Notice: Purchase of this instrument confers no express or implied license under any Oridion
Medical patent or Nellcor Puritan Bennett patent to use the instrument with any accessory that is
not manufactured or licensed by Oridion Medical 1987 Ltd.
Capnography is covered by one or more of the following United States patents: 6,491,643;
6,656,127; 6,168,958; 5,300,859 and 4,755,675 and their foreign equivalents
この製品は、次の日本における特許、1980816と海外での同等のものによって保護されて
居ります。
Microstream
Smart BiteBloc™ and NIV Line™ are trademarks of Oridion Medical 1987 Ltd.
Nellcor Puritan Bennett Incorporated is an affiliate of Tyco Healthcare.
The following are trademarks of Nellcor Puritan Bennett Incorporated: Oxiband
OxiCliq
Pulse Oximetry is covered by one or more of the following US patents and foreign equivalents:
4,802,486; 4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,485,847; 5,743,263; 5,865,736;
6,035,223; 6,298,252; 6,463,310; 6,591,123; 6,675,031; 6,708,049; 6,801,797; Re 35,122.
®
and FilterLine® are registered trademarks and Smart CapnoLine™, CapnoLine™,
13-1Spare Parts List ............................................................... 13-1
v
Table of Contents
SECTION 1: INTRODUCTION
1.1Manual Overview
1.2Warnings, Cautions, and Notes
1.3Product Description
1.1MANUAL OVERVIEW
This manual contains information for servicing the handheld capnograph/
pulse oximeter (the monitor). Only authorized service personnel should
service this product. The Handheld Capnograph/Pulse Oximeter
Operator’s Manual (the operator’s manual) is an integral part of the
service procedures. Before servicing the monitor, read the operator’s
manual carefully for a thorough understanding of how to operate the unit.
Only use the tools and test equipment as specified in this manual. Only
use original spare parts available from your local distributor.
Warning: Incorrect procedures may harm the patient, or
damage the monitor.
1.2WARNINGS, CAUTIONS, AND NOTES
1.2.1Warning
1.2.2Caution
1.2.3Note
A warning precedes an action that may result in injury or death to the
patient or user. Warnings are boxed and highlighted in boldface type.
A caution precedes an action that may result in damage to, or malfunction
of, the monitor. Cautions are highlighted in boldface type.
A note gives information that requires special attention.
1-1
Section 1: Introduction
1.3PRODUCT DESCRIPTION
The handheld capnograph/pulse oximeter is used to continuously monitor
end-tidal carbon dioxide (EtCO
2), and pulse rate. The monitor is for attended monitoring only and
(SpO
must be used in the continuous presence of a qualified healthcare
provider. The monitor can be used on adult, pediatric, and infant/neonatal
patients. It is intended for use where continuous, noninvasive monitoring
of these parameters is desired, including hospital and mobile use (when
protected from excessive moisture such as direct rainfall). The monitor
operates on AC power or a rechargeable Nickel Metal Hydride battery
pack.
2), respiratory rate, oxygen saturation
1-2
Figure 1-1: Handheld Capnograph/Pulse Oximeter
The monitor is operated using a four-key keypad. Measurements are
displayed on the front panel on the 7-segment LED digital displays (LED)
and on the LCD graphic display (LCD) as shown in Figure 1-2. Refer to the
operator’s manual for complete operating instructions.
Figure 1-2: LCD Graphic Display and 7-Segment Digital Displays
SECTION 2: REQUIRED EQUIPMENT
2.1Required Equipment
2.1REQUIRED EQUIPMENT
You will need the following equipment to disassemble, replace parts,
check, adjust, or calibrate the monitor.
For disassembly and replacement:
• Phillips-head screwdriver (medium)
• Flat-head screwdriver (small)
• IC Extractor for PLCC socket
• Threadlocker glue e.g. Loctite
For checks, adjustments and calibration:
®
222 super screw lock
• Calibration Gas (5% CO
2 in air)
• FilterLine or FilterLine H and a connecting means
• DS-100A - O
XIMAX Durasensor
®
, Nellcor
• Leak Test Jig
• Vacuum manometer with a range of 0-500 mBar
e.g. MPL-3200-DM, Digital Manometer MicroPneumatic Logic, Inc.
• Communication Adapter Kit (only necessary to verify print functions)
• Seiko DPU-414 printer (only necessary to verify print functions)
• Flow Meter (range 0-50 ml/min) e.g. Top Track Mass Flow Meter,
model: 822-13-OV1-PV1-V1, Sierra Instruments, Inc.
Warning: Observe ESD (electrostatic discharge) precautions
when handling, adjusting or performing any procedure with
the monitor’s internal components.
2-1
SECTION 3: CLEANING
3.1Cleaning
3.1CLEANING
Always clean the outer surface of the monitor before servicing.
Warning: Always wear latex or surgical gloves when cleaning
or servicing the monitor after hospital use.
Warning: Do not allow liquids to drip inside the housing.
Warning: Do not clean internal components.
Warning: Do not use preparations based on: phenols, halogen
producing compounds, strong organic acids, or oxygen
producing compounds.
3.1.1Procedure
1.Unplug the unit from AC power.
2.With a damp cloth, gently wipe the outer case of the monitor with a
disinfectant based on aldehyde, alcohol or quaternary ammonium
producing compounds.
3-1
SECTION 4: SERVICE MODE
4.1Introduction
4.2Service Level 1
4.3Service Level 2
4.1INTRODUCTION
The monitor supports a service mode built into the instrument’s firmware.
To access the service mode, refer to Table 4-1: "Accessing Service Mode
and Changing Parameter Settings".
The architecture of the monitor’s service mode divides the service mode
into two levels as described in the sections below.
• Service Level 1
• Service Level 2
4.2SERVICE LEVEL 1
The Service Level 1 provides information and tools on three service
screens to test and calibrate specific functions without disassembling the
monitor.
• Information Screen
• Event Messages Screen
• Interface Test Screen
From these display screens the service technician can check the operating
hours counter and event messages, test the Human Interface functions
(HMI), set the light detection threshold, adjust the LCD contrast middle
point and do a calibration check and a calibration process, if necessary.
Refer to Table 4-1: "Accessing Service Mode and Changing Parameter
Settings".
4.2.1Information Screen
The information screen displays the monitor’s main board serial number,
the number of operating hours, and the SpO
versions.
4.2.2Event Messages Screen
The monitor automatically performs a series of built-in tests (BITs) at
start-up and during operation to check sub-systems and modules for
errors. The event messages screen lists the last five event messages, the
time each occurred relative to the unit’s operating hours, and a
recommended service action. There are three event messages and each
has a specific response as follows:
2 and CO2 boards’ software
Event Message:Action:
2 boardReplace CO2 board
CO
2
boardReplace SpO2 board
SpO
FlowReplace CO
2 board
4-1
Section 4: Service Mode
Note:To exit the service mode and save event messages, turn off the
4.2.3Interface Test Screen
The interface test screen shows four parameters (listed below) that
provide tools for the user to test, adjust, and calibrate the unit’s
interfacing systems.
•HMI Test
• Light Detector
• Contrast Cal. (Calibration)
monitor. Event messages will be erased when exiting the service
mode by a long press of .
•CO
2 Calib. (Calibration)
4.2.3.1 HMI Test
When the HMI Test is turned on, the monitor performs a self-test of the
following sub-systems:
• Buzzer (sounds for one second)
• LCD graphic display (screen shows a checkerboard pattern)
• 7-segment digital display (8s scroll left to right)
• Alarm bar (lights red, yellow, off)
• Backlight (turns on and off)
• Alarm silence indicator (light turns on and off)
When the HMI test is completed the status field returns automatically to
OFF.
4.2.3.2 Light Detector
The Light Detector parameter allows the service technician to set the
threshold value of the phototransistor. The default value is 1 and the
range is from 1 to 5 with 1 as the most sensitive (backlight on at all times)
and 5 as the least sensitive (backlight off). The threshold value should be
set according to the lighting conditions of the working environment. A
higher threshold value saves on battery power consumption.
4-2
4.2.3.3 Contrast Cal. (calibration)
The Contrast Cal. gives the service technician the option to change the
middle point default value of the contrast intensity of the LCD. Refer to
Table 4-1: "Accessing Service Mode and Changing Parameter Settings".
4.2.3.4 CO
The CO
2 Calib. (calibration)
2 Calib. allows the service technician to perform a calibration
check and calibration process to ensure the accuracy of the monitor.
Section 10: Performance Verification describes the calibration check
procedure.
4.2.4Accessing Service Mode and Service Level 1.
Table 4-1: "Accessing Service Mode and Changing Parameter Settings"
describes the process to access the service mode, change parameter
settings and how to change Contrast Cal. settings.
Table 4-1: Accessing Service Mode and
Changing Parameter Settings
ObjectiveActionResponse
Section 4: Service Mode
To access Service Mode During self-test,
long press
simultaneously
and
To access Event Messages screen
To access Interface Test screen
long press
long press
Self-test occurs and the
Information Screen
appears:
To change the
parameter setting or
value
To move to the next
parameter
short press
short press
4-3
Section 4: Service Mode
Table 4-1: Accessing Service Mode and
Changing Parameter Settings (Continued)
ObjectiveActionResponse
To change Contrast
short press
Cal. setting to set
To start Contrast Cal.short press“Contrast Cal.” flashes.
To adjust Contrast Cal. middle point
press
Screen lightens or
darkens
until desired
setting is reached.
(Pressing on the left
lightens the screen,
and pressing on the
right darkens the
screen.)
To accept new settingshort press
To reset and return to
measuring mode
(automatically erasing
the event messages
memory)
To exit service mode
from any service
screen without erasing
the event messages
memory
long press
Note: Reset time is
approximately one
minute.
Turn off the
monitor.
4-4
4.3SERVICE LEVEL 2
Section 4: Service Mode
The Service Level 2 gives the service technician tools on three service
screens:
• Default Alarm Limits
• Default Settings
• Flow Calibration
To access this service level, the service technician must enter a valid
password (1627), refer to Table 4-2: "Accessing Service Level 2".
Note: The password should not be disclosed to avoid unauthorized setting
of parameters.
Table 4-2: Accessing Service Level 2
ObjectiveActionResponse
To access
Service Level 2
Enter Password: 1627
Enter the 1st digitshort press (right)
Move to the next digitshort press
After entering the
service mode
long press (x3)
(x1)
Pressing on the
right arrow scrolls
the numbers up,
the left arrow
scrolls the numbers
down.
Password Screen
appears
Enter the 2nd digitshort press (right)
x6
Repeat the same procedure as above for the next two digits (2, 7).
4-5
Section 4: Service Mode
Table 4-2: Accessing Service Level 2 (Continued)
ObjectiveActionResponse
To accept the last digit
and access the Default Alarm Limits menu
In case you entered a
wrong digit:
To restart password
validation process
To return to the
Interface Test Screen
short press
long press
long press
To exit service mode
and save event
messages
Turn off the
monitor.
4-6
4.3.1Default Alarm Limits
The service technician can change the default alarm limits for either
patient mode (Adult or Neonatal) as described in Table 4-3: "Changing
Default Alarm Limits".
The monitor has the following alarms with adjustable level settings:
Section 4: Service Mode
•EtCO
•FiCO
•SpO
2 high and low levels (mmHg)
2 high level (mmHg)
2 high and low levels (%)
• Pulse rate high and low levels (bpm)
Warning: Changing default settings will permanently change the
user’s start-up value and should be verified with the responsible
clinical personnel.
Table 4-3: Changing Default Alarm Limits
ObjectiveActionResponse
To access the Default
short press
Alarm Limits menu
after entering last
password digit