This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 Canada.
Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor, OxiCliq, Dura-Y,
Max-Fast and O
XIMAX are trademarks of Nellcor Puritan Bennett Inc.
T o obtain information about a warranty, if any , contact Nellcor’ s Technical Services Department, or your local
representative.
Purchase of this instrument confers no express or implied license under any Nellcor Puritan Bennett patent to
use the instrument with any sensor that is not manufactured or licensed by Nellcor Puritan Bennett.
Covered by one or mor e of the foll owing U.S. Patents and for eign equivalents: 4,621,643; 4,653,498; 4,700,708;
4,770,179; Re. 35,122; 4,802,486; 4,869,254; 4,928,692; 4,934,372; 5,078 ,136; 5,351,685; 5,368,026; 5,485,847;
5,533,507; 5,662,106; and 5,853,364.
Table 19:Operating Status Codes .... ... .... ... ..............................................120
Table 20:Nurse Call Relay Pin States ......................................................123
Table 21:Rating of Nurse Call Relay ........................................................123
vi
Introduction
Warnings
Warnings are identified by the WARNING symbol shown above.
Warnings alert the user to potential serious outcomes (death, injury, or adverse
events) to the patient or user.
WARNING: The sensor extrapolates from the date and time provided by the
N-595 when recording the sensor event record to the sensor. The accuracy of
the date/time is the responsibility of the N-595. It is recommended that the
N-595 user set the time/date to the correct value before a sensor event
record-enabled sensor is connected, and that this date/time not be changed
while the sensor remains connected. Since a sensor with sensor event record
data can be transported from one monitor to another, having discrepancies in
the date/time between monitors and the sensor event record data will affect
the order the sensor event record data appears. To eliminate this possible
problem, all monitors within an institution should be set to the same time.
WARNING: Explosion hazard. Do not use the N-595 pulse oximeter in the
presence of flammable anesthetics.
WARNING: Do not spray, pour, or spill any liquid on the N-595, its
accessories, connectors, switches, or openings in the chassis.
WARNING: Before attempting to open or disassemble the N-595, disconnect
the power cord from the N-595.
WARNING: The LCD panel contains toxic chemicals. Do not ingest
chemicals from a broken LCD panel.
N-5951
Introduction
Cautions
WARNING: The use of accessories, OXI-MAX sensors, and cables other than
those specified may result in increased emission and/or decreased immunity
of the N-595 pulse oximeter.
WARNING: Do not silence the N-595 audible alarm or decrease its volume if
patient safety could be compromised.
Cautions are identified by the CAUTION symbol shown above.
Cautions alert the user to exercise care necessary for the safe and effective use of
the N-595 pulse oximeter.
Manual Overview
Caution: Observe ESD (electrostatic discharge) precautions when working
within the unit.
Caution: Observe ESD (electrostatic discharge) precautions when
disassembling and reassembling the N-595 and when handling any of the
components of the N-595.
Caution: When reassembling the N-595, tighten the screws that hold the
cases together to a maximum of 10 inch-pounds. Over-tightening could strip
out the screw holes in the top case, rendering it unusable.
Caution: When installing the Power Supply or the User Interface PCB,
tighten the seven screws to a maximum of 4 inch-pounds. Over-tightening
could strip out the screw holes in the bottom case, rendering it unusable.
This manual contains information for servicing the Nellcor model N-595 pulse
oximeter. Only qualified service personnel should service this product. Before
servicing the N-595, read the operator's manual carefully for a thorough
understanding of safe operation of the N-595.
WARNING: Explosion hazard. Do not use the N-595 pulse oximeter in the
presence of flammable anesthetics.
2
Description of N-595 Monitor
The N-595 Monitor is intended for the continuous non-invasive monitoring of
functional oxygen saturation of arterial hemoglobin (SpO
N-595 is intended for use with neonatal, pediatric, and adult patients during both
no-motion and motion conditions and for patients who are well or poorly
perfused, in hospitals, hospital-type facilities, intra-hospital transport, and home
environments. For prescription use only.
Note:Hospital use typically covers such areas as general care floors, operating
rooms, special procedure areas, intensive and critical care areas, within
the hospital plus hospital-type facilities. Hospital-type facilities include
physician office based facilities, sleep labs, skilled nursing facilities,
surgicenters, and sub-acute centers.
Intra-hospital transport includes transport of a patient within the hospital
or hospital-type facility.
Home Care use is defined as managed/used by a lay person (parent or
other similar non-critical caregiver) in the home environment.
Introduction
2) and pulse rate. The
Use with any particular patient requires the selection of an appropriate
O
XI-MAX oxygen sensor as described in the N-595 Operator's Manual.
Motion performance claims are applicable to models MAX-A, MAX-AL,
MAX-P, MAX-N, and MAX-I Nellcor O
XIMAX
oximetry sensors.
TM
Through the use of the four softkeys, the operator can access trend information,
select an alarm limit to be changed, choose the language to be used, adjust the
internal time clock, and change communications protocol. The N-595 can operate
on AC power or on an internal battery. The controls and indicators for the N-595
are illustrated and identified in Figure 1 and Figure 2.
N-5953
Introduction
Front Panel
1. SpO2 Sensor Port12. ADJUST DOWN Button
2. AC Power Indicator13. Neonate Mode Indicator
3. ON/STANDBY Button14. CONTRAST Button
4. Low Battery Indicator15. Fast Response Mode Indicator
2. AC Power Connector5. Supply Voltage Selector Switch
3. Data Port Connector
Figure 2: N-595 Rear Panel
4
Softkey Menu
Introduction
The N-595 hierarchy is outlined below. The user chooses the type of trend data to
view by selecting either Monitor trend or Sensor trend data in the Trend menu.
Sensor sub-menu choices differ depending on what type of in-sensor data is stored
in the sensor chip, such as, event or loop.
The menu structure includes BACK softkey options that allow the user to move
back to the previous menu level without exiting the Trend menu entirely. Trend
data must be compiled on entry/reentry to the T rends menu. When the softkeys are
available, both BACK and EXIT options are available. The BACK softkey goes to
the previous level and the EXIT softkey goes to the main menu. If only one space
is available the BACK Softkey is included, this may require going back one or
two levels to get to an EXIT softkey.
The BACK and EXIT softkeys are positioned on the right-most softkeys,
respectively.
The below menu structure identifies:
•BOLDFACE TYPE — softkey title as displayed on the monitor
•Underlined Text
— description of the softkey menu item
•Italicized Text — the destination of the BACK and EXIT softkeys
---SCALE (for current view, cycle through ±5, ±10, ±15, ±20, ±25,
±30, ±35, ±40 and ±50 of the max and min. values under the
cursor, default to 10 to 100 if there is no data point under the
cursor)
---AUTO (based on all of the graphed trend data: maximum value,
rounded up to nearest multiple of 10, minimum value, rounded
down to nearest multiple of 10 minus 10)
---BACK (back to Monitor menu)
--NEXT (Delete/Print1 Menu)
---DELETE
----“DELETE TRENDS?”
-----YES (to Main menu)
-----NO (back to Delete/Print1 menu)
---PRINT
---BACK (back to Monitor menu)
---EXIT (to Main menu)
--BACK (back to Trend menu)
-SENSOR (Sensor/Event Menu)
(if Event data is in the sensor, the following menu, the Screen will remain in
the appropriate state until the next menu selection is made)
--GRAPH (Graph Menu)
order; up/down also scroll through events in order)
---< (show previous graph, only available when there is a previous
graph)
---> (show next graph, only available when there is a next graph)
---PRINT
---BACK (back to Sensor menu)
--TABLE (Table Menu)
---^ (show previous table, only available when there is a previous
graph; bottom/top line repeats in new table)
---v (show next table, only available when there is a next graph;
bottom/top line repeats in new table)
---PRINT
---BACK (back to Sensor menu)
---EXIT (to Main menu)
(Sensor/Loop Menu)
will be displayed)
--VIEW (Sensor Trend View Menu)
---DUAL (shows SPO2+BPM)
---SPO2
---PULSE
--ZOOM (cycle through 2h. 1h, 30m, and 15m for current view)
--PRINT
--BACK (to Trend menu)
-EXIT (to Main menu)
SETUP (Setup Monitor Menu)
-VIEW (Setup View Menu)
--PLETH
--BLIP
--BACK (back to Setup menu)
--EXIT (to Main menu)
-SENSOR (Setup Sensor Menu)
(display events #1-N, in inverse chronological
(If continuous-Loop data is in the sensor, the following
6
Introduction
--DATA (On-screen options for SENSOR-R (Write-once Sensor) sensor
are: “SPO2, SPO2+BPM, DEFAULT.” On-screen options for
SENSOR-RW (rewritable sensor) are: “SPO2, SPO2+BPM,
DEFAUL T.” SELECT toggles SENSOR-R or SENSOR-RW
sensor type; up/down keys scroll through options in order.)
The SENSOR-R feature supports all of the current O
XIMAX
sensors.
---SELECT
---BACK (back to Setup Sensor menu)
---EXIT (to Main menu)
--MSG (Sensor Set Message Menu)
---BACK (back to Setup Sensor menu)
---EXIT (to Main menu)
-NEXT (Clock/Language Menu)
--CLOCK (Clock Menu)
---SET (Clock Set Menu)
----SELECT (press select to toggle through hours, minutes,
seconds, month, day, year; use up/down buttons to set each
selection)
----BACK (back to Clock/Language menu)
----EXIT (to Main menu)
--LANG (Language Setup Menu)
(use up/down buttons to toggle though
languages)
---BACK (back to Clock/Language menu)
--NEXT (Communication/Nurse Call Menu)
---COMM (Communication Port Configuration Menu)
----SELECT
----BACK (back to Communication/Language menu)
----EXIT (to Main menu)
---NCALL (Nurse Call Menu)
----NORM +
----NORM -
----BACK (back to Communication/Nurse Call menu)
----EXIT (to Main menu)
---NEXT (Analog/Mode Menu)
----ANALOG (Analog Voltage Select Menu)
-----0 VOLT
-----1 VOLT
-----STEP
-----BACK (back to Analog/Mode menu)
----MODE (Mode Menu)
-----BACK (back to Analog/Mode menu)
-----EXIT (to Main menu)
----BACK (back to Communication/Nurse Call menu)
----EXIT (to Main menu)
---BACK (back to Clock/Language menu)
--BACK (back to Setup menu)
-EXIT (to Main menu)
LIGHT (Turns the display backlight on or off)
N-5957
Introduction
Related Documents
To perform test and troubleshooting procedures and to understand the principles
of operation and circuit analysis sections of this manual, you must know how to
operate the monitor. Refer to the N-595 operator's manual. To understand the
various Nellcor approved O
individual O
The latest version of the operator’s manual and the service manual are posted on
the Internet at:
Spare Parts and Accessories are posted on the Internet at:
WARNING: Do not spray, pour, or spill any liquid on the N-595, its
accessories, connectors, switches, or openings in the chassis.
For surface-cleaning and disinfecting follow your institution's procedures or:
•The N-595 may be surface-cleaned by using a soft cloth dampened with
either a commercial, nonabrasive cleaner or a solution of 70% alcohol in
water, and lightly wiping the surfaces of the monitor.
•The N-595 may be disinfected using a soft cloth saturated with a 10%
solution of chlorine bleach in tap water.
Before attempting to clean an SpO
enclosed with the O
specific to that sensor.
Periodic Safety Checks
The N-595 requires no calibration.
The battery should be replaced at least every 2 years. See Battery Replacement on
page 67.
The following checks should be performed at least every 24 months by a qualified
service technician.
1. Inspect the equipment for mechanical and functional damage.
2. Inspect safety labels for legibility . If the labels are damaged, contact Nellcor’s
Technical Services Department, 1.800.635.5267, or your local Nellcor
representative.
2OXI-MAXsensor, read the directions for use
XI-MAX sensor. Each sensor model has cleaning instructions
Functional Checks
If the monitor has been visibly damaged or subjected to mechanical shock (for
example, if dropped), immediately perform the performance tests. SeePerformance Tests on page 11.
N-5959
Routine Maintenance
Battery
The following checks should be performed at least every 2 years by a qualified
service technician.
1. Perform the electrical safety tests detailed in Safety Tests on page 33. If the
unit fails these electrical safety tests, refer to Troubleshooting on page 51.
2. Inspect the fuses for proper value and rating (F1 & F2 = 0.5 amp, 250 volts).
Nellcor recommends replacing the instrument's battery every 2 years. When the
N-595 is going to be stored for 3 months or more, remove the battery prior to
storage. To replace or remove the battery, refer to Disassembly Guide on page 63.
If the N-595 has been stored for more than 30 days, charge the battery as
described in Battery Charge on page 12. A fully discharged battery requires 14
hours with the monitor turned off, or 18 hours if it is in use, to receive a full
charge. The battery is being charged whenever the instrument is plugged into AC.
Note:If power stored in the battery is too low, the unit will not operate even
when plugged into AC. If this occurs, leave the unit plugged in to allow
the battery to charge as described inBattery Charge on page 12. After
approximately 10 minutes, the battery should have enough charge to
allow the unit to operate on AC.
10
Performance Verification
Introduction
This section discusses the tests used to verify performance following repairs or
during routine maintenance. All tests can be performed without removing the
N-595 cover. All tests except the battery charge and battery performance tests
must be performed as the last operation before the monitor is returned to the user.
If the N-595 fails to perform as specified in any test, repairs must be made to
correct the problem before the monitor is returned to the user.
Equipment Needed
Table 1: Equipment Needed
EquipmentDescription
Digital Multimeter (DMM)Fluke Model 87 or equivalent
Durasensor
sensor
O
Safety AnalyzerMust meet current AAMI ESI/1993 & IEC
Pulse oximetry cableDOC-10
Data interface cableEIA-232 cable (optional)
StopwatchManual or electronic
Nellcor model SRC-MAX TesterProvides testing for DigiCal compatible
Performance Tests
The battery charge procedure should be performed before monitor repairs
whenever possible.
XIMAX
®
OXI-MAXoxygen
oxygen sensor
DS-100A
MAX-A
60601-1/1998 specifications
Monitors
Note:This section is written using Nellcor factory-set defaults. If your
institution has pre configured custom defaults, those values will be
displayed. Factory defaults can be restored (see Reset Softkey on
page 40).
N-59511
Performance Verification
Battery Charge
Perform the following procedure to fully charge the battery.
1. Connect the monitor to an AC power source.
2. Verify that the monitor is off and that the AC Power/Battery Charging
3. Charge the battery for at least 14 hours with the monitor turned of f or 18 hours
Power-Up Performance
The power-up performance tests verify the following monitor functions:
indicator is lit.
with the monitor turned on.
Power-On Self-Test
•Power-On Self-Test on page 12
•Power-On Defaults and Alarm Range Limits on page 13
1. Connect the monitor to an AC power source.
2. Verify that the monitor is off and that the AC Power/Battery Charging
indicator is lit.
3. Do not connect any cables to the monitor.
4. Turn on the N-595 by pressing the ON/STANDBY button. Observe the
monitor front panel. The monitor must perform the following:
•Within 2 seconds all LEDs are illuminated, all pixels on the LCD display
are illuminated, and the backlight comes on.
12
•The indicators remain lighted.
•The LCD display shows NELLCOR and the software version of the
N-595.
Performance Verification
Note:The software “Version” displayed in the example below is X.X.X.X.The
actual software version will be displayed on your monitor.
•A 1-second beep sounds, indicating proper operation of the speaker, and
all indicators turn off except the AC Power/Battery Charging indicator
and the LCD screen.
•The N-595 begins normal operation.
PLETH VIEW:
BLIP (MAGNIFIED) VIEW
Power-On Defaults and Alarm Range Limits
Note:When observing or changing alarm limits, a time-out is in effect
(approximately 10 seconds). If no action is taken within the time-out, the
monitor automatically returns to the monitoring display.
Note:The descriptions that follow are based on the assumption that Pleth view
is the view that has been selected.
N-59513
Performance Verification
The steps for changing an alarm limit are the same if the view being used is Blip
(Magnified) view .
Note:Power-on defaults will be the factory-set defaults or the defaults set by
your institution.
1. Turn on the N-595 by pressing the ON/STANDBY button.
LIMITS
2. Press the LIMITS softkey. Verify that the monitor emits a single beep and the
pleth view is replaced with a display of the alarm limits. The upper alarm limit
for %SpO
2 will indicate an alarm limit of “100” (or institutional default
setting) inside a box.
3. Press and hold the ADJUST DOWN button. Verify that the boxed number for
2 upper alarm limit reduces to a minimum of “86.”
%SpO
Note:A decimal point in the display indicates that the alarm limits have been
changed from factory default values.
14
SELECT
4. Press the SELECT softkey . Verify that the monitor emits a single beep and the
box moves to the %SpO
2 lower alarm limit of “85” (or your institutional
default setting).
5. Press and hold the ADJUST DOWN button and verify that the %SpO
2 lower
alarm limit display reduces to a minimum of “20.”
Performance Verification
6. Press and hold the ADJUST UP button and verify that the %SpO2 lower alarm
limit display cannot be raised past the upper alarm limit setting of “85.”
EXIT
LIMITS
SELECT
7. Press the EXIT softkey.
8. Press the LIMITS softkey
9. Press the SELECT softkey three times. Verify that the monitor emits a beep
after each keystroke. The Pulse upper alarm limit should be “170” and should
be boxed.
10. Press and hold the ADJUST DOWN button.
EXIT
LIMITS
11. Verify that the minimum displayed value is “41” for the BPM upper alarm
limit.
12. Press the EXIT softkey.
13. Press the LIMITS softkey.
N-59515
Performance Verification
SELECT
14. Press the SELECT softkey four times. Verify that the pulse rate lower alarm
limit display indicates an alarm limit of “40” and is boxed.
15. Press and hold the ADJUST DOWN button. Verify that the boxed pulse rate
lower alarm limit display reduces to a minimum of “30.”
16. Press and hold the ADJUST UP button and verify that the boxed pulse rate
lower alarm limit display cannot be adjusted above the pulse rate upper alarm
limit of “40.”
EXIT
LIMITS
SELECT
17. Press the EXIT softkey.
18. Press the LIMITS softkey.
19. Press the SELECT softkey two times. Verify that SatSeconds SAT-S alarm is
selected.
20. Press the ADJUST UP button repeatedly and verify that the SatSeconds alarm
display cycles from OFF through 10, 25, 50, 100, OFF.
16
21. Press the ON/STANDBY button to turn the monitor off.
22. Press the ON/STANDBY button to turn the monitor back on.
Performance Verification
LIMITS
23. Press the LIMITS softkey. Verify that the %SpO2 upper alarm limit display is
boxed and indicates an alarm limit of “100.”
24. Verify that the %SpO
2 lower alarm limit display is boxed and indicates an
alarm limit of “85.”
25. Verify that the SatSeconds SAT-S alarm is set to OFF.
26. Verify that the pulse rate upper alarm limit display is boxed and indicates an
alarm limit of “170.”
27. Verify that the pulse rate lower alarm limit display is boxed and indicates an
alarm limit of “40.”
28. Press the ON/STANDBY button to turn the monitor off.
Operational Setup
Operational setup procedures verify and set up the following parameters.
•Alarms and Alarm Silence on page 18
•Alarm Volume Control on page 20
•Pulse Tone Volume Control on page 20
•Nurse Call on page 21
•Analog Output on page 22
•Operation on Battery Power on page 23
N-59517
Performance Verification
Alarms and Alarm Silence
1. Connect the DOC-10 monitor cable to the monitor sensor port.
1. Sensor Port
SETUP
VIEW
PLETH
LIMITS
SELECT
2. Connect the DS-100 O
XI-MAX sensor to the DOC-10 cable and your finger.
3. Press the ON/STANDBY button to turn the monitor on.
4. Press the SETUP softkey.
5. Press the VIEW softkey.
6. Press the PLETH softkey. Verify that the %SpO
2 and pulse rate.
SpO
2 and BPM indicate your
7. Press the LIMITS softkey.
8. Press the SELECT softkey to select SpO
9. Press the ADJUST UP button until the SpO
2 lower alarm limit.
2 lower alarm limit indicates 99.
18
SELECT
10. Press the SELECT softkey three times to select pulse rate lower alarm limit.
11. Press the ADJUST UP button until the pulse rate lower alarm limit indicates
160.
12. Verify the following monitor reactions:
Performance Verification
•The plethysmograph waveform tracks your pulse rate.
•The pulse tone is heard.
•Your SpO
2 and pulse rate are flashing in the %SpO2 and BPM displays.
•The audible alarm sounds, indicating that both parameters have violated
the alarm limits.
13. Press and hold the ALARM SILENCE button until the BPM display indicates
“SEC.” Continue to press the ALARM SILENCE button and press the
ADJUST DOWN button until “60” is displayed in the %SpO
2 display.
14. Press the ALARM SILENCE button.
15. With the monitor’s alarm silenced, verify the following:
•The alarm remains silenced for 60 seconds.
•The ALARM SILENCE indicator lights.
•The %SpO
2and BPM displays continue to flash.
•The pulse tone is still audible.
•The audible alarm returns in approximately 60 seconds.
16. Press and hold the ALARM SILENCE button until the BPM display indicates
“SEC.” Continue to press the ALARM SILENCE button and press the
ADJUST DOWN button until “30” is displayed in the %SpO
2 display.
17. Press the ADJUST UP button and verify that the displays indicate 60 SEC,
90 SEC, 120 SEC, and OFF. Release the ADJUST UP button when the
display indicates “OFF.”
18. Press and release the ALARM SILENCE button. Verify that the monitor’s
ALARM SILENCE indicator flashes.
19. Wait approximately 3 minutes. Verify that the monitor’s alarm does not
return. After 3 minutes, the monitor’s alarm silence reminder beeps three
times, and will continue to do so at approximately 3-minute intervals.
N-59519
Performance Verification
Alarm Volume Control
After completing Alarms and Alarm Silence on page 18, perform the following
procedure.
1. Press and hold the ALARM SILENCE button and verify the following:
2. While still pressing the ALARM SILENCE button, press the ADJUST
•“OFF” is displayed for approximately 3 seconds.
•After 3 seconds, a steady tone is heard at the default alarm volume setting,
the %SpO
2 display indicates “VOL,” and the BPM display indicates the
default setting of 7.
DOWN button until an alarm volume setting of 1 is displayed. Verify that the
volume of the alarm has decreased but is still audible.
3. Continue pressing the ALARM SILENCE button and press the ADJUST UP
button to increase the alarm volume setting to a maximum value of 10. Verify
that the volume increases.
4. Continue pressing the ALARM SILENCE button and press the ADJUST
DOWN button until a comfortable audio level is attained.
5. Release the ALARM SILENCE button. The tone will stop.
Pulse Tone Volume Control
After completing Alarm Volume Control on page 20, perform the following
procedure.
1. Press the ADJUST UP button and verify that sound level of the beeping pulse
tone volume increases.
20
2. Press the ADJUST DOWN button and verify that the sound level of the
beeping pulse tone volume decreases until it is no longer audible.
3. Press the ADJUST UP button to return the beep volume to a comfortable
level.
Nurse Call
Performance Verification
4. Remove the OXI-MAXsensor from your finger. Disconnect the DOC-10
monitor cable and the O
XI-MAX sensor.
1. Data Port Connector
1. Connect the negative lead of a voltmeter to pin 5 and positive lead to pin 11 of
the data port connector (1) on the back of the monitor. Ensure that the audible
alarm is not silenced or turned off.
2. Connect the SRC-MAX tester to the DOC-10 sensor cable.
3. Connect the DOC-10 sensor cable to the monitor SpO
2 connector.
4. Turn on the monitor and wait for the monitor to complete POST.
Note:The monitor should indicate a %SpO
2 alarm of 75.
5. Verify an output voltage at pin s 5 and 11 between +5 to +12 VDC.
6. Press the ALARM SILENCE button. With no active audible alarm, the output
voltage at pins 5 and 11 must be between -5 to -12 VDC. This verifies the
RS-232 Nurse Call function.
7. With the instrument in an alarm condition, use a digital voltmeter (DVM) to
verify that there is no continuity (1 megohms or greater) between pins 8 and
15 and that there is continuity (60 ohms or less) between pins 7 and 15.
8. Press the SRC-MAX tester %SpO
N-59521
2 button to change the %SpO2 to 90.
Performance Verification
G
Analog Output
9. Use a DVM to verify that there is continuity between pins 8 and 15 and that
there is no continuity between pins 7 and 15. This verifies the solid state
Nurse Call function.
1 Data Port Connector
SETUP
NEXT
NEXT
NEXT
ANALO
1 VOLT
1. Connect the negative lead of a voltmeter to pin 10 and the positive to lead pin
6 of the data port connector (1) on the back of the monitor.
2. Press the SETUP softkey.
3. Press the NEXT softkey.
4. Press the NEXT softkey.
5. Press the NEXT softkey.
6. Press the ANALOG softkey.
7. Press the 1 VOLT softke y.
8. Verify that the monitor’s output voltage is +1.0 ± 0.025 VDC. This verifies
the analog SpO
2 function.
9. Leave the negative lead connected to pin 10 and verify 1.0 ± 0.025 VDC on
pins 13 and 14. This verifies the monitor’s BPM and Pleth function.
22
Loading...
+ 132 hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.