Nellcor N-600x User manual

Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor,
OxiCliq, Dura-Y, MAX-FAST, and O
XIMAX are trademarks of Nellcor Puritan Bennett Inc.
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 Canada. To obtain information about a warranty, if any, contact Nellcor’s Technical Services Department,
Bennett patent to use the instrument with any sensor that is not manufactured or licensed by Nellcor Puritan Bennett.

Contents

Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .i
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety Warnings ................................................................................1
Safety Cautions .................................................................................3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Intended Use .....................................................................................5
How to Use this Manual ....................................................................6
Symbols, Controls, Displays and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . 7
About the Front Panel .......................................................................7
About the Rear Panel ........................................................................8
About the Symbols ............................................................................9
About the Controls ..........................................................................10
About the Displays ..........................................................................11
Pleth Display .........................................................................11
Blip Display ..........................................................................12
Real-Time Trend Display ......................................................14
SpO2 and Pulse Rates .........................................................14
About the Visual Indicators .............................................................16
About the Audible Indicators ...........................................................19
Setting Up the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
List of Components .........................................................................23
Connecting to an AC Power Source ...................... ... ... ... ... .... ... ... ...24
Connecting an OXIMAX Sensor ...................... ... .... ... ... ... ... .... ... ... ...26
Operating the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Operating on Battery Power ......................... ...................................27
Low Battery Indicator ............... .... ... ................................................29
Description of Low and Critical Battery Conditions ...............30
Battery Fuel Gauge Indicator ................................................34
Using the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Overview ......................................................................................... 35
Menu Description .. ... .... ... ... ... ... .... ... ... ....................................... ... ...35
Menu Structure .....................................................................36
Parameter Ranges ..........................................................................40
Turning On the Monitor ...................................................................43
N-600x Operator’s Manual i
Contents
OXIMAX Sensor Attached .................................................... 45
No OXIMAX Sensor Attached .............................................. 49
Turning the Backlight On or Off ...................................................... 50
Adjusting the Screen Contrast ........................................................ 50
Adjusting the Backlight Brightness ................................................. 50
Selecting the Pleth View ................................................................. 51
Selecting the Blip View ................................................................... 51
Selecting the Real-Time Trend View .............................................. 52
Selecting the Trend Data Display ......................................... 53
Setting the Trend Time Scale Display .................................. 53
Setting the Trend Amplitude Scale Display .......................... 54
Setting the Pulse Beep Volume .......................................... ... ... ... ... 54
Setting the Alarm Volume .. .... ... ... ... ... .... ... ... ... .... ... ... ... ... .... ... ... ... ... 55
Setting the Date and Time...................................... ... ... ... .... ... ... ... .. 56
Setting the Alarm Silence Duration ........................... ... ... .... ... ... ... ... 58
Disabling Audible Alarms ................................................................ 59
Selecting the Standby Mode ........................................................... 60
Adult-Pediatric or Neonatal Settings ............................ ................... 61
Setting Patient Adult-Pediatric or Neonatal Modes .............. 62
Alarm Limit Changed Indicator ....................................................... 63
Setting Alarm Limits ............................... ... ... ... .... ... ... ... ... ................ 63
Setting SatSeconds Alarm Limit ........ .... ... ... ... .... ... ... ... ... .... ... ... ... ... 65
Setting Monitor Response Mode .............. ... ... .... ... ... ... ... .... ... ... ... ... 66
Selecting the Display Language ..................................................... 68
OXIMAX Sensor Messages ............................................................ 69
OXIMAX Sensor Adjust Condition Messages ...................... 70
OXIMAX Sensor Adjust Messages ....................................... 71
Using Monitor Trend Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Overview ................................... ................................................ ...... 73
Storing Trend Data ............................................................... 75
OXIMAX Sensor Type .................................................................... 76
Selecting the Trend Data Display Scale ......................................... 76
Reading the Trend Data Display ............................ ................... ...... 78
Dual Trend Data Display ................................................................. 79
SpO2 Trend Display ....................................................................... 80
Pulse Rate Trend Display ............................................................... 80
Histogram Trend Data Display .................. ... ... .... ... ... ... ... .... ... ... ... ... 81
Pulse Amplitude Trend Data Display .............................................. 82
Clearing Trend Information ............................................................. 83
Using OXIMAX Sensor Event Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Overview ................................... ................................................ ...... 85
Setting up OXIMAX Sensor Messages ........................... .... ... ... ... ... 87
ii
Setting In-Sensor Data Type ...........................................................88
OXIMAX Sensor Data Type ............................................................89
OXIMAX Sensor Event Record Data Available ...............................90
OXIMAX Sensor Event Record Not Available .................................91
OXIMAX Sensor Event Record Graphical Data ........................... ...92
Viewing and Printing OXIMAX Sensor Event History Data .............94
OXIMAX Sensor Tabular Event Data ..............................................96
Viewing and Printing In-Sensor Tabular Event History Data ...........97
Printing Monitor Trend Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Overview ......................................................................................... 99
Printing ..................................... .................... ................... ......99
Monitor Trend Data in ASCII Mode ...............................................102
Trend Data in Graph Mode ...........................................................103
Real-Time Display/Printout Format ...............................................103
Column Headings ...............................................................105
Data Source ........................................................................105
Software Version ........ ... .... ... ... ... .... ....................................106
Alarm Limits ........................................................................106
Monitor Mode ........................ ... ... .... ... ... ... .... .......................106
Response Mode ..................................................................107
Data Column Headings .......................................................107
Time ...................................... ..............................................108
Patient Data .................................... ... ... ... .... .......................108
Operating Status ......................... .... ... ... ... .... ... ... ... ... .... ... ... .108
Using the Data Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Overview ....................................................................................... 111
Connecting to the Data Port ..........................................................112
Data Port Pinouts ................................................................112
Data Port Setup .............................................................................114
Using the Nurse Call Interface ......................................................116
Setting Nurse Call RS-232 Polarity .......... .... ... ... ... ... .... ... ... .117
Setting Nurse Call Relays Normally Open/Closed ..............118
Calculating the Analog Voltage Output .........................................118
OXIMAX Sensors and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Overview ....................................................................................... 121
Selecting an OXIMAX Sensor .......................................................122
OXIMAX Sensor Features .............................................................126
Biocompatibility Testing ................................................................126
Optional Accessories ................... .......................................... ... ... .127
GCX Mounting Plate ...........................................................128
GCX Vertical Wall Mount Arm ............................................129
N-600x Operator’s Manual iii
Contents
GCX Roll Stand .................................................................. 130
Soft-Sided Carrying Case ................................................... 131
Performance Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133
Overview ................................... ................................................... .133
Performance Considerations .......................................... .... ... ....... 134
Dysfunctional Hemoglobins ................................................ 134
Anemia ............................................................................... 135
Saturation ........................................................................... 135
Pulse Rates ........................................................................ 135
OXIMAX Sensor Performance Considerations ............................. 135
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .139
Overview ................................... ................................................... .139
On-Screen Help ..................................... ... .................................... 140
Accessing Multiple Topics .................................................. 140
Accessing Single Topics .................................................... 143
Error Codes .................................................................................. 145
Prompts and Error Messages ....................................................... 148
Primary Speaker Failure ............................................................... 151
Help and Support ................................... ... ... ... .... ... ... ... ... .... .......... 153
EMI (Electromagnetic Interference) .............................................. 158
Obtaining Technical Assistance .................................................... 159
Returning your Monitor ....................................................... 159
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .161
Overview ................................... ................................................... .161
Service ................................... ................................................ ....... 161
Periodic Safety Checks ................................................................. 161
Cleaning ....................................... ................... ....................... ....... 162
Using SatSeconds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .163
Overview ................................... ................................................... .163
SatSeconds “Safety Net” .............................................................. 165
SatSeconds Display ...................................................................... 165
Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .167
Overview ................................... ................................................... .167
Neonate Default Settings .............................................................. 167
Adult Default Settings ................................................................... 168
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .171
Overview ................................... ................................................... .171
Automatic Calibration ......................................................... 172
iv
Functional versus Fractional Saturation .............................173
Measured versus Calculated Saturation .............................174
OXIMAX Technology .....................................................................175
Functional Testers and Patient Simulators ...................................176
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Performance .................................... ................... .................... .......177
Electrical ....................................... ................ ................ ................ .178
Environmental Conditions ..... ... .... .......................................... ... ... .180
Physical Characteristics ................................................................182
Compliance ................................................................................... 182
Manufacturer’s Declaration ...........................................................183
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
N-600x Operator’s Manual v
Contents
vi

Safety Information

Safety Warnings

Warnings are identified by the WARNING symbol shown above.
Warnings alert you to potential serious outcomes (death, injury, or adverse events) to the patient or user.
WARNING: The sensor extrapolates from the date and time provided by the N-600x when recording the sensor event record to the sensor . The accuracy of th e date/time is the responsibility of the N-600x. It is recommended that the N-600x user set the time/ date to the correct value before a sensor event record-enabled sensor is connected, and that this date/time not be changed while the sensor remains connected. Since a sensor with sensor event record data can be transported from one monitor to another, having discrepancies in the date/time between monitors and the sensor event record data will affect the order the sensor event record data appears. To eliminate this possible problem, all monitors within an institution should be set to the same time.
Safety Information
N-600x Operator’s Manual 1
Safety Information
WARNING: Explosion hazard. Do not use the N-600x pulse oximeter in the presence of flammable anesthetics or gases.
WARNING: Chemicals from a broken LCD display panel are toxic when ingested. Use caution when handling a pulse oximeter with a broken display panel.
WARNING: Pulse oximetry readings and pulse signals can be affected by certain environmental conditions, O
XIMAX sensor
application errors, and certain patient conditions. See the appropriate sections of this manual for specific safety information.
WARNING: The use of accessories, sensors, and cables other than those specified may result in increased emission and/or decreased immunity and inaccurate readings of the N-600x pulse oximeter.
WARNING: Failure to cover the OXIMAX sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.
2 N-600x Operator’s Manual

Safety Cautions

Safety Information
Cautions are identified by the CAUTION symbol shown above.
Cautions alert you to exercise care necessary for the safe and effective use of the N-600x pulse oximeter.
Caution: When connecting the N-600x to any instrument, verify proper operation before clinical use. Both the N-600x and the instrument connected to it must be connected to a grounded outlet. Accessory equipment connected to the pulse oximeter’s data interface must be certified according to IEC Standard 60950 for data-processing equipment or IEC Standard 60601-1 for electromedical equipment. All combinations of equipment must be in compliance with IEC Standard 60601-1-1 systems requirements. Anyone who connects additional equipment to the signal input port or signal output port (N-600x data port connector) configures a medical system and is therefore responsible for ensuring that the system complies with the requirements of system standard IEC Standard 60601-1-1 and the electromagnetic compatibility system standard IEC S tandard 60601- 1-2. The N-600x accuracy may degra de if it is connected to secondary I/O devices when the instrument is not connected to earth reference.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
Caution: Dispose of battery in accordance with local requirements and regulations.
N-600x Operator’s Manual 3
Safety Information
4 N-600x Operator’s Manual

Introduction

WARNING: The N-600x is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.

Intended Use

The N-600x pulse oximeter is indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO with neonatal, pediatric, and adult patients who are well or poorly perfused, in hospitals, hospital-type facilities, intra-hospital transport, and home environments. For prescription use only.
Introduction
2) and pulse rate. The N-600x is intended for use
Note:Hospital use typically covers such areas as general care floors, operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.
Note: Home Care use is defined as managed or used by a lay person (parent or other similar noncritical caregiver) in the home environment.
Use with any particular patient requires the selection of an appropriate oxygen O
XIMAX sensor as described in this manual.
N-600x Operator’s Manual 5
Introduction

How to Use this Manual

All users should read this manual thoroughly . More experienced users of the N-600x can directly go to the topics for the information they require.
This manual is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
6 N-600x Operator’s Manual

Symbols, Controls, Displays and Indicators

Symbols, Controls, Displays and Indicators

About the Front Panel

1. SpO2 OXIMAX Sensor Port, page 26.
2. Low Battery Indicator, page 17. 13. ADJUST DOWN Button, page 10.
3. AC Power Indicator, page 16. 14. Neonate Mode Indicator, page 18.
4. ON/STANDBY Button, page 10.
5. Battery Fuel Gauge Indicator, page 34. 16. Fast Response Mode Indicator, page 18.
6. Waveform Display, page 12. 17. Softkeys, page 11.
7. SatSeconds
8. %SpO
9. Pulse Rate Display, page 16. 20. Interference Indicator, page 17.
10. Alarm Silence Indicator, page 17. 21. Pulse Search Indicator, page 17.
11. ALARM SILENCE Button, page 10. 22. Monitor Speaker.
TM
Indicator, page 18.
2 Display, page 16. 19. Data In-Sensor Indicator, page 18.
12. ADJUST UP Button, page 10.
15. HELP/CONTRAST Button, page 11.
18. Menu Bar, page 11.
Figure 1: Front Panel Buttons and Symbols
N-600x Operator’s Manual 7
Symbols, Controls, Displays and Indicators

About the Rear Panel

1. Equipotential Terminal (Ground).
2. AC Power Connector, page 24.
3. Data Port Connector, page 112.
4. Fuse Holder.
5. Supply Voltage Selector Switch, page 24.
Figure 2: Rear Panel Components
8 N-600x Operator’s Manual

About the Symbols

The symbols, located on the rear panel of the N-600x, are as follows.
Symbols, Controls, Displays and Indicators
Table 1: Symbols and Descriptions
Symbol
Description
See Instructions for Use
Fuse Replacement
Equipotential Terminal (ground)
Date of Manufacture
Data Interface
Type BF Applied Part - Not defibrillator proof
N-600x Operator’s Manual 9
Symbols, Controls, Displays and Indicators

About the Controls

ON/STANDBY Button Turns the monitor on and off.
Note: Pressing a button, except the ON/STANDBY button, should result in either a valid or an invalid key tone (refer to Table 3). If the key pressed fails to emit a tone, contact qualified service personnel.
ALARM SILENCE Button Silences current alarms for the alarm silence duration period. When an alarm has been silenced, pressing the button again reactivates, or “unsilences” the alarm. It is also used to view and adjust alarm silence duration and alarm volume. The ALARM SILENCE button clears “SENSOR OFF,” “LOW BATTERY,” and “SENSOR DISCONNECT” messages from the display.
ADJUST UP Button Increases variable parameters of the monitor.
ADJUST DOWN Button Decreases variable parameters in the monitor.
10 N-600x Operator’s Manual
HELP/CONTRAST Button Enables you to access the on-screen help and adjust the monitor screen contrast.
Softkey Menu Bar
Displays the current functions for each of the four softkey buttons.

About the Displays

Symbols, Controls, Displays and Indicators
Pressing and releasing the HELP/CONTRAST button launches the on-screen help.
Pressing and holding the HELP/CONTRAST button while simultaneously pressing the ADJUST UP and ADJUST DOWN buttons lightens or darkens the display screen.

Pleth Display

The pleth display is user selectable. Refer to Selecting the Pleth View on page 51.
N-600x Operator’s Manual 11
Symbols, Controls, Displays and Indicators
The pleth display includes a “wiper bar” plethysmographic waveform, menu bar, and current measured %SpO SatSeconds are enabled, the pleth display includes the SatSeconds indicator and SatSeconds setting. A decimal point after the %SpO pulse rate indicates the respective limits have been changed from the power on defaults (Overview on page 73). Plethysmographic waveforms with peak to peak amplitudes less than ten PAUs are associated.
Caution: Verify the movement of the blip bar or plethys­mographic waveform or beating heart before accepting any displayed data as a current measurement.
When the monitor is powered by the internal battery, the pleth display includes a horizontal battery fuel gauge positioned in the upper left corner which shows the remaining charge (operating hours) on the battery. If a monitor reporting low battery is connected to an AC power source, the battery fuel gauge displays the charging progress. The battery fuel gauge is cleared from the display once the monitor can provide at least 15 minutes of operating time.
2 and pulse rate. If
2 or

Blip Display

The blip display includes a pulse amplitude blip bar, current measured %SpO
2 and pulse rate, and current upper and lower %SpO2 and pulse
rate limits. If SatSeconds are enabled, the blip display includes the SatSeconds indicator and SatSeconds setting. Decimal points after the %SpO
2 or pulse rate indicate that the respective limits have been
changed from the power-on defaults.
12 N-600x Operator’s Manual
Symbols, Controls, Displays and Indicators
When the monitor is powered by internal battery, the blip display includes a horizontal battery fuel gauge positioned in the upper left corner that shows the remaining charge (operating hours) on battery. If a monitor reporting a low battery is connected to AC power, the battery fuel gauge shows the charging progress. The battery fuel gauge is cleared from the display once the monitor can provide at least 15 minutes of operating time on battery.
N-600x Operator’s Manual 13
Symbols, Controls, Displays and Indicators

Real-Time Trend Display

The real-time trend display includes %SpO2 and/or pulse rate trend data plots and current measured %SpO2 and pulse rates. The trend data plots are automatically updated as each new trend point is calculated, where the interval between calculations is based on the display time scale selected. If SatSeconds is enabled, the real-time trend display includes the SatSeconds indicator. Decimal points after the displayed %SpO2 or pulse rate indicate that the respective limits have been changed from the power-on defaults. Each time a pulse is detected by the oximeter, a heart icon flashes.

SpO2 and Pulse Rates

There are various matrixes within the N-600x algorithm. Some of these are used to assess the severity of conditions presented to the N-600x in measuring SpO individual matrices or combinations of these matrices are used to drive the LED indicators on the N-600x front panel.
14 N-600x Operator’s Manual
2 and pulse rate on a patient. These
Symbols, Controls, Displays and Indicators
The N-600x algorithm automatically extends the amount of data required for measuring SpO
2 and pulse rate depending on the
measurement conditions. During normal measurement conditions the averaging time is 6 to 7 seconds. During conditions such as those caused by low perfusion, interference (e.g., external interference like ambient light), or a combination of these, the N-600x algorithm automatically extends the amount of data required beyond 7 seconds. If the resulting dynamic averaging time exceeds 20 seconds, the pulse search indicator is lit solid and SpO
2 and Pulse Rate will continue to
be updated every second. As these conditions extend, the amount of data required continues to increase. If the dynamic averaging time reaches 40 seconds, the pulse search indicator begins flashing, the SpO
2 and pulse rate displays flash zeros indicating a loss-of-pulse
condition.
WARNING: Failure to cover the OXIMAX sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.
N-600x Operator’s Manual 15
Symbols, Controls, Displays and Indicators

About the Visual Indicators

Table 2: Visual Indicators
Indicator Description
%SpO2 Display Shows the hemoglobin oxygen saturation level. The display value
flashes zeros during loss-of-pulse alarms and flashes the SpO when the SpO monitor continues to update the display. If alarm limits have been changed from their power-on defaults, a decimal point (.) is displayed after the SpO
2 is outside the alarm limits. During Pulse Search, the
2 value (81.).
2 value
Pulse Amplitude
Indicator (blip bar)
Indicates pulse beat and shows the relative (non-normalized) pulse amplitude. As the detected pulse becomes stronger, more bars light with each pulse. This indicator is available only in the blip view.
Plethysmographic
Waveform Display
Displays a non-normalized waveform in real-time sensor signals. The relative pulsatile strength and quality of the incoming signals can be observed.
Pulse Rate Display Displays the pulse rate in beats per minute. It flashes during
loss-of-pulse alarms and when the pulse rate is outside of the alarm limits. During Pulse Search, the monitor continues to update the display. Pulse rates outside of the pulse rate range of 20 to 250 bpm are displayed as 0 and 250, respectively. If alarm limits have been changed from their power-on defaults, a decimal point (.) is displayed after the BPM value (112.).
AC Power
Indicator
Lights continuously when the N-600x is connected to an AC power source. The indicator shows that the battery is charging. It is off when the monitor is being powered by internal battery.
16 N-600x Operator’s Manual
Table 2: Visual Indicators
Indicator Description
Symbols, Controls, Displays and Indicators
Low Battery
Indicator
Battery Fuel Gauge
Indicator
Alarm Silence
Indicator
Interference
Indicator
Lights continuously when 15 or fewer minutes of battery capacity remain. Flashes when the battery capacity reaches a critically low condition.
Displays the battery charge remaining on the monitor. The battery fuel gauge consists of four bars, each corresponding to approximately 1.5 hours of operating time. All four bars are lit when the battery is fully charged. No bars are lit when a low battery condition exists. See Battery Fuel Gauge Indicator on page 34.
Lights continuously when an audible alarm has been silenced. It flashes when the alarm silence duration has been set to Off.
Lights whenever the N-600x algorithm detects the incoming signal quality is degraded.
Note: Degradation can be caused by ambient light, electrical noise, electro-surgical interference, patient activity, or other causes.
An intermittently lit Interference Indicator is common during patient monitoring, and indicates the N-600x algorithm is dynamically adjusting the amount of data required for measuring
2 and Pulse Rate. When lit continuously, the N-600x algorithm
SpO has extended the amount of data required for measuring SpO
2 and
Pulse Rate and consequently fidelity in tracking rapid changes in these values may be reduced.
Pulse Search
Indicator
Lights continuously prior to initial acquisition of a pulse signal and during prolonged and challenging monitoring conditions. The pulse search indicator flashes during a loss-of-pulse signal.
N-600x Operator’s Manual 17
Symbols, Controls, Displays and Indicators
Table 2: Visual Indicators
Indicator Description
Data In-Sensor
Indicator
SatSeconds
Indicator
Fast Response
Mode Indicator
Neonate Alarm
Limits Indicator
Lights to indicate that the attached OXIMAX sensor contains a patient sensor event record. The sensor event record information may be viewed or printed.
Fills in clockwise as the SatSeconds alarm management system detects a %SpO counterclockwise direction when %SpO
2 reading outside of the limit setting. Empties in
2 reading is within limits.
When the indicator is full, a medium priority alarm sounds. Determines the response time (2 to 4 seconds in fast mode and 5 to
7 seconds in normal mode) applied by the O calculation of SpO
2. The OXIMAX algorithm’s calculation of pulse
XIMAX algorithm in its
rate is unaffected by the response mode setting. The trending interval (2 seconds or 4 seconds) updated automatically by the monitor to roughly correspond with the SpO
2 calculation response
time. Displays when the alarm limits are set to neonate. No symbol
displays when the monitor is set to adult limits.
18 N-600x Operator’s Manual

About the Audible Indicators

Table 3: Audible Indicators
Function Description
Symbols, Controls, Displays and Indicators
Alarm Silence Reminder
Confirmation Tone Three beeps sound to indicate default settings
Invalid Button Press
Valid Button Press
High Priority Alarm
Medium Priority Alarm
Three beeps sound approximately every three minutes when alarms are silenced with the alarm silence duration set to OFF and the alarm silence reminder function is enabled.
have been saved or reset to factory defaults or trend data has been deleted.
Short, low been pressed that is inappropriate for the current state of the monitor.
Short, medium appropriate button has been pressed.
High loss
Note: If a High Priority Alarm is not silenced within 30 seconds by pressing the ALARM SILENCE Key, the monitor increases the urgency level of the audible alarm signal by alternating a piezo tone with the primary alarm tone. See Piezo Tone on page 20.
Medium
-pitched tone indicating a button has
-pitched tone indicating an
-pitched, fast-pulsing tone indicating
-of-pulse.
-pitched, pulsing tone indicating an
SpO2 or pulse rate limit violation.
Note: If a Medium Priority Alarm is not silenced within 2 minutes by pressing the ALARM SILENCE Key, the monitor increases the urgency level of the audible alarm signal by alternating a piezo tone with the primary alarm tone. See Piezo Tone on page 20.
N-600x Operator’s Manual 19
Symbols, Controls, Displays and Indicators
Table 3: Audible Indicators
Function Description
Low Priority Alarm
-pitched, slow-pulsing tone indicating an
Low
OXIMAX sensor disconnect, low battery, or
monitor failure. Note: If a Low Priority Alarm is not silenced
within 2 minutes by pressing the ALARM SILENCE Key, the monitor increases the urgency level of the audible alarm signal by alternating a piezo tone with the primary alarm tone. See Piezo Tone on page 20.
Piezo Tone A high-pitched piezo tone is sounded if there is
no user response to an audible alarm, or if the monitor detects a failure of the primary speaker. See High, Medium, and Low Priority Alarms in Ta ble 3.
Power
-On Self-Test
Pass
Pulse Beep Single beep sounds for each detected pulse.
Volume Setting Tone Continuous tone used when adjusting the alarm
One-second tone indicating the monitor has
been turned on and has successfully completed the power
The pitch of the pulse beep signal changes with a point level.
volume.
-on self-test.
-by-point rise or fall in the saturation
20 N-600x Operator’s Manual

Setting Up the Monitor

WARNING: To ensure patient safety, do not place the pulse oximeter in any position that might cause it to fall on the patient.
WARNING: As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
WARNING: Ensure that the speaker is clear of any obstruction. Failure to do so could result in an inaudible alarm tone.
WARNING: Disconnect the N-600x and Nellcor OXIMAX sensor from the patient during magnetic resonance imaging (MRI) scanning. Objects containing metal can become dangerous projectiles when subjected to the strong magnetic fields created by MRI equipment. Also, induced currents could potentially cause burns.
Setting Up the Monitor
WARNING: To ensure accurate performance and prevent device failure, do not subject the N-600x to extreme moistur e, such as direct exposure to rain. Such exposure may cause inaccurate performance or device failure.
WARNING: Do not use an N-600x pulse oximeter, OXIMAX sensor, cables, or connectors that appear damaged.
N-600x Operator’s Manual 21
Setting Up the Monitor
WARNING: Do not lift the pulse oximeter by the pulse oximetry cable or power cord because the cable or cord could disconnect from the pulse oximeter, causing the pulse oximeter to drop on the patient.
WARNING: The N-600x is not defibrillator-proof. However, it may remain attached to the patient during defibrillation or while an electrosurgical unit is in use, but the readings may be inaccurate during the defibrillation and shortly thereafter.
WARNING: In the USA, do not connect the pulse oximeter to an electrical outlet controlled by a wall switch, because the pulse oximeter may be accidentally turned off.
WARNING: Use only the Nellcor pulse oximetry cable DOC-10 with the N-600x pulse oximeter. Use of another pulse oximetry cable will have an adverse effect on performance. Do not attach any cable that is intended for computer use to the O
XIMAX sensor
port. Do not connect any device other than a Nellcor -approved OXIMAX sensor to the OXIMAX sensor connector.
WARNING: The N-600x should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the N-600x should be observed to verify normal operation in the configuration it is to be used.
Note: The monitor incorporates watchdog timers which reset the monitor in the event of software errors.
22 N-600x Operator’s Manual
Loading...
+ 182 hidden pages