SERVICE
MANUAL
PULSE
Caution:
NELLCOR®
OXIMETER
Federal
to
sale
by
MODEL
law
or
on
the
ULTRA
AND
N-6000
(U.S.)
order
CAP™
CAPNOGRAPH,
restricts
of a physician.
this
device
©
1993
Nellcor
Incorporated
Nellcor
25495
Hayward,
Incorporated
Whitesell
California
510
887-5858
1-800-NELLCOR
.
Street
94545
U.S.A.
028065A-0193
Corporate
Headquarters
European
Office
Asia/Pacific
Headquarters
Nellcor
25495
Hayward,
Whitesell
USA
Tel
510-887-5858
Incorporated
Street
California
94545
Nelleor
Hambakenwetering
5231
The
Tel.
European
Northwest
Nellcor
Hambakenwetering
5231
The
Tel.
Middle
Nelleor
Black-&-Decker-Strasse
W-6270
Germany
|
Tel.
Southern
Nellcor
3,
78353
France
Tel.
BV
DD
’s-Hertogenbosch
Netherlands
+31.73.426565
Regional
Europe
Europe
DD
‘s-Hertogenbosch
Netherlands
+31.73.426565
and
GmbH
Idstein
+49
6126.5930
Sarl
rue
du
Petit
Jouy-en-Josas
+33.1.39.46.96.58
BV
Central
Europe
Robinson
1
Offices
1
Europe
.
28
Cedex
Nellcor
Suite
Admiralty
18
Hong Kong
Tel.
Limited
1204C
Harcourt
+852.529.0363
Centre,
Road
Tower
1
European
Nellcor
University
Science
Coventry
United
Tel.
Nelleor
Interleuvenlann
B-3001
Belgium
Tel.
Local
(UK)
Limited
of
Park
CV4
7EZ
Kingdom
+44.203.690220
(Belgium)
Leuven
+32.16.400467
Offices
Warwick
NV/SA
62/8
.
ii
TABLE
OF
CONTENTS
11
12
13
2.1
2.2
2.3
2.4
2.5
3.1
3.2
3.3
3.4
Symbols
We
Introduction...
Introduction.........................
Warnings,
Description..................,..,............,...
1.8.1
13.2
13.3
13.4
1.3.5
1.3.6
1.3.7
18.8
1.3.9
1.3.10
1.3.11
13.12
Principles
Overview.........................
PulseOximetrySubsystem.......................
2.2.1
2.2.2
2.2.8
2.2.4 - Measured
Capnography
N20/02
24.1
2.4.2
2.4.3.
2.4.4
2.4.5
2.4.6
2.4.7
.
Factory
Circuit
Introduction.................
CO2
3.2.1
3.2.2
3.2.3
8.2.4
3.2.5
3.2.6
3.2.7
8.2.8
3.2.9
3.2.10
Oximetry
©
8.8.1
Detailed
3.4.1
AAA
Cautions,
Visible
Audible
Visible
StatusMessages..............................
Automatic
On-Screen
N-6000
Trend
Automatic
Battery
Noninvasive
C-LOCK™
C-LOCK
Automatic
Functional
Standard
Pressure
N20
O2
Water
BTPS/ATPS
Summary:
Calibrated
Analysis
Module
Sensor
CO2
Default
Memory......:...,...............,,,,,......
of
Subsystem
Compensation
Collision
Collision
Circuit
Module
bDigital.....................
Motor
Signal
Heater
Source....
Barometer...
StatusLEDs....................
Status
OximetryModule........................
Oximetry
Control........
Module
Oximetry
and Notes
and
Audible
Indicator.......................
and
Operation
Broadening
Vapor
Operation.........
22220
Audible
Self-Test
Menus..................,...,,,.......
and
Calibration..…..............,........,.,.......ss
..................
Oximetry
ECG
Synchronization
Operation.
ECG
Calibration..................
versus
versus
Gas
Reported
Amplifier..
LED
Module
Calculated
............................
............
Conditions
Broadening...........
Broadening.........................,.,...........ss
Effect...
Compensation.
Sensor
Details.
Hardware
Summary............
Block
Diagram
Module
...
Indicators.
Alarms.
and
Warm-Up
Custom
Synchronization
Fractional
CO2
μμ
Default
Sensors.........................
Saturation
Compensation...
Values...
ΕΟΟ
....
Analysis
Circuit
Analysis
Time..
Set-Up
for
Pulse
Saturation...
...
…
.
.
Oximetry..
....
...
...
レレ
くく
レト
issues
ニュ
レッ
くく
トー トト ャ ッッ て と
と ャ
て て と て て
ss
ρου
sense
とく
나
…
…
..
„12
..
..
…
....
ここ
..
‚...
„21
„22
ーー
.
..
..
.
..
..
..
..
..
..
…
..
..
…
...
…
...
ας
.
.
1-1
1-1
1-2
1-2
12
1-2
1-3
13
13
13
13
13
1-4
2-1
21
2-1
2-3
2-5
25
2-5
2-6
2-6
2-6
2-6
2-6
2-7
3-1
3-2
3-2
3-2
3-2
3-3
3-8
3-8
3-3
3-3
BA
3-4
3-4
3-6
3-6
TABLE
OF
CONTENTS
(continued)
3.5
3.6
4.1
4.2
4.3
4.4
5.1
5.2
5.3
6.1
6.2
”Disassempbly
7.1
7.2
7.3
8.1
8.2
8.3
9.1
Main
8.5.1
8.5.2
3.5.3
3.5.4
8.5.5
8.5.6
3.5.7
8.5.8
8.5.9
3.5.10
3.5.11
3.5.12
3.5.13
Power
3.6.1
3.6.2
3.6.3
3.6.4
Routine
Introduction............
Cleaning
Changing
Replacing
Packing
Överview...........................
Repacking
Repacking
Disassembly
Introduction............
6.2.1
6.2.2
6.2.3
6.2.4
6.2.5
6.2.6
6.2.7
6.2.8
6.2.9
Testing
Description.....................
COs
7.2.1
7.2.2
Calibration
Troubleshooting........................
Introduction.......
Advisory
Status
Spare
In
Processor
Microprocessor
Power.
Reset
Battery
Internal
Graphical
BEPROM..............
FLASH
Real
Time
Audio
Charging
Patient
Front
Supply-Charger
Power
AC
Rectifier
+5
Volt
C-LOCK
Maintenance
Instructions............
Mains
or
Changing
and
PCB
Control.........
and
Watehdog......................
Backup
and
Display
Memory.
Clock....
Generation
Indicator................
Isolation
Panel
Supply
and
Logic
QRS
Voltage
Shipping......
Circuit
Section.......,...............,....,......,
and
Memory
External
..................,....,,.....,..,.
....
Power
Display
and
Sync
Controller..
PCB
Charger
+12
volt
Display
Input
......................
Input
the
Fuse...
Details.
Retention............
Serial
Supply...
Communications..
Circuit
Theory
Battery
Power
Circuit...
....
In
Original
In
New
Guide........
Procedure
RemovinginstrumentCover...........................
Removing
Removing
Removing
Removing
Removing
Removing
Removing
Removing
and
Calibration
Carton..
Carton
Front Panel
Power
Transformer..........
.......
了
attery.
Oximetry
Main
Pull-Out
Speaker............................
............
and
C02
Processor
Assembly
Card
Supply-Charger
Display
ConventionS
Service
Breath
and
Screen
Detection
Accuracy
and
Display
and
Check
Conventions...................
Display
of
Messages.
Messages.....
Parts........
го
ас
6101...
sise
ーー
...
essences
Details.........
of
Charger...........
Supply
Subassembly..
PCB...........................
.............................,................,.....
Tray...
Assembly.
0
End-Tidal
Operation..
....
…
<
scccrcre*
CO2
ui
iii
Values..
…
ees
enten
ennen
eker
0040000400000
000000e0
..
nernnse
„
en
…
..
.
.
..
..
..
72
7-6
8-1
8-1
8-6
8-7
9-1
9-1
iv
TABLE
OF
CONTENTS
(continued)
10
10.1
1
11.1
11.2
113
114
185
11.6.
11.7.
¡ACA
Overview.........
Specifications.............
Physical/Environmental.........
Components
Controls
Connectors................
Performance.
Calibration.....
Airway
and
Adapter
11.8 再 jectrical.
2
12.1
2-1
2-2
3-1
3-2
3-3
3-4
3-5
8-6
8-7
4-1
4-2
6-1
7-1
Schematic
Introduction...................,......,.................
Oxyhemoglobin
Nondispersive
Overall
Oximetry
LRD
Input
Input
A:D
Support
North
European
Disassembly
Failed
Block
Module
Driver
Signal
Amplifier,
Conversion
American
LED
essences
and
User
Interface..
Indicators...........
EP O KO KO R O R K
0
0000
Koh
OK S KO
RS O O
Diagrams..
see
LIST
OF
FIGURES
Dissociation
Infrared
Diagram..........................
Overall
Cireuit に し に
Processing.
Synchronous
Circuits.......................,.......,,,...........
Circuits...................
Fuse
Аттапветейв.............
Fuse
Diagram..............
Атгапретейф...............
Digits
Curve..........................................
CO2
Block
に に に レート
Analyzer..
Diagram...
ーーー・
Detector,
and
Filter/Amplifier
линии
нии
ини
тни
KOK
KK P KKK
низине
ss
KK O RR
oP P een
K R
0.0000
öseeeeeea
PK O v
нение
наити
ek
6
εως
Kone
..
.
7-1
8-1
The
United
OXISENSOR;;
The
4,770,179;
Service
Troubleshooting
following
States
N-6000
Screen
are
trademarks
and
other
and
is
covered
4,802,486;
IndicatiomS
Guide
countries;
ULTRA
by
4,869,254; 4,928,692;
the
following
CAP,
of
Nellcor
C-LOCK,
which
レレ
ーー
patents:
トレ
ーー
トト
ーー
Incorporated:
which
is a
commercial
U.S.
4,934,372;
NELLCOR,
is
registered
Patent
and
スー
スト
ドー
くに
トー
which
in
the
trademark.
No.
4,621,643;
corresponding
トト
トト
トス
is
United
4,653,498;
patents
スレ
(レレ に レレ に トス
registered
States;
4,700,708;
in
other
countries,
レト
in
ストー
the
て
SYMBOLS
O/O
Sc
nx
el
On/Standby
Pulse
tone
volume
Audible
C-LOCK
Low
alarm
signal
battery
off
lost
FRONT
PANEL
eke
MENU
A
+
ni
Freeze
Menu
Audible
C-LOCK
Battery
alarm
in
in
use
on
use
κ
220/240V~
FRONT
PANEL
A
Battery
INSTRUMENT
Attention:
manual!
charging
ONLY
Refer
to
operator’s
Type
BF
equipment,
(patient
electrically
isolated)
vi
SYMBOLS
AS
=
MO
À
Battery
ECG
fuse
input
On/Off
BF
Type
(patient
equipment,
electrically
isolated)
BACK
PANEL
O
<>
RS232
Fuse
replacement
Eguipotential
RS232inputoutput
ground
port
110/120
BACK
VOLT
PANEL
INSTRUMENTS
Instrument
proof.
ONLY
not
anesthetic
VAN
Attention:
Refer
manual!
to
operator’s
vii
WARNINGS
The
N-6000 contains
service
.
[For
The
from
connection,
must
be
protection
N-6000
is a patient-connected
potentially
the
no
performed
against
fire
dangerous
procedures
user-serviceable
by
qualified
hazard,
personnel.
replace
medical
electrical
and
potentials
part
specifications
parts.
fuses
device.
For
protection
only
Isolated
or
ground
contained
with
against
the
same
patient
paths.
in
this
electrical
type
and
connectors
To
protect
manual
rating.
protect
the
integrity
must
hazard,
the
be
adhered
all
patient
of
this
to.
viii
1.1
INTRODUCTION
This
manual
and
capnograph,
personnel,
Note:
service
language
screen,
and
then
who
This
screens.
option
press
press
“LANGUAGE”
from
the
left.
or
“ENGLISCH.”
ㆍ
The
N-6000
time
monitoring
carbon
.
SpO2
SpO2
numerically,
shows
contains
information
model
have a technical
manual
the
If
to
Freeze
the
the
is
written
unit
English.
second
(LANGUE
Press
this
is a compact,
of
dioxide
is
noninvasively
sensors
(ETCO2),
which
while
the
patient’s
N-6000.
for
is
configured
To
reach
button
key
and
in
to
on
fourth
French
gain
oxygen
and
and
are
available
the
plethysmograph
relative
SECTION
I
Introduction
for
servicing
Service
of
this
background
an
N-6000
for
the
language
the
front
soft
keys
and
SPRACHE
access
microprocessor-controlled
saturation
respiratory
continuously
for a variety
pulse
strength
the
product
in
analog
configured
French
panel
from
to
the
language
(SpO2),
rate
measured
is
displayed
as
NELLCOR®
must
and
for
or
German,
selection
while
the
the
left
in
German)
pulse
(RR).
of
patient
measured
ULTRA
be
done
digital
by
electronics.
CAP™
qualified
English-language
the
technician
menu
main
at
selection
instrument
rate
using
from
monitoring
the
same
appears
(PR),
Nellcor's
sites.
the
time.
above
menu
and
used
for
inspired
reusable
SpO2
values
main
screen
The
CO2
as a waveform, A bar
by
the
pulse
oximetry
pulse
service
displays
must
and
change
monitoring
is
displayed,
legend
the
fourth
select
“ANGLAIS”
continuous
(ins.),
or
adhesive
are
displayed
graph
circuits.
oximeter
the
soft
key
real-
end-tidal
indicator
ㆍ
ETCOs
intubated
exhalation
This
manual
digital
N-6000
maintenance
12
Before
These
proper
WARNING:
service
Caution:
always
patient
operator’s
WARNINGS,
you
terms
operation
technician.
Cautions
in
boldface
Note: A note
is
determined
patient’s
is
is
intended
monitoring
or
repair.
read
the
identify
A
warning
gives
by
positioning
airway
displayed
for
(15
mm
numerically,
use
by
equipment.
manual
CAUTIONS,
N-6000
of
this
and
be
AND
service
information
instrument.
familiar
manual,
that
describes a condition
WARNINGS
are
describes a condition
type.
information
that
an
diameter).
while
authorized
Service
with
NOTES
it
pertains
that
always
warrants
that
in
may
infrared
The
CO2
is
service
personnel
instrument
is
important
to
technician
may
result
boldface
result
special
attention.
mainstream
amount
of
displayed
personnel,
must
have
operation
to
understand
and
in
injury
type
and boxed.
in
damage
CO2
sensor
CO2
present
as a waveform.
trained
read
patient
to
in
ahd
before
the
safety,
to the
the
instrument.
servicing
understood
attempting
following
patient,
across
at
the
and
an
end
of
analog
the
terms.
indicates
operator
Cautions
and
or
are
1-1
1.3
DESCRIPTION
The
following
1.3.1
Visible
The
N-6000
.
C02
*
provides a qualitative
9
Trends
paragraphs
and
features
and
plethysmographic
are
displayed:
respiratory
ㆍ
When
the
switch, a battery
1.3.2
Audible
Indicator
describe
Audible
Indicators
numeric
displays
indication
acquired
SpO2
rate
for
trend
alone;
trends;
all
SpOz,
operator connects
charging
the
N-6000
of
SpO2,
waveform
of
pulse
parameters,
CO2
trend
CO2,
pulse
the
monitor
indicator
lights.
and
list
PR,
RR,
displays:
strength
and
the
alone;
rate,
to
AC
SpO2
and
power
important
and
ETCO2:
and a pulse
at
the
oximetry
following
and
CO2
respiratory
and turns
features.
amplitude
five
trend
trends;
rate
on
the
bar
sensor.
screens
pulse
trends.
rear-panel
graph,
can
rate
and
which
be
on/off
Pulse
pitch
decreases.
can
1.3.3
The
moves
apnea,
priority
the
button
ALARM
function
Note:
Note:
rate
of
this
watch
Visible
monitor
outside
or
of
alarm
alerts
SILENCE
disabled.
are
always
The
When
audible
1.3.4
A
Status
status
alarm
message
identified
and
tone
This. early
the
patient
and
has
when a sensor
the
alarm
temporarily. A flashing
the
operational.
SpO2
the
Messages
by a module
troubleshooting
oxygen
changes
warning
and
Audible
both
visible
an
adjustable
state.
operator
indicator
Visible
and
ETCO2
French
function
is
displayed
the
problem.
saturation
with
listen
are
indicated
variation
system
for
encourages
SpO2
in
changes
Alarms
and
limit
gets
disconnected.
Pressing
that
the
warns
alarms
numeric
language
audible
(operator-defined),
the
ALARM
alarm
that
appear
is
chosen
alarms.
The
ALARM
SILENCE
tone
one
on
indicators
and
disabled, a single tone
in
case
an
error
identifier
and
(See
an
error
section
VIII
audibly
SpO2,
tone
SILENCE
has
or
more
the
monitor
change
one
occurs
condition
code
for
with a tone
rising
prompt
as
saturation
corrective
simultaneously.
These
when
and
alarms
the
monitor
pattern
button
indicator
been
silenced
parameters
temporarily. A steady
sereen,
from
green
or
more
of
the
every
three
occurs.
number
more
that
information.)
that
increases
action
are
activated
detects
of
the
audible
turns
adjacent
have
had
and
unlike
to
red
parameters
minutes.
An
error
assists
signals
since
off
the
to
the
their
audible
when
condition
service
each
and
the
pulse.
falling
The
as
clinician
when a variable
loss
of
pulse
or
alarm
audible
ALARM
audible
depend
portion
SILENCE
ON
of
alarm
on
the
alarms,
an
alarm
have
had
personnel
the
will
occurs.
error
be
in
it
the
of
they
1.3.5
Automatic
The
monitor
test
takes
that
the
allowing
45
seconds.
approximately
program
the
Self-Test
automatically
mainstream
If
an
error
“Troubleshooting,”
1.8.6
On-Screen
The
on-screen
at
the
bottom
function
MENU
1.8.7
The
Once
1.3.8
The
button
N-6000
N-6000
configured,
Trend
N-6000
memory
the
memory
memory
continues
key
can
can
while
Menus
menu
of
the
is
pressed,
to
display
Default
power-on
Memory
stores
be
viewed
is
full,
be
viewed
the
memory,
is
for
more
guides
screen
and
default
the
custom
up
to
in
the
oldest
on
trend
and
Warm-Up
performs a series
15
seconds
data
COz
after
memory,
sensor
to
detected, a status
information.)
the
operator
just
above
the
the
screen
the
top level
Custom
24
hours
30-minute,
the
data
displays a new
Default
settings
alarm
data
of
trend
are
limits
sereen
are
being
menu,
can
2-hour, 4-hour,
automatically
and
Time
of
the
operator
and
internal
warm.
The
message
—
through
four
function keys
or
to
Set-Up
be
customized
will
always
data
for
can
be
printed.
diagnostic
turns
circuitry
CO2
sensor
appears.
all
system
menu
return
with
to
according
be
CO2,
SpO2,
8-hour,
erased
printed
with a printer.
tests
when
on
the
are
warm-up
(Refer
functions.
that
are
additional
the
main
in
place,
pulse
12-hour,
as
new
turned
monitor.
These
functioning
time
to
Section
Menu
used
to
select
functions.
monitoring
to
institutional
even
after
rate,
and
or
24-hour
data
are
Patient
on.
The
system
tests
properly,
is
approximately
8,
items
an
are
item.
Press
screen.
requirements,
power-down.
respiratory
segments.
stored.
Data
monitoring
self-
confirm
while
displayed
When
the
rate.
Trend
When
stored
a
in
1.3.9
Automatic
The
SpO2
subsystem
the
system
is
turned
whenever a new
calibrated.
1.3.10
Battery
If
external
90
minutes
minutes
ㆍ
by
By
choosing
power
on
using
(including
w/BATTERY
displays
.
battery
When
AC
the
the
life.
the
MAINS
MENU
operator
Calibration
of
on
oximetry
Operation
is
lost
its
rechargeable
the
GRAPHICS
GRAPHICS
all
waveforms,
OFF,
only
the
numerical
GRAPHICS
power,
button
reapplies
the
N-6000
and
periodically
sensor
or
transportable
is
internal
w/BATTERY
messages,
the
display
values
w/BATTERY
the
display
to
turn
the
AC power
is
fully
thereafter.
connected.
operation
battery.
w/BATTERY
remains
of
option
will
only
GRAPHIC
to
the
self-calibrating.
It
Additionally,
The
capnography
is
necessary,
This
operating
power
ON
etc.)
will
blank
SpO2
is
blank
w/BATTERY
in
the
continue
during
and
ETCO2,
selected
when
saver
SYSTEM
battery
to
the
option
monitor.
is
calibrated
it
subsystem
the
N-6000
time
option.
menu,
to
operate.
This
allows
be
OFF
unit
transfers
ON.
automatically
is
recalibrated
of
can
can
be
extended
the
display
By
choosing
operation.
up
to
and
the
to
The
display
automatically
the
N-6000
operate
up
graphics
The
N-6000
180
minutes
unit
is
battery
returns
whenever
is
factory
up
to
to
180
GRAPHICS
of
operating
power.
Press
when
on
1.3.11
Noninvasive
Oximetry
Sensors
Noninvasive
two
light-emitting
differences
infants,
1.3.12
children,
C-LOCK™
If a patient
quality
separate
signal
for
signals
from
synchronize
When
If
SIGNAL
N-6000
the
C-LOCK
ECG
LOST
is
not
NELLCOR
diodes
in
tissue
is
moving
thickness
and
adults.
ECG
Synchronization
or
measurements
that
reflect
the
ECG.
The
the
SpOz
measurements.
ECG
synchronization
signal
is
lost
symbol
being
used.
oximetry
(LEDs)
or
Refer
has
poor
of
oxygen
cardiac
N-6000
or
deteriorates
appears
on
sensors
as
light
sources.
skin
pigmentation.
to
specific
for
perfusion,
saturation.
activity:
uses
the
ECG
is
used,
to
the
the
display.
obtain
sensor
Pulse
C-LOCK
When
an
QRS
the
point
No
measurements
The
N-6000
Specific
directions
Oximetry
ECG
this
feature
optical
signal
complex
C-LOCK
that
symbol
it
IN
can
is
:
by
optical
adjusts
sensors
automatically
are
for
use
for
synchronization
is
used,
the
from
the
sensor
to
help
identify
USE
symbol
no
longer
displayed
be
if
used,
the
means
available
complete
can
enhance
N-6000
and
the
pulse
appears
the
C-LOCK
C-LOCK
alone,
using
for
for
neonates,
information.
signal
receives
an
electrical
and
on
the
display.
feature
of
two
the
2.1
OVERVIEW
This
section
2.2
PULSE
The
N-6000
plethysmography.
sensor
photodetector.
920
nm).
The
photodetector
absorption.
With
each
hemoglobin
The
N-6000
hemoglobin
In
principle, a pulse
Absorption
both
wavelengths
describes,
OXIMETRY
oximetry
has two
When
low-voltage
One
the light
heartbeat, a pulse
differs
measures
that
is
also
in
subsystem
It
includes
LED
measures
from
red
is
saturated
oximeter
measured
results
general
SUBSYSTEM
light-emitting
emits
from
deoxygenated
and
in a value
terms,
is
based
an
‘electro-optical
red
light
the
LEDs
the
light
of
oxygenated
infrared
with
oxygen.
measures
when
pulsatile
SECTION
Principles
operating
on
the
diodes
(nominal
passes
that
hemoglobin
absorption
for
the
through
passes
arterial.blood
the
light
arterial
arterial
II
of
Operation
principles
principles
sensor
(LEDs)
through,
absorption
and a microprocessor-based
as
660
nm)
the
flows
in
its
to
determine
blood
oxygen
for
the
N-6000.
of
spectrophotometry
light
sources,
and
the
other
sensor
which
relative
is
saturation
to
by
tissues
in
the
the
site,
part
indicates
the
sensor
red
and
percentage
and
tissue.
(SpO2).
and
and
one
photodiode
emits
infrared
of
the
light
red
and
infrared
site.
Oxygenated
infrared
of
functional
nonpulsatile
The
ratio
module. The
as
(nominal
is
absorbed.
absorption.
blood.
of
absorption
a
at
2.2.1
C-LOCK
Through
identifying
during
When
an
complex
with
length
relationship
random
If
an
optical
C-LOCK
C-LOCK
patient
provided
optical
is
the
patient’s
of
the
motion.
ECG
pulse
ECG
signal
detected,
signal
ECG
Synchronization
ECG
the
pulse
and
movement
with
an
from
the
an
physiology
delay
is
relatively
is
used
to
identify
is
not
provided,
alone
is
used
synchronization
synchronization,
synchronizing
and
when
the
ECG
signal,
sensor
optical
and
to
determine
pulse
the
and
an
is
with
the
stable.
“good”
or
if
that
resumes
detected
Through
pulses
the
N-6000
SpO2
measurements.
patient’s
N-6000
electrical
at
location
and
signal
the
pulse
when
an
uses
an
perfusion
receives
signal
the
of
the
C-LOCK
reject
deteriorates
rate
adequate
is
two
from
sensor
sensor.
ECG
nonsynchronized
and
to
ECG
signal
This
poor.
signals
the
site.
The
However,
synchronization,
so
that
initiate
ECG
signal
as a reference
enhances
that
ECG. A short
length
it
can
saturation
is
signal
reflect
for a given
artifacts
no
cardiac
time
of
this
that
such
longer
measurements.
available.
point
quality
activity:
after a QRS
delay
varies
patient,
time
as
be
used,
for
the
-
the
2-1
2.2.2
Automatic
The
oximetry
calibrated
connected.
differences
Each
sensor
determined
calibration
obtained
2.2.3
by
Functional
Calibration
subsystem
each time
Also,
the
in
tissue
is
calibrated
and
encoded
resistor
that
sensor.
versus
incorporates
it
is
turned
intensity
of
thickness
when
into a calibration
to
determine
Fractional
automatic
on,
at
periodic
the
sensors
and
pigmentation.
manufactured:
resistor.
the
appropriate
Saturation
calibration
intervals
LEDs
is
the
effective
calibration
thereafter,
adjusted
The
instrument’s
mechanisms.
and
automatically
mean
wavelength
coefficients
It
is
automatically
whenever a new
to
compensate
of
the
software
for
reads
the
measurements
red
this
sensor
for
LED
is
is
Because
those
expressed
N-6000
functional
the
N-6000
of
instruments
as a percentage
uses
two
SaO2.
carboxyhemoglobin
In
contrast,
Oximeter
measured
Measured
Consequently,
other
measurements
some
report
hemoglobin,
dysfunctional
to
instrument
can
functional
2.2.4
Measured
When
oxygen
saturation
(PaO2),
because
the
relationship
carbon
dioxide
the
versus
the
calculated
between
(PaCO2),
measures
that
measure
of
wavelengths,
It
does
not
or
methemoglobin.
laboratory
fractional
SaQ2-—~oxygenated
whether
hemoglobins
directly
must
measure
be
converted
compare
saturation
Calculated
is
calculated
calculated
saturation
PO2
the
functional
the
hemoglobin
it
measures
detect
instruments
or
not
N-6000
functional
using
=
100 — (%carboxyhemoglobin + %methemogiobin
SaO2,
fractional
SaO2.
that
oxygenated
significant
such
that
hemoglobin
are
included.
measurements
SaO2.
the
following
it
may
produce
Functional
is
capable
and
amounts
as
the
of
Instrumentation
hemoglobin
is
If
it
measures
equation:
fractional
measurements
SaQ2
of
transporting
deoxygenated.
dysfunctional
expressed
available
with
those
fractional
saturation
Saturation
from a blood
value
may
may
and
saturation
concentrations
not
gas
measurement
differ
from
have
been corrected
(Figure
of
2,3-DPG
the
N-6000
2-1):
temperature,
and
of
the
SpO2
for
fetal
is
oxygenated
oxygen.
hemoglobin,
hemoglobin,
Laboratory
as a percentage
for
oxygen
of
another
Sa02,
~
partial
pressure
measurement.
the
effects
pH,
of
the
hemoglobin.
that
differ
hemoglobin
Because
yielding
such
282
CO-
of
all
transport.
instrument,
those
x
of
arterial
This
variables
partial
that
pressure
from
the
as
that
is
shift
of
2-2
100
—
pH
t
Temperature
+
PCO2
+
+
2,3-DPG
=
ο
я
504
5
©
.
の
a
+
VA
0
ifs
4
+
+
nf.
+
7
の
;
+
4
e
+
+
“e
¿e
΄
+
7
+
.
e
+
e
è
>
e
.”
>.
+
„==
„>“
ㆍ
+
+
+
+
|
est
n°
pH
Temperature
PCO2
2,3-DPG
50
PO:
(mmHg)
=.
100
2.3
CAPNOGRAPHY
The
N-6000
present
sensor
ventilator
elbow
into
Infrared
infrared
specific
the
versus
have
at
the
that
circuit
and
the
contact
spectroscopy
light.
wavelengths
infrared
the
similar
proportional
determined
The
CO2
detection
“optical
bench”),
wavelengths
During
reaches
selectively
second.
monitoring,
the
absorbed
Three
uses
nondispersive
end
of
exhalation
attaches
to a disposable
either
patient
with
Of
the
respired
the
wye.
can
normally
of
spectrum
wavelength
spectra
to
by
that
the
concentration
comparing
mechanism
has
an
corresponding
light
narrow-band
by
measurements
Figure
2-1:
SUBSYSTEM
infrared
(ETCO2).
between
infrared
of a molecule
of
have
infrared
first
infrared
0602.
the
The
CO2
gases.
be
used
respired
light),
is
overlapping
the
absorbance
used
to
an
passes
It
then
are
to
light
of
source
absorption
Oxyhemoglobin
spectroscopy
It
features a small
airway
adapter.
endotracheal
mainstream
measure
gases,
(i.e.,
have
(usually
unique
to.
absorption
the
absorbing
to
that
in
the
N-6000
that
peak
through
filter,
which
encounters
obtained
during
Dissociation
to
The
tube
sensor
the
concentration
only
CO2,
an
infrared
displayed
that
molecule.
molecule,
of a known
(commonly
is
optically
of
the
the
respiratory
was
selected
the
chopper
each
quantitatively
lightweight
airway
and
fits
as a graph
peaks.
filtered
CO2
rotation:
the
ventilator
on
top
N20,
and
spectrum).
However,
Because
an
standard.
referred
spectrum
gas
because
wheel,
adapter
of
of
any
of
unknown
to
in
Curve
measure
the
“mainstream”
is
inserted
circuit
the
adapter
molecule
water vapor
This
light
different
the
to
as
provide a narrow
(see
the
airway
it
passes
which
rotates
or
and
that
selectively
absorption
absorbed
molecules
absorption
concentration
the
CO2
Figure
adapter.
wavelengths
many
amount
of
CO2
into
the
between
by a molecule
sensor
2-2).
does
not
absorbs
pattern,
may
of
light
can
band
Next
times
the
absorb
be
or
of
that
CO2
come
is
it
are
each
ο
wheel
A
sample
and
measurement
then
reaches
the
is
made when
detector
(i.e.,
light
the
light
passes
passes
through
through
the
the
open
respiratory
area
gas).
of
the
chopper
.
A
reference+sample
containing a known
respiratory
ㆍ
A
reaches
The
ratio,
determine
from
the
The
N-6000
assumed
N-6000.
expressed
are
assumed
is
being
to
fully
The
N-6000
dynamically
several
ETCO?
The
value
obtainable
simply
value
ETCO2.
patient.
gas
and
dark
measurement
the
detector).
“reference+sample/sample”,
the
CO2
analysis
measures
conditions
When
displayed
saturated
spontaneous
value,
average
of
ETCO2
This
units
as % dry
to
be
detects
from
the
of
ETCO2
from
the
all
seen
results
measurement
CO2
concentration
gas
in
the
is
concentration
of
the
CO2
the
partial
of
33° C and
of % (by
gas.
By
convention,
measured
on
the
conditions
and
maximum
breaths
N-6000
that best
patient
breaths
within
in a stable
from a patient
front
counts
with
accurately
during
together
the
is
obtained
and
reference
made when
in
waveform
airway
fully
volume)
panel
(BTPS ) before
breaths
and
minimum
low
estimates
last 8 seconds
ETCO2
cell).
light
is
then
respiratory
(capnograph).
pressure
saturated.
are
displayed
all
readings
and
LED
in
when
ETCO2
detects
the
forced
to-display
exhalations.
value
when
then
reaches
strikes a solid
used
in
gas.
Respiratory
of
CO2
Barometric
on
of
corrected
units
of
being
displayed.
the
COz2
level
CO2
values.
values
and
true
the
and
that
followed
counts
alveolar
ETCO2
displays
best
light
passes
the
detector
area
the
sensor-specific
rate
in
the
patient
pressure
the
front
panel
CO2
posted
to
body
mmHg
crosses a threshold,
Hence,
all
breaths.
CO2
Unlike
value,
that
approximates
on
temperature
or
kPa,
when a ventilated
by a mechanical
value
conventional
the
maximum
through a reference
(i.e.,
light
passes
of
the
is
determined
airway
is
measured
LEDs,
the
the
values
is
the
N-6000
the
arterial
wheel
front-panel
maximum
(1.9.,
calibration
adapter
the
measured
(37°
C).
are
which
breath
capnometers
looks
for
value
for
PaCO
gas
through
no
light
equation
by
the
N-6000
at
normal
directly
also
is
value
the
the
by
value
LED
display
When
patient
with a higher
PCO2
converted
set
has
that
maximum
true
value
in
cell
to
the
is
the
2-4
Light
source
Figure
2-2:
!
Airway
adapter
Nondispersive
Infrared
filter
Chopper
wheel
Infrared
Reference
gas
cell
Open
area
Photodetector
CO2
Analyzer
24
N20/O2
COMPENSATION
Unless
affect
To
oxide
screen
The
oxide
in
20%
for
The
provides
there
The
provides
there
Finally,
barometric
changes.
compensated
the
ensure
with
menu
N-6000
and
O2
beyond
of
increase
every
first
is
zero
second
is
zero
barometric
the
the
for,
measurement
accurate
oxygen; and
measurements,
selections.
is
calibrated
20%
oxygen.
the
reference
in
O2.
20%
increase
compensation
necessary
error
under
compensation
correction
error
under
pressure
pressure
require
elevated
of
the
concentration
(2)
high
for
low O2
This
is
the
state
An
increase
in
N20.
level
assumes
correction
these
level
factor
for
these
changes
no
levels
the
N-6000
oxygen.
compensation,
default
decreases
in
N20
50%
factor
assumed
assumes
high
levels
assumed
also
user
intervention;
of
oxygen
of
Compensation
mode
the
increases
nitrous
for
high
conditions,
0%
nitrous
conditions.
affect
(O2)
CO2
by a collision
provides
which
where
displayed
oxide
levels
of
O2.
Once
measured
the
and
nitrous
two
levels
levels
COMP. = OFF.
the
of
oxide
N-6000
are
assumes
CO2
value
displayed
and
50%
N20.
and
60%
again,
CO2
automatically
oxide
(N20)
broadening
of
compensation:
user
selectable
the
reference
In
the
by
1.5%
of
CO2
value
oxygen.
When
oxygen. Selecting
when
values.
Selecting
this
correction
this
correction
However,
adjusts
in
the
effect.
via
state
N-6000,
reading
by
1.6%
changes
airway
(1)
nitrous
display
of
0%
nitrous
an
increase
for
every
of
reading
COMP. = N20
is
applied,
COMP. = O2
is
applied,
in
for
these
The
following
conditions,
2.4.1
Standard
The
accuracy
CO2
in
balance
38
mmHg),
errors
2.4.2
Ambient
energy
barometric
CO2
additional
from
mmHg
are
Pressure
barometric
exchanged
values
the
standard
for
paragraphs
as
pertains
Gas
specifications
air
and a barometric
required
Broadening
pressure
reported
residual
every
to
Conditions
(21%
O2),
for
barometric
pressure
in
molecular
is
automatically
on
both
errors
reference
5000-foot
contain
N-6000
for
Compensation
may
change
further
operation.
the
N-6000
at
airway
pressure
pressure
affects
collisions
the
main
be
observed
state
of
in
in-depth
refer
conditions
of
760
(altitude),
measured
alters
measured
monitoring
when
760
mmHg
altitude.
discussion
to
the
following
of
33
°C,
fully
mmHg.
CO2
the
Additional
N20,
02,
by a “pressure
absorption
and a compensation
screen
testing
(sea
level).
spectrum
and
the
reference
These
of
compensation
standard
saturated
corrections
and
broadening”
service screen.
gases
amount
conditions:
(water
water
vapor.
of
is
automatically
at
to
and
standard
test
vapor
and/or
CO2.
pressures
approximately
residual
effect,
In
the
applied
However,
gas
is
pressure
in
which
N-6000,
to
different
+0.5
of
2-5
2.4.3
N20
Collision
Broadening
Nitrous
“collision
absorption
eliminate
causes
standard
correct
both
mixture
2.4.4
Oxygen
exchanged
causes a decrease
Oxide
an
(N2O)
broadening”
spectrum
any
direct
increase
in
can
affect
effect
in
of
CO2.
The
absorption
measured
conditions. A software-selected
for
high
N20,
assuming a gas
the
N-6000
if
O2
Collision
(02)
the
02
can
in
molecular
main
monitoring
compensation
Broadening
affect
the
CO2
collisions
in
measured
conditions. A software-selected
high
O2,
assuming a gas
main
monitoring
compensation
2.4.5
Water
Water
vapor
respiratory
water
N-6000
are
to
vapor
is
calibrated
automatically
verify
calibration,
sereen
option
Vapor
also
has
gas
is
(with a vapor
composition
and
the
is
selected.
Effect
an
effect
at
standard
pressure
to
include
compensated,
CO2
values
airway
the
CO2
measurement
which
of
CO2
energy
infrared
infrared
of
approximately
exchanged
narrow
energy
(user-selectable)
composition
screen
option
measurement
alters
CO2
and
on
is
selected.
the
absorption
of
approximately
by a “collision
(user-selectable)
of
60%
O2
(0%
service
on
CO2
sereen
measurements.
conditions
of
38
mmHg)
this
effect,
while those
reported
and
on
CO2
on
the
by
bandpass
by
of
50%
the
service
-0.75%
02
compensation
N20).
are
compensated
of
33°C,
causes
values
the
service
service
both
in
filter
N20.
+0.8%
N20
N20
screen
spectrum
CO2
In
normal
fully
an
increase
reported
screen
screen
direct
absorption
molecular
used
in
However,
per
10%
increase
compensation
and
50%
are
compensated
broadening”
of
CO2.
per
10%
increase
option
values
for
use,
this
reported
it is
saturated.
in
on
the
are
not.
should
be
of
collisions
the
N-6000
N20
collision
in
option
O2.
CO2
values
effect
O2
collision
in
is
provided
on
gas
mixture
assumed
Under
measured
main
If
applying a dry
used
for
infrared
alters
is
chosen
the
broadening
N20
from
is
provided
reported
for
this
in
which
broadening
O2
from
to
correct
both
the
if
the
that
these
conditions,
CO2
of
6%.
monitoring
comparison.
and
by
to
on
gas
energy
standard
for
N-6000
O2
The
screen
test
gas
a
to
2.4.6
BTPS/ATPS
The
N-6000
temperature
adapter
barometric
conversion
fully
is
measured
pressure.
to
saturated
screen values
2.4.7
Summary:
CO2
water
a
dry
also
CO2
water
values
test
applied
values
reported
vapor
gas
reported
vapor,
Compensation
assumes
and
deep
that
measured
pressure
fully
at
33
°C,
By
convention,
lung conditions
with a water vapor
do
not
have
these
Reported
correction).
is
present
to
the
service
and,
if
CO2
on
the
That
in
the
screen
on
the
selected,
saturated
fully
saturated
of
pressure
corrections
Values
service
is,
CO2
airway
values.
main
monitoring
N20
and
respiratory
(ATPS).
with a water
CO2
values
body
temperature
of
47
applied.
on
the
Main
screen
are
values
adapter
reported
(at
screen
O2.
gases
are
That
is,
vapor
are
reported
and
mmHg,
at
Monitoring
uncorrected
on
the
25
°C).
Corrections
have
at
standard
it
is
assumed
pressure
on
the
pressure
ambient
and
for
BTPS/ATPS
service
all
corrections
airway
that
of
38
main
monitoring
fully
saturated
barometric
Service
screen
for
Screens
are
N20
applied:
conditions
the
gas
in
mmHg,
at
(BTPS):
pressure.
conditions
accurate
and
Oz,
if
BTPS/ATPS,
of
airway
the
airway
ambient
screen
after
37
Service
(including
assuming
selected,
°C,
are
2.5
FACTORY
Each
CO2
sensor
temperatures.
components
calibrated
automatically
and
over
CALIBRATED
is
individually
The
N-6000
electronics
multiple
compensated
temperatures,
SENSOR
factory-calibrated
sensor
is
at a known
by
temperature
over
temperature-regulated
constant
so
that
temperature.
any
deviation
calibration
coefficients
multiple
to
42
°C,
Additionally,
from
gas
concentrations
which
the
42
°C
stored
keeps
the
the
sensor
set-point
with each
and
sensor
is
is
sensor
multiple
optical
factory-
unit.
3.1
INTRODUCTION
This
section
provides
details
of
circuit
SECTION
Circuit
operation.
Analysis
Refer
HI
to
Figure
3-1,
the
overall
bleck
diagram.
i
Oximetry
Sensor
Heater/Motor/IR
CO,
Sensor
LED
Drive
Output
Sensor
Output
Drive
Source | Module
Sensor
Analog
Oximetr
>
>
Mo.
tule.
CO»
Optical
Isolators
»
Serial
Data
isolated
DC
T
i
・
Serial
|
1
|
Power
1
1
I
.
Data
>
Charger
Power
Transformer
Main
Processor
Battery
and
Supply
Å
AC
Isolated
Audio
isolated
Power
Signals
Keyboard
y
Display
—|
Battery
>
>
speaker
|
Serial
Loud-
Port
3.2
CO2
MODULE
The
CO2
module
information
modes
The
thermistor,
and
mainstream
acquires
to
the
send
sensor
sensor
and
non-volatile
Figure
CIRCUIT
main
DETAILS
CO2
data
processor,
information/status.
has
six
basic
memory.
3-1:
Overall
from
the
The
module
components:
AC
Block
mainstream
accepts
an
IR
Mains
Power
Diagram
CO2
sensor
commands
source,
and
from
the
detector,
communicates
processor
chopper
motor,
the
to
change
heater,
3.2.1
Sensor
Operation
The
IR
source
based
on
components: a filter,
The
filter
frequency,
other
molecules
The
airway
tube
and
within
The
path
The
energy
The
can
third
The
the
chopper
as
second
to
detector
be
determined
(dark)
CO2
condensation
because
the
allows
the
gas
the
sensor's
wheel
the
wheel
phase
pass
phase
sensor
the
sensor’s
amount
emits
energy
of
the
only
attenuation
normally
sample
ventilator
light
is
rotates.
is
an
to
the
detector.
output
is
by
provides a zero
temperature
from
forming
output
that
is
energy
energy
it
receives.
airway
gas
of a wavelength
effects
present
is a disposable
circuit.
The
path.
split
into
three
The
first
opening
sampled
comparing
during
is
elevated
on
the
is
affected
providing
the
directed
sample,
due
to
the
in
respired
airway
sensor
sections
phase
each
change
reference
to
airway
by
toward
In
the
the
light
path
and a chopper
within
presence
the
of
gas.
adapter
clips
or
to
phases
the
that
airway
has a reference
no
attenuation.
phase.
in
42°
windows,
temperature.
The
detector
for
recalibrating
C.
There
and
detector.
The
between
the
wheel.
CO2
absorption
CO2
molecules
is
inserted
adapter,
that
are
presented
cell
filled
The
third
amount
output
are
of
CO2
from
the
detector.
two
reasons
to
maintain a stable
detector
are
generates a voltage
source
band
and
to
much
between
putting
the
sequentially
with a known
phase
phase
is
present
one
for
doing
dark,
in
to
detector
detector
pass.
At
greater
the
endotracheal
respired
to
amount
allowing
the
gas
phase
this:
two.
to
temperature
are
this
than
for
gas
the
light
of
CO2.
no
sample
The
keep
three
The
sensor
maximum
sensor
requires
As
the
3.2.2
The
CO2
and
barometric
3.2.3
The
module
stored
with
EEPROM
waveform.
its
internal
power-up
available
reaches
some
motor
CO2
Module
module
Digital
in
U600.
the
main
U608,
EEPROM
A/D
determination
sequence
heater
temperature,
current;
reaches
speed,
Hardware
has
six
pressure
is
controlled
The
microprocessor
processor.
and
the
U603
converter
in
other
is
optimized
output
by
disabling
main
the
heater
the
heater
sections:
can
the
be
is
heater,
is
digital,
transducer/amplifier.
by
U602,
an
80C552
uses
The
microprocessor
sensor
sections
EEPROM.
stores
to
determine
of
calibration
sensor
the
hardware.
to
keep
power
varied
in
software
temporarily
the
demands
re-enabled,
motor
microprocessor,
its
on-board
communicates
The
external
constants
temperature,
©
supply
to
disabled,
followed
control,
serial
port
A/D
for
the
loading
match
on
by
the
and
the
the
power
activation
signal
amplifier,
running
te
communicate
serially
to
converter
module.
barometric
to a minimum.
power
motor
is
supply
of
the
available.
started.
IR
heater,
at
11
MHz.
A/D
converter
is
used
to
digitize
The
microprocessor
pressure, and
The
As
the
This
are
minimized.
source.
source
The
program
asynchronously
U605,
the
sensor
uses
fault
is
3-2
3.2.4
Motor
The
sensor’s
The
sense
signal.
motor.
signal
processor
3.2.5
Signal
The
“bench”
output
A/D
converter. A positive
microprocessor
to
the
3.2.6
Heater
The
heater
The
FET
the
sensor
temperature
regulate
inhibited
active
processor’s
Control
chopper
coil
is
This
signal
The
motor
to
regulate
via
the
Amplifier
hybrid
and
bias
to
sensor
which
output
signal
is
converted
by
it
to
42°
by a comparator
as
the
sensor
internal
motor
amplified
is
control
the
MTRGATE
circuit
the
detector
used
hybrid
speed
consists
to
generate a synchronized
by
to
U607
properly,
bias
set
the
operating
servos
consists
is
filtered by
the
of a PWM
L600
to a proportional
the
microprocessor. The
C.
As a safety
circuit
temperature
A/D
converter
of
two
the
processor
controls
the
equivalent
signal.
contains
so
that
voltage
is
point
sensor
signal
signal
and
C614
software
precaution,
monitoring
reaches
for
fault
fixed
coils
for
synchronizing
the
amplitude
of a 30
the
necessary
its
output
applied
of
to
the
photodetector. A negative
to
be
from
before
voltage
by the
regulates
the
PWM
the
48°
C.
The
determination.
on
the
stator
drive
voltage,
of
the
ms
period.
circuitry
is
zero-referenced
the
sensor
referenced
the
processor
being
to
applied
bench
the
output
feeding
temperature
heater output
and
three
magnets
and
to
processor
drive
The
to
activity
signal
motor
amplify
derived
can
the
and
using a filtered
bias
zero
volts.
switching a FET
to
the
heater.
hybrid.
power
voltage,
This
applied
the
heater
this
voltage
voltage
override
is
available
on
the
rotor.
generate a motor
to
the
chopper
from
the
sense
be
disabled
sensor
in
the
PWM
voltage
power
The
is
to
the
output
by
the
detector
range
for
the
output
is
from
also
applied
transistor.
temperature
translated
sensor
can
will
to
to
be
become
the
clock
the
in
to
a
3.2.7
Source
A
current
located
source
regulator
in
the
voltage
determination.
3.2.8
Barometer
The
barometer
achieved
The
the
3.2.9
On
reset
turns
by
sensing
transducers
processor's
Status
LEDs
power-up,
and
the
on,
indicating a sensor
temperature,
LED
is
turned
Ifthe
CO2
module
exists,
the
RED
formed
sensor.
is
brought
circuit
by
The
source
back
generates a voltage
voltage
differential
internal
the
initialization
allowing
off,
LED
green
and
is
not
will
A/D
LED
sequence
the
sensor
the
green
receiving
flash
U615
and
is
controlled
to
the
microprocessor's
variations
voltage
is
converter
flashes
five
is
warm-up
to
start
LED
is
commands
at a rate
R645
provides constant
by the
proportional
at
the
pressure
amplified
for
translation
times
momentarily,
taking
place.
condition.
measuring.
turned
on.
from
of 2 Hz.
microprocessor
internal
to
the
transducer
and
zero-referenced
to a barometric
After
the
The
light
remains
When
the
main
the
processor,
current
A/D
of
via
Q608
converter
atmospheric
with a constant
before
pressure
indicating
green
sensor
LED
on
is
that
until
or a sensor
83.3
mA
and
for
pressure.
being
reading.
the
flashes,
the
ready
to
to
the
IR
Q606.
The
fault
This
is
current
presented
processor
the
red
sensor
reaches
measure,
fault
condition
source
applied.
to
has
LED
the
red