous tone that is used when adjusting
The pitch of the pulse beep signal changes with
-by-point rise or fall in the saturation
a point
level.
the alarm volume.
OxiMax N-60015
Page 26
Blank Page
Page 27
Setting up the N-600
WARNING: To ensure patient safety, do not place the pulse
oximeter in any position that might cause it to fall on the patient.
WARNING: As with all medical equipment, carefully route
patient cabling to reduce the possibility of patient entanglement
or strangulation.
WARNING: Ensure that the speaker is clear of any obstruction.
Failure to do so could result in an inaudible alarm tone.
WARNING: Disconnect the N-600 and Nellcor OxiMax sensor
from the patient during magnetic resonance imaging (MRI)
scanning. Induced current could potentially cause burns.
WARNING: To ensure accurate performance and prevent device
failure, do not subject the N-600 to extreme moisture, such as
direct exposure to rain. Such exposure may cause inaccurate
performance or device failure.
WARNING: Do not use an N-600 pulse oximeter, OxiMax sensor,
cables, or connectors that appear damaged.
WARNING: Do not lift the pulse oximeter by the pulse oximetry
cable or power cord because the cable or cord could disconnect
from the pulse oximeter, causing the pulse oximeter to drop on
the patient.
OxiMax N-60017
Page 28
Setting up the N-600
WARNING: The N-600 is not defibrillator-proof. However, it
may remain attached to the patient during defibrillation or while
an electrosurgical unit is in use, but the readings may be
inaccurate during the defibrillation and shortly thereafter.
WARNING: In the USA, do not connect the pulse oximeter to an
electrical outlet controlled by a wall switch, because the pulse
oximeter may be accidentally turned off.
WARNING: Use only the Nellcor pulse oximetry cable DOC-10
with the N-600 pulse oximeter. Use of another pulse oximetry
cable will have an adverse effect on performance. Do not attach
any cable that is intended for computer use to the OxiMax sensor
port. Do not connect any device other than a Nellcor-approved
OxiMax sensor to the OxiMax sensor connector.
18
WARNING: The N-600 should not be used adjacent to or stacked
with other equipment. If adjacent or stacked use is necessary, the
N-600 should be observed to verify normal operation in the
configuration it is to be used.
Page 29
List of Components
Setting up the N-600
QuantityItem
1N-600 Pulse Oximeter
1Nellcor OxiMax Sensor or Assortment Pack
1
1N-600 Operator’s Manual (applicable to
1Power Cord (applicable to country of sale)
1Quick Guide
DOC
country of sale) and/or Compact Disc
Connecting the N-600 to AC Power
WARNING: In the USA, do not connect the pulse oximeter to an
electrical outlet controlled by a wall switch, because the pulse
oximeter may be accidentally turned off.
Caution: The SUPPLY VOLTAGE SELECTOR switch must be
set to the correct voltage (115 or 230) to avoid equipment damage
and ensure battery charging.
-10 Pulse Oximetry Cable
OxiMax N-60019
Page 30
Setting up the N-600
Caution: Use only the hospital-grade power cord provided by
Nellcor.
1 — Power Connector2 — Supply Voltage Selector
1.Set the SUPPLY VOLTAGE SELECTOR (2) switch to the
applicable voltage.
20
2.Plug the female connector end of the power cord into the N-600
POWER CONNECTOR (1) on the rear of the N-600.
3.Plug the male connector of the power cord into a properly
grounded AC outlet.
4.Verify that the N-600’s
Note: If the
AC Power Indicator is not lit, check:
AC Power Indicator is lit.
•the power cord
•the SUPPLY VOLTAGE SELECTOR switch
•the user-accessible fuses
•the AC power outlet
Page 31
Connecting an OxiMax Sensor to the N-600
The OxiMax sensor type is displayed at the bottom of the display
when an OxiMax sensor is connected to the N-600 or when the N-600
completes POST with an OxiMax sensor attached.
Caution: Use only Nellcor-approved OxiMax sensors and pulse
oximetry cables.
Note: Physiological conditions, medical procedures, or external
agents that may interfere with the N-600’s ability to detect and
display measurements include dysfunctional hemoglobin,
arterial dyes, low perfusion, dark pigment, and externally
applied coloring agents, such as nail polish, dye, or pigmented
cream.
Setting up the N-600
1 — SpO
1.Connect a DOC-10 pulse oximetry cable to the SpO
2OxiMax Sensor Port
2OxiMax
sensor port (1) of the N-600.
2.Connect a Nellcor OxiMax SpO
2 sensor to the other end of the
DOC-10 pulse oximetry cable.
OxiMax N-60021
Page 32
Blank Page
Page 33
Battery Operation
WARNING: Dispose of battery in accordance with local
requirements and regulations.
Operating the N-600 on Battery Power
The N-600 has an internal battery that can be used to power the
N-600 during transport or when AC power is not available. A new,
fully charged battery will provide at least two hours of monitoring
time under the following conditions:
•No audible alarms sound
•No analog or serial output devices are attached to the
N-600
The N-600 cannot operate with a fully discharged battery. Before
attempting to turn on an N-600 whose battery charge has been
completely depleted, first plug the N-600 into an AC outlet to allow
the battery to charge for a few minutes. The N-600 may then be
powered on.
To charge a low or dead battery, connect the N-600 to AC power. A
full charge of a dead battery takes 14 hours while the N-600 is turned
off. A full charge of a dead battery takes 18 hours while the N-600 is
in operation (monitoring a patient).
When all of the following conditions are present for 15 minutes, the
N-600 will automatically shut down:
•N-600 is running on battery power
•No buttons have been pressed
OxiMax N-60023
Page 34
Battery Operation
Note: Whenever the N-600 is connected to AC power, the battery is
Low Battery Indicator
The Low Battery Indicator lights and a low priority alarm begins to
sound when approximately 15 minutes of monitoring time is
available on the existing battery charge. Refer to Table 2 for a
description of the low and critical battery conditions.
•No pulse has been detected (for example, when a patient
is not connected to the OxiMax sensor or the OxiMax
sensor is disconnected from the N-600)
•No alarms are present (other than low battery or a
non-correctable error)
being charged. Therefore, it is recommended that the N-600
remain connected to AC power when not in use. This will
ensure a fully charged battery whenever it is needed.
24
If the N-600 is not on AC power, a low battery audible alarm can be
canceled by pressing the
Indicator and display screen message will continue to be displayed.
Plugging the N-600 into AC power will silence the audible alarm, but
the
Low Battery Indicator will stay lit as long as the battery is in the
low voltage condition. After the 15-minute period of low battery
condition, a high priority alarm will sound for about 10 seconds
before the N-600 shuts off.
If the N-600 backlight is turned off during a low battery condition, the
backlight cannot be turned back on until the low battery condition is
corrected.
It is recommended that qualified service personnel replace the
internal battery every 24 months. Replaced batteries should be
disposed of in accordance with local ordinances.
Alarm Silence button. The Low Battery
Page 35
Battery Operation
Caution: If the N-600 pulse oximeter is to be stored for a period of
three months or longer, notify service personnel to remove the
battery from the pulse oximeter prior to storage. Recharge the
battery when the battery has not been charged for two or more
months.
Caution: The pulse oximeter default settings will return to
factory default setting if the battery becomes fully discharged or
is replaced. Qualified service personnel will have to reset the
institutional defaults, following the instructions in the N-600 Service Manual.
Note: If the AC voltage selector switch on the N-600 rear panel does
not match your AC voltage source, the N-600 may run on
battery power, even though it is plugged into AC power, which
will eventually result in a low priority alarm and a lighted
Battery
Indicator. Ensure that the switch setting matches your
Low
AC voltage.
Note: As the battery is used and recharged over a period of time, the
amount of time between the onset of the low battery alarm and
the N-600 shut-off may become shorter
OxiMax N-60025
Page 36
Battery Operation
Table 2: Low Battery and Critical Battery
State
1No No Yes
2No No No
Critical
Battery
Low
Battery
ACOperation
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY
message
Audible alarm
Error code
Effect of
key
Shutdown
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY
message
2- normal
-on
-off
-off
-off
-none
Alarm Silence
-normal
-N/A
2-normal
-off
-off
-off
26
Audible alarm
Error code
Effect of
key
Shutdown
-none
Alarm Silence
-normal
- N/A
-off
Page 37
Battery Operation
Table 2: Low Battery and Critical Battery (Continued)
State
3No Yes No
Critical
Battery
Low
Battery
ACOperation
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY
message
Audible alarm
Error code
Effect of
SILENCE
silences audio alarm, second
press cancels LOW
BATTERY message. The
LED stays on until Low
Battery Condition is
corrected.
Shutdown
2-normal
-off
-on
-on
-low priority
-logged
ALARM
key-First press
-N/A
OxiMax N-60027
Page 38
Battery Operation
Table 2: Low Battery and Critical Battery (Continued)
State
4No Yes Yes
Critical
Battery
Low
Battery
ACOperation
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY
message
Audible alarm
Error code
Effect of
key
until low battery condition
is corrected.
Shutdown
Note: Connecting AC
2-normal
-on
-on
-off
-off
-logged
Alarm Silence
-N/A. The LED stays on
-N/A
functions the same as
Alarm Silence key
in state 3.
28
5Not used
Page 39
Battery Operation
Table 2: Low Battery and Critical Battery (Continued)
State
6Yes Yes No
7YesYesYes
Critical
Battery
Low
Battery
ACOperation
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY
Audible alarm
Error code
Effect of
Shutdown
SpO
AC/Battery charge LED
LOW BATTERY LED
2-not displayed
(flashing)
message
-on
-displayed and
logged
Alarm Silence
key
-none
-after 10 seconds
2-not displayed
(flashing)
-off
-on
-high priority
-on
-on
LOW BATTERY
message
Audible alarm
Error code
logged
Effect of
key
Shutdown
OxiMax N-60029
-on
-high priority
-displayed and
Alarm Silence
-N/A
-after 10 seconds
Page 40
Blank Page
Page 41
Using the N-600
Introduction
The parameters of the N-600 are preset to factory default settings. See
Factory Defaults on page 145. The factory default parameters may be
changed to institutional default parameters by following the
procedures in the N-600 Service Manual.
Table 3 lists the parameters, ranges available, and the factory default
setting. The parameters may be set on an individual basis, by the
clinician, and these settings will remain in effect until the N-600 is
turned off.
EnglishEnglish
Dutch, Finnish,
French, German,
Norwegian,
Portuguese,
Spanish, Italian,
Swedish
0 to 1044
Normal or FastNormalNormal
Page 43
Table 3: Parameter Ranges (Continued)
Using the N-600
Parameter
RS-232 Level
Nurse Call
Ranges/
Selections
Normally High,
Normally Low
Polarity
SatSecondsOff, 10, 25, 50,
100
Sensor Event
Date Format
(SENSOR-R
and
SpO
2,
SpO
2+Pulse Rate,
Default (default is
factory default)
SENSOR-RW
Sensor
Yes, NoYesYes
Messages
Enabled
Trend DisplayDual, %SpO
2,
Pulse, Histogram,
Amplitude
Trend Scale48, 36, 12,8, 4, 2,
1 hours, 30, 15
minutes, 40, 20
seconds
Factory Adult
Defaults
Factory
Neonate
Defaults
Normally lowNormally low
OffOff
DefaultDefault
%SpO2%SpO2
2 hours2 hours
OxiMax N-60033
Page 44
Using the N-600
Turning On the N-600
Before using the N-600 in a clinical setting, you must verify that the
N-600 is working properly and is safe to use. Proper working
condition will be verified each time the N-600 is turned on as
described in the following procedure.
Caution: If any indicator or display element does not light when
the pulse oximeter is turned on, do not use the pulse oximeter.
Instead, contact qualified service personnel, your local Nellcor
representative, or Nellcor’s Customer Services Department.
Note: Physiological conditions, medical procedures, or external
agents that may interfere with the N-600’s ability to detect and
display measurements, include dysfunctional hemoglobin,
arterial dyes, low perfusion, dark pigment, and externally
applied coloring agents such as nail polish, dye, or pigmented
cream.
Note: The N-600 automatically starts the Power-On Self-Test
(POST), which tests the N-600 circuitry and functions.
Caution: During POST (immediately after power-up), confirm
that all indicators light, all display segments turn on, and the
pulse oximeter speaker sounds a one-second tone.
1.Turn on the N-600 by pressing the
2.Ensure that all of the front panel indicators illuminate.
On/standby button.
34
Page 45
Using the N-600
3.Once the display test portion of POST is complete, the N-600
software version is displayed for approximately five seconds.
Note: The software version shown above is only a sample. Check
your N-600 for the software version installed.
Software version numbers are often needed when calling Nellcor’s
Customer Services Department or your local Nellcor representative
for technical assistance. Write down the software version number and
have it available prior to requesting technical assistance.
4.If the N-600 detects an internal problem during POST, an error
tone sounds and the N-600 displays an error code (EEE) and the
corresponding number. See Troubleshooting on page 123.
5.Upon successful completion of the POST, the N-600 sounds a
one-second tone indicating that the N-600 has passed the test.
OxiMax N-60035
Page 46
Using the N-600
WARNING: If you do not hear the POST pass tone, do not use
the pulse oximeter.
WARNING: Ensure that the speaker is clear of any obstructions.
Failure to do so could result in an inaudible alarm tone.
Note: In addition to serving as the POST pass verification, the POST
OxiMax Sensor Attached
When an OxiMax sensor is attached to the N-600, a “DATA TYPE: . .
.” message is displayed briefly at the bottom of the N-600 display. For
a sensor containing data, the message identifies the sensor data type.
For a blank sensor, the message identifies the N-600’s current data
type setting that will be used to write data to the sensor. The data type
settings are SPO2 and SPO2+BPM.
pass tone also functions as an audible confirmation that the
speaker is performing properly. If the speaker does not
function, the alarm warning sounds cannot be heard.
36
Note: The type of data recorded is only displayed when data is
present in the OxiMax sensor.
The N-600 displays zeros in the %SpO
2 and Pulse Rate displays
while the N-600 is searching for a valid pulse. For optimal
performance, allow the N-600 to search and lock onto a pulse for
approximately 10 seconds in non-interference conditions.
Page 47
Using the N-600
When a valid pulse is detected, the N-600 enters the Monitoring
Mode and displays patient parameters.
Look for movement of the blip bar or of the plethysmographic
waveform indicating that the N-600 is displaying real-time data.
Listen for the pulse beep tone. If the pulse beep tone does not sound
with each pulse, it is an indication that the pulse beep volume is set to
zero, the speaker is malfunctioning, or the signal is corrupted.
When an OxiMax sensor is attached to the N-600 and is applied to a
patient, if the N-600 loses the pulse signal, the N-600 will display “---
& ---” (three dashes and three dashes) and remain in Pulse Search
Mode for five seconds before displaying the poor signal condition
screen. The poor signal condition screen is part of the N-600's Sensor
Messages feature. For more information about OxiMax Sensor
Messages, refer OxiMax Sensor Messages on page 56.
OxiMax N-60037
Page 48
Using the N-600
No OxiMax Sensor Attached
Upon successful completion of the POST, the N-600 sounds a
one-second tone indicating that the N-600 has passed POST.
or
38
The N-600 displays dashes (---) and the Pulse Search Indicator is
not lit, indicating that the N-600 failed to detect an OxiMax sensor.
Page 49
Turning the Backlight On or Off
Note: When the backlight is off, any of the following conditions will
turn on the backlight:
•pressing any of the softkeys
Using the N-600
•pressing the
•pressing the Alarm Silence button
•any alarm
LIGHT
With the N-600 in the normal monitoring mode, press the
softkey.
Adjusting Screen Contrast
With the N-600 in the normal monitoring mode:
1.Press the
2.Press the
contrast is obtained.
Contrast button.
Adjust Up or Adjust Down button until the desired
Contrast button
LIGHT
3.Press the
OxiMax N-60039
Contrast button.
Page 50
Using the N-600
Selecting the Pleth View
The pleth view displays the pleth waveform, %SpO2, and pulse rate
data. Refer to Principles of Operation on page 149, for a description
of the pleth waveform.
With the N-600 in the normal monitoring mode:
SETUP
VIEW
PLETH
1.Press the
2.Press the
3.Press the
SETUP softkey.
VIEW softkey.
PLETH softkey.
40
Page 51
Selecting the Blip View
Displays SpO2, pulse rate, blip bar, and limits in a larger format for
easier viewing.
With the N-600 in the normal monitoring mode:
Using the N-600
SETUP
VIEW
BLIP
1.Press the
2.Press the
3.Press the
SETUP softkey.
VIEW softkey.
BLIP softkey.
Setting the Pulse Beep Volume
With the N-600 in the normal monitoring mode:
1.Press and hold the
Adjust Up/Adjust Down button to increase/
decrease pulse beep volume.
OxiMax N-60041
Page 52
Using the N-600
Setting the Alarm Volume
The Alarm Volume display allows the user to adjust the volume of
alarm tones.
With the N-600 in the normal monitoring mode:
1.Press the
displayed and sounds on the N-600.
2.While continuing to press the
hold the
the volume.
Alarm Silence button until the alarm volume level is
Adjust Up/Adjust Down button to increase/decrease
Alarm Silence button, press and
42
Page 53
Setting the Date and Time
WARNING: The sensor extrapolates from the date and time
provided by the N-600 when recording the sensor event record to
the sensor. The accuracy of the date/time is the responsibility of
the N-600. It is recommended that the N-600 user set the time/
date to the correct value before a sensor event record-enabled
sensor is connected, and that this date/time not be changed while
the sensor remains connected. Since a sensor with sensor event
record data can be transported from one monitor to another,
having discrepancies in the date/time between monitors and the
sensor event record data will affect the order the sensor event
record data appears. To eliminate this possible problem, all
monitors within an institution should be set to the same time.
With the N-600 in the normal monitoring mode:
Using the N-600
SETUP
NEXT
CLOCK
SET
OxiMax N-60043
1.Press the
2.Press the
3.Press the
4.Press the
SETUP softkey.
NEXT softkey.
CLOCK softkey.
SET softkey.
Page 54
Using the N-600
5.Press the SELECT softkey to select:
TIME HOURS : MINUTES : SECONDS (16:46:05)
DATE DAY - MONTH - YEAR (30-JAN-02)
EXIT
6.Use the
selected value.
7.Press the
Adjust Up or Adjust Down buttons to change the
EXIT softkey.
44
Page 55
Setting Alarm Silence Duration
The Alarm Silence Duration display allows the user to adjust the
alarm silence duration.
With the N-600 in the normal monitoring mode:
Using the N-600
1.Press the
setting is displayed. Alarm silence durations that are available are
OFF, 30, 60, 90, and 120 seconds.
2.Press and hold the
button to increase the alarm silence duration setting.
3.Press and hold the
button to decrease the alarm silence duration setting.
Note: Releasing the
alarm silence duration.
Alarm Silence button until the alarm silence duration
Alarm Silence button and the Adjust Up
Alarm Silence button and the Adjust Down
Adjust Up or Adjust Down button sets the
OxiMax N-60045
Page 56
Using the N-600
Disabling Audible Alarms
Setting the alarm silence duration to OFF means that the N-600 will
produce no audible alarms.
Note: The ability to set the alarm silence duration to OFF can be
enabled or disabled by qualified service personnel as described
in the N-600 Service Manual. The current copy of the service
manual is available on the Internet at:
WARNING: Do not silence the audible alarm function or
decrease the audible alarm volume if patient safety could be
compromised.
With the N-600 in the normal monitoring mode:
46
1.Press the
setting is displayed.
Alarm Silence button until the alarm silence duration
Page 57
2.While pressing the Alarm Silence button, press and hold the
Adjust Up button until OFF is displayed. Release the buttons.
Selecting Standby Mode
The standby mode allows the N-600 to retain the alarm limit settings
that are in effect while monitoring a patient. The N-600 must be
powered by AC power to enter the standby mode.
Using the N-600
Usually the standby mode is used when a patient has to leave the
N-600 for a period of time and will return to the same N-600.
To place a N-600 in the standby mode:
1.The N-600 should be monitoring a patient.
2.The N-600 alarm limits should be configured to the patient being
monitored.
3.Disconnect the sensor from the N-600.
4.Press the
alarms.
Alarm Silence button. This silences the audible
OxiMax N-60047
Page 58
Using the N-600
5.Press the Alarm Silence button. This disables the alarm
messages.
The N-600 is now in standby. To return to normal monitoring,
connect the sensor to the N-600 and the patient.
Adult-Pediatric or Neonatal Settings
The clinician can set the N-600’s operating mode to adult-pediatric or
neonatal by using the
in the N-600 until the N-600 is turned off. The factory default
power-on setting is for adult-pediatric patients. This default setting
can be changed to neonatal by qualified service personnel using the
procedures indicated in the N-600 Service Manual.
LIMITS softkey. The setting will only remain
48
Refer to Table on page 145, for neonate factory default limit settings.
Refer to Table on page 145, for adult factory default limit settings.
WARNING: Each time the pulse oximeter is used, check alarm
limits to ensure that they are appropriate for the patient being
monitored.
Page 59
Setting Patient Adult-Pediatric/Neonatal Mode
With the N-600 in the normal monitoring mode:
Using the N-600
LIMITS
1.Press the
2.The N-600 will display the
LIMITS softkey.
Adult Limits or Neonate Limits
screen, depending on the patient setting being used.
ADULT or
NEO
3.Press the
NEO or ADULT softkey to select ADULT LIMITS or
Neonate Limits as applicable for the patient being monitored.
OxiMax N-60049
Page 60
Using the N-600
Alarm Limit Changed Indicator
Alarm limits that have been changed from the institutional or factory
default settings are identified by a decimal point (.) after the displayed
reading (%SpO
2 or BPM). The changed parameter is also identified
by a decimal point on the alarm limits screen.
Setting Alarm Limits
50
The Alarm Limit display allows the user to adjust the upper and lower
saturation and pulse rate limits. It also allows the user to adjust the
SatSeconds limit.
The Alarm Limit display is accessed by pressing the
LIMITS softkey
on the Main menu.
The Alarm Limit display includes the alarm limit table and current
measured %SpO
indicate whether the N-600 is in
2 and pulse rate. The title of the alarm limit table will
Adult or Neonate mode. If
SatSeconds are enabled, the Alarm Limit display also includes the
SatSeconds Indicator. Decimal points after the displayed %SpO2 or
pulse rate indicate that the respective limits have been changed from
the power-on defaults.
With the N-600 in the normal monitoring mode:
Page 61
Using the N-600
LIMITS
ADULT or
NEO
1.Press the LIMITS softkey. Current alarm limits are displayed.
or
2.Press the
ADULT or NEO softkey to select Adult-Pediatric or
Neonatal alarm limits screen.
SELECT
3.Press the
SELECT softkey as required to select the parameter to
be adjusted.
4.Use the
Adjust Up or Adjust Down buttons to increase or
decrease the selected limit parameter.
OxiMax N-60051
Page 62
Using the N-600
5.Repeat steps 3, 4, and 5 as necessary to complete the alarm limits
setup.
EXIT
6.To accept the changes, let the display time-out or press the
softkey to exit the display and return to normal monitoring.
Note: Limit changes will only be in effect as long as the N-600
remains turned on. When the N-600 is turned off, the
institutional or factory default limits will be restored into the
N-600. When the N-600 is turned on, the institutional or
factory default limits will be in effect. Factory or institutional
defaults are selected by qualified service personnel following
the procedure in the N-600 Service Manual.
Setting SatSeconds Alarm Limit
Refer to SatSeconds on page 141, for a description of the SatSeconds
function.
With the N-600 in the normal monitoring mode:
LIMITS
1.Press the
LIMITS softkey. Current alarm limits are displayed.
EXIT
52
SELECT
2.Press the
SELECT softkey twice to select %SpO2 SAT-S.
Page 63
Using the N-600
3.Use the Adjust Up or Adjust Down buttons to select the limit.
The choices are 10, 25, 50, or 100 seconds or OFF.
EXIT
4.Press the
EXIT softkey to save your choice.
Setting the N-600 Response Mode
The purpose of the response mode is to set the response time of the
N-600 algorithm calculation of the SpO
affect the N-600 algorithm’s calculation of pulse rate). The trending
interval (two- or four-seconds) is updated automatically by the N-600
to roughly correspond with the SpO
The response mode programs the N-600 algorithm to display N-600
trend information at two-second intervals (Fast Mode) or four-second
intervals (Normal Mode).
The response mode display screen includes the current SpO
mode setting and the current measured %SpO
in the fast mode, the screen displays the fast mode symbol.
With the N-600 in the normal monitoring mode:
2 (the response mode does not
2 calculation response time.
2 response
2 and pulse rate. When
SETUP
NEXT
NEXT
NEXT
MODE
OxiMax N-60053
1.Press the
2.Press the
3.Press the
4.Press the
5.Press the
SETUP softkey.
NEXT softkey.
NEXT softkey.
NEXT softkey.
MODE softkey.
Page 64
Using the N-600
Note: When the N-600 is in the fast response mode the N-600 may
produce more SpO
2 and pulse rate alarms than the user is
accustomed to seeing.
EXIT
6.Use the
response mode.
7.Press the
Adjust Up or Adjust Down buttons to select the desired
EXIT softkey.
54
Page 65
Selecting the Display Language
The N-600 can be programmed to display the information in various
languages. The languages available are ENGLISH, DANSK
(Danish), FRANCAIS (French), DEUTSCH (German), ITALIANO
(Italian), ESPAÑOL (Spanish), NEDERLANDS (Dutch), NORSK
(Norwegian), PORTUG (Portuguese), SUOMI (Finnish) and
SVERIGE (Swedish).
With the N-600 in the normal monitoring mode:
Using the N-600
SETUP
NEXT
LANG
1.Press the
2.Press the
3.Press the
4.Use the
language.
SETUP softkey.
NEXT softkey.
LANG softkey.
Adjust Up or Adjust Down buttons to select the desired
OxiMax N-60055
Page 66
Using the N-600
EXIT
5.Press the EXIT softkey.
Note: The selected language will be displayed until the N-600 is
turned off. The selected language can be set as a default by
qualified service personnel following the procedures in the
N-600 Service Manual.
OxiMax Sensor Messages
OxiMaxsensor messages consist of sensor adjust condition messages
and sensor adjust messages which, when enabled, are displayed when
the N-600 is not able to display saturation. When OxiMax sensor
messages are displayed, it is an indication that the OxiMax sensor is
functioning correctly, but the site to which the OxiMax sensor applies
or the application method is not optimal for calculating %SpO
Condition messages are followed by action messages. Up to three
condition messages may be displayed on the “
CONDITION
condition display may be dismissed by using the
exited, the OxiMax sensor message screen will not return until a new
condition occurs.
2.
POOR SIGNAL
” display in priority order, highest on top. The
EXIT softkey. Once
56
If the
HELP softkey is pressed from the Condition message display,
the action messages are displayed. Action messages are linked to the
sensor type; action messages will be displayed for the type of OxiMax
sensor connected to the N-600. Up to five action messages may be
displayed. Multiple screens may be required to display all of the
messages. When multiple screens are required, navigation between
screens can be accomplished through the
NEXT, BACK, and EXIT
softkeys.
Page 67
OxiMax sensor messages may be disabled. Refer to Troubleshooting
on page 123 for selecting the OxiMax Sensor Messages, Enable/
Disable function.
OxiMax Sensor Adjust Condition Messages
•Condition 1 — SENSOR OFF?
•Condition 2 — SMALL PULSES
•Condition 3 — WEAK SIGNAL
Using the N-600
•Condition 4 — INTERFERENCE
•Condition 5 — EXCESS INFRARED LIGHT
•Condition 6 — INTERFERENCE
•Condition 7 — HIGH PULSE AMPLITUDE
OxiMax N-60057
Page 68
Using the N-600
OxiMax Sensor Adjust Messages
•Message 1 — ALTERNATE SITE?
•Message 2 — COVER SENSOR SITE?
•Message 3 — EAR/FOREHEAD SENSOR?
•Message 4 — NASAL/EAR SENSOR?
•Message 5 — OxiMax ADHESIVE SENSOR
•Message 6 — SECURE CABLE
•Message 7 — HEADBAND
•Message 8 — WARM SITE
•Message 9 — BANDAGE ASSEMBLY
58
•Message 10 — NAIL POLISH
•Message 11 — SENSOR TOO TIGHT?
•Message 12 — REPOSITION SENSOR
•Message 13 — ISOLATE INTERFERENCE SOURCE
•Message 14 — CLEAN SENSOR SITE
Page 69
N-600 Trend
N-600 Trend Data
The trend displays allow the user to view trend data. Two types of
trend data can be viewed:
N-600 trend data can be viewed anytime patient trend is stored in the
N-600. N-600 trend displays are accessed by pressing the
softkey on the main menu and selecting the
The N-600 trend submenu allows you to choose which trend data are
displayed:
•N-600 trend data stored in the N-600
•Patient event data stored in the OxiMax sensor
(single-patient-use OxiMax sensors only) and can be
used with the sensor event record feature.
TREND
MONITR softkey option.
•Saturation and pulse rate (Dual)
•Saturation
•Pulse rate
•Pulse amplitude
•Histogram
The N-600 can graphically display trend data for SpO
both. Trend data is stored at two- or four-second intervals. When the
2, pulse rate, or
TREND softkey is pressed, “READING TRENDS . . .” is
displayed at the bottom of the N-600 screen, indicating that the N-600
is formatting the trend data to be displayed.
OxiMax N-60059
Page 70
N-600 Trend
The N-600 stores up to 48 hours of four-second trend data or 24 hours
of two-second trend data. The amount of trend data displayed on the
screen is determined by using the
ZOOM softkey. The settings
available are 20 and 40 seconds, 15 or 30 minutes, and 1, 2, 4, 8, 12,
24, 36, or 48 hours. All trend data are displayed in a graphical format
except the 20- and 40-second trend displays, which are shown in
tabular format.
The trend display is scrolled, that is, the data displayed can be moved
throughout the 48 hours of trend data. Selecting the one-hour trend
display allows you to view one hour of trend information. By using
the scrolling feature, any one hour of trend data can be viewed over
the 48 hours of trend information. The
the display to the left and the
Adjust Up button scrolls the display to
Adjust Down button scrolls
the right.
When the data are displayed, the most recent readings are on the right
side of the graph. The numbers below %SpO
2 indicate the highest and
lowest parameter values at the cursor position (vertical dotted line on
the display). See Table 4 on page 64.
60
.
Trend data information may be retrieved through the N-600 data port
or cleared using options available in a display menu.
Caution: N-600 trend data will be lost if the main battery fails or
is removed.
Page 71
Trend Data Operation
While the N-600 is turned on, it stores the N-600 %SpO2 and pulse
rate readings in memory every two seconds (fast response mode) or
four seconds (normal response mode), regardless of whether the
N-600 is monitoring a patient or not. The N-600 can store up to 48
hours of trend data in the normal response mode or 24 hours of trend
data in the fast response mode. The 48/24 hours of stored trend data
are available for downloading to a printer or a portable computer. Up
to 50 alarm limit changes can be stored in the trend data. If more than
50 alarm limit changes occur during the 48/24 hours of trend data
collection, then the additional alarm limit changes take the space
reserved for trend data.
Caution: Change alarm limits only as needed.
Trend memory always contains the most recent 48 or 24 hours of
data, with newly collected data overwriting the oldest data on a
preemptive basis. The N-600 continues to record data points as long
as the N-600 is turned on. Blank data points are collected when the
N-600 is not connected to a patient. Blank data overwrites older
patient data if the memory becomes full. Therefore, to retain old
patient data, it is important to turn off your N-600 when you are not
monitoring a patient. Download the trend memory before it fills up
and overwrites the old data with new or blank data. Figure 3
illustrates when the new trend data starts to overwrite the old trend
data.
N-600 Trend
Figure 3: Trend Overwriting
OxiMax N-60061
Page 72
N-600 Trend
Selecting the Trend Data Display Scale
The trend scale is the amount of trend data displayed on the screen.
With the N-600 in the normal monitoring mode:
TREND
MONITR
VIEW
ZOOM
1.Press the
2.Press the
3.Press the
TREND softkey.
MONITR softkey.
VIEW softkey.
4.Press any of the trend softkeys (
select
HIST (histogram) or AMP (amplitude), press the NEXT
softkey and then the
5.Press the
ZOOM softkey. The Zoom menu is displayed.
HIST or AMP softkeys.
DUAL, SPO2, or PULSE). To
62
TIME
Pressing the
TIME softkey cycles the displayed trend time scale
Note: The 20-second and 40-second trend displays are in tabular
format. The below display starts out in the normal response
mode (left side of the display) and switches to the fast response
mode.
SCALE
AUTO
BACK
Pressing the
SCALE softkey cycles the displayed trend amplitude
scale through ±5 points, ±10 points, ±15 points, ±20 points, ±25
points, ±30 points, ±35 points, ±40 points and ±50 points above and
below the data point under the cursor. The saturation graphical N-600
trend display vertical scale default setting is from 10 to 100 if there is
no data under the cursor. The pulse rate graphical N-600 trend display
vertical scale is from 5 to 250 if there is no data under the cursor.
Pressing the
AUTO softkey presets the amplitude of the graphed
trend data. The maximum trend data point is rounded up to the nearest
multiple of 10, this value is the top of the graph display. The
minimum trend data point is rounded down to the next multiple of 10.
Then 10 is subtracted from the rounded down number, this value is
the bottom of the trend graph.
Pressing the
BACK softkey returns the N-600 to the Monitor menu.
OxiMax N-60063
Page 74
N-600 Trend
Reading the Trend Data Display
Table 4 identifies the components of the trend data display.
Table 4: Reading Trend Display
ItemDescription
64
1Amount of trend data displayed on the screen. Settings available
are 20 and 40 seconds, 15 and 30 minutes, 1, 2, 4, 8, 12, 24, 36,
and 48 hours.
2Date represented by the cursor (item 5).
3No trend data recorded during this time.
4Time represented by the cursor (item 5).
5
6Highest and lowest reading at the cursor position.
7Trend data that is being displayed (%SPO2, BPM, or PAU
- can be moved left or right using the Adjust Up (right)
Cursor
Adjust Down (left) buttons.
or
[pulse amplitude units]).
Page 75
Dual Trend Data Display
The dual trend data display displays both oxygen saturation (%SpO2)
levels and pulse rate (bpm) trend data.
With the N-600 in the normal monitoring mode:
N-600 Trend
TREND
MONITR
VIEW
DUAL
1.Press the
2.Press the
3.Press the
4.Press the
is displayed.
TREND softkey.
MONITR softkey.
VIEW softkey.
DUAL softkey. The dual trend (%SpO2and Pulse Rate)
SpO2 Trend Display
With the N-600 in the normal monitoring mode:
TREND
MONITR
1.Press the
2.Press the
OxiMax N-60065
TREND softkey.
MONITR softkey.
Page 76
N-600 Trend
VIEW
SPO2
3.Press the VIEW softkey.
4.Press the
Pulse Rate Trend Display
With the N-600 in the normal monitoring mode:
TREND
1.Press the TREND softkey.
SPO2 softkey. SpO2 trend data is displayed.
66
MONITR
VIEW
PULSE
2.Press the
3.Press the
4.Press the
MONITR softkey.
VIEW softkey.
PULSE softkey. The pulse trend data is displayed.
Page 77
Histogram Trend Data Display
The histogram displays trend data for the percent of oxygen blood
saturation (SpO
the trend data stored over the period of time indicated on the display.
Refer to Selecting the Trend Data Display Scale on page 62, to set up
the desired trend data scale.
Pulse amplitude cannot be displayed on the histogram display.
With the N-600 in the normal monitoring mode:
2) and pulse rate (bpm). The data displayed represents
N-600 Trend
TREND
MONTIR
VIEW
NEXT
HIST
1.Press the
2.Press the
3.Press the
4.Press the
5.Press the
TREND softkey.
MONITR softkey.
VIEW softkey.
NEXT softkey.
HIST softkey. The Histogram trend data is displayed.
OxiMax N-60067
Page 78
N-600 Trend
Pulse Amplitude Trend Data Display
The pulse amplitude trend data display shows the amplitude of the
patient’s pulse rate over the period of time indicated on the display.
Refer to Selecting the Trend Data Display Scale on page 62, to setup
the desired trend data scale.
With the N-600 in the normal monitoring mode:
TREND
MONITR
VIEW
NEXT
AMP
1.Press the
2.Press the
3.Press the
4.Press the
5.Press the
data is displayed.
TREND softkey.
MONITR softkey.
VIEW softkey.
NEXT softkey.
AMP softkey. The pulse amplitude units (PAU) trend
68
The PAU reading (12 : 20) indicates the pulse amplitude units (upper
and lower) at the cursor position (dashed line). The cursor is moved
right or left using the
Adjust Up (right) and Adjust Down (left)
buttons.
Page 79
Clearing Trend Information
With the N-600 in the normal monitoring mode:
N-600 Trend
TREND
MONITR
NEXT
DELETE
YES
1.Press the TREND softkey.
2.Press the
3.Press the
4.Press the
Note: Press the
MONITR softkey.
NEXT softkey.
DELETE softkey.
NO softkey and then the EXIT softkey to leave this
function without deleting trend data.
5.Press the
YES softkey.
All the trend data is cleared and the N-600 sounds three beeps.
OxiMax N-60069
Page 80
Blank Page
Page 81
OxiMax Sensor Event Record
WARNING: The sensor extrapolates from the date and time
provided by the N-600 when recording the sensor event record to
the sensor. The accuracy of the date/time is the responsibility of
the N-600. It is recommended that the N-600 user set the time/
date to the correct value before a sensor event record-enabled
sensor is connected, and that this date/time not be changed while
the sensor remains connected. Since a sensor with sensor event
record data can be transported from one monitor to another,
having discrepancies in the date/time between monitors and the
sensor event record data will affect the order the sensor event
record data appears. To eliminate this possible problem, all
monitors within an institution should be set to the same time.
The adhesive OxiMax sensors are capable of storing patient event
data. A sensor event record allows alarm event history to travel with
the patient on the sensor’s memory chip for quick assessment at every
point of care where OxiMax monitors are used.
Patient (event) data is stored on the memory chip of adhesive OxiMax
sensors (single-patient-use OxiMax sensors only). The event data is
stored (recorded) with the limit/threshold settings that were active at
the time of the event on the recording monitor. These events can be
viewed on the next OxiMax sensor monitor when the patient moves to
a new point of care.
An event occurs when the %SpO
lower alarm limit for at least 15 seconds. The first OxiMax sensor
event record event will be stored in the OxiMax sensor after the
OxiMax sensor has been attached to a patient for five minutes and
every five minutes thereafter. The maximum number of events that
can be stored in an OxiMax sensor is 100.
OxiMax N-60071
2 value exceeds either the upper or
Page 82
OxiMax Sensor Event Record
Event records can only be viewed after an OxiMax sensor containing
patient data (event records) has been connected to an OxiMax
monitor. Event records are designed to view patient events from prior
areas of care or transport (history) while monitor trend should be used
to view data or events from a patient currently being monitored. The
monitor’s SENSOR EVENT RECORD Indicator will light when an
OxiMax sensor containing event data is connected to the OxiMax
monitor.
Patient event data is accessed by pressing the
main menu and selecting the
record can be viewed in graphical form (
table (
TABLE).
Note: Once the OxiMax sensor event record type is set up in the
OxiMax sensor and event data is stored in the OxiMax sensor,
the OxiMax sensor event record type cannot be reset. The
monitor’s type set up can be changed at any time.
Recording and viewing of OxiMax sensor event record is only
available on OxiMax compatable monitors. The OxiMax sensors may
function on older technology monitors but the OxiMax sensor event
record feature is not available.
Refer to the N-600 Service Manual for the procedure to disable the
storage of sensor event record on an OxiMax sensor.
Setting In-Sensor Data Type
The In-Sensor Data Type display allows the user to set the type of
trend data to be recorded in an OxiMax sensor. OxiMax sensors can be
set to record either SpO
TREND softkey on the
SENSOR softkey option. Sensor event
GRAPH) or in a summary
2 or SpO2+BPM.
72
Note:The OxiMax sensor data type can only be set when an OxiMax
sensor is not connected to the N-600.
With the N-600 turned on and no cable attached to the SpO
2OxiMax
sensor port:
Page 83
OxiMax Sensor Event Record
R
SETUP
SENSO
DATA
1.Press the SETUP softkey.
2.Press the
3.Press the
SENSOR softkey.
DATA softkey.
Note: OxiMax sensor data type settings are displayed on the N-600 as
shown in the figure below (in-sensor data type). If no sensor is
connected, both sensor types and the full set of options for each
are displayed. If a sensor is connected, only the sensor data
type for that sensor is displayed.
Note: The SENSOR-R feature supports all the current OxiMax
sensors. The SENSOR-RW feature is only applicable to
sensors with a read/write chip installed.
SELECT
OxiMax
4.Use the SELECT softkey to toggle between SENSOR-R and
SENSOR-RW.
OxiMax N-60073
Page 84
OxiMax Sensor Event Record
5.Use the Adjust Up or Adjust Down button to select the OxiMax
sensor data type.
are:
SENSOR-R and SENSOR-RW selections
EXIT
6.Press the
OxiMax Sensor Type
When an OxiMax sensor is connected to the N-600, a “SENSOR
TYPE: …” message is displayed for four to six seconds at the bottom
of the display. The message identifies the type (model) of OxiMax
sensor connected to the N-600. Type is used in the determination of
action messages in the OxiMax sensor message(s) function. This
display is the first message displayed when an OxiMax sensor is
connected to the N-600.
•SpO
2
•SpO2+BPM
•DEFAULT
EXIT softkey to set the OxiMax sensor type.
74
Page 85
OxiMax Sensor Data Type
When an OxiMax sensor with no previously recorded patient data is
connected to the N-600, a “DATA TYPE: . . .” message is displayed
briefly at the bottom of the display, this message is displayed after the
OxiMax sensor type message. The message identifies the N-600's
current data type setting that will be used to write data to the OxiMax
sensor. The data type setting options are EVENT/SPO
SPO
2+BPM.
The user can change the setting by referring to Setting In-Sensor Data Type on page 72. The OxiMax sensor event record type must be set
prior to connecting the OxiMax sensor to the N-600.
OxiMax Sensor Event Record
2 and EVENT/
OxiMax Sensor Event Record Data Available
When an OxiMax sensor containing data (single-patient-use OxiMax
sensors only) is connected to the N-600, the S
Indicator on the N-600 front panel blinks at a medium priority flash
TREND
rate to indicate that the OxiMax sensor attached to the N-600 contains
patient event data. The indicator blinks for approximately 60 seconds
SENSOR
OxiMax N-60075
or until the OxiMax sensor is disconnected or until the sensor trend
data is displayed by pressing
TREND, then SENSOR.
ensor Event Record
Page 86
OxiMax Sensor Event Record
A corresponding “DATA IN SENSOR” message is also displayed at
the bottom of the display. After four to six seconds, if all the data has
been read from the OxiMax sensor, the message is replaced with the
main menu.
If data is still being read from the OxiMax sensor, after four to
six seconds, the
READING TRENDS message with an ABORT option.
DATA IN SENSOR message is replaced with a
76
Selecting the
ABORT softkey stops the recording of additional data
in the OxiMax sensor and accessing or viewing the data that is in the
OxiMax sensor.
Sensor event record can be viewed by accessing the
SENSOR
menu.
TREND/
The Sensor Event Record Indicator comes on steady when
OxiMax sensor memory is full and stays on until the OxiMax sensor is
disconnected.
Page 87
OxiMax Sensor Event Record Not Available
If the user selects the TREND/SENSOR option when a connected
OxiMax sensor (single-patient-use OxiMax sensors only) does not
contain data, because no events were recorded to the OxiMax sensor
memory chip in the prior monitoring situation, a “NO DATA”
message is displayed on the default trend or event graph.
A sample event display in which no data are available is shown
below. The message will be cleared when the graph or summary is
exited.
OxiMax Sensor Event Record
OxiMax Sensor Event Record Graphical Data
Graphical representations of patient event history is only available on
single-patient-use OxiMax sensors. Graphed data points are the
minimum or maximum %SpO
throughout the duration of an event (%SpO
alarm threshold for at least 15 seconds) and continuing every
30 seconds until the actual %SpO
threshold.
The duration of an event is determined by the number of data points
in the event. Each data point is stored at 30-second intervals.
Events end for one of four reasons:
•The %SpO
OxiMax N-60077
2 returns to or above the alarm limit
2 value for each 30-second interval
2 continuously below
2 value equals or exceeds the alarm
Page 88
OxiMax Sensor Event Record
•Loss of pulse
•The OxiMax sensor is disconnected
•The OxiMax sensor is off the patient
The graph title shows the data type (
EVENT GRAPH) in the upper
left corner. The number of the displayed event and the total number of
events recorded in the OxiMax sensor are shown to the right of the
title (example, 2/2). The date and time of the displayed event are
shown in the upper center and upper right corner.
The type of data displayed in the graph is indicated to the left of the
vertical axis (%SpO
2). Below this is the range of values (min/max)
during the event. The duration of the event is shown below the range
value. The vertical axis of the graph is labeled to show the magnitude
scale of the graphed data. The horizontal axis is not labeled but
automatically scales to accommodate the number of 30-second
intervals during the event. The alarm threshold (lower than %SpO
2
alarm limit) is represented by a horizontal dotted line across the
graph. The first data point is always the alarm threshold.
Events are displayed one at a time, one per graph. Graphs are
displayed in chronological sequence with the most recent event
shown first when accessing the graphical OxiMax sensor event
display. The user can move between events by using the two left-most
softkeys which are labeled with left- and right-facing arrow icons,
respectively. At the beginning of an event sequence, event 1of 2
events, the left-arrow soft key is blank; at the end of a sequence, event
2 of 2 events, the right-arrow soft key is blank.
78
The
Adjust Up and Adjust Down buttons on the N-600 front panel
can also be used to move through events.
Page 89
OxiMax Sensor Event Record
R
The PRINT softkey allows the user to print the displayed event
graph. The
BACK softkey takes the user back to the previous
TREND/SENSOR sub-menu level.
Viewing and Printing OxiMax Sensor Event History Data
With the N-600 in the normal monitoring mode, you must connect a
printer, capable of printing graphs, to the N-600 data port connector
to print OxiMax sensor event history data.
TREND
SENSO
GRAPH
The N-600 protocol must be set to
GRAPH to print the in-sensor
event history data. Refer to Printing N-600 Trend Information on
page 85. To view and print in-sensor event history data:
1.Connect an OxiMax sensor containing patient data to the N-600.
2.Press the
3.Press the
4.Press the
TREND softkey.
SENSOR softkey.
GRAPH softkey.
Note: Use the left and right arrow softkeys to scroll through the pages
of the event graph.
OxiMax N-60079
Page 90
OxiMax Sensor Event Record
PRINT
EXIT
5.Press the PRINT softkey to print the displayed screen.
6.Press the
A sequence of %SpO
EXIT softkey.
2 + BPM (saturation plus pulse rate)
“dual-view” event graphs are shown below. The dual-view graph is
the same as a single graphical event history graph except the graphs
are compressed horizontally to allow both %SpO
2 and pulse rate
graphs to be shown for the same event.
80
Page 91
OxiMax Sensor Tabular Event Data
The OxiMax sensor tabular event data is a listing of all events
recorded on the OxiMax sensor’s memory chip.
OxiMax Sensor Event Record
The table title shows in the upper left corner. Below the table title is a
six-column table with left-to-right column headings of event number
(#), date (
%SPO
DATE), event start time (START), event duration (DUR),
2 minimum and maximum values during the event (%SPO2),
and pulse rate minimum and maximum values during the event
(BPM).
Event data are listed in chronological order with the most recent event
shown first, at the top of the list, when the tabular Event Summary
display is first accessed. Four events can be displayed
simultaneously; the table must be scrolled to view additional events.
The user can move to the next screen view of the table, the next three
events (the previously displayed bottom or top event is retained as the
fourth event for context when a table is scrolled), using the two
left-most softkeys which are labeled with left- and right-facing arrow
OxiMax N-60081
Page 92
OxiMax Sensor Event Record
R
icons, respectively. At the beginning of an event sequence, Event 1
of 5 events, the left-arrow soft key is blank; at the end of a sequence,
Event 5 of 5 events, the right-arrow soft key is blank, indicating you
have reached the beginning or end of the table.
The
Adjust Up and Adjust Down buttons on the N-600 front panel
can be used to move through the Event Summary table line by line.
PRINT softkey allows the user to print the displayed event
The
graph.
The
BACK softkey takes the user back to the previous TREND/
SENSOR
sub-menu level.
Viewing and Printing In-Sensor Tabular Event History Data
The N-600 should be in the normal monitoring mode.
To view and print in-sensor tabular event history data:
82
TREND
SENSO
TAB LE
1.Press the
2.Press the
3.Press the
TREND softkey.
SENSOR softkey.
TABLE softkey.
Page 93
OxiMax Sensor Event Record
PRINT
BACK
4.Press the PRINT softkey to print the data.
5.Press the
BACK softkey.
OxiMax N-60083
Page 94
Blank Page
Page 95
Printing
Printing N-600 Trend Information
Trend information (N-600 and in-sensor event history) may be sent to
a personal computer or to a serial printer.
Note: The protocol settings must be set to ASCII MODE for printing
text data or
GRAPH MODE for printing graphical data.
1 — Data Port Connector
With the N-600 in the normal monitoring mode:
1.Connect the serial printer to the N-600’s data port connector (1),
using Nellcor printer cable part number 036341.
2.Turn on the printer.
SETUP
OxiMax N-60085
3.Press the
SETUP softkey.
Page 96
Printing
R
NEXT
NEXT
COMM
4.Press the NEXT softkey.
5.Press the
6.Press the
7.Set the
Up
NEXT softkey.
COMM softkey.
BAUD rate to the appropriate number using the Adjust
button.
86
SELECT
EXIT
TREND
MONITR
SENSO
8.Press the
9.Set the
graph printing using the
10. Press the
11. Press the
12. Press the
SELECT softkey to select PROTOCOL.
PROTOCOL to ASCII for text printing or GRAPH for
Adjust Up button.
EXIT softkey.
TREND softkey.
MONITR softkey for N-600 trend printing or press the
SENSOR softkey for in-sensor event history data printing.
Page 97
Printing
NEXT
PRINT
13. Press the NEXT softkey.
14. Press the
PRINT softkey.
ASCII printout:
GRAPH printout:
OxiMax N-60087
Page 98
Printing
N-600 Trend Data in ASCII Mode
Refer to Printing N-600 Trend Information on page 85 for the
procedure to print trend information.
The format of data displayed when a trend printout is shown in
Figure 4. “
Readings are displayed in 2- or 4-second intervals depending on the
response mode selected. The values on each row are an average of the
response mode selected period.
At the end of the printout an “Output Complete” line indicates that
the transmission was successful. If the “Output Complete” line is not
present, a corruption of the data may have been detected and the data
should be ignored.
TREND” is displayed in the top row.
88
Figure 4: ASCII Mode Printout
Once a trend printout has begun, it cannot be aborted without turning
off the N-600 or the printer.
Page 99
Trend Data in Graph Mode
Refer to Printing N-600 Trend Information on page 85 for the
procedure to print trend information. See Figure 5.
The graph mode disables all printout functions except trend data.
Graph mode trend printouts are formatted for a Seiko DPU-414 and
Okidata 320 serial printer.
Printing
Figure 5: Graph Mode Printout
Real-Time Display/Printout Format
Real-time data is continuously sent to the data port on the back of the
N-600. Patient data can be obtained through the data port by
connecting the N-600 data port to a PC or serial printer. When a
real-time printout or display is being transmitted to a printer or PC, a
new line of data is displayed every two seconds. Column headings are
displayed or printed after every 25 lines, or if one of the values in the
column heading changes. Readings are displayed at four-second
intervals if the SpO
two-second intervals when the SpO
Data cannot be obtained if the N-600 is operating on battery power.
Note: If the data output stops transmitting, turn the power off and
back on again or, if the N-600 is connected to a PC, send an
XON (Ctrl-q) command to reset the N-600.
OxiMax N-60089
2 response mode is set to normal and at
2 response mode is set to fast.
Page 100
Printing
An example of a real-time output is shown in Figure 6.
90
Figure 6: Real-Time Printout
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.