Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor,
OxiCliq, Dura-Y, MAX-FAST, and O
XIMAX are trademarks of Nellcor Puritan Bennett Inc.
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est conforme à la norme NMB-001 Canada.
To obtain information about a warranty, if any, contact Nellcor’s Technical Services Department,
or your local representative.
Purchase of this instrument confers no express or implied license under any Nellcor Puritan
Bennett patent to use the instrument with any sensor that is not manufactured or licensed by
Nellcor Puritan Bennett.
Covered by one or more of the following U.S. Patents and foreign equivalents: 4,621,643;
4,653,498; 4,700,708; 4,770,179; Re. 35,122; 4,802,486; 4,869,254; 4,928,692; 4,934,372; 5,078,136;
5,351,685; 5,368,026; 5,485,847; 5,533,507; 5,662,106; and 5,853,364.
Warnings are identified by the WARNING symbol shown above.
W arnings alert the user to potential serious outcomes (death, injury, or
adverse events) to the patient or user.
WARNING: The sensor extrapolates from the date and time
provided by the N-595 when recording the sensor event record to
the sensor. The accuracy of the date/time is the responsibility of
the N-595. It is recommended that the N-595 user set the time/
date to the correct value before a sensor event record-enabled
sensor is connected, and that this date/time not be changed while
the sensor remains connected. Since a sensor with sensor event
record data can be transported from one monitor to another,
having discrepancies in the date/time between monitors and the
sensor event record data will affect the order the sensor event
record data appears. To eliminate this possible problem, all
monitors within an institution should be set to the same time.
WARNING: Explosion hazard. Do not use the N-595 pulse
oximeter in the presence of flammable anesthetics or gases.
WARNING: Chemicals from a broken LCD display panel are
toxic when ingested. Use caution when handling a pulse oximeter
with a broken display panel.
N-5951
Safety Information
WARNING: Pulse oximetry readings and pulse signals can be
affected by certain environmental conditions, O
XIMAX sensor
application errors, and certain patient conditions. See the
appropriate sections of this manual for specific safety
information.
WARNING: The use of accessories, sensor s, and cables other
than those specified may result in increased emission and/or
decreased immunity and inaccurate readings of the N-595 pulse
oximeter.
WARNING: Failure to cover the OXIMAX sensor site with opaque
material in high ambient light conditions may result in inaccurate
measurements.
Cautions
2
Cautions are identified by the CAUTION symbol shown above.
Cautions alert the user to exercise care necessary for the safe and
effective use of the N-595 pulse oximeter.
Safety Information
Caution: When connecting the N-595 to any instrument, verify
proper operation before clinical use. Both the N-595 and the
instrument connected to it must be connected to a grounded
outlet. Accessory equipment connected to the pulse oximeter’s
data interface must be certified according to IEC Standard 950
for data-processing equipment or IEC Standard 60601-1 for
electromedical equipment. All combinations of equipment must
be in compliance with IEC Standard 60601-1-1 systems
requirements. Anyone who connects additional equipment to the
signal input port or signal output port (N-595 data port
connector) configures a medical system and is therefore
responsible for ensuring that the system complies with the
requirements of system standard IEC Standard 60601-1-1 and
the electromagnetic compatibility system standard IEC S tandard
60601-1-2. The N-595 accuracy may degrade if it is connected to
secondary I/O devices when the instrument is not connected to
earth reference.
Caution: Federal law (U.S.A.) restricts this device to sale by or on
the order of a physician.
Caution: Dispose of battery in accordance with local
requirements and regulations.
N-5953
Introduction
WARNING: The N-595 is intended only as an adjunct in patient
assessment. It must be used in conjunction with clinical signs and
symptoms.
Intended Use for the N-595
The N-595 pulse oximeter is indicated for the continuous
non-invasive monitoring of functional oxygen saturation of arterial
hemoglobin (SpO
with neonatal, pediatric, and adult patients during both no motion and
motion conditions and for patients who are well or poorly perfused, in
hospitals, hospital-type facilities, intra-hospital transport, and home
environments. For prescription use only.
Note: Hospital use typically covers such areas as general care floors,
operating rooms, special procedure areas, intensive and critical
care areas, within the hospital plus hospital-type facilities.
Hospital-type facilities include physician office based
facilities, sleep labs, skilled nursing facilities, surgicenters, and
sub-acute centers.
2) and pulse rate. The N-595 is intended for use
Intra-hospital transport includes transport of a patient within
the hospital or hospital-type facility .
Home Care use is defined as managed/used by a lay person
(parent or other similar non-critical caregiver) in the home
environment.
Use with any particular patient requires the selection of an
appropriate oxygen O
Operator's Manual.
N-5955
XIMAX sensors as described in this
Introduction
Motion performance claims are applicable to models MAX-A,
MAX-AL, MAX-P, MAX-N, and MAX-I Nellcor O
oximetry sensors.
How to Use this Manual
All users should read this manual thoroughly . More experienced users
of the N-595 will be able to go to the topics for the information they
require.
The current copy of this manual is available on the internet at:
2 Display, page 1218. Data In Sensor Indicator, page 13
12. ADJUST DOWN Button, page 10
14. CONTRAST Button, page 10
17. Menu Bar, page 10
Figure 1: Front Panel Buttons and Symbols
N-5957
Description of Controls, Indicators, and Symbols
Identification of Rear Panel Components
1. Equipotential Terminal (Ground) 4. Fuse Holder
2. AC Power Connector, page 175. Supply Voltage Selector Switch, page 17
3. Data Port Connector, page 91
Figure 2: Rear Panel Components
N-595 Symbols
8
The symbols that are located on the rear panel of the N-595 are as
follows:
See Instructions for Use
Fuse Replacement
Equipotential Terminal (ground)
Date of Manufacture
Data Interface
There is one symbol located on the front panel of the N-595.
Type BF Applied Part - Not defibrillator proof.
Description of Controls
Description of Controls, Indicators, and Symbols
Note: A button press, except the ON/STANDBY button, should
result in either a valid or an invalid key tone (refer to Table 1
on page 14). If the key pressed fails to emit a tone, contact
qualified service personnel.
The ON/ST ANDBY button. Used to turn the N-595 monitor on or off.
The ALARM SILENCE button. Used to silence current alarms for the
alarm silence duration period. When an alarm has been silenced,
pressing the button again reactivates, or “unsilences” the alarm. It is
also used to view and adjust alarm silence duration and alarm volume.
N-5959
Description of Controls, Indicators, and Symbols
The ALARM SILENCE button clears “SENSOR OFF,” “LOW
BATTERY,” and “SENSOR DISCONNECT” messages from the
display.
The ADJUST UP button. Used to increase variable parameters of the
monitor.
The ADJUST DOWN button. Used to decrease variable parameters
in the monitor.
The CONTRAST button. Used in conjunction with the ADJUST UP
and ADJUST DOWN buttons to lighten or darken the display screen.
The softkey buttons have multiple uses depending on the legend
displayed above the button.
Description of Displays and Indicators
The type of display is user selectable. Refer to Selecting the Pleth
View on page 34.
10
The pleth display includes a “wiper bar” plethysmographic
waveform, menu bar, and current measured %SpO
2 and pulse rate. If
SatSeconds are enabled, the pleth display includes the SatSeconds
indicator and SatSeconds setting. A decimal point after the %SpO
2 or
Description of Controls, Indicators, and Symbols
pulse rate indicate that the respective limits have been changed from
the power on defaults (Monitor Trend Data on page 53).
The blip display includes a pulse amplitude blip bar, current measured
%SpO
2 and pulse rate, and current upper and lower %SpO2 and pulse
rate limits. If SatSeconds are enabled, the blip display includes the
SatSeconds indicator and SatSeconds setting. Decimal points after the
%SpO
2 or pulse rate indicate that the respective limits have been
changed from the power-on defaults.
There are various matrixes within the O
XIMAX algorithm. Some of
these, are used to assess the severity of conditions presented to the
N-595 in measuring SpO
2 and pulse rate on a patient. These
individual matrices or combinations of these matrices are used to
drive the LED indicators on the N-595 front panel.
The OxiMax algorithm automatically extends the amount of data
required for measuring SpO2 and pulse rate depending on the
measurement conditions. During normal measurement conditions the
averaging time is 6-7 seconds. During challenging measurement
conditions which could be caused by low perfusion, motion, external
interference like ambient light, or a combination of these, the
OXIMAX algorithm automatically extends the amount of data
required beyond 7 seconds. If the resulting dynamic averaging time
exceeds 20 seconds, the pulse search indicator is lit solid and SpO2
and Pulse Rate will continue to be updated every second. As these
conditions become even more challenging, the amount of data
required continues to extend. If the dynamic averaging time reaches
40 seconds, the pulse search indicator begins flashing, the SpO2 and
pulse rate displays flash zeros indicating a loss-of-pulse condition.
N-59511
Description of Controls, Indicators, and Symbols
WARNING: Failure to cover the OXIMAX sensor site with opaque
material in high ambient light conditions may result in inaccurate
measurements.
The %SpO
2 Display. Shows the hemoglobin oxygen saturation level.
The display value flashes zeros during loss-of-pulse alarms and
flashes the SpO
2 value when the SpO2 is outside the alarm limits.
During Pulse Search, the monitor continues to update the display. If
alarm limits have been changed from their power-on defaults, a
decimal point (.) is displayed after the SpO
2 value (81.).
The Pulse Amplitude Indicator (blip bar). Indicates pulse beat and
shows the relative pulse amplitude. As the detected pulse becomes
stronger, more bars light with each pulse. This indicator is available
only in the blip view.
The Pulse Rate Display. Shows the pulse rate in beats per minute. It
flashes during loss-of-pulse alarms and when the pulse rate is outside
of the alarm limits. During Pulse Search, the monitor continues to
update the display. Pulse rates outside of the pulse rate range (20 to
250 bpm) are displayed as the closest value within the range. If alarm
limits have been changed from their power-on defaults, a decimal
point (.) is displayed after the BPM value (112.).
The AC Power Indicator. Lights continuously when the N-595 is
connected to AC power. It also indicates that the battery is charging.
It is off when the monitor is being powered by internal battery.
12
The Low Battery Indicator. Lights continuously when 15 or fewer
minutes of battery capacity remain. Flashes when the battery capacity
reaches critical condition.
The Alarm Silence Indicator. Lights continuously when an audible
alarm has been silenced. It flashes when the alarm silence duration
has been set to Off.
The Motion Indicator. The motion indicator is lit whenever the
O
XIMAX algorithm detects the presence of artifacts
its severity or the impact on the SpO
2 or pulse rate values. When the
1
independent of
Description of Controls, Indicators, and Symbols
motion indicator and the pulse search indicator are simultaneously lit,
it is an indication that the artifact is significant and/or has been
persistent.
The Pulse Search Indicator. Lights continuo usly prior to initial
acquisition of a pulse signal and during prolonged and challenging
monitoring conditions. The pulse search indicator flashes during a
loss-of-pulse signal.
The Data In-Sensor Indicator. Lights to indicate that the attached
O
XIMAX sensor contains a patient sensor event record. The sensor
event record information may be viewed or printed.
The SatSeconds Indicator. Fills in clockwise as the SatSeconds alarm
management system detects a %SpO
setting. Empties in counterclockwise direction when %SpO
2 reading outside of the limit
2 reading
is within limits. When the indicator is full, a medium priority alarm
will sound.
The Fast Response Mode Indicator. The response mode setting
dictates the response time (2 to 4 seconds in fast mode and 4 to 7
seconds in normal mode) applied by the O
calculation of SpO
2. The OXIMAX algorithm’s calculation of pulse rate
XIMAX algorithm in its
is unaffected by the response mode setting. The trending interval
(2-seconds or 4-seconds) is updated automatically by the monitor to
roughly correspond with the SpO
2 calculation response time.
The Neonate Alarm Limits Indicator. This symbol is displayed when
the alarm limits are set to neonate. No symbol is displayed when the
monitor is set to adult limits.
1
Artifacts are events contained in the in-sensor data.
N-59513
Description of Controls, Indicators, and Symbols
Description of Audible Indicators
Table 1 identifies the audible indicators of the N-595 indicators.
Table 1: Audible Indicators
FunctionDescription
Alarm Silence
Reminder
Confirmation ToneThree beeps sound to indicate that default
Invalid Button Press
Valid Button Press
High Priority Alarm
Medium Priority
Alarm
Low Priority Alarm
Three beeps that sound approximately every 3
minutes when alarms are silenced with the
alarm silence duration set to Off and the alarm
silence reminder function is enabled.
settings have been saved or reset to factory
defaults or trend data has been deleted.
A short, low
button has been pressed that is not appropriate
for the current state of the monitor.
A short, medium
an appropriate button has been pressed.
A high
loss
-of-pulse.
A medium
SpO
2 or pulse rate limit violation.
A low
-pitched tone indicating that a
-pitched tone indicating that
-pitched, fast-pulsing tone indicating
-pitched, pulsing tone indicating an
-pitched, slow-pulsing tone indicating an
OXIMAX sensor disconnect, low battery, or
monitor failure.
Power
-On Self-Test
Pass
A 1-second tone indicating that the N-595 has
been turned on and has successfully completed
the power
-on self-test.
14
Pulse BeepA single beep sounds for each detected pulse.
Volume Setting ToneA continuous tone that is used when adjusting
The pitch of the pulse beep signal changes with
-by-point rise or fall in the saturation
a point
level.
the alarm volume.
Setting up the Monitor
WARNING: To ensure patient safety, do not place the pulse
oximeter in any position that might cause it to fall on the patient.
WARNING: As with all medical equipment, carefully route
patient cabling to reduce the possibility of patient entanglement
or strangulation.
WARNING: Ensure that the speaker is clear of any obstruction.
Failure to do so could result in an inaudible alarm tone.
WARNING: Disconnect the N-595 and Nellcor OXIMAX se nsor
from the patient during magnetic resonance imaging (MRI)
scanning. Induced current could potentially cause burns.
WARNING: To ensure accurate performance and prevent device
failure, do not subject the N-595 to extreme moisture, such as
direct exposure to rain. Such exposure may cause inaccurate
performance or device failure.
WARNING: Do not use an N-595 pulse oximeter, OXIMAX sensor,
cables, or connectors that appear damaged.
WARNING: Do not lift the pulse oximeter by the pulse oxime try
cable or power cord because the cable or cord could disconnect
from the pulse oximeter, causing the pulse oximeter to drop on
the patient.
N-59515
Setting up the Monitor
WARNING: The N-595 is not defibrillator-proof. However, it
may remain attached to the patient during defibrillation or while
an electrosurgical unit is in use, but the readings may be
inaccurate during the defibrillation and shortly thereafter.
WARNING: In the USA, do not connect the pulse oximeter to an
electrical outlet controlled by a wall switch, because the pulse
oximeter may be accidentally turned off.
WARNING: Use only the Nellcor pulse oximetry cable DOC-10
with the N-595 pulse oximeter. Use of another pulse oximetry
cable will have an adverse effect on performance. Do not attach
any cable that is intended for computer use to the O
XIMAX sensor
port. Do not connect any device other than a Nellcor -approved
OXIMAX sensor to the OXIMAX sensor connector.
16
WARNING: The N-595 should not be used adjacent to or stacked
with other equipment. If adjacent or stacked use is necessary, the
N-595 should be observed to verify normal operation in the
configuration it is to be used.
List of Components
Setting up the Monitor
QuantityItem
1
1Nellcor O
1
1
1Power Cord (applicable to country of sale)
2
1Sensor Accuracy Grid
1Quick Guide
N
DOC
N
country of sale) and/or Compact Disk
Fuses, 0.5 A, 250 volts, slow
20 mm)
Connecting the N-595 to AC Power
WARNING: In the USA, do not connect the pulse oximeter to an
electrical outlet controlled by a wall switch, because the pulse
oximeter may be accidentally turned off.
-595 Pulse Oximeter
XIMAX Sensor or Assortment Pack
-10 Pulse Oximetry Cable
-595 Operator’s Manual (applicable to
-blow, IEC (5 x
Caution: The SUPPLY VOLTAGE SELECTOR switch must be
set to the correct voltage (1 15 or 230) to avoid equipment dama ge
and ensure battery charging.
N-59517
Setting up the Monitor
Caution: Use only the hospital-grade power cord provided by
Nellcor.
1. Power Connector2. Supply Voltage Selector
1. Set the SUPPLY VOLTAGE SELECTOR (2) switch to the
applicable voltage.
2. Plug the female connector end of the power cord into the N-595
POWER CONNECTOR (1) on the rear of the monitor.
3. Plug the male connector of the power cord into a properly
grounded AC outlet.
4. Verify that the monitor’s AC POWER INDICATOR is lit.
Note: If the AC POWER INDICATOR is not lit, check:
•the power cord
•the SUPPLY VOLTAGE SELECTOR switch
•the user-accessible fuses
•the AC power outlet
18
Connecting an OXIMAX Sensor to the N-595
The OXIMAX sensor type is displayed at the bottom of the display
when an O
completes POST with an O
XIMAX sensor is connected to the N-595 or when the N-595
XIMAX sensor attached.
Setting up the Monitor
Caution: Use only Nellcor-approved O
XIMAX sensors and pulse
oximetry cables.
Note: Physiological conditions, medical procedures, or external
agents that may interfere with the monitor’s ability to detect
and display measurements include dysfunctional hemoglobin,
arterial dyes, low perfusion, dark pigment, and externally
applied coloring agents, such as nail polish, dye, or pigmented
cream.
1. SpO2 O
XIMAX Sensor Port
1. Connect a DOC-10 pulse oximetry cable to the SpO
2OXIMAX
sensor port (1) of the monitor.
2. Connect a Nellcor O
XIMAX SpO2 sensor to the other end of the
DOC-10 pulse oximetry cable.
N-59519
Battery Operation
WARNING: Dispose of battery in accordance with local
requirements and regulations.
Operating the N-595 on Battery Power
The N-595 monitor has an internal battery that can be used to power
the monitor during transport or when AC power is not available. A
new, fully charged battery will provide at least 2 hours of monitoring
time under the following conditions:
•No audible alarms sound
•No analog or serial output devices are attached to the
N-595
The monitor cannot operate with a fully discharged battery. Before
attempting to turn on an N-595 monitor whose battery charge has
been completely depleted, first plug the monitor into an AC outlet to
allow the battery to charge for a few minutes. The monitor may then
be powered on.
To charge a low or dead battery, connect the monitor to AC power. A
full charge of a dead battery takes 14 hours while the monitor is
turned off. A full charge of a dead battery takes 18 hours while the
monitor is in operation (monitoring a patient).
When all of the following conditions are present for 15 minutes, the
N-595 will automatically shut down:
•Monitor is running on battery power
•No buttons have been pressed
N-59521
Battery Operation
Note: Whenever the monitor is connected to AC power, the battery is
Low Battery Indicator
The Low Battery Indicator lights and a low priority alarm begins to
sound when approximately 15 minutes of monitoring time is
available on the existing battery charge. Refer to Table 2 for a
description of the low and critical battery conditions.
•No pulse has been detected (for example, when a patient
is not connected to the O
XIMAX sensor or the OXIMAX
sensor is disconnected from the monitor)
•No alarms are present (other than low battery or a
non-correctable error)
being charged. Therefore, it is recommended that the monitor
remain connected to AC power when not in use. This will
ensure a fully charged battery whenever it is needed.
22
If the monitor is not on AC power, a low battery audible alarm can be
canceled by pressing the ALARM SILENCE button. The low battery
indicator and display screen message will continue to be displayed.
Plugging the monitor into AC power will silence the audible alarm,
but the low battery indicator will stay lit as long as the battery is in the
low voltage condition. After the 15-minute period of low battery
condition, a high priority alarm will sound for about 10 seconds
before the monitor shuts off.
If the monitor backlight is turned off during a low battery condition,
the backlight cannot be turned back on.
It is recommended that qualified service personnel replace the
internal battery every 24 months. Replaced batteries should be
disposed of in accordance with local ordinances.
Battery Operation
Caution: If the N-595 pulse oximeter is to be stored for a period of
3 months or longer , notify service personnel to r emove the battery
from the pulse oximeter prior to storage. Recharge the battery
when the battery has not been charged for 2 or more months.
Caution: The pulse oximeter default settings will return to
factory default setting if the battery becomes fully discharged or
is replaced. Qualified service personnel will have to reset the
institutional defaults, following the instructions in the service
manual.
Note: If the AC voltage selector switch on the monitor rear panel
does not match your AC voltage source, the monitor may run
on battery power, even though it is plugged into AC power,
which will eventually result in a low priority alarm and a
lighted low battery indicator. Ensure that the switch setting
matches your AC voltage.
Note: As the battery is used and recharged over a period of time, the
amount of time between the onset of the low battery alarm and
the instrument shut-off may become shorter.
Table 2: Low Battery and Critical Battery
State
1No No Yes
N-59523
Critical
Battery
Low
Battery
ACOperation
2- normal
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY
message
Audible alarm
Error code
Effect of ALARM
SILENCE key
Shutdown
-off
-none
-N/A
-on
-off
-off
-normal
Battery Operation
Ta ble 2: Low Battery and Critical Battery
State
2No No No
3No Yes No
Critical
Battery
Low
Battery
ACOperation
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY
message
Audible alarm
Error code
Effect of ALARM
SILENCE key
Shutdown
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY
message
2-normal
-off
-off
-off
-off
-none
-normal
- N/A
2-normal
-off
-on
-on
24
Audible alarm
Error code
Effect of ALARM
SILENCE key
silences audio alarm, second
press cancels LOW
BATTERY message (LED)
stays on until Low Battery
Condition is corrected.
Shutdown
-low priority
-logged
-First press
-N/A
Table 2: Low Battery and Critical Battery
Battery Operation
State
4No Yes Yes
5Not used
Critical
Battery
Low
Battery
ACOperation
SpO
AC/Battery charge LED
LOW BATTERY LED
LOW BATTERY
message
Audible alarm
Error code
Effect of ALARM
SILENCE key
stays on)
Shutdown
Note: Connecting AC
2-normal
-on
-on
-off
-off
-logged
-N/A (LED
-N/A
functions the same as
ALARM SILENCE
key in state 3.
6Yes Yes No
N-59525
SpO
2-not displayed
AC/Battery charge LED
LOW BATTERY LED
(flashing)
LOW BATTERY
message
Audible alarm
Error code
logged
Effect of ALARM
SILENCE key
Shutdown
-on
-high priority
-displayed and
-none
-after 10 seconds
-off
-on
Battery Operation
Ta ble 2: Low Battery and Critical Battery
State
7YesYesYes
Critical
Battery
Low
Battery
ACOperation
2-not displayed
SpO
AC/Battery charge LED
LOW BATTERY LED
(flashing)
LOW BATTERY
message
Audible alarm
Error code
logged
Effect of ALARM
SILENCE key
Shutdown
-on
-high priority
-displayed and
-N/A
-after 10 seconds
-on
-on
26
Using the Monitor
Introduction
The parameters of the N-595 monitor are preset to factory default
settings. See Factory Defaults on page 139. The factory default
parameters may be changed to institutional default parameters by
following the procedures in the N-595 service manual.
Table 3 lists the parameters, ranges available, and the factory default
setting. The parameters may be set on an individual basis, by the
clinician, and these settings will remain in effect until the N-595 is
turned off.
EnglishEnglish
German, Dutch,
Portuguese,
Spanish, Italian,
Swedish
0 to 1044
Normal or FastNormalNormal
Table 3: Parameter Ranges
Using the Monitor
Parameter
RS-232 Level
Nurse Call
Polarity
SatSecondsOff, 10, 25, 50,
Sensor Event
Date Format
(SENSOR-R
and
SENSOR-RW
Sensor
Messages
Enabled
Trend DisplayDual, %SpO
Trend Scale48, 36, 12,8, 4, 2,
Ranges/
Selections
Normally High,
Normally Low
100
SpO
2,
SpO
2+Pulse Rate,
Default (default is
factory default)
Yes, NoYesYes
2,
Pulse, Histogram,
Amplitude
1 hours, 30, 15
minutes, 40, 20
seconds
Factory Adult
Defaults
Normally lowNormally low
OffOff
DefaultDefault
%SpO2%SpO2
2 hours2 hours
Factory
Neonate
Defaults
Turning On the Monitor
Before using the N-595 in a clinical setting, you must verify that the
monitor is working properly and is safe to use. Proper working
condition will be verified each time the N-595 is turned on as
described in the following procedure.
Caution: If any indicator or display element does not light when
the pulse oximeter is turned on, do not use the pulse oximeter.
Instead, contact qualified service personnel, your local Nellcor
representative, or Nellcor’s Technical Services Department.
N-59529
Using the Monitor
Note: Physiological conditions, medical procedures, or external
agents that may interfere with the monitor’s ability to detect
and display measurements, include dysfunctional hemoglobin,
arterial dyes, low perfusion, dark pigment, and externally
applied coloring agents such as nail polish, dye, or pigmented
cream.
Note: The monitor automatically starts the Power-On Self-Test
(POST), which tests the monitor circuitry and functions.
Caution: During POST (immediately after power-up), confirm
that all indicators light, all display segments turn on, and the
pulse oximeter speaker sounds a one-second tone.
1. Turn on the N-595 by pressing the ON/STANDBY button.
2. Ensure that all of the front panel indicators illuminate.
30
3. Once the display test portion of POST is complete, the N-595
software version is displayed for approximately 5 seconds.
Note: The software version shown above is only a sample. Check
your monitor for the software version installed.
Software version numbers are often needed when calling Nellcor’s
Technical Services Department or your local Nellcor representative
for technical assistance. Write down the software version number and
have it available prior to requesting technical assistance.
Using the Monitor
4. If the N-595 detects an internal problem during POST, an error
tone sounds and the monitor displays an error code (EEE) and the
corresponding number (see Troubleshooting on page 117).
5. Upon successful completion of the POST, the N-595 sounds a
one-second tone indicating that the monitor has passed the test.
WARNING: If you do not hear the POST pass tone, do not use
the pulse oximeter.
WARNING: Ensure that the speaker is clear of any obstructions.
Failure to do so could result in an inaudible alarm tone.
Note: In addition to serving as the POST pass verification, the POST
pass tone also functions as an audible confirmation that the
speaker is performing properly. If the speaker does not
function, the alarm warning sounds cannot be heard.
OXIMAX Sensor Attached
When an OXIMAX sensor is attached to the monitor, a “DAT A TYPE: .
. .” message is displayed briefly at the bottom of the monitor display.
For a sensor containing data, the message identifies the sensor data
type, For a blank sensor, the message identifies the monitor’s current
N-59531
Using the Monitor
data type setting that will be used to write data to the sensor. The data
type settings are SPO2 and SPO2+BPM.
Note: The type of data recorded is only displayed when data is resent
in the O
XIMAX sensor.
The monitor displays zeros in the %SpO
2 and Pulse Rate displays
while the N-595 is searching for a valid pulse. For optimal
performance, allow the monitor to search and lock onto a pulse for
approximately 10 seconds in non-motion conditions.
When a valid pulse is detected, the monitor enters the Monitoring
Mode and displays patient parameters.
32
Look for movement of the blip bar or of the plethysmographic
waveform indicating that the monitor is displaying real-time data.
Listen for the pulse beep tone. If the pulse beep tone does not sound
with each pulse, it is an indication that the pulse beep volume is set to
zero, the speaker is malfunctioning, or the signal is corrupted.
When an O
XIMAX sensor is attached to the monitor and is applied to a
patient, if the monitor loses the pulse signal, the monitor will display
“--- & ---” (3 dashes and 3 dashes) and remain in Pulse Search Mode
for 5 seconds before displaying the poor signal condition screen. The
poor signal condition screen is part of the N-595's Sensor Messages
feature. For more information about OXIMAX Sensor Messages, refer OXIMAX Sensor Messages on page 49.
No OXIMAX Sensor Attached
Upon successful completion of the POST, the N-595 monitor sounds
a one-second tone indicating that the monitor has passed POST.
Using the Monitor
or
The monitor displays dashes (---) and the Pulse Search indicator is
not lit, indicating that the monitor failed to detect an O
N-59533
XIMAX sensor.
Using the Monitor
Turning the Backlight On or Off
Note: When the backlight is off, any of the following conditions will
turn on the backlight:
•pressing any of the softkeys
•pressing the CONTRAST button
•pressing the ALARM SILENCE button
•any alarm
LIGHT
With the monitor in the normal monitoring mode, press the LIGHT
softkey.
Adjusting Screen Contrast
With the monitor in the normal monitoring mode:
1. Press the CONTRAST button.
2. Press the ADJUST UP or ADJUST DOWN button until the
desired contrast is obtained.
3. Press the CONTRAST button.
Selecting the Pleth View
34
The pleth view displays the pleth waveform, %SpO2, and pulse rate
data. Refer to Principles of Operation on page 143, for a description
of the pleth waveform.
With the monitor in the normal monitoring mode:
Using the Monitor
SETUP
VIEW
PLETH
1. Press the SETUP softkey.
2. Press the VIEW softkey.
3. Press the PLETH softkey.
Selecting the Blip View
Displays SpO2, pulse rate, blip bar, and limits in a larger format for
easier viewing.
With the monitor in the normal monitoring mode:
SETUP
VIEW
N-59535
1. Press the SETUP softkey.
2. Press the VIEW softkey.
Using the Monitor
BLIP
3. Press the BLIP softkey.
Setting the Pulse Beep Volume
With the monitor in the normal monitoring mode:
1. Press and hold the ADJUST UP/ADJUST DOWN button to
increase/decrease pulse beep volume.
Setting the Alarm Volume
The Alarm Volume display allows the user to adjust the volume of
alarm tones.
With the monitor in the normal monitoring mode:
36
1. Press the ALARM SILENCE button until the alarm volume level
is displayed and sounds on the monitor.
2. While continuing to press the ALARM SILENCE button, press
and hold the ADJUST UP/ADJUST DOWN button to increase/
decrease the volume.
Setting the Date and Time
Using the Monitor
WARNING: The sensor extrapolates from the date and time
provided by the N-595 when recording the sensor event record to
the sensor. The accuracy of the date/time is the responsibility of
the N-595. It is recommended that the N-595 user set the time/
date to the correct value before a sensor event record-enabled
sensor is connected, and that this date/time not be changed while
the sensor remains connected. Since a sensor with sensor event
record data can be transported from one monitor to another,
having discrepancies in the date/time between monitors and the
sensor event record data will affect the order the sensor event
record data appears. To eliminate this possible problem, all
monitors within an institution should be set to the same time.
With the monitor in the normal monitoring mode:
SETUP
N-59537
1. Press the SETUP softkey.
Using the Monitor
NEXT
CLOCK
SET
2. Press the NEXT softkey.
3. Press the CLOCK softkey.
4. Press the SET softkey.
5. Press the SELECT softkey to select:
TIME HOURS : MINUTES : SECONDS (16:46:05)
DATE DAY - MONTH - YEAR (30-JAN-02)
6. Use the ADJUST UP or ADJUST DOWN buttons to change the
selected value.
EXIT
7. Press the EXIT softkey.
Setting Alarm Silence Duration
The Alarm Silence Duration display allows the user to adjust the
alarm silence duration.
With the monitor in the normal monitoring mode:
38
Using the Monitor
1. Press the ALARM SILENCE button until the alarm silence
duration setting is displayed. Alarm silence durations that are
available are OFF, 30, 60, 90, and 120 seconds.
2. Press and hold the ALARM SILENCE button and the ADJUST
UP button to increase the alarm silence duration setting.
3. Press and hold the ALARM SILENCE button and the ADJUST
DOWN button to decrease the alarm silence duration setting.
Note: Releasing the ADJUST UP or ADJUST DOWN button sets the
alarm silence duration.
Disabling Audible Alarms
Setting the alarm silence duration to OFF means that the monitor will
produce no audible alarms.
Note: The ability to set the alarm silence duration to OFF can be
enabled or disabled by qualified service personnel as described
in the service manual. The current copy of the service manual
is available on the Internet at:
WARNING: Do not silence the audible alarm function or
decrease the audible alarm volume if patient safety could be
compromised.
With the monitor in the normal monitoring mode:
1. Press the ALARM SILENCE button until the alarm silence
duration setting is displayed.
40
2. While pressing the ALARM SILENCE button, press and hold the
ADJUST UP button until OFF is displayed. Release the buttons.
Selecting Standby Mode
The standby mode allows the monitor to retain the alarm limit
settings that are in effect while monitoring a patient. The monitor
must be powered by AC power to enter the standby mode.
Usually the standby mode is used when a patient has to leave the
monitor for a period of time and will return to the same monitor.
To place a monitor in the standby mode:
1. The monitor should be monitoring a patient.
2. The monitor alarm limits should be configured to the patient
being monitored.
Using the Monitor
3. Disconnect the sensor from the monitor.
4. Press the ALARM SILENCE button. This silences the audible
alarms.
5. Press the ALARM SILENCE button. This disables the alarm
messages.
N-59541
Using the Monitor
The monitor is now in standby. To return to normal monitoring,
connect the sensor to the monitor and the patient.
Adult-Pediatric or Neonatal Settings
The clinician can set the monitor’s operating mode to adult-pediatric
or neonatal by using the LIMITS softkey. The setting will only
remain in the monitor until the monitor is turned off. The factory
default power-on setting is for adult-pediatric patients. This default
setting can be changed to neonatal by qualified service personnel
using the procedures indicated in the service manual.
Refer to Table 11 on page 139, for neonate factory default limit
settings. Refer to T able 12 on page 140, for adult factory default limit
settings.
WARNING: Each time the pulse oximeter is used, check alarm
limits to ensure that they are appropriate for the patient being
monitored.
Setting Patient Adult-Pediatric/Neonatal Mode
With the monitor in the normal monitoring mode:
42
Using the Monitor
LIMITS
1. Press the LIMITS softkey.
2. The monitor will display the ADULT LIMITS or NEONATE
LIMITS screen, depending on the patient setting being used.
ADULT or
NEO
3. Press the NEO or ADULT softkey to select ADULT LIMITS or
NEONATE LIMITS as applicable for the patient being
monitored.
N-59543
Using the Monitor
Alarm Limit Changed Indicator
Alarm limits that have been changed from the institutional or factory
default settings are identified by a decimal point (.) after the displayed
reading (%SpO
2 or BPM). The changed parameter is also identified
by a decimal point on the alarm limits screen.
Setting Alarm Limits
44
The Alarm Limit display allows the user to adjust the upper and lower
saturation and pulse rate limits. It also allows the user to adjust the
SatSeconds limit.
The Alarm Limit display is accessed by pressing the LIMITS softkey
on the Main menu.
The Alarm Limit display includes the alarm limit table and current
measured %SpO
2 and pulse rate. The title of the alarm limit table will
indicate whether the instrument is in Adult or Neonate monitoring
mode. If SatSeconds are enabled, the Alarm Limit display also
includes the SatSeconds indicator. Decimal points after the displayed
%SpO
2 or pulse rate indicate that the respective limits have been
changed from the power-on defaults.
With the monitor in the normal monitoring mode:
Using the Monitor
LIMITS
ADULT or
NEO
1. Press the LIMITS softkey. Current alarm limits are displayed.
or
2. Press the ADULT or NEO softkey to select Adult-Pediatric or
Neonatal alarm limits screen.
SELECT
3. Press the SELECT softkey as required to select the parameter to
be adjusted.
4. Use the ADJUST UP or ADJUST DOWN buttons to increase or
decrease the selected limit parameter.
5. Repeat steps 3, 4, and 5 as necessary to complete the alarm limits
setup.
EXIT
6. To accept the changes, let the display time-out or press the EXIT
softkey to exit the display and return to normal monitoring.
N-59545
Using the Monitor
Note: Limit changes will only be in effect as long as the monitor
remains turned on. When the monitor is turned off, the
institutional or factory default limits will be restored into the
monitor. When the monitor is turned on, the institutional or
factory default limits will be in effect. Factory or institutional
defaults are selected by qualified service personnel following
the procedure in the service manual.
Setting SatSeconds Alarm Limit
Refer to Describing SatSeconds on page 135, for a description of the
SatSeconds function.
With the monitor in the normal monitoring mode:
LIMITS
SELECT
EXIT
1. Press the LIMITS softkey. Current alarm limits are displayed.
2. Press the SELECT softkey twice to select %SpO
2 SAT-S.
3. Use the AJDUST UP or ADJUST DOWN buttons to select the
limit. The choices are 10, 25, 50, or 100 seconds or OFF.
4. Press the EXIT softkey to save your choice.
46
Setting Monitor Response Mode
The purpose of the response mode is to set the response time of the
XIMAX algorithm calculation of the SpO2 (the response mode does
O
not affect the O
trending interval (2- or 4-seconds) is updated automatically by the
monitor to roughly correspond with the SpO
time.
XIMAX algorithm’s calculation of pulse rate). The
Using the Monitor
2 calculation response
SETUP
NEXT
NEXT
NEXT
The response mode programs the O
XIMAX algorithm to display
monitor trend information at 2-second intervals (Fast Mode) or
4-second intervals (Normal Mode).
The response mode display screen includes the current SpO
mode setting and the current measured %SpO
2 and pulse rate. When
2 response
in the fast mode, the screen displays the fast mode symbol.
With the monitor in the normal monitoring mode:
1. Press the SETUP softkey.
2. Press the NEXT softkey.
3. Press the NEXT softkey.
4. Press the NEXT softkey.
MODE
5. Press the MODE softkey.
Note: When the monitor is in the fast response mode the monitor may
produce more SpO
N-59547
2 and pulse rate alarms than the user is
Using the Monitor
accustomed to seeing, and may be inappropriate in challenging
measurement conditions.
6. Use the ADJUST UP or ADJUST DOWN buttons to select the
desired response mode.
EXIT
7. Press the EXIT softkey.
Selecting the Display Language
The N-595 can be programmed to display the information in various
languages. The languages available are English, Francais (French),
Deutsch (German), Italiano (Italian), Espanol (Spanish), Nederlands
(Dutch), Port (Portuguese) and Sverige (Swedish).
With the monitor in the normal monitoring mode:
SETUP
NEXT
1. Press the SETUP softkey.
2. Press the NEXT softkey.
48
Using the Monitor
LANG
EXIT
3. Press the LANG softkey.
4. Use the ADJUST UP or ADJUST DOWN buttons to select the
desired language.
5. Press the EXIT softkey.
Note: The selected language will be displayed until the monitor is
turned off. The selected language can be set as a default by
qualified service personnel following the procedures in the
service manual.
OXIMAX Sensor Messages
OXIMAX sensor messages consist of sensor adjust condition messages
and sensor adjust messages which, when enabled, are displayed when
the monitor is not able to display saturation. When O
messages are displayed, it is an indication that the O
functioning correctly , but the site to which the O
or the application method is not optimal for calculating %SpO
Condition messages are followed by action messages. Up to three
condition messages may be displayed on the “POOR SIGNAL
CONDITION” display in priority order, highest on top. The condition
display may be dismissed by using the EXIT softkey. Once exited, the
N-59549
XIMAX sensor
XIMAX sensor is
XIMAX sensor applies
2.
Using the Monitor
OXIMAX sensor message screen will not return until a new condition
occurs.
If the HELP softkey is pressed from the Condition message display,
the action messages are displayed. Action messages are linked to the
sensor type; action messages will be displayed for the type of O
XIMAX
sensor connected to the monitor. Up to five action messages may be
displayed. Multiple screens may be required to display all of the
messages. When multiple screens are required, navigation between
screens can be accomplished through the NEXT, BACK, and EXIT
softkeys.
O
XIMAX sensor messages may be disabled. Refer to OXIMAX Sensor
Message Setup on page 127 for selecting the O
Messages, Enable/Disable function.
OXIMAX Sensor Adjust Condition Messages
•Condition 1 — SENSOR OFF?
•Condition 2 — SMALL PULSES
•Condition 3 — WEAK SIGNAL
XIMAX Sensor
50
•Condition 4 — MOTION INTERFERENCE
•Condition 5 — EXCESS INFRARED LIGHT
•Condition 6 — ELECTRICAL/LIGHT
•Condition 7 — HIGH PULSE AMPLITUDE
OXIMAX Sensor Adjust Messages
•Message 1 — ALTERNATE SITE?
•Message 2 — COVER SENSOR SITE?
•Message 3 — EAR/FOREHEAD SENSOR?
•Message 4 — NASAL/EAR SENSOR?
Using the Monitor
INTEREFERENCE
•Message 5 — O
XIMAX ADHESIVE SENSOR
•Message 6 — SECURE CABLE
•Message 7 — HEADBAND
•Message 8 — WARM SITE
•Message 9 — BANFAGE ASSEMBLY
•Message 10 — NAIL POLISH
•Message 11 — SENSOR TOO TIGHT?
•Message 12 — REPOSITION SENSOR
•Message 13 — ISOLATE INTERFERENCE SOURCE
•Message 14 — CLEAN SENSOR SITE
N-59551
Monitor Trend
Monitor Trend Data
The trend displays allow the user to view trend data. Two types of
trend data can be viewed:
•Monitor trend data which are stored in the monitor
•Patient event data which are stored in the O
(single-patient-use O
with the sensor event record feature.
Monitor trend data can be viewed anytime patient trend is stored in
the monitor. Monitor trend displays are accessed by pressing the
TREND softkey on the main menu and selecting the MONITR
softkey option. The monitor trend sub-menu allows you to choose
which trend data are displayed:
•Saturation and pulse rate (Dual)
•Saturation
•Pulse rate
•Pulse amplitude
•Histogram
The N-595 can graphically display trend data for SpO
both. Trend data is stored at 2- or 4-second intervals. When the
TREND softkey is pressed, “READING TRENDS . . .” is displayed
at the bottom of the N-595 screen, indicating that the monitor is
formatting the trend data to be displayed.
XIMAX sensors only) and can be used
XIMAX sensor
2, pulse rate, or
N-59553
Monitor Trend
The monitor stores up to 48 hours of 4-second trend data or 24 hours
of 2-second trend data. The amount of trend data displayed on the
screen is determined by using the ZOOM softkey. The settings
available are 20 and 40 seconds, 15 or 30 minutes, and 1, 2, 4, 8, 12,
24, 36, or 48 hours. All trend data are displayed in a graphical format
except the 20- and 40-second trend displays, which are shown in
tabular format.
The trend display is scrolled, that is, the data displayed can be moved
throughout the 48 hours of trend data. Selecting the 1-hour trend
display allows you to view one hour of trend information. By using
the scrolling feature, any one hour of trend data can be viewed over
the 48 hours of trend information. The AJUST DOWN button scrolls
the display to the left and the ADJUST UP button scrolls the display
to the right.
When the data are displayed, the most recent readings are on the right
side of the graph. The numbers below %SpO
2 indicate the highest and
lowest parameter values at the cursor position (vertical dotted line on
the display). See Table 4 on page 57.
.
Trend data is further explained in Specifications on page 147.
Trend data information may be retrieved through the N-595 data port
or cleared using options available in a display menu.
Caution: Monitor trend data will be lost if the main battery fails
or is removed.
54
Trend Data Operation
Whenever the N-595 is turned on, it stores the monitor %SpO2 and
pulse rate readings in memory every 2 or 4 seconds (regardless of
whether the N-595 is monitoring a patient or not). The N-595 can
store up to 48 hours of 4-second trend data or 24 hours of 2-second
trend data. The 48/24 hours of stored trend data are available for
downloading to a printer or a portable computer . Up to 50 alarm limit
changes can be stored in the trend data. If more than 50 alarm limit
changes occur during the 48/24 hours of trend data collection, the
additional alarm limit changes will take space reserved for trend data.
Caution: Changing alarm limit settings uses up trend memory
space. Change alarm limits only as needed.
Note: Trend memory always contains the most recent 48 hours of
Monitor Trend
data, with newly collected data overwriting the oldest data on a
rolling basis. The N-595 continues to record data points as long
as the monitor is powered on, with “blank” data points
collected if no O
XIMAX sensor is connected to the monitor or
patient. “Blank” data will over-write older patient data if the
memory becomes full. Therefore, if you want to save old
patient data, it is important that you turn your monitor of f when
you are not monitoring a patient, and that you download the
trend memory before it fills up and over-writes the old data
with new data (or “blank” data).
Selecting the Trend Data Display Scale
The trend scale is the amount of trend data displayed on the screen.
With the monitor in the normal monitoring mode:
TREND
MONITR
N-59555
1. Press the TREND softkey.
2. Press the MONITR softkey.
Monitor Trend
VIEW
ZOOM
TIME
3. Press the VIEW softkey.
4. Press any of the trend softkeys (DUAL, SPO2, or PULSE). To
select HIST (histogram) or AMP (amplitude), press the NEXT
softkey and then the HIST or AMP softkeys.
5. Press the ZOOM softkey. The Zoom menu is displayed.
Pressing the TIME softkey cycles the displayed trend time scale
through 48 hours, 36 hours, 12 hours, 8 hours, 4 hours, 2 hours, 1
hours, 30 minutes, 15 minutes, 40 seconds and 20 seconds.
Note: The 20-second and 40-second trend displays are in tabular
format. The below display starts out in the normal response
mode (left side of the display) and switches to the fast response
mode.
56
SCALE
Pressing the SCALE softkey cycles the displayed trend amplitude
scale through ±5 points, ±10 points, ±15 points, ±20 points, ±25
points, ±30 points, ±35 points, ±40 points and ±50 points above and
below the data point under the cursor. The saturation graphical
monitor trend display vertical scale default setting is from 10 to 100 if
there is no data under the cursor. The pulse rate graphical monitor
Monitor Trend
trend display vertical scale is from 5 to 250 if there is no data under
the cursor.
AUTO
Pressing the AUTO softkey presets the amplitude of the graphed
trend data. The maximum trend data point is rounded up to the nearest
multiple of 10, this value is the top of the graph display. The
minimum trend data point is rounded down to the next multiple of 10.
Then 10 is subtracted from the rounded down number, this value is
the bottom of the trend graph.
BACK
Pressing the BACK softkey returns the monitor to the Monitor menu.
Reading the Trend Data Display
Table 4 identifies the components of the trend data display.
Table 4: Reading Trend Display
ItemDescription
1Amount of trend data displayed on the screen. Settings available
are 20 and 40 seconds, 15 and 30 minutes, 1, 2, 4, 8, 12, 24, 36,
and 48 hours.
2Date represented by the cursor (item 5).
3No trend data recorded during this time.
4Time represented by the cursor (item 5).
N-59557
Monitor Trend
Table 4: Reading Trend Display
ItemDescription
5
6Highest and lowest reading at the cursor position.
7Trend data that is being displayed (%SPO2, BPM, or PAU
Cursor
(right) or ADJUST DOWN (left) buttons.
[pulse amplitude units]).
Dual Trend Data Display
The dual trend data display displays both oxygen saturation (%SpO2)
levels and pulse rate (bpm) trend data.
With the monitor in the normal monitoring mode:
TREND
MONITR
1. Press the TREND softkey.
2. Press the MONITR softkey.
- can be moved left or right using the ADJUST UP
58
VIEW
DUAL
3. Press the VIEW softkey.
4. Press the DUAL softkey. The dual trend (%SpO
is displayed.
2and Pulse Rate)
SpO2 Trend Display
With the monitor in the normal monitoring mode:
Monitor Trend
TREND
MONITR
VIEW
SPO2
1. Press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
4. Press the SPO2 softkey. SpO
Pulse Rate Trend Display
2 trend data is displayed.
With the monitor in the normal monitoring mode:
TREND
MONITR
VIEW
N-59559
1. Press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
Monitor Trend
PULSE
4. Press the PULSE softkey. The pulse trend data is displayed.
Histogram Trend Data Display
The histogram displays trend data for the percent of oxygen blood
saturation (SpO
the trend data stored over the period of time indicated on the display.
Refer to Selecting the Trend Data Display Scale on page 55, to set up
the desired trend data scale.
Pulse amplitude cannot be displayed on the histogram display.
2) and pulse rate (bpm). The data displayed represents
60
TREND
MONTIR
VIEW
NEXT
With the monitor in the normal monitoring mode:
1. Press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
4. Press the NEXT softkey.
Monitor Trend
HIST
5. Press the HIST softkey. The Histogram trend data is displayed.
Pulse Amplitude Trend Data Display
The pulse amplitude trend data display shows the amplitude of the
patient’s pulse rate over the period of time indicated on the display.
Refer to Selecting the Trend Data Display Scale on page 55, to setup
the desired trend data scale.
With the monitor in the normal monitoring mode:
TREND
MONITR
VIEW
NEXT
N-59561
1. Press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
4. Press the NEXT softkey.
Monitor Trend
AMP
5. Press the AMP softkey. The pulse amplitude units (PAU) trend
data is displayed.
The PAU reading (12 : 20) indicates the pulse amplitude units (upper
and lower) at the cursor position (dashed line). The cursor is moved
right or left using the ADJUST UP (right) and ADJUST DOWN (left)
buttons.
Clearing Trend Information
With the monitor in the normal monitoring mode:
62
TREND
MONITR
NEXT
DELETE
YES
1. Press the TREND softkey.
2. Press the MONITR softkey.
3. Press the NEXT softkey.
4. Press the DELETE softkey.
Note: Press the NO softkey and then the EXIT softkey to leave this
function without deleting trend data.
5. Press the YES softkey.
Monitor Trend
All the trend data is cleared and the monitor sounds three beeps.
N-59563
O XIM AX Sensor Event Record
WARNING: The sensor extrapolates from the date and time
provided by the N-595 when recording the sensor event record to
the sensor. The accuracy of the date/time is the responsibility of
the N-595. It is recommended that the N-595 user set the time/
date to the correct value before a sensor event record-enabled
sensor is connected, and that this date/time not be changed while
the sensor remains connected. Since a sensor with sensor event
record data can be transported from one monitor to another,
having discrepancies in the date/time between monitors and the
sensor event record data will affect the order the sensor event
record data appears. To eliminate this possible problem, all
monitors within an institution should be set to the same time.
The adhesive O
XIMAX sensors are capable of storing patient event
data. A sensor event record allows alarm event history to travel with
the patient on the sensor’s memory chip for quick assessment at every
point of care where O
Patient (event) data is stored on the memory chip of adhesive O
sensors (single-patient-use O
XIMAX monitors are used.
XIMAX
XIMAX sensors only). The event data is
stored (recorded) with the limit/threshold settings that were active at
the time of the event on the recording monitor. These events can be
viewed on the next O
XIMAX sensor monitor when the patient moves to
a new point of care.
An event occurs when the %SpO
lower alarm limit for at least 15 seconds. The first O
event record event will be stored in the O
XIMAX sensor has been attached to a patient for five minutes and
O
2 value exceeds either the upper or
XIMAX sensor
XIMAX sensor after the
every five minutes thereafter. The maximum number of events that
can be stored in an O
XIMAX sensor is 100.
N-59565
OXIMAX Sensor Event Record
Event records can only be viewed after an OXIMAX sensor containing
patient data (event records) has been connected to an O
monitor. Event records are designed to view patient events from prior
areas of care or transport (history) while monitor trend should be used
to view data or events from a patient currently being monitored. The
monitor’s SENSOR EVENT RECORD indicator will light when an
XIMAX sensor containing event data is connected to the OXIMAX
O
monitor.
Patient event data is accessed by pressing the TREND softkey on the
main menu and selecting the SENSOR softkey option. Sensor event
record can be viewed in graphical form (GRAPH) or in a summary
table (TABLE).
XIMAX
Note: Once the O
XIMAX sensor and event data is stored in the OXIMAX sensor,
O
the O
XIMAX sensor event record type cannot be reset. The
monitor’s type set up can be changed at any time.
Recording and viewing of O
available on O
XIMAX comparable monitors. The OXIMAX sensors may
function on older technology monitors but the O
record feature is not available.
Refer to the N-595 service manual for the procedure to disable the
storage of sensor event record on an O
Setting In-Sensor Data Type
The In-Sensor Data Type display allows the user to set the type of
trend data to be recorded in an O
set to record either SpO
Note:The O
XIMAX sensor data type can only be set when an OXIMAX
sensor is not connected to the monitor.
XIMAX sensor event record type is set up in the
XIMAX sensor event record is only
XIMAX sensor event
XIMAX sensor.
XIMAX sensor. OXIMAX sensors can be
2 or SpO2+BPM.
66
With the monitor turned on and no cable attached to the SpO
sensor port:
2OXIMAX
OXIMAX Sensor Event Record
R
SETUP
SENSO
DATA
1. Press the SETUP softkey.
2. Press the SENSOR softkey.
3. Press the DATA softkey.
Note: O
XIMAX sensor data type settings are displayed on the monitor
as shown in the figure below (in-sensor data type). If no sensor
is connected, both sensor types and the full set of options for
each are displayed. If a sensor is connected, only the sensor
data type for that sensor is displayed.
Note: The SENSOR-R feature supports all the current O
XIMAX
sensors. The SENSOR-RW feature is only applicable to
O
XIMAX sensors with a read/write chip installed.
SELECT
4. Use the SELECT softkey to toggle between SENSOR-R and
SENSOR-RW.
N-59567
OXIMAX Sensor Event Record
5. Use the ADJUST UP or ADJUST DOWN button to select the
O
selections are:
XIMAX sensor data type. SENSOR-R and SENSOR-RW
EXIT
6. Press the EXIT softkey to set the O
OXIMAX Sensor Type
When an OXIMAX sensor is connected to the monitor, a “SENSOR
TYPE: …” message is displayed for 4 to 6 seconds at the bottom of
the display . The message identifies the type (model) of O
connected to the monitor. Type is used in the determination of action
messages in the O
the first message displayed when an O
the monitor.
•SpO
2
•SpO2+BPM
•DEFAULT
XIMAX sensor message(s) function. This display is
XIMAX sensor type.
XIMAX sensor
XIMAX sensor is connected to
OXIMAX Sensor Data Type
When an OXIMAX sensor with no previously recorded patient data is
connected to the O
displayed briefly at the bottom of the display, this message is
68
XIMAX monitor, a “DATA TYPE: . . .” message is
OXIMAX Sensor Event Record
displayed after the OXIMAX sensor type message. The message
identifies the monitor's current data type setting that will be used to
write data to the O
EVENT/SPO
XIMAX sensor. The data type setting options are
2 and EVENT/SPO2+BPM.
The user can change the setting by referring to Setting In-Sensor Data Type on page 66. The O
prior to connecting the O
XIMAX sensor event record type must be set
XIMAX sensor to the monitor.
OXIMAX Sensor Event Record Data Available
When an OXIMAX sensor containing data (single-patient-use OXIMAX
sensors only) is connected to the monitor, the Sensor Event Record
indicator on the monitor front panel blinks at a medium priority flash
TREND
rate to indicate that the O
XIMAX sensor attached to the monitor
contains patient event data. The LED blinks for approximately 60
SENSOR
seconds or until the O
XIMAX sensor is disconnected or until the sensor
trend data is displayed by pressing TREND, then SENSOR.
A corresponding “DATA IN SENSOR” message is also displayed at
the bottom of the display. After 4 to 6 seconds, if all the data has been
read from the O
XIMAX sensor, the message is replaced with the main
menu.
N-59569
OXIMAX Sensor Event Record
If data is still being read from the OXIMAX sensor , after 4 to 6 seconds,
the DATA IN SENSOR message is replaced with a READING
TRENDS message with an ABORT option.
Selecting the ABORT softkey stops the recording of additional data
in the O
XIMAX sensor.
O
Sensor event record can be viewed by accessing the TREND/
SENSOR menu.
The SENSOR EVENT RECORD LED comes on steady when
O
XIMAX sensor memory is full and stays on until the OXIMAX sensor is
disconnected.
XIMAX sensor and accessing or viewing the data that is in the
OXIMAX Sensor Event Record Not Available
If the user selects the TREND/SENSOR option when a connected
XIMAX sensor (single-patient-use OXIMAX sensors only) does not
O
contain data, because no events were recorded to the O
memory chip in the prior monitoring situation, a “NO DATA”
message is displayed on the default trend or event graph.
70
XIMAX sensor
OXIMAX Sensor Event Record
A sample event display in which no data are available is shown
below. The message will be cleared when the graph or summary is
exited.
OXIMAX Sensor Event Record Graphical Data
Graphical representations of patient event history is only available on
single-patient-use O
minimum or maximum %SpO
throughout the duration of an event (%SpO
alarm threshold for at least 15 seconds) and continuing every 30
seconds until the actual %SpO
threshold.
XIMAX sensors. Graphed data points are the
2 value for each 30-second interval
2 value equals or exceeds the alarm
2 continuously below
The duration of an event is determined by the number of data points
in the event. Each data point is stored at 30-second intervals.
Events end for one of four reasons:
•The %SpO2 returns to or above the alarm limit
•Loss of pulse
•The O
N-59571
XIMAX sensor is disconnected
OXIMAX Sensor Event Record
The graph title shows the data type (EVENT GRAPH) in the upper
left corner. The number of the displayed event and the total number of
events recorded in the O
title (example, 2/2). The date and time of the displayed event are
shown in the upper center and upper right corner.
The type of data displayed in the graph is indicated to the left of the
vertical axis (%SpO
during the event. The duration of the event is shown below the range
value. The vertical axis of the graph is labeled to show the magnitude
scale of the graphed data. The horizontal axis is not labeled but
automatically scales to accommodate the number of 30-second
intervals during the event. The alarm threshold (lower than %SpO
alarm limit) is represented by a horizontal dotted line across the
graph. The first data point is always the alarm threshold.
•The OXIMAX sensor is off the patient
XIMAX sensor are shown to the right of the
2). Below this is the range of values (min/max)
2
72
Events are displayed one at a time, one per graph. Graphs are
displayed in chronological sequence with the most rece nt ev ent
shown first when accessing the graphical O
XIMAX sensor event
display. The user can move between events by using the two left-most
softkeys which are labeled with left- and right-facing arrow icons,
respectively. At the beginning of an event sequence, event 1 of 2
events, the left-arrow soft key is blank; at the end of a sequence, event
2 of 2 events, the right-arrow soft key is blank.
The ADJUST UP and ADJUST DOWN buttons on the monitor panel
can also be used to move through events.
The PRINT softkey allows the user to print the displayed event graph.
The BACK softkey takes the user back to the previous TREND/
SENSOR sub-menu level.
OXIMAX Sensor Event Record
R
Viewing and Printing OXIMAX Sensor Event History Data
With the monitor in the normal monitoring mode. You must connect a
printer, capable of printing graphs, to the monitor data port connector
to print O
The monitor protocol must be set to GRAPH to print the in-sensor
event history data. Refer to Printing Monitor Trend Information on
page 79. To view and print in-sensor event history data:
XIMAX sensor event history data.
TREND
SENSO
GRAPH
1. Connect an O
XIMAX sensor containing patient data to the monitor .
2. Press the TREND softkey.
3. Press the SENSOR softkey.
4. Press the GRAPH softkey.
Note: Use the left and right arrow softkeys to scroll through the pages
of the event graph.
PRINT
EXIT
5. Press the PRINT softkey to print the displayed screen.
6. Press the EXIT softkey.
N-59573
OXIMAX Sensor Event Record
A sequence of %SpO2 + BPM (saturation plus pulse rate)
“dual-view” event graphs are shown below. The duel-view graph is
the same as a single graphical event history graph except the graphs
are compressed horizontally to allow both %SpO
graphs to be shown for the same event.
2 and pulse rate
74
OXIMAX Sensor Tabular Event Data
The OXIMAX sensor tabular event data is a listing of all events recorded on
the
OXIMAX sensor’s memory chip.
OXIMAX Sensor Event Record
The table title shows in the upper left corner. Below the table title is a
six-column table with left-to-right column headings of event number
(#), date (DATE), event start time (START), event duration (DUR),
%SPO
2 minimum and maximum values during the event (%SPO2),
and pulse rate minimum and maximum values during the event
(BPM).
Event data are listed in chronological order with the most recent event
shown first, at the top of the list, when the tabular Event Summary
display is first accessed. Four events can be displayed
simultaneously; the table must be scrolled to view additional events.
The user can move to the next screen view of the table, the next three
events (the previously displayed bottom or top event is retained as the
fourth event for context when a table is scrolled), using the two
left-most softkeys which are labeled with left- and right-facing arrow
icons, respectively. At the beginning of an event sequence, Event 1
of 5 events, the left-arrow soft key is blank; at the end of a sequence,
N-59575
OXIMAX Sensor Event Record
R
Event 5 of 5 events, the right-arrow soft key is blank, indicating you
have reached the beginning or end of the table.
The ADJUST UP and ADJUST DOWN buttons on the monitor panel
can be used to move through the Event Summary table line by line.
The PRINT softkey allows the user to print the displayed event graph.
The BACK softkey takes the user back to the previous TREND/
SENSOR sub-menu level.
Viewing and Printing In-Sensor Tabular Event History Data
The monitor should be in the normal monitoring mode.
To view and print in-sensor tabular event history data:
TREND
SENSO
TABLE
PRINT
1. Press the TREND softkey.
2. Press the SENSOR softkey.
3. Press the TABLE softkey.
4. Press the PRINT softkey to print the data.
76
OXIMAX Sensor Event Record
BACK
5. Press the BACK softkey.
N-59577
Printing
Printing Monitor Trend Information
Trend information (monitor and in-sensor event history) may be sent
to a personal computer or to a serial printer.
Note: The protocol settings must be set to ASCII MODE for printing
text data or GRAPH MODE for printing graphical data.
1. Data Port Connector
With the monitor in the normal monitoring mode:
1. Connect the serial printer to the monitor’s DATA PORT
connector (1), using Nellcor printer cable part number 036341.
2. Turn on the printer.
SETUP
NEXT
N-59579
3. Press the SETUP softkey.
4. Press the NEXT softkey.
Printing
R
NEXT
COMM
SELECT
5. Press the NEXT softkey.
6. Press the COMM softkey.
7. Set the BAUD rate to the appropriate number using the ADJUST
UP button.
8. Press the SELECT softkey to select PROTOCOL.
9. Set the PROTOCOL to ASCII for text printing or GRAPH for
graph printing using the ADJUST UP button.
EXIT
TREND
MONITR
10. Press the EXIT softkey.
11. Press the TREND softkey.
12. Press the MONITR softkey for monitor trend printing or press the
SENSOR softkey for in-sensor event history data printing.
SENSO
80
Printing
NEXT
PRINT
13. Press the NEXT softkey.
14. Press the PRINT softkey.
ASCII printout:
GRAPH printout:
Monitor Trend Data in ASCII Mode
Refer to Printing Monitor Trend Information on page 79 for the
procedure to print trend information.
N-59581
Printing
The format of data displayed when a trend printout is shown in
Figure 3. “TREND” is displayed in the top row.
Readings are displayed in 2- or 4-second intervals depending on the
response mode selected. The values on each row are an average of the
response mode selected period.
At the end of the printout an “Output Complete” line indicates that
the transmission was successful. If the “Output Complete” line is not
present, a corruption of the data may have been detected and the data
should be ignored.
Once a trend printout has begun, it cannot be aborted without turning
off the N-595 or the printer.
Trend Data in Graph Mode
Refer to Printing Monitor Trend Information on page 79 for the
procedure to print trend information. See Figure 4 on page 83.
82
Figure 3: ASCII Mode Printout
The graph mode disables all printout functions except trend data.
Graph mode trend printouts are formatted for a Seiko DPU-414 and
Okidata 320 serial printer.
Figure 4: Graph Mode Printout
Real-Time Display/Printout Format
Real-time data is continuously sent to the data port on the back of the
N-595. Patient data can be obtained through the data port by
connecting the monitor data port to a PC or serial printer. When a
real-time printout or display is being transmitted to a printer or PC, a
new line of data is displayed every 2 seconds. Column headings are
displayed or printed after every 25 lines, or if one of the values in the
column heading changes. Readings are displayed at 4-second
intervals if the SpO
intervals when the SpO2 response mode is set to fast.
2 response mode is set to normal and at 2-second
Printing
Data cannot be obtained if the N-595 is operating on battery power.
Note: If the data output stops transmitting, turn the power off and
back on again or, if the monitor is connected to a PC, send an
XON (Ctrl-q) command to reset the monitor.
N-59583
Printing
An example of a real-time output is shown in Figure 5 on page 84.
84
Figure 5: Real-Time Printout
Column Headings
Data Source
Printing
Every 25th line of the data consists of a column heading.
A column heading is also output whenever a value of the column
heading is changed. There are three column-heading lines shown in
the printout. Using the top row as the starting point there are 25 lines
before the second row of column headings is printed. The third row of
column headings was displayed because the operator changed the
SpO
2 lower alarm limit from 85 percent to 80 percent.
Data in the highlighted box above represents the model number of the
monitor, in this case the N-595.
Software Version
The next data field tells the user the software level (Version 1.0.0.0)
and a software verification number (CRC: XXXX). Neither of these
numbers should change during normal operation. The numbers may
change if the monitor is serviced and receives a software upgrade.
N-59585
Printing
Alarm Limits
Monitor Mode
The last data field in the top line indicates the upper and the lower
alarm limits for %SpO
above the lower alarm limit for SpO
2 and for the pulse rate (PR). In the example
2 is 85% and the upper alarm
limit is 100%. Pulse Rate alarm limits are 40 and 170 bpm. The
SatSeconds alarm limit (0SAT-S) displays the SatSeconds alarm
setting. In this example SatSeconds is set to off.
Response Mode
86
The monitor mode (ADULT or NEONATE) is identified on the
printout.
The response mode (NORMAL or FAST) is identified on the
printout.
Data Column Headings
Actual column headings are in the second row of the column heading
line. Patient data presented in the chart, from left to right, is the:
Printing
•time the patient data were obtained
Time
Patient Data
•current %SpO
2 value
•current Pulse Rate (BPM)
•current Pulse Amplitude (PA)
•operating status of the N-595.
The Time column represents the N-595 real-time clock.
Patient data are highlighted in the display above. Parameter values are
displayed directly beneath the heading for each parameter. In this
N-59587
Printing
Operating Status
example the %SpO2 is 100, and the pulse rate is 190 beats per minute.
The “*” next to the 190 indicates that 190 beats per minute is outside
of the alarm limits, indicated in the top row, for pulse rate. If no data
for a parameter is available, three dashes (- - -) will be displayed.
PA represents pulse amplitude. The number can range from 0 to 254.
There are no alarm parameters for this value. It can be used for
trending information as an indication of a change in pulse volume,
relative pulse strength, or circulation.
The Status column indicates alarm conditions and operating status of
the N-595. In this example, the PH means that the pulse rate upper
alarm limit (Pulse High) has been exceeded. A complete listing of the
status codes is listed below. As many as four codes can be displayed
at one time in the Status column.
88
CodeMeaning
AOAlarm Off
ASAlarm Silence
LBLow Battery
LMLoss of Pulse w/ Motion
LPLoss of Pulse
MOPatient MOtion
PHPulse Rate Upper Limit Alarm
PLPulse Rate Lower Limit Alarm
PSPulse Search
Printing
CodeMeaning
SHSaturation Upper Limit Alarm
SLSaturation Lower Limit Alarm
SDSensor Disconnect
SOSensor Off
Note:An OXIMAX sensor disconnect will also cause three dashes
(- - -) to be displayed in the patient data section of the display
or printout.
N-59589
Loading...
+ hidden pages
You need points to download manuals.
1 point = 1 manual.
You can buy points or you can get point for every manual you upload.