Nellcor N-600x User manual

Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor,
OxiCliq, Dura-Y, MAX-FAST, and O
XIMAX are trademarks of Nellcor Puritan Bennett Inc.
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 Canada. To obtain information about a warranty, if any, contact Nellcor’s Technical Services Department,
Bennett patent to use the instrument with any sensor that is not manufactured or licensed by Nellcor Puritan Bennett.

Contents

Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .i
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Safety Warnings ................................................................................1
Safety Cautions .................................................................................3
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Intended Use .....................................................................................5
How to Use this Manual ....................................................................6
Symbols, Controls, Displays and Indicators . . . . . . . . . . . . . . . . . . . . . . . . . . 7
About the Front Panel .......................................................................7
About the Rear Panel ........................................................................8
About the Symbols ............................................................................9
About the Controls ..........................................................................10
About the Displays ..........................................................................11
Pleth Display .........................................................................11
Blip Display ..........................................................................12
Real-Time Trend Display ......................................................14
SpO2 and Pulse Rates .........................................................14
About the Visual Indicators .............................................................16
About the Audible Indicators ...........................................................19
Setting Up the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
List of Components .........................................................................23
Connecting to an AC Power Source ...................... ... ... ... ... .... ... ... ...24
Connecting an OXIMAX Sensor ...................... ... .... ... ... ... ... .... ... ... ...26
Operating the Battery . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Operating on Battery Power ......................... ...................................27
Low Battery Indicator ............... .... ... ................................................29
Description of Low and Critical Battery Conditions ...............30
Battery Fuel Gauge Indicator ................................................34
Using the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
Overview ......................................................................................... 35
Menu Description .. ... .... ... ... ... ... .... ... ... ....................................... ... ...35
Menu Structure .....................................................................36
Parameter Ranges ..........................................................................40
Turning On the Monitor ...................................................................43
N-600x Operator’s Manual i
Contents
OXIMAX Sensor Attached .................................................... 45
No OXIMAX Sensor Attached .............................................. 49
Turning the Backlight On or Off ...................................................... 50
Adjusting the Screen Contrast ........................................................ 50
Adjusting the Backlight Brightness ................................................. 50
Selecting the Pleth View ................................................................. 51
Selecting the Blip View ................................................................... 51
Selecting the Real-Time Trend View .............................................. 52
Selecting the Trend Data Display ......................................... 53
Setting the Trend Time Scale Display .................................. 53
Setting the Trend Amplitude Scale Display .......................... 54
Setting the Pulse Beep Volume .......................................... ... ... ... ... 54
Setting the Alarm Volume .. .... ... ... ... ... .... ... ... ... .... ... ... ... ... .... ... ... ... ... 55
Setting the Date and Time...................................... ... ... ... .... ... ... ... .. 56
Setting the Alarm Silence Duration ........................... ... ... .... ... ... ... ... 58
Disabling Audible Alarms ................................................................ 59
Selecting the Standby Mode ........................................................... 60
Adult-Pediatric or Neonatal Settings ............................ ................... 61
Setting Patient Adult-Pediatric or Neonatal Modes .............. 62
Alarm Limit Changed Indicator ....................................................... 63
Setting Alarm Limits ............................... ... ... ... .... ... ... ... ... ................ 63
Setting SatSeconds Alarm Limit ........ .... ... ... ... .... ... ... ... ... .... ... ... ... ... 65
Setting Monitor Response Mode .............. ... ... .... ... ... ... ... .... ... ... ... ... 66
Selecting the Display Language ..................................................... 68
OXIMAX Sensor Messages ............................................................ 69
OXIMAX Sensor Adjust Condition Messages ...................... 70
OXIMAX Sensor Adjust Messages ....................................... 71
Using Monitor Trend Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .73
Overview ................................... ................................................ ...... 73
Storing Trend Data ............................................................... 75
OXIMAX Sensor Type .................................................................... 76
Selecting the Trend Data Display Scale ......................................... 76
Reading the Trend Data Display ............................ ................... ...... 78
Dual Trend Data Display ................................................................. 79
SpO2 Trend Display ....................................................................... 80
Pulse Rate Trend Display ............................................................... 80
Histogram Trend Data Display .................. ... ... .... ... ... ... ... .... ... ... ... ... 81
Pulse Amplitude Trend Data Display .............................................. 82
Clearing Trend Information ............................................................. 83
Using OXIMAX Sensor Event Records . . . . . . . . . . . . . . . . . . . . . . . . . . . . .85
Overview ................................... ................................................ ...... 85
Setting up OXIMAX Sensor Messages ........................... .... ... ... ... ... 87
ii
Setting In-Sensor Data Type ...........................................................88
OXIMAX Sensor Data Type ............................................................89
OXIMAX Sensor Event Record Data Available ...............................90
OXIMAX Sensor Event Record Not Available .................................91
OXIMAX Sensor Event Record Graphical Data ........................... ...92
Viewing and Printing OXIMAX Sensor Event History Data .............94
OXIMAX Sensor Tabular Event Data ..............................................96
Viewing and Printing In-Sensor Tabular Event History Data ...........97
Printing Monitor Trend Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
Overview ......................................................................................... 99
Printing ..................................... .................... ................... ......99
Monitor Trend Data in ASCII Mode ...............................................102
Trend Data in Graph Mode ...........................................................103
Real-Time Display/Printout Format ...............................................103
Column Headings ...............................................................105
Data Source ........................................................................105
Software Version ........ ... .... ... ... ... .... ....................................106
Alarm Limits ........................................................................106
Monitor Mode ........................ ... ... .... ... ... ... .... .......................106
Response Mode ..................................................................107
Data Column Headings .......................................................107
Time ...................................... ..............................................108
Patient Data .................................... ... ... ... .... .......................108
Operating Status ......................... .... ... ... ... .... ... ... ... ... .... ... ... .108
Using the Data Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Overview ....................................................................................... 111
Connecting to the Data Port ..........................................................112
Data Port Pinouts ................................................................112
Data Port Setup .............................................................................114
Using the Nurse Call Interface ......................................................116
Setting Nurse Call RS-232 Polarity .......... .... ... ... ... ... .... ... ... .117
Setting Nurse Call Relays Normally Open/Closed ..............118
Calculating the Analog Voltage Output .........................................118
OXIMAX Sensors and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 121
Overview ....................................................................................... 121
Selecting an OXIMAX Sensor .......................................................122
OXIMAX Sensor Features .............................................................126
Biocompatibility Testing ................................................................126
Optional Accessories ................... .......................................... ... ... .127
GCX Mounting Plate ...........................................................128
GCX Vertical Wall Mount Arm ............................................129
N-600x Operator’s Manual iii
Contents
GCX Roll Stand .................................................................. 130
Soft-Sided Carrying Case ................................................... 131
Performance Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .133
Overview ................................... ................................................... .133
Performance Considerations .......................................... .... ... ....... 134
Dysfunctional Hemoglobins ................................................ 134
Anemia ............................................................................... 135
Saturation ........................................................................... 135
Pulse Rates ........................................................................ 135
OXIMAX Sensor Performance Considerations ............................. 135
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .139
Overview ................................... ................................................... .139
On-Screen Help ..................................... ... .................................... 140
Accessing Multiple Topics .................................................. 140
Accessing Single Topics .................................................... 143
Error Codes .................................................................................. 145
Prompts and Error Messages ....................................................... 148
Primary Speaker Failure ............................................................... 151
Help and Support ................................... ... ... ... .... ... ... ... ... .... .......... 153
EMI (Electromagnetic Interference) .............................................. 158
Obtaining Technical Assistance .................................................... 159
Returning your Monitor ....................................................... 159
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .161
Overview ................................... ................................................... .161
Service ................................... ................................................ ....... 161
Periodic Safety Checks ................................................................. 161
Cleaning ....................................... ................... ....................... ....... 162
Using SatSeconds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .163
Overview ................................... ................................................... .163
SatSeconds “Safety Net” .............................................................. 165
SatSeconds Display ...................................................................... 165
Factory Default Settings . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .167
Overview ................................... ................................................... .167
Neonate Default Settings .............................................................. 167
Adult Default Settings ................................................................... 168
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .171
Overview ................................... ................................................... .171
Automatic Calibration ......................................................... 172
iv
Functional versus Fractional Saturation .............................173
Measured versus Calculated Saturation .............................174
OXIMAX Technology .....................................................................175
Functional Testers and Patient Simulators ...................................176
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 177
Performance .................................... ................... .................... .......177
Electrical ....................................... ................ ................ ................ .178
Environmental Conditions ..... ... .... .......................................... ... ... .180
Physical Characteristics ................................................................182
Compliance ................................................................................... 182
Manufacturer’s Declaration ...........................................................183
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 197
N-600x Operator’s Manual v
Contents
vi

Safety Information

Safety Warnings

Warnings are identified by the WARNING symbol shown above.
Warnings alert you to potential serious outcomes (death, injury, or adverse events) to the patient or user.
WARNING: The sensor extrapolates from the date and time provided by the N-600x when recording the sensor event record to the sensor . The accuracy of th e date/time is the responsibility of the N-600x. It is recommended that the N-600x user set the time/ date to the correct value before a sensor event record-enabled sensor is connected, and that this date/time not be changed while the sensor remains connected. Since a sensor with sensor event record data can be transported from one monitor to another, having discrepancies in the date/time between monitors and the sensor event record data will affect the order the sensor event record data appears. To eliminate this possible problem, all monitors within an institution should be set to the same time.
Safety Information
N-600x Operator’s Manual 1
Safety Information
WARNING: Explosion hazard. Do not use the N-600x pulse oximeter in the presence of flammable anesthetics or gases.
WARNING: Chemicals from a broken LCD display panel are toxic when ingested. Use caution when handling a pulse oximeter with a broken display panel.
WARNING: Pulse oximetry readings and pulse signals can be affected by certain environmental conditions, O
XIMAX sensor
application errors, and certain patient conditions. See the appropriate sections of this manual for specific safety information.
WARNING: The use of accessories, sensors, and cables other than those specified may result in increased emission and/or decreased immunity and inaccurate readings of the N-600x pulse oximeter.
WARNING: Failure to cover the OXIMAX sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.
2 N-600x Operator’s Manual

Safety Cautions

Safety Information
Cautions are identified by the CAUTION symbol shown above.
Cautions alert you to exercise care necessary for the safe and effective use of the N-600x pulse oximeter.
Caution: When connecting the N-600x to any instrument, verify proper operation before clinical use. Both the N-600x and the instrument connected to it must be connected to a grounded outlet. Accessory equipment connected to the pulse oximeter’s data interface must be certified according to IEC Standard 60950 for data-processing equipment or IEC Standard 60601-1 for electromedical equipment. All combinations of equipment must be in compliance with IEC Standard 60601-1-1 systems requirements. Anyone who connects additional equipment to the signal input port or signal output port (N-600x data port connector) configures a medical system and is therefore responsible for ensuring that the system complies with the requirements of system standard IEC Standard 60601-1-1 and the electromagnetic compatibility system standard IEC S tandard 60601- 1-2. The N-600x accuracy may degra de if it is connected to secondary I/O devices when the instrument is not connected to earth reference.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
Caution: Dispose of battery in accordance with local requirements and regulations.
N-600x Operator’s Manual 3
Safety Information
4 N-600x Operator’s Manual

Introduction

WARNING: The N-600x is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.

Intended Use

The N-600x pulse oximeter is indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO with neonatal, pediatric, and adult patients who are well or poorly perfused, in hospitals, hospital-type facilities, intra-hospital transport, and home environments. For prescription use only.
Introduction
2) and pulse rate. The N-600x is intended for use
Note:Hospital use typically covers such areas as general care floors, operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility.
Note: Home Care use is defined as managed or used by a lay person (parent or other similar noncritical caregiver) in the home environment.
Use with any particular patient requires the selection of an appropriate oxygen O
XIMAX sensor as described in this manual.
N-600x Operator’s Manual 5
Introduction

How to Use this Manual

All users should read this manual thoroughly . More experienced users of the N-600x can directly go to the topics for the information they require.
This manual is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
6 N-600x Operator’s Manual

Symbols, Controls, Displays and Indicators

Symbols, Controls, Displays and Indicators

About the Front Panel

1. SpO2 OXIMAX Sensor Port, page 26.
2. Low Battery Indicator, page 17. 13. ADJUST DOWN Button, page 10.
3. AC Power Indicator, page 16. 14. Neonate Mode Indicator, page 18.
4. ON/STANDBY Button, page 10.
5. Battery Fuel Gauge Indicator, page 34. 16. Fast Response Mode Indicator, page 18.
6. Waveform Display, page 12. 17. Softkeys, page 11.
7. SatSeconds
8. %SpO
9. Pulse Rate Display, page 16. 20. Interference Indicator, page 17.
10. Alarm Silence Indicator, page 17. 21. Pulse Search Indicator, page 17.
11. ALARM SILENCE Button, page 10. 22. Monitor Speaker.
TM
Indicator, page 18.
2 Display, page 16. 19. Data In-Sensor Indicator, page 18.
12. ADJUST UP Button, page 10.
15. HELP/CONTRAST Button, page 11.
18. Menu Bar, page 11.
Figure 1: Front Panel Buttons and Symbols
N-600x Operator’s Manual 7
Symbols, Controls, Displays and Indicators

About the Rear Panel

1. Equipotential Terminal (Ground).
2. AC Power Connector, page 24.
3. Data Port Connector, page 112.
4. Fuse Holder.
5. Supply Voltage Selector Switch, page 24.
Figure 2: Rear Panel Components
8 N-600x Operator’s Manual

About the Symbols

The symbols, located on the rear panel of the N-600x, are as follows.
Symbols, Controls, Displays and Indicators
Table 1: Symbols and Descriptions
Symbol
Description
See Instructions for Use
Fuse Replacement
Equipotential Terminal (ground)
Date of Manufacture
Data Interface
Type BF Applied Part - Not defibrillator proof
N-600x Operator’s Manual 9
Symbols, Controls, Displays and Indicators

About the Controls

ON/STANDBY Button Turns the monitor on and off.
Note: Pressing a button, except the ON/STANDBY button, should result in either a valid or an invalid key tone (refer to Table 3). If the key pressed fails to emit a tone, contact qualified service personnel.
ALARM SILENCE Button Silences current alarms for the alarm silence duration period. When an alarm has been silenced, pressing the button again reactivates, or “unsilences” the alarm. It is also used to view and adjust alarm silence duration and alarm volume. The ALARM SILENCE button clears “SENSOR OFF,” “LOW BATTERY,” and “SENSOR DISCONNECT” messages from the display.
ADJUST UP Button Increases variable parameters of the monitor.
ADJUST DOWN Button Decreases variable parameters in the monitor.
10 N-600x Operator’s Manual
HELP/CONTRAST Button Enables you to access the on-screen help and adjust the monitor screen contrast.
Softkey Menu Bar
Displays the current functions for each of the four softkey buttons.

About the Displays

Symbols, Controls, Displays and Indicators
Pressing and releasing the HELP/CONTRAST button launches the on-screen help.
Pressing and holding the HELP/CONTRAST button while simultaneously pressing the ADJUST UP and ADJUST DOWN buttons lightens or darkens the display screen.

Pleth Display

The pleth display is user selectable. Refer to Selecting the Pleth View on page 51.
N-600x Operator’s Manual 11
Symbols, Controls, Displays and Indicators
The pleth display includes a “wiper bar” plethysmographic waveform, menu bar, and current measured %SpO SatSeconds are enabled, the pleth display includes the SatSeconds indicator and SatSeconds setting. A decimal point after the %SpO pulse rate indicates the respective limits have been changed from the power on defaults (Overview on page 73). Plethysmographic waveforms with peak to peak amplitudes less than ten PAUs are associated.
Caution: Verify the movement of the blip bar or plethys­mographic waveform or beating heart before accepting any displayed data as a current measurement.
When the monitor is powered by the internal battery, the pleth display includes a horizontal battery fuel gauge positioned in the upper left corner which shows the remaining charge (operating hours) on the battery. If a monitor reporting low battery is connected to an AC power source, the battery fuel gauge displays the charging progress. The battery fuel gauge is cleared from the display once the monitor can provide at least 15 minutes of operating time.
2 and pulse rate. If
2 or

Blip Display

The blip display includes a pulse amplitude blip bar, current measured %SpO
2 and pulse rate, and current upper and lower %SpO2 and pulse
rate limits. If SatSeconds are enabled, the blip display includes the SatSeconds indicator and SatSeconds setting. Decimal points after the %SpO
2 or pulse rate indicate that the respective limits have been
changed from the power-on defaults.
12 N-600x Operator’s Manual
Symbols, Controls, Displays and Indicators
When the monitor is powered by internal battery, the blip display includes a horizontal battery fuel gauge positioned in the upper left corner that shows the remaining charge (operating hours) on battery. If a monitor reporting a low battery is connected to AC power, the battery fuel gauge shows the charging progress. The battery fuel gauge is cleared from the display once the monitor can provide at least 15 minutes of operating time on battery.
N-600x Operator’s Manual 13
Symbols, Controls, Displays and Indicators

Real-Time Trend Display

The real-time trend display includes %SpO2 and/or pulse rate trend data plots and current measured %SpO2 and pulse rates. The trend data plots are automatically updated as each new trend point is calculated, where the interval between calculations is based on the display time scale selected. If SatSeconds is enabled, the real-time trend display includes the SatSeconds indicator. Decimal points after the displayed %SpO2 or pulse rate indicate that the respective limits have been changed from the power-on defaults. Each time a pulse is detected by the oximeter, a heart icon flashes.

SpO2 and Pulse Rates

There are various matrixes within the N-600x algorithm. Some of these are used to assess the severity of conditions presented to the N-600x in measuring SpO individual matrices or combinations of these matrices are used to drive the LED indicators on the N-600x front panel.
14 N-600x Operator’s Manual
2 and pulse rate on a patient. These
Symbols, Controls, Displays and Indicators
The N-600x algorithm automatically extends the amount of data required for measuring SpO
2 and pulse rate depending on the
measurement conditions. During normal measurement conditions the averaging time is 6 to 7 seconds. During conditions such as those caused by low perfusion, interference (e.g., external interference like ambient light), or a combination of these, the N-600x algorithm automatically extends the amount of data required beyond 7 seconds. If the resulting dynamic averaging time exceeds 20 seconds, the pulse search indicator is lit solid and SpO
2 and Pulse Rate will continue to
be updated every second. As these conditions extend, the amount of data required continues to increase. If the dynamic averaging time reaches 40 seconds, the pulse search indicator begins flashing, the SpO
2 and pulse rate displays flash zeros indicating a loss-of-pulse
condition.
WARNING: Failure to cover the OXIMAX sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.
N-600x Operator’s Manual 15
Symbols, Controls, Displays and Indicators

About the Visual Indicators

Table 2: Visual Indicators
Indicator Description
%SpO2 Display Shows the hemoglobin oxygen saturation level. The display value
flashes zeros during loss-of-pulse alarms and flashes the SpO when the SpO monitor continues to update the display. If alarm limits have been changed from their power-on defaults, a decimal point (.) is displayed after the SpO
2 is outside the alarm limits. During Pulse Search, the
2 value (81.).
2 value
Pulse Amplitude
Indicator (blip bar)
Indicates pulse beat and shows the relative (non-normalized) pulse amplitude. As the detected pulse becomes stronger, more bars light with each pulse. This indicator is available only in the blip view.
Plethysmographic
Waveform Display
Displays a non-normalized waveform in real-time sensor signals. The relative pulsatile strength and quality of the incoming signals can be observed.
Pulse Rate Display Displays the pulse rate in beats per minute. It flashes during
loss-of-pulse alarms and when the pulse rate is outside of the alarm limits. During Pulse Search, the monitor continues to update the display. Pulse rates outside of the pulse rate range of 20 to 250 bpm are displayed as 0 and 250, respectively. If alarm limits have been changed from their power-on defaults, a decimal point (.) is displayed after the BPM value (112.).
AC Power
Indicator
Lights continuously when the N-600x is connected to an AC power source. The indicator shows that the battery is charging. It is off when the monitor is being powered by internal battery.
16 N-600x Operator’s Manual
Table 2: Visual Indicators
Indicator Description
Symbols, Controls, Displays and Indicators
Low Battery
Indicator
Battery Fuel Gauge
Indicator
Alarm Silence
Indicator
Interference
Indicator
Lights continuously when 15 or fewer minutes of battery capacity remain. Flashes when the battery capacity reaches a critically low condition.
Displays the battery charge remaining on the monitor. The battery fuel gauge consists of four bars, each corresponding to approximately 1.5 hours of operating time. All four bars are lit when the battery is fully charged. No bars are lit when a low battery condition exists. See Battery Fuel Gauge Indicator on page 34.
Lights continuously when an audible alarm has been silenced. It flashes when the alarm silence duration has been set to Off.
Lights whenever the N-600x algorithm detects the incoming signal quality is degraded.
Note: Degradation can be caused by ambient light, electrical noise, electro-surgical interference, patient activity, or other causes.
An intermittently lit Interference Indicator is common during patient monitoring, and indicates the N-600x algorithm is dynamically adjusting the amount of data required for measuring
2 and Pulse Rate. When lit continuously, the N-600x algorithm
SpO has extended the amount of data required for measuring SpO
2 and
Pulse Rate and consequently fidelity in tracking rapid changes in these values may be reduced.
Pulse Search
Indicator
Lights continuously prior to initial acquisition of a pulse signal and during prolonged and challenging monitoring conditions. The pulse search indicator flashes during a loss-of-pulse signal.
N-600x Operator’s Manual 17
Symbols, Controls, Displays and Indicators
Table 2: Visual Indicators
Indicator Description
Data In-Sensor
Indicator
SatSeconds
Indicator
Fast Response
Mode Indicator
Neonate Alarm
Limits Indicator
Lights to indicate that the attached OXIMAX sensor contains a patient sensor event record. The sensor event record information may be viewed or printed.
Fills in clockwise as the SatSeconds alarm management system detects a %SpO counterclockwise direction when %SpO
2 reading outside of the limit setting. Empties in
2 reading is within limits.
When the indicator is full, a medium priority alarm sounds. Determines the response time (2 to 4 seconds in fast mode and 5 to
7 seconds in normal mode) applied by the O calculation of SpO
2. The OXIMAX algorithm’s calculation of pulse
XIMAX algorithm in its
rate is unaffected by the response mode setting. The trending interval (2 seconds or 4 seconds) updated automatically by the monitor to roughly correspond with the SpO
2 calculation response
time. Displays when the alarm limits are set to neonate. No symbol
displays when the monitor is set to adult limits.
18 N-600x Operator’s Manual

About the Audible Indicators

Table 3: Audible Indicators
Function Description
Symbols, Controls, Displays and Indicators
Alarm Silence Reminder
Confirmation Tone Three beeps sound to indicate default settings
Invalid Button Press
Valid Button Press
High Priority Alarm
Medium Priority Alarm
Three beeps sound approximately every three minutes when alarms are silenced with the alarm silence duration set to OFF and the alarm silence reminder function is enabled.
have been saved or reset to factory defaults or trend data has been deleted.
Short, low been pressed that is inappropriate for the current state of the monitor.
Short, medium appropriate button has been pressed.
High loss
Note: If a High Priority Alarm is not silenced within 30 seconds by pressing the ALARM SILENCE Key, the monitor increases the urgency level of the audible alarm signal by alternating a piezo tone with the primary alarm tone. See Piezo Tone on page 20.
Medium
-pitched tone indicating a button has
-pitched tone indicating an
-pitched, fast-pulsing tone indicating
-of-pulse.
-pitched, pulsing tone indicating an
SpO2 or pulse rate limit violation.
Note: If a Medium Priority Alarm is not silenced within 2 minutes by pressing the ALARM SILENCE Key, the monitor increases the urgency level of the audible alarm signal by alternating a piezo tone with the primary alarm tone. See Piezo Tone on page 20.
N-600x Operator’s Manual 19
Symbols, Controls, Displays and Indicators
Table 3: Audible Indicators
Function Description
Low Priority Alarm
-pitched, slow-pulsing tone indicating an
Low
OXIMAX sensor disconnect, low battery, or
monitor failure. Note: If a Low Priority Alarm is not silenced
within 2 minutes by pressing the ALARM SILENCE Key, the monitor increases the urgency level of the audible alarm signal by alternating a piezo tone with the primary alarm tone. See Piezo Tone on page 20.
Piezo Tone A high-pitched piezo tone is sounded if there is
no user response to an audible alarm, or if the monitor detects a failure of the primary speaker. See High, Medium, and Low Priority Alarms in Ta ble 3.
Power
-On Self-Test
Pass
Pulse Beep Single beep sounds for each detected pulse.
Volume Setting Tone Continuous tone used when adjusting the alarm
One-second tone indicating the monitor has
been turned on and has successfully completed the power
The pitch of the pulse beep signal changes with a point level.
volume.
-on self-test.
-by-point rise or fall in the saturation
20 N-600x Operator’s Manual

Setting Up the Monitor

WARNING: To ensure patient safety, do not place the pulse oximeter in any position that might cause it to fall on the patient.
WARNING: As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
WARNING: Ensure that the speaker is clear of any obstruction. Failure to do so could result in an inaudible alarm tone.
WARNING: Disconnect the N-600x and Nellcor OXIMAX sensor from the patient during magnetic resonance imaging (MRI) scanning. Objects containing metal can become dangerous projectiles when subjected to the strong magnetic fields created by MRI equipment. Also, induced currents could potentially cause burns.
Setting Up the Monitor
WARNING: To ensure accurate performance and prevent device failure, do not subject the N-600x to extreme moistur e, such as direct exposure to rain. Such exposure may cause inaccurate performance or device failure.
WARNING: Do not use an N-600x pulse oximeter, OXIMAX sensor, cables, or connectors that appear damaged.
N-600x Operator’s Manual 21
Setting Up the Monitor
WARNING: Do not lift the pulse oximeter by the pulse oximetry cable or power cord because the cable or cord could disconnect from the pulse oximeter, causing the pulse oximeter to drop on the patient.
WARNING: The N-600x is not defibrillator-proof. However, it may remain attached to the patient during defibrillation or while an electrosurgical unit is in use, but the readings may be inaccurate during the defibrillation and shortly thereafter.
WARNING: In the USA, do not connect the pulse oximeter to an electrical outlet controlled by a wall switch, because the pulse oximeter may be accidentally turned off.
WARNING: Use only the Nellcor pulse oximetry cable DOC-10 with the N-600x pulse oximeter. Use of another pulse oximetry cable will have an adverse effect on performance. Do not attach any cable that is intended for computer use to the O
XIMAX sensor
port. Do not connect any device other than a Nellcor -approved OXIMAX sensor to the OXIMAX sensor connector.
WARNING: The N-600x should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the N-600x should be observed to verify normal operation in the configuration it is to be used.
Note: The monitor incorporates watchdog timers which reset the monitor in the event of software errors.
22 N-600x Operator’s Manual

List of Components

Setting Up the Monitor
Quantity Item
1 1 Nellcor O
1 1
1 Power Cord (applicable to country of sale) 2
1 Quick Guide
-600x Pulse Oximeter
N
XIMAX Sensor or Assortment Pack
DOC
-10 Pulse Oximetry Cable
N
-600x Operator’s Manual (applicable to
country of sale) and/or compact disc
Fuses, 0.5 A, 250 volts, slow (5 x 20 mm)
-blow, IEC
N-600x Operator’s Manual 23
Setting Up the Monitor

Connecting to an AC Power Source

WARNING: In the USA, do not connect the pulse oximeter to an electrical outlet controlled by a wall switch, because the pulse oximeter may be accidentally turned off.
Caution: The Supply Voltage Selector switch must be set to the correct voltage (115 or 230) to avoid equipment damage and ensure battery charging.
Caution: Use only the hospital-grade power cord provided by Nellcor.
1. Set the Supply Voltage Selector switch to the applicable voltage.
2. Plug the female connector end of the power cord into the power connector on the rear of the monitor.
3. Plug the male connector of the power cord into a properly grounded AC outlet.
24 N-600x Operator’s Manual
Setting Up the Monitor
Caution: Ensure the pulse oximeter is properly grounded when operating on AC power. If you are uncertain whether the AC outlet is properly grounded, disconnect the pulse oximeter from the outlet and use the battery power. Contact a qualified electrician to examine the outlet for ground connections.
4. Verify the monitor’s AC power indicator is lit.
Note: If the AC power indicator is not lit, check the:
power cord
supply voltage selector switch
•user-accessible fuses
AC power outlet
Note: The monitor can be operated with a depleted battery when connected to an AC power outlet. A warning message displays and must be cleared by pressing the ALARM SILENCE button before the monitor can be used for patient monitoring.
N-600x Operator’s Manual 25
Setting Up the Monitor

Connecting an OXIMAX Sensor

The OXIMAX sensor type is shown at the bottom of the display when
XIMAX sensor is connected to the N-600x or when the monitor
an O completes POST with an O
Note: Sensor LED light emissions fall within Class 1 level, according to IEC 60825-1:2001.
XIMAX sensor attached.
Caution: Use only Nellcor-approved O
XIMAX sensors and pulse
oximetry cables.
Note:Physiological conditions, medical procedures, or external
agents that may interfere with the monitor’s ability to detect and display measurements include dysfunctional hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring agents, such as nail polish, dye, or pigmented cream.
1. Firmly connect a DOC-10 pulse oximetry cable to the SpO
XIMAX Sensor Port of the monitor.
O
2
2. Connect a Nellcor O
XIMAX SpO2 sensor to the opposite end of the
DOC-10 pulse oximetry cable.
26 N-600x Operator’s Manual

Operating the Battery

WARNING: Dispose of battery in accordance with local requirements and regulations.

Operating on Battery Power

The N-600x monitor has an internal battery that can be used to power the monitor during transport or when AC power is not available. A new, fully charged battery provides at least 7 hours of monitoring time under the following conditions:
No audible alarms sound
No analog or serial output devices are attached to the N-600x
Operating the Battery
Default display brightness setting
The monitor cannot be used when the battery is depleted unless the monitor is connected to an AC power source. A warning message displays and must be cleared by pressing the ALARM SILENCE button before the monitor can be used for patient monitoring.
The pleth and blip displays include a battery fuel gauge indicator that shows the remaining charge (operating hours). When the monitor is fully charged, all four bars are lit on the indicator.
Caution: If the N-600x monitor is stored for a period of three months or longer, notify service personnel to place the monitor in "Shelf-mode" prior to storage. The monitor can be placed in "Shelf-mode" by qualified service personnel using the procedures indicated in the N-600x Service Manual. Recharge the battery when it has not been charged for three or more months.
N-600x Operator’s Manual 27
Operating the Battery
Caution: Replace the battery if fewer than four bars are lit after fully charging the battery. To charge a low or fully depleted battery, connect the monitor to AC power outlet. A full charge of a fully depleted battery takes 8 hours to charge while the monitor is turned off. A full charge of a fully depleted battery takes 12 hours while the monitor is in the normal operating mode.
Caution: If the monitor is operated on an AC power source with a depleted battery and the AC power is subsequently lost, the monitor will shut down immediately.
When all of the following conditions are present for 15 minutes, the N-600x automatically shuts down:
Monitor is running on battery power
No buttons have been pressed
No pulse has been detected (for example, when a patient is not connected to the O
XIMAX sensor or the OXIMAX
sensor is disconnected from the monitor)
No alarms are present (other than low battery or a non-correctable error)
Note:Whenever the monitor is connected to AC power source, the battery is being charged. We recommend the monitor remain connected to an AC power source when not in use. This ensures a fully-charged battery when the monitor is needed.
28 N-600x Operator’s Manual

Low Battery Indicator

The Low Battery Indicator lights and a low priority alarm begins to sound when approximately 15 minutes of monitoring time is available on the existing battery charge. Refer to Table 4 for a description of the low and critical battery conditions.
A low battery audible alarm can be cancelled by pressing the ALARM SILENCE button. The low battery indicator and display screen message continues to display. Connecting the monitor to an AC power source silences the audible alarm, but the low battery indicator remains lit as long as the battery is in the low voltage condition. After the 15-minute period of low battery condition, a high priority alarm sounds for about 10 seconds before the monitor shuts off.
If the monitor backlight is turned off during a low battery condition, the backlight cannot be turned back on.
Operating the Battery
Nellcor recommends that a qualified service personnel replace the internal battery every 24 months. Replaced batteries should be disposed of in accordance with local ordinances.
Caution: The pulse oximeter default settings will return to factory default setting if the battery becomes fully discharged or is replaced. Qualified service personnel will have to reset the institutional defaults, following the instructions in the N-600x Service Manual.
Note: If the AC voltage selector switch on the monitor rear panel
does not match your AC voltage source, the monitor may run on battery power, even though it is connected to an AC power source, which eventually results in a low priority alarm and a lighted low battery indicator. Ensure that the switch setting matches your AC voltage.
Note: As the battery is used and recharged over time, the amount of time between the onset of the low battery alarm and the instrument shut-off may become shorter.
N-600x Operator’s Manual 29
Operating the Battery

Description of Low and Critical Battery Conditions

Table 4: Low and Critical Battery Conditions
State
1No No Yes
2No No No
Critical Battery
Low Battery
AC Power
Operation
2-normal
SpO AC/Battery charge LED LOW BATTERY LED LOW BATTERY
message Audible alarm Error code Effect of ALARM
SILENCE key Shutdown SpO AC/Battery charge LED LOW BATTERY LED LOW BATTERY
message
-off
-off
-none
-normal
-N/A
2-normal
-off
-on
-off
-off
-off
Audible alarm Error code Effect of ALARM
SILENCE key Shutdown
30 N-600x Operator’s Manual
-off
-none
-normal
- N/A
Operating the Battery
State
3No Yes No
4No Yes Yes
Critical Battery
Low Battery
AC Power
Operation
2-normal
SpO AC/Battery charge LED LOW BATTERY LED LOW BATTERY
message Audible alarm Error code Effect of ALARM
SILENCE key silences audio alarm, second press cancels LOW BATTERY message (LED) stays on until Low Battery Condition is corrected.
Shutdown SpO
-on
-low priority
-logged
-First press
-Imminent
2-normal
-off
-on
AC/Battery charge LED LOW BATTERY LED LOW BATTERY
message Audible alarm Error code Effect of ALARM
SILENCE key stays on)
Shutdown
-off
-off
-logged
-N/A (LED
-N/A
-on
-on
Note: Connecting AC functions the same as ALARM SILENCE key in state 3.
N-600x Operator’s Manual 31
Operating the Battery
State
Critical Battery
Low Battery
AC Power
5 Not used 6Yes Yes No
Operation
2-not displayed
SpO AC/Battery charge LED LOW BATTERY LED
(flashing) LOW BATTERY
message Audible alarm Error code
-on
-high priority
-displayed and
logged Effect of ALARM
SILENCE key Shutdown
-none
-after 10 seconds
-off
-on
32 N-600x Operator’s Manual
Operating the Battery
State
Critical Battery
Low Battery
AC Power
Operation
7 Yes Yes Yes SpO2 - displayed.
AC/Battery Charge LED ­on
LOW BATTERY LED-on (flashing)
LOW BATTERY message ­on
The Battery Fuel Gauge Indicator shows a fully depleted battery (no bars lit).
Warning message in the pleth window: UNIT WILL SHUT DOWN IF AC POWER LOST
Audio alarm - low priority Error code - logged Affect of Silence key - One
press silences the audible alarm. Pressing the Affect Silence key twice cancels the LOW BATTERY message, removes the warning message and restores default Pleth (or Blip) display (LED continues to FLASH until Low Battery condition is not true, Battery Fuel Gauge Indicator shows charging progress)
Shutdown - N/A
N-600x Operator’s Manual 33
Operating the Battery

Battery Fuel Gauge Indicator

The N-600x has a battery fuel gauge indicator which displays the battery power remaining on the monitor . The indicator appears on the the pleth and blip display screens. When the monitor is fully charged, all four bars are lit on the battery fuel gauge indicator. The battery fuel gauge indicator capacities are described below.
Table 5: Battery Fuel Gauge Indicator Levels
Level Description
Indicates 100% battery capacity remaining.
Indicates 75% battery capacity remaining.
Indicates 50% battery capacity remaining.
Indicates 25% battery capacity remaining.
Indicates 0% battery capacity remaining.
Note: The levels in Table 5 are based on a brand new battery. As a battery is used and recharged over time, it may provide only 75% of the capacity of brand new battery. For example, a battery that is two­years old may provide only 75% (3 bars) of the capacity of a new battery.
Caution: If the battery is fully depleted, and the AC power is lost, the monitor will shut down.
34 N-600x Operator’s Manual

Using the Monitor

Overview

This section describes menu navigation, power on/off and display options, parameter ranges, O configuring default settings suitable for your environment.

Menu Description

The N-600x is outlined below. You can choose the type of trend data to view by selecting either Monitor trend or Sensor trend data in the Trend menu. Sen sor sub-menu choices dif fer depending o n what typ e of in-sensor data is stored in the sensor chip, such as, event or loop.
Using the Monitor
XIMAX sensor attachments, and
The menu structure includes BACK softkey options that enable you to move back to the previous menu level without exiting the Trend menu entirely. Trend data must be compiled on entry/reentry to the Trends menu. When the softkeys are available, both BACK and EXIT options are available. The BACK softkey goes to the previous level and the EXIT softkey goes to the main menu. If only one space is available the BACK softkey is included, this may require going back one or two levels to get to an EXIT softkey.
The BACK and EXIT softkeys are positioned on the right-most softkeys, respectively. The below menu structure identifies:
BOLDFACE TYPE — softkey title as displayed on the
monitor
Underlined Text
Italicized Text — the destination of the BACK and EXIT softkeys
N-600x Operator’s Manual 35
— description of the softkey menu item
Using the Monitor

Menu Structure

Main Menu
LIMITS (Limits Menu)
- SELECT
- NEO
- ADULT
EXIT (to Main menu)
TREND (Trend Menu)
- MON (Monitor Menu)
--VIEW (Monitor Trend View Menu)
---DUAL
---SPO2
---PULSE
---NEXT (History/Amplitude Menu)
----HIST (Delete/Print2 Menu)
-----DELETE (Delete Trends)
------“DELETE TRENDS
-------YES (return to Main menu)
-------NO (back to Delete/Print menu)
-----PRINT
-----BACK (back to Hist/Amp menu)
-----EXIT (to Main menu)
----AMP (Amplitude Menu)
-----BACK (back to Hist/Amp menu)
-----EXIT (to Main menu)
----BACK (back to Monitor Trend View menu)
----EXIT (to Main menu)
--ZOOM (Monitor Trend Zoo m Menu)
---TIME (Cycle through 48h, 36h, 24h, 12h, 8h, 4h, 2h,
1h, 30m, 15m, 40s, 20s for current view)
---SCALE (Cycle through ±5, ±10, ±15, ±20, ±25, ±30, ±35, ±40 and ±50 (units of BPM or %SpO
depending on the data displayed) of the max and
min. values under the cursor, default to 10 to 100 for SAT trend graph and 5-250 for Pulse trend graph if there is no data point under the cursor for current view)
---AUTO (Based on all of the graphed trend data: maximum value, rounded up to nearest multiple of 10, minimum value, rounded down to nearest multiple of 10 minus 10)
---BACK (back to Monitor menu)
2,
36 N-600x Operator’s Manual
Using the Monitor
Menu Structure
(continued...)
--NEXT (Delete/Print1 Menu)
---DELETE
----“DELETE TRENDS?
-----YES (to Main menu)
-----NO (back to Delete/Print1 menu)
---PRINT
---BACK (back to Monitor menu)
---EXIT (to Main menu)
--BACK (back to Trend menu)
- SENSOR (Sensor/Event Menu) (if Event data is in the sensor, the following menu, the Screen will
remain in the appropriate state until the next menu selection is made)
--GRAPH (Graph Menu)
chronological order; up/down also scroll through events in order)
---< (show previous graph, only available when there is a
previous graph)
---> (show next graph, only available when there is a next
graph)
---PRINT
---BACK (back to Sensor menu)
--TABLE (Table Menu)
---^ (show previous table, only available when there is a
previous graph; bottom/top line repeats in new table)
---v (show next table, only available when there is a next
graph; bottom/top line repeats in new table)
---PRINT
---BACK (back to Sensor menu)
--BACK (back to Trend menu)
---EXIT (to Main menu) (Sensor/Loop Menu)
the following will be displayed)
--VIEW (Sensor Trend View Menu)
---DUAL (shows SPO2+BPM)
---SPO2
---PULSE
--ZOOM (cycle through 2h, 1h, 30m, and 15m for current
view)
(display events #1-N, in inverse
(If continuous-Loop data is in the sensor,
N-600x Operator’s Manual 37
Using the Monitor
Menu Structure
(continued...)
--PRINT
--BACK (to Trend menu)
- EXIT (to Main menu) SETUP (Setup Monitor Menu)
- VIEW (Setup View Menu)
--PLETH
--BLIP
- TREND
---VIEW (RT
Trend View Menu)
----DUAL
----SPO2
----PULSE
----BACK
---ZOOM (RT
Trend View Menu)
----TIME
----SCALE
----AUTO
----BACK
--BACK (back to Setup menu)
--EXIT (to Main menu)
---SENSOR (Setup Sensor Menu)
--DATA (On-screen options for SENSOR-R (Write-once
Sensor) sensor are: “SPO2, SPO2+BPM, DEFAULT.” On-screen options for SENSOR-RW (rewritable sensor) are: “SPO2, SPO2+BPM, DEFAULT.” SELECT toggles SENSOR-R or SENSOR-RW sensor type; up/down keys scroll through options in order.) The SENSOR-R feature supports all of the current O
XIMAX sensors.
---SELECT
---BACK (back to Setup Sensor menu)
---EXIT (to Main menu)
--MSG (Sensor Set Message Menu)
---BACK (back to Setup Sensor menu)
---EXIT (to Main menu)
- NEXT (Clock/Language Menu)
--CLOCK (Clock Menu)
---SET (Clock Set Menu)
----SELECT (press select to toggle through hours,
minutes, seconds, month, day, year; use up/down buttons to set each selection)
38 N-600x Operator’s Manual
Using the Monitor
Menu Structure
(continued...)
----BACK (back to Clock/Language menu)
----EXIT (to Main menu)
---BACK (back to Clock/Language menu)
---EXIT (to Main menu)
--LANG (Language Setup Menu)
toggle though languages)
---BACK (back to Clock/Language menu)
--NEXT (Communication/Nurse Call Menu)
---COMM (Communication Port Configuration Menu)
----SELECT
----BACK (back to Communication/Language menu)
----EXIT (to Main menu)
---NCALL Nurse Call Menu)
----NORM +
----NORM -
----BACK (back to Communication/Nurse Call menu)
----EXIT (to Main menu)
---NEXT (Analog/Mode Menu)
----ANALOG (Analog Voltag e Select Menu)
-----0 VOLT
-----1 VOLT
-----STEP
-----BACK (back to Analog/Mode menu)
----MODE (Mode Menu)
-----BACK (back to Analog/Mode menu)
-----EXIT (to Main menu)
----BACK (back to Communication/Nurse Call menu)
----EXIT (to Main menu)
---BACK (back to Clock/Language menu)
--BACK (back to Setup menu)
- EXIT (to Main menu)
LIGHT (Light Menu)
- OFF (Turns display backlight off)
- EXIT (to Main menu)
(use up/down buttons to
N-600x Operator’s Manual 39
Using the Monitor

Parameter Ranges

The parameters of the N-600x monitor are preset to factory default settings. See Factory Default Settings on page 167. The factory default parameters may be changed to institutional default parameters by following the procedures in the N-600x Service Manual.
Table 6 lists the parameters, ranges available, and the factory default setting. The parameters may be set on an individual basis, by the clinician, and these settings remain in effect until the monitor is turned off.
Table 6: Parameter Ranges
Parameter
%SpO
2
Upper Alarm Limit
2
%SpO Lower Alarm Limit
Pulse Rate Upper Alarm Limit
Pulse Rate Lower Alarm Limit
Alarm Silence Duration
Alarm Volu m e
Alarms
Ranges/ Selections
Lower Alarm Limit plus 1 to 100%
20% to Upper Alarm Limit minus 1
Lower Alarm Limit plus 1 to 250 bpm
30 bpm to Upper Alarm Limit minus 1
Alarms 30, 60, 90, 120 seconds
1 to 10 7 7
Allow Off Yes/No
-
Factory Adult Defaults
100% 95%
85% 85%
170 bpm 190 bpm
40 bpm 90 bpm
60 60
No
Factory Neonate Defaults
No
Off Reminder ­Yes/No
40 N-600x Operator’s Manual
Yes
Yes
Table 6: Parameter Ranges
Using the Monitor
Parameter
Backlight Brightness
Data Port Baud Rate
Data Port Mode
Default Display Format
Display Contrast
Language English, Danish,
Ranges/ Selections
0 to 10
2400, 9600, 19200
ASCII, GRAPH, OXINET, CLINICAL, PHILIPS, SPACELBS, MARQ (GE Marquette), DATEX (Datex­Ohmeda)
Pleth, Blip Real-Time T rend
Low to high Medium Medium
Dutch, Finnish, French, German, Norwegian, Portuguese, Spanish, Italian, Swedish
Factory Adult Defaults
9600 9600
ASCII ASCII
Pleth Pleth
English English
Factory Neonate Defaults
Limits Adult, Neonate A dult Neonate On AC
Power On Battery
Power Pulse Beep
Vol ume Pulse Rate
Lower Alarm Limit
N-600x Operator’s Manual 41
0 to 10 4 4
30 bpm to Upper Alarm Limit minus 1
10 10
88
40 bpm 90 bpm
Using the Monitor
Table 6: Parameter Ranges
Parameter
Pulse Rate Upper Alarm Limit
Real-Time Trend
Ranges/ Selections
Lower Alarm Limit plus 1 to 250 bpm
Saturation, Dual, Pulse Rate
Display Real-Time
Trend Scale
48, 36, 24, 12, 8, 4, 2, 1 hours, 30 minutes
Response
Normal or Fast Normal Normal
Mode
-232 Level
RS Nurse Call
Normally High, Normally Low
Polarity SatSeconds Off, 10, 25, 50,
100
Allow
Yes/No Yes Yes
SatSeconds
Factory Adult Defaults
Factory Neonate Defaults
170 bpm 190 bpm
Saturation Saturation
30 minutes 30 minutes
Normally Low Normally Low
Off Off
Sensor Event Date Format (SENSOR-R and
SpO
2,
SpO
2+Pulse Rate,
Default (default is factory default)
Default Default
SENSOR-R W
Sensor
Yes/No Yes Yes Messages Enabled
Trend Display
Dual, %SpO
2,
Pulse, Histogram,
%SpO2 %SpO2
Amplitude Trend Scale 48, 36, 24, 12, 8,
2 hours 2 hours 4, 2, 1 hours, 30, 15 minutes, 40, 20 seconds
42 N-600x Operator’s Manual

Turning On the Monitor

Before using the monitor in a clinical setting, verify the monitor is safe and working properly. Proper working condition is verified each time the N-600x is turned on as described in the following procedure.
Caution: If any indicator or display element does not light when the pulse oximeter is turned on, do not use the pulse oximeter. Instead, contact qualified service personnel, your local Nellcor representative, or Nellcor’s Technical Services Department.
Note: Physiological conditions, medical procedures, or external
agents that may interfere with the monitor’s ability to detect and display measurements, include dysfunctional hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring agents such as nail polish, dye, or pigmented cream.
Note: The monitor automatically launches the Power-On Self-Test (POST), which tests the monitor circuitry and functions.
Using the Monitor
Caution: During POST (immediately after power-up), confirm that all indicators light, all display segments turn on, and the pulse oximeter speaker sounds a sequence of three ascending tones. After the POST process complete, verify that a single one­second tone sounds.
1. Turn on the monitor by pressing the ON/STANDBY button.
2. Ensure all of the front panel indicators illuminate.
N-600x Operator’s Manual 43
Using the Monitor
3. Once the display test portion of POST completes, the software version displays for approximately five seconds and a sequence of three ascending tones sound.
Note:The software version shown above is only a sample. Check your monitor for the current software version installed.
The software version is often needed when calling Nellcor’s Technical Services Department or your local Nellcor representative for technical assistance. Record the software version number and have it available prior to contacting technical assistance.
4. If the monitor detects an internal problem during the POST process, an error tone sounds and the monitor displays an error code (EEE) and the corresponding number (see Troubleshooting on page 139).
5. Upon successful completion of the POST, the monitor sounds a one-second tone indicating the monitor has passed the test.
44 N-600x Operator’s Manual
WARNING: If you do not hear the POST pass tone, do not use the pulse oximeter.
WARNING: Ensure the speaker is clear of any obstructions. Failure to do so could result in an inaudible alarm tone.
Note:In addition to serving as the POST pass verification, the POST
pass tone also functions as an audible confirmation that the speaker is performing properly. If the speaker does not function, the alarm warning sounds cannot be heard.

OXIMAX Sensor Attached

WARNING: Do not use any other cables to extend the length of the DOC-10 Pulse Oximetry Cable. Increasing the length of the DOC-10 cable will degrade signal quality and may lead to inaccurate measurements.
Using the Monitor
When an O
XIMAX sensor is connected to the monitor, a “SENSOR
TYPE: . . .” message is displayed 4 to 6 seconds at the bottom of the monitor display. The message identifies the type (model) of O
XIMAX
sensor connected to the monitor. The type is used to determine the action messages in the O display is the first message displayed when an O
XIMAX sensor message(s) function. This
XIMAX sensor is
connected to the monitor.
N-600x Operator’s Manual 45
Using the Monitor
For a sensor containing data, the message identifies the sensor data type. For a blank sensor, the message identifies the monitor’s current data type setting used to write data to the sensor. The data type settings are SpO
2 and SPO2+BPM.
Note:The type of data recorded is only displayed when data is present in the O
XIMAX sensor.
The monitor displays zeros in the %SpO
2 and Pulse Rate displays
while the N-600x is searching for a valid pulse. For optimal performance, allow the monitor to search and lock onto a pulse for approximately 5 to 10 seconds.
When a valid pulse is detected, the monitor enters the Monitoring Mode and displays patient parameters.
46 N-600x Operator’s Manual
Using the Monitor
Notice the movement of the blip bar or plethysmographic waveform or beating heart indicating the monitor is displaying real-time data. Listen for the pulse beep tone. If the pulse beep tone does not sound with each pulse, it is an indication the pulse beep volume is set to zero, the speaker is malfunctioning, or the signal is corrupt.
N-600x Operator’s Manual 47
Using the Monitor
When an OXIMAX sensor is attached to the monitor and applied to a patient, the monitor may lose a pulse signal. If a pulse signal is lost, an alarm sounds and a poor signal condition message displays on the
monitor screen. At this point, the monitor displays [
--- / ---]
(3 dashes / 3 dashes) and remains in the Pulse Search Mode for five seconds before displaying the poor signal condition screen. The poor signal condition screen is part of the N-600x's Sensor Messages feature. For more information about O
XIMAX Sensor Messages, refer
OXIMAX Sensor Messages on page 69.
48 N-600x Operator’s Manual

No OXIMAX Sensor Attached

Upon successful completion of the POST process, the monitor sounds a one-second tone indicating that the monitor has passed the POST.
or
Using the Monitor
The monitor displays dashes [ - - - ] and the Pulse Search indicator is not lit, indicating the monitor failed to detect an O
N-600x Operator’s Manual 49
XIMAX sensor.
Using the Monitor

Turning the Backlight On or Off

You can turn off the backlight by pressing the LIGHT softkey and then pressing OFF.
Note: Any of the following conditions turn on the backlight:
pressing any of the softkeys
pressing and holding the HELP/CONTRAST button
pressing the ALARM SILENCE button
•any alarm

Adjusting the Screen Contrast

1. With the monitor in the normal monitoring mode, press and hold the HELP/CONTRAST button while pressing the ADJUST UP or ADJUST DOWN button until the desired contrast is obtained.
2. Press the HELP/CONTRAST button to return to the normal monitoring mode.

Adjusting the Backlight Brightness

1. With the monitor in the normal monitoring mode, press the LIGHT softkey.
2. Press the ADJUST UP or ADJUST DOWN button until the desired backlight brightness is obtained.
50 N-600x Operator’s Manual

Selecting the Pleth View

The pleth view displays the pleth waveform, %SpO2, and pulse rate data. Refer to Principles of Operation on page 171, for a description of the pleth waveform.
1. With the monitor in the normal monitoring mode, press the SETUP softkey.
2. Press the VIEW softkey.
3. Press the PLETH softkey. The pleth view displays.
Using the Monitor

Selecting the Blip View

The blip view displays the SpO2, pulse rate, blip bar, and limits in a larger format for easier viewing.
1. With the monitor in the normal monitoring mode, press the SETUP softkey.
2. Press the VIEW softkey.
N-600x Operator’s Manual 51
Using the Monitor
3. Press the BLIP softkey. The blip view displays.

Selecting the Real-Time Trend View

The real-time trend view displays the %SpO2 and/or pulse rate trend data. The real-time trend submenu enables you to:
select the trend data display,
set the trend time scale display, and
set the trend amplitude scale display.
1. With the monitor in the normal monitoring mode, press the SETUP softkey.
2. Press the VIEW softkey.
3. Press the TREND softkey. The real-time trend view displays.
52 N-600x Operator’s Manual

Selecting the Trend Data Display

1. Press the SETUP softkey.
2. Press the VIEW softkey.
3. Press the TREND softkey.
4. Press the VIEW softkey.
5. Press any of the trend softkeys (DUAL, SPO2, or PULSE).
Using the Monitor

Setting the Trend Time Scale Display

1. Press the SETUP softkey.
2. Press the VIEW softkey.
3. Press the TREND softkey.
4. Press the ZOOM softkey.
5. Press the TIME softkey to cycle the displayed trend time scale through 48 hours, 36 hours, 24 hours, 12 hours, 8 hours, 4 hours, 2 hours, 1 hour, and 30 minutes.
N-600x Operator’s Manual 53
Using the Monitor

Setting the Trend Amplitude Scale Display

1. Press the SETUP softkey.
2. Press the VIEW softkey.
3. Press the TREND softkey.
4. Press the ZOOM softkey.
5. Press the SCALE softkey to cycle the trend amplitude scale display through ±5 points, ±10 points, ±15 points, ±20 points, ±25 points, ±30 points, ±35 points, ±40 points and ±50 points above and below the newest - rightmost - trend data point.
Note: You can set the trend amplitude scale to AUTO by pressing the AUTO softkey. The maximum trend data point is rounded up to the nearest multiple of 10, shown at the top of the graph display. The minimum trend data point is rounded down to the next multiple of 10. 10 is then subtracted from the rounded down number. This value is located at the bottom of the trend graph.

Setting the Pulse Beep Volume

With the monitor in the normal monitoring mode, press and hold the ADJUST UP or ADJUST DOWN button to increase or decrease the pulse beep volume.
54 N-600x Operator’s Manual

Setting the Alarm Volume

The Alarm Volume display enables you to adjust the volume of alarm tones.
1. With the monitor in the normal monitoring mode, press the ALARM SILENCE button until the alarm volume level displays and sounds on the monitor.
Using the Monitor
2. While continuing to press the ALARM SILENCE button, press and hold the ADJUST UP or ADJUST DOWN button to increase or decrease the volume.
N-600x Operator’s Manual 55
Using the Monitor

Setting the Date and Time

WARNING: The sensor extrapolates from the date and time provided by the N-600x when recording the sensor event record to the sensor . The accuracy of the date/time is the r esponsibility of the N-600x. It is recommended that the N-600x user set the time/ date to the correct value before a sensor event record-enabled sensor is connected, and that this date/time not be changed while the sensor remains connected. Since a sensor with sensor event record data can be transported from one monitor to another, having discrepancies in the date/time between monitors and the sensor event record data will affect the order the sensor event record data appears. To eliminate this possible problem, all monitors within an institution should be set to the same time.
1. With the monitor in the normal monitoring mode, press the
SETUP softkey.
2. Press the NEXT softkey.
3. Press the CLOCK softkey.
4. Press the SET softkey.
56 N-600x Operator’s Manual
Using the Monitor
5. Press the SELECT softkey to select the TIME and DA TE fields as shown in the graphic below.
TIME HOURS : MINUTES : SECONDS (16 : 46 : 05)
DATE DAY - MONTH - YEAR (02 - JAN - 06)
6. Use the ADJUST UP or ADJUST DOWN buttons to change the selected value.
7. Press the EXIT softkey.
N-600x Operator’s Manual 57
Using the Monitor

Setting the Alarm Silence Duration

The Alarm Silence Duration display enables you to adjust the alarm silence duration.
1. With the monitor in the normal monitoring mode, press the ALARM SILENCE button until the alarm silence duration setting displays. The alarm silence durations available are 30, 60, 90, and 120 seconds.
2. Press and hold the ALARM SILENCE button and the ADJUST UP button to increase the alarm silence duration setting.
3. Press and hold the ALARM SILENCE button and the ADJUST DOWN button to decrease the alarm silence duration setting.
Note: Releasing the ADJUST UP or ADJUST DOWN button sets the alarm silence duration.
58 N-600x Operator’s Manual

Disabling Audible Alarms

Setting the alarm silence duration to OFF disables all audible alarms.
Note: The ability to set the alarm silence duration to OFF can be enabled or disabled by qualified service personnel as described in the N-600x Service Manual.
The N-600x Service Manual is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_SuppProductManuals.html
WARNING: Do not disable the audible alarm function or decrease the audible al arm volume if the patient’s safety could be compromised.
1. With the monitor in the normal monitoring mode, press the
ALARM SILENCE button until the alarm silence duration setting displays.
Using the Monitor
2. While pressing the ALARM SILENCE button, press and hold the ADJUST UP button until OFF displays. Release the buttons.
N-600x Operator’s Manual 59
Using the Monitor

Selecting the Standby Mode

The standby mode enables the monitor to retain the alarm limit settings that are in effect while monitoring a patient. The monitor must be connected to an AC power source to enter the standby mode.
Normally , the standby mode setting is used when a patient has to temporarily leave the monitor.
1. Verify the N-600x is monitoring a patient and the alarm limits are configured to the patient being monitored.
2. Disconnect the sensor from the monitor.
3. Press the ALARM SILENCE button to silence the audible alarms.
4. Press the ALARM SILENCE button to disable the alarm messages.
The monitor is now in the standby mode. Reconnect the sensor to the monitor and the patient to return to normal monitoring.
60 N-600x Operator’s Manual

Adult-Pediatric or Neonatal Settings

The clinician can set the monitor’s operating mode to adult-pediatric or neonatal by using the LIMITS softkey. This setting remains active until the monitor is turned OFF. The factory default power-on setting is for adult-pediatric patients. This default setting can be changed to neonatal by qualified service personnel using the procedures indicated in the N-600x Service Manual.
Refer to Table 15, for neonate factory default limit settings. Refer to Table 16, for adult factory default limit settings.
WARNING: Each time the pulse oximeter is used, check alarm limits to ensure that they are appropriate for the patient being monitored.
Using the Monitor
N-600x Operator’s Manual 61
Using the Monitor

Setting Patient Adult-Pediatric or Neonatal Modes

1. With the monitor in the normal monitoring mode, press the LIMITS softkey. The monitor displays the ADULT LIMITS or NEONATE LIMITS screen, depending on the patient setting used.
NEONATE LIMITS
%SPO2
UPPER
LOWER
SAT-S
SELECT
2. Press the NEO or ADULT softkey to select ADULT LIMITS or NEONATE LIMITS depending on the patient being monitored.
100
NEO ADULT
95 85
BPM
190
90
%SP02
100
BPM
EXIT
---
---
NEO
62 N-600x Operator’s Manual

Alarm Limit Changed Indicator

Alarm limits that have been changed from the institution or factory default settings are identified by a decimal point (.) after the displayed reading (%SpO
2 or BPM). The changed parameter is also identified
by a decimal point on the alarm limits screen.

Setting Alarm Limits

The Alarm Limit display enables you to adjust the upper and lower saturation and pulse rate limits. It also enables you to adjust the SatSeconds limit.
Using the Monitor
The Alarm Limit display is accessed by pressing the LIMITS softkey on the main menu.
The Alarm Limit display includes the alarm limit table and current measured %SpO
2 and pulse rate. The title of the alarm limit table
indicates whether the instrument is in Adult or Neonate monitoring mode. If SatSeconds are enabled, the Alarm Limit display also includes the SatSeconds indicator. Decimal points after the displayed %SpO
2 or pulse rate indicate that the respective limits have been
changed from the power-on defaults.
N-600x Operator’s Manual 63
Using the Monitor
1. Press the LIMITS softkey. The current alarm limits display.
OR
UPPER
LOWER
SAT-S
SELECT
NEONATE LIMITS
%SPO2
100
BPM
190
95 85
NEO ADULT
90
%SP02
100
BPM
EXIT
---
---
NEO
2. Press the ADUL T or NEO softkey to select the Adult-Pediatric or Neonatal alarm limits screen.
3. Press the SELECT softkey to select the parameter to be adjusted.
4. Use the ADJUST UP or ADJUST DOWN buttons to increase or decrease the selected limit parameter.
5. Repeat steps 2, 3, and 4 as necessary to complete the alarm limits setup.
64 N-600x Operator’s Manual
6. Wait for the display to time-out to accept the changes or press the EXIT softkey to close the display and return to the normal monitoring mode.
Note: Limit changes are in effect as long as the monitor remains turned on. When the monitor is turned off, the institutional default limits are restored. When the monitor is turned on, the institutional default limits are effected. Factory or institutional defaults are selected by qualified service personnel by following the procedures outlined in the N-600x Service Manual.

Setting SatSeconds Alarm Limit

Refer to Using SatSeconds on page 163, for more information on the SatSeconds feature.
Note: The ability to adjust the SatSeconds Alarm limit can be enabled or disabled by a qualified service personnel as described in the N-600x Service Manual.
Using the Monitor
1. With the monitor in the normal monitoring mode, press the LIMITS softkey. The current alarm limits display.
2. Press the SELECT softkey twice to select %SpO
2 SAT-S.
3. Use the ADJUST UP or ADJUST DOWN buttons to select the limit. The choices are 10, 25, 50, 100 seconds or OFF.
4. Press the EXIT softkey to save your selection.
N-600x Operator’s Manual 65
Using the Monitor

Setting Monitor Response Mode

The purpose of the response mode is to set the response time of the N-600x algorithm calculation of the SpO not affect the N-600x algorithm’s calculation of pulse rate). The trending interval (2- or 4-seconds) is updated automatically by the monitor to roughly correspond with the SpO time.
The response mode automatically programs the N-600x algorithm to record and display monitor trend information at 2-second intervals (Fast Mode) up to 24 hours or 4-second intervals (Normal Mode) up to 48 hours.
2 (the response mode does
2 calculation response
The response mode display screen includes the current SpO mode setting and the current measured %SpO
2 and pulse rate. When
2 response
in the fast mode, the screen displays the fast mode symbol.
1. With the monitor in the normal monitoring mode, press the SETUP softkey.
2. Press the NEXT softkey three times.
66 N-600x Operator’s Manual
Using the Monitor
3. Press the MODE softkey.
Note: When the monitor is in the fast response mode, the monitor may produce more SpO
2 and pulse rate alarms than you are
accustomed to experiencing.
4. Use the ADJUST UP or ADJUST DOWN buttons to select the desired response mode.
5. Press the EXIT softkey.
N-600x Operator’s Manual 67
Using the Monitor

Selecting the Display Language

The N-600x can be programmed to display the information in various languages. The languages available are:
ENGLISH DANSK (Danish) DEUTSCH (German) ESPAÑOL (Spanish) FRANCAIS (French) ITALIANO (Italian) NEDERLANDS (Dutch) NORSK (Norwegian) PORTUG (Portuguese) SUOMI (Finnish) SVERIGE (Swedish)
1. With the monitor in the normal monitoring mode, press the SETUP softkey.
2. Press the NEXT softkey.
3. Press the LANG softkey.
68 N-600x Operator’s Manual
4. Use the ADJUST UP or ADJUST DOWN buttons to select the desired language.
5. Press the EXIT softkey.
Note: The selected language displays until the monitor is turned OFF. The selected language can be set as a default by qualified service personnel by following the procedures outlined in the N-600x Service Manual.

OXIMAX Sensor Messages

OXIMAX sensor messages consist of sensor adjust condition messages and sensor adjust messages which, when enabled, are displayed when the monitor is not able to display saturation. When O messages are displayed, it is an indication that the O functioning properly, but the site to which the O or the application method is not optimal for calculating %SpO Condition messages are followed by action messages. Up to three condition messages may be displayed on the “POOR SIGNAL CONDITION” display in priority order, highest on top. The condition display may be dismissed by using the EXIT softkey. Once closed, the O
XIMAX sensor message screen will not return until a new
condition occurs.
Using the Monitor
XIMAX sensor
XIMAX sensor is
XIMAX sensor applies
2.
N-600x Operator’s Manual 69
Using the Monitor
If the HELP softkey is pressed from the Condition message display, the action messages are displayed. Action messages are linked to the sensor type; action messages will be displayed for the type of O
XIMAX
sensor connected to the monitor. Up to five action messages may be displayed. Multiple screens may be required to display all of the messages. When multiple screens are required, navigation between screens can be accomplished by using the NEXT, BACK, and EXIT softkeys.
O
XIMAX sensor messages may be disabled. Refer to OXIMAX Sensor
Messages on page 69 for selecting the O
XIMAX Sensor Messages,
Enable/Disable function.

OXIMAX Sensor Adjust Condition Messages

Condition — SENSOR OFF?
Condition — SMALL PULSES
Condition — WEAK SIGNAL
Condition — INTERFERENCE
Condition — EXCESS INFRARED LIGHT
Condition — INTERFERENCE
Condition — HIGH PULSE AMPLITUDE
70 N-600x Operator’s Manual

OXIMAX Sensor Adjust Messages

Message — ALTERNATE SITE?
Message — COVER SENSOR SITE?
Message — EAR/FOREHEAD SENSOR?
Message — NASAL/EAR SENSOR?
Using the Monitor
Message — O
XIMAX ADHESIVE SENSOR
Message — SECURE CABLE
Message — HEADBAND
Message — WARM SITE
Message — BANDAGE ASSEMBLY
Message — NAIL POLISH
Message — SENSOR TOO TIGHT?
Message — REPOSITION SENSOR
Message — ISOLATE INTERFERENCE SOURCE
Message — CLEAN SENSOR SITE
N-600x Operator’s Manual 71
Using the Monitor
72 N-600x Operator’s Manual

Using Monitor Trend Data

Overview

The trend displays enable you to view trend data. Two types of trend data can be viewed:
Monitor trend data stored in the monitor
Using Monitor Trend Data
Patient event data stored in the O (single-patient-use O
XIMAX sensors only) and can be used
XIMAX sensor
with the sensor event record feature.
Monitor trend data can be viewed anytime patient trend is stored in the monitor. Monitor trend displays are accessed by pressing the TREND softkey on the main menu and selecting the MONITR softkey option. The monitor trend submenu enables you to choose which trend data are displayed:
Saturation and pulse rate (Dual)
Saturation
•Pulse rate
Pulse amplitude
Histogram
The N-600x can graphically display trend data for SpO
2, pulse rate, or
both. Trend data is stored at 2- or 4-second intervals and linked to the response mode. When the TREND softkey is pressed, “READING TRENDS” displays at the bottom of the N-600x screen, indicating that the monitor is formatting the trend data to be displayed.
N-600x Operator’s Manual 73
Using Monitor Trend Data
The monitor stores up to 48 hours of 4-second trend data or 24 hours of 2-second trend data. The amount of trend data displayed on the screen is determined by using the ZOOM softkey. The settings available are 20 and 40 seconds, 15 or 30 minutes, and 1, 2, 4, 8, 12, 24, 36, or 48 hours. All trend data are displayed in a graphical format except the 20 and 40 second trend displays, which are shown in tabular format.
The trend display can be viewed throughout the 48 hours of trend data. Selecting the 1-hour trend display allows you to view one hour of trend information. By using the scrolling feature, any one hour of trend data can be viewed up the 48 hours of trend information. The ADJUST DOWN button scrolls the display to the left and the ADJUST UP button scrolls the display to the right.
When the data displays, the most recent readings are on the right side of the graph. The numbers below %SpO
2 indicate the highest and
lowest parameter values at the cursor position (vertical dotted line on the display). See Table 7.
Trend data is further explained in Specifications on page 177.
Trend data information may be retrieved through the N-600x data port or cleared using options available in a display menu.
Caution: Monitor trend data will be lost if the main battery fails or is removed.
74 N-600x Operator’s Manual

Storing Trend Data

Using Monitor Trend Data
Whenever the N-600x is turned on, it stores the monitor %SpO2 and pulse rate readings in memory every 2- or 4-seconds (regardless of whether the N-600x is monitoring a patient or not). The N-600x can store up to 48 hours of 4-second trend data or 24 hours of 2-second trend data. The 48/24 hours of stored trend data can be downloaded to a printer or a portable computer. Up to 50 alarm limit changes can be stored in the trend data. If more than 50 alarm limit changes occur during the 48/24 hours of trend data collection, the additional alarm limit changes will take space reserved for trend data.
Caution: Changing alarm limit settings uses up trend memory space. Change alarm limits only as needed.
Note: Trend memory always contains the most recent 48 hours of
data, with newly collected data overwriting the oldest data on a rolling basis. The N-600x continues to record data points as long as the monitor is powered on, with “blank” data points collected if no
O
XIMAX sensor is connected to the monitor or patient. “Blank” data
overwrites older patient data if the memory becomes full. Therefore, if you want to save old patient data, it is important that you turn your monitor off when you are not monitoring a patient, and that you download the trend memory before it fills up and overwrites the old data with new data (or “blank” data).
N-600x Operator’s Manual 75
Using Monitor Trend Data

OXIMAX Sensor Type

When an OXIMAX sensor is connected to the monitor, a “SENSOR TYPE:...” message is displayed for 4 to 6 seconds at the bottom of the display. The message identifies the type (model) of O connected to the monitor. Type is used in the determination of action messages in the O the first message displayed when an O the monitor.
XIMAX sensor
XIMAX sensor message(s) function. This display is
XIMAX sensor is connected to

Selecting the Trend Data Display Scale

The trend scale is the amount of trend data displayed on the screen.
1. With the monitor in the normal monitoring mode, press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
4. Press any of the trend softkeys (DUAL, SPO2, or PULSE). To select HIST (histogram) or AMP (amplitude), press the NEXT softkey and then the HIST or AMP softkeys.
76 N-600x Operator’s Manual
Using Monitor Trend Data
5. Press the ZOOM softkey. The Zoom menu displays.
Pressing the TIME softkey cycles the displayed trend time scale through 48 hours, 36 hours, 12 hours, 8 hours, 4 hours, 2 hours, 1 hour, 30 minutes, 15 minutes, 40 seconds and 20 seconds.
Note: The 20-second and 40-second trend displays are in tabular format. The display below begins in the normal response mode (left side of the display) and switches to the fast response mode.
Pressing the SCALE softkey cycles the displayed trend amplitude scale through ±5 points, ±10 points, ±15 points, ±20 points, ±25 points, ±30 points, ±35 points, ±40 points and ±50 points above and below the data point under the cursor. The saturation graphical monitor trend display vertical scale default setting is from 10 to 100 if there is no data under the cursor. The pulse rate graphical monitor trend display vertical scale is from 5 to 250 if there is no data under the cursor.
Pressing the AUTO softkey presets the amplitude of the graphed trend data. The maximum trend data point is rounded up to the nearest multiple of 10, this value is the top of the graph display. The minimum trend data point is rounded down to the next multiple of 10. Then 10 is subtracted from the rounded down number, this value is the bottom of the trend graph.
Pressing the BACK softkey returns the monitor to the Monitor menu.
N-600x Operator’s Manual 77
Using Monitor Trend Data

Reading the Trend Data Display

The following table identifies the components of the trend data display.
Table 7: Reading Trend Display
Item Description
1 Amount of trend data displayed on the screen. Settings available
are 20 and 40 seconds, 15 and 30 minutes, 1, 2, 4, 8, 12, 24, 36,
and 48 hours. 2 Date represented by the cursor (item 5). 3 No trend data recorded during this time. 4 Time represented by the cursor (item 5). 5
Cursor
- can be moved left or right using the ADJUST UP
(right) or ADJUST DOWN (left) buttons. 6 Highest and lowest reading at the cursor position. 7 Trend data that is being displayed (%SPO2, BPM, or PAU
[pulse amplitude units]).
78 N-600x Operator’s Manual

Dual Trend Data Display

The dual trend data display shows both oxygen saturation (%SpO2) levels and pulse rate (bpm) trend data.
1. With the monitor in the normal monitoring mode, press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
Using Monitor Trend Data
4. Press the DUAL softkey . The dual trend (%SpO displays.
2 and Pulse Rate)
N-600x Operator’s Manual 79
Using Monitor Trend Data

SpO2 Trend Display

1. With the monitor in the normal monitoring mode, press the
2. Press the MONITR softkey.
3. Press the VIEW softkey.
TREND softkey.
4. Press the SPO

Pulse Rate Trend Display

1. With the monitor in the normal monitoring mode, press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
2 softkey. SpO2 trend data displays.
80 N-600x Operator’s Manual
4. Press the PULSE softkey. The pulse rate trend data displays.

Histogram Trend Data Display

The histogram displays trend data for the percent of oxygen blood saturation (SpO the trend data stored over the period of time indicated on the display. Refer to OXIMAX Sensor Type on page 76, to set up the desired trend data scale. Pulse amplitude cannot be displayed on the histogram display.
2) and pulse rate (bpm). The data displayed represents
Using Monitor Trend Data
1. With the monitor in the normal monitoring mode, press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
4. Press the NEXT softkey.
N-600x Operator’s Manual 81
Using Monitor Trend Data
5. Press the HIST softkey. The histogram trend data displays.

Pulse Amplitude Trend Data Display

The pulse amplitude trend data display shows the amplitude of the patient’s pulse rate over the period of time indicated on the display. Refer to OXIMAX Sensor Type on page 76 to setup the desired trend data scale.
1. With the monitor in the normal monitoring mode, press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
4. Press the NEXT softkey.
82 N-600x Operator’s Manual
5. Press the AMP softkey. The pulse amplitude units (PAU) trend data displays.
The PAU reading (12 : 20) indicates the pulse amplitude units (upper and lower) at the cursor position (dashed line). The cursor moves right or left using the ADJUST UP (right) and ADJUST DOWN (left) buttons.

Clearing Trend Information

Using Monitor Trend Data
1. With the monitor in the normal monitoring mode, press the TREND softkey.
2. Press the MONITR softkey.
3. Press the NEXT softkey.
4. Press the DELETE softkey.
Note: Press NO and then EXIT to close this function without deleting the trend data.
5. Press the YES softkey.
All the trend data clears and the monitor sounds three beeps.
N-600x Operator’s Manual 83
Using Monitor Trend Data
84 N-600x Operator’s Manual

Using OXIMAX Sensor Event Records

Using OXIM AX Sensor Event Records

Overview

WARNING: The sensor extrapolates from the date and time provided by the N-600x when recording the sensor event record to the sensor . The accuracy of th e date/time is the responsibility of the N-600x. It is recommended that the N-600x user set the time/ date to the correct value before a sensor event record-enabled sensor is connected, and that this date/time not be changed while the sensor remains connected. Since a sensor with sensor event record data can be transported from one monitor to another, having discrepancies in the date/time between monitors and the sensor event record data will affect the order the sensor event record data appears. To eliminate this possible problem, all monitors within an institution should be set to the same time.
The adhesive O
XIMAX sensors are capable of storing patient alarm
event data. A sensor event record allows alarm event history to travel with the patient on the sensor’s memory chip for quick assessment at every point of care where O
XIMAX monitors are used.
Patient (alarm event) data is stored on the memory chip of adhesive
O
XIMAX sensors (single-patient-use OXIMAX sensors only). The alarm
event data is stored (recorded) with the limit/threshold settings that were active at the time of the event on the recording monitor. These events can be viewed on the next O
XIMAX sensor monitor when the
patient moves to a new point of care.
An event occurs when the %SpO
2 value exceeds either the upper or
lower alarm limit for at least 15 seconds. Alarm events are grouped and recorded to the memory chip every 5 minutes. The maximum number of events that can be stored in an O
XIMAX sensor is typically
100.
N-600x Operator’s Manual 85
Using OXIMAX Sensor Event Records
Event records can only be viewed after an OXIMAX sensor containing patient alarm data (event records) has been connected to an O monitor with SENSOR enabled. Event records are designed to view patient events from prior areas of care or transport (history) while monitor trend should be used to view data or events from a patient currently being monitored.
The monitor’s SENSOR EVENT RECORD indicator lights when an
O
XIMAX sensor containing event data is connected to the OXIMAX
monitor.
Patient alarm event data is accessed by pressing the TREND softkey on the main menu and selecting the SENSOR softkey option. Sensor event record can be viewed in graphical form (GRAPH) or in a summary table (TABLE).
XIMAX
Note: Once the O O
XIMAX sensor and event data is stored in the OXIMAX sensor, the XIMAX sensor event record type cannot be reset. The monitor’s type
O
XIMAX sensor event record type is setup in the
set up can be changed at any time.
Recording and viewing of O available on O
XIMAX sensors may function on older technology monitors but
The O
XIMAX sensor event record feature is not available.
the O
XIMAX compatible monitors with SENSOR enabled.
XIMAX sensor event record is only
Refer to the N-600x Service Manual for specific instruction on how to disable the storage of the sensor event recorded in an O
XIMAX sensor.
86 N-600x Operator’s Manual

Setting up OXIMAX Sensor Messages

The OXIMAX sensor message setup display allows you to enable or disable the O the “SENSOR NOT POSTING” nor the “RECOMMENDED ACTION” messages display.
1. With the monitor in the normal monitoring mode, press the SETUP softkey.
2. Press the SENSOR softkey.
3. Press the MSG softkey.
XIMAX sensor message feature. When disabled, neither
Using OXIMAX Sensor Event Records
4. Press the ADJUST UP or ADJUST DOWN button to toggle the ENABLE message.
5. Press the EXIT softkey.
N-600x Operator’s Manual 87
Using OXIMAX Sensor Event Records

Setting In-Sensor Data Type

The In-Sensor Data Type display enables you to adjust the type of patient alarm event trend data to be recorded in an O
XIMAX sensors can be set to record either SpO2 or SpO2+BPM.
O
XIMAX sensor.
Note: The O
XIMAX sensor data type can only be set when an OXIMAX
sensor is not connected to the monitor.
1. With the monitor turned on and no cables attached to the SpO
O
XIMAX sensor port, press the SETUP softkey.
2
2. Press the SENSOR softkey.
3. Press the DATA softkey.
Note: O
XIMAX sensor data type settings are displayed on the monitor
as shown in the figure below (in-sensor data type). If no sensor is connected, both sensor types and the full set of options for each are displayed. If a sensor is connected, only the sensor data type for that sensor displays.
Note: The SENSOR-R feature supports all the current O
XIMAX
sensors. The SENSOR-RW feature is only applicable to OXIMAX sensors with a read/write chip installed.
88 N-600x Operator’s Manual
Using OXIMAX Sensor Event Records
4. Use the SELECT softkey to toggle between SENSOR-R and SENSOR-RW.
5. Use the ADJUST UP or ADJUST DOWN button to select the
XIMAX sensor data type. SENSOR-R and SENSOR-RW
O
selections are:
•SpO
•SpO2+BPM
•DEFAULT
6. Press the EXIT softkey to set the O

OXIMAX Sensor Data Type

When an OXIMAX sensor with no previously recorded patient data is connected to the O displayed briefly at the bottom of the display after the O type message. The message identifies the monitor's current data type setting that will be used to write data to the O type setting options are EVENT/SPO
2
XIMAX sensor type.
XIMAX monitor, a “DATA TYPE: . . .” message is
XIMAX sensor
XIMAX sensor. The data
2 and EVENT/SPO2+BPM.
You can change the setting by referring to Setting up OXIMAX Sensor Messages on page 87. The O set prior to connecting the O
N-600x Operator’s Manual 89
XIMAX sensor event record type must be
XIMAX sensor to the monitor.
Using OXIMAX Sensor Event Records

OXIMAX Sensor Event Record Data Available

When an OXIMAX sensor containing patient alarm data (single-patient-use O the Sensor Event Record indicator on the monitor front panel blinks at a medium priority flash rate to indicate that the O attached to the monitor contains patient event data. The LED blinks for approximately 60 seconds or until the O disconnected or until the sensor trend data is displayed by pressing TREND, then SENSOR.
A corresponding “DATA IN SENSOR” message is also displayed at the bottom of the display. After 4 to 6 seconds, if all the data has been read from the O menu.
XIMAX sensors only) is connected to the monitor,
XIMAX sensor, the message is replaced with the main
XIMAX sensor
XIMAX sensor is
If data is still being read from the O
XIMAX sensor after 4 to 6 seconds,
the DATA IN SENSOR message is replaced with a READING TRENDS message which includes an ABORT option.
Selecting the ABORT softkey stops the recording, accessing or viewing of additional data into the O
90 N-600x Operator’s Manual
XIMAX sensor .
Using OXIMAX Sensor Event Records
Sensor event records can be viewed by accessing the TREND/ SENSOR menu.
The SENSOR EVENT RECORD LED comes on steady when
O
XIMAX sensor memory is full and stays on until the OXIMAX sensor is
disconnected.

OXIMAX Sensor Event Record Not Available

If you select the TREND/SENSOR option when a connected OXIMAX sensor (single-patient-use O data, because no events were recorded to the O chip in the prior monitoring situation, a TREND/SENSOR option is unavailable.
A sample event display without data is shown below. The message clears when the graph or summary closes.
XIMAX sensors only) does not contain
XIMAX sensor memory
N-600x Operator’s Manual 91
Using OXIMAX Sensor Event Records

OXIMAX Sensor Event Record Graphical Data

Graphical representations of patient event history is only available on single-patient-use O minimum or maximum %SpO throughout the duration of an event (%SpO alarm threshold for at least 15 seconds) and continuing every 30 seconds until the actual %SpO threshold.
The duration of an event is determined by the number of data points in the event. Each data point is stored at 30-second intervals.
Events end due to one of the following reasons:
XIMAX sensors. Graphed data points are the
2 value for each 30-second interval
2 value equals or exceeds the alarm
2 continuously below
•%SpO
2 returns to or above the alarm limit
Loss of pulse
O
XIMAX sensor is disconnected
O
XIMAX sensor is removed from the patient
The graph title shows the data type (EVENT GRAPH) in the upper left corner. The number of the displayed event and the total number of events recorded in the O
XIMAX sensor are shown to the right of the
title (example, 2/2). The date and time of the displayed event are shown in the upper center and upper right corner.
92 N-600x Operator’s Manual
Loading...