Nellcor N-595 User manual

N-595
Pulse Oximeter Operator’s Manual
Nellcor Puritan Bennett Inc. is an affiliate of Tyco Healthcare. Nellcor, Oxiband, Durasensor, OxiCliq, Dura-Y, MAX-FAST, and O
XIMAX are trademarks of Nellcor Puritan Bennett Inc.
This ISM device complies with Canadian ICES-001. Cet appareil ISM est conforme à la norme NMB-001 Canada. To obtain information about a warranty, if any, contact Nellcor’s Technical Services Department,
or your local representative. Purchase of this instrument confers no express or implied license under any Nellcor Puritan
Bennett patent to use the instrument with any sensor that is not manufactured or licensed by Nellcor Puritan Bennett.
Covered by one or more of the following U.S. Patents and foreign equivalents: 4,621,643; 4,653,498; 4,700,708; 4,770,179; Re. 35,122; 4,802,486; 4,869,254; 4,928,692; 4,934,372; 5,078,136; 5,351,685; 5,368,026; 5,485,847; 5,533,507; 5,662,106; and 5,853,364.
Contents
Contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .i
Figures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .vi
Tables . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . vi
Safety Information and Introduction
Safety Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
Warnings ........................................................................................... 1
Cautions ............................................................................................ 2
Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
Intended Use for the N-595 .................................... ... ... ... ... .... ... ... ... ..5
How to Use this Manual ....................................................................6
Using the N-595
Description of Controls, Indicators, and Symbols . . . . . . . . . . . . . . . . . . . . . . 7
Identification of Front Panel Buttons and Symbols ...........................7
Identification of Rear Panel Components ..........................................8
N-595 Symbols ...................... ... .... ... ....................................... ... ... ... ..8
Description of Controls ......................................................................9
Description of Displays and Indicators ............................................ 10
Description of Audible Indicators .....................................................14
Setting up the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
List of Components .........................................................................17
Connecting the N-595 to AC Power .......................................... ... ...17
Connecting an O
XIMAX Sensor to the N-595 ..................................19
Battery Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 21
Operating the N-595 on Battery Power ........................................... 21
Low Battery Indicator ............................................. ... ... ... ... .... ... ... ...22
Using the Monitor . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 27
Introduction ....................................................................................27
N-595 i
Contents
Turning On the Monitor ................................................................... 29
O
XIMAX Sensor Attached ..................................................... 31
No O
XIMAX Sensor Attached .................... ... ... ... ... .... ... ... ... ... 33
Turning the Backlight On or Off ...................................................... 34
Adjusting Screen Contrast .............................................................. 34
Selecting the Pleth View ................................................................. 34
Selecting the Blip View ................................................................... 35
Setting the Pulse Beep Volume .......................................... ... ... ... ... 36
Setting the Alarm Volume .. .... ... ... ... ... .... ... ... ... .... ... ... ... ... .... ... ... ... ... 36
Setting the Date and Time . .... ... ... ... ... .... ... ... ... .... ... ... ... ... .... ... ... ... ... 37
Setting Alarm Silence Duration ....................................................... 38
Disabling Audible Alarms ................................................................ 39
Selecting Standby Mode ................................................................. 41
Adult-Pediatric or Neonatal Settings ............................ ................... 42
Setting Patient Adult-Pediatric/Neonatal Mode ....................42
Alarm Limit Changed Indicator ....................................................... 44
Setting Alarm Limits ............................... ... ... ... .... ... ... ... ... ................ 44
Setting SatSeconds Alarm Limit ........ .... ... ... ... .... ... ... ... ... .... ... ... ... ... 46
Setting Monitor Response Mode .............................................. ... ... 47
Selecting the Display Language ..................................................... 48
O
XIMAX Sensor Messages .............................................................. 49
O
XIMAX Sensor Adjust Condition Messages ........................ 50
O
XIMAX Sensor Adjust Messages ........................................ 51
Monitor Trend . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .53
Monitor Trend Data ........................................................................ 53
Trend Data Operation ........................................................... 55
Selecting the Trend Data Display Scale ......................................... 55
Reading the Trend Data Display ............................ ................... ...... 57
Dual Trend Data Display ................................................................. 58
SpO2 Trend Display ....................................................................... 59
Pulse Rate Trend Display ............................................................... 59
Histogram Trend Data Display .................. ... ... .... ... ... ... ... .... ... ... ... ... 60
Pulse Amplitude Trend Data Display .............................................. 61
Clearing Trend Information ............................................................. 62
Sensors and Accessories
OXIMAX Sensor Event Record . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .65
Setting In-Sensor Data Type .......................................................... 66
O
XIMAX Sensor Type ...................................................................... 68
O
XIMAX Sensor Data Type ............................................................. 68
O
XIMAX Sensor Event Record Data Available ................................ 69
ii
Contents
O
XIMAX Sensor Event Record Not Available ..................................70
O
XIMAX Sensor Event Record Graphical Data ................................71
Viewing and Printing O O
XIMAX Sensor Tabular Event Data ...............................................75
XIMAX Sensor Event History Data ...............73
Viewing and Printing In-Sensor Tabular Event History Data ...........76
Printing . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 79
Printing Monitor Trend Information ..................................................79
Monitor Trend Data in ASCII Mode .................................................81
Trend Data in Graph Mode .............................................................82
Real-Time Display/Printout Format .................................................83
Column Headings .................................................................85
Data Source ..........................................................................85
Software Version ..................................... .... ... ... ... ... .... ... ... ...85
Alarm Limits ..........................................................................86
Monitor Mode ........................ ... ... .... ... ... ... .... .........................86
Response Mode ....................................................................86
Data Column Headings .........................................................87
Time ...................................... ................................................87
Patient Data .................................... ... ... ... .... .........................87
Operating Status ......................... .... ... ... ... .... ... ... ... ... .... ... ... ...88
Using the Data Port . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91
Overview ......................................................................................... 91
Connecting to the Data Port ............................................................91
Data Port Pinouts ..................................................................92
Data Port Setup ...............................................................................93
Using the Nurse Call Interface ........................................................95
Setting Nurse Call RS-232 Polarity .......... .... ... ... ... ... .... ... ... ...96
Setting Nurse Call Relays Normally Open/Closed ................97
Calculating the Analog Voltage Output ...........................................97
OXIMAX Sensors and Accessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 99
OXIMAX Sensor Event Record Data ................................................99
Selecting an O
O
XIMAX Sensor Features ..............................................................103
Biocompatibility Testing ................................................................103
Optional Accessories ................................................ ... ... ... .... ... ... .103
GCX Mounting Plate ...........................................................105
GCX Poly-Mount (vertical wall mount with
GCX Poly-Mount (horizontal wall mount with rail adapter) .107
GCX Poly-Mount Roll Stand ...............................................108
GCX Utility Basket ..............................................................109
N-595 iii
XIMAX Sensor ...........................................................99
19-inch channel) ...........................................................106
Contents
Soft-Sided Carrying Case ................................................... 110
Performance Considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 111
Performance Verification ................................................ .... ... ... ... . 111
N-595 Monitor Performance Considerations ................................ 111
Dysfunctional Hemoglobins ................................................ 112
Anemia ............................................................................... 112
Saturation ........................................................................... 112
Pulse Rates ........................................................................ 112
O
XIMAX Sensor Performance Considerations .............................. 113
Troubleshooting
Troubleshooting . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .117
Error Codes .................................................................................. 117
Prompts and Error Messages ....................................................... 119
Corrective Action .......................................................................... 122
EMI (Electro-magnetic Interference) ......................... .................... 125
Obtaining Technical Assistance .................................................... 126
O
XIMAX Sensor Message Setup ................................................... 127
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .129
Returning the N-595 ..................................................................... 129
Service ................................... ................................................ ....... 129
Periodic Safety Checks ................................................................. 130
Cleaning ....................................... ................... ....................... ....... 130
Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .131
N-595 Menu Description ............................................................... 131
Technical Information
SatSeconds . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .135
Describing SatSeconds ................................................................ 135
SatSeconds “Safety Net” .............................................................. 137
SatSeconds Display ...................................................................... 137
Factory Defaults . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .139
Neonate Default Settings .............................................................. 139
Adult Default Settings ................................................................... 140
iv
Contents
Principles of Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 143
Oximetry Overview .......................................................................143
Automatic Calibration ..........................................................144
Functional versus Fractional Saturation .............................144
Measured versus Calculated Saturation .............................145
O
XIMAX Technology ......................................................................145
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 147
Performance .................................................................................147
Electrical ....................................... ................ ................ ................ .148
Environmental Conditions ..... ... .... .......................................... ... ... .149
Physical Characteristics ................................................................151
Compliance ................................................................................... 152
Manufacturer’s Declaration ...........................................................155
Index . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 167
N-595 v
Contents

Figures

Figure 1: Front Panel Buttons and Symbols .............................7
Figure 2: Rear Panel Components ...........................................8
Figure 3: ASCII Mode Printout ...............................................82
Figure 4: Graph Mode Printout ...................................... .........83
Figure 5: Real-Time Printout ...................................................84
Figure 6: Data Port Pin Layout ................................................93
Figure 7: GCX Mounting Plate ..............................................105
Figure 8: GCX Poly-Mount (vertical wall mount with
19-inch channel) ....................................................106
Figure 9: GCX Poly-mount (horizontal wall mount with
rail adapter) ............................................................107
Figure 10: GCX Poly-mount Roll Stand ..................................108
Figure 11: GCX Utility Basket .................................................109
Figure 12: Soft-Sided Carrying Case ...................................... 110
Figure 13: Alarm Response with SatSeconds ........................136
Figure 14: Oxyhemoglobin Dissociation Curve .......................145
vi

Tables

Contents
Table 1: Audible Indicators ....................................................14
Table 2: Low Battery and Critical Battery ..............................23
Table 3: Parameter Ranges ......... .... ... ... ... .............................27
Table 4: Reading Trend Display ............................................57
Table 5: Data Port Pinouts ........ ... .... ... ... ... .... ... ......................92
Table 6: Analog Pinouts ............ .............................................97
Table 7: Nellcor OXIMAX Sensor Models and
Patient Sizes .................. .... ... ... ... .... ... ... ... ... .... .......101
Table 8: OXIMAX Sensor Features .....................................1 03
Table 9: Error Codes .................................... ... ... ... ... .... ... ... .118
Table 10: Prompt/Error Messages ........... ... .... ... ... ... ... .... ... ... .120
Table 11: Neonate Factory Defaults ......................................139
Table 12: Adult Factory Defaults ...........................................140
Table 13: Electromagnetic Emissions ...... ... .... ... ... ... ... .... ... ... .155
Table 14: Electromagnetic Immunity .......................... .... ... ... .156
Table 15: Electromagnetic Immunity, RF Portable
Equipment ....................................... .......................159
Table 16: Recommended Separation Distances ...................161
Table 17: Cables ......... ... ... .... ... ... ... .... ... ... ..............................162
N-595 vi

Safety Information

Warnings

Warnings are identified by the WARNING symbol shown above.
W arnings alert the user to potential serious outcomes (death, injury, or adverse events) to the patient or user.
WARNING: The sensor extrapolates from the date and time provided by the N-595 when recording the sensor event record to the sensor. The accuracy of the date/time is the responsibility of the N-595. It is recommended that the N-595 user set the time/ date to the correct value before a sensor event record-enabled sensor is connected, and that this date/time not be changed while the sensor remains connected. Since a sensor with sensor event record data can be transported from one monitor to another, having discrepancies in the date/time between monitors and the sensor event record data will affect the order the sensor event record data appears. To eliminate this possible problem, all monitors within an institution should be set to the same time.
WARNING: Explosion hazard. Do not use the N-595 pulse oximeter in the presence of flammable anesthetics or gases.
WARNING: Chemicals from a broken LCD display panel are toxic when ingested. Use caution when handling a pulse oximeter with a broken display panel.
N-595 1
Safety Information
WARNING: Pulse oximetry readings and pulse signals can be affected by certain environmental conditions, O
XIMAX sensor
application errors, and certain patient conditions. See the appropriate sections of this manual for specific safety information.
WARNING: The use of accessories, sensor s, and cables other than those specified may result in increased emission and/or decreased immunity and inaccurate readings of the N-595 pulse oximeter.
WARNING: Failure to cover the OXIMAX sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.

Cautions

2
Cautions are identified by the CAUTION symbol shown above.
Cautions alert the user to exercise care necessary for the safe and effective use of the N-595 pulse oximeter.
Safety Information
Caution: When connecting the N-595 to any instrument, verify proper operation before clinical use. Both the N-595 and the instrument connected to it must be connected to a grounded outlet. Accessory equipment connected to the pulse oximeter’s data interface must be certified according to IEC Standard 950 for data-processing equipment or IEC Standard 60601-1 for electromedical equipment. All combinations of equipment must be in compliance with IEC Standard 60601-1-1 systems requirements. Anyone who connects additional equipment to the signal input port or signal output port (N-595 data port connector) configures a medical system and is therefore responsible for ensuring that the system complies with the requirements of system standard IEC Standard 60601-1-1 and the electromagnetic compatibility system standard IEC S tandard 60601-1-2. The N-595 accuracy may degrade if it is connected to secondary I/O devices when the instrument is not connected to earth reference.
Caution: Federal law (U.S.A.) restricts this device to sale by or on the order of a physician.
Caution: Dispose of battery in accordance with local requirements and regulations.
N-595 3

Introduction

WARNING: The N-595 is intended only as an adjunct in patient assessment. It must be used in conjunction with clinical signs and symptoms.
Intended Use for the N-595
The N-595 pulse oximeter is indicated for the continuous non-invasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO with neonatal, pediatric, and adult patients during both no motion and motion conditions and for patients who are well or poorly perfused, in hospitals, hospital-type facilities, intra-hospital transport, and home environments. For prescription use only.
Note: Hospital use typically covers such areas as general care floors,
operating rooms, special procedure areas, intensive and critical care areas, within the hospital plus hospital-type facilities. Hospital-type facilities include physician office based facilities, sleep labs, skilled nursing facilities, surgicenters, and sub-acute centers.
2) and pulse rate. The N-595 is intended for use
Intra-hospital transport includes transport of a patient within the hospital or hospital-type facility .
Home Care use is defined as managed/used by a lay person (parent or other similar non-critical caregiver) in the home environment.
Use with any particular patient requires the selection of an appropriate oxygen O Operator's Manual.
N-595 5
XIMAX sensors as described in this
Introduction
Motion performance claims are applicable to models MAX-A, MAX-AL, MAX-P, MAX-N, and MAX-I Nellcor O oximetry sensors.

How to Use this Manual

All users should read this manual thoroughly . More experienced users of the N-595 will be able to go to the topics for the information they require.
The current copy of this manual is available on the internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_Supp/ProductManuals.html
XIMAX
TM
6

Description of Controls, Indicators, and Symbols

Identification of Front Panel Buttons and Symbols

1. SpO2 OXIMAX Sensor Port, page 19
2. AC Power Indicator, page 12 13. Neonate Mode Indicator, page 13
3.
ON/STANDBY Button, page 9
4. Low Battery Indicator, page 12 15. Fast Response Mode Indicator, page 13
5. Waveform Display, page 10 16. Softkeys, page 10
6. SatSeconds
7. %SpO
8. Pulse Rate Display, page 12 19. Motion Indicator, page 12
9. Alarm Silence Indicator, page 12 20. Pulse Search Indicator, page 13
10. ALARM SILENCE Button, page 9 21. Speaker
11. ADJUST UP Button,page 10
TM
Indicator, page 13
2 Display, page 12 18. Data In Sensor Indicator, page 13
12. ADJUST DOWN Button, page 10
14. CONTRAST Button, page 10
17. Menu Bar, page 10
Figure 1: Front Panel Buttons and Symbols
N-595 7
Description of Controls, Indicators, and Symbols

Identification of Rear Panel Components

1. Equipotential Terminal (Ground) 4. Fuse Holder
2. AC Power Connector, page 17 5. Supply Voltage Selector Switch, page 17
3. Data Port Connector, page 91
Figure 2: Rear Panel Components
N-595 Symbols
8
The symbols that are located on the rear panel of the N-595 are as follows:
See Instructions for Use
Fuse Replacement
Equipotential Terminal (ground)
Date of Manufacture
Data Interface
There is one symbol located on the front panel of the N-595.
Type BF Applied Part - Not defibrillator proof.

Description of Controls

Description of Controls, Indicators, and Symbols
Note: A button press, except the ON/STANDBY button, should
result in either a valid or an invalid key tone (refer to Table 1 on page 14). If the key pressed fails to emit a tone, contact qualified service personnel.
The ON/ST ANDBY button. Used to turn the N-595 monitor on or off.
The ALARM SILENCE button. Used to silence current alarms for the alarm silence duration period. When an alarm has been silenced, pressing the button again reactivates, or “unsilences” the alarm. It is also used to view and adjust alarm silence duration and alarm volume.
N-595 9
Description of Controls, Indicators, and Symbols
The ALARM SILENCE button clears “SENSOR OFF,” “LOW BATTERY,” and “SENSOR DISCONNECT” messages from the display.
The ADJUST UP button. Used to increase variable parameters of the monitor.
The ADJUST DOWN button. Used to decrease variable parameters in the monitor.
The CONTRAST button. Used in conjunction with the ADJUST UP and ADJUST DOWN buttons to lighten or darken the display screen.
The softkey buttons have multiple uses depending on the legend displayed above the button.

Description of Displays and Indicators

The type of display is user selectable. Refer to Selecting the Pleth View on page 34.
10
The pleth display includes a “wiper bar” plethysmographic waveform, menu bar, and current measured %SpO
2 and pulse rate. If
SatSeconds are enabled, the pleth display includes the SatSeconds indicator and SatSeconds setting. A decimal point after the %SpO
2 or
Description of Controls, Indicators, and Symbols
pulse rate indicate that the respective limits have been changed from the power on defaults (Monitor Trend Data on page 53).
The blip display includes a pulse amplitude blip bar, current measured %SpO
2 and pulse rate, and current upper and lower %SpO2 and pulse
rate limits. If SatSeconds are enabled, the blip display includes the SatSeconds indicator and SatSeconds setting. Decimal points after the %SpO
2 or pulse rate indicate that the respective limits have been
changed from the power-on defaults.
There are various matrixes within the O
XIMAX algorithm. Some of
these, are used to assess the severity of conditions presented to the N-595 in measuring SpO
2 and pulse rate on a patient. These
individual matrices or combinations of these matrices are used to drive the LED indicators on the N-595 front panel.
The OxiMax algorithm automatically extends the amount of data required for measuring SpO2 and pulse rate depending on the measurement conditions. During normal measurement conditions the averaging time is 6-7 seconds. During challenging measurement conditions which could be caused by low perfusion, motion, external interference like ambient light, or a combination of these, the OXIMAX algorithm automatically extends the amount of data required beyond 7 seconds. If the resulting dynamic averaging time exceeds 20 seconds, the pulse search indicator is lit solid and SpO2 and Pulse Rate will continue to be updated every second. As these conditions become even more challenging, the amount of data required continues to extend. If the dynamic averaging time reaches 40 seconds, the pulse search indicator begins flashing, the SpO2 and pulse rate displays flash zeros indicating a loss-of-pulse condition.
N-595 11
Description of Controls, Indicators, and Symbols
WARNING: Failure to cover the OXIMAX sensor site with opaque material in high ambient light conditions may result in inaccurate measurements.
The %SpO
2 Display. Shows the hemoglobin oxygen saturation level.
The display value flashes zeros during loss-of-pulse alarms and flashes the SpO
2 value when the SpO2 is outside the alarm limits.
During Pulse Search, the monitor continues to update the display. If alarm limits have been changed from their power-on defaults, a decimal point (.) is displayed after the SpO
2 value (81.).
The Pulse Amplitude Indicator (blip bar). Indicates pulse beat and shows the relative pulse amplitude. As the detected pulse becomes stronger, more bars light with each pulse. This indicator is available only in the blip view.
The Pulse Rate Display. Shows the pulse rate in beats per minute. It flashes during loss-of-pulse alarms and when the pulse rate is outside of the alarm limits. During Pulse Search, the monitor continues to update the display. Pulse rates outside of the pulse rate range (20 to 250 bpm) are displayed as the closest value within the range. If alarm limits have been changed from their power-on defaults, a decimal point (.) is displayed after the BPM value (112.).
The AC Power Indicator. Lights continuously when the N-595 is connected to AC power. It also indicates that the battery is charging. It is off when the monitor is being powered by internal battery.
12
The Low Battery Indicator. Lights continuously when 15 or fewer minutes of battery capacity remain. Flashes when the battery capacity reaches critical condition.
The Alarm Silence Indicator. Lights continuously when an audible alarm has been silenced. It flashes when the alarm silence duration has been set to Off.
The Motion Indicator. The motion indicator is lit whenever the O
XIMAX algorithm detects the presence of artifacts
its severity or the impact on the SpO
2 or pulse rate values. When the
1
independent of
Description of Controls, Indicators, and Symbols
motion indicator and the pulse search indicator are simultaneously lit, it is an indication that the artifact is significant and/or has been persistent.
The Pulse Search Indicator. Lights continuo usly prior to initial acquisition of a pulse signal and during prolonged and challenging monitoring conditions. The pulse search indicator flashes during a loss-of-pulse signal.
The Data In-Sensor Indicator. Lights to indicate that the attached
O
XIMAX sensor contains a patient sensor event record. The sensor
event record information may be viewed or printed.
The SatSeconds Indicator. Fills in clockwise as the SatSeconds alarm management system detects a %SpO setting. Empties in counterclockwise direction when %SpO
2 reading outside of the limit
2 reading
is within limits. When the indicator is full, a medium priority alarm will sound.
The Fast Response Mode Indicator. The response mode setting dictates the response time (2 to 4 seconds in fast mode and 4 to 7 seconds in normal mode) applied by the O calculation of SpO
2. The OXIMAX algorithm’s calculation of pulse rate
XIMAX algorithm in its
is unaffected by the response mode setting. The trending interval (2-seconds or 4-seconds) is updated automatically by the monitor to roughly correspond with the SpO
2 calculation response time.
The Neonate Alarm Limits Indicator. This symbol is displayed when the alarm limits are set to neonate. No symbol is displayed when the monitor is set to adult limits.
1
Artifacts are events contained in the in-sensor data.
N-595 13
Description of Controls, Indicators, and Symbols

Description of Audible Indicators

Table 1 identifies the audible indicators of the N-595 indicators. Table 1: Audible Indicators
Function Description
Alarm Silence Reminder
Confirmation Tone Three beeps sound to indicate that default
Invalid Button Press
Valid Button Press
High Priority Alarm
Medium Priority Alarm
Low Priority Alarm
Three beeps that sound approximately every 3 minutes when alarms are silenced with the alarm silence duration set to Off and the alarm silence reminder function is enabled.
settings have been saved or reset to factory defaults or trend data has been deleted.
A short, low button has been pressed that is not appropriate for the current state of the monitor.
A short, medium an appropriate button has been pressed.
A high loss
-of-pulse.
A medium SpO
2 or pulse rate limit violation.
A low
-pitched tone indicating that a
-pitched tone indicating that
-pitched, fast-pulsing tone indicating
-pitched, pulsing tone indicating an
-pitched, slow-pulsing tone indicating an
OXIMAX sensor disconnect, low battery, or
monitor failure.
Power
-On Self-Test
Pass
A 1-second tone indicating that the N-595 has been turned on and has successfully completed the power
-on self-test.
14
Pulse Beep A single beep sounds for each detected pulse.
Volume Setting Tone A continuous tone that is used when adjusting
The pitch of the pulse beep signal changes with
-by-point rise or fall in the saturation
a point level.
the alarm volume.

Setting up the Monitor

WARNING: To ensure patient safety, do not place the pulse oximeter in any position that might cause it to fall on the patient.
WARNING: As with all medical equipment, carefully route patient cabling to reduce the possibility of patient entanglement or strangulation.
WARNING: Ensure that the speaker is clear of any obstruction. Failure to do so could result in an inaudible alarm tone.
WARNING: Disconnect the N-595 and Nellcor OXIMAX se nsor from the patient during magnetic resonance imaging (MRI) scanning. Induced current could potentially cause burns.
WARNING: To ensure accurate performance and prevent device failure, do not subject the N-595 to extreme moisture, such as direct exposure to rain. Such exposure may cause inaccurate performance or device failure.
WARNING: Do not use an N-595 pulse oximeter, OXIMAX sensor, cables, or connectors that appear damaged.
WARNING: Do not lift the pulse oximeter by the pulse oxime try cable or power cord because the cable or cord could disconnect from the pulse oximeter, causing the pulse oximeter to drop on the patient.
N-595 15
Setting up the Monitor
WARNING: The N-595 is not defibrillator-proof. However, it may remain attached to the patient during defibrillation or while an electrosurgical unit is in use, but the readings may be inaccurate during the defibrillation and shortly thereafter.
WARNING: In the USA, do not connect the pulse oximeter to an electrical outlet controlled by a wall switch, because the pulse oximeter may be accidentally turned off.
WARNING: Use only the Nellcor pulse oximetry cable DOC-10 with the N-595 pulse oximeter. Use of another pulse oximetry cable will have an adverse effect on performance. Do not attach any cable that is intended for computer use to the O
XIMAX sensor
port. Do not connect any device other than a Nellcor -approved OXIMAX sensor to the OXIMAX sensor connector.
16
WARNING: The N-595 should not be used adjacent to or stacked with other equipment. If adjacent or stacked use is necessary, the N-595 should be observed to verify normal operation in the configuration it is to be used.

List of Components

Setting up the Monitor
Quantity Item
1 1 Nellcor O
1 1
1 Power Cord (applicable to country of sale) 2
1 Sensor Accuracy Grid 1 Quick Guide
N
DOC N
country of sale) and/or Compact Disk
Fuses, 0.5 A, 250 volts, slow 20 mm)
Connecting the N-595 to AC Power
WARNING: In the USA, do not connect the pulse oximeter to an electrical outlet controlled by a wall switch, because the pulse oximeter may be accidentally turned off.
-595 Pulse Oximeter
XIMAX Sensor or Assortment Pack
-10 Pulse Oximetry Cable
-595 Operator’s Manual (applicable to
-blow, IEC (5 x
Caution: The SUPPLY VOLTAGE SELECTOR switch must be set to the correct voltage (1 15 or 230) to avoid equipment dama ge and ensure battery charging.
N-595 17
Setting up the Monitor
Caution: Use only the hospital-grade power cord provided by Nellcor.
1. Power Connector 2. Supply Voltage Selector
1. Set the SUPPLY VOLTAGE SELECTOR (2) switch to the applicable voltage.
2. Plug the female connector end of the power cord into the N-595 POWER CONNECTOR (1) on the rear of the monitor.
3. Plug the male connector of the power cord into a properly grounded AC outlet.
4. Verify that the monitor’s AC POWER INDICATOR is lit.
Note: If the AC POWER INDICATOR is not lit, check:
the power cord
the SUPPLY VOLTAGE SELECTOR switch
•the user-accessible fuses
the AC power outlet
18
Connecting an OXIMAX Sensor to the N-595
The OXIMAX sensor type is displayed at the bottom of the display when an O completes POST with an O
XIMAX sensor is connected to the N-595 or when the N-595
XIMAX sensor attached.
Setting up the Monitor
Caution: Use only Nellcor-approved O
XIMAX sensors and pulse
oximetry cables.
Note: Physiological conditions, medical procedures, or external
agents that may interfere with the monitor’s ability to detect and display measurements include dysfunctional hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring agents, such as nail polish, dye, or pigmented cream.
1. SpO2 O
XIMAX Sensor Port
1. Connect a DOC-10 pulse oximetry cable to the SpO
2 OXIMAX
sensor port (1) of the monitor.
2. Connect a Nellcor O
XIMAX SpO2 sensor to the other end of the
DOC-10 pulse oximetry cable.
N-595 19

Battery Operation

WARNING: Dispose of battery in accordance with local requirements and regulations.
Operating the N-595 on Battery Power
The N-595 monitor has an internal battery that can be used to power the monitor during transport or when AC power is not available. A new, fully charged battery will provide at least 2 hours of monitoring time under the following conditions:
No audible alarms sound
No analog or serial output devices are attached to the N-595
The monitor cannot operate with a fully discharged battery. Before attempting to turn on an N-595 monitor whose battery charge has been completely depleted, first plug the monitor into an AC outlet to allow the battery to charge for a few minutes. The monitor may then be powered on.
To charge a low or dead battery, connect the monitor to AC power. A full charge of a dead battery takes 14 hours while the monitor is turned off. A full charge of a dead battery takes 18 hours while the monitor is in operation (monitoring a patient).
When all of the following conditions are present for 15 minutes, the N-595 will automatically shut down:
Monitor is running on battery power
No buttons have been pressed
N-595 21
Battery Operation
Note: Whenever the monitor is connected to AC power, the battery is

Low Battery Indicator

The Low Battery Indicator lights and a low priority alarm begins to sound when approximately 15 minutes of monitoring time is available on the existing battery charge. Refer to Table 2 for a description of the low and critical battery conditions.
No pulse has been detected (for example, when a patient is not connected to the O
XIMAX sensor or the OXIMAX
sensor is disconnected from the monitor)
No alarms are present (other than low battery or a non-correctable error)
being charged. Therefore, it is recommended that the monitor remain connected to AC power when not in use. This will ensure a fully charged battery whenever it is needed.
22
If the monitor is not on AC power, a low battery audible alarm can be canceled by pressing the ALARM SILENCE button. The low battery indicator and display screen message will continue to be displayed. Plugging the monitor into AC power will silence the audible alarm, but the low battery indicator will stay lit as long as the battery is in the low voltage condition. After the 15-minute period of low battery condition, a high priority alarm will sound for about 10 seconds before the monitor shuts off.
If the monitor backlight is turned off during a low battery condition, the backlight cannot be turned back on.
It is recommended that qualified service personnel replace the internal battery every 24 months. Replaced batteries should be disposed of in accordance with local ordinances.
Battery Operation
Caution: If the N-595 pulse oximeter is to be stored for a period of 3 months or longer , notify service personnel to r emove the battery from the pulse oximeter prior to storage. Recharge the battery when the battery has not been charged for 2 or more months.
Caution: The pulse oximeter default settings will return to factory default setting if the battery becomes fully discharged or is replaced. Qualified service personnel will have to reset the institutional defaults, following the instructions in the service manual.
Note: If the AC voltage selector switch on the monitor rear panel
does not match your AC voltage source, the monitor may run on battery power, even though it is plugged into AC power, which will eventually result in a low priority alarm and a lighted low battery indicator. Ensure that the switch setting matches your AC voltage.
Note: As the battery is used and recharged over a period of time, the
amount of time between the onset of the low battery alarm and the instrument shut-off may become shorter.
Table 2: Low Battery and Critical Battery
State
1No No Yes
N-595 23
Critical Battery
Low Battery
AC Operation
2- normal
SpO AC/Battery charge LED LOW BATTERY LED LOW BATTERY
message Audible alarm Error code Effect of ALARM
SILENCE key Shutdown
-off
-none
-N/A
-on
-off
-off
-normal
Battery Operation
Ta ble 2: Low Battery and Critical Battery
State
2No No No
3No Yes No
Critical Battery
Low Battery
AC Operation
SpO AC/Battery charge LED LOW BATTERY LED LOW BATTERY
message Audible alarm Error code Effect of ALARM
SILENCE key Shutdown SpO AC/Battery charge LED LOW BATTERY LED LOW BATTERY
message
2-normal
-off
-off
-off
-off
-none
-normal
- N/A
2-normal
-off
-on
-on
24
Audible alarm Error code Effect of ALARM
SILENCE key silences audio alarm, second press cancels LOW BATTERY message (LED) stays on until Low Battery Condition is corrected.
Shutdown
-low priority
-logged
-First press
-N/A
Table 2: Low Battery and Critical Battery
Battery Operation
State
4No Yes Yes
5 Not used
Critical Battery
Low Battery
AC Operation
SpO AC/Battery charge LED LOW BATTERY LED LOW BATTERY
message Audible alarm Error code Effect of ALARM
SILENCE key stays on)
Shutdown Note: Connecting AC
2-normal
-on
-on
-off
-off
-logged
-N/A (LED
-N/A
functions the same as ALARM SILENCE key in state 3.
6Yes Yes No
N-595 25
SpO
2-not displayed
AC/Battery charge LED LOW BATTERY LED
(flashing)
LOW BATTERY
message Audible alarm Error code
logged Effect of ALARM
SILENCE key Shutdown
-on
-high priority
-displayed and
-none
-after 10 seconds
-off
-on
Battery Operation
Ta ble 2: Low Battery and Critical Battery
State
7 Yes Yes Yes
Critical Battery
Low Battery
AC Operation
2-not displayed
SpO AC/Battery charge LED LOW BATTERY LED
(flashing)
LOW BATTERY
message Audible alarm Error code
logged Effect of ALARM
SILENCE key Shutdown
-on
-high priority
-displayed and
-N/A
-after 10 seconds
-on
-on
26

Using the Monitor

Introduction

The parameters of the N-595 monitor are preset to factory default settings. See Factory Defaults on page 139. The factory default parameters may be changed to institutional default parameters by following the procedures in the N-595 service manual.
Table 3 lists the parameters, ranges available, and the factory default setting. The parameters may be set on an individual basis, by the clinician, and these settings will remain in effect until the N-595 is turned off.
Table 3: Parameter Ranges
Parameter
%SpO2 Upper Alarm Limit
%SpO2 Lower Alarm Limit
Pulse Rate Upper Alarm Limit
Pulse Rate Lower Alarm Limit
Alarm Silence Duration
Alarm Volume 1 to 10 7 7
Ranges/ Selections
Lower Alarm Limit plus 1 to 100%
20% to Upper Alarm Limit minus 1
Lower Alarm Limit plus 1 to 250 bpm
30 bpm to Upper Alarm Limit minus 1
Alarms 30, 60, 90, 120 seconds
Factory Adult Defaults
100% 95%
85% 80%
170 bpm 190 bpm
40 bpm 90 bpm
60 60
Factory Neonate Defaults
N-595 27
Using the Monitor
Table 3: Parameter Ranges
Parameter
Alarms
Data Port Baud Rate
Data Port Mode ASCII, GRAPH,
Default Display Format
Default Trend Display
Ranges/ Selections
Allow Off Yes/No
Off Reminder ­Yes/No
2400, 9600, 19200
OXINET, CLINICAL, AGILENT (HP Agilent), SPACELBS, MARQ (GE Marquette), DATEX (Datex­Ohmeda)
Pleth, Blip Pleth Pleth
Saturation, Pulse Rate, Dual, Histogram
-
Factory Adult Defaults
Yes
Yes 9600 9600
ASCII ASCII
Saturation Saturation
Factory Neonate Defaults
Yes
Yes
28
Display Contrast
Language English, French,
Limits Adult, Neonate Adult Neonate Pulse Beep
Volu m e Response
Mode
Low to high Medium Medium
English English German, Dutch, Portuguese, Spanish, Italian, Swedish
0 to 10 4 4
Normal or Fast Normal Normal
Table 3: Parameter Ranges
Using the Monitor
Parameter
RS-232 Level Nurse Call Polarity
SatSeconds Off, 10, 25, 50,
Sensor Event Date Format (SENSOR-R and SENSOR-RW
Sensor Messages Enabled
Trend Display Dual, %SpO
Trend Scale 48, 36, 12,8, 4, 2,
Ranges/ Selections
Normally High, Normally Low
100 SpO
2,
SpO
2+Pulse Rate,
Default (default is factory default)
Yes, No Yes Yes
2,
Pulse, Histogram, Amplitude
1 hours, 30, 15 minutes, 40, 20 seconds
Factory Adult Defaults
Normally low Normally low
Off Off
Default Default
%SpO2 %SpO2
2 hours 2 hours
Factory Neonate Defaults

Turning On the Monitor

Before using the N-595 in a clinical setting, you must verify that the monitor is working properly and is safe to use. Proper working condition will be verified each time the N-595 is turned on as described in the following procedure.
Caution: If any indicator or display element does not light when the pulse oximeter is turned on, do not use the pulse oximeter. Instead, contact qualified service personnel, your local Nellcor representative, or Nellcor’s Technical Services Department.
N-595 29
Using the Monitor
Note: Physiological conditions, medical procedures, or external
agents that may interfere with the monitor’s ability to detect and display measurements, include dysfunctional hemoglobin, arterial dyes, low perfusion, dark pigment, and externally applied coloring agents such as nail polish, dye, or pigmented cream.
Note: The monitor automatically starts the Power-On Self-Test
(POST), which tests the monitor circuitry and functions.
Caution: During POST (immediately after power-up), confirm that all indicators light, all display segments turn on, and the pulse oximeter speaker sounds a one-second tone.
1. Turn on the N-595 by pressing the ON/STANDBY button.
2. Ensure that all of the front panel indicators illuminate.
30
3. Once the display test portion of POST is complete, the N-595 software version is displayed for approximately 5 seconds.
Note: The software version shown above is only a sample. Check
your monitor for the software version installed.
Software version numbers are often needed when calling Nellcor’s Technical Services Department or your local Nellcor representative for technical assistance. Write down the software version number and have it available prior to requesting technical assistance.
Using the Monitor
4. If the N-595 detects an internal problem during POST, an error
tone sounds and the monitor displays an error code (EEE) and the corresponding number (see Troubleshooting on page 117).
5. Upon successful completion of the POST, the N-595 sounds a
one-second tone indicating that the monitor has passed the test.
WARNING: If you do not hear the POST pass tone, do not use the pulse oximeter.
WARNING: Ensure that the speaker is clear of any obstructions. Failure to do so could result in an inaudible alarm tone.
Note: In addition to serving as the POST pass verification, the POST
pass tone also functions as an audible confirmation that the speaker is performing properly. If the speaker does not function, the alarm warning sounds cannot be heard.

OXIMAX Sensor Attached

When an OXIMAX sensor is attached to the monitor, a “DAT A TYPE: . . .” message is displayed briefly at the bottom of the monitor display. For a sensor containing data, the message identifies the sensor data type, For a blank sensor, the message identifies the monitor’s current
N-595 31
Using the Monitor
data type setting that will be used to write data to the sensor. The data type settings are SPO2 and SPO2+BPM.
Note: The type of data recorded is only displayed when data is resent
in the O
XIMAX sensor.
The monitor displays zeros in the %SpO
2 and Pulse Rate displays
while the N-595 is searching for a valid pulse. For optimal performance, allow the monitor to search and lock onto a pulse for approximately 10 seconds in non-motion conditions.
When a valid pulse is detected, the monitor enters the Monitoring Mode and displays patient parameters.
32
Look for movement of the blip bar or of the plethysmographic waveform indicating that the monitor is displaying real-time data. Listen for the pulse beep tone. If the pulse beep tone does not sound with each pulse, it is an indication that the pulse beep volume is set to zero, the speaker is malfunctioning, or the signal is corrupted.
When an O
XIMAX sensor is attached to the monitor and is applied to a
patient, if the monitor loses the pulse signal, the monitor will display “--- & ---” (3 dashes and 3 dashes) and remain in Pulse Search Mode for 5 seconds before displaying the poor signal condition screen. The poor signal condition screen is part of the N-595's Sensor Messages
feature. For more information about OXIMAX Sensor Messages, refer OXIMAX Sensor Messages on page 49.

No OXIMAX Sensor Attached

Upon successful completion of the POST, the N-595 monitor sounds a one-second tone indicating that the monitor has passed POST.
Using the Monitor
or
The monitor displays dashes (---) and the Pulse Search indicator is not lit, indicating that the monitor failed to detect an O
N-595 33
XIMAX sensor.
Using the Monitor

Turning the Backlight On or Off

Note: When the backlight is off, any of the following conditions will
turn on the backlight:
pressing any of the softkeys
pressing the CONTRAST button
pressing the ALARM SILENCE button
•any alarm
LIGHT
With the monitor in the normal monitoring mode, press the LIGHT softkey.

Adjusting Screen Contrast

With the monitor in the normal monitoring mode:
1. Press the CONTRAST button.
2. Press the ADJUST UP or ADJUST DOWN button until the desired contrast is obtained.
3. Press the CONTRAST button.

Selecting the Pleth View

34
The pleth view displays the pleth waveform, %SpO2, and pulse rate data. Refer to Principles of Operation on page 143, for a description of the pleth waveform.
With the monitor in the normal monitoring mode:
Using the Monitor
SETUP
VIEW
PLETH
1. Press the SETUP softkey.
2. Press the VIEW softkey.
3. Press the PLETH softkey.

Selecting the Blip View

Displays SpO2, pulse rate, blip bar, and limits in a larger format for easier viewing.
With the monitor in the normal monitoring mode:
SETUP
VIEW
N-595 35
1. Press the SETUP softkey.
2. Press the VIEW softkey.
Using the Monitor
BLIP
3. Press the BLIP softkey.

Setting the Pulse Beep Volume

With the monitor in the normal monitoring mode:
1. Press and hold the ADJUST UP/ADJUST DOWN button to increase/decrease pulse beep volume.

Setting the Alarm Volume

The Alarm Volume display allows the user to adjust the volume of alarm tones.
With the monitor in the normal monitoring mode:
36
1. Press the ALARM SILENCE button until the alarm volume level
is displayed and sounds on the monitor.
2. While continuing to press the ALARM SILENCE button, press
and hold the ADJUST UP/ADJUST DOWN button to increase/ decrease the volume.

Setting the Date and Time

Using the Monitor
WARNING: The sensor extrapolates from the date and time provided by the N-595 when recording the sensor event record to the sensor. The accuracy of the date/time is the responsibility of the N-595. It is recommended that the N-595 user set the time/ date to the correct value before a sensor event record-enabled sensor is connected, and that this date/time not be changed while the sensor remains connected. Since a sensor with sensor event record data can be transported from one monitor to another, having discrepancies in the date/time between monitors and the sensor event record data will affect the order the sensor event record data appears. To eliminate this possible problem, all monitors within an institution should be set to the same time.
With the monitor in the normal monitoring mode:
SETUP
N-595 37
1. Press the SETUP softkey.
Using the Monitor
NEXT
CLOCK
SET
2. Press the NEXT softkey.
3. Press the CLOCK softkey.
4. Press the SET softkey.
5. Press the SELECT softkey to select:
TIME HOURS : MINUTES : SECONDS (16:46:05)
DATE DAY - MONTH - YEAR (30-JAN-02)
6. Use the ADJUST UP or ADJUST DOWN buttons to change the selected value.
EXIT
7. Press the EXIT softkey.

Setting Alarm Silence Duration

The Alarm Silence Duration display allows the user to adjust the alarm silence duration.
With the monitor in the normal monitoring mode:
38
Using the Monitor
1. Press the ALARM SILENCE button until the alarm silence
duration setting is displayed. Alarm silence durations that are available are OFF, 30, 60, 90, and 120 seconds.
2. Press and hold the ALARM SILENCE button and the ADJUST
UP button to increase the alarm silence duration setting.
3. Press and hold the ALARM SILENCE button and the ADJUST
DOWN button to decrease the alarm silence duration setting.
Note: Releasing the ADJUST UP or ADJUST DOWN button sets the
alarm silence duration.

Disabling Audible Alarms

Setting the alarm silence duration to OFF means that the monitor will produce no audible alarms.
Note: The ability to set the alarm silence duration to OFF can be
enabled or disabled by qualified service personnel as described in the service manual. The current copy of the service manual is available on the Internet at:
http://www.mallinckrodt.com/respiratory/resp/Serv_SuppProductManuals.html
N-595 39
Using the Monitor
WARNING: Do not silence the audible alarm function or decrease the audible alarm volume if patient safety could be compromised.
With the monitor in the normal monitoring mode:
1. Press the ALARM SILENCE button until the alarm silence duration setting is displayed.
40
2. While pressing the ALARM SILENCE button, press and hold the ADJUST UP button until OFF is displayed. Release the buttons.

Selecting Standby Mode

The standby mode allows the monitor to retain the alarm limit settings that are in effect while monitoring a patient. The monitor must be powered by AC power to enter the standby mode.
Usually the standby mode is used when a patient has to leave the monitor for a period of time and will return to the same monitor.
To place a monitor in the standby mode:
1. The monitor should be monitoring a patient.
2. The monitor alarm limits should be configured to the patient
being monitored.
Using the Monitor
3. Disconnect the sensor from the monitor.
4. Press the ALARM SILENCE button. This silences the audible
alarms.
5. Press the ALARM SILENCE button. This disables the alarm
messages.
N-595 41
Using the Monitor
The monitor is now in standby. To return to normal monitoring, connect the sensor to the monitor and the patient.
Adult-Pediatric or Neonatal Settings
The clinician can set the monitor’s operating mode to adult-pediatric or neonatal by using the LIMITS softkey. The setting will only remain in the monitor until the monitor is turned off. The factory default power-on setting is for adult-pediatric patients. This default setting can be changed to neonatal by qualified service personnel using the procedures indicated in the service manual.
Refer to Table 11 on page 139, for neonate factory default limit settings. Refer to T able 12 on page 140, for adult factory default limit settings.
WARNING: Each time the pulse oximeter is used, check alarm limits to ensure that they are appropriate for the patient being monitored.
Setting Patient Adult-Pediatric/Neonatal Mode
With the monitor in the normal monitoring mode:
42
Using the Monitor
LIMITS
1. Press the LIMITS softkey.
2. The monitor will display the ADULT LIMITS or NEONATE
LIMITS screen, depending on the patient setting being used.
ADULT or NEO
3. Press the NEO or ADULT softkey to select ADULT LIMITS or
NEONATE LIMITS as applicable for the patient being monitored.
N-595 43
Using the Monitor

Alarm Limit Changed Indicator

Alarm limits that have been changed from the institutional or factory default settings are identified by a decimal point (.) after the displayed reading (%SpO
2 or BPM). The changed parameter is also identified
by a decimal point on the alarm limits screen.

Setting Alarm Limits

44
The Alarm Limit display allows the user to adjust the upper and lower saturation and pulse rate limits. It also allows the user to adjust the SatSeconds limit.
The Alarm Limit display is accessed by pressing the LIMITS softkey on the Main menu.
The Alarm Limit display includes the alarm limit table and current measured %SpO
2 and pulse rate. The title of the alarm limit table will
indicate whether the instrument is in Adult or Neonate monitoring mode. If SatSeconds are enabled, the Alarm Limit display also includes the SatSeconds indicator. Decimal points after the displayed %SpO
2 or pulse rate indicate that the respective limits have been
changed from the power-on defaults.
With the monitor in the normal monitoring mode:
Using the Monitor
LIMITS
ADULT or NEO
1. Press the LIMITS softkey. Current alarm limits are displayed.
or
2. Press the ADULT or NEO softkey to select Adult-Pediatric or
Neonatal alarm limits screen.
SELECT
3. Press the SELECT softkey as required to select the parameter to
be adjusted.
4. Use the ADJUST UP or ADJUST DOWN buttons to increase or
decrease the selected limit parameter.
5. Repeat steps 3, 4, and 5 as necessary to complete the alarm limits
setup.
EXIT
6. To accept the changes, let the display time-out or press the EXIT
softkey to exit the display and return to normal monitoring.
N-595 45
Using the Monitor
Note: Limit changes will only be in effect as long as the monitor
remains turned on. When the monitor is turned off, the institutional or factory default limits will be restored into the monitor. When the monitor is turned on, the institutional or factory default limits will be in effect. Factory or institutional defaults are selected by qualified service personnel following the procedure in the service manual.

Setting SatSeconds Alarm Limit

Refer to Describing SatSeconds on page 135, for a description of the SatSeconds function.
With the monitor in the normal monitoring mode:
LIMITS
SELECT
EXIT
1. Press the LIMITS softkey. Current alarm limits are displayed.
2. Press the SELECT softkey twice to select %SpO
2 SAT-S.
3. Use the AJDUST UP or ADJUST DOWN buttons to select the limit. The choices are 10, 25, 50, or 100 seconds or OFF.
4. Press the EXIT softkey to save your choice.
46

Setting Monitor Response Mode

The purpose of the response mode is to set the response time of the
XIMAX algorithm calculation of the SpO2 (the response mode does
O not affect the O trending interval (2- or 4-seconds) is updated automatically by the monitor to roughly correspond with the SpO time.
XIMAX algorithm’s calculation of pulse rate). The
Using the Monitor
2 calculation response
SETUP
NEXT
NEXT
NEXT
The response mode programs the O
XIMAX algorithm to display
monitor trend information at 2-second intervals (Fast Mode) or 4-second intervals (Normal Mode).
The response mode display screen includes the current SpO mode setting and the current measured %SpO
2 and pulse rate. When
2 response
in the fast mode, the screen displays the fast mode symbol.
With the monitor in the normal monitoring mode:
1. Press the SETUP softkey.
2. Press the NEXT softkey.
3. Press the NEXT softkey.
4. Press the NEXT softkey.
MODE
5. Press the MODE softkey.
Note: When the monitor is in the fast response mode the monitor may
produce more SpO
N-595 47
2 and pulse rate alarms than the user is
Using the Monitor
accustomed to seeing, and may be inappropriate in challenging measurement conditions.
6. Use the ADJUST UP or ADJUST DOWN buttons to select the desired response mode.
EXIT
7. Press the EXIT softkey.

Selecting the Display Language

The N-595 can be programmed to display the information in various languages. The languages available are English, Francais (French), Deutsch (German), Italiano (Italian), Espanol (Spanish), Nederlands (Dutch), Port (Portuguese) and Sverige (Swedish).
With the monitor in the normal monitoring mode:
SETUP
NEXT
1. Press the SETUP softkey.
2. Press the NEXT softkey.
48
Using the Monitor
LANG
EXIT
3. Press the LANG softkey.
4. Use the ADJUST UP or ADJUST DOWN buttons to select the
desired language.
5. Press the EXIT softkey.
Note: The selected language will be displayed until the monitor is
turned off. The selected language can be set as a default by qualified service personnel following the procedures in the service manual.

OXIMAX Sensor Messages

OXIMAX sensor messages consist of sensor adjust condition messages and sensor adjust messages which, when enabled, are displayed when the monitor is not able to display saturation. When O messages are displayed, it is an indication that the O functioning correctly , but the site to which the O or the application method is not optimal for calculating %SpO Condition messages are followed by action messages. Up to three condition messages may be displayed on the “POOR SIGNAL CONDITION” display in priority order, highest on top. The condition display may be dismissed by using the EXIT softkey. Once exited, the
N-595 49
XIMAX sensor
XIMAX sensor is
XIMAX sensor applies
2.
Using the Monitor
OXIMAX sensor message screen will not return until a new condition occurs.
If the HELP softkey is pressed from the Condition message display, the action messages are displayed. Action messages are linked to the sensor type; action messages will be displayed for the type of O
XIMAX
sensor connected to the monitor. Up to five action messages may be displayed. Multiple screens may be required to display all of the messages. When multiple screens are required, navigation between screens can be accomplished through the NEXT, BACK, and EXIT softkeys.
O
XIMAX sensor messages may be disabled. Refer to OXIMAX Sensor
Message Setup on page 127 for selecting the O
Messages, Enable/Disable function.

OXIMAX Sensor Adjust Condition Messages

Condition 1 — SENSOR OFF?
Condition 2 — SMALL PULSES
Condition 3 — WEAK SIGNAL
XIMAX Sensor
50
Condition 4 — MOTION INTERFERENCE
Condition 5 — EXCESS INFRARED LIGHT
Condition 6 — ELECTRICAL/LIGHT
Condition 7 — HIGH PULSE AMPLITUDE

OXIMAX Sensor Adjust Messages

Message 1 — ALTERNATE SITE?
Message 2 — COVER SENSOR SITE?
Message 3 — EAR/FOREHEAD SENSOR?
Message 4 — NASAL/EAR SENSOR?
Using the Monitor
INTEREFERENCE
Message 5 — O
XIMAX ADHESIVE SENSOR
Message 6 — SECURE CABLE
Message 7 — HEADBAND
Message 8 — WARM SITE
Message 9 — BANFAGE ASSEMBLY
Message 10 — NAIL POLISH
Message 11 — SENSOR TOO TIGHT?
Message 12 — REPOSITION SENSOR
Message 13 — ISOLATE INTERFERENCE SOURCE
Message 14 — CLEAN SENSOR SITE
N-595 51

Monitor Trend

Monitor Trend Data

The trend displays allow the user to view trend data. Two types of trend data can be viewed:
Monitor trend data which are stored in the monitor
Patient event data which are stored in the O (single-patient-use O with the sensor event record feature.
Monitor trend data can be viewed anytime patient trend is stored in the monitor. Monitor trend displays are accessed by pressing the TREND softkey on the main menu and selecting the MONITR softkey option. The monitor trend sub-menu allows you to choose which trend data are displayed:
Saturation and pulse rate (Dual)
Saturation
•Pulse rate
Pulse amplitude
Histogram
The N-595 can graphically display trend data for SpO both. Trend data is stored at 2- or 4-second intervals. When the TREND softkey is pressed, “READING TRENDS . . .” is displayed at the bottom of the N-595 screen, indicating that the monitor is formatting the trend data to be displayed.
XIMAX sensors only) and can be used
XIMAX sensor
2, pulse rate, or
N-595 53
Monitor Trend
The monitor stores up to 48 hours of 4-second trend data or 24 hours of 2-second trend data. The amount of trend data displayed on the screen is determined by using the ZOOM softkey. The settings available are 20 and 40 seconds, 15 or 30 minutes, and 1, 2, 4, 8, 12, 24, 36, or 48 hours. All trend data are displayed in a graphical format except the 20- and 40-second trend displays, which are shown in tabular format.
The trend display is scrolled, that is, the data displayed can be moved throughout the 48 hours of trend data. Selecting the 1-hour trend display allows you to view one hour of trend information. By using the scrolling feature, any one hour of trend data can be viewed over the 48 hours of trend information. The AJUST DOWN button scrolls the display to the left and the ADJUST UP button scrolls the display to the right.
When the data are displayed, the most recent readings are on the right side of the graph. The numbers below %SpO
2 indicate the highest and
lowest parameter values at the cursor position (vertical dotted line on the display). See Table 4 on page 57.
.
Trend data is further explained in Specifications on page 147.
Trend data information may be retrieved through the N-595 data port or cleared using options available in a display menu.
Caution: Monitor trend data will be lost if the main battery fails or is removed.
54

Trend Data Operation

Whenever the N-595 is turned on, it stores the monitor %SpO2 and pulse rate readings in memory every 2 or 4 seconds (regardless of whether the N-595 is monitoring a patient or not). The N-595 can store up to 48 hours of 4-second trend data or 24 hours of 2-second trend data. The 48/24 hours of stored trend data are available for downloading to a printer or a portable computer . Up to 50 alarm limit changes can be stored in the trend data. If more than 50 alarm limit changes occur during the 48/24 hours of trend data collection, the additional alarm limit changes will take space reserved for trend data.
Caution: Changing alarm limit settings uses up trend memory space. Change alarm limits only as needed.
Note: Trend memory always contains the most recent 48 hours of
Monitor Trend
data, with newly collected data overwriting the oldest data on a rolling basis. The N-595 continues to record data points as long as the monitor is powered on, with “blank” data points collected if no O
XIMAX sensor is connected to the monitor or
patient. “Blank” data will over-write older patient data if the memory becomes full. Therefore, if you want to save old patient data, it is important that you turn your monitor of f when you are not monitoring a patient, and that you download the trend memory before it fills up and over-writes the old data with new data (or “blank” data).

Selecting the Trend Data Display Scale

The trend scale is the amount of trend data displayed on the screen.
With the monitor in the normal monitoring mode:
TREND
MONITR
N-595 55
1. Press the TREND softkey.
2. Press the MONITR softkey.
Monitor Trend
VIEW
ZOOM
TIME
3. Press the VIEW softkey.
4. Press any of the trend softkeys (DUAL, SPO2, or PULSE). To select HIST (histogram) or AMP (amplitude), press the NEXT softkey and then the HIST or AMP softkeys.
5. Press the ZOOM softkey. The Zoom menu is displayed.
Pressing the TIME softkey cycles the displayed trend time scale through 48 hours, 36 hours, 12 hours, 8 hours, 4 hours, 2 hours, 1 hours, 30 minutes, 15 minutes, 40 seconds and 20 seconds.
Note: The 20-second and 40-second trend displays are in tabular
format. The below display starts out in the normal response mode (left side of the display) and switches to the fast response mode.
56
SCALE
Pressing the SCALE softkey cycles the displayed trend amplitude scale through ±5 points, ±10 points, ±15 points, ±20 points, ±25 points, ±30 points, ±35 points, ±40 points and ±50 points above and below the data point under the cursor. The saturation graphical monitor trend display vertical scale default setting is from 10 to 100 if there is no data under the cursor. The pulse rate graphical monitor
Monitor Trend
trend display vertical scale is from 5 to 250 if there is no data under the cursor.
AUTO
Pressing the AUTO softkey presets the amplitude of the graphed trend data. The maximum trend data point is rounded up to the nearest multiple of 10, this value is the top of the graph display. The minimum trend data point is rounded down to the next multiple of 10. Then 10 is subtracted from the rounded down number, this value is the bottom of the trend graph.
BACK
Pressing the BACK softkey returns the monitor to the Monitor menu.

Reading the Trend Data Display

Table 4 identifies the components of the trend data display.
Table 4: Reading Trend Display
Item Description
1 Amount of trend data displayed on the screen. Settings available
are 20 and 40 seconds, 15 and 30 minutes, 1, 2, 4, 8, 12, 24, 36,
and 48 hours. 2 Date represented by the cursor (item 5). 3 No trend data recorded during this time. 4 Time represented by the cursor (item 5).
N-595 57
Monitor Trend
Table 4: Reading Trend Display
Item Description
5
6 Highest and lowest reading at the cursor position. 7 Trend data that is being displayed (%SPO2, BPM, or PAU
Cursor (right) or ADJUST DOWN (left) buttons.
[pulse amplitude units]).

Dual Trend Data Display

The dual trend data display displays both oxygen saturation (%SpO2) levels and pulse rate (bpm) trend data.
With the monitor in the normal monitoring mode:
TREND
MONITR
1. Press the TREND softkey.
2. Press the MONITR softkey.
- can be moved left or right using the ADJUST UP
58
VIEW
DUAL
3. Press the VIEW softkey.
4. Press the DUAL softkey. The dual trend (%SpO is displayed.
2 and Pulse Rate)

SpO2 Trend Display

With the monitor in the normal monitoring mode:
Monitor Trend
TREND
MONITR
VIEW
SPO2
1. Press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
4. Press the SPO2 softkey. SpO

Pulse Rate Trend Display

2 trend data is displayed.
With the monitor in the normal monitoring mode:
TREND
MONITR
VIEW
N-595 59
1. Press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
Monitor Trend
PULSE
4. Press the PULSE softkey. The pulse trend data is displayed.

Histogram Trend Data Display

The histogram displays trend data for the percent of oxygen blood saturation (SpO the trend data stored over the period of time indicated on the display. Refer to Selecting the Trend Data Display Scale on page 55, to set up the desired trend data scale.
Pulse amplitude cannot be displayed on the histogram display.
2) and pulse rate (bpm). The data displayed represents
60
TREND
MONTIR
VIEW
NEXT
With the monitor in the normal monitoring mode:
1. Press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
4. Press the NEXT softkey.
Monitor Trend
HIST
5. Press the HIST softkey. The Histogram trend data is displayed.

Pulse Amplitude Trend Data Display

The pulse amplitude trend data display shows the amplitude of the patient’s pulse rate over the period of time indicated on the display. Refer to Selecting the Trend Data Display Scale on page 55, to setup the desired trend data scale.
With the monitor in the normal monitoring mode:
TREND
MONITR
VIEW
NEXT
N-595 61
1. Press the TREND softkey.
2. Press the MONITR softkey.
3. Press the VIEW softkey.
4. Press the NEXT softkey.
Monitor Trend
AMP
5. Press the AMP softkey. The pulse amplitude units (PAU) trend data is displayed.
The PAU reading (12 : 20) indicates the pulse amplitude units (upper and lower) at the cursor position (dashed line). The cursor is moved right or left using the ADJUST UP (right) and ADJUST DOWN (left) buttons.

Clearing Trend Information

With the monitor in the normal monitoring mode:
62
TREND
MONITR
NEXT
DELETE
YES
1. Press the TREND softkey.
2. Press the MONITR softkey.
3. Press the NEXT softkey.
4. Press the DELETE softkey.
Note: Press the NO softkey and then the EXIT softkey to leave this
function without deleting trend data.
5. Press the YES softkey.
Monitor Trend
All the trend data is cleared and the monitor sounds three beeps.
N-595 63

O XIM AX Sensor Event Record

WARNING: The sensor extrapolates from the date and time provided by the N-595 when recording the sensor event record to the sensor. The accuracy of the date/time is the responsibility of the N-595. It is recommended that the N-595 user set the time/ date to the correct value before a sensor event record-enabled sensor is connected, and that this date/time not be changed while the sensor remains connected. Since a sensor with sensor event record data can be transported from one monitor to another, having discrepancies in the date/time between monitors and the sensor event record data will affect the order the sensor event record data appears. To eliminate this possible problem, all monitors within an institution should be set to the same time.
The adhesive O
XIMAX sensors are capable of storing patient event
data. A sensor event record allows alarm event history to travel with the patient on the sensor’s memory chip for quick assessment at every point of care where O
Patient (event) data is stored on the memory chip of adhesive O sensors (single-patient-use O
XIMAX monitors are used.
XIMAX
XIMAX sensors only). The event data is
stored (recorded) with the limit/threshold settings that were active at the time of the event on the recording monitor. These events can be viewed on the next O
XIMAX sensor monitor when the patient moves to
a new point of care.
An event occurs when the %SpO lower alarm limit for at least 15 seconds. The first O event record event will be stored in the O
XIMAX sensor has been attached to a patient for five minutes and
O
2 value exceeds either the upper or
XIMAX sensor
XIMAX sensor after the
every five minutes thereafter. The maximum number of events that can be stored in an O
XIMAX sensor is 100.
N-595 65
OXIMAX Sensor Event Record
Event records can only be viewed after an OXIMAX sensor containing patient data (event records) has been connected to an O monitor. Event records are designed to view patient events from prior areas of care or transport (history) while monitor trend should be used to view data or events from a patient currently being monitored. The monitor’s SENSOR EVENT RECORD indicator will light when an
XIMAX sensor containing event data is connected to the OXIMAX
O
monitor.
Patient event data is accessed by pressing the TREND softkey on the main menu and selecting the SENSOR softkey option. Sensor event record can be viewed in graphical form (GRAPH) or in a summary table (TABLE).
XIMAX
Note: Once the O
XIMAX sensor and event data is stored in the OXIMAX sensor,
O the O
XIMAX sensor event record type cannot be reset. The
monitor’s type set up can be changed at any time.
Recording and viewing of O available on O
XIMAX comparable monitors. The OXIMAX sensors may
function on older technology monitors but the O record feature is not available.
Refer to the N-595 service manual for the procedure to disable the storage of sensor event record on an O
Setting In-Sensor Data Type
The In-Sensor Data Type display allows the user to set the type of trend data to be recorded in an O set to record either SpO
Note:The O
XIMAX sensor data type can only be set when an OXIMAX
sensor is not connected to the monitor.
XIMAX sensor event record type is set up in the
XIMAX sensor event record is only
XIMAX sensor event
XIMAX sensor.
XIMAX sensor. OXIMAX sensors can be
2 or SpO2+BPM.
66
With the monitor turned on and no cable attached to the SpO sensor port:
2 OXIMAX
OXIMAX Sensor Event Record
R
SETUP
SENSO
DATA
1. Press the SETUP softkey.
2. Press the SENSOR softkey.
3. Press the DATA softkey.
Note: O
XIMAX sensor data type settings are displayed on the monitor
as shown in the figure below (in-sensor data type). If no sensor is connected, both sensor types and the full set of options for each are displayed. If a sensor is connected, only the sensor data type for that sensor is displayed.
Note: The SENSOR-R feature supports all the current O
XIMAX
sensors. The SENSOR-RW feature is only applicable to
O
XIMAX sensors with a read/write chip installed.
SELECT
4. Use the SELECT softkey to toggle between SENSOR-R and
SENSOR-RW.
N-595 67
OXIMAX Sensor Event Record
5. Use the ADJUST UP or ADJUST DOWN button to select the
O
selections are:
XIMAX sensor data type. SENSOR-R and SENSOR-RW
EXIT
6. Press the EXIT softkey to set the O

OXIMAX Sensor Type

When an OXIMAX sensor is connected to the monitor, a “SENSOR TYPE: …” message is displayed for 4 to 6 seconds at the bottom of the display . The message identifies the type (model) of O connected to the monitor. Type is used in the determination of action messages in the O the first message displayed when an O the monitor.
•SpO
2
•SpO2+BPM
•DEFAULT
XIMAX sensor message(s) function. This display is
XIMAX sensor type.
XIMAX sensor
XIMAX sensor is connected to

OXIMAX Sensor Data Type

When an OXIMAX sensor with no previously recorded patient data is connected to the O displayed briefly at the bottom of the display, this message is
68
XIMAX monitor, a “DATA TYPE: . . .” message is
OXIMAX Sensor Event Record
displayed after the OXIMAX sensor type message. The message identifies the monitor's current data type setting that will be used to write data to the O EVENT/SPO
XIMAX sensor. The data type setting options are
2 and EVENT/SPO2+BPM.
The user can change the setting by referring to Setting In-Sensor Data Type on page 66. The O prior to connecting the O
XIMAX sensor event record type must be set
XIMAX sensor to the monitor.

OXIMAX Sensor Event Record Data Available

When an OXIMAX sensor containing data (single-patient-use OXIMAX sensors only) is connected to the monitor, the Sensor Event Record indicator on the monitor front panel blinks at a medium priority flash
TREND
rate to indicate that the O
XIMAX sensor attached to the monitor
contains patient event data. The LED blinks for approximately 60
SENSOR
seconds or until the O
XIMAX sensor is disconnected or until the sensor
trend data is displayed by pressing TREND, then SENSOR.
A corresponding “DATA IN SENSOR” message is also displayed at the bottom of the display. After 4 to 6 seconds, if all the data has been read from the O
XIMAX sensor, the message is replaced with the main
menu.
N-595 69
OXIMAX Sensor Event Record
If data is still being read from the OXIMAX sensor , after 4 to 6 seconds, the DATA IN SENSOR message is replaced with a READING TRENDS message with an ABORT option.
Selecting the ABORT softkey stops the recording of additional data in the O
XIMAX sensor.
O
Sensor event record can be viewed by accessing the TREND/ SENSOR menu.
The SENSOR EVENT RECORD LED comes on steady when
O
XIMAX sensor memory is full and stays on until the OXIMAX sensor is
disconnected.
XIMAX sensor and accessing or viewing the data that is in the

OXIMAX Sensor Event Record Not Available

If the user selects the TREND/SENSOR option when a connected
XIMAX sensor (single-patient-use OXIMAX sensors only) does not
O
contain data, because no events were recorded to the O memory chip in the prior monitoring situation, a “NO DATA” message is displayed on the default trend or event graph.
70
XIMAX sensor
OXIMAX Sensor Event Record
A sample event display in which no data are available is shown below. The message will be cleared when the graph or summary is exited.

OXIMAX Sensor Event Record Graphical Data

Graphical representations of patient event history is only available on single-patient-use O minimum or maximum %SpO throughout the duration of an event (%SpO alarm threshold for at least 15 seconds) and continuing every 30 seconds until the actual %SpO threshold.
XIMAX sensors. Graphed data points are the
2 value for each 30-second interval
2 value equals or exceeds the alarm
2 continuously below
The duration of an event is determined by the number of data points in the event. Each data point is stored at 30-second intervals.
Events end for one of four reasons:
The %SpO2 returns to or above the alarm limit
Loss of pulse
•The O
N-595 71
XIMAX sensor is disconnected
OXIMAX Sensor Event Record
The graph title shows the data type (EVENT GRAPH) in the upper left corner. The number of the displayed event and the total number of events recorded in the O title (example, 2/2). The date and time of the displayed event are shown in the upper center and upper right corner.
The type of data displayed in the graph is indicated to the left of the vertical axis (%SpO during the event. The duration of the event is shown below the range value. The vertical axis of the graph is labeled to show the magnitude scale of the graphed data. The horizontal axis is not labeled but automatically scales to accommodate the number of 30-second intervals during the event. The alarm threshold (lower than %SpO alarm limit) is represented by a horizontal dotted line across the graph. The first data point is always the alarm threshold.
•The OXIMAX sensor is off the patient
XIMAX sensor are shown to the right of the
2). Below this is the range of values (min/max)
2
72
Events are displayed one at a time, one per graph. Graphs are displayed in chronological sequence with the most rece nt ev ent shown first when accessing the graphical O
XIMAX sensor event
display. The user can move between events by using the two left-most softkeys which are labeled with left- and right-facing arrow icons, respectively. At the beginning of an event sequence, event 1 of 2 events, the left-arrow soft key is blank; at the end of a sequence, event 2 of 2 events, the right-arrow soft key is blank.
The ADJUST UP and ADJUST DOWN buttons on the monitor panel can also be used to move through events.
The PRINT softkey allows the user to print the displayed event graph. The BACK softkey takes the user back to the previous TREND/ SENSOR sub-menu level.
OXIMAX Sensor Event Record
R

Viewing and Printing OXIMAX Sensor Event History Data

With the monitor in the normal monitoring mode. You must connect a printer, capable of printing graphs, to the monitor data port connector to print O
The monitor protocol must be set to GRAPH to print the in-sensor event history data. Refer to Printing Monitor Trend Information on page 79. To view and print in-sensor event history data:
XIMAX sensor event history data.
TREND
SENSO
GRAPH
1. Connect an O
XIMAX sensor containing patient data to the monitor .
2. Press the TREND softkey.
3. Press the SENSOR softkey.
4. Press the GRAPH softkey.
Note: Use the left and right arrow softkeys to scroll through the pages
of the event graph.
PRINT
EXIT
5. Press the PRINT softkey to print the displayed screen.
6. Press the EXIT softkey.
N-595 73
OXIMAX Sensor Event Record
A sequence of %SpO2 + BPM (saturation plus pulse rate) “dual-view” event graphs are shown below. The duel-view graph is the same as a single graphical event history graph except the graphs are compressed horizontally to allow both %SpO graphs to be shown for the same event.
2 and pulse rate
74

OXIMAX Sensor Tabular Event Data

The OXIMAX sensor tabular event data is a listing of all events recorded on
the
OXIMAX sensor’s memory chip.
OXIMAX Sensor Event Record
The table title shows in the upper left corner. Below the table title is a six-column table with left-to-right column headings of event number (#), date (DATE), event start time (START), event duration (DUR), %SPO
2 minimum and maximum values during the event (%SPO2),
and pulse rate minimum and maximum values during the event (BPM).
Event data are listed in chronological order with the most recent event shown first, at the top of the list, when the tabular Event Summary display is first accessed. Four events can be displayed simultaneously; the table must be scrolled to view additional events. The user can move to the next screen view of the table, the next three events (the previously displayed bottom or top event is retained as the fourth event for context when a table is scrolled), using the two left-most softkeys which are labeled with left- and right-facing arrow icons, respectively. At the beginning of an event sequence, Event 1 of 5 events, the left-arrow soft key is blank; at the end of a sequence,
N-595 75
OXIMAX Sensor Event Record
R
Event 5 of 5 events, the right-arrow soft key is blank, indicating you have reached the beginning or end of the table.
The ADJUST UP and ADJUST DOWN buttons on the monitor panel can be used to move through the Event Summary table line by line.
The PRINT softkey allows the user to print the displayed event graph.
The BACK softkey takes the user back to the previous TREND/ SENSOR sub-menu level.
Viewing and Printing In-Sensor Tabular Event History Data
The monitor should be in the normal monitoring mode.
To view and print in-sensor tabular event history data:
TREND
SENSO
TABLE
PRINT
1. Press the TREND softkey.
2. Press the SENSOR softkey.
3. Press the TABLE softkey.
4. Press the PRINT softkey to print the data.
76
OXIMAX Sensor Event Record
BACK
5. Press the BACK softkey.
N-595 77

Printing

Printing Monitor Trend Information

Trend information (monitor and in-sensor event history) may be sent to a personal computer or to a serial printer.
Note: The protocol settings must be set to ASCII MODE for printing
text data or GRAPH MODE for printing graphical data.
1. Data Port Connector
With the monitor in the normal monitoring mode:
1. Connect the serial printer to the monitor’s DATA PORT
connector (1), using Nellcor printer cable part number 036341.
2. Turn on the printer.
SETUP
NEXT
N-595 79
3. Press the SETUP softkey.
4. Press the NEXT softkey.
Printing
R
NEXT
COMM
SELECT
5. Press the NEXT softkey.
6. Press the COMM softkey.
7. Set the BAUD rate to the appropriate number using the ADJUST UP button.
8. Press the SELECT softkey to select PROTOCOL.
9. Set the PROTOCOL to ASCII for text printing or GRAPH for graph printing using the ADJUST UP button.
EXIT
TREND
MONITR
10. Press the EXIT softkey.
11. Press the TREND softkey.
12. Press the MONITR softkey for monitor trend printing or press the SENSOR softkey for in-sensor event history data printing.
SENSO
80
Printing
NEXT
PRINT
13. Press the NEXT softkey.
14. Press the PRINT softkey.
ASCII printout:
GRAPH printout:

Monitor Trend Data in ASCII Mode

Refer to Printing Monitor Trend Information on page 79 for the procedure to print trend information.
N-595 81
Printing
The format of data displayed when a trend printout is shown in Figure 3. “TREND” is displayed in the top row.
Readings are displayed in 2- or 4-second intervals depending on the response mode selected. The values on each row are an average of the response mode selected period.
At the end of the printout an “Output Complete” line indicates that the transmission was successful. If the “Output Complete” line is not present, a corruption of the data may have been detected and the data should be ignored.
Once a trend printout has begun, it cannot be aborted without turning off the N-595 or the printer.

Trend Data in Graph Mode

Refer to Printing Monitor Trend Information on page 79 for the procedure to print trend information. See Figure 4 on page 83.
82
Figure 3: ASCII Mode Printout
The graph mode disables all printout functions except trend data. Graph mode trend printouts are formatted for a Seiko DPU-414 and Okidata 320 serial printer.
Figure 4: Graph Mode Printout
Real-Time Display/Printout Format
Real-time data is continuously sent to the data port on the back of the N-595. Patient data can be obtained through the data port by connecting the monitor data port to a PC or serial printer. When a real-time printout or display is being transmitted to a printer or PC, a new line of data is displayed every 2 seconds. Column headings are displayed or printed after every 25 lines, or if one of the values in the column heading changes. Readings are displayed at 4-second intervals if the SpO intervals when the SpO2 response mode is set to fast.
2 response mode is set to normal and at 2-second
Printing
Data cannot be obtained if the N-595 is operating on battery power.
Note: If the data output stops transmitting, turn the power off and
back on again or, if the monitor is connected to a PC, send an XON (Ctrl-q) command to reset the monitor.
N-595 83
Printing
An example of a real-time output is shown in Figure 5 on page 84.
84
Figure 5: Real-Time Printout

Column Headings

Data Source

Printing
Every 25th line of the data consists of a column heading.
A column heading is also output whenever a value of the column heading is changed. There are three column-heading lines shown in the printout. Using the top row as the starting point there are 25 lines before the second row of column headings is printed. The third row of column headings was displayed because the operator changed the SpO
2 lower alarm limit from 85 percent to 80 percent.
Data in the highlighted box above represents the model number of the monitor, in this case the N-595.

Software Version

The next data field tells the user the software level (Version 1.0.0.0) and a software verification number (CRC: XXXX). Neither of these numbers should change during normal operation. The numbers may change if the monitor is serviced and receives a software upgrade.
N-595 85
Printing

Alarm Limits

Monitor Mode

The last data field in the top line indicates the upper and the lower alarm limits for %SpO above the lower alarm limit for SpO
2 and for the pulse rate (PR). In the example
2 is 85% and the upper alarm
limit is 100%. Pulse Rate alarm limits are 40 and 170 bpm. The SatSeconds alarm limit (0SAT-S) displays the SatSeconds alarm setting. In this example SatSeconds is set to off.

Response Mode

86
The monitor mode (ADULT or NEONATE) is identified on the printout.
The response mode (NORMAL or FAST) is identified on the printout.

Data Column Headings

Actual column headings are in the second row of the column heading line. Patient data presented in the chart, from left to right, is the:
Printing
time the patient data were obtained

Time

Patient Data

current %SpO
2 value
current Pulse Rate (BPM)
current Pulse Amplitude (PA)
operating status of the N-595.
The Time column represents the N-595 real-time clock.
Patient data are highlighted in the display above. Parameter values are displayed directly beneath the heading for each parameter. In this
N-595 87
Printing

Operating Status

example the %SpO2 is 100, and the pulse rate is 190 beats per minute. The “*” next to the 190 indicates that 190 beats per minute is outside of the alarm limits, indicated in the top row, for pulse rate. If no data for a parameter is available, three dashes (- - -) will be displayed.
PA represents pulse amplitude. The number can range from 0 to 254. There are no alarm parameters for this value. It can be used for trending information as an indication of a change in pulse volume, relative pulse strength, or circulation.
The Status column indicates alarm conditions and operating status of the N-595. In this example, the PH means that the pulse rate upper alarm limit (Pulse High) has been exceeded. A complete listing of the status codes is listed below. As many as four codes can be displayed at one time in the Status column.
88
Code Meaning
AO Alarm Off
AS Alarm Silence LB Low Battery
LM Loss of Pulse w/ Motion
LP Loss of Pulse
MO Patient MOtion
PH Pulse Rate Upper Limit Alarm
PL Pulse Rate Lower Limit Alarm PS Pulse Search
Printing
Code Meaning
SH Saturation Upper Limit Alarm
SL Saturation Lower Limit Alarm SD Sensor Disconnect SO Sensor Off
Note:An OXIMAX sensor disconnect will also cause three dashes
(- - -) to be displayed in the patient data section of the display or printout.
N-595 89
Loading...