To contact Nellcor Puritan Bennett’s representative: In the United States, call 1-800-NELLCOR or 510 463-4000;
outside the United States, call your local Nellcor Puritan Bennett representative.
Caution: Federal law (U. S.) restricts this device to sale by or on the order of a physician.
Nellcor Puritan Bennett Italia S.r.L.
Via Edison 6
20090 Assago (MI)
Italy
Tel. +39.2.4577161
Nellcor Puritan Bennett Finland Oy
Kappelitie 8
02200 Espoo
Finland
Tel. +358.9.270.92.900
To obtain information about a warranty, if any, for this product, contact Nellcor Puritan Bennett
Technical Services or your local Nellcor Puritan Bennett representative.
Nellcor Puritan Bennett, Durasensor, Oxisensor II, Oxichip, C-Lock, and the Nellcor Puritan Bennett
knob configuration are trademarks of Nellcor Puritan Bennett Incorporated.
SureTemp is a trademark of Welch Allyn, Inc.
Covered by one or more of the following U. S. Patents and foreign equivalents: 4,621,643; 4,653,498;
4,700,708; 4,770,179;,4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,368,224; 5,632,555; Re.35,122.
TABLE OF CONTENTS
List of Figures
List of Tables
List of Figures..................................................................................................v
List of Tables..................................................................................................vi
This manual contains information for servicing the NPB-3900 series of patient
monitors. Only qualified service personnel should service this product. Before
servicing the NPB-3900, read the operator’s manual carefully for a thorough
understanding of operation.
1.2 WARNINGS, CAUTIONS, AND NOTES
This manual uses three terms that are important for proper operation of the
monitor: Warning, Caution, and Note.
1.2.1 Warning
A warning precedes an action that may result in injury or death to the patient or
user. Warnings are boxed and highlighted in boldface type.
1.2.2 Caution
A caution precedes an action that may result in damage to, or malfunction of, the
monitor. Cautions are highlighted in boldface type.
1.2.3 Note
A note gives information that requires special attention.
1.3 NPB-3900 PATIENT MONITOR DESCRIPTION
The purpose and function of the NPB-3900 series of patient monitors is to
monitor: ECG; heart rate; noninvasive blood pressure (systolic, diastolic, and
mean arterial pressures); functional arterial oxygen saturation; and temperature
for adult and pediatric patients in all hospital areas and hospital-type facilities.
They may be used during hospital transport and in mobile, land-based
environments, such as ambulances.
Refer to the NPB-3900 operator’s manual for a description of the NPB-3900
controls, indicators, and operation. The physical and operational characteristics
of the monitors are described in the operator’s manual and Section 9,
Specifications, of this manual.
The parameter measurements for each model in the NPB-3900 series are
indicated in Table 1-1.
1-1
Section 1: Introduction
ModelParameter
NPB-3910XX
NPB-3920XXX
NPB-3930XXX
NBP-3940XXXX
1.4 P-3900 INTRODUCTION
The P-3900 is an optional, standalone printer designed for use with the
NPB-3900 patient monitor. The P-3900 communicates with the monitor using a
null-modem cable connected between each device’s RS-232 connector. The
P-3900 contains an internal battery, which, when fully charged, will operate the
printer for 3 hours (typical, at 25°C, producing fifteen 20-second printouts per
hour). The P-3900 can be connected to AC power using an external power
supply. The P-3900 uses the same type power supply as the NPB-3900 monitor,
the PS-120V or PS-240V.
Table 1-1: Model Configuration
NIBPSpO2TEMPECG
The P-3900 does not have an On/Off switch. The printer can sense when it has
an established communications link with the monitor. At that time, the green
LINKED indicator on the front panel lights, indicating that the printer is ready
for operation. See the NPB-3900 operator’s manual for more information
regarding use of the printer.
1.5 RELATED DOCUMENTS
To perform test and troubleshooting procedures and to understand the principles
of operation and circuit analysis sections of this manual, you must know how to
operate the monitor. Refer to the NPB-3900 operator’s manual. To understand
the various Nellcor Puritan Bennett sensors, ECG leads, blood pressure cuffs,
and temperature probes that work with the monitor, refer to the operator’s
manual and individual directions for use that accompany these accessories.
1-2
SECTION 2: ROUTINE MAINTENANCE
2.1 Cleaning
2.2 Periodic Safety and Functional Checks
2.3 Battery
2.4 Environmental Protection
2.1 CLEANING
WARNING: Do not spray, pour, or spill liquid on the NBP-3900, its
accessories, connectors, switches, or openings in the chassis. Do not immerse
the NPB-3900 or its accessories in liquid or clean with caustic or abrasive
cleaners.
To clean the NPB-3900, dampen a cloth with a commercial, nonabrasive cleaner
and wipe the exterior surfaces lightly. Do not allow any liquids to come in
contact with the power connector or switches. Do not allow any liquids to
penetrate connectors or openings in the instrument. For cables, sensors, and
cuffs, follow the cleaning instructions in the directions for use that accompany
these accessories.
2.2 PERIODIC SAFETY AND FUNCTIONAL CHECKS
Nellcor Puritan Bennett recommends that the following checks be performed at
least every 2 years by a qualified service technician.
1.Inspect the exterior of the NPB-3900 for damage.
2.Inspect labels for legibility. If the labels are not legible, contact Nellcor
Puritan Bennett’s Technical Services Department or your local Nellcor
Puritan Bennett representative.
3.Verify that the unit performs properly as described in paragraph 3.3.
4.Perform the electrical safety tests detailed in paragraph 3.4. If the unit fails
these electrical safety tests, do not attempt to repair. Contact Nellcor Puritan
Bennett’s Technical Services Department or your local Nellcor Puritan
Bennett representative.
2.3 BATTERY
If the NPB-3900 has not been used for a long period of time, the battery will
need charging. To charge the battery, connect the NPB-3900 to an AC source as
described in the Setup & Use section of the operator’s manual.
NOTE: Storing the NBP-3900 for a long period without charging the battery
may degrade the battery capacity. A complete battery recharge when
not using the monitor requires 8 hours. The battery may be recharged
while the monitor is in use; in which case, the battery will require 14
hours to be recharged. The battery may require a full charge/discharge
cycle to restore normal capacity.
2-1
Nellcor Puritan Bennett recommends that the NPB-3900’s sealed, lead-acid
battery be replaced at 2-year intervals. Refer to Section 6, Disassembly Guide.
2.4 ENVIRONMENTAL PROTECTION
Follow local governing ordinances and recycling plans regarding disposal or
recycling batteries and other device components.
Section 3: Performance Verification
2-3
SECTION 3: PERFORMANCE VERIFICATION
3.1 Introduction
3.2 Equipment Needed
3.3 Performance Tests
3.4 Safety Tests
3.1 INTRODUCTION
This section discusses the tests used to verify performance following repairs or
during routine maintenance. All tests can be performed without removing the
NPB-3900 covers.
If the NPB-3900 fails to perform as specified in any test, repairs must correct the
problem before the monitor is returned to the user.
3.2 EQUIPMENT NEEDED
Table 3-1 lists the equipment required for performance verification.
Table 3-1: Required Test Equipment
EquipmentDescription
Digital multimeter (DMM)Fluke Model 87 or equivalent
Sensor extension cableEC-8
Durasensor® finger clip sensor
Oxisensor® II adhesive sensor
DS-100A
D-25
ECG cableCE-10
ECG electrodesStandard
ECG leadsLE series
NIBP hoseSHBP-10
NIBP cuffSCBP series
Temperature probeWelch Allyn SureTemp® (blue capped
equivalent
NIBP simulatorBio-Tek “BP Pump” or equivalent
Thermometer Calibrator KeyWelch Allyn Model 767
Safety analyzerBio-Tek 601 Pro or equivalent
StopwatchManual or electronic
3-1
Section 3: Performance Verification
3.3 PERFORMANCE TESTS
The battery charge and battery performance tests should be performed before
monitor repairs whenever the battery is suspected as being a source of a problem.
All other tests should be performed following monitor repairs. Before
performing the battery performance test, ensure that the battery is fully charged
(paragraph 3.3.1).
This section is written using Nellcor Puritan Bennett factory-set power-up
defaults. If your institution has reconfigured custom defaults, those values will
be displayed.
3.3.1 Battery Charge
Perform the following procedure to fully charge the battery.
1. Connect the monitor to an AC power source using the PS-120 or PS-240
external power supply and power cord, if needed.
2. Verify that the EXTERNAL POWER indicator is lit.
3. Charge the battery for at least 8 hours. The battery may require a complete
charge/discharge cycle to restore its normal capacity.
4. To check for a full charge, perform the procedure in paragraph 3.3.2
“Battery Performance Test.”
3.3.2 Battery Performance Test
The monitor is specified to operate typically on battery power for a minimum of
4 hours, at 25°C, with one NIBP measurement every 15 minutes. Before
performing this test, ensure that the battery is fully charged (paragraph 3.3.1).
1. Connect the Nellcor Puritan Bennett SRC-2 pulse oximeter tester to the
monitor via the EC-8 sensor cable. Connect the NIBP simulator to the
monitor via the SHBP-10 hose.
3. Set the NIBP simulator to simulate a pressure setting of 120/80 mmHg
and heart rate of 80 bpm.
4. Ensure that the monitor is not connected to AC power.
Section 3: Performance Verification
5. With the NPB-3900 turned off, press the ON/STANDBY button and
verify that the battery icon appears at the bottom of the display after the
power-on self-test is completed. The boxes in the battery icon should all
be filled, indicating the battery is charged.
6. Verify that the monitor is responding to the SpO2 simulator signal and
that the audible alarm is sounding. Use the knob to select the SpO
2
Menu and permanently silence the SpO2 audible alarm.
7. Use the knob to select the NIBP Menu and set the Automatic
Measurement Interval to 15 minutes. Exit the menu and press the front
panel NIBP button to manually initiate the first NIBP measurement.
Subsequent NIBP measurements will be taken automatically every 15
minutes.
8. The monitor must operate for at least 4 hours before the monitor
automatically powers down due to low battery condition.
9. Verify that the low battery alarm occurs 15-30 minutes before the battery
fully discharges.
10. Allow the monitor to operate until it automatically powers down due to
low battery condition. Verify that an audible alarm sounds when the
monitor automatically shuts down. Press the alarm silence button to
terminate this audible alarm.
11. If the monitor passes this test, immediately recharge the battery
3.3.3 Power-On Self-Test
1. Connect the monitor to an AC power source using the PS-120 or PS-240
2. Do not connect any input cables to the monitor.
3. Observe the monitor front panel. With the monitor off, press the
(paragraph 3.3.1, steps 1–3).
power supply and power cord, and verify that the EXTERNAL POWER
indicator is lit.
ON/STANDBY button. The monitor must perform the following sequence.
a.The monitor emits a beep.
b.A few seconds later, the display backlight illuminates, but the display
is blank.
c.The Nellcor Puritan Bennett logo then appears for a few seconds,
with the version numbers of the boot and operational software
displayed in the lower left corner of the display. (The upper version
number corresponds to the boot software, the lower version number
corresponds to the operational software.)
d.A beep signals the end of the power-on self-test. The power-on self-
test takes approximately 10 seconds to complete.
e.Upon successful completion of the power-on self-test, the display will
be in the normal monitoring screen configuration. No vital-sign
numeric values or trend values will be displayed.
3-3
Section 3: Performance Verification
3.3.4 Hardware and Software Tests
Hardware and software testing includes the following tests applicable to the
indicated models in the series.
•3.3.4.1SpO
•3.3.4.2Operation with an ECG Simulator — NPB-3930, NPB-3940
•3.3.4.3Verification of Pneumatic System — NPB-3900
•3.3.4.4 Operation with a Temperature Simulator — NPB-3920, NPB-3940
•3.3.4.5General Operation — NPB-3900
3.3.4.1 SpO
2 Testing (NPB-3900)
2 testing includes the following tests.
SpO
•3.3.4.1.1 Alarms and Alarm Silence
•3.3.4.1.2 Heart Rate Tone Volume Control
•3.3.4.1.3 Dynamic Operating Range
•3.3.4.1.4 LED Excitation Test
3.3.4.1.1 Alarms and Alarm Silence
1. Connect the SRC-2 pulse oximeter tester to the EC-8 sensor extension cable
and connect the cable to the monitor. Set the SRC-2 as follows:
2. Press the ON/STANDBY button to turn the monitor on. After the normal
power-up sequence, verify that the SpO
2% display initially indicates zero
or is blank.
NOTE: The pulse bar may occasionally indicate a step change as the monitor is
in the pulse search mode.
3. Move the modulation switch on the SRC-2 to LOW.
4. Verify the following monitor reaction:
a. The pulse bar begins to track the artificial pulse signal from the SRC-2.
b. Initially, zero is displayed in the SpO
2 frame, or it is blank.
c. After about 10 to 20 seconds, the monitor displays saturation and heart
rate as specified by the tester. Verify that the values are within the
following tolerances:
Oxygen Saturation Range 79% to 83%
Heart Rate Range 35 to 41 bpm
3- 4
Section 3: Performance Verification
d. The audible alarm sounds and both the SpO2% and HEART RATE
displays flash, indicating both parameters have violated the default
alarm limits.
e. The heart rate tone is heard. (Heart rate tone source, found in the
Heart Rate Menu, should be set to “SpO2”.)
5. Press the ALARM SILENCE button on the front panel of the monitor. The
audible alarm is temporarily silenced.
6. Verify the following:
a.The audible alarm remains silenced.
b.The “slashed bell” icon appears in each numeric frame on the display.
c.The SpO
d.The heart rate tone remains audible.
e.The audible alarm returns in approximately
60 seconds.
3.3.4.1.2 Heart Rate Tone Volume Control
Connect the SRC-2 pulse oximeter tester to the EC-8 sensor extension cable and
connect the cable to the monitor. Set the SRC-2 as follows:
1. Power on the monitor and verify that the SpO2 and heart rate values are
correctly displayed. Press the ALARM SILENCE button on the front panel
of the monitor to temporarily silence the audible alarm.
2. Verify that the heart rate tone source, found in the Heart Rate Menu, is set
to “SpO
2”. Press the VOLUME button on the front panel of the monitor.
Within 3 seconds of having pressed the button, rotate the knob CW and
verify that the beeping heart rate tone sound level increases.
2% and HEART RATE displays continue flashing.
POSITION
3. Rotate the knob CCW and verify that the beeping heart rate tone decreases
until it is no longer audible. Rotate the knob CW to return the beep volume
to a comfortable level. (Note that 3 seconds after the last button-press or
rotation of the knob, function of the knob reverts to moving the highlight
on the display screen.)
3-5
Section 3: Performance Verification
3.3.4.1.3 Dynamic Operating Range
The following test sequence verifies proper monitor operation over a range of
input signals.
1. Connect the SRC-2 to the NPB-3900 (using an EC-8) and turn the
NPB-3900 on.
2. Place the SRC-2 in the RCAL 63/LOCAL mode.
3. Set the SRC-2 as indicated in Table 3-2. Verify that the NPB-3900
readings are within the indicated tolerances. Allow the monitor several
seconds to stabilize the readings.
NOTE: A “*” indicates values that produce an alarm. Press the ALARM
SILENCE button to temporarily silence the audible alarm.
NOTE: For the pulse rate setting of 201 bpm, the pulse rate tolerance of 195 to
4.Turn the monitor off.
3.3.4.1.4 LED Excitation Test
This procedure uses normal system components to test circuit operation. A
Nellcor Puritan Bennett Oxisensor II adhesive sensor, model D-25, is used to
examine LED intensity control. The red LED is used to verify intensity
modulation caused by the LED intensity control circuit.
1. Connect an EC-8 sensor extension cable to the monitor.
2. Connect a D-25 sensor to the sensor extension cable.
3. Press the ON/STANDBY button to turn the monitor on.
4. Leave the sensor open with the LED and photodetector visible.
5. After the monitor completes its normal power-up sequence, verify that the
207 bpm is greater than the ±3 bpm accuracy specification of the
monitor, due to the performance characteristics of the SRC-2 tester.
sensor LED is brightly lit.
3- 6
6. Slowly move the sensor LED in proximity to the photodetector element of
the sensor. Verify, as the LED approaches the optical sensor, that the LED
intensity decreases.
7. Open the sensor and notice that the LED intensity increases.
8. Repeat step 6 and the intensity again decreases. This variation is an
indication that the microprocessor is in proper control of LED intensity.
9. Turn the NPB-3900 off.
3.3.4.2 Operation with an ECG Simulator (NPB-3930, NPB-3940)
1. With the monitor off, connect the ECG leads to the appropriate jacks on
the ECG tester.
2. Connect the leads to the CE-10 ECG cable. Connect the CE-10 to the ECG
input port on the NPB-3900. Set the ECG tester as follows:
Heart rate:30 bpm
Amplitude:1 millivolt
Lead select:II
Section 3: Performance Verification
Normal sinus rhythm
Adult mode
NOTE: The accuracy of the monitor’s ECG measurements is ±5 bpm. In the
procedure below, add the tolerance of the simulator to the acceptable
range of readings.
3. Press the ON/STANDBY button to turn the monitor on. After the normal
power-up sequence, verify the following monitor reactions:
a.After at least five heartbeats, the monitor displays a heart rate of
30 ±5 bpm.
b.The audible alarm sounds and the HEART RATE display flashes,
indicating the heart rate is below the default lower alarm limit.
4. Press the ALARM SILENCE button. Verify that the audible alarm is
silenced.
5. Increase the heart rate setting on the ECG simulator to 240 bpm.
6. After at least five heartbeats, verify that the monitor displays a heart rate
of 240 ±5 bpm.
7. Verify that the audible alarm sounds and the HEART RATE display
flashes, indicating that the heart rate is above the default upper alarm limit.
8. Press the ALARM SILENCE button to silence the alarm.
3-7
Section 3: Performance Verification
9. Decrease the heart rate setting on the ECG simulator to 120 bpm.
10. After at least five heartbeats, verify that the monitor displays a heart rate
of 120 ±5 bpm.
11. Disconnect the LL lead from the ECG simulator. Verify that the “Leads
Off” alarm message appears, three dashes are displayed in the HEART
RATE display, and a low priority audible alarm sounds.
12. Reconnect the LL lead to the ECG simulator. Verify that the “Leads Off”
alarm message no longer appears and that the audible alarm is silenced.
13. Repeat steps 11 and 12 for the LA and RA leads.
14. Turn the monitor off.
3.3.4.3 Verification of Pneumatic System (NPB-3900)
Tests in paragraphs 3.3.4.3.1 through 3.3.4.3.5 verify the functionality of the
NPB-3900 pneumatic system. These tests were designed to use the Bio-Tek “BP
Pump” noninvasive blood pressure simulator. The internal test volume of the
Bio-Tek simulator is 250 cm
3
, which is used to calculated the inflation/deflation
rate periods. The Bio-Tek simulator, or any equivalent NIBP simulator, is
required to perform these tests.
The NPB-3900 must be placed in Diagnostic Mode, with the NIBP Test screen
active for each of the NIBP tests. For a detailed explanation of the Diagnostic
Mode, refer to Section 4, Power-up Defaults Menu and Diagnostic Mode.
Each of the tests described in paragraphs 3.3.4.3.1 through 3.3.4.3.5 must be
performed to verify pneumatic system functionality. These tests can be
performed individually (in any order) or sequentially. Prior to performing any of
these tests, perform the following setup procedure. If these tests are performed in
sequence, this procedure needs to be performed once prior to the first test.
1. Turn on the Bio-Tek simulator and press the MODE button to place the
simulator in test mode. The simulator screen will indicate “Internal Cuff”
and “Pressure Gauge”.
2. Connect the simulator hose to the NIBP connector on the NPB-3900.
3. Follow the procedure described in Section 4 to place the NPB-3900 in
Diagnostic Mode with the NIBP Test screen active.
3.3.4.3.1 Pressure Transducer Accuracy
The pressure transducer accuracy test verifies the pressure accuracy of the
NBP-3900 pressure transducer.
3- 8
Section 3: Performance Verification
1. Confirm that the Bio-Tek simulator is in test mode. The simulator should
display “Pressure Gauge”. Confirm that the simulator is set up for the
internal cuff.
2. Confirm that the NIBP Test screen is active on the NPB-3900. Press, then
release, the SPEAKER button on the NPB-3900 to verify that the valve is
closed.
3. Press the CONTRAST button on the NPB-3900, then the ZERO button on
the simulator, to perform an offset adjustment so that the simulator and
NBP-3900 both display a pressure of 0 mmHg.
4. Press the SELECT button on the simulator until the simulator displays
“Pressure Source Set Test Pressure”. Use the UP/DOWN buttons on the
simulator to adjust for 250 mmHg.
5. Press the START PUMP button on the simulator. The simulator will begin
to pressurize. The current pressure in mmHg is displayed on both the
simulator and NPB-3900 displays.
6. Allow 15-20 seconds for the pressure to stabilize. The pressure displayed
on the NPB-3900 and the simulator should be within 5 mmHg of one
another to complete the test successfully.
7. Press the STOP PUMP button on the simulator to stop the test.
8. Press and hold the SPEAKER button until the NPB-3900 displays a
9. Additional NIBP tests may be performed at this time. If no further NIBP
3.3.4.3.2 Pneumatic Leakage
The pneumatic leakage test verifies the integrity of the pneumatic system. A
timer/stopwatch is required for this test.
1. Ensure that the Bio-Tek simulator is in test mode. The simulator should
2. Ensure that the NIBP Test screen is active on the NPB-3900. Press, then
3. Press the CONTRAST button on the NPB-3900, then the ZERO button on
pressure of 0 mmHg.
tests are to be conducted, turn the NPB-3900 off. Normal monitoring
operation will return the next time the monitor is turned on.
display “Pressure Gauge”. Confirm that the simulator is set up for the
internal cuff.
release, the SPEAKER button on the NPB-3900 to verify that the valve is
closed.
the simulator, to perform an offset adjustment so that the simulator and
NBP-3900 both display a pressure of 0 mmHg.
3-9
Section 3: Performance Verification
4. Press the NIBP button on the NPB-3900 to activate the pump. Hold the
5. Press and hold the SPEAKER button until the NPB-3900 displays a
6. Additional NIBP tests may be performed at this time. If no further NIBP
3.3.4.3.3 Inflation Rate
The inflation rate test verifies the inflation rate of the NPB-3900. A
timer/stopwatch is required for this test.
1. Ensure that the Bio-Tek simulator is in test mode. The simulator should
button until the NPB-3900 displays a pressure of approximately 250 mmHg.
Allow 15-20 seconds for the pressure to stabilize. Record the pressure
displayed on the monitor, and initiate a 1-minute timer. After 1 minute,
again record the pressure displayed. The test is successfully completed if
the pressure has dropped by 6 mmHg, or less, during the 1-minute period.
pressure of 0 mmHg.
tests are to be conducted, turn the NPB-3900 off. Normal monitoring
operation will return the next time the monitor is turned on.
display “Pressure Gauge”. Confirm that the simulator is set up for the
internal cuff.
3.3.4.3.4 Deflation Rate
2. Ensure that the NIBP Test screen is active on the NPB-3900. Press, then
release, the SPEAKER button on the NPB-3900 to verify that the valve is
closed.
3. Press the CONTRAST button on the NPB-3900, then the ZERO button on
the simulator, to perform an offset adjustment so that the simulator and
NBP-3900 both display a pressure of 0 mmHg.
4. Press the NIBP button on the NPB-3900 to activate the pump, and
simultaneously start the timer. Hold the NIBP button until the monitor
displays a pressure of 280 mmHg. When a pressure of 280 mmHg is
reached, stop the timer. The test is successfully completed if the inflation
time is between 1 and 6 seconds.
5. Press and hold the SPEAKER button until the NPB-3900 displays a
pressure of 0 mmHg.
6. Additional NIBP tests may be performed at this time. If no further NIBP
tests are to be conducted, turn the NPB-3900 off. Normal monitoring
operation will return the next time the monitor is turned on.
The deflation rate test verifies the deflation rate of the NPB-3900. A timer/stop
watch is required for this test.
3- 10
1. Ensure that the Bio-Tek simulator is in test mode. The simulator should
display “Pressure Gauge”. Confirm that the simulator is set up for the
internal cuff.
Section 3: Performance Verification
2. Ensure that the NIBP Test screen is active on the NPB-3900. Press, then
release, the SPEAKER button to verify that the valve is closed.
3. Press the CONTRAST button on the NPB-3900, then the ZERO button on
the simulator, to perform an offset adjustment so that the simulator and
NBP-3900 both display a pressure of 0 mmHg.
4. Press the NIBP button on the NPB-3900 to activate the pump. Hold the
button until the NPB-3900 displays a pressure of 280 mmHg. Initiate a
1-minute timer, and simultaneously press and hold the ALARM SILENCE
button on the NPB-3900. This will cause the pneumatic system to deflate at
a rate of 3 mmHg/sec ±1.5 mmHg/sec. After 1 minute, record the pressure
displayed on the NPB-3900. The test is successfully completed if the
monitor displays a pressure reading of 10 mmHg to 190 mmHg.
5. Press and hold the SPEAKER button until the NPB-3900 displays a
pressure of 0 mmHg.
6. Additional NIBP tests may be performed at this time. If no further NIBP
tests are to be conducted, turn the NPB-3900 off. Normal monitoring
operation will return the next time the monitor is turned on.
3.3.4.3.5 Over-pressure
The over-pressure test verifies the functionality of the over-pressure relief
system of the NPB-3900.
1. Ensure that the Bio-Tek simulator is in test mode. The simulator should
display “Pressure Gauge”. Confirm that the simulator is set up for the
internal cuff.
2. Ensure that the NIBP Test screen is active on the NPB-3900. Press, then
release, the SPEAKER button on the NPB-3900 to verify that the valve is
closed.
3. Press the CONTRAST button on the NPB-3900, then the ZERO button on
the simulator, to perform an offset adjustment so that the simulator and
NBP-3900 both display a pressure of 0 mmHg.
4. Press the SELECT button on the simulator until the simulator displays
“Overpressure Test”. Press the START TEST button on the simulator. The
simulator will pressurize the system until the monitor’s over-pressure relief
system activates, including the warning display screen. The simulator will
display the pressure value that caused the NPB-3900 over-pressure relief
system to activate. The test is successfully completed if the simulator
displays a pressure reading of 280 mmHg to 330 mmHg.
5. Press and hold the SPEAKER button to ensure that the NPB-3900 displays a
pressure of 0 mmHg.
6. Additional NIBP tests may be performed at this time. If no further NIBP
tests are to be conducted, turn the NPB-3900 off. Normal monitoring
operation will return the next time the monitor is turned on.
3-11
Section 3: Performance Verification
3.3.4.4 Operation with a Thermometer Calibration Key (Models 3920 and 3940)
1. Remove the probe from its holder.
2. Insert the calibration key in the temperature input port
3. Press the ON/STANDBY button to turn the monitor on. After the normal
power-up sequence, verify that the temperature reads 36.3 ±0.1 °C (or 97.3
±0.2 °F).
4. Turn the monitor off.
3.3.4.5 General Operation
The following tests provide an overall performance check of the system:
•3.3.4.4.1Operation with a Human Subject
•3.3.4.4.2Serial Interface Test
•3.3.4.4.3Printer Verification
3.3.4.5.1 Operation with a Human Subject
Patient monitoring involves connecting the monitor to a human subject for a
qualitative test.
1. Connect an EC-8 sensor extension cable to the monitor. Connect a
Nellcor Puritan Bennett Durasensor finger clip sensor, model DS-100A,
to the sensor extension cable. Clip the DS-100A to the subject as
described in the sensor directions for use.
T
on the NPB-3900.
2. Connect a CE-10 ECG cable to the NPB-3930 or NPB-3940. Connect
ECG leads to the cable. Connect ECG electrodes to the leads. Apply ECG
electrodes to the subject according to the lead’s and electrodes’ directions
for use.
3. Connect an SHBP-10 blood pressure hose to the monitor. Apply the
appropriate SCBP series blood pressure cuff to the subject according to
the cuff directions for use.
4. Connect a blue-capped SureTemp oral thermometer probe to the
NPB-3920 or NPB-3940. Place the probe in its holder in the module on
the rear of the monitor.
5. Press the ON/STANDBY button to turn the monitor on and verify that
the monitor is operating.
6. The monitor should stabilize on the subject’s physiological signals in
about 15 to 30 seconds. Verify that the saturation and heart rate are
reasonable for the subject.
7. Press the NIBP button on the front panel of the monitor. Verify that the
blood pressure values are reasonable for the subject.
3- 12
8. Remove the temperature probe from its holder. Following the directions
3.3.4.5.2 Serial Interface Test
Perform the following procedure to test the serial port voltages. The test is
RS-232
qualitative and only verifies that the serial interface port is powered correctly,
and that the “nurse call” signal is operational. The serial connector is a male
DB-9 located on the monitor’s rear panel, identified with the RS-232 symbol.
1. Turn the monitor ON.
2. Set up the DMM with the function set to “VDC” at a range of 10 volts.
3. Connect the DMM negative lead to connector pin 5 (GND), or the shell of
4. Connect the DMM positive lead to the following pins, in turn, and verify
Section 3: Performance Verification
in the NPB-32900 operator’s manual, apply a new probe cover and take
the subject’s temperature. Verify that the temperature measurement is
reasonable for the subject.
the RS-232 connector.
the voltage values listed in Table 3-3. (Voltage for pin 9 is that listed for the
“no alarm” condition.)
Printer verification consists of connecting the printer to the monitor and the
monitor to a human subject for a qualitative test.
1. Connect the output of the appropriate power supply, PS-240V or PS-120V,
to the labeled connector in the rear of the printer. When the printer’s external
15V - 1A
power supply is connected, the printer front-panel charging LED is lighted.
POSITION
RS-232
2. Connect the serial cable between the labeled connectors in the rear panels of
the monitor and the printer.
3. The printer front-panel communication LED is lighted when the RS-232
communications link is completed.
4. Rotate the monitor knob to highlight the setup icon . Press the knob and
ensure Communications Selection is (Printer).
5. Connect an EC-8 sensor extension cable to the monitor. Connect a NellcorPuritan Bennett Durasensor oxygen transducer, model DS-100A, to the
sensor extension cable. Attach the DS-100A to the subject as described in
the sensor directions for use.
6. Press the ON/STANDBY button to turn the monitor on and verify that the
monitor is operating.
7. The monitor should stabilize on the subject’s physiological signal in about
15 to 30 seconds. Verify that the saturation and heart rate is reasonable for
the subject.
3- 14
8. Press the printer CONTINUOUS BUTTON. Verify that the printout contains
vital signs across the top of the paper, and that aSpO
marks, occupies the center portion of the paper. Press the CONTINUOUS
BUTTON again to terminate printout.
9. Disconnect the sensor and shut off the monitor.
3.4 SAFETY TESTS
NPB-3900 safety tests consist of the following Leakage Currents elements,
performed in accordance with IEC 601-1.
3.4.1 Protective Ground Continuity
NOTE: The NPB-3900 does not require an isolated Earth Ground terminal,
neither is one installed. No Protective Ground Continuity check is
required.
3.4.2 Electrical Leakage
NPB-3900 leakage current tests consist of the following elements, performed in
accordance with IEC 601-1, clause 19:
Section 3: Performance Verification
2 waveform, with grid
•Patient Leakage Current
•Patient Leakage Current, with Mains Voltage on the Applied Part
3.4.2.1 Patient Leakage Current
This test measures patient leakage current in accordance with IEC 601-1, clause
19, for Class II, type CF equipment. Patient leakage current in this test is
measured from any individual patient connection to earth (power ground).
NOTE: This test requires a test cable for each patient connector. For example,
the ECG test cable consists of the ECG cable connector, with all the
conductors shorted together, connected to a test lead from the electrical
safety analyzer. Test cables for SpO
in a similar manner.
1. Configure the electrical safety analyzer as recommended by the analyzer
operating instructions.
2. Connect the appropriate external power supply input power cord to the
analyzer as recommended by the analyzer operating instructions. Connect
the external power supply output cord to the monitor.
3. Connect the ECG test cable to the ECG connector on the NPB-3930 /3940
and the appropriate input connector on the analyzer. Turn on the
NPB-3930/3940.
2 and temperature can be configured
4. Perform the test as recommended by the analyzer operating instructions.
Patient leakage current is measured under various conditions of the AC
mains. For each condition, the measured leakage current must not exceed
that indicated in Table 3-4.
3-15
Section 3: Performance Verification
5. Repeat the test for the SpO2 and temperature patient connections, using the
appropriate test cables.
Table 3-4: Current Test
Allowable Leakage
Test Condition
Current (milliamps)
Normal polarity0.01
Normal polarity; Neutral (L2) open0.05
Reverse polarity0.01
Reverse polarity; Neutral (L2) open0.05
3.4.1.3 Patient Leakage Current, with Mains Voltage on the Applied Part
This test measures patient leakage current in accordance with IEC 601-1, clause
19, for Class II, type CF equipment. In this test, 110% of mains voltage is
applied between each patient connection and earth (power ground). Patient
leakage current is then measured from any individual patient connection to earth.
NOTE: This test requires the same test cables for each patient connector as
described in section 3.4.1.2.
WARNING: AC mains voltage will be present on the applied part
terminals during this test. Exercise caution to avoid electrical shock hazard.
1. Configure the electrical safety analyzer as recommended by the analyzer
operating instructions.
2. Connect the monitor’s appropriate external power supply input power cord
to the analyzer as recommended by the analyzer operating instructions.
Connect the external power supply output cord to the monitor.
3. Connect the ECG test cable to the ECG connector on the NPB-3930/3940
and the appropriate input connector on the analyzer. Turn on the
NPB-3930/3940.
4. Perform the test as recommended by the analyzer operating instructions.
Patient leakage current is measured with normal and reverse mains polarity.
For each condition, the measured leakage current must not exceed that
indicated in Table 3-5.
5. Repeat the test for the SpO2 and temperature patient connections, using the
appropriate test cables.
Table 3-5: Leakage Current
Allowable Leakage
Test Condition
Current (milliamps)
Normal polarity0.05
Reverse polarity0.05
3- 16
SECTION 4: POWER-UP DEFAULTS MENU AND DIAGNOSTIC MODE
4.1 Introduction
4.2 Power-up Defaults Menu
4.3 Restoring Factory Settings
4.4 Diagnostic Mode
4.1 INTRODUCTION
This section discusses use of the Power-up Defaults Menu to configure power-on
default settings, and the Diagnostic Mode to obtain service-related information
about the monitor.
4.2 POWER-UP DEFAULTS MENU
The purpose of the Power-up Defaults Menu (Table 4-1) is to allow the
authorized user to create a “power-up default” for each setting in the NPB-3900.
Power-up defaults are the settings in effect each time the NPB-3900 is powered
on. Once the Power-up Defaults Menu is entered, physiological monitoring is
terminated. The screen layouts do NOT display any information associated with
normal monitoring operation.
Use the following procedure to configure the power-up default settings for the
NPB-3900 monitor.
1. While in normal monitoring mode, adjust each accessible setting on the
monitor as desired, using the techniques described in the operator’s manual.
Such settings include alarm limits, choice of display type for the graphic
frame, and ECG lead select.
2. Use the knob to invoke the Set-up Menu (choose the screwdriver icon found
along the bottom of the display).
3. Select the menu item “Enter Power-Up Defaults Menu”. Once selected, a
pop-up box appears with the text “Enter 3-Digit Passcode”. Use the knob to
enter the passcode, 2 1 5. This passcode is set at the factory and may not be
changed.
4. The Power-up Defaults Menu is now present. The available menu items are
explained in the table that follows. Make changes to these menu items as
desired.
Table 4-1: Power-Up Defaults Menu
MENU ITEM*CHOICES**EXPLANATION
Accept Current
Settings?
“Yes”
“No”
If “Yes” is chosen, the current
NPB-3900 settings become the
power-up defaults.
4-1
Section 4: Power-up Defaults Menu and Diagnostic Mode
Table 4-1: Power-Up Defaults Menu - (Continued)
MENU ITEM*CHOICES**EXPLANATION
Permanent Audible
“Make Available”
Alarm Silence
“Deny Access”
Alarm Suspend“Make Available”
“Deny Access”
Auto-Set Limits
“Make Available”
“Deny Access”
If “Make Available” is chosen,
the caregiver may permanently
silence the audible alarm for a
particular parameter via the
Alarm/Limits Menu. Some
institutions may wish to prevent
audible alarms from being
permanently silenced. If so,
“Deny Access” should be
selected.
If “Make Available” is chosen,
the caregiver may invoke the
Alarm Suspend Mode by
pressing and holding the Alarm
Silence button for 2 seconds.
Some institutions may wish to
prevent Alarm Suspend from
being invoked. If so, “Deny
Access” should be selected.
If “Make Available” is chosen,
the caregiver may invoke the
Auto-Set Limits function via
the Alarm/Limits Menu. Some
institutions may wish to prevent
Auto-Set Limits from being
invoked. If so, “Deny Access”
should be selected.
4-2
Section 4: Power-up Defaults Menu and Diagnostic Mode
Table 4-1: Power-Up Defaults Menu - (Continued)
MENU ITEM*CHOICES**EXPLANATION
Language
Enter Diagnostic
Mode
“English”
“Français”
“Deutsch”
“Español”
“Italiano”
“Portugués”
“Japanese”
“Chinese”
“Russian”
“Yes”
“No”
The language selected will be
used for all the text shown on
the display; the selected
language will be effective the
next time the monitor is
powered up.
If “Yes” is chosen, the Powerup Defaults Menu is exited and
the Diagnostic Menu appears.
DoneWhen selected, the Power-up
Defaults Menu is immediately
exited and the user is instructed
to power down the monitor.
* The choice in effect at the time the screen is accessed is shown
in parentheses following the menu item.
** Bold type indicates the choice when the factory-set default menu
appears. The highlighting is displayed in reverse video.
5. After making any desired changes to the menu items, choose the menu item
“Accept current settings?”, select “YES”, then select “Done”.
6. Upon selecting “Done”, a Notice screen will appear, with the directions that
the monitor must be powered off, and that any changes made to the power-up
defaults will be in effect next time the unit is powered up.
4-3
Section 4: Power-up Defaults Menu and Diagnostic Mode
Turn off the monitor at this time. Any changes
made to the power-up defaults will be in effect
the next time the monitor is turned on.
4.3 RESTORING FACTORY SETTINGS
CAUTION: In addition to restoring factory defaults, this procedure will
also clear the contents of trend memory.
NOTICE
NOTE: Read this procedure completely before performing the first step.
The following technique can be used to restore the monitor’s power-up default
settings which were originally established at the factory:
1. With the monitor powered off, simultaneously press the Volume and
2. While continuing to press the Volume and Contrast buttons, power-up the
3. Continue to keep the Volume and Contrast buttons depressed until the
4.4 DIAGNOSTIC MODE
The purpose of Diagnostic Mode is to allow factory, field-service, and hospital
biomedical technicians access to a series of test and system-related information
screens for the purpose of verifying NPB-3900 performance or troubleshooting
problems.
To access the Diagnostic Mode, first invoke the Power-up Defaults Menu as
described in section 4.2. Then, select the menu item, “Enter Diagnostic Mode”.
Choose “Yes”. The Power-up Defaults Menu will be exited and the Diagnostic
Menu will appear.
Contrast buttons on the front keypad.
monitor.
power-up diagnostic sequence is complete. When the normal monitoring
screen appears, release the two buttons.
4-4
Section 4: Power-up Defaults Menu and Diagnostic Mode
DIAGNOSTIC MENU
Error Codes
System Information
System A/D Values
NIBP Test
Return
The Diagnostic Menu lists the test and system-related information screens.
Selection of an item in the menu will invoke that test or information screen. The
test and information screens that appear in the Diagnostic Menu are as follows:
•Error Codes
4.4.1 Error Codes
•System Information
•System A/D Values
• NIBP Test
• Return
This screen displays the 10 most recent error code types, logged by the
NPB-3900. After 10 error code types have been logged, the oldest error code
type will be deleted as new error code types are added. Adjacent to each error
code will be an entry which is the number of occurrences of that error. This
means that if there are many occurrences of one type of error code, that one error
code won’t “overwrite” the other 9 error codes.
Next to the occurrence field is the time and date of the most recent occurrence of
the error code. Error codes may not be changed or reset in this screen. When in
the Error Code screen, the “Return” item is always highlighted; a press of the
knob will return the user to the Diagnostic Menu. Rotating the knob while in the
Error Code screen will have no effect.
NOTE: Refer to Section 5.6.2 for more detail on error codes.
4-5
Section 4: Power-up Defaults Menu and Diagnostic Mode
4.4.2 System Information
SYSTEM INFORMATION
Monitor On-Time1563
Backlight On-Time871
Battery Deep
Discharges
152
System Software
Version
2 Software VersionNellcor MP204/205
SpO
Return
V 2.01
V1.1.0.6 10/06/95
This screen displays several system-related items:
• Monitor On-time: Displays the number of hours, rounded to the nearest
hour, that the Main PCB has been operational. This value may not be reset.
(See Note 1.)
NOTE 1:Monitor on-time, backlight on-time, and battery deep
discharge values are stored in nonvolatile memory.
When a new Main PCB is installed, this value will be
set at zero.
•Backlight On-time: Displays the number of hours, rounded to the nearest
hour, that the LCD Backlight has been operational. This value may be reset
to zero, for instance, when a technician changes the backlight or installs a
new LCD. (See Note 1.)
•Battery Deep Discharges: Displays the number of deep-discharge cycles
seen by the battery. The monitor records a deep discharge cycle when the
battery voltage reaches 5.6 V, the voltage at which a “Low Battery” alarm is
issued. This value may be reset to zero, for instance, when a technician
installs a new battery. (See Note 1.)
4-6
•System software version: Displays the revision level of the system
software. This revision level is also momentarily shown on the LCD as part
of the Copyright screen. This value may not be changed by the user.
• SpO
2 software version: Displays the revision level of the software of the
SpO
2 EEPROM module. This value may not be changed by the user.
When in the System Information screen, the knob may be rotated to select any of
the “changeable” items. If one of those items is selected, a press of the knob will
cause a pop-up menu to appear. The first item in the pop-up will read “Make no
change”; the second item in the pop-up will read “Reset to zero”. Exiting the
screen is accomplished in the normal manner, by selecting “Return”.
4.4.3 System A/D Values
g
s
R
R
Section 4: Power-up Defaults Menu and Diagnostic Mode
SYSTEM A/D VALUES
1.1.47812.1.478
2.0.23513.0.235
3.3.65214.3.652
4.0.78215.0.782
5.4.01216.4.012
6.0.04517.0.045
7.2.14918.2.149
8.1.02519.1.025
9.0.47820.0.478
10.1.36921.
11.0.702
2 S1S018SpO2 S2S0010
SpO
Return
This screen displays the current value of each analog-to-digital (A/D) channel, in
volts. Some of the channels are for AC-coupled signals (such as ECG input), so
the numbers on the screen will be constantly changing when an input signal is
present. These AC-coupled values indicate whether basic functionality of the
channel is present, but no significance can be derived from the values of the
numbers displayed. However, other A/D channels read DC voltages, (for
example, power supply voltages and battery voltage) so those voltage values
provide useful diagnostic information.
The Primary and Secondary Status messages from the SpO
2 module will be
displayed and updated at the rate of about once per second. Presence of the
correct SpO
the SpO
2 message indicates that, at a basic level, communication between
2 module and the main monitor processor is working correctly. None of
the displayed values may be changed or reset in this screen. When in the System
A/D screen, the “Return” item is always highlighted; a press of the knob will
return the user to the Diagnostic Menu. Rotating the knob while in the System
A/D screen will have no effect. The A/D channel designators are shown in
Table 4-2.
Section 4: Power-up Defaults Menu and Diagnostic Mode
Table 4-2: A/D Channel Designators - (Continued)
5.PRESSURE XDUCER 116.GROUND REFERENCE
6.PRESSURE XDUCER 217.(+5 VDC POWER SUPPLY) X 0.8
7.NIBP OSCILLATORY18.ADC MID-SCALE VALUE
8.ECG LEADS OFF19.ADC FULL-SCALE VALUE
9.TEMP 2(59° TO 93°)20.ADC ZERO-SCALE VALUE
10.ISOLATED VOLTAGE REF21.(NOT USED)
11.ISOLATED VOLTAGE ZERO
SpO2 S1 S018SpO2 S2 S010
4.4.4 NIPB Test
WARNING: Never apply an attached blood pressure cuff to a patient while
the monitor is in Diagnostic Mode. Injury could result.
NIBP TEST
Pressure
179
(mmHg)
Valve: OPEN
Press “NIBP” to activate pump;
release to stop pump.
Press “Volume” to open valve; release
to close valve.
Press “Alarm Silence” to open
proportional valve and deflate at 3
mmHg/s; release to close valve.
Press “Contrast” to perform offset
adjust.
Return
4-8
An NIBP Test screen is provided to facilitate troubleshooting problems and
perform verification testing for the NIBP subsystem. Typically, when these tests
are performed, the pneumatic system is connected to an external pressure-
Section 4: Power-up Defaults Menu and Diagnostic Mode
reading device and a closed reference volume. The NIBP Test screen provides a
real-time numeric display of the pressure in the pneumatic system, means for
controlling the pump and valve, and a display indicating whether the valve is
open or closed.
The NIBP Test screen elements are described below.
• Pressure Display: The real-time value of the system pneumatic pressure is
displayed in mmHg. The value is updated at the rate of approximately two
times per second.
•Valve Display: A display indicates whether the valve is open or closed.
•Activate pump: While the NIBP button is pressed, the pump will run. If
system pressure reaches the hardware over-pressure protection point (280 to
330 mmHg), the safety valve will open and the pump will be disabled, until
the pressure falls below the safety threshold.
•Deflate: For as long as the Alarm Silence button is pressed, the valve will
open and bleed off pressure at the rate of 3 ±1.5 mmHg/sec. It is useful to
control the bleed rate to 3 mmHg/sec to facilitate certain AAMI SP10 tests.
Any time the bleed rate falls below 3 mmHg/sec, the valve will open and
remain at maximum as long as the button is pressed.
•Open Valve: While the Volume button is pressed, the valve opens and
remains at maximum as long as the button is pressed.
•Offset Adjust: A momentary press of the Contrast button will invoke the
“zero calibration” routine that is performed immediately prior to each blood
pressure measurement. This routine looks at the pressure in the system, and
if the pressure is non-zero, an offset is applied which causes the system
pressure to be displayed as “zero”.
When in the NIBP Test screen, the “Return” item is always highlighted; a press
of the knob will return the user to the Diagnostic Menu. Rotating the knob while
in the NIBP Test screen will have no effect.
4-9
SECTION 5: TROUBLESHOOTING
5.1 Introduction
5.2 How to Use this Section
5.3 Who Should Perform Repairs
5.4 Replacement Level Supported
5.5 Obtaining Replacement Parts
5.6 Troubleshooting Guide
5.7 Troubleshooting the Oximetry Function
5.8 P-3900 Troubleshooting Guide
5.1 INTRODUCTION
This section explains how to troubleshoot the NPB-3900 if problems arise.
Tables are supplied that list possible monitor difficulties, along with probable
causes, and recommended actions to correct the difficulty.
5.2 HOW TO USE THIS SECTION
Use this section in conjunction with Section 3, Performance Verification, and
Section 7, Spare Parts. To remove and replace a part you suspect is defective,
follow the instructions in Section 6, Disassembly Guide. The circuit analysis
section in the Technical Supplement offers information on how the monitor
functions.
5.3 WHO SHOULD PERFORM REPAIRS
Only qualified service personnel should open the monitor housing, remove and
replace components, or make adjustments. If your medical facility does not have
qualified service personnel, contact Nellcor Puritan Bennett Technical Services
or your local Nellcor Puritan Bennett representative.
5.4 REPLACEMENT LEVEL SUPPORTED
The replacement level supported for this product is to the printed circuit board
(PCB) and major subassembly or component level. Once you isolate a suspected
PCB, follow the procedures in Section 6, Disassembly Guide, to replace the PCB
with a known good PCB. Check to see if the trouble symptom disappears and
that the monitor passes all performance tests. If the trouble symptom persists,
swap back the replacement PCB with the suspected malfunctioning PCB (the
original PCB that was installed in the monitor before you started
troubleshooting) and continue troubleshooting as directed in this section.
5-1
Section 5: Troubleshooting
5.5 OBTAINING REPLACEMENT PARTS
Nellcor Puritan Bennett Technical Services provides technical assistance
information and replacement parts. To obtain replacement parts, contact Nellcor
Puritan Bennett or your local Nellcor Puritan Bennett representative. Refer to
parts by the part names and part numbers listed in Section 7, Spare Parts.
5.6 TROUBLESHOOTING GUIDE
Problems with the NPB-3900 are separated into the categories indicated in Table
5-1. Refer to the paragraph indicated for further troubleshooting instructions.
NOTE: Taking the recommended actions discussed in this section will correct
the majority of problems you may encounter. However, problems not
covered here can be resolved by calling Nellcor Puritan Bennett
Technical Services or your local representative.
Problem AreaRefer to
Table 5-1: Problem Categories
Paragraph
1. Power
5.6.1
•No power-up
•Fails power-on self-test
•Powers down without apparent cause
2. Error Messages5.6.2
3. Buttons/Knob
5.6.3
•Monitor does not respond properly to buttons
4. Display/Audible Tones
5.6.4
•Display does not respond properly
•Tones do not sound properly
5. Operational Performance
5.6.5
•Displays appear to be operational, but monitor
shows no readings
•Suspect readings
•Printer not responding
All of the categories in Table 5-1 are discussed in the following paragraphs.
5-2
5.6.1 Power
Section 5: Troubleshooting
Table 5-2 lists recommended actions to address power problems.
Table 5-2: Power Problems
ConditionRecommended Action
1. With external
power supply
connected, the
green
EXTERNAL
POWER
indicator
on the front panel
is not lit.
1. Ensure that the external power supply input (PS120V or PS-240V) is plugged into an operational
AC outlet of the appropriate voltage and frequency.
2. Disconnect the power supply output cable from the
monitor. Measure the voltage across pins 1 and 4 of
output connector. If the open circuit voltage does
not measure approximately 17 ±3 V~ RMS, replace
the power supply.
3. If the battery is severely discharged or shorted, the
EXTERNAL POWER indicator will not light.
Connect the external power supply to an AC outlet
and to the monitor. Allow the battery to charge for
30 minutes. If the EXTERNAL POWER indicator
still does not light, replace the battery.
4. Inside the monitor, check the ribbon cable and
ensure that it is properly connected to the main
PCB.
5. The EXTERNAL POWER indicator is embedded in
the keypad. Ensure that the keypad is plugged into
Main PCB. If the connection is good, replace the
keypad.
2.The NPB-3900
fails to power-up
when the
ON/STANDBY
button is pressed
... with the
monitor
connected to
external power
supply
6. If the problem persists, replace main PCB.
1. Connect the appropriate external power supply (PS120V or PS-240V) to the monitor. Ensure that the
external power supply input is plugged into an
operational AC outlet of the appropriate voltage and
frequency. Ensure that the green EXTERNAL
POWER indicator is lit. If the indicator is not lit,
follow the steps described in Condition 1, above.
2. Ensure that the keypad is plugged into Main PCB. If
the connection is good, replace keypad.
3. If the problem persists, replace the main PCB.
5-3
Section 5: Troubleshooting
Table 5-2: Power Problems - (Continued)
ConditionRecommended Action
3. The NPB-3900
fails to power-up
when the
ON/STANDBY
button is pressed
with the monitor
not connected to
external power
supply.
4. The NPB-3900
turns on, then
shuts off and
sounds an alarm
and no error code
is displayed.
1. First, follow the steps described in Condition 2,
above, to ensure that the monitor will operate when
connected to an external power supply.
2. Check fuse F301 located on the Main PCB, near the
battery cable connector. Replace fuse if necessary.
3. Recharge the battery as directed in paragraph 3.3.1.
If the battery fails to hold a charge, replace the
battery.
4. If the problem persists, replace the main PCB.
1. Press the alarm silence button to terminate the
audible alarm. Ensure that the external power supply
is connected and the green EXTERNAL POWER
indicator is lit. If the monitor operates successfully,
the battery may be discharged, or the battery fuse
may be blown.
2. Recharge the battery as directed in paragraph 3.3.1.
If the battery fails to hold a charge, replace the
battery.
3. Check fuse F301 located on the Main PCB, near the
battery cable connector. Replace the fuse if
necessary.
5.6.2 Error Codes
4. If problem persists, replace the main PCB.
When the NPB-3900 detects an error condition, the monitor shows an error code
on the display screen. If such an error occurs during monitoring operation, an
audible alarm tone will sound, as well. Press the ALARM SILENCE button to
terminate the audible alarm tone.
When an error code appears on the display, a number in hexadecimal
representation indicates the nature of the error. Additionally, Diagnostic Mode
may be used to gain access to an error code record, stored in nonvolatile
memory, of the last 10 error codes encountered by the monitor. See Section 4 for
further details on Diagnostic Mode.
Each error code corresponds to a particular problem in the monitor.
Recommended actions to take when an error code is encountered are listed in the
sections that follow.
As an aid to troubleshooting, the NPB-3900 provides the capability for
technicians to print out a copy of the error log.
5-4
Generating an Error Log printout
1. Connect a P-3900 printer to the monitor, and its power supply to an
appropriate source. Refer to the operator’s manual.
2. Use the Setup button and displayed menu to verify that the Printer mode is
the selected option for the Communications Selection item. (It is the factoryset default value.)
3. Turn monitor power OFF.
4. Simultaneously press the contrast button and the On/Standby button to
power up the monitor. Keep the contrast button depressed until the
monitoring screen appears (after 10 seconds). The error code printout is
generated automatically.
If error codes listed on the Diagnostic Mode error code screen or on the error log
printout are in the range from 1 to 65 (hex), a hardware problem has been
detected. Refer to Table 5-3, Serviceable Hardware Codes for additional
information on these codes.
5.6.2.1 Serviceable Hardware Error Codes
Section 5: Troubleshooting
In Table 5-3 are error codes that correspond to hardware problems, and the
recommended actions to take should such an error be encountered.
Table 5-3: Serviceable Hardware Error Codes
Hex
ExplanationRecommended Action
Code
1Improper shutdown.1. Cycle power.
2. If this error persists, return
monitor for service.
2NIBP Sensor Error. The
two pressure
1. Check for blocked hoses in the
pneumatic system.
transducers do not
agree.
3
NIBP Pressure Violation
Error. The pressure on
the cuff could not be
removed by normal
means. A fault has been
detected in the NIBP
system that could not be
handled by releasing
pressure by normal
means.
2. Replace Main PCB.
1. Cycle power.
2. Check for blocked hoses in the
pneumatic system.
detected on the powerup settings region of
Flash memory.
64The SpO2 module is
sending an error
messages to the host
CPU.
65The SpO2 module is
not communicating
with the host CPU.
1. Turn Power Off
2. Turn power back on while
pressing both the Contrast and
Volume buttons. See Section 4.3.
3. All user selections must be
restored.
4. If error persists, replace Main
PCB.
Cycle power. If problem persists,
replace Main PCB.
Cycle power. See Section 5.7.
If problem persists, replace main
PCB.
5.6.2.2 Other Error Codes
If an error code occurs that is not listed in Section 5.6.2.1, take the following
actions:
1. Turn the monitor off, then on again.
2. If the error code still appears, take the monitor out of service and contact
Nellcor Puritan Bennett Technical Services or your local Nellcor Puritan
Bennett representative for advice on remedial action.
3. If the monitor powers up and the error code does not recur, enter the
Diagnostic Mode and invoke the Error Code screen. Examine the record of
the last 10 error codes and determine if the same error code occurred
previously.
4. If the Error Code screen indicates that the same error has occurred
previously, take the monitor out of service and contact Nellcor Puritan
Bennett Technical Services or your local Nellcor Puritan Bennett
representative for advice on remedial action.
5. If the Error Code screen indicates no previous occurrences of this error, the
monitor may be returned to service.
5-7
Section 5: Troubleshooting
As a reference, Table 5-4 lists the general categories for other error codes. The
error code categories are shown only in hexadecimal format.
Table 5-4: Error Code Categories
Code (hex)Explanation
500xxxxinternal user interface error
501xxxxremote serial port error
502xxxxdate and time error
503xxxxNIBP error
504xxxxfront end error
505xxxxalarm error
506xxxxaudio error
507xxxxrecorder error
508xxxxtrend error
Table 5-5 lists recommended actions to address problems with the knob and
front-panel buttons.
5-8
Section 5: Troubleshooting
Table 5-5: Buttons/Knob Problems
ConditionRecommended Action
1. The NPB-3900 fails to
power-up when the
ON/STANDBY button is
pressed.
2. The NPB-3900 powers-up,
but some/one of the other
buttons does not respond.
3. When the knob is rotated,
no highlight appears on the
display screen, and/or the
monitor does not respond to
knob presses.
5.6.4 Display/Audible Tones
Table 5-6 lists recommended actions to address problems with the display and
audible tones.
Take steps as noted in section 5.6.1.
1. Ensure that the keypad is plugged into
the Main PCB. If the connection is
good, change the keypad.
2. If the problem persists, change the Main
PCB.
1. Ensure that the encoder cable is plugged
into the Main PCB. If the connection is
good, change the encoder.
2. If the problem persists, replace the Main
PCB.
Table 5-6: Display/Audible Tones Problems
ConditionRecommended Action
1. System powers-up
and…
•LCD screen is
totally black or white.
Or,
•LCD screen is
illuminated, but no
data is visible.
Or,
•LCD screen has
data, but is not
illuminated.
1. Adjust the LCD screen contrast by pressing
the contrast button momentarily, then turning
the knob four revolutions in each direction.
Turning the knob clockwise should brighten
the screen; turning the knob counterclockwise should darken the screen.
2. Ensure that the backlight cable is connected
to the main PCB.
3. Ensure that the LCD connector is properly
connected to the main PCB.
4. If problem persists, replace main PCB.
5. If problem persists, replace LCD assembly.
5-9
Section 5: Troubleshooting
Table 5-6: Display/Audible Tones Problems - Continued
ConditionRecommended Action
2. NPB-3900 responds
to button press, but
key press tone fails to
sound.
3. Audible alarm does
not sound.
1.Ensure that the speaker cable is connected to
the main PCB.
2.If the problem persists, replace the speaker
assembly.
3.If the problem persists, replace the Main
PCB.
1. Verify alarm volume setting in the
Alarm/Limits menu, and test operation of the
alarm tone by pressing the volume button
while the alarm volume setting is displayed.
2. Ensure that the speaker cable is connected to
the Main PCB.
3. If the problem persists, replace the speaker
assembly.
4. If the problem persists, replace the Main
PCB.
5.6.5 Operational Performance
Table 5-7 lists recommended actions to address problems related to operational
performance.
ConditionRecommended Action
1.The monitor appears to
be operational, but the
physiological values are
suspect or nonexistent.
Table 5-7: Operational Performance Problems
1.Replace each patient cable (or hose) with a
known-good cable.
2. Ensure that the Patient Connector PCB is
properly connected to the main PCB.
Ensure that the hoses in the pneumatic
system are properly connected, and that the
NIBP pump motor is connected to the Rear
Connector PCB.
3. If the problem persists, replace the Patient
Connector PCB.
4.If the problem persists, replace Main PCB.
5-10
5.7 TROUBLESHOOTING THE OXIMETRY FUNCTION
5.7.1 Introduction
The oximetry functional hardware is embedded on the NBP-3900 main printed
circuit board. (Note that “oximetry” and “SpO
manual.) This section assumes that the NBP-3900 has been thoroughly checked
and that all indications point to an oximetry malfunction.
5.7.2 Fault Evaluation
Table 5-8 provides fault indications and possible solutions.
Table 5-8: Fault Evaluation
IndicationAction
Section 5: Troubleshooting
2” are used interchangeably in this
NPB-3900 gives error code
64 or 65, indicating
communications problem
w/ SpO2 module.
Check 5-volt digital power supply.
Check processor clock Y1.
Check TXD buffer U5.
Sensor may be disconnected or damaged.
Patient Connector PCB may be defective.
The sensor may be improperly applied to
the patient or may be damaged. Try
another sensor. Try an SRC-2 pulse
oximeter tester to check oximetry
functionality.
Cover the sensor to eliminate the
possibility of ambient light interference.
The patient’s perfusion may be too poor
for the instrument to detect an acceptable
pulse. Try using C-Lock ECG
synchronization, if available.
Check + and –5 volt analog power
supplies.
Waveform output incorrect
Check for proper LED drive function.
Check the signal path from the
photodetector input to the A:D converter.
Sensor or interconnecting cables may be
damaged.
Noise may be present.
5-11
Section 5: Troubleshooting
5.7.3 Waveforms
Figures 5-1 through 5-5 are typical waveforms as measured at various test points
(labeled TP) on the oximetry module. These waveforms are valuable in tracing
signals and locating faults. The user must use a Nellcor Puritan Bennett SRC-2
pulse oximeter tester. Contact Nellcor Puritan Bennett’s Technical Services
Department or your local Nellcor Puritan Bennett representative if you have
difficulty replicating these waveform examples.
TP5 ADC Input
TP6 Red Filter Output
TP2 IR Filter Output
5-13
Section 5: Troubleshooting
5.7.3.3 SpO2 Module with an SRC-2 Waveform
5.7.3.3.1 SRC-2 Settings
Rate:
Light:
Modulation:
RCAL/MODE:
112
High2
HIGH
RCAL 63/LOCAL
5.7.3.3.2 Trace Descriptions
Figure 5-3: SpO2 Module with an SRC-2
CHNL 2 :
CHNL 1 :
TP6 Red Filter Output
TP2 IR Filter Output
5-14
Section 5: Troubleshooting
5.7.3.4 SpO2 Module with an SCR-2 LED Drive Current Test at TP7 Waveform
5.7.3.4.1 SRC-2 Settings
Rate:
Light:
Modulation:
RCAL/MODE:
112
High2
HIGH
RCAL 63/LOCAL
Figure 5-4: SpO2 Module with SRC-2 Drive Current Test at TP7
5-15
Section 5: Troubleshooting
5.7.3.5 SpO2 Module with SCR-2 Waveform
5.7.3.5.1 SCR-2 Settings
Rate:
Light:
Modulation:
RCAL/MODE:
112
High2
HIGH
RCAL 63/LOCAL
5.7.3.5.2 Trace Description
Serial Port TXD Signal, U4 Pin 25
Figure 5-5: SpO2 Module with SRC-2
5-16
5.8 P-3900 TROUBLESHOOTING GUIDE
Table 5-9 lists recommended actions to address printer problems.
Table 5-9: P-3900 Troubleshooting Guide
ConditionRecommended Action
Section 5: Troubleshooting
1. With external
power supply
connected, the
green
EXTERNAL
POWER
indicator on the
front panel is not
lit.
1. Ensure that the external power supply input (PS120V or PS-240V) is plugged into an operational
AC outlet of the appropriate voltage and frequency.
2. Disconnect the power supply output cable from the
printer. Measure the voltage across pins 1 and 4 of
output connector. If the open circuit voltage does
not measure approximately 17 ±3 V~ RMS, replace
the power supply.
3. If the battery is severely discharged, the
EXTERNAL POWER indicator may not light.
Connect the external power supply to an AC outlet
and to the printer. Allow the battery to charge for 30
minutes.
4. Open the printer enclosure and ensure that all
connectors are properly seated.
5. If the problem persists, replace the battery.
6. If the problem persists, replace the Printer PCB.
7. The EXTERNAL POWER indicator is embedded in
the Front Panel. If the problem persists, replace the
Front Panel.
2. No printout
occurs when
control buttons
are pressed on
P-3900 front
panel.
1. Open the printer paper door and verify that the
printer is properly loaded with paper.
2. Ensure that the green LINKED indicator is lit. If not,
first determine that the serial cable is properly
connected between printer and NPB-3900 monitor.
Next, on the NPB-3900 monitor, check the “Set-up
Menu” and verify that “Communications Selection”
is set to “Printer”. If the indicator is still not lit,
replace the serial cable with a known-good cable.
3. Connect a known-good external power supply to an
AC outlet and to the printer. If both the green front
panel indicators are lit (EXTERNAL POWER and
LINKED), then the printer should operate. If the
EXTERNAL POWER indicator still fails to light,
follow the steps outlined in Condition 1.
4. Open the printer enclosure and ensure that all
connectors are properly seated.
5. If the problem persists, replace the Printer PCB.
3. Printer paper will
advance, but
paper remains
blank when
printing should
be present.
6. If the problem persists, replace the printer
mechanism.
1. Open the printer door and verify that paper is
oriented correctly. The PAPER icon adjacent to the
door release button illustrates proper orientation of
the paper roll. (The paper is thermally sensitive on
one side only; if the roll is installed backwards,
printing will not occur.)
2. If the problem persists, replace the printer
mechanism.
5-18
SECTION 6: DISASSEMBLY GUIDE
6.1 Introduction
6.2 How to Use this Section
6.3 Disassembly Flow Charts
6.4 Closed Case Disassembly
6.5 Front Case Disassembly
6.6 Rear Case Disassembly
6.7 Main PCB Disassembly
6.1 INTRODUCTION
WARNING: Performance Verification. Do not place the NPB-3900 into
operation after repair or maintenance has been performed, until all
Performance Tests and Safety Tests listed in Section 3 of this service
manual have been performed. Failure to perform all tests could result in
erroneous monitor readings.
The NPB-3900 can be disassembled down to all major component parts,
including:
•PCBs
•battery
•cables
•function buttons
•chassis enclosures
The following tools are required:
WARNING: Before attempting to open or disassemble the NPB-3900,
disconnect the power supply from the NPB-3900.
WARNING: High voltage is generated by the LCD backlight driver.
Exercise caution when operating monitor with covers open.
Caution: Observe ESD (electrostatic discharge) precautions when working
within the unit.
6-1
Section 6: Disassembly Guide
Caution: If internal battery cable has been disconnected, pay particular
attention to polarity of the cable before reattaching. If battery cable
polarity is reversed, it is likely that circuit damage will occur.
NOTE: Some spare parts have a business reply card attached. When you receive
these spare parts, please fill out and return the card.
6.2 HOW TO USE THIS SECTION
The step-by-step procedures that are used to access replaceable parts of the
NPB-3900 are illustrated in the Disassembly Flow Charts in paragraphs 6.3,
Figures 6-1, 6-2, 6-3 and 6-4. As indicated in the flow charts, the monitor
consists of two main assemblies, the Front Case Assembly, and Rear Case
Assembly. The Main PCB assembly is separable from the front case assembly.
The circles on the flow charts contain reference designators that point to specific
steps in the Disassembly Procedures. The Disassembly Procedures, paragraphs
6.4, 6.5, and 6.6 contain detailed disassembly instructions, accompanied by
illustrations.
The rectangular boxes on the flow charts represent the various spare components
or subassemblies. The digits appearing in these boxes are their respective part
numbers. Section 7, Spare Parts, contains a complete listing of the available
spare parts.
The Disassembly Flow Charts are organized so that minimum disassembly is
required to remove and replace defective items.
Further important disassembly information may be found in the diagrams in the
Appendix.
6.3 DISASSEMBLY FLOW CHARTS
The charts have been developed for use with all models of the NPB-3900 family.
Therefore, some disassembly procedures will not be applicable to model
configurations of less than the full complement of functions. In most cases, the
relevant subassemblies will not have been installed, and it will be apparent that
the pertinent procedures will not apply. However, when a temperature-measuring
function is involved, a different hardware configuration of the rear-case
assembly is shown in the flow chart.
The charts have been developed to provide service personnel with the most
direct route to a replaceable item after a troubleshooting analysis has led to a
probable cause, traced to hardware sources.
6-2
A1A3
y
Battery
)
048936
g
048939
048938
Switch
048992
y
y
y
y
y
y
Batter
Cover
048935
A2
A4
Batter
048987
Batter
Pads
048937
Temp
Model
?
NPB-3900
YN
Cover(Temp
Temp
Housin
Temp
Grommet
Probe
Section 6: Disassembly Guide
B1
SpO2
Hood
048942
B2
Rear Cover
Gasket
048991
Rear Case
Assembl
Fig. 6-3
B3
Batter
Cable
048940
Front Case
Assembl
Fig. 6-2
Main PCB
Assembl
Fig. 6-4
Figure 6-1: Top Level Disassembly Flow Chart
6-3
Section 6: Disassembly Guide
y
Display
Shield
048944
C1
C2
Front Case
Assembl
C3
Spring
Retainer
048990
Speaker
048948
C4
LCD
Display
048943
Display
Window
048945
Front
Cover
048947
Encoder
291186
Knob
044727
C5
Keypad
048946
Figure 6-2: Front Case Disassembly Flow Chart
6-4
y
Rear Conn.
PCB
048956
D1
D3
Section 6: Disassembly Guide
Rear Case
Assembl
D2
Pump
Clamp
048989
NIBP
Pump
048949
Handle
048941
Foot
Cushion
048950
Rear
Cover
048951
Figure 6-3: Rear Case Disassembly Flow Chart
6-5
Section 6: Disassembly Guide
y
NIBP
Pneumatics
048957
E1
Main PCB
Assembl
E2
Main
PCB
048952…54
Fuse
F301
048970
E3
Figure 6-4: Main Board Disassembly Flow Chart
Patient
Conn. PCB
048962…65
Connector
Panel
048966…69
6-6
6.4 CLOSED CASE DISASSEMBLY
The paragraphs in this section describe and photographically illustrate
procedures for disassembling the NPB-3900 to enable removal and replacement
of suspected defective assemblies/components.
The sequence supports the guides in the previous paragraphs of this section.
The illustrations may also contain juxtaposed photographs of the relevant spares.
Section 6: Disassembly Guide
See Figure 6-1. If there is no apparent reason to replace the battery, begin with
procedure B1. If the battery needs replacement, and there is no temperature
module (Models 3910 and 3930), begin with procedure A1.
6-7
Section 6: Disassembly Guide
Step A1
Procedure
To remove the Battery from Models 3910 and 3930, when a Temperature
module is not installed:
Use a Phillips head screwdriver to remove the two screws fastening the battery
cover to the rear case.
Remove the battery cover.
Illustration
Battery
Pad
6-8
Section 6: Disassembly Guide
Step A2
Procedure
Disconnect the spade terminal connectors from the battery terminals.
Caution: Pay particular attention to polarity of the battery cable before
reattaching. If battery cable polarity is reversed, it is likely that circuit
damage will occur.
Remove battery.
As required, remove the battery cushions on the inside of the battery
compartment and battery cover.
Illustration
Foot
Cushions
Battery
Cover
Step A3
Procedure
To remove the battery from Models 3920 and 3940, when a temperature
module is installed:
Illustration
Temp
Rear View
Models
3920, 3940
Module
6-9
Section 6: Disassembly Guide
Step A3
Procedure
Use a Phillips head screwdriver to remove the two screws fastening the
temperature module housing to the rear case assembly.
Illustration
Step A3
Procedure
Remove the temperature module housing and the battery cover.
Illustration
6-10
Section 6: Disassembly Guide
Procedure
The probe sensor contact switch is mounted on the outer side of the battery
cover.
As required, disconnect the leads from the probe switch to the rear panel
connector. Remove the switch.
Illustration
Temperature
Probe
Probe
Sensor
Switch
Grommet
Spare Probe
Sensor Switch
Step A4
Procedure
Disconnect the spade terminal connectors from the battery terminals.
Caution: Pay particular attention to polarity of the battery cable before
reattaching. If battery cable polarity is reversed, it is likely that circuit
damage will occur.
Remove battery.
As required, remove the battery cushions on the inside of the battery
compartment and battery cover.
Illustration
Battery Cover
Plate
Foot
Cushions
6-11
Section 6: Disassembly Guide
Step B1
Procedure
To separate the Front and Rear Case Assemblies:
Remove the SpO
2 connector hood by squeezing the sides to release the detents
holding the hood in place.
Illustration
SpO2
Connector
Hood
Step B2
Procedure
Use a Phillips head screwdriver to remove the four screws fastening the
front to rear case assemblies.
6-12
Separate the main front and rear case assemblies.
If the rear cover gasket seal is to be replaced, remove it.
Illustration
Rear Case
Assembly
Front Case
Assembly
Section 6: Disassembly Guide
Step B2
Procedure
Disconnect the ribbon cable and its connector from the PCB assembly.
Illustration
Ribbon Cable
and Connector
Procedure
To remove the Main PCB Assembly from the Front Case Assembly:
Use a Phillips head screwdriver to remove screws holding the Main PCB
Assembly in place in the Front Cover Assembly.
For reassembly, note the two guides in the cover and the corresponding two
notches in the PCB.
Illustration
6-13
Section 6: Disassembly Guide
Step B3
Procedure
Disconnect the remaining connectors at the Main PCB.
Disconnect the battery cable spade terminals from the main PCB assembly. If
the battery cable is to be removed, the cable must also be disconnected from the
battery, as described in Procedure A.
Caution: Pay particular attention to polarity of the battery cable before
reattaching. If battery cable polarity is reversed, it is likely that circuit
damage will occur.
Illustration
Luer Connector
to NIBP pump
tubing
Battery
Cable
6-14
Section 6: Disassembly Guide
Step B3
Procedure
Unscrew the NIBP Luer connector. See Illustration in Step B2.
There are now three separate items:
Front Cover Assembly
Rear Case Assembly
Main PCBPCB Assembly
Illustration
Rear Case
Assembly
Main PCB
Assembly
Front Case
Assembly
6-15
Section 6: Disassembly Guide
6.5 FRONT CASE DISASSEMBLY
See Figure 6-2.
Step C1
To remove the Display:
Use a Phillips head screwdriver to unfasten the four corner screws and remove
the display shield.
The four screws also hold the LCD assembly in place.
Speaker
Connector
Procedure
Illustration
Backlight
Connector
Keypad Cable
and Connector
Knob Encoder
Assembly
6-16
Section 6: Disassembly Guide
Step C2
Procedure
Remove the LCD Assembly, providing access to the Display Window.
Remove Display Window by carefully prying up one corner, then peeling back.
Illustration
Display
Window
Shield
LCD Assembly
Speaker
Spare
Display
Window
6-17
Section 6: Disassembly Guide
y
Step C3
Procedure
To remove the speaker:
Remove the retaining spring clip.
Remove the speaker.
Illustration
Spare Speaker,
Leads Attached
Step C4
Procedure
To remove the NPB Knob and Encoder:
From the front, remove the knob by grasping the sides of the knob firmly and
pulling straight back from the monitor. (The knob is friction fit on the stem of
the encoder assembly.)
Use a 9/16” hex socket to unscrew the fastening nut on the outside of the front
case. The encoder may now be pulled away from the front case.
Illustration
Nut
Spare Knob
Encoder
Assembl
6-18
Section 6: Disassembly Guide
Step C5
Procedure
To remove the Keypad
The keypad is attached with an adhesive to the front panel. From the front side
of the panel, carefully pry up one corner of the keypad from the cover, and peel
away from the cover.
Carefully, thread the cable out through the slot in the cover.
Illustration
Keypad,
Integral Cable
and Connector
Spare Knob
6-19
Section 6: Disassembly Guide
6.6 REAR CASE DISASSEMBLY
See Figure 6-3.
Step D1
To remove a Rear Connector PCB:
Use a Phillips head screwdriver to remove the two screws holding the Rear
Connector PCB to the rear cover.
From outside the rear cover, use 3/16 socket driver to remove the two standoff
fasteners of the RS-232 connector.
From inside the rear cover, remove the Rear Connector PCB Assembly.
NIBP Pump,
Hose and
Lead Attached
Procedure
Illustration
Rear Connector
Board Assembly
6-20
Section 6: Disassembly Guide
Step D2
Procedure
To remove NIBP Pump:
Use Phillips head screwdriver to unfasten screw holding clamp to rear cover.
Disconnect power lead from Rear Connector PCB.
Remove Clamp and Pump.
Illustration
Pump
Power Lead
NIBP
Pump
Step D3
Procedure
To remove Handle and Foot Cushions:
Each end of the handle is friction-fit onto a cross-shaped boss. Use flat-bladed
screwdriver to carefully pry one end of the handle.
When the end of the handle has begun to loosen from the boss, use the same
technique to begin to pry up the other end.
Alternate prying action between each end of the handle until the handle is free of
the rear case.
Foot cushions are attached with an adhesive to the bottom surface of the rear
cover, and can be removed by lifting one end of the foot and peeling off.
6-21
Section 6: Disassembly Guide
6.7 MAIN PCB DISASSEMBLY
See Figure 6-4.
Step E1
To remove NIBP Pneumatic Assembly from the Main PCB Assembly:
Pull tubing from barbed fitting on rear of NIBP panel connector.
Pull tubing from fittings on the pressure sensors and valve.
Pressure
Sensor
Procedure
Illustration
NIBP
Pneumatic
Assembly
Group
Step E2
Procedure
To separate the Patient Connector PCB Assembly from the Main PCB
Assembly:
Use wire cutters to remove the two Tinnerman fasteners securing the Patient
Connector PCB Assy to the underside of the Main PCB.
Disconnect the Patient Connector PCB Assy by pulling it straight up from the
Main PCB.
The battery fuse F301, located near the battery cable connector on the Main
PCB, may be replaced if necessary.
Illustration
Patient
Connector
PCB
Main PCB
Assembly
Assembly
Foot of Molded
Connector Panel
after removal of
Tinnerman nut
6-22
Section 6: Disassembly Guide
Step E3
Procedure
To separate the Patient Connector PCB from the Connector Panel:
Use Phillips head screwdriver to remove four screws fastening the two
assemblies together. NOTE: Two of the screws are accessible on the face of the
Connector PCB; two are accessible through access holes in the PCB.
Illustration
Connector
Panel
Patient
Connector
PCB
6-23
SECTION 7: SPARE PARTS
7.1 Introduction
7.2 Top Level Assembly
7.3 Front Case Assembly
7.4 Rear Case Assembly
7.5 Main PCB Assembly
7.6 P-3900 Printer
7.1 INTRODUCTION
Spare parts, along with part numbers, are listed in the tables that follow. “Item
No.” corresponds to the callout numbers in Figures A 29 through A-31 that are
found in the Appendix. The “Step Ref.” corresponds to the disassembly
procedures described in Section 6.
Note: The Main PCB for all of the NPB-3900 models has jumpers which must
be set correctly so that the User Interface software is configured to
support the measuring parameters of each particular model . There are
two jumpers, marked “JP101” and “JP102”, located immediately below
the NIBP valve. Ensure that the jumpers are installed as noted in the table
which follows:
Spare parts, along with part numbers, are listed in the table that follows. “Item
No.” corresponds to the callout numbers in Figure A-32 found in the Appendix.
Table 7-5: P-3900 Printer
Item
No.
DescriptionNPB
Part No.
1Battery, 12 V Rechargeable048972
2PCB Assembly048971
3Chassis048973
4Cover048974
5Cable, Battery to PCB048976
6Cable, Power Conn. to PCB048977
7Panel, Front048978
8AR-42 Printer Mechanism (complete)048979
9Door Assembly048980
Pack the monitor carefully. Failure to follow the instructions in this section may
result in loss or damage to the monitor. If the original shipping carton is not
available, use another suitable carton. North American customers may call
Nellcor Puritan Bennett Technical Services to obtain a shipping carton.
Prior to shipping the monitor, contact Nellcor Puritan Bennett’s Technical
Services Department or your local Nellcor Puritan Bennett representative for a
returned goods authorization (RGA) number. Mark the shipping carton and any
shipping documents with the RGA number. European customers not using RGA
numbers, should return the product with a detailed, written description of the
problem.
Return the monitor by any shipping method that provides proof of delivery.
To pack the monitor for return, disconnect all cables. It is not necessary to return
sensors, patient cables, NIBP hose and cuff, temperature probe, or external
power supply.
8.2 REPACKING IN ORIGINAL CARTON
If available, use the original carton and packing materials. Pack the monitor as
follows:
1. Place the monitor in original packaging.
2. Place in shipping carton and seal carton with packaging tape.
3. Label carton with shipping address, return address and RGA number, if
applicable.
8.3 REPACKING IN A DIFFERENT CARTON
If the original carton is not available, use the following procedure to pack the
monitor:
1. Place the monitor in a plastic bag.
2. Locate a corrugated cardboard shipping carton with at least 200 pounds per
square inch (psi) bursting strength.
3. Fill the bottom of the carton with at least 2 inches of packing material.
4. Place the bagged unit on the layer of packing material and fill the box
completely with packing material.
5. Seal the carton with packing tape.
8-1
Section 8: Packing for Shipping
6. Label the carton with the shipping address, return address, and RGA
Operating:5 to 95% RH, non-condensing
Storage:5 to 95% RH, non-condensing
Water Resistance:
IEC 529 Classification IPX1 (Protected
against vertically dripping water)
Altitude:
0 - 10,000 ft ( 0 - 3050 m)
Electromagnetic
Compatibility
9.5 MEASURING PARAMETERS
9.5.1 ECG Measurement/Display
Heart Rate Range:
Heart Rate Accuracy:
9-2
Radiated and conducted electromagnetic
energy per CISPR 11, Class B
20 - 250 BPM
±5 BPM
Bandwidth:
Normal Monitoring:0.5 Hz to 40 Hz
Section 9: Specifications
Extended Low Frequency
Response:
Leads:
Display Sweep Speeds:
Pacemaker Detection:
ECG Size (sensitivity):
Lead Off Detection:
Input Impedance:
CMMR (common mode
rejection ratio):
Input Dynamic Range:
Defibrillator Discharge
Recovery:
Standards:
0.05 Hz to 40 Hz (user selectable)
3 Lead (user selectable)
12.5, 25, and 50 mm/sec
Indicator on waveform display (user
selectable)
0.5, 1, 2, 4 mV/cm
Detected and displayed
> 5 MΩ
> 90 dB at 50 Hz or 60 Hz
±5 mV AC, ±300 mV DC
<5 sec per IEC 601-2-27
<8 sec per AAMI EC13-1992
Meets the performance standards of
ANSI/AAMI EC13-1992. Instead of a 1 mV
standardizing voltage (section 3.2.2.9), a
fixed, 1 cm reference bar is always present in
the ECG display, along with the ECG size
setting expressed in mV/cm. The following
information references particular sections of
ANSI/AAMI EC13-1992.
Leads-off sensing waveform.
Section 3.1.2.1(b)
Tall T-wave rejection.
Section 3.1.2.1(c)
Heart rate averaging
Section 3.1.2.1(d)
Response to irregular rhythm.
Section 3.1.2.1(e)
Applied currents less than 0.25 microamps.
T-wave of 0.6 mV amplitude will not affect
heart rate determination.
Averages six of the most recent eight detected
R-R intervals excluding the longest and
shortest of the eight intervals.
a) Ventricular bigeminy: the NPB-3900
counts both large and small QRS complexes
to display a rate of 80 bpm.
9-3
Section 9: Specifications
b) Slow alternating ventricular bigeminy: the
NPB-3900 inconsistently counts the large Twave following the first Q-wave and the
smaller QRS complexes, Thus causing the
rate to vary between 38 and 80. With slightly
smaller T-wave, the rate was 30 bpm
consistently.
c) Rapid alternating ventricular bigeminy: the
NPB-3900 generally counts only the first
QRS complex of each pair to display a rate of
60 bpm, with infrequent counting of the
second complex, resulting in a momentary
increase to 70 bpm.
d) Bi-directional systoles: the NPB-3900
counts both the positive and negative phases
of the large complexes due to the long
interval between them. It also counts the
small complexes, for an averaged heart rate of
135 bpm, with variation between 127 and 157
bpm due to inconsistent counting.
Heart rate meter response
time.
Section 3.1.2.1(f)
Time to alarm for
tachycardia. 3.1.2.1(g)
a) Change from 80 to 120 BPM: 3 sec
b) Change from 80 to 40 BPM: 7 sec
Waveform 4(a)
Amplitude
0.5 mV
1mV
2mV
Waveform 4(b)
Amplitude
1 mV
2mV
4mV
Pacemaker pulse rejection without over/undershoot. 3.1.4.1
a. For single (ventricular-only) pacemaker pulses alone, with 0.1 and 2.0
ms. pulse-widths and ±2 mV and ± 700 mV pulse-amplitudes, the
NBP3900 correctly displays heart rate as zero bpm (Asystole).
9-4
Section 9: Specifications
b) For single (ventricular-only) pacemaker pulses with normally paced QRS-
T, with 0.1 and 2.0 ms. pulse-widths and ±2 mV and ± 700 mV pulseamplitudes, the NBP3900 correctly displays heart rate of the QRS-T
rhythm (60 bpm for the specified test waveform).
c) For single (ventricular-only) pacemaker pulses with ineffectively paced
QRS pattern, with 0.1 and 2.0 ms. pulse-widths and ±2 mV and ± 700 mV
pulse-amplitudes, the NBP3900 correctly displays heart rate of the
underlying QRS-T rhythm (30 bpm).
d) For atrial/ventricular pacemaker pulses alone, with 0.1 and 2.0 ms. pulse-
widths and ±2 mV and ± 340 mV pulse-amplitudes, the NBP3900
correctly displays heart rate of zero bpm (Asystole).
e) For atrial/ventricular pacemaker pulses with normally paced QRS-T, with
0.1 and 2.0 ms. pulse-widths and ±2 mV and ± 700 mV pulse-amplitudes,
the NBP3900 correctly displays heart rate of the QRS-T rhythm (60 bpm),
except for the case of 2.0 ms width and -700 mV amplitude, which causes
a displayed heart rate of 120 bpm.
f) For atrial/ventricular pacemaker pulses with ineffectively paced QRS
pattern, with 0.1 and 2.0 ms. pulse-widths and ±2 mV and ± 340 mV
pulse-amplitudes, the NBP3900 correctly displays heart rate of the
underlying QRS-T rhythm (30 bpm).
Pacemaker pulse rejection with over/undershoot. 3.1.4.2
a) For single (ventricular-only) pacemaker pulses alone, with 0.1 and 2.0 ms.
pulse-widths and ±2 mV and ± 120 mV pulse-amplitudes, the NBP3900
correctly displays heart rate of zero bpm (Asystole), except for cases with
2.0ms width and ± 2mV amplitude and 25% over/undershoot (time
constant t0 is 55 ms.) which cause a displayed heart rate of 60 bpm.
b) For single (ventricular-only) pacemaker pulses with normally paced QRS-
T, with 0.1 and 2.0 ms. pulse-widths and ±2 mV and ± 700 mV pulseamplitudes, the NBP-3900 correctly displays heart rate of the QRS-T
rhythm (60 bpm), except for the case of 2.0 ms width and +700 mV
amplitude, which causes a displayed heart rate of 120 bpm.
c) For single (ventricular-only) pacemaker pulses with ineffectively paced
QRS pattern, with 0.1 and 2.0 ms. pulse-widths and ±2 mV and ± 120 mV
pulse-amplitudes, the NBP3900 correctly displays heart rate of the
underlying QRS-T rhythm (30 bpm).
d) For atrial/ventricular pacemaker pulses alone, with 0.1 and 2.0 ms. pulse-
widths and ±2 mV and ± 120 mV pulse-amplitudes, the NBP3900
correctly displays heart rate of zero bpm (Asystole), except for cases with
2.0ms width and ± 2mV amplitude and 25% over/undershoot (time
constant t0 is 55 ms.) which cause a displayed heart rate of 60 bpm.
9-5
Section 9: Specifications
e) For atrial/ventricular pacemaker pulses with normally paced QRS-T, with
0.1 and 2.0 ms. pulse-widths and ±2 mV and ± 700 mV pulse-amplitudes,
the NBP3900 correctly displays heart rate of the QRS-T rhythm (60 bpm
for the specified test waveform).
f) For atrial/ventricular pacemaker pulses with ineffectively paced QRS
pattern, with 0.1 and 2.0 ms. pulse-widths and ±2 mV and ±120 mV pulseamplitudes, the NBP3900 correctly displays heart rate of the underlying
QRS-T rhythm (30 bpm).
Systolic:60 to 250 mmHg
Mean Arterial Pressure:30 to 235 mmHg
Diastolic:20 to 220 mmHg
Pulse Rate Range:
Blood Pressure
Accuracy:
40 to 200 BPM
Mean error and standard deviation per
ANSI/AAMI SP10, 1992
Pulse Rate Accuracy:
Standards:
9.5.3 Temperature Measurement/Display
Technique:
Range:
Accuracy:
Measurement Time:
9-6
Greater of ±2 BPM or ±2% of pulse rate
value
Meets performance standards of ANSI/AAMI
SP10-1992
Welch-Allyn SureTemp
o
84
F to 108oF (28.9oC to 42.2oC)
o
±0.2
F, (±0.1°C)
®
Thermistor Probe
Oral - approximately 4 seconds
Rectal - approximately 15 seconds
9.5.4 SpO2 Measurement/Display
Range:
Pulse Rate:20–250 BPM
% Saturation:0–100%
Accuracy:
Pulse Rate:±3 BPM
2:70–100%: ±2 digits
SpO
Accuracies are expressed as plus or minus “X” digits (saturation percentage
points) between saturations of 70-100%. This variation equals plus or minus
one standard deviation (1SD), which encompasses 68% of the population. All
accuracy specifications are based on testing the subject monitor on healthy
adult volunteers in induced hypoxia studies across the specified range. Adult
accuracy is determined with Oxisensor II D-25 sensors. Accuracy for neonatal
readings is determined with Oxisensor II N-25 sensors. In addition, the
neonatal accuracy specification is adjusted to take into account the theoretical
effects of fetal hemoglobin in neonatal blood on oximetry measurements.
Section 9: Specifications
0–69% Unspecified
Pulse Rate (optically derived)
Accuracies are expressed as plus or minus “X” bpm across the display range.
This variation equals plus or minus 1 Standard Deviation, which encompasses
68% of the population.
9.6 TRENDS
Type:
Memory Storage:
Data interval:
Tabular Format:
Display interval:
9.7 P-3900 PRINTER (OPTIONAL)
Type:
Size:
20–250 bpm ±3 bpm
Tabular
12 hours, nonvolatile
20 seconds: (Stored data point is the average
over 20-second interval)
One table for all variables
Six fields per row (time and 5 vital signs)
Per NIBP measurement, or 15 minutes for no
The 9-pin connector mounted on the rear panel provides an access port for a
serial (RS-232) interface to the P-3900 Printer, or to a suitably configured
personal computer. Alternatively, qualified service personnel can use the
connector to send a Nurse Call signal.
NOTE: The “Communications Selection” item in monitor’s “Set-up Menu”
must be set to “Printer” if the P-3900 is to be used; or, must be set to
“Trend Xfer” if trend data is to be exported to a personal computer.
(Set-up Menu is opened by selecting the screwdriver icon found along
the bottom of the display.)
Pin 9 of the RS-232 serial interface connector provides an “Alarm Out” signal.
Any time there is an alarm condition active in the NPB-3900, pin 9 will go to
plus RS-232 level voltage (> +5 VDC), if “Nurse Call Signal” is set to ON in the
Set-up Menu. Any time there is no active alarm condition, pin 9 will be at minus
RS-232 level voltage (< -5 VDC). If in the Set-up Menu “Nurse Call Signal” is
set to OFF, pin 9 will always be at the minus RS-232 level voltage. In order to
make use of the Alarm Out signal, pin 9 should be connected to a high-
impedance circuit (> 1000Ω) and protected against transient voltages.
10-1
Section 10: RS-232 Interface
10.3 EXPORTING TREND DATA
In order to download trend data from the NPB-3900, communication software,
such as PROCOMM™, should be installed in the external computer. The
transfer protocol should be set as follows:
Baud Rate:19,200
Data Bits:8
Start Bit:1
Stop Bits:1
Parity:None
Connect the NPB-3900 to the serial port of the computer using a null modem
cable. Start the communication program on the computer and enter terminal
emulation mode. To initiate the transfer, type tr (lower case is necessary),
followed by a carriage return <cr>.
If the command is not accepted, the response to an invalid command is ??,
followed by a carriage return <cr>.
In response to a valid command, the NPB-3900 will send a comma-delimited
ASCII text file comprising the entire contents of the NPB-3900’s trend memory.
Each line is divided into five main groups, separated by a space <sp> and ending
with a carriage return <cr> and line feed <lf>. The format for each line is:
RECORD<sp>DATE<sp>TIME<sp>ALARMS<sp>VITALS<cr><lf>
The fields within each group are identified and defined as follows:
RECORD:
record number,
Format:
2 characters
no leading zero suppression
right justified
DATE:
day, month, year,
Format:
day and month: 2 characters
year: 4 characters
no leading zero suppression
right justified
10-2
Section 10: RS-232 Interface
TIME:
hours, minutes, seconds,
Format:
2 characters
no leading zero suppression
right justified
Each field in this group is either:
0: corresponding vital sign was not in alarm state
or,
1: corresponding vital sign was in alarm state
NOTE: In order to maintain a consistent trend data format between the
NPB-3900 and NPB-4000 patient monitors, the NPB-3900 maintains a
field for “respiration rate alarm” and “temperature alarm”, even though
the NPB-3900 does not have an alarm for respiration rate or temperature.
For the NPB-3900, the fields for “respiration rate alarm” and
“temperature alarm” will always have a value of “0”.
NOTE: In order to maintain a consistent trend data format between the
NPB-3900 and NPB-4000 patient monitors, the NPB-3900 maintains a
field for “respiration rate”, even though the NPB-3900 does not measure
respiration rate. For the NPB-3900, the field for “respiration rate” will
always have a value of “0”.
If no vital sign was measured during a 20-second trend interval, characters in the
corresponding field will be blank (<sp><sp><sp><sp>,). If the vital sign
displayed dashes during a 20-second trend interval, characters in the
corresponding field will contain dashes (----,).
Example of several lines of a trend file:
10-4
APPENDIX - TECHNICAL SUPPLEMENT
A-1 General
A-2 Block Diagram
A-3 Isolated Patient Connection Section
A-4 Temperature Measurement Circuit
A-5 ECG Inputs
A-6 On/Off Power Control
A-7 Audio Volume and Speaker Drive
A-8 Power Supplies
A-9 NIBP Section
A-10 System A/D
A-11 Buttons and Lights
A-12 SpO
A-13 Microcontroller
A-14 Program Storage/Execution
A-15 DRAM Control
A-16 Real Time Clock
A-17 UART Operation
A-18 FPGA Glue Logic
A-1 GENERAL
This section contains descriptions of the principles of operation of the major
functional modules of the monitor, including the overall block diagram, power
supply, isolated front end, NIBP control, the SpO
microcontroller.
A-2 BLOCK DIAGRAM
The Monitor (see Figure A-1) contains an isolated front-end section, powered by
an isolated power supply, and in which the signals from SpO
ECG sensors are processed. The plastic tubing provides sufficient isolation for
signals from the cuff in NIBP monitoring.
2
2 processing module, and
2, temperature, and
A single A/D converter is used to digitize processed temperature, NIBP, and
ECG inputs; the SpO2 module produces digitized data.
A microcontroller, Intel 386, requests and receives instructions from a flash
memory. The processor has a 16-bit data bus, and uses 19 of the 24-bit address
bus. These, and eight control signals, are used to read and write to the DRAM,
flash memory, UART, and FPGA (programmable gate array). Other interface
connections are made through the I/O port signals, timer signals, and interrupt
signals.
A-1
Appendix - Technical Supplement
The FPGA provides signals for control and data to the LCD. Bias voltage and
backlight power for the LCD are provided by the power supply section. The
FPGA processes front-panel button and Nellcor Puritan Bennett knob operations.
Circuit details for these blocks are contained in this section, of the manual. This
section provides a brief theory of operation of the circuits noted in the block
diagram. Portions of the schematic diagrams are reproduced for some
paragraphs.
Temp Probe
Switch
TEMP
Probe
ECG Input
Connector
SPO2
Input
Connector
NIBP
Fitting
ISOLATED FRONT END
Pressure Sensor
Pressure Sensor
Proportiona
Keypad/Knob
RTC
FPGADRAM
CNTL
FLASH ADDRESSFPGA ADDRESS
FLASH
ON/OFF Control
and Watchdog
DRAM CNTL
ADDRESS BUS
FPGA DATA
16 Bit Bi-directional bus
FLASH DATA
ON/OFF
Control
Failure
Alarm
Alarm
Volume Control
Audio Driver
6V 4AH Battery
LCD CNTL/DATA
DRAM ADDRESS
Back Light Power
Supply (300 VRMS)
+5 V
DRAM DATA
Speake
Battery Charger
LCD Display
Bias
Voltage
r
Charger
Connecto
Backlight
Power
RS-232
Connecto
r
External
16V RMS
Battery
AC INPUT
Charger
r
FRONT END CNTL
A/D
SPO2 SERIAL CNTL
PS CLK
MSTR
CLK
OSC
A/D
CNTL
386EX
C STEP
FLASH CNTL
UART/RS232
+5 V Regulator
ISO PS
Heater
TEMP
XFMR
DIG
OPTO
LIN
ECG
OPTO
SPO2
Pump
#1
#2
l
Valve
DIG
OPTO
NIBP
Amplifiers
Ven
t
LCD Bias Regulator
+3.3 V Regulator
Figure A-1: NPB-3900 Block Diagram
A-3 ISOLATED PATIENT CONNECTION SECTION
A-3.1 General
The connections to the patient consist of:
ECG (three directly connected leads)
Temperature (probe)
2 (DS-100A sensor, etc.)
SpO
NIBP (pneumatic cuff)
A-2
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.